Breas Medical PV 501-2, PV 501 Service Manual

Service Manual PV 501/501-2
Table of contents
1 INTRODUCTION ....................................................................................................... 1
1.1 The scope of this manual ............................................................................ 1
1.2 Intended use of the ventilator ...................................................................... 1
1.4 Intended audience ....................................................................................... 1
1.5 Service personnel's training requirements .................................................. 2
2 OPERATING MANUAL
3 MAINTENANCE SERVICE INSTRUCTIONS ........................................................... 1
3.1 Verifying the components and software installed ........................................ 1
3.2 Maintenance Service Schedule................................................................... 1
3.3 Special safety precautions .......................................................................... 2
3.4 Equipment required ..................................................................................... 3
3.5 Replacement parts required ........................................................................ 3
3.6 Maintenance instructions ............................................................................. 3
3.6.1 Registration ................................................................................. 3
3.6.2 Information from the user ............................................................ 3
3.6.3 Validity of documentation ............................................................ 3
3.7 External checks ........................................................................................... 4
3.7.1 External damage and wear ......................................................... 4
3.7.2 Power cables and plugs.............................................................. 4
3.7.3 Minimum function check ............................................................. 4
3.8 Internal Checks ........................................................................................... 4
3.8.1 Open hood and rear panel .......................................................... 4
3.8.2 Cleaning...................................................................................... 4
3.8.3 Cables and connectors ............................................................... 4
3.8.4 Component fastening.................................................................. 4
3.8.5 Mains supply ............................................................................... 5
3.8.6 Remove and check / replace the motor unit. .............................. 5
3.8.7 Grease the ball screw ................................................................. 5
3.8.8 Replace the membranes in the check valves ............................. 5
3.8.9 Leakage check of hoses and bellows ......................................... 5
3.8.10 Reassemble the motor unit ......................................................... 5
3.8.11 Checking the offset and gain of the pressure transducer ........... 5
3.8.12 Checking the instrument accuracy.............................................. 5
3.8.13 Internal battery test ..................................................................... 5
3.8.14 Reassemble the casing .............................................................. 6
3.9 Electrical safety ........................................................................................... 6
3.10 Final checks before delivery........................................................................ 6
3.10.1 Function check/leakage .............................................................. 6
3.10.2 Check the tidal volume/frequency ............................................... 6
3.10.3 Calculating the tidal volume/patient pressure ............................. 7
3.10.4 Checking the pressure limit ........................................................ 7
3.10.5 Checking low pressure alarm/alarm mute .................................. 8
3.10.6 Trigger ........................................................................................ 8
Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 1 – 0
Service Manual PV 501/501-2
3.10.7 Battery operation ........................................................................ 8
3.10.8 Checking accessories ................................................................. 8
3.10.9 Cleaning the PEEP adapter ........................................................ 9
3.10.10 Set the correct values for the patient ........................................ 10
4 REPLACEMENT PARTS .......................................................................................... 1
4.1 OVERVIEW DIAGRAM ............................................................................... 1
4.2 Exploded drawing, Base Plate Assembly PV 501 ....................................... 2
4.3 Exploded drawing, Base Plate Assembly PV 501-2 .................................... 3
4.4 Exploded drawing, Motor Unit Assembly PV 501........................................ 4
4.5 Exploded drawing, Motor Unit Assembly PV 501-2 ..................................... 5
4.6 Exploded drawing, Rear Panel Assembly PV 501 ...................................... 6
4.7 Exploded drawing, Rear Panel Assembly PV 501-2 ................................... 7
4.8 Exploded drawing, Hood Assembly 501 ...................................................... 8
4.9 Exploded drawing, Hood Assembly 501-2 .................................................. 9
4.10 Part Number List ....................................................................................... 10
5 FUNCTIONAL DIAGRAMS ....................................................................................... 1
5.1 Pneumatic diagram ..................................................................................... 1
5.2 Functional block diagram ............................................................................ 2
6 DISMANTLING AND ASSEMBLING THE PV 501/PV501-2 ..................................... 1
