B&K Precision XLN60026 User Manual

Model: XLN3640, XLN6024, XLN8018, XLN10014
High Power Programmable
Click for models XLN15010, XLN30052 & XLN60026
USER MANUAL
DC Power Supply
The following safety precautions apply to both operating and maintenance personnel and must be observed during all phases of operation, service, and repair of this instrument. Before applying power, follow the installation instructio ns and become familiar with the operating instructions for this instrument.
Failure to comply with these prec a utio ns or with specific warnings elsewher e in this manual violates safety standards of design, manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply with these requirements . This is a Safety Class I instrument.
GROUND THE INSTRUMENT
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. This instrument is grounded through the ground conductor of the supplied , three-conductor ac power cable. The power cable must be plugged into an approved three-conductor electrical outlet. Do not alter the ground connection. Without the protective ground connection, all accessible conductive parts (including control knobs) can render an electric shock. The power jack and mating plug of the power cable meet IEC safety standards.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes. Operatio n of any electri cal instrument in such an environme nt constitutes a definite safety hazard.
KEEP AWAY FROM LIVE CIRCUITS
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must be made by qualified maintenance personnel. Disconnect the power cord before removing the inst rument covers and replacing components. Under certain conditions, even with the power cable removed, dangerous vol t ages may exist. To avoid injuries, always disconnect power and discharge circuits before touching them.
DO NOT SERVICE OR ADJUST ALONE
Do not attempt any internal service or adjust ment unless another perso n, capable of rendering first aid and resuscitation, is present.
DO NOT SUBSTITUTE PARTS OR MODIFY THE INSTRUMENT
tive ground connection, all accessible conductive parts (including control knobs) can render an electric shock. The power jack and mating plug of the
before
Before connecting the line cord to the AC mains, check the rear panel AC line voltage indicator. Applying a line voltage other than the
fuses. For continued fire
protection, replace fuses only with those of the specified voltage and
This product uses components which can be damaged by
static discharge (ESD). To avoid damage, be sure to follow
ocedures for handling, storing and transporting parts and
Do not install substitute parts o r perform any unauthorized modificat ions to this instrument. Return the instrument to B&K Pre c isio n for service and repair to ensure that safety features are maintained.
WARNINGS AND CAUTIONS
WARNING and CAUTION statements, such as the following examples,
denote a hazard and appear throughout this manual. Follow all instructions contained in these statements.
A WARNING statement calls attention to an operating procedure, practice, or condition, which, if not follo we d correctly, could result in injury or de a th to personnel.
A CAUTION statement calls attention to an operating procedure, practice, or condition, which, if not follo we d correctly, could result in damage to or destruction of part or all of the product.
WARNING:
WARNING:
CAUTION:
CAUTION:
Do not alter the ground connection. Without the protec
power cable meet IEC safety standards. To avoid electrical shock hazard, disconnect power cord
removing covers. Refer servicing to qualified personnel.
indicated voltage can destroy the AC line current ratings.
electro­proper pr subassemblies which contain ESD-sensitive compo nents.
Store/Move/Maintain
Storage
When this device is not in use, properly package it and store it in an environment suitable for storage (if present in a good preserving environment, the packaging process can be waived).
Freight
While moving this product, move it by using the original packaging to pack this product in advance. If the packaging material is lost, use an equivalent buffer material to replace it in packaging; and with external marks indicating “fragile & water-prevention”.
Maintenance
Please return the power supply to factory for any repair, service, or maintenance.
Disposal
When the device is in an unusable condition and can’t be repaired, please discard it according to your company’s disposal procedures or local legal procedures. Don’t discard arbitrarily to avoid polluting the environment.
