BMA Automation FCCBMA2000, FCCBMA1000 User Manual

Concentration Meter
DynFAS MW
Hardware Manual
User's Guide
ID No. 54877BA2
Rev. No. 00 01.07.2009
The units supplied should not be repaired by anyone other than BMA Schaltanlagen Service engineers or technicians by BMA Schaltanlagen.
In case of operation trouble, please address to our central service department.
The complete user’s guide consists of two parts, namely the hardware and software instruc­tions.
The hardware manual comprises:
¾ mechanical components ¾ installation ¾ electrical installation ¾ technical data ¾ electrical and mechanical drawings
The software manual comprises:
¾ operation of the control unit ¾ parameter description ¾ basic setting ¾ calibration ¾ error messages
The present part is the hardware instruction.
Subject to change without prior notice.
BMA Schaltanlagen GmbH Am Alten Bahnhof 5 38122 Braunschweig Germany
Tel. +49 531 804 261 Fax +49 531 804 269 schaltanlagen@bma-de.com www.bma-worldwide.com
DynFAS MW
Table of Contents
Table of Contents
Page
Chapter 1. Safety Instructions 7
1.1 Identification and warning notices 7
1.2 General Instructions 8
1.3 General Safety Instructions 9
Chapter 2. General Information 11
2.1 Use and Function 11
2.2 Frequency approval 12
2.3 Intended Use 14
2.4 Definitions 15
Chapter 3. System Description 16
3.1 Principle of Measurement 16
3.2 Calculation of Measured Values 17
3.3 Temperature Compensation 18
3.4 Mechanical Components 19
3.4.1 Control Units 20
3.4.2 Flow Cell 23
3.4.3 Container Probe 24
3.4.4 High-Frequency Cable 28
3.5 Pipeline Measurement Configuration 30
3.6 Container Measurement Configuration 31
Chapter 4. Getting Started 32
4.1 Transport 32
4.2 Installation 32
4.2.1 Flow Cell Installation 32
4.2.2 Container Probe Installation 34
4.2.3 Installing the Control Unit 36
4.3 Connecting the Control Unit 37
4.3.1 Connecting the HF Cable 37
4.3.2 Pin Configuration of the Connector Strip 40
4.3.3 Digital Outputs, Relays 44
Chapter 5. Service Instructions 45
5.1 General Information 45
5.2 Wearing Parts 45
5.3 Instrument Cleaning 45
5.4 Battery 46
5.5 Fuse Replacement 46
Chapter 6. Technical Data 47
6.1 Control Unit 47
6.2 Technical Data Sensors 50
6.3 Technical Data HF-Cable 52
6.4 Format of Serial Data Output RS 232 and RS 485 53
DynFAS MW
5
Chapter 7. Certificates 55
7.1 EC Declaration of Conformity 55
7.2 Frequency Approval 57
Chapter 8. Technical Drawings 63
8.1 Dimensional Drawing of Control Unit Hous ing 63
8.1.1 Control Unit Standard CS-100 63
8.1.2 Control Unit High Dynamic CH-200 64
8.2 Electrical Wiring Diagram 65
8.3 Dimensional Drawings Flow Cells 66
8.3.1 Type FC-050-016 66
8.3.2 Type FC-065-040 67
8.3.3 Type FC-080-016 68
8.3.4 Type FC-100-016 69
8.3.5 Type FC-150-016 70
8.3.6 Type FC-020-150 71
8.3.7 Type FC-025-300 72
8.3.8 Type FC-030-150 73
8.3.9 Type FC-040-150 74
8.3.10 Type FC-060-150 75
8.4 Dimensional Drawings Probes 76
8.4.1 Type P-065-006 76
8.4.2 Type P-080-016 77
8.4.3 Type P-100-016 78
8.4.4 Type P-150-016 79
8.4.5 Type P-025-150 80
8.4.6 Type P-030-150 81
8.4.7 Installation Situation in Pipelines 82
8.5 Dimensional Drawings Flush Probes 83
8.5.1 Type PF-065-006 83
8.5.2 Type PF-080-016 84
8.5.3 Type PF-100-016 85
8.5.4 Type PF-150-016 86
8.5.5 Type PF-025-150 87
8.5.6 Installation Situation in Pipelines 88
8.6 Installation Sheets for DynFAS MW (Probe) 89
8.7 Installation Sheets for DynFAS MW (Flush Probe) 91
6
DynFAS MW
Chapter 1 Safety Instructions
Chapter 1. Safety Instructions
1.1 Identification and warning notices
The term BMA Schaltanlagen in this User’s Manual stands for the company BMA Schaltanlagen GmbH.
Please observe the warnings and safety instructions given in this User’s Manual to rule out physical injury and property damage. They are identified by the following symbols: DANGER, WARNING, CAUTION or NOTICE.
Indicates an imminently dangerous condition. Failure to follow the instructions will lead to death or serious injury.
Indicates a potentially dangerous condition. Failure to follow the instructions may lead to death or serious injury.
Indicates a potentially dangerous condition. Failure to follow the instructions may lead to slight injury or a medium-degree inju ry.
Indicates a situation which may cause property damage if the in­structions are not followed.
IMPORTANT
Paragraphs with this symbol provide important information on the product or how to work with the product.
DynFAS MW
Tip
Includes application tips and other useful information.
Further Symbols
Warning sign: no intervention, change nothing
7
Chapter 1 Safety Instructions
Instruction: Disconnect from mains supply
Instruction: Wear safety shoes
1.2 General Instructions
The most important safety measures are summarised in this op­eration manual. It supplements the applicable regulations which have to be studied by the personnel in charge.
Please keep in mind:
¾ the national safety and accident prevention regulations ¾ the national installation regulations (e.g. EN 60079) ¾ the generally accepted engineering rules ¾ the information on transportation, installation, operation, ser-
vice and maintenance
¾ the safety instructions and information in this User’s Manual
and the enclosed technical drawings and wiring diagrams
¾ the characteristic data, limit values and the information on the
operating and environmental conditions on the type labels an d data sheets
¾ the signs on the devices
8
DynFAS MW
Chapter 1 Safety Instructions
1.3 General Safety Instructions
IMPORTANT
The instrument housings are protected according to protection type IP 65 and are suitable for outdoor application. The instru­ment has been tested by the manufacturer and is delivered in a condition that allows safe and reliable operation.
The measuring systems have to be protected against direct sun rays and rain during outdoor applications e. g. by a suitable pro­tective cover.
IMPORTANT
Never change the installation and the parameter settings without a full knowledge of these operating instruction s, as well as a full knowledge of the behavior of the connected controller and the possible influence on the operating process to be controlled.
Ambient conditions
The systems may be used only in technically good order and only according to regulations!
Only authorized persons who have been trained, have the proper qualification and have received the necessary instructions may work with the systems! Installations and modifications on the sys­tems which may affect the operational safety are not permitted!
IMPORTANT
All systems components require non corrosive ambient conditions during transport, storage and starting up.
Electrical shock hazard: Disconnect power to rule out any contact with live parts during
installation and when servicing. Turn off power supply before opening the instrument. NEVER work
on open and live instruments.
DynFAS MW
9
Chapter 1 Safety Instructions
Attention! Possible danger, damage to property! Concerns the sys­tem type CH-200-024 Control Unit High Dynamic 24 V DC (Id.-No. 54878-02): When connecting the 24 V DC auxiliary power, the + and – Poles should be connected correctly. There is no reverse voltage protec­tion!
Qualified Persons
Spare fuses must match the rating specified by the device manu­facturer. Short-circuiting or manipulation is not permitted.
IMPORTANT
The DynFAS MW and all ancillary units have to be connected to mains via grounded connection.
IMPORTANT
The concentration measuring system DynFAS MW is to be serviced and repaired solely by qualified persons.
Persons are qualified if they have acquired adequ ate knowledge in the area concerned in the course of their professional education, and if they are familiar with the pertinent national occupational safety regulations, accident prevention regulations, directives and acknowledged rules of technology. They must be capable of as­sessing the result of their work safely; moreover, they need to be familiar with the contents of this User’s Manual.
10
IMPORTANT
If liquid gets inside the instrument, cut off the power supply. The instrument has to be checked and cleaned by an authorized ser­vice center.
DynFAS MW
Chapter 2 General Information
Chapter 2. General Information
2.1 Use and Function
The DynFAS MW has been designed as a concentration measuring system and may only be used for this purpose. If it is used in any manner, which is not described in this user’s manual, the protec­tion of the device is impaired and all warranty claims are void.
