Bruker RS Service Manual

RS
Service Manual
NMR Spectroscopy
Copyright © by Bruker Corporation
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form, or by any means without the prior consent of the publisher. Product names used are trademarks or registered trademarks of their respective holders.
© May 2011: Bruker Corporation
Fällanden, Switzerland
ZTKS0156 / Z31820 / Rev.: 03
For further technical assistance on the NMR magnet system, please do not hesitate to contact your nearest BRUKER dealer or contact us directly at:
BRUKER Corporation
Industriestrasse 26
CH - 8117 Fällanden
Switzerland
Phone:+ 41 44 825 91 11
Fax: + 41 44 825 96 96
E-mail:service@bruker.ch
Internet:www.bruker.com
0 Contact ...........................................................................................................9
1 Introduction..................................................................................................11
2 Safety............................................................................................................15
3 Transportation .............................................................................................31
4 Assembling ..................................................................................................37
5 Operation......................................................................................................79
6 Troubleshooting ..........................................................................................87
7 Maintenance...............................................................................................113
8 Disassembling ...........................................................................................121
A Appendix ....................................................................................................133
ZTKS0156 / Z31820 / Rev.: 03
3
4
ZTKS0156 / Z31820 / Rev.: 03
Table of Contents
0 Contact ...........................................................................................................9
1 Introduction..................................................................................................11
1.1 General Information................................................................................................ 11
1.2 Limitation of Liability ............................................................................................... 11
1.3 Customer Service................................................................................................... 11
1.4 Warranty................................................................................................................. 11
1.5 Copyright ................................................................................................................ 11
1.6 General View.......................................................................................................... 12
2 Safety............................................................................................................15
2.1 Approved Persons.................................................................................................. 15
2.2 Customer Responsibilities...................................................................................... 16
2.3 Key Words.............................................................................................................. 17
2.4 Residual Risks........................................................................................................ 18
2.4.1 Persons .................................................................................................................. 18
2.4.2 Intended Use .......................................................................................................... 18
2.4.3 Safety Devices ....................................................................................................... 19
2.4.4 Spare Parts ............................................................................................................ 19
2.4.5 Signs and Labels.................................................................................................... 20
2.4.6 Technical Risks ...................................................................................................... 20
2.5 Personal Protective Equipment .............................................................................. 26
2.6 Description of Signs and Labels............................................................................. 27
2.7 Safety Devices ....................................................................................................... 28
2.8 Behavior in Danger and Emergency Situations...................................................... 30
2.9 Fire Department Notification................................................................................... 30
3 Transportation .............................................................................................31
3.1 Safety ..................................................................................................................... 31
3.2 Packaging............................................................................................................... 31
3.2.1 Disposal.................................................................................................................. 32
3.3 Transport Inspection............................................................................................... 32
3.4 Transportation by Fork Lift / Pallet Jack ................................................................. 33
3.5 Transportation with a Crane ................................................................................... 34
3.6 Storing .................................................................................................................... 35
4 Assembling ..................................................................................................37
4.1 Safety ..................................................................................................................... 37
4.2 Assembling Workflow ............................................................................................. 39
4.3 Assembling Work ................................................................................................... 40
4.3.1 Preparing the Cryostat ........................................................................................... 40
4.3.1.1 Removing the Transportation Locks....................................................................... 40
ZTKS0156 / Z31820 / Rev.: 03
5
Table of Contents
4.3.1.2 Mounting the Getter Container ............................................................................... 43
4.3.1.3 Checking the Feedthrough Plate ............................................................................ 44
4.3.1.4 Assembling the Nitrogen Tube ............................................................................... 44
4.3.1.5 Assembling the RT Tube ........................................................................................ 46
4.3.1.6 Checking and Fitting the Drop-off Plate.................................................................. 49
4.3.2 Evacuating the Cryostat ......................................................................................... 50
4.3.2.1 Mounting the Vacuum Valve................................................................................... 50
4.3.2.2 Mounting the Vacuum Valve at Evacuated Cryostat .............................................. 51
4.3.2.3 Generating the Vacuum.......................................................................................... 52
4.3.2.4 Rebuilding the Vacuum .......................................................................................... 53
4.3.2.5 Removing the Vacuum Valve ................................................................................. 53
4.3.3 Mounting the Cryogenic Refrigerator Parts ............................................................ 54
4.3.3.1 Mounting the Cold Head......................................................................................... 54
4.3.3.2 Mounting the Rotary Valve Column........................................................................ 56
4.3.3.3 Mounting the Rotary Valve and Flex Lines............................................................. 61
4.3.3.4 Mounting the Compressor ...................................................................................... 67
4.3.3.5 Mounting the Helium Level Sensor......................................................................... 69
4.3.3.6 Mounting the Helium Flow System ......................................................................... 70
4.3.3.7 Mounting the Temperature Control......................................................................... 71
4.3.4 Cooling down from 300 K to 80 K........................................................................... 73
4.3.5 Cooling down to 4.2 K ............................................................................................ 77
5 Operation......................................................................................................79
5.1 Safety ..................................................................................................................... 79
5.2 Energizing............................................................................................................... 80
5.2.1 Inserting the Current Lead ...................................................................................... 80
5.2.2 Energizing Procedure............................................................................................. 81
5.2.3 Cycling Procedure .................................................................................................. 82
5.2.4 Shimming Procedure .............................................................................................. 83
5.2.5 Removing the Current Lead ................................................................................... 84
5.2.6 Complete Energizing .............................................................................................. 84
5.3 Set into Operation................................................................................................... 85
6 Troubleshooting ..........................................................................................87
6.1 Safety ..................................................................................................................... 87
6.2 Problem .................................................................................................................. 88
6.2.1 During Transportation............................................................................................. 88
6.2.2 During Assembling ................................................................................................. 88
6.2.3 During Cool Down .................................................................................................. 90
6.2.4 During Energizing and Shimming........................................................................... 94
6.2.5 During Operation of the Magnet Stand ................................................................... 96
6.2.6 During Standard Operation..................................................................................... 98
6.2.7 During De-energizing and Warming up ................................................................ 101
6.2.8 During Operation of the Cryogenic Refrigerator ................................................... 102
6.3 Troubleshooting Work .......................................................................................... 107
6
ZTKS0156 / Z31820 / Rev.: 03
Table of Contents
6.3.1 After a Quench ..................................................................................................... 107
6.3.2 Replacement of the Helium Level Sensor ............................................................ 109
6.3.3 Procedure in case of an alarm signal................................................................... 110
6.3.4 Procedure in case of Cryogenic Refrigerator failure ............................................ 111
7 Maintenance...............................................................................................113
7.1 Safety ................................................................................................................... 113
7.2 Cleaning ............................................................................................................... 114
7.3 Maintenance Timetable ........................................................................................ 114
7.4 Maintenance Work at the Cryogenic Refrigerator ................................................ 116
7.4.1 Service Kit ............................................................................................................ 116
7.4.2 Flex Lines ............................................................................................................. 117
7.4.3 Replacement of the Rotary Valve and the Cold Head.......................................... 118
8 Disassembling ...........................................................................................121
8.1 Safety ................................................................................................................... 121
8.2 Disassembling Work............................................................................................. 123
8.2.1 De-energizing ....................................................................................................... 123
8.2.2 Warming up from 4.2 K to 300 K .......................................................................... 124
8.2.3 Breaking the Vacuum ........................................................................................... 125
8.2.4 Removing of the Helium Flow System ................................................................. 126
8.2.5 Removing the Cryogenic Refrigerator Parts......................................................... 126
8.2.6 Removing of the RT Tube and the Nitrogen Tube ............................................... 127
8.2.7 Assembling of the Transportation Lock ................................................................ 129
8.2.8 Remount the Magnet Stand ................................................................................. 131
8.2.9 Prepare the Cryostat for Transportation............................................................... 131
A Appendix ....................................................................................................133
A.1 Warning Signs ...................................................................................................... 135
A.2 Figures ................................................................................................................. 137
A.3 Tables................................................................................................................... 141
A.4 Glossary / Abbreviations ...................................................................................... 143
A.5 Technical Data USR Service................................................................................ 145
A.6 Index..................................................................................................................... 153
ZTKS0156 / Z31820 / Rev.: 03
7
Table of Contents
8
ZTKS0156 / Z31820 / Rev.: 03

0Contact

Manufacturer:
Bruker BioSpin AG
Industriestrasse 26
CH-8117 Faellanden
Switzerland
Phone: +41-44-825-91-11
Fax: +41-44-825-96-96
http://www.bruker.com
E-mail: service@bruker.ch
Please refer to the Model No., Serial No. and Internal Order No. in all correspondence regarding the NMR system or components thereof.
ZTKS0156 / Z31820 / Rev.: 03
9
Contact
10
ZTKS0156 / Z31820 / Rev.: 03

1 Introduction

1.1 General Information

This manual contains important information about the handling of the supplied magnet system as a part of the magnet system used for NMR analyses. The compliance with all safety and handling instructions, the applicable local accident prevention and general safety regulations are necessary for safe work.
The manual is part of the product. It must be kept in the immediate vicinity of the magnet system and unimpeded access must be ensured at any time. Read the manual carefully before handling the magnet system or its components.

1.2 Limitation of Liability

The information in this manual will take into account the current state of the technology.
The manufacturer assumes no liability for damages resulting from:
non-compliance with the instructions and all applicable documentation,
use for purposes not intended,
not sufficiently approved persons,
arbitrary changes or modifications and
use of not approved spare parts or accessories.

1.3 Customer Service

Technical support is provided by Bruker service via telephone or e-mail. For contact information see page 9 of this document.

1.4 Warranty

The warranty terms can be found in the sales documents of the magnet system and in the Terms and Conditions of Bruker BioSpin AG.

1.5 Copyright

No part of this publication may be reproduced, stored in a retrieval system, or transmit­ted, in any form or by any means without the prior consent of the publisher. Product names used are trademarks or registered trademarks of their respective holders.
ZTKS0156 / Z31820 / Rev.: 03
11
Introduction
1. Cryogenic Refrigerator Compressor (noise protection cover optionally)
2. Cryogenic Refrigerator Flex Lines
3. Rotary Valve (RV) covered by the Rotary Valve Column (RVC)
4. Connection Line
5. Helium flow system
6. Current lead turret
7. Cryogenic Refrigerator Cold Head
8. Helium fill-in turret with helium fill-in port
9. RT bore
10. RT vessel
11. Magnet Stand

1.6 General View

12
Figure 1.1: General view RS Cryostat of the Magnet System
ZTKS0156 / Z31820 / Rev.: 03
Introduction
The heart of the NMR magnet system is a superconducting magnet located inside the helium vessel, which is filled with liquid helium. The helium vessel is surrounded by a radiation shield (RS) both cooled by a Cryogenic Refrigerator. The outer casing, the room temperature (RT) vessel (10) contains the helium vessel and the radiation shield. The vacuum inside the RT vessel reduces thermal conduction. RT vessel, helium vessel, radiation shield, helium turrets, helium flow systems and the RT bore together build the cryostat of the NMR magnet system. The RT bore (9) allows the access to the magnetic centre of the NMR magnet system.
The cryostat is mounted on a magnet stand (8). The isolators in the magnet stand absorb floor vibrations. Different heights and isolators are available optionally.
The helium turrets (5, 8) build the interface of the helium vessel to the outside. The current lead turret (5) builds the interface for energizing the magnet coil and for diagnostic. The helium fill-in turret (8) with access to the siphon and connecting plug for the helium level sensor is used to fill the helium vessel. It is marked with a yellow label. The helium flow system (6) connects the two helium turrets.
The Cryogenic Refrigerator is a closed loop helium expansion cycle. It consists of a Compressor (1), two Flex Lines (2), a Remote Rotary Valve inside the Rotary Valve Column (RVC; 3), a Cold Head (7) and a Connection Line (4) between Remote Rotary Valve and Cold Head. The Remote Rotary Valve, Flex Lines, Connection Line and Cold Head are covered with an applicable noise protection cover. For the Compressor a noise protection is available optionally. Depending on the customers site restrictions several opportunities for the additional equipment of the Cryogenic Refrigerator are possible. Refer to the order subscription and to the supplied manuals of the supplied equipment.
ZTKS0156 / Z31820 / Rev.: 03
13
Introduction
14
ZTKS0156 / Z31820 / Rev.: 03

2 Safety

The supplied cryostat of the magnet system was designed and manufactured according to best available technical knowledge and practice, archived in over 50 years of experi­ence of Bruker Corporation. International standards for quality and approval recom­mended for cryostats of superconducting magnets were certified.
Nevertheless non-compliance with the following instructions and safety advice may cause serious hazards and property damage.

