Boston Scientific U125, U128, V272, V273, V172 Technical Manual

...
Page 1
CAUTION: Federal law (USA)
restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures.
PHYSICIAN’S TECHNICAL MANUAL
VALITUDE™, VALITUDE™X4,
INTUA
,INVIVE
CARDIAC RESY NCHRONIZATION THERAPY PAC E MAK ER
Model U125, U128, V272, V273, V172, V173
Page 2
Page 3
Table of Con tents
Additional Information..................................................................................................................... 1
Device Description.......................................................................................................................... 1
Related Information ........................................................................................................................ 3
Indications and Usage.................................................................................................................... 4
Contraindications ............................................................................................................................ 4
Warnings ........................................................................................................................................ 5
Precautions..................................................................................................................................... 7
Supplemental Precautionary Information ..................................................................................... 20
Post-Therapy Pulse Generator Follow Up .......................................................................... 20
Minimizing Pacemaker/S-ICD Interaction ........................................................................... 21
Transcutaneous Electrical Nerve Stimulation (TENS)......................................................... 23
Electrocautery and Radio Frequency (RF) Ablation............................................................ 24
Ionizing Radiation ................................................................................................................ 26
Elevated Pressures ............................................................................................................. 27
Potential Adverse Events ............................................................................................................. 28
Mechanical Specications ............................................................................................................ 31
Items Included in Package ........................................................................................................... 34
Symbols on Packaging................................................................................................................. 34
Characteristics as Shipped ........................................................................................................... 38
X-Ray Identier............................................................................................................................. 39
Federal Communications Commission (FCC) .............................................................................. 41
Pulse Generator Longevity ........................................................................................................... 42
Warranty Information .................................................................................................................... 46
Product Reliability......................................................................................................................... 46
Patient Counseling Information ....................................................................................................47
Patient Handbook................................................................................................................ 48
Lead Connections......................................................................................................................... 48
Page 4
Implanting the Pulse Generator .................................................................................................... 52
Step A: Check Equipment ................................................................................................ 52
Step B: Interrogate and Check the Pulse Generator ........................................................ 52
Step C: Implant the Lead System ..................................................................................... 53
Step D: Take Baseline Measurements ............................................................................. 54
Step E: Form the Implantation Pocket.............................................................................. 57
Step F: Connect the Leads to the Pulse Generator.......................................................... 58
Step G: Evaluate Lead Signals......................................................................................... 62
Step H: Program the Pulse Generator.............................................................................. 63
Step I: Implant the Pulse Generator ................................................................................. 64
Step J: Complete and Return the Implantation Form ....................................................... 65
Bidirectional Torque Wrench......................................................................................................... 66
Follow Up Testing ......................................................................................................................... 68
Predischarge Follow Up ...................................................................................................... 68
Routine Follow Up............................................................................................................... 68
Explantation .................................................................................................................................. 69
Page 5
ADDITIONAL INFORMATION
For additional reference information, go to www.bostonscientic.com/ifu.
DEVICE DESCRIPTION
This manual contains information about the VALITUDE, INTUA, and INVIVE families of cardiac resynchronization therapy pacemakers (CRT-Ps) (specic models are listed in "Mechanical Specications" on page 31).
NOTE: Specic features discussed in this manual may not apply to all models. References to names of non-quadripolar devices also apply to the corresponding quadripolar devices. References to “ICD” include all types of ICDs (e.g., ICD, CRT-D, S-ICD).
Therapies
These pulse generators provide a variety of therapies, including:
Cardiac Resynchronization Therapy (CRT), which treats heart failure by resynchronizing ventricular contractions through biventricular electrical stimulation
Bradycardia pacing, including adaptive rate pacing, to detect and treat bradyarrhythmias
Leads
The pulse generator has independently programmable outputs and accepts one or more of the following leads, depending on the model:
•OneIS-1
1
unipolar or bipolar atrial lead
One IS-1 unipolar or bipolar right ventricular lead
One LV-1 unipolar or bipolar left ventricular lead
1. IS-1 refers to the international standard ISO 5841-3:2013.
1
Page 6
One IS-1 unipolar or bipolar left ventricular lead
One compatible IS4
2
quadripolar left ventricular lead
3
The pulse generator and the leads constitute the implantable portion of the pulse generator system.
PRM System
These pulse generators can be used only with the ZOOM LATITUDE Programming System, which is the external portion of the pulse generator system and includes:
Model 3120 Programmer/Recorder/Monitor (PRM)
Model 3140 ZOOM Wireless Transmitter
Model 2869 ZOOMVIEW Software Application
Model 6577 Accessory Telemetry Wand
You can use the PRM system to do the following:
Interrogate the pulse generator
Program the pulse generator to provide a variety of therapy options
Access the pulse generator’s diagnostic features
Perform noninvasive diagnostic testing
Access therapy history data
Store a 12 second trace of the ECG/EGM display from any screen
Access an interactive Demonstration Mode or Patient Data Mode without the presence of a pulse generator
Print patient data including pulse generator therapy options and therapy history data
2. IS4 refers to the international standard ISO 27186:2010.
3. Compatibility with St. Jude Quartet IS4 lead has been demonstrated through chronic in-vivo animal GLP testing.
2
Page 7
Save patient data
You can program the pulse generator using two methods: autom atically using Indications-Based Programming (IBP) or manually.
RELATED INFORMATION
Refer to the lead’s instruction manual for implant information, general warnings and precautions, indications, contraindications, and technical specications. Read this material carefully for implant procedure instructions specic to the chosen lead congurations.
Refer to the PRM system Operator’s Manual or ZOOM Wireless Transmitter Reference Guide for specific information about the PRM or ZOOM Wireless Transmitter such as setup, maintenance, and handling.
LATITUDE N XT is a remote monitoring system that provides pulse generator data for clinicians. These pulse generators are designed to be LATITUDE NXT enabled; availability varies by region.
LATITUDE NXT is available for the following devices: VALITUDE, INTUA, INVIVE.
Physicians/Clinicians—LATITUDE NXT enables you to periodically monitor both patient and device
status remotely and automatically. The LATITUDE NXT system provides patient data that can be used as part of the clinical evaluation of the patient.
