BW Technologies GasAlert MicroClip Quick Start Guide

BW Technologies LP (BW) warrants this product to be free from defects in material and workmanship under normal use and service for a period of two years, beginning on the date of shipment to the buyer. This warranty extends only to the sale of new and unused products to the original buyer. BW’s warranty obligation is limited, at BW’s option, to refund of the purchase price, repair, or replacement of a defective product that is returned to a BW authorized service center within the warranty period. In no event shall BW’s liability hereunder exceed the purchase price actually paid by the buyer for the Product. This warranty does not include:
a) fuses, disposable batteries or the routine replacement of parts due to the normal wear and tear of the product arising from
use;
b) any product which in BW’s opinion, has been misused, altered, neglected or damaged by accident or abnormal conditions of
operation, handling or use;
c) any damage or defects attributable to repair of the product by any person other than an authorized dealer, or the installation
The obligations set forth in this warranty are conditional on:
THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. BW SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT OR RELIANCE OR ANY OTHER THEORY. Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
of unapproved parts on the product; or
a) proper storage, installation, calibration, use, maintenance and compliance with the product manual instructions and any other
applicable recommendations of BW;
b) the buyer promptly notifying BW of any defect and, if required, promptly making the product available for correction. No
goods shall be returned to BW until receipt by the buyer of shipping instructions from BW; and
c) the right of BW to require that the buyer provide proof of purchase such as th e original invoice, bill of sale or packing slip to
establish that the product is within the warranty period.
Limited Warranty & Limitation of Liability

Contacting BW Technologies by Honeywell

USA: 1-888-749-8878 Canada: 1-800-663-4164 Europe: +44 (0) 1295 700300 Other countries: +1-403-248-9226
Email us at: Visit BW Technologies by Honeywell’s web site at: www.gasmonitors.com
info@bwtnet.com
GasAlertMicroClip

Introduction

This quick reference guide provides basic information for the GasAlertMicroClip. Refer to the user manual on the accompanying CD-ROM for complete operating instructions. The GasAlertMicroClip gas detector (“the detector”) warns of hazardous gas at levels above user-selectable alarm setpoints.
The detector is a personal safety device. It is your responsibility to respond properly to the alarm.
Note
The detector is shipped with English as the default displayed language. Additional languages provided are Portuguese, Spanish, German, and French. The screens for the additional languages are displayed on the detector and in the corresponding Quick Reference Guide.

Safety Information - Read First

Use the detector only as specified in this guide, otherwise the protection provided by the detector may be impaired.
Read the following Cautions before using the detector.
a Cautions
Warning: Substitution of components may impair
Intrinsic Safety.
Caution: For safety reasons, this equipment
must be operated and serviced by qualified personnel only. Read and understand the user manual completely before operating or servicing.
Charge the detector before first-time use. BW
recommends the detector be charged after every workday.
Calibrate the detector before first-time use and
then on a regular schedule, depending on use and sensor exposure to poisons and contaminants. BW recommends at least once
1
GasAlertMicroClip
Quick Reference Guide
every 180 days (6 months).
The combustible sensor is factory calibrated to
50% LEL methane. If monitoring a different combustible gas in the % LEL range, calibrate the sensor using the appropriate gas.
Only the combustible gas detection portion of
this instrument has been assessed for performance by CSA International.
Calibrate only in a safe area that is free of
hazardous gas.
It is recommended that the combustible sensor
be checked with a known concentration of calibration gas after any known exposure to contaminants/poisons (sulfur compounds, silicon vapors, halogenated compounds, etc.).
BW recommends to “bump test” the sensors,
before each day’s use, to confirm their ability to respond to gas by exposing the detector to a gas concentration that exceeds the alarm setpoints. Manually verify that the audible and visual alarms are activated. Calibrate if the readings are not within the specified limits.
Caution: High off-scale readings may indicate an
explosive concentration.
Any rapid up-scaling reading followed by a
declining or erratic reading may indicate a gas concentration beyond upper scale limit, which
2
may be hazardous.
Extended exposure of the GasAlertMicroClip to
certain concentrations of combustible gases and air may stress a detector element, which can seriously affect its performance. If an alarm occurs due to high concentration of combustible gases, recalibration should be performed, or if needed, the sensor replaced.
Protect the combustible sensor from exposure to
lead compounds, silicones, and chlorinated hydrocarbons. Although certain organic vapors (such as leaded gasoline and halogenated hydrocarbons) may temporarily inhibit sensor performance, in most cases, the sensor will recover after calibration.
For use only in potentially explosive
atmospheres where oxygen concentrations do not exceed 20.9% (v/v).
Parts of the GasAlertMicroClip
GasAlertMicroClip

Parts of the GasAlertMicroClip

Item Description
1 Visual alarm bars (LED) 2 Liquid crystal display (LCD) 3 Audible alarm 4 Sensors 5 Pushbutton 6 Charging connector / IR interface 7 Alligator clip
3
GasAlertMicroClip
Quick Reference Guide

Display Elements

4
Item Description
1 2 Automatically span sensor 3 Gas cylinder 4 Gas identifier bars 5 Battery life indicator 6 Stealth mode 7 Numeric value 8 Automatically zero sensor
Alarm condition