BW Technologies BWC3R-M, BWC3R-M30200, BWC2R-H515, BWC2R-X, BWC2R-H User manual

...
BW Clip
PushbuttonPushbutton
Sensor and
Sensor and
and gas type
and gas type
Audible
Audible
LED
LED
Infrared port
Alligator
Product label
2 OR 3 YEAR H2S, CO, O2, SO2 SINGLE-GAS DETECTOR
BW Clip is a portable, single-gas detector manufactured by BW Technologies by Honeywell (BW). BW Clip continuously measures the
concentration of a specific gas in the ambient environment, and activates an alarm when concentrations exceed alarm setpoints. It is your responsibility to respond
appropriately to the alarms.
Part Number 50115903-067 | OM-EN-PALO-A8 | BW Clip Operator Manual
Getting Started
visual alarms
visual alarms
Safety Information: Read First
a WARNING
1. Substitution of components may impair Intrinsic Safety.
2. Periodically test the response of the sensor by exposing the detector to a
target gas concentration that exceeds the low alarm setpoint. Manually verify that the audible and visual alarms are activated.
3. Detectors in hibernation mode are not functional.
a CAUTION
1. Activate the detector before the activation date on the package.
2. This product is a gas detector, not a measurement device.
3. Ensure that the sensor grill is free of dirt, debris, and is not obstructed.
4. Clean the exterior with a soft, damp cloth.
5. Bump test the detector in a normal atmosphere that is free of hazardous gas.
6. For optimal performance, periodically zero the sensor in a normal atmosphere
(20.9% v/v O
) that is free of hazardous gas.
2
Informations de sécurité – À lire au préalable
AVERTISSEMENT
Le remplacement d’un composant de l’appareil peut compromettre la sécurité
1.
intrinsèque du détecteur.
Testez périodiquement la réponse du capteur en exposant le détecteur à une
2.
concentration de gaz supérieure au seuil d’alarme basse. Vérifiez manuellement que les alarmes sonores et visuelles sont activées.
3. Détecteurs en mode veille prolongée ne sont pas fonctionnels.
MISE EN GARDE
1. Activez le détecteur avant la date d’activation sur l’emballage.
2. Ce produit est un détecteur de gaz, pas un dispositif de mesure.
3. Veillez à assurer que la grille du capteur est exempt de saleté, les
débris, et ne soit pas obstrué.
4. Nettoyez l’extérieur de l’appareil à l’aide d’un chiffon doux et humide.
5. Veillez à effectuer un test fonctionnel du capteur dans une atmosphère normale
qui est exempte de gaz dangereux
6. Pour des performances optimales, zéro périodiquement le capteur dans une
atmosphère normale (20,9% v/v O2) qui est exempte de gaz dangereux.
LCDLCD
alarm
alarm
Bubble label
Bubble label
with product name
with product name
LCD and Symbols
Pushbutton prompt. When this symbol is displayed, press and hold the pushbutton until the symbol disappears.
BW Clip O2 models only. Oxygen concentration is measured as percent by volume.
BW Clip H concentration is measured in parts per million.
Low alarm and high alarm. These symbols are displayed when a gas concentration that exceeds alarm setpoints is detected.
Non-compliance warning. This symbol is displayed when a diagnostic or compliance test fails; when the operating life
countdown reaches 24 hours or less; and for up to 30 days after operating life expiry. It is also displayed when certain notifications are enabled through Fleet Manager II via an IntelliDoX station.
Real-time clock. This symbol is displayed when real-time clock is enabled through Fleet Manager II via an IntelliDoX station.
End of operating life warning. This symbol is displayed when the operating life countdown reaches 24 hours or less.
Peak reading. This symbol is displayed when a maximum gas exposure event has occurred within the past 24 hours.
Bump test reminder. If the detector is configured to
bump test reminder, this symbol is displayed when a bump test is due.
BW Clip O than 24 hours have elapsed since the last successful zero
procedure was performed. indicates a failed test.
