INTRODUCTION
The Blancett B220210 Magnetic Pick-up with Preamplifi er was
designed to directly replace a standard pick-up within a turbine
fl ow meter when signal amplifi cation is required. The pick-up is a
combination of a standard Blancett magnetic pick-up and a solid
state electronic preamplifi er in a single stainless steel housing.
OPERATING PRINCIPLE
As fl uid fl ows through the meter the turbine rotates at a speed
proportional to the fl uid velocity. As each turbine blade passes
through the magnetic fi eld at the base of the pickup, an AC voltage
pulse is generated in the pick-up coil. The voltage signal generated
by these pulses becomes amplifi ed within the pick-up. The output
signal generated from the pick-up is now in the form of a square
wave. The frequency of this amplifi ed signal is used to represent
the fl ow rate of fl uid passing through the turbine fl ow meter.
Figure 1
Model 1100 fl ow meter
with preamplifi er pick-up and connecting cable
2 Form No. B910-002 07/09

SPECIFICATIONS
Supply Voltage: +5 to 30 VDC
Input Current: 15 mA
Current Sinking: 25 mA maximum
Pulse Out: 0 to 10 V peak-to-peak
Frequency Range: 5Hz to 10kHz
Transmit Distance: 1500 feet maximum
Temperature: -20 °F to +160 °F (-29 °C to +71 °C)
Housing: Stainless steel and epoxy encapsulated
Electrical
Connection: 3-pin straight Amphenol connector supplied with
10 feet of cable standard. (Part No. B220-219, sold
separately.)
INSTALLATION and WIRING INSTRUCTIONS
The B220210 is simply threaded into the hole (⅝″-18UNF) in the
turbine meter until the pick-up bottoms out. Tighten the pick-up
fi nger tight. Do not over tighten.
Install the 3-pin connector in the proper orientation and connect the
wire ends of the cable using appropriate common wiring practices.
Refer to Table 1 and Figure 2 for wiring specifi cations.
Pin Description Wire Color
A Power: +5 to 30 VDC @ 15 mA Red
B Common: Power return Black
C Output Signal: 0 to 10 V (Square Wave) Green or Clear
Table 1
Figure 2
Pin confi guration
Form No. B910-002 07/09 3

STATEMENT OF WARRANTY
Blancett Flow Meters, Division of Racine Federated Inc. warrants to the
end purchaser, for a period of one year from the date of shipment from
the factory, that all fl ow meters manufactured by it are free from defects in
materials and workmanship. This warranty does not cover products that
have been damaged due to defects caused by misapplication, abuse, lack
of maintenance, modifi ed or improper installation. Blancett’s obligation
under this warranty is limited to the repair or replacement of a defective
product, at no charge to the end purchase, if the product is inspected by
Blancett and found to be defective. Repair or replacement is at Blancett’s
discretion. A return goods authorization (RGA) number must be obtained
from Blancett before any product may be returned for warranty repair or
replacement. The product must be thoroughly cleaned and any process
chemicals removed before it will be accepted for return.
The purchaser must determine the applicability of the product for its desired
use and assumes all risks in connection therewith. Blancett assumes no
responsibility or liability for any omissions or errors in connection with the
use of its products. Blancett will under no circumstances be liable for any
incidental, consequential, contingent or special damages or loss to any
person or property arising out of the failure of any product, component or
accessory.
All expressed or implied warranties, including the implied warranty of
merchantability and the implied warranty of fi tness for a particular
purpose or application are expressly disclaimed and shall not apply to
any products sold or services rendered by Blancett.
The above warranty supersedes and is in lieu of all other warranties, either
expressed or implied and all other obligations or liabilities. No agent or
representative has any authority to alter the terms of this warranty in any
way.
8635 Washington Avenue • Racine, Wisconsin 53406
Toll Free: 800.235.1638
Phone: 262.639.6770 • Fax: 262.417.1155
www.blancett.com • info@blancett.com
BLANCETT and QUIKSERT are registered trademarks of
Racine Federated Inc.
UL is a registered trademark of Underwriters Laboratories.
© 2009 Racine Federated Inc.
Form No. B910-002 07/09
Printed in USA