Broda Broda Centric Positioning Wheelchair (30VT) Manual User manual

Page 1
Operating Manua l
Model:
30VT Semi-Recliner
Caregiver: Before using this chair, read this manual thoroughly and save for future reference.
Page 2
Contents
FOREWORD ................................................................................................................................................................. 1
GENERAL INFORMATION ........................................................................................................................................ 2
1 DEFINITIONS ...................................................................................................................................................... 3
2 SAFETY REQUIREMENTS ................................................................................................................................ 4
2.1 Training ........................................................................................................................................................ 4
2.2 Application ................................................................................................................................................... 4
2.3 Shipping and Storage ................................................................................................................................... 4
2.4 Pre-Service Check ........................................................................................................................................ 5
2.5 Hazards ........................................................................................................................................................ 5
2.5.1 Position of Chair - “Danger of Falling” ................................................................................................... 5
2.5.2 Position of Seat Tilt - “Danger of Tipping” ............................................................................................. 5
2.5.3 Location of Chair - “Danger of Tipping or Fa lling Objects” ................................................................... 5
2.5.4 Total Lock Wheel Brakes - “Danger of Falling” ..................................................................................... 5
2.5.5 Re-Positioning of Resident - “Danger of Clamping” ............................................................................... 6
2.5.6 Unintended Mo vement - “Danger o f Falling or Collision” ...................................................................... 6
2.5.7 Resident Clothing - “Risk of Injury to Resident’s Skin” ......................................................................... 6
2.5.8 Improper Restraint Use - “Risk of Serious Injury” .................................................................................. 6
2.6 Improper Use................................................................................................................................................ 7
2.7 Cleaning ....................................................................................................................................................... 7
2.8 Maintenance ................................................................................................................................................. 8
2.9 Resident Specific Instructions .................................................................................................................... 10
3 OPERATION AND MOVEMENT ..................................................................................................................... 11
3.1 Seat Tilt ...................................................................................................................................................... 11
3.2 Back Recline (Model 30VT only) .............................................................................................................. 11
3.3 Elevating Leg Rest ..................................................................................................................................... 11
3.4 Flip Down Footrest .................................................................................................................................... 12
3.5 Removable Wings ...................................................................................................................................... 12
3.6 Shoulder Bolsters ....................................................................................................................................... 13
3.7 Height Adjustable Armrests ....................................................................................................................... 13
3.8 Adjustable Seat Height ............................................................................................................................... 14
3.9 Casters ........................................................................................................................................................ 14
3.10 Accessories (Options) ................................................................................................................................ 14
4 INSPECTION AND FUNCTIONAL TESTING ................................................................................................ 15
4.1 Inspection ................................................................................................................................................... 15
4.2 Functional Testing ...................................................................................................................................... 15
5 TECHNICAL INFORMATION .......................................................................................................................... 16
5.1 Specifications ............................................................................................................................................. 16
5.2 Labels ......................................................................................................................................................... 17
5.3 BRODA Seating Basic 30VT Parts Schematic Drawing ........................................................................... 18
5.4 Broda Seating Basic 30VT Parts Schematic Legend ................................................................................. 19
6 COMMON TROUBLESHOOTING AND REPAIR .......................................................................................... 20
6.1 Troubleshooting for Gas Cylinders ............................................................................................................ 20
7 WARRANTY ...................................................................................................................................................... 22
ii
Page 3

GENERAL INFORMATION

This document provides guidance on the safe and effective operation of the BRODA chair.
Information in this manual must be followed at all times.
Anyone involved with the operation or maintenance of the BRODA chair, including the resident’s family members, must read this operating manual before using the chair.
The resident’s primary caregiver is responsible for ensuring that anyone who is unfamiliar with, unwilling, or unable to adhere to the safety and operating instructions, is not permitted to operate or move the chair.
A copy of this instruction manual must always be available.
BRODA accepts no liability for damages, injury or accidents caused by operating errors, improper maintenance, or disregard of the instructions in this manual, including any resident specific instructions.
BRODA reserves the right to make changes to the specifications, dimensions, functions, or components of its products without notice. Product representations in this manual may vary from delivered products.
Each BRODA chair has a unique identifying serial number that must be maintained on the chair as well as with any equipment records.
2
Page 4

