BW Technologies LP (BW) warrants the product to be free from defects in material and workmanship under normal use and service for a period of
Limited Warranty and Limitation Liability
two years, beginning on the date of shipment to the buyer. This warranty extends only to the sale of new and unused products to the original buyer.
BW’s warranty obligation is limited, at BW’s option, to refund of the purchase price, repair or replacement of a defective product that is returned to
a BW authorized service center within the warranty period. In no event shall BW’s liability hereunder exceed the purchase price actually paid by the
buyer for the Product.
This warranty does not include:
a) fuses, disposable batteries or the routine replacement of parts due to the normal wear and tear of the product arising from use;
b) any product which in BW’s opinion, has been misused, altered, neglected or damaged, by accident or abnormal conditions of operation,
handling or use;
c) any damage or defects attributable to repair of the product by any person other than an authorized dealer, or the installation of unapproved
parts on the product; or
The obligations set forth in this warranty are conditional on:
a) proper storage, installation, calibration, use, maintenance and compliance with the product manual instructions and any other applicable
recommendations of BW;
b) the buyer promptly notifying BW of any defect and, if required, promptly making the product available for correction. No goods shall be
returned to BW until receipt by the buyer of shipping instructions from BW; and
c) the right of BW to require tha t the buyer provide proof of purchase such as the original invoice, bill of sale or packing slip to establish that the
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential
damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this warranty is held invalid or
unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
Email us at: info@gasmonitors.com
Visit BW Technologies by Honeywell’s website at: www.gasmonitors.com
Page 3
GasAlert Extreme
Introduction
This quick reference guide provides basic information to operate the GasAlert Extreme gas detector. For complete operating
instructions, refer to the GasAlert Extreme User Manual provided on the CD-ROM. The GasAlert Extreme gas detector
(“the detector”) is designed to warn of hazardous gas levels
above user defined alarm setpoints.
The detector is a personal safety device. It is your responsibility
to respond properly to the alarm.
Note
If you have a multi-language detector, the detector is
shipped with English as the default displayed
language. Additional languages provided are French,
German, Spanish, and Portuguese. The screens for
the additional languages are displayed on the
detector and in the corresponding quick reference
guide.
Safety Information - Read First
Use the detector only as specified in this quick reference guide
and user manual, otherwise the protection provided by the
detector may be impaired.
Read the following Cautions before using the detector.
a Cautions
• Warning: Substitution of components may impair
Intrinsic Safety.
• Warning: To prevent ignition of flammable or
combustible atmospheres, disconnect power before
servicing.
• Warning: Use only the Panasonic CR-2PE/BN. Refer to
Replacing the Battery or Sensor.
• Warning: To reduce the risk of ignition of a flammable
atmosphere, batteries must only be changed in a safe
area free of hazardous gas.
• Calibrate only in a safe area that is free of hazardous
gas in an atmosphere of 20.9% oxygen.
• Use only a sensor specifically designed for the GasAlert
Extreme model.
• Do not deactivate the detector during a work shift.
1
Page 4
GasAlert Extreme
Quick Reference Guide
Deactivating the detector resets the time-weighted
average (TWA), short-term exposure limit (STEL), and
maximum gas exposure values to 0.
• If the detector is damaged or parts are missing, contact
BW Technologies by Honeywell
• Do not use the detector if it is damaged. Before using
the detector, inspect the case. Look for cracks and
missing parts.
• If the detector has been disassembled, ensure the front
and rear shells are properly aligned and fastened before
activating the detector. Refer to Maintenance
• BW recommends to bump test the sensor before each
day’s use to confirm its ability to respond to gas by
exposing the detector to a gas concentration that
exceeds the alarm setpoints. Manually verify that the
audible and visual alarms are activated. For detectors
with the calibration feature, calibrate the sensor if the
reading is not within the specified limits.
• Calibrate the detector before first-time use, and then at
least once every 180 days. For HCN detectors, calibrate
once every 30 days.
immediately.
.
2
• Ensure the sensor screen is not blocked.
