BMS HARTIP 2000 Operation Manual

Page 1
BMS Bulut Makina Sanayi Ve Ticaret Ltd. Şti.
İkitelli Organize Sanayi Bölgesi Dolapdere Sanayi Sitesi
Ada 4 No : 7-9 Başakşehir / İSTANBUL
Tel : 0 212 671 02 24 / 25 Fax : 0 212 671 02 26
OPERATION MANUAL
KULLANMA KILAVUZU
Page 2
CONTENT
1.1. History ................................................................................................................ 2
1.2. Leeb Hardness Test (definition) .......................................................................... 2
1.3. Notation of Leeb’s Hardness .............................................................................. 2
2. Features and Applications .................................................................................... 3
2.1. Introduction ........................................................................................................ 3
2.2. Specifications ..................................................................................................... 3
2.3. Applications ........................................................................................................ 3
3. Layout and Key-pad Description .......................................................................... 4
3.1. Layout of HARTIP 2000...................................................................................... 4
3.2. Function of Key .................................................................................................. 5
3.3. Special Features of Impact Devices ................................................................... 5
4. Symbols and Illustrations ................................................................................... 5
4.1. Symbols and Illustrations ................................................................................... 5
5. Measurement and Conversion Table .................................................................. 6
6. Preparation before Measuring.............................................................................. 6
6.1. Requirements for the sample ............................................................................. 6
6.2. Samples with Curved Surfaces .......................................................................... 8
7. Operation ............................................................................................................ 8
7.1. Switch on the tester ............................................................................................ 8
7.2. Parameter Setup ................................................................................................ 9
7.2.1. Probe Setup ..................................................................................................... 10
7.2.2. Materials Selection ........................................................................................... 10
7.2.3. Hardness Scale (Conversion) .......................................................................... 10
7.2.4. Mean Time ........................................................................................................ 11
7.2.5. Storage .............................................................................................................. 11
7.2.6. Printing Setup .................................................................................................... 11
7.2.7. Impact Direction ................................................................................................ 11
7.3. Operation ......................................................................................................... 12
7.3.1. Take measurement ........................................................................................... 12
7.3.2. Load spring force.............................................................................................. 12
7.3.3. Release ............................................................................................................ 12
7.4. Data storing and re-reading .............................................................................. 13
7.5. Statistics ........................................................................................................... 13
7.5.1. View statistics value ......................................................................................... 13
7.5.2. Delete the non-realistic value ........................................................................... 13
7.6. Print-Out ........................................................................................................... 14
7.6.1. Printing stored values ....................................................................................... 14
7.6.2. Printing on line ................................................................................................. 14
7.7. Calibration (Compensation) .............................................................................. 14
7.7.1. Calibration for probe D ..................................................................................... 14
7.7.2. Calibration for probe DL ................................................................................... 15
7.8. Restore factory settings and check serial no. ................................................... 16
8. Maintenance and Repair .................................................................................... 16
8.1. Maintenance of the Impact Device ................................................................... 16
Page 3
2
Forewords
1.1. History
The Leeb measuring method was first brought into measurement technology in 1978. It is defined as the quotient of an
impact body’s rebound velocity over its impact velocity, multiplied by 1000. Harder materials produce a higher rebound
velocity than softer materials. For a specific group of material (e.g. steel, aluminum. etc.), Leeb hardness value represents a direct relationship to its hardness properties. For ordinary metal, conversion curves of hardness HL versus other standard static hardness (HB, HV, HRC, etc.) are available, enabling you to convert HL into other hardness values.
1.2. Leeb Hardness Test (definition)
An impact body with a spherical test tip made of tungsten carbide is propelled against the sample surface by a spring force and then rebounds back. At a distance of 1mm from the sample surface, the impact and rebound velocity of the impact body are measured by the following method: A permanent magnet embedded in the impact body, when passing through the coil in its coil holder, induces in the coil an electric voltage proportional to the velocities of the magnet. Leeb hardness is expressed by the following formula:
1000*
Vi
Vr
HL
Where: HL is Leeb Hardness Vr is the rebound velocity of the impact body Vi is the impact velocity of the impact body The voltage characteristic of output signal, when the impact body passes through the induction coil is illustrated in the following figure:
Voltage characteristic of output signal
A Leeb’s Hardness Tester measures the hardness of sample material in terms of Hardness Leeb (HL), which can be
converted into other Hardness units (Rockwell B and C, Vicker, Brinell and Shore D).
