All rights reserved. No part of this manual may be reproduced or transmitted in any form
or by any means including printing, photocopying, microfilm, electronic systems etc. with
out our prior written permission. Brand names, registered trademarks etc. used in this
manual, even if not explicitly marked as such, are not to be considered unprotected by
trademarks law. They are the property of their respective owner.
This manual is the original documentation for the IFS 125M spectrometer.
Index ...........................................................................................................223
Bruker Optik GmbHIFS 125M User Manual
7
Table of Contents
8
IFS 125M User ManualBruker Optik GmbH
1Introduction
1.1About this manual
This manual is a complete documentation about the IFS 125M FT-IR spectrometer.
Depending on the degree of hazard important safety information and safety instructions
are classified as follows:
➣ Indicates a hazardous situation which, if not avoided, will result in death or serious
(possibly irreversible) injury.
➣ Indicates a hazardous situation which, if not avoided, could result in death or seri-
ous (possibly irreversible) injury.
DANGER
WARNING
CAUTION
➣ Indicates a hazardous situation which, if not avoided, may result in minor or mod-
erate (reversible) injury or major material damage.
NOTE
➣ Hazard, which could result in material damage if the appropriate safety instruc-
tions are not observed.
Information given to the user to facilitate spectrometer operation, and make the best
i
possible use of the spectrometer.
Bruker Optik GmbHIFS 125M User Manual
9
Introduction 1
1.2Terms
This manual uses both the term spectrometer or IFS 125M when the spectrometer is
described.
1.3Safety
Always observe the instructions described in this manual to ensure user safety and to
avoid property damage. Keep this manual for further reference available at any time.
Improper use or failure to follow the safety instructions can result in serious injuries and/
or property damage. Any non-observance infringes the intended use (i.e. spectroscopic
measurements) of the spectrometer. In this case Bruker Optik GmbH does not assume
any liability.
It is the operator's duty to plan and implement all necessary safety measures and to
supervise their observance. Moreover, the operator must ensure that the spectrometer is
in proper condition and fully functioning. A safe and faultless operation can only be guar
anteed if the spectrometer is transported, stored, installed, operated and maintained
properly according to the procedures and instructions described in this manual.
-
Never remove or deactivate any supporting safety systems during spectrometer operation. Objects and/or material not required for operation must not be kept near the spectrometer operating area.
The spectrometer has been developed according to the EN 61010-1:2010 (IEC 610101:2010+Cor.:2011) safety regulations for electrical measuring, control and laboratory
devices.
1.3.1Warning labels
When operating the spectrometer you have to observe a number of safety instructions
which are highlighted by the appropriate warning label. The warning labels and their
meaning are described in the following.
All warning labels on the spectrometer must always be kept legible. Immediately replace
worn or damaged labels.
LabelDefinition
General Hazard:
This warning symbol indicates general hazard. Observe the safety
instructions and follow the precautions described to avoid personal
injury and/or property damage.
10
Table 1.1: Warning labels
IFS 125M User ManualBruker Optik GmbH
LabelDefinition
Frostbite:
This warning symbol indicates cryogenic liquids (e.g. liquid nitrogen)
required to operate the spectrometer (e.g. cooling detector).
Exposure to these liquids or cooled components causes frostbite
effects. Handle the liquids with utmost care. Observe the safety
instructions for operating with cryogenic liquids.
Hot Surface:
This warning symbol refers to components and surfaces which can
become very hot during spectrometer operation. Do not touch these
components and surfaces. Risk of skin burn! Be careful when operat
ing near hot components and/or surfaces.
Laser Radiation:
This warning symbol indicates the existence of laser radiation. Never
look directly into the laser beam, or use any kind of optical instruments
to look into the beam as this may cause permanent eye damage.
Introduction 1
-
Table 1.1: Warning labels
1.3.1.1In case of hazardous sample material
Besides the dangers described above, there can also be hazards caused by the sample
material. Depending on the type of hazardous substances used, you have to observe
the specific substance-relevant safety instructions. Put on the specific warning label on
the corresponding module position. The label must be legible and permanently discern
ible.
The following list exemplifies types of hazardous sample material:
LabelDefinition
Infectious Material
This warning symbol indicates the possible existence of biologically
dangerous and infectious material. When working with this kind of
material always observe the prevailing laboratory safety regulations
and take necessary precautions and disinfection measures (e.g. wear
ing protective clothing, masks, gloves etc.). Non-observance may
cause severe personal injury or even death.
For information on how to use, dilute and efficiently apply disinfectants,
refer to the Laboratory Biosafety Manual: 1984 by WHO - World Health
Organization.
-
-
Table 1.2: Warning labels in case of hazardous sample material
Bruker Optik GmbHIFS 125M User Manual
11
Introduction 1
LabelDefinition
Table 1.2: Warning labels in case of hazardous sample material
Radioactive Material
This warning symbol indicates the possible existence of radioactivity.
When working with radioactive material always observe the safety reg
ulations and take necessary protective measures. Wear protective
clothing, e.g. masks and gloves. Non-observance may cause severe
personal injury or even death.
Corrosive Substances
This warning symbol indicates the possible existence of corrosive substances. When working with corrosive substances always observe the
laboratory safety regulations, and take protective measures (e.g. wear
protective masks and gloves). Non-observance may cause severe
personal injury or even death.
-
1.3.2Waste disposal
Dispose all waste produced (chemicals, infectious and radioactively contaminated substances etc.) according to the prevailing laboratory regulations. Detergents and cleaning
agents must be disposed according to the special waste regulations.
1.4General information
1.4.1Protective earthing
To avoid personal injuries and/or damage caused by electrical power, the supplied spectrometer power cord is equipped with a safety plug. Connect this safety plug only to an
earthed power socket. Make sure that the earthed power socket used complies with IEC
(International Electrotechnical Commission
1.4.2Qualified personnel
1
).
12
Initial installation and all maintenance and repair work not described in this manual
should only be performed by Bruker service personnel. Make sure that the spectrometer
is only operated and maintained (i.e. only maintenance work that is described in this
manual) by authorized operating personnel trained in the spectrometer operation and all
relevant safety aspects.
1. International standards organization for electrical and electronic-related technologies.
IFS 125M User ManualBruker Optik GmbH
All repairs, adjustments and alignments on any spectrometer component must be performed in accordance with the safety regulations and standards applied in the country
where the instrument will be installed.
1.4.3Intended use
The spectrometer and its components should only be used according to the instructions
described in the manual or advised by a Bruker engineer.
In case of accessories or components made by other manufacturers and used in connection with the spectrometer, Bruker Optik GmbH does not assume any liability for safe
operation and proper functioning.
1.5Questions and concerns
Introduction 1
If you have questions or concerns about safety, operating the spectrometer, or if you
need assistance with software problems or replacement parts, contact Bruker at the
numbers listed below:
•Service hotline hardware:+49 (0) 72 43 504-2020
•Service hotline software:+49 (0) 7243 504-2030
•Fax: +49 (0) 72 43 504-2100
•E-mail:service@brukeroptics.com
•Internet:www.brukeroptics.com
Bruker Optik GmbHIFS 125M User Manual
13
Introduction 1
14
IFS 125M User ManualBruker Optik GmbH
2General
The IFS 125M spectrometer is a mobile high resolution FT-IR spectrometer and has
been designed to perform field operation and precise stratospheric analysis, generate
atmospheric emission studies, and remotely monitor atmospheric pollutants.
Atmospheric pollution can be measured in transmittance, using the sun as active IR
radiation source.
For stratospheric transmittance measurements, the solar radiation is collected by a solar
tracker system, which automatically follows the position of the solar source. It has been
proven that this technique can also be used with the moon as IR source, despite the
orders of magnitude lower radiation power.
2.1Design
•Compact optics housing
•Detachable interferometer scanning arm
•Purgeable enclosure for interferometer, scanner and detectors
2.2Optics
•Spectral range starting from the mid-infrared (MIR) region at 450 up to the ultraviolet range of 45000 cm
•Spectral resolution of 0.008 cm-1 (optionally: 0.0035 cm-1) in the MIR region
The spectrometer is equipped with a mobile (separate) electronics unit which uses
Ethernet connectivity, and a 24
station, laptop etc.), with the OPUS spectroscopy software installed, can be used to control the spectrometer and perform data processing.
Analog to Digital Conversion (ADC) electronics allow high resolution measurements
from the MIR through the near-infrared (NIR) region, up to the visible and ultraviolet
(VIS/UV) spectral ranges.
Functions such as checking measurement parameters allow fast and reliable FT-IR
spectroscopic measurements. Optical components, e.g, source and beamsplitter, are
electronically coded so that the spectrometer firmware can recognize them.
-1
6
V DC power supply. Any modern data system (PC work-
In addition, the permanent online check of spectrometer components facilitates error
diagnostics and maintenance.
Bruker Optik GmbHIFS 125M User Manual
15
General 2
2.4Data acquisition
The data acquisition is based on a free running delta-sigma A/D converter with 24 bit
dynamic range. The series of time equidistant samples is converted into samples equi
distant in optical path difference, according to James Brault’s scheme. A digital signal
processor does the conversion in real time.
During natural phenomena such as sunset and sunrise in which the absorbance path is
considerably extended, the available measurement time is limited. Therefore, the dual
channel data acquisition technique in connection with a dichroic beamsplitter allows par
allel data acquisition with two dedicated and optimized detectors in one or two different
spectral ranges.
-
-
16
IFS 125M User ManualBruker Optik GmbH
3Installation
Unpacking and initial installation of the spectrometer is done by Bruker service engineers. The operating company has to provide an installation site that meets the site
requirements described in this chapter.
3.1Delivery contents
The IFS 125M spectrometer allows upgrading with additional components and/or accessories. Depending on the spectrometer configuration ordered, the delivery contents may
not include all optional components listed below.
