BTG TCT-2301 User Manual

Enter
Sample
Total Consistency
Transmitter
TCT-2300
User Manual
TCT-2301
©BTG 2004
M2007/3en
Disclaimer
The contents of this document are subject to revision without notice due to continued progress in methodology, design, and manufacturing. BTG shall have no liability for any error or damages of any kind resulting from the use of this document. ©BTG 2004.
BTG Pulp & Paper Sensors AB
P.O. Box 602
SE-661 29 Säffle
Sweden
Phone: +46 533 426 00
www.btg-group.com
M2044/0en TCT-2302 Table of contents I
©BTG 2004
Table of contents
1 Product introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.2 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
1.3 Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3.1 The Peak Method. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.4 System description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
1.5 Typical applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
1.6 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
1.7 Type sign explanations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
1.8 CE-declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2 Safety recommendations . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.1 General safety recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.2 Product specific safety recommendations . . . . . . . . . . . . . . . . . . . . . . . . 14
2.2.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.2.2 Selecting a transmitter model and assembly parts . . . . . . . . . . . . . . . . 14
2.2.3 Installing mounting details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.2.4 Installing the transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.2.5 Installing the electronics box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.2.6 Starting up and servicing the transmitter . . . . . . . . . . . . . . . . . . . . . . . . 15
3 Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1 Before you start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
3.1.1 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.1.2 Installation and operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.2 Installing the transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2.1 Positioning the transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2.2 Probe clearance requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.2.3 Sampling valve. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.2.4 Installing the ball valve assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.2.5 Mounting the probe in the valve assembly. . . . . . . . . . . . . . . . . . . . . . . 24
3.3 Installing the electronics box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.3.1 Choosing a site for the electronics box . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.3.2 Electronics box design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.3.3 Connecting the measuring probe to the electronics box . . . . . . . . . . . . 29
II TCT-2302 Table of contents M2044/0en
©BTG 2004
3.3.4 External wiring dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.3.5 Main power supply connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.3.6 Transmitter input and output, measuring range connections . . . . . . . . . 34
3.3.7 Digital communication using the Hart®-protocol . . . . . . . . . . . . . . . . . . . 34
3.4 Probe cleaning system (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.4.1 Dimensions and installation clearance. . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.4.2 Installing the PCD-1000 upgrade kit . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
3.4.3 Installing the PCD-1000 probe cleaning device . . . . . . . . . . . . . . . . . . . 40
3.4.3.1 Probe cleaning device connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
3.4.4 Controlling the cleaning device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
3.5 Checklist before start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
4 Operating instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.1 Starting the transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.1.1 Adjusting TAG / Serial no. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.1.2 Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.1.3 Select measuring unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
4.1.4 Setting measuring span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
4.2 Clear water adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
4.2.1 At start up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
4.2.2 Selecting the correct ”mode” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
4.2.3 Adjustment of measuring gap angle. . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
4.3 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
4.3.1 General remarks on calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
4.3.1.1 Criteria for calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
4.3.2 General remarks on sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
4.3.3 Calibration procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
4.3.3.1 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
4.3.3.2 Take sample. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4.3.3.3 Load lab value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4.3.3.4 Calculate calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.3.4 Multi-point calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.3.5 Calibration with given constants. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.3.6 Calibration table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4.3.7 Calibration curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4.4 Other functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
4.4.1 Setting time constant (damping) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
4.4.2 Setting alarm limits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
4.4.3 Offset adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
4.4.4 Manual loading of probe signal settings. . . . . . . . . . . . . . . . . . . . . . . . . 59
M2044/0en TCT-2302 Table of contents III
©BTG 2004
4.5 Evaluation and documentation of calibration . . . . . . . . . . . . . . . . . . . . .60
4.5.1 Trend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
4.5.2 Documenting calibrations with printer . . . . . . . . . . . . . . . . . . . . . . . . . . 61
4.5.3 Documenting calibration using the ”back-up card” . . . . . . . . . . . . . . . . 64
4.5.4 Documenting calibration using a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
4.6 Electronics box display menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
4.6.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
4.6.2 Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
4.6.2.1 How to change settings and collect samples . . . . . . . . . . . . . . . . . . . . 70
4.7 Hand-held Terminal SPC-1000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
4.7.1 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4.7.2 Code protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
4.8 Digital communication according to the Hart®-protocol . . . . . . . . . . . . . .80
5 Service instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
5.1 Maintenance planning and quality assurance . . . . . . . . . . . . . . . . . . . . .81
5.1.1 Inspection of the accuracy/calibration of the consistency transmitter . . 81
5.1.2 Calibration recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
5.1.3 Maintenance of the transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
5.1.4 Functional description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
5.2 Service electronics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
5.2.1 Electronics troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
5.2.2 Electronics card LED functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
5.2.3 Function and testpoints of electronics card . . . . . . . . . . . . . . . . . . . . . . 87
5.2.4 Changing the electronics card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
5.2.5 Change of power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
5.2.6 Changing the front panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
5.2.7 Check analog out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
5.2.7.1 Calibrate analog out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
5.2.8 Checking communication to and from SPC-1000/A. . . . . . . . . . . . . . . . 93
5.2.9 Checking range inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
5.2.9.1 Change range input using digital input (hardware) . . . . . . . . . . . . . . . . 94
5.2.9.2 Change range using software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
5.2.10 Checking the alarm output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
5.3 Service probe and valve arrangement . . . . . . . . . . . . . . . . . . . . . . . . . . 96
5.3.1 Pull out the probe from the ball valve. . . . . . . . . . . . . . . . . . . . . . . . . . . 96
5.3.2 Check clear water value and clean of probe . . . . . . . . . . . . . . . . . . . . . 97
5.3.3 Change the probe with fiber connection. . . . . . . . . . . . . . . . . . . . . . . . . 99
5.3.4 Replacing ball valve seals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
5.3.5 Inserting the probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
IV TCT-2302 Table of contents M2044/0en
©BTG 2004
5.4 Service software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
5.4.1 Software revisions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
5.4.2 Troubleshooting functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
5.4.2.1 Alarm & diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
5.4.3 Software Fault feedback SPC-1000. . . . . . . . . . . . . . . . . . . . . . . . . . . 104
5.5 Evaluating calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
5.6 Troubleshooting, general . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .108
5.7 Maintenance of the cleaning system. . . . . . . . . . . . . . . . . . . . . . . . . . . 109
6 Parts list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
6.1 Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
6.2 Valve arrangement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113
6.3 TCT-2301 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
6.4 Cleaning device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
6.5 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .117
1.1 - General
M2007/3en TCT-2301 Product introduction 1
©BTG 2004
1 Product introduction
Total consistency transmitter
Measures total consistency, independent of varying content of filler and fines, with a unique measuring method – enables accurate consistency measurement
Virtually insensitive to flow variations and temperature changes – for high accuracy and reproducibility
No moving parts and non-clogging probe design – gives low maintenance costs
Simple calibration and documentation – provides easy start up and usage
Simple installation – for low installation costs
Probe can be retracted without shutting down the process
1.1 General
The TCT-2301 is an in-line transmitter for measuring total consistency of pulp suspensions in the range 0.01-3.5%.
Based on BTG's patented Peak Method (see section 1.3) of optical anal­ysis, the TCT-2301 is virtually insensitive to variations in filler and fines content, making it particularly suitable for use with recycled and me­chanical pulp suspensions.
The transmitter is mounted in-line, without any special bypass arrange­ment, for optimized functioning of the control loop. The instrument's unique probe, without electronic components, makes the transmitter in­sensitive to variations in temperature and vibration. The probe is con­nected to the wall-mounted electronics unit by an optical fiber.
An integrated fail-safe system means the probe can be installed and re­tracted safely without shutting down the process, even in high pressure applications.
The transmitter's non-clogging and self-cleaning features ensure trouble­free and accurate operation with easy maintenance.
1.2 - Technical data
2 TCT-2301 Product introduction M2007/3en
©BTG 2004
The TCT-2301 employs modern microprocessor technology, including fast data collection with a separate signal processor, digital data process­ing with advanced signal analysis.
The TCT-2301 is served by the SPC-1000 hand-held terminal, which works not only with the TCT-2301, but also with all instruments in the BTG family of smart transmitters.
The SPC-1000 can be connected at any point on the 4-20 mA output signal loop. Communication with a DCS is by means of a superimposed digitized signal according to a standard Hart® protocol. In addition, the unit is prepared for fieldbus digital communication.
BTG has adopted the OPC communication standard, which enables any OPC equipped system direct access to TCT-2301 transmitters.
1.2 Technical data
Enter
Sample
Terminal
Total Consistency
Transmitter
TCT-2300
Fig 1 Smart optical low
consistency transmitter TCT­2301 with SPC­1000 hand-held terminal
Typ e TCT-2301 in-line smart optical consistency
transmitter for pulp slurries. One analog output for total consistency.
Manufacturer BTG, Säffle, Sweden
Measuring principle Light scattering using the Peak-method.
Performed by light transmission of NIR, 880 nm
Measuring range Total consistency 0.01 to max. 3.5% (0.1 - 35 g/l)
depending on fiber type and filler content
Measuring accuracy ± 0.004%
Repeatability 0.0001%
1.2 - Technical data
M2007/3en TCT-2301 Product introduction 3
©BTG 2004
Process conditions
Process pressure PN 16 (232 psi at 68 °F)
Media temperature Max. 150ºC (302°F)
Max. ambient temperature
Probe: 150ºC (302 °F) Electronics box: 50ºC (122 °F)
Process pH 4-9
Storage temperature Min. - 20ºC (70°F)
Standardization
Standardization Quality-assured in accordance with ISO 9001.
Designed in accordance with relevant CE standards. See section 1.8: CE-declaration of conformity
Probe
Mounting Mounted through a ball valve assembly.
Material Stainless steel to SS 2343, equiv. to ANSI 316, DIN
X5 CrNiMo 18.22, B.S.316 S16/En 58J
Connection Optical connection to electronics box.
Weight Probe 3.3 kg (7.2 lb)
Valve assembly approx. 3.8 kg (8.4 lb)
1.2 - Technical data
4 TCT-2301 Product introduction M2007/3en
©BTG 2004
Electronics Box
Supply voltage Supply voltage 84 - 264 VAC, 50/60 Hz, max. 50 VA
- automatic setting. Output voltage 24 VDC, regulated. Output power max. 40 VA, max. constant power 1800 mA at an ambient temperature of 50 °C
(122°F). Approved according to UL, CSA, VDE.
Output signal - analog 4-20 mA. Isolated against mains supply / protective
earth. Built in insulation amplifier Current limited to 21 mA. Loop load signal: voltage supply / load
21.6 VDC / max. 1030 24 VDC / max. 1100
26.4 VDC / max. 1270 Min. (nom.) load for communication with hand-held terminal 250 Ω.
Input signal - analog 0 / 4-20 mA
Output signal - digital Superimposed signal over
4-20 mA current loop according to standard HART
®
protocol and BELL 202 modem . Follows HART
®
universal commands.
Power consumption Max 600 mA incl. hand-held terminal
(excl. 200 mA) at 24 VDC
Measuring ranges (analog)
Four separate, individually programmable, externally connectable with binary-coded switch. Also accessible via the communications link.
Input – digital (measuring ranges)
2 x 2 optically isolated contact inputs.
Alarm function (digital)
Optocoupler output. Provides alarm signal on low/ high LED curr./cons.level /internal temp.
Electronic Enclosure Made of polycarbonate thermoplastic with a transpar-
ent cover. Metal-plated internally to protect against interference.
Protection rating IP65, NEMA 4x, UL, CSA
Weight Electronics box 2.4 kg (5.3 lb)
1.2 - Technical data
M2007/3en TCT-2301 Product introduction 5
©BTG 2004
Communications
Option 1: Using the BTG SPC-1000 handheld terminal jacked
into the electronics box. The digital communication is superimposed over the 4-20 mA current loop.
Option 2: Directly with the DCS. See “Output signal - digital”.
Allows HART
®
universal commands.
Option 3: RS-485 for connection to OPC-server. OPC program
provided by BTG. Note: Only for DCS/OCS systems with OPC capabilities.
Option 4: Prepared for fieldbus communication e.g. Profibus or
Fieldbus Foundation.
Signal transmission 10 m (33 ft) quartz glass optic fibers included,
protected in a flexible conduit. On request 20 (66 ft), 30 (98 ft) and 40 m (131 ft).
Accessories
Hand-held terminal, SPC-1000
See data sheet D218.55
Smart Lab Sampler, SLS-1000
See data sheet D218.62
Documentation program for PC
Probe cleaning device: PCD-1000 See data sheet D2020.
PCD-1000 cleaning device (option)
Weight 9.5kg/20.9lb including valve and probe.
Dimensions See page 2.
Ambient temp. 50ºC/122°F
Air supply/conn. Filtered 4-7bar (58-102 psi).
Ø6/4mm (0.236/0.157)
Water conn. Mechanically cleaned, 4-10bar (58-145 psi).
Inlet Ø8,5mm (0.335in) Outlet Ø17mm (0.669in)
Supply voltage 24V DC supply from
TCT-2301 electronics box.
Min. 4 x 0.75
mm² (AWG18)
1.3 - Measuring principle
6 TCT-2301 Product introduction M2007/3en
©BTG 2004
1.3 Measuring principle
1.3.1 The Peak Method
The TCT-2301 employs the patented peak method for measuring total consistency of pulp suspensions. This technology is based on the fact that suspensions contain both large and small particles. Large particles are typically fibers and small particles are fillers and fines.
Close study of a suspension sample shows that the high number of small particles in the suspension remains relatively constant over time. On the other hand, there are far fewer large particles and their number varies sig­nificantly over time.
The large particles form a relatively transparent network, within which the small particles move freely. A narrow light beam directed through the suspension is generally affected by both the large and small particles. There is however an interval when only small particles affect the light beam, and this is know as the "Peak" period.
During the other time periods, the light beam is affected by both the large and small particles. Fig 2 and fig 3 illustrate this principle.
In fig 2, a light beam passes through the pulp flow (1). The transmitted light is detected by a photo detector. The detector signal is amplified (5) and then treated by the signal processor (6).
8
1
9
2
5
4
3
DC
PEAK
6
7
Fig 2 Operating principle
1 Pulp flow 2 Measuring gap with light beam 3 Optical fiber from light source 4 Optical fiber to detector 5 Amplifier 6 Digital signal processor 7 Microprocessor 8 Input signals 9 Output signals
2143
1.3 - Measuring principle
M2007/3en TCT-2301 Product introduction 7
©BTG 2004
Three signals are computed by the detector signal: The mean value, VDC, the peak value, VP and the AC-component, VAC. Typical time graph showing the detector signal is displayed in fig 3.
The total consistency is obtained by adding the further treated VP and the V
DC
values.
