B&K Precision RFM3000 User Manual

Safety Summary
The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
• Read and understand the safety and operational information in this manual.
• Apply all the listed safety precautions.
• Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating the instrument at an incorrect line voltage will void the warranty.
• Make all connections to the instrument before applying power.
• Do not operate the instrument in ways not specied by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design, manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply with these requirements.
2
Category rating
The IEC 61010 standard denes safety category ratings that specify the amount of electrical energy available and the voltage impulses that may occur on electrical conductors associated with these category ratings. The category rating is a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of the circuit to be tested, which denes the voltage impulses expected and required insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected to the
mains supply. The voltages in the environment are typically derived from a limited-energy trans­former or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the mains
supply at a standard wall outlet or similar sources. Example measurement environments are portable tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be connected to the mains
installation of a building. Examples are measurements inside a building’s circuit breaker panel or the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to the primary
power entering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what is specied in this manual for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the instrument’s category rating to maintain the instrument’s category rating. Failure to do so will lower the category rating of the measuring system.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be 115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This instrument is grounded through the ground conductor of the supplied, three-conductor AC line power cable. The power cable must be plugged into an approved three-conductor electrical outlet. The power jack and mating plug of the power cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive parts (including control knobs) may provide an electric shock. Failure to use a properly-grounded approved outlet and the recommended three-conductor AC line power cable may result in injury or death.
3
Unless otherwise stated, a ground connection on the instrument’s front or rear panel is for a reference of potential only and is not to be used as a safety ground. Do not operate in an explosive or ammable atmosphere.
Do not operate the instrument in the presence of ammable gases or vapors, fumes, or nely-divided particulates.
The instrument is designed to be used in oce-type indoor environments. Do not operate the instrument
• In the presence of noxious, corrosive, or ammable fumes, gases, vapors, chemicals, or nely-divided particulates.
• In relative humidity conditions outside the instrument’s specications.
• In environments where there is a danger of any liquid being spilled on the instrument or where any liquid can condense on the instrument.
• In air temperatures exceeding the specied operating temperatures.
• In atmospheric pressures outside the specied altitude limits or where the surrounding gas is not air.
• In environments with restricted cooling air ow, even if the air temperatures are within specications.
• In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature range is 0∘C to 40∘C and 20% to 80% relative humidity, with no condensation allowed. Measurements made by this instrument may be outside specications if the instrument is used in non-oce-type environments. Such environments may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise, strong electric elds, or strong magnetic elds.
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the instrument, remove the instrument’s power cord, remove the instrument from service, label it as not to be operated, and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual. Not for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a life-support device or system.
4
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must be made by qualied service-trained maintenance personnel who are aware of the hazards involved when the instrument’s covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may exist when the covers are removed. To avoid injuries, always disconnect the power cord from the instrument, disconnect all other connections (for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there are no hazardous voltages present on any conductors by measurements with a properly-operating voltage-sensing device before touching any internal parts. Verify the voltage-sensing device is working properly before and after making the measurements by testing with known-operating voltage sources and test for both DC and AC voltages. Do not attempt any service or adjustment unless another person capable of rendering rst aid and resuscitation is present. Do not insert any object into an instrument’s ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit exists.
Fuse replacement must be done by qualied service-trained maintenance personnel who are aware of the instrument’s fuse requirements and safe replacement procedures. Disconnect the instrument from the power line before replacing fuses. Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings specied in this manual or on the back of the instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a re. Failure to use the specied fuses will void the warranty.
Servicing
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to B&K Precision for service and repair to ensure that safety and performance features are maintained.
For continued safe use of the instrument
• Do not place heavy objects on the instrument.
• Do not obstruct cooling air ow to the instrument.
• Do not place a hot soldering iron on the instrument.
• Do not pull the instrument with the power cord, connected probe, or connected test lead.
• Do not move the instrument when a probe is connected to a circuit being tested.
Compliance Statements
Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with separate collection systems)
This product is subject to Directive 2002/96/EC of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste. Please utilize your local WEEE collection facilities in the disposition of this product and otherwise observe all applicable requirements.
5
Safety Symbols
Symbol Description
indicates a hazardous situation which, if not avoided, will result in death or serious injury.
indicates a hazardous situation which, if not avoided, could result in death or serious injury
indicates a hazardous situation which, if not avoided, will result in minor or moderate injury
Refer to the text near the symbol.
Electric Shock hazard
Alternating current (AC)
Chassis ground
Earth ground
This is the In position of the power switch when instrument is ON.
This is the Out position of the power switch when instrument is OFF.
is used to address practices not related to physical injury.
6
Contents
1 General Information 9
1.1 Organization 9
1.2 Description 10
1.3 Features 10
1.4 Front Panel 11
1.5 Rear Panel 13
1.6 Touch Screen Display 14
2 Installation 17
2.1 Contents 17
2.2 Input Power & Fuse Requirements 17
2.3 Connections 18
2.4 Preliminary Check 18
3 Getting Started 20
3.1 Initialization 20
3.2 Taking Measurements 22
3.2.1 Continuous Mode 22
3.2.2 Pulse Mode 23
3.2.3 Statistical Mode 24
4 Operation 26
4.1 Control Menus 26
4.2 Parameter Date Entry and Selection 26
4.2.1 Numerical Data Entry & Drop Down Menus 26
4.3 Menu Reference 27
4.3.1 Main Menu 27
4.3.2 Measure > 28
4.3.3 Display > 29
4.3.4 Stat. Mode > 29
4.3.5 Channel > 30
4.3.6 Channel Settings 31
4.3.7 Time > 34
4.3.8 Trigger > 34
4.3.9 Pulse Def. > 36
4.3.10 CH# Pulse Def 36
4.3.11 Favorites > 38
4.3.12 System > 38
5 Application Notes 42
5.1 Introduction to Pulse Measurements 42
5.1.1 Measurement Fundamentals 42
5.1.2 Diode Detection 44
5.1.3 Pulse Denitions 46
5.1.4 Standard IEEE Pulse 46
5.1.5 Automatic Measurements 47
5.1.6 Automatic Measurement Criteria 47
5.1.7 Automatic Measurement Terms 48
5.1.8 Automatic Measurement Sequence 49
5.1.9 Average Power Over an Interval 52
5.1.10 Statistical Mode Automatic Measurements 53
5.2 Measurement Accuracy 54
6 Maintenance 55
6.1 Safety 55
6.2 Cleaning 55
6.3 Inspection 55
6.4 Lithium Battery 56
6.5 Software Upgrade 56
7 Service Information 57
8 LIMITED THREE-YEAR WARRANTY 58
8
General Information
This instruction manual provides you with the information you need to install, operate, and maintain the RFM3000 RF Power Meter. Section 1 is an introduction to the manual and the instrument.
1.1 Organization
The manual is organized into ve sections and two Appendices, as follows:
Section 1
General Information presents summary descriptions of the instrument and its principal features, acces-
sories, and options.
Section 2
Section 3
Section 4
Section 5
Installation provides instructions for unpacking the instrument, setting it up for operation, connecting
power and signal cables, and initial power-up.
Getting Started describes the controls and indicators and the initialization of operating parameters. Several practice exercises are provided to familiarize yourself with essential setup and control procedures.
Operation describes the display menus and procedures for operating the instrument locally from the front panel.
Application Notes provides supplementary information about RFM3000 and sensor operation, ad­vanced features, pulse measurement information, and measurement accuracy.
General Information 10
1.2 Description
The RFM3000 provides design engineers and technicians the utility of traditional benchtop instruments, the exibility and performance of modern USB RF power sensors, and the simplicity of a multi-touch display built with advanced technology.
As a benchtop meter, the RFM3000 provides a standalone solution for capturing, displaying, and analyzing peak and average RF power in both the time and statistical domains through an intuitive, touch screen display.
The RFM3000 RF Power Meter utilizes up to four RFP Series Sensors with industry-leading performance and capabilities either independently or for synchronized multi-channel measurements of CW, modulated, and pulsed signals.
Providing the ultimate exibility, the RFM3000 sensors can be disconnected and independently used as standalone instruments.
1.3 Features
• Compatible with B&K Precision’s RFP3000 Series USB RF Power Sensors
• Capture/display/analyze peak and average power
• Independent or synchronous multi-channel measurements (up to 4 channels)
• Trigger synchronization
• Supports SCPI-1999.0
• Sensor verication test source
• Display 16 common power measurements
• Ethernet:10/100/1000 BaseT; HiSLIP
• HDMI output for mirror display
• Sensors can be used as standalone instruments
General Information 11
1.4 Front Panel
Refer to table 1.1 for a description of each of the illustrated items. The function and operation of all controls, indicators, and connectors are the same on the standard and optional models.
Figure 1.1 Front Panel
General Information 12
Item Description
Four sensor inputs are located on the front and rear panels of the instrument. These are stan-
1 USB Host
dard USB 2.0 Type A receptacles designed to accept only RFP3000 Series Sensors or standard USB keyboards, mice, and ash drives.
Four sensor trigger inputs are located on the front and rear panels of the instrument. These are standard SMB receptacles designed to accept only BK Precision power sensor trigger ca-
2 Sync Ports
bles.
Do not attempt to connect anything other than RFP3000 sensors and trigger cables!