6.1 Opening the hood ........................................................................................ 1
6.2 Replacing pushbutton boards ...................................................................... 2
6.3 Replacing the CPU board ............................................................................ 3
6.4 Replacing the potentiometer board ............................................................. 4
6.5 Opening the rear panel................................................................................ 5
6.6 Replacing the MDA board ........................................................................... 5
6.7 Removing the motor unit ............................................................................. 6
6.8 Reassembling the motor unit ....................................................................... 7
6.9 Assembling the hood and rear panel ........................................................... 7
7 MOTOR UNIT ............................................................................................................ 1
7.1 Construction ................................................................................................ 1
7.2 Lubricating the ball screw ............................................................................ 2
7.3 Replacing the membranes in the check valves ........................................... 3
7.4 Hoses and bellows, leakage check ............................................................. 3
7.5 Wiring diagram, Motor unit .......................................................................... 4
8 ELECTRONICS ......................................................................................................... 1
8.1 Function, Construction ................................................................................ 1
8.2 PV 501/PV 501-2's circuit boards ................................................................ 2
8.3 Pushbutton board ........................................................................................ 3
8.4 Potentiometer board .................................................................................... 3
8.5 MDA board .................................................................................................. 4
8.5.1 Checking the pressure transducer offset and gain ..................... 5
8.5.2 Checking the instrument accuracy .............................................. 5
8.6 CPU board ................................................................................................... 6
8.7 Terminal diagram, test points ....................................................................... 7
8.8 Measuring points on the motor unit ........................................................... 10
1 – 1 BREAS MEDICAL Doc. No.: 1523EN Issue:N-1
Service Manual PV 501/501-2
8.9 Internal Battery test ................................................................................... 11
8.10 Replacing the internal power batteries (lead acid cells) ............................ 11
8.11 Replacing the alarm batteries (Ni-Cd) ....................................................... 12
8.12 Electrical safety ......................................................................................... 13
8.12.1 Mains ........................................................................................ 13
8.12.2 Isolation .................................................................................... 13
8.12.3 Leakage current ........................................................................ 13
8.12.4 Leakage current from the cabinet ............................................. 13
8.12.5 Patient leakage ......................................................................... 13
8.12.6 Leakage current at mains voltage at patient connected part .... 13
8.13 Wiring Diagrams PV 501/501-2 ................................................................. 14
8.14 Functional block diagram PV6 T1 ............................................................. 30
8.15 COMPONENT LOCATION DRAWINGS PV 501 ...................................... 31
8.15.1 Pushbutton Boards PV 501, Versions B & E ............................ 31
8.15.2 Potentiometer Board PV 501 ................................................... 31
8.15.3 CPU Boards PV 501, Version B & E ......................................... 32
8.15.4 MDA Boards PV 501 ................................................................. 33
8.16 COMPONENT LOCATION DRAWINGS PV 501-2 ................................... 35
8.16.1 Pushbutton Boards PV 501-2 ................................................... 35
8.16.2 Potentiometer Board PV 501-2 ................................................. 36
8.16.3 CPU Boards PV 501-2 .............................................................. 37
8.16.4 MDA Boards PV 501-2 XA ........................................................ 38
8.17 List of components PV 501 ....................................................................... 39
8.18 List of components PV 501-2, Rev. XD ..................................................... 44
9 FAULT TRACING...................................................................................................... 1
9.1 Fault tracing guide ....................................................................................... 1
9.2 Error codes .................................................................................................. 3
9.3 Recalling error codes .................................................................................. 3
9.4 Erasing the error code memory ................................................................... 3
9.5 ERROR CODE TABLE................................................................................ 4
9.6 Fault tracing using the error codes .............................................................. 5
10 ENGINEERING CHANGE HISTORY PV501 AND PV501-2 .................................... 1
10.1 Service record for BREAS PV501/PV501-2
Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 1 – 2
Service Manual PV 501/501-2 INTRODUCTION