1 DEFINITIONS

“BRODA” means BRODA Enterprises Inc. doing business as BRODA Seating.
“BRODA chair” refers to Model 30VT.
“Long-Term Care Institution” refers to a nursing home, hospital, or o ther healthcare facility that provides health and personal care to its residents on a long-term basis.
“Resident” refers to an individual living in a long-term care insti t ution under the care of professional caregivers.
“Professional Caregiver” refers to doctors, nurses, therapists, nursing aids, healthcare aids, and other specialists who work in a long-te rm care institution and provide health and personal care to its residents.
“Caregiver” refers to any person in a long-term care institution who is appropriately trained to provide care or services to the resident or the chair used by the resident and may include the resident’s family members or guardian.
“Seat Tilt” refers to changing the relative angle between the chair’s seat and the chair frame or ground without changing the relative angle between the back and the seat.
“Back Recline” refers to changing the relative angle between the chair’s back and the seat.
“Footrest Elevation” and “Leg Rest Elevation” refers to changing the relative angle between the chair’s footrest and/or leg rest and the seat.
“Transfer(s)” refers to the movement of a resident into or out of the chair with the assistance of their caregiver(s).
“Mechanical Transfer(s)” refers to the movement of a resident into or out of the chair with the assistance of their caregiver(s) using a patient lift or other assistive device that bares the weight of the resident.
3
Page 5

2 SAFETY REQUIREMENTS

2.1 Training

Before the chair is put into service, this manual must be read thoroughly by the caregiver(s) directly responsible for the resident’s care.
After the chair is put into service, this manual must be re ad thoroughly by any new caregivers prior to operating or moving the chair.
For the purpose of this manual, a resident’s family member who shares responsibility for their care is considered a caregiver and is subject to the same competency before being permitted by the resident’s primary caregiver to operate of move the chair.
Prior to first use, the customer must arrange for an in-service on the operation and safety requirements in this manual, must be given to the resident’s caregivers by the local BRODA representative who supplied the chair.
The primary caregiver must maintain a list of caregivers who have read this manual and who they have authorized to operate and move the chair.
The resident’s primary caregiver is responsible for ensuring that anyone who is unfamiliar with, unwilling, or unable to adhere to the safety and operating instructions, is not permitted to operate or move the chair.
The operations of the chair must be performed by the resident’s primary caregiver who is responsible for seating. All The operations and adjustments performed should be done in a manner to ensure the overall safety, comfort and well-being of the resident, caregiver and thir d party. All operations and adjustments required for the resident should be determined by the resident’s primary caregiver who is responsible for seating.

2.2 Application

BRODA chairs are intended exclusively for residents of long-term care institutions who are under the care of professional caregivers. The suitability of a BRODA chair must be determined by a qualified caregiver who is familiar with the s eating need s of the intended resident. An y other use of the chair is excluded from possible liability claims.
The Basic chairs are not to be used in the shower. The frame and components will rust, and will void the warranty.
The chairs are not explosion resistant and must not be used where there are flammable gases or liquids present (e.g., anesthetics, volatile solvents and cleaners, etc.)
BRODA chairs are designed for use with specific BRODA parts and accessories. The use of non-BRODA parts or accessories with a BRODA will void the warranty and is excluded from possible liability claims.
BRODA chairs may only be used as desc ribed in this manual and with proper regard for recognized healthcare and workplace safety and accident prevention practices.

2.3 Shipping and Storage

BRODA chairs should be shipped and stored in an upright position and not stacked higher tha n 3 boxes. No other materials should be shipped or stored on top of a BRODA box. BRODA boxes should not be placed on pallets.
BRODA chairs should be shipped and stored at temperatures between -20ºC and 40ºC. BRODA chairs should not be used until they are between 0ºC and 30ºC.
4
Page 6
BRODA chairs should be kept in a clean, dry environment. Do not leave BRODA chairs outdoors as it may cause the paint to peel.
Upon receipt, the shipping carton must be immediately examined for damage. Any damage should be noted on the delivery receipt and a request for inspection by the transportation company should be made. Next, the shipping carton should be opened and the chair must be examined for concealed shipping damage. If the chair appears to be damaged, do not use the chair. File a concealed damage report with the transportation company.

2.4 Pre-Service Check

BRODA chairs are delivered fully assembled. If the chair does not appear to be ready to use upon receipt, immediately contact your supp lie r a nd do not put the chair into service until any concerns have been resolved.
Visually inspect the chair for damage, missing parts, and loose fasteners prior to testing the chair’s functions. Functional testing must be successfully completed after visual inspection and before use. These obligations apply to the chair’s first use and to all subsequent uses (Section 4: Inspection and Functional Testing).

2.5 Hazards

2.5.1 Position of Chair - “Danger of Falling”

After a resident is transferred into a chair, assess the amount of tilt required. We recommend that the chair’s seat be tilted sufficiently to prevent the resident from sliding or falling forward off the chair. The amount of seat tilt used should be determined by the resident’s caregiver who is responsible for seating.
We recommend that the resident’s feet be correctly positioned on the footrests and slightly to fully elevated to prevent the resident from sliding or fa lling forward off the chair. The amount of elevation used should be determined by the resident’s caregiver who is responsible for seating.

