Buhler TC-Double 6111 X2, TC-Double X2 Series Original Instructions Manual

Analysentechnik
Installation and Operation Instructions
Original instructions
Sample gas cooler
TC-Double (+) X2
BE440022 10/2018
Tel. +49 (0) 21 02 / 49 89-0, Fax: +49 (0) 21 02 / 49 89-20
E-Mail: analyse@buehler-technologies.com
Internet: www.buehler-technologies.com
Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen Tel. +49 (0) 21 02 / 49 89-0, Fax: +49 (0) 21 02 / 49 89-20 Internet: www.buehler-technologies.com E-Mail: analyse@buehler-technologies.com
Read this instruction carefully prior to installation and/or use. Pay at­tention particularly to all advises and safety instructions to prevent in­juries. Bühler Technologies can not be held responsible for misusing the product or unreliable function due to unauthorised modifications.
All rights reserved. Bühler Technologies GmbH 2018
Document information
Document No......................................................... BE440022
Version..........................................................................10/2018
TC-Double (+) X2
Contents
1 Introduction..................................................................................................................................................................................................................... 3
1.1 Intended use .........................................................................................................................................................................................................3
1.2 Overview ................................................................................................................................................................................................................3
1.3 Type plate ..............................................................................................................................................................................................................4
1.4 Scope of delivery.................................................................................................................................................................................................. 4
1.5 Ordering instructions ........................................................................................................................................................................................ 5
1.5.1 TC-Double X2..........................................................................................................................................................................................5
1.5.2 TC-Double+ X2 ...................................................................................................................................................................................... 6
2 Safety instructions......................................................................................................................................................................................................... 7
2.1 Important notices................................................................................................................................................................................................ 7
2.2 General hazard warnings .................................................................................................................................................................................8
3 Transport and storage ................................................................................................................................................................................................10
4 Installation and connection....................................................................................................................................................................................... 11
4.1 Installation site requirements........................................................................................................................................................................ 11
4.2 Installation ...........................................................................................................................................................................................................11
4.2.1 Flow adapter connection (optional)...............................................................................................................................................11
4.2.2 Connecting the moisture detector (option).................................................................................................................................11
4.2.3 Connecting the peristaltic pump (option) ...................................................................................................................................11
4.2.4 Sample gas pump connection gas lines (optional) ..................................................................................................................12
4.2.5 Connecting the heat exchanger .....................................................................................................................................................12
4.3 Electrical connections .......................................................................................................................................................................................13
4.4 Signal outputs.................................................................................................................................................................................................... 14
5 Operation and control ................................................................................................................................................................................................ 16
5.1 Description of functions.................................................................................................................................................................................. 16
5.2 Use of menu functions .....................................................................................................................................................................................17
5.2.1 Lock Menu............................................................................................................................................................................................. 17
5.2.2 Menu navigation overview..............................................................................................................................................................18
5.3 Description of menu functions .................................................................................................................................................................... 20
5.3.1 Main menu.......................................................................................................................................................................................... 20
5.3.2 Submenu 1 ...........................................................................................................................................................................................20
5.3.3 Submenu 1 (Global Settings) ...........................................................................................................................................................21
5.3.4 Set favourite menu........................................................................................................................................................................... 24
6 Maintenance..................................................................................................................................................................................................................25
7 Service and repair.........................................................................................................................................................................................................26