Patients—A key component of the system is the LATITUDE Communicator, an easy-to-use, in-home
monitoring device. The Communicator automatically reads implanted device data from a compatible Boston Scientic pulse generator at times scheduled by the physician. The Communicator sends this data to the LATITUDE NXT secure server through a standard analog telephone line or over a cellular data network. The LATITUDE NXT server displays the patient data on the LATITUDE NXT Web site, which is readily accessible over the Internet to authorized physicians and clinicians.
Refer to t he LATITUDE NXT Clinician Manual for more information.
3
Page 8
INTENDED AUDIENCE
This literature is intended for use by professionals trained or experienced in device implant and/or follow-up procedures.
INDICATIONS AND USAGE
Boston Scientic c ardiac resynchronization therapy pacemakers (CRT-Ps) a re indicated for patients with moderate to severe heart failure (NYHA Class III/IV) including left ventricular dysfunction (EF 35%) and QRS duration 120 ms and remain symptomatic despite stable optimal pharmacological therapy (OPT) for heart failure.
Atrial tracking modes are also indicated for patients who may benet from maintenance of AV synchrony.
Adaptive-rate pacing is indicated for patients exhibiting chronotropic incompetence and who would benetfrom increased pacing rates concurrent with increases in physical activity.
CONTRAINDICATIONS
These Boston Scientic pulse generators have t he following contraindications:
This device is contraindicated in patients who have a separate implanted cardioverter debrillator (ICD) with transvenous leads.
Unipolar pacing or use of the Respiratory Sensor with a Subcutaneous Implantable Cardioverter Debrillator (S-ICD) is contraindicated because it may cause inappropriate therapy or inhibition of appropriate S-ICD therapy.
Single-chamber atrial pacing is contraindicated in patients with impaired AV nodal conduction.
Atrial tracking modes are contraindicated in patients with chronic refractory atrial tachyarrhythmias (atrial brillation or utter), which might trigger ventricular pacing.
Asynchronous pacing is contraindicated in the presence (or likelihood) of competition between paced and intrinsic rhythms.
4
Page 9
WARNINGS
General
Labeling knowledge. Read this manual thoroughly before implantation to avoid damage to the pulse
generator and/or lead. Such damage can result in patient injury or death.
For single patient use only. Do not reuse, reprocess, or resterilize. Reuse, reprocessing, or
resterilization may compromise the structural integrity of the device and/or lead to device failure which, in turn, may result in patient injury, illness, or death. Reuse, reprocessing, or resterilization may also create a risk of contamination of the device and/or cause patient infection or cross-infection, including, but not limited to, the transmission of infectious disease(s) from one patient to another. Contamination of the device may lead to injury, illness, or death of the patient.
Backup debrillation protection. Always have external debrillation equipment available during
implant and electrophysiologic testing. If not terminated in a timely fashion, an induced ventricular tachyarrhythmia can result in the patient’s death.
Separate pulse generator. Using multiple pulse generators could cause pulse generator interaction,
resulting in patient injury or a lack of therapy delivery. Test each system individually and in combination to help prevent undesirable interactions ("Minimizing Pacemaker/S-ICD Interaction" on page 21).
Safety Core operation. In response to applicable nonrecoverable or repeat fault conditions, the pulse
generator will switch irreversibly to Safety Core operation. Safety Core pacing is unipolar, which may interact with an ICD ("Minimizing Pacemaker/S-ICD Interaction" on page 21).
Handling
Do not kin k leads. Do not kink, twist, or braid the lead with other leads as doing so could cause lead
insulation abrasion damage or conductor damage.
5
Page 10
Handling the lead without Connector Tool. For leads that require the use of a Connector Tool, use caution handling the lead terminal when the Connector Tool is not present on the lead. Do not directly contact the lead terminal with any surgical instruments or electrical connections such as PSA (alligator) clips, ECG connections, forceps, hemostats, and clamps. This could damage the lead terminal, possibly compromising the sealing integrity and result in loss of therapy or inappropriate therapy, such as a short within the header.
Handling the terminal while tunneling. Do not contact any other portion of the IS4–LLLL lead terminal, other than the terminal pin, even when the lead cap is in place.
Programming and Device Operations
Atrial tracking modes. Do not use atrial tracking modes in patients with chronic refractory atrial
tachyarrhythmias. Tracking of atrial arrhythmias could result in ventricular tachyarrhythmias.
Atrial-only modes. Do not use atrial-only modes in patients with heart failure because such modes do not provide CRT.
Lead Safety Switch. Lead Safety Switch should be programmed Off for patients with an ICD. Unipolar pacing due to Lead Safety Switch is contraindicated for patients with an ICD.
RAAT testing. Unipolar pacing due to RAAT is contraindicated and should be programmed off for patients with an ICD. The RAAT feature performs automatic threshold testing in a unipolar pacing con guration.
Ventricular sensing. Left ventricular lead dislodgement to a position near the atria can result in atrial oversensing and left ventricular pacing inhibition.
Sensitivity settings and EMI. If programmed to a xed atrial Sensitivity value of 0.15 mV, or a xed sensitivity value of 2.0 mV or less in a unipolar lead conguration in any chamber, the pulse generator may be more susceptible to electromagnetic interference. This increased susceptibility should be taken into consideration when determining the follow-up schedule for patients requiring such a setting.
6
Page 11
Post-Implant
Protected environments. Advise patients to seek medical guidance before entering environments that
could adversely affect the operation of the active implantable medical device, including areas protected by a warning notice that prevents entry by patients who have a pulse generator.
Magnetic Resonance Imaging (MRI) exposu re. Do not expose a patient to MRI scanning. Strong
magnetic elds may damage the pulse generator and/or lead system, possibly resulting in injury to or death of the patient.
•Diathermy.Do not subject a patient with an implanted pulse generator and/or lead to diathermy since
diathermy may cause brillation, burning of the myocardium, and irreversible damage to the pulse generator because of induced currents.
PRECAUTIONS
Clinical Considerations
STAT PACE. STAT PACE will initiate unipolar pacing. Unipolar pacing due to STAT PACE may cause
inappropriate therapy or inhibition of appropriate S-ICD therapy.