Fig 3 Output signal
V
CW
Clear water
V
P
Peak value
V
DC
Direct current signals
V
AC
Alternating current signals
TC Total Consistency
Light transmission Volts
TC
Time
1.4 - System description
8 TCT-2301 Product introduction M2007/3en
©BTG 2004
1.4 System description
The system consists of a probe (1) connected by an optical fiber to an electronics box (2) The probe contains only optical fibers. A ball valve arrangement that is welded to the pipe system make it possible to remove the probe without interrupting the process.
The electronics box contains electronics card, power supply, electrical and optical connections. To set up the transmitter a hand-held terminal (3) type SPC-1000 must first be connected (see fig 4).
The hand-held terminal allows the operator to set up and monitor the sys­tem. It includes a liquid crystal display (LCD) with a touchscreen key­pad.
1.5 Typical applications
Control of flotation systems or wash stages for de-inking of recycled fibers
Control of screening and cyclone cleaning processes
Monitoring of TMP/CTMP steam and hot-water recovery systems
CRC
1
3
5
2
8
4
9
7
6
Enter
Sample
Terminal
TotalConsistency
Transmitter
TCT-2300
Fig 4 Typical control loop for
consistency control of the TCT­2301
1 Measuring probe 2 Electronics box 3 Hand-held terminal SPC-1000 4 3/4 wire system for analog output 5 Controller/recorder (DCS) 6 Dilution water valve 7 Pulp chest with sufficient mixing 8 Stock pump 9 BTG sampling valve type MPS-1000
1.6 - Dimensions
M2007/3en TCT-2301 Product introduction 9
©BTG 2004
1.6 Dimensions
300 12"
469
18.5"
53
2.1"
121
4.8"
121
4.8"
Ø50
2”
251
9.9"
108
4.3"
Min Ø100
4"
Fig 5 TCT-2301 Dimensions
1 Electronics box 2 Probe with retraction
system
1
2
1.7 - Type sign explanations
10 TCT-2301 Product introduction M2007/3en
©BTG 2004
1.7 Type sign explanations
1. Manufacturer
2. CE-marking
The TCT-2301 is approved according to CE directives.
3. CSA-marking
The TCT-2301 electronics box is approved according to Australian C­TIC N1619 directives.
4. ETL-marking
The TCT-2301 electronics box is approved by ETL
5. Warning sign
The device is designed for industrial use. Installation, handling and ser­vice must only be carried out by trained and authorized personnel and ac­cording to relevant standards. Read the manual for detailed information and pay special attention to the warning signs!
BTG Pulp & Paper Sensors AB Box 602 661 29 Säffle SWEDEN
N1619
TYPE
www.btg-group.com
VOLTAGE
APPARENT POWER
NO
FREQUENCY
INSTALLATIONCATEGORY
FACTORYIDENTIFICATION
Fig 6 Type sign
1 Manufacturer 2CE-marking 3 C-TIC marking 4 ETL-marking 5 Warning sign 6 Type specification 7 Voltage 8 Apparent power
9 Installation category 10 Frequency 11 Manufacturing number 12 Factory Identification
1
234
6
7
8
5
91011 12
TCT-2301
100-240 VAC
50-60 Hz
50 W III
Always refer to the type sign when ordering spare parts.
1.7 - Type sign explanations
M2007/3en TCT-2301 Product introduction 11
©BTG 2004
6. Type specification
7. Voltage
Automatically adjustable between 100-240V AC.
8. Apparent power
According to CE directive. Max. continuous output.
9. Installation category
According to CE directive. Resistant to transients.
10.Frequency
The unit operates both 50 and 60 Hz.
11.Manufacturing number
BTG internal product identification number.
12.Factory identification
1.8 - CE-declaration of conformity
12 TCT-2301 Product introduction M2007/3en
©BTG 2004
1.8 CE-declaration of conformity
CE-Declaration of Conformity
According to EN 45014
Manufacturer´s Name
BTG Pulp & Paper Sensors AB
Manufacturer´s Address
P.O. Box 602 S- 661 29 SÄFFLE, Sweden
declares that the product:
Product Name
Total Consistency Transmitter
Model Number
TCT-2300/2301/2302/with FCI-1000 complies with the amendments and requirements of the: Machinery Directive 89/392/EEC
Low Voltage Directive 73/23/EEC
EMC Directive 89/336/EEC
PED Directive 97/23/EEC
and conforms with the following product standards and PED conformity assessment procedure
Safety
EN 292/1-2
LVD
EN 61010-1
PED Pressure equipment for piping.
In accordance to Article 1,
Table 9, art.3 (3), Annex II
(Sound eng. practise)
Must not bear CE-marking
EMC
EN 50081-2 Aug.93
EN 61000-6-2 Apr.02
Quality System
ISO 9001
monitored by
Lloyd`s Register Quality Assurance
Säffle Mars 2002
Tom Gustavsson, MD
2.1 - General safety recommendations
M2007/3en TCT-2301 Safety recommendations 13
©BTG 2004
2 Safety recommendations
2.1 General safety recommendations
Installation, handling and service must only be carried out by trained and authorized personnel and according to valid standards.
The product is designed for industrial use.
Installation category: III
Shock protection: Class I
Pollution degree: 2
IP Code: IP65 / NEMA 4X
The product complies with the EEC directives and relevant standards:
To meet the EMC directive the following precautions have to be taken:
1. All wiring has to be shielded.
Power supply electronics: shielded, coverage≥ 80% Interface cable: coverage 100% aluminum
2. All units are tested as a complete system to conform with rele­vant CE directives and their standards.
When using the unit in other combinations, BTG cannot guarantee the CE directive conformity.
The unit in combination with customer installed external devices may conform with EMC and safety requirements when properly installed and using an adequate CE marked equipment.
The system operator is responsible for the CE directive conformi­ty. Conformity must be checked by inspection.
Take precautions when handling equipment in pressurized/hot lines.
See section 1.2: Technical data for detailed
information.
2.2 - Product specific safety recommendations
14 TCT-2301 Safety recommendations M2007/3en
©BTG 2004
2.2 Product specific safety recommendations
These safety recommendations are based on a risk analysis carried out in accordance with the requirements of the machinery and low voltage di­rective in order to comply with European standards for CE marking.
2.2.1 General
The possibility of risk from a consistency transmitter in operation only arises when covers have been removed during installation and servicing, due to hazardous electrical voltage.
Read these safety recommendations before installing the transmitter. Follow the recommendations when installing the transmitter, starting up and servicing. Use warning signs for safety information!
These safety recommendations apply to the transmitter fitted with an AC powered electronics box. Mounting parts, such as the weld-in studs, are dealt with in accordance with the pressure vessel standards of the respec­tive countries.
! For good personal and functional safety: Use only parts which
have been manufactured or approved by BTG.
2.2.2 Selecting a transmitter model and assembly parts
Pressurized parts must be suited to the current maximum pipe pressure in relation to temperature. Select a material for the parts that are in contact with the medium so that corrosion does not occur.
2.2.3 Installing mounting details
Welding and subsequent inspection should take place in accordance with current standards and regulations.
2.2.4 Installing the transmitter
Construct a platform or use safe and approved ladders if the
transmitter is located high up. This platform will make it easier to fit and start-up the transmitter and carry out service in the future.
The rubber quality of the transmitter seal (O-ring) should be selected
to suit the current medium. The O-ring supplied is suitable for this application.
Screw the valve assembly firmly onto the weld-in stud.
Be sure that the pipe is empty before opening it!
2.2 - Product specific safety recommendations
M2007/3en TCT-2301 Safety recommendations 15
©BTG 2004
Install the probe into the valve assembly. Fit the two screws locking the probe in place. Check that these screws are properly tightened, as well as the four screws holding the valve together. See section 3.2.5: Mounting the probe in the valve assembly for details.
When the probe has been installed, carry out test pressurization using water in the pipeline. The correct test pressure depends on standards and regulations in each country. In some lines for which inspection is required, a pressure test must be carried out before the product can be commissioned.
2.2.5 Installing the electronics box
A hazardous voltage is used to drive the electronics box. The electronics box may only be connected by a qualified electrician in accordance with instructions. The earth ground should be connected correctly and checked.
It should be possible to switch off the voltage in the distribution centre, for example. A clear warning, such as a sign, should let other people know that work is in progress and the switch must not be touched.
Use only metallic cable fittings with shield connection for the electronics box.
2.2.6 Starting up and servicing the transmitter
Read through the following points before starting up and servicing the transmitter. Bear also in mind that the points mentioned above contain important information for these operations.
Before removing the probe from the valve assembly, check carefully that the valve is closed. Hot or corrosive liquid flowing out under
pressure may cause serious chemical burn injuries!
Take care when opening the cover of the electronics box with built­in power supply unit. It contains live parts that may cause electric shocks. Live parts are protected against normal contact provided that connections are made correctly.
When the probe is exposed to dangerous basic or acidic corrosive media, it should be removed from the pipeline regularly for inspection. Replace any damaged seals. If the probe’s pressurized parts have corroded, check that the material is correct for the application. Leakage may cause personal injury or damage to
equipment due to corrosion or burning!
Check that the screws/nut holding the valve/probe together are mounted and properly tightened.
DANGER! High voltage inside the electronics box. Connections may only be carried out by qualified personnel.
To prevent electrical shock — connect the protective earth PE properly.
2.2 - Product specific safety recommendations
16 TCT-2301 Safety recommendations M2007/3en
©BTG 2004
3.1 - Before you start
M2007/3en TCT-2301 Installation instructions 17
©BTG 2004
3 Installation instructions
3.1 Before you start
This section contains instructions for planning and implementing the in­stallation of the BTG TCT-2301 in-line consistency transmitter.
If you are in doubt whether the model you plan to install is the same in all respects as the model described in this manual, or you have any questions about installation, please contact your BTG sales engineer.
When you are satisfied that your TCT-2301 has been correctly installed and are ready to power up the system for calibration, please turn to sec­tion 3.5: Checklist before start.
Double check items in the list before powering up. This list can help en­sure a trouble-free initialization of your system.
This section introduces the essential components of the system.
This introductory section is followed by:
section 3.2: Installing the transmitter
This section contains detailed instructions for installing the trans­mitter.
section 3.3: Installing the electronics box
This section describes how to connect mains power supply, inputs and outputs.
section 3.4: Probe cleaning system (option)
This section describes how to install the PCD-1000.
section 3.5: Checklist before start
When you are satisfied that your TCT-2301 has been correctly in­stalled and are ready to power up the system for calibration, please turn to this section.
3.1 - Before you start
18 TCT-2301 Installation instructions M2007/3en
©BTG 2004
3.1.1 Visual inspection
This product was inspected and tested prior to shipment. However, even the best products can sustain shipping damage that will only be seen if the product is inspected. Before proceeding check the probe assembly, the hand-held terminal, and the electronics box for shipping damage. Look for loose screws, wires or electronic components.
3.1.2 Installation and operating conditions
The Consistency Transmitter is a precision tool and must be properly in­stalled to ensure reliable service. Please, read the following instructions and user information before installing the transmitter.
The transmitter is manufactured to provide accurate and reliable mea­surements over a long period of time.
! Correct installation will ensure maximum operating life of the
transmitter.
For safety precautions please refer to section 2. Product type sign explanations see section 1.7
3.2 - Installing the transmitter
M2007/3en TCT-2301 Installation instructions 19
©BTG 2004
3.2 Installing the transmitter
3.2.1 Positioning the transmitter
To make an accurate and reliable measurement it is critical to find the correct installation point.
Fig 4 on page 8 shows a typical installation. Your BTG sales engineer can help you select the location that will give the best results in relation to your control strategy.
! Minimum recommended flow velocity: 1 m/s (3.3 ft/s).
! To minimize time lag, locate the probe close to the dilution point.
Warning: To avoid personnel injuries, burning injuries, aggressive chemical injuries etc., it is extremely important that the pipe is depressurized and empty before opening it for installation or removal of the equipment.
Fig 7 Recommended
distance between pump and consistency transmitter. Can also be placed in the cone after the pump. Turbulent flow will give the best measuring conditions for this type of transmitter.
0.1m
(4 in)
Fig 8 Installation of the
probe in relation to different pumps
3.2 - Installing the transmitter
20 TCT-2301 Installation instructions M2007/3en
©BTG 2004
Important recommendations:
1. Do not install the probe where there is trapped air or where air may be sucked into the process.
2. If there is risk of flocculation in the pulp suspension, install the probe where the flow rates exceeds 1 m/s (3.3 ft/sec). This applies particularly to long-fiber pulp suspensions.
3. Install the probe so that it is protected from direct mechanical dam­age.
4. Protect the probe from heavy vibration such as cavitation or unbal­anced pumps. One method is to install a rubber bellows in the line.
ABC
90°
Flow
Direction
TCT­2300
Fig 9 Installation of the
probe near a pipe elbow
! Alternative A is the
most suitable. Avoid alternative C.
3.2 - Installing the transmitter
M2007/3en TCT-2301 Installation instructions 21
©BTG 2004
If the probe is installed in a horizontal pipe, always install it at the side of the pipe rather than at the top or bottom to avoid problems with trapped air or dirt. Remember the clearance requirements, so that the probe can be mounted in the ball valve arrangement.
5. The ball valve assembly should be mounted at a 90° angle to the pipe line.
6. The ball valve assembly should be mounted on the side of the pipe line. A small angle downwards makes it possible for the pulp to evac­uate when the probe is retracted and also makes the insertion of the probe easier. See fig 10 below.
Fig 10 Installation of the
probe in a horizontal pipe.
3.2 - Installing the transmitter
22 TCT-2301 Installation instructions M2007/3en
©BTG 2004
3.2.2 Probe clearance requirements
The installed length of the probe, inserted into the ball valve arrange­ment, is approximately 470 mm (18.5 in) long. You will also need a min­imum distance of 300 mm (11.8 in) to remove the probe from the pulp line. Leave sufficient room around the probe to perform this task.
The probe weighs approximately 3.3 kg (7.2 lbs).
! Remember to provide access room for an operator, as well as the
full depth of the probe assembly and the electronics box.
3.2.3 Sampling valve
The sampling valve should be installed close to the transmitter. This helps to ensure that the laboratory sample is representative of the sample measured by the transmitter.
Installing an appropriate BTG sampling valve, will ensure reliable re­sults.
! Install the sampling valve up or downstream of the transmitter but
not closer than 0.5 m (1.5 ft) to the transmitter to avoid disturbances.