The output of the built-in 50 MHz programmable test source is available from a Type-N
3 RF Output
connector located on the front, or optionally on the rear panel of the instrument. This test source is used to verify basic performance of sensors used with the RFM3000.
4 Display Screen
5
6
7
8
Color touch screen display for the measurement and trigger channels, screen menus, status messages, text reports, and help screens.
Favorites key. (This function is not fully implemented at this time). Enables the user to setup a customized menu to allow grouping frequently used menu items into one convenient menu.
Store image key saves a screen image of the meter to local storage. The images can be copied to an external USB storage device.
Used to assist navigating between items on the display and in the menus. Unless the user is in digit editing numeric entry mode.
Used for incrementing or decrementing numeric parameters, or scrolling through multi-line or multi-page displays.
Selects an on-screen item or menu and completes a numeric or picklist entry
Toggles the instrument between “on” (fully powered) and “standby” (o, except for certain low-power internal circuits) modes. Entering standby mode will perform a save of the current instrument state before shutdown. Pressing and holding the On/Standby key for several sec­onds will force standby mode if the instrument has become non-responsive. In this case, no context save is performed
Table 1.1 Front Panel
General Information 13
1.5 Rear Panel
Refer to table 1.2 for a description of each of the illustrated items. The function and operation of all controls, indicators, and connectors are the same on the standard and optional models.
Figure 1.2 Rear Panel
General Information 14
Item Description
Four sensor inputs are located on the front and rear panels of the instrument. These are stan-
1 USB Host
2 Sync Ports
3 RF Output
4 Trig In
dard USB 2.0 Type A receptacles designed to accept only RFP3000 Series Sensors or standard USB keyboards, mice, and ash drives.
Four sensor trigger inputs are located on the front and rear panels of the instrument. These are standard SMB receptacles designed to accept only BK Precision power sensor trigger ca­bles.
Do not attempt to connect anything other than RFP3000 sensors and trigger cables!
The output of the built-in 50 MHz programmable test source is available from a Type-N connector located on the front, or optionally on the rear panel of the instrument. This test source is used to verify basic performance of sensors used with the RFM3000.
BNC input for connecting an external trigger signal to the power meter. Voltage range is±5 volts, but the input impedance is 1 Megohm to allow use of common 10x oscilloscope probe for a±50 volt input range.
5 Multi-I/O
6 LAN Ethernet
7 HDMI
8 AC Line Input
9 HDMI Cooling air intake.
10 GPIB
BNC input/output for exible use. May serve as a status or alarm output, signal level moni­tor, or settable voltage source
LAN connector for remote control and rmware updates. Allows DHCP or xed (IP / Sub­net) setting mode. LAN parameters can be congured through the menu.
HDMI receptacle for connecting an external monitor to mirror front panel display. The image resolution will be 800 x 480 and will be stretched to t the external full display size.
A multi-function power input module is used to house the AC line input, main power switch, and safety fuse. The module accepts a standard AC line cord, included with the power meter. The power switch is used to shut o main instrument power. The safety fuse may also be ac­cessed once the line cord is removed. The instrument’s power supply accepts 90 to 264 VAC, so no line voltage selection switch is necessary.
24-pin GPIB (IEEE-488) connector for connecting the power meter to the remote control General Purpose Instrument Bus. GPIB parameters can be congured through the menu.
Replace fuse only with specied type and rating:
1.0A-T (time delay type), 250 VAC.
Table 1.2 Rear Panel
1.6 Touch Screen Display
The RFM3000 can be controlled through the touch screen display and by use of the front panel buttons. Table 1.3 describes the dierent areas of the display layout of the RFM3000. Figure 1.3 shows the Graph Display mode of the instrument using the Pulse Measure mode with a menu exposed. Figure 1.4 shows the same measure mode with the
General Information 15
menus hidden. The Text Display mode of the instrument provides a table view of measured parameters. Parameters depend on the Measure mode selected. See Section Menu Referencefor more information on the display format.
Figure 1.3 Display
Figure 1.4 Display Hidden Menu
General Information 16
Item Description
Indicates the measurement acquisition status of the unit. In Pulse mode, the sample rate and
1 Status Bar
number of sweeps per second are also shown. In Statistical mode, it indicates the gating set­ting in use, run time, and number of points.
Displays a table of measurements for each channel that is enabled on the meter. In Pulse and
2 Parameters
Continuous mode, measurements indicated are for power levels at each marker and the aver­age power between the markers. For Statistical mode, the measurements are Average power, Maximum power, and Peak-to-Average Ratio or Crest Factor.
This area indicates which channels are ON and their individual scale and vertical center.
3 Channel Status
The base RFM3000 model only permits two sensors to be active at any one time. With the RFM300-4CH option, four sensors can be active at any one time.
4 Main Display
5 Menu Bar
This area will show a plot when in Graph Display mode or a table of parameters when in Text Display mode for the measurement mode selected.
Select to show and to hide the on-screen menus.
6 Menu Parth Used to navigate the menu structure. Shows the menu that will be displayed when selected.
7
Current
Menu/Home
8 Horizontal Scale
9 Measure Mode
Displays the name of the current menu and provides a home shortcut to the top-level Main menu. When in the Main or top level menu, this eld is not available.
For Pulse and Continuous mode, indicates the time per division for the waveform display. In Statistical mode, the horizontal scale for the CCDF graph is in dBr (dB relative).
Indicates and allows selection of the current Measurement mode. Modes available are Contin­uous, Pulse, and Statistical.
10 Display Mode
Indicates and allows selection of the current Display mode in use. Modes available are Graph and Text.
The Graph Display mode for Continuous Measure mode will be a at trace. It is best to use Text Display mode for continuous signal measurements.
Table 1.3 Display
Installation
This section contains unpacking and repacking instructions, power requirements, connection descriptions, and preliminary checkout procedures.
2.1 Contents
Please inspect the instrument mechanically and electrically upon receiving it. Unpack all items from the shipping carton, and check for any obvious signs of physical damage that may have occurred during transportation. Report any damage to the shipping agent immediately. Save the original packing carton for possible future reshipment. Every power supply is shipped with the following contents:
• RFM3000 RF Power Meter
• Line Cord
• Information Card (describes where to download the latest manual, software, utilities)
2.2 Input Power & Fuse Requirements
The RFM3000 is equipped with a switching power supply that provides automatic operation from a line voltage input within: The supply has a universal AC input that accepts line voltage input within:
Voltage: 100 - 240 VAC (+/- 10 %)
Frequency: 43 to 63 Hz
Input Power: 70 VA MAX.
Before connecting the instrument to the power source, make certain that a 1.0-ampere time delay fuse (type T) is installed in the fuse holder on the rear panel.
Before removing the instrument cover for any reason, position the input module power switch to o (0 = OFF; 1 = ON) and disconnect the power cord.
Connect the power cord supplied with the instrument to the power receptacle on the rear panel. See gure ??
The included AC power cord is safety certied for this instrument operating in rated range. To change a cable or add an extension cable, be sure that it can meet the required power ratings for this instrument. Any misuse with wrong or unsafe cables will void the warranty.
Installation 18
2.3 Connections
Sensor(s)
Note:
Trigger
Note:
The Sync cable must be connected to the Sync port corresponding to the USB port for the sensor Channel in use.
Compatible sensors can be connected to any of the USB ports on the front or rear panel. The base RFM3000 model only permits two sensors to be active at any one time. With the RFM3000-4CH model, four sensors can be active at any one time. Sensors become active when plugged into a USB port or immediately if already plugged in when the RFM3000 powers up.
The base RFM3000 model only permits two sensors to be active at any one time.
With the RFM3000-4CH option, four sensors can be active at any one time.
Most triggered applications can use the RF signal applied to the sensors for triggering. For measure­ments requiring external triggering, connect the external trigger signal to the Trig In BNC connector on the rear panel and connect a Sync cable from the Sync connector on the meter to the Multi I/O port on the sensor.
Remote
If the instrument is to be operated remotely using the GPIB (IEEE-488) bus, connect the instrument to the bus using the rear panel GPIB connector and appropriate cable. For Ethernet control, connect to the rear panel LAN connector. In most cases, it will be necessary to congure the interface using the System > I/O > Cong menu.
2.4 Preliminary Check
The preliminary check veries that the instrument is operational and has the correct software installed. It should be performed before the instrument is placed into service.
To perform the preliminary check, proceed as follows:
1. Press the lower half (marked "O") of the power switch in the center of the power module on the rear panel.
2. Connect the AC (mains) power cord to a suitable AC power source; 90 to 264 VAC, 47 to 63 Hz.
– The power supply will automatically adjust to voltages within this range.
3. Press the upper half (marked "—") of the power switch in the center of the power module on the rear panel, it will enter standby mode.
4. Press the ON/STBY key on the front panel to turn the instrument on. The cooling fan and display backlight should turn on.
5. A bootup screen should appear that shows the boot status. After a self-check, the instrument will execute the application program. There will be some temporary dialogs indicating application initialization and channel updating. After several moments, a screen similar to gure 2.1 should be displayed.
Installation 19
Figure 2.1 Power-On Display
6. On the front panel, press the select key to bring up the on-screen menu. From the Main menu, use the touch screen or the navigation keys on the front panel to browse to System > Reports > Conguration and select Show. A display similar to gure 2.2 will appear.