Pacemaker-mediated tachycardia (PMT). Programming minimum PVARP less than retrograde V–A
conduction may increase the likelihood of a PMT.
Rate Adaptive Pacing in Heart Failure Patients. The clinical benet of Rate Adaptive Pacing in heart
failure patients has not been studied. Rate Adaptive Pacing should be used with medical discretion if the patient develops an indication such as chronotropic incompetence. Patients with heart failure may have hemodynamic compromise at rapid sensor-driven rates, and the physician may wish to program less aggressive rate adaptive parameters in accordance with patient condition. Rate Adaptive Pacing may be helpful for heart failure patients with coexisting bradyarrhythmic conditions. It is not recommended for patients who exhibit only heart failure-induced chronotropic incompetency.
7
Page 12
Sterilization and Storage
If package is damaged. The blister trays and contents are sterilized with ethylene oxide gas before
nal packaging. When the pulse generator and/or lead is received, it is sterile provided the container is intact. If the packaging is wet, punctured, opened, or otherwise damaged, return the pulse generator and/or lead to Boston Scientic.
If device is dropped. Do not implant a device which has been dropped while outside of its intact shelf package. Do not implant a device which has been dropped from a height of more than 24 inches (61 cm) while within its intact shelf package. Sterility, integrity and/or function cannot be guaranteed under these conditions and the device should be returned to Boston Scientic for inspection.
Storage tem perature and equilibration. Recommended storage tem peratures are 0°C–50°C (32°F–122°F). Allow the device to reach a proper temperature before using telemetry communication capabilities, programming or implanting the device because temperature extremes may affect initial device function.
•Devicestorage.Store the pulse generator in a clean area away from magnets, kits containing magnets, and sources of EMI to avoid device damage.
•Usebydate.Implant the pulse generator and/or lead before or on the USE BY date on the package label because this date reects a validated shelf life. For example, if the date is January 1, do not implant on or after January 2.
Implantation
Expected benets. Determine whether the expected device benets provided by programmable options
outweigh the possibility of more rapid battery depletion.
Evaluate patient for surgery. There may be additional factors regarding the patient’s overall health and medical condition that, while not related to device function or purpose, could render the patient a poor candidate for implantation of this system. Cardiac health advocacy groups may have published guidelines that may be helpful in conducting this evaluation.
8
Page 13
Lead compatibility. Prior to implantation, conrm the lead-to-pulse generator compatibility. Using
incompatible leads and pulse generators can damage the connector and/or result in potential adverse consequences, such as undersensing of cardiac activity or failure to deliver necessary therapy.
Telemetry wand. Make sure a sterile telemetry wand is available should loss of ZIP telemetry occur.
Verify that the wand can easily be connected to the programmer and is within reach of the pulse generator.
Line-powered equipment. Exercise extreme caution if testing leads using line-powered equipment
because leakage current exceeding 10 µA can induce ventricular brillation. Ensure that any line-powered equipment is within specications.
Replacement device. Implanting a replacement device in a subcutaneous pocket that previously housed
a larger device may result in pocket air entrapment, migration, erosion, or insufcient grounding between the device and tissue. Irrigating the pocket with sterile saline solution decreases the possibility of pocket air entrapment and insufcient grounding. Suturing the device in place reduces the possibility of migration and erosion.
Do not bend the lead near the lead-header interface. Insert the lead terminal straight into the lead
port. Do not bend the lead near the lead-header interface. Improper insertion can cause insulation or connector damage.
Absence of a lead. The absence of a lead or plug in a lead port may affect device performance. If a
lead is not used, be sure to properly insert a plug in the unused port, and then tighten the setscrew onto the plug.
Dual chamber device without a functional RV lead. If a dual-chamber device is programmed to
AAI(R), ensure that a functional RV lead is present. In the absence of a functional RV lead, programming to AAI(R) may result in undersensing or oversensing.
9
Page 14
Electrode connections. Do not insert a lead into the pulse generator connector without taking the following precautions to ensure proper lead insertion:
Insert the torque wrench into the preslit depression of the s eal plug before inserting the lead into the
port, to release any trapped uid or air.
Visually verify that the setscrew is sufciently retracted t o allow insertion. Use the torque wrench
to loosen the setscrew if necessary.
Fully insert each lead into its lead port and then tighten the setscrew onto the terminal pin.
Do not suture directly over lead. Do not suture directly over the lead body, as this may cause structural damage. Use the suture sleeve to secure the lead proximal to the venous entry site to prevent lead movement.
Diaphragmatic stimulation. Patients should be tested for diaphragmatic stimulation by pacing the LV lead through the pulse generator at 7.5 V and adjusting the lead congurations and lead position as necessary. PSA testing at higher outputs (e.g., 10.0 V) may also be considered to better characterize stimulation margins. The probability of diaphragmatic stimulation increases when a pacing system includes an LV lead because of this lead’s proximity to the phrenic nerve.
Device Programming
Device communication. Use only the designated PRM and software application to communicate
with this pulse generator.
STAT PACE settings. When a pulse generator is programmed to STAT PACE settings, it will continue to pace at the high-energy STAT PACE values if it is not reprogrammed. The use of STAT PACE parameters will likely decrease device longevity.
10
Page 15
Biventricular pacing therapy. This device is intended to provide biventricular pacing therapy.
Programming the device to provide RV-only pacing, or programming the RV pace amplitude below the pacing threshold (resulting in LV-only pacing), is not intended for the treatment of heart failure. The clinical effects of LV-only or RV-only pacing for the treatment of heart failure have not been established.
Pacing and sensing margins. Consider lead maturation in your choice of Pacing Amplitude, pacing
Pulse Width, and Sensitivity settings.
An acute Pacing Threshold greater than 1.5 V or a chronic Pacing Threshold greater than 3 V can result in loss of capture because thresholds may increase over time.
An R-Wave Amplitude less than 5 mV or a P-Wave Amplitude less than 2 mV can result in undersensing because the sensed amplitude may decrease after implantation.
Pacing Lead Impedance should be greater than the programmed Low Impedance Limit and less than 2000 (or the programmed High Impedance Limit).
