25
Message/Icon
The strip is not valid
Time off
Empty Battery
Verify the BSICare insertion
Grey Battery
Value lower than 20 mg/
dL (1,1 mmol/L)
Value higher than 600
mg/dL (33,3 mmol/L
Insufcient blood
volume, repeat the test
Error messages
Error
The strip is used or not valid
90 seconds without applying the blood sample have
exceeded
Flat battery
Wrong BSICare insertion
When starting the measurement, the battery icon stays
grey
The result is lower than the
reading threshold
The result is higher than the
reading threshold
Insufcient blood volume
Solution
Please replace the strip with a new one
Please disconnect the BSICare meter and repeat
the procedure. If the error remains, please contact
the customer service
Please check if the BSICare is correctly inserted
Replace the battery
If the error remains, please check the compatibility of the device or contact the customer service
If the result does not match your feelings, please
repeat the test. If you get “Lo” again, please contact your physician for further information
Wash your hands thoroughly and repeat the test
Please contact the physician immediately, repeat
the test in the hospital
Repeat the test using a new strip and apply a bigger drop of blood
Lo
Hi