1 INTRODUCTION

1.1 The scope of this manual

This manual provides information for the maintenance and service of both the PV 501 and PV 501-2 ventilators. The PV 501-2 is a further development of the PV 501. The main differences are that the PV 501-2 delivers a maximum patient pressure of 60 mbar (compared to the PV 501’s 50 mbar) and that theTrigger setting is adjustable between -
-2 to 6 mbar (compared to the PV 501’s -4 to 6 mbar).
The PV 501and PV 501-2 are designed to give many years of trouble-free breathing assistance to the user provided that preventative maintenance is done at the specified intervals described in this manual. Correctly performed maintenance will increase the ventilator's service life considerably.
All points to be checked and service instructions for the PV 501 and the PV 501-2 are described in this manual. Where information is only relevant to one of the models it is clearly marked which model the text/figure refers to, otherwise all information is valid for both models.
It is also important that any peripheral equipment is checked at the same time as the maintenance service is done.
Also included is a reference copy of the Operation Manual.

1.2 Intended use of the ventilator

PV 501 and the PV 501-2 are volume-controlled, pressure-limited ventilators, specially designed for long term ventilator assistance at home.
The internal batteries are connected automatically should the mains supply fail or is dis­connected. Fully-charged batteries will give a running time of 4 hours.
Thanks to its design, the PV 501 and the PV 501-2 are easy to operate. It does not require any gas, and together with its low weight, it is ideal for breathing therapy both in hospital and while travelling.
The PV 501and the PV 501-2 are not intended for intensive care and are aimed for adults and children weighing about 30 kg or more.

1.3 Design and function of the ventilator

The PV 501and the PV 501-2 are built around a bellows that is compressed and drawn out by a ball screw driven by an electronically controlled servo motor. The micro-proces­sor controlled electronics calculate the correct speed and running time for the motor by reading the settings for tidal volume, frequency and the I/E-relationship. The setting lim­its for pressures and the trigger level are monitored. If the mains supply or the external battery supply should fail during operation, the internal batteries are automatically con­nected and an indication of this is given on the front panel. If the battery supply voltage drops too low, both audible and visual alarms are given. The ventilator's modular design makes maintenance easy.

1.4 Intended audience

This Service Manual is intended for technicians who have medical/technical training and knowledge about the construction and function of the ventilator.
It is not intended for clinic personnel or patients.
Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 1 – 1
INTRODUCTION Service Manual PV 501/501-2
Breas Medical reserves the right to make changes to the product and the contents of this manual without prior notice.

1.5 Service personnel's training requirements

Thanks to their simple construction using a modular system, no special competence is required other than general medical technical training on ventilators.
Always contact BREAS MEDICAL if there are any questions or if training is required.
All service must, however, be performed according to the instructions in this manual.
1 – 2 BREAS MEDICAL Doc. No.: 1523EN Issue:N-1
Service Manual PV 501/501-2 MAINTENANCE SERVICE INSTRUCTIONS

3 MAINTENANCE SERVICE INSTRUCTIONS

All routine maintenance checks and additional se rvice instructions for the PV 501and PV 501-2 are described in this chapter. For information about fault-trac­ing, detailed drawings, board schematics, spare parts etc, please refer to the respective chapters in this Service Manual.
The checks described in the Patient Instruction delivered with the ventilator should be followed by the patient and/or care provi ders.

3.1 Verifying the components and software installed

Check the Engineering Change History document in the Appendices section (Chapter 10) for a history of all the changes made and at which serial No they were introduced.
If in any doubt, read the component designation on the motor unit, circuit boards and PROMs as upgrades can have been made but not recorded.

3.2 Maintenance Service Schedule

IMPORTANT!
A complete maintenance service (as described in this chapter) must be done
every 12th month. If the ventilator is used for continuous operation (24 hours
per day) a complete maintenance service must be done every 6th month.
Interval Action
At every 500th operation hour or more frequent if necessary (especially in town environments).
Every 6 months if the ventilator is used for continu­ous operation i.e. 24 hours per day, otherwise every 12th month.
Additional service action (when required)
At 20.000 running hours Replace motor unit and hoses. The internal batteries should be replaced every 24
months.
Replace the alarm batteries every 5th year counted from date of delivery or when required.
Replace the patient air inlet filter. Done by the patient/care provider. Replace during maintenance service.
Complete maintenance service accord­ing to this instruction.
Replace the internal batteries Replace the alarm batteries.
Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 3 – 1
MAINTENANCE SERVICE INSTRUCTIONS Service Manual PV 501/501-2
Every 12th month or every 6th month if the ventilator is used
for continuous operation (24 hours per day)
Replace the patient air inlet filter. 2 Clean the ventilator air intake filter 2
Motor Unit
Lubricate the ball screw 7.2 Replace the check valve membranes in bellows end cover 7.3 Leakage test of motor unit and tubes 7.4
Electronics
Checking the pressure sensor offset and gain. 8.5.1 Check operation using internal battery 8.9 Check operation using external battery 2 Check electrical safety levels 8.13
Internal Batteries (lead-ac id)
The internal batteries should be replaced every 24 months or when neces­sary.
Section No.
8.10
See Ch./
Alarm Batteries
Replace every 5th year counted from date of delivery or when last replaced.
Accessories (where applicable)
Inspect patient circuit 2 Replace membrane in exhalation valve 2 Clean PEEP adapter, replace O-ring 2
Every 20000 operation hours
Replace the complete Motor unit 6.6

