B&K Precision DAS30, DAS50, DAS60 User Manual

Safety Summary
The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
Read and understand the safety and operational information in this manual.Apply all the listed safety precautions.Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating the instrument
Make all connections to the instrument before applying power.Do not operate the instrument in ways not specied by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design, manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply with these requirements.
Category rating
The IEC 61010 standard denes safety category ratings that specify the amount of electrical energy available and the voltage impulses that may occur on electrical conductors associated with these category ratings. The category rating is a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of the circuit to be tested, which denes the voltage impulses expected and required insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected to the
mains supply. The voltages in the environment are typically derived from a limited-energy trans­former or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the mains
supply at a standard wall outlet or similar sources. Example measurement environments are portable tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be connected to the mains
installation of a building. Examples are measurements inside a building’s circuit breaker panel or the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to the primary
power entering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what is specied in this manual for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the instrument’s category rating to maintain the instrument’s category rating. Failure to do so will lower the category rating of the measuring system.
2
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be 115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This instrument is grounded through the ground conductor of the supplied, three-conductor AC line power cable. The power cable must be plugged into an approved three-conductor electrical outlet. The power jack and mating plug of the power cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive parts (including control knobs) may provide an electric shock. Failure to use a properly-grounded approved outlet and the recommended three-conductor AC line power cable may result in injury or death.
Unless otherwise stated, a ground connection on the instrument’s front or rear panel is for a reference of potential only and is not to be used as a safety ground. Do not operate in an explosive or ammable atmosphere.
Do not operate the instrument in the presence of ammable gases or vapors, fumes, or nely-divided particulates.
The instrument is designed to be used in oce-type indoor environments. Do not operate the instrument
In the presence of noxious, corrosive, or ammable fumes, gases, vapors, chemicals, or nely-divided particulates.In relative humidity conditions outside the instrument’s specications.In environments where there is a danger of any liquid being spilled on the instrument or where any liquid can condense
on the instrument.
In air temperatures exceeding the specied operating temperatures.In atmospheric pressures outside the specied altitude limits or where the surrounding gas is not air.In environments with restricted cooling air ow, even if the air temperatures are within specications.In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature range is 0∘C to 40∘C and 20% to 80% relative humidity, with no condensation allowed. Measurements made by this instrument may be outside specications if the instrument is used in non-oce-type environments. Such environments may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise, strong electric elds, or strong magnetic elds.
3
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the instrument, remove the instrument’s power cord, remove the instrument from service, label it as not to be operated, and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual. Not for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a life-support device or system.
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must be made by qualied service-trained maintenance personnel who are aware of the hazards involved when the instrument’s covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may exist when the covers are removed. To avoid injuries, always disconnect the power cord from the instrument, disconnect all other connections (for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there are no hazardous voltages present on any conductors by measurements with a properly-operating voltage-sensing device before touching any internal parts. Verify the voltage-sensing device is working properly before and after making the measurements by testing with known-operating voltage sources and test for both DC and AC voltages. Do not attempt any service or adjustment unless another person capable of rendering rst aid and resuscitation is present. Do not insert any object into an instrument’s ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit exists.
Fuse replacement must be done by qualied service-trained maintenance personnel who are aware of the instrument’s fuse requirements and safe replacement procedures. Disconnect the instrument from the power line before replacing fuses. Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings specied in this manual or on the back of the instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a re. Failure to use the specied fuses will void the warranty.
Servicing
4
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to B&K Precision for service and repair to ensure that safety and performance features are maintained.
For continued safe use of the instrument
Do not place heavy objects on the instrument.Do not obstruct cooling air ow to the instrument.Do not place a hot soldering iron on the instrument.Do not pull the instrument with the power cord, connected probe, or connected test lead.Do not move the instrument when a probe is connected to a circuit being tested.
Safety Symbols
Symbol Description
indicates a hazardous situation which, if not avoided, will result in death or serious injury.
indicates a hazardous situation which, if not avoided, could result in death or serious injury
indicates a hazardous situation which, if not avoided, will result in minor or moderate injury
Refer to the text near the symbol.
