Black Box TS1002A, TS1000A-R5, TS1003A User Manual

CUSTOMER
SUPPORT
INFORMATION
Order toll-free in the U.S. 24 hours, 7 A.M. Monday to midnight Friday: 877-877-BBOX FREE technical support, 24 hours a day, 7 days a week: Call 724-746-5500 or fax 724-746-0746 Mail order: Black Box Corporation, 1000 Park Drive, Lawrence, PA 15055-1018 Web site: www.blackbox.com • E-mail: info@blackbox.com
AUGUST 2000
TS1002A TS1003A
OMNIScanner
TX
RX
Fiber
OMNI
850 nm
1300/1310 nm
Scanner
I
N
M
O
OMNI
M
O
RX
Fiber
N
TX
850 nm
1300/1310 nm
Remote
I
S
S
A
P
S
U
T
A
T
S
IL
A
F
D
R
A
U
G
IT
U
C
R
I
C
E
G
R
A
H
C
def
abc
%
1
ghi
4
prs
7
A
3
2
mno
jkl
6
5
wxy
tuv
9
8
qz
?
0
/a
1
FCC INFORMATION
FEDERAL COMMUNICATIONS COMMISSION
AND
CANADIAN DEPARTMENT OF COMMUNICATIONS
RADIO FREQUENCY INTERFERENCE STATEMENTS
This equipment generates, uses, and can radiate radio frequency energy and if not installed and used properly, that is, in strict accordance with the manufacturer’s instructions, may cause interference to radio communication. It has been tested and found to comply with the limits for a Class A computing device in accordance with the specifications in Subpart J of Part 15 of FCC rules, which are designed to provide reasonable protection against such interference when the equipment is operated in a commercial environment. Operation of this equipment in a residential area is likely to cause interference, in which case the user at his own expense will be required to take whatever measures may be necessary to correct the interference.
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
This digital apparatus does not exceed the Class A limits for radio noise emission from digital apparatus set out in the Radio Interference Regulation of the Canadian Department of Communications.
Le présent appareil numérique n’émet pas de bruits radioélectriques dépassant les limites applicables aux appareils numériques de la classe A prescrites dans le Règlement sur le brouillage radioélectrique publié par le ministère des Communications du Canada.
2
OMNISCANNER
NORMAS OFICIALES MEXICANAS (NOM)
ELECTRICAL SAFETY STATEMENT
INSTRUCCIONES DE SEGURIDAD
1. Todas las instrucciones de seguridad y operación deberán ser leídas antes de que el aparato eléctrico sea operado.
2. Las instrucciones de seguridad y operación deberán ser guardadas para referencia futura.
3. Todas las advertencias en el aparato eléctrico y en sus instrucciones de operación deben ser respetadas.
4. Todas las instrucciones de operación y uso deben ser seguidas.
5. El aparato eléctrico no deberá ser usado cerca del agua—por ejemplo, cerca de la tina de baño, lavabo, sótano mojado o cerca de una alberca, etc..
6. El aparato eléctrico debe ser usado únicamente con carritos o pedestales que sean recomendados por el fabricante.
7. El aparato eléctrico debe ser montado a la pared o al techo sólo como sea recomendado por el fabricante.
8. Servicio—El usuario no debe intentar dar servicio al equipo eléctrico más allá a lo descrito en las instrucciones de operación. Todo otro servicio deberá ser referido a personal de servicio calificado.
9. El aparato eléctrico debe ser situado de tal manera que su posición no interfiera su uso. La colocación del aparato eléctrico sobre una cama, sofá, alfombra o superficie similar puede bloquea la ventilación, no se debe colocar en libreros o gabinetes que impidan el flujo de aire por los orificios de ventilación.
10. El equipo eléctrico deber ser situado fuera del alcance de fuentes de calor como radiadores, registros de calor, estufas u otros aparatos (incluyendo amplificadores) que producen calor.
11. El aparato eléctrico deberá ser connectado a una fuente de poder sólo del tipo descrito en el instructivo de operación, o como se indique en el aparato.
3
NOM STATEMENT
12. Precaución debe ser tomada de tal manera que la tierra fisica y la polarización del equipo no sea eliminada.
13. Los cables de la fuente de poder deben ser guiados de tal manera que no sean pisados ni pellizcados por objetos colocados sobre o contra ellos, poniendo particular atención a los contactos y receptáculos donde salen del aparato.
14. El equipo eléctrico debe ser limpiado únicamente de acuerdo a las recomendaciones del fabricante.
15. En caso de existir, una antena externa deberá ser localizada lejos de las lineas de energia.
16. El cable de corriente deberá ser desconectado del cuando el equipo no sea usado por un largo periodo de tiempo.
17. Cuidado debe ser tomado de tal manera que objectos liquidos no sean derramados sobre la cubierta u orificios de ventilación.
18. Servicio por personal calificado deberá ser provisto cuando:
A: El cable de poder o el contacto ha sido dañado; u
B: Objectos han caído o líquido ha sido derramado dentro del aparato; o
C: El aparato ha sido expuesto a la lluvia; o
D: El aparato parece no operar normalmente o muestra un cambio en su
desempeño; o
E: El aparato ha sido tirado o su cubierta ha sido dañada.
4
OMNISCANNER
TRADEMARKS
The trademarks mentioned in this manual are the sole property of their owners.
5
CONTENTS
CONTENTS
1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.1 The Digital Solution for Fiber and Copper . . . . . . . . . . . . . . . . . . 9
1.2 The Graphical User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3 OMNIRemote . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.4 OMNIFiber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.5 ScanLink Tools Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.6 OMNIScanner’s Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.7 OMNIScanner’s Keypad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.8 OMNIRemote’s LED Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
1.9 OMNIRemote Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
1.10 When to Use OMNIRemote. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
1.11 Connecting OMNIScanner and OMNIRemote . . . . . . . . . . . . . . 15
1.12 OMNIFiber’s LED Indicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
1.13 Attaching the OMNIScanner Fiber Adapters . . . . . . . . . . . . . . . . 17
1.14 Editing with OMNIScanner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
1.14.1 Predefined Lists. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
1.14.2 Editing Numeric Fields with the Spin Control . . . . . . . . . . 19
1.14.3 Editing Alphanumeric Fields . . . . . . . . . . . . . . . . . . . . . . . . 20
1.15 Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
1.16 Product Versions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2. Getting Started: Twisted Pair Networks . . . . . . . . . . . . . . . . . . . . . . . . 23
2.1 Charging OMNIScanner’s Battery. . . . . . . . . . . . . . . . . . . . . . . . . . 23
2.2 OMNIScanner’s Project Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
2.3 Certification of Network Installations . . . . . . . . . . . . . . . . . . . . . . . 24
2.4 Setting the Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
2.5 Project Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
2.6 Running an Autotest for Twisted Pair Cabling. . . . . . . . . . . . . . . . 28
2.6.1 Saving Twisted Pair Autotest Results . . . . . . . . . . . . . . . . . . . 29
2.6.2 Viewing Twisted Pair Autotest Results . . . . . . . . . . . . . . . . . . 32
2.7 Typical Test Configurations for Twisted Pair Cabling . . . . . . . . . . 36
2.7.1 Basic Link. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
2.7.2 Basic Link with 110 Block and Modular 8 Jacks . . . . . . . . . . 36
2.7.3 Full Channel with Modular 8 Jacks . . . . . . . . . . . . . . . . . . . . 37
2.7.4 Permanent Link. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
3. Getting Started: Fiber Optic Networks . . . . . . . . . . . . . . . . . . . . . . . . . 39
3.1 Charging OMNIScanner’s Battery. . . . . . . . . . . . . . . . . . . . . . . . . . 39
3.2 OMNIScanner’s Project Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
3.3 Certification of Fiber Installations. . . . . . . . . . . . . . . . . . . . . . . . . . 40
3.4 Setting the Reference for Fiber Optic Cabling. . . . . . . . . . . . . . . . 41
6
OIMNISCANNER
3.5 Project Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
3.6 Running an Autotest for Fiber Optic Cabling . . . . . . . . . . . . . . . . 45
3.6.1 Preparing an Autotest . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3.6.2 Loopback Autotest Using OMNIFiber Main. . . . . . . . . . . . . 47
3.7 Running a Fiber Autotest . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
3.8 Fiber Autotest Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
4. Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
4.1 Configuring OMNIScanner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
4.2 Set Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Set Reference with the OMNIFiber Adapters . . . . . . . . . . . . . . . . 54
4.3 Autotests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Viewing Details for Copper and Fiber Autotests . . . . . . . . . . . . . . 57
4.4 Cables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
4.5 Projects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Editing a Circuit ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
4.6 Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
4.7 Autotest Run Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
4.8 Autotest Save Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
4.9 User Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
4.10 Adapter Usage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
4.11 Remote Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
4.12 Fiber Adapter Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
4.13 Format MMC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
5. Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
5.1 Performing Individual Measurements with OMNIScanner . . . . . 75
5.1.1 Measuring Twisted Pair Cabling . . . . . . . . . . . . . . . . . . . . . . 75
5.1.2 Measuring Fiber Optic Cabling . . . . . . . . . . . . . . . . . . . . . . . 75
5.1.3 Measuring Coaxial Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . 76
5.2 General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
5.3 Twisted Pair Measurements with OMNIScanner . . . . . . . . . . . . . . 78
5.3.1 Wiremap. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
5.3.2 Length/Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
5.3.3 NEXT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
5.3.4 Attenuation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
5.3.5 Return Loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
5.3.6 ELFEXT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
5.3.7 ACR/Bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
5.3.8 Resistance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
5.3.9 Power Sum NEXT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
5.3.10 Power Sum ELFEXT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
7
CONTENTS
5.3.11 Power Sum ACR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
5.3.12 Office Locator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
5.4 Fiber Optic Measurements with OMNIFiber . . . . . . . . . . . . . . . . . 92
5.4.1 Length/Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
5.4.2 Loss. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
5.4.3 Power Meter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
5.5 Coaxial Measurements with OMNIScanner . . . . . . . . . . . . . . . . . . 95
Coax Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
6. Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
6.1 OMNIScanner’s Diagnostic Functions . . . . . . . . . . . . . . . . . . . . . . 97
6.2 Wiremap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Wiremap Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
6.3 TDNXT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
6.3.1 Using TDNXT to Diagnose a Link. . . . . . . . . . . . . . . . . . . . . 101
6.3.2 TDNXT Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
6.4 TDRL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
6.4.1 Using TDRL to Diagnose a Link . . . . . . . . . . . . . . . . . . . . . . 106
6.4.2 TDRL Examples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
7. Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
7.1 Autotest Results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
7.2 Viewing Autotest Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
7.3 Results Stored on a MultiMedia Card . . . . . . . . . . . . . . . . . . . . . . . 113
8. Fiber Optic Reference Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
8.1 Establishing the Reference. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
8.2 The Launch/Jumper Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
8.3 The Reference Methods. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
8.3.1 Method B: One Jumper Reference Method . . . . . . . . . . . . . 116
8.3.2 Method A: Two Jumper Reference Method . . . . . . . . . . . . . 117
8.3.3 Method C: Three Jumper Reference Method. . . . . . . . . . . . 118
8.3.4 Method A: Two Jumper Reference Method
in Loopback Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
9. Uploading and Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
9.1 ScanLink Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
10. Updating Scanner Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
10.1 Downloading Software from the WWW . . . . . . . . . . . . . . . . . . . . 122
10.2 Updating OMNIScanner and OMNIRemote Software
from the PC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
8
OMNISCANNER
11. Multi Media Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
11.1 Portable Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
11.2 Multi Media Card Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . 125
11.3 The MMC Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
12. Universal Serial Bus Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
12.1 Installing the USB Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
12.2 Uploading OMNIScanner Test Results to the PC . . . . . . . . . . . . 129
12.3 Downloading Configurations to OMNIScanner . . . . . . . . . . . . . 130
12.4 Remote Controlling OMNIScanner . . . . . . . . . . . . . . . . . . . . . . . 131
Appendix A: Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Appendix B: Batteries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
B.1 Batteries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
B.2 Battery Charging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
B.3 Battery Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
Appendix C: OMNIFiber PASS/FAIL Criteria. . . . . . . . . . . . . . . . . . . . . 143
Appendix D: Cable Pinouts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
D.1 Serial Cable DB9 Female to DB9 Male Pinouts . . . . . . . . . . . . . . . 144
D.2 Serial Cable DB9 Female to DB25 Male Pinouts . . . . . . . . . . . . . . 144
D.3 Serial Cable DB25 Female to DB25 Male Pinouts . . . . . . . . . . . . . 145
D.4 Cable Pinout Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
9
CHAPTER 1: Introduction
1. Introduction
1.1 The Digital Solution for Fiber and Copper
OMNIScanner is the most advanced hand-held certification and diagnostic tool for the verification of copper links. The Category 7, 6, 5E (Enhanced), 5 and ISO Class F, E, D, and C tester has an extended frequency range of 300 MHz. OMNIScanner provides full software compliance with TIA and ISO/IEC requirements for measurement reporting and analysis (including Length, Delay, ACR, NEXT, ELFEXT, Return Loss, Resistance, PSNEXT, PSACR, PSELFEXT, and Attenuation). In addition, OMNIScanner measures Length, Delay, Impedance, and Resistance on coaxial cabling.