3.3 Special safety precautions

Avoid working with the mains c onnect ed when the casi ng is r emoved. Always tes t run using internal battery power.
Insulate the wires for the inter nal batte ri es when they are di sconne cted to prev ent short circuits.
Explosive gases or liquids must not be kept or used close to the ventilator.
8.11
Always follow good ESD practices when working with the ventilator.
3 – 2 BREAS MEDICAL Doc. No.: 1523EN Issue: N-1
Service Manual PV 501/501-2 MAINTENANCE SERVICE INSTRUCTIONS
3.4 Equipment required
Test lung (1L) with exhalation valve.
Measuring equipment for tidal vol ume and minut e volume/frequency. (Timeter, Spirometeor or equivalent).
Voltmeter.
3.5 Replacement parts required
Have the following spare parts at hand:
Part No.
000 139 Internal battery (lead) 000 036 Alarm battery (Ni-Cd) 001 141 Motor unit, BEI 000 557 Grease (BREAS 283 AZ) 000 004 Motor unit, Maxon 000 248 Filter, patient air 002 123 Service kit, check valves 002 178 Membrane kit, exhalation valve
Description
3.6 Maintenance instructions
3.6.1 Registration
Fill out the registration part of the Service Record. Check that the following mark­ings can be read:
Make, model designation and serial number.
All warning texts on labels.
Any inventory control markings.
Document the current patient settings.
Note the number of running hours. Does the motor unit need to be replaced?
Check any comments or events made on the previous service record.

3.6.2 Information from the user

Before starting the service request the following information from the patient:
Has the ventilator functioned problem-free, such as running failure?
How does the patient check the function of the ventilator? How often?
How often is the filter changed?
What is the patient's need of filters unt il the next service point?

3.6.3 Validity of documentation

Check that the patient instructions are up-to-date.
Check if any modification or update of the ventilator is to be performed during the service.
Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 3 – 3
MAINTENANCE SERVICE INSTRUCTIONS Service Manual PV 501/501-2

3.7 External checks

3.7.1 External damage and wear

Clean the surface with a mild detergent.
Clean the filter for the machine ventila tor air inlet. Use a vacuum cleaner or wash and dry the filter before fitting.
Check that there is no visible damage on the surfaces and other components.
Check that all texts for controls etc. are readable.
Turn all knobs and check that their operation feels OK.
Check that nothing is loose, (including the handles).
Check that the cover for the adjustment panel is secu red properly.

3.7.2 Power cables and plugs

Check the power cable and its plugs, and the power socket in the rear panel.
Check that the cable securing clamp is undamaged.
Check other external battery cables, where applicable.

3.7.3 Minimum function check

Connect the ventilator to the mains and check that the LED for Mains/Charging is lit.
Start the ventilator and check that it oper ates normally and no abnormal noise is heard. Check that the lighting for the press u re gauge is on.
Check the condition of the internal batteries. They should, when fully charged, be able to run the PV 501 for at least 2 hours. If not they must be replaced.

3.8 Internal Checks

3.8.1 Open hood and rear panel

Remove the mains power cable.
Open the hood and rear panel. Remove stabiliser bar, see chapter 6.

3.8.2 Cleaning

Remove any dirt or dust that has collected in the ventil ator, also from the folds in the bellows.

3.8.3 Cables and connectors

Check all cables and thei r connectors for an y damage. Check at the fr ont and rear panels where cables can be pinched.

3.8.4 Component fastening

Check that the batteries are properly fastened.
Check that the motor unit is properly fastened.
Check that all parts on the front and rear panels are properly fastened.
3 – 4 BREAS MEDICAL Doc. No.: 1523EN Issue: N-1
Service Manual PV 501/501-2 MAINTENANCE SERVICE INSTRUCTIONS

3.8.5 Mains supply

Check that the power inlet socket is undamaged and that it is properly fastened.
Check that the touch protection cover is undamaged and that it is properly fas­tened over the power inlet socket.
Check that the transformer is properly fastened.
Check the wiring to the transformer.
3.8.6 Remove and check / replace the motor unit.
The motor unit must be checked at each service. At 20.000 running hours the entire motor unit assembly must be replaced. See Chapter 6 for removing the motor unit.
Checking the motor unit
Check that the motor is properly fastened.
Check that the shaft coupling screws are properly tightened.
Check that the encoders are undamaged and properly fastened.
3.8.7 Grease the ball screw
Use only BREAS grease type 283 AZ. See chapter 7.

3.8.8 Replace the membranes in the check valves

The membranes must be changed at each maintenance service. See chapter 7.

3.8.9 Leakage check of hoses and bellows

Check the function of the outlet check valves. See chapter 7.

3.8.10 Reassemble the motor unit

Before reassembling the motor unit, check if the internal batteries need to be replaced as this is best done with the motor unit removed, see 3.8.6.
Reassemble the motor unit. See Chapter 6.

3.8.11 Checking the offset and gain of the pressure transducer

See Chapter 8, Section 8.5.1 for detail ed inf o rmation.