Figure 2.2 Conguration Report Display
Getting Started
This chapter will introduce the user to the RFM3000 RF Power Meter. The chapter will identify display organization, list the conguration of the instrument after initializing, and provide practice exercises for front panel operation. For additional information, see Section 4 Operation.
3.1 Initialization
The steps below initialize the RFM3000 and prepare it for normal operations. Step 3 should only be performed when you wish to set the meter operations to a known state. This is typically done when you rst power on the instrument or at the start of a new test.
1. If the main power is o, press the power switch located on the rear panel. See gure 1.2.
2. Press the key to turn on the RFM3000.
3. After a self-check, the instrument will execute the application program.
– There will be some momentary dialogs indicating application initialization and channel updating.
– After the last dialog the main screen will be on the display.
When selecting a sensor for an exercise or a measurement, be sure you know the power range of the sensor. Operation beyond the specied upper power limit may result in a permanent change of characteristics or burnout.
4. Connect the sensor USB cable to the Channel 1 input on the front or rear of the instrument.
– When the sensor is connected or disconnected, the instrument will momentarily show a channel update dialog.
Note:
Connecting the Sync cable from the Multi I/O port on the sensor to the corresponding Sync port on the instrument for the sensor in use is necessary if using an external trigger or when performing measurements across multiple chan­nels
5. Use the nagivation keys to navigate the menus.
– The touch screen can be used to navigate the menus.
6. From the Main menu, select the Measure menu, and navigate to the Meas. Settings option.
7. Select Initialize.
– This will load the default operating parameters listed in table 3.1.
– This table only shows the parameters that are aected by initialization.
Getting Started 21
Parameter Default
Measure Mode Select Graph
Parameters Related to the Measure Menu Measurement Measurement Run
Parameters Related to the Display Menu View Graph
Parameters Related to the Channel > Channel # > Menus Channel1 On Channel2 On Channel3 On(4 CH option) Channel4 On(4 CH option) Vertical Scale 10 dB/Div Vertical Center -20.00 dBm Averaging 8 Units dBm Video BW HIGH Peak Hold OFF dB Oset 0 dB
Parameters Related to the Time Menu Timebase 100 uS/div Position 5.0 divisions Trigger Delay 0.0 uS
Parameters Related to the Trigger Menu Holdo 0 uS Trigger Mode AUTOPKPK Trigger Slope Positive Trigger Source CH1
Parameters Related to the Markers Menu Marker 1 -300 uS Marker 2 300 uS
Parameters Related to the Pulse Def. > CH# Pulse Def > Menus Distal 90% Mesial 50% Proximal 10%
Pulse Units Watts Start Gate 5.00% End Gate 95%
Parameters Related to the Stat Mode Menu Cursor Mode Power
Table 3.1 Default
Getting Started 22
3.2 Taking Measurements
To perform accurate measurements, the following is a minimum list of things you should know about the signal that you wish to measure.
Signal Frequency
The center frequency of the carrier must be known to allow sensor frequency response compensation.
Modulation Bandwidth
If the signal is modulated, know the type of modulation and its bandwidth. Note that power sensors respond only to the amplitude modulation component of the modulation, and constant envelope modulation types such as FM can be considered a CW carrier for power measurement purposes.
3.2.1 Continuous Mode
Continuous mode is best for measuring repetitive signals. Since this mode performs a continuous measurement, it does not dierentiate between the times a pulsed or periodic signal is o, and the times it is on. If you wish to make measurements that are synchronous with a period of a waveform, consider using Pulse mode instead.
Continuous mode is best for the following types of measurement:
• Moderate signal level (above -40 dBm for Peak sensors and -60 dBm for CW sensors).
• Signal that is CW or continuously modulated with a modulation bandwidth that is less than the VBW of the sensor in use.
• Signal modulation may be periodic, but only non-synchronous measurements are needed (overall average and peak power).
• Noise-like digitally modulated signals such as CDMA and OFDM when only average measurements are needed.
The measured result is the average power of the signal. Since the graphic display would basically just show a straight line, measurements in Continuous mode are best viewed using the Text Display mode. Figure 3.1 shows a two-channel measurement displaying an average, minimum, and maximum power in Continuous mode.
Figure 3.1 Average
Getting Started 23
3.2.2 Pulse Mode
For periodic or pulsed signals, it is often necessary to analyze the power for a portion of the waveform, or a certain region of a pulse or pulse burst. For these applications, the RFM3000 Series has a triggered Pulse mode.
The trigger signal can be either internal, triggered from a rising or falling edge on the measured signal; or external, triggered from a rear-panel BNC input. The trigger level and polarity are both programmable, as is the trigger delay time and trigger holdo time. Displays of both pre- and posttrigger data are available, and an auto-trigger mode can be used to keep the trace running when no trigger edges are detected.
An auto peak-to-peak trigger level setting can be chosen to automatically set the trigger level based on the currently applied signal. The timebase can be set from 5 ns/div to 50 ms/div. The RFM3000’s graphical display has 10 horizontal and 8 vertical divisions. Vertical units can be set in dBm, Watts, and dB Volts. Setting vertical resolution does not aect the sensitivity of the instrument and is provided for ease of viewing.
Programmable markers can be moved to any portion of the trace that is visible on the screen. They can be used to mark regions of interest for detailed power analysis. The instrument can display power at each marker, as well as average, minimum, and maximum power in the region between the two markers. This is very useful for examining the power during a TDMA or GSM burst when only the modulated portion in the center region of a timeslot is of interest.
By adjusting trigger delay and other parameters, it is possible to measure the power of specic timeslots within the burst. Trigger holdo allows burst synchronization even if there is more than one edge in the burst that may satisfy the trigger level. Set the holdo time to slightly shorter than the burst’s repetition interval to guarantee that triggering occurs at the same point in the burst each sweep. Figure 3.2shows marker measurements for pulses on CH1 and CH2.
Figure 3.2 Marker Measurements
Pulse mode is only available when using RFP Series power sensors and is the best choice for most pulse modulated and periodic signals. Pulse mode requires a repeating signal edge that can be used as a trigger, or an external trigger pulse that is synchronized with the modulation cycle.
Pulse mode performs measurements that are synchronous with the trigger (that is the measurements are timed or gated) so that the same portion of the waveform is measured on each successive modulation cycle. Multiple modulation cycles may be averaged together, and measurement intervals may span both before and after the trigger.
Getting Started 24
Pulse mode is best for the following types of measurements:
• Moderate signal level (above about -40 dBm except when modulation is o).
• The signal is periodic.
• A time snapshot of a single event is needed (minimum single-shot time is 200 nanoseconds).
• Typical modulation and signal types: LTE, 5G, RADAR, SatCom, TCAS, Bluetooth, Wireless LAN.
3.2.3 Statistical Mode
Certain modulated signals are completely random and provide no event that can serve as a trigger for measurements. CDMA or OFDM are common examples. The RFM3000’s Statistical mode was designed to provide measurements for these types of signals.
Statistical mode is only available when using peak power sensor. It is the best choice for analyzing signals with a high crest factor, that are noise-like with random or infrequent peaks, or are modulated in a random, non-periodic fashion. Statistical mode yields information about the probability of occurrence of various power levels without regard for when those power levels occurred.
In Statistical mode the instrument continuously samples the input signal and processes all of the samples to build power histograms. Many digitally modulated spread-spectrum formats use bandwidth coding techniques or many individual modulated carriers to distribute a source’s digital information over a wide bandwidth, and temporally spread the data for improved robustness against interference. When these techniques are used, it is dicult to predict when peak signal levels will occur. Analysis of millions of data points gathered during a sustained measurement of several seconds or more can yield the statistical probabilities of each signal level with a high degree of condence.
Statistical mode is best forthe following types of measurements:
• Moderate signal level (above about -40 dBm except when modulation is o).
• Noise-like digitally modulated signals such as CDMA (and all its extensions) or OFDM when probability information is helpful in analyzing the signal.
• Any signal with random, infrequent peaks, when you need to know the peak-to-average ratio or Crest Factor and just how infrequent those peaks are.
Complementary Cumulative Distribution Function (CCDF)
The statistical analysis of the current sample population is displayed using a normalized Complementary Cumulative Distribution Function (CCDF) presentation shown in gure ??. The CCDF is the probability of occurrence of a range of peak-to-average power ratios on a log-log scale. CCDF is non-increasing in y-axis and the maximum power sample lies at 0%. A cursor allows measurement of power or percentage at a user-dened point on the CCDF. As with all other graphical displays, the trace can be easily scaled and zoomed. The statistical data may be presented as a table in Text Display mode. The CCDF is a useful tool for analyzing communication signals that have a Gaussian-like distribution (CDMA, OFDM) where signal compression can be observed at rarely occurring peaks. It is most often presented graphically using a log-log format where the x-axis represents the relative oset in dB from the average power level and the log-scaled y-axis is the percent probability that power will exceed the x-axis value.
Getting Started 25
CCDF CCDF with Cursor Menu
Figure 3.3 CCDF
In a non-statistical peak power measurement, the peak-to-average ratio is the parameter that describes the headroom required in linear ampliers to prevent clipping or compressing the modulated carrier. The meaning of this ratio is easy to visualize in the case of simple modulation in which there is close correspondence between the modulating waveform and the carrier envelope. When this correspondence is not present, the peak-to-average ratio alone does not provide adequate information.