BMA Schaltanlagen only guarantees that the devices comply with the published specifications. The DynFAS MW may only be inst alled in an undamaged, dry and clean condition. Alteration work and modification of the system components are not permitted.
The DynFAS MW does not qualify as a “safety regulated measure­ments.
Conformance with standards
Protection type
Misuse warning
The standards and regulations customary to the DynFAS MW are listed in the system instructions in chapter 2.2 f requency approval and in chapter 7.1 EC-conformity certificate.
The degree of protection of the DynFAS MW according to IEC 60529 is max. IP 65.
The following is contrary to the intended use and, therefore, has to be prevented:
¾ Use under other conditions and prerequisites than those speci-
fied by the manufacturer in the technical documents, data sheets, operating and installation instructions and in other spe­cifications.
¾ Use after repair by persons who have not been authorized by
BMA Schaltanlagen.
¾ Use in a damaged or corroded state. ¾ Operation with open or inadequately closed cover. ¾ Operation with inadequately tightened adapters and screwed
cable glands.
¾ Operation without observing the safety precautions defined by
the manufacturer.
Authorized Persons
DynFAS MW
¾ Tampering with or bypassing existing saf et y in stallations. Authorized persons are persons, who are foreseen for certain ac-
tivities as a consequence of statutory provisions, or who have been approved by BMA Schaltanlagen for carrying out certain activities.
11
Chapter 2 General Information
2.2 Frequency approval
The DynFAS MW complies with part 15 of the FCC1 Rules. These devices fulfill the requirements regarding immunit y to interference and emitted interference and are licensed for operation.
FCC­Approval plates
12
1
FCC ... Federal Communications Commission
DynFAS MW
Chapter 2 General Information
The DynFAS MW complies with the R&TTE regulations 1999/5/EG and fulfill herein all requirements for this type of high-frequency device. The devices bear the identification of conformity according to the CE symbol, No. 0682 of the certification office. The certifi­cate can be found in chapter 7.2 Frequency Approval.
The DynFAS MW is a system for concentration measurement using microwave technology. The emitted microwaves have a very low activity and are, therefore, not at all hazardous to human beings or the environment. Also, the product is not affected at all by the microwaves.
IMPORTANT
The DynFAS MW has been manufactured in compliance with the safety requirements for microwave devices. If special legal provi­sions exist regarding the use of microwaves, it will be the respon­sibility of the user to adhere to them.
IMPORTANT
Any change in frequency or any other manipulation on the micro­wave device will result in a loss of the frequency approval and w ill be prosecuted.
The microwave modules do not include any replaceable compo­nents and must not be opened.
DynFAS MW
13
Chapter 2 General Information
2.3 Intended Use
The measuring system DynFAS MW can be used to determine the concentration of nearly all materials which can be dissolved or sus­pended in water using microwave technology. The following sensor and control unit versions are available:
1. The container probes have been designed for installation into pipelines with a nominal width of 200 mm and in containers, for example, crystallizers. The probe is installed such that both measuring rods (transmitter and receiver) are immersed into the product being measured.
2. The Flow Cell is a tubular probe, with microwave transmitter and receiver being firmly welded onto the outside of t he pipe. The inside of the pipe is Teflon-coated. The flow cell is in­stalled into the existing pipeline system inline or into a bypass.
The control unit is available in two versions: The Standard Model CS-100 and the high dynamic version CH-200. The control unit CH-200 can only be used when sufficiently large microwave at­tenuation (min. 40 dB) is present. The Standard Model CS-100 should be used for lower microwave attenuation.
During operation, the concentration measuring device DynFAS MW send out electromagnetic radiation in the frequency range between
2.4 GHz and 2.5 GHz (range restrictions depending on local regu-
lations in your country). The microwaves which emerge are not dangerous to human beings and the environment (power emission < 10 mW). The microwaves are emitted from the microwave win­dow; the product is not changed by the microwaves.
To ensure proper function of the meter, please pay attention to the following:
14
Tip
¾ The material being measured must not be electrically conduc-
tive, i.e. the ohmic resistance is infinite.
¾ The product must not contain any gas bubbles, or gas bubbles
have to be compressed with adequate pressure when carrying out measurements in pipelines.
¾ The ion concentration, e.g. salt content, has to be nearly con-
stant.
¾ The total attenuation of microwave signals must be at least 40
dB for the control unit CH-200. For details, see Chapter 3.4.1 the control units
DynFAS MW
Chapter 2 General Information
2.4 Definitions
Attenuation Weakening of microwave signals, microwave measurement
effect.
Container flush probe
Factory setting All parameters have been set to standard values by the manu-
Flow cell Tubular probe for simple integration into the ex isting pipeline
HF cable High-frequency cable.
Microwaves Electromagnetic waves in a certain frequency range.
Phase Phase or phase shift. Microwave measurement effect.
Quad cable Combination of four HF cables of equal length in a corrugated
Softkeys Buttons associated with the software.
Container probe with flushing device.
facturer. In most cases this simplifies calibration of the device significantly. Despite factory setting, calibration always has to be performed.
system.
tube.
TC Temperature compensation.
DynFAS MW
15
Chapter 3 System Description
r
r
Chapter 3. System Description
3.1 Principle of Measurement
The microwaves that spread between the rods pass through the product being measured; their propagation speed is slowed down (= phase shift) and their intensity is damped (= attenuation). Figure 3-1 illustrates the principle of measurement: the propaga­tion speed of microwaves passing through the product being measured is slowed down (phase shift) and their intensity (at­tenuation) is reduced, relative to a reference signal.
HF- sources
Figure 3-1:
Schematic diagram:
Change of microwave
by product
Transmitte
Reference signal
Reference path
Phase comparison
--> Phase
Product
Receive
Amplitude comp
--> Attenuation
Measured value: Concentration %TS
Measurement signal
Phase shift
Prerequisite is that the product being measured shows some dielectric properties. In general, water is a very distinct dielec­tric fluid. The water or dry mass concentration, respectively, can therefore be determined by measuring the phase shift and/or attenuation.
The concentration to be detected in the product is therefore de­pendent in good approximation linear on phase shift and at­tenuation. For this reason we can measure the concentration or the Brix content of the product using a linear calibration (see chapter 3.2 Calculation of Measured Values ).
16
DynFAS MW
Chapter 3 System Description
+ϕ⋅
3.2 Calculation of Measured Values
The microwave measuring phase and attenuation are calibrated after an automatic plausibility analysis.
During calibration, the phase and/or the attenuation or a concen­tration value (or density value) are assigned by sampling. The calibration is full automatic and the sample taking is supported by the control unit.
Which of the parameters, either phase, attenuation or both are used for the calibration depends on the size and interference of the measuring effect. For example, the attenuat ion is significantly more sensitive to electrolytic conductivity (salt content).
In many cases, the mere phase measurement is recommended and is calculated in good approximation by a linear calibration as follows:
=
Con concentration A, C coefficients of respective calibration function ϕ phase
The DynFAS MW allows you to calibrate, display and output two concentrations Con1 and Con2. You have to enter the calibration coefficients separately for concentration 1 and 2. For more infor­mation please refer to the Software Manual.
CACon
DynFAS MW
17
Chapter 3 System Description
θΔ⋅
+
θΔ⋅
+
3.3 Temperature Compensation
Temperature compensation (TC) is necessary if the product tem­perature varies. In general, we recommend connecting a tempera­ture compensation, i.e. a temperature signal (0/4...20 mA or Pt 100) to the control unit and, if necessary, to enable the com­pensation in the control unit. The control unit is designed such that the required TC’s can be calculated automatically. The variat ion in temperature where temperature compensation becomes absolutely essential is dependent on the product and on the water content. In first approximation, ± 2°C should be set as fluctuation limit.
Tip
TC has to be carried out whenever you are working with.
The TC corrects the phase and attenuation before the measured value calculation (calibration), in most applications according to the following formulae (linear compensation, additive).
ϕ=ϕ
=
C
meascomp
CDD
ϕ
Dmeascomp
Where
= measured phase
ϕ
meas
ϕ
= compensated phase
comp
= measured attenuation
D
meas
= compensated attenuation
D
meas
= temperature coefficient
C
ϕ
= temperature coefficient
C
D
Δθ = measured temperature (T
) – reference temp. (T
meas
Ref
)
Depending on the selected function (additive, multiplicative, lin ear, quadratic), the required temperature coefficients appear on the Calibration menu. Temperature coefficients that are not used are set to zero.
If you select two-range calibration (split concentration), separate TC’s have to be entered for both concentration ranges. The coeffi­cients are entered in the course of calibration.
