This manual contains a description of the equipment, the vessel, installation, operating and
maintenance instruction for the BioFlo 3000 System manufactured by the New Brunswick
Scientific Co., Inc., 44 Talmadge Road, P.O. Box 4005, Edison, New Jersey 08818-4005,
U.S.A..
1.2DESCRIPTION OF EQUIPMENT
BioFlo 3000 is a versatile bioreactor that provides a fully equipped fermentation and cell
culture system in one compact package.* It can be employed for batch or continuous culture
with microprocessor control of pH, DO2, agitation, temperature, pump feed, antifoam and
vessel level.
* NOTE: This manual only contains a description of the vessel, installation, operating
and maintenance instructions for the fermentation system. For cell culture
system please read CelliGen Plus instruction manual (M1176-0050).
1.3DESCRIPTION OF VESSEL
The vessels are designed for working volumes of 1.25, 2.5 and 5.0 liters. It consists of a
stainless steel head plate, a flanged glass tube (thick walled) vessel body which is
detachable from the stainless steel bottom-dished head. The dished head is jacketed for
circulation of temperature controlled water. Four sterilizable polypropylene compression
ports are provided in the glass wall for the addition of antifoam and nutrients, as well as for
vessel overflow in continuous culture studies. Ports are provided in the headplate for:
Inoculation, base, and acid addition; a thermowell for a resistance temperature detector; a
foam probe; a sparger; a harvest tube; a sampling tube; an exhaust condenser;
dissolved oxygen and pH electrodes. The drive bearing housing is also located on the
headplate (Fig. 1, 2).
1.4AGITATION SYSTEM
A removable agitation servo motor located on the top of the bearing housing is connected
to the agitation shaft with a multi-jaw coupling. It can be easily disconnected while
autoclaving the vessel and replaced after sterilization. The motor will provide agitation
speed range of 20 to 1200 RPM (see important note in section 2.6). A PID control loop
holds the setting to within 1 RPM.
1.4.1When Dissolved Oxygen (D.O.) is in the PID mode and the agitation is in the D.O. mode
the agitation is cascade-connected to the D.O. so that the agitation speed is varied between
the minimum (set by user) and the maximum setpoint of agitation (set by user) to maintain
the set percentage of the D.O..
1.5TEMPERATURE CONTROL
The culture temperature may be selected in the range from 5C above coolant temperature to
80C ( .1C) and is controlled by a microprocessor based PI (Proportional and Integral)
controller. The media temperature is sensed by an RTD (Resistance Temperature Detector)
submerged in the thermowell (Fig. 3).
1.6AERATION
M1217-0050 4
Air and oxygen can be introduced into the medium through the ring sparger and the flow
rate is controlled by a needle valve located on the right hand side of the control cabinet
(Fig. 4).
Percentage of oxygen introduced into the medium can be determined manually by the user.
For high density culture, 100% of oxygen can be applied.
Oxygen can be introduced into the medium automatically through D.O. cascade system.
When DO2 loop is in PID mode and "2 GAS" is in "DO" mode, oxygen will be introduced
automatically into the inlet gas flow to maintain the set percentage of the D.O..
When DO2 loop is in PID mode, agitation loop is in DO mode and 2-Gas is in AgO2DO
mode, oxygen can also be introduced automatically into the medium after agitation speed
reaches its maximum setting value, yet the dissolved oxygen is still lower than the set point.
1.7pH CONTROL
pH is controlled in the range of 2.00-12.00 ( 0.01). The pH is sensed by a glass electrode.
Control is maintained by PID controller which operates two peristaltic pumps, connected to
acid and base addition ports (Fig. 5). Available as an option is the ability to provide control
with a deadband of 0.1 about a setpoint (see switch 2 settings at the end of section 2.6).
This will provide for only allowing the pumps to be activated when you are either 0.1 pH
units above or below the setpoint chosen.
1.8D.O. CONTROL
D.O. is controlled in the range of 5-95% ( 1%). It is sensed by the D.O. electrode and
control is maintained by the PID controller which can change the speed of agitation and the
percentage of oxygen in aeration.
The D.O. probe used on BioFlo 3000 can be Phoenix polarographic or Ingold
polarographic (Fig. 6, 7).
1.9FOAM CONTROL
Foam is controlled during batch fermentation by the antifoam probe which is located in the
headplate. The controller operates the antifoam addition pump that adds chemical
defoamer into the vessel (Fig. 8).
1.10EXHAUST SYSTEM
The exhaust gases pass into the exhaust condenser where moisture is removed and returned
to the vessel. The air remaining passes into the 0.2 m exhaust filter (Fig. 9).