Lead impedance values and Lead Safety Switch. If properly functioning leads with stable measured impedance values near the programmed impedance limits are used, consider programming Lead Safety Switch Off o r changing the impedance limits to avoid undesirable switching to a Unipolar Lead Conguration.
Proper programming of the lead conguration. If the Lead Conguration is programmed to Bipolar when a unipolar lead is implanted, pacing will not occur.
Programming for supraventricular tachyarrhythmias (SVTs). Determine if the device and programmable options are appropriate for patients with SVTs because SVTs can initiate unwanted device therapy.
AV Delay. To ensure a high percentage of biventricular pacing, the programmed AV Delay setting must be less than the patient’s intrinsic PR interval.
Adaptive-rate pacing. Rate Adaptive Pacing should be used with care in patients who are unable to tolerate increased pacing rates.
11
Page 16
Ventricular refractory periods (VRPs) in ad aptive-rate pacing. Adaptive-rate pacing is not limited by refractory periods. A long refractory period programmed in combination with a high MSR can result in asynchronous pacing during refractory periods since the combination can cause a very small sensing window or none at all. Use Dynamic AV Delay or Dynamic PVARP to optimize sensing windows. If you are programming a xed AV Delay, consider the sensing outcomes.
MTR/MSR programming. The pulse generator’s MTR and MSR should be programmed to a rate lower than a concomitant S-ICD’s lowest tachycardia detection zone.
Atrial Tachy Response (ATR). ATR should be programmed to On if the patient has a history of atrial tachyarrhythmias. The delivery of CRT is compromised because AV synchrony is disrupted if the ATR mode switch occurs.
Threshold test. During a manual LV Threshold test, RV Backup Pacing is unavailable.
Left ventricular pacing only. The clinical effect of LV pacing alone for heart failure patients has not
been studied.
Atrial oversensing. Take care to ensure that artifacts from the ventricles are not present on the atrial channel, or atrial oversensing may result. If ventricular artifacts are present in the atrial channel, the atrial lead may need to be repositioned to minimize its interaction.
ATR entry count. Exercise care when programming the Entry Count to low values in conjunction with a short ATR Duration. This combination allows mode switching with very few fast atrial beats. For example, if the Entry Count was programmed to 2 and the ATR Duration to 0, ATR mode switching could occur on 2 fast atrial intervals. In these instances, a short series of premature atrial events could cause the device to mode switch.
ATR exit count. Exercise care when programming the Exit Count to low values. For example, if the Exit Count was programmed to 2, a few cycles of atrial undersensing could cause termination of mode switching.
12
Page 17
Proper programming w ithout an atrial lead. If an atrial lead is not implanted (port is plugged instead),
or an atrial lead is abandoned but remains connected to the header, device programming should be consistent with t he number and type of leads actually in use.
Atrial sensing programmed to Off. When atrial sensing is programmed to Off in a DDI(R) or DDD(R)
mode, any atrial pacing that occurs will be asynchronous. Additionally, features that require atrial sensing may not function as expected.
High atrial rates. Sensing high atrial rates may impact device longevity. Therefore, the Atrial Sense
lead conguration will be seeded to Off when programming from an atrial sensing mode to a non-atrial sensing mode.
Cross-chamber artifacts. Sensitivity adjustments associated with Smart Blanking may not be sufcient
to inhibit detection of cross-chamber artifacts if the cross-chamber artifacts are too large. Consider other factors that impact the size/amplitude of cross-chamber artifacts including lead-placement, pacing output, and programmed Sensitivity settings.
Sensor signal artifacts. If Respiratory Sensor signal artifacts are observed on EGMs, and the leads
are otherwise shown to be performing appropriately, consider programming the sensor to Off to prevent oversensing.
Single pass VDD leads. When a single pass VDD lead is used with a dual-chamber device, the atrial
electrodes may not be in contact with the atrial wall. In this case, the measured depolarization signal has a relatively low Amplitude and could require a more sensitive setting.
Left ventricular lead conguration. Proper programming of the LV coronary venous Lead Conguration
is essential for proper LV lead function. Program the Lead Conguration in accordance with the number of electrodes on the LV lead; otherwise, erratic LV sensing, loss of LV pacing, or ineffective LV pacing might occur.
Quadripolar pacing conguration. When an LVRing4>>RV pacing conguration is programmed with
an IS4-LLLL lead, the LV tip may be used as the anode rather than the RV ring. When programming to this conguration, evaluate the pacing threshold and ensure no extracardiac stimulation is present.
13
Page 18
Left Ventricular Protection Period (LVPP). Use of a long LVPP reduces the maximum LV pacing rate and may inhibit CRT at higher pacing rates.
Sensing adjustment. Following any Sensitivity parameter adjustment or any modication of the sensing lead, always verify appropriate sensing. Programming Sensitivity to the highest value (lowest sensitivity) may result in undersensing of cardiac activity. Likewise, programming to the lowest value (highest sensitivity) may result in oversensing of non-cardiac signals.
Sensitivity in unipolar lead conguration. T he amplitude and prevalence of myopotential noise is increased in unipolar lead congurations, as compared to bipolar lead congurations. For patients with a unipolar lead conguration and myopotential oversensing during activity involving the pectoral muscles, the programming of Fixed Sensitivity is recommended.
Use of Patient Triggered Monitor. Use care when using Patient Triggered Monitor, because the following conditions will exist while it is enabled:
All other magnet features, including asynchronous pacing, are disabled. The Magnet feature will
not indicate magnet position.
Device longevity is impacted. To help reduce the longevity impact, PTM only allows storage of one
episode, and PTM is automatically disabled after 60 days if data storage was never triggered.
Once the EGM is stored (or 60 days elapses), PTM is disabled and the device Magnet Response
automatically will be set to Pace Async. However, if a magnet is used, the pulse generator will not revert to asynchronous operation until the magnet is removed for 3 seconds and placed on the device again.
14
Page 19
Environmental and Medical Therapy Hazards
Avoid electromagnetic interference (EMI). Advise patients to avoid sources of EMI. The pulse
generator may inhibit pacing due to oversensing, or may switch to asynchronous pacing at the programmed pacing rate or at the magnet rate in the presence of EMI.
