The information in this manual may be altered without notice.
BRUKER BIOSPIN accepts no responsibility for actions taken as a result of use of
this manual. BRUKER BIOSPIN accepts no liability for any mistakes contained in
the manual, leading to coincidental damage, whether during installation or
operation of the instrument. Unauthorised reproduction of manual contents,
without written permission from the publishers, or translation into an other
language, either the entire manual or a part of it, is forbidden.
This manual describes the units as they are at the date of printing. On request, the
manufacturer shall supply circuit diagrams, lists of components, descriptions,
calibrating instructions and any other information for use by qualified personnel of
the user, in charge of repairing the parts of the unit which have been stated by the
manufacturer to be "repairable". Such supply shall in no event constitute
permission to modify or repair the units or approval of the same.
All rights reserved for the units, circuits, processes and appellations mentioned
herein.
This unit is not designed for any type of use which is not specifically described in
this manual. Such use may be hazardous.
This document contains technical information about the "BCU-Xtreme External
Temperature Controller" (P/N:W1215349).
The temperature of the gas delivered by the BCU-Xtreme (P/N:W1212749) may
be to too cold for some experiments or some NMR probes, requiring strong
heating of the cold gas. The new External Temperature Controller Device
(P/N:W1215349) was designed to gradually adjust the outlet temperature of the
BCU-Xtreme from -80°C to nearly 0°C. It is compatible with all BCU-Xtreme
manufactured until 2010.
Safety1.2
The device is slightly magnetic. The controller must always be handled and
installed at a safe distance from a magnet.
Controller Installation1.3
The controller is delivered with several meters of 6mm and 8mm HDPE plastic
hoses which are connected to the BCU-Xtreme. Optimal controller placement is
near the BCU-Xtreme or on top of the cooling unit.
Main gas inlet
Temperature
control knob
Figure 1.1. External temperature controller device top view
Technical Manual Version 001BRUKER BIOSPIN5 (13)
Page 6
Device Information
Variable outlet temperature
-80°C to approx. 0°C
Probe adapter
By-pass inlet
MAIN GAS
INLET
Multi-turn knob
for outlet
temperature
control
To probe adapter
To BCU gas in
10%-100% of
inlet gas flow
W1215349
To By-pass out
BCU-XtremeT° Control
GAS IN
BY-PASS OUT
90%-0% of inlet
gas flow
Figure 1.2. Connections of Controller
Hoses Connections1.4
1. Connect the dry gas supply hose from the temperature control unit BVT or
MAS unit to the "MAIN GAS INLET" port of the device with 8mm tubing.
2. Connect a piece of 8mm tubing between port "To BCU GAS IN" and "GAS IN"
of the BCU-Xtreme.
3. Connect a 6mm tubing between port "To By-PASS" and "to BY-PASS OUT"
of BCU-Xtreme.
4. Connect a 6mm tubing between port "To PROBE ADAPTER" and "PROBE ADAPTER" of BCU-Xtreme.
Check all hoses connections for leaks.
Device Operation1.5
The outlet temperature control knob is located on the lid.
The knob allows you to gradually adjust the outlet temperature from -80°C to
approximately to 0°C. The knob actuates a couple of valves that divide the inlet
gas flow between the BCU-Xtreme and the adapter located at the end of the
transfer line. The cold gas mixed with room gas at temperature gives the desired
result.
The gas flow delivered to BCU-Xtreme can be varied between 10% to 100% of
the device inlet gas flow and the gas flow to the BCU-Xtreme adapter varies
inversely from 90% to 0%.
6 (13)BRUKER BIOSPINTechnical Manual Version 001
Page 7
Getting Started
Warning : A minimal gas flow (> 1l/min) must always pass through the BCUXtreme, otherwise an error occurs and the compressors will be stopped.
When the temperature control knob is turned completely towards "WARMER"
position, there is still a fraction of the gas flow that passes through the BCUXtreme. However, if the gas flow through the BCU-Xtreme is too low it causes the
shutdown of the compressors. If this happens, turn slightly back the knob
towards position "COLDER" to increase the gas flow through the BCU-Xtreme.
When the knob is fully in "WARMER" or "COLDER" position do not force or
damage may occur to the gas mixer valves.
Getting Started1.6
It is recommended to place the temperature knob in the intermediate position
between "COLDER" and "WARMER" and then start the operation of the BCUXtreme to avoid startup problems (low gas flow).
Wait 20-30 minutes until the cooling power is steady. The outlet temperature can
be changed with the control knob after the initial cooldown phase.
If the BCU-Xtreme is used even after a longer stop, there is no need to readjust
the knob position before restarting the unit.
Technical Specifications1.7
Table 1.1.Device specification
DimensionsHeight : 60 mm
Width : 160 mm
Length : 240 mm
Weight800 grs
Inlet gas flow10 l/min minimal up to 50 l/minIf the inlet gas flow is too low,
Gas connections8 mm and 6 mmQuick coupling type
Inlet gas pressure5 bar max.Dry air or dry nitrogen gas
Temperature control range-80 °C to 0 °C approx.With BCU-Xtreme (P/N:W1212749)
Stainless steel enclosure
the BCU-Xtreme may be shutdown
Technical Manual Version 001BRUKER BIOSPIN7 (13)
Page 8
Device Information
8 (13)BRUKER BIOSPINTechnical Manual Version 001
Page 9
Figures
1 Device Information5
Figure 1.1. External temperature controller device top view ..................................... 5
Figure 1.2. Connections of Controller ...................................................................... 6