1.3. Notation of Leeb’s Hardness
When measuring the hardness of a sample material using the traditional static hardness testing method, a change of applied pressure will result in a change in the hardness reading. This will also happen during a Leeb’s Hardness test when
one changes the impact device. In hardness measurement of the same test sample with different impact devices, the Leeb’s
hardness values obtained will vary.
For example: 720HLD≠720HLC Because different converting curves are obtained from different impact devices, when converting hardness HL into another hardness values, the notation for the converted hardness value should include the impact device used.
For example: Hardness HV converted from hardness HL using impact device D+15 should be written as 22, 8 HV LD+15. Where: 22=Hardness value HL 8=Hardness value HV L=Leeb’s Method D+15=Impact device
Hardness HRC converted from hardness L using impact device D should be written as 35, 9 HRCLD. Where: 35=Hardness value HL 9=Hardness value HRC L=Leeb’s Method D=Impact device
Page 4
3
2. Features and Applications
2.1. Introduction
HARTIP 2000 is an innovative portable Leeb hardness tester with our new patent technology which makes HARTIP 2000 a universal impact direction hardness tester. It is no need to set up impact direction when taking measurement by any angle. Therefore, HARTIP 2000 offers a linear accuracy comparing to the angle compensating method. HARTIP 2000 is also a cost saving hardness tester and has many other features.
2.2. Specifications
Principle Leeb hardness measurement Accuracy ±0.3% @ HL=800 Repeatability ±2HL Display Digital LCD with backlight Impact direction Universal angle type Hardness scale HL/HRC/HRB/HB/HV/HS/σb Measuring range HL170-960 / HRC17-70 / HRB13-109 / HB20-655 / HV80-940 / HS32-99.5 / σb(rm)255-2639N/mm2 Impact device D(U) (External) /DC, DL, D+15, G, C(External, optional) Materials 10 common metal materials Memory 300 data can be stored and re-readable Statistics Calculated automatically Recalibration Allowed by user Indicator Low battery Communication interface RS232 to micro-printer, Bluetooth (optional) to Bluetooth micro-printer Auto power off Auto Power supply 1.5V AA alkaline battery x 2 Working environment -10ºC ~+45ºC Dimension (mm) 124x67x30 Net weight (g) 240 Standards Conforming to ASTM A956, DIN50156, GB/T 17394-1998
2.3. Applications
Hardness tests on installed machines or steel structures: e.g. on heavy and large work-piece or on permanently installed system parts. Rapid testing of multiple measuring areas for examination of hardness variations over larger regions. Measuring hardness for produced parts at production line. Identifying metallic material stored in a warehouse. Ineffectiveness analysis of permanent parts, pressure -vessel, turbo generator.
Page 5
4
3. Layout and Key-pad Description
3.1. Layout of HARTIP 2000
PROBE
Page 6
5
3.2. Function of Key
:
Read the memory
:
Power On Power Off
:
Menu Increase the value Turn the page forth
:
Change parameter Decrease the value Turn the page back
:
Delete the current reading Delete the stored values Press for 3 seconds to activate/deactivate direction indicator
:
Confirm the setup View the statistics values
3.3. Special Features of Impact Devices
Type
Brief Description
D
Universal standard unit for majority of hardness testing assignments.
DC
Extremely short impact device, other specs identical with type D.
- highly confined spaces
- holes and cylinders
- internal measurements on assembled machines
D+15 Slim front section
Application:
- grooves and recessed surfaces.
DL
Extremely slim front section
Application:
- extremely confined spaces
- base of grooves
C
Reduced impact energy ( compared with type D).
Application:
- surface hardened components, coatings
- minimum layer thickness: 0.2mm.
- thin walled or impact sensitive components (small measuring indentation).
G
Increased impact energy(approx. 9 times that of type D)
Application:
- Brinell hardness range only
- heavy cast and forged parts with lower demands on surface finish.
