BTX Gemini X2, Gemini SC User Manual

Page 1
Gemini Series
Twin Waveform
Electroporation Systems
User’s Manual
www.btxonline.com
5507-002 REV 1.0
Page 2
Safety Information
Please read the following safety precautions to ensure proper use of your generator. If the equipment is used in a manner not specified, the protection provided by the equipment may be impaired.
To Prevent Hazard or Injury
Caution
Protective Ground
Terminal
Caution Risk of
Electric Shock
Use Proper Line Cord
Use only the specified line cord for this product and make sure line cord is certified for country of use.The operating voltage range for the BTX Gemini Twin Wave Series is 100-240 vac, 50-60 Hz.
Ground the Product
This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making any connections to the input or output terminals of the product, ensure that the product is properly grounded.
Make Proper Connections
Make sure all connections are made properly and securely. Any signal wire connections to the unit must be no longer than 3 meters.
Observe All Terminal Ratings
Review the operating manual to learn the ratings on all connections.
Use Proper Fuse
Use only specified fuses with product.
Avoid Exposed Circuitry
Do not touch any electronic circuitry inside of the product.
High Voltage Risk
󰀜󰀙󰀙󰀙   
   
 
   
   
   
 󰀚󰀡󰀙󰀙󰀛󰀠󰀛󰀛󰀠󰀠󰀞󰀚󰀞󰀙󰀡󰀡󰀢󰀜󰀡󰀢󰀢󰀢
Do Not Operate with Suspected Failures
If damage is suspected on or to the product do not operate the product. Contact qualified service personnel to perform inspection.
Orient the Equipment Properly
Do not orient the equipment so that it is difficult to operate the disconnection device.
Place Product in Proper Environment
Review the operating manual for guidelines for proper operating environments.
Observe All Warning Labels on Product
Read all labels on product to ensure proper usage.
Gemini Series Electroporator User’s Manual
2
Page 3
A Message from BTX
Thank you for investing in a Gemini System.
Since its founding in 1983, the main focus of BTX has been in the area of applying controlled electric fields for genetic engineering applications. Because of this, we quickly established a reputation as the technological leader in the fields of electroporation and electrofusion. Our systems have been installed in many prestigious institutes around the globe where they are used successfully for high efficiency transfection, transformation and cell fusion applications. We offer a variety of waveforms, electrodes and chamber options to provide you with the tools to achieve your goals.
We are vested in your success. To that end, the BTX technical support team constantly tracks published literature for any reference to electroporation and electrofusion. We extract the pertinent experimental conditions and yields from these papers to help us in our efforts to help you. In addition to tracking publications, we are available to you for support at any time for advice in experimental design, product recommendations, troubleshooting, and any other relevant technical advice.
We thank you again for your investment and we look forward to assisting you in any way we can.
Finally, please read this manual carefully before attempting to operate the electroporation system. If you have any questions about the unit or about particular applications, please contact us:
BTX
84 October Hill Rd Holliston, MA 01746 USA
Toll Free: 1-800-272-2775 International Callers: 508-893-8999 Fax: 508-429-5732 Web: www.btxonline.com Email: [email protected]
For any customers outside the US or Canada, please call your local BTX dealer or call us directly.
Gemini Series Electroporator User’s Manual
3
Page 4
Table of Contents
Safety Information .......................................................................................................................2
Introduction .................................................................................................................................3
Table of Contents ........................................................................................................................4
Product Overview ........................................................................................................................5
Electroporation Overview ............................................................................................................6
General Optimization Guide for Electroporation .................................................................... 7-8
Wave Forms ..................................................................................................................................... 7
Field Strength ................................................................................................................................... 7
Pulse Length ..................................................................................................................................... 8
Number of Pulses ............................................................................................................................. 8
Electroporation Buffer ...................................................................................................................... 8
DNA/RNA Concentrations ................................................................................................................. 8
Applications ........................................................................................................................... 9-11
Mammalian Cell Transfection ............................................................................................................ 9
In Vivo, In Utero, In Ovo ................................................................................................................... 9
Bacteria and Yeast Transformation .................................................................................................... 9
Plant and Insect Transfection ............................................................................................................ 9
References ........................................................................................................................................ 9
General References ...................................................................................................................... 9-11
Glossary of Terms ......................................................................................................................12
Electroporation Buffers .............................................................................................................13
Unpacking the System ...............................................................................................................14
Touchscreen Button Reference ..................................................................................................15
Software Setup .................................................................................................................... 16-17
Setting Time/Date ........................................................................................................................... 16
Setting Audible Alarm Preferences .................................................................................................. 16
Setting Backlight Preferences .......................................................................................................... 17
Displaying Device Information ........................................................................................................ 17
Preset Protocols ................................................................................................................... 18-20
Using Preset Protocols .................................................................................................................... 18
Customizing a Preset Protocol ................................................................................................... 19-20
Performing Experiments ...................................................................................................... 20-26
Creating New Protocols .................................................................................................................. 21
Creating an Exponential Decay Wave Protocol ..................................................................... 21-23
Creating a Square Wave Protocol ........................................................................................ 24-26
Using Specialty Protocols .................................................................................................... 27-29
Using an HT Plate Handler ......................................................................................................... 27-28
Using Specialty Electrodes ......................................................................................................... 28-29
Managing Protocols ............................................................................................................. 30-32
Saving a Copy of a Protocol ............................................................................................................ 30
Renaming a Protocol ...................................................................................................................... 30
Deleting a Protocol ......................................................................................................................... 31
Password Protecting a Protocol ....................................................................................................... 32
Protocol Manager Software ................................................................................................ 33-35
Installation ..................................................................................................................................... 33
Overview ........................................................................................................................................ 33
Upload– Generator to PC ................................................................................................................ 33
Download– PC to Generator ........................................................................................................... 34
Upload Log Files ............................................................................................................................. 35
Remote Control Software ..........................................................................................................36
Generator Specifications ..................................................................................................... 37-38
Maintenance ..............................................................................................................................39
Upgrading Gemini Series Software ..................................................................................... 40-42
SB Virtual Commport Driver Installation ..................................................................................... 40-41
Device Updater Program ................................................................................................................. 42
Error Messages & Troubleshooting ..................................................................................... 43-45
Ordering Information .......................................................................................................... 46-47
Warranty Information ................................................................................................................48
Declaration of Conformity .........................................................................................................49
Gemini Series Electroporator User’s Manual
4
Page 5
Product Overview
The BTX Gemini X2 is a highly advanced twin waveform generator incorporating both square and exponential decay waves in a single unit. The BTX Gemini X2 has been designed with these waveform combinations to enable researchers to easily and efficiently electroporate eukaryotic cells and prokaryotic cells in all forms with one easy to use setup. The BTX Gemini X2, which can be operated via PC or remote control, boasts a wide range of voltage (5-3000 v, 1 v or 5 v increments), pulse length (10 μS – 1S, 1 μS resolution), time constant options (which include capacitance choices from 25 to 3275 μF in LV and 10, 25, 35, 50, 60, 75, 85 μF in HV), along with multiple pulsing options with both the square waveform and the exponential decay waveform, and unparalleled pulse delivery accuracy. The BTX Gemini X2, with over 1,000 custom protocol storage, monitors and displays pre-pulse sample resistance as well as delivered voltage values and records logs of all experiment parameters internally, which can be downloaded to a computer for analysis and QC. The generator is controlled through a color LCD touch screen interface and incorporates USB communications. The pulse can also be activated by a foot switch. The enhanced safety features of the BTX Gemini X2 protect users as well as precious samples. The BTX Gemini X2 is designed to give researchers the ultimate flexibility in their experiments, making it possible to perform electroporation on tissues and organs in vivo (as well as in utero, in ovo, ex plant) on adherent cells, and cells in suspension in either single cuvettes or 96 well plates. The Safety Dome allows researchers to safely work with electroporation cuvettes, up to two at a time, giving researchers the ability to experiment on sample volumes from 20 μl up to 800 μl. This state of the art system comes complete with the twin waveform generator for cell electroporation in all forms, dozens of preprogrammed protocols for commonly electroporated cells, Safety Dome, 30 cuvettes, cuvette stand, user manual, two year warranty, unlimited application support and the same high quality researchers have come to expect from BTX.
The BTX Gemini SC is a twin waveform generator incorporating both square and exponential decay waves in a single unit. These waveform combinations enable researchers to easily and efficiently electroporate eukaryotic and prokaryotic cells in suspension with one easy to use setup. The BTX Gemini SC boasts a wide range of voltage (10-3000 v, 5 v or 10 v increments), pulse length (50 μS – 5 mS, 50 μS resolution), time constant options (which include capacitance choices from 25 to 3275 μF in LV and 10, 25 or 50 μF in HV), multiple pulsing option with the square waveform and unparalleled pulse delivery accuracy. The BTX Gemini SC, with unlimited custom protocol storage, monitors and displays pre-pulse sample resistance as well as delivered voltage values. The generator is controlled through a color LCD touch screen interface. The enhanced safety features of the BTX Gemini SC protect users as well as precious samples. The Safety Dome allows researchers to safely work with electroporation cuvettes, up to two at a time, giving researchers the ability to experiment on sample volumes from 20 μl up to 800 μl. This affordable system comes complete with the dual waveform generator for suspension cell electroporation, dozens of preprogrammed protocols for commonly electroporated cells, Safety Dome, 30 cuvettes, cuvette stand, user manual, two year warranty, unlimited application support and the same high quality researchers have come to expect from BTX.
What is the difference between the Gemini X2 and the Gemini SC?
The Gemini X2 is designed to give researchers the ultimate control and flexibility in their experiments by making it possible to perform electroporation on tissues and organs in vivo (as well as in utero, in ovo, ex plant) on adherent cells, as well as cells in suspension in either single cuvettes or 96 well plates, can be controlled remotely via footswitch or PC, and it offers storage of pulse data. Because of the Gemini X2’s broad range of use, the specifications are wide­ranging, making the Gemini X2 the most versatile electroporation system available today.
The Gemini SC is designed for researchers working to electroporate cells in suspension in cuvettes. This system cannot accommodate multiple pulsing with the exponential decay waveform, remote operation, specialty electrodes or 96 well options. For this reason, some of the specifications of the Gemini SC are not as extensive as what is available in the Gemini X2.
Gemini Series Electroporator User’s Manual
5
Page 6
Electroporation Overview
DNA Delivery Into Cells Using Electroporation
General Electroporation Discussion
Electroporation is the use of a transmembrane electric field pulse to induce microscopic pathways (pores) in a bio-membrane. Their presence allows molecules, ions, and water to pass from one side of the membrane to the other. When the electric field is applied the ions inside and outside the cell membrane migrate. As the charge builds up on either side of the membrane the membrane weakens and the pathways form permitting material outside of the cell to enter. If the electric field is promptly removed the pathways close and the membrane reseals. If the electric field duration is too long the pathways increase and the cell is killed. Efficient electroporation depends on proper selection of electric field waveforms. The electropores are located primarily on the membrane areas which are closest to the electrodes. The pathways form in about a microsecond and seal in seconds to minutes. The duration of the electric field is tens of microseconds to tens of milliseconds.
The use of electroporation was described by Neumann in the early 1980’s. The routine use of electroporation became very popular with researchers through the 1980’s because it was found to be a practical way to place drugs, or other molecules into cells. In the late 1980’s, scientists began to use electroporation for applications in multi-cellular tissue.
In the early 1990’s Lluis Mir of the Institute Gustave-Roussy was the first to use electroporation in a human trial to treat external tumors.