1
Table of Contents
1. Preface ...................................................................................................... 1
1.1 Products Outline ................................................................................ 1
1.2 Features ............................................................................................. 2
1.3 Specifications .................................................................................... 4
2. Cautions Before Using ............................................................................ 9
2.1 Check and Confirm Accessories before Using .................................. 9
2.2 Operation Instructions ....................................................................... 9
2.3 Ambient Environment ....................................................................... 9
2.4 Storage ............................................................................................. 10
2.5 Power-line voltage ........................................................................... 10
2.6 Fuses ................................................................................................ 10
2.7 Warm-up Time ................................................................................. 11
2.8 Power-off procedure ........................................................................ 11
2.9 Cautions in Operation ...................................................................... 11
3. Front Panel Operation .......................................................................... 12
3.1 XLN3640/XLN6024/XLN8018/XLN10014 Panel ......................... 12
3.1.1 Front Panel ................................................................................. 12
3.1.2 Rear Panel .................................................................................. 30
4. Operation Instructions ......................................................................... 34
4.1 Voltage Setting................................................................................. 34
4.2 Current Setting ................................................................................ 34
4.3 Over-voltage Protection OVP .......................................................... 34
4.4 Over-current Protection OCP .......................................................... 34
4.5 Voltage Output ................................................................................. 35
2
4.6 Control Voltage Output with Rotary knob ....................................... 35
4.7 Timer Function ................................................................................ 35
4.8 Series (cascade) / Parallel Mode Setting ......................................... 36
4.8.1 Parallel Connection Setting ........................................................ 36
4.8.2 Series Mode Setting ................................................................... 38
4.8.3 Error Message of Series/Parallel Connection ............................ 41
4.9 External Tuning Setting ................................................................... 42
4.10 Timer of Current Flow ..................................................................... 43
4.11 Programmable Capability (SCPI Command Only) ......................... 44
4.12 Multi-unit Connection mode (RS485) ............................................. 50
5. Protection and Error Messages............................................................ 54
5.1 Over-voltage Protection (OVP) ....................................................... 54
5.2 Over-current Protection (OCP) ........................................................ 54
5.3 Overpower Protection (OPP) ........................................................... 54
5.4 Constant Voltage Protection (CV TO CC) ....................................... 55
5.5 Constant Current Protection (CC TO CV)....................................... 55
5.6 Over-temperature Protection (OTP) ................................................ 55
5.7 Low Voltage Protection (ACD) ....................................................... 56
5.8 Error Input Message ........................................................................ 56
6. Remote Interface communication protocol ........................................ 57
6.1 Prefaces ........................................................................................... 57
6.2 Parameters Definition ...................................................................... 57
6.3 The Error/Event List ........................................................................ 58
6.4 Remote Communication Protocol ................................................... 59
6.5 SCPI Conformity Information ......................................................... 62
6.5.1 Common SCPI commands ......................................................... 62
6.5.2 SCPI Command subsystem ........................................................ 63
3
6.6 State Bit Definition .......................................................................... 74
6.7 LAN Communication (-GL versions) .............................................. 75
Using Web Server........................................................................................... 75
Main Page (Home) .................................................................................. 75
Using Telnet ............................................................................................ 78
Using Sockets .......................................................................................... 78
7. Assemble Accessories ............................................................................ 79
7.1 Assemble Rack Mount Brackets ..................................................... 79
7.2 Assembly of Output Protective Cover
(XLN6024/XLN8018/XLN10014) .................................................. 80
7.3 Assembly of Remote Sense Protective Cover
(XLN6024/XLN8018/XLN10014) .................................................. 81
8. Accessories ............................................................................................. 82
9. Service Information .............................................................................. 84
10. Limited Three-year Warranty ............................................................. 85
4
1. Preface
1.1 Products Outline
B&K Precision models XLN3640/XLN6024/XLN8018/XLN10014 are programmable DC power supplies with single outputs that offer the maximum power output up to 1440 watts (0 -- 36 V/40 A or 0 -- 60 V/24 A or 0 -- 80 V/18 A or 0 -- 100 V/14.4 A). With a 16-bit D/A, A/D converter embedded, the power supplies come with the resolution of 1mV in voltage setting and 1mA in current setting. By connecting up to 4 power supplies in parallel or series, a maximum power output up to 5760 watts can be generated. With four XLN10014 connected in series, the maximum output voltage can reach 400 V. With four XLN3640 connected in parallel, the output current can reach up to 160 A.