Electric Shock hazard
Alternating current (AC)
Chassis ground
Earth ground
This is the In position of the power switch when instrument is ON.
This is the Out position of the power switch when instrument is OFF.
is used to address practices not related to physical injury.
5
Contents
1 Introduction 10
1.1 Recording Modes 10
1.1.1 Direct mode 10
1.1.2 Memory mode 10
1.1.3 Go No-Go mode 10
1.1.4 File mode 10
1.1.5 Power Analysis mode 10
2 Getting Started 11
2.1 Measurement Types 11
2.2 Powering the Recorder 11
2.3 Connections 11
2.3.1 Analog Inputs 12
2.4 Ground the Recorder 12
2.5 Thermal Printer 13
2.5.1 Loading Paper 13
2.5.2 Storing Printed Recordings 13
2.6 Power-up Conguration 13
2.7 Routine Maintenance 13
2.8 Battery 14
2.8.1 To recharge the battery 14
2.8.2 Safety precautions 14
2.9 Oset calibration 14
2.10 Default calibration 15
2.11 Screen brightness 15
2.11.1 Locking of the recorder 16
2.12 User mode 16
2.13 Update of the internal software 16
3 Home Screen 17
4 Setup Screen 19
4.1 Logic channels 20
4.2 Real time on-screen display of the measurements 21
4.3 XY display 24
4.4 Digital display 24
4.5 Trigger menu 25
4.6 Memory output key 25
4.7 Record key 26
4.8 Screen copy 27
5 FILE MANAGEMENT 28
5.1 Save setup les 29
5.2 Load setup les 29
5.3 Save data acquisitions 29
5.4 Recovery of the data acquisition les 30
5.5 Transfer via FTP 30
5.6 Display under SeframViewer 31
5.7 Managing with SeframPilot 31
6
6 Direct Mode 33
6.1 Launching the printing 33
6.2 Conguration of print 34
6.2.1 Relaunching the printing 35
6.2.2 Writing information 35
7 Memory Mode 36
7.1 Setup and start of the data acquisition 36
7.2 Sampling period 37
7.3 Internal memory, blocks 37
7.4 Trigger position 38
7.5 Double Trigger mode 38
7.6 Recording 38
7.7 Memory output 40
8 File Mode 42
8.1 Setup and triggering of the data acquisition 42
8.1 Restrictions 43
9 Go/No Go Mode 44
9.1 Setup and launching of the data acquisition 44
9.2 Blocks: parting of the internal memory in blocks 44
9.3 Real time save: simultaneous record of the data acquisition into a le 44
9.4 Start: start condition of the data acquisition 45
9.5 Automatic: automatic stop when the block is full 45
9.5.1 Creation of the template 45
9.6 Use of the template 46
10 Power Analysis Mode 47
10.1 Power Analysis 47
10.2 Network setup 48
10.3 Wiring and check 49
10.4 Fresnel diagram 50
10.5 Warning messages 50
10.6 Recording additional channels 50
10.7 Real time display of the signals 50
10.8 Power analysis setup 52
10.9 Parameter acquisition le 52
10.10 Record launch 52
10.10.1 Manual launch 52
10.10.2 Automatic trigger 53
10.10.3 Parametric trigger 53
10.11 You can select a stop condition (recording time or manual stop) 53
10.11.1 Trigger according to an inrush 53
10.11.2 In version 1.1 of the rmware or later 54
10.11.2.1 Trigger according to a duration or a date 54
10.12 Recordable parameters 54
10.12.1 Voltage and Current 54
10.12.2 Power 54
10.12.3 Power Factor 54
10.13 Record 55
10.14 Measurement process 56
7
11 Channel Setup 57
11.1 Analog channels 58
11.2 Type of measurement V/ t 59
11.2.1 Example 60
11.2.2 Type of measurement
 60
11.2.3 Example 61
11.3 Additional functions between channels 62
12 Triggers 64
12.1 Analog channel trigger 64
12.1.1 Single threshold 64
12.1.2 Multiple thresholds 65
12.1.3 Example 66
12.1.4 Parasites 66
12.1.5 Trigger on Logic Channels 66
13 Math Functions 68
13.1 Denitions 68
13.2 Types of calculations 68
13.2.1 Math Functions 69
14 Input - Output 71
14.1 Additional Input/Output Connectors 71
14.2 Logic inputs 71
14.2.1 Use 72
14.3 Alarm outputs 72
14.3.1 Use 72
14.4 Power supply output 72
14.5 External voltage box input/output interface 73
15 Ethernet Interface 75