OMNIFiber expands the tester’s capabilities to certify installed fiber links against TIA and ISO standards. Fiber cabling can be verified at 850 and 1300 nm wavelengths.
1.2 The Graphical User Interface
OMNIScanner’s graphical user interface consists of a backlit LCD display, quick­access menus, arrow keys, and descriptive menu options that guide you through certifying or troubleshooting cabling systems.
1.3 OMNIRemote
For test and certification of twisted-pair cabling, the OMNIScanner requires the OMNIRemote unit at the opposite end of the cabling link under test. OMNIScanner communicates with OMNIRemote to obtain status and perform measurements from both ends of the link.
OMNIRemote is equipped with a speaker to generate tones, and several colored LEDs to indicate overall test results and status.
1.4 OMNIFiber
OMNIFiber transforms OMNIScanner and OMNIScanner2 into a fiber optic tool for testing and certification of multimode and single-mode fiber premise networks. OMNIFiber consists of two identical test heads that attach directly onto OMNIScanner and OMNIRemote to become a part of each unit. Using the 1­Button Autotest feature, OMNIFiber measures loss (multimode and single-mode) and length (multimode) over two fibers simultaneously. The measurements are
10
OMNISCANNER
compared against the selected LAN standard for instant PASS/FAIL analysis. OMNIFiber measures loss at 850 and 1300 nm over multimode fiber and at 1310 nm over single-mode fiber.
Up to 1000 Autotests can be saved to nonvolatile memory. Saved results can be easily uploaded to the PC with Scanlink software.
1.5 Scanlink Tools Software
PC-based software expands OMNIScanner’s user interface and aids with the management of projects. Scanlink Tools consist of the OMNIScanner Configuration software, the Scanlink Upload software, and the OMNIScanner Remote Control software. These programs allow you to store, retrieve, manage, edit, and print network cabling information. Scanlink Tools; provide the ability to create professional reports, remote control OMNIScanner, and define cable and project configurations that can be downloaded to OMNIScanner.
1.6 OMNIScanner’s Display
Scanner Display Function
1-2 or 12/36: The paired numbers in the display indicate wire pairs. For
example, 1-2 indicates the wire pair of pins 1 and 2. 12/36 indicates two wire pairs tested, the first pair of pins 1 and 2, and the second pair of pins 3 and 6.
Pop-up lists: Pop-up lists are available by pressing ENTER when the
cursor is located on user-definable fields. A predefined selection list will appear, and the rest of the display will be inactive.
Spin Control: The spin control is available on editable fields that allow
for increasing or decreasing numeric values by depressing the up-arrow or down-arrow key.
Scrollbars: Use the up-arrow and down-arrow key on the keypad to
move up and down in a scrollable screen.
Graph: Use the left-arrow or right-arrow keys to move the graph
cursor on the frequency axis. Us the up-arrow or down­arrow to zoom in and out of graphs.
11
1.7 OMNIScanner’s Keypad
ON/OFF
When turned on, OMNIScanner will display the project screen.
Backlight
Press the Backlight key to toggle the backlight between its two levels of brightness. Press and hold the Backlight key to display the Contrast screen. Use the up-arrow and down-arrow keys on the keypad to adjust the display contrast. In addition to contrast adjustments, the screen displays hardware version, software version, serial­number information, and the date and time when the unit was last factory­calibrated.
CHAPTER 1: Introduction
ON/OFF
FUNCTION KEYS
AUTOTEST
BACKLIGHT
ESCAPE
ARROW KEYS
ENTER
HELP KEY
DB9 SERIAL
CONNECTOR
ADAPTER
CONNECTOR
AC ADAPTER
MMC CARD
SOCKET
USB PORT
ALPHANUMERIC
KEYS
UPPER/LOWER CASE
SWITCH KEY
12
OMNISCANNER
Autotest
Press the Autotest key to run a custom or predefined Autotest.
ESC
Press the Escape key to cancel the current operation.
Arrow keys
Press the left, right, up, or down arrow key to navigate within a list, menu, graph, or editable field.
ENTER
When in a selection field, press the ENTER key to open the listbox and then select the highlighted item as the new default. When in an editable field, press the ENTER key to start the editing process, and press ENTER again when finished with the editing process.
Alphanumeric Keys
Use the 0 through 9 keys to enter numeric and alphabetic characters in editable fields. Alphabetic and special characters are shown above each numbered key on the keypad. The 1 key allows entry of the following characters: - (hyphen), . (period), and , (comma).
The 0 key allows entry of an empty space.
The alphanumeric keys also allow direct selection of items in numbered lists.
Use this key to toggle between upper- and lower-case letters.
Use this key to quickly display a window listing the functionality of the
function keys in the current display.
1
ESC
13
CHAPTER 1: Introduction
Function Keys
The functions of the 4 keys located below the display vary depending on the current display. The assigned functions are:
Setup Measurements Diagnostics Results
New Edit Delete Save Options
Save Save with Graphs Save to MMC
Save with Graphs to MMC Copy Project to MMC
Copy Results to MMC Copy Circuit ID
Erase a letter Clear a field List Circuit IDs
Detail Retest Graph Pair
Next Page Previous Page Select
Set Default Switch to 850 nm Switch to 1300 nm
Set Reference Continue Test
Use Method A Use Method B Use Method C
Switch to Continuous Wave Switch to Modulated (2 kHz) Wave
Switch to OMNIScanner Info Switch to OMNIRemote Info
14
OMNISCANNER
1.8 OMNIRemote’s LED Indicators
LED Color Description
PASS Green Overall Autotest Result is
PASS.
Status Green OMNIRemote is ready for
(flashing) testing. Green OMNIRemote is active.
(consistent) FAIL Red Overall Autotest Result is FAIL Circuit Orange Hazard condition is detected; Guard LED will be lit in 5-second
intervals until the hazard
condition disappears. REMOVE CABLE FROM HAZARDOUS SOURCE! Charge Green OMNIRemote is inactive, the
AC adapter is present, and the
battery pack is being charged.
Green OMNIRemote battery is fully (flashing) charged. Red OMNIRemote battery is low. (flashing)
1.9 OMNIRemote Keys
ON/OFF
When turned on, OMNIRemote will blink the LEDs.
λ(Lambda)
When OMNIFiber is attached to OMNIRemote, the remote unit can be used as a light source. With the OMNIFiber adapter attached, press the Lambda key to activate the light source and provide continuous or modulated light at 850 or 1300/1310 nm. The green LED of the appropriate wavelength will blink rapidly when modulated light is transmitted. If the transmitted light is continuous, the green LED will not blink. (See OMNIFiber’s LED indicators in this chapter for further information.)
15
1.10 When To Use OMNIRemote
Most measurements require the use of OMNIRemote at the far end of the cable. For other measurements, OMNIRemote can be connected, but its use is optional. OMNIScanner detects the presence of the remote unit and automatically configures it for the measurement to be performed.
Twisted pair measurements requiring the use of OMNIRemote:
• Autotest
• Wiremap
• NEXT and Power Sum NEXT
• Attenuation
• Return Loss
• ELFEXT and Power Sum ELFEXT
• ACR and Power Sum ACR
Twisted pair measurements not requiring the use of OMNIRemote:
• Length/Delay/Skew
• Resistance
1.11 Connecting OMNIScanner and OMNIRemote
OMNIScanner and OMNIRemote are used together to perform measurements, which allow the scanner to verify installed links or assist in the diagnosis of physical layer network problems. To accomplish these measurements, OMNIScanner and OMNIRemote are connected at either end of a network-cabling link. The end at which the scanner is connected, is referred to as the “Scanner end,” and the end at which the remote unit is connected, is referred to as the “Remote end.”
CHAPTER 1: Introduction
16
OMNISCANNER
OMNIScanner2 displays test results for both ends of the link. Press or
to toggle the display between Scanner and Remote test results.
NOTE
Refer to the graphics in Chapter 2, Getting Started: Typical Test Configurations, for instructions on connecting OMNIScanner and OMNIRemote for basic link and full channel tests.
Indicates
OMNI
Scanner
test
results.
Indicates
OMNI
Remote
test
results.
Toggles the display from Scanner to Remote test results.
17
1.12 OMNIFiber’s LED Indicators
LED Color Description
850 nm Green OMNIFiber is
transmitting at 850 nm
850 nm Green OMNIFiber is
(flashing) transmitting
modulated light at 850 nm
850 nm Orange Part 1 of Two Way
Test completed,
switch fibers 1300/ Green OMNIFiber is 1310 nm transmitting at 1300/
1310 nm 1300/ Green OMNIFiber is 1310 nm (flashing) transmitting
modulated light at
1300/1310 nm 1300/ Orange Part 1 of Two Way 1310 nm test completed,
switch fibers
1.13 Attaching the OMNIScanner Fiber Adapters
NOTE
Both OMNIFiber adapters have dust caps to protect the ST connectors from dirt. Keep the ST connectors covered when the Fiber Adapters are not in use. This helps to ensure that contaminants from handling fiber do not affect test measurements.