1.11SAMPLING SYSTEM
1.11.1System I
This system has a sampler which is attached to a sampling tube extending to the lower
portion of the vessel. The sampler has a rubber suction bulb to facilitate collection of
representative samples without contamination. A 25mL screw cap container serves as a
reservoir (Fig. 10).
1.11.2System II
This system consists of a sample line and a peristaltic pump. Use "ON/OFF" mode of the
M1217-0050 5
pump to operate the pump (Fig. 11).
1.12SPECIFICATIONS
Vessela. Total Volume1.6 Liters/3.3 Liters/6.6 Liters
Temperaturea. IndicationDigital display in 0.1C increments.
Agitationa. DrivePermanent magnet servo motor with
Exhausta. Filter0.2 m interchangeable cartridge.
Aerationa. 2-Gas SystemAir, O2, delivered to ring sparger.
(setting range: 4C-80C. Temperature under
that of coolant can be obtained with optional
refrigeration unit.)
c. ControlPID control employing PWM of heater
& cooling water.
d. SensorPlatinum RTD.
high torque output.
b. Range20-1200 RPM (see important note in
section 2.6)
c. SensorOptical photoplastic disc 1000 lines/rev
with Quadrature output.
d. ControlMicroprocessor based with PID.
e. ImpellersSix-blade turbine impeller.
f. IndicationDigital display in 1 RPM increments.
b. CondenserStainless steel counterflow water cooled
in headplate.
b. Flowmeter0-10 SLPM mass flowmeter.
c. SpargerRing sparger
d. Inlet Filter0.2 m interchangeable cartridge.
b. Range2-12 pH
c. ControlPID
d. ProbepH Ingold
c. ProbePolarographic (Ingold or Phoenix)
d. ControlPID via superboard/Agit, 2-Gas
* NOTE: The numbers shown in the table are the flowrates when the control
mode of the feed pump is "ON". When the feed pump is set in other
modes (manual, base, acid, lvl 1, lvl 2 and lvl 3) the flowrates will
be 20% less than the numbers shown in the table.
1.13OPERATING CONTROLS AND DISPLAYS
The main operating controls of the BioFlo 3000 is the membrane keypad. There are twenty
keys on the keypad. The function of "Screen" key is to select the screen. There are three
screens which can be selected on the display. The first one is the "Master" screen, the other
two are: a "Calibration" and a "Gases" screen (Fig. 12, 14, 15, 16).
Besides above three screens, there is a mode selection screen which displays only after
the power switch is turned on. The "Mode Selection" (Fig. 13) will display on the
screen for 20 seconds. Operator can select cell culture or fermentation by pressing "1"
or "2" within this period of time. Then the screen will automatically turn to the "Master
Screen".
MODE SELECTION
1. CELL CULTURE
2. FERMENTATION
Your Selection: 2
Figure 13
On the top of each selective screen, three screen names are listed, the one in brackets is the
screen currently displaying, the other two screens are to be selected. Press "Screen" key, the
middle one will blink, this means that this screen is ready to be selected, press "Enter" key,
the selection of a new screen is done. If another screen is to be selected, don't press "Enter"
key but arrow () key first, then press "Enter" key.
The function of the "Alter" key is to select the loop and the mode of each loop. When
M1217-0050 7
making these selections, move the cursor to the loop or the mode to be changed, then press
the "Alter" key to get the desired selection and then press "Enter" key. If you have a loop
set already, you cannot change into that loop.
NOTE:The control mode of each loop are shown in the following table.
FEED 1 LVL 1
TOTO
LOOPAGIT 1TEMP 1pH 1DO2 1 AIRFEED 5 LVL 3
OFFOFFOFFOFFMONOFFOFF
PIDPRIMEPIDPIDMANADD
DOBASE
PIDACIDHRVST
CONTLVL1
MODELVL2
LVL3
ON
Figure 14
SCREEN<2-GAS>MASTERCALIBRATION
Oxygen EnrichmentAirO2
Output %75.524.5
ModeDO
Control GainsP0.25I2.50
Figure 15
NOTE:
1.Mode can be off, Do, AgO2DO or manual. DO mode can be selected when
DO2 loop is in PID mode. AgO2DO mode can be selected when DO2 loop
is in PID mode and agitation is in DO mode.
2.P (Proportional) I (Integral) value can be selected by the operator. Move
cursor to O2, press code "9XXXX", then P I value can be changed.*
* NOTE: P I values have been set in the factory before shipping.