Standard components:• IFS 125M spectrometer
• Electronics unit
• External power supply unit with low-voltage cable and power cord, including
country-specific power plug
• Data cable (100Base-T cross-over
cable)
• PC compatible data system (including
Windows or Linux operating system)
• OPUS spectroscopy software (spectrometer control, data acquisition, manipulation and evaluation)
• IFS 125M user manual
•Spare parts
Optional components:• Other OPUS application software pack-
ages (e.g. QUANT, IDENT) and the
respective user manuals
• Accessories
Table 3.1: Delivery contents
Bruker Optik GmbHIFS 125M User Manual
17
Installation 3
3.2Packaging
Damaged packaging
Injury and/or property damage
➣ Contact shipping company.
➣ Do not put spectrometer into operation.
3.3Transportation
CAUTION
Due to its weight (approx. 120 kg) the spectrometer should be carried by at least 3 persons. When transporting the spectrometer, use the original packaging to avoid damages.
3.4Site requirements
The operating company has to provide an installation site that meets the following site
requirements:
Environment:• Temperature range: 18 - 35°C
Electromagnetic interferences:
• Humidity (non-condensing): less than 80% (relative
humidity)
• The spectrometer is shielded against electromagnetic interferences.
• Do not locate the spectrometer near potential inductive electrical interference, high energy pulses and
high electric or magnetic fields.
• Use lines which are separated from those which are
connected to machines (e.g. elevators, large electric
motors, etc.) with large electrical surges or bursts.
18
Table 3.2: Site requirements
IFS 125M User ManualBruker Optik GmbH
Installation 3
NOTE
Power supplya:
The spectrometer power supply unit automatically
adapts to the most common power sources.
• Valid voltage range: 100 V AC to 240 V AC
• Valid frequency range: 50 to 60 Hz
• Additionally, any power supply via any 24 V DC
source is possible (load 5
A)
No uninterruptible power supply:
Property damage.
➣ Always ensure uninterruptible power supply (UPS)
and/or a line conditioner.
How to interrupt power
supply:
• Either switch off the spectrometer using the ON/OFF
switch, which is located on the front side of the
mobile electronics unit or
• unplug the power cable from the primary power
receptacle (AC inlet), which is located at the power
supply unit or
• unplug the power cable from the mains outlet
Space:• Stable and horizontal base which is able to bear the
weight of at least 160
kg
• For exact dimensions refer to appendix A.
Temperature variations:• Less than 1°C per hour and not more than 4°C per
day
• Avoid temperature variations as they can impair the
results of long-term measurements
Vibration:Do not locate the spectrometer in high traffic areas,
nor near vibration sources (ventilation hoods, air con
-
ditioners etc.).
Table 3.2: Site requirements
a. If you connect external accessories to the spectrometer interface such as R232 (serial), Ethernet or
LPT1 (parallel), these accessories must comply with the requirements outlined in EN 61010 or EN
60950. This means that you always have to connect a SELV circuit (Safety Extra Low Voltage Circuit) to
a corresponding SELV circuit and not to any other circuit type.
Bruker Optik GmbHIFS 125M User Manual
19
Installation 3
Figure 3.1:Connection ports on electronics unit
A
B
C
D
C
3.5Connection ports
The connection ports are located on the mobile electronics unit and electronics boxes.
Definition
ACAN bus connector
BMain power switch
CLow-voltage sockets (both ports are con-
nected via a y-cable)
DEthernet port
Table 3.3: Connection port definition on electronics unit
20
IFS 125M User ManualBruker Optik GmbH
Installation 3
Figure 3.2:Connection ports on left and right electronics box
A
B
C
D
E
Definition
ALight switch board port (LSB)
BSolar tracker port (not used)
CCAN bus connector (all equivalent)
DOuter motor port (OMOT)
EDigital detector connectionport (DDC)
Table 3.4: Connection port definition of electronics boxes
Bruker Optik GmbHIFS 125M User Manual
21
Installation 3
3.6Connecting procedures
3.6.1Connecting cables to the mobile electronics unit
On panel:Connect......with...
Electronics:LASERcable from HeNe laser
(located in upper panel)
ICONICON at flange panel
Sourcesreceptacle at calibration
source
OMOTOMOT at left electronics
box
Flange:
DDC0
a
DDCa at right electronics
box
a
DDC1
DDCa at right electronics
box
LSB122LSB at the left electronics
box
b
CAN
b
CAN
Table 3.5: Connecting cables to the mobile electronics unit
a. All DDC receptacles are equivalent, no specific order has to be observed.
b. All CAN receptacles are equivalent, no specific order has to be observed.
CAN at left electronics box
CAN at right electronics box
22
IFS 125M User ManualBruker Optik GmbH
3.6.2Overview on cables connected
Figure 3.3:Overview on cables connected
Installation 3
Bruker Optik GmbHIFS 125M User Manual
23
Installation 3
CAUTION
3.7Connecting spectrometer to power supply
1Switch off spectrometer.
☞The switch must be in the OFF
position.
2Use a y-cable and connect one end to
the 24
V in port, and the other to the
DC port.
3Connect the power cord to the C5
socket of the external power supply.
4Connect the power plug of the power
cord to the mains socket outlet.
Power plug of the power cord not
connected to a socket outlet with
earthing contact.
Personal injury and property damage
➣ Connect power plug only to a socket
with earthing contact.
Safety Instructions
To ensure a safe operation of the external power supply unit, observe the following
safety instructions:
24
•Operate the external power supply unit only in a dry environment.
•Make sure that the external power supply unit is not exposed to direct sunlight.
Avoid temperatures above +50°C. Ensure that there is sufficient air circulation.
IFS 125M User ManualBruker Optik GmbH
•Properly position the external power supply unit to avoid the danger of tripping
over it.
•Do not put heavy objects on the external power supply unit.
•Do not place the external power supply unit on a hot surface.
•If the external power supply unit is damaged disconnect it instantly from the supply circuit. Never put a damaged external power supply unit into operation. Only
authorized technicians are allowed to repair the external power supply unit.
3.7.1Connecting spectrometer to a PC
1The main power switch must be in the OFF
position.
Installation 3
2Connect one end of the cross-over data cable
3Connect the other end of the cross-over data cable to the RJ-45 socket of
the computer network interface card. (In case of different connection topolo
gies, see chapter 9.2)
Table 3.6: Connecting spectrometer to a PC
3.7.1.1IP address
The spectrometer IP address has been set to 10.10.0.1 by Bruker. If the delivery content
of your spectrometer configuration does not include a Bruker PC, you have to assign an
appropriate IP address to the type of PC to be used. For detailed information on how to
assign an IP address refer to chapter
to the ETH socket on the mobile electronics
unit.
-
9.3.
Bruker Optik GmbHIFS 125M User Manual
25
Installation 3
BA
3.7.2Connecting accessories via CAN bus
1Connect the CAN bus cable to the CAN bus
connector located at the mobile electronics
unit.
Table 3.7: Connecting accessories via CAN bus
3.7.3Connecting spectrometer to purge gas supply line
The spectrometer has one purge gas inlet to purge the interferometer and detector compartment. This is only required if the spectrometer is operated in a humid environment.
Tools required and not included in the delivery content: stiff purge hose (PVC), outer
diameter: 6
1Pull off the plug from the purge gas inlet.
2• Insert one end of the hose into the purge
mm.
gas inlet (A).
• Connect the other end of the hose to your
purge gas supply line. The supplied gas
must be dry and oil-free (see chapter
for details).
• Control the purge gas flow by the purge gas
controller (B).
5.11
26
Table 3.8: Connecting spectrometer to purge gas supply line
IFS 125M User ManualBruker Optik GmbH
3.8Inserting beamsplitter
During initial installation, the beamsplitter needs to be inserted. All beamsplitter types
available for the spectrometer are electronically coded, enabling the spectrometer firm
ware to auto-detect the beamsplitter type currently installed.
The data about the beamsplitter type are transferred to the spectroscopy software.
Inadequate handling:
Beamsplitter may become useless.
➣ Handle the beamsplitter with utmost care to ensure long service life.
➣ Observe the handling instructions described in chapter 3.8.1.
Installation 3
-
NOTE
3.8.1Handling instructions
•Do not touch the beamsplitter surface as the beamsplitter may become useless.
Always hold the beamsplitter on the two knobs (A in figure
•The surface of some beamsplitters (especially KBr beamsplitters) is hygroscopic. Never expose them to humidity, water or aqueous solutions. Always
store the beamsplitter in a sealed container and in dry environment, e.g. in the
beamsplitter storage box. Fix the protective covers onto the beamsplitter
(figure
3.4b). The covers provide mechanical protection, but do not protect
against humidity!
•Do not try to loosen or fasten the screws (B in figure 3.4a) as this will impair the
optical quality of the beamsplitter.
•Do not try to clean the beamsplitter. This will definitely damage the beamsplitter
irreversibly.
•Do not expose the beamsplitter (especially beamsplitters made of KBr) to large
temperature changes. In particular, a cold KBr beamsplitter must not be exposed
to a warmer ambience, as this will lead to condensation of humidity.
3.4a).
Bruker Optik GmbHIFS 125M User Manual
27
Installation 3
Figure 3.4a: Beamsplitter
A
B
Figure 3.4b: Beamsplitter with protective
cover
CAUTION
3.8.2Inserting procedure
1Use a cross-head screwdriver (size PH2) and
2Carefully lift off the lid.
Table 3.9: Inserting beamsplitter
rotate the 2 screws of the top lid counterclock
wise.
➣You have to rotate over the snap-in resis-
tance. Then, the screw is pushed
upwards, and you can loosen it.
Open lid:
Risk of exposure to laser light, which can
cause permanent eye damage.
➣ Do not stare into the laser beam.
-
28
IFS 125M User ManualBruker Optik GmbH
Installation 3
3Open the beamsplitter locking door by moving
the release lever.
4• Hold the beamsplitter - with two hands - on
the knobs and carefully insert the beamsplit
ter straight downwards into the fixture, with
the electrical contacts facing to the scanner
side and without catching an edge.
• The two guiding pins have to engage into
the slots located on the block.
• Push down the beamsplitter completely, until
you feel resistance.
-
5Close the beamsplitter locking door by moving
the release lever.
6The spectrometer beeps when it detects the new beamsplitter and automati-
cally starts scanning.
• Put on the lid again.