2.5.2 Position of Seat Tilt - “Danger of Tipping”

We recommend that the chair’s seat be tilted sufficiently to prevent an agitated resident from tippin g the chair forward or backward, or from slumping and sliding in the chair. The amount of seat tilt used should be determined by the resident’s caregiver who is responsible for seating. Always ensure that the resident is properly positioned before operating the seat tilt.

2.5.3 Location of Chair - “Danger of Tipping or Falling Objects”

We recommend that when a resident has been moved to their destination, the chair is placed where the resident cannot reach handrails or other objects, fixed or movable. This is to prevent the resident from pulling the chair over or pulling themselves off the seating surface and to prevent the resident from pulling movable objects onto the chair and themselves.
We recommend that the chair be used in a supervised area to prevent untrained residents, caregivers, or third parties from unaut horized operation, movement, or unsafe actions such as sitting or leaning on the reclined back, e l evated footrest, or the armrests. These actions, if not prevented, put the chair at risk of tipping or damage to the chair.
We recommend that a chair only be located on a level surface to minimize the risk of tipping over.

2.5.4 Total Lock Wheel Brakes - “Danger of Falling”

The special casters found on the BRODA chair have total lock brakes which prevent the wheels from turning and swiveling. The brakes must always be applied when:
1) the chair is not in use;
2) a resident is being transferred (moved) into or out of the chair; and,
3) the chair is not being moved by a caregiver.
It is important to note that if the wheel locks are applied while the patient is in the chair, that the caregiver does not leave the patient unattended, especially those patients who have the capability or tendency to move the chair and/or those who may be agitated. This could cause harm to the patient if they attempt to move the chair while the wheel locks are applied.
5
Page 7
Note that removing and attaching the footrest may be easier for the caregiver with the brakes applied
Failure to follow these instructions will unnecessarily increase the risk of serious falls by residents, caregivers, or third part i es caused by t he chair unintentionally moving.

2.5.5 Re-P ositioning of Resident - “Danger of Clamping”

BRODA chairs offer the benefits of seat tilt, back recline, leg rest/footrest elevatio n, and moveable arms. During the movement of any of these functions, the following safety measures must be observed:
1) The resident’s arms must be positioned on the armrests or inside the chair frame with their hands
on their body or on.
2) The resident’s feet must be correctly positioned on the footrest.
3) All of the chair’s brakes have been applied.
4) Only one caregiver at a time attempts to operate the chair’s functions.
5) Only one chair function is operated at a time.
6) The rear wheels are in the trailing position, behind the chair frame.
7) The residents’ and caregivers’ body are clear of all pinch points before operating the chair’s
functions.
Failure to follow these safety measures can put the residents’ or caregivers’ limbs at risk of injury. Resid e nts who may be unaware of their body position or unable to maintain a safe body position are at the most risk of the danger from clamping and caregivers should be more cautious with these residents. A second caregiver may be required to ensure the safety of these residents during these operations.

2.5.6 Unint ended Movement - “Danger of Falling or Collision”

We recommend BRODA chairs for indoor use within a long-term care i ns t itution and where there is not enough slope to cause the chairs to move unaided. Chairs used where the surfa ce is u neven or sloped are at risk of unintended movement and could become a serious danger to the resident, caregiver(s), or a third party. We recommend that BRODA chairs are located away from stairwells, elevators, and exterior doorways within a long­term care institution.
Outdoor use is appropriate only under the strict supervision and full atte ntio n of a caregiver who is physically capable of preventing any unintended movement over any surfaces that are to be traveled on. We recommend that a second caregiver assist when the chair is moved over surfaces that could cause significant unintended movement.

2.5.7 Reside nt Clothing - “Ris k of Inj ury to Resident’s Skin”

We recommend that residents only be seated while they are fully dressed in clothing that meets the needs of their specific condition. If after being fully dressed, a resident’s bare arms, legs, or body could still come into direct contact with the vinyl straps or vinyl pads, we recommend the use of a covering, such the BRODA terry cloth covered seat and/or back pad or a folded cloth bed sheet to prevent direct contact. Direct contact of bare skin on the straps over a period of time could cause moisture on the resident, and/or cause the skin to stick to the straps. Prio r to the operation or movement of a chair with a covering, the caregiver must ensure any covering placed on the chair does not come into contact with any moving parts and is securely placed, so not to slip.

