BW Technologies GasAlertClip Extreme 2 Year, GasAlertClip Extreme 3 Year, GasAlert Clip Extreme Instruction Sheet

Page 1
2 OR 3 YEAR GAS DETECTOR
50104153-003/05-2012
GasAlert Clip Extreme
The GasAlertClip Extreme 2 or 3 year gas detector (“the detector”) is a personal safety device that warns when hazardous gas exceeds factory set alarm setpoints. The detector stores and transmits gas alarm event data. It is your responsibility to respond properly to the alarms.
Oxygen (O2) Percent by volume (%)
Carbon monoxide (CO) Parts per million (ppm)
Hydrogen sulfide (H2S) Parts per million (ppm)
Sulfur dioxide (SO2) Parts per million (ppm)
a Safety Information—Read First
Warning: Substitution of components may impair Intrinsic Safety. Warning: Bump test the O2 sensor before each day’s use to confirm its
ability to respond to gas by exposing the detector to a gas concentration that exceeds the alarm setpoints. Manually verify that the audible and visual alarms are activated.
 Do not activate the detector after the date on the package.  This product is a gas detector, not a measurement device.  Perform a self-test each day prior to use.  Ensure the sensor grill is free of dirt, debris, and is not obstructed.  Calibrate and/or bump test the detector in a normal atmosphere
(20.9% O2) that is free of hazardous gas.
 Periodically test the response of the sensor to gas by exposing the
detector to a target gas concentration that exceeds the low alarm setpoint. Manually verify that the audible and visual alarms are activated.
 Periodically calibrate the GasAlertClip Extreme
oxygen (O2) detector.
c e Recycling
This instrument contains a lithium battery. Do not mix with the solid waste stream. Spent batteries should be disposed of by a qualified recycler or hazardous materials handler.
About This Instruction Sheet
Title GasAlertClip Extreme Instruction Sheet
Description This Instruction Sheet is a reference for the
activation, operation and maintenance of the GasAlertClip Extreme 2 or 3 year gas detector.
Part Number 50104153-003
Release Date 05-2012
At a Glance
Liquid cr ystal display(LCD)
Audible alarm
Senso r and sensor g rill
LCD and Elements
C
Pushbutton Functions
Activate detector
Perform self-test
Calibrate oxygen sensor
Perform auto zero
View maximum gas exposure
Transmit data
Infrar ed downlo ad port
LED visual alarms
Alligat or clip
Pushbutton
Self-test status
High and low alarm setpoints
Data transmission
Maximum exposure in alarm
Gas type
Alarm condition
Detector life-remaining indicators
Months-days-hours since last maximum gas exposure alarm
Press and hold C for 5 seconds
Press and hold C for 1 second when LCD displays Te st
Press and hold C for 3 seconds
Press and hold C for 3 seconds
Press C within 24 hours of a gas alarm
Press C when LCD displays Prn and
© 2012 BW TECHNO LOGIES BY HONEYWELL. A LL RIGHTS RESERV ED.
Display gas alarm setpoints
Press C during normal operation
Page 2
Activate the Detector
1. Move to a normal atmosphere (20.9% O2) that is free of hazardous gas.
2. Press and hold the Pushbutton C for 5 seconds.
3. The detector beeps and vibrates once.
4. The LEDs flash once.
5. The detector performs self-test. This make take several minutes. For more information, refer to Self-Test.
6. When Self-Test is successful, the detector beeps and vibrates once and the LEDs flash once. The detector then displays the Detector Life-Remaining Clock.
Note: Once the detector is activated, it cannot be deactivated except
after a battery life-ending alarm. For more information, refer to
Automatic Battery Test and Safety Shutdown Mode.
Self-Test
The detector performs self-test before any other function. Self-test must be performed in a normal atmosphere (20.9% O2) that is free of hazardous gas. After self-test is performed, wait 30 seconds before using the detector to ensure that it accurately detects gas.
When is self-test required?
 Prior to each day’s use; and  When the LCD displays Test.
Perform Self-Tes t
1. Move to a normal atmosphere (20.9% O2) that is free of hazardous gas.
2. Press and hold C for 1 second.
3. Confirm that the following tests are performed: a) The detector beeps and vibrates once. b) The LEDs flash once. c) The LCD displays all elements. d) The LCD flashes Test while sensor integrity and battery life
are tested. e) The LCD displays low and high alarm setpoints. f) If an alarm has occurred in the last 24 hours, the LCD displays
the maximum gas exposure value and the hours that have elapsed
since the exposure. g) The LCD flashes Prn and .