• Do not cover the sensor grill with hands.
To avoid possible damage to the detector, adhere to the
following:
• Do not expose the detector to electrical shock and/or
severe continuous mechanical shock.
• The oxygen GasAlert Extreme detector is classified by
Underwriters Laboratories Inc. up to an atmosphere of
21% oxygen.
• Do not attempt to disassemble, adjust, or service the
detector unless instructions for that procedure are
contained in the user manual, and/or that part is listed
as a replacement part. Use only BW Technologies by
Honeywell replacement parts. Refer to Replacement
Parts and Accessories.
• The detector warranty will be voided if customer
personnel or third parties damage the detector during
repair attempts. Non-BW Technologies by Honeywell
repair/service attempts void this warranty.
• Products may contain materials that are regulated for
transportation under domestic and international
dangerous goods regulations. Return product in
compliance with appropriate dangerous goods
regulations. Contact freight carrier for further
instructions.
Page 5
Informations de sécurité – À lire au préalable
Utiliser le détecteur dans le seul but indiqué dans ce mémento
et dans le manuel d'utilisation, sous peine de porter préjudice à
la protection assurée par le détecteur.
Lire les Mises en garde suivantes avant d'utiliser le détecteur.
a Mises en garde
• Avertissement : Le remplacement d'un composant de
l'appareil peut compromettre la sécurité intrinsèque du
détecteur.
• Avertissement :Pour éviter l'inflammation
d'atmosphères inflammables ou combustibles, couper
l'alimentation électrique avant tout entretien.
• Avertissement : Utiliser uniquement des piles
Panasonic CR-2PE/BN. Se reporter à la section
Replacing the Battery or Sensor.
• Avertissement : Pour réduire le risque d'ignition dans
les atmosphères inflammables, les piles doivent être
changées in a dans une zone sûre, exempte de gaz
dangereux.
• Veillez à effectuer l'étalonnage dans une zone sûre,
exempte de gaz dangereux, dans une atmosphère
contenant 20,9 % d'oxygène.
• Utiliser uniquement les capteurs spécialement conçus
pour le modèle GasAlert Extreme.
• Ne pas désactiver le détecteur au cours d'une période
de travail. La désactivation du détecteur permet de
remettre à zéro la moyenne pondérée dans le temps
(TWA), la limite d'exposition à court terme (STEL) et les
Informations de sécurité – À lire au préalable
GasAlert Extreme
valeurs d'exposition maximale au gaz.
• Si le détecteur est endommagée ou que des pièces sont
manquantes, contacter immédiatement BW
Technologies by Honeywell.
• Ne pas utiliser le détecteur s'il est endommagé. Avant
d'utiliser le détecteur, inspecter le boîtier. S'assurer
qu'aucune pièce ne manque et vérifier l'absence de
fissures.
• Si le détecteur a été démonté, s'assurer que les coques
avant et arrière sont correctement alignées et serrées
avant d'activer le détecteur. Se reporter à la section
Maintenance
• Avant chaque utilisation quotidienne, BW recommande
d'effectuer un test fonctionnel du capteur afin de vérifier
qu'il réagisse bien aux gaz présents, en exposant le
détecteur à une concentration de gaz supérieure aux
seuils d'alarme. Vérifier manuellement que les alarmes
sonore et visuelle sont activées. Pour les détecteurs
équipés de la fonction d'étalonnage, étalonner le
capteur si la valeur ne respecte pas les limites
spécifiées.
• Étalonner le détecteur avant sa première utilisation,
puis une fois tous les 180 jours. Pour les détecteurs
HCN, effectuer un étalonnage tous les 30 jours.
.
3
Page 6
GasAlert Extreme
Quick Reference Guide
• Vérifier que le filtre du capteur n'est pas bouché.
• Ne pas couvrir la grille du capteur avec vos mains.
• Ne pas placer le détecteur près de la bouche ou des
épaules.
Pour éviter des dommages possibles sur le détecteur,
respecter les consignes suivantes :
• Ne pas soumettre le détecteur à des chocs électriques
ni à d'importants chocs mécaniques continus.