NOTE
Both OMNIFiber adapters have 160-pin connector covers. Keep the 160­pin connectors covered when the Fiber Adapters are not in use.
To attach a fiber adapter to OMNIScanner, proceed as follows:
1. Remove the 160-pin connector cover from the OMNIFiber adapter by gently pressing the black pushbutton latches. The cover will be released.
CHAPTER 1: Introduction
18
OMNISCANNER
2. Push the OMNIFiber adapter onto OMNIScanner’s 160-pin connector until it snaps into position.
3. To remove OMNIFiber, use the pushbutton latches to release the secured adapter.
4. Protect OMNIFiber’s 160-pin connector by snapping the cover back in place.
1.14 Editing with OMNIScanner2
OMNIScanner includes sophisticated easy-to-use editing methods to accommodate a variety of user-definable items. Different fields have different editing procedures. Please follow the described procedures to make editing with the OMNIScanner effortless.
NOTE
All editable fields can be easily recognized by their dotted borders.
19
1.14.1 P
REDEFINED
L
ISTS
To quickly and easily select Projects, Autotests, Cables, etc., OMNIScanner2 uses predefined lists.
1. Use the up-arrow or down-arrow keys to move the cursor to the field to be edited and press ENTER.
A pop-up list will be displayed while the main screen becomes inactive.
2. Use the up-arrow or down-arrow key to move the cursor to the appropriate item or value and press ENTER to select it as the new default.
1.14.2 E
DITING
N
UMERICFIELDS WITH THESPINCONTROL
1. Use the up-arrow or down-arrow key to move the cursor to the field to be edited. The entire field will be highlighted in black.
2. Press ENTER. Notice that the field is now highlighted in white and the main screen in the background turns gray to indicate that the edit mode has been launched.
3. Press the up-arrow or down-arrow key on your keypad to scroll through the string of numbers.
CHAPTER 1: Introduction
20
OMNISCANNER
4. Press ENTER or ESC to accept the value and terminate edit mode.
1.14.3 E
DITINGALPHANUMERICFIELDS
To start the editing procedure for fields that contain alphabetic and numeric characters, use the alphanumeric keypad.
1. Use the up-arrow or down-arrow key to move the cursor to the field to be edited. The entire field will be highlighted in black.
2. Press ENTER. Notice that the field is now highlighted in white and the main screen in the background turns gray to indicate that the edit mode has been launched. The first letter in the field is highlighted in black.
3. Press any numerical key on your keypad to display a small drop-down list containing letters to choose from. Example: Press 8 key to choose t, u, and v or the number 8.
NOTE
Alphabetic and special characters are printed above each numbered key on the keypad. For example, ABC is printed above the 2 key.
21
4. Use the up-arrow or down-arrow key to move to the appropriate character, or simply press 8 on your keypad repeatedly until the appropriate letter or number appears in the highlighted field. Press the key to toggle between uppercase and lowercase letters.
5. Press the right-arrow to close the drop-down list and highlight the next letter
to be edited. Notice that OMNIScanner does not overwrite existing
characters. To delete single characters use the function key. To clear
an entire field use the function key.
6. Press ENTER to accept the field and exit the edit mode.
1.15 Technical Support
If you have technical questions, call Black Box Technical Support at 724-746-5500.
NOTE
Before calling Technical Support, please have your serial number and the hardware and software version numbers available.
1.16 Product Versions
To display the software version number, the hardware version number, the serial number, and the factory calibration data for OMNIScanner, press and hold the Backlight key. The CONTRAST screen will be displayed, which contains information that you may need to provide when calling Technical Support.
To display software, hardware, and serial number information for OMNIRemote, connect the units with the supplied test cable and press the function key.
CHAPTER 1: Introduction
22
OMNISCANNER
Select Remote Info to open the REMOTE INFO screen, which contains the date the reference was last set, the hardware and software information, the serial number, the battery indicator, and the factory calibration data.
The serial number can also be found on the back of each unit.
To display information for OMNIFiber, attach the OMNIFiber adapters to OMNIScanner and OMNIRemote, connect the units with the supplied fiber launch cable and press the (Setup) function key.
Select Fiber Into to open the Fiber Info screen, which contains the serial numbers and the date the adapters were last factory calibrated.
23
CHAPTER 2: Getting Started: Twisted Pair Networks
2. Getting Started: Twisted Pair Networks
2.1 Charging OMNIScanner’s Battery
To operate the OMNIScanner and OMNIRemote with battery power, the battery must first be charged. Until the charging cycle is complete, the units can run with the AC adapter plugged in. (See Appendix B, Batteries for more information about the batteries.)
2.2 OMNIScanner’s Project Screen
OMNIScanner’s project screen is designed to allow you to perform all typical work­flow operations. The project, the test specifications, and the cable types are all selectable from the OMNIScanner Project screen.
Press the ON/OFF key to power the unit on and display the project screen with
its three user-definable items: Project, Autotest, and Cable.
The first item Project name should be highlighted. Use the down-arrow to move the highlight to the Autotest or Cable field. Press ENTER when a filed is highlighted to change the current Project, Autotest, or Cable. OMNIScanner will auto-sense the Adapter.
Battery and Memory status are graphically indicated by gauges with a display ranging from empty (left-hand side) to full (right-hand side).
Press (Setup) to select Autotests configure Cables, and more.
Press (Measurements) to measure a cable’s performance, using a variety of tests.
24
OMNISCANNER
Press (Diagnostics) to view and manage stored test results.
Press (Results) to view and manage stored test results.
2.3 Certification of Network Installations
OMNIScanner is designed to efficiently certify cabling installations through its Autotest feature. Test results, test specifications, and cable types are organized into Projects. The Project, its test specifications, and its cable types are selectable from the OMNIScanner Project screen.
OMNIScanner contains a number a predefined projects, grouped by test standards to allow easy and fast operation.
The typical order of operations performed when certifying a cable installation is as follows:
Preparation
Before initial certification, do the following:
1. Run a self-test by setting the reference for OMNIScanner and OmniRemote periodically—for example, weekly.
2. Select the appropriate Project, Autotest specification, and Cable type.
Certification
Do the following for each installed cable to be tested:
3. Attach OMNIScanner to one end of the link, which is typically at the patch panel in a wiring closet.
4. Attach OmniRemote to the corresponding outlet at the other end of the cable link.
5. Run Autotest.
6. Save test results.
Completion
At the end of the day or shift, the test results are uploaded using Scanlink Tools software.
25
CHAPTER 2: Getting Started: Twisted Pair Networks
Scanlink Tools supports the following:
• Upload test results.
• Storage of test results to files (Scanlink compatible format)
• Print test reports
• Export results in CSV format
• Setup Projects
• View Graphs
• Edit Autotest Information
The following sections provide detail on the certification process.
2.4 Setting the Reference
A reference must be set when OMNIScanner2 and OMNIRemote are used as a set for the first time. After the initial reference setting, the process should be repeated periodically to insure unit verification. OMNIScanner stores reference data for up to 5 remote units by serial number.
Modular 8
Channel Adapter
Modular 8 Link Adapter
26
OMNISCANNER
1. To set the reference, connect OMNIScanner and OMNIRemote using the test cable and adapter supplied with the scanner (as shown).
2. Turn OMNIScanner on.
3. Press (Setup).
4. Select Set Reference.
5. On finding OMNIRemote, the scanner reports: Acquiring Reference Data. A gauge will indicate the progress.
6. Once the reference values has been recorded, OMNIScanner displays the REMOTE INFO screen.
7. Press ESC to return to OMNIScanner’s SETUP screen.
2.5 Project Selection
A project name must be selected for test-result management. Autotests and Cables were previously assigned to a specific project through the use of the OMNIScanner Configuration utility. (See Chapter 9, Uploading and Printing, for further information about OMNIScanner Configuration.)
1. Use the up-arrow or down-arrow to highlight Project and press ENTER.
2. Select the appropriate Project from the drop-down list and press ENTER.
27
CHAPTER 2: Getting Started: Twisted Pair Networks
3. Use the up-arrow or down-arrow to highlight Autotest and press ENTER.
4. Select the appropriate Autotest from the pop-up list and press ENTER.
5. Use the up-arrow or down-arrow to highlight Cable and press ENTER.
6. Select the cable type to be tested from the pop-up list and press ENTER.
28
OMNISCANNER
The Adapter field will automatically change to reflect the adapter that is currently attached. OMNIScanner auto-senses the adapter. If the adapter is changed, OMNIScanner will automatically adjust the content in the fields Autotest: and Cable: to reflect the new setup.
Every Project has certain Autotests and cables assigned already. These associated
lists will appear in the pop-up lists for each item. If the Autotest or Cable you want
to use is not in a pop-up list, open (Setup) to select the appropriate item.
(See Chapter 3, Setup: Autotest, Cable, or Project, for further information.)
NOTE
2.6 Running an Autotest for Twisted Pair Cabling
1. Connect OMNIScanner to the near end of the twisted-pair cable you are testing and connect OMNIRemote to the far end.
(See the Typical Test Configurations for Twisted Pair Cabling section in this chapter for more information.)
2. Press the key. When testing twisted-pair cable, OMNIScanner will immediately verify that OMNIRemote is attached. If OMNIRemote is not attached, OMNIScanner will wait until it is attached before running the Autotest.
3. Press ESC at any time to cancel the Autotest.
Modular 8 Jack
Modular 8
(RJ-45)
Link Adapter
29
CHAPTER 2: Getting Started: Twisted Pair Networks
4. The Autotest screen is displayed while Autotest is running.
The Autotest and the Cable name are displayed at the top of the screen. A gauge indicates the test progress. Tests will be listed with their individual
(PASS) or (FAIL) results as soon as a test is completed.
Once all tests have been performed, the overall test Result: is displayed as PASS or FAIL.
NOTE
The list of tests varies depending on the kind of Autotest that was performed.
An Autotest will PASS if all the selected tests meet the specified criteria. An
Autotest will FAIL if one or more selected tests failed.
2.6.1 S
AVINGTWISTEDPAIRAUTOTESTRESULTS
1.Press (Save) from the Autotest screen to open the SAVE AUTOTEST screen, where the Autotests can be stored into OMNIScanner’s memory.
2. The Project field contains the name of the project.
30
OMNISCANNER
3. To assign a unique name to the Autotest, enter a Circuit ID into the Circuit ID: field. Circuit IDs can be set up to automatically increment. (See Chapter 3, Setup: Editing a Circuit ID, for further information.)
If an incremental Circuit ID was created in the PROJECT Setup screen, the Circuit ID: field will already contain the next available Circuit ID that can be used as a default.
4. To change the name using OMNIScanner’s edit feature, highlight the field, and press ENTER to initiate the editing mode. (See Section 1.9 for more information.)