2.5.8 Improper Restraint Use - “Risk of Serious Injury”

We recommend that alternatives to physical restraints be used with residents while seated in the chair except under the specific instructions of the resident’s primary caregiver and with permission of the resid ent’s family o r guardian. Physical restraints have been identified as a common cause of serious injury to residents while they are seated. We recommend that the primary caregiver responsible for seating first consider the re-positioning options available in the chair to reduce the risks of slidin g, falling, or self-injury.
If a physical restraint is determined to be appropriate to prevent sliding or falling out of the chair, we recommend that the BRODA thigh belt be used. If a physical restraint is determined to be appropriate to prevent self-inj ur y from resident movement, we recommend the BRODA HSP padding package. In all cases, it is the resident’s primary caregiver that must take responsibility for the safety of the resident if restraints are used.
6
Page 8

2.6 Improper Use

As outlined , the improper use of the chair is dangerous to the resident, caregivers, or third parties, and can consist of, but is not limited to the following:
1) Unauthorized operation of the chair’s functions.
2) Unauthoriz ed mo ve ment of the chair.
3) Inappropriate use of the chair for a resident who has not been assessed by a qualified caregiver
responsible for their seating.
4) Failure to frequently reposition the resident in the chair
5) Attempting to operate of multiple chair functions simultaneously by one or more caregivers.
6) Attempting to move the chair with the b rake(s) applied.
7) Leaving the resident unattended in the chair near other objects.
8) Leaving an agitated resident in the chair in an unsupervised area.
9) Leaving a resident unattended.
10) Leaving a resident in a chair on a sloping surface.
11) Leaving a chair unattend ed on a sloping surface.
12) Using non-BRODA accessories on the chair.
13) Using the chair at temperatures below 0ºC.
14) Using the chair as a shower or bathing chair.
15) Using the cha i r for any use other than i t s intended purpose.

2.7 Cleaning

BRODA chairs should be wiped clean with soap and water. Diluted household strength ammonia or chlo rine based cleaner, as well as a hospital grade cleaner may be used if necessary.
Refer to the following guide for cleaning certain staining agents (Applie s to the Permablok3® vinyl, as per the vinyl manufacturer’s recommendations) The frame and components of the chair can be cleaned using the same procedure as the vinyl. The steps below are also located in the Cleaning Guide found on the back of the Broda chair
Staining Agent Cleaning Procedure
Eye Shadow / Mascara Step 1 Grease / Suntan Lotion / Chocolate St e p 1 Blood / Bodily Fluids Step 2 Red Lipstick Step 2 Oil Base Paint / Tar / Asphalt Step 3 Other Tough Stains Step 4 or 5*
Step 1: Remove excess spill with a damp cloth. Clean with a 1:1 mix of Ivory® liquid and water. Rinse with clean water and dry.
Step 2: Use a straight application of concentrated cleaners such as Formula 409® or Fantastik® Spray Cleaner. Then wipe with a clean cloth. If usin g a hosp ita l grade cleaner, follow the diluting instructions on the label.
Step 3: Use a 1:1 mix of ammonia and water or a 1:4 mix of bleach and water. Rinse with clean water and dry.
Step 4: Use a straight application of naphtha (lighter fluid). Rinse thoroughly with clean water and pat surface dry.
Step 5: Use a 1:1 mix of isopropyl alcohol and water. If the stain persists, use straight alcohol. Rinse thoroughly
with clean water and pat dry. If the stain remains, use a 1:1 mix of acetone and water. Rinse with clean water and pat surface dry.
*Note: for cleani ng that requires steps 4 o r 5 – Use a soft cotton cloth saturated with the cleaning material and rub the stain in circles 10 times. Pat dry with another soft cotton cloth, and check results.
7
Page 9
For best cleaning results, certain parts and padding on the chair should be removed: Instructions for removal of the following parts are located in this manual: Ensure that the entire chair including removable and non re mo vab le parts is thoroughly cleaned.
Pay close attention to the frame, sockets, fasteners, parts and casters, as well as the padding for an even, thorough clean.
Remove parts for cleaning if desired:
1) Wings
3) Footrest
4) Padding
Thoroughly clean all removable parts and padding, as well as the frame and components around the parts and padding that have been removed.
BRODA chairs should not be cleaned with petroleum based cleaners. Any petroleum based products that come in contact with any vinyl surface should be removed as quickly as possible. Petroleum based products make vinyl brittle and will damage the seating surface and cushions. Metal parts and cushions should be wiped completely dry after cleaning. Do not launder vinyl padding. Terry cloth seat and back covers can be laundered and tumble dried.
Do not allow the chair to air dry. Leaving the chair to air dry and not ensuring the entire chair is completely dry may cause the frame and components to rust. Solvents such as those found in spray lubricants should not be used on BRODA Chairs as they can damage moving parts.
Frequency and method of cleaning the chairs should be determined by facility infection control protocols. If visibly soiled, thoroughly clean the chair immediately as per the cleaning instructions in this manual.
The cleaning instructions in this manual are guidelines only. Results may vary under actual conditions. The information does not relieve the user of proper and safe use of the product and all cleaning agents and consideration for the overall cleaning maintenance of the chair. Cleaning and care instructions must be followed in conjunction with facility infection control protocols.
BRODA will not accept warranty or liability claims on chairs that have not been cleaned according to the instructions or cared for in proper regard for patient, caregiver and third party safety and hygiene.. The use of certain agents may be harmful to the surface appearance and lifespan of the vinyl. BRODA assumes no responsibility resulting fr om the use of such cleaning agents to t he vinyl.
Touch up paint is available for the p owder coated frame. Please call Broda’s Customer Service Department if touch up paint is required. Do not leave BRODA chairs outdoors as the frame is not UV protected. Leaving the chairs outdoors could cause the paint to peel.
Formula 409® is a trademark of the Clorox Company. Fantastik™® is a trademark of DowBr ands, Inc. Trademark of the Dow Chemical Company. Ivory® is a trademark of Proctor and Gamble.