3.8.12 Checking the instrument accuracy

Check using a manometer conne cted to the exh. valve connec tion t hat the i ndic at­ing instrument gives the correct indicati on and that the needle moves "softly". See Chapter 8, Section 8.5.1.1.
NOTE! Make small pressure changes, the indicating instrument should be "slower" compared with the manometer.

3.8.13 Internal battery test

The internal batteries should be fully charged or charged for at l east 12 hours with the ventilator turned off before this test is carried out. The batteries, when fully charged, should operate the PV 501 for at least 2 hours, if not t he batteries must be replaced. See Chapter 8, Section 10.
Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 3 – 5
MAINTENANCE SERVICE INSTRUCTIONS Service Manual PV 501/501-2
3.8.14 Reassemble the casing
See Chapter 6 for information.

3.9 Electrical safety

Follow the instructions for the following checks in Chapter 8.
Check mains supply value
Check isolation against the value obtained at the delivery inspection
Check leakage current. Do not forget to reverse the polarity.
Check patient leakage current.
Leakage current at mains voltage at patient connected part

3.10 Final checks before delivery

3.10.1 Function check/leakage

Start the ventilator with the pat ient circuit and reservoir bag connected as shown in the figure below. Check that everything seems normal. Create a pressure of approximately 30 cm H
0 and listen for any leakage.
2

3.10.2 Check the tidal volume/frequency

Set: Tidal vol ume: 1L
Frequency: 12 I/E: 1:2
Measure the tidal volume and that the volume per minute is cor rect. (Accuracy ± 10%).
When checking using a volume monitor, an exhalation valve with a peep valve connector can be required. The volume monitor is then connected to the peep valve outlet on the exhalation valve.
LOW PRESSURE
HIGH PRESSURE
LOW BATTERY
FUNCTION FAIL URE
TIDAL VOLUME
Litre
Setting error
Setting error Setting errorSetting error
PATIENT PRESSUREALARMS
Peep valve connector
EXHALATION
VALVE
BPM
BPMcm HO BPMBPM
POWER
PATIEN T
AIR
ON
OFF
LOW PRESSURE
cm HO cm HOcm HO
POWER
AC/CHARGING
EXTERNAL BATTERY
INTERNAL BATTERY
TRIGGER
Off
Off
OffOff
TRIGGER
cm HO
2
HIGH PRESSURE
2222
Signal
Signal SignalSignal
I/E BREATH RATE
cm HO
cm HO
Ratio
Ratio
cm HOcm HO
RatioRatio
2222
mbar
Volume monitor
Test lung / Reservoir bag
Fig. 3-1 Tidal volume check
3 – 6 BREAS MEDICAL Doc. No.: 1523EN Issue: N-1
Service Manual PV 501/501-2 MAINTENANCE SERVICE INSTRUCTIONS

3.10.3 Calculating the tidal volume/patient pressure

The tidal volume delivered is always slightly less than the tidal volume setting. This difference is due to the effect of the patient pressure on the compliance of the bellows.
The diagram below shows these differences at different patient pressures. Settings: I/E = 1:2, Breath rate = 15 breaths/min. with the patient circuit con-
nected.
Set volume/l itr e s
Fig. 3-2 Volume/Patient pressure diagramme
Example: A patient needs a tidal volume of 0,7 litres and has an airway pressure of 20 cm
O. Start from 0.7 on the axis for “ Delivered volume”. Draw a horizont al line unti l
H
2
it crosses the line for 20 cm H
O patient pressure. From this point, go straight
2
down and read off the value the tidal vol ume setti ng should be. In this case the ventilator setting is 0.8 litres.
Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 3 – 7
MAINTENANCE SERVICE INSTRUCTIONS Service Manual PV 501/501-2
1

3.10.4 Checking the pressure limit

Set: Low pressure 10 cmH2O
High pressure 40 cmH
Check the accuracy (±10%). Check that LED indicati on is given as well as an audible alarm.
Also check that the solenoid valve for high pressure "clicks". You can clearly hear the solenoid valve activate at the following inhalation.
O
2

3.10.5 Checking low pressure alarm/alarm mute

Turn the vent ilator on without anything connected to the patient air outlet. Wait 15 sec until the alarm for "Low pressure" is activated. Press the alarm mute button and check that the alarm is heard again after approx. 2 min.

3.10.6 Trigger

Set the trigger knob to -1 c m H20. Create a negative pressure an d check that ther e is a triggered inhalation. The green LED should light.

3.10.7 Battery operation

Disconnect the mains cable while the ventilat or is running. Check that the ventila­tor automatically changes over to internal battery and an alarm is given.
Reconnect the mains power and check that indication "MAINS" comes on.