• Le détecteur d'oxygène GasAlert Extreme est classé par
l'organisme Underwriters Laboratories Inc. apte à une
utilisation jusque dans une atmosphère de 21 %
d'oxygène.
• Ne pas essayer de démonter, ajuster ou entretenir le
détecteur à moins que des instructions soient
spécifiées dans le manuel d'utilisation et/ou que la
pièce en question soit répertoriée comme pièce de
rechange. Utiliser des pièces de rechange BW
Technologies by Honeywell uniquement. Se reporter à
la rubrique Pièces de rechange et accessoires.
• La garantie du détecteur sera annulée si le personnel du
client ou des tiers endommagent le détecteur lors d'une
tentative de réparation. Les tentatives de réparation/
entretien non effectuées par BW Technologies by
Honeywell annulent la présente garantie.
• Les produits peuvent contenir des matériaux qui sont
réglementés pour le transport en vertu des règlements
nationaux et internationaux de marchandises
dangereuses. Retourner le produit conformément à la
réglementation sur les marchandises dangereuses
4
appropriées. Contactez transporteur pour plus
d'instructions.
Page 7
Parts of the GasAlert Extreme
Parts of the GasAlert Extreme
GasAlert Extreme
ItemDescription
1Visual alarms bars (LEDs)
2Liquid crystal display (LCD)
3Pushbuttons
4Audible alarm
5Sensor and sensor grill
6Infrared (IR) communication port
7Alligator clip
5
Page 8
GasAlert Extreme
Quick Reference Guide
Display Elements
Note
Press and hold (until the backlight activates) any
button to activate the backlight for 6 seconds. When
the backlight option is enabled, the backlight
automatically activates for 3 seconds whenever there
is insufficient light to view the LCD. The detector is
shipped with the backlight option enabled.
The backlight does not activate when stealth mode is
enabled.
6
ItemDescription
1Numeric value
2Gas cylinder
3Automatic span sensor
4Passcode lock
5Set alarm setpoints and user options
6Maximum gas exposure
7Alarm conditions
8Battery
9Data transmission
10Alarm or alarm setpoint
11Automatically zero sensor
12Optional datalogger indicator
13Parts per million (ppm)
14Percentage by volume (% vol.)
15
Percentage by lower explosive limit (% LEL)
(future use)
Page 9
GasAlert Extreme
Pushbuttons
Pushbuttons
PushbuttonDescription
• To activate the detector, press A.
A
E
D
C
• To deactivate the detector, press A and hold for 5 seconds.
• To enable/disable the confidence beep, while the detector is deactivated press and hold C. While
holding C, press A to enable or disable the confidence beep during startup.
• To decrement the displayed value or to scroll down, press E.
• To enter the user options menu, press E and D simultaneously and hold until OPTN and then EXIT
displays (5 seconds).
• To initiate calibration and define alarm setpoints, press E and C simultaneously until CAL. displays.
• To increment the displayed value or to scroll up , press D.
• To view the TWA, STEL, and maximum gas exposures, press D and C simultaneously.
• To save a displayed value, press C.
• To clear TWA, STEL, and maximum gas exposures, press and hold C for 6 seconds.
• To acknowledge a latched alarm, press C.
Activating the Detector
To activate the detector, press A. Once the detector passes the self-test, the detector enters normal operation.
Deactivating the Detector
To deactivate the detector, press A and hold for 5 seconds.A deactivation passcode can be enabled to ensure only authorized deactivations occur. This option can only be enabled at the factory and must be specified at time of order, and cannot be disabled by the
customer. Refer to the GasAlert Extreme User Manual.
7
Page 10
GasAlert Extreme
Quick Reference Guide
Calibration
ProcedureScreenProcedureScreen
1. In a safe area free of hazardous gas and
in an atmosphere of 20.9% oxygen, press
E and C simultaneously and hold for 5
seconds. The detector be eps and vibrates
four times. After CAL. displays on the LCD,
the detector then beeps on ce more.
