Any trademarks mentioned in this Instruction Leaflet are the protected property of their rightful
owners.
All rights reserved, including the right to translate the document.
The content of the Instruction Leaflet is the intellectual property of Brain Products GmbH. No part
of the Instruction Leaflet may be reproduced or distributed in any form (by printing, photocopying or any other method) without the express written permission of Brain Products GmbH.
List of figures ................................................................................................................................................. 4
1.1 Intended use .................................................................................................................................... 5
1.2 Product identification information ................................................................................................... 6
1.3 Explanation of the markings on the devices ...................................................................................... 7
Section 2 Notes on the safe use of the adapters .................................................................................................... 12
Section 3 Technical data ........................................................................................................................................... 13
3.1 MOVE2actiCAP Adapter and MOVE2BCIPack Adapter ...................................................................... 13
MOVE2X Adapter | Instruction Leaflet | Version 003 | April, 2 2012
Page 4
List of figures
Section 1 Overview
1-1 MOVE2actiCAP Adapter 8
1-2 MOVE2BCIPack Adapter 9
1-3 MOVE2actiCHamp Adapter 10
Section 6 Block diagrams
6-1 MOVE2actiCAP Adapter: connection between MOVE transmitter and actiCAP system 20
6-2 Connecting two MOVE2actiCAP Adapters for recording 128 channels from one subject 21
4
6-3 MOVE2BCIPack Adapter: connection between MOVE transmitter and BCI active electrodes (on test
subject side) 22
6-4 MOVE2V-Amp Adapter (E105b): connection between MOVE receiver and V-Amp system (on amplifier
side) 22
6-5 MOVE2actiCHamp Adapter: connection between MOVE receiver and actiCHamp system 23
6-6 Connecting two MOVE2actiCHamp Adapters for recording 128 channels 24
6-7 Trigger cable connection for a 128-channel MOVE2actiCHamp system 25
6-8 Trigger cable connection for a 160-channel MOVE2actiCHamp system 26
6-9 Connection between MOVE receiver and QuickAmp system 27
6-10 Connection between MOVE receiver and BrainAmp system 28
MOVE2X Adapter | Instruction Leaflet | Version 003 | April, 2 2012
Page 5
Section 1Overview
1.1Intended use
The MOVE2actiCAP Adapter is used to connect the active electrodes of the actiCAP active electrode system to the MOVE wireless transmitter.
The MOVE2BCIPack Adapter is used to connect the active electrodes of the BCI package for the
V-Amp to the MOVE wireless transmitter.
The MOVE2actiCHamp Adapter is used to connect the MOVE receiver to the actiCHamp amplifier. This allows the wireless acquisition of EEG signals using active electrodes in combination
with the actiCHamp.
5
The MOVE2actiCAP, MOVE2BCIPack and MOVE2actiCHamp adapters are intended exclusively
for research purposes. The devices must not be used for therapy, diagnosis, the monitoring of
vital bodily functions or any other diagnostic or therapeutic purposes.
The MOVE2actiCAP Adapter and the MOVE2BCIPack Adapter are applied parts and may be
used in the patient environment (test subject environment).
It is the responsibility of the user to ensure correct use of the actiCHamp, BrainAmp, V-Amp,
QuickAmp and/or MOVE systems.
The devices must be used in the application environment defined by the manufacturer by appropriate specialist personnel or personnel suitable for the utilization of the devices in accordance with the safety regulations and instructions for use set out in the associated Operating
Instructions.
MOVE2X Adapter | Instruction Leaflet | Version 003 | April, 2 2012
GMDN collective term:CT497, Amplifiers and associated devices
Accessories:Battery charger
Rechargeable batteries
Active reference electrode
Active ground electrode
Electrode splitter cable for actiCAP
Electrode splitter cable for V-Amp
Ground Y-cable for MOVE2actiCAP
Ground Y-cable for MOVE2actiCHamp
Reference Y-cable for MOVE2actiCAP
BrainAmp connection cable
actiCHamp connection cable
actiCHamp ground cable
MOVE 64-channel actiCHamp trigger cable
MOVE 128-channel actiCHamp trigger cable
MOVE 160-channel actiCHamp trigger cable
actiCHamp2MOVE trigger cable
MOVE2V-Amp Adapter (E105b)
MOVE velcro pad
Ear clips
Page 7
1.3Explanation of the markings on the devices
The following labels are affixed to the devices:
Attention. Carefully read the Instruction Leaflet. Consult also the Operating Instructions of the EEG amplifier.