2.8 Maintenance

The maintenance on a BRODA chair will vary with the amount of use and the condition of the resident using the chair.
In regular use, after the initial inspection and functional testing, the chair should be inspected and tested bimonthly. We recommend visually inspecting for signs of wear, damage, loose or missing fasteners, and other safety concerns. Periodic testing of chair functions is also necessary. If a breakage, defect, or operational problem is detected, the chair must be successfully repaired, inspected and tested for function before it is returned to service.
The chair should be inspected and tested as often as each use if the chair is used by:
1) Aggressive or agitated residents.
8
Page 10
2) Residents who ha ve invo lu ntary movements.
3) Residents weighing over 220 lbs.
4) A facility with irregular or sloped surfaces.
5) Any unauthorized person.
Do not use lubricants that contain solvents. Solvents may compromise any rubber components on the chair. If necessary, white food grade grease or lubricant can be used on any sliding components.
9
Page 11

2.9 Resident Specific Instructions

The primary caregiver responsible for the specific resident’s seatin g shall add additional instructions necessary for the safe and effective use of the chair based on their professional experience and kno wledge of the resident’s specific conditions and requirements.
BRODA representatives are not professional caregivers and will not know the specific requirements of the individual using the c hair. BRODA relies on the knowledge, experience, and judgment of the resident’s professional caregiver to ensure the specific resident’s safety and comfort needs are satisfied while using the chair.
These instructions form an essential part of the Safety Requirements for using the chair and must be made available to all caregivers. Space is provided below to include these instructions.
BRODA Enterpr i ses Inc. 560 Bingemans Centre Drive Kitchener, Ontario Canada N2B 3X9 Phone: 1-519-746-8080 Toll Free Can/US: 1-800-668-0637 Fax: 1-519-746-8616 Email: sales@brodaseating.com URL: www.brodaseating.com
10
Page 12

3 OPERATION AND MOVEMENT

Gas cylinders allow the seat to be tilted to any position
achieved.
Gas cylinders allow the back to b e r eclined to any
Gas cylinders allow the leg rest to be positioned
1
1 1
Safety measures as described in Section 1 must be observed when operating the chair’s functions or moving the chair. Only authorized caregivers should attempt to operate or move a chair. The caregiver must use proper body mechanics when operating the chair, and be prepared to support the weight of the resident while operating the chair.

3.1 Seat Tilt

from most upright to the lowest wit hin i ts ra nge. Use seat tilt to safely position the resident, redistribute pressure and increase comfort.
To Activate the Seat Tilt
-Place your right hand on the bar to a ssist movement and squeeze the tilt handle with your le ft hand .
-Raise or lower the seat tilt until the desired angle is

3.2 Back Recline:

position from the most upright to the lowest within its range. Use back recline to position the resident, create pressure redistribution, increase comfort, and assist with daily care activities.
WARNING: Always tilt slightly before reclining to prevent resident from sliding.
To Activate the Back Recline
-Place your left hand on the top bar to assist movement and squeeze the recline hand l e with your right hand.
-Raise or recline the back until the desired angle is achieved.

3.3 Elevating Leg Rest

infinitely from horizontal to vertic a l. The leg rest elevates with the recline. It is also independently adjustable.
To Adjust the Leg Rest:
-Hold the leg rest to assist movement and squeeze the leg rest handle.
-Raise or lower the leg rest to the desired angle.
The leg rest elevates with the back recline but not wit h the seat tilt. If the back is in the recline p osition, the leg rest cannot be lowered independently.
11
Page 13

3.4 Flip Down Footrest

The flip down footrest is removable and length
left and right side.
Wings support the Shoulder Bolsters which provide
Length
1
2
adjustable.
To Attach the Footrest:
-Insert the right end of the footrest upper cross piece into the desired hole in the right side of the leg rest frame.
-Insert the left end of the footrest upper cross piece into the corresponding hole in the left sid e of the leg rest frame.
-Release the tension on the spring mechanism.
To Remove t he Footrest:
-Lift up the footrest slightly to discontinue contact between the straps on the footrest and the leg rest.
-Squeeze the footrest to the right to activate the spring mechanism, and release the