TC can be carried out via Pt 100 or via current input. This has to be defined on the Calibration menu. The Pt 100 temperature range is between –50°C and +200°C.
How to work with the temperature compensation is described in detail in the Software Manual.
18
DynFAS MW
Chapter 3 System Description
Figure 3-2:
Control Unit Standard
CS-100
3.4 Mechanical Components
The measurement system consists of a control unit, a probe and a set of special high frequency cables (in short HF-cable). The con­trol unit is available in two versions: the standard model CS-100 and the high dynamic version CH-200, see Figure 3-2 and 3-3.
DynFAS MW
Figure 3-3:
Control Unit
High Dynamic
CH-200
19
Chapter 3 System Description
The probes are available in different versions, as pipeline and con­tainer probe with and without flushing device (see Figure 3-4, 3-5 and 3-6).
Abb. 3-4 left:
Probe
Abb. 3-5 right:
Probe with Flushing
Figure 3-6:
nominal width 50 mm
Flow Cell
3.4.1 Control Units
The control units consist of evaluation analyser with microwave unit. The microwaves are generated, received and analyzed by the microwave unit. Signal processing and communication take place in the evaluation computer. For simple operation, the measuring system includes a display, 4 softkeys and an alphanumeric keypad. Different functions are assigned to the softkeys on the display.
20
DynFAS MW
Chapter 3 System Description
High Dynamic Version
CH-200
Differences between Control Unit Standard CS-100 and Control Unit High Dynamic CH-200
The Control Unit High Dynamic CH-200, has an additional HF am­plifier module in comparison to the standard model, whereby the wall housing is larger (dimensions see chapter 6.2 Technical Data control unit). Otherwise, the control units only differ in their appli­cations. Higher product attenuations are allowed for the high dynamic ver­sion of DynFAS MW. Therefore larger measuring paths can be irra­diated, for example measuring cells of larger nominal width can be used. The application of both control units is predetermined by the product attenuation. Up to an attenuation of 50 dB, CS-100 is used and beyond, CH-200. The CH-200 generally requires an attenua­tion of 40 dB. If this is lower, the software indicates an err or mes­sage.
An RS232 interface is included on the underside of the instrument.
Figure 3-7:
Front view of
CS-100
LCD display
Lock
Numerical
keypad
Cable feed-through
M 20 and M 16
HF connections for
signal cable reference cable
and cable feed-through
M-TxM-Rx R-Tx
R-Rx
Softkey buttons
LED’s
RS232 connection
RS232 9-pole Sub-D-connector
Figure 3-8:
Control Unit -
bottom view
DynFAS MW
High-frequency connections
21
Chapter 3 System Description
LED’s on the Front Panel
Five LED’s on the instrument front panel indicate the instrument status.
Figure 3-9:
LED’s on the front panel
of the Control Unit
LED Function RUN Instrument in measurement mode
Display is flashing when concentration average value is put on hold, e.g. if an error has occurred, if the measurement has been paused or stopped
ERR Error
Goes out after reset or if fault has been repaired
SIG 1 Display depending on the selected function of relay
1, possible functions: error, no product, limit value min., limit value max., measurement stopped
SIG 2 Display depending on the selected function of relay
2, possible functions: error, no product, limit value min., limit value max., measurement stopped.
COM Communication active, e.g. via RS 232
Terminal Block
The electrical connections of the DynFAS MW are located on a con­nector strip in the wall housing. The terminal block is accessible from the front after you have opened the cover. There, you also find the power cut-off switch and the fuses. The high-frequency connections are located on the outside of the housing. All other elements, especially the live elements (on the motherboard) are provided with a protection cap.
22
DynFAS MW
Chapter 3 System Description
3.4.2 Flow Cell
The flow cells are available with nominal widths from 50 to 150 mm (see Figure 3-10) and different flanges. For technical data please refer to chapter 6.2.
Figure 3-10:
Flow Cell
B
Pipeline with nominal width 50 ... 150 mm
B
A
A: High-frequency connections B: Process connection, flanges of different sizes
The flow cell consists of a sturdy stainless steel body. The micro­wave transmitter and receiver are firmly welded to the outside of the pipe. The entire product pipe is PTFE-coated and fulfills the special requirements for use in foodstuffs.
There are not objects extending into the pipe (such as measuring sensors). The flow cell can be installed in the pipeline via flange.
A
DynFAS MW
The flow cell has two HF connections to feed in and output micro­wave signals. Input and output can be allocated as needed (M-Tx, M-Rx). The microwave signals transmit the product over the entire pipeline cross-section.
23
Chapter 3 System Description
A
3.4.3 Container Probe
Two different container probe versions are available – either with or without flushing device (see Figure 3-11). For technical data please refer to chapter 6.2 Technical Data.
Probe
with flushing device
Probe
without flushing device
Figure 3-11:
Container probes
B
A
A: High-frequency connections B: Process connection, flanges of different sizes
The container probe has been specially designed for concentration measurements in containers. Both measuring rods are immersed into the product. Microwaves are emitted from one end of the rod and received by the other end of the rod; they are emitted only towards the opposite end of the rod. This direction characteristic of the probe minimizes the interfering influence of metal parts in the vicinity of the probe and allows installation if only little space is available. For example, the concentration of sugar strike can be measured continuously to find the suitable inoculation time.
The plastic rods meet the special requirements for application in foodstuffs.
24
DynFAS MW
Chapter 3 System Description
Two different probe types are available: ¾ The standard type is the container probe without flushing de-
vice
¾ The probe with flushing is employed in processes where incrus-
tations are likely to occur, for example, due to increased depo­sitions. The flushing device prevents any deposition on the mi­crowave exit windows. Long travel times are supported by con­tinuous crystal processes (i.e. VKT).
The flow direction of the product being measured should be verti­cal, as shown in Figure 3-12. This ensures that the product be­tween the measuring rods is representative, provided it is mixed thoroughly.
Product
Microwave measur in g f ield
Plastic rods
~ 190 mm
100 mm
Pt100
Container wall
Figure 3-12:
Probe
HF conne ct ions
Pt 100 Only the probe without flushing is provided with a Pt 100 and is
connected to the control unit via 4-wire cable. The wiring diagram for the Pt 100 is described in chapter 4.3.2 Pin Configuration of the Connector Strip. To reduce the danger of incrustation in the imme­diate vicinity of the measuring rods, the probe with flushing is not provided with a Pt 100.
DynFAS MW
25
Chapter 3 System Description
Warning, possible property damages! Do not open the cover screws on the front of the container probes, see Figure 3-13.
Probe flange
e.g. Probe DN65 / PN6
Figure 3-13:
The front of the probe
Flow
4
1
Strömung
ø
R-Tx
M-Rx
ø
1
30
R-Rx
M-Tx
do not remove!
PT100
ø
16
0
Probe with Flushing
The probe with flushing device has been designed for processes where depositions, for example, due to incrustations are likely to occur on the probe.
26
The flush probe has two flushing channels which keep the plastic rod free from incrustations; this ensures that the microwaves come into direct contact with the product being measured. All parts coming into contact with the product meet the specific re­quirements for application in foodstuffs. Figure 3-14 shows the probe design.
DynFAS MW
Chapter 3 System Description
Microwave measuring field
Plastic rods
Rinsing medium
~ 320 mm
100 mm
Container wall
Figure 3-14:
Probe with Flushing
Figure 3-15:
Rod head
with flushing pipe
4 x HF connections
2 x 3/8’’ Flushing connections, internal thread
The flushing slit width is the same for both probe rods and shown in Figure 3-15.
Flushing pipe
Plastic rods
F
lush
ng slit
i
.5
m
m
~ 1
DynFAS MW
27
Chapter 3 System Description
3.4.4 High-Frequency Cable
High-frequency cables (HF cable) are used to transmit microwaves between probe and evaluation electronics.
HF cables change their conductivity (for microwaves) relative to the temperature. Therefore, variations in the ambient temperature would create measurement errors. This error is compensated for by enabling the cable compensation. Influences of the ambient temperature on the signal cable are compensated for by means of the reference cable. The reference cable has the same length as the signal cable; during operation, it should be exposed to the same ambient temperature. Therefore, we recommend installing both cable types together in a corrugated tube; this also simplifies installation.
The HF cable quad (see Figure 3-16) consists of four individual HF cables of the same length, which each end with one HF plug con­nector (N-type). Available cable lengths: 2, 4, 6 and 10 m.
When connecting the Flow Cell, the reference cable is short­circuited to the probe side by means of N-connectors (see Figure 3-17).