300 12"
469
18.5"
53
2.1"
121
4.8"
121
4.8"
Ø50
2”
251
9.9"
108
4.3"
Min Ø100
4"
Fig 11 Probe. Dimensions in
mm/in
3.2 - Installing the transmitter
M2007/3en TCT-2301 Installation instructions 23
©BTG 2004
3.2.4 Installing the ball valve assembly
1. Mark the position on the pipe wall where the probe is to be located.
2. Drill a 50 mm / 2” inch hole in the pipe at the marked position.
3. Adapt the welded stud profile to the pipe. Adjust length according to fig 12.
4. Make sure the weld-in stud is aligned to the pipe. See fig 12.
5. Weld the stud to the pipe. A full penetration weld must be used.
! If the probe is delivered with a PCD-1000, refer to section 3.4: Probe
cleaning system (option).
6. Mount the ball valve arrangement. The slot for the crank handle must be located upwards since there is a scale on the probe that must be adjusted before the probe is mounted.
Fig 12 Mounting the flanged
weld-in stud
Fig 13 Mounting positions of
the probe.
3.2 - Installing the transmitter
24 TCT-2301 Installation instructions M2007/3en
©BTG 2004
3.2.5 Mounting the probe in the valve assembly
1. Make sure the measuring gap on the probe is at a 45° angle to the pulp flow. The angle may have to be changed after start-up.
If the angle is not correct, loosen the three screws that hold the probe flange and adjust the probe so that the punch mark on the probe is at the 45° line on the scale. Tighten the three screws properly. A suit­able allen key is included in the delivery.
! Angles other than 45° can be used in some installations. See
section 4.2.3: Adjustment of measuring gap angle.
45°
2
2
1
45°
2d
2c
1
90°
45°
45°
T
C
T
-
2
3
0
0
2b
2a
1
90°
45°
45°
T
C
T
-
2
3
0
0
Fig 14 Measuring gap angle
adjustment
1Punch mark 2 Pulp flow direction
Fig 15 Probe scale
adjustment
3.2 - Installing the transmitter
M2007/3en TCT-2301 Installation instructions 25
©BTG 2004
2. Apply a small amount of o-ring lubricant to the o-ring in the ball valve housing.
3. Make sure the o-ring on the probe flange is mounted.
4. Insert the probe, using a twisting motion to get the probe tip to enter the o-ring. Place the flexible insertion screw in the correct position and make sure the hook with its locking device is in the correct po­sition and secured.
5. Open the ball valve and insert the probe by cranking the handle.
6. Secure the probe in inserted position with the two security screws
! Secure the two security screws properly
7. When removing the probe, reverse the above procedures.
Fig 16 Mounting the probe in
the valve assembly
Avoid getting O-ring lubricant on the measuring gap
3.3 - Installing the electronics box
26 TCT-2301 Installation instructions M2007/3en
©BTG 2004
3.3 Installing the electronics box
3.3.1 Choosing a site for the electronics box
! Locate the electronics box close to the transmitter/ sampling valve
to achieve a correct and convenient connection of the SPC-1000 hand-held terminal for calibration and monitoring.
The hand terminal SPC-1000 can be positioned in a holder, located close to the electronics box. See fig 20.
Install the box in a position where it will be protected from mechanical damage.
Fig 17 Dimensions of the
electronics box
Fig 18 Electronics box
clearances
50 2"
300 12"
50
2"
200
8"
Enter
Sample
Terminal
3.3 - Installing the electronics box
M2007/3en TCT-2301 Installation instructions 27
©BTG 2004
BTG recommends that a roof or overhang is located above the electronics box to protect it against spray and pulp. If the box is located outdoors a roof will protect the unit from direct sunlight which can cause excessive operating temperatures. If possible, the SPC-1000 mounted in its holder should also be protected.
The electronics box is equipped with three attachment lugs and should be bolted to a wall.
200
8"
300
12"
75°
Fig 19 Protecting roof for
Electronics box
Fig 20 Wall mounted
brackets positioning the SPC-1000, for easier operation
Fig 21 SPC-1000 positioned
in the brackets
3.3 - Installing the electronics box
28 TCT-2301 Installation instructions M2007/3en
©BTG 2004
3.3.2 Electronics box design
! Electronics boxes manufactured by BTG are designed to meet
personal safety and interference requirements under the European CE directive and Canadian CSA requirements.
The plastic enclosures are made of polycarbonate thermoplastic. They are metal-plated internally to protect against interference.
Metal screw caps for attaching connection cable shield are an im­portant part of the interference suppression system.
The electronics box is supplied with an external jack connection for the hand terminal, SPC-1000/A, and the Smart Lab Sampler, SLS-1000.
Terminal
Total Consistency
Transmitter
TCT-2300
1.30% 12.53mA
Sample
Enter
+ALARM-
F3
F2
F1
+24
GND
+AOUT1- +AOUT2-
IN1
IN2
GND
+AIN-
NLPE
ALARM
MODE
PWR OK
DO1
DO2
Di1
IN1
IN2
GND
AB
COM+DO1DO2 DI1
GND
AB
6
8
Fig
22
Electronics box
1Display 2 Operating buttons 3 SPC-1000/A
connection 4LEDs 5 Power switch 6 Fuses 7 Electrical connections 8 Optical fiber connection
3
4
5
2
1
7
3.3 - Installing the electronics box
M2007/3en TCT-2301 Installation instructions 29
©BTG 2004
3.3.3 Connecting the measuring probe to the electronics box
The optical connection is on the lower right side of the electronics box. Remove the protective cover on the electronics box connector, and on the optical connector on the probe. Note the position for the optical connec­tor in fig 23. The connector fits only in one position. Make sure the O­ring is mounted on the probe connector. Push the connector all the way in. Try to turn the connector (this should not be possible) to verify that it is in the end position. Secure the connection properly with the plastic nut. Coil the extra optical fiber and secure it under the electronics box in order to avoid the risk of mechanical damage.
!
Make sure the optical connector from the probe is positioned correctly and pushed all the way in before securing it with the plastic nut.
! The optical fiber must not be coiled with a radius smaller than
50mm (2 in) Locate it so that is protected and not subject to risk of mechanical damage.
The standard optical fiber is 10 m (33 ft). It is possible to use optical fiber up to 40 m (131 ft) long. Contact your BTG sales engineer for more in­formation regarding optional fiber lengths.
!
Do not cut the optical fiber.
Fig 23 Optical connection
1 Aluminum washer 2 Electronics box 3 Rubber sealing 4 Electronics box optical
connector 5 O-ring 6 Plastic nut 7 Probe optical connector 8 Optical fiber to probe
3.3 - Installing the electronics box
30 TCT-2301 Installation instructions M2007/3en
©BTG 2004
3.3.4 External wiring dimensions
BTG recommends properly dimensioned wire, shielded and twisted in pairs, for connection between the electronics box and external equip­ment. See also section 2.1: General safety recommendations.
Power supply cable:
Electronics box with built-in power supply unit:
• Shielded (80%)
• 3 x 0.75 mm2 (3 x AWG18) with PE.
• Cable diameter: 7,5-13 mm
• In accordance with the IEC 227/245 standard
Output signal cables:
• Shielded (80%), twisted pair
• Min. 2 x 0.3 mm2 (2 x AWG24).
The size typically used is 2 x 0.75 mm2 (2 x AWG18)
• Cable diameter: 6-11 mm
Range select cable:
• Shielded (80%), twisted pair
• Min. 2 x 0.3 + 2 x 0.3 mm2 (2 x AWG24 + 2 x AWG24)
• Cable diameter: 6-11 mm
Alarm/AUX-in cable:
• Shielded (80%), twisted pair
• Min. 2 x 0.3 mm
2
(2 x AWG24)
• Cable diameter: 6-11 mm
!
Signal cables can be combined to one or two cables.
3.3 - Installing the electronics box
M2007/3en TCT-2301 Installation instructions 31
©BTG 2004
3.3.5 Main power supply connections
Important: Before installation, ensure that:
1. All power to the system has been turned off.
2. The 2AT fuses on the main card, marked F2 and F3, are in­stalled.
The conduit for the power supply to the electronics box should be installed in the leftmost hole.
Wiring the main power supply screw terminal:
The earth wire must be connected to the terminal indicated by the protective earth symbol PE.
See fig 26 on page 32.
3. Strip the cable covering (insulation) to expose the shortest pos­sible length of the connection wires.
! Unshielded long, looped wires may cause interference! This can
interfere with the signal!
4. Once the wires are inserted, tighten the screws to secure them. Mount the connector in its position.
5. To ensure interference-free operation a shielded cable should be used. Make sure the shield is correctly connected - fig 24.
6. Turn on the main power supply.
DANGER! High voltage within the electronics box. Connections may only be carried out by qualified personnel
To prevent electrical shock
- connect the protective earth properly PE
Fig 24 Connection of cable
shield, in special cable screw cap (metal)
1 Connection cable for power
supply, measuring range
etc. 2Nut 3 Sealing ring 4 Insert ring 5 Cable shield 6 Socket 7 Free wires
! Tighten nut fully for
maximum contact for the shield.
1
3
6
7
2
5
4
3.3 - Installing the electronics box
32 TCT-2301 Installation instructions M2007/3en
©BTG 2004
7. Switch on the main power on the electronics card. The power LED should now light up. See fig 25. If not, check first the con­nections and then the 2 AT fuse for the power supply located in­side the electronics box. Check the 24 V DC at terminal +24, GND - fig 26.
Notes on fig 26:
An external 2-pole switch close to the electronics box is required.
Standard output signal 4-20 mA with superimposed digital signal according to Hart®-protocol. See section 3.3.7.
Four different measuring ranges (RANGE 1-4) are available. The setting of the respective measuring range is arranged via a binary code switch (range setting - see fig 26 switch A/B) or DCS.
The unit is prepared for fieldbus. To access the communication an additional circuit card is required.
Terminal
Total Consistency
Transmitter
TCT-2300
Sample
Enter
+ALARM-
F3
F2
F1
+24
GND
+AOUT1- +AOUT2-
IN1
IN2
GND
+AIN-
NLPE
ALARM
MODE
PWR OK
DO1
DO2
Di1
IN1
IN2
GND
AB
COM+DO1 DO2DI1
GND
AB
Fig 25 Power connections –
electronics box
1 Fuses 2 Power connection
100-240 V AC 3 Power switch 4 24 VDC out
4-20mA output signals
Range setting
0-20 mA input signals
Alarm 5LEDs 6 Digital signals 7 Fieldbus 8RS-485 9 Optical fiber connection
10 SPC-1000/A
connection
523
1104 6 7 89
OUT 1 4-20mA min 250ohm
TCT-2301
S = Shield connected to cable-gland
* = Not used for TCT-2301
PE
L
N
+24V
Range A
GND + - + - IN1
Range B
GND
Line power 100-240 VAC
RS485
IN2 GND
AOUT 1 A OUT 2
+-+-
AIN ALARM
Power to
external
equipment
A Out 1
A Out 2 *
Aux In
Alarm
COM
+
DO1 DO2 DI1
S
S
AB
GND
Field bus
AB
PCD-1000 option
PCD-1000 option
AB RANGE 1 0 0 RANGE 2 1 0 RANGE 3 0 1 RANGE 4 1 1
0=
1=
Fig 26 Wiring diagram for
electronics box TCT-2301
3.3 - Installing the electronics box
M2007/3en TCT-2301 Installation instructions 33
©BTG 2004
Explanation for fig 26 TCT-2301 wiring diagrams.
! To enable communication using the Hart
®
-protocol, and use of a hand-held terminal, a 250 min. circuit resistance is required. A 250 resistor is connected at terminals A OUT+ and A OUT- at the factory. Remove the resistor when connecting the output signal.
Terminal Function
N Neutral
L Phase 85-264 V AC, 47-440 Hz, power fuse 2AT
PE,
Protective earth
+24 +24V DC (to optional PCD-1000)
GND Ground, 0V DC (to optional PCD-1000)
AOUT 1+ Output signal 4-20mA
AOUT 1- Output signal 4-20mA
AOUT 2+ Not used
AOUT 2- Not used
IN1 Range A for external range setting
IN2 Range B for external range setting
GND Ground, Common for range A and B
AIN+ Input signal 0-20mA (from external equipment)
AIN- Input signal 0-20mA (from external equipment)
ALARM+ Alarm from TCT
ALARM- Alarm from TCT
COM + Common (to optional PCD-1000)
DO 1 Digital output signal 1(to optional PCD-1000)
DO 2 Digital output signal 2
DI 1 Digital input signal 1
RS-485 GND Ground
RS-485 A Serial connection
RS-485 B Serial connection
Fieldbus GND Ground
Fieldbus A Fieldbus connection
Fieldbus B Fieldbus connection
3.3 - Installing the electronics box
34 TCT-2301 Installation instructions M2007/3en
©BTG 2004
3.3.6 Transmitter input and output, measuring range connections
Important: Before installation, ensure that:
1. The mains power supply is connected according to section
3.3.5.
2. The mains power is turned OFF.
3. To ensure interference-free operation, the cable fittings should be adapted for shielded cables. See fig 24 on page 31.
Wiring the signal screw terminals:
The screw terminals are in the lower section of the connection terminal fig 25 on page 32.
4. Strip the cable covering so the shield is exposed. Use the shortest possible length of the connection wires.
! Unshielded long, looped wires may cause signal interference! Strip
only enough insulation from the wires to insert them. The wires are connected according to fig 26 on page 32. Once the wires are inserted, tighten the screws to secure them.
5. Tighten the nut fully to ensure a good contact for the shield.
6. Turn on the power and test the functionality.
3.3.7 Digital communication using the Hart®­protocol
Certain functions are available as a digital signal under the Hart®-proto­col and BELL 202 modem standard.
! The transmitter complies to Hart
®
-protocol standard requirements. A BTG handheld terminal is used to set and calibrate the transmitter. As this unit provides additional functions not covered by the Hart
®
-protocol, a standard Hart®competitive terminal offers
only limited functionality.
Communication takes place as a superimposed signal over the 4-20 mA output signal loop.
! For details, please contact the Hart
®
Communication Foundation or
BTG.
To ensure disturbance-free operation it is essential that cable shields are correctly connected in the cable screw caps - see fig
24.
3.4 - Probe cleaning system (option)
M2007/3en TCT-2301 Installation instructions 35
©BTG 2004
3.4 Probe cleaning system (option)
This section describes how to install the PCD-1000 probe cleaning de­vice. The system can be delivered as a separate unit for rebuilding exist­ing units or together with a complete TCT-2301 transmitter. If the cleaning system is ordered with the TCT-2301 the probe will be mounted in the cleaning device at the BTG factory. The installation method is de­scribed for both cases.
3.4.1 Dimensions and installation clearance
Ensure that there is sufficient clearance to remove the probe and cleaning device from the valve assembly.
The installed device is approx. 700 mm (28 in) long. You will need at least 300 mm (12 in) to remove the device from the valve arrangement. The cleaning device weighs approx. 9.5 kg (21 lbs) including the TCT­2301 probe.