2. The LCD flashes AUTO-ZERO while
the detector zeroes the sensor. The detector beeps twice at the end of auto-zero.
3. If the detector is passcode protected,
PAS S flashes on the LCD. The correct
passcode must be entered before defining
the calibration gas concentration.
4. The LCD displays the current calibration gas concentration. Press C to accept
the current concentration, or press E or
D to change the concentration and press
C to confirm the new concentration.
8
5. When the LCD flashes , attach the
calibration cap and apply gas at a flow
rate of 500 to
1000 ml/min. The detector beeps three
times at the end of the span stage.
Remove the calibration cap and close the
valve on the gas cylinder.
6. After a successful calibration, the LCD
displays CAL. DUE. Press E or D to
define the next calibration due date. Press
C to save.
7. Press C to bypass defining the alarm
setpoints.
Or
Press E or D to change the alarm setpoint and press C to save the new value.
The detector beeps and vib rates four
times at the end of calibration.
a Warning
For correct flow and span rates, refer to the
GasAlert Extreme User Manual.
Page 11
Attach the Gas Cylinder to the Detector
ItemDescription
1Calibration cap
2Hose
3Regulator
4Gas cylinder
Attach the Gas Cylinder to the Detector
GasAlert Extreme
9
Page 12
GasAlert Extreme
Quick Reference Guide
Alarms
Refer to the following table for information about alarms and corresponding screens. During an alarm condition, the detector activates the backlight and the LCD displays the current ambient gas reading.
AlarmScreenAlarmScreen
Low Alarm
• Slow beep
• Slow flash
• ALARM flashes
• Slow vibrations
High Alarm
• Fast beep
• Fast flash
• ALARM flashes
• Fast vibrations
Sensor Alarm
• Slow beep
• Slow flash
• ALARM flashes
• Slow vibrations
10
TWA Alarm
• Slow beep
•Slow flash
• ALARM flashes
• Slow vibrations
STEL Alarm
• Fast beep
• Fast flash
• ALARM flashes
• Fast vibrations
Low Battery Alarm
• Confidence beep disabled: One
beep and flash every 5 seconds and
one quick vibration every minute
Confidence beep enabled: Confi-
dence beep deactivates. No beeps,
flashes, or vibrations.
• LOW displays
Page 13
GasAlert Extreme
AlarmScreenAlarmScreen
Automatic Shutdown Alarm
(Low battery)
• Eight beeps, flashes, and vibrations
• LOW displays
After Automatic Shutdown
(Low battery)
• No beep
• No flash or vibrations
• displays for a short time
High alarm and STEL alarm have the same priority. A high alarm and/or STEL alarm overrides a low alarm and/or TWA
alarm. To check STEL and TWA alarms specifically, press and hold C and D simultaneously.
The vibrator alarm is disabled at -20°C.
The high and low alarms deactivate when the gas concentration is lower than the low alarm setpoint. If the alarms are set
to latch, alarms persist until the gas concentration is below the low alarm setpoint and the alarms have been acknowledged
by pressing C. The TWA and STEL alarms deactivate by clearing the TWA and STEL peak exposure. Refer to the
GasAlert Extreme User Manual to clear TWA and STEL peak exposures.
Automatic Shutdown Alarm
(Calibration past)
• Eight beeps, flashes, and vibrations
Confidence Beep
• One beep every 5 seconds
• One quick vibration per minute
Note
If the detector enters low
battery alarm, the confidence
beep deactivates.
Note
Alarms
11
Page 14
GasAlert Extreme
Quick Reference Guide
User Options Menu
Note
When selecting a user option, Set flashes and the
LCD displays the opposite of what is currently
enabled.
To access the user options menu, press and hold E and D
simultaneously until the LCD displays OPTN.
To scroll through the options, press E or D. Press C to select
the option. The following are the available user options:
• EXIT: Exits the user options menu.
• CLCK (Clock): Define the date and time of the detector.
• PASS (Passcode): Enable or disable passcode protec-
tion. Prevent unauthorized access to the user options
menu, calibration, and defining alarm setpoints.