The XLN series provides a rotary control knob and numerical and function keys to make the instrument convenient and easy to use. Additionally, the power supplies provide a memory space for storage of 10 instrument settings that can be recalled directly. This feature offers an easy way to restore the application settings. I n addition, users can program to control when to cut off the output. This feature provides extra safety for burn-in and electroplating applications. The supplies also provide over voltage protection ( OVP), over current protection (OCP), and over power protection (OPP) features used to keep the output voltage and current within safety level and preventing damage to the UUT (Unit Under Test) due to excessive current. The key lock feature is added to avoid accidental setting changes to the XLN series. When the input power and the load change, the power suppl ies maintain a stable output due to load and line regulation of less than 0.05%; the transient time less than 1 ms. In remote mode, the supplies can output a new voltage/current setting 50 ms after receiving a command, which can increase the throughput on production lines.
1
1.2 Features
1) Output Voltage & Current
Voltage output range: 0 -- 36V (XLN3640) / 0 -- 60V (XLN6024) 0 -- 80V (XLN8018) / 0 -- 100V (XLN10014) Current output range: 0 -- 40A (XLN3640) / 0 -- 24A (XLN6024) 0 -- 18A (XLN8018) / 0 -- 14.4A (XLN10014) Power output range: 0 -- 1440W
2) Rotary knob, numerical keys and functions keys
The rotary knob can be used to rapidly change the output voltage s etting and simulate the surge of the voltage output. It offers a good solution for testing triggering circuits. Numerical keys allow for direct entry of parameters. Using function keys to switch modes makes the overall operation more convenient.
3) Precise voltage and current measurement
Besides the precise output, the XLN series also offers the capability to measure voltage & current accurately (read back), saving users the extra expense and space for extra measuring instruments.
4) Internal memory and timer function
The XLN series provides a memor y space for storage and retrieval of 10 instrument settings. The instruments provide one (1) timer with the resolution of 1 second. The timers are used to time the outputs. When the timer counts down to zero the power supply will automatically turn the output off. This feature is useful when the supply is providing power to the test object in a burn-in room where operators can precisely set the time when the equipment is to shut off.
5) OVP (over voltage protection), OCP (over current protection) and OPP (over power protection) and key lock functions
The over voltage p r ot e ction (OVP), over current protection (OCP) and over power protection (OPP) features limit the maximum output current and voltage to avoid damages to the unit under test (UUT). The key lock feature disables all keys except the CLR key. It prevents damaging the
2
UUT by accidentally entering the wrong settings.
6) Series & parallel connection mode The series-parallel connection mode of two or more units (maximum to 4
units) significantly increases the combined output power to a maximum of 5760 W. In parallel connection mode of four XLN3640 the maximum output is 36 V/160 A; and in series connection mode of four XLN10014, the maximum output is 400 V/14.4 A.
7) Multi-unit connection mode The RS 485 interface can be used to connect multiple power supplies in
series, up to maximum of 30 units. They can be controlled via USB interface with a computer.