7.1 Troubleshooting ................................................................................................................................................................................................26
7.1.1 Error messages on the display........................................................................................................................................................27
7.2 Safety instructions ............................................................................................................................................................................................28
7.3 Cleaning and removal of the heat exchanger...........................................................................................................................................29
7.4 Replacing the hoses of the peristaltic pump (option) ............................................................................................................................29
7.5 Replacing the filter element (option) ..........................................................................................................................................................29
7.6 Drying of the moisture detector (option)...................................................................................................................................................30
7.7 Calibration of the moisture detector (option)..........................................................................................................................................30
7.8 Replacing sample gas pump inlet and outlet valves (optional) ..........................................................................................................30
7.9 Replacing the O-ring on the bypass valve (optional) ...............................................................................................................................31
7.10 Replacing the bellow (optional) .....................................................................................................................................................................31
7.11 Spare parts and accessories ...........................................................................................................................................................................32
7.11.1 Consumables and accessories ........................................................................................................................................................32
8 Disposal........................................................................................................................................................................................................................... 33
9 Appendices.....................................................................................................................................................................................................................34
9.1 Gas Cooler Technical Data ..............................................................................................................................................................................34
9.2 Technical Data - Options .................................................................................................................................................................................34
9.3 Flow chart ............................................................................................................................................................................................................35
9.4 Performance curves..........................................................................................................................................................................................36
iBühler Technologies GmbHBE440022 ◦ 10/2018
TC-Double (+) X2
9.5 Heat exchanger ..................................................................................................................................................................................................37
9.5.1 Heat exchanger description ............................................................................................................................................................37
9.5.2 Heat exchanger overview.................................................................................................................................................................37
9.6 Dimensions (mm)..............................................................................................................................................................................................38
10 Attached documents...................................................................................................................................................................................................39
ii Bühler Technologies GmbH BE440022 ◦ 10/2018
TC-Double (+) X2
1 Introduction
1.1 Intended use
This unit is intended for industrial use in gas analysis systems. It's an essential component for conditioning the sample gas to protect the analysis instrument from residual moisture in the sample gas.
This device may be operated in a zone 2, explosion class IIC (ATEX & IECEx), explosive gas atmosphere which is rarely and only temporarily explosive and in Class I, Division 2 (US & Canada).
Optional mounting of various add-on units may limit the area of application of this device.
Cooler with certificate Canada and IECEx
The equipment shall be mounted within a tool-secured IP54 enclosure.
Cooler with certificate USA
The equipment shall be mounted within a tool-secured IP54 enclosure which is capable of accepting one or more of the ClassIDivision2 wiring methods per the National Electric Code (ANSI/NFPA70).
Cooler with certificate ATEX
The equipment shall be mounted within a IP54 enclosure in accordance with EN60079-15, and in a tool-secured enclosure which meets the requirements of EN60079-0 and EN60079-15.
Please note the specifications in the data sheet on the specific intended use, existing material combinations, as well as pressure and temperature limits.
1.2 Overview
The TC-Double X2 series was designed specifically for high cooling capacities, high ambient temperatures and to cool in two cycles to minimise wash out effects. The two cooling blocks on the TC-Double+ X2 can be set do different temperatures.
Any other use of this cooler is determined by the type of built-in passive pre-cooling, i.e. the first cooling level is not controlled electronically.
The Peltier cooler is distinguished by two types according to cooling capacity or operation temperature. This classification is re­flected in the type designation. The exact item number of the model defined by you is determined by the model code in the cat­egory ordering information.
Application Standard applications
Operation temperature 40 °C 50 °C 2 heat exchangers in series TC-Double 6111 X2 TC-Double 6112 X2 2 heat exchangers in series TC-Double+ 6111 X2 TC-Double+ 6112 X2
Additional components which every conditioning system should feature can optionally be integrated:
– Peristaltic pump for condensate separation,
– Filter,
– Moisture detector,
– Sample gas pump.
This allows for various configurations of cooler and options. Here the approach is to simplify creating a complete system in a cost-efficient way through pre-installed components with hoses connected. We further paid attention to easy access to wear parts and consumables.
3Bühler Technologies GmbHBE440022 ◦ 10/2018
TC-Double (+) X2
1.3 Type plate
Example:
TC-Double 230 V, 50/60Hz, 390W 101001081115 44966112121115000000 FM18ATEX0012X II 3 G Ex ec nC IIC T4 Gc IECEx FMG 18.0005X Ex ec nC IIC T4 Gc USA/Canada: CL I DIV 2 GP ABCD Tamb: 0°C to 60°C Year: 2018 Manual doc.no.: BX440022
Manufacturer and address
Type designation and voltage
Serial no., Item no.