This label indicates that defective devices must not be disposed of with household waste. Dispose of in accordance with national regulations or return the device and its accessories to the manufacturer.
Overview 7
This label indicates that the device is not suitable for MR.
The CE mark confirms that the device meets the requirements of the EMC Directive.
The CE mark should not be confused with the CE mark for medical devices which
is applied to medical products.
Page 8
1.4Packaging
1.4.1MOVE2actiCAP kit
The MOVE2actiCAP kit is supplied with the following components:
X One MOVE2actiCAP Adapter (see Figure 1-1)
X One Charger Set Box including two NiMH rechargeable batteries
X Two BrainAmp connection cables 30 cm
X One MOVE velcro pad
Overview 8
X One active reference electrode
X One pair of ear clips
X Instruction Leaflet
Figure 1-1. MOVE2actiCAP Adapter
Page 9
1.4.2MOVE2BCIPack kit
The MOVE2BCIPack kit is supplied with the following components:
X One MOVE2BCIPack Adapter (see Figure 1-2)
X One Charger Set Box including two NiMH rechargeable batteries
X One BrainAmp connection cable 30 cm
X One MOVE velcro pad
X One Move2V-Amp Adapter (E105b)
X One Y-trigger cable for V-Amp
Overview 9
X Instruction Leaflet
Figure 1-2. MOVE2BCIPack Adapter
Page 10
1.4.3MOVE2actiCHamp kit
The MOVE2actiCHamp kit is supplied with the following components:
X One MOVE2actiCAP kit (see Section 1.4.1 on page 8)
X One MOVE2actiCHamp Adapter (see Figure 1-3)
X Two BrainAmp connection cables 30 cm
X Two actiCHamp connection cables 20 cm 40-pin
X One actiCHamp ground cable 20 cm 4-pin
X One MOVE2actiCHamp trigger cable 3 m 15-pin
Overview 10
X Instruction Leaflet
Figure 1-3. MOVE2actiCHamp Adapter
Page 11
1.5Ordering codes
1.5.1Kits
MOVE2actiCAP kit:BP-04241-actiCAP
MOVE2BCIPack kit:BP-204-0004
MOVE2actiCHamp kit:BP-04241-CHAMP
Overview 11
1.5.2Components
MOVE2actiCAP Adapter:BP-04246
MOVE2BCIPack Adapter:BP-405-0001
MOVE2actiCHamp Adapter:BP-04247
MOVE2V-Amp Adapter (E105b):BP-04241-FA16
Charger Set Box including two NiMH rechargeable batteries:BP-1982-CH
BrainAmp connection cable 30 cm 50-pin:BP-02400-NN
actiCHamp connection cable 20 cm 40-pin:BP-09300
actiCHamp ground cable 20 cm 4-pin:BP-09301
MOVE 64-channel actiCHamp trigger cable 3 m 15-pin:BP-204-0003
MOVE 128-channel actiCHamp trigger cable 3 m 15-pin:BP-204-0002
MOVE 160-channel actiCHamp trigger cable 3 m 15-pin:BP-204-0001
Ground Y-cable for MOVE2actiCAP side 128-channel system:BP-09304
Ground Y-cable for MOVE2actiCHamp side 128-channel system:BP-09305
Reference Y-cable for MOVE2actiCAP side 128-channel system:BP-09306
Electrode splitter cable for actiCAP:BP-4242-32
Electrode splitter cable for V-Amp:BP-04242-18
actiCHamp2MOVE trigger cable 3 m 15-pin:BP-09302-FA
Active reference electrode:BP-04243-REF
Active ground electrode:BP-04243-GRD
MOVE velcro pad:BP-08420
Pair of ear clips:FMS-060212
MOVE2X Adapter Instruction Leaflet:BP-04249-e
Page 12
Section 2Notes on the safe use of the adapters
The adapters can be used to connect the active electrodes from the actiCAP system or the BCI
package respectively via the MOVE transmitter and receiver to the following amplifiers only:
actiCHamp, V-Amp, QuickAmp or BrainAmp.
For detailed information, please refer to the operating instructions of the EEG amplifier.
The adapters do not provide galvanic isolation between input and output. Therefore, you may
only connect isolated amplifier inputs to the adapter.
Please disconnect the splitter cables of the active electrodes (e.g. actiCAP) after operation.