It is necessary to know what fraction of time the power is above (or below) particular levels. For example, some digital modulation schemes produce narrow and relatively infrequent power peaks that can be compressed with minimal eect. The peak-to-average ratio alone would not reveal anything about the fractional time occurrence of the peaks, but the CCDF clearly shows this information. In Figure 3-8, assume a full length run of one hour plus has been made and the CCDF is analyzed. At 0% is the maximum peak power that occurred during the entire run. At 1% is the power level that was exceeded only 1% of the time during the entire run.
Note that this analysis does not depend upon any particular test signal, or upon synchronization with the modulating signal. The analysis can be done using actual communication system signals. Normal operation is not disturbed by the need to inject special test signals. This type of analysis is particularly suited to the situation in which the bit error rate (BER) or some other error rate measure is correlated with the percentage of time that the signal is corrupted. If known short intervals of clipping are tolerable, the CCDF can be used to determine optimum transmitter power output. The CCDF is also used to evaluate various modulation schemes to determine the demands that will be made on linear ampliers and transmitters and the sensitivity to non-linear behavior.
Operation
This section presents the control menus and procedures for operating the RFM3000 in the manual mode. All the display menus that control the instrument are illustrated and accompanied by instructions for using each menu item.
4.1 Control Menus
The menus that control the RFM3000 RF are accessed from the top-level MAIN menu. Display the MAIN menu by
selecting the icon on the display. The Menus and parameters may be selected by using the navigation keys or the touch screen.
Some menus have mode-dependent properties. Typically, one or more menu boxes in a submenu may change as the measurement mode is changed from Continuous to Pulse to Statistical. Section Menu Reference these menus are indicated for mode dependency.
Menus with more selections than what ts on the display can be scrolled with the touch screen or front panel buttons.
4.2 Parameter Date Entry and Selection
The RFM3000’s parameters can be changed in various ways depending on the type of parameter being addressed.
4.2.1 Numerical Data Entry & Drop Down Menus
Numerical data can be incremented/decremented by selecting the “+” and “–“ icons with the front panel keys or by touching them. Selecting the numeric setting brings up an on-screen numeric keypad as shown in gure 4.1.
Some parameters use a drop down menu to select a setting. The Trigger source setting in gure 4.1 is an example of a drop down menu. Use the arrow up and down icons to cycle through the settings or select the value to view all available option in the drop down menu.
Numeric Keypad Trig Source Drop Down Menu
Figure 4.1 Numerical Data Entry & Drop Down Menus
Operation 27
4.3 Menu Reference
4.3.1 Main Menu
To open the Main menu press the icon.
The Main Menu shown in gure 4.2 is the top most menu level from which all other menus originate.
Measure >
Display >
Stat. Mode >
Channel >
Time >
Trigger >
Markers >
Pulse Def. >
Favorites >
System >
When navigating any submenu the icon will appear to the right of the submenu name.
Press the icon to return to the main menu.
Figure 4.2 Main Menu
Operation 28
4.3.2 Measure >
To enter the Measure menu:
Press the icon then select the Measure menu.
The available setting vary depending on the mode selected. See gure 4.3 to view the available settings in each menu.
Measurement
Toogle the state of the data acquisition mode for taking measurements. If Measurement is set to Run, the RFM3000 immediately begins taking measurements (Continuous and Statistical modes), or arms its trigger and takes a measurement each time a trigger occurs (Pulse Mode). If set to Stop, the measurement will begin (or be armed) if Start is selected under the Single Sweep setting (Pulse Mode Only), and will stop once the measurement criteria (averaging) has been satised.
Meas. Clear/Stat Capture
Selecting Execute clears display traces, data buers, and clears averaging lters to empty. In Statistical Measure mode, the menu item is replaced with Stat Capture and selecting Reset clears the acquired sample population.
Single Sweep
Only available in Pulse mode. Select to start a single measurement cycle in Pulse mode when Measurement is set to Stop. Enough trace sweeps must be triggered to satisfy the channel averaging setting.
Meas. Settings
This will load the default operating parameters listed in table 3.1. Only the settings shown in the table are aected and all others remain in their current state.
Cont. Mode
Measure Menu
Pulse Mode
Measure Menu
Figure 4.3 Measure Menus
Stat. Mode
Measure Menu
Operation 29
4.3.3 Display >
To enter the Display menu press the icon then select the Display menu.
View
Toogle between Text and Graph view. Text view displays a table of measurements for the current measurement mode. In Continuous mode, selecting Text displays the enabled channel power as shown in gure 3.1. Selecting Graph displays the trace graphical view.
Envelope
Enables/disables the Envelope display mode. In Pulse and Modulated modes, the Enve­lope display is used to highlight the range of signal excursions. When Envelope display mode is On, the trace is drawn as a wide line. The line is lled in between the minimum and maximum power readings. A series of vertical pixels, representing the range of signal excursions or "envelope" of the signal will be illuminated for each horizontal trace pixel. Envelope display is only available for Peak Power sensors.
Key Beep
Enable/disable the audible key beep.
4.3.4 Stat. Mode >
To enter the Stat. Mode menu press the icon then select the Stat. Mode menu.
Cursor Pct./Cur Pow Ref
Sets the CCDF cursor to the desired probability. When Cursor mode is set to Power, the menu item changes to Cur Pow Ref and sets the desired power
Cursor Mode
Select the independent variable for the CCDF cursor. If Percent is selected, probability at the cursor’s intersection with the CCDF curve will be measured. If Power is selected, relative power at the cursor’s intersection with the CCDF curve will be measured.
Figure 4.4 Display Menu
Horiz Scale
Select the horizontal scale for the statistical graphic display.
Horiz Oset
Select the horizontal oset for statistical graphic display.
Figure 4.5 Stat. Mode
Operation 30
Stat. Gating
Select Freerun or Markers gating for statistical acquisition. If Freerun is selected, all the samples are acquired without regard of sweep acquisition. If Markers are selected, then only samples within the time marker interval on the Pulse mode triggered sweep will be included in the statistical sample population.
Term Count
Sets the terminal sample count for the CCDF acquisition.
Term Time
Sets the terminal running time for the CCDF acquisition.
Figure 4.6 Stat. Mode
Term Action
Select the action to take when either the terminal count is reached or terminal time has elapsed. Selecting Decimate will divide all sample bins by 2 and continue. The total sample count will be halved each time a decimation occurs. Selecting Restart clears the statistical sample population and starts a new one. Selecting Stop will stop accumulating samples and hold the result.
4.3.5 Channel >
To enter the Channel menu press the icon then select the Channel menu.
Channel #
Select the channels settings menu to be congured.
Note:
The base RFM3000 model only permits two sensors to be active at any one time. With the RFM-4CH option, four sensors can be active at any one time.
Figure 4.7 Channels
Operation 31
4.3.6 Channel Settings
To enter the Channel menu press the icon then select the Channel > Channel # menu.
Chan Enabled
Toggle the display of the trace and measurements for the selected channel. The chan­nel may still be used as a trigger source when set to O.
Vert Scale
Set the power or voltage vertical axis level for the trace display based on the untis as shown in table 4.8
Units Scale
dBm 0.1, 0.2, 0.5 1, 2, 5, 10, 20, 50 dB/div
Watts 1 pW to 500 MW/div in a 1-2-5 progression
Volts 1 µV to 100 kV/ div in a 1-2-5 progression
Table 4.1 Vertical Scale Range
Vert Center
Set the power or voltage, horizontal centerline, level of the graph for the specied channel in the selected channel untis.
Averaging
Only available in Pulse and Statistical modes. Set the number of traces averaged to­gether to form the measurement result on the selected channel. Averaging can be used to reduce display noise on both the visible trace, marker, and automatic pulse measure­ments.
Trace averaging is a continuous process in which the measurement points from each sweep are weighted (multiplied) by an appropriate factor and averaged into the existing trace data points.
The most recent data will always have the greatest eect on the trace waveform, and older measurements will be decayed at a rate determined by the averaging setting and trigger rate. This averaging technique is often referred to as “exponential” averaging because averaging imposes a rst-order Innite Impulse Response (IIR) exponential lter with a time constant of "n" where n is the Average (number of averages) setting.
Note:
Figure 4.8
Channel Setttings
Figure 4.9
Channel Setttings
For timebase settings of 200 ns/div and faster, the RFP3000 Series sensors acquire samples using a technique called equivalent time or random interleaved sampling (RIS). In this mode, not every pixel on the trace gets updated on each sweep, and the total number of sweeps needed to satisfy the average setting will be increased by the sample in­terleave ratio of that particular timebase. At all times the average trace is the average of all samples for each pixel, and the min/max are the lowest and highest of that same block of samples for each pixel.
Operation 32
Units
Select the channel units. The trace may be shown in units of dBm, Watts, or Volts. The Units selection determines the range of the scale values and also aects displayed text and measurement values.
Freq. Corr
Sets the measurement frequency for the RF signal that is applied to the sensor for the current measurement. The appropriate frequency calibration factor from the sensor’s calibration table will be interpolated and applied automatically. Application of this calibration factor compensates for the eect of variations in the atness of the sensor’s frequency response.