Under normal conditions, it is not necessary to reset
these values.
SCREEN<CALIBRATION> MASTERGASES
M1217-0050 8
Calibration
Value ZeroSpanFunction
pH7.007.004.00Read
D.O.1000100Read
NOTE:
1.Change function to zero, calibrate zero.
2.Change function to span, calibrate span.
Figure 16
There is a recorder output connector on the left hand side of the control cabinet. An
external recorder with 0-1 volt high impedance input may be connected to the output.
An RS232/422 computer interface has been provided. A 25 pin "D" connector is located
on the left hand of the control cabinet. An advanced fermentation software program is
available which will enable the operator to interface with an IBM or suitably compatible
MS-DOS computer. It will enable one to establish or change the set points for temp, pH,
D.O., agitation speed and pump flow rate. The operator may also read and log the current
values of those parameters (temp, pH, DO, air flow, pump flow rate, levels and agitation)
which are monitored. By having the data available in this form it can be stored, plotted
and, afterwards, transferred to other commonly available programs, manipulated and
analyzed in various ways.
M1217-0050 9
CHAPTER 2 - INSTALLATION
2.1INSPECTION
Unpack the BioFlo 3000 and carefully inspect for any apparent damage which may have
occurred during transit. A component list is enclosed. Check the component list against
the parts shipped. Report any missing parts to New Brunswick Scientific Co. Report any
obvious damage to the carrier and to New Brunswick Scientific Co., Inc..
2.2UTILITY REQUIREMENTS
Water: 20 PSIG max., 50m filtration
Gas:10 PSIG max.
Electrical Requirements: 100/120 Volt - 50/60 Hz - 10 Amp
220/240 Volt - 50/60 Hz - 6 Amp
NOTE: See Fig. 24 for fuse replacing.
2.3INSTALLATION OF CONSOLE
A. Position the BioFlo 3000 console on a firm level surface in an area where services are
readily available.
B. Level the horizontal surface of the base with four leveling glides if necessary.
C. Check the specifications plate on the rear of the unit. Connect the line cord to the
console and plug the line cord into a suitable electrical outlet (Fig. 17).
2.4POWER SWITCH
The main power switch which controls the power to the system is located on the left hand
rear of the console.
Prior to turning on the main switch make sure that:
A. The input water hose is connected at the rear of the unit and the water supply is
turned "ON". The drain line must also be connected.
B. The vessel base is in place and the quick connect plastic lines are connected to
the vessel in front.
2.5Recorder Output:0.0 - 1.0 Volt DC
PINS:1, 2 TEMP
3, 4 pH
5, 6 D.O.
7, 8 AGITATION
Grounds are pins 2, 4, 6 and 8.
M1217-0050 10
PIN #PIN FUNCTIONFOR 0-1 VDC OUTPUT
1TEMPERATURE +0-100C
2TEMPERATURE -
3pH +2-12 pH
4pH -
5D.O. +0-200%
6D.O. -
7AGITATION +0-1500 RPM
8AGITATION -
2.6RS232/422 COMPUTER INTERFACE
The RS232/422 Computer Interface is a 25 pin "D" type connector mounted on the left
hand side of the electrical cabinet. It is by means of this connector that an external
computer may communicate with and control the BioFlo 3000 Fermentor. Pin
designations are as follows:
PIN #PIN FUNCTIONPIN #PIN FUNCTION
7JUMPER23ICTS
(ONLY REQ.FOR
21RS422 CONFIG.)2TXD
3RXD4RTS
5CTS13ITXD
12IRXD25ITXD
24IRXD10IRTS
11ICTS22IRTS
Unless requested otherwise the baud rate is factory selected at 9600 and the connector is
configured as an RS232 port: i.e. no jumper between pin #7 and pin #21. Furthermore,
the machine has been assigned an address location of 0.
In addition to the two available connector configurations, there are four baud rates and
16 addresses which may be selected by changing dip switch settings. The switches to
effect both these types of changes are internally mounted in the same array on the
superboard.
If a change is requested, first turn the power off, then open the back door, the superboard
is located on the top of the shelf. The single array for both functions is an assembly of
eight dip switches labeled SW1.
To select one of the four available baud rates the first two switches in the array should be
set as shown in the following table:
BAUDS1-1S1-2
M1217-0050 11
1200ONON
2400OFFON
4800ONOFF
* 9600OFFOFF
Switch positions 3 and 4 are used as follows:
* Multidrop (8 data bits) and even parity checkONON
Multidrop (8 data bits) and no parity checkOFFON
No Multidrop (7 data bits) and even parity checkONOFF
No Multidrop (7 data bits) and no parity check OFFOFF
To select one of the sixteen addresses for the machine it is necessary to set the last four
switches in accordance with the following table:
*SW1 of the machine has been set as "OFF-OFF-ON-ON-ON-ON-ON-ON".