• Use the cross-head screwdriver. Push the screws downwards and rotate
them clockwise.
• Check the laser signals as described in chapter 6.2.
• Check whether a signal is detected, and the optics works properly. More
details are described in chapter
• If you use VIS and UV beamsplitters, the interferometer has to be realigned
by the OPUS Auto Align function (see chapter
Table 3.9: Inserting beamsplitter
5.6.
5.8.2).
Bruker Optik GmbHIFS 125M User Manual
29
Installation 3
30
IFS 125M User ManualBruker Optik GmbH
4Description
Figure 4.1:Spectrometer front side
A
B
C
E
F
G
D
This chapter describes all relevant spectrometer compartments.
The local indications right and left assume that the operator stands in front of the spec-
i
trometer. The same applies to the indication front and rear side, respectively.
4.1Spectrometer front side
Definition
ASample compartment lid
B
CLeft electronics box
DBeamsplitter lid
EDetector hoods
FRight electronics box
Bruker Optik GmbHIFS 125M User Manual
GScanner compartment
Table 4.1: Spectrometer front side
a. Can be used to observe the image of the external source, e.g.
Viewing port
the sun, on the aperture.
a
31
Description 4
Figure 4.2:Spectrometer rear side
A
B
C
D
E
F
G
4.2Spectrometer rear side
Definition
ADetector hoods
BDirect input port for collimated beam
CBeamsplitter lid
D
Viewing port
a
ESample compartment lid
FSource (for calibration purpose)
GStandard input port for collimated beam
Table 4.2: Spectrometer rear side
a. Can be used to observe the image of the external source, e.g.
the sun, on the aperture.
32
IFS 125M User ManualBruker Optik GmbH
4.3Sample compartment
Figure 4.3a: Sample compartment with lidFigure 4.3b: Sample compartment with gas cell
Figure 4.4a: Closed beamsplitter lidFigure 4.4b: Open beamsplitter lid
The sample compartment can be accessed by lifting the blue top lid. For calibration purpose the sample compartment can accommodate e.g. a 10 cm calibration gas cell. This
gas cell is fixed by 2 retaining clips.
Description 4
How to insert the gas cell is described in chapter 5.5.
4.4Beamsplitter lid
The spectrometer is equipped with a permanently-aligned Michelson interferometer,
cube-corner mirrors and a beamsplitter.
The interferometer has a 30° angle of incidence. This narrow angle of incidence makes
more effective use of the beamsplitter and reduces polarization effects.
How to change the beamsplitter is described in chapter 5.8.
Bruker Optik GmbHIFS 125M User Manual
33
Description 4
NOTE
CAUTION
4.4.1Beamsplitter types
The spectrometer can be configured with the following beamsplitter types:
Description
KBr beamsplitter (with
multilayer coating)
Hygroscopic!
➣ Avoid any contact to
humidity.
➣ Do not touch or clean
the surface.
➣ Always store in a dry
place
CaF
2
Harmful
➣ Do not inhale or swal-
low dust or particles.
➣ Avoid any eye and skin
contact.
➣ Observe the safety
data sheet.
Spectral regiona (cm-1)
4,800 - 450T301/2
NIR, 14,000 - 1,850T401/2
Part number
34
Quarz with dielectric coatings
Quarz with metallic coatings
Table 4.3: Beamsplitter types
a. The combination of light source, beamsplitter and detector defines the measurement range. It is lim-
The spectrometer can accommodate up to 2 internal detectors. In the standard configuration the spectrometer is equipped with a liquid N2-cooled MCT detector system for
MIR spectral range and high spectral sensitivity.
Description 4
Apart from the standard detector there is a large number of optional detectors available.
All detectors are mounted on dovetail slides which allow an easy interchange.
4.5.1Detector types
The following table contains the most important types of detectors to be used with the
spectrometer. For further detector options contact Bruker.
Description
DLATGS12,000 - 350D301/3
InGaAs 12,800 - 5,800D424/2
InSb, liquid-nitrogen
cooled
Harmful
➣ Observe the safety
instructions when cool
ing.
Spectral regiona (cm-1)
9,600 - 1,850D413/2 (12 h)
-
Part number
D413-L/2 (24 h)
Table 4.4: Detector types
Bruker Optik GmbHIFS 125M User Manual
35
Description 4
CAUTION
CAUTION
CAUTION
Description
MCT, narrow band, liquidnitrogen cooled
Harmful
➣ Observe the safety
instructions when cool
ing.
MCT, broad band, liquidnitrogen cooled
-
Harmful
➣ Observe the safety
instructions when cool
ing.
MCT, medium band, liquid-nitrogen cooled
-
Spectral regiona (cm-1)
5,000 - 780D313/2 (12 h)
7,000 - 420D315/2 (12 h)
6,000 - 600D316/2 (12 h)
Part number
D313-L/2 (24 h)
D315-L/2 (24 h)
D316-L/2 (24 h)
Harmful
➣ Observe the safety
instructions when cool
ing.
Si diode30,000 - 9,000D510/2
Table 4.4: Detector types
a. The combination of light source, beamsplitter and detector defines the measurement range. It is lim-
ited by the component with the most narrow range.
-
36
IFS 125M User ManualBruker Optik GmbH
4.6Scanner compartment
Figure 4.6a: Closed scanner compartment Figure 4.6b: Open scanner compartment
A short and a long version of the scanner arm are available. The scanner is driven by 2
drive units, i.e. 1 DC (direct current) motor and 1 linear drive unit. These two drive units
are also called the outer and inner scanner.
Description 4
The outer scanner ensures the continuous movement, while the main function of the
inner scanner is to compensate the fluctuations of the outer scanner. For low resolution
measurement the inner scanner is activated, only.
Based on the laser interference signal, the inner scanner compensates the fluctuations.
To compensate the tilting of the inner scanner this scanner uses a displacement sensor
(digital encoder).
The scanner movement is controlled by the HeNe laser signal. To avoid any malfunctions with regard to mechanical scanner movements make sure that the bearings are
always free of any visible deposits.
Bruker Optik GmbHIFS 125M User Manual
37
Description 4
Figure 4.7:Input ports
4.7Input ports
Ensuring electromagnetic shielding requires a conductive connection between accessory and spectrometer, and a complete metallic enclosure of the accessory without any
slits or openings larger than 4
with (metallic) blank flanges.
cm are required. Unused ports have always to be covered
If the port opening has a clear aperture of more than 4 cm, the port will not be shielded
any more according to electromagnetic compatibility as soon as the blank flange is
removed.
NOTE
Clear port aperture of port opening >4 cm:
No shield safety according to EMCa requirements.
➣ Accessory connected must ensure complete EMC shielding.
a. EMC: Electromagnetic Compatibility
38
IFS 125M User ManualBruker Optik GmbH
4.8Output port
Figure 4.8:Output port
Description 4
The output port is located on the right side of the spectrometer.
Bruker Optik GmbHIFS 125M User Manual
39
Description 4
Figure 4.9:Mobile electronics unit
4.9Mobile electronics unit
The spectrometer is equipped with a mobile electronics unit. This unit is connected to
the spectrometer via several cables and consists of different panels.
4.9.1Panel types
The following table contains the different types of panels. A detailed description of the
single panels is given in chapter
Panel typeLocation on front side of the electronics unit
Flange panel
Table 4.5: Panel types
10.
40
IFS 125M User ManualBruker Optik GmbH
Description 4
Panel typeLocation on front side of the electronics unit
Electronics panel
Power supply panel
Laser panel
Table 4.5: Panel types
Bruker Optik GmbHIFS 125M User Manual
41
Description 4
Figure 4.10: IFS 125M - Optical path
4.10Optical path
In standard operation, the parallel input beam is focused by an off-axis-paraboloidal
(OAP) mirror onto computer-controlled filter and aperture wheels. After passing the field
stop, the input beam is re-collimated onto the beamsplitter.
The optical filter and aperture are selected by the operator to optimize instrumental
response for the experimental conditions. After recombining the beam from the two arms
of the interferometer, a computer-controlled mirror stage selects between two detectors.
Alternatively, using an optional dichroic filter, the modulated beam can be split to two
detectors, thus enabling simultaneous collection from two data channels.
For calibration purposes, a mirror stage allows light from either a globar or tungsten
source to pass through a 10
selection between the calibration source and the collimated input is computer controlled.
cm gas cell and then follow the same optical path. The
42
IFS 125M User ManualBruker Optik GmbH
5Operation
If the IFS 125M spectrometer has been configured and connected to the PC, you can
start acquiring data. The spectrometer is completely software controlled
components (e.g. detector) are selected and controlled by using the OPUS software. For
details about the software and how to perform a data acquisition refer to the OPUS Ref
erence Manual.
This chapter describes the spectrometer-related aspects of operation. For information
on software-related aspects (e.g. parameter setting and data acquisition), refer to the
OPUS software Reference Manual.
5.1Operating safety
During operation the spectrometer generates laser radiation and high voltages in some
of the electronics.
Objects placed on spectrometer
Objects may fall inside and damage spectrometer components when any of the compartment covers is removed.
➣ Do not place any object on top of the spectrometer.
NOTE
1
, i.e. the optical
-
CAUTION
Spectrometer covers removed
Laser and source radiation could be emitted and cause health hazards and property
damage. The spectrometer‘s optics may be contaminated.
➣ Always keep the spectrometer covers closed and secured during operation.
1. Except for the beamsplitter.
Bruker Optik GmbHIFS 125M User Manual
43
Operation 5
5.2Switching spectrometer on and off
The spectrometer is turned on and off by the main power switch located on mobile electronics unit.
Main power switchStatusWhat is to be done?
OnTurn the main power switch in the ON
position.
OffTurn the main power switch in the
OFF position.
Table 5.1: Operating status
After powering up the spectrometer, wait at least 30 minutes before running any
i
measurements. This allows the electronics and the laser to stabilize.
44
IFS 125M User ManualBruker Optik GmbH
5.3Spectrometer operating phase
How to start?Switch on the spectrometer, using the ON/OFF switch
located on the front side of the mobile electronics unit.
Operation 5
What happens after the
spectrometer is
switched on?