2.1 Approved Persons

Bruker BioSpin AG identifies the following qualifications for personnel performing tasks on the magnet system or its components:
Approved Customer Personnel:
As a result of professional training by Bruker Service Personnel, experience and knowledge of applicable regulations these persons are qualified to perform the specific tasks on the magnet system and its components assigned to them in this manual. Approved Customer Personnel are qualified to identify possible hazards and risks associated with the tasks assigned to them and to perform all possible steps to eliminate or minimize these risks.
Bruker Service Personnel:
These persons are qualified by appropriate qualification and professional training and experience (including all necessary knowledge of applicable regulations and regulatory requirements) to perform specific tasks on the magnet system and its components. Bruker Service Personnel are qualified to identify possible hazards and risks and to perform all possible steps to eliminate or minimize these risks.
ZTKS0156 / Z31820 / Rev.: 03
15
Safety

2.2 Customer Responsibilities

The customer of the magnet system must obey the security advice and the rules for safety, accident prevention and environmental protection correctly for the magnet systems. Furthermore, the customer is responsible for keeping the magnet system in correct technical condition.
In particular:
The customer must identify additional dangers resulting from the working conditions at the site of the magnet system and provide applicable safety measures.
The customer must ensure that the site plan meets the specified conditions for operating the magnet system and supplied site planning document.
The customer must clearly mark the danger area around the magnet system and post the corresponding instruction plates.
The customer has to ensure the intended use of the magnet system.
The customer has to inform the local fire brigade about the special risks of the magnet system and how to react in the event of an incident.
The customer must clearly define the responsibilities for operation and main­tenance.
The customer must ensure that all employees working with the magnet system have read and understood the manual.
The customer has to provide the necessary personal protective equipment for his employees.
The customer has to instruct his employees at regular intervals on hazards and safety measures.
The customer has to instruct other persons not working on the magnet system but carrying out work in the same room, for instance cleaning staff or guards.
The customer has to consider the specific items of this cryostat equipped with a Cryogenic Refrigerator. The customer is responsible for obeying the advice given in this manual. In case the Cryogenic Refrigerator is not running correctly his immediate reaction is mandatory. In case of an unexpected alarm his immediate response is mandatory. For further instruction refer to chapter ”Troubleshooting” on page 6-87.
The customer must ensure that maintenance is performed according to the schedule listed in chapter ”Maintenance Timetable” on page 7-114.
16
ZTKS0156 / Z31820 / Rev.: 03

2.3 Key Words

Indicates a hazardous situation which, if not prevented, will result in death or serious injury.
Indicates a hazardous situation which, if not prevented, could result in death or serious injury.
Safety
DANGER
WARNING
CAUTION
Indicates a hazardous situation which, if not prevented, may result in moderate or minor injury.
NOTICE
Hazard, which could result in property damage.
Information and links for efficient and trouble-free handling and operation.
i
ZTKS0156 / Z31820 / Rev.: 03
17
Safety

2.4 Residual Risks

In the following section, the residual risks from the risk analysis according ISO 14971 are summarized. To prevent health hazards and hazardous situations obey all safety instruc­tions and warnings in the manual.

2.4.1 Persons

Risk of injury and property damage by not approved persons.
Incorrect handling of the magnet system from not approved persons may result in significant bodily injury and property damage.
Thus:
Work must only be carried out by approved persons with applicable qualifications. The necessary qualifications are specified in the beginning of the relevant chapter.
WARNING
In case of doubt, contact Bruker Service. Contact information see page 9 of this document.

2.4.2 Intended Use

The supplied magnet systems is designed and intended for NMR spectroscopy only.
Risk of damage to life and limb by incorrect use of the magnet system.
Incorrect use of the magnet system can lead to life-threatening situations and destruction of the magnet system.
Thus:
Only use the magnet system as intended.
Do not change the magnet system.
Do not exceed specified values for operating the magnet system.
Do not use inserts inside the RT bore not approved by Bruker Service.
WARNING
18
Damage claims from damages caused by other than the intended use of the magnet system are excluded and the customer is held liable.
ZTKS0156 / Z31820 / Rev.: 03

2.4.3 Safety Devices

Risk of damage to life and limb due to not sufficient safety devices.
Several safety devices ensure safe operation of the magnet system. They must always be in correct working condition.
Thus:
Do not block safety devices.
Do not remove safety devices.
Check the operational reliability of the safety devices before working on the magnet system.
Safety
WARNING

2.4.4 Spare Parts

Risk of injury and property damage from using incorrect or defective spare parts and accessories.
Incorrect or defective spare parts can cause serious injuries. They may cause damaging, malfunctioning and the destruction of the magnet system.
Thus:
Only use original equipment manufacturer spare parts.
Only use original equipment manufacturer accessories.
WARNING
ZTKS0156 / Z31820 / Rev.: 03
19
Safety

2.4.5 Signs and Labels

Risk of damage to persons and property due to not readable signs and labels.
Signs and labels with advice may become not readable.
Thus:
Maintain signs and labels in a readable state.
Replace damaged or not readable signs and labels immediately. New signs and labels can be obtained from Bruker Service.
WARNING

2.4.6 Technical Risks

Magnetic Field
Risk of damage to life and limb due to high magnetic fields.
A magnetic field of more than 0.5 mT (5 Gauss) is life-threatening for people with pacemakers or metal implants. Ferromagnetic tools in the magnetic field are significantly hazardous. Disks and electronic devices may be damaged. Duration of exposure (8 h/day) above the limit of 200 mT can cause damage to health.
Thus:
Mark the magnetic field of more than 0.5 mT (5 Gauss) before start up.
The workplace must be outside the 0.5 mT area.
Keep people with pacemakers and metal implants away from the identified area.
Keep disks, credit cards and electronic devices away from the identified area.
Do not use ferromagnetic tools or items within the identified area.
Only use non-ferromagnetic transportation dewars for the cryogenic agents.
Only use non-ferromagnetic ladders or steps.
WARNING
20
ZTKS0156 / Z31820 / Rev.: 03
Safety
Cryogenic Agents
WARNING
Risk of damage to life and limb due to cryogenic agents.
Risk of damage to life and limb due to not correct handling of liquid cryogenic agents and their dewars. Within the transition from liquid to gas, helium and nitrogen expands their volume, causing closed vessels to burst.
The evaporating cryogenic agents will displace the air. In case of not sufficient ven­tilation this may result in death by suffocation. Helium displaces oxygen in the upper part of the room, nitrogen displaces oxygen in the lower parts of the room.
Liquid and gaseous cryogenic agents are extremely cold. Contact with liquid or gaseous cryogenic agents will lead to cold burns.
Thus:
Only use cryogenic agents in well ventilated rooms. In case of doubt ask Bruker Service.
Wear an oxygen monitor on the body during service and maintenance work.
Prevent any skin contact with liquid or gaseous cryogenic agents.
Electricity
WARNING
Risk of damage to life and limb due to electricity.
Risk of damage to life and limb due to contact with electrical lines and damaged insulation.
Thus:
Work on electrical equipment must be done by an approved electrical technician.
Keep moisture away from electrical lines to prevent short-circuits.
Check the magnet system accidental ground before start up.
Switch power OFF before working on the Bruker Power Supply or further equip­ment.
ZTKS0156 / Z31820 / Rev.: 03
21
Safety
Quench
WARNING
Risk of suffocation during a quench of the magnet system.
A quench is the very fast de-energizing of the magnet by loss of its superconductivity. The stored magnetic energy is converted into heat and thus large quantities of helium evaporate. The evaporating helium will displace oxygen. In case of not sufficient ventilation this may result in death by suffocation.
Thus
The magnet system site must be well ventilated. In case of doubt contact Bruker Service.
The evaporating gas may resemble smoke. Never pour water on the magnet system.
Gas under Pressure
WARNING
Risk of injury due to gas under pressure inside the cryostat and further equipment.
The helium or the nitrogen vessel of the cryostat may get sealed off due to ice formation inside the helium or the nitrogen turrets in case of non-compliance with the instruction given in this manual. This may lead to overpressure and damage of the helium or the nitrogen vessel.
Manipulations of components with gas under pressure may lead to injury and property damage.
Thus:
In case of icing inside the helium or the nitrogen turrets contact Bruker Service immediately.
Release the pressure before working on components with gases under pressure.
Do not seal cryogenic agent vessels of the magnet system or the transport dewars.
Do not connect high pressure transport dewars for liquid cryogenic agents to the magnet system. Completely eliminate the high pressure from the transport dewars before connecting and transferring.
Keep the Cryogenic Refrigerator circuit closed at any time. Overpressure can release in the buffers of the Compressor, the Rotary Valve and the Cold Head.
22
ZTKS0156 / Z31820 / Rev.: 03
Safety
Low Temperatures
WARNING
Risk of injury due to low temperatures of liquids and metal parts.
Physical contact with extremely cold liquids and metal parts may cause serious inju­ries. Contact with the skin may cause cold burns. Contact with the eyes may cause blindness.
Thus:
Always wear protective goggles, protective gloves and protective clothes while handling with liquid cryogenic agents or metal parts in contact with liquid cryogenic agents.
Protect temperature sensitive components such as O-rings from contact with liquid cryogenic agents.
Spontaneous Ignition and Explosion
WARNING
Risk of injury from spontaneous ignition and explosion caused by liquid oxygen.
Pure oxygen condenses on extremely cold metal pieces. Together with oil it may ignite spontaneously. In case of fire the pure oxygen may cause an explosion.
Thus:
Do not smoke near the magnet system.
Do not use open flames near the magnet system.
Keep the environment around the magnet system clean.
Do not leave oily rags near the magnet system.
Risk of Slippage
WARNING
Risk of injury from slippage.
The accumulation of condensed water on the floor and ladders causes slippery surfaces.
Thus:
Always wear safety shoes with an anti slip sole.
Be careful using ladders.
Clean floor and ladders regularly.
ZTKS0156 / Z31820 / Rev.: 03
23
Safety
Risk of Tilting
WARNING
Risk of injury due to tilting of the magnet system.
The magnet system is very sensitive to lateral forces. It may tilt.
Thus:
Do not climb onto the magnet system.
Do not lean items against the magnet system.
Do not lean against the magnet system.
Do not move the magnet system on your own.
Risk of Heavy Weights
WARNING
Risk of damage to life and limb caused from heavy weights.
Lifting heavy weights is life-threatening due to falling or moving parts.
Thus:
Do not stay or work under a lifted magnet system.
All used lifting equipment must be approved to carry the weight (see Appendix, Technical Data).
Do not use damaged lifting equipment.
Do not use lifting equipment without updated check tag.
Lifting only with approved qualification.
Obey ergonomic guidelines while lifting heavy parts.
Protect parts against falling.
Always wear safety shoes with approved toe caps.
24
ZTKS0156 / Z31820 / Rev.: 03
Safety
Risk of Hot Surfaces
WARNING
Risk of injury from contact with hot or cold surfaces.
Surfaces of the Cryogenic Refrigerator parts may be hot. Skin contact with these surfaces can cause serious injuries.
Thus:
Any work at the Cryogenic Refrigerator parts must only be performed by Bruker Service.
Always wear protective gloves while handling Cryogenic Refrigerator parts.
Transportation
CAUTION
Risk of injury and property damage due to incorrect transportation.
The box may tilt, movement may get out of control. Thus persons may get injured and the cryostat or further equipment may be damaged.
Thus:
Be careful while unloading and moving the boxes.
Do not move the boxes arbitrarily.
Pay attention to all symbols on the boxes.
Pay attention to sharp edges and spikes of boxes and parts by using protective gloves while moving.
Move the boxes in an upright position.
Do not tilt the boxes.
Prevent crossing thresholds, even if they are only a few millimeters high.
Clean the transportation way before transporting the box.
Unpack shortly before assembling.
The cryostat or further equipment must be protected from rain and other bad weather conditions during transportation.
Exclusively move the cryostat in its original box.
Do not remove the tightening straps inside the box until assembling.
Only use the attachment points provided.
Ensure that the cryostat is always leveled during any transportation.
Transportation only with transportation locks attached.
Do not move the evacuated cryostat.
Do not move the cryostat after cool down.
ZTKS0156 / Z31820 / Rev.: 03
25
Safety

2.5 Personal Protective Equipment

The personal protective equipment must be worn at any time while working on the magnet system and to avoid health hazards.
Protective Goggles
Used to protect the eyes from injury due to flying cold liquids and parts.
Protective Gloves
Used to protect the hands from injury caused by contact with extremely cold liquids or surfaces and for protection from injury caused by rough edges.
Protective Clothes
Used to protect the body from injury caused by contact with extremely cold liquids or surfaces and for protection from wounds.
Safety Shoes
Used to protect the feet from injury from falling of heavy objects. An anti-slip sole protects from injury caused by slipping and falling on slippery floor and steps. Only use safety shoes with non-ferromagnetic toe caps.
Portable Oxygen Monitor and Alarm
Used to warn against low oxygen concentrations in surrounding air.
26
ZTKS0156 / Z31820 / Rev.: 03

2.6 Description of Signs and Labels

Signs and labels are always related to their immediately vicinity. The following signs and labels are found on the magnet system and in the vicinity.
Prohibition sign: No person with pacemakers!
People with pacemakers are endangered in the identified area of 0.5 mT (5 Gauss) and are not allowed to enter these areas.
Prohibition sign: No person with implants!
People with metallic implants are endangered in the identified area of 0.5 mT (5 Gauss) and are not allowed to enter these areas.
Prohibition sign: No watches or electronic devices!
Safety
Watches and electronic devices may be damaged in the identified area of 0.5 mT (5 Gauss).
Prohibition sign: No credit cards or other magnetic memory!
Credit cards and magnetic memory may be damaged in the identified area of 0.5 mT (5 Gauss).
Prohibition sign: Do not touch! Do not block!
Do not touch or block identified area.
Hazard warning sign: Strong magnetic field!
No magnetic memory.
No jewelry.
No metallic items.
Emergency exit!
Always keep the emergency exit clear.
Follow the arrows if necessary.
ZTKS0156 / Z31820 / Rev.: 03
Doors must pushed open in escape direction.
27
Safety
1. Quench Valves
2. One-way Valve of the helium vessel
3. Pressure Relief Valve of the cold head turret
4. Drop-off Plate