This preserves the batteries.
Please remove the batteries when the adapter is not used.
12
When combining with other devices to form a system, use the DIN EN 60601-1-1:2002-08 standard.
Do not operate the system within less than 3 meters of an operating cellular phone, similar radio transmitting devices or any other device producing powerful radio interference such as arc
welders, radio thermal treatment equipment, x-ray machines, or any other equipment that produces electrical sparks.
Explosion hazard. Do not use the system in the presence of a flammable anesthetic mixture
with air, oxygen or nitrous oxide.
Make sure that no fluids enter the adapters; therefore do not clean them under running water
or immerse them in water.
Take care when arranging the test subject and sensor cables to avoid risk of subject entanglement or strangulation.
Do not connect the system to a subject undergoing electrosurgery or defibrillation.
MOVE2X Adapter | Instruction Leaflet | Version 003 | April, 2 2012
Page 13
Section 3Technical data
3.1MOVE2actiCAP Adapter and MOVE2BCIPack Adapter
ParameterValue
Channels:MOVE2actiCAP Adapter:
64 referential inputs for EEG acquisition
1 input for REF electrode
1 input for GND electrode
MOVE2BCIPack Adapter:
16 referential inputs for EEG acquisition with an additional
REF electrode and an additional GND electrode built in the
splitter cable; REF and GND connectors are not necessary
13
Output range:1.65 V DC output for GND electrode
Power supply:2x 1.2 V AA NiMH rechargeable batteries
Battery capacity:Min. 2000 mAh
Battery lifetime:500 cycles
Operating time:6 hours (this value may vary due to battery capacity and
temperature)
Dimensions (L x W x H):75 mm x 50 mm x 35 mm
Environment:Operation: 0 °C to 40 °C, rel. humidity 30 to 95%, non-con-
densing
Storage: -35 °C to 65 °C
Air pressure: 700 hPa to 1050 hPa
Weight:Approx. 75 g /140 g (excluding/including batteries)
REF connector:For details, refer to both the actiCAP Operating Instructions
and the actiCHamp Operating Instructions.
GND connector:For details, refer to both the actiCAP Operating Instructions
and the actiCHamp Operating Instructions.
Common input connector:For details, refer to both the actiCAP Operating Instructions
and the actiCHamp Operating Instructions.
Indicators:The green LED indicates device operation.
The yellow LED indicates low battery status.
The yellow LED flashes once when the splitter cable is con-
nected and the adapter starts operation.
Switches:The adapter starts to operate when at least one splitter
cable is connected.
MOVE2X Adapter | Instruction Leaflet | Version 003 | April, 2 2012
Page 14
3.2MOVE2actiCHamp Adapter
ParameterValue
Channels:64 referential inputs for EEG acquisition
1 input for GND electrode
Dimensions (L x W x H):75 mm x 50 mm x 35 mm
Environment:Operation: 0 °C to 40 °C, rel. humidity 30 to 95%, non-con-
densing
Storage: -35 °C to 65 °C
Air pressure: 700 hPa to 1050 hPa
Technical data 14
Weight:Approx. 75 g
GND connector:For details, refer to the actiCHamp Operating Instructions.
Common input connector:For details, refer to the actiCHamp Operating Instructions.
Indicators:The green LED indicates device operation.
Trigger:The actiCHamp trigger cable connects the MOVE receiver
and the actiCHamp trigger input. Depending on the trigger
cable used, the bits of the actiCHamp trigger input are used
as follows:
For the MOVE 64-channel actiCHamp trigger cable:
Bit 1 to 7: Available as trigger lines to the computer.
Bit 8: Set by the first MOVE system if a sample is lost.
For the MOVE 128-channel actiCHamp trigger cable:
Bit 1 to 6: Available as trigger lines to the computer.
Bit 7: Set by the second MOVE system if a sample is lost.
Bit 8: Set by the first MOVE system if a sample is lost.
For the MOVE 160-channel actiCHamp trigger cable:
Bit 1 to 5: Available as trigger lines to the computer.
Bit 6: Set by the third MOVE system if a sample is lost.
Bit 7: Set by the second MOVE system if a sample is lost.
Bit 8: Set by the first MOVE system if a sample is lost.
The TTL level remains "high" (= +5 Volt) until the problem
has disappeared.