Fig 27 PCD-1000 probe
cleaning device
140
5,5"
~550
22"
Min 100
4"
700
28"
Min 300
12"
92
3,6”
133
5,3"
Fig 28 PCD-1000 dimensions
3.4 - Probe cleaning system (option)
36 TCT-2301 Installation instructions M2007/3en
©BTG 2004
3.4.2 Installing the PCD-1000 upgrade kit
This section describes how to upgrade an existing TCT-2301 transmitter with the probe cleaning device PCD-1000/A (accessory).
! Make sure the pipe is empty and pressureless before starting the
installation.
1. Remove the TCT-2301 probe from the existing valve arrangement, note the measuring angle, that is, the measuring gap angle in relation to the pulp flow.
2. Disconnect the optical connection from the TCT-2301 electronics box. Protect the optical connection with a plastic tube or similar.
3. Remove the handle and the flange with the angle scale.
4. Place the cleaning device on a workbench with the valve facing downwards.
! To keep all O-rings in position during disassembly/assembly,
always place the cleaning device in a upright position with the valve facing downwards.
5. Unscrew and remove the screw (2) and the nut (1).
6. Unscrew and remove the nuts (4) and the washers (5).
7. Loosen the air tube (9) from the nipple and remove the rear cylinder flanged end (6) together with the protective sleeve (3). Keep track of the O-rings (7 and 8).
2120
Fig 29 Sequence of installing
the PCD-1000 upgrade kit
23
5
4
6
7
8
9
1
3.4 - Probe cleaning system (option)
M2007/3en TCT-2301 Installation instructions 37
©BTG 2004
8. Remove cylinder flanged end (12) and pull out piston with guiding rod (13). Keep track of O-ring (11) and guide ring (10).
Alternatively, pull out the cylinder flanged end and piston in one piece and separate them afterwards.
9. Remove the rubber bushing (14) from the guiding rod.
10. Loosen the screws (16) and separate the piston (15) and guiding rod (17) from each other.
11. Insert the probe (18) in the piston (15) and the fiber cable in the guid­ing rod (17).
12. Make sure that the screw hole on the edge of the guiding rod is aligned with the measuring gap on the probe. Then join the piston and the guiding rod with the screws (16). Do not forget the O-rings (19).
13. Put the accompanying rubber bushing (14) over the fiber cable and push it into the tip of the guiding rod.
12
10
11
13 14
1615 17
18
19
19 17
3.4 - Probe cleaning system (option)
38 TCT-2301 Installation instructions M2007/3en
©BTG 2004
14. Push the probe and piston into the cylinder. Make sure the guide ring (10) is in position.
15. Mount the cylinder flanged end (12), the rear cylinder flanged end (6), and the protective sleeve (3), to the cylinder. Make sure that the O-rings (7, 8, and 11) are in their proper positions, and that the nipple (21) on the rear cylinder flanged end are posi­tioned on the same side of the cylinder as the nipple (20) on the front cylinder flanged end. The parts must be pulled over the optic cable.
16. Mount the nuts (4) and washers (5), but do not tighten them fully.
17. Connect the air tube (9) to the nipple (21) on the rear cylinder flanged end.
18. Loosen the screws (24) and turn the protective sleeve until the screw hole on the guiding rod is visible through the slot (25).
2120
24 25
22
23
26
3.4 - Probe cleaning system (option)
M2007/3en TCT-2301 Installation instructions 39
©BTG 2004
19. Connect the protective sleeve to the guiding rod with the screw (1) and nut (2). The slot on the protective sleeve is now aligned with the measuring gap.
20. Turn the probe, by turning the protective sleeve, until the measuring gap is pointing 45° towards the pulp flow. Use the angle scale on the rear cylinder flanged end for assistance.
21. Tighten the screws (24) to fasten the protective sleeve and to fix the angle of the probe.
22. Loosen the screws (23).
23. Turn the flushing valve (26) so that the inlet faces the measuring gap of the probe. The air cylinder shall at this stage have its quadratic ex­ternal cross section turned the same way as the flanged ends of the ball valve.
24. Tighten the screws (23) and the nuts (5).
25. If the existing TCT-2301 is mounted with a threaded weld-in stud be­tween the ball valve assembly and pipe, the new flanged weld-in stud (22) must be used. Weld the new flanged weld-in stud to the pipe in a new position ac­cording to section 3.2.4. Plug the old weld-in stud with a blind flange (accessory).
26. If the existing TCT-2301 is mounted with a flanged weld-in stud, re­move the delivered weld-in stud from the cleaning device and mount the PCD-1000 to the existing flanged weld-in stud.
If the transmitter is submitted to vibration we recommend using rein­forcement. See fig 28 on page 35.
Fig 30 Mounting the weld-in
stud
The threaded weld-in stud is not dimensioned to hold the PCD-1000
3.4 - Probe cleaning system (option)
40 TCT-2301 Installation instructions M2007/3en
©BTG 2004
3.4.3 Installing the PCD-1000 probe cleaning device
This section describes how to install the probe cleaning device PCD-1000/A that is delivered with a TCT-2301 probe. (Option)
! Make sure the pipe is empty and pressureless before starting the
installation
1. Loosen the screws (3) that holds the protective sleeve (4) fixed.
2. Loosen the nuts (2).
3. Turn the probe, by turning the protective sleeve, until the measuring gap is at a 45° angle to the pulp flow. The slot (3) on the protective sleeve is aligned with the measuring gap. Use the angle scale on the rear cylinder flanged end for assistance.
4. Tighten the screws (3) to fasten the protective sleeve (4) and to fix the angle of the probe.
5. Loosen the screws (1).
6. Turn the flushing valve (6) so that the inlet faces the measuring gap of the probe. The air cylinder shall at this stage have its quadratic ex­ternal cross section turned the same way as the flanged ends of the ball valve.
7. Tighten the screws (1) and nuts (2).
8. Remove the flanged weld-in stud from the cleaning device and weld the flanged weld-in stud to the pipe, according to section 3.2.4.
9. Remount the PCD-1000 to the flanged weld-in stud.
435
2
1
6
Fig 31 Mounting the weld in
stud
3.4 - Probe cleaning system (option)
M2007/3en TCT-2301 Installation instructions 41
©BTG 2004
If the transmitter is submitted to vibration we recommend using rein­forcement. See fig 28 on page 35.
3.4.3.1 Probe cleaning device connections
Three connections are needed to make the cleaning device operational.
Connection requirement:
Water: Use standard water quality.
Air: 4-8 bar (58-100psi) or at least half the process pressure.
1. Connect water to the membrane valve. The connection is 8.5mm (0.33 in)
2. Connect air to the solenoid valve by 6/4 (1/4 in) hose.
3. Connect the control signal and 24V DC supply from the TCT-2301 electronics box. See section 3.3.5: Main power supply connections.
3.4.4 Controlling the cleaning device
All settings to control the cleaning device are made from the TCT-2301 electronics box. Instructions can be found in section 4.6.2.1: How to change settings and collect samples.
Fig 32 Cleaning device
connections
Water inlet
Water outlet
Control signals and 24 V DC from TCT-2301 electronics unit
POWER
VALVE
VALVE
VALVE
DO1
+COM
GND
+24V
Air connection
3.5 - Checklist before start
42 TCT-2301 Installation instructions M2007/3en
©BTG 2004
3.5 Checklist before start
When you are confident that your TCT-2301 has been correctly installed, and are ready to power up the system for calibration and/or testing, run through this brief check list before powering up. Using this list will help ensure trouble-free initialization of your system.
1. Make a clear water adjustment before installing the probe in the ball-valve arrangement. See TCT-2301 operating instructions, section 4.2.
2. Check that the measuring gap on the probe is at the correct angle to the pulp flow.
3. Check that the transmitter is fully inserted and the two security screws are properly tightened.
4. Ensure that the optical fiber is protected from mechanical dam­age, and attached securely to the transmitter.
5. Check all wiring.
6. Switch on the mains power supply.
7. Connect the hand-held terminal to the electronics box. Switch it on and check that the display lights up.
Proceed to the TCT-2301 Operation instructions.
4.1 - Starting the transmitter
M2007/3en TCT-2301 Operating instructions 43
©BTG 2004
4 Operating instructions
4.1 Starting the transmitter
All settings are made from the hand-terminal, SPC-1000, or from the electronics box.
This chapter describes the handling of the transmitter and its possibilities
All functions can be accessed from the hand-terminal, while the display on the electronics box only provides certain functions, see description in section 4.6: Electronics box display menu.
Before getting started, the transmitter must be installed in the pulp line, the electronics box wired electrically and connected, and the power switched on according to section 3: Installation instructions.
The hand-terminal, SPC-1000, is very easy to handle. It is equipped with a touch-screen with buttons to make selections and settings. See section
4.7: Hand-held Terminal SPC-1000 for further information.
Connect the SPC-1000 to the front of the electronics box. Switch on the hand-terminal.
Select from the menu.
Before making any changes, always read all transmitter data .
Should a failure message appear on the screen, see section 5.4.3: Soft- ware Fault feedback SPC-1000 for action, if necessary.
All changes made must be downloaded to the transmitter to take effect.
Use . This function can be password protected. See section
4.7.2: Code protection.
Enter
Sample
Terminal
TotalConsistency
Transmitter
TCT-2300
Fig 33 TCT-2301 consistency transmitter
with electronics box and hand terminal
TCT
2301
v1.x
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Sensor: Select program below.
Backup: First select program below and then restart with backup card.
TCT
2300
v1.x
TCT
2300
v1.x
TCT
2300
v1.x
TCT
2300
v1.x
TCT
2300
v1.x
TCT
2300
v1.x
TCT
2300
v1.x
TCT
2300
v1.x
TCT
2301
v1.x
Read data
Send data
4.1 - Starting the transmitter
44 TCT-2301 Operating instructions M2007/3en
©BTG 2004
4.1.1 Adjusting TAG / Serial no.
The TAG is the transmitter identification name in the process. Maximum 8 identification characters can be entered (letters and/or digits).
The setting is done using followed by .
Enter the required TAG. The buttons have several characters, select the correct one by pressing several times, wait two seconds and the cursor will jump to the next position.
4.1.2 Measuring range
The transmitter can handle four different measuring ranges and calibra­tions. Different ranges make it possible to have different calibrations for different qualities and conditions in the process.
Connection of different ranges is done via binary inputs, see section
3.3.6: Transmitter input and output, measuring range connections.
Range 1 is always active if nothing is connected to the binary inputs.
Make sure changes are made in correct range, select using .
The active range is shown on the main menu, lower right ”Range: X”
Confi­gure
Tag.no 25NT1237
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Read data
Edit Range 1
Confi- gure
Cali- brate
4 mA
1.00%
20 mA
3.00%
Damp1sOffset
0.00%
Misc. Print
Note ! Trend
Calib­ration curve
Send data
Tag: 25NT1237 Range 1
--------------­0% 53.3% 100%
Cons. 1.30% TCT-2301
Confi­gure
Tag: 25NT1237 Range x
--------------- 0% 53.3% 100%
Cons. 1.30% TCT-2301
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Read data
Edit Range x
Confgure
Cali­brate
4 mA
1.00%
20 mA
3.00%
Damp1sOffset
0.00%
Misc. Print
Note ! Trend
Calib­ration curve
Send data
Edit Range X
Tag: 25NT1237 Range 1
--------------- 0% 53.3% 100%
Cons. 1.30% TCT-2301
Tag: 25NT1237 Range x
--------------­0% 53.3% 100%
Cons. 1.30% TCT-2301
TCT-2301
Edit Range X
4.1 - Starting the transmitter
M2007/3en TCT-2301 Operating instructions 45
©BTG 2004
4.1.3 Select measuring unit
Unit selection is made using , . Press to
switch between mg/l and %. End by pressing .
4.1.4 Setting measuring span
Try to keep the measuring span as narrow as possible.
Specify the consistency values represented by output signals of 4 mA and 20 mA respectively.
Confi­gure
Tx. Spec
Unit xxx
Send page
Tag: 25NT1237 Range x
--------------- 0% 53.3% 100%
Cons. 1.30% TCT-2301
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Read data
Edit Range x
Confi­gure
Cali­brate
4 mA
1.00%
20 mA
3.00%
Damp1sOffset
0.00%
Misc. Print
Note ! Trend
Calib­ration curve
Send data
4 mA
1.00%
Tag: 25NT1237 Range 1
--------------­0% 53.3% 100%
Cons. 1.30% TCT-2301
20 mA
3.00%
4.2 - Clear water adjustment
46 TCT-2301 Operating instructions M2007/3en
©BTG 2004
4.2 Clear water adjustment
4.2.1 At start up
A high-quality consistency measurement depends on a good zero refer­ence value. The zero reference value is the signal when the probe is put in Clear Water (CW).
! The water used for clear water adjustment must be of drinking
water quality.
The purpose of this adjustment is to establish the LED intensity setpoint in water to have a zero reference when measuring in water. It also provides a comparison value when cleaning the probe.
For LED intensity adjustment select .
The following menu appears, see fig 34. Then press
Fig 35 shows the following menu.
1. Fill a bucket of drinking water and put the probe in the water. Make
Confi­gure
Led int. adjust
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Led int. adjust
Tag 25NT1237
Tx. Spec
Mode 1
Menu
Devive TCT-2301 version 1.xx Boot version 1.xx DSP version xx
<Led int. adjust>: Led intensity adjust­ment / basic sensitivity settings.
<Tag>: Transmitter ID <Tx. Spec>: Transmitter spec.
Fig 34 Configure sub-menu
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
CW/DCO CW
LED intens
57.0%
Set CW
96.7%
Edit CW
96.7%
Menu
Fig 35 LED int. adjust sub-
menu.
Send page
LED current: 54.1% CW measured: 97.8%
4.2 - Clear water adjustment
M2007/3en TCT-2301 Operating instructions 47
©BTG 2004
sure that no air bubbles are trapped in the measuring gap on the probe. Remove any air bubbles with your fingers.
2. Make sure the transmitter is set for clear water, . If the key
is set on a dark signal adjustment will be made. This ad-
justment has been made in the factory and need not be repeated. The dark signal adjustment is made using the plastic plug in the optical connection instead of the probe.
3. Check that the CW measured value is 100% ± 5. If this is the case no further action is needed, continue with 4.3.2. Otherwise continue with point 4.
4. Change the LED intensity by pressing .
Change the LED intensity in steps by 2. Send the setting to the trans-
mitter by pressing and .
5. Check that the CW measured value is close to 100% (±2%). If not, repeat step 4.
6. When the CW measured value is stable press to store the
value. Send the setting to the transmitter by pressing
and .
7. Install the probe in the valve arrangement and lock it properly using the two screws.
! Check that the measuring gap on the probe faces the flow at the
correct angle. See section 3.2.5 for instruction on how to change the angle. See section 4.2.3 for recommended angle settings
! The key ”Edit CW xx.x%” is used to manually send a water value to
the transmitter if needed.