Operating life remaining in months (MM) or days (DD).
clip
sensor grill
sensor grill
S, CO and SO2 models only. Toxic gas
2
display a
indicates a failed test.
models only. This symbol is displayed when more
2
This instrument was factory inspected, tested and calibrated in accordance with the conditions and
BW Clip Factory Calibration Certificate
requirements of our registered Quality System, Operating Standards and Sales Agreements.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
BW Clip
2 OR 3 YEAR H2S, CO, O2, SO2 SINGLE-GAS DETECTOR
Activate a New Detector
1. Move to a normal atmosphere (20.9% v/v O2) that is free of hazardous gas.
2. Press and hold the pushbutton until a 5 second countdown is displayed, and then continue to hold until the countdown is complete.
3. When the countdown is complete, the LCD and LEDs turn on and then turn off. The detector performs an internal diagnostic test.
4. When the diagnostic test is successful, the alarm setpoints are
displayed and the LEDs ash.
5. The sensor stabilization countdown is displayed. During the countdown,
the detector vibrates continuously for 20 seconds. The time required to stabilize varies depending on sensor type. When the countdown reaches 0, the LEDs ash and the audible alarm beeps.
6. When the type of gas detected and the operating life countdown are displayed, the detector is in normal operating mode.
Normal Operating Mode
When the detector is in normal operating mode, the type of gas detected is permanently displayed. The remaining operating life is displayed until it is disrupted by a
pushbutton action, gas alarm, or error event.
Operating Life
When the detector is activated for the rst time, the
operating life countdown is displayed. The countdown displays remaining operating life in months until the
countdown reaches 90 days or less before expiry.
90 Days or Less Remaining
When the operating life countdown reaches 90 days or less before expiry, the countdown display changes to
days. The countdown displays the remaining operating
life in days until the countdown reaches 24 hours or less before expiry.
24 Hours or Less Remaining
When the operating life countdown reaches 24 hours or less before expiry, the countdown display changes to
hours, and the EXPIRY! warning and non-compliance symbol are displayed.
End of Operating Life
When the operating life countdown ends, the detector
deactivates and detector safety functions are disabled. It is possible to retrieve event logs for a limited time after
expiry. The EXPIRY! warning and non-compliance symbol
are displayed for up to 30 days after expiry.
Internal Diagnostic Tests
Activated detectors automatically perform one internal diagnostic test every 24 hours. If the internal diagnostic test fails, the diagnostic fail-safe begins.
Diagnostic Fail-Safe
1. The EXPIRY! warning and non-compliance symbol are displayed.
2. The detector beeps and vibrates.
3. The LEDs ash 2 times per second for 15 seconds.
4. OFF or an error code is displayed. The detector is deactivated.
It is possible to retrieve the event logs for a limited time after expiry. If you are unable to retrieve the event logs, contact BW.
Alarms and Warnings
An alarm is initiated when the sensor is exposed to a gas concentration that exceeds alarm setpoints. The alarm persists until the alarm
gas concentration returns to an acceptable range. You can use
Fleet Manager II via an IntelliDoX station to enable the display of the gas
concentration reading during an alarm. Battery life decreases rapidly when the detector is in alarm condition.
Low Alarm
Audible: 1 beep per second Visual: 1 ash per second Vibration: 1 vibration per second
High Alarm
Audible: 2 beeps per second Visual: 2 ashes per second Vibration: 2 vibrations per second
Over-Limit and Under-Limit Alarms
If the gas concentration is beyond the sensor range, OL (over-limit) or -OL (under-limit) is displayed.
Audible: 2 beeps per second Visual: 2 ashes per second Vibration: 2 vibrations per second
Automatic Zero Reminder for O2 Models
For O
models only, the Automatic Zero Reminder is
2
displayed when more than 24 hours have elapsed since
the last successful zero procedure was performed.
Peak Readings
The peak readings symbol is displayed when the sensor is exposed to a gas concentration that exceeds the alarm setpoints. It is no longer displayed when more than 24 hours have passed since the last alarm, or when a
successful bump test is performed via an IntelliDoX or MicroDock II station.
Non-compliance Symbol and LED Indicators
The non-compliance warning symbol is displayed when:
• an internal diagnostic test fails;
• a bump test or zero procedure fails;
• a bump test is due, if enabled;
the sensor is exposed to a gas concentration that exceeds alarm setpoints, if enabled.
The non-compliance warning symbol is no longer displayed when more
than 24 hours have passed since the last alarm, or when a successful bump test is performed via an IntelliDoX or MicroDock II station. The non­compliance warning symbol is permanently displayed when the service-life countdown reaches 24 hours or less before expiry, and for up to 30 days after expiry.