• STLH (Stealth): Enable or disable stealth mode. This
option disables the audible alarms, LEDs, and backlight.
The vibrator alarm is disabled at -20°C.
• BKLT (Backlight): Enable or disable the automatic
backlight. When enabled, the backlight automatically
activates for 3 seconds whenever there is insufficient
light to view the LCD. When stealth mode is enabled,
this option is not available.
• LTCH (Latching alarms): Enable or disable the latching
alarm option. When enabled, the low and high gas
alarms (audible, visual, and vibrator) persist until the
gas concentration is below the low alarm setpoint and
the alarms have been acknowledged by pressing C.
12
• ACAL (Automatic oxygen calibration): Enable or disable
automatic oxygen calibration at startup. Only applicable
to the GasAlert Extreme oxygen detector.
• PAST (Calibration past due): Enable or disable auto-
matic shutdown if calibration is past due (at startup).
• PORT, ESPA, DEUT, FRAN, or ENGL: The LCD can
display text in five different languages: Portuguese,
Spanish, German, French, or English. Only applicable to
detectors with the multi-language feature.
• RATE (Datalogger sampling rate option): Adjust the dat-
alogger sampling rate. Only applicable to detectors with
the datalogger option.
• SEND (Data transfer): Transfers the datalogs and event
logs from the detector to the PC. Only applicable to
detectors with the datalogger option.
Information can be sent in the following three ways:
EVNT (Event): Sends all of the event logs.
LAST: Sends all of the datalogs since the last time they
were downloaded.
ALL: Sends all of the datalogs saved on the detector.
Note
Confidence Beep is not enabled in the user options
menu. To enable the confidence beep, while the
detector is deactivated press and hold C. While
holding
C, press A. For more information refer to
the GasAlert Extreme User Manual.
Page 15
Maintenance
To keep the detector in good operating condition, perform the
following basic maintenance as required:
•
Calibrate, test, and inspect the detector at regular intervals
• Keep an operations log of all maintenance, calibrations,
and alarm events.
• Clean the exterior with a soft damp cloth. Do not use
solvents, soaps, or polishes.
• Do not immerse the detector in liquids.
Replacing the Battery or Sensor
a Warning
To avoid possible personal injury, adhere to the following:
• Replace the battery in a safe area, free of hazardous gas
immediately when the detector enters low battery alarm.
• Use only the Panasonic CR-2PE/BN battery.
• Wear an ESD wrist or heel strap when replacing the
sensor or battery. Avoid touching electronic
components on the detector PCB or shorting circuits on
the PCB.
• Use only the sensor specifically designed for the
GasAlert Extreme. Otherwise, the detector will not
monitor the target gas. Refer to Replacement Parts and
Accessories in the GasAlert Extreme User Manual.
•
After replacing a sensor, allow the new sensor 5 minutes
to stabilize before use. For an ETO or NO sensor, allow
the new sensor 2 hours to stabilize before use.
GasAlert Extreme
Maintenance
• Do not expose a sensor to vapors of organic solvents
such as paint fumes or organic solvents.
Note
.
When the battery is removed from the detector, the
clock reverts back to the default value. Refer to Clock
Option in the GasAlert Extreme User Manual.
To preserve the life of the battery, deactivate the detector when
not in use.
For additional information regarding problems caused by a sensor requiring calibration or replacement, refer to Troubleshooting in the GasAlert Extreme User Manual.
Replacing the Battery
To replace a battery, complete the following. Refer to Internal
Parts of the GasAlert Extreme and Internal Seals of the
GasAlert Extreme. Replace the battery in a safe area, free of
hazardous gas.
1. Deactivate the detector.
2. Remove the four machine screws on the rear shell
and remove the rear shell.
3. Remove the battery.
ec Warning
This instrument contains a lithium battery. Do not mix with
the solid waste stream. Spent batteries should be disposed
of by a qualified recycler or hazardous materials handler.