CW/DCO CW
CW/DCO DCO
LED intens
57.0%
Send page
Send change
Set CW
96.7%
Send page
Send change
Changing the LED intensity adjustment changes the setting of the transmitter. This must be done only when the probe is in clear water.
4.2 - Clear water adjustment
48 TCT-2301 Operating instructions M2007/3en
©BTG 2004
4.2.2 Selecting the correct ”mode”
The TCT-2301 has preset calibration curves for different fiber types. Fi­ber types are divided into two main groups, low-strength pulp (short fi­ber) and high-strength pulp (long fiber).
This setting is located under . See fig 34 on page 46.
Press the key to select the correct preset calibration curve de-
pending on the actual fiber type:
: This mode is for short-fiber pulps,
such as recycled and groundwood.
: This mode is for pulp types such as
kraft pulp, TMP, CTMP etc.
: This mode contains a linear curve
and is not normally used.
! Please contact BTG for further information on different pulp and
fiber types.
Confi­ gure
Mode 1
Mode 1
Mode 2
Mode 3
4.2 - Clear water adjustment
M2007/3en TCT-2301 Operating instructions 49
©BTG 2004
4.2.3 Adjustment of measuring gap angle.
It is important that the measuring gap faces the flow at the correct angle. Follow these guidelines.
Start with the standard setting, 45° facing the flow.
If the signal from the transmitter appears to be unstable despite using the angles above, make the following quick test to see if can be improved. Remove the probe from the line, see section 3.2.5: Mounting the probe in the valve assembly, and adjust the angle in 15° increments. (15°, 30°, 45°, 60°, 75°). Install the probe and observe the signal from the transmit­ter.
! Contact BTG for further technical information.
! Never go under a 15° or over a 75° angle to the measuring gap. This
could spoil the cleaning effect of the measuring gap and/or create a plugging effect.
4.2 - Clear water adjustment
50 TCT-2301 Operating instructions M2007/3en
©BTG 2004
! Definition:
At 0° the measuring gap points directly towards the flow. At 90° the measuring gap points towards the pipe wall.
The transmitter is now ready for calibration against pulp samples.
Fig 36 Probe angles relative
to the measuring gap
Pulp Flow
Pulp Flow
Pulp Flow
Pulp Flow
Pulp Flow
Normal position
4.3 - Calibration
M2007/3en TCT-2301 Operating instructions 51
©BTG 2004
4.3 Calibration
This section assumes that you have completed section 4.2, and studied the operation of the hand-held terminal as described in section 4.8.
4.3.1 General remarks on calibration
The transmitter should be calibrated for the specific application after it has been installed in the line and all connections have been made.
The measuring principle used by this transmitter is insensitive to differ­ent wood species and normally one calibration can be used with different pulp/stock qualities.
The amount of filler can vary in different pulp/stock qualities and this can affect the output signal. This can be handled by using different calibra­tions (Measuring range).
4.3.1.1 Criteria for calibration
1. The transmitter is installed and has been started up.
2. A clear water adjustment must have been made as per section 4.2: Clear water adjustment.
3. The hand-held terminal must be connected to the electronics box and switched on, and the transmitter data read into the terminal as per section 4.1: Starting the transmitter.
4. The correct mode must be selected as per section 4.2.2: Selecting the correct ”mode”.
5. The pulp must have representative consistency and flow values.
4.3.2 General remarks on sampling
Additional information can be found in Accurate Consistency — a hand-book in consistency measurement in pulp and paper processing”,
available from BTG.
To ensure acceptable precision in laboratory samples, we recommend use
of BTG sampling valves.
4-20mA MODEMPRINTER PWR EJECTON/OFF ME MORY CARD
SPC-1000
Communication Terminal
Read data
Edit Range 1
Confi- gure
Cali- brate
4 mA
0.50%
20 mA
2.00%
Damp1sOffset
0.00%
Misc. Print
Note ! Tr end
Calib­ration curve
Send data
Tag: 25NT1237 Range 1
--------------­0% 53.3% 100%
Cons. 1.30% TCT-2301
Cali­brate
The only way to calibrate the transmitter correctly is to take correct laboratory samples and adjust the transmitter accordingly.
4.3 - Calibration
52 TCT-2301 Operating instructions M2007/3en
©BTG 2004
Sampling procedure:
1. Install the sampling valve close to the transmitter. The valve should be installed so there is no interference from pipe bends, pumps, etc. Recommendations regarding turbulence damping zones are the same as for the transmitter itself.
2. Take at least two, preferably three, laboratory samples. Calculate the mean value; reject samples that deviate widely from the mean. Take an adequate volume for sampling – at least 500 ml (1/2 US quart) for a reliable sample.
3. Make sure that samples for calibration and subsequent checking are always taken in the same manner regardless of who does the sam­pling. Similarly, the laboratory procedure must always be exactly the same.
The design of, and compliance with, sampling routines is crucial in as­suring uniform treatment of samples. Calibration and future precision of measurement by the transmitter depend on this.
During calibration, the pulp flow past the Transmitter must be represen- tative for the position in which it is installed.
To simplify and improve the regular follow-up routines, the BTG Smart lab sampler - SLS-1000 is a useful aid.
4.3.3 Calibration procedure
Normal calibration is very simple, only one sample needs to be taken.
The functionality of the TCT-2301 means that a single point calibration or even the preprogrammed curve is often fully adequate.
Since only one sample is taken, it is very important that it is correct. Therefore, take out a large sample so that several consistency determina­tions can be made at the lab, and an average value can be downloaded to the transmitter.
! It is important to read data to the SPC each time it is connected. It
is also important to continuously send all changes.
4.3.3.1 Summary
1. Collect values using the SPC at the same time as pulp samples are taken from the line.
2. Evaluate the sample at the lab.
3. Enter the lab value in the SPC.
4. Calculate calibration constants.
5. Send the calibration values to the transmitter.
Tag: 25NT1237 Range x
--------------- 0% 53.3% 100%
Cons. 1.30% TCT-2301
4-20mA MODEM
PRINTER
PWR
EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Read data
Edit Range x
Confi- gure
Cali- brate
4 mA
1.00%
20 mA
3.00%
Damp1sOffset
0.00%
Misc. Print
Note ! Trend
Caliration curve
Send data
Tag: 25NT1237 Range 1
--------------­0% 53.3% 100%
Cons. 1.30% TCT-2301
Cali­brate
4.3 - Calibration
M2007/3en TCT-2301 Operating instructions 53
©BTG 2004
4.3.3.2 Take sample
The menu is located under the Calibration menu.
1. Select a free position in the table using , note the number on
the sampling can.
2. Select the number of measuring values the transmitter is to collect by
pressing . The time for collecting measuring values is the
time needed to take a sample from the line (it takes approximately 2 seconds to collect a value, so an average of 5 takes approximately 10 seconds).
3. Press .
4. Collect pulp samples from the line.
5. The handterminal displays a graph of signal stability during the sam­pling. Evaluate the samples displayed to see how stable the signal was.
6. If OK, press , if not press and take a new sample.
The value is now stored in the SPC and can now be sent to the transmit-
ter. From the main menu, press .
Evaluate the sample in the lab.
4.3.3.3 Load lab value
Read all data again from the transmitter to the SPC-1000.
Enter the consistency value as follows:
1. Go to menu and .
2. Step to correct position for the sample using .
3. Press to active the value (”ON”).
4. Press and enter the lab value.
Return to the main menu and .
Take sample
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Take sample menu
Load lab. cons.
Calc. new calibr.
Special func.
Menu
Action:
1. Do configure before calibration.
2. Push <take sample> and collect sample
3. After lab. evaluation push <Load lab. cons.>and enter your lab cons. value.
3. Push <Calc. new calibr.> to calculate the new calibration.
<Special func.>: Set or adjust sample values and lab values manually
Take sample menu
Edit sample X
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Sample Probe no signal
-->1 1.52% 2 1.77% 3 1.30% 4 1.09% 5 0.00% 6 0.00% 7 0.00% 8 0.00% 9 0.00%
Probe s. 36.85%
Menu
New calib ration
Take sample
Edit Sample 1
Avrg. of 15
Take sample
Avrg.
of x
Take sample
Accept Cancel
Send data
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
SampleProbeLab.-Active no.conscons.
1 1.52%1.66%ON 2 --OFF 3 --OFF 4 --OFF 5 --OFF 6 --OFF 7 --OFF 8 --OFF 9 --OFF
ON=Included OFF=Excluded
Menu
Edit sample X
ON/OFF
Load lab. cons.
Load lab. cons.
Cali­brate
Load lab. cons.
Edit sample X
ON/OFF
Send data
4.3 - Calibration
54 TCT-2301 Operating instructions M2007/3en
©BTG 2004
4.3.3.4 Calculate calibration
The calibration will not be effective until the calibration value is calcu­lated.
Go to menu and .
The calculation is presented in a graph.
To activate the calibration, send the calibration value to the transmitter.
Go to main menu and .
4.3.4 Multi-point calibration
When single-point calibration is not sufficient a multi-point calibration can be used.
Take samples for up to 9 calibration points under different conditions. Try to collect samples that cover the entire consistency span.
If the calibration points (the lab value) are spread over at least half the chosen measuring span, both calibration constants K0 and K1 will be af­fected. Otherwise only K1 will be affected and K0 will be 0.
The more points you sample, and the wider their spread within the measuring span, the more accurate the calibration will be.
Perform sampling as in single-point calibration, but take more samples and activate more than one sample in the ”Load lab menu”.
Calculate new K0 and K1.
4.3.5 Calibration with given constants
The two calibration constants K0 and K1 define the calibration curve.
It is a simple matter to enter new constants under , . Step
down through the menu and enter desired value.
Cali­brate
Calc new calibr.
Send data
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Probe signal %
2.0
Lab
0.5% 2.0 cons. %
Menu
Constants k0 = 26.68 k1 = 7.87
Correl.: r2= 1.0000
Cali­brate
Special func.
Edit row x
SPC-1000
Communication Terminal
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMO RY CARD
SPC-1000
Communication Terminal
No. Probe Lab.- Active
signal cons.
1 1.52% 1.66% ON 2 1.77% 1.86% ON 3 1.30% 1.34% ON 4 1.09% 1.10% ON 5 0.00% 0.00% OFF 6 0.00% 0.00% OFF 7 0.00% 0.00% OFF 8 0.00% 0.00% OFF 9 0.00% 0.00% OFF k0 = 0.1366 k1 = 0.8613
Menu
Edit row 1
Load lab. cons.
Calibr. curve
Load probe signal
ON/OFF
4.3 - Calibration
M2007/3en TCT-2301 Operating instructions 55
©BTG 2004
4.3.6 Calibration table
All values collected for calibration are stored in a single table.
Up to 9 samples can be stored in the table even if only one is used for the calibration (single-point calibration).
The samples can be activated (ON) in the calibration.
Columns in the load lab menu shown to the left are explained below:
4.3.7 Calibration curve
This function is located in the main menu under .
The probe signal is displayed as a function of the set measuring span xx
- xx%. Probe signal points within the predetermined measuring span are displayed. Note that activated calibration points outside the measuring span affect the correlation.
The two calibration constants K0 and K1, and the special correlation fac­tor r2, provide a guide. Note that the displayed probe signal and the mea­suring span are the same.
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Sample Probe Lab.- Active No. signal cons.
11.52%1.66%ON
21.77%1.86%ON
31.30%1.34%ON
41.09%1.10%ON
50.00%0.00%OFF
60.00%0.00%OFF
70.00%0.00%OFF
80.00%0.00%OFF
90.00%0.00%OFF ON=Included
OFF=Excluded
Menu
Edit sample 0
ON/OFF
Load lab. cons.
Sample no: 1 to 9
Probe signal: Value from probe.
Lab cons: Consistency. Evaluated from sample.
Active: If this sample is used in calibration Active is set to ON,
otherwise it is set to OFF.
4-20mA MODEMPRINTER PWR EJECTON/OFF ME MORY CARD
SPC-1000
Communication Terminal
Probe signal %
2.0
Lab
0.5% 2.0 cons. %
Menu
Constants k0 = 0.1366 k1 = 0.8613
Correl.: r2= 1.0000
Calib­ration curve
4.4 - Other functions
56 TCT-2301 Operating instructions M2007/3en
©BTG 2004
4.4 Other functions
4.4.1 Setting time constant (damping)
The time constant is set after calibration is complete. Set it so that the sig­nal is stable, normally at 2 to 10 seconds.
If you find that you have to set a very long time constant because the probe signal is unsteady, the transmitter is probably working in an unsta­ble, poorly mixed pulp flow. In such a case you should consider:
relocating the transmitter further from the source of turbulence.
improving the remixing system or the supply of dilution water, etc.
If the time constant is too long, the high precision level of the transmitter is reduced. Contact BTG for further advice.
Procedure:
1. Study the stability of the signal under MAIN MENU op-
tion .
2. Go to MAIN MENU option . Enter a suitable time con-
stant, e.g. 5 seconds. Press , .
3. Return to and study the result.
4. If necessary, adjust the time constant.
Do not set a time constant that is too long as this reduces the accuracy of the control system.
Trend
Damp Xs
Send data
Send change
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Read data
Edit Range 1
Confi­gure
Cali­brate
4 mA
0.50%
20 mA
2.00%
Damp1sOffset
0.00%
Misc. Print
Note ! Trend
Calib­ration curve
Send data
Tag: 25NT1237 Range 1 0% 53.3% 100% Cons. 1.30% TCT-2301
Damp 1s
Trend
4.4 - Other functions
M2007/3en TCT-2301 Operating instructions 57
©BTG 2004
4.4.2 Setting alarm limits
In the system there is one alarm. The alarm can be set on:
Cons (Output value, chan. 1)
Itemp (internal temperature)
AnaIn (analog input, 0-20mA)
LedCur (LED current, 0-1000%)
When the alarm is active it affects the digital output signal called Alarm and the value will be written to an event log in the transmitter.
The alarm output should be connected to mill equipment. See section
3.3.6: Transmitter input and output, measuring range connections for de- tails.
1. From main menu, press , .
2. Select alarm to set using . (One alarm only)
3. Select limits to activate the alarm using and .
The same units are used for the set value as for the selected alarm.
4. Activate alarm using .
4.4.3 Offset adjustment
Offset adjustment can be made in two ways:
A: Based on a laboratory sample
B: Based on a deviation detected from laboratory samples.
!
The entire calibration curve is moved on the basis of a single point, so it is most important that the laboratory sample value for that point is correct.
The offset adjustment, which is a zero point displacement, can be made on all types of calibrations.
A. Offset adjustment based on a laboratory sample
Procedure:
1. Go to MAIN MENU option .
Misc Alarm &
diag­nostics
Set alarm
High
x.x
Low
x.x
Alarm ON
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Take sample
Load lab.
0.00%
Calc. Offset
+5.00%
Menu
Sensor cons.= 3% acc.to original cal.