Moving away from the source of the EMI or turning off the source usually allows the pulse generator to return to normal operation.
Examples of potential EMI sources are:
Electrical power sources, arc welding or resistance welding equipment, and robotic jacks
High voltage power distribution lines
Electrical smelting furnaces
Large RF transmitters such as radar
Radio transmitters, including those used to control toys
Electronic surveillance (antitheft) devices
An alternator on a car that is running
Medical treatments and diagnostic tests in which an electrical current is passed through the body, such as TENS, electrocautery, electrolysis/thermolysis, electrodiagnostic testing, electromyography, or nerve conduction studies
Any externally applied device that uses an automatic lead detection alarm system (e.g., an EKG machine)
Hospital and Medical Environments
Mechanical ventilators. During mechanical ventilation, respiration-based trending may be misleading;
therefore, the Respiratory Sensor should be programmed to Off.
15
Page 20
Conducted electrical current. Any medical equipment, t reatment, therapy, or diagnostic test that introduces electrical current into the patient has the potential to interfere with pulse generator function.
External patient monitors (e.g., respiratory monitors, surface ECG monitors, hemodynamic
monitors) may interfere with the pulse generator’s impedance-based diagnostics (e.g., Respiratory Rate trend). To resolve suspected interactions with Respiratory Sensor-based diagnostics, deactivate the pulse generator’s Respiratory Sensor by programming it to Off.
Medical therapies, treatments, and diagnostic tests that use conducted electrical current (e.g.,
TENS, electrocautery, electrolysis/thermolysis, electrodiagnostic testing, electromyography, or nerve conduction studies) may interfere with or damage the pulse generator. Program the device to Electrocautery Protection Mode prior to the treatment, and monitor device performance during the treatment. After the treatment, verify pulse generator function ("Post-Therapy Pulse Generator Follow Up" on page 20).
•Internaldebrillation. Do not use internal defibrillation paddles or catheters unless the pulse generator is disconnected from the leads because the leads may shunt energy. This could result in injury to the patient and damage to the implanted system.
16
Page 21
External debrillation. Itcantakeupto15secondsforsensingto recover after an external shock is
delivered. In non-emergency situations, for pacemaker dependent patients, consider programming the pulse generator to an asynchronous pacing mode and programming the Respiratory Sensor to Off prior to performing external cardioversion or debrillation.
External debrillation or cardioversion can damage the pulse generator. To help prevent damage to the pulse generator, consider the following:
Avoid placing a pad (or paddle) directly over the pulse generator. Position the pads (or paddles) as far from the pulse generator as possible.
Position the pads (or paddles) in a posterior-anterior orientation when the device is implanted in the right pectoral region or an anterior-apex orientation when the device is implanted in the left pectoral region.
Set energy output of external debrillation equipment as low as clinically acceptable.
Following external cardioversion or debrillation, verify pulse generator function ("Post-Therapy Pulse Generator Follow Up" on page 20).
Lithotripsy. Extracorporeal shock wave lithotripsy (ESWL) may cause electromagnetic interference with or damage to the pulse generator. If ESWL is medically necessary, consider the following to minimize the potential for encountering interaction:
Focus the ESWL beam at least 15 cm (6 in) away from the pulse generator.
Depending on the pacing needs of the patient, program the Brady Mode to a non-rate-responsive
VVI or VOO mode.
Ultrasound energy. Therapeutic ultrasound (e.g., lithotripsy) energy may damage the pulse generator. If therapeutic ultrasound energy must be used, avoid focusing near the pulse generator site. Diagnostic ultrasound (e.g., echocardiography) is not known to be harmful to the pulse generator.
17
Page 22
Electrical interference. Electrical interference or “noise” from devices such as electrocautery and monitoring equipment may interfere with establishing or maintaining telemetry for interrogating or programming the device. In the presence of such interference, move the programmer away from electrical devices, and ensure that the wand cord and cables are not crossing one another. If telemetry is cancelled as a result of interference, the device should be re-interrogated prior to evaluating information from pulse generator memory.
Radio frequency (RF) interference. RF signals from devices that operate at frequencies near that of the pulse generator may interrupt ZIP telemetry while interrogating or programming the pulse generator. This RF interference can be reduced by increasing the distance between the interfering device and the PRM and pulse generator. Examples of devices that may cause interference in the 916.5 MHz frequency band include:
Cordless phone handsets or base stations
Certain patient monitoring systems
Central line guidewire insertion. Use caution when inserting guidewires for placement of other types of central venous catheter systems such as PIC lines or Hickman catheters in locations where pulse generator leads may be encountered. Insertion of such guidewires into veins containing leads could result in the leads being damaged or dislodged.
Home and Occupational Environments
Home appliances. Home appliances that are in good working order and properly grounded do not
usually produce enough EMI to interfere with pulse generator operation. There have been reports of pulse generator disturbances caused by electric hand tools or electric razors used directly over the pulse generator implant site.
18
Page 23
Magnetic elds. Advise patients that extended exposure to strong (greater than 10 gauss or 1 mTesla)
magnetic elds may t rigger the magnet feature. Examples of magnetic sources include:
Industrial transformers and motors
MRI scanners
Large stereo speakers
Telephone receivers if held within 1.27 cm (0.5 inches) of the pulse generator
Magnetic wands such as those used for airport security and in the Bingo game
Electronic Article Surveillance (EAS) and Security Systems. Advise patients to avoid lingering near
or leaning against antitheft and security gates or tag readers that include radio frequency identication (RFID) equipment. These systems may be found at the entrances and exits of stores, in public libraries, and in point-of-entry access control systems. These systems are unlikely to affect cardiac device function when patients walk through them at a normal pace. If the patient is near an electronic antitheft, security, or entry control system and experiences symptoms, t hey should promptly move away from nearby equipment and inform their doctor.
Cellular phones. Advise patients to hold cellular phones to the ear opposite the side of the implanted
device. Patients should not carry a cellular phone that is turned on in a breast pocket or on a belt within 15 cm (6 inches) of the implanted device since some cellular phones may cause the pulse generator to deliver inappropriate therapy or inhibit appropriate therapy.