When Non-compliance after gas exposure is enabled through
Fleet Manager II via an IntelliDoX station, the LEDs ash in an alternating
pattern when the non-compliance warning symbol is displayed. When the
symbol is rst displayed, the top LEDs ash. After 30 seconds, the side LEDs ash. After 30 seconds, the pattern repeats until the non-compliance
warning symbol is no longer displayed.
NOTE: When the detector is bump tested via a MicroDock II station, an abnormal test result may occur if non-compliance LEDs are ashing. To suspend the ashing for up to 10 minutes, press and release the pushbutton. When the ashing stops, insert the detector and perform the test. If the bump test fails, the LEDs resume ashing.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Recall Events and Settings
When the detector is in normal operating mode, press the pushbutton
at any time to scroll through events and settings that are not empty or disabled through Fleet Manager II via an IntelliDoX station. Events and
settings are displayed in this order:
1. Peak Reading, if available, and when a maximum gas exposure event
has occurred within the past 24 hours.
2. Time of Peak Reading, if Real-time Clock display is enabled.
3. Low Alarm Setpoint
4. High Alarm Setpoint
5. Real-time Clock, if enabled.
6. Firmware Version
7. Next Bump Due, if enabled.
8. Hibernation, for 2-year H
While an event or setting is displayed, press the pushbutton to scroll to the next available event or setting. If you do not press the pushbutton, or if you have reached the last available event or setting, the detector returns to
normal operation.
S and CO hibernation models only.
2
Zero the Sensor
Over time and through use, the sensor baseline at zero exposure may drift from the manufacturer’s baseline. For optimal performance of O BW recommends that you zero the O2 sensor once every 24 hours or
when the Automatic Zero Reminder is displayed. For all other models, BW recommends that you zero the sensor periodically.
models,
2
Zero Procedure
1. Move to a normal atmosphere (20.9% v/v O2) that is free of hazardous gas.
2. Press and hold the pushbutton until a 5 second countdown is displayed, and then continue to hold until the countdown is complete.
3. When the countdown is complete, the Zero procedure begins and ZErO is displayed.
4. When Zero is successful, PASS is displayed, and then the detector returns to normal operation.
5. If Zero is not successful, the non-compliance LED
ashes. FAIL and the non-compliance warning symbol
are displayed. Press the pushbutton to acknowledge
the result and return to normal operation. The non­compliance warning symbol is displayed and the non-
compliance LED ashes. Perform the Zero procedure
again. If the procedure fails again, contact BW.
Bump Tests
When the detector is congured to display a bump test reminder, the bump
test symbol is displayed when a bump test is due. For best results, bump
test the detector via an IntelliDoX or MicroDock II station. Bump tests performed via an IntelliDoX or MicroDock II station are logged as bump tests. Manual bump tests are logged as unsafe gas concentrations.
Event Logs
The detector stores the last 35 events that occurred, including peak
readings, bump tests and auto zeros. Each record contains:
• detector serial number, sensor type and life-remaining;
total number of events that have occurred;
event type and duration;
alarm level(s) in ppm or %;
time elapsed since the alarm occurred in days/hours/minutes;
duration of the alarm (minutes/seconds).
Use Fleet Manager II via an IntelliDoX or MicroDock II station to transfer event logs from the detector to a computer. When logs are transferred via an IntelliDoX station, the most recent 35 events are transferred. When logs are transferred via a MicroDock II station, the most recent 10 events are
transferred.
Optional Settings
Use Fleet Manager II via an IntelliDoX station to enable the following
optional settings for BW Clip detectors.
Display Gas Reading During Alarm
When Display gas reading during alarm is enabled, the
gas concentration is displayed when a gas alarm event is
detected. The gas concentration is displayed until the
alarm event ends.
Non-compliance After Gas Exposure
When Non-compliance after gas exposure is enabled, the LEDs ash in
an alternating pattern when the sensor is exposed to a gas concentration that exceeds alarm setpoints. When the gas exposure is detected, the top LEDs ash. After 30 seconds, the side LEDs ash. After 30 seconds, the pattern repeats until more than 24 hours have passed since the last alarm, or when a successful bump test is performed via an IntelliDoX or MicroDock II station.