Figure 3-16:
HF-cable quad
Never bend HF cables! The bending radius should not be less than 100 mm. After installation, fix cables with cable binders.
4 x N-connector
4 x N-
0.35 m
ø
18.5 mm
Corrugated tube
Lengths 2, 4, 6 and 10 m
4 x N-connector
0.35 m
28
DynFAS MW
Chapter 3 System Description
Figure 3-17:
HF-cable quad, at the
side of the probe
Sensor side
x
R
-
M
x
R
-
R
N-connector CC-100
R
-
T
x
M
-
T
x
Figure 3-17: The ends of the reference cable R-Rx and R-Tx are short-circuited with an N-connector.
For further technical data see chapter 6.3 Technical Data HF­Cable.
DynFAS MW
29
Chapter 3 System Description
3.5 Pipeline Measurement Configuration
The control unit is installed in the immediate vicinity of the con­tainer probe to keep the HF-cable between control unit and probe fairly short. The shorter the cable connection, the better the stabil­ity of the measurement. The standard length is 4 m and the maxi­mum length of the HF-cables is 10 m.
The flow cell is installed into the exist ing pipeline system inline or into a bypass. The orientation of the flow cell may either be verti­cal or horizontal. To rule out sedimentary depositions, vertical in­stallation in a riser is preferred (see Figure 3-18).
The flow cell should be installed fairly close to the sampling loca­tion to ensure representative sampling for calibration.
For possibly required product temperature compensation, a repre­sentative temperature signal (current signal or Pt 100) has to be connected to the control unit.
Figure 3-18:
Typical measurement
configuration
on a pipeline
Control Unit
Sampling
Distance: typical 4 m
HF-cable quad
Reference line with N-connector CC-100
Product flow
Orientation: HF-connections facing down
Pipeline inline or in bypass
30
DynFAS MW
Chapter 3 System Description
3.6 Container Measurement Configuration
The control unit is installed in the immediate vicinity of the probe to keep the HF-cable between control unit and probe fairly short. The shorter the cable connection, the better the stability of the measurement. The standard length is 2 or 4 m and the maximum length of the HF-cables is 10 m.
The control unit should be installed fairly close to the sampling location to ensure representative sampling for calibration. A repre­sentative temperature signal (current signal or Pt 100) should be connected to the control unit for possibly required product tem­perature compensation.
Our example below shows the measurement configuration on a discontinuous evaporation crystallizer. The probe is fixed to the container wall such that both measuring rods are immersed into the product.
Figure 3-19:
Typical system
configuration on a
evaporation
crystallizer
Distance typical 4 m
(no radiation heat)
a
a > 60 mm
Container wall
PT-100 HF quad cable
Product
HF connections
Microwave measuring field
Pt100
DynFAS MW
31
Chapter 4 Getting Started
Chapter 4. Getting Started
4.1 Transport
IMPORTANT
Risk of damage! System parts may get damaged during transportation!
Transport probe and control unit in their original packaging. Pro­tect parts against shocks. Especially the plastic rods of the con­tainer probes have to be protected against mechanical impact!
After unpacking, make sure all parts listed on the packing list have been delivered and show no sign of damage; if necessary, clean these parts.
If you detect any damage, please notify the forwarder and the manufacturer immediately.
The weight of the system components can amount to more than 25 kg depending on the model. You should wear safety shoes.
4.2 Installation
4.2.1 Flow Cell Installation
For installation of the flow cell please keep in mind: ¾ The flow cell is installed into the pipelin e system. Keep in mind
that material sampling should be possible for calibration di­rectly behind the flow cell.
¾ The flow cell should be installed in a vertical riser, if possible. It
has to be ensured that no material depositions occur on the pi­pe walls and no bubbles are present in the product. For hori­zontal installation, please observe the correct orientation of the HF-connections (see Figure 4-2).
32
¾ There should be a straight pipe section of at least 200 mm and
equal nominal width before and after the flow cell to ensure a fairly homogeneous flow profile and to rule out possibly occur­ring microwave reflections in the pipeline.
DynFAS MW
Chapter 4 Getting Started
¾ No gas bubbles should be present in the product. If gas bub-
bles cannot be ruled out, a pressure of at least 4 bar is re­quired in the pipeline to minimize the influence of gas bubbles. Please observe the max. permissible working pressure (see chapter 6.2 Technical Data Sensors)
¾ The high-frequency cable should preferably be connected to
the flow cell from below to prevent inflowing water from getting to the connecting sockets.
¾ The HF cable should not come into contact with the warm pipe-
lines.
Figure 4-1:
Installation in a
vertical riser
HF-connections
min. 200 mm
Flow cell
Orientation:
facing down
M-Rx
( M-Tx )
Wrong arrangement !
Sampling
Product flow
Flange 200 ... 250 Nm
HF-connections max. 2 Nm
M-Tx
( M-Rx )
Pipeline inline or in bypass
Correct arrangeme nt !
Figure 4-2:
Horizontal installation:
orientation of
HF-connections
DynFAS MW
33
Chapter 4 Getting Started
4.2.2 Container Probe Installation
For installation, please keep in mind: ¾ Select the installation site such that good mixing and a homo-
geneous product are ensured and no bubbles are present in the probe. A tap should be provided in the direct vicinity to allow representative sampling.
¾ The probe has to be flange-mounted on the container such that
the product being measured flows between both measuring rods. That means the fork (both measuring rods) has to be in­stalled at an angle of 90° to the material flow.
¾ The distance between the measuring rod tips and any metal-
ized walls (heating elements, stirrer, container wall) should be at least 60 mm.
¾ The following installation hole sizes in the fitting flange
are required for installation of the probe:
Flange DN 65 / PN 6 others
¾ For further installation dimensions please refer to chapter 8
(see installation sheets).
¾ Use the respective flat gasket (standard accessory) to compen-
sate for minor surface tolerances in the fitting flange.
Minimum installation hole size (mm) 100 ± 0.2 102 ± 0.5
Installation on Process Containers
Figure 3-19 shows the position of the container probe on the con­tainer. This position is also valid for the container flus h probe.
The assembly sheet in chapter 8 includes all the inf ormation re­quired for installation.
Installation in Pipelines
The container probes can be installed in pipelines with a nominal width 200 mm using an adapter flange. Please observe the posi­tion and orientation of the container probe (see the technical drawings in chapters 8.4.7 and 8.5.6 Installation Situation in Pipe­lines).
34
DynFAS MW
Chapter 4 Getting Started
Connection of the flushing pipes
The container probe with flushing device consists of two flushing devices with a 3/8 inch inner thread (DIN ISO 228-1). The flushing connections are subsequently sealed to this thread. A sealing to the probe cover is not permitted, for example with silicone.
Flush Parameters (only for container flush probe)
The degree of deposition or incrustation is essential for the flush parameters, i.e. flush frequency and duration. The flush parame­ters have to be adapted to the product and the process.
The following independent flushing parameters for products and processes have to be observed:
Flush solution water, condensation Temperature of flush solution Pressure Fittings 2 x 3/8 inch female screw thread
Supply pipe
Independent flushing parameters product and process, typical starting rates:
Interval every 2 hours Duration 12 seconds Temperature of flush solution
For measurements on the C-product the flushing intervals can be considerably reduced, e.g. every 6 hours for 30 seconds.
Maximum 120 °C
3 bar, max. 8 bar
(DIN ISO 228-1) 1/2 inch
average product temperature, mostly 65 ±5°C
Amount of water
DynFAS MW
The following is generally valid: the flushing devices can be flushed simultaneously or in shifts. The flushing parameters are valid for every flushing device.
Tip
The required flush duration has to take into account a possible inertness of the system, e.g. valve openings. The flush supply pipes have to be insulated well against heat to prevent that the flush solution is initially colder.
The amount of water per flushing connector is approx. 0.8 l/sec at a flushing pressure of 5 bars.
35
Chapter 4 Getting Started
4.2.3 Installing the Control Unit
For installation of the control unit, please keep in mind: ¾ Install the control unit in the vicinity of the microwave probe,
keeping in mind the length of the HF cable. HF cables are available in a length of 2, 4, 6 and 10 m; the standard cable length are 2 or 4 m.
¾ Protect the instrument against vibrations. ¾ For instrument installation you should foresee a cutoff device to
allow easy and quick disconnection of the device from the power supply.
¾ When installing the control unit on a crystallizer, use a distance
rail to minimize thermal radiation and heat conduction. See Figure 4-3.
¾ When the control unit is set-up outdoors, it has to be protected
from direct sunshine and rain for example by means of an ade­quately large protective roof.