Action:
1. Push <Take sample> and collect sample
2. After lab. evaluation push <Load lab.>and enter value
3. Push <Calc.> alt. enter the actual value at <Offset>.
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Take sample
Load lab.
0.00%
Calc. Offset
+5.00%
Menu
Sensor cons.= 3% acc.to original cal.
Action:
1. Push <Take sample> and collect sample
2. After lab. evaluation push <Load lab.>and enter value
3. Push <Calc.> alt. enter the actual value at <Offset>.
Send page
Offset x.xx%
4.4 - Other functions
58 TCT-2301 Operating instructions M2007/3en
©BTG 2004
2. Press , and save the consistency value to mem-
ory in the normal manner, taking correct laboratory samples at the same time.
3. Load the corresponding laboratory sample value, xx
%. The consistency according to the original curve appears in
the display, and after pressing the offset consistency
appears in the key in the offset menu.
4. Press to transfer the offset adjustment to the transmit-
ter.
5. The offset adjustment can be cancelled at if you enter
the value Ø (zero).
B. Offset adjustment based on a detected deviation
A deviation can be entered directly if a deviation from the laboratory val­ues is detected during regular monitoring of a transmitter’s display. Note however that the deviation should be verified by several laboratory sam­ples before carrying out zero point adjustment.
Procedure:
1. Go to the main menu function .
2. Press in the submenu and enter the detected deviation.
3. Transfer the value to the transmitter using . (Clear the
value by going back to the Offset function and entering 0.)
Take sample
Start
Load lab. cons. x,x%
Calc.
Offset x.xx%
Send page
Offset
0.00%
Offset x.xx%
Offset x.xx%
Send page
4.4 - Other functions
M2007/3en TCT-2301 Operating instructions 59
©BTG 2004
4.4.4 Manual loading of probe signal settings
This function is located under , .
It is a special function which can be useful in certain cases.
Procedure:
1. Go to MAIN MENU option , .
2. Step between values and load the desired probe signal set-
ting by pressing . Exit by pressing the key.
3. Press , to send calibration data to the transmitter.
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Take sample menu
Load lab. cons.
Calc. new calibr.
Special func.
Menu
Action:
1. Do configure before calibration.
2. Push <take sample> and collect sample
3. After lab. evaluation push <Load lab. cons.>and enter your lab cons. value.
3. Push <Calc. new calibr.> to calculate the new calibration.
<Special func.>: Set or adjust sample values and lab values manually
Cali­brate
Special func.
Cali­brate
Special func.
SPC-1000
Communication Terminal
4-20mA MODEMPRINTER PWR EJECTON/OFF ME MORY CARD
SPC-1000
Communication Terminal
No.ProbeLab.-Active signal cons.
11.52%1.66%ON
21.77%1.86%ON
31.30%1.34%ON
41.09%1.10%ON
50.00%0.00%OFF
60.00%0.00%OFF
70.00%0.00%OFF
80.00%0.00%OFF
90.00%0.00%OFF k0 = 0.1366 k1 = 0.8613
Menu
Edit row 1
Load lab. cons.
Calibr. curve
Load probe signal
ON/OFF
Edit row
x
Load probe signal
Menu
Send data
Send change
4.5 - Evaluation and documentation of calibration
60 TCT-2301 Operating instructions M2007/3en
©BTG 2004
4.5 Evaluation and documentation of calibration
When you have made a calibration it should be evaluated, documented and backed up. This data is needed for possible fault tracing, or if the trans­mitter or electronics card is replaced.
4.5.1 Trend
This function is in the main menu under .
The time scale of the display is changed by pressing . It starts
from the point at which the button is pressed.
The 4 - 20 mA output signal is displayed as 0 - 100%. This means that the resolution is relatively low - approximately 1% (0.16 mA) - but the display is still a useful tool when setting parameters such as the time con-
stant .
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
100| | | | Output | 12.53 mA | 53.3 % 50| | Cons. | 1.30 % | | | 0|________________________ 0 5
MenuScale
5
Start new trend
Trend
Scale 5
Damp 1s
4.5 - Evaluation and documentation of calibration
M2007/3en TCT-2301 Operating instructions 61
©BTG 2004
4.5.2 Documenting calibrations with printer
A standard printer is an invaluable aid that simplifies evaluation and doc­umentation of transmitter settings.
The printer must be EPSON-compatible with an RS232 serial input. The recommended model is SEIKO DPU 414. The printer should be set to a baud rate of 2400.
An alternative is to use a special PC program for print out. See section
4.5.4: Documenting calibration using a PC.
Documentation can also be made manually. Make a written record of all data listed on the printout shown in fig 37 on page 62, and draw a cali­bration graph.
!
Actual data is also available at , .
1. Plug the printer into the hand-held terminal socket marked ”PRINT­ER. Switch on the printer.
2. Select the measuring range to be printed .
3. In the MAIN MENU press .
Then select preferred printer type and a printout is made.
If the display reads ”Reply time out”, check that the printer is correctly plugged in, that the battery is charged or the printer is connected to an AC adapter and that the baud rate (2400) is correctly set.
A sample printout is shown in fig 37 on page 62.
Cali­brate
Load lab. cons.
Edit range X
Print
4-20mA MODEMPRINTER PWR EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
Read data
Edit Range 1
Confi­gure
Cali­brate
4 mA
1.00%
20 mA
4.00%
Damp1sOffset
0.00%
Misc. Print
Note ! Trend
Calib­ration curve
Send data
Tag: 25NT1237 Range 1
--------------­0% 53.3% 100%
Cons. 1.30% TCT-2301
Print
4.5 - Evaluation and documentation of calibration
62 TCT-2301 Operating instructions M2007/3en
©BTG 2004
.
.
BTG CONS. TRANSMITTTER TCT-2301
SPC software version 1.2en Software version 1.04 Boot version 1.01 DSP version 7g Date Sign
Tag 12NT1237 Serial. no. 123456/78/90 Configuration values: LED int = 50.0 % CW = 100.0 % DCO = 2800
---- Range no. 1 4 mA = 0.50 % 20 mA = 2.00 % Offset = 0.00 % Damp = 1 s Mode = 1 Calibration: k0 = 0.1366 k1 = 0.8613 No. Probe cons. Lab cons. Active 1 1.52 1.66 ON 2 1.77 1.86 ON 3 1.30 1.34 ON 4 1.09 1.10 ON 5 - 0.00 OFF 6 - 0.00 OFF 7 - 0.00 OFF 8 - 0.00 OFF 9 - 0.00 OFF
Events & Alarms:
Fig 37 Sample of calibration
documentation printout from printer or PC
11
8
7
4
3
1
9
5
13
12
* Note: The probe signal
values shown in this diagram
diagram are not printed on a normal printout.
15
0.00
1.00
3.00
2.00
0.00 1.00 2.00 3.00
Lab cons (%)
Probe signal (%)
14
10
6
2
1Tx. model 2 Software version 3 Operator data 4Tx. data 5LED int. 6 Clear water value 7 Range 8 Measuring span
9 Consistency offset 10 Damping time 11 Mode (fiber type) 12 Calibration constants 13 Calibration data 14 Data log.
4.5 - Evaluation and documentation of calibration
M2007/3en TCT-2301 Operating instructions 63
©BTG 2004
Fig 38 Calibration diagram for TCT
5
4.5
0
0.5
1
1.5
2
2.5
3
3.5
4
0 0.5 1 1.5 2 2.5 3 3.5
[%] Consistency
[%] Probe signal
4
4.5 - Evaluation and documentation of calibration
64 TCT-2301 Operating instructions M2007/3en
©BTG 2004
4.5.3 Documenting calibration using the ”back-up card”
All transmitter settings, calibration data and transmitter data can be stored on a special SRAM memory card. This is the ”back-up card”. The card replaces the standard ”sensor card” which holds the transmitter pro­gram. The back-up card can store up to 25 complete transmitter settings, including all four measuring ranges. For practical reasons, it is preferable to store only one mill section´s transmitters (TAG) on one card.
A. Storing the transmitter’s data on the back-up card (Store data).
1. With sensor card inserted make a standard complete calibration and adjustment of the relevant transmitter.
Download data from the transmitter.
2. Switch off the hand-held terminal and replace the sensor card with the back-up card.
3. Switch on the hand-held terminal. Press .
The message for this function is displayed only in English.
4. Select the relevant storage position. Data can be stored on every
fifth position e.g. 5, 10, 15, 20....up to 125, with a total of 25 po-
sitions. For example, in the case of storage position No. 30 press
one time (+25 in each step) = 25 and one time
(+5 in each step) = 5. 25+5=30
!
Transmitter data can only be stored at every fifth position (5, 10, 15 etc.).
5. Press to store the terminal’s transmitter adjustment for
Always switch off the hand-held terminal when changing a memory card. The terminal could be damaged in some cases if this is not done!
Read data
BACKUP CARD
+10 +1
4-20mA MODEM
PRINTER
PWR
EJECTON/OFF MEMORY CARD
SPC-1000
Communication Terminal
CE
STORE DATA
+10 +1 READ
DATA
+ 10/+ 1
STORE DATA: SPC 1000 ----> BACKUP CARD
READ DATA : BACKUP CARD --> SPC 1000
(1-125): 0
Fig 39 Store data menu
Backup card
SPC-1000
74414152
Copyright© 1995
BTG
All rights reserved
CE
STORE DATA
4.5 - Evaluation and documentation of calibration
M2007/3en TCT-2301 Operating instructions 65
©BTG 2004
the relevant transmitter (TAG) on the back-up card.
Then press for storage to be executed.
! Note the relevant TAG and the number of the storage position on a
separate identification list.
All four measuring ranges will be stored so each one must be clearly identified in the identification list.
6. Switch off the terminal and replace the back-up card with the sensor card.
Store the back-up card in a safe place protected from moisture. The card is an object of considerable value.
The back-up card can be protected against accidental or unwanted data input by using its write-protect. Activate the write-protect switch to block input.
B. Retrieving transmitter data stored on the back-up card.
1. With sensor card inserted, select .
2. Switch off the hand-held terminal and replace the sensor card with the back-up card.
3. Switch on the hand-held terminal. Press .
The message for this function is displayed only in English.
4. Select the relevant storage position. For example, in the case of storage position No. 30 press
one time (+25 in each step) = 25 and
one time (+5 in each step) = 5. 25+5=30
5. Press to access the relevant transmitter setting from the
back-up card and transfer this to the terminal’s memory. The relevant Tag number is shown in the identification list as per Item 5 section A: Storing the transmitter’s data on the back-up
card (Store data). Press .
6. Switch off the hand-held terminal and replace the back-up card with the sensor card.
Verify
TCT
2301
v1.0
BACKUP CARD
+10
+1
Read data
Verify
4.5 - Evaluation and documentation of calibration
66 TCT-2301 Operating instructions M2007/3en
©BTG 2004
7. Switch on the hand-held terminal. Press, .
! Do not ” ” from the transmitter. If so the values obtained
from the backup card will be erased.
8. The data for the respective measuring range is now available in the terminal for further processing.
4.5.4 Documenting calibration using a PC
If you use Windows 3.1 or later, you can use the ”SPC-1000 Documen­tation” software, available from BTG. If you have several BTG smart transmitters this is a practical way of storing all transmitter data within the same file.
Installation — Document PC:
The SPC-1000 Documentation program for Windows® comes with an automated installation program called setup.
! ”SPC-1000 Documentation” Version 1.21 or later must be used.
Installing the ”SPC-1000 Documentation” program:
1. Start Windows. Insert the installation diskette (marked ”SPC­1000 Documentation”) into drive A (or relevant drive).
2. Choose Run.
3. In the Command Line text box, type: a:\setup, then select OK.
4. Follow the instructions to install the program to your hard drive.
TCT
2301
v1.0
Read data
Fig 40 Store function of
the SPC-1000
SPC-1000 Documentation
program for Windows
Version 1.X
Copyright© 1995 BTG All rights reserved
4.5 - Evaluation and documentation of calibration
M2007/3en TCT-2301 Operating instructions 67
©BTG 2004
Connection and starting your program:
! Arrange a suitable library in the File manager to help you to keep
track of downloaded files (TAGs).
1. Plug the serial cable into a free comm port on your computer, using either a 25-pin or a 9-pin, and into the hand-held terminal (marked ”PRINTER”).
2. Double click on the BTG icon.
3. Double click on the SPC-1000 terminal icon.
4. Click on ”CommPort”, ”Settings” and select the connected comm- port. (normally selected automatically)
5. For more help click on the menu choice ”Help”.
6. Switch on the handheld terminal and press .
7. Send data to PC: Main menu option .
8. Press option and values as well as calibration curve are cop-
ied to PC.
9. Save (and print if necessary).
Fig 41 PC serial connections
9-Pin 25-Pin
Read data
Print
PC
4.6 - Electronics box display menu
68 TCT-2301 Operating instructions M2007/3en
©BTG 2004
4.6 Electronics box display menu
4.6.1 General information
The TCT-2301 electronics box is equipped with a display and four keys. The dimension of the display is one row with sixteen characters. This al­lows the operator to make some basic adjustments without using the handheld terminal SPC-1000. To make a complete calibration or other major settings the SPC-1000 must be used. Settings made using the elec­tronics box can be read from the SPC-1000 for documentation and print­out.
Fig 42 Electronics box front
panel with buttons and display
1 Display: 1 row
16 characters
2 Sample button: to store
probe signal for calibration or consistency value for offset adjustment.
3 Arrow up button: to
scroll in the menu, change a value or discard sample
4 Arrow down button: to
scroll in the menu, change a value or discard sample
5 Enter button: to select
menu options, verify entered values
6 Power button
1
2
5
4
3
6
4.6 - Electronics box display menu
M2007/3en TCT-2301 Operating instructions 69
©BTG 2004
4.6.2 Menu structure
! After 15 seconds, the transmitter always returns to the result
display regardless of where it was in the menu structure.
Press the arrow buttons to step between menu items.
Start up display
TCT-2301 Ver x.x
Shows transmitter model TCT-2301 and software version.
Result display 1
1.30% TCONS
Result display, shows consistency.
Result display 2
12.53mA TCONS
Result display, shows consistency output signal.
Range setting
RANGE 1
Current measuring range, can be changed to feed in values for other ranges (range 1, 2, 3 and 4).
Span setting
SPAN
Measuring span, low limit = 4mA high limit = 20mA.
Offset adjustment
OFFSET
Consistency offset if required.
Damping setting
DAMPING
Damping on output signal.
Calibration setting
CALIBRATION
Calibration constants for the transmitter.
Cleaning cycle
CLEANING
The cleaning system is an option for the TCT. All time settings to control the system are made here.
Language selection
LANGUAGE en
Language setting, in this case English.