When is the spectrometer ready-to-operate?
What is to be done if the
spectrometer is readyto-operate?
• The spectrometer starts booting. Booting and initializing take about 60 seconds.
• The SR LED (on the front side of the mobile electronics unit) turns red.
• The spectrometer beeps once and starts a self test.
• The SR LED turns from continuous red to sporadic
flashing red if the self-test has been successful.
As soon as the red ERR and SR LED (on the front side
of the mobile electronics unit) have turned off.
The HeNe reference laser needs about 20 minutes to
stabilize. The laser is ready, when the green LASER
LED stops flashing.
i By default, the scanner initializes automatically as
soon as the LASER LED lights continuously. If the
scanner does not initialize automatically see
chapter 7.5 Scanner not scanning.
Switch on the computer.
➣ There is no specific order required. You can also
switch on the computer, before switching on the
spectrometer.
i For information on how to install the computer and
how to set up signal and power cable connections
for the computer, monitor etc. refer to the computer
manual.
Table 5.2: Spectrometer operating phase
Bruker Optik GmbHIFS 125M User Manual
45
Operation 5
5.4Starting spectroscopy software
If the PC has finished booting, start the OPUS spectroscopy software which has been
pre-installed on the PC. How to start OPUS is described in the OPUS Reference Man
ual.
5.4.1Setting up the connection to the optics
1.On the OPUS Measure menu, select the Optic Setup and Service command.
2.Click the Optical Bench tab and select the IFS125 spectrometer type from the Con-figuration drop-down list.
3.Enter correct IP address (see chapter 3.7.1.1).
4.Click the Save Settings button.
➣ OPUS status light at the lower right end of the OPUS interface has to change
from gray to red, yellow or green. If the status light remains gray, see
chapter 7.7.
-
The settings saved remain, unless the IP address of the optics has been changed, or a
i
different spectrometer type is to be accessed.
46
IFS 125M User ManualBruker Optik GmbH
5.5Inserting accessory into sample compartment
A 10 cm gas cell can be inserted into the sample compartment for calibration purposes.
5.5.1Inserting procedure
1• Use a cross-head screwdriver (size: PH2)
and rotate the 4 screws of the top lid coun
terclockwise.
➣You have to rotate over the snap-in resis-
tance. Then, the screw is pushed
upwards, and you can loosen it.
• Carefully lift off the sample compartment
cover.
Operation 5
-
2• Insert the accessory into the sample com-
partment.
•
• Turn the 2 retaining clips by 90° to fix the
cell.
• Put on the sample compartment lid again.
• Use the cross-head screwdriver. Push the
screws downwards and rotate them clock
wise.
Table 5.3: Inserting accessory into the sample compartment
-
Bruker Optik GmbHIFS 125M User Manual
47
Operation 5
Figure 5.1:OPUS - Check Signal tab
B
A
5.6Checking the signal
Especially after you have replaced a spectrometer component (source, laser, beamsplitter, detector), it is advisable to check whether a signal is detected, and to check the signal intensity (signal amplitude).
If a liquid nitrogen-cooled detector is used, the detector must be cooled (chapter 5.9)
before a signal can be detected.
5.6.1Procedure
1.Make sure that the spectrometer sample compartment is empty, or contains a gas
cell the windows of which are transparent in the spectral range desired.
2.On the OPUS Measure menu, select the Advance Measurement command.
3.Select appropriate measurement parameters.
4.Click the Check Signal tab.
5.Activate the Interferogram option button.
48
IFS 125M User ManualBruker Optik GmbH
Definition
AAmplitude value:
Indicates the currently detected signal intensity. The amplitude is proportional
to the intensity of the incident IR radiation on the detector. It should not have a
short-term deviation of more than 200 ADC counts if the beam is directed onto
the thermally stable calibration source.
Long-term deviation (over several hours) can lead to larger deviations in the
signal intensity due to variations in the ambient temperature. The signal inten
sity should not deviate by more than 20% from the value recorded during the
delivery in the installation report.
Position:
The position represents the position (in HeNe fringe zero crossings) of the
interferogram maximum. This value must be constant when the OPUS Check Signal mode is selected (as long as the scanner is not stopped or re-initial
ized).
Operation 5
-
-
If it varies, check whether the spectrometer is subject to extreme vibrations. If
there are two interferogram maxima of nearly the same intensity, the position
may vary between two values.
BInterferogram
It indicates that a signal is detected.
Table 5.4: OPUS check signal tab
The position value can change any time you power up the spectrometer as the scanner
i
is always re-initialized during start-up. Thus, after powering up the spectrometer or reinitializing the scanner, select the Advanced Measurement command of OPUS and click
the Check Signal tab. Then, click the Save Peak Position button.
To verify the currently detected signal intensity, compare the amplitude value displayed
on the Check Signal tab with the amplitude value stated in the supplied test protocol
If there is not any signal detected or if the amplitude value displayed on the Check Signal
tab deviates significantly from amplitude value given on the supplied test protocol, check
the installation of the spectrometer component(s) you have replaced before.
1
.
1. The supplied test protocol documents the result of a factory-performed test. The test has been performed
with the spectrometer being optimally adjusted. You will find the test protocol in the blue folder supplied with
the spectrometer.
Bruker Optik GmbHIFS 125M User Manual
49
Operation 5
5.7Performing measurement
Before starting any type of measurement, the spectrometer has to be set up properly.
Make sure that the HeNe laser is stabilized, the beamsplitter inserted and the detector
cooled down (if you use a liquid nitrogen-cooled detector).
5.7.1Types of measurement
5.7.1.1Standard measurement using the internal source and the sample compartment
There are two types of measurement to be performed:
•sample measurement
•background measurement
A background measurement is a measurement which uses the empty gas cell or not any
sample at all, and is always performed before sample measurement. The purpose of the
background measurement is to detect how the ambient conditions, e.g. level of air
humidity, temperature etc., and the spectrometer itself have an influence on the spectro
scopic measurement result.
-
After the sample measurement has been completed, OPUS calculates the result sample
spectrum, dividing the sample spectrum (SSC
1
) by the background spectrum (RSC2).
The spectral bands, which result from the ambient conditions and/or the spectrometer,
are eliminated from the result sample spectrum.
5.7.1.2Atmospheric measurement using the input port for collimated beam
This type of measurement performs a sample measurement only. If there are narrow
absorption lines on a smoothly varying background, a quasi absorption spectrum can be
obtained as follows:
•use the OPUS Spectrum Calculator command to obtain -lg SSC
•use the OPUS Baseline Correction command (interactive mode) to bring the
baseline down to zero
5.7.2Measurement parameters
The measurement parameters are defined on the OPUS Measure menu. Select the
Advanced Measurement command and define the parameters on the different tabs. The
standard measuring parameters are provided in appendix
C.
3
50
1. SSC: Sample Single Channel, name of the data block added to the OPUS spectrum file
2. RSC: Reference Single Channel, name of the data block added to the OPUS spectrum file
3. Due to the nature of FT spectroscopy, absorbance values >=2 are unreliable and should not be included in
any analysis.
IFS 125M User ManualBruker Optik GmbH
5.7.3Measuring procedure
1.Start OPUS, the spectroscopy software.
2.Set up the connection between spectrometer and PC.
➣ This step is only required if not yet done.
☞ On the Measure menu, select the Optic Setup and Service command. Click the
Optical Bench tab and select the IFS125 spectrometer type from the Configuration drop-down list. The OPUS status light at the lower right end of the OPUS
interface has to be green or yellow before being able to start measurement.
3.Enter the measuring parameters.
☞ On the Measure menu, select the Advanced Measurement command.
☞ Either load the measuring experiment. On the Basic tab, click the Load button.
The experiments are stored in the <OPUS/xpm directory.
☞ Or set up all measurement parameters manually.
4.Check signal intensity.
☞ Click the Check Signal tab. The Interferogram option button must be activated to
have the interferogram displayed. The interferogram indicates that a signal is
detected. The amplitude value displayed above the interferogram shows the signal intensity currently detected.
☞ To check signal intensity compare the amplitude value displayed above the inter-
ferogram with the amplitude value indicated in the test protocol. The supplied
test protocol documents the result of a factory-performed test. The test has been
performed with the spectrometer being optimally adjusted. You find the test protocol in the folder supplied with the spectrometer.
Operation 5
Especially after switching on the spectrometer, or after a power failure, it is advisable to
i
check the intensity and position of the interferogram. Click the Save Peak Position but
ton once.
5.Measure background spectrum (if applicable).
☞ On the Basic tab, click the Background Single Channel button.
6.Position the sample (if applicable).
7.Start sample measurement.
☞ On the Basic tab, click the Sample Single Channel button.
Detailed information on data acquisition and evaluation are described in the OPUS Reference Manual.
-
Bruker Optik GmbHIFS 125M User Manual
51
Operation 5
5.8Changing beamsplitter
The basic spectrometer configuration is equipped withan MIR beamsplitter, i.e. KBr substrate with multilayer coating. To switch the spectral range the beamsplitter can be
changed.
All beamsplitters available for the spectrometer are electronically coded, enabling the
spectrometer firmware to auto-detect the beamsplitter type currently installed.
The data about the beamsplitter type are transferred to the spectroscopy software.
Inadequate handling:
Beamsplitter may become useless.
➣ Handle the beamsplitter with utmost care to ensure long service life.
➣ Observe the handling instructions described in chapter 5.8.1.
NOTE
5.8.1Handling instructions
•Do not touch the beamsplitter surface as the beamsplitter may become useless.
Always hold the beamsplitter on the two knobs (A in figure
•The surface of some beamsplitters (especially KBr beamsplitters) is hygroscopic. Never expose them to humidity, water or aqueous solutions. Always
store the beamsplitter in a sealed container and in dry environment, e.g. in the
beamsplitter storage box. Fix the protective covers onto the beamsplitter
(figure
5.2b). The covers provide mechanical protection, but do not protect
against humidity!
•Do not try to loosen or fasten the screws (B in figure 5.2a) as this will impair the
optical quality of the beamsplitter.
•Do not try to clean the beamsplitter. This will definitely damage the beamsplitter
irreversibly.