5. To change the Name using the list function, press (List). The CIRCUIT ID screen opens and displays all Circuit IDs available for the project. Use the left-arrow or right-arrow to quickly scroll through the pages. Use the up-arrow or down-arrow to highlight a Circuit ID and press ENTER to select it as the name for the Autotest to be saved and return to the SAVE AUTOTEST screen.
6. Press (Autotest Save Options) to open the SAVE OPTIONS screen. OMNIScanner: Select the Destination for saved test results (MMC or internal)
31
CHAPTER 2: Getting Started: Twisted Pair Networks
and select or unselect the graphs to be saved. (See Section 3.7, Autotest Options for more information.
Press (Save) to save the new settings and return to the SAVE AUTOTEST screen.
7. The SAVE AUTOTEST screen‘s save function key indicates the selected Save Option.
Press (Save) to save the test results to OMNIScanner’s internal memory.
Press (Save with Graphs) to save the test results with their graphs to OMNIScanner’s internal memory.
NOTE
Saving graphs will reduce the number of Autotests that can be stored.
Press (Save to MMC) to save the test results to the MultiMedia Card (OMNIScanner 2 only).
Press (Save with Graphs to MMC) to save the test results with their graphs to the MultiMedia Card (OMNIScanner 2 only).
NOTE
Saving graphs will reduce the number of Autotests that can be stored.
8. Press ESC to return to the Autotest screen without saving the Autotest.
9. To run the next Autotest, press the key.
32
OMNISCANNER
2.6.2 V
IEWINGTWISTEDPAIRAUTOTESTRESULTS
This viewing option is only available immediately after the Autotest was performed.
To view stored test results at a later time, press the (Results) function
key located in the Project screen. (See Chapter 6, Results, for more information.)
NOTE
The list of tests varies depending on the kind of Autotest that was performed.
1. To view detailed results of all the tests that were performed during Autotest, highlight a specific test in the Autotest screen using the up-arrow or down­arrow key and press ENTER.
2. Use the left-arrow or right-arrow to quickly browse through all other detailed test result screens and their results.
All Tests that can run during Autotest are displayed on the next page. (See Chapter 4, Measurements, for detailed descriptions of each available test.)
33
CHAPTER 2: Getting Started: Twisted Pair Networks
Press
to view
34
OMNISCANNER
Most Autotest result screens have a retest feature for diagnostic purposes.
Press (Retest) to run continuous tests for the same cable. The arrow
located on the Retest button will be turning for the duration of the test.
Press (Retest) once more to stop the continuous measurement.
OMNIScanner will display the last measured data.
NOTE
Retest invalidates an Autotest. After using the retest feature, the Autotest must be repeated before results can be saved.
The NEXT, Attenuation, Return Loss, Power Sum NEXT, ELFEXT, ACR, Power Sum ELFEXT, and Power Sum ACR test-result screens have a graph feature to display the graph associated with the test. To avoid misinterpretation, the limit line is shown in a gray color, while the actual measurement is represented in black.
Press (Graph) to display a graph for a pair or a pair combination.
The Length results screen has a (Detail) function key that allows you to view detailed test results for each pair.
35
CHAPTER 2: Getting Started: Twisted Pair Networks
Tests that require the use of OMNIRemote will have small icons in the upper
right corner of the titlebar to indicate whether the information is pertinent for
OMNIScanner or OMNIRemote.
To toggle between the two screens, press to open the screen that contains the information for OMNIScanner.
Press to display the information for OMNIRemote.
36
OMNISCANNER
The RETURN LOSS and NEXT Graph screens have (Diagnostic) function keys to help you with the interpretation of failed test results. (See Chapter 5, TDRL and TDNXT for detailed information.
2.7 Typical Test Configurations for Twisted Pair Cabling
2.7.1 B
ASICLINK
The graphic shows the connection of OMNIScanner to a Modular 8 Jack in the wiring closet, and OMNIRemote to a Modular 8 jack in the work area.
Use the supplied Link Adapters at both ends of the link to attach OMNIScanner and OMNIRemote to the Modular 8 jacks.
2.7.2 B
ASICLINK WITH
110 B
LOCK ANDMODULAR
8 J
ACK
The graphic shows the connection of OMNIScanner to an IDC (110) block in the wiring closet, and OMNIRemote to a Modular 8 jack in the work area.
Modular 8 Jack
Modular 8
(RJ-45)
Link Adapter
37
CHAPTER 2: Getting Started: Twisted Pair Networks
The Punch Down Block Adapter kits 110 A Link Adapter and 110 B Link Adapter are available (as optional accessories) to interface the OMNIScanner and OMNIRemote with IDC (110) punch-down blocks.
Use the 110 A or B Link Adapter to attach OMNIScanner to the 110 Punch Down block. At the other end of the link, use the 110 A or B Link Adapter to attach OMNIRemote to the Modular 8 jack.
2.7.3 F
ULLCHANNEL WITHMODULAR
8 J
ACKS
A Full Channel consists of all the link elements from the hub to the workstation.
For cables terminated with a Modular 8 plug, use the supplied modular 8 Channel Adapter to attach OMNIScanner and OMNIRemote to channel patch cords.
110 Block
Modular 8 Jack
Modular 8
Link Adapter
110 A Link Adapter
110 Block
User Cable
Modular 8
Channel Adapter
Modular 8 Jack
38
OMNISCANNER
NOTE
Best results for Category 6 channel tests will be obtained when using OMNIScanner software version 4.0 or higher.
2.7.4 P
ERMANENT
L
INK
For Permanent Link Autotests, connect the supplied Link adapters in the same manner as the Basic Link.
During permanent link Autotests, the effects of the link adapters are removed from the measurements.
The Permanent Link method measures the link only, from jack to jack (including the Modular 8 plug that plugs into the Jack).
Modular 8 Jack
Modular 8
(RJ-45)
Link Adapter
39
CHAPTER 3: Getting Started: Fiber Optic Networks
3. Getting Started: Fiber Optic Networks
3.1 Charging OMNIScanner’s Battery
In order to operate the OMNIScanner and OMNIRemote with battery power, the battery must first be charged. Until the charging cycle is complete, the units can run with the AC adapter plugged in. (See Appendix B, Batteries, for more information about the batteries.)
3.2 OMNIScanner’s Project Screen
The project screen is designed to allow you to perform all typical work flow operations. The Project, the test specifications, and the cable types are all selectable from the Project screen.
Press the ON/OFF key to power the unit on and display the project screen with
its three user definable items: Project, Autotest, and Cable.
The first item Project name will be highlighted. Use the down-arrow to move the highlight to the Autotest currently selected. Use the down-arrow to highlight the currently selected Cable type. OMNIScanner will auto-sense the Adapter.
Battery and Memory status are graphically indicated by gauges with a display ranging from empty (left hand side) to full (right hand side).
Press (Setup) to select Autotests, configure Cables and Projects, view Fiber Adapter information, edit User Options, and more.
Press (Measurements) to measure a cable’s performance, utilizing a variety of tests.
Press (Results) to view and manage stored test results.
40
OMNISCANNER
3.3 Certification of Fiber Installations
OMNIScanner and OMNIFiber are designed to efficiently certify fiber optic cabling installations through the Autotest feature. Test results, test specifications, and cable types are organized into Projects. The Project, its test specifications and cable types are selectable from the OMNIScanner Project screen.
OMNIScanner contains a number a predefined projects, grouped by test standards, to allow easy and fast operation.
The typical order of operations performed when certifying a cable installation is as follows:
P
REPARATION
Prior to initial certification, perform the following:
1. Attach an OMNIFiber adapter to OMNIScanner. This will be referred to as the OMNIFiber Main.
2. Attach an OMNIFiber adapter to OMNIRemote. This will be referred to as the OMNIFiber Remote.
3. Establish a reference power by setting the reference. See Chapter 8 - Fiber Optic Reference Methods for further information.
4. Select the appropriate Project, Autotest specification, and Cable type.
C
ERTIFICATION
Perform the following for each installed fiber to be tested:
5. Attach OMNIFiber Main to one end of the link, which is typically at the patch panel in a wiring closet.
41
CHAPTER 3: Getting Started: Fiber Optic Networks
6. Attach OMNIFiber Remote to the corresponding patch panel or outlet at the other end of the cable link.
7. Run Autotest.
8. Save test results.
C
OMPLETION
At the end of the day or shift, the test results are uploaded using Scanlink Tools software.
Scanlink Tools support the following for fiber Autotests:
• Upload test results
• Storage of test results to files (Scanlink proprietary format)
• Print test reports
• Export results in CSV format
• Setup Projects
• Edit Autotest Information
3.4 Setting the Reference for Fiber Optic Cabling
NOTE
To ensure consistent accurate measurements, the reference value must be set at least once each calendar day that the Fiber Adapters are used. The reference value expires each day at midnight.
1. To set the reference value, connect the OMNIFiber main and OMNIFiber Remote using launch cables and couplers. See Chapter 8, Fiber Optic Reference Methods, for further information.
2. Press (Setup) and then select Set Reference to record the reference value.
42
OMNISCANNER
3. Select the appropriate Reference Method: A, B, or C.
OMNIScanner will store a separate reference value for each wavelength. OMNIScanner will store reference data for up to 5 OMNIFiber pairs. See Chapter 8, Fiber Optic Reference Methods for further information.
NOTE
Once the reference value has been set, ensure that both launch cables remain attached to the fiber adapters. When setting the Reference using Loopback or Method B, ensure that the TX (transmit) connections are not disturbed.
43
CHAPTER 3: Getting Started: Fiber Optic Networks
4. Once the reference value has been set, OMNIScanner will display the FIBER INFO screen.
5. Press ESC to return to the SETUP screen.
NOTE
If the test configuration changes, a new reference value needs to be established.
3.5 Project Selection
A project name must be selected for test result management. Autotests and Cables were previously assigned to a specific project through the use of the OMNIScanner Configuration software. (See Chapter 9, Uploading and Printing for further information about OMNIScanner Configuration.)
The Project feature is very useful for grouping test results and eases uploading
and sorting of Autotests stored in the scanner.
1. Use the up-arrow or down-arrow to highlight Project: and press ENTER.
2. Select the appropriate Project from the pop-up list and press ENTER.
44
OMNISCANNER
3. Use the up-arrow or down-arrow to highlight Autotest: and press ENTER.
4. Select the appropriate Autotest from the pop-up list and press ENTER.
5. Use the up-arrow or down-arrow to highlight Cable: and press ENTER.
6. Select the cable type to be tested from the pop-up list and press ENTER.
The Adapter: field will automatically change to reflect that OMNIFiber is
currently attached.
OMNIScanner auto-senses the adapter. If the adapter is changed, OMNIScanner will automatically adjust the content in the fields Autotest: and Cable: to reflect the new setup.
Every Project has certain Autotests and cables assigned already. These associated lists will appear in the pop-up lists for each item. If the Autotest or Cable you want to use is not in a pop-up list, open (Setup) to select the appropriate item. (See Chapter 4, Setup: Autotest, Cable, or Project for further information.)