Figure 4-3:
View from above:
Installation of
control unit
on a crystallizer
Insulation container wall
Spacer
Control Unit
36
DynFAS MW
Chapter 4 Getting Started
4.3 Connecting the Control Unit
4.3.1 Connecting the HF Cable
For the connection of the sensor with the control unit, you will need a HF cable quad. Additionally, the flow cell requires an N­connector.
Prerequisite for a proper measurement is the correct installation of cables! Please keep in mind:
IMPORTANT
Make sure the cables do not get into contact with hot pipes over the entire length (corrugated tube and single cable section after splitting), e.g. direct contact with the device wall (not insulated). This alone guarantees that all single cables are subject to the sa­me ambient conditions and that the compensation of the cable drift works properly.
Never bend HF cables! The bending radius should not be less than 100 mm. After installation, fix the cables with cable binders to prevent the cable from slipping!
Connecting the Flow Cell
The HF-cable quad and the HF-connections on the control unit are labeled. Connect the flow cell to the control unit as shown in Figure 4-4 and make sure that you only connect cables with equal label­ing. The two connections on the flow cell are not labeled, the allo­cation of the cable connectors M-Tx and M-Rx is arbitrary. The ca­ble plugs R-Tx and R-Mx are connector to the N-connector (short­circuited).
DynFAS MW
37
Chapter 4 Getting Started
Control Unit, underside
Figure 4-4:
Connection of flow cell
version 1
M-Rx
M-Rx
Bending radius min. 100 mm
M-Tx
M-Tx
R-Tx
R-Rx
R-Rx
R-Tx
HF-cable quad (stand ard 4 m)
Flow cell
Reference line with N-connector
Flange 200 ... 250 Nm
HF-connections max. 2 Nm
M-Rx
( M-Tx )
R-Rx R-Tx
M-Tx
( M-Rx )
Connecting the Container Probes
The HF cables and the HF connections on the control unit and on the probe are labeled. Connect the flow cell to the control unit as shown in Figure 4-5, and make sure that you only connect cables with equal labeling.
Figure 4-5:
Connection of the
container probe to
the control unit
Microwave measuring field
Control Unit, underside
M-Rx
M-Rx
M-Tx R-Tx
ø100
R-Rx M-Rx
PT-100
HF quad cable
M-Tx
M-Tx
R-Tx
R-Rx
R-Rx
R-Tx
38
DynFAS MW
Chapter 4 Getting Started
IMPORTANT
When tightening the 21 mm screw nut, make sure that the con­nector is not twisted on the cable. If the connector is twisted rela­tive to the cable, the shielding may get damaged and this could result in mismatching and bad sealing.
Hand tighten all screwed connections of the HF cable (2 Nm = 0.2 kg/m)! Before tightening, carefully screw on the cable by hand. Caution! Threaded joint jams easily.
Occasionally you should check if the screwed connection is still properly tightened. If the installation is ex posed to vibrations, the screwed connection may come loose and this may result in inaccu­rate measurements or corrosion of the connections.
As long as the cables are not connected, the coaxial sockets have to be covered immediately with plastic caps and the cable connec­tors have to be protected by suitable provisions against moisture and dirt.
DynFAS MW
39
Chapter 4 Getting Started
4.3.2 Pin Configuration of the Connector Strip
Electrical shock hazard: Disconnect power to rule out any contact with live parts during
installation and when servicing. Turn off power supply before opening the instrument. NEVER work
on open and live instruments.
Temperature Signal Connection
A Pt 100 or a temperature current signal has to be connected to current input 1 or 2 if temperature fluctuations occur in the prod­uct and if a temperature dependence of the phase or attenuation measurement is likely to occur. The temperature sensor has to measure the material temperature in the vicinity of the microwave probe.
When taking the container probe into operation, connect the 4wire cable of the Pt 100 to the connector strip of the control unit as follows:
Figure 4-6:
Pt 100 connection
container probe
Contr o l Unit Connector strip
(23)
(11)
( ) Terminal no.
blue blue
white
white
PT-100 sensor in the container probe
40
DynFAS MW
Chapter 4 Getting Started
Other Connections
¾ Connect all desired input and output signals to the terminal
strip as shown below. Use the M feed-through to maintain the degree of protection.
¾ Check if the voltage indicated on the type plate matches your
local supply voltage.
¾ Connect the line cable to the terminals 3(L1), 2(N) and 1(PE). ¾ Check if the test switch (mains interru ption) is in position „on“
(see Figure 5-1).
¾ Close the instrument housing and turn on the power supply.
Attention! Possible danger, damage to property! Concerns the sys­tem type CH-200-024 Control Unit High Dynamic 24 V DC (Id.-No. 54878-02): When connecting the 24 V DC auxiliary power, the + and – Poles should be connected correctly. There is no reverse voltage protec­tion!
The line cross-section for the power supply must be at least
1.0 mm
2
.
On the connector strip of the control unit you find the following connections:
or / bzw.
DynFAS MW
Figure 4-7:
DynFAS MW
wiring diagram
- +
41
Chapter 4 Getting Started
Power supply: Terminals 3 (L1, +), 2 (N, -) and 1 (PE, )
For CS-100, depending on instrument version, see type label on the outer wall of the housing.
1.) 90 V - 265 V AC, 45 - 65 Hz
2.) 24 V DC: 18 … 36 V
24 V AC: -20%, +5%, 40 … 440 Hz For CH-200, depending on instrument version, see type label on
the outer wall of the housing.
1.) 90 V - 265 V AC, 45 - 65 Hz
2.) 24 V DC: 18 … 36 V, no reverse voltage protection
Current input no. 1 (terminals 20+ and 8-), insulated Input as 0/4 - 20 mA signal. e.g. for temperature compensation or
reference signal recording.
Current input no. 2 (terminals 22+ and 10-), not insulated Input as 0/4 - 20 mA signal. e.g. for temperature compensation or
reference signal recording.
Current output no. 1 (terminals 27+ and 15-), insulated Output as 4 - 20 mA signal. Output options: concentrations (1/2),
current inputs signals (1 / 2) and Pt 100 signal
Current output no. 2 (terminals 19+ and 7-), insulated Output as 0/4 - 20 mA signal. Output options: concentrations
1 and 2, current input signals 1 and 2 and Pt 100 signal
Pt 100 (terminals 23+ and 11-)
Connection for temperature measurement.
Digital input 1: DI1 (terminals 24+ and 12-)
Configuration options:
¾ no function ¾ measurement: start (closed) and stop (open)
Digital input 2: DI2 (terminals 25+ and 13-)
Configuration options:
42
¾ no function ¾ average value: hold (closed) and continue averaging (open) ¾ product selection: product 1 (open) and product 2 (closed)
DynFAS MW
Chapter 4 Getting Started
Digital input 3: DI3 (terminals 26+ and 14-)
Configuration options:
¾ no function ¾ start sampling, open: no action, closed: unique measurement
starts
¾ product selection
Relay 1: (terminals 4, 5 and 6) Changeover contacts (SPDT), insulated, configuration option:
¾ no function ¾ error message ¾ stop measurement ¾ limit value min. and max. ¾ no product
Relay 2: (terminals 16, 17 and 18)
Changeover contacts (SPDT), insulated, configuration option:
¾ no function ¾ error message ¾ stop measurement ¾ limit value min. and max. ¾ no product
RS485 interface (terminals 21 (RS1) and 9 (RS2))
Serial data interface for output of live data (all measuring data for every sweep, measuring cycle) the setup protocol and data log. Data format: 38400 baud, 8 data bits, 1 stop bit, no parity, no handshake.
RS232 interface (on instrument bottom)
9-pole SubD-connector. Serial data interface for output of live data (all measuring data for every sweep, measuring cycle) the setup protocol and data log. Data format: 38400 baud (Data transfer rate ), 8 data bits, 1 stop bit, no parity, no handshake.
DynFAS MW
43
Chapter 4 Getting Started
5 6
4.3.3 Digital Outputs, Relays
The status of the measurement is output via two relays:
¾ Error ¾ Alarm (alarm min. and max.) ¾ No product
Under menu item Plausibility, you may enter a min. attenuation for pause detection (e.g. for process pause, no product pre­sent); if this value is not reached, „no product“ is sign aled via a relay and the current output drops to 0 or 4 mA. A typical application is pause detection between t he discontinu-
ous evaporation crystal processes. ¾ Measurement stopped The respective switching status is also signaled via LED’s on the
front panel (LED’s: signal 1 and 2).