4. Symbols and Illustrations
4.1. Symbols and Illustrations
Symbol
Meaning
LD LG LC LD +15 LDL
Leeb hardness value obtained with impact device D Leeb hardness value obtained with impact device G Leeb hardness value obtained with impact device C Leeb hardness value obtained with impact device D+15 Leeb hardness value obtained with impact device DL
HLD HB HRB HRC HS HV
σb (N/mm2)
Leeb hardness value used with impact device D Brinell hardness value Rockwell B hardness value Rockwell C hardness value Shore hardness value Vicker hardness value Strength value
Page 7
6
5. Measurement and Conversion Table
Range for measurement and conversion:
IMPACT DEVICE D/DC
HLD: 170-960
MATERIALS
HRC
HRB
HB
HV
HS
σb (N/mm2)
Steel & cast steel
20.0-67.9
59.6-99.5
80-647
80-940
32.5-99.5
255-1710
Cold work tool steel
20.5-67.1
80-898
1170-2639
Stainless steel & High-temp. resistant steel
17-62.4
46.5-109
85-655
85-802
740-1725
Cast iron with lamellar graphite (GG)
21-59
24-100
93-334
90-698
Cast iron with nodular graphite (GGG)
21-60
24-100
131-387
96-724 Cast aluminum alloys
23-85
20-159
22-193
Copper-zinc alloys (Brass)
13-95.3
40-173
Copper-aluminum / Copper-tin alloys (Bronze)
14-100
60-290 Wrought copper alloys
14-100
45-315
Forging steel
142-651
IMPACT DEVICE DL
LDL: 560-950
MATERIALS
HRC
HRB
HB
HV
HS
σb (N/mm2)
STEEL
20.6-68.2
37.0-99.9
81-646
80-950
30.6-96.8
IMPACT DEVICE D+15
LD+15: 300-900
MATERIALS
HRC
HRB
HB
HV
HS
σb (N/mm2)
STEEL
19.3-67.9
80-638
80-937
33.3-99.3
CW. ST.
19.8-68.2
80-935
IMPACT DEVICE G L G: 300-750
MATERIALS
HRC
HRB
HB
HV
HS
σb (N/mm2)
STEEL
47.7-99.9
90-646
GC.IRON
92-326
NC.IRON
127-364
IMPACT DEVICE C
L C: 350-950
MATERIALS
HRC
HRB
HB
HV
HS
σb (N/mm2)
STEEL
20.0-70
80-683
80-996
31.9-102.3
CW. ST.
20.7-68.2
100-941
6. Preparation before Measuring
6.1. Requirements for the sample
The surface temperature of sample should be less than 120 C. The samples must feature a metallic smooth, ground surface, in order to eliminate erroneous measurements brought about by coarse grinding or lathe scoring. Roughness of the finished surface should not exceed values shown in following table:
Types of impact devices
Max surface roughness of sample Ra
D/DC/D+15/DL G C
2μm 7μm
0.4μm
Requirements for the weight of the sample For samples weighing over 5 kg and of compact shape, no support is needed. Samples weighing between 2-5 kg, and also for heavier samples with protruding parts or thin walls, should be placed on a solid support in such a manner that they do not bend or move by the impact force. Samples weighing less than 2 kg should be firmly coupled with a stable support weighing over 5 kg.
Page 8
7
For coupling purposes, The coupling surface between the sample and base plate should be flat, plane parallel and ground. A thin proper layer of coupling paste is to be applied to the contact surface of the sample. The sample should be firmly pressed against the surface of the base plate by moving it with a circular motion. The direction of impact should be perpendicular to the coupling surface.
For the coupling operation, the following prerequisites must be fulfilled: The contact surface of the sample and the surface of the base plate must be flat, plane parallel and ground. The direction of the test impact must be perpendicular to the coupled surface. Minimum thickness of the sample for coupling under various impact devices are shown in following table:
Types of impact devices
Minimum thickness
D/DC/D+15/DL
3mm
G
10mm
C
1mm
Proper Coupling: Proper coupling requires a little experience. Insufficiently coupled samples produce large variations of individual measurements, L-values which are too low and the operation is characterized by a rattling noise upon impact of the test tip. Example for coupling a test piece with a base plate:
Requirement for the surface hardened layer of the sample Surface-hardened steels, especially case-hardened steels, produce L-values which are too low when case-hardening depth is small because of their soft core .When measuring with impact devices D, D+15 or DL, depth of the hardened layer should be no less than 0.8 mm. When measuring with impact device C, the depth of the hardened layer should be no less than 0.2 mm.