2.7 Safety Devices

The cryostat of the supplied magnet system is equipped with the following safety devices:
Figure 2.1: Safety Devices of the RS Cryostat
28
ZTKS0156 / Z31820 / Rev.: 03
Safety
Quench Valve
The quench valves (1) are the safety devices of the helium vessel. They open with a defined pressure. In case of an accidental overpressure in the helium vessel the quench valves will release the pressure smoothly.
One-way Valves
The one-way valves at the helium flow system (2) keeps air and moisture from entering the helium vessel in case of an accidental underpressure inside the vessel.
Pressure Relief Valve
The pressure relief valve (3) is the safety device of the cold head turret. It opens with a defined pressure. In case of an accidental overpressure in the cold head turret the pressure relief valve will release the pressure smoothly.
Drop-off Plate
The drop-off plate (4) is a safety device of the RT vessel. If the vacuum breaks, the drop­off plate will open. In case of an accidental overpressure in the RT vessel the drop-off plate will release the pressure smoothly.
Cryogenic Refrigerator Parts
For information about the safety devices of the Cryogenic Refrigerator parts refer to the supplied separate manual.
ZTKS0156 / Z31820 / Rev.: 03
29
Safety

2.8 Behavior in Danger and Emergency Situations

Preparations
Keep the emergency exits free at all times.
Prepare and maintain an up-to-date list of emergency telephone numbers in the magnet system area.
In Case of Emergency
Leave the danger zone immediately.
Check for sufficient ventilation in the room before entering, especially if people are showing symptoms of suffocation.
Rescue persons from the danger zone.
Provide medical attention for people with symptoms of suffocation.
Start first aid immediately.
Call the responsible contact.
Call for medical assistance.
Call the fire department.
First Aid for Cold Burns
Help the injured persons to lie down comfortably in a warm room.
Loosen all clothing which could prevent blood circulation in the injured area.
Pour large quantities of warm water over the affected parts.
Cover the wound with dry and sterile gauze.
In case of contact of liquid cryogenic agents with the eyes rinse thoroughly with clean water.
Call for medical assistance.

2.9 Fire Department Notification

Inform the fire department about the potential risks of a magnet system, i.e. danger of ferromagnetic rescue equipment close to the magnet system.
30
Laboratory windows which are accessible during an emergency should be clearly identified with warning signs, visible from the outside.
Inform the fire department about the characteristics of a quench to prevent confu­sion with smoke.
Never pour water over the magnet system during a quench!
ZTKS0156 / Z31820 / Rev.: 03

3 Transportation

The cryostat is supplied in a wooden box on a pallet. It is secured inside with straps against tilting and moving.
Accessories such as the flow systems, level sensors and bore tubes are in the side compart­ment of the box.
The Cryogenic Refrigerator parts and the Flex Lines are supplied in boxes on a pallet.
The Magnet Stand is supplied in a wooden box on a pallet.

3.1 Safety

The transportation is carried out by Bruker Service or approved persons. However, it may happen that other persons have to receive the delivery of the shipping boxes. In this case it is essential to obey the instructions in this chapter and to inform these persons before.
WARNING
Heavy Weights (see page 2-24)
CAUTION
Transportation (see page 2-25)

3.2 Packaging

Figure 3.1: Packaging (without surrounding panels)
ZTKS0156 / Z31820 / Rev.: 03
31
Transportation

3.2.1 Disposal

Keep the original boxes for future transportation.
If no further transportation is planned, dispose of the boxes according to environ­mentally friendly regulations.

3.3 Transport Inspection

Investigate the delivery with regard to visible damage and completeness of delivery.
Transport control systems
The shipping and handling monitors (“Shock Watch“, “Tilt Watch“) on the boxes shows, if the boxes were kicked or tilted during transportation.
Checks
Shock Watch: Follow instructions on the label.
Tilt Watch: Follow instructions on the label.
In case of damage
Accept delivery with reservation.
Make a documentation of all observable damage and add it to the transport
documents.
Start complaint process.
Contact Bruker Service before installation.
The claim for damage expires after the fixed period.
i
Thus: Report damages to Bruker Service immediately after detection of damage. For contact information see page 9 of this document.
32
ZTKS0156 / Z31820 / Rev.: 03

3.4 Transportation by Fork Lift / Pallet Jack

1. Check the route of transport for the minimal height and width.
2. Check sufficient floor capacity on the route of transport. In case of doubt ask a stress analyst.
3. Check sufficient carrying capacity while using an elevator.
4. Position the forks between the bars of the box as shown in the figure. Make sure the side towards the operator is the one with the labels on it.
5. Make sure the forks of the fork lift are longer than the box and projects out of the back of the box as shown in the figure below.
6. Now lift the fork and move the box to the site.
Recommended for transporting the boxes to the installation site.
Approved Persons: Approved forklift / pallet jack operator
Precondition: The fork lift / pallet jack must be approved for the transportation weight
(see chapter Appendix, Technical Data).
Transport
Transportation
Figure 3.2: Transportation by forklift
ZTKS0156 / Z31820 / Rev.: 03
33
Transportation
1. Exclusively use the marked eyelets (1) as attachment points for the lifting equipment.
2. Use all
eyelets for the lifting equipment.
3. Follow the instructions on the label on the top of the cryostat. This label gives important information about correct attachment and transportation.
4. Check for correct fastening of the lifting equipment before lifting the cryostat.
5. Make sure that any movement of the crane is as slowly as possible to avoid any damage due to acceleration.
6. Check for correct leveling of the cryostat while hanging on the crane.

3.5 Transportation with a Crane

Recommended for lifting the cryostat out of the box.
Approved Persons: Approved crane operator
Precondition: The crane must be approved for the transportation weight (see chapter
Appendix, Technical Data).
Attachment Points
34
Figure 3.3: Attachment points for lifting equipment
ZTKS0156 / Z31820 / Rev.: 03

3.6 Storing

If it is necessary to store the cryostat and accessories before installation obey the following instructions:
Store the boxes in a closed, dry and dust-free room.
Store the boxes upright.
Do not tilt the boxes.
Do not unpack the supplied boxes.
Prevent mechanical vibrations to the boxes.
Storage temperature: 5 - 40 °C.
Storage humidity: less than 50% @ 23 °C.
Transportation
ZTKS0156 / Z31820 / Rev.: 03
35
Transportation
36
ZTKS0156 / Z31820 / Rev.: 03

4 Assembling

4.1 Safety

Approved Persons: Bruker Service only
Magnetic Fields (see page 2-20)
Cryogenic Agents (see page 2-21)
Electricity (see page 2-21)
Gas under Pressure (see page 2-22)
Spontaneous Ignition and Explosion (see page 2-23)
WARNING
Heavy Weights (see page 2-24)
Personal Protective Equipment:
Protective goggles
Protective gloves
Protective clothes
Safety shoes
Portable oxygen monitor and alarm
ZTKS0156 / Z31820 / Rev.: 03
37
Assembling
Necessary Tools and Equipment
Alignment tool 1.8 mm
Bruker Vacuum Valve (Material No. Z53420)
Vacuum pumping unit
Bruker Power Supply
Bruker ACD (Automatic Cooling Device)
Precooling tube ("L-tube")
Helium transfer line
Torque wrench 5 - 50 Nm
Clean gloves
Ethanol
Polyester tape
Vacuum grease
Screw grease MOLYKOTE P1000 (supplied with the magnet stand)
Tool kit: (supplied with the CRC package)
Wrenches for Flex Lines
PTFE spray
Check if there is sufficient free space above the cryostat to mount the bore tubes and the
i
helium level sensor.
If not sufficient, mount the cryostat according following instructions at a place with sufficient ceiling height, remove the magnet stand after assembling and transport the cryostat to the site using the transport box. If this is not possible, mount the magnet stand supports. For work at the bottom of the cryostat attach the magnet stand pillars. For inserting the bore tubes or the helium level sensor remove the magnet stand and lower the cryostat. Contact Bruker Service head quarter for further information.
38
ZTKS0156 / Z31820 / Rev.: 03
Mount the Magnet Stand
Prepare the Cryostat
Assemble
Bore Tubes
Check
Drop off plate
Mount the Cryogenic Refrigerator Parts
Rotary Valve Column
Monitoring
Generate Vacuum
Cooldown the Cryostat
Assemble vacuum valve
Remove vacuum valve
Precool (80K)Cooldown (4.2K)
Switch on
Cryogenic Refrigerator
Energize Magnet
Cycle/Shim Magnet
Remove current
lead
Insert Current
Lead
Remove the
transportation locks
Insert the getter
container
Sub-Procedure different
from St andard NMR
Standar d NMR Procedur e
Standard NMR
Sub-Proc edure
KEY:
Procedure different from
Standar d NMR
Start Monitoring
Pump an Flush
Evacuate
Cold head
He-Turrets and
Flow System
Feedthrough Plate
Compressor
Assembling

4.2 Assembling Workflow

Figure 4.1: Assembling Workflow
ZTKS0156 / Z31820 / Rev.: 03
39
Assembling
1. Remove the top and side panels of the box.
2. Loosen the fixing straps from the cryostat.
3. Lift the cryostat out of its box.
4. Attach the magnet stand to the cryostat. Follow the instructions in the manual of the supplied magnet stand respecting the warnings given there.
WARNING: Heavy Weights
Overview positions of the transportation lock (colored blue).