4. After self-test is performed, wait 30 seconds before using the detector to ensure that it accurately detects gas.
Self-Test Pass
When self-test is successful, the detector beeps and vibrates once. The LCD displays u to confirm that self-test was successful. Test displays again 22 hours after performing self-test to indicate that a new self-test is required.
Self-Test Fail
If self-test fails, the detector beeps 5 times and the LEDs flash before the LCD displays a blank screen. The LCD then returns to the normal operation screen and displays Tes t. Repeat self-test.
If self-test fails 3 consecutive times, the LCD displays a blank screen and the detector deactivates. For more information, refer to Safety Shutdown Mode.
Automatic Battery Test
The detector automatically tests the battery every 2 hours.
Battery Test Fail
If a battery test fails, another test is initiated 30 minutes later. After 5 consecutive battery test failures, the LCD displays a blank screen and the detector deactivates. For more information, refer to Safety
Shutdown Mode.
Detector Life-Remaining Clock
The detector life-remaining clock counts down the amount of operating time remaining in months, then days, then hours.
Detector Life-Ending Alarm
The detector life-ending alarm occurs when the battery life-remaining clock displays 0 hours. The detector continues to operate for a maximum of 8 hours after the detector life-ending alarm activates. To deactivate the detector, press C.
Audible: 1 beep every 30 seconds Visual: 1 flash every 30 seconds Vibration: 1 slow vibration every 30 seconds
Gas Alarms
Low Gas Alarm
Audible: 1 slow beep ever y second Visual: 1 slow flash ever y second Vibration: 1 slow vibration every second
High Gas Alarm
Audible: 2 fast beeps every second Visual: 2 fast flashes every second Vibration: 2 fast vibrations ever y second
Battery life decreases rapidly when the detector is in alarm condition. When alarm gas level returns to an acceptable range, the alarm deactivates.
Gas Alarm Setpoints
Gas alarm setpoints are factory configured and cannot be modified.
Display Setpoints
C during normal operation.
Press
Factory Setpoints
Detector Model Low Alarm High Alarm
GasAlertClip Extreme O
GasAlertClip Extreme CO 35 ppm 200 ppm
GasAlertClip Extreme H2S 10 ppm 15 ppm
GasAlertClip Extreme H2S 5 ppm 10 ppm
GasAlertClip Extreme SO25 ppm 10 ppm
Customize Setpoints
Detectors may be configured with customer-specified alarm setpoints. For more information, contact BW Technologies by Honeywell.
19.5 % 23.5%
2
50104153-003/05-2012 || GASALERTCLIP EXTR EME INSTRUCTION SHEET WWW.GASMONITORS.COM
Page 3
2 OR 3 YEAR GAS DETECTOR
Safety Shutdown Mode
The detector initiates safety shutdown mode when:
 Self-test fails 3 consecutive times;  Automatic battery test fails 5 consecutive times; or  The detector has not been manually deactivated within 8 hours of
detector life-ending alarm activation.
If the event logs have not been downloaded, contact BW Technologies by Honeywell.
What happens during safety shutdown?
 The LCD displays a blank screen when the detector initiates safety
shutdown mode.
 The detector beeps and vibrates.  The LEDs flash 2 times per second for 15 seconds.  The LCD displays OFF or an error code, depending on the reason for
the safety shutdown.
Auto Zero (H2S, SO2, and CO models only)
Auto zero can only be performed when the low alarm setpoint is below the setpoints shown in the following table:
Low Alarm
Model
GasAlertClip Extreme CO 25 ppm
GasAlertClip Extreme H2S 5 ppm
GasAlertClip Extreme SO23 ppm
Setpoint
Perform Auto Zero
1. Move to a normal atmosphere (20.9% O2) that is free of hazardous gas.
2. Press and hold C for 3 seconds. If self-test has not been in the previous 22 hours, the detector performs self-test.
3. The LCD displays ZEr ppm and the gas type.
4. The LCD displays low alarm setpoint and high alarm setpoint.
5. If an alarm has occurred in the past 24 hours, the LCD displays the maximum gas exposure value and the hours that have elapsed since exposure began.
6. The LCD flashes Prn and .
completed
Maximum Gas Exposure
When a gas exposure alarm initiates, the detector records the maximum gas exposure level and begins calculating the number of hours elapsed since the maximum exposure alarm began.
For each new exposure greater than the current maximum exposure, the detector resets the maximum gas exposure level to the new level and resets hours to 0.