Research has shown that the induction of pathways is affected by three major factors. First, cell-to-cell biological variability causes some cells to be more sensitive to electroporation than other cells. Second, for pathways to be induced, the product of the pulse amplitude and the pulse duration has to be above a lower limit threshold. Third, the number of pathways and effective pathway diameter increases with the product of “amplitude” and “duration.” Although other factors are involved, this threshold is now understood to be largely dependent on a fourth factor, the reciprocal of cell size. If the upper limit threshold is reached pore diameter and total pore area are too large for the cell to repair by any spontaneous or biological process, the result is irreversible damage to the cell or cell lysis. Because the mechanism of electroporation is not well understood, the development of protocols for a particular application has usually been achieved empirically, by adjusting pulse parameters (amplitude, duration, number, and inter-pulse interval).
Research shows that certain experimental conditions and parameters of electrical pulses may be capable of causing many more molecules to move per unit time than simple diffusion. There is also good evidence (Sukharev et al., 1992) that DNA movement is in the opposite direction.
An additional important consideration is when the voltage pulse is applied to the cells and medium that the amount of current that flows is dependent on the conductivity of the material in which the cells are located. Some material is quite conductive and severe heating will occur if the pulse duration is too long. Therefore long duration fields will kill cells by destroying the membrane and heating.
The electric field in which the cells are located is produced by two system components. The first is the voltage waveform generator and the second is the electrode which converts the voltage into the electric field.
As the charge accumulates at the membrane, which is a capacitance, the voltage across the membrane increases.
voltage = capacitance charge
As charge accumulates at the membrane, the voltage across the membrane increases. Neumann et al. (1989) described the equation that relates the transmembrane voltage (TMV) to electric field intensity:
where:
Pores in the membrane will begin to form as the voltage increases from its quiescent value of a few tenths of a volt to more than 0.5 volts. To produce a TMV of 1 volt across the membrane of a cell with 7μm radius, the required electric field intensity is:
2
E= =
3
The number of pores and effective pore diameter increase as the product of pulse amplitude and duration increase. At the upper limit threshold, pore diameter and total pore area become too large for the cell to repair by any spontaneus or biological process. The result is irreversible damage to the cell or cell lysis.
Another important point to consider is the generation of heat during electroporation. Heat production is directly related to current intensity which is, in turn, dependent on the conductivity of the material through which the electric field is applied. Standard saline solutions such as PBS and many tissue culture media are highly conductive and, thus will generate considerable amounts of heat when used in cell electroporation. Excessive heating can be detrimental to cell viability. The effects of heating can be reduced by using a low conductivity medium such as BTX’s Cytoporation medium to resuspend cells prior to electroporation.
Although electroporation is an effective method for introducing macromolecules onto cells, the biological mechanisms by which cells become electroporated are not completely understood. Therefore, the development of specific protocols for particular applicatons is usually achieved by empirical adjustment of pulse parameters (i.e. amplitude, duration, pulse number, and interpulse interval).
Gemini Series Electroporator User’s Manual
6
*
1
7 x 10
950 volts/cm
-4
Page 7
General Optimization Guide for Electroporation
As described, electroporation is the application of controlled direct current (DC) electrical pulses which are applied to living cells and tissues for a short duration of time. The pulse induces a transmembrane potential which causes the reversible breakdown of the cellular membrane. This action results in the permeation or “pore formation” of the cell membrane which allows small molecules (such as dye, oligonucleotides or peptides) and large molecules (such as proteins, DNA and RNA) to be introduced into the cell. During this process the cellular uptake of the molecules continues until the pores close, which can take milliseconds to minutes.
Optimization of the electroporation process involves several factors. Choosing the wave form, determining field strength and adjusting pulse length are just a few critical variables. Other parameters which play a crucial role in optimization include cell diameter, plasmid concentrations, temperature and electroporation buffer.
Wave Forms
Pulse shape generally falls into two categories, square wave or exponential decay wave:
Square Wave Pulse
Square wave pulses rise quickly to a set voltage level, maintain this level during the duration of the set pulse length and quickly turn off. Square waves yields higher efficiencies and viabilities in mammalian cells. Square wave EP in in vivo and ex vivo tissues, embryos, and plant protoplast applications yield better results in comparison to an exponential decay wave.
Exponential Decay Wave Pulse
Exponential decay waves generate an electrical pulse by allowing a capacitor to completely discharge. As a pulse is discharged into a sample, the voltage rises rapidly to the peak voltage set then declines over time. The powerful exponential decay wave pulse is routinely used for transformation of gram-negative and gram-positive, bacterial, yeast, plant tissues, insect cells and some mammalian cells.
Field Strength
The field strength is measured as the voltage delivered across an electrode gap and is expressed as kV/cm. Field strength is critical to surpassing the electrical potential of the cell membrane to allow the temporary reversible permeation or “pore formation” to occur in the cell membrane. Three factors should be considered for optimizing field strength:
1. Electrode Gap Size
2. Cell Diameter
3. Temperature
Cell Type Field Strength Ranges
Bacteria/Yeast: 3-24 kV/cm
Mammalian: 0.25-3 kV/cm
Plant: 3-12 kV/cm
Electrode Gap Size
The distance between electrodes, or “gap size” is important when optimizing your electroporation experiment. Field strength is calculated using voltage divided by gap size. For example, using a 4mm gap cuvette with 500V would provide a field strength of
1.25kV/cm. If instead of a 4mm gap cuvette, a 2mm gap cuvette was used, the voltage would have to be reduced by half or to 250V in order to maintain the same field strength of 1.25kV/ cm. It is possible to derive the voltage needed to accomplish electroporation if the desired field strength and gap size are known. The calculation for this is Field strength (kV) multiplied by gap size (cm) equals voltage. For example, if a user was certain that a 1.25 kV/cm field strength was required in a 1mm gap cuvette the calculation would be: 1.25kV x 0.1cm= 0.125kV or 125V.
Example: A field strength of 1.25 kV/cm
4mm gap cuvette = 500V
2mm gap cuvette = 250V
1mm gap cuvette = 125V
Cell Diameter
Generally, smaller cell sizes require higher voltages while larger cell diameters require lower voltages for successful cell membrane permeation.
Temperature
The temperature at which cells are maintained during electroporation effects the efficiency of the electroporation for several reasons. The majority of mammalian cell lines are effectively electroporated at room temperature. Samples which are pulsed at high voltage or exposed to multiple pulses and long pulse durations can cause the sample to heat up. These conditions cause increased cell death and lower the transfection efficiency. Maintaining the sample at lower temperatures can diminish the heating effects on cell viability and efficiency. Since electroporation causes the transient formation of pores, keeping the cells at a lower temperature following the pulse may allow the pores to remain open longer to allow more uptake of the exogenous molecules. Yet lower temperatures on other cell lines can be damaging and cause high cell mortality. This effect is specific to each cell line and should be considered during optimization studies. The standard pulse voltage used for cells at room temperature will need to be approximately doubled for electroporation at 4°C in order to effectively permeate the cell membrane.
Gemini Series Electroporator User’s Manual
7
Page 8
General Optimization Guide for Electroporation
Pulse Length
The pulse length is the duration of time the sample is exposed to the pulse. This is measured as time in micro to milliseconds ranges. Adjusting this parameter is dependent on the pulse wave form. The pulse length in a square wave system can be inputted directly. The pulse length in an exponential decay wave system is called the “time constant” which is characterized by the rate at which the pulsed energy (e) or voltage is decayed to 1/3 the original set voltage. This time constant is modified by adjusting the resistance and capacitance (RC) values in an exponential decay wave form. Time constant calculation T=RC, where T is time and R is resistance and C is capacitance.
The pulse length works indirectly with the field strength to increase pore formation and therefore the uptake of target molecules. Generally, during optimization of parameters an increase in voltage should be followed by an incremental decrease in pulse length. When decreasing the voltage, the reverse is true. Pulse length is a key variable that works hand in hand with voltage and needs to be considered when optimizing electrical parameters to maximize the results for a given cell type.
Number of Pulses
Electroporation is typically carried out as a single pulse for most cell types. However, other cell lines may require multiple pulses to achieve maximum transfection efficiencies. Usually lower voltages are used when applying multiple pulses in order to gradually permeate the cell membranes. This allows the transfer of molecules while avoiding damage to delicate or whole tissue samples. This method of multiple pulsing is critical for maximum gene delivery without causing tissue damage to in vivo, in utero and explant tissue environments. The use of multiple pulse will require the optimization of key electrical parameters including voltage and pulse length. Typically, for in vivo applications the use of lower voltages between 10-100 volts with pulse lengths ranging 30-50msec provides efficient transfection. The optimal voltage, pulse length and number of pulses will vary depending on the cell type and molecule (DNA or RNA) transfected.
resistance of the sample which in turn will affect the pulse length or time constant of the pulse. The volume of liquid in a cuvette has significant effect on sample resistance for ionic solutions, the resistance of the sample is inversely proportional to the volume of solution and pH. As the volumes are increased resistance decreases which increases the chance of arcing, while lowering the volume will increase the resistance and decrease the arc potential.
BTX now offers BTXpress High Performance Electroporation Solution, a low conductance buffer that achieves higher transfection efficiencies with minimal cell toxicity. The BTXpress buffer is a single buffer developed to facilitate high efficiency gene delivery into mammalian cells.
DNA/RNA Concentrations
Electroporation is typically thought of as a nucleic acid (DNA, mRNA, siRNA and miRNA) transfer method into prokaryotic and eukaryotic cells. Electroporation is not limited to just nucleic acid delivery, it can introduce proteins, antibodies, small molecules and fluorescent dyes. The standard range of DNA used for transfections is 5-20g/ml for most cell types; however in some instances increasing the DNA concentration as high as 50g/ ml improves transfection efficiency without changing other parameters. Determining the optimal DNA concentration through a DNA titration can be beneficial. The size of a molecule will have an effect on the electrical parameters used to transfect the cell. Smaller molecules (siRNA or miRNA) may need higher voltage with microsecond pulse lengths and larger molecules (DNA) may need lower voltages with longer pulse lengths. Buffers such as EDTA or Tris can drastically reduce the transfection efficiency. Therefore, we recommend resuspending DNA in distilled water. Finally, electroporating ligation mixtures into E.coli can cause arcing and reduced transformations. Diluting the ligation mixture a minimum of 1:5 with diH2O, dialysis, or ethanol precipitation can significantly improve transformation efficiencies and reduce the potential for arcing.
Electroporation Buffer
The buffers used for electroporation can vary depending on the cell type. Many applications use highly conductive buffers such as PBS (Phosphate Buffered Saline <30 ohms) and HBSS (Hepes Buffer <30 ohms) or standard culture media which may contain serum. Other recommended buffers are hypoosmolar buffers in which cells absorbs water shortly before pulse. This swelling of the cells results in lowering the optimal permeation voltage while ensuring the membrane is more easily permeable for many cells but can be damaging to others. Prokaryotic cells such as bacteria require the use of high resistance buffers (>3000 ohms) for this reason proper preparation and washing of the cells is essential to remove excess salt ions to reduce the chance of arcing. Ionic strength of an electroporation buffer has a direct affect on the
Gemini Series Electroporator User’s Manual
8
Page 9
Applications
Mammalian Cell Transfection
Electroporation is a highly flexible technique used to genetically modify mammalian cells. Whether you are studying up or down regulation of genes, specific protein expression. This method is non-toxic and requires no expensive reagents to successfully transfect your cells. Primary cells, stem cells or established cell lines can be electroporated with yield high transfection efficiencies and great cell survival rates.