Relay no. Error, alarm, no product,
measurement stopped, currentless status
4
Normal
4
1
com
5
6
com
2
16
17
18
com
16
17
18
com
The relays with changeover contacts can either be operated as make contact, terminals 4 & 5 (open at error, alarm ...) or as break contact, terminals 5 & 6 (closed at error, alarm ...).
44
DynFAS MW
Chapter 5 Service Instructions
Chapter 5. Service Instructions
5.1 General Information
The control unit has no wearing parts or components requirin g any special maintenance.
A malfunction of the measuring system is not always due to a de­fect in the instrument. Often the error is caused by incorrect op­eration, wrong installation, or irregularities in the product being measured.
If a malfunction occurs, anyway, the measuring system helps you to identify and eliminate errors by displaying error messages on the LCD, indicating operator errors and defects of the electronics.
Usually, faulty modules of the control unit cannot be repaired but have to be replaced. The microwave module is fixed with screws to a shielding cover and must not be opened.
5.2 Wearing Parts
The control unit consists of no wearing parts and components that need special attention.
The plastic rods of the Probes and the PTFE lining of the Flow Cell can eventually experience abrasion depending on the material be­ing measured. A lower to middle abrasion influences inconsiderably the measurement or is compensated by calibration. Therefore, you should check the wearing parts approximately every 2 years. The plastic rods of the Probe and the lining of the Flow Cell can be exchanged if abrasion is heavy. During excessive wear, the plastic rods of the Probe and the lining of the Flow Cells can be ex­changed. These Probes and the Flow cells have to be sent back to the company. An on-site exchange is not possible.
5.3 Instrument Cleaning
Clean all system components using a moistened cloth. Do not use any chemical cleaning agent. Parts coming into contact with the product (during regular operation) can be cleaned with hot water, taking into account the temperature limits (see chapter 6.2 Technical Data Sensors).
DynFAS MW
45
Chapter 5 Service Instructions
5.4 Battery
If the measuring system DynFAS MW is without power supply (power failure or disconnected from mains), the system clock is supplied with power by the Lithium battery on the CPU. The in­strument works correctly even with empty battery, only measured data which are output via one of the serial interfaces may become useless as a result of the faulty date and time information.
The service life of the battery, even under continuous load, is at least 8 years. To replace the battery, you have to disconnect the instrument from mains.
Battery type: 3 Volt Lithium cell (round cell battery), type CR2032.
5.5 Fuse Replacement
The mains fuse of the DynFAS MW is located in the wall housing. Replace the fuses only if the instrument is disconnected from mains. Be sure that the new fuses match the rating specified.
Use only fuses with correct rating:
For CS-100:
Instrument version with 90 ... 265 V AC: 2.0 A slow-blow Instrument version with 24 V AC/DC: 2.0 A slow-blow
For CH-200:
Instrument version with 90 ... 265 V AC: 2.0 A slow-blow Instrument version with 24 V DC: 6.3 A slow-blow
Spare fuses must match the rating specified by the device manu­facturer. Short-circuiting or manipulation is not permitted.
Netzteil
Test switch Motherboard
I
0
Fuses
Line connecto r
Protective cover
EEprom (can be pulled off)
Feed-through for
Figure 5-1:
Look inside the
instrument CS-100
46
line connector
Terminal strip
(can be pulled off)
Battery
DynFAS MW
Chapter 6 Technical Data
Chapter 6. Technical Data
General Specifications
Method Microwave transmission measurement Working frequency 2.4 – 2.5 GHz (ISM band), depending on local
regulations
Transmission power CS-100: < 0.1 mW (< -10 dBm)
CH-200: < 10 mW (< 10 dBm) All coaxial line power
Applications Concentration measurement in containers and
pipes
6.1 Control Unit
Control unit
Housing
Protection type IP 65 Weight CS-100: approx. 6.5 kg
Operating temperature
Storage temperature
Achievable accuracy
Display Dot matrix LC display, 114 mm x 64 mm, 240 x
Keyboard Freely accessible foil keypad, light-stable and
Power supply For CS-100 depending on instrument version:
Wall housing made of stainless steel, material
1.4571 (~316+Ti), see dimensional drawing in chapter 8.
CS-100: HxWxD: 300 x 323 x 140 mm CH-200: HxWxD: 400 x 338 x 170 mm
CH-200: approx. 8.0 kg
-20 ... +60°C ( 253 ...333 K ), Humidity 0 – 90% relatively, no condensation
-20 ... +80°C ( 253 ...353 K ), Humidity 0 – 90% relatively, no condensation
0.2 weight % (standard deviation) depending on product and sensor
128 pixels, with back-lighting, automatic contrast setting
weatherproof: alphanumeric keyboard and four softkeys (software-assigned buttons)
1.) 90 ... 265 V AC, 45 ... 65 Hz or
2.) 24 V DC: 18 ... 36 V 24 V AC: -20%, +5%, 40 ... 440 Hz
For CH-200 depending on instrument version:
1.) 90 ... 265 V AC, 45 ... 65 Hz or
2.) 24 V DC: 18 ... 36 V, no reverse voltage pro­tection
DynFAS MW
47
Chapter 6 Technical Data
Power consumption
Fuses For CS-100:
Battery type 3 V Lithium button cell, type CR2032 Measured value e.g. concentration, dry content
For CS-100: max. 30 VA (AC/DC), depending on configuration
For CH-200: max. (48/60) VA (AC/DC), depending on configu-
ration
2 x 2.0 A / slow-blow For CH-200:
2 x 2.0 A / slow-blow for 90 … 265 V AC or 2 x 6.3 A / slow-blow for 24 V DC
Inputs and Outputs
Cable cross-section
Cable feed-through
Sensor connection
min. 1.0 mm² (mains supply)
2 x M20x1.5 for cable 5...14 mm (depending on application)
4 x M16x1.5 for cable 5 ...8 mm (depending on application)
Inputs and outputs for signal and reference channel, 50 Ω N-socket
HF-cable
Current input 2 x current input 0/4 ...20 mA, ohmic resistance
Current output
Pt 100 connection
Cable lengths: 2, 4, 6 and 10 m; 50 Ω; both sides with 4 N connectors
50 Ω, 1x insulated, 1x instrument ground e.g. for temperature compensation
Current output 1: 4...20 mA, ohmic resistance max. 800 Ω , insulated current output 2: 0/4...20 mA, ohmic resistance max. 800 Ω , insulated e.g. for measured value or temperature output
Measuring range: -50 ... +200°C (223 ... 473 K); measurement tolerance: < 0.4°C
48
DynFAS MW
Chapter 6 Technical Data
Digital input 3 x digital inputs (DI1..3), for floating connectors
Configuration options: DI1: none, measurement start/stop DI2: none, measurement hold, product selection DI3: none, sampling, product selection
Function description:
1. Measurement (Start/Stop) open: measurement stopped closed: running
2. Hold measurement open: closed: values and current output are held
3. Product selection open: closed: with two DI’s: DI2 open & DI3 open closed & DI3 open: P2, DI2 open & DI3 closed: P3, DI2 closed &DI3 closed:
4. Start sampling open: closed: single measurement starts
measurement started or measurement
measurement running
measurement stopped, i.e. average
product 1 (P1)
P2;
no actions
: P1, DI2
P4
Relay outputs 2 x relays, insulated
Configuration options:
- Collective failure message
- Stop measurement
- Limit value (alarm min. and max.)
- No product
Load capacity: AC: max. 400VA DC: max. 90W AC / DC: max. 250V, max. 2A, non-inductive
150V: voltage must be grounded
The cable and insulation that are to be connected to these cables must correspond to a mains con­nection.
Restrictions for 24 V AC/DC mains supply, if the ground conductor is not connected to terminal 1 (PE):
AC: max. 50 V DC: max. 70 V
Serial interfaces
RS 232 on the underside of the instrument, RS 485 through terminal block
Data type: 38400 baud, no handshake, 8 data bits, 1 stop bit, no parity
DynFAS MW
49
Chapter 6 Technical Data
6.2 Technical Data Sensors
Flow Cells
Application Microwave flow cell with various nominal widths and
flanges for measurement on pipelines Material Stainless steel, PTFE lining Process coupling Process pressure Up to 20 bar (relative), depending on nominal width
Flange according to DIN EN 1092 Type 05 and ASA
and flange type, see table below Temperature range
Connections
Versions Nominal pipe widths from 50 ... 150 mm Dimensions See dimensional drawings in chapter 8
Product temperature: +10 ... +130°C (283 … 403 K)
Ambient temperature: -20 ... +60°C (253 … 333 K)
Storage temperature: +10 ... +80°C (283 … 353 K)
2 x HF connections: N-socket, 50 Ω
for HF-cable with max. 10 m length
Overview Flow Cells
Designation ID-No. Nominal
width
Flange Pressure
[bar]
[mm]
FC-050-016 54943 50 DN 50 / PN 16 16 FC-065-040 54944 65 DN 65 / PN 40 20 FC-080-016 54945 80 DN 80 / PN 16 16 FC-100-016 54946 100 DN 100 / PN 16 16 FC-150-016 54941 150 DN 150 / PN 16 16 FC-020-150 54947 50 ASA 2’’ / 150 PSI 16 FC-025-300 54948 65 ASA 2.5’’ / 300 PSI 20 FC-030-150 54949 80 ASA 3’’ / 150 PSI 16 FC-040-150 54950 100 ASA 4” / 150 PSI 16 FC-060-150 54951 150 ASA 6’’ / 150 PSI 16
50
DynFAS MW
Chapter 6 Technical Data
Probes
Application Probes with and without flushing
device for concentration measurement in process containers and pipelines with nominal width 200 mm.