15.1 General 75
15.1.1 Crossover cable 75
15.2 WIFI interface 75
15.3 FTP transfer 76
15.4 Managing with VNC 77
15.4.1 Changing the password 77
15.4.2 External software VNC Viewer 77
16 Programming 78
16.1 Message Format 78
16.2 Response Format 78
16.3 Message Syntax 78
16.3.1 Header 78
16.3.2 Header separator 78
16.3.3 One or several pieces of data 78
16.3.4 Data separator 79
16.3.5 Data 79
16.3.5.1 Alphanumeric data 79
16.3.5.2 Decimal digital data 79
16.3.5.3 Text 79
16.4 Message Response Format 79
16.4.1 Syntax 79
16.4.1.1 Single Header 79
16.4.2 Data 80
8
17 Specications 81
18 LIMITED TWO-YEAR WARRANTY 82
19 Service Information 83
9
Introduction
The DAS30, DAS50 and DAS60 programmable recorders measures and records voltage, current, temperature, etc. on up to 6 channels. The recorders also include digital input capabilities using the 12 logic input channels. An optional feature of the recorders is a port for measuring temperature with Pt100 or Pt1000 temperature sensors.
1.1 Recording Modes
Finally, the optional thermal printer adds the ability to print recorded signals (saved and in real-time).
1.1.1 Direct mode
Print measured signals to the thermal printer
Use Print readings and/or charts. Useful for recording long duration signals.
Enables Complex triggers of the drawing, action after the end of the drawing, simultaneous recording to
a le.
1
1.1.2 Memory mode
Quick acquisition into internal memory of the measured signals
Use quick short data acquisition (transient)
Enables complex triggers of the data acquisition, action after the end of the data acquisition, simulta-
neous record on le
1.1.3 Go No-Go mode
Quick acquisition into internal memory of the measured signals
Use quick short data acquisition (transient) for capture of non repetitive events
Enables complex triggers, particularly when outrunning a pre-recorded template from a channel, action
after the end of the data acquisition, simultaneous record on le
1.1.4 File mode
Quick acquisition into internal memory of the measured signals
Use quick protracted data acquisition (only restricted by the size of the saving means)
Enables complex triggers, action after the end of the data acquisition, acquisition of very big amounts
of data.
1.1.5 Power Analysis mode
single-phase network analysis three-phase analysis, Aaron mode, Star and Delta. (two-wattmeter method) . Data acquisition on internal memory.
1
Requires Printer option installed at time of order
10
Getting Started
You turn on the recorder by pressing the button at the top of the device. When the device is on, the ON/OFF button is lit blue. After launching the software, the recorder displays a homepage that species the hardware version, then switches auto­matically to the « Direct display » mode (oscilloscope).
You can display this homepage any time in the main menu by pressing . To turn o the device, you only have to press the ON/OFF button twice. If the device does not stop correctly, you must press the power button for 5 seconds and the recorder will shut o; in this case, the instrument conguration will not be saved.
2.1 Measurement Types
Voltage The voltage is measured between the red and black terminals of the inputs with wires and male
security pin plugs (according to CEI 1010).
Thermocouple Temp The voltage resulting from thermocouple eect must be measured between the red and black ter-
minals of the relevant input. To provide right measure values, connect the two ends of the ther­mocouple cord directly onto the quick fastening socket. Plug the two pin plugs onto the selected input with respect of the polarity. (Do not use welding male pin plugs: the welding would alter the thermocouple eect.)