3
Output Protection
OVP Adjustment Range
2--38 V
3--64 V
OVP Accuracy
200 mV
300 mV
Line Regulation
Current
8 mA
7 mA
Ripple/Noise (20Hz-20MHz)
1.3 Specifications
XLN3640/XLN6024 Specifications
Output Rating
Output Voltage 0--36 V 0--60 V Output Current 0--40 A 0--24 A Output Power 1440 W 1440 W
Voltage 4 mV 6 mV Current 4 mA 4 mA Load Regulation Voltage 8 mV 8 mV
XLN3640 XLN6024
Normal Mode Voltage ( 0.5 % of max. power)
Normal Mode Current 90 mA 70 mA
Programming Resolution
Programming 1 mV/1 mA 1.5 mV/1 mA Readback 1 mV/1 mA 1.5 mV/1 mA
Programming Accuracy ±(% output+offset)
Voltage 0.05 %+10 mV 0.05 %+15 mV Current 0.05 %+10 mA 0.05 %+18 mA
Readback Accuracy ±(% output+offset)
Voltage 0.05 %+10 mV 0.05 %+15 mV Current 0.05 %+10 mA 0.05 %+18 mA
5 mVrms/ 60 mVpp 6 mVrms/ 70 mVpp
4
Average Command Response Time
0.99
(Full load)
0.99
(Full load)
100--240 VAC
(Full load)
-15% -- +10%
when voltage under 95 VAC)
Power Cord, Terminal Blocks for Rapid Plug
Connector, Rackmount Kit, Manual
General
<50 ms <50 ms
Power Factor Correction Remote Sense Compensation 2V 2V
Rising Time at Full Load 15 ms 20 ms Rising Time at No Load 15 ms 20 ms Falling Time at Full Load 15 ms 20 ms Falling Time at No Load 1000 ms 1000 ms Standard Interface USB Transient Response Time 1 ms Efficiency 80 %
AC Line Rated Input Voltage
Tolerance/Variation in Voltage
Rated Frequency 47 Hz--63 Hz Maximum Rated Input Power 1700 VA Temperature Ratings(O) Operation (0 °C -- 40 °C) Temperature Ratings(S) Storage (-10 °C -- 70 °C) Dimensions(W*H*D) 16.5 x 1.7 x 17 inch(420 x 43.6 x 432 mm) Weight 19.8 lbs. (9 kg)
Standard Accessories Standard Interface USB
Optional Interface LAN & GPIB
(10% power de-rating mode
5
Normal Mode Voltage ( 0.5 % of max. power)
7 mVrms/ 80
mVpp
XLN8018/XLN10014 Specifications
Output Rate
Output Voltage Output Current
XLN8018 XLN10014
080 V 0100 V 018 A 014.4 A
Output Power 1440 W 1440 W
Output Protection
OVP Adjustment Range
485 V 5105 V
OVP Accuracy 400 mV 500 mV
Line Regulation
Voltage 8 mV 10 mV Current 4 mA 4 mA Load Regulation Voltage 10 mV 12 mV Current 6.5 mA 6 mA
Ripple/Noise (20Hz-20MHz)
8 mVrms/ 80 mVpp
Normal Mode Current 50 mA 40 mA
Programming Resolution
Programming 2 mV/1 mA 2.5 mV/1 mA Readback 2 mV/1 mA 2.5 mV/1 mA
Programming Accuracy ±(% output+offset)
Voltage 0.05 %+20 mV 0.05 %+25 mV Current 0.05 %+7 mA 0.05 %+6 mA
Readback Accuracy ±(% output+offset)
Voltage 0.05 %+20 mV 0.05 %+25 mV Current 0.05 %+7 mA 0.05 %+6 mA
General
6
Average Command Response Time
Power Factor Correction
0.99
(Full load)
0.99
(Full load)
Remote Sense
100~240 VAC
( Full load )
-15 %~+10 %
when voltage under 95 VAC )
Power Cord, Terminal Blocks for Rapid Plug
Connector, Rackmount Kit, Manual
<50 ms <50 ms
Compensation
2 V 2 V
Rising Time at Full Load 25 ms 30 ms Rising Time at No Load 25 ms 30 ms Falling Time at Full Load 25 ms 30 ms Falling Time at No Load 1000 ms 1000 ms Transient Response Time 1 ms Efficiency 80 %
AC Line Rated Input Voltage
Tolerance/Variation in Voltage
Rated Frequency Maximum Rated Input
Power
( 10 % power de-rating mode
47 Hz63 Hz
1700 VA
Temperature Ratings(O) Operation (0 °C -- 40 °C) Temperature Ratings(S) Storage (-10 °C -- 70 °C) Dimensions(W*H*D) 16.5 x 1.7 x 17 inch (420 x 43.6 x 432 mm) Weight 19.8 lbs. (9 kg)
Standard Accessories Standard Interface USB
Option Interface LAN & GPIB Specifications and information is subject to change without notice
7
Features of models XLN3640/XLN6024/XLN8018/XLN10014:
Graphical, easy to read LCD display Compact, high efficiency and power density 40 A output connector for quick connectivity Convenient numerical & function keys S tore and recall 10 instrument settings Timer (1 sec -- 100 hours) Programmable (SCPI command only) List mode supports up to 10 set s of program and maximum 150 steps in
total
Auxiliary 5 V/1 A output Built-in precise voltage and current measurement OVP, OCP, OPP and key-lock function Series & parallel connection setup (up to 4) Multi-unit connection mode via RS485 interface allows connection of up
to 30 power supplies.