Material Plastic rod, stainless steel
PT100 connection cable: Silicon / Teflon
Process coupling Flange according to DIN EN 1092 Type 05 DN65 / PN6,
DN 80, 100, 150 / PN16;
ASA flange 2.5’’, 3’’ / 150 PSI Process pressure Up to 16 bar (relative), depending on model Temperature range Product temperature: +10 ... +120°C (283 … 393 K)
Ambient temperature: -20 ... +60°C (253 … 333 K)
Storage temperature: +10 ... +80°C (283 … 353 K) Connections
Dimensions See dimensional drawings in chapter 8
4 x HF connections: N-socket, 50 Ω
for HF-cable with max. 10 m length
Accessory sealing washer
Material Klingersil C-4400 Thickness 3 mm
Overview Probes
Designation ID-No. Flange Pressure
[bar] P-065-006 54939-01 DN 65 / PN 6 6 P-080-016 54939-02 DN 80 / PN 16 16 P-100-016 54939-03 DN 100 / PN 16 16 P-150-016 54939-04 DN 150 / PN 16 16 P-025-150 54939-05 ASA 2.5’’ / 150 PSI 16 P-030-150 54939-09 ASA 3’’ / 150 PSI 16
DynFAS MW
PF-065-006 54940-01 DN 65 / PN 6 6 PF-080-016 54940-02 DN 80 / PN 16 16 PF-100-016 54940-03 DN 100 / PN 16 16 PF-150-016 54940-04 DN 150 / PN 16 16 PF-025-150 54940-05 ASA 2.5’’ / 150 PSI 16
51
Chapter 6 Technical Data
6.3 Technical Data HF-Cable
HF-Cable Quad
Material Corrugated tube: Polyamide (PA6)
Cable sheath: Polyethylene (PE) Protection type IP 66 Temperature Operating temperature: -30 ... +70°C (243 … 343 K)
Installation temperature: -20 ... +70°C (253… 343 K)
Summary HF-Cable Quad, Cable lengths
Length [m] Id No.
2.0 54856
4.0 54857
6.0 54858
10.0 54859
52
DynFAS MW
Chapter 6 Technical Data
6.4 Format of Serial Data Output RS 232 and RS 485
Header
Date·TimeFlagsStatusProductAttPhiR2TintIN1IN2Pt 100CCmC2C2m¶
Following lines
01.01.2005·00:00:000000→0→1→0.435.300.070.00.00.00.075.3675.000.000.00¶
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Column no. Description Format 1 Date and time DD.MM.YY·HH:MM:SS 2 Flags (for test purposes) 4 digits, HEX 3 Status: Information on quality of last
measurement 4 Product number X (1 to 4) 5 Attenuation [dB] X.XX 6 Phase [°/GHz] X.XX 7 Statistical spread of phase regression X.XX 8 Instrument temperature [temperature unit] X.X 9 Current input 1 [unit of current input] X.X 10 Current input 2 [unit of current input] X.X 11 Pt 100 temperature [temperature unit] X.X […] by selection of unit g/cm 12 Concentration 1 live X.XX [X.XXXX] 13 Concentration 1 averaged X.XX [X.XXXX] 14 Concentration 2 live X.XX [X.XXXX] 15 Concentration 2 averaged X.XX [X.XXXX]
0 : measurement OK < 0 : error
3
Special characters
” Tabulation “¶” Carriage return + Line feed “·” Blank character
DynFAS MW
53
Chapter 7 Certificates
7.1 EC Declaration of Conformity
Chapter 7. Certificates
DynFAS MW
55
Chapter 7 Certificates
56
DynFAS MW
Chapter 7 Certificates
7.2 Frequency Approval
DynFAS MW
57
Chapter 7 Certificates
58
DynFAS MW
Chapter 7 Certificates
DynFAS MW
59
Chapter 7 Certificates
60
DynFAS MW
Chapter 7 Certificates
DynFAS MW
61
Chapter 7 Certificates
62
DynFAS MW
Chapter 8 Technical Drawings
Chapter 8. Technical Drawings
8.1 Dimensional Drawing of Control Unit Housing
8.1.1 Control Unit Standard CS-100
DynFAS MW
63
Chapter 8 Technical Drawings
8.1.2 Control Unit High Dynamic CH-200
64
DynFAS MW
Chapter 8 Technical Drawings
8.2 Electrical Wiring Diagram
Line in for CS-100:
1. / 2. depending on instrument version
1. AC 90-265V, 45-65 Hz
2. DC 24 V (18-36 V) or AC 24 V -20% / +5%, 40-440 Hz ________________________________
Line in for CH-200:
1. / 2. depending on instrument version
1. AC 90-265V, 45-65 Hz
2. DC 24 V (18-36 V), no reverse voltage protection
DynFAS MW
65
Chapter 8 Technical Drawings
8.3 Dimensional Drawings Flow Cells
8.3.1 Type FC-050-016
66
DynFAS MW
Chapter 8 Technical Drawings
8.3.2 Type FC-065-040
DynFAS MW
67
Chapter 8 Technical Drawings
8.3.3 Type FC-080-016
68
DynFAS MW
Chapter 8 Technical Drawings
8.3.4 Type FC-100-016
DynFAS MW
69
Chapter 8 Technical Drawings
8.3.5 Type FC-150-016
70
DynFAS MW
Chapter 8 Technical Drawings
8.3.6 Type FC-020-150
DynFAS MW
71
Chapter 8 Technical Drawings
8.3.7 Type FC-025-300
72
DynFAS MW
Chapter 8 Technical Drawings
8.3.8 Type FC-030-150
DynFAS MW
73
Chapter 8 Technical Drawings
8.3.9 Type FC-040-150
74
DynFAS MW
Chapter 8 Technical Drawings
8.3.10 Type FC-060-150
DynFAS MW
75
Chapter 8 Technical Drawings
8.4 Dimensional Drawings Probes
8.4.1 Type P-065-006
76
DynFAS MW
Chapter 8 Technical Drawings
8.4.2 Type P-080-016
DynFAS MW
77
Chapter 8 Technical Drawings
8.4.3 Type P-100-016
78
DynFAS MW
Chapter 8 Technical Drawings
8.4.4 Type P-150-016
DynFAS MW
79
Chapter 8 Technical Drawings
8.4.5 Type P-025-150
80
DynFAS MW
Chapter 8 Technical Drawings
8.4.6 Type P-030-150
ll
DynFAS MW
81
Chapter 8 Technical Drawings
8.4.7 Installation Situation in Pipelines
82
DynFAS MW
Chapter 8 Technical Drawings
8.5 Dimensional Drawings Flush Probes
8.5.1 Type PF-065-006
DynFAS MW
83
Chapter 8 Technical Drawings
8.5.2 Type PF-080-016
84
DynFAS MW
Chapter 8 Technical Drawings
8.5.3 Type PF-100-016
DynFAS MW
85
Chapter 8 Technical Drawings
8.5.4 Type PF-150-016
86
DynFAS MW
Chapter 8 Technical Drawings
8.5.5 Type PF-025-150
DynFAS MW
87
Chapter 8 Technical Drawings
8.5.6 Installation Situation in Pipelines
88
DynFAS MW
Chapter 8 Technical Drawings
8.6 Installation Sheets for DynFAS MW (Probe)
DynF AS MW with Probe
- Installa t ion In Cryst a llizer -
top view into the crystallizer
evaluation unit
botten side
M-Rx
M-Tx
90 - 265 V AC
or 24 V AC/DC
(24 V DC)
output
0/4 - 20 mA
R-Tx
R-Rx
microwave measuring field
probe flange DN65 / PN6
14
installation position
ø
M-Rx
ø
1
30
R-Rx
190
PT100
container wall
Flow
Strömung
R-Tx
M-Tx
60
100
ø
insulation
ø
1
M-Tx
R-Tx
R-Rx M-Rx
sealing
do not remove!