4.3 Assembling Work

4.3.1 Preparing the Cryostat

Figure 4.2: Preparing the cryostat
4.3.1.1 Removing the Transportation Locks
Figure 4.3: Overview Transportation Locks
40
ZTKS0156 / Z31820 / Rev.: 03
Removing the upper Transportation Lock
1. Remove the water protection cover (1).
2. Remove the two M 12 nuts (2) and washers.
3. Release the three M 8 nuts (1).
4. Remove the threaded M 8 bolt (3).
5. Release the six M 5 x 14 screws (4).
6. Remove the transport top plate (5).
7. Remove the short top fixture flange (1).
8. Remove the intermediate flange (2).
9. Remove the top fixture flange (3).
Figure 4.4: Removing the Transportation Lock - step 1
Assembling
Figure 4.5: Removing the Transportation Lock - step 2
Figure 4.6: Removing the Transportation Lock - step 3
ZTKS0156 / Z31820 / Rev.: 03
41
Assembling
10. The threaded bar (1) remains until removing the lower transportation lock. Afterwards remove the threaded bar from the top.
11. Remove three nuts M 8 (2).
12. Remove the threaded bolts M 8 (1).
13. Loosen the twelve screws M 5 x 12 at the lower transportation lock plate (3). Secure the lower transportation lock plate against falling.
14. Remove lower transportation lock plate (3), the brass nut (4) and lower fixture flange (5).
15. Loosen the threaded bar from the brass nut (4).
Continue on top of the cryostat:
16. Remove the threaded bar from the top.
WARNING: Heavy Weights
Figure 4.7: Removing the Transportation Lock - step 4
Removing the lower Transportation Lock
Figure 4.8: Removing the Transportation Lock - step 5
Each transportation lock is adapted to only one cryostat, indicated by the magnet system
i
number in each part of the transportation lock. Do not mix-up transportation locks from other cryostat. Keep all transportation lock parts for future transportation. For storage use the compartment of the shipping box of the cryostat.
42
ZTKS0156 / Z31820 / Rev.: 03
4.3.1.2 Mounting the Getter Container
1. Check the clamp (2); retighten it if neces­sary.
2. Unpack the getter containers from their tin.
3. Insert the getter container (1) into the clamps (2) respecting the orientation as shown (flat side of the getter container atop).
4. Check the distance to the radiation shield.
5. Start generating vacuum no later than 4 hours after inserting the getter container.
If the cryostat is equipped with a getter box (1) like shown check if the getter box is:
1. mounted correctly and
2. not damaged.
i
Getter is a material, which absorbs helium gas to assure the quality of the vacuum for a
i
long time. After unpacking the getter starts to absorb moisture, reducing the capacity for absorbing helium inside the cryostat. Thus: Unpack getter container shortly before insertion. Close and evacuate the cryostat no later than 4 hours after the insertion of the getter container. Wear clean gloves for inserting to prevent contamination.
Assembling
Figure 4.9: Mounting the Getter Container
Figure 4.10: Checking the Getter Container
ZTKS0156 / Z31820 / Rev.: 03
43
Assembling
1. Check through the top opening of the cryostat that all connectors at the inner side of the feedthrough plate are plugged and the cables are not stressed or bent.
2. Check the feedthrough plate is correct assembled.
1. Mount the lower RS reduction flange (1) to the radiation shield clips.
2. Mount the upper RS reduction flange (1) to the radiation shield clips.
4.3.1.3 Checking the Feedthrough Plate
Figure 4.11: Checking and fitting the Feedthrough Plate
4.3.1.4 Assembling the Nitrogen Tube
Bottom of Cryostat
Figure 4.12: Assembling the Nitrogen Tube - step 1
Top of the Cryostat
44
Figure 4.13: Assembling the Nitrogen Tube - step 2
ZTKS0156 / Z31820 / Rev.: 03
Bottom of the Cryostat
3. Insert and fix the lower nitrogen contact flange (2). Be careful with the super insula­tion.
4. Clean the nitrogen tube with ethanol. Pre­vent any contamination (finger prints, dirt e.g.) to the nitrogen tube surface by using clean gloves. Prevent any damage (scrat­ches, buckling) to the nitrogen tube by hand­ling it carefully.
5. Insert the nitrogen (1) tube from the top.
6. Check the distance between the nitrogen and helium vessel by inserting the alignment tool (1) into the three check holes of the nitrogen contact flange (2). The insertion/ extraction of the alignment tool (1) should be smooth.
7. If the alignment tool does not insert/extract smoothly, remove the nitrogen tube and lower nitrogen contact flange and adjust the alignment with the alignment rods (1) (loosen/tighten).
8. Remount the nitrogen tube and nitrogen contact flange and check the alignment.
9. If the alignment is correct, fix the nitrogen contact flange.
Figure 4.14: Assembling the Nitrogen Tube - step 3
Assembling
Figure 4.15: Assembling the Nitrogen Tube - step 4
Figure 4.16: Assembling the Nitrogen Tube - step 5
ZTKS0156 / Z31820 / Rev.: 03
45
Assembling
10. Mount the upper nitrogen reduction flange (1).
11. Adjust the distance between the nitrogen and helium vessel by inserting the alignment tool into the three check holes of the nitro­gen contact flange (2).
12. Fix the upper nitrogen contact flange (1).
13. Check the alignment tool (2) for free movement. Remove the alignment tool.
14. Close the cuts of the super insulation with polyester tape without taping the insulation to other surfaces.
1. Check the O-ring slot (1) of the lower RT reduction flange.
2. Remove scratches, if necessary.
3. Clean the O-ring slot with ethanol.
4. Clean the O-ring with ethanol.
5. Grease the O-ring with vacuum grease and place it into the O-ring slot. Prevent contamination with dust and dirt.
6. Grease the twelve M 5 x 12 screws (1) with screw grease.
7. Insert the lower RT reduction flange from the bottom. Check the index. The lower RT reduction flange can be inserted in only one position.
8. Fix the lower RT reduction flange with the twelve screws (torque 12 Nm).
Top of the Cryostat
Figure 4.17: Assembling the Nitrogen Tube - step 6
4.3.1.5 Assembling the RT Tube
Bottom of the Cryostat
Figure 4.18: Assembling the RT Tube - step 1
46
Figure 4.19: Assembling the RT Tube - step 2
ZTKS0156 / Z31820 / Rev.: 03
Figure 4.20: Assembling the RT Tube - step 3
9. Check the alignment of the nitrogen tube and the lower RT reduction flange with a view through the nitrogen tube from the top. If the alignment is not correct, remove the lower RT reduction flange and adjust the alignment with the three lower alignment rods (loosen/tighten).
10. Insert the lower RT closure flange (1) to the RT reduction flange (2). Check the index. The lower RT closure flange can be inserted in only one position.
11. Fix it with only three of the six M5 x 12 screws.
12. Clean the RT tube carefully with ethanol. Prevent any contamination (finger prints, dirt e.g.) to the RT tube surface by using clean gloves. Prevent any damage (scratches, buckling) to the RT tube by handling it carefully.
13. Insert the RT tube from the top. In case the RT tube is insulated handle it carefully and prevent damage to the insula­tion. In case the RT tube is not symmetrical respect the correct position. Align the marks at the inner side of the RT tube with the mark at upper RT reduction flange.
14. Clean the O-ring of the lower RT closure flange with ethanol.
15. Grease the O-ring with vacuum grease.
16. Carefully remove the lower RT closure flange. Secure the RT bore tube against falling down.
17. Check the O-ring slot at the lower RT closure flange and clean it with ethanol. Remove scratches, if necessary.
Assembling
Figure 4.21: Assembling the RT Tube - step 4
Bottom of the Cryostat
Figure 4.22: Assembling the RT Tube - step 5
ZTKS0156 / Z31820 / Rev.: 03
47
Assembling
18. Mount the O-ring (1) to the RT tube.
19. Grease six M 5 x 12 screws with screw grease.
20. Mount the lower RT closure flange to the end of the RT tube. Push the RT closure flange with the RT tube inside the cryostat. Check the index. The lower RT closure flange can be inserted in only one position.
21. Fix the bottom RT closure flange with the six screws (torque 12 Nm).
22. Mount the upper RT reduction flange. Res­pect the correct position if an eccentrically upper RT reduction flange was used. Align the marks at the RT vessel and the upper RT reduction flange.
23. Grease twelve M 5 x 12 screws with screw grease.
24. Fix the upper RT reduction flange with the twelve screws (torque 12 Nm).
25. Grease six M 5 x 12 screws with screw grease.
26. Fix the upper RT closure flange with the six screws (torque 12 Nm). Respect the correct position if an eccen­trically upper RT closure flange was used. Align the marks at the RT closure flange with the marks at the RT vessel and the upper RT reduction flange.
Figure 4.23: Assembling the RT Tube - step 6
Figure 4.24: Assembling the RT Tube - step 7
Top of the Cryostat
Figure 4.25: Assembling the RT Tube - step 8
48
ZTKS0156 / Z31820 / Rev.: 03
4.3.1.6 Checking and Fitting the Drop-off Plate
1. Open the drop-off plate.
2. Clean the flange of the RT vessel and the O-ring with ethanol.
3. Grease the O-ring with vacuum grease.
4. Close the drop-off plate.
5. Check the drop-off plate is correct assembled.
Figure 4.26: Checking and fitting the Drop-off Plate
Assembling
ZTKS0156 / Z31820 / Rev.: 03
49
Assembling
1. Remove the protection cap of the vacuum flange at the bottom plate (2) of the cryostat.
2. Install the valve stem (10) onto the sealing plug (3) and tighten it slightly. Do not fix the valve body (1) to the cryostat yet.
3. Pull out the sealing plug (3) off the bottom plate at the valve stem (10).
4. Clean the sealing plug.
5. Check the sealing plug and the O-rings for damage. Replace damaged O-rings.
(4) O-ring 36.14 x 2.5 (Material No.
40692)
(5) O-ring 44 x 2.5 (Material No. 40693)
(6) O-ring 57 x 2.5 (Material No. 40695)
6. Grease the O-rings with vacuum grease.
7. Pull out the valve stem (10) of the valve body (1) until it snaps into place. The snap­ping is well defined and will be heard and felt.
8. Check if the sealing plug (3) inside the valve body is fully inserted.
9. Insert the Vacuum Valve into its seat inside the bottom plate (2).
10. Turn the Vacuum Valve in the desired posi­tion (KF 40 flange (9) looking outwards).
11. Insert the two half rings (7) as shown into the slots of the valve body (1).
12. Fix the half rings (7) with four M 6 x 12 (8) screws.

4.3.2 Evacuating the Cryostat

For evacuating the cryostat use the Bruker Vacuum Valve, Material No. Z53420.
Be careful handling the Vacuum Valve. Tilting of the Vacuum Valve during insertion may
i
damage the O-ring or required edges and surfaces of the Vacuum Valve. Thus pay attention to the axis during insertion of the Vacuum Valve.
4.3.2.1 Mounting the Vacuum Valve
50
Figure 4.27: Mounting the Vacuum Valve
ZTKS0156 / Z31820 / Rev.: 03
4.3.2.2 Mounting the Vacuum Valve at Evacuated Cryostat
1. Remove the protection cap of the vacuum flange at the bottom plate (2) of the cryostat.
2. Install the valve stem (10) onto the sealing plug (3) and tighten it slightly.
3. Turn the Vacuum Valve in the desired posi­tion (KF 40 flange (9) looking outwards).
4. Place the two half rings (7) as shown into the slots of the valve body (1).
5. Fix the half rings (7) with four M 6 x 12 (8) screws.
The sealing plug can not be moved until a correct vacuum was applied at the KF 40
i
flange at the Vacuum Valve.
Assembling
Figure 4.28: Mounting the Vacuum Valve at evacuated cryostat
ZTKS0156 / Z31820 / Rev.: 03
51
Assembling
1. Connect the vacuum pumping unit with the KF connector to the KF 40 flange (6) of the Vacuum Valve (2). Use a short pump line with large diameter.
2. Pull out the valve stem (7) of the valve body (2) to release the sealing plug (1) out of the bottom plate (3). The sealing plug (1) snaps into place. The snapping is well defined and will be heard and felt. The cryostat is open after this pro­cedure.
3. Pump and flush five times with dry nitrogen gas (only necessary for generating vacuum for the first time).
4. Generate a vacuum of 10
-4
bar with the
roughing pump.
5. Generate a vacuum of less than
5x10
-8
bar with a turbo pump (up to
48 hours).
6. The pumping unit has to be operating until liquid helium filled inside the helium vessel, respectively until the magnet inside is at a temperature of 77 K.
7. At the end of the helium filling procedure push the valve stem (7) slightly and fully into the valve body (2) to insert the sealing plug (1) in the vacuum flange at the bottom plate (3). The sealing plug (1) snaps in. The snapping is well defined and will be heard and felt.
8. Stop pumping.
4.3.2.3 Generating the Vacuum
Figure 4.29: Generating the Vacuum
52
ZTKS0156 / Z31820 / Rev.: 03
4.3.2.4 Rebuilding the Vacuum
1. Connect the vacuum pumping unit to the KF 40 flange (6) of the Vacuum Valve (2). Use a short pump line with large diameter.
2. Generate a vacuum of 10
-4
bar with the
roughing pump.
3. Generate a vacuum of less than
5 x 10
-8
bar with a turbo pump.
4. Pull out the valve stem (7) of the valve body (2) to release the sealing plug (1) out of the bottom plate (3). The sealing plug (1) snaps into place. The snapping is well defined and will be heard and felt. The cryostat is open after this pro­cedure.
5. Continue generating a vacuum of less than
5 x 10
-8
bar (up to 48 hours).
6. Push the valve stem (7) slightly and fully into the valve body (2) to insert the sealing plug (1) into the bottom plate (3). The sealing plug (1) snaps in. The snapping is well defined and will be heard and felt.
7. Stop pumping.
1. Ventilate the pump line at the pumping unit.
2. Remove the pump line at the KF 40 flange (9).
3. Release the valve stem (10) from the sealing plug (3) and pull out the valve stem (10) until it snaps in. The snapping is well defined and will be heard and felt.
4. Remove the four M 6 x 12 screws (8) from the half rings (7).
5. Remove the half rings (7) and the vacuum valve (1).
6. Mount the protective caps on the vacuum flange at the bottom plate (2) of the cryostat.
Assembling
Figure 4.30: Rebuilding the Vacuum
4.3.2.5 Removing the Vacuum Valve
Figure 4.31: Removing the Vacuum Valve
ZTKS0156 / Z31820 / Rev.: 03
53
Assembling
Remove the top and the front panel of the box of the Cryogenic Refrigerator box like shown.
1. Remove the service flange from the cold head turret.
Keep the service flange for further service or transportation inside the box of the cold head.
2. Check if the cold head turret inside is dry and clean. Clean it if necessary.
3. Clean the mating surface from cold head turret top flange.
4. The cold head turret flange is marked with an “X“ (1) for the following steps.