Page 15
Section 4Operation and maintenance
4.1Operating the adapters
4.1.1MOVE2actiCAP Adapter
Please note that in order to operate the MOVE2actiCAP Adapter, it is necessary to insert the
supplied rechargeable batteries in the battery compartment.
Before starting acquisition, insert fully charged batteries into the adapter.
Prepare the electrodes according to the Operating Instructions of the actiCAP active electrode
system and measure the impedances; make sure that impedances are always low. The
MOVE2actiCAP Adapter does not provide any impedance measurement. Therefore, you should
use the actiCAP ControlBox II or the actiCHamp amplifier to measure the impedances.
15
Connect the MOVE2actiCAP Adapter to the MOVE transmitter via the ribbon cables and then
connect the subject via the electrode splitter cable for actiCAP (see Figure 6-1 on page 20
connectors to the MOVE transmitter are labeled output 1 to 32 and output 33 to 64. The connectors to the splitter cables are labeled input 1 to 32 and input 33 to 64. Ensure that the right
group of channels is connected to the input/output connectors to avoid confusion.
The MOVE2actiCAP Adapter also requires that the reference and the ground electrode be connected to the test subject. You can attach the reference electrode to the subject by means of
an ear clip. Alternatively, if one of the electrode positions on the cap is used as the reference
site you can remove one EEG electrode from the cap and place the reference electrode there.
After recording, disconnect the actiCAP splitter cables from the MOVE2actiCAP Adapter and
check the electrode impedances again with the help of the actiCAP ControlBox II. If the yellow
LED indicates a low battery status, we recommend that you change the rechargeable batteries
as soon as possible. The electrodes will still provide valid signals but the level of noise may
increase rapidly. Empty batteries cause the actiCAP active electrodes to produce signals that
are corrupted by switching artifacts and therefore are not usable.
If your experimental setup requires more than 64 channels, you can combine two adapters to
form a system of 128 channels (Figure 6-2 on page 21
available from Brain Products on request (see also Section 1.5.2). The use of Y-cables ensures
that identical reference and ground potentials are provided for each amplifier.
If you use a MOVE2actiCAP 128-channel system (two MOVE2actiCAP Adapters) in a configuration where only one adapter is equipped with electrodes (e.g. only 64 channels are to be recorded), you must remove the ground and reference Y-cables and connect the electrodes
directly to the MOVE2actiCAP Adapter equipped with electrodes. Otherwise, all active electrodes may go into saturation (provide flat signals) because the reference electrode is not supplied with power.
). The required connection cables are
). The
MOVE2X Adapter | Instruction Leaflet | Version 003 | April, 2 2012
Page 16
4.1.2MOVE2BCIPack Adapter
Please note that in order to operate the MOVE2BCIPack Adapter, it is necessary to insert the
supplied rechargeable batteries in the battery compartment.
Before starting acquisition, insert fully charged batteries into the adapter.
Prepare the active electrodes and measure the impedances; make sure that impedances are
always low. The MOVE2BCIPack Adapter does not provide any impedance measurement.
Therefore, you should use the ImpBox to measure the impedances (for detailed information,
refer to the V-Amp Operating Instructions).
First, connect the MOVE2BCIPack Adapter to the MOVE transmitter via the ribbon cables supplied and then connect the test subject via the electrode splitter cable for V-Amp (Figure 6-3 on
page 22). The connectors to the MOVE transmitter are labeled output 1 to 32 and output 33 to
64. The connectors to the splitter cables are labeled input 1 to 32 and input 33 to 64. Ensure
that the right group of channels is connected to the input/output connectors to avoid confusion.
Operation and maintenance 16
After you have connected the MOVE2BCIPack Adapter to the MOVE transmitter, connect the
MOVE2V-Amp Adapter (E105b) to the MOVE receiver via the ribbon cable. Then connect the
MOVE2V-Amp Adapter (E105b) to the V-Amp amplifier using the single lead cables with touchproof connectors (see Figure 6-4 on page 22).
The MOVE2BCIPack Adapter requires that the electrode splitter cable for V-Amp be connected
to the adapter's first common input (channels 1 to 32). This splitter cable is equipped with
built-in reference and ground electrodes; it is therefore not necessary to use the adapter's separate inputs for the ground and reference electrodes. Also, please note that the adapter's second common input (channels 33 to 64) remains open.
After recording, disconnect the splitter cables from the MOVE2BCIPack Adapter and check the
electrode impedances again with the help of the ImpBox. If the yellow LED indicates a low battery status, we recommend that you change the rechargeable batteries as soon as possible.