After 24 hours of gas readings in the acceptable range, the detector resets maximum gas exposure level and hours to 0.
View Maximum Gas Exposure
1. Within 24 hours of receiving a gas alarm, press
2. The LCD displays the low and high alarm setpoints.
3. If a maximum gas exposure has occurred within 24 hours, the LCD displays the maximum gas exposure level and number of hours elapsed since the maximum exposure alarm began.
4. If the maximum gas exposure is beyond the detection range, the LCD displays OL (over limit).
C.
Calibrate Oxygen (O2 ) Detector
O2 detector calibration must be performed in a safe area that is free of hazardous gas in an atmosphere of 20.9% oxygen.
When is oxygen detector calibration required?
 Every 30 days; or
 When the LCD displays CAL and the detector life-remaining value. A calibration may also be performed if the O2 detector has not been used recently or is in a false alarm state.
Perform O2 Detector Calibration
1. Move to a normal atmosphere (20.9% O2) that is free of
hazardous gas.
2. Press and hold C for 3 seconds.
in the previous 22 hours, the detector performs self-test.
3. The detector beeps and flashes once. The LCD displays CAL and O2.
If self-test has not been completed
O2 Detector Calibration Pass
When O2 detector calibration is successful, the detector beeps and vibrates once, then returns to normal operation.
O2 Detector Calibration Fail
If O2 detector calibration fails, the detector does not beep or vibrate. Repeat O2 detector calibration. If the second attempt fails, contact BW Technologies by Honeywell.
Transmit Gas Event Data
The detector stores the last 10 events such as maximum gas exposure, bump test, and auto zero. The recorded data includes:
 detector serial number;
 detector life-remaining values (months/days/hours);
 self-test performed;
 total number of events that have occurred ;
 event type;
 duration of all events;
 gas type;
 alarm level(s) in ppm or %;
 time elapsed since the alarm occurred (days/hours /minutes); and
 duration of the alarm (minutes/seconds). Transmit data directly to the IR Handheld Printer. Use the IR DataLink to transmit gas event data to a personal computer (PC).
Transmit Data to a Personal Computer
1. Connect the IR DataLink to the computer.
2. Place the detector 2 inches from the IR DataLink.
3. On the detector, press C to access the transmission screen.
4. The LCD flashes Prn and . Within 5 seconds, press C to begin
data transmission.
5. During data transmission, the LCD displays Y and flashes. The
LCD displays a percentage value to indicate how much data remains to be transmitted.
6. The detector beeps and vibrates once when transmission is complete. For more information, refer to the IR DataLink User Manual.
Transmit Data to an IR Handheld Printer
1. Place the detector 2 inches from the IR Handheld Printer.
2. On the detector, press C to access the transmission screen.
3. The LCD flashes Prn and . Within 5 seconds, press C to begin
data transmission.
4. During data transmission, the LCD displays Y and flashes. The
LCD displays a percentage value to indicate how much data remains to be transmitted.
5. The detector beeps and vibrates once when transmission is complete. For more information, refer to the IR Handheld Printer User Manual.
50104153-003/05-2012 || GASALERTCLIP EXTR EME INSTRUCTION SHEET WWW.GASMONITORS.COM
Page 4
2 OR 3 YEAR GAS DETECTOR
General Specifications
Shelf life: 1 year before activation Weight : 76 g (2.7 oz.) Instrument dimensions:
28 x 50 x 81 mm (1.1 x 2.0 x 3.2 in.)
Operating temperatures :
H2S: -40 to +50°C (-40 to +122°F) CO: -30 to +50°C (-22 to +122°F) SO2: -30 to +50°C (-22 to +122°F) O2: -20 to +50°C (-4 to +122°F)
Internal vibrating operates to :
-15°C (+5°F)
Operating humidity: 5% to 95%
relative humidit y (non-condensing)
Audible alarm: 95 dB at 30 cm (1 ft.) Visual alarm: Flashing, wide-
angled alarm lens with quad red LEDs plus alarm LCD readout
Display: Liquid crystal display (LCD) Sensor type : Electrochemical cells Detection technique:
Instantaneous alarm
Battery: Lithium, non-replaceable Ingress protection : IP 66 /67 Alarm setpoints: Instant
low and instant high
Calibration:
H2S, CO, SO2: Not required O2: Self-c alibrating
Note: This product has been
classified for use in atmospheres not more than 21% v/v O2.