The switch settings of SW2 are:
SW2-1pH ControlpH Control
without dead bandwith dead band
SW2-2AgitationAgitation
20-1000 RPM20-1200 RPM
SW2-3 and SW2-4 can be set either "ON" or "OFF".
SW1-3 SW1-4
ONOFF
M1217-0050 12
IMPORTANT NOTE:
SW2-1 and SW2-2 have been set at "ON" position in the factory before shipping.
If 1200 RPM agitation speed is required, please consult with New Brunswick
Scientific Co. before taking action.
M1217-0050 13
CHAPTER 3 - PREPARATION & OPERATION
(For preparation and operation of Cell Culture, please read CelliGen Plus Manual
M1176-0050 - included with conversion kit - see p. 23)
3.1CLEANING OF VESSEL
A. Fill the vessel with a mild detergent and water solution. Let stand for one hour then
scour thoroughly. Use a brush on both inside and outside surfaces.
B. Drain the vessel and rinse several times with tap water. Repeat rinsing with distilled
water and let dry.
C. Sterilization (see 3.8).
3.2VESSEL ASSEMBLY
A. Slide impellers on the shaft of the bearing housing and clamp them down. Lower
impeller should be positioned about 1/4 inch above the bottom of the baffle. Upper
impeller one to one and half impeller diameters above lower impeller (Fig. 18).
* Lubricate vessel O-ring with silicone grease, position glass tube reactor and
secure clamping screws, perform the same for head plate. DO NOT OVERTIGHTEN.
Working From the Inside of the Coverplate.
B. Insert the sparger tube in the sparger port, (Fig. 19).
C. Insert the harvest tube in the harvest port, (Fig. 20).
Working From the Outside of the Coverplate.
D. Insert the thermowell tube in the temperature control port, (Fig. 3).
E. Insert the sampler assembly into the sample port (Fig. 10).
F. Install foam probe in the head plate, (Fig. 8).
G. Install the baffle assembly inside of the glass jar by pressing the two edged baffles
together.
H. Install four polypropylene ports or plugs in the wall of the glass jar. Be sure to use
a blind plug (without hole) when a port is not in use, (Fig. 21).
Place head plate on the flange of the vessel and lock it to the clamping ring with
knurled screws (see operation tips). Install pH and DO probes (coat probe with
glycerin before insertion into head plate). Prior to installation, probes should be
well prepared (see sec. 3.3, 3.4, 3.5 and 3.7).
NOTE: To avoid damage to the probes make sure that no interference exists between
the probes and the baffle assembly.
Place the vessel on the three pins located on the base of the cabinet.
Position the motor assembly on top of the bearing housing locating pin. Connect the motor
cable to the receptacle on the face of the cabinet.
M1217-0050 14
Connect cables from all probes to their respective sockets on the face of the cabinet.
Ground lead from the antifoam socket on the face of the cabinet is to be connected
to the pin in the headplate.
Connect exhaust condenser to the exhaust condenser port. Connect the exhaust filter on
the top of the condenser with flexible tubing.
Slide 2" long, 0.25"ID silicone tubing on the top of the sparger tube, then connect air filter
to it. One side of the filter is to be connected to the hose barb in the face of the cabinet with
flexible tubing.
3.3pH PROBE PREPARATION
Inspect probe for possible shipping damage. If damage is observed notify the Service
Department of the New Brunswick Scientific Co., immediately.
Check the level of the reference electrolyte. It should be about 1cm below the filling
orifice, which is closed with a rubber "T" stopper. To add reference electrolyte, take
the filling pipette (P0740-4820) and fill it with Viscolyte B (P0860-0130) Electrolyte.
Check the electrode tip for trapped air bubbles. To remove any air bubbles hold the
electrode upright and shake gently.
NOTE: The two chambers are filled with the same reference electrolyte. The
total volume of reference electrolyte held by the electrode is
approximately 30mL. During normal operation the two rubber
stoppers are to be removed.
3.4INSTALLATION OF THE pH ELECTRODE
CAUTION:
WEAR PROTECTIVE GLOVES WHEN INSTALLING ANY GLASS ELECTRODES.