Blast protection markings
(Atex, IECex and US/Canada)
Description
Year of manufacture
1.4 Scope of delivery
– Cooler
– Product documentation
– Connection-/mounting accessories (optional)
4 Bühler Technologies GmbH BE440022 ◦ 10/2018
TC-Double (+) X2
1.5 Ordering instructions
1.5.1 TC-Double X2
The item number is a code for the configuration of your unit. Please use the following model key:
4496 6 1 1 X 2 X 1 X X X X X X X 0 0 0 Product Characteristic
Gas cooler type
1 TC-Double 6111 X2: Ambient temperature 40 °C 2 TC-Double 6112 X2: Ambient temperature 60 °C
Certifications
2 for explosive areas
Supply voltage
1 115 V AC, 50/60 Hz 2 230 V AC, 50/60 Hz
Heat exchanger
1 1 0 Stainless steel, PTS, metric 1 1 5 Stainless steel, PTS-I, US fitting 1 2 0 Duran glass, PTG, metric 1 2 5 Duran glass, PTG, US fitting 1 3 0 PVDF, PTV, metric 1 3 5 PVDF, PTV-I, US fitting
Peristaltic Pumps
1)
0 without peristaltic pump 2 CPdouble X2 with hose nipple, angled 4 CPdouble X2 with screw connection
2)
Sample Gas Pumps
1) 3)
0 without sample gas pump
1 P1.3, PVDF
2 P1.3, with bypass valve
Moisture Detector/Filter
1) 2)
0 0 without filter, without moisture detector 0 1 without filter, 1 moisture detector with adapter
1 0 1 filter, without moisture detector 1 1 1 filter with built-in moisture detector
Status outputs
0 0 status output only
1 0 Analogue output option, add-on
1)
If option is selected, the maximum ambient temperature is limited to 50 ° C.
2)
Metric or US fitting, per heat exchanger.
3)
Factory installed tubing for suction operation.
5Bühler Technologies GmbHBE440022 ◦ 10/2018
TC-Double (+) X2
1.5.2 TC-Double+ X2
The item number is a code for the configuration of your unit. Please use the following model key:
4496 6 1 1 X 2 X 1 X X X X X X X 0 0 0 Product Characteristic
Gas cooler type
1 TC-Double+ 6111 X2: Ambient temperature 40 °C 2 TC-Double+ 6112 X2: Ambient temperature 60 °C
Certifications
2 for explosive areas
Supply voltage
1 115 V AC, 50/60 Hz 2 230 V AC, 50/60 Hz
Heat exchanger
1 2 2 Duran glass, PTG-2, metric 1 2 7 Duran glass, PTG-2, US fitting 1 3 2 PVDF, PTV-2, metric 1 3 7 PVDF, PTV-2-I, US fitting
Peristaltic pumps
1)
0 without peristaltic pump 2 CPdouble X2 with hose nipple, angled 4 CPdouble X2 with screw connection
2)
Sample Gas Pumps
1) 3)
0 without sample gas pump
1 P1.3, PVDF
2 P1.3, with bypass valve
Moisture Detector/Filter
1) 2)
0 0 without filter, without moisture detector 0 1 without filter, 1 moisture detector with adapter
1 0 1 filter, without moisture detector 1 1 1 filter with built-in moisture detector
Status outputs
0 0 status output only
1 0 Analogue output option, add-on
1)
If option is selected, the maximum ambient temperature is limited to 50 ° C.
2)
Metric or US fitting, per heat exchanger.
3)
Factory installed tubing for suction operation.
6 Bühler Technologies GmbH BE440022 ◦ 10/2018
TC-Double (+) X2
2 Safety instructions
2.1 Important notices
NOTICE
The device is suitable for hazardous areas.
This unit may only be used if:
– The product is being used under the conditions described in the operating- and installation instructions, used according to
the nameplate and for applications for which it is intended. Any unauthorized modifications to the unit will void the war­ranty provided by Bühler Technologies GmbH,
– The specifications and markings in the type plate must be observed,
– The threshold values in the data sheet and the instructions must be observed,
– Monitoring equipment / protection devices must be connected correctly,
– Service and repair work not described in these instructions are performed by Bühler Technologies GmbH,
– Genuine replacement parts must be used.