The electrodes will still provide valid signals but the level of noise may increase rapidly. Empty
batteries cause the active electrodes to produce signals that are corrupted by switching artifacts and that are not usable.
4.1.3MOVE2actiCHamp Adapter
The MOVE2actiCHamp Adapter is provided to adapt the MOVE receiver to the actiCHamp amplifier. This module is a passive connector. Please note that although the MOVE2actiCHamp
Adapter has a battery compartment, the use of batteries is neither necessary nor recommended for its operation.
Page 17
Prepare the electrodes according to the Operating Instructions of the actiCAP active electrode
system and measure the impedances using the actiCHamp amplifier (for detailed information,
refer to the actiCHamp Operating Instructions). Make sure that impedances are always low.
Connect the MOVE2actiCHamp adapter to the MOVE receiver via the ribbon cable and then
connect the adapter to the actiCHamp amplifier using the connection cable supplied (see
Figure 6-5 on page 23
33 to 64. The connectors to the amplifier cables are labeled output 1 to 32 and output 33 to 64.
If your experimental setup requires more than 64 channels, you can combine two adapters to
form a system of 128 channels (see Figure 6-6 on page 24
configuration for 128 channels including the trigger cable connection. The required connection
cables are available from Brain Products on request (see also Section 1.5.2). The use of Y-ca-
bles ensures that identical reference and ground potentials are provided for each amplifier.
Figure 6-8 on page 26 shows the combination of three adapters to form a system of 160 chan-
nels.
). The connectors to the MOVE receiver are labeled input 1 to 32 and input
). Figure 6-7 on page 25 shows the
Operation and maintenance 17
4.2Expanding the MOVE system for QuickAmp and BrainAmp
You can connect the MOVE receiver to the QuickAmp amplifier as shown in Figure 6-9 on
page 27.
You can connect the MOVE receiver to the BrainAmp amplifier as shown in Figure 6-10 on
page 28.
4.3Maintenance and cleaning
Please remove the batteries from the MOVE2actiCAP Adapter/MOVE2BCIPack Adapter if the
adapter is not used for an extended period.
Please disconnect the splitter cables of the active electrodes (e.g. actiCAP) after operation.
This preserves the batteries.
If you should notice any damage to the devices, please contact your local representative or
Brain Products directly in order to receive a return authorization.
Use a soft, slightly moist cloth to clean the adapter.
Never clean the adapter under running water. Never use aggressive or corrosive cleaning
agents.
Page 18
Never clean the devices when the test subject is connected to them or when they are connect-
ed to the power supply.
Operation and maintenance 18
Page 19
Section 5Warranty
Brain Products GmbH provides a warranty of 36 months as of the date of purchase in respect
of the adapters.
Consumables (e.g. batteries) are excluded from the manufacturer's warranty.
19
MOVE2X Adapter | Instruction Leaflet | Version 003 | April, 2 2012
Page 20
Section 6Block diagrams
Figure 6-1. MOVE2actiCAP Adapter: connection between MOVE transmitter and actiCAP system
20
MOVE2X Adapter | Instruction Leaflet | Version 003 | April, 2 2012
Page 21
Figure 6-2. Connecting two MOVE2actiCAP Adapters for recording 128 channels from one subject
Block diagrams 21
Page 22
Block diagrams 22
Figure 6-3. MOVE2BCIPack Adapter: connection between MOVE transmitter and BCI active electrodes (on test subject side)
Figure 6-4. MOVE2V-Amp Adapter (E105b): connection between MOVE receiver and V-Amp system (on amplifier side)
Page 23
Figure 6-5. MOVE2actiCHamp Adapter: connection between MOVE receiver and actiCHamp system
Block diagrams 23
Page 24
Figure 6-6. Connecting two MOVE2actiCHamp Adapters for recording 128 channels
Block diagrams 24
Page 25
Figure 6-7. Trigger cable connection for a 128-channel MOVE2actiCHamp system
Block diagrams 25
Page 26
Figure 6-8. Trigger cable connection for a 160-channel MOVE2actiCHamp system
Block diagrams 26
Page 27
Figure 6-9. Connection between MOVE receiver and QuickAmp system
Block diagrams 27
Page 28
Figure 6-10. Connection between MOVE receiver and BrainAmp system
Block diagrams 28