13
Page 16
GasAlert Extreme
Quick Reference Guide
4. Insert the new battery.
5. Re-assemble the detector. When assembling the
detector be aware of the following:
•Clean the seal on the front and rear shells with a
soft damp clean cloth. Do not use solvents,
soaps, or polishes. Refer to Internal Seals of the
GasAlert Extreme.
•Ensure the front and rear shells are properly
aligned to ensure a proper environmental seal.
•Torque the screws to 3-4 in-lbs in a crisscross
pattern to ensure a proper environmental seal.
Do not overtighten.
Replacing the Sensor
To replace a sensor, complete the following. Refer to Internal
Parts of the GasAlert Extreme and Internal Seals of the
GasAlert Extreme.
1. Deactivate the detector.
2. Remove the four machine screws on the rear shell
and remove the rear shell.
3. Remove the two machine screws from the PCB.
4. Remove the PCB. Place the PCB on a clean, nonconductive surface.
5. Replace the sensor.
Note
14
Allow the new sensor 5 minutes to stabilize before
use. For a new ETO or NO sensor, allow the new
sensor 2 hours to stabilize before use.
6. Re-assemble the detector. When assembling the
detector, perform the following:
•Clean the seal on the front and rear shells with a
soft damp clean cloth. Do not use solvents,
soaps, or polishes. Refer to Internal Seals of the
GasAlert Extreme.
•Ensure the front and rear shells are properly
aligned to ensure a proper environmental seal.
•Torque the screws to 3-4 in-lbs in a crisscross
pattern to ensure a proper environmental seal.
Do not overtighten.
Cleaning a Sensor Screen
Clean or replace the sensor screen as required. If replacement
sensor screens are required, refer to Replacement Parts and
Accessories in the GasAlert Extreme User Manual.
To clean a removed sensor screen, complete the following.
Refer to Internal Parts of the GasAlert Extreme and Internal
Seals of the GasAlert Extreme.
1. Deactivate the detector.
2. Remove the four machine screws on the rear shell
and remove the rear shell.
3. Remove the two machine screws from the PCB.
Page 17
4. Remove the PCB. Place the PCB on a clean, nonconductive surface.
5. Remove the screen.
6. Using a soft, clean brush, wash the screen with clean,
warm water.
7. Insert the sensor screen with the shiny side facing the
sensor grill.
Note
Ensure the screen is dry before reinserting into the
detector.
8. Re-assemble the detector. When assembling the
detector be aware of the following:
•When reinserting the sensor screen into the
detector, ensure the sensor screen is inserted
with the shiny side facing the sensor grill.
•Clean the seal around the edge of the front and
rear shells with a soft damp clean cloth. Do not
use solvents, soaps, or polishes. Refer to Inter-
nal Seals of the GasAlert Extreme.
•Ensure the front and rear shells are properly
aligned to guarantee a proper environmental
seal.
•Torque the screws to 3-4 in-lbs in a crisscross
pattern to ensure a proper environmental seal.
Do not overtighten.
GasAlert Extreme
Maintenance
Internal Seals of the GasAlert Extreme
ItemDescription
1Seal
2Front shell
3Rear shell
15
Page 18
GasAlert Extreme
Quick Reference Guide
Internal Parts of the GasAlert Extreme
ItemDescription
1Detector rear screws
2Rear shell
3Battery
4Main PCB screws
5Main PCB
6Sensor
7Sensor screen
8Front shell
16
Page 19
WEEE Directive and Battery Directive
Failure to comply with the following battery removal and disposal instructions may result in battery shorting, battery leakage, and/or other damage. Ensure a qualified technician
completes the following procedures.
Removal and Disposal of the Battery
Only a qualified technician should complete the following procedure. Refer to Internal Parts of the GasAlert Extreme
the detector.
Dispose of the battery according to local laws.
1. Press and hold C to deactivate the detector.
2. Remove the four machine screws on the rear shell
and remove the rear shell.
