2 OR 3 YEAR H2S, CO, O2, SO2 SINGLE-GAS DETECTOR
BW Clip is a portable, single-gas detector manufactured by BW Technologies by Honeywell (BW). BW Clip continuously measures
the concentration of a specic gas in the ambient environment, and activates an alarm when concentrations exceed alarm setpoints. It is your
responsibility to respond appropriately to the alarms.
BW Clip is a member of the GasAlert family of products. It is compatible with IntelliDoX and MicroDock II automatic test and calibration stations, as well as Fleet Manager II software version 4.0 or higher. For more information, visit the product website at www.honeywellanalytics.com
Safety Information: Read First
1. Substitution of components may impair Intrinsic Safety.
2. Periodically test the response of the sensor by exposing the detector
to a target gas concentration that exceeds the low alarm setpoint. Manually verify that the audible and visual alarms are activated.
3. Detectors in hibernation mode are not functional.
1. Activate the detector before the activation date on the package.
2. This product is a gas detector, not a measurement device.
3. Ensure that the sensor grill is free of dirt, debris, and is not obstructed.
4. Clean the exterior with a soft, damp cloth.
5. Bump test the detector in a normal atmosphere that is free of hazardous gas.
6. For optimal performance, periodically zero the sensor in a normal
atmosphere (20.9% v/v O2) that is free of hazardous gas.
Informations de sécurité – À lire au préalable
AVERTISSEMENT
1. Le remplacement d’un composant de l’appareil peut compromettre la sécurité intrinsèque du détecteur.
2. Testez périodiquement la réponse du capteur en exposant le détecteur à une concentration de gaz supérieure au seuil d’alarme basse. Vériez manuellement que les alarmes sonores et visuelles sont activées.
3. Détecteurs en mode veille prolongée ne sont pas fonctionnels.
MISE EN GARDE
1. Activez le détecteur avant la date d’activation sur l’emballage.
2. Ce produit est un détecteur de gaz, pas un dispositif de mesure.
3. Veillez à assurer que la grille du capteur est exempt de saleté, les débris, et ne soit pas obstrué.
4. Nettoyez l’extérieur de l’appareil à l’aide d’un chiffon doux et humide.
5. Veillez à effectuer un test fonctionnel du capteur dans une atmosphère
normale qui est exempte de gaz dangereux
6. Pour des performances optimales, zéro périodiquement le capteur
dans une atmosphère normale (20,9% v/v O2) qui est exempte de gaz dangereux.
BW Clip Factory Calibration Certicate
This instrument was factory inspected, tested and calibrated in accordance with the conditions and
requirements of our registered Quality System, Operating Standards and Sales Agreements.
Part Number 50115903-067 | OM-EN-PALO-A4 | BW Clip Operator Manual
Getting Started
visual alarms
visual alarms
LCDLCD
alarm
alarm
Bubble label
Bubble label
with product name
with product name
LCD and Symbols
Pushbutton prompt. When this symbol is displayed, press and hold the pushbutton until the symbol disappears.
BW Clip O2 models only. Oxygen concentration is measured as percent by volume.
BW Clip H2S, CO and SO2 models only. Toxic gas concentration is measured in parts per million.
Low alarm and high alarm. These symbols are
displayed when a gas concentration that exceeds
alarm setpoints is detected.
Non-compliance warning. This symbol is displayed when a diagnostic or compliance test fails; when the
operating life countdown reaches 24 hours or less; and for up to 30 days after operating life expiry. It is also displayed when certain notications are enabled through Fleet Manager II via an IntelliDoX station.
Real-time clock. This symbol is displayed when real­time clock is enabled through Fleet Manager II via an IntelliDoX station.
End of operating life warning. This symbol is displayed
when the operating life countdown reaches 24 hours
or less.
Peak reading. This symbol is displayed when a maximum gas exposure event has occurred within the past 24 hours.
Bump test reminder. If the detector is congured to
display a bump test reminder, this symbol is displayed when a bump test is due. indicates a failed test.
BW Clip O2 models only. This symbol is displayed when more than 24 hours have elapsed since the last
successful zero procedure was performed. indicates a failed test.
Operating life remaining in months (MM) or days (DD).
sensor grill
sensor grill
clip
BW Clip
2 OR 3 YEAR H2S, CO, O2, SO2 SINGLE-GAS DETECTOR
Activate a New Detector
1. Move to a normal atmosphere (20.9% v/v O2) that is free of hazardous gas.
2. Press and hold the pushbutton until a 5 second countdown is displayed, and then continue to hold until the countdown is complete.