Platinum RTD The Pt100 probe must be connected on the « I+ » and « I- » terminals.
The voltage from the Pt100 must be measured between the « + » and « -» terminals with one of the two following mountings: 2, 3 or 4 wires. The 3 or 4-wire mountings make the measure­ment independent from the resistance of the line.
Current You can make current measurements by shunt between the red and black terminals of the rel-
evant input. In this case, select the "current" type among the parameters of the relevant channel. Connect the measurement wires to the terminals of the shunt. The results are directly displayed in Amperes or milliamperes according to the caliber of the se­lected channel.
Logic Using the Logic input (25 pin connector) digital signals can included in the recording or used as
triggers. The logic connector also includes pins that can be used as output signals (alarms) for other equipment. Up to 16 signals are available for recording and triggering and 4 signals for alarm outputs.
2.2 Powering the Recorder
Each recorder includes an internal rechargeable Lithium-Ion battery. The recorder is powered and charged by a 15 V 4 A charger. The charger included with the recorder is powered by 100-240 VAC at 50-60 Hz. The recorder runs with or without the power supply connected and provides up to 8 hours of power when running on battery. When the power supply is connected a green LED lights beside the 25-pin logic input connector. To turn on the recorder, press the button labeled “ON/OFF” located on the top surface of the recorder. When on, the ring around the button lights up blue. Battery charge state is shown on the front screen. To turn o the recorder press the power button. A window will appear on-screen prompting for conrmation; press the green “Check Mark” to shut down the recorder. If the red “X” is pressed or if the green “check mark” is not pressed within 10 seconds, the recorder will not turn o.
2.3 Connections
11
DAS60 DAS50
DAS30
Figure 2.1 Recorder Connections
2.3.1 Analog Inputs
Isolated inputs have 2 shielded banana jack type terminals for each input:
Red Positive Input
Black Negative Input
Item Description
1 DC Power
2 PT100-PT1000 (optional)
3 Power Switch
4 USB Ports
5 Ethernet Port
6 Logic Signal Connector
7 Analog Inputs
8 Grounding Terminal
The inputs are capable of measuring500 VDC or 440 VAC.
For best results and to eliminate noise, twisting or shielding the leads is recommended. If the source of the signal you have to record has low internal impedance, use twisted wires. For high impedance signal sources, using shielded wires is recommended.
When gathering grounds from the various items on the measurement line, it is good to check that there is no voltage dierence between them, in order to prevent any shortcut. If any doubt, make a measurement with a voltmeter on a
low resistance (i.e. 1k) between the terminals.
2.4 Ground the Recorder
For safety, the unit should be connected to Earth ground. The recorder includes a terminal for this purpose located on the top surface of the DAS60 and on the right-hand side of the DAS30 and DAS50.
12
2.5 Thermal Printer
The DAS30, 50 and 60 recorders may be ordered with a thermal printer. The printer prints to 110mm wide thermal paper. Paper must always be loaded in the printer to avoid printing without paper and risking damage to the thermal print heads.
2.5.1 Loading Paper
Lay the device on the right sidePress the black button to open the door to the printerInstall the paper into the receptacle ("no de ref" must be on the side of the plugs) and insert the paper through the
slot in the door.
Close the lid securely
Note: If the printer outputs paper and the paper is blank, it is possible that the paper is loaded upside down.
Note: Only one face of the paper is heat-sensitive. Reversing the paper at loading will result in the absence of any
printing out.
2.5.2 Storing Printed Recordings
In order to maintain the quality of the paper recordings, the following is recommended:
Avoid using laminated sleevesStore away from light, in a dry and cool place.Store in a carboard sleeve
2.6 Power-up Conguration
At power-up, the recorder will be congured how it was when it was shut down. If it becomes inconsistent, the recorder may need to be repaired. Contact B&K Precision for service information. A default setup can be loaded from the “Setup Screen”.