PT100
6
0
M-Rx
PT100
HF cable quad
M-Tx
R-Rx
R-Tx
Control Unit
160
ø
(...) for CH-200
heating
300
(318)
220
(320)
DynFAS MW
fitting flange
ø
130 / 4 x M12
±
ø
100
current stream
0.2
0
6
area without metalic installations
89
Chapter 8 Technical Drawings
DynFAS MW with Probe
- Installa tion In C rystallizer -
top view into the crystallizer
microwave measuring field
probe flange DN80, 100, 15 0 / PN16
installation position
DN
80 100 150
container wall
190
PT100
ABC
160
200
18 M16
180 220
18 M16
240
28522M20
60
S
100
ø
sealing
insulation
evaluation unit
botten side
M-Tx
R-Tx
R-Rx M-Rx
B
ø
M-Rx
M-Rx
PT100
HF cable quad
Control Unit
(...) for CH-200
M-Tx
M-Tx
300
(318)
90 - 265 V AC
or 24 V AC/DC
(24 V DC)
output
0/4 - 20 mA
R-Tx
R-Rx
R-Rx
R-Tx
220
(320)
C
ø
fitting flange
M-Rx
A
ø
R-Rx
ø
R-Tx
A / 8 x S
±
ø
102
M-Tx
current stream
0,5
do not remove!
ø
B
PT100
0
6
area without metalic installations
heating
90
DynFAS MW
Chapter 8 Technical Drawings
8.7 Installation Sheets for DynFAS MW (Flush Probe)
DynFAS MW with Pro be with Flus hi ng
- Installation In Crystallizer -
top view into the crystallizer
evaluation unit
botten side
M-Rx
M-Tx
90 - 265 V AC
or 24 V AC/DC
(24 V DC)
output
0/4 - 20 mA
R-Tx
R-Rx
microwave measuring field
320
container wall
160
M-Tx
R-Tx
100
ø
sealing
R-Rx
M-Rx
160
ø
HF cable quad
M-Tx
M-Rx
Control Unit
(...) for CH-200
300
(318)
R-Rx
R-Tx
probe flange DN65 / PN6
insulation
220
(320)
installation position
ø
M-Rx
ø
R-Tx
M-Tx
5
5
0
3
1
R-Rx
Flow
14
Strömung
current stream
fitting flange
ø
130 / 4 x M12
±
0,2
ø
100
do not remove!
flushing connection: intern a l th r e a d 2 x G3/8’’ (DIN ISO 228-1 )
ø
160
heating
60
area without metalic installations
DynFAS MW
91
Chapter 8 Technical Drawings
DynFAS MW with Probe with Flushing
- Installation In Crystallizer -
top view into the crystallizer
microwave measuring field
container wall
320
100
160
ø
evaluation uni t
M-Tx
R-Tx
R-Rx M-Rx
B
ø
probe fl ange DN80, 100, 150 / PN16
DN
installation position
100 150
80
ABC 160 180 220 240
285 22 M20
200
S 18 M16 18 M16
sealing
insulation
botten side
HF cable quad
M-Rx
M-Tx
M-Tx
M-Rx
Control Unit
(...) for CH-200
300
(318)
90 - 265 V AC
or 24 V AC/DC
(24 V DC)
output
0/4 - 20 mA
R-Tx
R-Rx
R-Rx
R-Tx
220
(320)
C
ø
fitting flange
ø
M-Rx
A
R-Rx
ø
A / 8 x S
ø
102
R-Tx
5
5
±
0,5
M-Tx
current stream
do not remove!
ø
B
flushing connection: inte rn a l th read 2 x G3/8’’ (DIN ISO 228-1)
heating
60
area without metalic installations
92
DynFAS MW
Index
A
accuracy · 47 adapter flange · 34
Index
foodstuffs · 24 Frequency Approval · 13, 57 Fuse Replacement · 46
G
B
Battery · 46 bending radius · 28
C
Calculation of Measured Values · 17 CE symbol · 13 Certificates · 55 Components · 19 Connecting the Container Probes · 38 Connecting the Flow Cell · 37 Container Probe · 24 Container Probe Installation · 34 Control unit Housing · 63 Control Units · 20 cooling crystallizers · 18
D
Data format RS232 · 53 Data format RS485 · 53 Data transfer rate · 43 Digital Outputs · 44 Dimensional Drawings Flow Cells · 66 Dimensional Drawings Flush Probes · 83 Dimensional Drawings Probes · 76 distance rail · 36
gas bubbles · 14, 33
H
High-Frequency Cable · 2 8
I
Installation in Pipelines · 34 Installation Sheets · 89 Installation Situation in Pipelines · 82, 88 Instrument Cleaning · 45
L
LED’s · 22
M
Measurement Configuration · 30, 31 Microwaves · 15
O
Overview Flow Cells · 50 Overview Probes · 51
P
E
EC Declaration of Conformity · 55 Electrical Wiring Diagram · 65
F
Factory setting · 15 fitting flange · 34 Flow Cell · 23 Flow Cell Installation · 32 Flush Parameters · 35 flush probe · 26
DynFAS MW
Principle of Measurement · 16 Pt 100 connection · 40
R
Reference temperature · 18 Relays · 44 riser · 32 RS232 interface · 43
93
S
safety instructions · 7 salt content · 14 Softkeys · 15 symbols · 7
T
Technical Data · 47
Technical Data HF-Cable · 52 Technical Data Sensors · 50 Technical Drawings · 63 Temperature Compensation · 18 Transmission power · 47 Transport · 32
W
Wearing Parts · 45
94
DynFAS MW
Notes
DynFAS MW
95
Notes
96
DynFAS MW
Concentration Meters
DynFAS MW
Software Manual
User's Guide
Id. Nr. 54877BA2
Rev. No.: 00 01.07.2009
The units supplied should not be repaired by anyone other than BMA Schaltanlagen Service engineers or technicians authorized by BMA Schaltanlagen.
In case of operation trouble, please address to our central service department.
The complete user’s guide consists of two manuals, the hardware description and the software description.
The hardware manual comprises:
¾ mechanical components
¾ installation
¾ electrical installation
¾ radiation protection guidelines
¾ technical data
¾ electrical and mechanical drawings
The software manual comprises:
¾ operation of the evaluation unit
¾ parameter description
¾ basic setting
¾ calibration
¾ error messages
The present manual is the software description.
Subject to changes without prior notice.
BMA Schaltanlagen GmbH Am Alten Bahnhof 5 D-38122 Braunschweig
Tel. +49 531 804 261 Fax +49 531 804 269 schaltanlagen@bma-de.com www.bma-worldwide.com
DynFAS MW
Table of Contents
Page
Chapter 1. Communication with DynFAS MW 9
1.1 Brief Instructions 10
1.2 System Configuration 10
1.3 System Calibration 10
Chapter 2. Software Functions 11
2.1 Information on Menu Structure 11
2.2 Menu Structure 12
2.2.1 Start Menu 14
2.2.2 Diagnostic 14
2.2.3 Setup 16
2.2.4 Access Level 17
2.2.5 Language 17
2.2.6 Configuration 18
2.2.7 General Data 19
2.2.8 Measurement 19
2.2.9 Plausibility 20
2.2.10 Phase Measurement 21
2.2.11 Pause Detection 22
2.2.12 Calibration 24
2.2.13 System Adjust 24
2.2.14 Calibrate Concentration 25
2.2.15 Sample No. 26
2.2.16 Sample Data (expanded) 27
2.2.17 Advanced Settings 27
2.2.18 Calibr. manual 28
2.2.19 Input / Output 28
2.2.20 Current output 29
2.2.21 Current Output 1 29
2.2.22 Current Output 2 30
2.2.23 Current input 30
2.2.24 Current In 1 30
2.2.25 Current In 2 30
2.2.26 Pt 100 31
2.2.27 Digital Output 31
2.2.28 Digital Input 31
2.2.29 Service 32
2.3 Trend Display 34
Chapter 3. Configuration 35
3.1 Configuration Setup 35
3.1.1 General Data 35
3.1.2 Measurement 36
3.1.3 Plausibility 36
3.1.4 Microwave 36
3.1.5 Marker 37
3.1.6 Units 37
3.2 Start Calibration Coefficients 39
DynFAS MW
Loading...