In Vivo, In Utero, In Ovo
Square wave systems allow researchers to set the pulse lengths and number of pulses, which is critical to ensure viable cells and tissues while still maintaining efficient transfection both in vivo and ex vivo. Electroporation mediated gene and drug delivery has been shown to substantially increase intracellular uptake and expression of DNA, siRNA and miRNA in muscle, skin, liver, kidney, testis, retina, tumors, etc. In vivo electroporation has successfully been used in embryo applications, in utero and in ovo applications in addition to transfection of Zebra fish.
Bacteria and Yeast Transformation
Electroporation has long been recognized as the most efficient means of transforming both gram negative and gram positive bacteria and yeast. Gramnegative bacteria such as coli or Helicobacter pylori are generally easier to transform than grampositive bacteria (e.g. Streptococcus pneumoniae) due to their cell wall composition. Transformation efficiencies of 1x10 transformants/μg DNA are commonly seen for gram-negative bacteria, while for gram-positive bacteria, generally 1x10
10
6
transformants/μg DNA are achievable.
Plant and Insect Transfection
Electroporation of plant tissue can be used to generate transgenic crops that are useful in agricultural/horticultural applications.
Insect models are also widely used throughout the scientific community to study development and gene regulation and function. The ability to introduce genes or molecules is essential to researchers working with either of these two species. This is why researchers consistently turn to BTX for all of their electroporation needs.
References
Jonathan M. Dermott, J. M. Gooya, B. Asefa, S. R. Weiler, M. Smith, J. R. Keller. Inhibition of Growth by p205: A nuclear protein and putative tumor suppressor expressed during Myeloid Cell differentiation. Stem Cells 22:832-848. 2004
JonathanM. Quinlan, Wei-Yuan Yu, MarkA. Hornsey, David Tosh andJonathan MWSlack. In vitro culture of embryonic mouse intestinal epithelium: cell differen-tiation and introduction or reporter genes. BMC Developmental Biology 6:24. 2006
YangbingZhao, ZhiliZheng, Cyrille J. Cohen, Luca Gattinoni, Douglas C. Palmer, Nicholas P. Restifo, Steven A. Rosenberg, and RichardA. Morgan. High-efficiency transfection ofprimary human and mouse Tlymphocytes using RNA electropora-tion. Molecular Therapy (2006) 13, 151–159
William J. Buchser, Jose R. Pardinas, Yan Shi, John L. Bixby, and VanceP. Lemmon. 96-Well electroporation method for transfection of mammalian central neurons. BioTechniques Vol. 41, no. 5.2006
K. Regha, AjitK. Satapathy and Malay K. Ray. RecD plays an essential function during growth at low temperature in the Antarctic Bacterium Pseudomonas syringae Lz4W. Genetics 170: 1473-1484. August2005
Victor B. Busov, R. Meilan, D.W. Pearce, C. Ma, S. B. Rood, andS. H. Strauss. Activation Tagging of a dominant Gibberellin catabolism gene (GA 2-oxidase) from poplar that regulates tree stature. Plant Physiology, Vol. 132, pp. 1283-1291. July 2003
Jun Ishikawa, Kazuhiro Chiba, Haruyo Kurita, and Hiroyuki Satoh. Contribution of rpoB2 RNA Polymerase‚ Subunit Gene to Rifampin Resistance in Nocardia Species. Antimicrobial Agents Chemotherapy, 50(4):1342-1346. April 2006
Bindu Garg, Romesh C. Dogra, and Parveen K. Sharma. High-Efficiency Transformation of Rhizobium leguminosarum by Electroporation. Applied Environmental Microbiology. 65(6):2802-2804. June 1999
General References
In Vitro Electroporation
Kim, T. et. al., Mesoporous Silica-Coated Hollow Manganese Oxide Nanoparticles as Positive T1 Contrast Agents for Labeling and MRI Tracking of Adipose- Derived Mesenchymal Stem Cells. J. Am. Chem. Soc., 133, 2955–2961, 2011
Kataoka, N. et. al., Development of butanol-tolerant Bacillus subtilis strain GRSW2-B1 as a potential bioproduction, AMB Express, 1:10, 2011
Hutson, T.H. et. al., Optimization of a 96 well electroporation assay for post natal rat CNS neurons suitable for cost–effective medium­throughput screening of genes that promote neurite outgrowth. Frontiers in Molecular Neuroscience; 4(55): December 2011
Djouad, F. et. al., Activin A expression regulates multipotency of mesenchymal progenitor cells. Stem Cell Res & Therapy, 1(11), 2010
Sankaranarayanan , K. et. al., Electro-Molecular Therapy using Adult Mesenchymal Stem Cells. Proc. ESA Annual Meeting on Electrostatics, 13, 2010
Blackmore, M. et. al., High content screening of cortical neurons identifies novel regulators of axon growth. Molecular and Cellular Neuroscience, 44, 43 -54, 2010
Yao, S. et. al., Improvement of electroporation to deliver plasmid DNA into dental follicle cells. Biotechnol J. October ; 4(10): 1488–1496. 2009
Rinaldi, G., Development of Functional Genomic Tools in Trematodes: RNA Interference and Luciferase Reporter Gene Activity in Fasciola hepatica. PLoS One, 2(7); e260, July 2008
Nguyen, K. et al., T Cell Costimulation via the Integrin VLA-4 Inhibits the Actin-Dependent Centralization of Signaling Microclusters Containing the Adaptor SLP-76. Immunity, 28, 810–821, June 2008
Yang, C. et al., Dimeric heat shock protein 40 binds radial spokes for generating coupled power strokes and recovery strokes of 9 + 2 flagella. The Journal of Cell Biology, 180(2), pp. 403-415, January 28, 2008
Gemini Series Electroporator User’s Manual
9
Page 10
Applications
General References (continued)
In Vitro Electroporation (continued)
Zhao, Y. et al., High-Efficiency Transfection of Primary Human and Mouse T Lymphocytes Using RNA Electroporation. Molecular Therapy, Vo. 13, No. 1, January 2006
Heaney, J.D. et. al. Tissue-specific expression of a BAC transgene targeted to the Hprt locus in mouse embryonic stem cells. Genomics, 2004
Raoul, C. et al., Motoneuron Death Triggered by a Specific Pathway Downstream of Fas: Potentiation by ALS-Linked SOD1 Mutations. Neuron, Vol. 35, 1067-1083, September 12, 2002
Dawson, K. et al., Insulin-Regulated Trafficking of Dual-Labeled Glucose Transporter 4 in Primary Rat Adipose Cells, Biochemical and Biophysical Research Communications. 287, pp. 445–454, 2001
In Vitro Electroporation
Johannson, D. et. al., Intradermal Electroporation of Naked Replicon RNA Elicits Strong Immune Responses. PLoS ONE, 7(1): e29732, 2012
Daftarian, P. et. al., In vivo Electroporation and Non-protein Based Screening Assays to Identify Antibodies Against Native Protein Conformations. Hybridoma, 30(5); 2011
Hallengard, D. et. al., Comparison of plasmid vaccine immunization schedules using intradermal in vivo electroporation. Clinical Vaccine Immunology, 2011
Bolhassani, A. et. al., Improvement of different vaccine delivery systems for cancer therapy. Molecular Cancer, 10(3), 2011
Li, W. et. al., The Effects of Irreversible Electroporation (IRE) on Nerves. PLoS ONE, 6(4), 2011
Lladser, A. et. al., Intradermal DNA electroporation induces survivin­specific CTLs, suppresses angiogenesis and confers protection against mouse melanoma. Cancer Immunol Immunother, 59; 81-92, 2010
Shi, W. et al., Generation of sp3111 transgenic RNAi mice via permanent integration of small hairpin RNAs in repopulating spermatogonial cells in vivo. Acta Biochim Biophy Sci, 42(2): p 116, 2010
Haller, BK. et. al., Therapeutic efficacy of a DNA vaccine targeting the endothelial tip cell antigen delta-like ligand 4 in mammary carcinoma. Oncogene, 29, 4276-4286, 2010
Guo, Y. et. al., Irreversible Electroporation Therapy in the Liver: Longitudinal Efficacy Studies in a Rat Model of Hepatocellular Carcinoma. Cancer Research; 70(4) February 15, 2010
Roos, A. K. ,Skin Electroporation: Effects on Transgene Expression, DNA Persistence and Local Tissue Environment. PLoS ONE, 4(9) e7226, 2009
Brave, A. et. al., Late administration of plasmid DNA by intradermal electroporation efficiently boosts DNA-primed T and B cell responses to carcinoembryonic antigen. Vaccine, 27, 3692-3696, 2009
Roos, A. K. et. al., Optimization of Skin Electroporation in Mice to Increase Tolerability of DNA Vaccine Delivery to Patients. Molecular Therapy, 17(9), 1637-1642, Sep 2009
Danner, S. et. al., Seminiferous tubule transfection in vitro to define post-meiotic gene regulation. Reproductive Biology and Endocrinology, 7(67), 2009
Benton, C. et al., Modest PGC-1_ Overexpression in Muscle in Vivo Is Sufficient to Increase Insulin Sensitivity and Palmitate Oxidation in Subsarcolemmal, Not Intermyofibrillar, Mitochondria*. The Journal of Biological Chemistry, 283(7); pp. 4228–4240, February 15, 2008
Chesler, A. T., Selective Gene Expression by Postnatal Electroporation during Olfactory Interneuron Neurogenesis. PLoS ONE, 3(1): e1517, 2008
Rao, N. M. et al., Electroporation of Adult Zebrafish. S. Li (ed.), Electroporation Protocols: Preclinical and Clinical Gene Medicine. Methods in Molecular Biology, Vol. 423. p 289, 2008
Johnson, C. J. et al., Technical Brief: Subretinal injection and electroporation into adult mouse eyes. Molecular Vision, 14:2211­2226, 2008
Heller, L. et. al., Comparison of electrically mediated and liposome­complexed plasmid DNA delivery to the skin. Genetic Vaccines and Therapy, 6(16), 2008
Roos, A.K., et. al., Enhancement of Cellular Immune Response to a Prostate Cancer DNA Vaccine by Intradermal Electroporation. Molecular Therapy, 13(2), February 2006
Kong, X. C. et al., Inhibition of synapse assembly in mammalian muscle in vivo by RNA interference. EMBO Rep, 5(2): 183-188, January 2004
Pringle, I. A. et al., Duration of reporter gene expression from naled pDNA in the mouse lung following direct electroporation and development or wire electrodes for sheep lung electroporation studies. Molecular Therapy, 9, S56–S56, 2004
Mikata, K. et al., Inhibition of Growth of Human Prostate Cancer Xenograft by Transfection of p53 Gene: Gene Transfer by Electroporation. Molecular Cancer Therapeutics, Vol. 1, 247–252, February 2002
Pekarik, V. et al., Screening for gene function in chicken embryo using RNAi and electroporation. Nature Biotechnology, 21: 93-96, December 2002
Dujardin, N. et. al., In vivo assessment of skin electroporation using square wave pulses. J Controlled Release, 79, 219-227; 2002
Drabick, J.J. et. al., Cutaneous Transfection and Immune Responses to Intradermal Nucleic Acid Vaccination Are Significantly Enhanced by in Vivo Electropermeabilization. Molecular Therapy, 3(2), Feb 2001
In Utero Electroporation
Maiorano, N. A., et al., Promotion of embryonic cortico-cerebral neuronogenesis by miR-124. Neural Development, 4:40, 2009
Ex Vivo Electroporation
Deora, A.A. et. al., Efficient Electroporation of DNA and Protein into Confluent and Differentiated Epithelial Cells in Culture. Traffic, 8: 1304-1312, 2007
Thomas J-L. et. al., Electroporation, an alternative to biolostics for transfection of Bombyx mori embryos and larval tissues. Journal of Insect Science, 3:17, 2003
Dimitrov, D.S., and Sowers, A.E., (1990) Membrane electroporation ­fast molecular exchange by electroosmosis. Biochimica et Biophysica Acta 1022: 381-392.