Types of impact devices
Min. layer thickness for surface hardening
D/DC/D+15/DL
0.8mm
C
0.2mm
Application of the coupling paste
Mutual rubbing of both parts while firmly press the sample against the base plate.
A particular advanced of coupling is the possibility of obtaining a very uniform, rigid connection between the sample and the support, totally eliminating stresses at the sample surface. The resulting variation in measured values is very low.
Page 9
8
Surface of the test sample should not be magnetic. For test sample of curving surface with radius of curvature R less than 30mm, a small support ring should be used. Supporting the Samples during Testing
Types of impact
devices
Classification of samples
Heavy-weigt
medium-weight
light-weight
D/DC/D+15/DL
>5kg
2 - 5kg
0.05 – 2kg
G
>15 kg
5 - 15kg
0.5 – 5kg
C
>1.5kg
0.5 - 1.5kg
0.02 - 0.5kg
When measuring hardness with HARTIP 2000, the following has to be noticed: Despite the low mass of the impact body and low impact energy, a relatively large impact force within short duration is generated when the impact body hits the measuring surface.
Types of impact devices
D/DC/D+15/DL
G
C
Max. impact force
900N
2500N
500N
No particular precautions are necessary for heavy-weight samples with compact shape. Smaller and lighter samples or workpieces may yield or flex under this force, producing too-low L-values with excessively large variation. Even with big or heavy workpieces, it is possible for thin-wall regions or thinner protruding parts to yield upon impact. Depending on the frequency of the resilient yielding action, the measured L-value may be abnormally low or high. Under many situation, potential problems can be checked in the following manner: a) Medium-weight samples and also heavier samples with protruding parts or thin walls should be placed on a solid support in such a manner that they do not move or flex during the test impact. b) Light-weight samples should be rigidly “coupled” with a non-yielding support such as a heavy base plate. Clamping in a vice is of no value, since the samples become exposed to stress and because complete rigidity is never attained. As a rule, the measured L-values would be too small and show excessive variations.
6.2. Samples with Curved Surfaces
Impact testers only work properly, if the impact body has a certain position in the guide tube at the moment of impacting the test surface. In the normal position, automatically present when testing flat and convex-cylindrical samples (such as round samples), the spherical test tip is located exactly at the end of the guide tube. However, when testing spherically or cylindrically shaped concave surfaces, the impact body remains further within the guide tube or protrudes further therefore. Thus, with such types of curved surfaces, it is to be observed that radii of curvature do not drop below the values indicated in the following Fig. Curved surfaces should always be tested with the small support ring.
Impact device types D, D+15 and C Rmin =30mm Impact device type G Rmin =50mm For impact devices D, D+15 and C, special support rings are available to accommodate smaller radii on convex or concave surface.
7. Operation
7.1. Switch on the tester
Press the key to switch on the tester and press the key again to switch off the tester. When the tester is switched on, the tester will enter into measuring mode.
Page 10
9
7.2. Parameter Setup
Operation Diagram
Page 11
10
7.2.1. Probe Setup
Press key to enter the Probe setup menu.
Press key to change probe between D, DL, D15, G and C.
Press key to confirm the setting or press key again to change probe.
7.2.2. Materials Selection
The material selected is prior to the conversion from HL value to other scales.
Press key to enter into the menu until “MATE” displays on LCD.
Press key to change material from M1M2M3 ... M11 with each pressing key .
Press key to confirm the setting or press key again to change other parameters.
M1:
Steel & Cast Steel
M2:
Cold Work Tool Steel
M3:
Stainless Steel & High-temp. Resistant Steel
M4:
Cast Iron with Lamellar Graphite (GG)
M5:
Cast Iron with Nodular Graphite (GGG)
M6:
Cast Aluminum Alloys
M7:
Copper-Zinc Alloys (Brass)
M8:
Copper-Aluminum / Copper-Tin Alloys (Bronze)
M9:
Wrought Copper Alloys
M0:
Forging Steel
7.2.3. Hardness Scale (Conversion)
Hardness scale is based on the material selected. Not every material has same conversion. For example, for steel, it has conversions to HRC, HRB, HB, HV, HSD; but for cast iron, only has conversions to HB.