4.6 - Electronics box display menu
70 TCT-2301 Operating instructions M2007/3en
©BTG 2004
4.6.2.1 How to change settings and collect samples
Result display
1.30% TCONS
When the result display is shown, samples can be collected for calibration purposes.
Press the sample and collect a sample from the pipe
for lab evaluation.
NO:5 *
The display indicates the position in the calibration table where the sample will be stored. The first empty position in the table is selected. Mark the lab sample with this number. Value X will be counted down for each signal stored. An average will then be calculated.
NO:5 1.30% OK?
Save the sample by pressing enter , discard the
sample by pressing or .
SAMPLE SAVED
Enter .
NOT SAVED
or .
SAMPLE TAB FULL
If all nine samples in the calibration table are already used, select the sample to be replaced with a new probe signal value.
9
Press enter and select which calibration sample to
replace using the or buttons. Verify with
enter . Press or to return to result display
without replacing any calibration sample.
NO:9 *
The display indicates the position, take out the lab sample at the same time.
NO:9 1.35% OK?
Enter to save the value, or to discard.
4.6 - Electronics box display menu
M2007/3en TCT-2301 Operating instructions 71
©BTG 2004
! Always mark the lab sample with the number indicated on the
display. The handheld terminal SPC-1000 must be used if you are making a complete calibration.
Range setting
RANGE 1
Press enter to select another range.
1
Press or to specify range. Verify by pressing
enter. Note! The transmitter is still operating on the range set by the external switch.
Span setting
SPAN
Press enter to change span.
0.50-2.00 %
The first number is blinking, press or to change
the number, press enter when ready and go to next
number and repeat the procedure.
1.00-2.50 %
When the last number is blinking, press or to
change and enter when ready.
SPAN
The result will be stored in the transmitter and the display will return to the menu.
Offset adjustment
OFFSET
Press the sample in this position and collect a
sample from the pipe for lab evaluation.
SAMPLE *
1.77% OK?
Press enter to save the value, press or to
discard the value.
4.6 - Electronics box display menu
72 TCT-2301 Operating instructions M2007/3en
©BTG 2004
SAMPLE SAVED
Enter .
NOT SAVED
or .
OFFSET
When the lab evaluation is ready, step down to offset and
press enter .
1.77% +0.00%
Calculate the offset, this means the lab value minus the saved value. If the lab value is 1.69 the offset in this case will be 1.69–1.77=–0.08%.
1.77% -0.00%
The + sign is blinking, press or to change sign,
press enter when ready.
The first digit starts to blink, press enter twice to
step to the last digit.
1.77% -0.08%
Press or to change the digit.
End by pressing enter .
OFFSET
The display will return to the menu.
Damping setting
DAMPING
Press enter if the damping value is to be changed.
00s
Press or to change value,
verify with enter .
Calibration setting
CALIBRATION
Calibration constants for the transmitter. Press enter
and then or to step between calibration
constants.
4.6 - Electronics box display menu
M2007/3en TCT-2301 Operating instructions 73
©BTG 2004
K0 0.2
Press enter if the value is to be changed. Press
or to change value, verify with enter .
K1 1.15
Press enter if the value is to be changed.
Press or to change value, verify with enter .
Cleaning cycle (option)
CLEANING
Press enter to change the cleaning cycle.
Use the arrow buttons to step between
Cycle, Flush and Resume.
CYCLE 0000 m
Press enter to change the cycle time. This is the time
in minutes between each cleaning.
0000 m
Press or to change value,
verify with enter . (Set to 0 if not used.)
FLUSH 00s
Press enter to change the flush time. This is the time
in seconds that it takes to retract the probe and to flush the measuring gap clean.
00s
Press or to change value,
verify with enter . (Set to 0 if not used.)
RESUME 00s
Press enter to change the resume time. This time is
to allow the probe enter back into the pipe before the TCT-2301 starts to measure again. Also set in seconds.
00s
Press or to change value,
verify with enter . (set to 0 if not used)
4.6 - Electronics box display menu
74 TCT-2301 Operating instructions M2007/3en
©BTG 2004
Language selection
LANGUAGE en
Selected language is indicated by the lower case letters.
English
Press enter to change language.
Scroll with the or keys to the desired language.
Verify with enter .
Svenska
Español
Italiano
Deutsch
Português
Suomi
Français
4.7 - Hand-held Terminal SPC-1000
M2007/3en TCT-2301 Operating instructions 75
©BTG 2004
4.7 Hand-held Terminal SPC-1000
When communicating with the sensor, the hand-held terminal is normal­ly placed in the special terminal holder installed on a wall close to the electronics box.
When the terminal is not in the wall terminal holder, place it on your left forearm (for right-handed users) and hold it securely using the hole in the handle. This is the most practical and safest way of using it, you can work properly without damaging the terminal.
The telephone jack on the spiral cable can be parked in the hole in the rubber transport casing.
You may need to upgrade your SPC-1000 to use it with the TCT-2301.
Check the sign on the back of the SPC-1000 to see if you have the re­quired version SPC-1000/A.
If you have a previous version of SPC-1000, an upgrade kit is available. You can also send it to BTG for upgrading.
SPC-1000/A version can be used with all BTG smart transmitters.
! Treat the terminal with care. Do not abuse it physically, submerge
it or spray it with water. Avoid storing it in a humid environment for long periods.
Damp
0% 61.7% 100%
Cons. 3.35%
Confi-
gure
Trend
Note
1
Edit Range
20 mA
4.0%
Print
4mA
Misc.
2.0%
data
Read
Tag
2s
Offset
0%
brate
Cali-
Send data
Range 2
Calib­ration
curve
MEK-2300
PRINTER PWR ON/OFF EJECT
C
ommunicationTerminal
SPC-1000
MEMORYCARD4-20mAMODEM
Fig 43 Handling the SPC-1000
terminal
Make sure that your hand-held terminal is upgraded to a SPC-1000/A
Fig 44 Type sign
4.7 - Hand-held Terminal SPC-1000
76 TCT-2301 Operating instructions M2007/3en
©BTG 2004
4.7.1 Power supply
The hand-held terminal can be powered in three different ways:
12 V DC supply from the electronics box
12 V DC supply from an AC adapter
9 V DC from an internal battery.
12 V DC supply from an electronics box.
This is the normal way of providing power for transmitter calibration and process monitoring. The terminal’s telephone jack is connected to the electronics box jack connection.
This also provides a communications link with the transmitter. When this supply option is chosen, the display lighting comes on for best visibility.
12 V DC supply from an AC adapter.
This type of power supply is suitable during simulation and documenta­tion such as when the terminal is used at a desk.
A portable AC adapter is supplied with the hand-held terminal. The AC adapter is available for 110 V AC or 220 V AC. The output supply is 12 V DC.
The AC adapter is supplied with a cord for plugging in to the mains pow­er supply. In some cases, the molded plug may not fit your outlet. Re­place the plug with a suitable standard plug if this is the case.
Fig 45 Hand-held terminal
SPC-1000 Connections
1 4-20mA modem
connection 2 Printer connection 3 Power connection 4 Power ON/OFF 5Eject 6 Memory card slot
Communication Terminal
SPC-1000
PRINTER PWR ON/OFF EJECT MEMORY CARD4-20mAMODEM
Sensor card
142 6
3
5
Check that the AC adapter is suited to power supply voltage (110 or 220 VAC).
4.7 - Hand-held Terminal SPC-1000
M2007/3en TCT-2301 Operating instructions 77
©BTG 2004
The AC adapter’s 12 V connection is a socket located on top of the ter­minal. This is marked ”PWR”. (12 V DC/0.4 A AC adapter, DC jack with 2 mm connection pins.) The + connection goes to the casing and the – connection goes to the center pin.
The display lighting also comes on when the 12 V power supply is used.
9 V DC supply from an integral battery.
The battery power supply is used, for example, when the terminal is con­nected to the 4 - 20 mA signal loop at a location other than at the elec­tronics box.
An internal standard 9 V DC battery is located behind a cover on the back of the terminal. To provide maximum operating time, a type 6F22 high­power alkaline 9 V DC battery should be used. A battery of this type will operate for approximately 3 hours.
! Switch off the terminal immediately after use to maximize operating
time.
Rechargeable batteries or manganese batteries provide much shorter op­erating times. Display lighting does not come on in battery mode in order to maximize battery life and operating time. Display readability is how­ever sufficient in normal room lighting.
4.7 - Hand-held Terminal SPC-1000
78 TCT-2301 Operating instructions M2007/3en
©BTG 2004
4.7.2 Code protection
A security code can be entered to prevent unauthorized adjustment of transmitter settings. All functions in the hand-held terminal can still be accessed. Code protection prevents the settings on the hand-held terminal from reaching the transmitter if the security code is not entered. A per­sonal code can be entered and, if this is lost, a general code is available.
! General code: 42600.
A. Initial activation of the security code / input of a personal code
1. This should be done at the first opportunity during installation.
Go to in the main menu. Press followed
by , . ”Load code” will be displayed.
Enter 42600 (applies only on this occasion). Press .
2. Enter your personal code. ”New value” will appear. Enter your
personal code, e.g. 123 (max 8 digits) and press .
3. Press to get . Your new code is now activated.
Finish by pressing .
Data can now be transferred to the transmitter. If you wish to
check that the code has been entered, press
and . The signal transfer will not take place without the
input instruction for the code being displayed.
Send data
Turn code ON/OFF
Menu Send
Change
Enter Menu
Enter Menu
Code OFF
Code ON
Menu
Send data
Send all
4.7 - Hand-held Terminal SPC-1000
M2007/3en TCT-2301 Operating instructions 79
©BTG 2004
B. Changing your personal code
Go to , . Press ”Load code” will be displayed.
Enter your old personal code (or the general BTG code). ”New value” will then be displayed. Enter your new code. Press Enter.
C. Deactivating the code you have entered
Press , . Press the ”button” and enter your per-
sonal code.
D. Activating the code you have entered
Press , . Press the ”button” so that it displays
”code on”.
Send data
Turn code ON/OFF
Change code
Send data
Turn code on/off
Code on/off
Send data
Turn code on/off
Code on/off
4.8 - Digital communication according to the Hart®-protocol
M2007/3en TCT-2301 Operating instructions 80
©BTG 2004
4.8 Digital communication according to the Hart®­protocol
Certain functions are available as a digital signal according to the Hart®­protocol and BELL 202 modem standard.
! The transmitter complies to Hart
®
-protocol standard requirements. To set and calibrate the transmitter the BTG hand-held terminal is required. Since additional functions are not covered by the Hart
®
-
protocol, a standard Hart
®
competitive terminal can only be used
for certain functions.
Communication takes place as a superimposed signal over the 4-20 mA output signal loop.
! For details, please contact the Hart
®
Communication Foundation or
BTG.
5.1 - Maintenance planning and quality assurance
M2007/3en TCT-2301 Service instructions 81
©BTG 2004
5 Service instructions
5.1 Maintenance planning and quality assurance
In April 1993, BTG Källe Inventing AB attained ISO 9001 certification for development, design and production of process control equipment for the pulp and paper industry. Factory repairs are made in accordance with the appropriate requirements for new products. Every transmitter is test­ed according to a specified program. The result is documented and in­cluded in the delivery.
5.1.1 Inspection of the accuracy/calibration of the
consistency transmitter
Inspection should be made against laboratory samples with the trans­mitter mounted in the pipe and running in operational mode. Comparison of lab samples can generate several types of error. The trans­mitter should be calibrated against many lab samples and adjusted ac­cordingly. Thus, comparison to one lab sample only can cause error.
The sample can be extracted from the pipe in a number of ways. Dif- ferent types of sampling valves, respectively common ball valves, result in a various number of serious sources of error, which will not be dis­cussed here. The guiding rule for good sampling is generally that the
samples shall be extracted from the pipe in a uniform, repeatable way in accordance with the original calibration. The sample volume must not be too small. At least 0.5 liter / 0.13 US gallon is necessary. The sample, after extraction from the pipe, must be treated in a uni­form, repeatable way. The lab analysis is to be in accordance with es-
tablished standards, e.g., SCAN-C 17:64 or valid TAPPI standard. Rapid analysis with the sample dried on a heating plate is not a reliable method.
Should a deviation be discovered, remove the transmitter from the pipe and clean the measuring gap. If this does not improve the situation, a recalibration (or possibly a zero point offset) of the transmitter may be necessary.
A printout, or other automatic documentation, is an excellent comple­ment to the ordinary protocol for documentation purposes.
5.1 - Maintenance planning and quality assurance
82 TCT-2301 Service instructions M2007/3en
©BTG 2004
5.1.2 Calibration recommendations
Calibration recommendations regarding comparison to lab samples and previous calibrations.
Twice per year:
Transmitters used for simpler measurements, e.g. for indicating purposes
Every month:
Transmitters used for measuring/control function of vital impor­tance to the process
Once a week or after a considerable change of process variables:
Transmitters used for measuring/control function of vital impor­tance to the process or used for debiting purposes
! The above recommendations are intended as a general guide and
should be adapted based on actual experience.
5.1.3 Maintenance of the transmitter
Maintenance needs will depend on the transmitter position, media influ­ence, and ambient conditions.
Regular maintenance includes:
Cleaning the measuring gap.
Inspection of wetted rubber details and metal parts for damage twice per year if exposed to aggressive chemicals.
Long-term maintenance includes:
Replacement of seals, typically after 5 years of operation.
5.1 - Maintenance planning and quality assurance
M2007/3en TCT-2301 Service instructions 83
©BTG 2004
5.1.4 Functional description
A light shines through the pulp suspension from the electronics card via the optical fiber cable. The light that passes through the suspension is measured and analyzed. The final measuring value is then converted into an analog output signal and shown on the display. The display also shows the value of the analog output signal generated by the instrument.
Fig 46 Electronics schema tic
diagram
1. Measuring gap inserted in the pulp suspension.
2. Processor card fitted in the electronics box.
3. Optical unit where the light beam through the fiber suspension is measured.
4. LED intensity control.
5. Amplifier for the incoming signal.
6. A/D converter of the measuring signal.
7. Digital signal processor (DSP) for handling and analysis of the signal to the measuring value used for consistency calculation.
8. Dual port memory for exchange of data between DSP and H8 pro­cessor.
9. Main processor handling calculations, communications etc.
10. Buttons and display positioned on the front of the electronics box.
5.2 - Service electronics
84 TCT-2301 Service instructions M2007/3en
©BTG 2004
5.2 Service electronics
5.2.1 Electronics troubleshooting
Symptom Probable cause Solution
1. Instrument does not power up
1.1. No power to instrument Check line voltage
1.2. Fuse on card blown Change fuse. See fig 48 on page 88, F2 and F3 for in­coming power and F1 for 24 V power on the card.
1.3. Bad power supply Check testpoint ”TP 15” and points 1.1 and 1.2 first. If still a problem change power supply. See section 5.2.5.
1.4. Internal power supply on
electronics card is broken
Change the electronics card. See section 5.2.4.
