6) GENERAL SPECIFICATIONS
Display: 3-5/6 digits 6000 counts + 3 digits 999 counts dual display LCD
Polarity: Automatic
Update Rate: 5 per second nominal;
Operating Temperature: 0C to 40C
Relative Humidity: Maximum relative humidity 80% for temperature up to 31C
decreasing linearly to 50% relative humidity at 40C
Pollution degree: 2
Storage Temperature: -20C to 60C, < 80% R.H. (with battery removed)
Altitude: Operating below 2000m
Temperature Coefficient (T.C.): nominal 0.15 x (specified accuracy)/ C @(0C --
18C or 28C -- 40C), or otherwise specified
Sensing: Average sensing for models 905s & 906s;True RMS for model 907s
Safety: Double insulation per IEC/UL/EN61010-1 Ed. 3.0, IEC/EN61010-2-030 Ed. 1.0,
IEC/EN61010-2-033 Ed. 1.0, IEC/UL/EN61010-031 Ed. 1.1 and CAN/CSA-C22.2 No.
61010-1-12 Ed. 3.0 to Category II 1000V, CAT III 600V and CAT IV 300V AC & DC
Transient Protection: 6kV (1.2/50s surge)
Terminals (to COM) Measurement Category:
V / A / mAA : Category II 1000V, CAT III 600V and CAT IV 300V AC & DC.
E.M.C. : Meets EN61326-1:2006 (EN55022, EN61000-3-2, EN61000-3-3,
EN61000-4-2, EN61000-4-3, EN61000-4-4, , EN61000-4-5, EN61000-4-6,
EN61000-4-8, EN61000-4-11)
In an RF field of 3V/m:
Capacitance function is not specified
Other function ranges:
Total Accuracy = Specified Accuracy + 100 digits
Performance above 3V/m is not specified
Overload Protection:
uA & mA : 0.4A/1000V DC/AC rms, IR 30kA, F fuse
A : 11A/1000V DC/AC rms, IR 20kA, F fuse
V : 1100V DC/AC rms
mV, & others : 1000V DC/AC rms
Power Supply: 1.5V AAA Size battery X 2
Power Consumption: 5.4mA typical
Low Battery: Below approx. 2.4V
APO Timing: Idle for 34 minutes
APO Consumption: 10A typical
Dimension: L186mm X W87mm X H35.5mm; L198mm X W97mm X H55mm with