•Do not expose the beamsplitter (especially beamsplitters made of KBr) to large
temperature changes. In particular, a cold KBr beamsplitter must not be exposed
to a warmer ambience, as this will lead to condensation of humidity.
5.2a).
52
IFS 125M User ManualBruker Optik GmbH
Operation 5
Figure 5.2a: Beamsplitter
A
B
Figure 5.2b: Beamsplitter with protective
cover
CAUTION
5.8.2Changing procedure
If you change the beamsplitter frequently, purging the spectrometer with dry air or nitro-
i
gen is highly recommended to reduce the air humidity level inside the spectrometer. Dryair purging considerably reduces the spectrometer down-time after changing the beam
splitter.
1Use a cross-head screwdriver (size PH2) and
2Carefully lift off the lid.
rotate the 2 screws of the top lid counterclock
wise.
➣You have to rotate over the snap-in resis-
tance. Then, the screw is pushed
upwards, and you can loosen it.
Open lid:
Risk of exposure to laser light, which can
cause permanent eye damage.
➣ Do not stare into the laser beam.
-
-
Bruker Optik GmbHIFS 125M User Manual
Table 5.5: Changing beamsplitter
53
Operation 5
3
4• Hold the beamsplitter - with two hands - on
5The spectrometer beeps when it detects the new beamsplitter and automati-
cally starts scanning.
• Put on the lid again.
• Use the cross-head screwdriver. Push the screws downwards and rotate
them clockwise.
• Check whether a signal is detected, and the optics works properly. More
details are described in chapter
• Unlock the release lever, move it further up.
• Hold the beamsplitter - with two hands - on
the knobs.
• Carefully remove the beamsplitter straight
upwards without catching an edge.
the knobs and insert the new beamsplitter,
with the electrical contacts facing to the
scanner side. The two guiding pins have to
engage into the slots located on the block.
• Push down the beamsplitter completely, until
you feel resistance.
• Lock the release lever, move it further down.
5.6.
Table 5.5: Changing beamsplitter
If the beamsplitter has been changed, the peak position changes and must therefore be
i
saved. Select the Advanced Measurement command of OPUS and click the Check Sig
nal tab. Then, click the Save Peak Position button.
If you use VIS and UV beamsplitters, the interferometer has to be realigned by the
OPUS Auto Align function as follows:
1.Check signal intensity (chapter 5.6).
2.Once an interferogram is displayed, click the Auto Align button.
➣ If you check Store mode and activated the Spectrum option button, you can
observe the progress visually.
➣ The alignment is most sensitive at high wavenumbers and large apertures. For
example, to align the VIS beamsplitter use a blue or green optical filter and an
aperture of 1.85 mm. Make sure that the detector is not overloaded.
➣ Never use the Auto Align function when no (or only a very small) interferogram is
displayed or when working at low wavenumbers. This may lead to severe misalignment.
-
54
IFS 125M User ManualBruker Optik GmbH
5.9Cooling detector
As the operating temperature of the MCT detector used in the spectrometer is significantly below room temperature, the detector is filled with liquid nitrogen to be able to
achieve the required operating temperature.
The MCT detector used inside the spectrometer has to be cooled in regular time intervals.
Hold time of MCT detector:Typically 16 hours
Delivery content of detector:• Funnel
Cooling medium required:• Liquid nitrogen
Operation 5
• Protective dewar plug
• About 1l/filling (16 hours)
Indications of a weakening cooling
effect:
Table 5.6: Detector features
• Decreased signal intensity
• No signal at all
• OPUS status light becomes red
• Instrument status message in OPUS
reads: Detector not ready.
CAUTION
Improper use of liquid nitrogen:
Can cause frostbites, damage to eye tissue, and asphyxiation.
➣ Avoid any skin contact.
➣ Always wear face shields or safety goggles.
➣ Use liquid nitrogen only in well-ventilated rooms.
➣ Observe the safety instructions described in chapter 5.9.1.
Bruker Optik GmbHIFS 125M User Manual
55
Operation 5
5.9.1Safety instructions when working with liquid nitrogen
Avoid spillage of liquid
nitrogen:
Rooms need to be well ventilated:• To be able to prevent excessive nitrogen
Avoid skin contact:Due to the extremely low temperature of
Wear face shield or
safety goggles:
When spilled on a surface, liquid nitrogen
tends to cover the surface completely and
cools it immediately.
gas concentrations from being available.
• Excessive nitrogen gas concentrations
in the air reduce the concentration of
oxygen and can cause asphyxiation.
• Nitrogen gas cannot be detected by
human senses.
• As nitrogen gas is odorless there is the
risk of being inhaled as if it were normal
air.
liquid nitrogen (-196°C), any skin contact
can cause severe frostbites.
• The sensitive eye tissue can be damaged if exposed to this cold gas, even if
exposed for a short period of time.
• Use only safety goggles with side
shields.
Table 5.7: Safety instructions on liquid nitrogen
56
IFS 125M User ManualBruker Optik GmbH
5.9.2Cooling procedure
CAUTION
1Remove the protective cap from the 2 detec-
2Insert the supplied funnel into the detector fill-
Operation 5
tor hoods.
ing port and slowly fill in liquid nitrogen. At the
beginning the liquid nitrogen will evaporate
and escape.
Filling cold nitrogen into warm container:
Nitrogen boils and splashes. Risk of frostbites.
➣ Fill in nitrogen slowly into the container.
➣ Keep away from boiling and splashing
nitrogen, and its escaping gases.
3• Wait until the funnel is empty before refilling. If the liquid nitrogen does not
escape any more, the dewar has cooled down to the liquid nitrogen tem
perature.
• Fill the funnel again with liquid nitrogen. Avoid spilling the liquid on the
spectrometer housing.
• Repeat this procedure until the detector dewar has been filled to maximum.
-
i As a rule of thumb for the standard MCT detector: the maximum dewar
capacity is about the quantity of 3 funnel fillings. Note that the first 3 funnel
fillings will evaporate almost completely. Avoid overfilling as the liquid
would then flow out of the filling port.
• Remove the funnel and put on the protective cap again.
4
• Wait about 15 minutes before starting the measurement to allow the detector to stabilize.
• Check whether a signal is detected and the optics works properly. More
details are described in chapter
5.6.
Table 5.8: Cooling detector
Bruker Optik GmbHIFS 125M User Manual
57
Operation 5
NOTE
5.10Changing detector
The basic spectrometer configuration is equipped with a liquid-nitrogen-cooled MCT
detector, mounted on a dovetail guide.
5.10.1Changing procedure
When changing the detector you have to mind the order described in the following.
Removing covers
1Removing scanning arm cover:
• Open the clips on each side.
• Put your hand below the clip and push the
clip upwards.
• Pull the clip towards you until you feel resistance.
i All the clips have to be in their outermost
2
Table 5.9: Changing detector - Removing covers
• Two persons are required to lift off the cover
• Put your hands below the hangers.
• Carefully lift off the cover from the scanning
position, otherwise you get problems when
lifting off the cover.
of the long scanning arm, one at the front
and one at the back end.
arm.
Cover not lift off properly.
Risk of scanner damage.
➣ Lift off the cover straight and far
upwards.
58
IFS 125M User ManualBruker Optik GmbH
Operation 5
3Unscrew the Teflon-made counter nuts from
the 2 detector hoods.
i The Teflon-made counter nuts are only
available if you use a liquid nitrogencooled detector.
4
5Use a flat-tip screwdriver to loosen the two
6Completely remove the trough out of the sam-
• Use a cross-head screwdriver (size: PH2)
and rotate the 4 screws of the top lid coun
terclockwise.
➣You have to rotate over the snap-in resis-
tance. Then, the screw is pushed
upwards, and you can loosen it.
• Carefully lift off the sample compartment
cover.
knurled thumb screws, located on the bottom
of the trough.
ple compartment.
-
7Repeat step 1 to remove the cover of the detector and interferometer block.
Table 5.9: Changing detector - Removing covers
Bruker Optik GmbHIFS 125M User Manual
59
Operation 5
NOTE
8• Put your hands below the flange of the input
Table 5.9: Changing detector - Removing covers
port and the diagonally located hanger.
• Carefully lift off the cover from the detector
and interferometer block.
Cover not lift off properly.
Risk of detector, scanner or mirror damage.
➣ Lift off the cover straight and far
upwards.
60
IFS 125M User ManualBruker Optik GmbH
Removing detector and inserting new one
A
B
1In case of room-temperature detectors:
Operation 5
remove the shim plate next to the detector.
This type of detector does not have a sepa
rate preamplifier.
• Gently tilt the detector to the left and right to
loosen it from the dovetail guide.
• Carefully pull upwards the preamplifier (A)
and the detector (B) simultaneously. Avoid
any damage on the mirrors or other compo
nents next to the detector.
-
-
2• Insert the new detector and preamplifier
simultaneously.
• Insert the detector precisely into the dovetail
guide. Make sure that no cable blocks the
optical path.
• Push the detector downwards until it locks
into place. The electrical connections are
established automatically if the detector has
been completely pressed down.
Table 5.10: Changing detector - Removing detector and inserting new one
Bruker Optik GmbHIFS 125M User Manual
61
Operation 5
3In case of room-temperature detectors:
• Fasten the shim plate (see marking in figure)
to mount the detector.
• Make sure that no cable blocks the optical
path.
4• Switch on the spectrometer.
• Check whether a signal1 is detected and the optics works properly. More
details are described in chapter 5.6.
The detector signal may be optimized by slightly adjusting the last parabolic
mirror in front of the detector.
Table 5.10: Changing detector - Removing detector and inserting new one
1. If applicable, the detector has to be cooled first.
62
IFS 125M User ManualBruker Optik GmbH
Remounting covers
AB
NOTE
1• Put on the cover of the detector and interferometer block.
• Close the clips on each side.
2• Insert the trough into of the sample compart-
3• Use a cross-head screwdriver (size: PH2)
Operation 5
ment.
• Use a flat-tip screwdriver to fasten the two
knurled thumb screws, located on the bot
tom of the trough.
and rotate the 4 screws of the top lid clock
wise.