3.10.8 Checking accessories Patient circuit

Inspect the patient circuit and replace it if necessary.
Replacing the membrane assembly in the exhalation valve
This instruction applies to BREAS exhalation valves, see figure below.
Fig. 3-3 Mounting the exhalation valve
Remove the PEEP adapter , if installed. See "Cleaning the PEEP adapter".
3 – 8 BREAS MEDICAL Doc. No.: 1523EN Issue: N-1
Service Manual PV 501/501-2 MAINTENANCE SERVICE INSTRUCTIONS
Unscrew and remove the complete membrane assembly (1).
Clean the inside of the exhalation valve using a moist rag or wash in hot water using a washing-up liquid. Let it air dry.
Screw on the new membrane assembly
If a PEEP valve is to be used, fit the O-ring seal for the PEEP valve as shown in fig. 3-5.
Connect the exhalation valve to a test lung. Check that no leakage occurs during the exhalation phase.

3.10.9 Cleaning the PEEP adapter

1
2
3
Fig. 3-4 Mounting the exhalation valve with PEEP adapter
Remove the plastic nut (1) holding the PEEP adapter.
Pull the adapter (2) up from the exhalation valve.
Clean using a moist rag or wash in hot water using a washing-up l iqui d. Let the adapter air dry.
Check that the O-ring seal (3) is properly in place on the exhalation valve (under the edge of the membrane assembly cover as shown in the figure below).
Note! Do not fit the O-r ing to the membrane assembly before screwing it on.
Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 3 – 9
MAINTENANCE SERVICE INSTRUCTIONS Service Manual PV 501/501-2
Fig. 3-5 Mounting the O-ring for the PEEP adapter
Fit the PEEP adapter to the exhalation valve and screw on the plastic nut
Check any other accessories.

3.10.10Set the correct values for the patient

Adjust the settings to the prescribed values for the patient.
Change the air filter. Check that the patient has enough filters to last until the next service.
The patient settings should be checked by a physician at least once a year.
3 – 10 BREAS MEDICAL Doc. No.: 1523EN Issue: N-1
Service Manual PV 501/501-2 REPLACEMENT PARTS

4 REPLACEMENT PARTS

4.1 OV ERVIEW DIAGRAM

The illustration belo w shows the PV501/PV501-2, seen f rom above, with the ho od and rear panel unscrewed and laid flat against the work surface.
The figure below shows the main units making up the PV501/PV501-2. The fol­lowing pages contain detailed exploded drawings for each of the main units and model. Last in this chapter is a complete list of part numbers.
1
Pos. No.
1 Hood assembly 2 Motor unit 3 Base plate assembly 4 Rear panel assembly
Component
2
3
4
Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 4 – 1
REPLACEMENT PARTS Service Manual PV 501/501-2

4.2 Exploded drawing, Base Plate Assembly PV 501

4 – 2 BREAS MEDICAL Doc. No.: 1523EN Issue: N-1
Service Manual PV 501/501-2 REPLACEMENT PARTS

4.3 Exploded drawing, Base Plate Assembly PV 501-2

Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 4 – 3
REPLACEMENT PARTS Service Manual PV 501/501-2

4.4 Exploded drawing, Motor Unit Assembly PV 501

4 – 4 BREAS MEDICAL Doc. No.: 1523EN Issue: N-1
Service Manual PV 501/501-2 REPLACEMENT PARTS

4.5 Exploded drawing, Motor Unit Assembly PV 501-2

Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 4 – 5
REPLACEMENT PARTS Service Manual PV 501/501-2

4.6 Exploded drawing, Rear Panel Assembly PV 501

4 – 6 BREAS MEDICAL Doc. No.: 1523EN Issue: N-1
Service Manual PV 501/501-2 REPLACEMENT PARTS

4.7 Exploded drawing, Rear Panel Assembly PV 501-2

Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 4 – 7
REPLACEMENT PARTS Service Manual PV 501/501-2

4.8 Exploded drawing, Hood Assembly 501

∂ρ#ελλερ
4 – 8 BREAS MEDICAL Doc. No.: 1523EN Issue: N-1
Service Manual PV 501/501-2 REPLACEMENT PARTS

4.9 Exploded drawing, Hood Assembly 501-2

Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 4 – 9
REPLACEMENT PARTS Service Manual PV 501/501-2