Gemini Series Electroporator User’s Manual
10
Page 11
Applications
General References (continued)
Ex Vivo Electroporation (continued)
Deora, A.A. et. al., Efficient Electroporation of DNA and Protein into Confluent and Differentiated Epithelial Cells in Culture. Traffic, 8: 1304-1312, 2007
Thomas J-L. et. al., Electroporation, an alternative to biolostics for transfection of Bombyx mori embryos and larval tissues. Journal of Insect Science, 3:17, 2003
Dimitrov, D.S., and Sowers, A.E., (1990) Membrane electroporation ­fast molecular exchange by electroosmosis. Biochimica et Biophysica Acta 1022: 381-392.
Sukharev SI, Klenchin VA, Serov SM, Chernomordik LV and Chizmadzhev YA, (1992) Electroporation, and electrophoretic DNA transfer into cells: The effect of DNA interaction with electropores, 1992, Biophys J. 63: 1320-1327.
Nickoloff, Jac A., ed. (1995) Plant Cell Electroporation and Electrofusion Protocols, Methods in Molecular Biology, Volume 55. (Humana Press, Totowa, New Jersey).
E. A. Disalvo and S.A. Simon, eds. (1995) Permeability and Stability of Lipid Bilayers (CRC Press, Boca Raton), p 105-121.
Chang, D.C., Chassy, B.M., Saunders,J.A. and Sowers, A.E., eds. (1992) Guide to Electroporation and Electrofusion, (Academic press, San Diego), 581 pp.
Neuman, E., Sowers, A.E., and Jordan, C.A.., eds. (1989) Electroporation and Electrofusion in Cell Biology, (Plenum Press, New York) 581 pp.
Bartoletti, D. C., Harrison, G. I., & Weaver, J. C. (1989). The number of molecules taken up by electroporated cells: quantitative determination. FEBS Lett., 256, 4-10.
Djuzenova, C. S., Zimmermann, U., Frank, H., Sukhorukov, V. L., Richter, E., & Fuhr, G. (1996). Effect of medium conductivity and composition on the uptake of propidium iodide into electropermeabilized myeloma cells. Biochim.Biophys.Acta, 1284, 143-
152.
Klenchin VA, Sukharev SM, Chernomordik LV, Chizmadzhev YA, Electrically induced DNA uptake by cells is a fast process involving DNA electrophoresis, 1991, Biophys J. 60:804-811 Neumann, E., Kakorin, S., & Toensing, K. (1999). Fundamentals of electroporative delivery of drugs and genes. Bioelectrochem.Bioenerg., 48, 3-16.
Neuman, E., Toensing, K., Kakorin, S., Budde, P., & Frey, J. (1998). Mechanism of electroporative dye uptake by mouse B cells. Biophys.J., 74, 98-108. Sukharev, S. I., Klenchin, V. A., Serov, S. M., Chernomordik, L. V., & Chizmadzhev, Y. (1992). Electroporation and electrophoretic DNA transfer into cells. The effect of DNA interaction with electropores. Biophys.J., 63, 1320-1327.
Wolf, H., Rols, M. P., Boldt, E., Neumann, E., & Teissie, J. (1994). Control by pulse parameters of electric field-mediated gene transfer in mammalian cells. Biophys.J., 66, 524-531.
Zerbib, D., Amalric, F., & Teissie, J. (1985). Electric field mediated transformation: isolation and characterization of a TK+ subclone. Biochem.Biophys.Res.Commun., 129, 611-618.
Gemini Series Electroporator User’s Manual
11
Page 12
Glossary of Terms
Capacitance – The ability of a body to store an electrical charge. Any object that can be electrically charged exhibits capacitance.
Capacitor – A device that stores energy in the form of an electric field. A capacitor consists of two metal plates insulated from each other by a dielectric (insulating, usually a plastic material such as Mylar) material. In an ideal capacitor, no conduction current flows between the plates after the capacitor is completely charged. Capacitors can be fixed, variable, or adjustable.
Cell Form – The format in which cells present for experimentation. Cell forms include suspension, in vivo, in ovo, ex plant, adherent, whole organism, etc.
Dielectric Breakdown – The reversible breakdown of bi­lipid layer membranes as a result of the application of a DC electroporation pulse. A sufficiently high field strength may increase the membrane potential past a critical point leading to the breakdown of the membrane.
Dielectric Constant – For a given dielectric (nonmetallic) material, the ratio of electrical capacitance of a dielectric-filled capacitor to a vacuum capacitor of identical dimensions.
Divergence – The deviation of field lines (e.g. electric field lines) from parallel, homogeneous conditions. A highly divergent field is a very inhomogeneous field where the value and direction of the field change drastically in the area under consideration.
Electrolytic – A fluid containing charged molecules is called an electrolyte. Electrolytic properties are associated with such a fluid, such as the ability to conduct current.
Electroporation – The application of high electric field pulses of short duration to create temporary pores (holes) in the membranes of cells.
Electroporation Cuvette – Square chambers with electrodes on two sides, usually measuring 1 mm, 2 mm or 4 mm in gap, for the purpose of electroporating cells in suspension. The cell type being electroporated typically determine the gap. Prokaryotic cells typically uses 1 mm or 2 mm gap cuvettes, while eukaryotic cells will typically use 2 mm or 4 mm gap sizes.
Exponential Decay Waveform – This waveform is mainly used for transforming cells during electroporation. In this type of pulse the set voltage is released from the capacitor and decays rapidly and exponentially over time (millisecs). The delivered pulse, is characterized by two parameters: the field strength (kV/cm) and the time constant. These parameters can be adjusted by varying voltage and capacitance settings to achieve a wide pulse gradient.
Field Strength – The potential difference between two points (electrodes) (in Volts) divided by the distance between the electrodes (called gap, and expressed in cm). Expressed as V/cm or kV/cm. This is true only if the electric field is homogeneous as it is in parallel plate electrodes.
Gap – The distance between electrodes.
Homogeneous Electric Field – The direction and field strength
are constant.
Hydrostatic Pressure – The pressure in liquids at rest.
Inhomogeneous Electric Field – Direction and strength of the
electric field vary.
Number of Pulses – The number of pulses the sample will be exposed to.
Osmotic Pressure – The applied pressure required to prevent the flow of solvents of different concentration across a semi­permeable membrane.
Pore – A small, mostly transient opening in a cell wall caused by the application of a brief high electric field pulse.
Potential Difference – The difference (in Volts) between points in an area between electrodes.
Protocols – The method for performing an experiment.
Pressure Gradient – The difference in pressure between two
points in a medium.
Pulse Interval – The time between multiple pulses.
Pulse Length/Pulse Duration – The length of time an electric
signal is applied.
Specialty Electrodes – Electrodes used with the BTX Gemini X2 to perform electroporation on a wide variety of cell forms. Some applications include in vivo, in ovo, in utero, or ex plant electroporation. BTX offer many types of specialty electrodes.
Square Waveform – This waveform is typically used for eukaryotic cells. It is characterized by the voltage delivered, the duration of the pulse, the number of pulses and the length of the interval between pulses.
Time Constant – (represented by the Greek letter tau, T) is the amount of time required for the actual voltage of the delivered pulse to decrease to a value 1/e of the true peak pulse.
Transfection – The introduction of nucleic acids into animal cells. Stable transfections result in integration of nucleic acids into host chromosomes and the inheritance of associated traits in progeny cells. Transient transfections result in temporary expression of exogenous nucleic acids.
Transformation – The introduction of nucleic acids into microorganisms and plant cells.
Turgor Pressure – The pressure in capillaries.
Voltage – The difference of electric potential between two
electrodes (expressed in volts (V) or kilovolts (kV)).
Waveforms – The shape of time-varying electric signals.
Gemini Series Electroporator User’s Manual
12
Page 13
Electroporation Buffers
There are many kinds of buffers and medium used for electroporating cells. Typically, we recommend using medias without serum or antibiotics.
The following is a list of the most commonly used buffers/ medium:
BTXpress – is a single buffer solution, developed to quickly and
efficiently deliver genes into mammalian cells that were previously considered “hard to transfect” by chemical and other non-viral methods. This solution, in combination with BTX electroporators, provides researchers with the versatility needed for success across a broad range of cell types while maintaining critical cell viability. Transfection using this high performance electroporation solution is equally effective in delivering DNA as well as siRNA into mammalian cells.
PBS – is a buffer solution commonly used in biological research. It is a water-based salt solution containing sodium chloride, sodium phosphate, and, in some formulations, potassium chloride and potassium phosphate. The buffer’s phosphate groups help to maintain a constant pH. The osmolarity and ion concentrations of the solution usually match those of the human body (isotonic).
HEPES – is widely used in cell culture, largely because it is better at maintaining physiological pH despite changes in carbon dioxide concentration (produced by cellular respiration) when compared to bicarbonate buffers, which are also commonly used in cell culture.
RPMI – is a form of medium used in cell culture and tissue culture. It has traditionally been used for growth of Human lymphoid cells. This medium contains a great deal of phosphate and is formulated for use in a 5% carbon dioxide atmosphere.
Water & 10% glycerol – Typically used for bacteria
Cytoporation Media T – is a buffer designed for larger volume
cell electroporation as it incorporates a low conductivity of 0.08 S/ cm to reduce heating of solution during electroporation.
Cytoporation Media T4 – is a buffer designed for larger volume cell electroporation as it incorporates a low conductivity of 3.45 mS/cm to reduce heating of solution during electroporation.
Opti-MEM – is an improved Minimal Essential Medium (MEM) that allows for a reduction of Fetal Bovine Serum supplementation by at least 50% with no change to growth rate or morphology. Opti-MEMt can be used with a variety of suspension and adherent mammalian cells, including Sp2, AE-1, CHO, BHK-21, HEK, and primary fibroblasts.
MEM Eagle – is suitable for a diverse spectrum of mammalian cell types. Various formulations available with either Hank’s or Earle’s salts.
DMEM – MEM is used in a wide range of mammalian cell culture applications. The high glucose version is well suited to high density suspension culture. The low glucose formula is used for adherent dependent cells.
CytoMix – is a composition of cytokines for the highly efficient and reproducible expansion of human multipotent mesenchymal stromal cells (MSCs).
Gemini Series Electroporator User’s Manual
13
Page 14
Unpacking the System
The shipping carton in which your BTX Gemini Twin Waveform Electroporation System is packed has been specifically designed to provide maximum protection to the instrument during transportation and normal handling conditions. Upon receipt, the carton should be examined for any external damage resulting from shipment.
Open the carton and carefully remove the BTX Gemini Twin Waveform Electroporator and inspect the unit for any apparent damage. Save the carton and packing materials for future transportation and shipping requirements.
Packing Data
Check the packing slip to ensure that all items ordered and listed are included in the shipment. Inform BTX immediately if any parts are missing or damaged.
Power Source
As received, the instrument is ready for use with either 100-240 V AC, 50/60 HZ.
The power requirements are 350 watts. In the USA, the power cord has a standard three prong plug.
Installation
Once you have determined that the components of the system have not sustained any obvious damage in shipment, proceed with the installation. The location of the BTX Gemini Twin Waveform Electroporator should be a dry, level, sturdy surface free from extremes in ambient temperature, dust or chemical exposures. Unpack the safety dome, cuvette rack and disposable cuvette chambers.