3.5 Removable Wings

upper body lateral support to the resident.
To Remove t he Wing:
-Depress the top button (1) and pull the wing from the receptacle. The wing may be moved out of the way without its full removal.
-To completely remove the wing, depress the bottom button (2) and remove from the receptacle.
To Insert the Wing:
- Insert the lower end of the wing into the lower receptacle.. Then insert the upper e nd into the uppe r receptacle. Ensure the wing is secured in place.
Adjustment Increments
12
Page 14

3.6 Shoulder Bolsters

The Shoulder Bolsters provide upper torso support, The armrest height can easily be adjusted in order to
accommodate various residents. The armrest height
2
1
protection and containment and accommodate asymmetries in the chair user. Their unique design allows for proper elbow access to the arm rest. The Shoulder Bolster is always used in c onjunction with the side wing.
To Attach the Shoulder Bolster s:
-Feed both Shoulder Bolster straps between the inner strap of the wing and the fra me of the chair.
-Secure the b uckles and cinch the straps to secure the bolster to the wing. The bolster can be slid up and down the wing if desired.
Note: The Shoulder Bolster can be inverted on the wing in order to provide support to narrower residents.

3.7 Height Adjustable Armrests

should be adjusted to provide correct arm support for the resident as determined by the caregiver.
To Adjust Armrest Height:
-Using a 5/32” Allen wrench, loos en, but do not remove the set screws.
-Raise or lower the armrest to the desired height and tighten the bolts. WARNING: Do not place fingers anywhere underneath the armrest while adjusting as this could cause pinching.
-Secure the o utside flap on the side panel. Change the height of the other arm as desired.
13
Page 15

3.8 Adjustable Seat Height

To adjust the s eat height The Basic chairs feature three 5” total lock casters and
2
1
-Using 7/16 wrenches, remove the bolts.
-Raise or lower the front of the seat to correspond with the holes at the desired seat height.
- Insert the bolts through the desired corresponding holes and ti ghten the bolts.

3.9 Casters

one 5” directional locking (tracking) caster. (dark grey pedal) The tracking feature allows for easy maneuverability.
To Operate the Casters:
Step on/off the caster pedal to lock/unlock the caster.
Always ensure the brakes are unlocked when wheeling the chair.
For chairs with rear mag wheels, lock and unlock using the handbrake.

3.10 Accessories (Options)

We recommend BRODA accessories for use only on BRODA chairs. The correct installation of the BRODA accessory is very important. We recommend that the BRODA chair is ordered with the required accessories already installed. Only author ized caregivers who tho roughly und erstand the accessory installation procedures should attempt to install their BRODA accessories without assistance fro m their lo c a l BRODA representative. BRODA does not recommend use of other manufacturer’s accessories on BRODA chairs.
14
Page 16

4 INSPECTION AND FUNCTIO NAL TESTING

Safety measures as described in Section 1 must be observed when inspecting or testing a chair. Only authorized caregivers or maintenance staff should attempt to inspect or test a chair.

4.1 Inspection

We recommend regular visual inspection for signs of wear, damage, loose or missing fittings, and other safety concerns. If a breakage, defect, or operational problem is detected, the chair must be repaired inspected and tested for function before it is returned to service.
We recommend that the chair be inspected as often as each use if there is any reason to be concerned about the possibility of increased wear or loose or missing fittings. At a minimum, in regular use the chair should be inspected on a bimonthly basis. The visua l inspection procedure should include at least the observation of all the fittings (fasteners):
The visual inspection procedure should include at least the observation of the following parts:
1) The cylinders that position the sea t tilt, back recline, and independent leg rest elevation.
2) The vinyl straps used in the seat, back, leg rest, footrest and wings
3) The cushio ns
4) The black handles and cables.
5) The armrest height adjus tment pins
The visual inspection procedure should include the observation of any installed accessories. The visual inspection should be performed by the facility, or if in a private residence, by the indi vidual r e sponsible for the chair.