4.10 Part Number List

Part No. DescriptionNotes
2 Base plate, complete 4 Motor unit, Maxon - complete 5 Hood, complete, SE 6 Complete, Eng 7 Complete, Ger 8 Rear panel, complete, Eng 9 Rear panel, complete, Ger 10 Hood, complete, Eng 11 Hood, complete, Ger 12 Rear panel, blank 13 Potentiometer box 14 Magnet - potentiometer box 15 Air-hose connection - patient exhale 17 Side unit - motor bracket 18 Motor plate 19 Rear panel - Swedish text 20 Hood 21 Bottom plate 22 Angle bracket - front / hood 24 Cooler angle element 25 Angle bracket - reading prong (l imit position) 26 Ball screw - complete 27 Motor bracket 29 Panel - potentiometer panel Swedish text 30 Panel - patient air Swedish text 31 Rubber mount, shock absorbing (opto-switch) 32 Pressure sensor 33 Circuit board - CPU, PV501 34 Circuit board- MDA, PV501 35 Circuit board- potentiometer board 36 Battery Ni/Cd or Ni/Mh alarm back-up 44 Decal - "PATIENT PRESSURE" 45 Decal - "ON / OFF" 50 Cylomer - G12 51 Cable marking - plus 52 Cable marking - minus 53 Transf ormer - 115V 54 Opening - front 55 Cross bar 56 Spacing- motor bracket 57 Software
4 – 10 BREAS MEDICAL Doc. No.: 1523EN Issue: N-1
Service Manual PV 501/501-2 REPLACEMENT PARTS
58 Battery - large 24 V lead 64 Hour counter - 24V 65 Flexible tube - inlet (2.5") 66 WAGO - 6-pole angled 67 WAGO - 2-pole angled 68 Reference pressure - valve 70 Reference pressure - reducer sleeve 71 Mains inlet 72 Touch protection 73 Air-filter box 74 Air-hose connection - inhale 75 Chassis pinning device XLR 76 Fuse holder 77 Screen guard - 60x60 78 Filter holder 79 Contact housing - black 80 Contact housing - red 82 Marking point - black 83 Marking point - red 84 Motor 85 Motor connection 86 Angle bracket- reading prong (column plate) 87 Air-hose connection - drive pack 88 Bellows side unit - fixed 89 Angle element - limit position (plastic) 90 Non-return valve 91 Opto switch - limit position 92 Opto switch - column plate 93 Column plate 94 Bellows 95 WAGO - 8-pole angled 96 Nipple - LCN-M5-PK3 97 O-ring - air-hose connection drive pack 98 Bellows side unit - moveable 99 Nipple - SCN-M10-PK4 100 Guide - large 101 Guide - small 102 Push-button board - on / off 103 Push-button board - battery test 104 Scale - graduated 105 O-ring - air-hose connection inlet 106 Panel - potentiometer panel English text 107 Ribbon cable 108 Instrument
Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 4 – 11
REPLACEMENT PARTS Service Manual PV 501/501-2
109 Magnetic valve 110 Cyl, isolator 111 Decal - "ON / OFF" 112 Decal - "PATIENT PRESSURE" 113 Flexible tube - out (5") 114 Angle pipe - out 115 Pivot 116 Arrow - large guide 117 Arrow - small guide 118 Cover - large guide 119 Cover - small guide 121 Nipple - CN-M5-PK3 122 Rear panel - English text 123 Insulation - side unit left 124 Insulation - side unit right 125 Insulation - bottom 126 Insulation - hood 127 Panel - patient air English text 128 Mains cable 130 Packing - bellows side unit 134 Decal - "EIN / AUS" 135 Decal - "BEATMUNGSDRÜCK" 136 Rear panel - German text 137 Panel - potentiometer panel German text 138 Panel - patient air German text 139 Battery pack, complete 140 Text plate (IT) 141 Panel, pot. panel (IT) 142 Panel, patient air (IT) 143 Text plate (SP) 144 Panel, pot. panel (SP) 145 Panel, patient air (SP) 146 End cover for motor console, new version 147 Motor console, new version 148 Motor coupling, Rotex 149 Bellows end cover fixed, new version 150 Bellows end cover moving, new version 156 Spacer sleeve, moving end cover 157 Opto-switch, Optek 158 Seal 160 Shrink tubing - end position 184 115/230 switch, PV501-2 215 Rubber foot 220 Hook - WAGO-terminal block