Connect the safety dome, or in the case of the BTX Gemini X2 system, specialty electrodes or HT plate handler, into the connectors at the bottom right-hand side of front panel.
Connect the mains/power cord to into the back panel at the bottom left.
Power up the system by pushing the rocker switch located on the back panel at the bottom left. The display will flash the BTX logo. Once the software initializes, the Main Menu screen will appear.
You are now ready to begin your work.
Touchscreen Display
USB Serial Input Footswitch Input
(switch sold separately)
Gemini SC
High Voltage Output
Fuse Holder
Main Power Switch
Universal Power Input
Gemini Series Electroporator User’s Manual
14
Page 15
Touchscreen Button Reference
Accept
Used to accept the parameters/settings on a screen and advance to the next screen in the menu, also used in place of a double tap on various icons.
Back
Used to go back one screen.
Cancel
Used to cancel any changes/entries on a screen and return to the previous screen in the menu.
Exponential Decay Wave Indicator
Used to indicate when a protocol utilizes exponential decay wave pulses.
File Options
Access the File Option menu that is used to save, rename and delete protocols.
Home/Main Menu
Access Preset Protocols, User Protocols and Settings.
Page Down
Used to page down in a display list.
Page Up
Used to page up in a display list.
Stop Protocol
Used during the pulse delivery sequence to stop the progress of the protocol.
Square Wave Indicator
Used to indicate when a protocol utilizes square wave pulses.
Touch Screen Locked
Indicates that the touch screen is currently locked. Press the icon and enter password to unlock the touch screen.
Touch Screen Unlocked
Indicates that the touch screen is currently unlocked. Pressing the icon twice will allow the user to password protect the protocol.
Pre-Pulse Resistance Measurement
Used to measure the resistance of the sample prior to delivering the DC pulse.
Run Protocol
Used to deliver the pulse protocol to the sample.
Scroll Down
Used to scroll down in a display list.
Scroll Up
Used to scroll up in a display list.
Settings
Access the settings menu used to adjust the following parameters: Date and Time, Audible Alarms, Backlighting, and software updates. Displays device information.
Gemini Series Electroporator User’s Manual
15
Page 16
Software Setup
Setting Time/Date
After initializing, the home screen will appear.
Tap the Gears icon
MAINSETTINGSDATE / TIME
Tap Time/Date icon.
Setting Audible Alarm Preferences
On the home screen, tap the Gears icon
Tap Audible Alarms icon.
Tap Date icon and use the keypad to type the date.
Tap Time icon and use the keypad to type the time.
Tap the Green Check icon to save and return to the settings screen.
NOTE: You may change the date and time format by tapping
✐
icons to the right of the given values.
Select preferred audible alarms by tapping the icons to activate or deactivate.
Tap the Green Check icon to save and return to the settings screen.
AUDIBLE ALARMS
Gemini Series Electroporator User’s Manual
16
Page 17
Software Setup
Setting Backlight Preferences
On the home screen, tap the Gears icon
Tap Backlight icon. Tap Device Information icon.
Displaying Device Information
On the home screen, tap the Gears icon.
Select preferred brightness by tapping the icons indicating % brightness.
Tap the Green Check icon to save and return to the settings screen.
(not shown)
BACKLIGHT
BACKLIGHT - KEYPAD
View device information.
Tap the Green Check icon to return to the settings screen.
INFO
Gemini Series Electroporator User’s Manual
17
Page 18
Preset Protocols
Using Preset Protocols
On the home screen, tap the Preset Protocols icon.
Double tap desired cell type.
PROTOCOL TYPEPROTOCOL SELECT
Using Preset Protocols (continued)
Review parameters.
Tap Omega icon to measure pre pulse load resistance.
RUN
With load measurement OK press the Circle icon to run protocol.
Scroll through available cells.
Double tap desired cells.
Once protocol is complete, data regarding your pulse is displayed and stored in the systems logs (BTX Gemini X2 only) for future use.
You may continue pulsing, go back to the protocol select screen, or to the home screen.
Gemini Series Electroporator User’s Manual
18
Page 19
Preset Protocols
Customizing a Preset Protocol
On the home screen, tap the Preset Protocols icon.
Double tap desired cell type.
Customizing a Preset Protocol (continued)
Use keyboard to name copy of protocol.
Tap the Green Check icon to save.
KEYBOARD
The message board will turn green and alert the user that the method has been copied.
Tap the Home icon.
Scroll through available cells.
Tap desired cells.
Tap Folder icon.
On the home screen, tap User Protocols icon.
Gemini Series Electroporator User’s Manual
19
Page 20
Preset Protocols
Customizing a Preset Protocol (continued)
Scroll through user protocols to locate the newly saved method.
Double tap the newly saved method.
Tap the parameter(s) requiring customization.
Proceed to modify selected parameters.
Tap the Green Check icon to proceed to the run screen.
PROTO EDIT
Gemini Series Electroporator User’s Manual
20
Page 21
Performing Experiments
Creating New Protocols
On the home screen, tap User Protocols icon.
Double tap New Protocol.
Creating an Exponential Decay Wave Protocol
Tap the Exponential icon.
Tap the Green Check icon to save.
PROTO SET TYPEEDIT SCREEN
Double tap your newly saved protocol.
Use keyboard to name protocol.
Tap the Green Check icon to save.
Tap the Voltage icon.
Use the numeric keypad to set voltage.
Tap the Green Check icon to save.
SET VOLUME
Gemini Series Electroporator User’s Manual
21
Page 22
Performing Experiments
Creating New Protocols (continued)
Creating an Exponential Decay Wave Protocol (continued)
Tap the Res. Cap. icon.
Tap the Resistance (ohms) icon.
Use the numeric keypad to set resistance value.
Tap the Pulses Interval icon.
Tap the # Pulses icon. Use the numeric keypad to set the # of pulses.
NOTE: Only the BTX Gemini X2 can perform multiple pulsing
✐
with Exponential Decay Wave, the BTX Gemini SC cannot perform multiple pulsing with Exponential decay waves.
SET RES CAP
EDIT SCREEN
Tap the Capacitance (uF) icon.
NOTE: LV and HV mode capacitance values differ. Use the
✐
numeric keypad to set resistance value.
EDIT SCREEN
SET PULSES INTVL
Tap the Interval (sec) icon.
Use the numeric keypad to set the interval between pulses.
Tap the Green Check icon to save.
NOTE: Intervals must only be set when performing
✐
multiple pulsing.
Gemini Series Electroporator User’s Manual
22
Page 23
Performing Experiments
Creating New Protocols (continued)
Creating an Exponential Decay Wave Protocol (continued)
Tap the Gap icon.
Use the numeric keypad to set the gap between electrodes.
Tap the Green Check icon to save.
View your protocol.
To run, tap the Omega icon.
With load measurement OK, press the Go icon to run protocol.
Tap the Green Check icon to proceed to the run screen.
EDIT SCREEN
SET PROTO GAP
Once protocol is complete, data regarding your pulse is displayed and stored in the systems logs (BTX Gemini X
You may continue pulsing, go back to set-up screen, or to the home screen.
NOTE: See the Protocol Manager user manual for instructions
✐
Gemini Series Electroporator User’s Manual
23
on viewing and downloading log data.
2
only) for future use.
Page 24
Performing Experiments
Creating New Protocols (continued)
Creating a Square Wave Protocol
Tap the Square Wave icon.
Tap the Green Check icon to save.
Double tap your newly saved protocol.
Use the numeric keypad to set voltage
Tap the Green Check icon to save.
EDIT SCREEN
SET VOLTAGE
Tap the Duration icon.
Tap the Voltage icon.
Use the numeric keypad to set the pulse length.
Tap the Green Check icon to save.
EDIT SCREEN
SET PULSE TIME
Gemini Series Electroporator User’s Manual
24
Page 25
Performing Experiments
Creating New Protocols (continued)
Creating a Square Wave Protocol (continued)
Tap the Pulses Interval icon.
Tap the # Pulses icon.
Use the numeric keypad to set the # of pulses.
Tap the Gap icon
Use the numeric keypad to set the gap between electrodes.
Tap the Green Check icon to save.
EDIT SCREEN
SET PROTO GAP
Tap the Interval (sec) icon.
Use the numeric keypad to set the pulse interval.
Tap the Green Check icon to save.
Tap the Green Check icon to proceed to the run screen.
Gemini Series Electroporator User’s Manual
25
Page 26
Performing Experiments
Creating New Protocols (continued)
Creating a Square Wave Protocol (continued)
View your protocol.
To run, tap the Omega icon.
With load measurement OK, press the Go icon to run protocol.
Once protocol is complete, data regarding your pulse is displayed and stored in the systems logs (BTX Gemini X2 only) for future use.
You may continue pulsing, go back to the set-up screen, or to the home screen.
Gemini Series Electroporator User’s Manual
26
Page 27
Using Specialty Protocols
Using an HT Plate Handler, Gemini X2 only
Plug the HT Plate Handler into the front ports of the BTX Gemini X2.
Scroll through available cells.
Double tap desired cells.
Review parameters.
Tap Plate Handler icon.
Review parameters.
Set the HT plate handler with the correct number of pre-pulse resistance measurement and DC pulses.
Tap the Go icon to run protocol.
Using the numeric keypad icon, type the number of columns being pulsed.
Tap the Green Check icon to save.
EDIT SCREEN
SET PLATE COLS
NOTE: Note the message box. It will indicate how many
✐
pulses should be set in the HT plate handler. This number will be a combination of both pre-pulse resistance measurements and DC pulses.
The Square icon can be used to stop the protocol at any time.
NOTE: Generator will display readings of the 1st and last pulse
✐
Gemini Series Electroporator User’s Manual
27
of each column.
Page 28
Using Specialty Protocols
Using an HT Plate Handler, Gemini X2 only (continued)
Once protocol is complete, data regarding your pulse is displayed and stored in the systems logs for future use.
You may continue pulsing, go back to the settings screen or protocol select screen, or to the home screen.
Using Specialty Electrodes, Gemini X2 Only
Plug the specialty electrodes into the front ports of the BTX Gemini X2
Scroll through available protocols.
Double tap desired selection.
Review parameters.
Position electrodes on area being electroporated.
To run tap the Omega icon.
With load measurement OK, press the Go icon to run protocol.
Scroll through available cells.
Double tap desired cells.
While the pulse is being delivered, the Square icon can be used to stop the protocol at any time.
NOTE: Generator will display readings after each pulse.
✐
Gemini Series Electroporator User’s Manual
28
Page 29
Using Speciality Protocols
Using Specialty Electrodes, Gemini X2 Only
Once protocol is complete, data regarding your pulse is displayed and stored in the systems logs for future use.
You may continue pulsing, go back to the settings screen or the protocol select screen, or to the home screen.
Gemini Series Electroporator User’s Manual
29
Page 30
Managing Protocols
All Preset and User-Defined Protocols are stored in the Protocol List for ongoing use. Over time, you may want to manage previously created Protocols to keep your Protocol List easy to navigate. In addition, the BTX Gemini X2 allows you to export Protocol information to an externally connected generator or to a computer that will be used to control operation. This section of the Guide describes the file management facilities available on the BTX Gemini X2.
Saving a Copy of a Protocol
Follow these instructions to save a copy of a Protocol under a new name:
1. From Protocol Configuration screen, choose the Protocol Name button to enter the Mode Selection screen.
2. From the list of available Protocols, select (press once) the Protocol you want to save a copy of, then choose the File Options button to enter the File Options screen.
3. Choose Save As/Copy to access the Protocol Name Entry screen. The current Protocol Name will be displayed in the text entry box.