1. Apply a small amount of silicone grease or glycerol to the electrode body.
2. Install the pH electrode as shown in Figure 5.
3.5PROBE MAINTENANCE AND STORAGE
pH
1. Check the level of the filling solution. It should be about 1cm below the filling orifice.
To add solution, see Section 3.3.
2. Check for any trapped air bubbles in the electrode's tip to remove bubbles, hold
electrode upright and shake electrode gently.
3. The probe should be stored standing upright. The electrode tip should be immersed
in the solution of 3 molar KC1 or a buffer solution between pH 4 and pH 7. The two
rubber "T" stoppers should be inserted. At no time should the electrode be allowed
to rest on the tip.
3.6pH ELECTRODE CALIBRATION
M1217-0050 15
NOTE: pH electrode is calibrated before autoclaving vessel.
1. Connect electrode to the pH connector using the appropriate cable.
2. Turn ON the main power switch.
3. Display calibration screen.
NOTE: The pH measuring system is calibrated using two external buffer
solution of known pH.
4. Immerse pH electrode into pH 7.00 buffer solution and allow a few minutes for the
system to equilibrate.
5. Set the pH function zero.
6. Set the display to read 7.00.
7. Rinse the pH electrode with distilled water.
8. Immerse pH electrode into a second pH buffer solution which is several pH units
above or below pH 7.00 (eg. 4.00) and allow a few minutes for the system to
equilibrate.
9. Set the pH function "Span".
10. Set the display to read the value of the second buffer solution.
11. Repeat steps 5-10 using the same buffer solutions.
NOTE: The pH calibration should be checked after autoclaving immediately prior
to inoculation. This is performed by taking a sample from the vessel and
comparing the value of pH displayed on the screen with that of an external
pH meter. Any discrepancy should be adjusted with the function set to
"Zero".
3.7OPERATION OF DISSOLVED OXYGEN (DO) PROBE
1. Remove protective cap from electrode end. The membrane is delicate and care must
be exercised to prevent accidental damage. Never rest probe on membrane.
2. To insure stable output, the probe should be subjected to two or three sterilization
(autoclaving) cycles prior to use. The probe will be operable after the second cycle
but will be more stable with additional sterilizations. The shorting plug should be
installed on the probe during autoclaving or sterilization.
3. Install probe into vessel head plate assembly (Fig. 6,7). If an Ingold probe is to be
used, wrap the adapter threads with teflon tape, screw into headplate and tighten
with a wrench. Carefully insert the probe into the adapter. Finger tighten enough
to compress the o-ring to insure a tight seal.
3.8STERILIZATION PROCEDURE
Remove the motor drive from the top of the vessel and place it on the motor mount at the
top of the cabinet. Place the housing cover on the top of the bearing housing before
sterilization (Fig. 22).
Disconnect air line of the inlet filter side.
Disconnect all probes and remove probe cables.
Rubber T-stoppers should be in place in the pH probe during autoclaving. (Rubber
bands can be placed around them to prevent blow out during sterilization.)
Remove the sampler rubber bulb then insert glass wool into sampler port and close
M1217-0050 16
sampler valve.
Sterilize the complete assembly consisting of jar, headplate and components of
headplate by inserting in an autoclave.
NOTE: For continuous culture, vessel should not be sterilized empty. Use at least
100 mL of sterilized water. Probe tips must be moist during sterilization.
After autoclaving medium will be brought from the medium reservoir,
sterilized separately. For batch fermentation the medium is sterilized
with the vessel.
*Autoclave at a temperature of 121C at 15 PSIG for 25 minutes.
CAUTION:
WHEN AUTOCLAVING THE GLASS VESSEL. IT MUST BE VENTED AT ALL TIMES.
DURING STERILIZATION OF THE GLASS REACTOR VESSEL, CONTENTS, AND
ELECTRODES IT IS IMPORTANT THAT THE PRESSURE BUILT UP IN THE AUTOCLAVE IS
RELEASED ONLY WHEN TEMPERATURE HAS DROPPED BELOW 90C. SLOW EXHAUST
IS REQUIRED.
CAUTION:
AT NO TIME SHOULD YOU SUBJECT THE PROBE TO CONDITIONS WHICH WOULD
CAUSE THE ELECTROLYTE TO BOIL. PARTICULAR CARE SHOULD BE EXERCISED
DURING THE COOLING CYCLE FOLLOWING AUTOCLAVING WHEN LOCALIZED
BOILING TENDS TO OCCUR. EITHER MAINTAIN A SLOW COOLING RATE OR
PRESSURIZE THE AUTOCLAVE DURING THIS PERIOD. CONSULT YOUR AUTOCLAVE
MANUFACTURER FOR INFORMATION REGARDING A PRESSURE BALANCING FEATURE
AND PROPER AUTOCLAVING TECHNIQUE TO ELIMINATE LOCALIZED BOILING
DURING THE COOLING CYCLE.