4.3.3 Mounting the Cryogenic Refrigerator Parts

Figure 4.32: Cryogenic Refrigerator Parts - General View
4.3.3.1 Mounting the Cold Head
Figure 4.33: Mounting the Cold Head - step 1
54
Figure 4.34: Mounting the Cold Head - step 2
ZTKS0156 / Z31820 / Rev.: 03
Assembling
5. Carefully remove the cold head from its box and place it upside down on a clean and stable work surface. Keep the box for future transportation.
6. Check if the copper flange (1) is assembled correctly and clean the surface.
7. Check if the insulation (2) is assembled cor­rectly.
8. Clean and grease the two O-rings (3) before mounting them on the cold head flange.
9. Assemble the three M 5 adjustment screws (4) to the cold head flange. The thread should not be visible at the O-ring side of the flange.
10. Turn the cold head carefully and set it on the copper flange.
11. Mount the cold head in the cold head turret respecting the orientation. Match the marks “X” on the cold head flange top side and on the cold head turret top flange.
12. Insert the cold head until rigid contact to the bottom of the cold head turret. The assembly is correct, if there is a gap of about 2 mm between cold head flange and turret flange.
The cold head tubes are sensitive against forces in any direction. Do not remove the
i
cold head insulation. Handle the cold head carefully.
Figure 4.35: Mounting the Cold Head - step 3
Figure 4.36: Mounting the Cold Head - step 4
ZTKS0156 / Z31820 / Rev.: 03
55
Assembling
13. Set the M 5 adjustment screws (1) to the turret flange and tighten them by one turn. The cold head is lifted 0.8 mm.
14. Fix the adjustment screws (1) with the three nuts on top of the cold head flange.
15. Attach the three M 6 fixing screws (3). Do not tighten these screws yet.
16. Slightly grease the O-ring and sealing sur­face of the KF 16 flange before assembling the pressure relief valve (2).
17. Assemble the pressure relief valve (2) to the KF 16 flange.
18. Slightly grease the O-ring and sealing sur­face of the KF 16 flange before mounting the stop valve (2).
19. Assemble the stop valve (2) to the KF 16 flange and open it.
1 Noise protection cover
2 Stand plates (each 3x)
3 Base supports (3x)
4 Noise protection cap
5 Base plate
6 Profile rail
7 Adapter plate
8 Front plates
9 Sliding blocks
10 Accessories
11 Profile cap
12 Velcro® fastener
13 Flex line positioning disc
14 Stand plates (2x)
Figure 4.37: Mounting the Cold Head - step 5
4.3.3.2 Mounting the Rotary Valve Column
Figure 4.38: Rotary Valve Column - overview
56
ZTKS0156 / Z31820 / Rev.: 03
Figure 4.39: Mounting the Rotary Valve Column - step 1
1. Put the base plate (1) upside down.
2. Arrange the three base supports (2) as shown in the figure with the tapered bores (3) on topside.
3. Mount the three base supports to the base plate with three M 8 x 25 screws (1) per sup­port.
4. Tighten the screws (torque 20 Nm).
5. Attach the spikes (1) into the middle thread of the three base supports.
Assembling
Figure 4.40: Mounting the Rotary Valve Column - step 2
Figure 4.41: Mounting the Rotary Valve Column - step 3
The spikes of the base plate should be protected with the supplied plastic cap nuts until
i
the rotary valve column is in its final position to prevent damage to the floor surface. These plastic cap nuts have to be removed before operating the magnet system.
ZTKS0156 / Z31820 / Rev.: 03
57
Assembling
6. Adjust the spikes to a height of 30 mm.
7. Fix the spikes with counter nuts.
8. Mount the stand plate (1) from the bottom side with three M 12 x 35 screws with washers (3) from the bottom side. Do not tighten these screws yet.
9. Continue with the same procedure for the second stand plate. Do not tighten these screws yet.
Figure 4.42: Mounting the Rotary Valve Column - step 4
Figure 4.43: Mounting the Rotary Valve Column - step 5
Figure 4.44: Mounting the Rotary Valve Column - step 6
58
ZTKS0156 / Z31820 / Rev.: 03
Figure 4.45: Mounting the Rotary Valve Column - step 7
10. Consider the bottom side of the profile with the two thread inserts (1).
11. Put two sliding blocks in each side track of the profile rail.
12. Put the assembled profile rail in-between the stand plates.
13. Attach the profile rail using four M 8 x 35 screws with washers (1) to the two stand plates. Do not tighten these screws yet.
14. Fix the profile rail to the base plate with the two M 8 x 35 screws with washers (2). Tighten the screws (torque 20 Nm).
15. Tighten the four M 8 x 35 screws (1) of the stand plates (torque 20 Nm).
16. Tighten the six M 12x35 screws of the stand plates (torque 50 Nm).
17. Put one sliding blocks (1) into the track at one side of the profile rail.
18. Attach two M 8 x 35 screws in the sliding block.
19. Adjust the sliding block on approximately half of the profile rail.
20. Tighten the two M 8 x 35 screws.
Assembling
Figure 4.46: Mounting the Rotary Valve Column - step 8
Figure 4.47: Mounting the Rotary Valve Column - step 9
ZTKS0156 / Z31820 / Rev.: 03
59
Assembling
21. Mount the flex line positioning disc (1) bet­ween the two screws on the sliding block as shown in the figure.
22. Put two sliding blocks (1) into the tracks on both sides of the profile rail.
23. Plug in the plastic cap (2) on the top side of the profile.
24. Assemble the front plate (1) to the two adap­ter plates (2) using six M 8 x 35 screws with washers (3) (torque 20 Nm).
25. Mount the preassembled adapter (1) using four M 8 x 35 screws with washers (2) to the two sliding blocks.
26. Adjust it as shown in the figure.
400 R: X = 166 mm 500 R: X = 206 mm
27. Fix the four screws (torque 20 Nm).
Figure 4.48: Mounting the Rotary Valve Column - step 10
Figure 4.49: Mounting the Rotary Valve Column - step 11
Figure 4.50: Mounting the Rotary Valve Column - step 12
60
Figure 4.51: Mounting the Rotary Valve Column - step 13
ZTKS0156 / Z31820 / Rev.: 03
4.3.3.3 Mounting the Rotary Valve and Flex Lines
1. Put down the Rotary Valve Column on the clean floor.
2. Mount the rotary valve (1) with four M 6 x 35 screws to the front plate (2) (torque 20 Nm).
3. Remove the flex lines from their box.
4. Thread the high (2) and low (3) pressure flex lines through the base plate and the flex line positioning disc (1) as shown.
5. Thread the motor cord (4) through the base plate and the flex line positioning disc (1) as shown.
6. Clean the male parts of the hose couplings (1) from the high and low pressure sockets of the rotary valve with ethanol.
7. Clean the sealing surface (2) with ethanol.
8. Check and clean the O-ring (3) with ethanol.
9. Check the correct position of the O-ring (3) before closing the coupling.
The Aeroquip® couplings of the connection line, rotary valve and cold head and the flex
i
lines are protected against damage and dust with plugs. Keep them for future transportation in the rotary valve box.
Assembling
Figure 4.52: Mounting the Rotary Valve and Flex Lines - step 1
Figure 4.53: Mounting the Rotary Valve and Flex Lines - step 2
Figure 4.54: Mounting the Rotary Valve and Flex Lines - step 3
ZTKS0156 / Z31820 / Rev.: 03
61
Assembling
10. Clean the thread (1) of the female part of the hose coupling of the flex lines with ethanol.
11. Clean the sealing surface (2) of the female part of the hose coupling with ethanol.
12. Grease the outside thread of the male part with PTFE-spray.
Figure 4.55: Mounting the Rotary Valve and Flex Lines - step 4
The thread surfaces of the house couplings have to be greased to reduce friction. But
i
grease on the sealing surfaces of the house couplings may enter the helium cycle and thus degrade the performance of the Cryogenic Refrigerator. Thus the sealing surfaces have to be protected during greasing. The easiest way to obey this advice is to connect (one turn) the male and female part as shown in the next figure or to use the supplied plugs of the Aeroquip® couplings. Alternatively a tissue may be used.
62
Figure 4.56: Mounting the Rotary Valve and Flex Lines - step 5
ZTKS0156 / Z31820 / Rev.: 03
Figure 4.57: Mounting the Rotary Valve and Flex Lines - step 6
13. Grease the outside thread of the female part behind the cap nut with PTFE-spray.
14. Connect the low pressure flex line to the low pressure socket of the rotary valve (short socket towards front plate).
15. Connect the high pressure flex line to the high pressure socket of the rotary valve (long socket behind front socket).
16. Tighten the cup nuts as shown and loosen them with 1/4 rotation.
17. Connect the motor cord to the socket of the rotary valve respecting the indexing.
18. Put up the rotary valve column carefully. Arrange the flex lines inside the hole of the base plate carefully. Prevent sharp bends to the flex lines and the motor cord. Prevent damage to the flex line or motor cord insula­tion.
19. Close the flex line positioning disc with the supplied Velcro® fastener.
Assembling
Figure 4.58: Mounting the Rotary Valve and Flex Lines - step 7
Figure 4.59: Mounting the Rotary Valve and Flex Lines - step 8
The sound-insulating material of the noise protection is pressure-sensitive. Damaging
i
the sound insulating material will reduce its performance. Thus: Prevent pressure on the sound insulating material. Prevent contact between the rotary valve column assembly during insertion into the noise protection cover.
ZTKS0156 / Z31820 / Rev.: 03
63
Assembling
20. Consider the weld of the noise protection cover as shown in the figure.
21. Put the noise protection cover (1) over the rotary valve. The weld in the noise protec­tion cover (2) should be on the front side of the rotary valve column (3) facing towards the cryostat.
22. Fix the noise protection cover with three special nuts M 6 from the bottom side to the base plate (3).
Figure 4.60: Mounting the Rotary Valve and Flex Lines - step 9
64
Figure 4.61: Mounting the Rotary Valve and Flex Lines - step 10
ZTKS0156 / Z31820 / Rev.: 03
Figure 4.62: Mounting the Rotary Valve and Flex Lines - step 11
23. Put the noise protection cap on the top of the rotary valve column.
24. Attach the insulation tube (1) to the connec­tion line of the rotary valve. Respect the orientation: put the female part towards the rotary valve.
25. If necessary cut away 40 mm on the male side of the insulation to allow free access to the Aeroquip® coupling.
26. Align the rotary valve column (1) relatively to the magnet stand and the cryostat (4) as shown. The connection line (2) between rotary valve column (1) and cold head (3) should not be stressed.
27. Remove the plastic cup nuts from the spikes of the rotary valve column.
28. Check the rotary valve column is levelled horizontally. If necessary adjust the levelling using the three spikes of the rotary valve column.
Assembling
Figure 4.63: Mounting the Rotary Valve and Flex Lines - step 12
Figure 4.64: Mounting the Flex Lines and the Connection Line: Alignment
ZTKS0156 / Z31820 / Rev.: 03
65
Assembling
29. Connect the connection line (3) to the cold head (2) following the instructions given in Figures 4.56-4.58 above. Prevent any torque to the connection line during tightening.
30. Loose the cup nut of the connection line with 1/4 rotation.
31. Assemble the two lower parts of the noise protection cover (1) on the cold head turret.
32. Check the groove inside the lower part of the noise protection cover is below the connection line (2).
33. Attach the two M 4 screws (3) to assemble the two lower parts of the noise protection cover.
34. Fix the lower parts of the noise protection cover with the four M 5 screws (1) to the cold head turret.
35. Attach the noise protection cover (1) over the cold head respecting the position.
Figure 4.65: Mounting the Noise Protection Cover - step 1
66
Figure 4.66: Mounting the Noise Protection Cover - step 2
Figure 4.67: Mounting the Noise Protection Cover - step 3
ZTKS0156 / Z31820 / Rev.: 03
4.3.3.4 Mounting the Compressor
1. Loosen the fixing straps from the compres­sor.
2. Remove the ring screws (1).
3. Use the supplied ramp (2) to roll out the compressor.
WARNING: Heavy Weights
For further information refer to the supplied manual of the Cryogenic Refrigerator.
The compressor is very sensitive against deviation from the power supply of more than
i
5 %. The compressor is also very sensitive against deviation from the pressure inside the helium circuit of more than 2 %. This can cause failure and damage. Thus: An applicable circuit breaker or fused disconnect switch must be provided on site to prevent deviant voltage situation. Check the compressor pressure while switched OFF. The gauges should read:
- 50 Hz compressor: 14.5 +/- 0.34 bar (210 +/- 5 PSI)
- 60 Hz compressor: 14.5 +/- 0.34 bar (210 +/- 5 PSI)
If the pressure is too high, vent helium to the set value. If the pressure is too low, add helium of high purity (He 5.0, 99.999%) to the set value. Contamination with other gases in the helium cycle will severely degrade the cold head‘s function and it will require factory servicing.
Assembling
Figure 4.68: Mounting the Compressor - step 1
ZTKS0156 / Z31820 / Rev.: 03
67
Assembling
4. Check the compressor pressure at the gauges on the front panel (1).
5. Prepare the compressor location. Ambient temperature must be in the range 7 to 38 °C (45 to 100 °F), compressor must be in a levelled position (within 5°).
6. Connect the low pressure flex line (2) to the compressor. Tighten cap nuts and loosen them with 1/4 rotation.
7. Connect the high pressure flex line (3) to the compressor. Tighten cap nuts and loosen them with 1/4 rotation.
8. Only compressor with water cooling option: connect the water lines (5) to the compres­sor respecting the flow direction.
9. Connect the motor cord (6) to the compres­sor.
10. Connect the compressor to the main power with the attached power cable (4).
11. If supplied mount the noise protection cover.
12. Cut the supplied Velcro® fastener in pieces of 400 mm to arrange and fix the flex lines and the motor cord.
Figure 4.69: Mounting the Compressor - step 2
Do not interchange the flex lines and their corresponding socket. Otherwise the
i
Cryogenic Refrigerator will not start.
68
ZTKS0156 / Z31820 / Rev.: 03
4.3.3.5 Mounting the Helium Level Sensor
1. Clean the O-ring (1) of the helium fill-in turret with ethanol.
2. Grease the O-ring (1) with vacuum grease.
3. Clean the sealing surface for the O-ring.
4. Mount the O-ring (1) into the sealing sur­face.
5. Insert the helium level sensor (1). Make sure the helium level sensor inserted into the cor­rect hole of the siphon.
6. Connect the connector (2) of the helium level sensor (1) with the connector (3) of the helium flow system (4).
7. Insert the helium level sensor (1) into the gap of the baffle (5) and insert the assembly into the helium fill-in turret.
8. Fix the black nut (1).
9. Calibrate the helium level sensor in the TopSpin Service Web. Refer to the supplied manual of TopSpin.
Figure 4.70: Mounting the Helium Level Sensor - step 1
Assembling
Figure 4.71: Mounting the Helium Level Sensor - step 2
Figure 4.72: Mounting the Helium Level Sensor - step 3
ZTKS0156 / Z31820 / Rev.: 03
69
Assembling
1. Clean the O-ring of the current lead turret (1) with ethanol.
2. Grease the O-ring (1) with vacuum grease.
3. Clean the sealing surface for the O-ring.
4. Mount the O-ring (1) into the sealing sur­face.
5. Mount the helium flow system (1) as shown with the one-way valve (2) showing backwards and the flexible side at the helium fill-in turret (7).
6. Check the correct position of the globes in the quench valves (6, 9) of the helium fill-in turret (7) and the current lead turret (8).
7. Connect the stop valve (4) with the supplied tube (3) with the connector at the one-way valve (2).
4.3.3.6 Mounting the Helium Flow System
Figure 4.73: Mounting the Helium Flow System - step 1
70
Figure 4.74: Mounting the Helium Flow System - step 2
ZTKS0156 / Z31820 / Rev.: 03
4.3.3.7 Mounting the Temperature Control
The temperature of the refrigerated radiation shield inside the cryostat during usual ope­ration must be controlled and in case of failure the operator will receive an alarm. Set the alarm temperature level to 90 K at MICS and at the External Temperature Control Unit.
The alarm will be given in four different ways:
Alarm message at the computer of the magnet system via MICS.
Alarm message by email generated via MICS.
Alarm message at the display of the External Temperature Control Unit.
Audible alarm from the External Temperature Control Unit.
The cryostat has two PT 100 temperature sensors for monitoring. One sensor is con­nected to the BSMS board of the console with supplied VTA MAG-RS unit and Cable Set BSVT, controlled with MICS. The other sensor is connected to the External Temperature Control Unit with the supplied cable.
Assembling
The two PT 100 sensors should be used in a parallel manner to prevent any lack of
i
monitoring due to failure of the sensors or the power supply. Thus: connect the console to another electrical circuit than the External Temperature Control Unit.
ZTKS0156 / Z31820 / Rev.: 03
71
Assembling
1. Connect the VTA MAG-RS (figure below) with the S1 and the S3 / H1 port at the feed­through plate respecting the number of pins.
2. Connect the VTA MAG-RS to the BSMS board at the console with the supplied BSVT cable.
3. Connect the S2 port of the feedthrough plate to the External Temperature Control Unit with the supplied cable.
For further steps follow the instructions given in the supplied manual of the External Temperature Control Unit.
Figure 4.75: Mounting the Temperature Control - step 1
Figure 4.76: Mounting the Temperature Control - step 2
Installation has been completed correctly, if both readings of MICS and the External
i
Temperature Control Unit show the same temperature values.
72
ZTKS0156 / Z31820 / Rev.: 03