4. On the Protocol Name screen, type the new name for your Protocol using the onscreen keyboard. Choose the Symbols/ Numbers button (.?123) to display the list of available numbers and symbols that can be included in the Protocol Name. To switch the keyboard back to letters mode, choose the Letters button (ABC). [15 characters maximum]
5. When finished entering the new Protocol Name, choose Accept to save your changes and return to the File Options screen.
6. Choose BACK to return to the Mode Selection screen
Renaming a Protocol
Follow these instructions to rename a previously saved user protocol:
1. From the Protocol Configuration screen, choose the Protocol Name button to enter the Protocol Mode screen.
2. From the list of available Protocols, select (press once) the Protocol you want to rename, then choose the File Options button to enter the File Options screen.
3. Choose Rename Protocol to access the Protocol Name Entry screen. The current Protocol Name will be displayed in the text entry box.
4. On the Protocol Name screen, type the new name for your Protocol using the onscreen keyboard. Choose the Symbols/ Numbers button (.?123) to display the list of available numbers and symbols that can be included in the Protocol Name. To switch the keyboard back to letters mode, choose the Letters button (ABC). [15 characters maximum]
5. When finished entering the new Protocol Name, choose Accept to save your changes and return to the File Options screen.
6. Choose Back to return to the Protocol.
FILE OPTIONS
Gemini Series Electroporator User’s Manual
30
Page 31
Managing Protocols
Deleting a Protocol
NOTE: Preset protocols cannot be deleted.
✐
On the home screen, tap the User Protocols icon.
Tap the protocol you wish to delete.
Tap Folder icon.
Confirm delete by tapping the Delete icon again.
Protocol has been removed from the menu.
Tap Delete icon.
Gemini Series Electroporator User’s Manual
31
Page 32
Managing Protocols
Password Protecting a Protocol
On the home screen, tap the User Protocols icon.
Double tap protocol you wish to password protect.
Confirm password protection by tapping the Open Lock icon again.
Use the numeric keypad to set the password.
Tap the Green Check icon to save.
LOCK
View the protocol.
Tap the Open Lock icon.
The protocol is now locked against modifications.
To Unlock press the lock icon. Type in the password then Tap the Green Check icon.
Gemini Series Electroporator User’s Manual
32
Page 33
Protocol Manager Software
Installation
Insert the generator CD into the CDROM, a file containing the CD’s contents will open automatically. Navigate to the Protocol Manager folder, under Applications. Right click on “Protocol Manager”. Select “Send To” then “Desktop” to create a shortcut on your desktop.
NOTE: Compatible with Windows XP & Windows 7.
✐
1. Connect the generator to the PC using a USB cable.
2. Turn generator on.
3. Start Protocol Manager.
4. Select the proper COM port.
Overview
Protocol Manager allows the user to:
• Upload protocols from the Gemini X2 generator to a PC.
• Download protocols from PC to the Gemini X2 generator.
• Upload log files from the Gemini X2 generator to PC.
Upload– Generator to PC
1. Select Protocol(s)
2. Click ‘Save Protocol’
NOTE: Hit ‘Refresh’ to update the list after creating or editing
✐
protocols on the Gemini X2 generator.
Log File
Control
PC to Generator Protocol Control
3. Select the drive and folder location
4. Click ‘Save’
Generator to PC Protocol Control
Gemini Series Electroporator User’s Manual
33
Page 34
Protocol Manager Software
Download - PC to Generator
1. Click ‘Choose Protocol’
2. Select Protocol(s) to download
3. Click ‘Open’
✐
NOTE: If a protocol already exists on the Gemini X2 generator, Protocol Manager will not overwrite the le if you attempt to re-download it.
4. Protocol name being downloaded is shown
Gemini Series Electroporator User’s Manual
34
Page 35
Protocol Manager Software
Upload Log Files
1. On the main screen, click ‘Pulse Data’
2. Select your desired files or folders and then click ‘Transfer’.
NOTE: Files should be backed up periodically. The generator’s
✐
3. Select the drive or folder location.
SD card can store up to approximately 100,000 les.
✐
NOTE: Log Files are saved as a tab delimited text le. They can be opened with any common text editor like Notepad or by using spreadsheet software such as Excel.
Troubleshooting Protocol Manager
1. Do not upload/download Protocols while generator is running. Program works best while generator is idle on Protocol setup screen.
2. Order of Operations:
A. Connect USB.
B. Turn generator on.
C. Start Protocol Manager program.
Gemini Series Electroporator User’s Manual
35
Page 36
Remote Control Software
Installation
Insert the generator CD into the CDROM. A file containing the CD’s contents will open automatically. Navigate to the Remote
Screen Interface folder, and then to the Remote Screen Interface Support Files Installer folder. Run the “setup.exe”
file. Follow the instructions provided on the installation screens. After the installation is complete, copy the folder Remote Screen Interface Application from the CD onto the PC in a location of your choice.
NOTE: during the installation, it will ask you to provide a
✐
Overview
Remote Screen Interface allows you to control the Gemini remotely. When you click on the computer screen, you control the generator.
While the Gemini generator is connected to the computer via USB, open the Control Panel and select System and Security, and then
System. Click Device Manager on the left hand menu. Under Ports, search for the “USB serial port for BTX Gemini.” This will show
the COM Port to which the generator is connected, as shown below.
location to install your application les. There is a known issue with this installer where it will not actually create this directory.
Once the COM port is known, open the Remote Screen Interface application. Choose the correct COM port and click Connect.
You should now be able to control your Gemini generator remotely.
Gemini Series Electroporator User’s Manual
36
Page 37
Generator Specifications
Generator Specifications Gemini X2 Gemini SC
Square Wave Pulse Ranges
Voltage Range
LV Mode
HV Mode
Voltage Accuracy
LV Mode
HV Mode
Pulse Length Range
LV Mode
HV Mode
Multiple Pulsing
LV Mode
HV Mode
Pulse Interval
Capacitance in µF
LV Mode
HV Mode
Voltage Droop
Exponential Decay Pulse Ranges
Voltage Range
LV Mode
HV Mode
Voltage Accuracy
LV Mode
HV Mode
Time Constant Range
LV Mode
HV Mode
Multiple Pulsing
Multiple Pulsing
Pulse Interval
Capacitance in µF
LV Mode
HV Mode
Available Internal Resistance Selections
All Modes
Sample Resistance (Load) Exponential Decay or Square
LV Mode
HV Mode
Other Specifications
Charging Time
Monitoring
Display
Controls
Programmability
Arc Control
5-500 in 1 V steps 10-500 V in 5 V steps 505-3000 in 5 V steps 510-3000 in 10 V steps
5% 5% 5% 5%
10-999 µs in 1 µs step, 1-999 ms in 1 ms step 0.05-10 ms in 0.05 ms steps, 10-100 ms in 1 ms step 10-600 µs / 1 µs step 50 µs-5 ms in 50 µs steps
1-10 pulses per sample 1 to 10 1-3 pulses per sample 1 to 2
0.1 s to 10 s 0.1 s to 10 s
3775 µF 3775 µF 85 µF 85 µF 20% Load < 1 K Ω, 10% Load > 1 K Ω 20% Load < 1 K Ω, 10% Load > 1 K Ω
5-500 in 1 V steps 10-500 in 5 V steps 505-3000 in 5 V steps 510-3000 / 10 V steps
5% 5% 5% 5%
1 ms to 5.158 s 1.25 ms to 3.275 s / 1.25 ms
0.5 ms to 133.875 ms 0.5 ms to 50 ms / 0.5 ms
1 to 2 depending on device’s current amplitude limit No 5 to 30 s N/A
25-3275 µF in 25 µF steps 25-3275 µF in 25 µF steps 10, 25, 35, 50, 60, 75, 85 µF 10, 25, 50 µF
LV 25-1575 Ω in25 Ω steps, HV 50-1575 Ω in 25 Ω steps 50-1000 Ω in 50 Ω steps
PL ≤ 100ms, load ≥ 8-9 Ω, PL >100ms, load ≥100 Ω Load ≥ 10 Ω Load ≥ 40 Ω 510-2500 Load ≥ 20 Ω, 2500 to 3000 V, Load ≥ 600 Ω
LV < 7 s, HV < 4 s LV < 7 s, HV < 4 s Pulse Voltage, Width, Droop % and Sample Resis-
tance
4.3 in color display 4.3 in color display Touch Screen Touch Screen Storage over 1000 Protocols Storage over 1000 Protocols Yes Yes
Pulse Voltage, Width, Droop % and Sample Resis­tance
Gemini Series Electroporator User’s Manual
37
Page 38
Generator Specifications
Generator Specifications (V7) Gemini X2 Gemini SC
Other Spcifications (continued)
Interlock Switches Ye s Ye s
Pre-Pulse Sample resistance check
Pulse Over Current Protection Ye s Yes
PC control Ye s No
PC communications Ye s No
Log Report Ye s No
Remote Operation Ye s No
Foot Switch Control Ye s No
Power Ratings 35 W idle and 350 W pulsing 35W idle and 350 W pulsing
Input Voltage ratings 100 to 240 VAC 100 to 240 VAC
CE, ETL Marking, CB Scheme, EuRoHS
Dimensions (H x W x D) inches 12.75 x 11.25 x 8.5 12.75 x 11.25 x 8.5
Weight ~16 lb ~16 lb
Warranty 2 years 1 year
Atmospheric Specifications
Operating Temperature 4˚C to 40˚C (40˚F to 104˚F)
Storage Temperature -10˚C to 70˚C (14˚F to 158˚F)
Operating Humidity See Chart Below
Storage Humidity 20% to 80% RH, non condensing
Mode of Operation Continuous
Classification Class I
Pollution Degree 1
Installation Category II
Supplier Name BTX
Supplier Address 84 October Hill Rd., Holliston, MA 01746
Supplier Phone Number 508-893-8999
Regulatory Certifications CE, ETL (UL, CSA), WEEE, EU RoHS & CB Scheme
Safety Declarations ANSI/UL 61010-1 ; CAN/CSA C22.2 No. 61010-1;
EMC Declarations IEC 61326-1
FCC 47 CFR Part 15 B
Ye s Ye s
Ye s Ye s
IEC 61010-1; CENELEC EN 61010-1; CB
Gemini Series Electroporator User’s Manual
38
Page 39
Maintenance
The BTX Gemini Twin Wave Series generators require no special maintenance other than keeping them clean.
To clean the exterior surfaces, use a lint-free cloth to remove loose dust. Use care to avoid scratching the clear display window. For more efficient cleaning, use a soft cloth dampened (not soaked) with an aqueous solution of 75% isopropyl alcohol, or a mild detergent.
Fuse Replacement
Key Information
1. Make sure the power cord is disconnected from the main supply before servicing the fuse.
2. Use only Type 3AG, 1/4 x 11/4 in, 5 amp, 250 volt; Type T (time delay) fuses.
Turn off power and remove power cord from power module. Use a straight blade screw-driver to pry open the access door. Remove the fuse holder and then remove the fuses from this holder as shown in figure below. Replace fuses, and then replace the fuse holder.
Battery Replacement
1. Remove screw and swing cover to the side.
2. Slide battery out from under clip.
3. Install Lithium Coin Battery: 3V, 16mm CR1620 by sliding under clip with positive side facing out.
CAUTION: Use only same type and rated battery. Observe
✐
polarity when installing.
Gemini Series Electroporator User’s Manual
39
Page 40
Upgrading Gemini Series Software
USB Virtual Commport Driver Installation
When you connect the BTX Gemini X2 to a computer via USB for the first time, Windows will seek to install a driver for communication. The following section details the installation of the Virtual CommPort Driver supplied with the BTX Gemini X2.