Real-time Clock Display
When Real-time Clock Display is enabled, you can display the current time by pressing the pushbutton to Recall Events and Settings, and then continuing to press the pushbutton until the time is displayed.
Time may be displayed in 12-hour or 24-hour clock format.
Hibernation
Hibernation is a factory installed feature that is available for 24-month
H2S and CO detectors. When Hibernation is activated, the operating life countdown is suspended. Use Hibernation to deactivate the detector for
7 days or longer, up to a maximum of 12 months. When Hibernation is
activated, all detector safety functions are disabled.
Activate Hibernation via BW Clip Hibernation Case
1. Have ready a BW Clip Hibernation Case. To purchase a Case, contact BW or an authorized distributor.
2. Move to a normal atmosphere (20.9% v/v O2) that is free of hazardous gas.
3. Verify that the detector is in normal operating mode.
4. Press the pushbutton to Recall Events and Settings, and then continue to press the pushbutton until Hib is displayed.
5. While Hib is displayed, press the pushbutton until a 5 second countdown is displayed. When the countdown is complete, the 20-second Hibernation counter is displayed.
6. While the Hibernation counter is displayed, place the detector into a
BW Clip Hibernation Case and then close the case rmly. The detector
enters Hibernation mode. All detector safety functions are disabled.
7. To reactivate the detector, open the case.
Activate Hibernation through Fleet Manager II via IntelliDoX
1. Use Fleet Manager II to activate Hibernation on Insertion via an
IntelliDoX station, and then insert the detector in the congured station. The detector event logs are retrieved, and the detector enters
Hibernation mode. All safety functions are disabled, and the detector
may be removed from the IntelliDoX station.
2. To reactivate the detector, press and hold the pushbutton until a 5 second countdown is displayed, and then continue to hold until the
LCD and LEDs turn on and off. The detector performs a self-diagnostic
test. When the test is successful, the type of gas detected and the remaining operating life are displayed. The detector is in normal operating mode.
3. For more information, refer to the operator manuals for the IntelliDoX
station and Fleet Manager II software version 4.0 or higher.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
BW Clip
2 OR 3 YEAR H2S, CO, O2, SO2 SINGLE-GAS DETECTOR
About this Publication
This publication is an operator guide for BW Clip portable single-gas detectors manufactured by BW Technologies by
Honeywell. It is intended for use with the
following models:
24- and 36-month H
24-month O2 detector
24-month SO2 detector
S and CO Detectors
2
Copyright, Notice, Trademarks
While this information is presented in
good faith and believed to be accurate,
BW Technologies by Honeywell (BW) disclaims the implied warranties of
merchantability and tness for a particular purpose and makes no express warranties except as may be stated in its written
agreement with and for its customers.
In no event is BW liable to anyone for any
indirect, special or consequential damages. The content in this document is subject to change.
Maximum Operating Life
24-month Detector
2 years af ter acti vation, assuming
2 minutes of alarm time per day.
Note: 24 -month H2S and CO detectors that use hi bernat ion may extend the service period of the detector up to an additional year. The se rvic e life will end once the detector reaches 24 months of actual operation. Hiber nation provides a maximum service life of 36 months.
36-month Detector
3 years af ter acti vation, assuming
1 minute of alarm time per day.
Detection Range
S 0 to 100 ppm
H
2
CO 0 to 300 ppm
0 to 25% by volume
O
2
0 to 100 ppm
SO
2
Factory Alarm Setpoints
Low Alarm High Alarm H2S 10 ppm 15 ppm CO 35 ppm 200 ppm O2 19.5% vol 23.5% vol SO2 5 ppm 10 ppm
Use Fleet M anager II via an IntelliDoX
station to adjust alarm setpoints. For more information, refer to the operator
manuals for Fleet Manager II software version 4.0 and the IntelliDoX automatic
test and calibration station.
Ratings and Certifications
The BW Clip is in conformity with the following standards:
UL 913 8th Edition UL 60079-0:2013, UL 60079-11:2013
CSA C22.2 No. 157-92:2012, CSA C22.2 No. 60079-0:2011,
CSA C22.2 No. 60079-11:2014
EN 60079-0:2012 +A11:2013, EN 60079-11:2012
IEC 60079-0:2011 IEC 60079-11:2011
ABNT NBR IEC 60079-0:2008 ABNT NBR IEC 60079-11:2009
UL
Classified by UL to both US and
Canadian Standards as intrinsically safe
for Class I, Division 1, Group A , B, C, D and Class I, Zone 0, Group IIC.