3. When the countdown is complete, the LCD and LEDs turn on and then turn off. The detector performs an internal diagnostic test.
4. When the diagnostic test is successful, the alarm setpoints are displayed and the LEDs ash.
5. The sensor stabilization countdown is displayed. During the countdown, the detector vibrates continuously for 20 seconds. The time required to stabilize varies depending on sensor type. When the countdown reaches 0, the LEDs ash and the audible alarm beeps.
6. When the type of gas detected and the operating life countdown are displayed, the detector is in normal operating mode.
Normal Operating Mode
When the detector is in normal operating mode, the type of gas detected is permanently displayed. The remaining operating life is displayed until it is disrupted by a
pushbutton action, gas alarm, or error event.
Operating Life
When the detector is activated for the rst time, the
operating life countdown is displayed. The countdown displays remaining operating life in months until the
countdown reaches 90 days or less before expiry.
90 Days or Less Remaining
When the operating life countdown reaches 90 days or less before expiry, the countdown display changes to
days. The countdown displays the remaining operating
life in days until the countdown reaches 24 hours or less before expiry.
24 Hours or Less Remaining
When the operating life countdown reaches 24 hours or less before expiry, the countdown display changes to
hours, and the EXPIRY! warning and non-compliance symbol are displayed.
End of Operating Life
When the operating life countdown ends, the detector
deactivates and detector safety functions are disabled. It is possible to retrieve event logs for a limited time after expiry. The EXPIRY! warning and non-compliance symbol are displayed for up to 30 days after expiry.
Internal Diagnostic Tests
Activated detectors automatically perform one internal diagnostic test every 24 hours. If the internal diagnostic test fails, the diagnostic fail-safe begins.
Diagnostic Fail-Safe
1. The EXPIRY! warning and non-compliance symbol are displayed.
2. The detector beeps and vibrates.
3. The LEDs ash 2 times per second for 15 seconds.
4. OFF or an error code is displayed. The detector is deactivated.
It is possible to retrieve the event logs for a limited time after expiry. If you are unable to retrieve the event logs, contact BW.
Alarms and Warnings
An alarm is initiated when the sensor is exposed to a gas concentration that exceeds alarm setpoints. The alarm persists until the alarm
gas concentration returns to an acceptable range. You can use
Fleet Manager II via an IntelliDoX station to enable the display of the gas
concentration reading during an alarm. Battery life decreases rapidly when the detector is in alarm condition.
Low Alarm
Audible: 1 beep per second Visual: 1 ash per second Vibration: 1 vibration per second
High Alarm
Audible: 2 beeps per second Visual: 2 ashes per second Vibration: 2 vibrations per second
Over-Limit and Under-Limit Alarms
If the gas concentration is beyond the sensor range, OL (over-limit) or -OL (under-limit) is displayed.
Audible: 2 beeps per second Visual: 2 ashes per second Vibration: 2 vibrations per second
Automatic Zero Reminder for O2 Models
For O2 models only, the Automatic Zero Reminder is displayed when more than 24 hours have elapsed since
the last successful zero procedure was performed.
Peak Readings
The peak readings symbol is displayed when the sensor is exposed to a gas concentration that exceeds the alarm setpoints. It is no longer displayed when more than 24 hours have passed since the last alarm, or when a
successful bump test is performed via an IntelliDoX or MicroDock II station.
Non-compliance Symbol and LED Indicators
The non-compliance warning symbol is displayed when:
• an internal diagnostic test fails;
• a bump test or zero procedure fails;
• a bump test is due, if enabled;
the sensor is exposed to a gas concentration that exceeds alarm
setpoints, if enabled.
The non-compliance warning symbol is no longer displayed when more
than 24 hours have passed since the last alarm, or when a successful bump test is performed via an IntelliDoX or MicroDock II station. The non­compliance warning symbol is permanently displayed when the service-life countdown reaches 24 hours or less before expiry, and for up to 30 days after expiry.
When Non-compliance after gas exposure is enabled through
Fleet Manager II via an IntelliDoX station, the LEDs ash in an alternating
pattern when the non-compliance warning symbol is displayed. When the
symbol is rst displayed, the top LEDs ash. After 30 seconds, the side LEDs ash. After 30 seconds, the pattern repeats until the non-compliance
warning symbol is no longer displayed.
NOTE: When the detector is bump tested via a MicroDock II station, an abnormal test result may occur if non-compliance LEDs are ashing. To suspend the ashing for up to 10 minutes, press and release the pushbutton. When the ashing stops, insert the detector and perform the test. If the bump test fails, the LEDs resume ashing.