Note:
The power sensor has no way to determine the carrier frequency of the applied signal, so the user must always enter the frequency.
Filter State
Available in Continuous mode only. Sets the current value of the integration lter. The lter can be set to O, On, or Auto.
O, provides no ltering, and can be used at high signal levels when absolute minimum settling time is required.
On, allows a user-specied integration time to be entered for use.
Auto, uses a variable amount of ltering, which is set automatically by the power me­ter based on the current signal level to a value that gives a good compromise between measurement noise and settling time at most levels.
Figure 4.10
Channel Setttings
Filter Time
Available in Continuous mode only. Sets the length of the integration lter. The lter is a “sliding window” which averages samples taken within a time window whose duration is set by this eld. All samples within the time window are equally weighted.
Duty Cycle
Available in Continuous mode only. Sets the duty cycle in percent for calculated CW pulse power measurements. Setting the duty cycle to 100% is equivalent to a CW measurement.
Video BW
Sets the sensor video bandwidth of the selected channel. HIGH is appropriate for most measurements. The actual bandwidth depends upon the sensor model. LOW bandwidth oers additional noise reduction for CW or signals with very low modulation bandwidth. If LOW bandwidth is used on signals with fast modulation, measurement errors may result if the sensor cannot track the fast-changing envelope of the signal.
Figure 4.11
Channel Setttings
Operation 33
Peak Hold
Set the operating mode of the selected channel’s peak hold function.
When set to OFF, peak values are not held.
When set to instantaneous (INST) instantaneous peak readings are held until reset by a new acquisition or cleared manually. This setting is used when it is desirable to hold the highest peak over a long measurement interval without any decay.
When set to average (AVG) peak readings are held for a short time, and then decayed towards the average power at a rate proportional to the Averaging setting. This is the best setting for most signals, because the peak will always represent the peak power of the current signal, and the resulting peak-to-average ratio will be correct shortly after any signal level changes.
dB Oset
Sets a measurement oset in dB for the selected channel. This is used to compen­sate for external couplers, attenuators, or ampliers in the RF signal path ahead of the power sensor.
Zero
Performs a zero oset null adjustment. The sensor does not need to be connected to any calibrator for zeroing. This action removes the eect of small, residual power o­sets, and should be performed prior to low-level measurements. The procedure is often performed in-system. There should be no RF signal applied to the sensor input prior to zeroing.
Figure 4.12
Channel Setttings
Fixed Cal
Performs a single point sensor gain calibration of the selected channel at 0 dBm and the current frequency setting. This requires a calibrated 0 dBm (1.00 mW) signal source at the current measurement frequency. This procedure calibrates the sensor’s gain at a single point. At other levels, that gain setting is combined with stored linearity factors to compute the actual power.
The built-in test source of the RFM3000 is not a suciently calibrated source for performing a xed calibration. An external calibration source is required. Note that xed calibration is NOT REQUIRED for USB power sensors.
Operation 34
4.3.7 Time >
To enter the Time menu press the icon then select the Time tab.
Timebase
Controls the timebase, horizontal scale, of the Trace View. The Timebase pulldown menu permits selection of xed timebase ranges from 5 ns/div to 50 ms/div (sensor series dependent) in a 1-2-5 progression.
Position
Sets the location of the trigger point on the acquired trace waveform. The Trig Delay setting is in addition to this setting, and will cause the trigger position to appear in a dierent location.
Figure 4.13
Time Setttings
Trig Delay
The trigger delay time is set in seconds with respect to the trigger. Positive values means that the trace display shows a time interval after the trigger event. This positions the trigger event to the left of the trigger point on the display, and is useful for viewing events during a pulse, or some xed delay time after the rising edge trigger. Negative trigger delay mean that the trace display shows a time interval before the trigger event, and is useful for looking at events preceding the trigger edge.
4.3.8 Trigger >
To enter the Trigger menu press the icon then select the Trigger tab.
Trigger Holdo
Set the trigger holdo time. Trigger holdo is used to disable the trigger for a spec­ied amount of time after each trigger event. The holdo time starts immediately after each valid trigger edge and will not permit any new triggers until the time has expired. When the holdo time is up, the trigger re-arms, and the next valid trig­ger event (edge) will cause a new sweep. This feature is used to help synchronize the power meter with burst waveforms such as a TDMA or GSM frame. The trigger hold­o resolution is 0.01 microseconds, and it should be set to a time that is longer than the burst duration but shorter than the frame repetition interval.
Trigger Level
Sets the threshold level for the trigger signal used in the Auto and Normal trigger modes. The trigger level can be entered numerically or changed by using arrow keys. The trigger level range has a range that is sensor model dependent (see the sensor specications for your specic sensor model).
The trigger range is automatically adjusted to include the dB Oset parameter set for the source channel. For example, if the trigger level = 10 dBm and the dB Oset is changed
Figure 4.14
Trigger Setttings
Operation 35
from 0 to 20 dB, then the oset-adjusted trigger level will be displayed to the user as 30 dBm. Likewise, the maximum trigger level range will be extended to 40 dBm. The trigger level set point and setting range are both shifted upward by 20 dB.
Trigger Mode
Set the trigger mode for synchronizing data acquisition with pulsed signals.
Normal mode will cause a sweep to be triggered each time the power level crosses the preset trigger level in the direction specied by the trigger slope setting. If there are no edges that cross this level, no data acquisition will occur.
Auto mode operates in much the same way as Normal mode but will automatically generate a trace if no trigger edges are detected for a period of time (100 to 500 milliseconds, depending upon timebase). This will keep the trace updating even if the pulse edges stop.
The Auto PK-PK mode operates the same as Auto mode but will adjust the trigger level to halfway between the highest and lowest power or voltage levels detected. This aids in maintaining synchronization with a pulse signal of varying level.
The Freerun mode forces unsynchronized traces at a high rate to assist in locating the signal.
Trigger Source
Set the trigger source used for synchronizing data acquisition. The CH # settings use the signal from the associated sensor. Ext setting uses the signal applied to the rear panel TRIG IN connector.
The trigger source can be any of the resource channels (CH1, CH2, etc.), or the Ext(ernal) trigger input signal. The Ind(ependent) trigger setting allows each connected sensor to trigger independently from its own RF input.
The external trigger is attached to the Trig In BNC connector on the rear of the RFM3000 Power Meter and requires a TTL signal level, minimum pulse width of 10 ns, and maximum frequency of 50 MHz.
Note:
Connecting the Sync cable from the Multi I/O port on the sensor to the corresponding Sync port on the instrument for the sensor in use is necessary if using an external trigger or when performing measurements across multiple chan­nels.
Trigger Slope
Set the trigger slope or polarity. When set to +, trigger events will be generated when a signal’s rising edge crosses the trigger level threshold. When – is selected, trigger events are generated on the falling edge of the pulse.
Operation 36
Markers >
To enter the Markers menu press the icon then select the Markers > tab.
Marker #
Set the time position of marker 1 or 2 relative to the trigger. Note that time markers must be positioned within the time limits of the trace window in the graph display. If a time outside of the display limits is entered, the marker will be placed at the rst or last time position as appropriate.
ΔTime
Displays the result of Marker 2 - 1 in seconds. This item is read only.
Figure 4.15
Marker Setttings
4.3.9 Pulse Def. >
To enter the Pulse Def menu press the icon then select the Pusle Def. > tab.
CH # Pulse Def
Select the channel to be congured.
Figure 4.16
Pulse Def Menu
4.3.10 CH# Pulse Def
To enter the CH# Pulse Def menu press the icon then select the CH# Pusle Def. > tab of the channel to be congured.
Operation 37
Distal
Sets the pulse amplitude percentage that denes the end of a rising edge or beginning of a falling edge transition. Typically, this is 90% voltage or 81% power relative to the top level of the pulse. This setting is used when making automatic pulse risetime and falltime calculations.
Mesial
Sets the pulse amplitude percentage that denes the midpoint of a rising or falling edge transition. Typically, this is 50% voltage or 25% power relative to the top level of the pulse. This setting is used when making automatic pulse width and duty cycle calculations.
Proximal
Sets the pulse amplitude percentage that denes the beginning of a rising edge or end of a falling edge transition. Typically, this is 10% voltage or 1% power relative to the top level of the pulse. This setting is used when making automatic pulse risetime and falltime calculations.
Pulse Unites
Controls whether the distal, mesial, and proximal thresholds are computed as voltage or power percentages of the top/bottom amplitudes. If Volts is selected, the pulse tran­sition thresholds are computed as voltage percentages. If Watts are selected, they are computed as power percentages. units setting.
Many pulse measurements call for 10% to 90% voltage (which equates to 1% to 81% power) for risetime and falltime measurements, and measure pulse widths from the half-power (–3 dB, 50% power, or 71% voltage) points. The Pulse Units setting is independent of the channel’s display
Figure 4.17
CH# Pulse Def Menu
Start Gate
Sets the beginning of the pulse measurement region as a percentage of the pulse width. The Start Gate has a continuous range of 0.0% to 40.0% of the pulse width and may be entered numerically or varied using the up or down arrows.