Erecting electrical systems in explosive areas requires compliance with the following national regulations:
– IEC/EN60079-14
– National electric code (NEC)
– Canadian electric code (CEC)
Additional national regulations pertaining to initial operation, operation, maintenance, repairs and disposal must be observed.
These operating instructions are a part of the equipment. The manufacturer reserves the right to change performance-, specific­ation- or technical data without prior notice. Please keep these instructions for future reference.
Signal words for warnings
DANGER
Signal word for an imminent danger with high risk, resulting in severe injuries or death if not avoided.
WARNING
Signal word for a hazardous situation with medium risk, possibly resulting in severe injuries or death if not avoided.
CAUTION
Signal word for a hazardous situation with low risk, resulting in damaged to the device or the property or minor or medium injuries if not avoided.
NOTICE
Signal word for important information to the product.
7Bühler Technologies GmbHBE440022 ◦ 10/2018
TC-Double (+) X2
Warning signs
These instructions use the following warning signs:
Warns of a general hazard General notice
Warns of voltage Unplug from mains
Warns not to inhale toxic gasses Wear respiratory equipment
Warns of corrosive liquids Wear a safety mask
Warns of explosive areas Wear gloves
Warns of hot surfaces
2.2 General hazard warnings
The equipment must be installed by a professional familiar with the safety requirements and risks.
Be sure to observe the safety regulations and generally applicable rules of technology relevant for the installation site. Prevent malfunctions and avoid personal injuries and property damage.
The operator of the system must ensure:
– Safety notices and operating instructions are available and observed,
– The respective national accident prevention regulations are observed,
– The permissible data and operational conditions are maintained,
– Safety guards are used and mandatory maintenance is performed,
– Legal regulations are observed during disposal.
Maintenance, Repair
Please note during maintenance and repairs:
– Repairs to the unit must be performed by Bühler authorised personnel.
– Only perform conversion-, maintenance or installation work described in these operating and installation instructions.
– Always use genuine spare parts.
Always observe the applicable safety and operating regulations in the respective country of use when performing any type of maintenance.
DANGER Electrical voltage
Electrocution hazard.
a) Disconnect the device from power supply.
b) Make sure that the equipment cannot be reconnected to mains unintentionally.
c) The device must be opened by trained staff only.
d) Regard correct mains voltage.
8 Bühler Technologies GmbH BE440022 ◦ 10/2018
TC-Double (+) X2
DANGER Toxic, corrosive gas/condensate
Sample gas/condensate may be hazardous to health.
a) If necessary, ensure a safe gas/condensate discharge.
b) Always disconnect the gas supply when performing maintenance or repairs.
c) Protect yourself from toxic/corrosive gasses/condensate when performing mainten-
ance. Wear appropriate protective equipment.
DANGER Explosion hazard
Life and explosion risk may result from gas leakage due to improper use.
a) Use the devices only as described in this manual.
b) Regard the process conditions.
c) Check tubes and hoses for leakage.
WARNING Risk of breakage
a) Protect the equipment against being hit.
b) Protect the device against falling objects.
9Bühler Technologies GmbHBE440022 ◦ 10/2018
TC-Double (+) X2
3 Transport and storage
Only transport the product inside the original packaging or a suitable alternative.
The equipment must be protected from moisture and heat when not in use. It must be stored in a covered, dry and dust-free room at a temperature of -20 °C to 60 °C (-4 °F to 140 °F).
10 Bühler Technologies GmbH BE440022 ◦ 10/2018
TC-Double (+) X2
4 Installation and connection
4.1 Installation site requirements
The unit is only intended for wall-mounted use in enclosed areas. Adequate protection from the weather must be provided when used outdoors.
Install the unit leaving enough room below the cooler to discharge the condensate. Leave room above for the gas supply.