Follow-up Testing
Pacing threshold testing. If the patient’s condition or drug regimen has changed or device parameters
have been reprogrammed, consider performing a pacing threshold test to conrm adequate margins for pace capture.
19
Page 24
Follow-up considerations for patients leaving the country. Pulse generator follow-up considerations should be made in advance for patients who plan to travel or relocate post-implant to a country other than the country in w hich their device was implanted. Regulatory approval status for devices and associated programmer software congurations varies by country; certain countries may not have approval or capability to follow specic products.
Contact Boston Scientic, using the information on the back cover, for help in determining feasibility of device follow-up in the patient’s destination country.
Explant and Disposal
Incineration. Be sure that the pulse generator is removed before cremation. Cremation and incineration
temperatures might cause the pulse generator to explode.
Device handling. Before explanting, cleaning, or shipping the device, complete the following actions to prevent overwriting of important therapy history data:
Program the pulse generator Brady Mode to Off
Program Ventricular Tachy EGM Storage to Off
Clean and disinfect the device using standard biohazard handling techniques.
SUPPLEMENTAL PRECAUTIONARY INFORMATION
Post-Therapy Pulse Generator Follow Up
Following any surgery or medical procedure with the potential to affect pulse generator function, you should perform a thorough follow-up, which may include the following:
Interrogating the pulse generator with a programmer
Reviewing clinical events and fault codes
Reviewing the Arrhythmia Logbook, including stored electrograms (EGMs)
20
Page 25
Reviewing real-time EGMs
Testing the leads (threshold, amplitude, and impedance)
Reviewing respiratory sensor-based diagnostics
Verifying battery status
Programming any permanent brady parameter to a new value and then reprogramming it back to the
desired value
Saving all patient data
Verifying the appropriate nal programming prior to allowing the patient to leave the clinic
Minimizing Pacemaker/S-ICD Interaction
These pulse generators are compatible for use with a Subcutaneous Implantable Cardioverter Debrillator (S-ICD) when implanted with bipolar leads and programmed to a bipolar pacing conguration.
A pacemaker can interact with an S-ICD in the following ways:
If during a tachyarrhythmia the pacemaker is not inhibited and the pacing pulses are detected by the
rate-sensing circuit of the S-ICD, the S-ICD could interpret the pacing pulses as a normal rhythm. The S-ICD would not detect the arrhythmia and therefore would not deliver therapy.
Pacemaker failure to sense or to capture could result in two independent signals (intrinsic and pacing
pulses) to the S-ICD. This could cause the S-ICD’s rate measurement to be faster than the actual heart rate. As a result, the S-ICD could deliver unnecessary therapy.
If the S-ICD counts both the pacing pulses and the resultant ventricular depolarizations, the S-ICD’s rate
measurement would be faster than the actual heart rate. This could result in unnecessary S-ICD therapy.
21
Page 26
In Safety Mode, these pulse generators use a unipolar pacing and sensing conguration. Safety Mode is compatible for use with an S-ICD because the congured parameters mitigate the potential pacemaker and S-ICD interactions as follows:
Sensing is AGC at 0.25 mV. The AGC sensing is able to effectively sense an intrinsic rhythm faster than the Safety Mode LRL of 72.5 bpm. As a result, pacing is inhibited and does not interfere with S-ICD tachyarrhythmia detection.
When pacing is necessary, the elevated output of 5.0 V and 1.0 ms reduces the risk of not capturing.
If double detection of the pace pulse and the resulting depolarization were to occur, it would not result in unnecessary S-ICD therapy provided the S-ICD tachy threshold is more than twice the Safety Mode LRL (145 ppm).
To help minimize device-device interaction of a bipolar pacemaker when an S-ICD is already implanted, follow these precautionary measures:
Use bipolar pacing leads with close electrode spacing in both chambers. Signicant spacing between electrodes may increase the likelihood that the S-ICD will detect the pacing pulses.
Consider programming the pacemaker to (1) the lowest Amplitude allowable for safe capture in the chronic state, (2) the maximum Sensitivity (the lowest programmable level) while maintaining an adequate safety margin, and (3) the minimum cardiac rate acceptable for the patient.
In addition to the above steps, perform the following testing to assess device-device interaction:
Use the S-ICD features, such as markers, real-time e lectrograms (EGMs), and/or beeping tones, to help evaluate potential for pacemaker interaction due to oversensing by the S-ICD.
NOTE: If a single chamber pacemaker is implanted with an atrial lead, perform testing in both unipolar and bipolar congurations.
22
Page 27
Ventricular brillation and all of the patient’s ventricular tachycardias should be induced while the S-ICD
is activated and the pacemaker is programmed to an asynchronous mode at maximum Amplitude and Pulse Width. This should provide the greatest opportunity for inhibition of arrhythmia detection due to detection of pacemaker pacing pulses. The pacemaker leads might have to be repositioned to eliminate detection of the pacing pulses by the S-ICD.
Temporarily deactivate the patient’s S-ICD when (1) evaluating pacing and sensing thresholds, (2) when using an external temporary pacemaker during implant, and (3) when reprogramming an implanted pacemaker.
Following any S-ICD discharge, reinterrogate the pacemaker to ensure that the S-ICD shock did not damage the pacemaker.
If implanting an S-ICD in a patient who has a pacemaker already implanted, refer to the S-ICD manual for implantation considerations.
Refer to the Warnings section for additional information regarding pacemaker and S-ICD interactions.
Transcutaneous Electrical Nerve Stimulation (TENS)
CAUTION: TENS involves passing electrical current through the body, and may interfere with pulse generator
function. If TENS is medically necessary, evaluate the TENS therapy settings for compatibility with the pulse generator. The following guidelines may reduce the likelihood of interaction:
Place the TENS electrodes as close together and as far away from the pulse generator and leads
as possible.
Use the lowest clinically-appropriate TENS energy output.
Consider cardiac monitoring during TENS use, especially for pacemaker-dependent patients.
Additional steps can be taken to help reduce interference during in-clinic use of TENS:
If interference is suspected during in-clinic use, turn off the TENS unit.