End Gate
Sets the end of the pulse measurement region as a percentage of the pulse width. The End Gate has a continuous range of 60.0% to 100.0% of the pulse width and may be entered numerically or varied using the up or down arrows. The Gate settings dene the measurement interval for the following power related pulse measurements: Pulse Average, Pulse Peak, Pulse Minimum, and Pulse Droop/Tilt. Pulse timing measurements between mesial crossings such as width and period are not aected. The purpose of the Pulse Gate setting is to exclude edge transition eects from the pulse power measurements.
Operation 38
4.3.11 Favorites >
To enter the Favorites menu press the icon then select the Favorites > tab. This function is not fully implemented at this time). Enables the user to setup a customized menu to allow grouping frequently used menu items into one convenient menu.
4.3.12 System >
To enter the System menu press the icon then select the System > tab.
The System menu displays the available system-level features and functionality.
Seonsr Data >
I/O Cong >
Calibration >
Exit >
Reports >
Update Software
cm
Figure 4.18 System Menu
Sensor Data >
To enter the Sensor Data menu press the icon then select the System > Sensor Data > tab.
Press Show to display information about the selected sensor in a pop-up log. See gure 4.19.
Figure 4.19 CH1 Information
Figure 4.20 Sensor Data
Operation 39
I/O Cong >
To enter the I/O Cong menu press the icon then select the System > I/O Cong > tab.
The RFM3000 supports remote communication over LAN and GPIB (optional).
GPIB Address
Set and View the current GPIB address in use for instruments equipped with GPIB option.
To increae the GPIB address press the icon or the icon do decrease the ad­dress. Pressing the number box located between the increase and decrease icon will open the numeric keypad. The numeric keypad can be used to instantaneously change the GPIB address
LAN
Figure 4.21 I/O Cong
To enter the I/O Cong menu press the icon then select the System > I/O Cong > LAN > tab.
DHCP/AutoIP
Set the state of DHCP/AutoIP system for the Ethernet port.
If DHCP/AutoIP is enabled (On), the instrument will attempt to obtain its IP Address and Subnet Mask, a DHCP (dynamic host conguration protocol) server on the net­work. If no DHCP server is found, the instrument will select its own IP Address and Subnet Mask values using the AutoIP protocol.
If DHCP/AutoIP is disabled (O), the instrument will use the IP Address and Subnet Mask values that have been set by the user.
IP Address
Set the Internet Protocol (IP) address of the Ethernet adapter. If DHCP/AutoIP mode is enabled, this menu is read-only.
Subnet Mask
Figure 4.22 LAN
Set the subnet mask for the Ethernet adapter. If DHCP/AutoIP mode is enabled, this menu is read-only.
MAC Address
Displays the MAC address for the Ethernet adapter. This menu item is read-only.
Operation 40
Calibration >
To enter the Calibrator menu press the icon then select the System > Calibrator > tab.
Cal Output
Enable/disable the output of the built-in 0 dBm 50 MHz test source.
Note:
The built-in test source of the RFM3000 is not a suciently calibrated source for per­forming a xed calibration. An external calibration source is required.
Figure 4.23 Calibrator
Exit >
To enter the Exit menu press the icon then select the System > Exit > tab.
Exit to Desktop
Exits the RFM3000 Power Meter Main application to access the OS Desktop.
Shut Down
Shuts down power to the PMX40 putting the meter in standby mode and is the same as pressing the ON/Standby button on the front panel.
Reports >
To enter the Reports menu press the icon then select the System > Reports > tab.
Conguration
Select Show to display an About dialog with conguration information for the RFM3000 Power Meter like that shown in gure 2.2.
Figure 4.24 Exit
Figure 4.25 Reports
Operation 41
Update Software
To view the Update Sofware option press the icon then select the System > tab.
Update Software
Select Go to search the connected USB drive for the *.tar software update le and up­date or re-install the version found. If no valid le is found, the dialog in gure 4.27 appears.
Figure 4.26
Update Software
Figure 4.27 Update Error
Application Notes
This section provides supplementary material to enhance your knowledge of the RFM3000 operation, advanced fea­tures, and measurement accuracy. Topics covered in this section include pulse measurement fundamentals, automatic measurement principles, and an analysis of measurement accuracy.
5.1 Introduction to Pulse Measurements
5.1.1 Measurement Fundamentals
The following is a brief reaview of the power measurement fundamentals.
Unmodulated Carrier Power
The average power of an unmodulated carrier consisting of a continuous, constant amplitude sinewave signal is also termed continuous wave (CW) power. For a known value of load impedance R, and applied voltage 𝑉 power is:
2
𝑉
𝑟𝑚𝑠
𝑃 =
𝑅
𝑤𝑎𝑡𝑡𝑠
, the average
𝑟𝑚𝑠
Power meters designed to measure CW power can use thermoelectric-based sensors which respond to the heating eect of the signal or diode detectors which respond to the voltage of the signal. With careful calibration accurate measurements can be obtained over a wide range of input power levels.
Modulated Carrier Power
The average power of a modulated carrier which has varying amplitude can be measured accurately by a CW type power meter with a thermoelectric detector, but the lack of sensitivity will limit the range. Diode detectors can be used at low power, square-law response levels. At higher power levels the diode responds in a more linear manner and signicant error results.
Pulse Power
Pulse power refers to power measured during the on time of pulsed RF signals gure 5.1. Traditionally, these signals have been measured in two steps: (1) thermoelectric sensors measure the average signal power, (2) the reading is then divided by the duty cycle to obtain pulse power, 𝑃
𝑃
Where Duty Cycle:
𝐷𝑢𝑡𝑦 𝐶𝑦𝑐𝑙𝑒 =
𝑝𝑢𝑙𝑠𝑒
𝑝𝑢𝑙𝑠𝑒
:
𝐴𝑣𝑒𝑟𝑎𝑔𝑒 𝑃 𝑜𝑤𝑒𝑟
=
𝐷𝑢𝑡𝑦 𝐶𝑦𝑐𝑙𝑒
𝑃𝑢𝑙𝑠𝑒 𝑊𝑖𝑑𝑡ℎ
𝑃𝑢𝑙𝑠𝑒 𝑃 𝑒𝑟𝑖𝑜𝑑
Pulse power provides useful results when applied to rectangular pulses, but is inaccurate for pulse shapes that include distortions, such as overshoot or droop (Figure 5.2).
Application Notes 43
Figure 5.1 Pulsed RF Signal
Figure 5.2 Distorted Pulsed Signal
Peak Power
The RFM3000 makes power measurements in a manner that overcomes the limitations of the pulse power method and provides both peak power and average power readings for all types of modulated carriers. The fast-responding diode sensors detect the RF signal to produce a wideband video signal, which is sampled with a narrow sampling gate. The video sample levels are accurately converted to power on an individual basis at up to a 100 MSa/sec rate. Since this power conversion is corrected based upon the sensor’s linearity correction table, these samples can be averaged to yield average power without restriction to the diode square-law region.
If the signal is repetitive, the signal envelope can be reconstructed using an internal or external trigger. The envelope can be analyzed to obtain waveshape parameters including, pulse width, duty cycle, overshoot, rise time, fall time, and droop. In addition to time domain measurements and simple averaging, the RFM3000 has additional capabilities that allow it to perform statistical analysis on a complete set of continuously sampled data points.
Data can be viewed and characterized using a CCDF presentation format. These analysis tools provide invaluable information about peak power levels and their frequency of occurrence, and are especially useful for non-repetitive signals, such as those used in 5G and Wi-Fi applications.
Application Notes 44
5.1.2 Diode Detection
Wideband diode detectors are the dominant power sensing device used to measure pulsed RF signals. Several diode characteristics must be compensated to make meaningful measurements. These include the detector’s nonlinear ampli­tude response, temperature sensitivity, and frequency response characteristic. Additional potential error sources include detector mismatch, signal harmonics, and noise.
Detector Response
The response of a single-diode detector to a sinusoidal input is given by the diode equation:
𝑖 = 𝐼𝑠(𝑒𝑎𝑣− 1)
where:
𝑖 = 𝑑𝑖𝑜𝑑𝑒𝑐𝑢𝑟𝑟𝑒𝑛𝑡 𝑣 = 𝑛𝑒𝑡𝑣𝑜𝑙𝑡𝑎𝑔𝑒𝑎𝑐𝑟𝑜𝑠𝑠𝑡ℎ𝑒𝑑𝑖𝑜𝑑𝑒 𝐼𝑠= 𝑠𝑎𝑡𝑢𝑟𝑎𝑡𝑖𝑜𝑛𝑐𝑢𝑟𝑟𝑒𝑛𝑡 𝛼 = 𝑐𝑜𝑛𝑠𝑡𝑎𝑛𝑡
An ideal diode response curve is plotted in gure 5.3
Figure 5.3 Ideal Diode Response
The curve indicates that for low microwave input levels (Region A), the single-diode detector output is proportional to the square of the input power. For high input signal levels (Region C), the output is linearly proportional to the input. In between these ranges (Region B), the detector response lies between square-law and linear.
For accurate power measurements over all three regions illustrated in gure 5.3, the detector response is pre-calibrated over the entire range. The calibration data is stored in the instrument and recalled to adjust each sample of the pulse power measurement.
Application Notes 45
Temperature Eects
The sensitivity of microwave diode detectors (normally Low Barrier Schottky diodes) varies with temperature. However, ordinary circuit design procedures that compensate for temperature-induced errors adversely aect detector bandwidth. A more eective approach involves sensing the ambient temperature during calibration and recalibrating the sensor when the temperature drifts outside the calibrated range.