4.3.4 Cooling down from 300 K to 80 K

1 Current lead turret
2 Helium fill-in turret
3 ACD (Automatic Cooling
Device)
4 Transportation dewar with
liquid nitrogen
5 Transportation dewar with
warm nitrogen gas
WARNING
Cryogenic Agents (see page 2-21)
Gas under Pressure (see page 2-22)
Preconditions
Magnet Stand mounted (refer to the manual of the Magnet Stand).
Helium flow system mounted (see chapter ”Mounting the Helium Flow System” on page 4-70).
Cryostat is evacuated with a vacuum less than 5 x 10
Assembling
-8
bar.
Cryogenic Refrigerator mounted completely.
Transportation dewar with sufficient dry and warm nitrogen gas.
Transportation dewar with sufficient warm helium gas.
ACD (Automatic Cooling Device) mounted on a transportation dewar with sufficient liquid nitrogen.
Figure 4.77: Assembly for cooling down from 300 K to 80 K with ACD
ZTKS0156 / Z31820 / Rev.: 03
73
Assembling
1 Current lead
2 One way valve
3 Stop valve
4 Cold Head
5 Precooling tube
6 Quench valves
Procedure cooling down the helium vessel:
1. Check the correct setting of the globes inside the quench valves (6).
Figure 4.78: Helium Flow System during cool down
2. Remove the connector and the tube to the stop valve.
3. Check the helium flow system is mounted like shown in the figure above.
4. Check the cold head (4) is in the lifted position (refer to ”Mounting the Cold Head” on page 4-54).
5. Close the vacuum valve (refer to ”Removing the Vacuum Valve” on page 4-53) and remove the vacuum pumping unit from the vacuum valve.
6. Connect the stop valve at the cold head flange with the vacuum pumping unit.
7. Apply a vacuum of 0.01 bar inside the cold head turret.
8. Pump and flush the cold head turret two times with warm helium gas.
9. Close the stop valve at the cold head flange.
10. Connect the stop valve at the cold head turret with a transportation dewar with warm helium gas. Apply a pressure of 0.1 bar (1.5 PSI) at the transportation dewar.
11. Open the stop valve at the cold head flange.
12. Remove the plug on top of the helium fill-in turret.
13. Insert the precooling tube into the helium fill-in turret carefully. Check the precooling tube is completely inserted. Check the O-ring sealing the precooling tube properly.
14. Connect the precooling tube (5) to a pressure gauge.
15. Connect the vacuum pumping unit at the one way valve (2).
16. Apply a vacuum of 0.01 bar inside the helium vessel.
74
17. Pump and flush the helium vessel one time at room temperature with warm nitrogen gas to make sure that the helium vessel is dry before cooling down.
18. Connect the vacuum pumping unit to the vacuum valve and continue to generate vacuum (refer to ”Generating the Vacuum” on page 4-52).
19. Insert the auxiliary shorting plug into the helium fill-in turret (right turret).
20. Remove the shorting plug out of the current lead turret (left turret).
ZTKS0156 / Z31820 / Rev.: 03
Assembling
31. Turn the three M 5 adjustment screws (1) at the cold head flange (3) counter clockwise until the thread is not longer visible at the bottom side of the cold head flange.
32. Tighten the three M 6 screws (2) (torque less than 1 Nm) to push down the cold head in operating position. The assembly is cor­rect, if there is a gap of about 2 mm between cold head flange (3) and turret flange (4).
33. Tighten the three M 5 adjustment screws (1) (torque less than 1 Nm) against the cold head flange (3).
34. Fix the three M 5 adjustment screws (1) with the nut.
21. Insert the current lead into the current lead turret respecting the correct position. The current lead can only be inserted in one position.
22. Remove the auxiliary shorting plug out of the helium fill-in turret.
23. Connect the ACD with the current lead (1).
24. Connect the current lead to the Bruker power supply.
25. Check resistance of the magnet coils and shims at room temperature (see Appendix Technical Data).
26. Check the zero readings of the PT 100 temperature sensors (see Appendix Technical Data).
27. Check zero reading of the helium level sensor.
28. Check the connection between the PT 100 connector at the current lead and the ACD. The ACD must show approximately room temperature (~293 K).
29. Connect the precooling tube (5) with the liquid nitrogen transportation dewar.
30. Adjust the pressure of the liquid nitrogen transportation dewar to 0.2 - 0.3 bar (1 - 2 PSI). => liquid nitrogen transfer starts.
Precooling with liquid nitrogen reduces the consumption of liquid helium during cooling
i
down.
Time for cooling down to 80 K with liquid nitrogen will take 18 to 24 hours.
i
Figure 4.79: Adjusting the Cold Head
35. Set the Cryogenic Refrigerator into operation. Operating properly, the Cryogenic Refrigerator will emit a rhythmic squeak or chirp. This noise is an indication of the proper flow of the gas inside the system.
ZTKS0156 / Z31820 / Rev.: 03
75
Assembling
36. Check at the high and low pressure gauges on the compressor front panel if the mean values are different and if an oscillation of the needles is noticeable. This difference will decrease as the cold head cools down. The typical pressure difference are approximately: 50 Hz compressor: 14.5 to 16.5 bar (210 to 240 PSI) with a 0.34 to 0.7 bar (5 to 10 PSI) amplitude on the needles. 60 Hz compressor: 14.5 to 16.5 bar (210 to 240 PSI) with a 0.34 to 0.7 bar (5 to 10 PSI) amplitude on the needles. For further information refer to the supplied manual of the Cryogenic Refrigerator.
37. Check the two PT 100 temperature sensors for monitoring the cooling down.
38. The ACD will stop the liquid nitrogen transfer automatically at a temperature of 80 K to prevent the accumulation of liquid nitrogen inside the helium vessel. The magnet’s temperature monitored with the two PT 100 temperature sensors should be between 80 K and 90 K.
In case of cooling down without Automatic Cooling Device (ACD):
i
- connect the two PT 100 temperature sensors to a measuring device (not supplied)
- check the PT 100 temperature sensor: temperature difference between upper and
lower temperature sensor maximal 25 K. Adjust the cooling speed with the liquid nitrogen flow.
- stop liquid nitrogen transfer at a temperature of 80 K at the lower PT 100 temperature
sensor.
39. Switch off the ACD and remove the connection to the liquid nitrogen transportation dewar at the precooling tube (5).
40. Remove the one way valve (2) from the helium flow system and connect the nitrogen gas transportation dewar to the helium flow system.
41. Blow in nitrogen gas with a pressure of 0.1 - 0.15 bar (1 PSI) and check if liquid nitrogen spills out of the precooling tube (5). In case of liquid nitrogen spilling out warm up the system to 90 K to make sure that no liquid nitrogen is left inside the helium vessel.
42. Check the temperature of the PT 100 temperature sensors.
43. Remove the nitrogen gas transportation dewar at the helium flow system and remount the one way valve (2).
44. Connect the precooling tube (5) to a pressure gauge.
Keep the helium flow system closed at all times if the cryostat is below room tem-
i
perature to prevent moisture entering the helium vessel. If necessary open the helium flow system as short as possible. Do not open both helium turrets simultaneously. Keep the cold head turret pressurized at all times if the cryostat is below room tem­perature.
76
ZTKS0156 / Z31820 / Rev.: 03

4.3.5 Cooling down to 4.2 K

1 Connector
2 One way valve
3 Stop valve
4 Cold Head
5 Quench valves
Preconditions
Helium vessel filled with cold nitrogen gas.
Current lead inserted into the current lead turret.
Power supply connected to the current lead.
Upper and lower PT 100 temperature sensors show 80 - 90 K.
Assembling
Vacuum pumping unit connected; cryostat evacuated with vacuum less than
Stop valve (3) at the cold head flange connected with the transportation dewar with
Transportation dewar with sufficient dry and warm helium gas ready.
Transportation dewar with sufficient liquid helium ready.
Procedure
1. Close the vacuum valve (refer to ”Removing the Vacuum Valve” on page 4-53)
2. Connect the one way valve (2) at the helium flow system with the vacuum pumping
-8
5x10
warm helium gas.
and remove the vacuum pumping unit from the vacuum valve.
unit.
bar).
Figure 4.80: Helium Flow System during usual operation
3. Pump and flush the helium vessel with warm helium gas 5 times:
4. Connect the vacuum pumping unit to the vacuum valve and continue to generate
5. Check the zero reading of the helium level sensor.
6. Insert the magnet side of the helium transfer line completely into the helium fill-in
ZTKS0156 / Z31820 / Rev.: 03
- Fill the helium vessel with dry helium gas until the quench valves open.
- Pump to the following pressures: 0.8, 0.5, 0.3, 0.1, 0.01 bar (11, 7, 4, 1.5, 0.1 PSI).
- Use short pumping times!
vacuum (refer to ”Generating the Vacuum” on page 4-52).
turret.
77
Assembling
7. Insert the other side of the helium transfer line into the helium transportation dewar slowly. Fast insertion of the warm transfer line causes evaporation of helium and may cause overpressure inside the transportation dewar.
8. Adjust a pressure of 0.1 - 0.2 bar (1.5 - 3 PSI) at the helium transportation dewar. => Helium transfer starts.
Time for cooling down to 4.2 K with helium will take 4 to 6 hours.
i
9. Use the IBT temperature sensor to monitor the cooling down (see chapter Appendix Tech n i c a l D a t a ) .
10. Stop helium transfer at a helium level of 100 % at the helium level sensor and 4.2 K at the IBT temperature sensor.
11. Remove the pumping unit and the vacuum valve (refer to ”Removing the Vacuum Valve” on page 4-53).
12. Close the stop valve at the cold head turret and remove the helium gas transporta­tion dewar.
13. Mount the supplied connector (1) to the helium flow system and connect the con­nector with the stop valve at the cold head flange (3).
14. Open the stop valve at the cold head flange (3).
If the magnet system will be energized during the next 24 h, the current lead can remain
i
in the current lead turret. Otherwise continue.
15. Remove the plug out of the turret cap of the helium fill-in turret.
16. Insert the auxiliary shorting plug into the helium fill-in turret.
17. Remove the current lead out of the current lead turret.
18. Insert the shorting plug into the current lead turret with the shorting plug tool respec­ting the correct position. The shorting plug can only be inserted in one position.
19. Close the current lead turret with the baffle tube, O-ring and screw cap.
20. Remove the auxiliary shorting plug out of the helium fill-in turret.
21. Close the helium fill-in turret immediately with the plug of the turret cap.
22. Record the filling session (see Appendix, Technical Data).
23. Observe the temperature of the outer radiation shield at the External Temperature Control Unit. The radiation shield will reach its usual operation temperature after four days operating. Turn on the alarm from the console and the External Temperature Control Unit if the magnet system is at set operation temperature.
78
ZTKS0156 / Z31820 / Rev.: 03

5Operation

5.1 Safety

Approved Persons:
Bruker Service, Approved Customer Personnel
Magnetic Fields (see page 2-20)
Cryogenic Agents (see page 2-21)
Electricity (see page 2-21)
Gas under Pressure (see page 2-21)
WARNING
Personal Protective Equipment:
Protective goggles
Protective gloves
Safety shoes
Portable oxygen monitor and alarm
Necessary Tools and Equipment
Bruker Power Supply
Auxiliary Shorting Plug
Current Lead
Shorting Plug Tool
ZTKS0156 / Z31820 / Rev.: 03
79
Operation
1 Current lead
2 Connector
3 One way valve
4 Stop valve
5 Cold Head
6 Quench valves

5.2 Energizing

Precondition
Cryostat at 4.2 K for at least 12 hours.
Helium level at the helium level sensor is above the "minimum level during energizing" (see Appendix Technical Data of the supplied User Manual).
Both PT 100 temperature sensors show the same set value.
Cryogenic Refrigerator operating correctly.