4.2 Functional Testing

We recommend that the chair should be tested for operation of the chair’s f unc tions without a resident in the chair. The testing may be as often as each use if there is any reason to be concerned about the possibility of
increased wear or damage to the chair’s components. At a minimum, in regular use the chair should be tested for functions as described in Section 3.
If the careg i ver or mainte nance department performing the functional testing believes that a ny function is not operating correctly, the chair should be taken out of service until a satisfactory functional test can be completed.
The caregiver or maintenance department performing the testing should be aware that the seat tilt, back recline, and footrest (leg rest) elevation operations will be more difficult without a resident in the chair.
When performing repairs or maintenance, do not use lubricants t hat contain so lvents. Solvents will damage man y of the moving components in the chair. If necessary, a white, food grade grease (lubricant) may be used on the sliding component s i n the chair. Do not use spray lubricants on any pa rt of the chair.
15
Page 17

5 TECHNICAL INFORMATION

1. Back Recline:
Up to 90º (infinitely adjustable) Model
2. Seat Tilt:
Up to 25º of seat tilt (infinitely adjustable)
3. Frame:
16 gauge powder coated tubular steel
4. Seat/Back Surface:
5. Arms:
Adjustable armrest height
6. Leg Rest:
Leg rest elevates with back recline
7. Footrest:
Flip down removable/adjustable footrest
8. Wings:
Removable win gs
9. Padding:
Removable standard padding – neckrest, wings, shoulder bolsters, foot pad and lower
10. Casters:
Three 5” total lock casters, one 5” directional locking caster
11. Seat Depth:
17”
12. Seat Height:
Adjustable 19.5” – 21.5”
13. Seat Width:
20” (18” also standard)
14. Overall Width:
29”
15. Overall Weight:
Approx. 90 lbs standard features
All dimensions are ± 0.5”.

5.1 Specifications

Specifications Based on 20” Wide Basic Chair 30VT
BRODA Comfort Tension Seating Strapping is 1.5” wide, by .130” thick; contains fungicide, has shape memory retention, is fire retardant
Leg rest adjusts independently of back recline Leg rest adjusts to any angle between horizontal and vertical
side pads The 34 ounce healthcare vinyl fabric covering is manufactured to meet the following requirements: DIN 75 200/DIN 53 438; MVSS 302, M2;CAL 117, and Wyzenbeck Heavy Duty Abrasion Test; has an antimicrobial, antibacterial, anti-stain and anti-static finish; Cold crack of -20ºC. Polyurethane foam meets CAL 117-2013.
Removable seat and back pads are covered with an 8 ounce soft vinyl which meets CAN2-162 fla mmability standard for hospital textiles. Polyurethane foam meets CAL 117-2013. Removable, washable, pre-shrunk fitted cotton terry covers are provided for the seat and back pads. Terry covers provide absorption, and vapor and heat transfer.
Non-marking grey rubber tires.
16
Page 18

5.2 Labels

Manufacturer’s Label (White)
Serial Number Label (Grey)
TRACKING
For Proper Positi on i ng
BACK RECLINE
MADE IN CANADA
Tilt Label (Tan)
SEAT TILT
SQUEEZE HANDLE
BRODA uses certain labels to assist caregivers to ide ntify items which permit the opera tion of the chair’s functions.
MADE IN CANADA / FABRIQUE AU CANADA
560 Bingemans Centre Drive, Kitchener, ON Canada, N2B 3X9 PHONE 1 – 519 746-8080 FAX 1-519-746-8616
Recline Label (Blue)
SQUEEZE HANDLE
Tilt and Recline Label (White) Model 30VT
STEP 1: Tilt Seat STEP 2: Recl in e B ack
1-800-668-0637
SEATING
PRODUCT # SERIAL #
Footrest Attachment/Detach ment Label (White)
Tracking Caster (Green)
CASTER
17
Page 19

5.3 BRODA Seating Basic 30VT Parts Schematic Drawing

18
Page 20

5.4 Broda Seating 30VT Parts Schematic Legend

Number Item Description Qty 1 BASE-BC Base Frame Basic (width) (color) 1 2 CASTER-5”DL Caster 5 Directional Lock (color) 1 3 CASTER-5”TL Caster 5 Total Lock (color) 3 4 SET-SCREW Screw 5/16-18X3/8 Set Screw, S.S 8 5 CABLE-HAND Cable and Handle Assembly (cable length 48) 3 6 ARM-TOP-BC Arm Top Frame Basic (color) 2 7 AR Armrest Soft 2 8 SEAT-BC Seat Frame Basic (width) (color) 1 9 LGRT Legrest Frame (width) (color) 1 10 & 12 WING-STRAP Wing Strapped Frame Removable (left/right) 2
11 DETENT-PIN Spring for Arm Bottom and Wings 4
13 & 16 F-SINGLE Footrest Frame Single (3, 4 or 5 strap) (width) (color) 1 14 F-SPRING Footrest Spring 3.25 Long 1 15 F-ROD Frame Foot Rod (width) 1 17 CYL-LOCK-KIT Cylinder (CYL-300N) and actuator with lever 3 18 BACK-RCLN Back Frame Recline (width) (color) 1
19
Page 21

6 COMMON TROUBLESHOOTING AND R E PAIR

Symptom
Problem
Solution
The chair is not reclining or tilting or
The pin in the end of the cylinder is
Ensure the cable is attached to the The chair (tilt, recline or leg rest)
The pin in the end of the cylinder
See Adjustment Steps below:
Adjustment Steps:
-Loosen the lock nut located on the stem of the gas
Please have the chair serial number ready when contacting BRODA or your local area representative for parts or assistance for your chair. The serial number can be found on a grey sticker on the rear of the chair.