1. Connect the Gemini X2 generator to the computer via USB. Open the control panel and select System and Security and then System. Click Device Manager on the left hand menu.
Mating Face
Pin # Signal
1 +5V
2 - Data
3 + Data
4 GND
2. Two Flash update applications will be displayed in the device manager, under Other Devices. Right click on either of the flash update applications and select Update Driver Software.
3. Select Browse my computer for driver software
Gemini Series Electroporator User’s Manual
40
Page 41
Upgrading Gemini Series Software
USB Virtual Commport Driver Installation (continued)
4. Click Browse and navigate to the generator drivers folder located on the CD supplied with the generator. Select the file BTX USB-CDC.inf. Make sure the Include Subfolders box is checked and then click Next.
5. If a warning message regarding the driver publisher is displayed, choose Install this driver software anyway.
6. Click Close after the software has been successfully installed.
Gemini Series Electroporator User’s Manual
41
Page 42
Upgrading Gemini Series Software
Device Updater Program
NOTE: Prior to upgrading, users will need to install USD
✐
1. Upload the latest software versions to your desktop (format is filename.srec). There are two .srec files that both need to be updated: ‘BTX Gemini vx.x.x’ and a ‘BTX Gemini Pwr vx.x.x’.
2. Disconnect all I/O devices and then connect the Gemini X2 generator to PC using a USB cable.
3. On the Set up screen of the Gemini X2 the Upgrade Software button twice to enter Boot Loader Mode.
✐
4. From the CD provided with the Gemini X2 generator, open the Device Updater Application.
drivers as well as the bootloader driver le. Reference ‘Virtual Comport Driver Installation’.
generator, press
NOTE: To exit boot loader mode without upgrading the software, power cycle the generator.
5. Click Open Update File and browse to the ‘BTX Gemini vx.x.x.srec’. file that was previously saved to the desktop.
6. Click Start Update.
7. After the update is complete, press Reboot and then power cycle the Gemini X2 generator from the power switch in the back of the unit.
8. Repeat steps 5 through 7 with the ‘BTX Gemini Pwr vx.x.x.srec’.
Gemini Series Electroporator User’s Manual
42
Page 43
Error Messages & Troubleshooting
Display Error Message Error Description User Troubleshooting
Power Loss Error Power loss during pulsing. Power lost while protocol was running.
PFC Over Temp Pfc module over temperature. Hardware error. Please contact the manufacturer.
Interlock Open Electrode interlock open. Connect electrode connectors to the front of the generator and make sure the
IGBT Switch Fault IGBT switches fault. Hardware error. Please contact the manufacturer.
Charge Voltage Max Error Charging voltage above the expected voltage. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
Charge Current Max Error Charging current above the expected current. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
LV Cap Bank Max Low voltage capacitor bank voltage above expected value. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
LV Cap Bank Min Low voltage capacitor bank voltage below expected value. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
HV Cap Bank Max High voltage capacitor bank voltage above expected value. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
HV Cap Bank Min High voltage capacitor bank voltage below expected value. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
Charge Time Max Charging capacitor bank to selected voltage exceeded max time estimated. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
Pulse Droop Error Pulse voltage droop exceeded estimated droop. A) Increase the sample resistance. This can be achived by:
Pulse Voltage Overshoot Pulse voltage monitor (p-vmon) pulse overshoot. Hardware error. Please contact the manufacturer
Pulse Voltage Mon too Low Pulse voltage monitor (p-vmon) below min voltage. Hardware error. Please contact the manufacturer
Pulse Voltage Mon too High Pulse voltage monitor (p-vmon) above max voltage. Hardware error. Please contact the manufacturer
Pulse Current Mon too Low Pulse current monitor (p-imon) below min current. Hardware error. Please contact the manufacturer
Pulse Current Mon too High Pulse current monitor (p-imon) above max current. Hardware error. Please contact the manufacturer
Arc_Detected Arc detected during pulsing. Arc could happen due to the one or more of the following conditions:
No_HB_from_GUI Communication error. Hardware error. Please contact the manufacturer
Sample Resistance Out Of Range
Over Current Pulse Abort Pulse aborted due to pulse over current. The over current protection feature is added to prevent sample arcing.
Idle V I Max Voltage or current in idle state is higher than expected. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
Charge Cap Bank A Max Charge capacitor bank A voltage is above tolerance. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
Charge Cap Bank A Min Charge capacitor bank A voltage is below tolerance. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
Charge Cap Bank B Max Charge capacitor bank B voltage is above tolerance. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
Charge Cap Bank B Min Charge capacitor bank B voltage is below tolerance. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
Charge Cap Bank C Max Charge capacitor bank C voltage is above tolerance. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
Charge Cap Bank C Min Charge capacitor bank C voltage is below tolerance. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
Dump Time Max The capacitor bank did not fully discharge in the estimated time. Turn off the unit for a few minutes then turn back on. Re-run the protocol if the
Invalid Protocol The generator was requested to run an invalid protocol. Verify that the used protocol is within the allowed specifications of the generator
System Processing (Busy) The control command is invalid or was issued while the generator was busy. Allow more time between events
TC_Error The measured rc time while delivering an exponential decay pulse is invalid. Hardware error. Please contact the manufacturer
Pulse Duration Error The pulse duration did not meet the expected duration. Hardware error. Please contact the manufacturer
Unit Type Sel Error The unit hardware and software do not match the corect type. Hardware error. Please contact the manufacturer
Error: 0X# Multiple errors. Multiple error. Please contact the manufacturer
The measured sample load is out of range. Increase the sample resistance. This can be achived by
cuvett safety dome cover is closed.
issue recurs, contact the manufacturer.
issue recurs, contact the manufacturer.
issue recurs, contact the manufacturer.
issue recurs, contact the manufacturer.
issue recurs, contact the manufacturer.
issue recurs, contact the manufacturer.
issue recurs, contact the manufacturer.
1. Increasing the gap size
2. Reducing the volume
3. Use a less conductive buffer B) Reduce the pulse duration
1. Sample resistance is very low for the voltage selected
2. Gap size is too small for the voltage selected
3. Pulse duration is too long
1. Increasing the gap size
2. Reducing the volume
3. Replace the buffer with less conductive type of buffer
This could occur when the sample resistance changes during the pulse:
1. Sample resistance is very low for the voltage selected
2. Gap size is too small for the voltage selected
3. Pulse duration is too long
issue recurs, contact the manufacturer
issue recurs, contact the manufacturer
issue recurs, contact the manufacturer
issue recurs, contact the manufacturer
issue recurs, contact the manufacturer
issue recurs, contact the manufacturer
issue recurs, contact the manufacturer
issue recurs, contact the manufacturer
Gemini Series Electroporator User’s Manual
43
Page 44
Troubleshooting
NOTE: It is recommended that you save your protocols
✐
Out of Range
A value was entered or encountered in a protocol that was beyond the generator’s limits.
Power Failure Notification
If power is interrupted during operation, an alarm will sound to alert the user to the interruption. When power is restored, an onscreen message is displayed to indicate the interruption.
periodically by uploading them to a PC.
Low Battery Indication
Time/Date area will turn red when battery is low.
Gemini Series Electroporator User’s Manual
44
Page 45
Experiment Troubleshooting
Arcing
Verify electrical component functionality. Verify properties of cell sample (do cells need to be washed? Is the buffer appropriate for application?). Verify properties of transfectant molecule (Is the DNA well purified?) Try reducing the voltage or increase sample volume until arcing is no longer a problem.
Low (or no) Transfection Efficiency, or Incorporation
(Verify physical, biological, chemical parameters) Verify delivery
of the pulse and pulse parameters. Is the voltage correct? Chamber gap? Pulse length or appropriate instrument settings? Number of pulses? If so, follow Optimization Guidelines outlined.
Low Viability
Verify physical, biological, chemical parameters. Is the voltage, chamber gap, pulse length (time constant), pulse number and other instrument settings correct? If so, reduce voltage, pulse length, or number of pulses and re-optimize protocol to improve viability as outlined .
Voltage Drop
A drop in output voltage accompanies pulse delivery into highly conductive samples (for example, PBS). Thus the displayed voltage may in these situations be less than that expected, given 5% full scale accuracy and the monitoring accuracy of 5%.
Gemini Series Electroporator User’s Manual
45
Page 46
Ordering Information
Order No. Description UOM
Electroporation Systems
45-2001 BTX Gemini SC Generator only 1 ea
45-2002 BTX Gemini SC Electroporation System.
Includes Gemini SC Generator, Cuvettes - 610, 620, 640 pack of 30 (10 ea), Safety Dome 1, and 660 Cuvette Rack
45-2006 BTX Gemini X2 Generator only 1 ea
45-2007 BTX Gemini X2 Electroporation System.