Be sure to maintain the approved ambient temperature. Do not obstruct the convection of the cooler. The vents must have enough room to the next obstacle. The distance must especially be a minimum of 10 cm on the air outlet side.
Ensure adequate ventilation when installing in enclosed housings, e.g. analyser cabinets. If the convection is inadequate, we re­commend aerating the cabinet or installing a fan to lower the inside temperature.
4.2 Installation
Run the gas supply to the cooler with a downward slope. The gas inputs are marked in red and additionally labelled “IN”. The op­timal gas connection is detailed in chapter Connecting the heat exchanger.
If a large amount of condensate accumulates, we recommend using a condensate trap with automatic condensate drain before the cooler.
Glass vessels and automatic condensate drains are available for draining condensate for external mounting below the unit. When using automatic condensate drains, the sample gas pump must be installed ahead of the cooler (pressure operation) to ensure proper function of the condensate drain.
If the sample gas pump is located at the cooler outlet (suction operation), we recommend using glass condensate traps or peri­staltic pumps.
Connecting the condensate drains
Depending on the material, build a connecting line with fittings and tubing or hose between the heat exchanger and condens­ate drain. For stainless steel the condensate drain can be suspended directly to the connecting tube, for hoses the condensate drain must be secured separately using a clamp.
The condensate drain can be mounted directly to the heat exchanger.
Condensate lines must always be installed with a slope and a minimum inside diameter of DN 6/8 (1/4”).
4.2.1 Flow adapter connection (optional)
When ordering the cooler with the
option moisture detector without filter
, it will be factory installed inside a flow adapter.
The connection between the heat exchanger outlet and the flow adapter inlet does not have tubing included. The connection G1/4 or NPT1/4" (flow adapter marked NPT) for the gas inlet/outlet must be carefully and properly connected using a suitable screw connection. Here the direction of flow is not relevant.
4.2.2 Connecting the moisture detector (option)
When ordering the cooler with
moisture detector option
, it will be factory installed inside a flow adapter, or for the
filter option
installed and connected in the filter head.
4.2.3 Connecting the peristaltic pump (option)
Coolers ordered with attached peristaltic pump already have it installed and wired. Heat exchangers ordered at the same time are already installed and connected to the peristaltic pump.
The connection via the condensate outlet screw connection on the pump must be carried out carefully and properly.
NOTICE
Installing peristaltic
pumps
CPsingle / CPdouble limits the maximum permissible
oper-
ating pressure
in the system!
Operating pressure ≤ 1 bar
11Bühler Technologies GmbHBE440022 ◦ 10/2018
TC-Double (+) X2
4.2.4 Sample gas pump connection gas lines (optional)
On coolers ordered with attached sample gas pump these are already installed and wired. Add-on parts ordered at the same time are already installed and connected to the sample gas pump.
Avoid mixed-material installation, i.e. metal piping to plastic bodies. If this cannot be avoided in isolated applications, screw the metal connections into the pump body with care, never use force.
Lay the lines so the line at the inlet and outlet remains flexible for an adequate distance.
The pumps are marked
IN
for inlet and
OUT
for outlet at the mounting ring. Be sure the gas line connections are tight.
4.2.5 Connecting the heat exchanger
To minimise gas wash out in the cooler, the two (identical) heat exchangers must be operated in series. This should be done as follows:
1. Gas inlet line to red gas inlet on heat exchanger no. 2 (pre-cooling).
2. Connection between the heat exchanger no. 2 gas outlet and the red gas inlet on heat exchanger no. 1 (after-cooling).
3. Secure final gas outlet line to the heat exchanger no. 1 gas outlet.
Air IN
Air OUT
Heat exchanger 2
Heat exchanger 1
Condensate outlet
Flow chart:
1 Heat exchanger 2 / pre-cooling 4 Filter (optional) 2 Heat exchanger 1 / after-cooling 5 Sample gas pump (optional) 3 Moisture detector (option) 6 Condensate pump (optional)
12 Bühler Technologies GmbH BE440022 ◦ 10/2018
TC-Double (+) X2
The gas inputs are marked in red.