M1217-0050 17
If after autoclaving, most of the liquid has exhausted from the vessel, the autoclave is
exhausting too quickly. Adjust the autoclave to exhaust more slowly. Attach a piece
of tubing with some non-absorbent material such as glass wool or non-absorbent
cotton and some foil wrapped on the ends to one of the addition ports. This helps the
vessel to vent more easily during autoclaving. Immediately crimp the foil and close
off the vent tubing to maintain sterility.
3.9DISSOLVED OXYGEN ELECTRODE (D.O.) CALIBRATION
NOTE: D.O. electrode is calibrated after autoclaving vessel.
IMPORTANT NOTE:
When the system is operated for the first time, or when the electrode has been
disconnected from the voltage source (amplifier or polarization module) for
longer than 5 to 10 minutes, the electrode must be connected to the operating
O2 amplifier for polarization purpose prior to calibration.
The electrode is polarized and ready for operation after six hours of
polarization time.
1. There are two methods of obtaining zero on the BioFlo 3000. Use either Method 1
or Method 2.
Method 1
a.Remove the D.O. electrode cable from the D.O. electrode.
b. Display calibration screen.
c.Set the D.O. function to "Zero".
d. Set the display to read zero by setting "Zero" to 0.
e.Re-connect the D.O. electrode cable to the D.O. electrode.
Method 2
a.Connect the D.O. electrode cable to the D.O. electrode.
b. Set agitation speed to 500 RPM.
c.Sparge nitrogen into the vessel via the filter on the head plate until the D.O.
display is stable for approximately 10 minutes (this may take up to 30
minutes).
d. Display calibration screen.
e.Set the function to "Zero".
f.Set the display to read zero by setting "Zero" to 0.
2. Setting the "Span".
a.Set the agitation rate of 500 RPM and agitation mode PID.
b. Vigorously sparge air or oxygen into the vessel via the filter on the head plate
until the display is stable for approximately 10 minutes (this may take up to
30 minutes).
c.Set the function to "Span".
d. Set the display to read 100 by setting "Span" to 100.
3.10PREPARATION FOR OPERATION
A. Position and secure the vessel on BioFlo 3000 console. Connect the heat exchanger
and exhaust condenser.
B. Carefully place the DC servo motor on the vessel assembly.
M1217-0050 18
C. Add glycerin to the thermowell and insert (RTD) temperature probe.
D. Turn on the water valve, adjust water pressure to 20 PSI.
E. Connect air and O2 to the unit, adjust gas pressure to 10 PSI.
CAUTION:
NEVER ATTACH TUBINGS MADE OF HARD MATERIAL, SUCH AS TEFLON, DIRECTLY
TO THE PLASTIC HOSE BARBS ON THE GAS ENTRY.
IN CASE HARD MATERIAL TUBING IS GOING TO BE USED, ATTACH A SHORT SOFT
TUBING (NBS P/N P0740-2430) BETWEEN THE HOSE BARB AND THE HARD TUBING.
F. Turn on the power switch.
G. Select fermentation mode by pressing key "2".
H. On master screen, set temp loop control mode to "Prime" for 1 minute.
I.Set agitation to desired speed and then set mode to PID.
J.Set temp to the desired working temperature set mode to PID.
K. Remove plugs and tape (if any) and shorting cap from the pH probe and connect the
pH cable to the pH probe.
L. Remove the protecting cap from the D.O. probe and connect the D.O. cable to the
D.O. probe.
M. When the vessel reaches desired working temperature, calibrate the D.O. probe.
N. Set pH and D.O. to the desired set points and set pH and D.O. mode to PID.
O. Set agitation to D.O. if applying D.O. cascade system (see 3.11).
P. Set 2-Gas mode to D.O. or AgO2DO, if applying oxygen enrichment/D.O. cascade
system (see 3.11.2). Otherwise, set 2-Gas mode to manual.
Q. Adjust air flow rate to the desired value.
3.11D.O. CASCADE SYSTEM
The system is designed to control D.O. by PID control of agitation speed and oxygen
output. When D.O. actual value is above the D.O. setpoint, the agitation speed will
automatically decrease until the D.O. setpoint is reached.
As the D.O. level drops below the setpoint, there are three ways to pick it up.