45
CHAPTER 3: Getting Started: Fiber Optic Networks
3.6 Running An Autotest for Fiber Optic Cabling
3.6.1 P
REPARING ANAUTOTEST
1. Connect OMNIFiber Main to the near end of the RX/TX fiber pair you are testing and connect OMNIFiber Remote to the far end. Depending on the method that was used to obtain the reference, connect your fibers as follows:
METHOD A
To run an Autotest when Method A was used to set the Reference proceed as follows:
1. Do not disconnect the launch cables from the OMNIFiber Adapters.
2. Remove the connector adapters (sleeves) that connect the two launch cables.
3. Connect the OMNIFiber Main to one end of the duplex link under test, and the OMNIFiber Remote to the opposite end.
2 m Launch Cable
Patch Panel
TX
850 nm
1300/1310 nm
RX
1300/1310 nm
850 nm
RX
TX
2 m Launch Cable
METHOD A
Cable under Test
Wall Plate
46
OMNISCANNER
4. Press .
METHOD B
To run an Autotest when Method B was used to set the Reference proceed as follows:
1. Do not disconnect the black connectors which are attached to the TX port of the OMNIFiber Adapters.
2. Disconnect the black connectors that are attached to the RX ports of the OMNIFiber Adapters.
3. Connect the white connectors to the RX ports on the OMNIScanner fiber adapters.
4. Connect the OMNIFiber Main to one end of the duplex link under test and the OMNIFiber Remote to the opposite end.
5. Press .
METHOD C
To run an Autotest when Method C was used to set the Reference, proceed as follows:
2 m Launch Cable
Patch Panel
TX
850 nm
1300/1310 nm
RX
1300/1310 nm
850 nm
RX
TX
2 m Launch Cable
METHOD B
Cable under Test
Wall Plate
47
CHAPTER 3: Getting Started: Fiber Optic Networks
1. Do not disconnect the launch cables that are attached to the OMNIFiber adapters.
2. Remove the middle launch cable and the connector adapters (sleeves) that were attached to the middle launch cable.
3. Connect the OMNIFiber Main to one end of the duplex link under test and the OMNIFiber Remote to the opposite end.
4. Press .
3.6.2 L
OOPBACKAUTOTEST USING
OMNIF
IBERMAIN
This configuration is used to test single fibers, reels and short cables when both ends of the fiber under test are accessible from the same location. Note: The loopback Autotest assumes that reference Method A (2 jumper method) was used to make the reference measurement.
To run a Loopback Autotest, proceed as follows:
1. Do not disconnect the black launch cable, which is attached to the TX port of the OMNIFiber adapter (it is circled in the graphic).
2 m Launch Cable
Patch Panel
TX
850 nm
1300/1310 nm
RX
1300/1310 nm
850 nm
RX
TX
2 m Launch Cable
METHOD C
Cable under Test
Wall Plate
48
OMNISCANNER
2. Do not disconnect the coupler from the black launch cable (it is circled in the graphic).
3. Disconnect the white launch cable from the RX port of the OMNIFiber adapter and the connector adapter (sleeve). Leave the two white connectors dangling.
4. Attach the cable to be tested to the RX port on the OMNIScanner Fiber adapter and the connector adapter (sleeve).
5. Press .
3.7 Running a Fiber Autotest
Some Autotests do not have a fixed loss limit. The Optical Link Budget (OLB) is calculated based on the wavelength, the length of the cable, and the number of connectors and splices.
OMNIScanner will automatically display the CONNECTIONS screen if you are
running an equation based Autotest (TIA 568A, ISO 11801, or 1000BASE-F).
1. With the Connections: field highlighted, press ENTER. Use the up-arrow or down-arrow key to adjust the number of connections in your link. To determine the total number of connections, simply count all of the connections (connector pairs) on one fiber. OMNIScanner will assume that both fibers of a duplex link have the same number of connections.
1300/1310 nm
850 nm
Cable under Test
RX
TX
2 m Launch Cable
Fixed Connection
Fixed Connection
49
CHAPTER 3: Getting Started: Fiber Optic Networks
2. To change the number of Splices, use the down-arrow to highlight the field and press ENTER. Use the up-arrow or down-arrow key to adjust the number of splices in your link. To determine the total number of splices, simply count all of the splices on one fiber. OMNIScanner will assume that both fibers of a duplex link have the same amount of splices.
3. After editing the fields, press to run the Autotest.
OMNIScanner starts the Autotest after the number of connections and splices
has been entered.
NOTE
If you are running a Fiber Test Options for further information.)
4. Swap the Launch Cable’s connectors as indicated in the graphic. Do not disconnect the Launch Cables from the OMNIFiber adapters!
5. Press to complete the Autotest.
50
OMNISCANNER
3.8 Fiber Autotest Results
Once the Autotest is completed the Autotest screen will be displayed.
The arrows at the top of the screen indicate the direction of the light.
‚ points to the result measured from OMNIRemote to OMNIScanner.
ˇ points out the result measured from OMNIScanner to OMNIRemote.
The overall test result for each fiber is located right below the arrows,
(PASS) or (FAIL).
The name of the Autotest is displayed at the top of the screen followed by the
name of the fiber.
When a bidirectional test was performed an icon for OMNIScanner or
OMNIRemote will appear next to the fiber‘s
name.
Press or to view results for the fibers with light transmission in the opposite direction. The arrows flip vertically to indicate that there are two results for each fiber, one from TX to RX and one from RX to TX.
2 m Launch Cable
TX
850 nm
RX
1300/1310 nm
850 nm
RX
TX
2 m Launch Cable
Cable under Test
1300/1310 nm
51
CHAPTER 3: Getting Started: Fiber Optic Networks
NOTE
Values measured with OMNIFiber will be compared against the preprogrammed PASS/FAIL limits. OMNIScanner will report independent PASS/FAIL indication for both fibers.
The Autotest Results screen displays the (Length) of multimode fiber at 850 and 1300 nm, or the loss of single-mode fiber at 1310 nm.
OMNIScanner displays the margin in the RESULTS screen. It is the worst case margin for all wavelengths on each fiber in both directions. A positive margin indicates that the measured loss was less than the PASS/FAIL value, a negative margin indicates a fail and shows how much the measurement exceeded the maximum allowable PASS/FAIL amount.
The number of Connections and Splices is displayed at the bottom of the screen (if applicable).
To store the Autotest Result for the fiber listed on the right side, press the
(Save) function key located on the right side.
To store the Autotest Result for the fiber listed on the left side, press the (Save) function key located on the left side.
52
OMNISCANNER
4. Setup
4.1 Configuring OMNIScanner
OMNIScanner’s easy-to-use setup feature allows you to define, edit and preset many items. They are conveniently organized and can easily be accessed by pressing the assigned number on your alphanumeric keypad, or by highlighting the item and then pressing ENTER.
From the Project screen, press the (SETUP) screen.
53
4.2 Set Reference
Connect OMNIScanner and OMNIRemote through their test connector ports using the test cable and the MOD 8 adapter. Select Set Reference to display the SET REFERENCE screen. OMNIScanner will immediately start to communicate with OMNIRemote.
While the reference data is being established, a gauge indicates the progress.
OMNIScanner notifies you when the process has been completed. The Remote Info screen will be displayed.
NOTE
Set Reference is only needed to periodically verify and establish the signal reference for attenuation. The function does not affect any other measurements.
Press ESC to return to the Setup screen.
CHAPTER 4: Setup
54
OMNISCANNER
S
ET
R
EFERENCE WITH THE
OMNIF
IBERADAPTERS
NOTE
The fiber reference expires each day at midnight. It must be set each calendar day that OMNIFiber adapters are used. The reference value should also be set every time a fiber is removed from either OMNIFiber TX port.
A reference value must be established before a loss measurement can be calculated.
With the OMNIFiber adapters attached to OMNIScanner2 and OMNIRemote, and the launch cable(s) in place, select Set Reference to display the Select Method screen.
NOTE
The Select Method screen will not be displayed if a fiber is attached to the TX and RX ports of the same OMNIFiber adapter (Loopback).
To select a Reference Method A, B, or C, press the appropriate function key.
See Chapter 8, Fiberoptic Reference, for detailed information about establishing a reference value with fiberoptic launch cables.
55
The Set Reference screen will be displayed while the reference data is being established.
OMNIScanner notifies you when the process has been completed. The Fiber Info screen will be displayed.
Press ESC to return to the Setup screen.
4.3 Autotests
Select Autotests to display the Autotests screen. It contains a list of standard Autotest specifications supported by the OMNIScanner2. Standards are grouped by TIA, ISO, VENDOR, MICROTEST, IEEE, ATM, and AS/NZ.
CHAPTER 4: Setup
56
OMNISCANNER
If there are Autotest specifications in a folder, it will appear full ; otherwise it will be empty .
Highlight a folder containing items and press ENTER to display the list of Autotests contained in that folder.
Use the up-arrow or down-arrow key to highlight a specific Autotest.
Copper tests are marked with a twisted pair icon.
Fiber optic tests are marked with a fiber icon.
Press (Set Default) to select it as the Autotest to run.
Press (Detail) to review the Autotest’s limits.
57
V
IEWINGDETAILS FORCOPPER ANDFIBERAUTOTESTS
For twisted-pair cabling, the Autotest Detail screen consists of four pages.
Use the and function keys to browse through the pages.
The screen titles show the standard being used to set the limits (for example, Cat 6 Link).
The first page displays the Configuration of the Autotest (Basic Link, Channel, or Permanent Link). Wiremap, Length, Delay, and Skew limits are also displayed. The Wiremap numbers show the pairs that are being tested.
The second page shows limits for Resistance, NEXT, Attenuation, and Return Loss.
CHAPTER 4: Setup
58
OMNISCANNER
The third page lists limits for ELFEXT, ACR, and Power Sum NEXT measurements.
The fourth page shows limits for Power Sum ELFEXT and Power Sum ACR. This screen also informs you whether or not OMNIRemote is required to perform the selected Autotest.
Tests that are performed during an Autotest are clearly marked . Tests that are not required when performing an Autotest have no icon.
Disabled tests may or may not show limits in the description field. An precedes the test name for tests being performed during the Autotest, but the results are not used to determine PASS or FAIL.
Defined test limits will be displayed right below the test name. If a test has no limit, the limit description field will be empty.
i
59
For fiberoptic cabling the View Autotest Detail screen shows the standard being used in the titlebar.
The tests that are performed during an Autotest are clearly marked .
If an precedes the test name, the test will be performed for information purposes only. The results are not used to determine PASS or FAIL.
The View Autotest Detail screen displays the current scanner configuration for multimode fibers. Use the function key to view the limits for single-mode fibers.
When viewing limits for equation-based Autotests like TIA 568A, ISO 11801, and 100BASE-F, the Loss per km (kilometer) will be displayed.
ii
CHAPTER 4: Setup
60
OMNISCANNER
For all other tests, the Loss for 850, 1300, and/or 1310 is displayed.
Press ESC to return to the Setup screen.