6.1 Troubleshooting for Gas Cylinders

The use of gas charged springs (cyli nders) on BRO DA products allows the caregiver to easily make adjustments to the tilt, recline, and leg rest with a minimal effort. The gas cylinders contain Nitrogen gas, it is not flammable nor is it toxic. The cylinders provide a lifti ng force which counterbalances the weight of the occupant thereby reducing the amount of weight the caregiver needs to lift.
leg rest is not moving when the handle is squeezed
seems to slowly slide out of positio n
cylinder next to the actuator (17mm or 11/16” wrench)
-From the above instruct i on determine whether you want to wind the stem further into the actuat or or furthe r out of the actuator. Wind the stem further into the
actuator if the chair i s not tilting or reclining, (wind clockwise) and wind it further out of the actuator if the chair is slowly sliding out of position. (wind counter-clockwise)
-Wind the stem in the correct direction one half turn.
-Check the f unction of the cylinder and repeat the above step until the cylinder functions properly.
-Tighten the lock nut loosened in the first step.
-If adjusting the cylinder does not correct the problem, the cylinder may need to be replaced. Please contact BRODA at or your local area representative.
not being depressed when the handle is pulled.
may be still partially depressed.
Note: Cylinder does not need to be removed from the chair to make this adjustment.
20
cylinder actuator. If cable is attached, see Adjustment
Steps below.
Page 22
Note: When adjusting cylinders with tools that come in
contact with the cylinder rod s, a clo th or some other protective material must be used to protect the sliding surface of the rod.If the rod becomes scratched or marred it will damage the seal on the cylinder and the cylinder will fail.
Damaging a cylinder in this fashion voids the manufacturer’s warranty.
21
Page 23

7 WARRANTY

BRODA SEATING
WARRANTY
BRODA provides a Three Year Warranty on the chair frame, and a One Year Warranty on all other component s subject to the follo wing conditions:
No warranty is provided on seat pads or cloth covers. The chair frames are guaranteed for three years against structural defects or failure. All other parts (except
seat pads and cloth covers) including but not limited to strapping, cushions, gas springs and attachments, casters, wheels, brakes and armrests are guaranteed for one year against defects in materials and workmanship based on normal institutional use. The guarantee does not cover malicious or deliberate damage or damage from misuse. The guarantee does not cover use of BRODA chairs in a shower or pool, with the exception of the BRODA Commode/Shower Chair.
Modifications to BRODA products or the use of non-BRODA supplied parts voids the warranty. This warranty does not cover shipping damage (see below).
BRODA will provide new or refurbished parts for installation by the owner at no cost following confirmation by the local BRODA Representative or the BRODA Head Office Customer Service Representative. On request, defective parts must be returned to the factory within thirty days of receipt of the replacement parts by the owner. If the defective parts are not returned to BRODA on request, the owner will bear the cost of the replacement parts on invoice from BRODA.
Warranty does not include on-site labor for the installation of warranty parts or warranty repairs. The owner may return to BRODA products for warranty replacement or repair by shipping items prepaid and insured to the factory. Warranty completed at the factory includes both materials and labor. The decision to repair or replace parts is at the discretion of BRODA. All returns to the factory require prior authorization from BRODA.
BRODA retains the right to make d esign and application changes without notice. All orders will be filled with BRODA's current models unless otherwise specified by the purchaser.
BRODA chairs are designed for patient mobility, positionin g, and comfort in spe cialty seati ng, ho wever , the application of BRODA products shall remain the responsibility of the purchaser or user.
This warranty is not transferable.
RETURNS
BRODA Seating will not accept any returns without a prior Returned Goods Authorization Number. Please contact our Head Office Customer Service Representative at 1-800-668-0637 for assistance. Returns must be insured when shipped.
DAMAGED FREIGHT
NOTIFY THE CARRIER OF ANY DAMAGE IMMEDIATELY
It is the responsibility of the person receiving the goods to examine cartons and goods before accepting receipt. Note all damages on the bill of lading and file a claim if necessary. Notify the carrier of any concealed damaged within 48 hours of receipt.
BRODA insures all products for in transit damage, failure to notify the carrier of in transit damage voids both the insurance and the BRODA warranty. If you require assistance, contact our Head Office at 1-800-668-
0637.
SALES TAX
Most BRODA products are G.S.T. Zero Rated and Exempt from Canadian Provincial Sales Tax. Purchaser s may be required to check with their Provincial or State Tax Office for purchaser tax payment.
22
Loading...