This process can be made automatic by collecting calibration data over a wide temperature range and saving the data in a form that can be used by the power meter to correct readings for ambient temperature changes.
Frequency Response
The carrier frequency response of a diode detector is determined mostly by the diode junction capacitance and the device lead inductances.
The frequency response will vary from detector to detector and cannot be compensated readily. Power measurements must be corrected by constructing a frequency response calibration table for each detector.
Mismatch
Sensor impedance matching errors can contribute signicantly to measurement uncertainty, depending on the mismatch between the device under test (DUT) and the sensor input. This error cannot be easily calibrated out, but can be minimized by employing an optimum matching circuit at the sensor input.
Signal Harmonics
Measurement errors resulting from harmonics of the carrier frequency are leveldependent and cannot be calibrated out. In the square-law region of the detector response (Region A, Figure 5-3), the signal and second harmonic combine on a root mean square basis. The eects of harmonics on measurement accuracy in this region are relatively insignicant. However, in the linear region (Region C, gure 5.3), the detector responds to the vector sum of the signal and harmonics. Depending on the relative amplitude and phase relationships between the harmonics and the fundamental, measurement accuracy may be signicantly degraded. Errors caused by even-order harmonics can be reduced by using balanced diode detectors for the power sensor. This design responds to the peak-to-peak amplitude of the signal, which remains constant for any phase relationship between fundamental and even-order harmonics. Unfortunately, for odd-order harmonics, the peak-to-peak signal amplitude is sensitive to phasing, and balanced detectors provide no harmonic error improvement. Measurement errors resulting from harmonics of the carrier frequency are leveldependent and cannot be calibrated out. In the square-law region of the detector response (Region A, gure 5.3), the signal and second harmonic combine on a root mean square basis. The eects of harmonics on measurement accuracy in this region are relatively insignicant. However, in the linear region (Region C, gure 5.3), the detector responds to the vector sum of the signal and harmonics. Depending on the relative amplitude and phase relationships between the harmonics and the fundamental, measurement accuracy may be signicantly degraded. Errors caused by even-order harmonics can be reduced by using balanced diode detectors for the power sensor. This design responds to the peak-to-peak amplitude of the signal, which remains constant for any phase relationship between fundamental and even-order harmonics. Unfortunately, for odd-order harmonics, the peak-to-peak signal amplitude is sensitive to phasing, and balanced detectors provide no harmonic error improvement.
Noise
For low-level signals, detector noise contributes to measurement uncertainty and cannot be calibrated out. Balanced detector sensors improve the signal-to-noise ratio by 3 dB, because the signal is twice as large.
Application Notes 46
5.1.3 Pulse Denitions
IEEE Std 194™-1977 Standard Pulse Terms and Denitions “provides fundamental denitions for general use in time domain pulse technology.” Several key terms dened in the standard are reproduced in this subsection, which also denes the terms appearing in the RFM3000 text mode display of automatic measurement results.
5.1.4 Standard IEEE Pulse
The key terms dened by the IEEE standard are abstracted and summarized below. These terms are referenced to the standard pulse illustrated ingure 5.4
Figure 5.4 Stanard IEEE Pulse
Note:
IEEE Std 194™-1977 Standard Pulse Terms and Denitions has been superseded by IEEE Std 181™-2003. Many of the terms used below have been deprecated by the IEEE. However, these terms are widely used in the industry. For this reason, they are retained.
Application Notes 47
Term Denition
Base Line
Top Line The portion of a pulse waveform which represents the second nominal state of a pulse.
First Transition
Last Transition
Proximal Line
Distal Line
Mesial Line
The two portions of a pulse waveform which represent the rst nominal state from which a pulse departs and to which it ultimately returns.
The major transition of a pulse waveform between the base line and the top line (com­monly called the rising edge).
The major transition of a pulse waveform between the top of the pulse and the base line (commonly called the falling edge).
A magnitude reference line located near the base of a pulse at a specied percentage (normally 10%) of pulse magnitude.
A magnitude reference line located near the top of a pulse at a specied percentage (normally 90%) of pulse magnitude.
A magnitude reference line located in the middle of a pulse at a specied percentage (normally 50%) of pulse magnitude.
Table 5.1 Pule Terms
5.1.5 Automatic Measurements
The RFM3000 automatically analyzes the waveform data in the buers and calculates key waveform parameters. The calculated values are displayed in text mode when you press the TEXT/GRAPH system key.
5.1.6 Automatic Measurement Criteria
Automatic measurements are made on repetitive signals that meet the following conditions:
Amplitude
– The dierence between the top and bottom signal amplitudes must exceed 6 dB to calculate waveform timing
parameters (pulse width, period, duty cycle). The top-to-bottom amplitude dierence must exceed 13 dB to measure rise and fall time.
Timing
To measure pulse repetition frequency and duty cycle, there must be at least three signal transitions. The interval between the rst and third transition must be at least 1/5 of a division (1/50 of the screen width). For best accuracy on rise and fall time measurements, the timebase should be set so the transition interval is at least one-half division on the display.
Application Notes 48
5.1.7 Automatic Measurement Terms
The following terms appear in the RFM3000 Text display in Pulse mode. The Text column lists the abbreviated forms that appear on the display screen.
Text TERM DEFINITION
Width Pulse Width The interval between the rst and second signal crossings of the mesial line.
Rise Risetime The interval between the rst signal crossing of the proximal line to the rst
signal crossing of the distal line.
Fall Falltime The interval between the last signal crossing of the distal line to the last signal
crossing of the proximal line.
Period Pulse Period The interval between two successive pulses (reciprocal of the Pulse Repetition
Frequency).
PRFreq Pulse Repetition The number of cycles of a repetitive signal that take place in one Frequency
second.
Duty C Duty Cycle The ratio of the pulse on-time to o-time. Otime O-time The time a repeti-
tive pulse is o (equal to the pulse period minus the pulse width).
Peak Peak Power The maximum power level of the captured waveform.
Pulse Pulse Power The average power level across the pulse width, dened by the intersection of
the pulse rising and falling edges with the mesial line.
Avg Average Power The equivalent heating eect of a signal. IEEETop Top Amplitude The ampli-
tude of the top line (see IEEE denitions). IEEEBot Bottom Amplitude The amplitude of the base line (see IEEE denitions). Skew Skew The time between the mesial level of a pulse on one channel and a pulse on a second channel.
EdgeDly Edge Delay The time between the left edge of the display and the rst mesial transition level
of either slope on the waveform.
Table 5.2 Automatic Measurement Terms
Application Notes 49
5.1.8 Automatic Measurement Sequence
The automatic measurement process analyzes the captured signal data in the following sequence:
1. Approximately 500 samples of the waveform (equivalent to one screen width) are scanned to determine the maximum and minimum sample amplitudes.
2. The dierence between the maximum and minimum sample values is calculated and stored as the Signal Amplitude.
3. The Transition Threshold is computed as one-half the sum of the maximum and minimum sample amplitudes.
4. The processor locates each crossing of the Transition Threshold.
5. Starting at the left edge of the screen, the processor classies each Transition threshold crossing according to whether it is positive-going (– +) or negative-going (+ –). Because the signal is repetitive, only three transitions are needed to classify the waveform, as follows:
Type Sequence Description
0 none No crossings detected
1 Not used
2 + – One falling edge
3 – + One rising edge
4 + – + One falling, followed by one rising edge
5 – + – One rising, followed by one falling edge
6 + – + – Two falling edges
7 – + – + Two rising edges
Table 5.3 Transition Threshold Crossing
Figure 5.5 Step Waveforms
Application Notes 50
6. If the signal is Type 0, (No crossings detected) no measurements can be performed and the routine is terminated, pending the next reload of the data buers.
7. The process locates the bottom amplitude (baseline) using the IEEE histogram method. A histogram is generated for all samples in the lowest 12.8 dB range of sample values. The range is subdivided into 64 power levels of 0.2 dB each. The histogram is scanned to locate the power level with the maximum number of crossings. This level is designated the baseline amplitude. If two or more power value have equal counts, the lowest is selected.
8. The process follows a similar procedure to locate the top amplitude (top line). The power range for the top histogram is 5 dB and the resolution is 0.02 dB, resulting in 250 levels. The level-crossing histogram is computed for a single pulse, using the samples which exceed the transition threshold. If only one transition exists in the buer (Types 2 and 3), the process uses the samples that lie between the edge of the screen and the transition threshold (see gure
5.6). For a level to be designated the top amplitude, the number of crossings of that level must be at least 1/16 the
number of pixels in the pulse width; otherwise, the peak value is designated the top amplitude.
Figure 5.6 Time Interpolation
9. The process establishes the proximal, mesial, and distal levels as a percentage of the dierence between top amplitude and bottom amplitude power. The percentage can be calculated on a power or voltage basis. The proximal, mesial, and distal threshold values are user settable from 1% to 99%, with the restriction that the proximal < mesial < distal. Normally, these values will be set to 10%, 50%, and 90%, respectively.