4.4 Cables
Select Cables to display the Cables screen. It contains a list of cable manufacturers.
If there are cables in a folder the manufacturer folder will appear full, otherwise it will be empty.
Highlight a cable manufacturer that is marked by a folder containing items
and press ENTER to display the list of cables in that manufacturer’s folder.
61
CHAPTER 4: Setup
Use the or to highlight the desired cable.
Copper cables are marked with a twisted pair icon.
Multimode fiber cables are marked with a multimode fiber icon.
Single-mode fiberoptic cables are marked with a single-mode fiber icon.
Coaxial cables are marked with the COAX icon.
Press (Set Default) to choose it as the cable to be tested.
Selecting the correct cable type will ensure accurate measurements. If you cannot find or do not know the name of your cable manufacturer, simply select a cable from the Category folder, or create a new cable (see below).
Press (Delete) to permanently remove a custom cable.
NOTE
Default cables cannot be removed.
With a twisted pair or a coax cable highlighted, press (Edit) to edit an existing custom cable. Cable Name, NVP, or Impedance can be changed.
NOTE
Default cables cannot be edited.
62
OMNISCANNER
With a twisted pair or a coaxial cable highlighted, press (Edit) to edit an existing custom cable. Cable Name, NVP, or Impedance can be changed.
NOTE
Default cables cannot be edited.
With a twisted pair or coax cable highlighted, press (New) to create a new cable based on the NVP and Impedance values of the highlighted default cable. Name, NVP, and Impedance are definable.
NOTE
A new or edited cable will be saved in the Custom Cables folder.
For an accurate Length test, the NVP (Nominal Velocity of Propagation) must be set correctly. The value is usually specified by the cable vendor, and OMNIScanner has a large library of cables with the manufacturer’s specified NVP.
O
63
CHAPTER 4: Setup
To determine the NVP of a cable with a known length, press (Measure) to capture the length of the cable using the displayed NVP. Use the down-arrow to highlight the NVP field and press ENTER. Use the up-arrow or down-arrow key to increase or decrease the value by 1% (percent) and adjust the NVP until the known length is displayed. The Length changes accordingly.
NOTE
The known cable must have a length of at least 50 feet or 15 meters.
To change the impedance, use the down-arrow to highlight the field and press ENTER. Use the up-arrow or down-arrow key to select the appropriate Impedance
in (ohms) from the predefined list.
To setup whether or not the cable is shielded, use the down-arrow key to highlight Shield: and press ENTER. Use the up-arrow or down-arrow key to make the appropriate selection. If “Yes” is selected, the wiring for the shield will be measured during the Autotest and used to determine if the Wiremap will PASS or FAIL.
With a or fiberoptic cable highlighted, press (Edit) to edit an existing custom cable, or press (New) to create a new fiber based on the GRI values of the highlighted cable. Cable Name and GRI @ 850 and 1300 can be viewed and changed.
NOTE
Default fibers cannot be edited.
64
OMNISCANNER
A fiber cable’s Graded Refractive Index (GRI) differs depending on the supplier. To receive accurate length results, OMNIScanner allows you to define the GRI values for both wavelengths.
To change the fiber’s mode, use the up-arrow or down-arrow key to highlight the mode. Press Enter to select Multimode or Single-mode.
Press (Save) to save the edited cable into the Custom Cable folder.
Press ESC to return to the SETUP screen.
4.5 Projects
Select Projects to display the Projects screen. It contains a list of all projects that are currently stored in OMNIScanner. Predefined default projects are: TIA, ISO, VENDOR, MICROTEST, IEEE, ATM, AS/NZ, ANSI. OMNIScanner’s project feature allows you to organize test specifications, cable types, and test results into Projects.
If there are tests stored in a project, the folder will appear full ; otherwise it will be empty .
Use the up-arrow or down-arrow key to highlight a specific Project.
65
Press (Set Default) to select it as the Project to be used and to return to the Setup screen.
Press (Delete) to permanently remove a project.
NOTE
Default projects cannot be deleted. Only projects without stored Autotests can be deleted.
Press (Edit) to edit an existing project name and its Circuit ID.
The project name will appear in the Name: field. The first and last Circuit ID fields contain a default circuit ID. (See Editing a Circuit ID for further information.)
Press (List) to display the Circuit ID screen. It contains a scrollable list of all Circuit IDs created for the project.
CHAPTER 4: Setup
66
OMNISCANNER
Press (New) to create a new project.
New Project will be displayed in the Name: field. Press ENTER to edit the name. The first and last Circuit ID fields contain the default Circuit ID from the Project that was highlighted when the (New) key was pressed.
NOTE
New and edited project names are placed at the bottom of the list in the PROJECTS screen. OMNIScanner can store up to 16 projects.
E
DITING ACIRCUIT
ID
To edit or create a Circuit ID, highlight an item in the PROJECTS screen and press
(New) or (Edit).
Use the down-arrow to highlight the First Circuit ID: field and press ENTER.
The Circuit ID field consists of 16 characters, which allows the sophisticated construction of Circuit IDs. Each character can be incremented.
The example shows how to quickly create an incremental Circuit ID.
Enter MT-F01 (MICROTEST - Floor 01) -PP01 (-Patch Panel 01) -P01 (Position
01) in the First Circuit ID: field.
Press ENTER to terminate the edit mode.
Use the down-arrow key to highlight the Last Circuit ID: field and press ENTER.
67
Enter MT-F12 (MICROTEST-Floor 12) -PP02 (Patch Panel 02) -P48 (Position
48) in the Last Circuit ID field. Press ENTER to terminate the edit mode.
The Circuit IDs will increment from right to left. The positions will change first from 01 to 48. Next the Patch Panels will increase from 01 to 02. Last, the floors will increase from 01 to 12.
Press (List) to review the continuous list of circuit IDs that were just created.
MT-F01-PP01-P01 (1st Circuit ID to be used)
MT-F01-PP01-P02 (2nd Circuit ID to be used)
MT-F01-PP01-P03 (3rd Circuit ID to be used)
MT-F01-PP02-P01 (49th Circuit ID to be used)
MT-F02-PP01-P01 (97th Circuit ID to be used)
MT-F12-PP02-P48 (last Circuit ID to be used)
Press (Save) to save the Circuit ID and return to the Projects screen.
NOTE
To quickly edit the Last Circuit ID, press (Copy) once you have completed the First Circuit ID. This function will copy the First Circuit ID into the Last Circuit ID field. Now edit the Last Circuit ID field.
Press ESC to return to the SETUP screen.
CHAPTER 4: Setup
68
OMNISCANNER
4.6 Configurations
Select Configurations to display the CONFIGURATIONS screen. It contains a list of all configurations which are currently stored on the MultiMedia Card that can be read by OMNIScanner2. (See Chapter 11, MultiMedia Card for more information about the MMC).
The predefined Default Configuration is always present and can be used to reset OMNIScanner to the factory default. It contains the following projects: TIA, ISO/IEC, VENDOR, Microtest, IEEE, ATM, AS/NZ, ANSI.
OMNIScanner2 only: Use the up-arrow or down-arrow key to highlight a specific Configuration.
Press (Set Default) to select it as the Configuration to be used and to return to the SETUP screen.
While the selected Configuration is being loaded onto OMNIScanner2, a gauge indicates the progress. Do not
remove the MultiMedia Card from the MultiMedia
Card reader during the data transfer.
4.7 Autotest Run Options
Select Autotest Run Options to display the RUN OPTIONS screen. With the presented options you are able to further customize the way OMNIScanner will perform an Autotest.
Press if you want OMNIScanner to abort the Autotest if a FAIL occurs. The Abort on Fail box will be checked .
69
Use the down-arrow to highlight the other checkbox.
Press if you want OMNIScanner to automatically move to the failed test result. The Jump to Fail box will be checked .
Use the down-arrow to advance to the Report: field and press ENTER. Select Margin if you want OMNIScanner to report the result, measured in dB, for the worst margin. Select Worst dB if you want the worst overall dB value to be reported, as in OMNIScanner software versions prior to v4.0. Press ENTER to set the default.
NOTE
Attenuation results in dB will always be reported for the worst value, regardless of what is selected in the Report: listbox.
Use the down-arrow to highlight the Fiber Test Options checkbox. You will be able to define the way fiber optic Autotests are performed.
Press if you want OMNIFiber to test the fiber link in both directions. The Test in Both Directions box will be checked .
Press (Save) to save your changes and return to the SETUP screen.
Press ESC to discard all changes and return to the SETUP screen.
4.8 Autotest Save Options
Select Autotest Save Options to display the SAVE OPTIONS screen. This screen contains several options that allow you to customize the way OMNIScanner will save Autotests.
CHAPTER 4: Setup
70
OMNISCANNER
OMNIScanner2 only: Press ENTER to change the Destination: from MMC (MultiMediaCard™) to Internal. Use the up-arrow and down-arrow keys to highlight the desired destination and press ENTER to select it as the default destination. Select MMC if you want to collect test results on the MultiMediaCard. (See Chapter 11, MultiMediaCard for more information about MMC.) Select Internal if you want to save the results to OMNIScanner’s internal memory. In either case, results can be transferred to the Scanlink Tools software program. (See Chapter 9, Uploading and Printing, for more information.)
In the Save Graph Options section, you will be able to define which of the graphs will be saved.
If a checkbox is checked , OMNIScanner will store an Autotest with the graph information for that particular test.
NOTE
Saving graphs reduces the number of Autotests that can be stored.
Depending on the performed Autotest, NEXT/PSNEXT, ELFEXT/PSELFEXT, Return Loss, and Attenuation graphs can be saved.
Use the up-arrow and down-arrow keys to highlight the Save Graph Option checkboxes.
Press to select the graph to be saved.
Press (Save) to save your changes and return to the SETUP screen.
Press ESC to discard the changes and return to the SETUP screen.
71
CHAPTER 4: Setup
4.9 User Options
Select User Options to display the USER OPTIONS screen. This screen contains several user-definable preferences.
Press ENTER to switch the Units from Meters to Feet. Use the up-arrow and down-arrow keys to highlight the desired measurement unit and press ENTER to select it as the default measurement unit.
Use the right-arrow to advance to the Auto Shutoff field and press ENTER. Select YES if you want OMNIScanner to automatically shut off after 10 minutes when operated with battery power. Select NO if you don’t want to use the function. Press ENTER to set the default.
To edit the Date use the right-arrow key to highlight the portions you want to change. Press ENTER to edit the highlighted segment. Use the up-arrow and down-arrow keys to change the displayed values in increments of 1. Press ENTER when done.
To edit the Time use the right-arrow key to highlight the portions you want to change. Press ENTER to edit the highlighted segment. Use the up-arrow and down-arrow keys to change the displayed values in increments of 1. Press ENTER when done.
Use the right-arrow to highlight the Owner Name. Press ENTER to start the edit mode. You can enter up to 19 characters for the Owner Name. They will appear on printed reports. Press ENTER to end the edit mode.