If pacing inhibition is observed, use a magnet to pace asynchronously.
23
Page 28
Do not change TENS settings until you have veried that the new settings do not interfere with pulse generator function.
If TENS is medically necessary outside the clinical setting (at-home u se), provide patients with the following instructions:
Do not change the TENS settings or electrode positions unless instructed to do so.
End each TENS session by turning off the unit before removing the electrodes.
If the patient experiences symptoms of lightheadedness, dizziness, or loss of consciousness during TENS use, they should turn off the TENS unit and contact their physician.
Follow these steps to use the PRM to evaluate pulse generator function during TENS use:
1. Observe real-time EGMs at prescribed TENS output settings, noting when appropriate sensing or interference occurs.
NOTE: Patient triggered monitoring may be used as an additional method to conrm device function during TENS use.
2. When nished, turn off the TENS unit.
You should also perform a thorough follow-up evaluation of the pulse generator following TENS, to ensure that device function has not been compromised ("Post-Therapy Pulse Generator Follow Up" on page 20).
For additional information, contact Boston Scientic using the information on the back cover.
Electrocautery and Radio Frequency (RF) Ablation
CAUTION: Electrocautery and RF ablation may induce ventricular arrhythmias and/or brillation, and may
cause asynchronous pacing, inhibition of pacing, and/or a reduction in pulse generator pacing output possibly leading to loss of capture. RF ablation may also c ause ventricular pacing up to the MTR and/or changes in pacing thresholds. Additionally, exercise caution when performing any other type of cardiac ablation procedure in patients with implanted devices.
24
Page 29
If electrocautery or RF ablation is medically necessary, observe the following to minimize risk to the patient and device:
Depending on the pacing needs of the patient, enable the Electrocautery Protection Mode, program
to an asynchronous pacing mode, or use a magnet to switch to asynchronous pacing. An option for patients with intrinsic rhythm is to program the Brady Mode to VVI at a rate below the intrinsic rate to avoid competitive pacing.
Have temporary pacing and external debrillation equipment available.
Avoid direct contact between the electrocautery equipment or ablation catheters and the pulse generator
and leads. RF ablation close to the lead electrode may damage the lead-tissue interface.
Keep the path of the electrical current as far away as possible from the pulse generator and leads.
If RF ablation and/or electrocautery is performed on tissue near the device or leads, monitor pre- and
post-measurements for sensing and pacing thresholds and impedances to determine the integrity and stability of the system.
For electrocautery, use a bipolar electrocautery system where possible and use short, intermittent, and
irregular bursts at the lowest feasible energy levels.
RF ablation equipment may cause telemetry interference between the pulse generator and PRM. If
device programming changes are necessary during an RF ablation procedure, turn off the RF ablation equipment before interrogation.
When the procedure is nished, cancel the Electrocautery Protection Mode in order to reactivate the previously programmed therapy modes.
25
Page 30
Ionizing Radiation
CAUTION: It is not possible to specify a safe radiation dosage or guarantee proper pulse generator function
following exposure to ionizing radiation. Multiple factors collectively determine the impact of radiation therapy on an implanted pulse generator, including proximity of the pulse generator to the radiation beam, type and energy level o f the radiation beam, dose rate, total dose delivered over the life of the pulse generator, and shielding of the pulse generator. The impact of ionizing radiation will also vary from one pulse generator to another and may range from no changes in function to a loss of pacing.
Sources of ionizing radiation vary signicantly in their potential impact on an implanted pulse generator. Several therapeutic radiation sources are capable of interfering with or damaging an implanted pulse generator, including those used for the treatment of cancer, such as radioactive cobalt, linear accelerators, radioactive seeds, and betatrons.
Prior to a course of therapeutic radiation treatment, the patient’s radiation oncologist and cardiologist or electrophysiologist should consider all patient management options, including increased follow-up and device replacement. Other considerations include:
Maximizing shielding of the pulse generator within the treatment eld
Determining the appropriate level of patient monitoring during treatment
Evaluate pulse generator operation during and following the course of radiation treatment to exercise as m uch device functionality as possible ("Post-Therapy Pulse Generator Follow Up" on page 20). The extent, timing, and frequency of this evaluation relative to the radiation therapy regimen are dependent upon current patient health, and therefore should be determined by the attending cardiologist or electrophysiologist.
Many pulse generator diagnostics are performed automatically once per hour, so pulse generator evaluation should not be concluded until pulse generator diagnostics have been updated and reviewed (at least one hour after radiation exposure). The effects of radiation exposure on the implanted pulse generator may remain undetected until some time following exposure. For this reason, continue to monitor pulse generator function closely and use caution when programming a feature in t he weeks or months following radiation therapy.
26
Page 31
Elevated Pressures
The International Standards Organization (ISO) has not approved a standardized pressure test for implantable pulse generators that experience hyperbaric oxygen therapy (HBOT) or SCUBA diving. However, Boston Scientic developed a test protocol to evaluate device performance upon exposure to elevated atmospheric pressures. The following summary of pressure testing should not be viewed as and is not an endorsement of HBOT or SCUBA diving.
CAUTION: Elevated pressures due to HBOT or SCUBA diving may damage the pulse generator. During laboratory testing, all pulse generators in the test sample functioned as designed when exposed to more than 1000 cycles at a pressure up to 5.0 ATA. Laboratory testing did not characterize the impact of elevated pressure on pulse generator performance or physiological response while implanted in a human body.
Pressure for each test cycle began at ambient/room pressure, increased to a high pressure level, and then returned to ambient pressure. Although dwell time (the amount of time under elevated pressure) may have an impact on human physiology, testing indicated it did not impact pulse generator performance. Pressure value equivalencies are provided below (Table 1 on page 27).
Table 1. Pressure Value Equivalencies
Pressure value equivalencies
Atmospheres Absolute 5.0 ATA
Sea water depth
a
40 m (130 ft)
Pressure, absolute 72.8 psia
Pressure, gauge
b
58.1 psig
27
Page 32
Table 1. Pressure Value Equivalencies (continued)
Pressure value equivalencies
Bar 5.0
kPa Absolute 500
a. All pressures were derived assuming sea water density of 1030 kg/m3. b. Pressure as read on a gauge or dial (psia = psig + 14.7 psi).