2. No information on display
2.1. No power to instrument See point 1 above
2.2. Display cable is loose Check cable. See section 5.2.6.
2.3. Display is broken Change the front panel. See section 5.2.6.
3. Display buttons don’t work.
3.1. Display/button cable is loose
Check cable. See section 5.2.6.
3.2. One ore more buttons is broken
Change front panel. See section 5.2.6.
4. No output signal 4.1. No power to instrument See point 1 above
4.2. Not measuring Check if value in display
4.3. Bad connections Check ”AOUT” connection
4.4. Bad analog out configuration
Check calibration. See section 5.2.7.
4.5. Analog output circuit is broken
Change the electronics card. See section 5.2.4.
5. Can’t change range from digital input
5.1. Bad connection Check connection. See section 5.2.9.
5.2. Configured for software
range setting
Change to hardware set­ting. See section 5.2.9.
6. The SPC does not work
6.1. Wrong program in the SPC
Switch off the SPC and change to correct sensor card and program.
5.2 - Service electronics
M2007/3en TCT-2301 Service instructions 85
©BTG 2004
6.2. Wrong SPC version SPC-1000/A is required (an upgraded version of SPC­1000 is needed as in sec­tion 4.1)
6.3. Not 250 Ω in 4-20mA
measure loop
If connected to the DCS, ensure the loop has a resis­tance of at least 250Ω. If not connected to DCS there has to be a resistor of 250connected between ”+AOUT” and ”-AOUT”.
6.4. The analog output signal
is not between 4-20 mA
See point 4. above.
6.5. Connector bad or loose Check the connector from SPC to electronics card. See section 5.2.6.
6.6. SPC/Hart communication
circuit is interrupted
Change the whole electron­ics card. See section 5.2.4.
7. Clear water calibra­tion can’t be made
7.1. Optical cable is not connected into electronics box.
Connect the probe optical connector. See section 3.3.3.
7.2. Optical cable is damaged Change the complete probe
and optical cable. See sec­tion 5.3.3.
7.3. LED circuits on card is interrupted.
Check testpoint ”TP3” with ”LED intens” set to 100% according to section 5.2.3. If values are no good change the electronics card as in section 5.2.4.
Symptom Probable cause Solution
5.2 - Service electronics
86 TCT-2301 Service instructions M2007/3en
©BTG 2004
5.2.2 Electronics card LED functions
1. DI1 Digital in 1 (not used)
2. DO2 Digital out 2 (not used)
3. DO1 Digital out 1. Lit when PCD-1000 function is activated. (Option)
4. PWR OK. Lit when power to the card is on
5. MODE. Indicate program function. – Slow flash (approx. 2 second interval). The program is running normal. – Fast flash (approx. 2 second interval). Indicates that the processor is running but doesn't have any software. Contact your BTG office. – No flash (lit or dark). Processor or program is not running.
6. ALARM. Lit when alarm function is activated.
7. IN2 and IN1 Indicates selected range, see below. (Lit=1 Dark=0)
8. Power switch on incoming power supply. Lit when 24 V DC is sup­plied to the card (before fuse F1).
75
4
Fig 47 LEDs on the
electronics card
1DI1 2DO2 3DO1 4 PWR OK 5MODE 6ALARM 7 IN2 and IN1 8 Power switch
8
2
1
6
3
IN1 IN2 Range
001
102
013
114
5.2 - Service electronics
M2007/3en TCT-2301 Service instructions 87
©BTG 2004
5.2.3 Function and testpoints of electronics card
TP Label Min. Max Unit Comment
F1 1 A Fuse for 24 V DC on the card. 250V 1A.
F2 1 A Fuse for input power. 250V 1A.
F3 1 A Fuse for input power. 250V 1A.
P10/P3+24 210 240 mA Current consumption at 24VDC. Measured at P10
when powered by 220VAC. Measured at P3/+24V when powered by 24VDC. Deviation (210 to 240 mA) depend mainly on status of LCD backlight, current loop output and DSP.
P3 +24/GND 23.5 24.5 V Output voltage of MOD1 24VDC power supply.
Measured at P3/+24 and P3/GND.
TP17 +7V 7 8 V System supply from DC-DC converter.
TP9 +5V 4.95 5.05 V Regulated system supply from DC-DC converter.
TP15 +24V 23 26 V Supply to current loop transmitter.
TP11 +12V 11.5 12.5 V Supply to external hand terminal.
TP13 GND-RS485 20 M Isolation between RS485 IF and other circuits
(measured at TP13 to P3,GND).
TP12 Vin-RS485 6 7 V Supply to RS485 IF from DC-DC converter
(measured at TP13 to TP12).
TP14 +5V-RS485 4.9 5.1 V Regulated supply to RS485 IF (measured at TP13
to TP14).
J3,1 GND-A/D 990 1010 W Isolation between optronics and system supply
(measured at P3/GND to J3,1).
TP5 Vin-A/D 6 7 V Supply to optronics from DC-DC converter
(measured at J3,1 to TP5).
TP6 Vdig-A/D 4.9 5.1 V Supply to digital circuitry on optronics module.
J3,6 +Vana-A/D 4.9 5.1 V Supply to analog circuitry on optronics module.
P6,2 A/D - input .99 1.01 V Offset voltage, negative A/D input.
P6,4 A/D + input 1.1 3.5 V Positive A/D input, value depends on current IR
intensity return. No light gives approximate 1.1 V.
TP3 LED SET 0 4 V IR intensity setpoint from D/A to IR output
regulator.
TP4 LED ACT 0 4 V Actual LED current (4V corresponds to 50 mA,
which is maximum) to obtain intensity set by LED SET (not directly comparable - LED SET is IR intensity and LED ACT is LED current).
TP8 2.5Vref 2.495 2.505 V A/D reference voltage.
P3, 3-4
+AOUT- 2 21 mA Analog output signal, depending on measured
value and span.
5.2 - Service electronics
88 TCT-2301 Service instructions M2007/3en
©BTG 2004
TP15 +24V
Fuse 250V 1A F1
P3 Conn. block
Fuse 250V 1A F2
Fuse 250V 1A F3
TP14 +5V
–RS485
TP13 GND
–RS485
TP9 +5V
TP12 Vin–RS485
TP4 LED–ACT
TP10 GND
TP3 LED–SET
TP11 +12V
TP2 DSP
P3
P13
P6
P10
P8
P5
COM+ DO1 DO2 DI1
TP17 +7V
TP16 1.23Vref
TP18 Q2 gate
TP1 TEST
P10 Conn. block
TP5 Vin–A/D
TP7 SDO
TP8 2.5Vref
J3,6 Vana–A/D
J3,1 GND–A/D
TP6 Vdig–A/D
Fig48 M
easurement points on
the electronics card
5.2 - Service electronics
M2007/3en TCT-2301 Service instructions 89
©BTG 2004
5.2.4 Changing the electronics card
If the circuit card is faulty the complete card must be replaced.
It is important to save all card information in the SPC before removing the card. All information about calibration and other settings are in the instrument. If no previous print out or backup is available, make one be­fore removing the card.
! Before replacing the card, disconnect line voltage to the card (it is
not sufficient to cut it off using the switch on the card).
All numbers refer to fig 49 on page 90.
1. Unscrew the lower cover on the box (2).
2. Check that all LEDs are off (12). If not, power is still on and should be disconnected.
3. Unscrew the optical holder (4) and loosen the optical connection. Put the protective cover over the optical connection. The cover is included with delivery.
4. Screw out the inner optical holder (3) that is screwed to the optical module on the circuit card.
5. Loosen all connection blocks (13) to which cables are connected. Screws need not be screwed out as the contact strip can be pulled out from the card.
6. Loosen the four screws on the front (1) and carefully lift out the front plate. Loosen the display switch (5) by pulling it straight out from the card. Also loosen the SPC-connection (11) by pulling it straight out from the card. It may be necessary to loosen the locking flap using a flat screw driver.
7. Unfasten the screws (7) holding the card to the box.
8. The card can now be removed – carefully, due to the tight space. Lift first the upper part with display connection (5).
Insert the new card by following the instructions above in reverse.
Ensure the optical connection is inserted in the correct position (see fig 23 on page 29).
Install the connection pipe (3) before screwing down the card in the box. Check that optical unit, intermediate washer, box wall and connection
Tools required:
Screwdriver (medium size), Star screwdriver (medium size), Block wrench 19 mm.
DANGER! High voltage within the electronics box. Connections may only be carried out by qualified personnel.
5.2 - Service electronics
90 TCT-2301 Service instructions M2007/3en
©BTG 2004
pipe butt against each other (important for grounding). See fig 23 on page
29.
When everything is mounted and power switched on, information will be shown on the display.
Fig 49 Changing the
electronics card
1 Screws for
upper cover (4x)
2 Screws for
lower cover (2x) 3 Connection pipe 4 Plastic nut 5 Display connection 6 12 V DC connection 7 Screws for electronics
card (13x) 8 Power supply 9 Screws for power
supply (4x)
10 85-240 V AC
connection
11 SPC-1000 connection 12 LEDs 13 External connection
blocks
4
5
7
11
12
13
2
3
1
8
6
10
9
Connect the optical connection pipe carefully as the card can easily be damaged. Hold the optical connection part off the card while the optical connection pipe is fastened.
5.2 - Service electronics
M2007/3en TCT-2301 Service instructions 91
©BTG 2004
The new card will not have any information about calibration and other settings.
Connect SPC with the reinstalled safety copy and push ,
to the instrument.
If no safety copy is available see section 4: Operating instructions.
! Perform a new water calibration as in section 5.3.2: Check clear
water value and clean of probe.
5.2.5 Change of power supply
If there is something wrong with the 24V-power supply, the whole part must be changed.
All numbers refer to fig 49 on page 90.
Before opening the electronics box, disconnect line voltage to the unit.
! Remove line voltage to the card (it is not sufficient to cut it off with
the switch on the card).
1. Unscrew lower cover (2).
2. Check that all LEDs are switched off (12). If not, the power is still on and should be disconnected.
3. Open the protection front cover.
4. Loosen the four screws on the front (1) and carefully lift out the front plate. Loosen the display switch (5) by pulling it straight out from the card. Also loosen the SPC-connection (11) by pulling it straight out from the card. It may be necessary to loosen the locking flap using a flat screw driver.
5. Loosen connectors (6 & 10) from the power supply card.
6. Unscrew the four tall nuts (9) with the socket wrench.
7. Lift the power supply unit (8) straight out.
To mount the new power supply unit, reverse the above procedures.
Tools required:
Star screwdriver (medium size), Socket wrench or Block wrench 7 mm
Send data
Send all
DANGER! High voltage within the electronics box. Connections may only be carried out by qualified personnel.
5.2 - Service electronics
92 TCT-2301 Service instructions M2007/3en
©BTG 2004
5.2.6 Changing the front panel
If there is something wrong with the display or buttons, the whole front cover has to be replaced.
! The instrument will work without display and buttons.
All numbers refer to fig 49 on page 90.
1. Unscrew lower cover (2).
2. Switch off power supply button (12).
3. Open the protection front cover.
4. Unscrew the upper cover (1) and lift the front carefully. Loosen dis­play contact (5) by lifting it carefully from the electronics card.
5. Unscrew the SPC-connector from the front cover.
To mount the new front cover, reverse the above procedures.
5.2.7 Check analog out
The display shows the analog output value. Use a digital multimeter to measure the current on connector ”P3” pin 3 and 4, labeled ”+AOUT–”.
If the measured value doesn’t correspond to the expected value, the cal­ibration has to be redone.
5.2.7.1 Calibrate analog out
All numbers refer to fig 49 on page 90.
1. Unscrew lower cover (2) on electronics box.
2. Connect the DMM set to mA (the smallest range for 20 mA) in series connection with the resistor on ”AOUT+”, connector ”P3 pin 3”.
3. From the SPC main menu, push and . Enter ”1632”
Tools required:
Star screwdriver (medium size), Block wrench 15 mm,
Tools required:
Star screwdriver (medium size), Screwdriver (medium size), Digital multimeter (DMM), SPC-1000/A
DANGER! High voltage within the electronics box. Connections may only be carried out by qualified personnel.
Misc. Factory
setting
5.2 - Service electronics
M2007/3en TCT-2301 Service instructions 93
©BTG 2004
and push .
4. Push key . Enter ”4”, push , and
then .
Read value on DMM. If the value is not correct, adjust using the keys
or , and then push and then to
obtain the new signal. Keep repeating these actions until exactly 4.00 mA is read on the DMM. (The test signal ”Test value” is active for approximate 15 seconds before the output signal is reset to the value generated by the instrument.)
5. To calibrate 20 mA, push key and enter 20. Push ,
and then .
6. Read value on DMM. If not correct repeat point 4 but use
or to adjust.
7. Disconnect DMM and replace cables and the lower cover.
5.2.8 Checking communication to and from SPC-1000/A
First make sure you are using an upgraded version of the hand terminal, type SPC-1000/A
Check that a resistor (249 ohms) is fitted between terminals ”AOUT+” and ”AOUT–”. If the signal is output to a DCS, the resistor must be po­sitioned in series with the DCS.
Connect SPC-1000 and switch on the hand terminal.
Go to main menu and select , .
If no error message is displayed, communication works. For an explanation of error message, see section 5.4.3: Software Fault feedback SPC-1000.
If there is a problem, check the current signal on analog out, measured in series between ”AOUT+” and ”AOUT–”. The current should be between 2 and 21 mA to get a working communication.
Enter/ menu
Test value
Enter/ menu
Send page
Send change
4 up 4 down Send
page
Send change
Test value
x
Enter/ menu
Send page
Send change
20 up 20 down
Read data
Yes
5.2 - Service electronics
94 TCT-2301 Service instructions M2007/3en
©BTG 2004
5.2.9 Checking range inputs
There are 4 different ranges to configure in the instrument. The ranges can be switched in 2 ways, using either digital input or software.
5.2.9.1 Change range input using digital input (hardware)
Create a short circuit between terminals ”IN1” and ”GND”. Check that LED ”IN1” is lit up and that ”Range” on the main menu on the hand ter­minal changes to ”2”.
Then create a short circuit between terminal ”IN2” and ”GND”. Check that LED ”IN2” is lit up and that ”Range” is changed to ”3”on the main menu.
If the range not is affected by the digital input see next section 5.2.9.2.
5.2.9.2 Change range using software
It is possible to change range using the SPC-1000. The instrument must be in Range Software (R.SW) Mode.
1. To change range, push , from main menu on SPC.
2. Push to change.
If range is chosen from digital input (hardware) the Mode button has R.HW. If range is chosen from SPC (software) the Mode button has R. SW. There is an other button at the right, Range: x, for range selection.
! If Mode button is in state R.SW the range cannot be set from digital
input.
Confi­gure
Tx. Spec.
Mode
Loading...