The USER OPTIONS screen displays the Language that is currently selected. To change the language, use the right-arrow key to highlight the field. Press ENTER
72
OMNISCANNER
and use the up-arrow and down-arrow keys to highlight a language. Press ENTER once more to select it as the default language.
Press to save all settings and return to the SETUP screen.
4.10 Adapter Usage
Select Adapter Usage to display the ADAPTER USAGE screen.
The number of Autotests that have been performed with the Adapter that is currently attached to OMNI: and Remote: is displayed below each adapter’s name (Adapter ID String).
Press ESC to return to the SETUP screen.
4.11 Remote Info
Select Remote Info to display the REMOTE INFO screen.
The date the Reference value for OMNIScanner and OMNIRemote was last set is shown at the top of the screen.
The Hardware and Software Versions, and the Serial number of the remote unit are displayed.
73
CHAPTER 4: Setup
The Battery status is represented as a gauge. Factory calibration* data is displayed at the bottom of the screen.
You may need to provide Serial Number and Factory Calibration data when having repairs or refurbishments done, or when calling for Technical Support.
*NOTE
OMNIScanner and OMNIRemote should be calibrated annually with specialized equipment. Contact Technical Support for information on calibration and service requirements.
4.12 Fiber Adapter Information
Select Fiber Info to display the FIBER INFO screen.
The serial numbers and the date the adapters were last factory calibrated are displayed.
74
OMNISCANNER
The currently used reference method(s) are displayed on the bottom of the screen.
Press ESC to return to the SETUP screen.
4.13 Format MMC
OMNIScanner2 only: When using the MultiMediaCard for the first time, it may require formatting.
Insert the MMC into OMNIScanner2’s MMC slot.
Select Format MMC to open the FORMAT MMC screen. Formatting will start immediately, a gauge will indicate the progress.
The formatting time depends on the capacity of the MultiMediaCard; 8 MB will require 2 minutes.
75
CHAPTER 5: Measurements
5. Measurements
NOTE
The list of tests in the Measurement screen changes depending on the adapter that is currently attached to OMNIScanner.
5.1 Performing Individual Measurements with OMNIScanner
OMNIScanner allows you to perform tests on twisted pair, fiber optic, and coaxial cabling.
5.1.1 M
EASURINGTWISTED
P
AIRCABLING
OMNIScanner allows you to run various individual tests. The tests are conveniently grouped into two screens and can be accessed separately with the alphanumeric keypad.
Press the (Measurements) function key to display the MEASUREMENTS screen.
Press the (Page Down) function key to scroll to the second screen containing MEASUREMENTS.
Press the (Page Up) function key to return to the first screen containing MEASUREMENTS.
5.1.2 M
EASURINGFIBEROPTICCABLING
With the OMNIFiber adapter attached, press the (Measurements) function key to display the MEASUREMENTS screen for fiberoptic cabling.
76
OMNISCANNER
Highlight the test you want to perform using the up-arrow or down-arrow key on your keypad and press ENTER, or direct-dial a test by pressing its affiliated number on the OMNIScanner keypad.
5.1.3 M
EASURINGCOAXIALCABLING
With the coax adapter attached, press the (Measurements) function key to display the MEASUREMENTS screen for coaxial cabling.
Press ENTER or direct-dial the test by pressing its associated number on the OMNIScanner keypad.
5.2 General Information
Most test result screens have a retest feature for diagnostic purposes.
Press (Retest) to run continuous tests for the same cable.
Press (Retest) once more to stop the continuous measurement. OMNIScanner will display the last computed data.
77
CHAPTER 5: Measurements
Tests that require the use of OMNIRemote will have small icons in the upper right corner of the titlebar indicating whether the information on the screen is pertinent for OMNIScanner or OMNIRemote.
Toggle between the screens by pressing to open the screen that contains
the information for OMNIScanner, and to display the information for OMNIRemote.
NOTE
The PASS/FAIL criteria are based on the Autotest currently selected in the Project screen.
Several measurements will display certain results in a graph format. When a graph can be displayed, the (Graph) function key is available.
Press (Graph) to display test results for the highlighted pair or pair combination.
Use the up-arrow or down-arrow key to magnify the area of the graph where the cursor is located. The graph can be magnified 1x, 2x, 4x, or 10x.
Use the left-arrow or right-arrow keys to move the cursor within the graph.
78
OMNISCANNER
Use the cursor to display values for the desired frequency.
5.3 Twisted Pair Measurements with OMNIScanner
5.3.1 W
IREMAP
The Wiremap function tests twisted-pair cabling for proper wiring. Your cabling configuration is checked for shield continuity, opens, shorts, crossed pairs, split pairs, and reversed pairs. For each pair, Wiremap displays the length.
The screen displays a graphical representation of the wiring in your cable, where the near end wiring is shown on the left side, and the far end wiring on the right side of the screen.
1X
2X
10X
4X
79
CHAPTER 5: Measurements
• A reversed number indicates a split pair.
• A straight line between pairs indicates a good pair.
• No line between wire pairs indicates an open.
• An X between wire pairs indicates a reversed pair.
• A | between wire pairs indicates a short.
• The measured length of the pairs 1-2, 3-6, 4-5, and 7-8 are shown on the left side.
In the example, OMNIScanner displays split pairs on 1 - 2 and 4 - 5 although they are wired straight through, a reversed pair 3 - 6, an open on wire 7, and a short on wire 8 to the Shield.
Press ESC to exit the WIREMAP screen.
5.3.2 L
ENGTH/DELAY
The Length/Delay function measures a cable’s length, propagation delay, and skew. OMNIScanner will determine whether the cable is open or shorted.
The accuracy of the NVP (Nominal Velocity of Propagation) setting determines the exactness of length measurements. The NVP specified by cable manufacturers is used for all the cables in OMNIScanner’s library. When creating or editing cables, the NVP value can be customized. (See Section 3.5, Setup: Cables, for further information.)
Delay is a measure of the time required for a signal to travel from one end of the circuit to the other. OMNIScanner measures the delay from when the pulse was sent and when the reflection was received. Knowing the speed that the pulse
1
80
OMNISCANNER
propagates down the cable (NVP), OMNIScanner then calculates the length of the cable.
Skew is the difference between the delay on the longest and shortest wirepairs.
The Length/Delay screen shows the test result information for all pairs.
The Limit is displayed at the top of the screen.
The pairs are displayed in the first column followed by small icons to indicate whether the pair is open or short.
The measured Length and the Delay are displayed for each pair. The result for the shortest pair is tagged with a PASS or FAIL indicator.
The length result can be displayed in meters or feet (see Chapter 4: Setup User Options).
Delay is displayed in nS (nanoseconds).
The Skew value and the Skew limit are displayed right below the pairs.
Press ESC to exit the LENGTH/DELAY screen.
5.3.3 NEXT
The NEXT (Near End Crosstalk) function measures the electromagnetic coupling that occurs between pairs at the near end. OMNIScanner and OMNIRemote allow NEXT to be measured at both ends of the link without having to swap ends.
OMNIRemote functions as a terminator while OMNIScanner transmits a signal which is monitored on the receive pair. The received signal is compared to the transmitted signal and the difference is reported as the NEXT value in dB. When
81
CHAPTER 5: Measurements
measuring the NEXT for the OMNIRemote end, the roles of the two units are reversed. OMNIScanner provides the termination; OMNIRemote transmits and receives the signal.
In the NEXT screen, the worst case margin for all near and far pair combinations is shown at the top of the screen.
The pair combinations are displayed in the column on the left side, followed by the measured NEXT value, margin/limit, and frequency. The result for each pair combination is tagged with a PASS or FAIL indicator.
Press or to toggle between the NEXT results for OMNIScanner and OMNIRemote.
Press (Graph) to display the NEXT graph for the highlighted pair combination. NEXT is displayed for the frequency indicated by the cursor position.
82
OMNISCANNER
Press (Pairs) to view the NEXT graphs for other pair combinations.
Press (Diagnostics) to display the TDNXT screen, where cabling and/or connection faults can be identified. (See Chapter 6, Diagnostics, for further information.
Press ESC to exit the NEXT screen.
5.3.4 A
TTENUATION
The Attenuation function measures signal loss in a cable as a function of frequency. OMNIRemote generates and injects a signal at the far end of the cable. OMNIScanner measures the signal at the other end of the cable to determine cable loss.
In the ATTENUATION screen, the worst case margin for all pairs is shown at the top of the screen.
The pairs are displayed in the column on the left side, followed by the Attenuation in decibels (dB), the Attenuation margin/limit values, and the frequency. The result for each pair is tagged with a PASS or FAIL indicator.
Press (Graph) to display the Attenuation graph for the highlighted pair. The Attenuation is displayed for the frequency indicated by the cursor position.
Press (Pairs) to view Attenuation results for other pairs.
83
CHAPTER 5: Measurements
Press ESC to exit the ATTENUATION screen.
NOTE
Attenuation always reports the worst dB value regardless of the Report: setting in the RUN OPTIONS screen.
5.3.5 R
ETURNLOSS
The Return Loss function measures the uniformity of a cable’s impedance. OMNIScanner injects a signal into the cable and measures the reflected signal strength. Return Loss measures the difference between the transmitted signal delivered to the cable and the reflected signal at the input port of the cable. The reflected signal is the result of impedance mismatches in the link under test.
In the RETURN LOSS screen, the worst case margin for all near end and far pairs is shown at the top of the screen.
The pairs are displayed in the column on the left side, followed by the Return Loss in decibels (dB), the Return Loss margin/limit, and the frequency. The result for each pair is tagged with a PASS or FAIL indicator.
84
OMNISCANNER
Press or to toggle between the Return Loss results for OMNIScanner and OMNIRemote.
Press (Graph) to display the Return Loss graph for the highlighted pair. The Return Loss is displayed for the frequency indicated by the cursor position.
Press (Pairs) to view Return Loss graphs for other pairs.
Press (Diagnostics) to display the TDRL screen, where cabling and/or connection faults can be identified. (See Chapter 6, Diagnostics, for further information.)
Press ESC to exit the RETURN LOSS screen.
5.3.6 ELFEXT
ELFEXT (Equal Level Far End Cross Talk) measures the electromagnetic coupling at the far end and normalizes the result to account for the cable’s attenuation. ELFEXT is reported in dB and results are computed for twelve pair combinations at the OMNIScanner and OMNIRemote end of the cabling.
In the ELFEXT screen, the worst case margin for all near and far pair combinations is shown at the top of the screen.
The pair combinations are displayed in the column on the left side, followed by the ELFEXT value, the margin/limit, and the frequency. The result for each pair combination is tagged with a PASS or FAIL indicator.
85
CHAPTER 5: Measurements
The scrollbar indicates that there are more test results. Use the down-arrow on OMNIScanner’s keypad to scroll through the list and view results for the other six pair combinations.
Press or to toggle between the ELFEXT results for OMNIScanner2 and OMNIRemote.
Press (Graph) to display the ELFEXT graph for the currently highlighted pair combination. ELFEXT is displayed for the frequency indicated by the cursor position.