Prior to SCUBA diving or starting an HBOT program, the patient’s attending cardiologist or electrophysiologist should be consulted to fully understand the potential consequences relative to the patient’s specichealth condition. A Dive Medicine Specialist may also be consulted prior to SCUBA diving.
More frequent device follow-up may be warranted in conjunction with HBOT or SCUBA diving. Evaluate pulse generator operation following high pressure exposure ("Post-Therapy Pulse Generator Follow Up" on page 20). The extent, timing, and frequency of this evaluation relative to the high pressure exposure are dependent upon current patient health, and should be determined by the attending cardiologist or electrophysiologist.
If you have additional questions, or would like more detail regarding the test protocol or test results specicto HBOT or SCUBA diving, contact Boston Scientic using the information on the back cover.
POTENTIAL ADVERSE EVENTS
Based on the literature and on pulse generator and/or lead implant experience, the following alphabetical list includes the possible adverse events associated with implantation of products described in this literature:
Air embolism
Allergic reaction
Bleeding
Bradycardia
28
Page 33
Cardiac tamponade
Chronic nerve damage
Component failure
Conductor coil fracture
•Death
Electrolyte imbalance/dehydration
Elevated thresholds
•Erosion
Excessive brotic tissue growth
Extracardiac stimulation (muscle/nerve stimulation)
Fluid accumulation
Foreign body rejection phenomena
Formation of hematomas or seromas
Heart block
Inability to pace
Inappropriate pacing
Incisional pain
Incomplete lead connection with pulse generator
Infection including endocarditis
Lead dislodgment
Lead fracture
Lead insulation breakage or abrasion
Lead perforation
29
Page 34
Lead tip deformation and/or breakage
Local tissue reaction
Loss of capture
Myocardial infarction (MI)
Myocardial nec rosis
Myocardial trauma (e.g., tissue damage, valve damage)
Myopotential sensing
Oversensing/undersensing
Pacemaker-mediated tachycardia (PMT)
Pericardial rub, effusion
Pneumothorax
Pulse generator migration
Shunting current during debrillation with internal or external paddles
Syncope
Tachyarrhythmias, which include acceleration of arrhythmias and early, recurrent atrial brillation
Thrombosis/thromboemboli
Valve damage
Vasovagal response
Venous occlusion
Venous trauma (e.g., perforation, dissection, erosion)
Worsening heart failure
Patients may develop psychological intolerance to a pulse generator system and may experience the following:
Dependency
30
Page 35
•Depression
Fear of premature battery depletion
Fear of device malfunction
Additionally, potential adverse events associated with the implantation of a coronary venous lead system include:
Allergic reaction to contrast media
Breakage/failure of implant instruments
Prolonged exposure to uoroscopic radiation
Renal failure from contrast media used to visualize coronary veins
MECHANICAL SPECIFICATIONS
The following mechanical specications and material specications apply to VALITUDE devices.
All VALITUDE models have a case electrode surface area of 35.05 cm² and all VALITUDE X4 models have a case electrode surface area o f 34.58 cm². Usable battery capacity is 1.6 Ah and residual usable battery capacity at Explant is 0.10 Ah.
Mechanical specications specic to each model are listed below.
31
Page 36
Table 2. Mechanical Specications - VALITUDE CRT-Ps
Model Dimensions
WxHxD(cm)
Mass (g) Volume (cm
U125 4.45 x 6.13 x 0.75 30.6 16.2
U128 4.45 x 6.17 x 0.75 33.0 17.6
3
) Connector Type
RA:IS-1;RV:IS-1;
LV: IS -1
RA:IS-1;RV:IS-1;
LV: IS4
VALITUDE devices include ZIP telemetry operating with a transmit frequency of 402 to 405 MHz.
Material specications are shown below:
Case: hermetically sealed titanium
Header: implantation-grade polymer
Power Supply (VALITUDE): lithium-carbon monouoride cell; Boston Scientic; 402294
The following mechanical specications and material specications apply to INTUA and INVIVE devices.
All INTUA and INVIVE models have a case electrode surface area of 35.98 cm². Usable battery capacity is
1.45 Ah and residual usable battery capacity at Explant is 0.09 Ah.
Mechanical specications specic to each model are listed below.
32
Page 37
Table 3. Mechanical Specications - INTUA CRT-Ps
Model Dimensions
WxHxD(cm)
Mass (g) Volume (cm
V272 4.45 x 6.10 x 0.75 34.0 15.0
V273 4.45 x 6.10 x 0.75 34.0 15.0
3
) Connector Type
RA: IS-1, RV: IS-1,
LV: LV- 1
RA: IS-1, RV: IS-1,
LV: IS -1
Table 4. Mechanical Specications - INVIVE CRT-Ps
Model Dimensions
WxHxD(cm)
Mass (g) Volume (cm
V172 4.45 x 6.10 x 0.75 34.0 15.0
V173 4.45 x 6.10 x 0.75 34.0 15.0
3
) Connector Type
RA:IS-1,RV:IS-1,
LV: LV -1
RA:IS-1,RV:IS-1,
LV: IS -1
INTUA and INVIVE devices include ZIP telemetry operating with a transmit frequency of 916.5 MHz.
Material specications are shown below:
Case: hermetically sealed titanium
Header: implantation-grade polymer
Power Supply (INTUA and INVIVE): lithium-manganese dioxide cell; Boston Scientic; 402125
33
Page 38
ITEMS INCLUDED IN PACKAGE
The following items are included with the pulse generator:
One torque wrench
Product literature
NOTE: Accessories (e.g., wrenches) are intended for one-time use only. They should not be resterilized or reused.
SYMBOLS ON PACKAGING
The following symbols may be used on packaging and labeling (Table 5 on page 34):
Table 5. Symbols on packaging
Symbol
34
Description
Reference number
Package contents
Pulse generator
Page 39
Table 5. Symbols on packaging (continued)
Symbol
Description
Torque wrench
Literature enclosed
Serial number
Use by
Lot number
Date of manufacture
Sterilized using ethylene oxide
35
Loading...