4 – 12 BREAS MEDICAL Doc. No.: 1523EN Issue: N-1
Service Manual PV 501/501-2 REPLACEMENT PARTS
248 Filter, air (5 pcs/packet) 287 Patient circuit (L-C) 378 Motor (BEI) 381 Decal "Pression de service" Fr 382 Decal "Reglage/Arret" Fr 383 Panel, pot. panel (Fr) 384 Panel, patient air (Fr) 385 Text plate (Fr) 386 Decal "PATIENTTRYK" DK 387 Decal "TIL/FRA" DK 388 Panel, pot. panel (Dk) 389 Text plate (DK) 390 Decal "PATIENT DRUK" NL 391 Decal "TIL/FRA" NL 392 Panel, pot. panel (NL) 393 Panel, patient air (NL) 394 Text plate (NL) 395 Bearing sleeve (PUR drive screw) 396 Bearing sleeve for bearing (drive screw) 397 Flange (PUR drive package)' 398 Flange (drive package) 458 Membrane, check valve 515 Serial No. decal with EAN code 516 Filter for machine ventilation 600 Tape - potentiometer box 601 Cellular rubber border 602 Air-hose 4 x 6 - frosted 604 Fuse TT 315mA 605 Fuse - T160mA 606 Fuse - F3,15A 611 M3 x 8 with washer 612 M3 - tooth nut 613 Heat conducting paste - HTC10S 614 Shrink tubing - 6.4 mm 615 Shrink tubing- 3.2 mm 616 Shrink tubing - 2.4 mm 617 Shrink tubing - 12.7 mm 618 Shrink tubing- 1.6 mm 632 Clips for ribbon cable 633 Strain relief - 4 mm 634 Strain relief - 3.5 mm 635 Cable tie brace 636 Cable tie- 8" 637 Cable tie- 4"
Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 4 – 13
REPLACEMENT PARTS Service Manual PV 501/501-2
638 Cable tie- 14" 639 5 mm plate spacing 640 Y-connection 641 Blue Festo tubing 642 Mains cable straps 643 Black screw 201 - RXS 3.5 x 19 644 Clear tubing 9 x 12 645 Clear tubing 4 x 7 646 Clear tubing 3 x 6 647 Clear tubing 3 x 5 648 SPAX 4 x 20 649 SPAX 3 x 12 650 SGA 22 651 Seeger - AK-26 , or Sgh 32 652 M6 x 25 - bolt 653 M6 x 22 - washer 654 M6 - flat washer 655 M6 - locking nut 656 M5 x 80 F 657 M5 x 50 F 658 M5 x 20 socket head cap screw 659 M5 x 16 F 660 M5 x 16 661 M5 - toothed plate connector 662 M5 - flat washer 663 M5 - nut 664 M5 - locking nut 665 M4 x 8 F 666 M4 x 8 667 M4 x 45 668 M4 x 30 669 M4 x 20 670 M4 x 16 F 671 M4 x 16 672 M4 x 12 673 M4 x 10 F 674 M4 - wing nut 675 M4 - toothed plate connector 676 M4 - flat washer 677 M4 - nut 678 M4 - locking nut 679 M3 x 8 680 M3 x 6 F 681 M3 x 6
4 – 14 BREAS MEDICAL Doc. No.: 1523EN Issue: N-1
Service Manual PV 501/501-2 REPLACEMENT PARTS
682 M3 x 16 F 683 M3 x 12 black socket head cap screw 684 M3 x 12 685 M3 x 10 F 686 M3 x 10 687 M3 - toothed plate connector 688 M3 - flat plate 689 M3 - nut 690 M3 - threaded bar 691 M2 x 10 692 M10 - toothed plate connector 693 KFXS 2.9 x 9.5 F 694 Fibre washer 699 Hook - for WAGO-terminal block 700 Packing - battery base 701 Plastic spacing - 4 x 15 702 Plastic spacing - 4 x 10 703 Plastic spacing - 3 x 10 704 Blind plug - angle bracket hole 705 Double-sided tape 706 Mains cable discharger 708 O-ring - potentiometer board 709 Fuse - 1.25 A 711 Cable tie - mini 713 M4 x 6 socket head cap screw 714 IC - MDA 716 Pressure sensor 723 Magnetic strip (0.01 m) 727 Cable tie, driller 740 Screw MFZ steel M4*10 Black, chrome Poz 741 M3 x 6 F - black, PV501-2 753 O-ring, check valve 795 Metal cable tie 796 Washer, support washer (drive unit) 797 Sliding bearing sleeve (drive unit) 798 Sliding bearing sleeve (PUR drive unit) 799 Spacer sleeve (drive unit) 800 Seal (drive unit) 1141 Drive unit complete (BEI) 1145 Rubber washer for transformer 1146 Metal washer for transformer 1410 Transformer, PV501-2 1636 Opening front, PV501-2 1720 Lock/keys, PV501-2
Doc. No.: 1523EN Issue: N-1 BREAS MEDICAL 4 – 15
Loading...
+ 77 hidden pages