➣Press down the screw and rotate it once.
-
-
4• Insert the Teflon-made counter nuts (A).
i The Teflon-made counter nuts are only
available if you use a liquid nitrogencooled detector.
• Put on the protective caps (B).
5• Two persons are required to put down the
cover of the long scanning arm, one at the
front and one at the back end.
• Put your hands below the hangers.
Cover not put down properly.
Risk of scanner damage.
➣ Put down the cover straight and far
downwards.
Bruker Optik GmbHIFS 125M User Manual
Table 5.11: Changing detector - Remounting covers
63
Operation 5
6Close the clips on each side.
Table 5.11: Changing detector - Remounting covers
64
IFS 125M User ManualBruker Optik GmbH
5.11Purging the spectrometer
Purging the spectrometer reduces the level of unwanted atmospheric interferents like
water vapor and carbon dioxide inside the spectrometer. Especially water vapor in ambi
ent air absorbs IR radiation.
Additionally, purging protects humidity-sensitive components (e.g. KBr beamsplitter)
from damage. A KBr beamsplitter must not be exposed to humidity.
Spectrometer permanently open:
Water vapor, CO2 and other atmospheric gases cause unwanted absorptions in the
infrared spectrum range.
➣ Open the spectrometer only if necessary in order to prevent water vapor or other
gases from entering and adhering to the spectrometer inner walls.
Operation 5
-
NOTE
When to purge?If the ambient air humidity level inside the spectrome-
ter is too high.
Purge gas:• dry (dew point < -40°C), oil-free and dust-free air
• nitrogen gas
Maximum pressure:1800 hPa
Sustained gas flow rate:at least 1 l/s
Purging duration:continuously
How to connect purge
gas supply line:
Table 5.12: Purging features
see chapter 3.7.3
WARNING
Not using adequate gases for purging:
Some spectrometer components become hot during operation. Risk of fire and/or
explosion.
➣ Never use flammable gases for purging.
Bruker Optik GmbHIFS 125M User Manual
65
Operation 5
66
IFS 125M User ManualBruker Optik GmbH
6Maintenance
The IFS 125M spectrometer is a low-maintenance instrument, i.e. the operator can
replace components with a limited service life without requiring the assistance of the
Bruker service personnel.
The following maintenance procedures are described in this chapter:
•Checking laser signal
•Replacing source
•Replacing window
•Evacuating detector
•Checking drive cable
•Transporting spectrometer
•Cleaning housing
6.1General maintenance considerations
Perform only the maintenance procedures described in this chapter. Strictly observe the
relevant safety precautions. Any failure to do so may cause property damage or per
sonal injury. In this particular case Bruker does not assume any liability.
-
Maintenance procedures not described in this manual should only be performed by a
Bruker service engineer.
The following precautions must be observed to ensure user and property safety:
•Disconnect power supply before performing any maintenance procedures.
•Be careful if the spectrometer covers are removed and the spectrometer is
switched on to avoid contact with potentially harmful voltages.
CAUTION
Spectrometer still connected to power supply and covers removed:
Laser and source radiation could be emitted and cause health hazards and property
damage.
➣ Always disconnect spectrometer from power supply before starting any mainte-
nance work.
Bruker Optik GmbHIFS 125M User Manual
67
Maintenance 6
6.1.1Electrostatic discharge
Electronic components (e.g. semiconductor chips, printed circuit boards) are very susceptible to electrostatic discharges. Even discharge which may not be perceptible to the
operator can damage electronic components. Therefore, it is important that the operator
is grounded to the spectrometer housing before any electronic component is touched
inside the spectrometer.
Electrical grounding can be accomplished by using a grounding wrist strap (not supplied)
or touching a grounded object (e.g. radiator). The grounding wrist strap is the most effec
tive and preferred grounding method.
6.2Checking laser signal
To reliably scan and correctly determine the optical path difference it is very important to
have proper laser signals at laser A and laser B detectors. The signals should be
checked once a month.
-
6.2.1Amplitude
The amplitude of the two modulated signals must be in the range between 2 Vpp to
i
18
Vpp. In addition, the absolute value of both signals must be in the range +/-10.0 VDC.
Any signal above or below will be distorted or clipped, leading to wrong detection of the
laser zero crossings. When severely clipped, two or more zero crossings may be
detected and scanning becomes impossible.
68
IFS 125M User ManualBruker Optik GmbH
6.2.2Quick check in OPUS
1Check the modulated signals in OPUS:
Maintenance 6
☞On the Measure menu, select the Optics
Diagnostics command.
☞On the Instrument Status dialog click the
Interferometer icon.
☞On the dialog shown, click the Service
Info button to open the Scanner diagnos-
tics page.
i Laser A (or B) Amplitude [mV] is the
detected peak-to-peak voltage. Laser A (or
B) Offs [mV] is the detected mean (or DC)
voltage.
For both laser signals applies the following:
• -10,000 < Offs - Ampl/2
Offs + Ampl/2 <10,000
Ampl >2,000
The values are only determined at the current
scanner position, which is usually close to
zero path difference.
2The laser beam must be properly transferred to the two laser detectors. Nor-
mally, there is no re-adjustment required by the user.
3The laser amplifier has two gain settings,
which are selected by a corresponding bit in
the EEPROM of the beamsplitter. This kind of
setting can only be accessed via the WEB
interface. Details are described in
chapter
If the signal amplitude is too small and the low
gain is in effect (amplifier off), write the high
gain setting permanently into the beamsplitter
(and vice versa).
Table 6.1: Quick check in OPUS
8.3.3 Edit hardware configuration.
Bruker Optik GmbHIFS 125M User Manual
69
Maintenance 6
CAUTION
4To make the gain setting changes come into
effect you have to reset the controller. Alterna
tively, you can re-establish the contact to the
beamsplitter by the electronics as follows:
• Carefully lift off the interferometer compartment lid.
Open lid:
Risk of exposure to laser light, which can
cause permanent eye damage.
➣ Do not stare into the beam.
• Open the beamsplitter looking door once.
This is necessary to make the gain setting
changes come into effect.
-
5If the signal for one beamsplitter is too low (even with high gain setting) while
the others have good signal, this particular beamsplitter has to be replaced.
Details on changing a beamsplitter are described in chapter 5.8.
6If the signal is too low for all beamsplitters while the optical alignment is cor-
rect, the laser power itself may have dropped or one of the optical components in the laser beam may have been contaminated. The fiber patch cable
could be broken or misaligned. Contact Bruker service.
Table 6.1: Quick check in OPUS
70
IFS 125M User ManualBruker Optik GmbH
6.2.3Signal decrease over optical path
The decrease of the peak-to-peak signal amplitude over the full optical path should be
less than 35%.
Maintenance 6
1
Connect a dual channel oscillographa to the
TEST adapter located on the mobile electronics unit.
to the Tes t connector located on the electron
ics unit.
3Connect the other end of the laser signal test cable to the oscillograph.
4• On the OPUS Measure menu, select the Optic Setup and Service com-
mand.
• Click the Control Panel tabb.
Table 6.2: Signal decrease over optical path
-
Bruker Optik GmbHIFS 125M User Manual
71
Maintenance 6
5
• Scroll to Scanner Control.
• Select Front Short Adjust Mode. Write down the peak-to-peak amplitudes
(LASA and LASB) detected on the oscillograph.
• Select Fast Adjust Mode to check the signal amplitude over the full path
length.
• Select Back Short Adjust Mode.
• Wait until the scanner commutes around the end position.
• Compare the amplitudes detected on the end position with the amplitudes
on the front position. If the amplitude decrease is more than 35%, you have
to readjust the laser (see chapter
Table 6.2: Signal decrease over optical path
a. Firmware version June 2012 or later display the laser signal amplitudes (see chapter 6.2.2, step 1) in all
adjust modes, too. Therefore, connecting an oscillograph is not mandatory.
b. Alternatively, you can access the spectrometer by the WEB interface. Details are described in
chapter
8.3.5.
6.2.4).
72
IFS 125M User ManualBruker Optik GmbH
6.2.4Readjusting laser
NOTE
6.2.4.1Removing covers
1Removing scanning arm cover:
Maintenance 6
• Open the clips on each side.
• Put your hand below the clip and push the
clip upwards.
• Pull the clip towards you until you feel resistance.
i All the clips have to be in their outermost
position, otherwise you get problems when
lifting off the cover.
2
3Unscrew the Teflon-made counter nuts from
• Two persons are required to lift off the cover
of the long scanning arm, one at the front
and one at the back end.
• Put your hands below the hangers.
• Carefully lift off the cover from the scanning
arm.
Cover not lift off properly.
Risk of scanner damage.
➣ Lift off the cover straight and far
upwards.
the 2 detector hoods.
i The Teflon-made counter nuts are only
available if you use a liquid nitrogencooled detector.
Table 6.3: Readjusting laser - Removing covers
Bruker Optik GmbHIFS 125M User Manual
73
Maintenance 6
NOTE
CAUTION
4• Use a cross-head screwdriver (size: PH2)
5Use a flat-tip screwdriver to loosen the two
and rotate the 4 screws of the top lid coun
terclockwise.
➣You have to rotate over the snap-in resis-
tance. Then, the screw is pushed
upwards, and you can loosen it.
• Carefully lift off the sample compartment
cover.
knurled thumb screws, located on the bottom
of the trough.
-
6Completely remove the trough out of the sam-
ple compartment.
7Repeat step 1 to remove the cover of the detector and interferometer block.
8• Put your hands below the flange of the input
port and the diagonally located hanger.
• Carefully lift off the cover from the detector
and interferometer block.
Cover not lift off properly.
Risk of detector damage.
➣ Lift off the cover straight and far
upwards.
74
Open lid:
Risk of exposure to laser light, which can cause permanent eye damage.
➣ Do not stare into the laser beam.
Table 6.3: Readjusting laser - Removing covers
IFS 125M User ManualBruker Optik GmbH
6.2.4.2Connecting oscillograph to spectrometer
1Connect one end of the laser signal test cable
to the Tes t connector located on the electron
ics unit.