Press key consecutively to enter into the menu until “CONV” displays on LCD.
Press key to change the hardness scale from HLDHRCHBHVHSσb with each pressing key
.
Press key to confirm the setting or press key again to go to next item of menu.
Page 12
11
7.2.4. Mean Time
With HARTIP2000, the statistics values can be calculated automatically after setup mean time.
Press key consecutively to enter into menu until “AVER” displays on LCD.
Press key to select mean time from 0345 circularly by each pressing key .
Press to confirm the setting or press key again to go to next item of menu.
7.2.5. Storage
The HARTIP2000 has a memory capacity of 300 data. The stored values can be re-readable on LCD.
Press key consecutively to enter into menu until “SAVE” displays on LCD.
Press key to change the setting between “OFF” and “ON” alternatively.
Press key to confirm the setting or press key again to go to next item of menu. When select “ON”, a “S” appears on LCD which means the function of storage is activated.
7.2.6. Printing Setup
Press key consecutively to enter into the menu until “PORT” displays on LCD.
Press key to change the setting between “OFF”, “232” and “bLU” alternatively.
Press key to confirm the setting or press key again to go to first item of menu. When selecting
“232” or “BLU”, an indicator “ ” will appears on left side of LCD which means the communication function is activated for printing. “232” means communicating by RS232 port, “BLU” means communicating by Bluetooth.
7.2.7. Impact Direction
The instrument is designed to take measurement at any impact direction without individually setting up impact direction on the instrument during measurement. The instrument will recognize the angle of probe and make compensation automatically. But you can still review the direction arrow to make sure if the impact direction of probe recognized by instrument is right or not.
Press to switch on the tester / during measuring mode
Press to active the function to display impact direction.
Page 13
12
Take measurement at different impact direction to see if the arrow direction on LCD is compatible to your probe angle. If you find the arrow direction on LCD is not compatible to your actual probe angle, you can calibrate the instrument to make impact direction compatible. Please refer to 7.7Calibration (Compensation).
Please note: once the instrument is switched off, the function of displaying impact direction will be deactivated automatically.
7.3. Operation
7.3.1. Take measurement
Switch on the tester, the instrument will go into measuring mode automatically. If the parameters are needed to change, please refer to 7.2.
7.3.2. Load spring force
Hold the impact device with left hand while push the loading tube with right hand toward to the end. Then loose the force and let the loading tube back to original position.
7.3.3. Release
Place the impact device against the object to be measured. Then press the release button on top of the impact device with finger of right hand. The measuring value will display on LCD.
Page 14
13
Please note:During the measurement, the impact device must be placed vertically with a little force against the surface of workpiece. Otherwise, it may affect the accuracy.
7.4. Data storing and re-reading
Set the storage function of tester to be activated. Please refer to 7.2.5. A “S” displays on LCD. At this time, all measuring values will be stored automatically into the memory.
After storage is activated, press key after “MEMR” displays and then the last stored value will display.
Press key or to turn the page and view measuring values. When “EOF” or “bOF” displays with turning page, it reminds you it is at the end.
Clear the memory Delete the stored data from memory.
Under the re-readable mode, press and hold the key until “dELE” displays on LCD. Then press key to confirm the deletion.
7.5. Statistics
With this function, the statistics values will be calculated automatically by the tester. Please refer to 7.2.4 Mean Time to activate this function.
7.5.1. View statistics value
After mean time was setup in the tester, the mean time number will display on LCD. With each measurement, the
number will increase one by one until all finished. Then press key consecutively to view value of average, maximum, minimum.
7.5.2. Delete the non-realistic value
When mean time was setup, in order to avoid wrong calculation, if there are random values occurs, you should
delete the random value. Press key , the current value will be deleted.
Page 15
14
7.6. Print-Out
The tester is designed to print our measuring values by two ways. One is print-out all from memory and the other is print-out online while measuring.
7.6.1. Printing stored values
Activate the storage function on the tester, there is a “S” displays on the right side of LCD. Please refer to 7.2.5 Storage. Connect the tester to the printer with cable and then power on the printer. For Bluetooth connection, the printing cable is not necessary. (The printer and cable will be ordered optionally)
Press key , a “R” displays on LCD which means the tester is in the re-readable mode, at this time, the last value will display on LCD.