Press (Pairs) to view ELFEXT graphs for other pair combinations.
Press ESC to exit the ELFEXT screen.
5.3.7 ACR/B
ANDWIDTH
Attenuation to Crosstalk Ratio (ACR) is an indication of how much larger the received signal is compared to the Near End Crosstalk (NEXT) on the same pair. ACR is calculated from the measured NEXT and Attenuation. The ACR value is computed at each measured Attenuation frequency. The worst-case ACR is reported for each pair combination.
86
OMNISCANNER
Bandwidth is a measure of the usable frequency range of the link under test. For most networks, this is the highest frequency at which the signal can easily be distinguished from background noise. Depending on the selected Autotest, the worst pair to pair ACR is typically 3 dB.
In the ACR screen, the Bandwidth is displayed at the top of the screen.
The pair combinations are located in the column on the left side, followed by the measured ACR, the ACR margin/limit, and the frequency. The result for each pair combination is tagged with a PASS or FAIL indicator.
Press or to toggle between ACR results for OMNIScanner and OMNIRemote.
Press (Graph) to view the graph for the currently highlighted pair combination. The values for Attenuation and NEXT are shown at the top of the screen. Attenuation and NEXT graphs are displayed for the frequency indicated by the cursor position.
87
CHAPTER 5: Measurements
Press (Pairs) to view Attenuation/NEXT graphs for other pair combinations.
Press ESC to exit the ACR screen.
5.3.8 R
ESISTANCE
Resistance is measured in a loop through each pair in the cable. In a twisted-pair cable, Resistance is measured by using OMNIRemote to short the far end of the pair under test.
The RESISTANCE screen displays the Resistance Limit and the test results in ohms for each pair. The result for each pair is tagged with a PASS or
FAIL indicator.
Press ESC to exit the RESISTANCE screen.
5.3.9 P
OWERSUM
NEXT
Like NEXT, Power Sum NEXT is a measurement of electromagnetic coupling between adjacent pairs. While NEXT reports test results for each pair combination, Power Sum NEXT is calculated from NEXT for individual pair combinations and takes into account the crosstalk from multiple transmission pairs. When reporting Power Sum NEXT, OMNIScanner displays results for pairs, rather than pair combinations. The PS NEXT measurement is required to assure proper operation of networks employing simultaneous transmission on more than one pair in a link (Gigabit Ethernet, for example).
In the POWER SUM NEXT screen, the worst case margin for all near and far pairs is shown at the top of the screen.
88
OMNISCANNER
The pairs are displayed in the column on the left side, followed by the PS NEXT in decibels (dB), the margin/limit, and the frequency. The result for each pair is tagged with a PASS or FAIL indicator.
Press or to toggle between the PS NEXT results for OMNIScanner and OMNIRemote.
Press (Graph) to display the PS NEXT graph for the currently highlighted pair. PS NEXT is displayed for the frequency indicated by the cursor position.
Press (Pairs) to view PS NEXT graphs for other pairs.
Press ESC to exit the POWER SUM NEXT screen.
5.3.10 P
OWERSUM
ELFEXT
Power Sum ELFEXT is calculated from the summation of the individual ELFEXT effects on each pair by the other three pairs.
In the POWER SUM ELFEXT screen, the worst case margin for all near and far pairs is shown at the top of the screen.
89
CHAPTER 5: Measurements
The pairs are displayed in the column on the left side, followed by the PSEFLEXT in decibels (dB), the margin/limit, and the frequency. The result for each pair is tagged with a PASS or FAIL indicator.
Press or to toggle between the PS ELFEXT results for OMNIScanner and OMNIRemote.
Press (Graph) to display the PS ELFEXT graph for the currently highlighted pair. PS ELFEXT is displayed for the frequency indicated by the cursor position.
Press (Pairs) to view PS ELFEXT graphs for other pairs.
Press ESC to exit the POWER SUM ELFEXT screen.
5.3.11 P
OWERSUM
ACR
Power Sum ACR is calculated from the Attenuation of the disturbing pair subtracted from the Power Sum NEXT for the pair. The worst case Power Sum ACR is reported for each pair.
90
OMNISCANNER
The POWER SUM ACR screen displays the pairs in the column on the left side, followed by the PS ACR in decibels (dB), the margin/limit, and the frequency. The result for each pair is tagged with a PASS or FAIL indicator.
Press or to toggle between Power Sum ACR results for OMNIScanner and OMNIRemote.
Press (Graph) to view the graph for the currently highlighted pair combination. Attenuation and PSNEXT are displayed for the frequency indicated by the cursor position.
Press (Pairs) to view Attenuation and PS NEXT graphs for other pairs.
Press ESC to exit the POWER SUM ACR screen.
5.3.12 O
FFICELOCATOR
The Office Locator function allows you to find the termination of an office cable drop from a wiring closet. It eliminates the need to trace cables through walls and ceilings.
91
CHAPTER 5: Measurements
1. Connect the cable to be tested to OMNIScanner.
2. Insert the RJ-45 Office Locator adapters into wall outlets in the offices you wish to identify. You can connect any or all of the twelve Office Locators before beginning your test.
3. At the patch panel, run the Office Locator function to identify which office is connected to a given port.
OMNIScanner will identify office locations and display the ID of the Office Locator adapter found at the far end of the attached cable.
If you are testing a cable that does not have an Office Locator attached at the other end, OMNIScanner will display a question mark.
92
OMNISCANNER
Press ESC to exit the OFFICE LOCATOR screen.
5.4 Fiber Optic Measurements with OMNIFiber
5.4.1 L
ENGTH/DELAY
The Length/Delay function measures a fiber’s length and propagation delay. Length measurements are derived from the propagation delay results using the GRI (Graded Refractive Index) at the appropriate wavelength.
Propagation Delay is the measure of the time required for a signal to advance from one end of the link to the other. When measuring propagation delay, OMNIScanner measures the round trip delay through both fibers, and then divides the result by two, assuming that the cables are equal in length.
Propagation Delay cannot be measured on a single-mode fiber; this measurement is available for multimode fiber only.
OMNIScanner measures the full length of the fibers under test.
93
CHAPTER 5: Measurements
The LENGTH/DELAY screen displays the GRI @ 1300 nm (nanometers).
The length limit and the actual derived length are displayed in meters or feet.
The delay is displayed in nanoseconds (nS).
Press ESC to exit the LENGTH/DELAY screen.
5.4.2 L
OSS
The Loss function measures the signal loss in a fiberoptic cable. The adapters inject a signal into the fiber cable and OMNIScanner and OMNIRemote measure the received signals at 850 and 1300 nm for multimode fiber cables and 1310 for single-mode fiber cables.
The LOSS screen displays the measured power in dBm (decibels, one milliwatt reference) for each wavelength.
The measured loss and limit values in decibel (dB) are displayed for both wavelengths.
94
OMNISCANNER
For dual fiber testing, the arrows at the top of the screen point out the direction of the light and indicate which fiber the result is for.
NOTE
When measuring single-mode fiber cables, OMNIScanner displays results for the wavelength; the upper (850 nm) section will be blank.
Press ESC to edit the LOSS screen.
5.4.3 P
OWERMETER
The Power Meter function reports the overall received power. Use it to quickly verify and correct cabling and equipment problems.
The Power Meter screen shows the currently selected wavelength and light source mode.
The overall power is displayed.
OMNIFiber is designed to provide a consistent, calibrated light source for effectively measuring signal loss in fiberoptic cables. The transmitted light can be operated continuously or modulated.
When testing loss, use the continuous light source. When identifying fiber cables with a tone locator, use the 2 kHz modulated mode.
Press (Continuous) or (Modulated) to change the light source.
Press or to change the wavelength of the OMNIFiber main transmitted and received wavelengths. Either main or remote can be used as the light source with the main power meter.
95
CHAPTER 5: Measurements
When OMNIFiber is attached to OMNIRemote, the remote unit can be used as a light source. With the OMNIFiber adapter attached, press the lambda key to activate the light source and provide continuous or modulated light at 850 or 1300/1310 nm. The green LED of the appropriate wavelength will blink rapidly when modulated light is transmitted. If the transmitted light is continuous, the green LED will not blink.
Press to find the reference value, place it in the appropriate field, and start calculating the loss.
Press ESC to exit the Power Meter screen.
5.5 Coaxial Measurements with OMNIScanner
C
OAXTESTS
In coaxial cable, an insulated wire runs through the middle of the cable and a second wire, called the shield, surrounds the insulation of the inner wire like a sheath. The outer insulation jacket wraps around the second wire.
The Coax Tests screen displays the measured Length, Impedance, and Termination.
The Length function measures the cable length. OMNIScanner will display the overall length in meters or feet. If NEF (No End Found) is displayed instead of a value, the cable is either too long or terminated in its characteristic impedance.
Impedance defines the relationship of voltage and current in an electrical circuit.
96
OMNISCANNER
Each Cable or transmission line is designed to have specific impedance, which allows proper matching to the currents of the transmitter and receiver. OMNIScanner measures the characteristic impedance of the coax cable in ohms.
The Termination function displays whether a cable is open or shorted
.
When the field Length displays NEF, the Termination will display .
97
6. Diagnostics
6.1 OMNIScanner’s Diagnostic Functions
OMNIScanner has several advanced diagnostic test functions to help you with the identification of network cabling faults. Diagnostics allows you to locate opens, breaks, and short circuits as well as NEXT and Return Loss cabling faults quickly and easily.
Press the (Diagnostics) function key to display the Diagnostics screen.
6.2 Wiremap
Wiremap identifies the location of opens, breaks, and pairs that are shorted. This allows quick identification of the fault location.
Select Wiremap to display the WIREMAP screen. The screen displays the wire connections between OMNIScanner and OMNIRemote. The pairs that are being tested are defined by the selected test standard. Shield continuity is tested if a shielded cable type is selected. (See Chapter 4, Setup for further information about shielded cables.)
For each of the 8 conductors and the Shield, Wiremap can indicate proper pin termination on each end, continuity to OMNIRemote, shorts between any two or more conductors, crossed pairs, reversed pairs, split pairs, and any other miswiring. For each pair, Wiremap displays the length. When a pair is faulty the length indicates the distance to the fault.
W
IREMAPDISPLAYS
The OMNIScanner icon on the left side indicates the wiring at the OMNIScanner end of the cable link.
CHAPTER 6: Diagnostics
98
OMNISCANNER
The OMNIRemote icon on the right side indicates the wiring at the OMNIRemote end of the cable link.
Correct connectivity:
The measured lengths of pairs 1-2, 3-6, 4-5, and 7-8 are shown on the left side of the Wiremap.
If the selected cable is shielded or screened, shield continuity will be measured, and if the shield fails the test, Wiremap will display the error.
Wire is open:
Wiremap checks the continuity of each wire.
In this example, there is no line between pin 5 at each end, which indicates a break in connection. Use the number on the left side of the screen to find the exact location of the open. For example if 36 ft is displayed, the open can be found 36 feet away from OMNIScanner.
Loading...