Display: 3-5/6 digits 6000 counts.
Polarity: Automatic
Update Rate: 5 per second nominal;
Operating Temperature: -10C to 50C
Relative Humidity: Non condensing (≦10C)
≦90% RH (at 10~30C)
≦75% RH (at 30~40C)
≦45% RH (at 40~50C)
Pollution degree: 2
Storage Temperature: -20C to 60C, < 80% R.H. (with battery
removed)
Altitude: Operating below 2000m; Storage below 12000m
Temperature Coefficient: nominal 0.10 x (specified accuracy)/ C
@(-10C -- 18C or 28C -- 50C), or otherwise specified
Sensing: True RMS
Safety: Double insulation per UL/IEC/EN61010-1 Ed. 3.0,
IEC/EN61010-2-033 Ed. 1.0, CAN/CSA C22.2 No. 61010-1 Ed. 3.0,
IEC/EN61010-2-032 Ed. 3.0 & IEC/EN61010-031 Ed. 1.1 to CAT III
1000V and CAT IV 600V AC & DC
Transient Protection: 8.0kV (1.2/50s surge)
Overload Protections:
Current & Hz functions via jaws: 1000ADC/AAC rms at <400Hz
Voltage & 3-Phase Rotation functions via terminals: 1100VDC/VAC
rms
Other functions via terminals: 1000VDC/VAC rms
E.M.C.: Meets EN61326-1:2006 (EN55022, EN61000-3-2, EN61000-33, EN61000-4-2, EN61000-4-3, EN61000-4-4, , EN61000-4-5,
EN61000-4-6, EN61000-4-8, EN61000-4-11):
ACA, DCA, and DC+ACA Functions, in an RF field of 1V/m:
Total Accuracy = Specified Accuracy + 60 digits at around
200MHz~350MHz
Ohm and DCA Functions, in an RF field of 1V/m:
Total Accuracy = Specified Accuracy + 80 digits
Other Functions, in an RF field of 3V/m:
Total Accuracy = Specified Accuracy + 20 digits
Power Supply: 1.5V AA Size battery X 2
Power Consumption: Typical 13mA for Current functions of