5.2.1 Inserting the Current Lead

1. Prepare the Bruker Power Supply and main power cable.
Figure 5.1: Helium Flow System during energizing
2. Remove the plug out of the turret cap of the helium fill-in turret.
3. Insert the auxiliary shorting plug into the helium fill-in turret.
4. Remove the shorting plug out of the current lead turret.
5. Insert the dry current lead into the current lead turret respecting the correct position. The current lead can only be inserted in one position.
6. Connect the main power cable and control cables of the Bruker Power Supply with the current lead.
7. Connect the short circuit cable to the main power cable until energizing.
8. Remove the auxiliary shorting plug out of the helium fill-in turret.
9. Close the helium fill-in turret with the plug of the turret cap.
10. Switch the Bruker Power Supply ON.
80
ZTKS0156 / Z31820 / Rev.: 03

5.2.2 Energizing Procedure

1. Check the helium level is above "minimum level during energizing" (see Appendix
Technical Data of the supplied User Manual) regularly during energizing. Refill the helium vessel if necessary.
2. Check the settings (current and inductivity) of the Bruker Power Supply:
- Check the main heater current. The current of the main heater must be the same for the magnet coil and the Bruker Power Supply.
- Check the shim heater current. The current of the shim heater must be the same for the shim heater and the Bruker Power Supply.
3. Make the main coil heater switch test.
4. Open the main coil heater switch.
5. Start the shim heater automatic.
6. Switch the Z and Z
7. Set the sense voltage to the first value of the energizing assignment (see Appendix
Technical Data of the supplied User Manual).
2
-Shim heater ON.
Operation
8. Set the current to the first value of the energizing assignment (see Appendix
Technical Data of the supplied User Manual).
9. Wait until the sense current at the Bruker Power Supply is 0 V.
10. Continue in the same way with the other values and the overshoot of the energizing
assignment.
11. Check the sense voltage is exactly 0 V for 5 seconds.
12. Close the main coil heater switch.
13. Wait 5 minutes.
=> Main coil heater switch is closed. Main coil is now persistent.
14. Set the current to 0 A at the Bruker Power Supply.
15. Wait until the Bruker Power Supply shows 0 A.
16. Remove the main power cable from the Bruker Power Supply and short-circuit it.
17. The shim automatic must remain switched ON for 12 h after energizing.
Recommended for magnet systems of 500 MHz and more: Continue with cycling, see chapter ”Cycling Procedure” on page 5-82.
ZTKS0156 / Z31820 / Rev.: 03
81
Operation
Interrupt Energizing
1. Set the current of the Bruker Power Supply to the value of the coil current.
2. Wait until the voltage is 0 V.
3. Close the main heater switch.
4. Wait 5 minutes.
Continue Energizing
1. Set the current of the Bruker Power Supply to the value of the coil current. Wait until
2. Open the main heater switch.
3. Set sense voltage to 30 mV.
4. Shim heater automatic: Z and Z
5. Continue according to the energizing assignment.
=> Main heater switch is closed. Main coil is now persistent.
the current of the Bruker Power Supply is reached.
2
heater ON.

5.2.3 Cycling Procedure

Cycling only recommended for magnet systems of 500 MHz and more.
i
1. Check the time between energizing and cycling was at least 12 hours.
2. Check the helium level is above "minimum level during energizing" (see Appendix Technical Data of the supplied User Manual). Refill the helium vessel if necessary.
3. Cycle all Z shims with currents according the cycling assignment (see Appendix Technical Data of the supplied User Manual).
4. Allow the shims 2 minutes to hold these currents.
5. Discharge shims.
6. After energizing/de-energizing the shims at least two times start the shim automatic.
7. Stop shim automatic if procedure has been completed.
8. Wait at least 12 hours before shimming.
Continue with shimming, see chapter ”Shimming Procedure” on page 5-83.
82
ZTKS0156 / Z31820 / Rev.: 03

5.2.4 Shimming Procedure

1. Check the time between cycling was at least 12 hours.
2. Check the settings (current and inductivity) of the Bruker Power Supply. The current
must be the same for the shim coil and the Bruker Power Supply.
3. Switch the shim heater ON.
4. Wait 5 seconds (default setting of the Bruker Power Supply).
5. Set the shim currents according to the shimming assignment (see Appendix
Technical Data of the supplied User Manual).
6. Wait until the current is reached.
7. Leave the shim heater open for 1 minute with the correct shim current before
closing it.
8. Switch the shim heater OFF.
9. Wait 5 seconds (default setting of the Bruker Power Supply).
10. Set the current to 0 A.
Operation
11. Wait until the supply current is 0 A.
12. Continue in the same way with all shims.
ZTKS0156 / Z31820 / Rev.: 03
83
Operation
1 Connector
2 One way valve
3 Stop valve
4 Cold head
5 Quench valves

5.2.5 Removing the Current Lead

1. Check the helium level is above "minimum level during energizing" (see Appendix Technical Data of the supplied User Manual). Refill the helium vessel if necessary.
2. Insert the auxiliary shorting plug into the helium fill-in turret.
3. Check the current at all Bruker Power Supply cables is 0 A.
4. Switch the Bruker Power Supply OFF.
5. Remove all cables from the current lead.
6. Remove the current lead as quickly as possible.
7. Insert the dry warm shorting plug with the shorting plug tool into the current lead tur­ret respecting the correct position. The shorting plug can only be inserted in one position.
8. Close the current lead turret with the turret cap. Be careful with the baffles.
9. Remove the auxiliary shorting plug out of the helium fill-in turret.
10. Close the helium fill-in turret immediately with the plug of the turret cap.

5.2.6 Complete Energizing

1. Refill helium to a value of 95 % at the helium level sensor.
2. Check the helium flow system correctly mounted as shown.
3. Tighten the three M 6 screws (torque less than 1 Nm) again to push down the cold head in operating position. The assembly is correct, if there is a gap of about 2 mm between cold head flange and turret flange.
4. Tighten the three M 5 adjustment screws (torque less than 1 Nm) and fix them at the cold head flange with the nuts.
84
Figure 5.2: Helium Flow System during usual operation
ZTKS0156 / Z31820 / Rev.: 03

5.3 Set into Operation

Set the magnet stand into operation by switching the pneumatic controller to UP position.
For any work at the magnet system like main­tenance or refill of helium stop the magnet stand by switching the pneumatic controller to DOWN position.
Mount the further equipment of the supplied magnet system respecting their manuals.
Figure 5.3: Start the Magnet Stand
Operation
Figure 5.4: Stop the Magnet Stand
ZTKS0156 / Z31820 / Rev.: 03
85
Operation
86
ZTKS0156 / Z31820 / Rev.: 03

6 Troubleshooting

Troubleshooting must be performed only with approved qualification.
In case of doubts or problems not specified in the following list contact Bruker Service immediately. For contact information see page 9 of this manual.

6.1 Safety

Approved Persons
Bruker Service, Approved Customer Personnel
WARNING
Magnetic Fields (see page 2-20)
Cryogenic Agents (see page 2-21)
Electricity (see page 2-21)
Gas under Pressure (see page 2-22)
Spontaneous Ignition and Explosion (see page 2-23)
Personal protective equipment:
Protective goggles
Protective gloves
Protective clothes
Safety shoes
ZTKS0156 / Z31820 / Rev.: 03
87
Troubleshooting

6.2 Problem

6.2.1 During Transportation

Indicator Possible reason Solution By
Tilt Watch / Shock Watch activated.
Visible damage. Careless transporta-
Careless transporta­tion.
tion.

6.2.2 During Assembling

1. Accept delivery with reservation.
2. Remark the extent of damage in the transport documents.
3. Start complaint process.
1. Accept delivery with reservation.
2. Remark the extent of damage in the transport documents.
3. Start complaint process.
Approved Customer Personnel
Approved Customer Personnel
Indicator Possible reason Solution By
Ceiling height too low for assembling on magnet stand.
Ceiling height too low for inserting the helium level sensor.
Helium bore tube and radiation shield are not concentric.
Continued on next page
88
Site does not meet the required condi­tions.
Site does not meet the required condi­tions.
Alignment is not correct.
Choose another site that meets the required condi­tions.
Insert the helium level sensor before mounting the magnet stand.
Check fixation of the alignment rods.
Bruker Service
Bruker Service
Bruker Service
ZTKS0156 / Z31820 / Rev.: 03
Troubleshooting
Continued from page before
Indicator Possible reason Solution By
Helium bore tube and radiation shield are not concentric.
Vacuum valve collides with the magnet stand.
Vacuum in RT vessel does not reach
-8
5x10
bar within
48 hours.
Alignment rod is loose or broken.
Reduction flange is not concentric.
Vacuum valve collides with the magnet stand.
O-rings may be damaged.
Defective pumping line.
Defective pumping unit.
Replace alignment rod
a
.
Bruker Service
Check orientation. Bruker Service
Turn the vacuum valve. Be
Bruker Service careful if the RT vessel is evacuated.
Check and clean O-rings and
Bruker Service slots; replace O-rings if necessary:
of the vacuum valve
of the drop-off plate
of the reduction and
sealing flanges
of the bottom plate
Check pumping line between
a
Bruker Service the cryostat and vacuum pumping unit; change if necessary.
Check pumping unit:
A pressure below 1 x 10
-9
Bruker Service
bar must be reached with a closed sealing plug.
Room temperature bore tube has scrat­ches or dust on the sealing surfaces.
Moisture within the RT vessel.
Check sealing surfaces on the room temperature bore tube: No scratches and no dust should be visible.
Pump and flush the RT vessel several times with dry nitro-
Bruker Service
Bruker Service
gen gas.
Super insulation touches RT vessel or bore tube or radiation shield.
Super insulation was not fixed correctly during assembly.
Fix super insulation on the outer radiation shield with
polyester tape
a
. Carefully
prevent any connection
Bruker Service
between different vessels or bore tubes in the cryostat.
a. For this work the bottom plate has to be removed. Check the suspension tubes of the helium vessel are not broken. Install the safety device for fall protection (not supplied). Contact Bruker Service Headquarter for further information.
ZTKS0156 / Z31820 / Rev.: 03
89
Troubleshooting