Safety Specifications
Maximum Operating Life,
24-month detector: 2 years
after activation, assuming 3 to 5 minutes of alarm time per day
Maximum Operating Life,
36-month detector: 3 years
after activation, assuming 1.5 minutes of alarm time per day
Detection Range:
H2S: 0 to 100 ppm CO: 0 to 300 ppm SO2: 0 to 100 ppm O2: 0 to 30% by volume
Ratings and Certifications
Classified by UL to both U.S. and
Canadian Standards as intrinsically safe for Class I, Division 1, Group A , B, C, D and Class I, Zone 0, Group IIC
ATE X:
CE 0539 II 1G Ex ia IIC T4 Ga IP66/67 DEMKO 08 ATEX 0814213
IECEx:
Ex ia IIC T4 Ga IP66/67 IECEx UL 08.0018
CE:
European Conformity ABS Type Approved VA-348-169-X
EC Declaration of Conformity
www.gasmonitors.com/ Declarations_of_Conformity
Event Logging Specifications
Number of stored events : Up to 10
events encountered. If more than 10 events are encountered, older events are replaced by the newer events.
Data transmission methods : Via
infrared port to thermal printer, or via IR DataLink to computer (for ordinary locations only)
Information transmit ted : Serial
number; life remaining; self-tests performed; total number and duration of all events encountered
Last 10 events: Ma ximum exposure;
MicroDock II bump check; O2 calibration, or auto zero
Data shown for maximum
exposures and bump checks:
 Gas type and alarm level in ppm
or %
 Time elapsed since the alarm
occurred in days, hours, and minutes
 Duration of alarm in minutes and
seconds
 Data transmission time: 45 seconds
plus 10 seconds per record
This equipment has been tested and
found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC rules and ICES­003 Canadian EMI requirements. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful
interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
 Reorient or relocate the receiving
antenna.
 Increase the separation bet ween
the equipment and the receiver.
 Connect the equipment into an
outlet on a circuit different from that to which the receiver is connected.
 Consult the dealer or an
experienced radio/TV technician for help.
Limited Warranty and Limitation of Liability
BW Technologies (BW) warrants this
product to be free from defects in material and workmanship under normal use and ser vice for a period of two or three years (depending upon detector), beginning on the date of activation. This Warranty is valid only if the detector is activated by the date on the package. This warranty extends only to the sale of new and unused products to the original buyer.
BW’s warranty obligation is limited, at
BW’s option, to refund of the purchase price, repair, or replacement of a defective product that is returned to a BW authorized service center within the warrant y period. In no event shall BW’s liability hereunder exceed the purchase price actually paid by the buyer for the Product.
This warranty does not include:
a) fuses, disposable batteries or
the routine replacement of parts
due to the normal wear and tear
of the product arising from use;
b) any product which in BW’s
opinion, has been misused,
altered, neglected or damaged by
accident or abnormal conditions
of operation, handling or use; or
c) any damage or defects attributable
to repair of the product by any
person other than an authorized
dealer, or the installation of
unapproved parts on the product
The obligations set forth in this
warranty are conditional on: d) proper storage, installation,
calibration, use, maintenance
and compliance with the product
manual instructions and any other
applicable recommendations of BW;
e) the buyer promptly notifying BW
of any defect and, if required,
promptly making the product
available for correction. No goods
shall be returned to BW until
receipt by the buyer of shipping
instructions from BW; and
f) the right of BW to require that
the buyer provide proof of
purchase such as the original
invoice, bill of sale or packing
slip to establish that the product
is within the warranty period.
THE BUYER AGREES THAT THIS
WARRANTY IS THE BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. BW SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT OR RELIANCE OR ANY OTHER THEORY.
Since some countries or states do
not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this warrant y is held invalid or unenforceable by a court of competent jurisdiction, such holding will not af fect the validity or enforceabilit y of any other provisions.
Warranty Registration
Register the product warranty at:
www.gasmonitors.com/support
Contact BW Technologies by Honeywell
Corporate Headquarters
2840 - 2 Avenue SE Calgar y AB Canada T2A 7X9
America
405 Barclay Blvd. Lincolnshire IL USA 60069
Europe
Hatch Pond House 4 Stinsford Road Nuffield Industrial Estate Poole Dorset BH17 0RZ United Kingdom
Tel ephon e
Toll Free Canada: 1-800-663-4164 Toll Free Europe: + 44(0)1295 700300 Toll Free USA: 1-888-749-8878 Other Countries: +1-403 -248-9226
Email
info@gasmonitors.com
Web
www.gasmonitors.com
50104153-003/05-2012 || GASALERTCLIP EXTR EME INSTRUCTION SHEET WWW.GASMONITORS.COM