Includes Gemini X2 Generator, Cuvettes - 610, 620, 640 pack of 30 (10 ea), Safety Dome 2, and 660 Cuvette Rack
45-2008 BTX Gemini X2 HT Electroporation System. In-
cludes Gemini X2 Generator, Cuvettes - 610, 620, 640 pack of 30 (10 ea), Safety Dome 2, HT 200 Plate Handler, 1 x 2 mm gap HT Plate, 1 x 4 mm HT Plate and 660 Cuvette Rack
Cuvettes
45-0124 Cuvette Plus, 1 mm gap, 90 µl, pk/50, Gray 50 pk
45-0125 Cuvette Plus, 2 mm gap, 400 µl, pk/50, Blue 50 pk
45-0126 Cuvette Plus, 4 mm gap, 800 µl, pk/50, Yellow 50 pk
45-0140 Bulk Cuvette, 1 mm, 24 pk/100, Gray Case 2400 ea
45-0141 Bulk Cuvette, 2 mm gap, 24 pk/100, Blue Case 2400 ea
45-0142 Bulk Cuvette, 4 mm gap, 24 pk/100, Yellow Case 2400 ea
HT Plates (GeminX2 use only)
45-0462 25 Well Plate, 4 mm gap, 250 µl 1 ea
45-0463 25 Well Plate, 4 mm gap, 250 µl, pk/6 6 pk
45-0466 25 Well Plate, 2 mm gap, 125 µl 1 ea
45-0467 25 Well Plate, 2 mm gap, 125 µl, pk/6 6 pk
45-0450 96 Well Plate, 2 mm gap, 125 µl 1 ea
45-0452 96 Well Plate, 4 mm gap, 250 µl 1 ea
Reagents
45-0802 BTXpress, 5 ml 1 ea
45-0803 BTXpress, 5 ml with 2 mm gap, Cuvettes pk/50 1 ea
45-0804 BTXpress, 5 ml with 4 mm gap, Cuvettes pk/20 1 ea
45-0805 BTXpress, 10 ml 1 ea
45-0806 BTXpress, 10 ml with 2 mm gap, Cuvettes pk/50 1 ea
45-0807 BTXpress, 10 ml with 4 mm gap, Cuvettes pk/40 1 ea
47-0002 Cytoporation Media T 500 ml 1 ea
47-0003 Cytoporation Media T4 500 ml 1 ea
1 ea
1 ea
1 ea
Order No. Description UOM
Accessories
45-0400 HT 100 Plate Handler Manual 1 ea
45-0401 HT 200 Plate Handler Auto 1 ea
45-2020 BTX Safety Dome for SC 1 ea
45-2021 BTX Safety Dome forX2 1 ea
45-2030 BTX Gemini X2 Footswitch 1 ea
45-0208 Cuvette Rack 1 ea
45-0465 HT 25 Well Adapter Plate 1 ea
45-0468 Plate Handler Pins, pk/25 25 ea
45-0469 Plate Handler Pins, pk/100 100 ea
5012-017 Pliers for Plate Handler Removing Pins 1 ea
4500-012 25 Well Plate Seal 2 ea
4500-015 96 Well Plate Seal 2 ea
Cables
45-0216 Connection Cable Cable, 10 ft, Banana to
Micrograbber
45-0204 Tweezertrode Cable / Single Adaptor Cable for
Tissue Slice Electrode Positive
45-0503 Mini Micro Grabber Adapter Cables for Tissue Slice
Chamber / L Shaped Needle Electrodes
45-0087 Micrograbber to Banana Adapter Set 45-0217 1 pr
45-2031 BTX USB Cable, 2 m 1 ea
45-2032 BTX USB Cable, 5 m 1 ea
45-0217 Electrode Cable for Flat Electrode, 10 ft BTB 1 ea
45-0088 Banana Splice F/F Adapter Set, 45-0216/45-0217 1 pr
45-0090 Adapter Set Banana to Pin Tip 1 ea
45-0089 Banana to Square Post Adapter Set for 45-0217 1 pr
1 ea
1 ea
1 ea
Gemini Series Electroporator User’s Manual
46
Page 47
Ordering Information
Order No. Description UOM
45-0101 Caliper Electrode, 1.0 x 1.0 cm Kit 1 Ea
45-0102 Caliper Electrode, 2.0 x 2.0 cm, 1.5 x 1.5 cm Kit 1 Ea
45-0103 Microslide 450, 0.5 mm gap, 20 µl, pk/10 10 Pk
45-0104 Microslide 450-1, 1 mm gap, 40 µl, pk/10 10 Pk
45-0105 Microslide 453, 3.2 mm gap, 650 µl 1 Ea
45-0106 Microslide 453-10, 10 mm gap, 2.0 ml 1 Ea
45-0107 Meander Fusion Chamber, 0.2 mm gap, pk/4 4 Pk
45-0108 Flat Electrode/Divergent Field, 1 mm
(needs 45-0217)
45-0217 Electrode Cable for Flat Electrode, 10 ft BTB 1 Ea
45-0109 Flatpack Chambers, 1.83 mm gap, pk/50 50 Pk
45-0110 Flatpack Chambers, 0.56 mm gap, pk/50 50 Pk
45-0113 Genetrodes Straight, 5 mm, Gold Tip 1 Ea
45-0160 Genetrodes Straight, 5 mm, Gold Tip Kit 1 Ea
45-0114 Genetrodes Straight, 10 mm, Gold Tip 1 Ea
45-0161 Genetrodes Straight, 10 mm, Gold Tip Kit 1 Ea
45-0115 Genetrodes L-Shape, 5 mm, Gold Tip 1 Ea
45-0162 Genetrodes L-Shape, 5 mm, Gold Tip Kit 1 Ea
45-0116 Genetrodes L-Shape, 3 mm, Gold Tip 1 Ea
45-0163 Genetrodes L-Shape, 3 mm, Gold Tip Kit 1 Ea
45-0117 Genetrodes L-Shape, 1 mm, Gold Tip 1 Ea
45-0164 Genetrodes L-Shape, 1 mm, Gold Tip Kit 1 Ea
45-0203 Genetrodes/Genepaddle Holder with Shaft 1 Ea
45-0216 Genetrodes/Genepaddle Cable, 10 ft, BTMG 1 Ea
45-0122 Genepaddles, 3 x 5 mm 1 Ea
45-0169 Genepaddle, 3 x 5 mm Kit 1 Ea
45-0123 Genepaddles, 5 x 7 mm 1 Ea
45-0170 Genepaddle, 5 x 7 mm Kit 1 Ea
45-0167 2-Needle Array, 10 mm Kit 1 Ea
45-0205 2-Needle Array Handle, 10 mm (needs 45-0120) 1 Ea
45-0120 2-Needle Array, 10 mm, pk/6 (needs 45-0205) 1 Ea
45-0168 2-Needle Array, 5 mm Kit 1 Ea
45-0206 2-Needle Array Handle, 5 mm (needs 45-0121) 1 Ea
45-0121 2-Needle Array, 5 mm pk/6 (needs 45-0206) 1 Ea
45-0510 Needle L-Shaped Pt Electrode, 3 mm Kit 1 Ea
45-0509 Needle L-Shaped Pt Electrode, 3 mm
(needs 45-0508)
45-0513 Petri, 7 mm, Tissue Chamber Kit 1 Ea
45-0505 Petri Dish Tissue Chamber, 5 x 5 mm Kit 1 Ea
45-0504 Petri Dish Tissue Chamber, 5 x 5 mm
(needs 45-0216)
45-0506 Petri Dish Tissue Chamber, 15 x 15 mm
(needs 45-0216)
45-0507 Petri Dish Tissue Chamber, 15 x 15 mm Kit 1 Ea
45-0100 Petri Dish Electrode, 2 mm gap, 90 mm Well Dish Kit 1 Ea
45-0130 Petri Pulser for 6 Well Plates, 35 mm Well Kit 1 Ea
45-0490 Tissue Slice Chamber, 7 x 7 mm Kit 1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
Order No. Description UOM
45-0491 Tissue Slice Chamber, 7 x 7 mm Dish
(needs 45-0492, 45-0503, 45-0204)
45-0492 Tissue Slice Wand, (+) 7 mm
(needs 45-0491, 45-0503, 45-0204)
45-0500 Tissue Slice Chamber, 10 x 10 mm Kit 1 Ea
45-0501 Tissue Slice Chamber, 10 x 10 mm Dish
(needs 45-0502, 45-0503, 45-0204)
45-0502 Tissue Slice Wand, (+) 10 mm
(needs 45-0501, 45-0503, 45-0204)
45-0503 Mini Micro Grabber Adapter Cables for Tissue Slice
Chamber / L Shaped Needle Electrodes
45-0530 Adherent Cell Electrode, 3 mm gap
(needs 45-0204)
45-0531 Adherent Cell Electrode, 3 mm Kit 1 Ea
45-0486 PT Tweezertrodes, 1 mm Diameter Kit 1 Ea
45-0487 PT Tweezertrodes, 3 mm Diameter Kit 1 Ea
45-0489 PT Tweezertrodes, 5 mm Diameter Kit 1 Ea
45-0488 PT Tweezertrodes, 7 mm Diameter Kit 1 Ea
45-0165 SS Tweezertrode, 7 mm Diameter Kit 1 Ea
45-0118 SS Tweezertrode, 7 mm Diameter (needs 45-0204) 1 Ea
45-0166 SS Tweezertrode, 10 mm Diameter Kit 1 Ea
45-0119 SS Tweezertrode, 10 mm Diameter
(needs 45-0204)
45-0524 PT Tweezertrode, 1 mm Flat (needs 45-0204) 1 Ea
45-0525 PT Tweezertrode, 1 mm Flat Kit 1 Ea
45-0204 Tweezertrode Cable / Single Adaptor Cable
for Tissue Slice Electrode Positive
Specialty Electrodes (AgilePulse Systems Only)
47-0090 Electrode Adapter Box for Agile Pulse In Vivo 1 Ea
47-0000 Parallel-Needle Array Handle for AP In Vivo 1 Ea
47-0040 4-Needle Array, 4 mm gap, 2 mm Length,
AP In Vivo (ID*)
47-0043 3-Needle Array, 4 mm gap, 3 mm Length,
AP In Vivo (IM)
47-0045 4-Needle Array, 4 mm gap, 5 mm Length,
AP In Vivo (IM*)
47-0050 6-Needle Array, 4 mm gap, 2 mm Length,
AP In Vivo (ID*)
47-0060 6-Needle Array, 6 mm gap, 2 mm Length,
AP In Vivo (ID)
47-0070 6-Needle Array, 6 mm gap, 10 mm Length,
AP In Vivo (IM*)
47-0080 6-Needle Array, 6 mm gap, 12 mm Length,
AP In Vivo (IM)
47-0086 6-Needle Array, 6 mm gap, 16 mm Length,
AP In Vivo (IM)
47-0204N 5 ml Chamber for Agile Pulse Max 1 Ea
47-0090 Electrode Adapter Box for Agile Pulse In Vivo 1 Ea
47-0206 10 ml Chamber for Agile Pulse Max 1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
1 Ea
Gemini Series Electroporator User’s Manual
47
Page 48
Warranty Information
RESEARCH ONLY
BTX
84 October Hill Rd Holliston, MA 01746, USA
Phone: 1-508-893-8999 Fax: 1-800-429-5732 Web: www.btxonline.com
Warranty
BTX warranties the Gemini X2 & BTX Gemini SC Twin Waveform Electroporation Systems for a period of two years from the date of purchase. At its option, BTX will repair or replace the unit if it is found to be defective as to workmanship or materials. This warranty does not extend to any instrumentation which has been (a) subjected to misuse, neglect, accident or abuse, (b) repaired or altered by anyone other than BTX without BTX express and prior approval, (c) used in violation of instructions furnished by BTX. This warranty extends only to the original customer purchaser. IN NO EVENT SHALL BTX BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES. Some states do not allow exclusion or limitation of incidental or consequential damages so the above limitation or exclusion may not apply to you. THERE ARE NO IMPLIED WARRANTIES OF MERCHANTABILITY, OR FITNESS FOR A PARTICULAR USE, OR OF ANY OTHER NATURE. Some states do not allow this limitation on an implied warranty, so the above limitation may not apply to you. Without limiting the generality of the foregoing, BTX shall not be liable for any claims of any kind whatsoever, as to the equipment delivered or for non-delivery of equipment, and whether or not based on negligence. Warranty is void if the BTX Gemini X2 & BTX Gemini SC instrument is changed in any way from its original factory design or if repairs are attempted without written authorization by BTX. Warranty is void if parts, connections or electrodes not manufactured by BTX are used with the BTX Gemini X2 & BTX Gemini SC instrument. If a defect arises within the warranty period, promptly contact BTX, 84 October Hill Road, Building 7, Holliston, Massachusetts, USA 01746-1388 using our toll free number 1-800-272-2775 (US Only) or 508-893-8999 (E-mail: [email protected]). Goods will not be accepted for return unless an RMA (Returned Materials Authorization) number has been issued by our customer service department. The customer is responsible for shipping charges. Please allow a reasonable period of time for completion of repairs, replacement and return. If the unit is replaced, the replacement unit is covered only for the remainder of the original warranty period dating from the purchase of the original device. This warranty gives you specific rights, and you may also have other rights, which vary from state to state.
will be glad to help at no charge.
Repair service will be billed on the basis of labor and materials. A complete statement of time spent and materials used will be supplied. Shipment to BTX should be prepaid. Your bill will include return shipment freight charges.
Disassembly by the user is prohibited. Service should only be carried out by experienced BTX technicians.
Repair Facilities and Parts
BTX stocks replacement and repair parts. When ordering, please describe parts as completely as possible, preferably using our part numbers. If practical, enclose a sample photo or drawing.
Caution Notice
The BTX Gemini X2 & BTX Gemini SC systems are intended for laboratory use only and can be used in research and development applications. These systems have been designed to meet the standards for electromagnetic compatibility (EMC) intended for laboratory equipment applications.
The BTX Gemini X2 & BTX Gemini SC systems comply with the applicable sections of IEC 61010-1:2012 3rd Ed – Safety requirements for electrical equipment for measurement, control, and laboratory use. This product should not be used in the presence of a flammable atmosphere such as an anesthetic mixture with air, oxygen, or nitrous oxide.
Out of Warranty Service
Proceed exactly as for Warranty Service above. If our service department can assist you by phone or other correspondence, we
Gemini Series Electroporator User’s Manual
48
Loading...