Press and hold key until “PALL” displays on LCD.
Press key to print out all data.
7.6.2. Printing on line
Activate the communication on the tester, there is a “ ” displayed on LCD. Please refer to 7.2.6 Printing Setup. Connect the tester to the printer with cable. Take measurements, all values will be printed out one by one with each measurement.
7.7. Calibration (Compensation)
The measurement compensation is used for calibration of the instrument. After the instrument is used for some time, the ball tip on impact body may be worn out or the instrument recognize wrong angle of probe which would lead inaccuracy. In order to compensate such error, the tester is designed to re-calibrate by user.
Press key in turn to display “CALD” on the LCD. Press key in turn, “ESC”, “CLR” and “CAL” will display in the lower right corner of LCD in turn.
7.7.1. Calibration for probe D
Set probe as D. Please refer to 7.2.1 Probe Setup.
When “ESC” displays, press to give up calibration and return to measuring mode.
When “CLR” displays, press to clear last calibration (compensation) value. To make sure the calibration is correct, please clear last calibration before each calibration.
When “CAL” displays, press to enter into calibration mode. An indicator “C” will display on the middle right of LCD.
Page 16
15
At the mode of calibration, you are asked to take 3 measurements (impact direction: downward) on standard test block in total. Please follow up the direction indication on LCD to take measurements.
During the measurements, you can delete the unsatisfied reading by press key . After taking 3
measurements on test block, press key to display average value. At this time, the arrow disappears and an indicator “AVE” is displayed on the right bottom of LCD.
If the average value differs from the standard value of test block, you can press and hold key for 3 seconds, “CALD” will flash once on the screen, and then an indicator “ADJ” will display on the right bottom of the
LCD, you can press key or to increase or decrease the value on LCD until the value is same as the value of test block.
Press key to finish the calibration and turn back to normal measuring mode.
7.7.2. Calibration for probe DL
Set probe as DL. Please refer to 7.2.1 Probe Setup.
Press key in turn to display “CALD” on the LCD. Press key in turn, “ESC”, “CLR” and “CAL” will
display in the lower right corner of LCD in turn. When “CAL” displays, press to enter into calibration mode. An indicator “C” will display on the middle right of LCD.
Take 3 measurements (impact direction: downward) on standard test block in total. During the measurements,
you can delete the unsatisfied reading by press key . After taking 3 measurements on test block, press key
to display average value. At this time, the arrow disappears and an indicator “AVE” is displayed on the
right bottom of LCD.
If the average value differs from the LDL value of test block, you can press and hold key for 3 seconds,
“CALD” will flash once on the screen, and then an indicator “ADJ” will display on the right bottom of the LCD, you
can press key or to increase or decrease the value on LCD until the value is same as the value of test block.
Press key to finish the calibration and turn back to normal measuring mode.
Page 17
16
Please note:When the calibration is only for impact direction instead of value, just make the same operation and don not change value on LCD. After you finish the whole operation of calibration, the impact direction will be also calibrated.
7.8. Restore factory settings and check serial no.
Restore factory settings
1. Press and hold button while turning on the tester, "ESC" will display on the LCD after release .
2. Press consecutively, the "ESC" and "FSET" will display on the LCD in turn.
3. Press key to restore factory settings when "FSET" displays on the LCD, and then the tester will return to the measuring mode. Check serial no.
1. Press and hold button while turning on the tester, serial no. will display on the LCD.
2. Press , the tester will return to the measuring mode.
8. Maintenance and Repair
Do your best to avoid shock, heavy dust, damp, strong magnetic field, and oil stain.
8.1. Maintenance of the Impact Device
The devices do not require any particular care other than periodic cleaning of the impact body and the guide tube after performing approximately 1000-2000 tests. During cleaning, the following procedures need to be observed:
Unscrew support ring and remove impact body from guide tube. Clean off any dirt and metallic dust from the impact body and the spherical test tip. Clean guide tube with the special brush provided. Do not apply oil to any parts for the impact device. Please make sure to keep the spring of impact device at releasing position, do not let the spring pressed by locking impact body after working and being storage.
Loading...