ATEX
CE 0539 II 1G Ex ia IIC T4 Ga IP66/67 DEMKO 14 ATEX 1356
CE
European Conformity
IECEx
Ex ia IIC T4 Ga IP66/67 IECEx UL 14.0063
INMETRO
Ex ia IIC T4 Ga DNV 15.0 024
KTL
15-KA4BO-0306
1
Specications
Shelf Life
S One (1) year before activatio n
H
2
CO One (1) year before activation O2 Six (6) months before activation SO2 One (1) year before activation
Instrument Weight alligator clip included
92 grams (3.2 ounces)
Instrument Dimensions alligator clip included
41 x 50 x 87 millimeters (1.6 x 2.0 x 3.4 inches)
Operating Temperatures
S -40° to +50°C (- 40° to +122° F)
H
2
CO -30° to +50°C (-22° to +122°F)
-20° to +50°C (-4° to +122°F)
O
2
- 30° to +5 0°C (-22° to +122°F)
SO
2
Internal Vibrating
Operates to -10°C (+14°F)
Intrinsic Safety Temperature Range
-40 °C to +50 °C (-40 °F to +122 °F)
Internal Vibrating
5% to 95% relative humidity (non­condensing)
Audible Alarm
≈ 95 dB at 10 cm (3.9 in)
Visual Alarm
Flashing, wide-angled alarm lens with red
LEDs plus alarm LCD re adout
Display
Liquid crystal display (LCD)
Sensor Type
Electrochemical cells
Battery
Lithium, non-replaceable
Ingress Protection
IP 66/67
Alarm Setpoints
Instant l ow and inst ant high
Calibration
S not required
H
2
CO not required
not required
SO
2
Perform the Zero procedure
O
2
once ever y 24 hours or when the
Automatic Zero Reminder is displayed.
Intended Use
This pro duct is classied for use in
hazardous atmospheres that are not
more than 21% v/v O
2
Recycling
This instrument contains a lithium battery. Do not mix with the solid waste stream. Spent batteries should be disposed of by a qualied
recycler or hazardous materials handler.
Products may contain materials that are regulated for transportation under domestic and international dangerous goods regulations. Return product in compliance with appropriate dangerous goods regulations. Contact freight carrier for further instructions.
Limited Warranty and Limitation of Liability
BW Technologies (BW) warrants this product to be free from defects in material
and workmanship under normal use and service for a period of two or three years
(depending upon detector), beginning on the
date of activation. 24-month H detectors are covered for up to an additional
12 months when hibernation is used, limited
by a total of 24 months of detector operation. This Warranty is valid only if the detector is activated by the date on the package. This warranty extends only to the sale of new and
unused products to the original buyer. BW’s warranty obligation is limited, at BW’s
option, to refund of the purchase price, repair,
or replacement of a defective product that is returned to a BW authorized service center within the warranty period. In no event shall BW’s liability hereunder exceed the purchase
price actually paid by the buyer for the Product.
This warranty does not include:
• fuses, disposable batteries or the routine replacement of parts due to the normal wear and tear of the product arising from use;
• any product which in BW’s opinion, has been misused, altered, neglected or damaged by accident or abnormal conditions of operation, handling or use; or
S and CO
2
• any damage or defects attributable to repair of the product by any person other than an authorized dealer, or the
installation of unapproved parts on the
product.
The obligations set forth in this warranty are conditional on:
1) proper storage, installation, calibration,
use, maintenance and compliance with the product manual instructions and any other applicable recommendations of BW;
2) the buyer promptly notifying BW of any
defect and, if required, promptly making the product available for correction.
No goods shall be returned to BW until receipt by the buyer of shipping instructions from BW; and
3) the right of BW to require that the buyer
provide proof of purchase such as the original invoice, bill of sale or packing slip
to establish that the product is within the warranty period.
THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. BW SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT OR RELIANCE OR ANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty,
or exclusion or limitation of incidental or
consequential damages, the limitations and
exclusions of this warranty may not apply to every buyer. If any provision of this warranty is held invalid or unenforceable by a court of
competent jurisdiction, such holding will not
affect the validity or enforceability of any other provisions.