M1217-0050 19
3.11.1Agitation/D.O. Cascade System
To set up, proceed as follows:
a.Set agitation and D.O. loop to PID mode.
b. Set agitation to the minimum RPM you wish for D.O. control.
c.Change agitation mode to D.O..
d. Set agitation speed to maximum RPM you wish for D.O. control.
e.D.O. is now controlled by regulation of the agitation speed between the limits
that have been set (minimum/maximum setpoints).
3.11.2Oxygen Enrichment/D.O. Cascade System
To setup, proceed as follows:
a.Set D.O. loop to PID mode.
b. Set 2-Gas to D.O. mode.
c.Select proper PI value*
d. D.O. is now controlled by regulation of the percentage of oxygen output between
0 and 100.
* NOTE: P.I. values have been set in the factory before shipping. Under normal
conditions, it is not necessary to reset these values.
3.11.3Agitation/Oxygen/D.O. Cascade System
To setup, proceed as follows:
a.Set agitation and D.O. loop to PID mode.
b. Set agitation to the minimum RPM you wish for D.O. control.
c.Change agitation mode to D.O.
d. Set agitation speed to maximum RPM you wish for D.O. control.
e.Set 2-Gas to "AgO2DO" mode.
f.Select proper PI value (see note of 3.11.2.C).
g. D.O. is now controlled by the regulation of agitation speed first. After the maximum
set speed is reached, then D.O. is controlled by the regulation of the percentage of
oxygen output.
3.12SAMPLING PROCEDURE
To sample the culture proceed as follows: (Fig. 10)
1. Check to be sure that the sample bottle is loose not sealed against sampler gasket.
2. Close the valve on the sampler tube, if open.
3. Squeeze the bulb and then see that sample bottle is sealed against sampler gasket.
Open valve and obtain desired volume of sample. Close valve.
4. Remove bottle from sampler. Place the cap from a new bottle on the bottle containing
sample and install the new bottle in the sampler and make sure that the sample bottle
is firmly sealed against the sampler gasket. Use aseptic techniques.
5. Repeat step 1 through 4 until desired number of samples are taken.
3.13SHUT-DOWN PROCEDURES
To shut-down the system proceed as follows:
1. Shut-off gas flow.
M1217-0050 20
2. Set the agitation, temperature to "OFF".
3. Turn off the power.
4. If the system is not to be used for several days, disconnect power plug. Remove and
clean vessel, and associated components as outlined in previous steps.
NOTE: DO NOT WASH THE FILTERS OR GET THEM WET.
3.14OPERATION TIPS
A. Glass Vessel Assembly
Recommendations for prevention of cracking glass during assembly and autoclaving:
Cracking of glass due to overtightening of assembly screws will occur during
tightening, not during autoclaving.
Therefore,
1. Prior to autoclaving, tighten screws "finger tight". If a wrench is applied
at this point, turn each nut 1/2 revolution not to exceed 1 revolution.
2. Place unit in autoclaving making certain that the exhaust filter(s) is not
wet or clogged. Also, loosen the inoculation diaphragm cap to achieve
further ventilation.
3. After autoclaving, tighten inoculation port cap. Further tightening of
nuts can be done at this time with air flowing through vessel.
NOTE: If you use a wrench in step 1, DO NOT TIGHTEN AGAIN in step 3.
(One revolution maximum after finger tight.)
B. Exhaust Condenser/Exhaust Filters
The inner assembly of condenser can be removed for cleaning.
Pass warm water and detergent through the top of the condenser, (NOT through
quick disconnects), twice.
Run clear water through once.
Blow out with air. Autoclave.
Clean exhaust condenser after each fermentation. This is most critical when
operating as a chemostat for protracted fermentation times.
C. Install a Double Filter System
Double exhaust and double inlet filters are recommended
Attach a "Y" fitting to the top of the condenser with a piece of tubing. Attach
an exhaust filter on each branch. This allows you the flexibility to exchange
sterilized filters during a run should one filter become clogged, (pinch off
unused line with a clamp).
D. OTR (Oxygen Transfer Rate) Calculation
OTR = 30,000 m
V T m.moles/L/hr
Where m = number of moles of Sodium Sulfite Na2SO3
Weight of 1 mole of Na2SO3 = 126.04 gm
V = vessel working volume L
T = time taken from DO curve at two points of 50% DO min.
Procedures:
M1217-0050 21
Operate machine at 37C, 1000 RPM, air flow = (1/2 - 1 V) LPM.
Calibrate DO to 100 and allow it to stabilize.
Add Na2SO3 (with a small amount about 5gm of Cu SO4).