10. The process determines horizontal position, in pixels, at which the signal crosses the mesial value. This is done to a resolution of 0.1 pixel, or 1/5000 of the screen width. Ordinarily, the sample values do not fall precisely on the mesial line, and it is necessary to interpolate between the two nearest samples to determine where the mesial crossing occurred. This process is demonstrated in the example above (gure 5.6):
Application Notes 51
Item dBm mW
Mesial value 10.0 10.0
Sample n 8.0 6.3
Sample n+1 11.0 12.6
Table 5.4 Interpolation Crossing
The interpolated crossing time, 𝑡𝑥, is calculated from:
𝑡𝑥= 𝑡𝑛+
𝑚𝑒𝑠
𝑃
𝑛+1
− 𝑃
𝑛
𝑛
𝑃
− 𝑃
where 𝑃 is in watts and n is the number of the sampling interval, referenced to the trigger event. For this example:
𝑡𝑥= 𝑡𝑛+
10.0 − 6.3
12.6 − 6.3
𝑡𝑥= 𝑡𝑛+ 0.6
11. The processor computes the rise and/or fall times of waveforms that meet the following conditions:
• The waveform must have at least one usable edge (Types 2 through 7).
• The signal peak must be at least 13 dB greater than the minimum sample value.
The rise time is dened as the time between the proximal and distal crossings (– +).
The fall time is dened as the time between the distal and proximal crossings (+ –).
If no samples lie between the proximal and distal values for either edge (rise or fall), the risetime for that edge is set to 0 seconds.
12. The processor calculates the output values according to the following denitions:
a)
b)
c)
d)
e)
Pulse Width
Rise time
Fall time
Period Cycle
Pulse Repetition
Interval between mesial points
See Step 11
See Step 11
time between mesial points
Reciprocal of Period Frequency
f)
g)
h)
i)
j)
k)
l)
m)
n)
Duty Cycle Pulse
O-time
Peak Power
Pulse Power
Overshoot
Average Power
Top Amplitude
Bottom Amplitude
Skew
Pulse Width/Period
(Period) - (Pulse Width)
Maximum sample value (See Step 1)
Average power in the pulse (between the mesial points)
(Peak Power) - (Top Amplitude)
See Step 13
See Step 8
See Step 7
See Step 14
Application Notes 52
5.1.9 Average Power Over an Interval
13. The average power of the signal over a time interval is computed by:
a)
b)
summing the sample powers in the interval
dividing the sum by the number of samples
This process calculates Pulse Power, Average Power, and the average power between markers.
Since each sample represents the power in a nite time interval, the endpoints are handled separately to avoid spreading the interval by one-half pixel at each end of the interval (see gure ??). For the interval in gure ??, the average power is given by:
𝑎𝑣𝑔
1
=
(𝑃0+ 𝑃𝑛) +
2
𝑃
1
𝑛 − 1
𝑛−1
𝑛=1
𝑃
𝑛
Figure 5.7 Sampling Interval
14. The processor calculates the delay between the two measurement channels. The time reference for each channel is established by the rst signal crossing (starting from the left edge of the screen) which passes through the mesial level. The signal excursion must be at least 6 dB.
Application Notes 53
5.1.10 Statistical Mode Automatic Measurements
When operating in Statistical mode, the RFM3000 has a unique text format display that is available when the TEXT/GRAPH system key is pressed. A sample of the text display is shown in gure 5.8.
Figure 5.8 Statistical Mode Text Display
In the Statistical mode the following ve automatic measurements are displayed in the RFM3000 Text display for both input channels and both trigger channels. The Text column lists the abbreviated forms that appear on the display screen. In the Statistical mode the following ve automatic measurements are displayed in the PMX40 Text display for both input channels and both trigger channels. The Text column lists the abbreviated forms that appear on the display screen.
TEXT TERM DEFINITION
Avg Average Power The unweighted average of all linear power samples occurring since acqui-
sition started.
Peak Peak Power The highest power sample occurring since acquisition was started.
Min Minimum Power The lowest power sample occurring since acquisition was started. In loga-
rithmic units a reading below the clip level will display as down arrows.
Pk2Avg Pk/Avg Ratio The ratio (in dB) of the Peak Power to the Average Power.
Table 5.5 Statistical Automatic Measurements
Application Notes 54
The following six cursor measurements display the set position (independent variable) and measured value (dependent variables) where the movable cursor intersects the measurement trace.
The position or value measurement text for each dependent variable is displayed in the color of its channel. The independent variable is white.
Note that the intersection of the movable cursors and the CCDF traces can be moved outside the visible display area. This does not aect the measurements in any way.
TEXT TERM DEFINITION
Cursor Pwr Cursor Powe Cursor Mode - Power Ref
Reference The reference power level in dBr set by the user to dene the measure-
ment point on the normalized CCDF for probability in percent.
Cursor Mode - Percent The measured power level in dBr of the normalized CCDF at the Proba­bility in percent specied by the user.
Cursor Pct Cursor Percentage Cursor Mode - Power Ref
The measured probability in percent of the normalized CCDF at the refer­ence power level specied by the user.
Cursor Mode - Percent The probability in percent set by the user to dene the measurement point on the normalized CCDF for power level in dBr.
Total Time The total time in Hours:Minutes:Seconds that the data acquisition has
been running.
Points The total number of data samples in MSa that has been acquired for
each channel in the current run.
Table 5.6 Cursor Measurements
Note:
The total number of data samples is aected by the terminal settings. If Terminal Action is set to decimate, then the sample count will be halved each time the Terminal Count or Time is reached. This should have very little visible eect on the CCDF values, since the entire population is decimated uniformly. If Terminal Action is set to restart, then the sample count will be cleared to zero each time the Terminal Count or Time is reached.
5.2 Measurement Accuracy
The measurement accuracy of the RFM3000 is completely contingent upon the USB sensor with which it is being used. Please reference the sensor datasheet and/or associated uncertainty calculator for measurement uncertainties associated with a specic sensor.
Maintenance
This section presents procedures for maintaining the RFM3000.
6.1 Safety
The RFM3000 has been designed in accordance with international safety standards, general safety precautions must be observed during all phases of operation and maintenance. Failure to comply with the precautions listed in the Safety Summary located in the front of this manual could result in serious injury or death. Service and adjustments should be performed only by qualied service personnel.
6.2 Cleaning
Painted surfaces can be cleaned with a commercial spray-type window cleaner or a mild detergent and water solution.
Note:
When cleaning the instrument, do not allow cleaning uid to enter the fan intake and exhaust vents. Avoid using chemical cleaning agents that can damage painted or plastic surfaces.
6.3 Inspection
If the RFM3000 malfunctions, perform a visual inspection of the instrument. Inspect for signs of damage caused by excessive shock, vibration, or overheating. Inspect for broken wires, loose electrical connections, or accumulations of dust or other foreign matter.
Correct any problems you discover, reboot the instrument, and observe the self-test results (see gure 6.1). If the malfunction persists of the instrument fails the performance verication, contact BK Precision for service.
Figure 6.1 Self-Test Results
Maintenance 56
6.4 Lithium Battery
The RFM3000 contains one Lithium “coin cell” battery to provide for non-volatile storage of the instrument state. This is located on the Main Printed Circuit assembly. It should have a life of 5-10 years. When replacement is necessary, the battery must be disposed of in strict compliance with local environmental regulations.
6.5 Software Upgrade
Instrument operating software has been loaded into the Model RFM3000 at the factory, including the BK Precision Model RFM3000 Application Software. The Application Software will be updated from time to time to correct errors and add new features. Users can upgrade their software by downloading it from the BK Precision webpage, bkprecision.com. Copy the upgrade le(s) into the root directory on a USB drive, and plug the drive into one of the instrument’s USB ports (front or rear). From the RFM3000 application, select the System menu, and select Go under Update Software.
Note:
When loading new software into the Model RFM3000, some or all stored instrument congu­rations and preset operating selections may be lost. Contact BK Precision for information on which les may be aected.
Service Information
Warranty Service: Please go to the support and service section on our website at bkprecision.com to obtain an RMA #. Return the product in the original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device.
Non-Warranty Service: Please go to the support and service section on our website at bkprecision.com to obtain an RMA #. Return the product in the original packaging to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Customers not on an open account must include payment in the form of a money order or credit card. For the most current repair charges please refer to the service and support section on our website. Return all merchandise to B&K Precision Corp. with prepaid shipping. The at-rate repair charge for Non-Warranty Service does not include return shipping. Return shipping to locations in North America is included for Warranty Service. For overnight shipments and non-North American shipping fees please contact B&K Precision Corp.
Include with the returned instrument your complete return shipping address, contact name, phone number and description of problem.
B&K Precision Corp.
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
bkprecision.com
714-921-9095
LIMITED THREE-YEAR WARRANTY
B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of three years from date of purchase. B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt. To help us better serve you, please complete the warranty registration for your new instrument via our website www.bk­precision.com
Exclusions: This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized alterations or repairs. The warranty is void if the serial number is altered, defaced or removed.
B&K Precision Corp. shall not be liable for any consequential damages, including without limitation damages resulting from loss of use. Some states do not allow limitations of incidental or consequential damages. So the above limitation or exclusion may not apply to you. This warranty gives you specic rights and you may have other rights, which vary from state-to-state.
B&K Precision Corp.
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
www.bkprecision.com
714-921-9095
Version: June 29, 2021
Loading...