Average measuring time per measurement is 50 ms Standard USB interface Optional interfaces: GPIB, LAN (order models XLN3640-GL,
XLN6024-GL, XLN8018-GL, XLN10014-GL)
8
2. Cautions Before Using
2.1 Check and Confirm Accessories before Using
After receiving this product, please verify the items received in accordance with the ones liste d belo w:
1. The appearance of the products is without scratc h or other dam ag es.
2. Standard parts as shown in parts list of section 8.
2.2 Operation Instructions
In order to avoid damaging the instrument due to improper operation, be sure to read this user manual. To maintain the specified accuracy, factory calibration should be performed annually.
2.3 Ambient Environment
1. Do not locate or operate this product in an environment with dust,
vibration, or corrosive gas and do not expose this product directly to the sunlight. Operate it in an environment with temperature 0--40 & relative humidity 20%--80%. Pause the operation when ambient temperature is over 40 temperatures drops to the acceptable temperature range. Operating temperature over the above range would damage the instrument.
2. This product is equipped with one blow-out type cooling fan on the
back board and three in-take cooling fans on inner side of front board. Provide room for good ventilation near the cooling fans and keep the boards with a space above 10cm away from wall. To maintain good accuracy, do not block the ventilation holes in the front and the rear parts of the unit.
3. Although the product is designed with filters to minimize noise from
AC power source, it is recommended that it be operated in a low power noise environment with proper earth ground. If the power noise is unavoidable, please install a power filter.
o
C; undo the operation only after the ambient
o
C
9
2.4 Storage
The storage temperature range of this product is within -10ºC - 70ºC and R.H. should be within 80% without moisture condensing. If not operating this product for a long time interval, pack it with original packaging or similar one and put it in a dry place without expos ure to direct sunlight.
2.5 Power-line voltage
Rated AC power source connected to this product is within 100 V-240 V (refer to the Product Specification for details). Before connecting to external power source, be sure that t he power switch is in OFF state and verify the suitability of power cable (including the extension line). It should be compatible with the rated voltage/current and should be firmly connected.
Warning :
The power cable attached with this product is certified for safety. To change a cable or add an extension cable, be sure that it can meet the required power ratings of this product. Any misuse with an additional cable
would void the warranty of this product.
2.6 Fuses
This product is a switching mode power supply. The fuse installed inside is a multi-barrier protection hardware design. It should not break under normal operation. In case the fuse does melt, it indicates another malfunction that causes the fuse to break. In this case, it is suggested to send this product back to service.
Warning :
Any disassembling of the casing or changing the fuse not performed by an authorized service technician will void
the warranty of the instrument.
10
2.7 Warm-up Time
The XLN series is fully operable upon switching the power on. However, to reach the specified equipment accuracy, please allow the supply to warm up for at least 30 minutes.
2.8 Power-off procedure
When the supply is not in use, be sure to turn the power switch on the panel to the OFF p os ition to turn off the power. After the pow e r s wi t c h i s t urned to the OFF position, the inner fans will still run for approximately 10-15 seconds to carry on the inside el ectric capacitor discharge p roces s per safe ty code requirement. Once the discharge process is complete, this product will carry out the autom atic shut -down process
2.9 Cautions in Operation
A. While connected in series, each power supplies should be in power-on
state and output should be "ON". In case there is any one supply that is in power-off st ate or output is " OFF", the associated output current will flow over the output bypass diode of the power-off unit and burn it out.