On glass heat exchangers the correct position of the seal is important when connecting the gas lines (see image). The seal con­sists of a silicone ring with a PTFE sleeve. The PTFE side must face the glass thread.
Glass
PTFE
4.3 Electrical connections
The operator must install an external separator for the device which is clearly assigned to this device.
This separator
– must be located near the device,
– must be easy for the operator to reach,
– must comply with IEC 60947-1 and IEC 60947-3,
– must separate all live conductors and the status output, and
– must not be attached to the power feed.
The mains supply of the device must be fused according to the specifications under technical data.
Potential equalization/static charge
Static charges can result in incendive sparking. Avoid static charges. Any conductive parts of the cooler must be grounded!
The housing has a connection for an earth/equipotential bonding conductor. Ensure the housing is adequately earthed (min­imum conductor cross-section 4 mm2).
Particularly observe the requirements of EN 60079-14!
WARNING Hazardous electrical voltage
The device must be installed by trained staff only.
CAUTION Wrong mains voltage
Wrong mains voltage may damage the device. Regard the correct mains voltage as given on the type plate.
13Bühler Technologies GmbHBE440022 ◦ 10/2018
TC-Double (+) X2
Connection via plug
The device is equipped with connectors according to EN 175301-803 for mains and status output. If the cables are mounted prop­erly, they cannot be interchanged. Please make sure that the connectors are remounted correctly after connecting the cable. The following figures show the pin assignment with respect to the numbers printed on the connector.
The supply line cross-sections must be suitable for the rated current. Use a maximum line cross-section of 1.5mm² (AWG16) and a cable diameter of 8-10mm (0.31-0.39 inch).
Plug numbering
Electric supply S1
Alarm contact S2
Analogue output S3
1 - N/C 2 - N/C 3 - GND 4 - 4-20 mA out 5 - shield
*
*
* When using an alarm contact connect to ≥33VAC or ≥70VDC PE voltage.
The clamping areas for S1 and S2 are 8-10mm (0.31–0.39 inch) in diameter.
4.4 Signal outputs
The device has various status signals. The maximum switching power of the alarm outputs is 250VAC/150VDC, 2A, 50 VA, each.
An alarm is triggered by the alarm contact / status output (S2) if the temperature in block 1 is outside the specified limits. It does not indicate if the alarm was triggered due to excess temperature or insufficient temperature.
Description of signal outputs
Function / contact type Description
Regard­ing S2)
internal changeover con­tact: max. 250VAC / 150VDC, 2A, 50VA
the following device statuses can be indicated via two switching outputs:
Contact between 3 and 2 closed (alarm)
– No mains voltage and/or actual temperature outside
the alarm thresholds
Contact between 3 and 1 closed (ok)
– Mains voltage attached + actual temperature within the
alarm thresholds
With temperature signal option
Regard­ing S3)
4-20mA analogue output (R
Load
<500Ω)
Signalling of actual temper­ature in block 1 (please use shielded cables)
T
Cooler
= -20°C (-4°F) -> 4mA/ 2V
T
Cooler
= 5°C (41°F) -> 9mA/ 4,5 V
T
Cooler
= 60°C (140°F) -> 20mA/ 10 V
The front film on the display module has three LEDs:
Colour Marking Function
Red S2 High/low temperature
Device error Yellow S1 Reserved Green OP Normal operation
The LEDs OP and S2 indicate the device status similar to S2.
14 Bühler Technologies GmbH BE440022 ◦ 10/2018
TC-Double (+) X2
If the option “temperature signal” is built in, the unit has a signal output via the analogue output to indicate the actual cooler temperature.
The temperature signal can be read via the panel plug (S3) using the M12x1 connector. This plug is located next to the moisture detector connectors at the top of the cooler.
The temperature (CH1) which continuously appears in the display corresponds to the temperature in block 1 (after-cooling). The analogue temperature signal also corresponds to the temperature in block 1.
15Bühler Technologies GmbHBE440022 ◦ 10/2018