3. Remove the two machine screws from the PCB.
4. Remove the PCB. Place the PCB on a clean, nonconductive surface.
5. Remove the battery.
6. Dispose of the battery according to local laws.
for parts of
WEEE Directive and Battery Directive
GasAlert Extreme
Specifications
Instrument dimensions: 2.8 x 5.0 x 9.5 cm
(1.1 x 2.0 x 3.75 in.)
Weight: 82 g (2.9 oz.)
Operating temperature:
S, SO2, HCN: -40°C to +50°C (-40°F to +122°F)
H
2
CO: -30°C to +50°C (-22°F to +122°F)
NH3 (high range): -20°C to +40°C (-4°F to +104°F)
Other gases: -20°C to +50°C (-4°F to +122°F)
Operating humidity:
S, SO2, Cl2, HCN, NO2, NH3, PH3, ETO, NO, O3:
CO, H
2
15% to 90% relative humidity (non-condensing)
Cl2: 10% to 95% relative humidity (non-condensing)
ClO
: 15% to 95% relative humidity (non-condensing)
2
: 0% to 99% relative humidity (non-condensing)
O
2
Alarm setpoints: All setpoints automatically display during the
start up self-test.
Sensor type: Plug-in electrochemical cells
Calibration: Auto zero, set span, and span sensor
Oxygen sensor: Automatic span upon activation (enable/dis-
able)
Alarm conditions: TWA alarm, STEL alarm, low alarm, high
Audible alarm: 95 dB at 1 ft. (0.3 m) typical
Visual alarm: Red light-emitting diode (LED)
Display: Alpha-numeric liquid crystal display (LCD)
Backlight: Automatically activates for 3 seconds whenever
there is insufficient light to view the LCD and during alarm conditions (unless the backlight is disabled in user options). Any
pushbutton reactivates the backlight for 6 seconds.
Self-test: Initiated at activation
: 0-50.0 ppm (0.1 ppm increments)
2
: 0-100 ppm (1 ppm increments)
3
(high range): 0-400 ppm
3
: 0-1.00 ppm (0.01 ppm increments)
2
: 0-1.00 ppm (0.01 ppm increments)
3
18
User field options: automatic backlight, passcode protection,
latching alarm, calibration due lock, language selection, enable/
disable automatic oxygen calibration, datalog interv
transfer option, confidence beep, calibration interval, stealth
mode, set alarm setpoints, calibration span concentration
Year of manufacture: The detector’s year of manufacture is
determined from the serial number. The second and third number after the first letter determines the year of manufacture.
al, data
E.g. J410-H001000 = 2010 year of manufacture
Battery test: Every 0.5-second
Battery: 3 V lithium Panasonic CR-2PE/BN
Event logs: A maximum of 10 events are stored. When 10
events have been recorded, the next event overwrites the oldest event log.
Warranty: 2 years including sensors (1 year Cl2, NH3, O3,
ETO, and ClO2 sensor)
Intrinsic Safety:
Classified by UL to both U.S. and Canadian Standards as intrinsically safe for Class I, Division 1, Group A, B, C, D and Class
II, Group E, F, G
European Explosives Protection
CE 0539 g II 1G
Ex ia IIC T4 Ga DEMKO 04 ATEX 0336363X
IECEx Ex ia IIC T4 Ga IECEx UL 08.0001X
ABS Type Approved: VA-348-169-X
KTL 12-KB4BO-0199X
1
EAC Ex Certificate: RU С-GВ.ГБ05.В.01115
Page 21
Inmetro: Ex ia IIC T4 Ga DNV 12.0137 X
Ratings and Certifications
The GasAlertExtreme is in conformity with the following standards:
This equipment has been tested and found to comply with the
limits for a Class B digital device, pursuant to Part 15 of the
FCC Rules and ICES-003 Canadian EMI requirements. These
limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if
not installed and used in accordance with the instructions, may
cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined
by turning the equipment off and on, the user is encouraged to
try to correct the interference by one or more of the following
measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from
that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for
help.
19
Page 22
GasAlert Extreme
Quick Reference Guide
20
Page 23
GasAlert Extreme
Specifications
21
Page 24
GasAlert Extreme
Quick Reference Guide
22