2Connect the other end of the laser signal test cable to the oscillograph.
3Use the Direct Control Panel on the spectrometer diagnostics page to move
the scanner. Click Stop Mode and then Back Short Adjust mode.
4The scanner moves to the outermost position
of the scanning arm.
Maintenance 6
-
5• Adjust the laser signal by the two screws on
top of the laser deflection unit.
i The screws are very sensitive. Turn slightly
and do not apply any force to the beamsplitter block. If the screws are too tight, the
central locking screw must be loosened a
bit.
• Adjust in the way that both laser signals are
at a maximum.
• Fix the central locking screw just finger tight.
Make sure that both laser signals remain at
their maximum.
Table 6.4: Readjusting laser - Connecting oscillograph to spectrometer
Bruker Optik GmbHIFS 125M User Manual
75
Maintenance 6
Figure 6.1:Beamsplitter area for the laser beam
A
Figure 6.2a: Adjustment screws of the laser
deflection unit
6.2.5Phase
The phase between laser A and B signals should be 90 +/- 45°. Check in regular scanning mode (after pressing Reinit Scanner) with a dual channel oscillograph connected to
the test adapter.
Make sure that the laser beam passes properly through the lower rectangular area (A in
figure
6.1) of the beamsplitter, which is specifically coated for the laser beam.
To slightly adjust the beam height turn all 3 adjustment screws of the laser deflection unit
to the same direction. Observe the laser signals on an oscillograph. Maximize the ampli
tude as described in chapter 6.2.4.
There is no way to adjust the phase, i.e. if the phase is out of specification, the particular
beamsplitter has to be replaced (chapter
5.8).
-
76
IFS 125M User ManualBruker Optik GmbH
6.2.6Remounting covers
AB
1• Put your hands below the flange of the input port and the diagonally
located hanger.
• Put on the cover of the detector and interferometer block.
➣Make sure that the two eyelets at the inner edge of the cover are
catched by the two guiding posts which are located on the console.
• Close the clips on each sidea.
2• Insert the trough into of the sample compart-
Maintenance 6
ment.
• Use a flat-tip screwdriver to fasten the two
knurled thumb screws, located on the bot
tom of the trough.
-
3• Use a cross-head screwdriver (size: PH2)
and rotate the 4 screws of the top lid clock
wise.
➣Press down the screw and rotate it once.
4• Insert the Teflon-made counter nuts (A).
i The Teflon-made counter nuts are only
available if you use a liquid nitrogencooled detector.
• Put on the protective caps (B).
Table 6.5: Adjusting laser - Remounting covers
-
Bruker Optik GmbHIFS 125M User Manual
77
Maintenance 6
NOTE
5• Two persons are required to put down the
6Close the clips on each side.
cover of the long scanning arm, one at the
front and one at the back end.
• Put your hands below the hangers.
Cover not put down properly.
Risk of scanner damage.
➣ Put down the cover straight and far
downwards.
Table 6.5: Adjusting laser - Remounting covers
a. If the pretension of any clip is too high or too low, it can be adjusted by turning the metallic screw nut with
a pipe wrench in the appropriate direction.
78
IFS 125M User ManualBruker Optik GmbH
6.3Replacing source
Standard source type:MIR source
Maintenance 6
Where is the source
located?
When has the source to be
replaced?
How is it indicated that the
source is defective?
Tools required:None
Handling instructions:The source becomes hot during spectro-
Has the source to be
aligned?
Table 6.6: Source features
6.3.1Source types
The spectrometer can be equipped with the following source types:
On the rear spectrometer side.
Only if it is defective.
The status light in the lower right end of
the OPUS interface becomes red. The
error message Source is broken or not connected is displayed.
meter operation. Avoid any skin contact.
All sources used are pre-aligned, and do
not have to be aligned after maintenance.
Source typeEstimated lifetime
(hours)
MIR source, globar, 12V9000Q328/7
NIR/VIS source, tungsten, 12V9000Q428/7
Blackbody source, emitting area:
100mm x 100mm
Table 6.7: Source types
Part number
S310
Bruker Optik GmbHIFS 125M User Manual
79
Maintenance 6
6.3.2Handling instructions for tungsten source
•Do not touch the tungsten bulb of the NIR source with your fingers as this contaminates the bulb and significantly reduces the service life.
•Wear gloves when changing the bulb and observe the operating instructions
indicated on the source packing.
CAUTION
NIR/VIS source operated out of the spectrometer:
Risk of UV radiation.
➣ Do not operate the NIR/VIS source out of the spectrometer.
➣ Operate according to the safety instructions to avoid risk of UV radiation.
80
IFS 125M User ManualBruker Optik GmbH
6.3.3Replacing procedure
CAUTION
NOTE
The replacing procedure is identical for all source types used in the spectrometer.
1Switch off the source:
Maintenance 6
☞On the OPUS Measure menu, select the
Optic Setup and Service command.
☞Click the Control Panel tab.
☞On the Internal SRC fieldclick OFF.
2
3• Loosen the knurled thumb screw (about one
4• Insert the replacement source into the holder.
Turn the knob on the source housing counterclockwise to remove the screw. Pull off the
cover.
Hot source:
Risk of skin burn.
➣ Wait until the light source has cooled
down to room temperature before you
replace it.
turn) of the clamping bar.
• Press the source downwards while rotating
the clamping bar side.
• Take the source out of the holder.
➣ The globar is very brittle. Do neither knock against nor drop it.
• The two pins of the holder must engage into the holes of the source.
• While pressing the source downwards, rotate the clamping bar over the
source.
• Tighten the knurled thumb screw of the clamping bar (about one turn).
5• Mount the source housing.
• Fasten the knob clockwise.
Table 6.8: Replacing source
Bruker Optik GmbHIFS 125M User Manual
81
Maintenance 6
WARNING
6• Click the OPUS status light on the lower
Table 6.8: Replacing source
6.4Replacing window
right end of the interface to reset the source
operating hour meter. The Instrument Status
window opens.
• Click the source icon and then the Service Info button. The Source diagnostics page
opens.
• Click the Reset button.
• Close the HTML view.
Where are the windows
located?
When have the windows to
be replaced?
Tools required:Window flange, Allen key
Handling instructions:• Handle with great care as windows are
On the rear and right side of the
interferometer cover
In case of too much dirt, cracking and
haze
very fragile and easily crack under the
influence of mechanical pressure.
• Do not touch the window surface to
avoid contaminations which can
decrease the infrared transparency.
• Avoid generating dust when the window
has been cracked.
• Observe the safety instructions on the
window packaging.
Non-observance of safety instructions on packaging:
Serious health problems or even death.
➣ Never clean the windows.
➣ Observe the safety instructions.
Table 6.9: Window features
82
Window material used:• MIR: KBr
• NIR/VIS/UV: Quartz or CaF
IFS 125M User ManualBruker Optik GmbH
2
6.4.1Window material
CAUTION
CAUTION
CAUTION
CAUTION
NOTE
The following table contains the most important window material which can be used in
connection with the spectrometer. For further possible window material, contact Bruker
service.
Window materialPropertiesPart number
Maintenance 6
Barium fluoride (BaF2)
Harmful
➣ Observe the safety
data sheet.
Calcium fluoride (CaF2)
Harmful
➣ Observe the safety
data sheet.
Calcium fluoride (CaF2)
Harmful
➣ Observe the safety
data sheet.
49.5 mm diameterF131-17
45 x 3 mm, wedge 6 mradF162-3
49.5 mm diameterF131-3
Calcium fluoride (CaF2)
90 x 3 mmF165-3/M
Harmful
➣ Observe the safety
data sheet.
Potassium bromide (KBr)
KBr window exposed to
moisture:
Property damage.
➣ Avoid any exposure to
moisture.
Table 6.10: Window material
Bruker Optik GmbHIFS 125M User Manual
45 x 5 mm, wedge 6 mradF162-5-5
83
Maintenance 6
NOTE
NOTE
CAUTION
CAUTION
Window materialPropertiesPart number
Potassium bromide (KBr)
KBr window exposed to
moisture:
Property damage.
➣ Avoid any exposure to
moisture.
Potassium bromide (KBr)
KBr window exposed to
moisture:
Property damage.
➣ Avoid any exposure to
moisture.
Quartz45 x 2 mmF162-1
Quartz90 x 3 mmF165-1/M
Zinc selenide (ZnSe)
49.5 mm diameterF131-5
90 x 10 mmF166-5/M
45 x 3 mm, wedge 6 mradF162-11
Toxic
➣ Observe the safety
data sheet.
Zinc selenide (ZnSe)
Toxic
➣ Observe the safety
data sheet.
Table 6.10: Window material
49.5 mm diameterF131-11
84
IFS 125M User ManualBruker Optik GmbH
6.4.2Replacing procedure
Figure 6.3:Window assembly
A
B
C
D
E
F
G
1• Use a hexagon socket setscrew (2mm) to
2Check whether a signal is detected and the optics works properly. More
details are described in chapter
Table 6.11: Replacing window
Maintenance 6
loosen the screws from the window cap.
Rotate the screws counterclockwise.
• Loosen the screws from the fixation ring.
• Replace window as described in
chapter
5.6.
6.4.3.
6.4.3Example of window assembly
Figure 6.3 exemplifies the assembly of a quartz window which consists of the following
components:
Bruker Optik GmbHIFS 125M User Manual
Definition
AWindow flange
BO-ring for the window flange (A)
Table 6.12: Parts of window assembly
85
Maintenance 6
Definition
CWindow
DDistance ring (only required for thin win-
dow types)
EO-ring for fixation ring (F)
FFixation ring
GWindow port cover
Table 6.12: Parts of window assembly
•Insert the O-ring (B) into the window flange (A).
•Put the window (C) onto the O-ring in the window flange. Always wear gloves.
•Position the distance ring (D), if applicable.
•Position the O-ring (E) into the fixation ring (F).
•Put the fixation ring (F) onto the window assembly.
•Use a hexagon socket setscrew (2 mm) to gradually tighten the screws one after
the other.
CAUTION
Damaged window:
Risk of breaking.
➣ Never insert a window which shows signs of damage.