The DO reading drops very quickly. At DO = 50, start counting time.
When DO returns to 50, stop counting time.
M1217-0050 22
CHAPTER 4 - MAINTENANCE
4.1GENERAL
Preventive maintenance is performed to keep equipment in proper working condition.
When periodically performed, it will result in longer life for the equipment and reduce
time lost due to equipment failure.
4.2CONSOLE CLEANING
At least once a month, clean all metal parts of unit. Use a damp cloth moistened with water
or mild detergent. If a detergent is used, remove all excess by clean water washing.
4.3PERIODIC INSPECTION
At three month intervals perform the following checks and inspections with all switches
"OFF" and incoming power disconnected.
1. Check the fuse(s) for clean contact.
2. Check all controls and accessible items (switches, knobs, fuse holders, screws,
nuts and bolts) to make sure they are properly tightened. Tighten any item which
is loose.
3. Check that all controls are free of dust and operate easily.
4. Check that all O-rings in the headplate and impeller are intact. Replace those that
are not.
4.4VESSEL AND TUBING CLEANING
After each run clean the vessel, headplate and associated parts. The tubings and filters
should be replaced.
4.5PROBE MAINTENANCE AND STORAGE
D.O. PROBE
To clean D.O. probe, use soft facial tissue.
Check the D.O. probes teflon membrane. (Make sure there is no wrinkling or
punctures.)
Check the vent tubes, make sure they are clean.
The probe should be stored standing upright with the shorting plug in place and
the membrane isolated from the air environment. At no time should the probe be
allowed to rest on the electrode's membrane.
As soon as the probe is filled with electrolyte, it begins aging and its output can be
expected to decline gradually. Depending on the operating conditions, a filled probe
should last through 15 to 30 sterilization cycles, and/or 6 months.
The probe should not be stored at a temperature greater than 60C.
pH PROBE
M1217-0050 23
Check the level of the filling solution. It should be about 1cm below the filling
orifice. To add solution, see section 3.3.
The probe should be stored standing upright. The electrode tip should be immersed
in a solution of 3-Molar KCL or a Buffer Solution between pH 4 and pH 7. The two
rubber "T" stoppers should be inserted. At no time should the electrode be allowed
to rest on the tip.
NOTE: The rubber T-stoppers should be left in place during autoclaving,
but removed during regular usage.
Certain types of Mycelia can grow on the liquid junction. These growths can be
prevented from forming by the addition of a few drops of Formaldehyde (up to 1%)
to the 3-Molar KCL solution.
4.6THE AGITATOR BEARING HOUSING (Fig. 23)
Every 3-6 months, the ball bearings and the shaft seals in the bearing housing should be
checked and cleaned. Replace the worn out bearings and shaft seals.
4.7REPLACEMENT PART LIST
Vessel (1.25L) AssemblyM1169-2000
Vessel (2.5L) AssemblyM1169-2001
Vessel (5L) AssemblyM1169-2003
Inlet FilterP0200-0491 (37mm Disc)
Exhaust FilterP0200-0490 (50mm Disc)
RTD AssemblyM1169-8002
pH Probe (1.25L, 2.5L)P0720-5325
pH Probe Cable (1.25L, 2.5L)P0720-2095
pH Probe (5L)P0720-5021
pH Probe Cable P0720-2095
pH Electrode K9 CapP0720-5317
pH Probe ElectrolyteP0860-0130
DO Probe (1.25L) (Phoenix)P0720-5440
DO Probe (2.5L) (Phoenix)P0720-5450
DO Probe (5L) (Phoenix)P0720-5460
DO Probe Cable (Phoenix)P0720-2202
DO Probe (1.25L) (Ingold)P0720-5560
DO Probe (2.5L) (Ingold)P0720-5561
DO Probe (5L) (Ingold)P0720-5561
DO Cable (Ingold)P0720-2202
DO Service Kit, IngoldP0720-5569
DO Electrode Cap, IngoldP0720-5567
Teflon Washer for Ingold DO ProbeP0100-9780
O-Ring for Ingold DO ProbeP0280-9163
Seal Washer for Ingold DO ProbeM1016-0890
Antifoam ProbeF5-137
Hose Connect PlugM1151-9518
Hose Connect OpenM1151-9519
Seal, Vessel Side PortM1151-9419
Nut, Vessel Side PortM1151-9420
Fuse, 10 Amp, 115VEF-143
Fuse, 6 Amp, 220VP0380-3140
Glass Vessel 1 1/4LM1151-9904
Glass Vessel 2 1/2LM1152-9903