6.2.3 During Cool Down

Indicator Possible reason Solution By
Precooling with liquid nitrogen conti­nue too slowly.
Precooling with liquid nitrogen conti­nue too quickly.
Vacuum in RT vessel does not reach
-8
5x10
bar.
Empty transport dewar.
Transfer pressure too low.
Transport dewar is leaky; no transfer pressure may be applied.
Transfer pressure too high.
O-rings may be leaky.
O-rings may be fro­zen due to contact with liquid nitrogen.
Refill or change transport dewar.
Increase transfer pressure slightly.
Check transfer dewar and change if necessary.
Stop precooling. Adjust correct transfer pressure.
Check and clean O-rings and slots; change O-rings if necessary:
of the vacuum valve
of the drop-off plate
of the reduction and sealing flanges
of the bottom plate
a
1. Stop precooling.
2. Warm up O-ring with warm air
3. Wait until the vacuum recovers.
4. Prevent liquid nitrogen from splashing on O-rings.
Bruker Service
Bruker Service
Bruker Service
Bruker Service
Bruker Service
Bruker Service
90
Defective pumping line.
Defective pumping unit.
a. see note on page before
Continued on next page
Check pumping line between the cryostat and vacuum pumping unit; change if necessary.
Check pumping unit:
A pressure below5 x 10
-8
bar must be reached with a closed sealing plug.
Bruker Service
Bruker Service
ZTKS0156 / Z31820 / Rev.: 03
Troubleshooting
Continued from page before
Indicator Possible reason Solution By
RT vessel becomes cold and wet.
Cold spot in the RT-bore.
The helium flow system becomes very cold and icy during pumping and flushing with helium gas.
After some intervals of pumping and flus­hing it is not possible to reach a vacuum in the range of
0.001 bar.
Vacuum is broken or less than
-6
1x10
bar.
Cold leak after transportation.
Alignment not correct.
liquid nitrogen remains in the helium vessel, boiling off strongly during pumping.
The globes in the quench valves are not fitting correctly in the O-rings and thus the quench valves are leaky.
Liquid nitrogen remains in the helium vessel, boiling off strongly during pumping.
Do not remove pumping unit until liquid helium fill­in is finished.
Continue as in problem
Vacuum in RT vessel does not reach
-8
5x10
bar above
1. Stop cool down.
2. Warm up cryostat.
1. Stop cool down.
2. Warm up cryostat.
3. Align the vessels.
1. Stop pumping.
2. Carefully remove all liquid nitrogen through the precooling tube.
3. Check with the dipstick to be sure that the helium vessel is completely empty of liquid nitrogen and of frozen nitrogen (nitrogen ice).
1. Stop pumping.
2. Remove frozen air and frozen moisture with warm helium gas.
3. Slightly grease the O-rings and check the position of the globes.
4. Check with the dipstick to be sure that the helium vessel is completely empty of liquid nitrogen and of nitrogen ice.
1. Stop pumping.
2. Carefully remove all liquid nitrogen through the precooling tube.
3. Check with the dipstick to be sure that the helium vessel is completely empty of liquid nitrogen and of frozen nitrogen (nitrogen ice).
Bruker Service
Bruker Service
Bruker Service
Bruker Service
Bruker Service
Bruker Service
Continued on next page
ZTKS0156 / Z31820 / Rev.: 03
91
Troubleshooting
Continued from page before
Indicator Possible reason Solution By
Nitrogen ice in the helium vessel.
Transfer of liquid helium does not start.
Pumping intervals during pumping and flushing were too long and remaining nitrogen was boiling off and got frozen.
The transport dewar is empty.
The transfer pres­sure in the transport vessel is too low.
The transfer dewar is leaky, there is no transfer pressure built up.
1. Warm up the magnet coil with warm helium gas through the precooling tube until the whole coil is warmer than 90 K.
2. Repeat pumping and flushing and carefully check with the dipstick to be sure that the helium vessel is completely empty of liquid nitrogen and of frozen nitrogen (nitrogen ice).
Refill or change transport dewar.
Increase the transfer pres­sure.
Check the transport dewar for leakage. Tight all connec­tions.
Bruker Service
Bruker Service
Bruker Service
Bruker Service
The cool down of the magnet coil does not continue although helium is transferred.
Continued on next page
The siphon or the helium transfer line are blocked with ice.
The helium transfer line is defective.
The extension is not mounted on the helium transfer line.
Check the siphon and helium transfer line for blockages, remove ice with warm helium gas.
Check the helium transfer line for icing. If there are cold spots visible, replace the helium transfer line.
Mount the extension piece on the helium transfer line. Check the helium transfer line to be inserted completely into the siphon.
Bruker Service
Bruker Service
Bruker Service
92
ZTKS0156 / Z31820 / Rev.: 03
Troubleshooting
Continued from page before
Indicator Possible reason Solution By
The zero reading of the helium level sen­sor can not be adju­sted at the beginning of cooling down with liquid helium.
The helium level does not reach 100% after cooling down.
After cool down the helium boil off is higher than specified (up to 5 times).
The helium level sensor is not con­nected correctly with the connector in the helium flow system.
The helium level sensor is defective.
The transport dewar is empty, no more helium is transferred.
The helium level sensor is disturbed by the transfer line’s extension piece.
Usual behavior. A few days are necessary for the radiation shields and the insulation to reach scheduled temperatures.
Check the connection in the helium fill-in turret between helium level sensor and con­nector.
Check the helium level sensor with the 0% calibration plug.
Refill or change transport dewar.
1. Stop the liquid helium
transfer.
2. Remove the transfer line.
3. Measure the helium level
after some minutes without the transfer line.
Wait a few days and check helium boil off. The presence of the current lead in the current lead turret during energizing and shimming helps to cool down the radiation shield due to higher helium flow.
Bruker Service
Bruker Service
Bruker Service
Bruker Service
Bruker Service
Temperature of the radiation shield decreases too slowly (if T
> 250 K after
RS
pre-cooling with liquid nitrogen)
2 days after cool down the T
RS
is still
higher than set value (alarm of External Temperature Control Unit)
Cryogenic Refrigera­tor not operating.
Cryogenic Refrigera­tion operating not correct
Cryogenic Refrigera­tor performance not sufficient.
Alarm default set­tings of the External Temperature Control Unit or the console not correct.
Cold head not moun­ted correctly.
Start Cryogenic Refrigerator. Bruker Service
See ”During Operation of the
Bruker Service Cryogenic Refrigerator” on page 6-102.
See ”During Operation of the
Bruker Service Cryogenic Refrigerator” on page 6-102.
Check set values (see
Bruker Service separate manual of the Exter­nal Temperature Control Unit and of the console)
See chapter ”Mounting the
Bruker Service Cryogenic Refrigerator Parts” on page 4-54.
ZTKS0156 / Z31820 / Rev.: 03
93
Troubleshooting

6.2.4 During Energizing and Shimming

Indicator Possible reason Solution By
The current lead can not be inserted completely into the connector.
Main coil heater test fails.
Setting of sense voltage fails.
The connector is covered with ice. (frozen moisture or nitrogen ice).
The shorting plug was not removed.
The orientation of the current lead is not correct.
Defective power supply.
Connector or cables defective.
The main coil heater switch is "OFF". The main coil switch is not opened.
The main coil heater current is not cor­rect. The main coil switch is not opened.
Carefully remove the ice with warm helium gas. To remove small ice spots use the dipstick or the precooling tube as tubing for the warm helium gas.
Remove the shorting plug with the shorting plug tool.
Turn the current lead carefully until it can be inserted correctly into the connector.
Replace the power supply Bruker Service
Clean connectors or change cables if necessary.
Switch the main coil heater to "ON" and check the main coil heater current to be adjusted correctly.
Adjust main coil heater current correctly.
Bruker Service
Bruker Service
Bruker Service
Bruker Service
Bruker Service
Bruker Service
Current lead can not be removed.
Continued on next page
94
The auxiliary shor­ting plug is inserted in the current lead turret by mistake and makes a short circuit across the main coil.
The connector is covered with ice (frozen moisture or nitrogen ice).
Remove the auxiliary shor­ting plug and insert it in the helium fill-in turret.
Carefully remove the ice with warm helium gas over the helium flow system. To remove small ice spots from the connector use the dipstick or the precooling tube as tubing for the warm helium gas.
Bruker Service
Bruker Service
ZTKS0156 / Z31820 / Rev.: 03
Troubleshooting
Continued from page before
Indicator Possible reason Solution By
Shorting plug cannot be removed.
The magnet system quenches
The main coil switch can not be closed on field.
The connector is covered with ice (frozen moisture or nitrogen ice).
Loss of superconductivity.
The helium level was too low for energi­zing, cycling, shimming, de-energizing.
The power supply is defective. The main current is oscillating.
The helium level is too low for energi­zing. The main coil switch is not covered with liquid helium.
Carefully remove the ice with warm helium gas. To remove small ice spots use the dip­stick or the precooling tube as tubing for the warm helium gas.
See chapter ”After a Quench” on page 6-107.
See chapter ”After a Quench” on page 6-107.
Replace the power supply. Bruker Service
Never try to energize the magnet with less than the "minimum level during energi­zing" in the helium vessel.
Bruker Service
Bruker Service
Bruker Service
Bruker Service
Shim current can not be set correctly.
Shims do not affect the NMR signal.
Continued on next page
The power supply is defective. The main current is oscillating.
The control cable is not connected cor­rectly to the current lead or to the power supply.
Switch „Main Coil/ OFF/Shim Coil“ in wrong position.
Shim heater current is not correct. The shim switches are not opened.
Replace the power supply. Bruker Service
Connect the control cable cor­rectly to current lead and power supply.
Change the switch position Bruker Service
Set the shim heater current to the specified value (see Appendix Technical Data)
Bruker Service
Bruker Service
ZTKS0156 / Z31820 / Rev.: 03
95
Troubleshooting
Continued from page before
Indicator Possible reason Solution By
Magnet system does not reach specifica­tion.
Magnetic material inside RT bore tube.
Large ferromagnetic parts are in the vicinity of the magnet system.
Carefully clean the RT bore tube.
1. Keep the maximum pos­sible distance between the magnet system and ferromagnetic parts.
2. Repeat shimming.

6.2.5 During Operation of the Magnet Stand

In case of doubt contact Bruker Service and refer to the supplied manual of the Magnet Stand
Indicator Possible reason Solution By
The NMR spectrum shows massive disturbances.
The pneumatic con­troller is in DOWN position.
Magnet system has direct mechanical contact with the floor via accessories.
Switch pneumatic controller to UP position.
Identify and eliminate contact point. Arrange lines in loose S- or U shapes
If the problem is still not solved, contact Bruker Service.
Bruker Service
Bruker Service
Approved Customer Personnel
Approved Customer Personnel
Approved Customer Personnel
96
Magnet system has physical contact to the magnet stand.
Piston of the isolator is not centric or touches its casing.
T-safety brakets touches the pillar.
Floor vibrations in vertical direction.
Floor vibrations in horizontal and ver­tical direction.
Continued on next page
Check leveling; adjust if necessary.
Align magnet stand. Bruker Service
Align magnet stand. Bruker Service
Replace elastomeric isolators with air damped isolators.
Replace air damped isolators with air piston isolators.
Bruker Service
Bruker Service
Bruker Service
ZTKS0156 / Z31820 / Rev.: 03
Troubleshooting
Continued from page before
Indicator Possible reason Solution By
The isolator of the magnet stand does not reach the opera­ting position.
Magnet system achieves working position jerkily.
Pneumatic con­troller in DOWN position.
The pressure of the gas supply is too low.
The magnet system is not leveled cor­rectly.
Defective leveling valve.
Defective membrane of an isolator.
Piston is not centric or touching its casing.
Switch pneumatic controller to UP position.
Check the pressure of the gas supply. It must be in the range of 5to8bar (70to112psi).
If the problem is still not solved, contact Bruker Service.
Stop the pneumatic isolators. Check the leveling of the cryostat.
Exchange leveling valve or isolator.
Exchange leveling valve or isolator.
Align magnet stand. Bruker Service
Approved Customer Personnel
Approved Customer Personnel
Approved Customer Personnel
Bruker Service
Bruker Service
Bruker Service
Audible loss of air Defective membrane
or defective leveling valve of an isolator.
Hose connector is defective or loose.
Velocity of lifting or lowering too high.
Wrong adjustment of the flow control valve.
Exchange leveling valve or isolator.
Insert hoses correctly and tighten screws.
Close restrictor of the flow control valve completely; then open it a half turn.
Bruker Service
Bruker Service
Bruker Service
ZTKS0156 / Z31820 / Rev.: 03
97
Troubleshooting

6.2.6 During Standard Operation

Indicator Possible reason Solution By
The helium boil off decrease to zero.
The atmospheric pressure is increa­sing.
The helium flow system is covered with ice.
The helium flow system or the sus­pension tubes are blocked with ice.
Usual behavior. Watch helium boil off daily.
Contact Bruker Service immediately! Do not try to remove ice of the helium flow system!
If the problem is still not solved, contact Bruker Service.
WARNING:
Cryogenic Agents
Quench
Blow in warm helium gas carefully through an applicable tube. Do not insert it more than 600 mm from the top of the helium turrets.
Approved Customer Personnel
Approved Customer Personnel
Approved Customer Personnel
Bruker Service
The helium boil off is too high.
Continued on next page
The helium level meter is permanently on (service mode) or used often.
The atmospheric pressure is decrea­sing.
Switch off helium level sensor. Reduce helium level measu­rement (during measuring of the helium level an amount of helium boils off due to the heat input of the helium level sensor; for reporting the helium boil off a weekly measurement is sufficient).
Usual behavior. Watch helium boil off weekly.
If the problem is still not solved, contact Bruker Service.
Approved Customer Personnel
Approved Customer Personnel
Approved Customer Personnel
98
ZTKS0156 / Z31820 / Rev.: 03
Troubleshooting
Continued from page before
Indicator Possible reason Solution By
Continue of: The helium boil off is too high.
RS temperature shield to high.
Vacuum reduced. Rebuild vacuum, see chapter
”Rebuilding the Vacuum” on page 4-53
The radiation baffles are not inserted in the current lead turret.
Cryogenic Refrigera­tion not sufficient.
Alarm default set­tings not correct.
Vacuum reduced. Rebuild vacuum, see chapter
Insert the radiation baffles into the current lead turret.
See chapter ”During Opera-
tion of the Cryogenic Refri­gerator” on page 6-102.
Check set values (see separate manual of the Exter­nal Temperature Control Unit).
If the problem is still not solved, contact Bruker Service.
”Rebuilding the Vacuum” on page 4-53.
Bruker Service
Bruker Service
Approved Customer Personnel
Approved Customer Personnel
Approved Customer Personnel
Bruker Service
Quench Loss of
superconductivity.
Cold spots within the RT bore.
RT vessel is wet and cold.
Not correct helium level warning out of supplied NMR soft­ware.
Helium level at constant level, no change during days.
Continued on next page
Alignment not cor­rect.
Vacuum reduced. Contact Bruker Service
Helium level sensor defective.
Helium level sensor defective.
See chapter ”After a
Quench” on page 6-107
Contact Bruker Service immediately!
Contact Bruker Service. Approved
immediately!
Contact Bruker Service immediately!
Contact Bruker Service immediately!
Approved Customer Personnel
Customer Personnel
Approved Customer Personnel
Approved Customer Personnel
Approved Customer Personnel
ZTKS0156 / Z31820 / Rev.: 03
99
Troubleshooting
Continued from page before
Indicator Possible reason Solution By
WARNING:
Low Temperature
Not correct helium level warning out of supplied NMR soft­ware.
Helium level at con­stant level, no change during days.
Helium level sensor defective.
Helium level sensor defective.
Replace helium level sensor (see chapter ”Replacement
of the Helium Level Sensor” on page 6-109)
Replace helium level sensor (see chapter ”Replacement
of the Helium Level Sensor” on page 6-109)
Bruker Service
Bruker Service
100
ZTKS0156 / Z31820 / Rev.: 03
Loading...