➣ Never insert a window thinner than specified.
86
IFS 125M User ManualBruker Optik GmbH
6.5Evacuating detector
Maintenance 6
Where are the detectors
located?
What detector type has to
be evacuated?
When has the detector to
be evacuated?
How is it indicated that the
detector has to be evacu
ated?
If the detector hold timea considerably
decreases (i.e. a hold time of less than
70% of specified hold time).
Most liquid-nitrogen cooled detectors are
mounted in re-pumpable vacuum dewars
(except for those which are permanently
sealed). Usually, the detector dewar holds
its vacuum for several months.
• Condensation water exists on the detec-
-
tor outer surface.
• Outer surface of detector is iced.
• Broad absorption around 3,250 cm-1 (ice
band)
uum pump
• Vacuum adapter
More details are described in
chapter
Table 6.13: Detector features
a. The hold time indicates how long the cooling effect of the liquid nitrogen lasts.
6.5.1.
Bruker Optik GmbHIFS 125M User Manual
87
Maintenance 6
A
B
C
D
6.5.1Evacuating tools required
Too lDefinition
Different pump models possibleTurbo molecular pump/oil-free high-vac-
uum pump (which achieves an ultimate
vacuum of <10
Adapter
-5
hPa)
Adapter with flexible hose and flange:
• A: adapter
• B: flange
• C: flexible metal hose
• D: NW 25 flange
Table 6.14: Evacuation tools
The evacuating equipment is not included in the delivery scope of the MCT detector.
Bruker offers evacuation equipment, and the service of evacuating (Part No: SD128). In
the latter you have to send the MCT detector to Bruker.
88
IFS 125M User ManualBruker Optik GmbH
6.5.2Evacuating procedure
closed
open
A BC D E FGH
Figure 6.4:Vacuum adapter, cross section
The following cross-section of the adapter aims to better illustrate the evacuating procedure.
Maintenance 6
ACoupling nut
BO-ring retainer
CO-ring
DUnion nut (to remove the valve closure of
EConnecting piece for vacuum pump
FVacuum adapter
GWasher and O-ring packing
HKnob
Table 6.15: Parts of vacuum adapter
Before evacuating the dewar, the dewar must not contain any more liquid nitrogen. The
i
dewar temperature must be about room temperature (maximum 60°C).
Evacuating the dewar takes several hours. Therefore, it is recommended to evacuate
the dewar over night. The final pressure in the dewar has to be less than 10-5hPa.
Definition
the detector dewar)
Bruker Optik GmbHIFS 125M User Manual
89
Maintenance 6
1Turn the main power switch in the OFF posi-
2Remove the MCT detector from the spectrometer. Detailed information are
given in chapter
3Connect the connecting piece of the vacuum
4Pull the knob to the open position.
tion to switch off the spectrometer.
5.10.
adapter to the vacuum pump.
The connecting piece has an outer diameter
of 9.7mm.
5Loosen the coupling nut.
6Carefully slip the vacuum adapter over the connecting piece of the detector
dewar and fasten the coupling nut finger-tight.
The O-ring must engage properly into the corresponding groove at the
detector port. Additional tightening is not necessary
7Push the knob into the closed position until
the union nut of the vacuum adapter is in con
tact with the dewar evacuation valve.
8• Screw the union nut into the dewar evacua-
tion valve by turning the knob clockwise; 2 to
3 rotations are sufficient. Do not tighten the
union nut completely as this would prevent it
from being unscrewed later.
• Evacuate the vacuum adapter by using the
vacuum pump.
-
90
Table 6.16: Evacuating detector
IFS 125M User ManualBruker Optik GmbH
Maintenance 6
9Pull the knob to the open position in order to
open the dewar evacuation valve and evacu
ate the dewar.
10• If the desired vacuum has been achieved,
gently push the knob to the closed position
in order to close the dewar evacuation valve.
• Purge the section between vacuum pump
and vacuum adapter.
11Rotate the knob several turns counterclock-
wise in order to screw the union nut off the
dewar evacuation valve. Be careful not to
purge the dewar.
-
12Pull the knob to the open position.
13• Loosen the coupling nut and remove the
vacuum adapter from the connecting piece
of the dewar.
• Reinstall the MCT detector into the spectrometer (chapter 5.10)
14Check whether a signal is detected and the optics works properly. More
details are described in chapter
Table 6.16: Evacuating detector
5.6.
Bruker Optik GmbHIFS 125M User Manual
91
Maintenance 6
Figure 6.5:Scanner drive cable unit
A
BC
6.6Checking drive cable
The drive cable for the moving mirror assembly must occasionally be checked for tension. Especially, when installing a new cable it is recommended to check the cable tension periodically to avoid slipping.
Definition
ACapstan
BCounter screw
CThumb screw
Table 6.17: Parts of scanner drive cable unit
92
IFS 125M User ManualBruker Optik GmbH
6.6.1Checking procedure
NOTE
1Removing scanning arm cover:
Maintenance 6
• Open the clips on each side.
• Put your hand below the clip and push the
clip upwards.
• Pull the clip towards you until you feel resistance.
i All the clips have to be in their outermost
position, otherwise you get problems when
lifting off the cover.
2
3There must be no signs of cable slipping at
Table 6.18: Checking drive cable
• Two persons are required to lift off the cover
of the long scanning arm, one at the front
and one at the back end.
• Put your hands below the hangers.
• Carefully lift off the cover from the scanning
arm.
Cover not lift off properly.
Risk of scanner damage.
➣ Lift off the cover straight and far
upwards.
the capstan when the scanner turns at high
velocity in long scan mode.
Bruker Optik GmbHIFS 125M User Manual
93
Maintenance 6
viewpoint
second
tool
6.6.2Adjusting procedure
1Loosen the counter nut. Use a tool and turn it
2• Adjust the drive cable tension by the thumb
counterclockwise.
screw.
• Turn the thumb screw counterclockwise to
increase the tension.
• Be careful not to overtighten the thumb
screw to prevent the cable from tearing.
3• Fix the tension by the counter nut.
• You may hold the screw in place by a second tool.
4Remount the cover onto the scanning arm.
Table 6.19: Adjusting drive cable
94
IFS 125M User ManualBruker Optik GmbH
6.7Transporting spectrometer
NOTE
When transporting the spectrometer you have to fix the scanner on the scanning arm,
remove the beamsplitter and demount the complete scanning arm. Mind the order
described in the following chapters.
Moving the spectrometer within a building can be done without demounting the scanning
i
arm.
6.7.1Removing covers
1Removing scanning arm cover:
• Open the clips on each side.
• Put your hand below the clip and push the
clip upwards.
• Pull the clip towards you until you feel resistance.
Maintenance 6
i All the clips have to be in their outermost
position, otherwise you get problems when
lifting off the cover.
2• Two persons are required to lift off the cover
of the long scanning arm, one at the front
and one at the back end.
• Put your hands below the hangers.
• Carefully lift off the cover from the scanning
arm.
Cover not lift off properly.
Risk of scanner damage.
➣ Lift off the cover straight and far
upwards.
Table 6.20: Transporting spectrometer - Removing cover from scanning arm
Bruker Optik GmbHIFS 125M User Manual
95
Maintenance 6
3Unscrew the Teflon-made counter nuts from
the 2 detector hoods.
i The Teflon-made counter nuts are only
available if you use a liquid nitrogencooled detector.
4
5Use a flat-tip screwdriver to loosen the two
6Completely remove the trough out of the sam-
• Use a cross-head screwdriver (size: PH2)
and rotate the 4 screws of the top lid coun
terclockwise.
➣You have to rotate over the snap-in resis-
tance. Then, the screw is pushed
upwards, and you can loosen it.
• Carefully lift off the sample compartment
cover.
knurled thumb screws, located on the bottom
of the trough.
ple compartment.
-
96
7Repeat step 1 to remove the cover of the detector and interferometer block.
Table 6.20: Transporting spectrometer - Removing cover from scanning arm
IFS 125M User ManualBruker Optik GmbH
8• Put your hands below the flange of the input
NOTE
Table 6.20: Transporting spectrometer - Removing cover from scanning arm
6.7.2Moving and fixing scanner
Maintenance 6
port and the diagonally located hanger.
• Carefully lift off the cover from the detector
and interferometer block.
Cover not lift off properly.
Risk of detector damage.
➣ Lift off the cover straight and far
upwards.
1• Manually move the scanner from its initial
position to the end of the scanning arm.
➣There is a high friction when moving the
scanner.
• Alternatively, use the Direct Control Panel
command on the spectrometer diagnostics
page to move the scanner. Click Stop Mode
and then Back Short Adjust Mode.
➣ Once the scanner has reached the back
position, click Stop Mode.
• Do not touch the scanner on the red-colored
surfaces (top side or spring leaves).
• Hold the scanner either at the side edges
(see arrow), or at the back.
Table 6.21: Transporting spectrometer - Moving and fixing scanner
Bruker Optik GmbHIFS 125M User Manual
97
Maintenance 6
2Loosen the fixing screw counterclockwise to
3Rotate the fixing screw clockwise into the fix-
remove it from its stand-by position.
ture on the scanner.
4When the screw is fixed, the outer scanner
cannot be moved any more.
5To fix the inner scanner use the supplied
clamp.
➣To position the clamp make sure that the
clamp screws fit into the boreholes (see
arrow), located on the inner scanner
housing.
➣Fasten the clamp screws (see arrows).
Pay attention, as the clamp touches the
top side of the movable structure, but
does not press it down! This avoids defor
mation of the spring leaves.
-
98
Table 6.21: Transporting spectrometer - Moving and fixing scanner
IFS 125M User ManualBruker Optik GmbH
Maintenance 6
6To fix the small retaining clip of the clamp
make sure that the clip is exactly positioned
below the boreholes (see arrow), located on
clamp.
➣The top of the scanner and the clamp
must be in touch, or must have a very
small gap in between.
Fasten the clip screws.
Table 6.21: Transporting spectrometer - Moving and fixing scanner
6.7.3Removing beamsplitter
Remove the beamsplitter to prevent it from breaking while demounting the scanning
arm, or during transport.