B. While in parallel connection mode, the output voltage of each power
supplies should be set to equal values. If the set ting value of each unit is not the same, the hi ghe r o ut pu t voltage will feed back to the smaller unit and destroy its inner parts.
C.
When the AC input vol tage is lower than the full-load voltage which is
100 VAC, the supplies will activate an inner over temperature protector and cut off the output in response to the condition. To ensure that the entire test process can be complete smoothly, confirm that the input AC voltage is within the specified range.
11
3. Front Panel Operation
1 2 3 4 5 6 7
8
9
10
11
12 13 14
3.1 XLN3640/XLN6024/XLN8018/XLN10014 Panel
15
3.1.1 Front Panel
(1) Power switch:
Please consult the “Cautions before use” section before turni ng on power switch.
(2) Display:
192x32 Graphic LCD Module
(3) Current setting
Press
to set up the current limit.
(4) Voltage setting
Press
to set up the output voltage.
:
:
(5) Dot/Local :
This button is applied as a decimal point. Or push this button after entering REMOTE online state to revert back to LOCAL mode (unit-operation mode). Or press this button to release after entering LOCK mode.
12
(6) ESC/CLR :
In “Menu Setting” status, use this “Down” key to mo ve cursor
Under Memory Setting status, use this key to store setting to the selected memory set.
Setting” status, use this “Down” key to move cursor
Under Memory Setting status, use this key to recall setting from the selected memory set.
=
36
. FF
5
1
.
=
.
.
Press this button to clean up numerical setting or jump to the previous screen.
(7) Numerical keys
They are used to directly input the voltage or current value or choose t he setting option in Menu screen.
(8) Down/Right/Store
This key is a multi-function key for the following three functions:
Down:
to the next item. Right: Under “Output” status, use this key to move cursor right. Store:
(9) Up/Left/Recall :
This key is a multi-function key for the following three functions:
Up:
Left: Under “Output” status, use this key to move cursor left. Recall:
In “Menu
to the up item.
- :
:
(10) Display :
In “Menu Setting”, press the display to show voltage and current or power and load r esistance as
shown below:
to return to main screen or toggle
13
.
1= .
5
FF
.
3=6
.
(11) Output :
=
FF
.
.
Control the On/Off of the output power.
(12) The rotary knob:
Use this knob to adjust voltage or current (press cursor display first). This is adjustable when output is ON.
first to let
(13) Enter
This ke y is the confirmation key o f current or voltage setti ng value; or press mode) or current (at CC mode).
(14) Mem
Press this key to enter access the storage memory. Users can then use the numerical key or knob to select the target memory set to save or recall the configuration by pressing the STORE or RECALL key. Ten sets are available in selection.
(15) Menu :
Use this key to enter system parameter settings. There are eight (8) major items under operation. Users may press
through the menu list or the numerical keys to enter the corresponding item number in the menu list.
:
under output status to dynamically adjust voltage (at CV
:
,
to scroll
14
L=
= 1
T
=L
FF
1 . SYSTEM SETTING 2 . OUTPU T S ETTING 3 . PROTECTION
4
.
F
5
.
6
.
L T T F
LL L
T
7 . TIMER CONTROL 8 . CALIBRATION 9 . CHAIN SETTING
1. SYSTEM SETTING:
Pressing following “SYSTEM SETTING” menu.
REMOTE CONTROL:
key in the first page of Menu Setting will enter the
Choose the remote interface (USB/GPIB/ETHERNET)
*GPIB and ETHERNET available only with on models with “-GL” suffix
*USB control requires installing USB drivers first. Download USB driver from
www.bkprecision.com
*USB interface is a virtual COM port. T he settings are:
Baudrate : 57600 bps
15
Loading...