2.7 Routine Maintenance
Maintenance simply consists of cleaning the outside the device. The recorder should be periodically cleaned.
Use water and soap to clean the metal surfacesnever use any product that contains petroleum products, benzine, or alcohol. These may damage screen printed
details.
Wipe all surfaces with a soft lint-free clothUse some antistatic screen cleaner for the touchscreen
13
2.8 Battery
The device is equipped with a lithium-ion battery (Li-ion) that should last at least 200 full charge cycles. It is shipped charged. Upon reciept or if the device has remained unused for more than one month, check its status and recharge if necessary. When the battery is nearly discharged, the device will close all open les and shut itself o. To extend the life of the battery:
Avoid complete dischargeDo not store the batteries too long without using themStore the battery charged to 40%Do not fully charge or discharge the battery before storage.
Note: Any service on the battery requires dismantling of the device and should be done by a B&K sta.
2.8.1 To recharge the battery
1. Connect the provided external power supply on the jack plug of the device
2. Connect AC power to the charger
When the internal charger starts loading the battery; the green light of battery charge will light. Charging may be done when it is on or o. When the battery is full, the green light turns o automatically. The recorder can run from AC power indenitely.
2.8.2 Safety precautions
Never heat up or expose the battery to reNever shunt parts of the battery: risk of explosion!Do not puncture the batteryDo not dismantle the battery packDo not reverse the polarities of the batteryThis battery pack includes a protective item that shall not be damaged or removedDo not store the pack in a place exposed to excessive heatDo not store the device in a vehicle in direct sunlight
Warning: When the charger is connected to the device, the metal frame is grounded.
2.9 Oset calibration
Calibration of the recorder inputs for voltage and thermocouple osets can help remove any osets present on each channel. All enabled analog channels are calibrated when the oset calibration is performed. Disable any channels that should not be calibrated.
14
1. Let the device work for 20 minutes (ambient temperature 68 - 77 F)
2. Enable all desired channels in the channel conguration menu
3. Short the terminals of the active channels
4. Go to the setup menu
1. Press
2. Press
5. From setup, go to calibration
1. Press
2. Press
6. Press . The recorder will perform the oset adjustment and save the values when complete.
2.10 Default calibration
To return the instrument to the factory calibration values, use the “Recover factory calibration” function.
1. Go to the setup menu
1. Press
2. Press
2. From setup, go to calibration
1. Press
2. Press
3. Press “Recover factory calibration” and conrm the operation by pressing the green check mark.
2.11 Screen brightness
Screen brightness is set in the setup screen under “Additional Options”.
1. Go to the setup menu
1. Press
2. Press
2. From setup, press
3. Press
15
4. Use the slider to adjust screen brightness.
5. Press the “Return” button to accept the change.
2.11.1 Locking of the recorder
The touchscreen of the recorder can be disabled by locking the recorder. Press and then the green check mark to lock the recorder. The recorder will enter lock mode in 10 seconds. To unlock the recorder, press 5 times. This button is located at the top right of the screen.
2.12 User mode
User mode disables most conguration settings leaving other menus available.
The password is: Sefram.
2.13 Update of the internal software
Software updates to the recorders may be available. These updates are available on the Sefram website in the “Software Updates” section. Only update the recorder when connected to power and/or with a charged battery.
http://www.sefram.com/en/software-updates.html
1. Copy the le to USB stick
2. Plug the USB stick into one of the USB connectors
3. Press
4. Press
5. Press
6. Select the update le
7. Initiate the update and wait till it is complete. The internal software automatically copies the les required for the new version.
8. Restart the recorder
16
Home Screen
The home screen provides access to all of the settings and functions of the recorder. It is accessed by pressing .
Item Description Item Description
1 Direct Mode 11 Logic Setup
2 Memory Mode 12 X-Y Plotting
3 File Mode 13 Start Recording
4 Go/No Go Mode 14 Chart Mode
5 Power Analysis Mode” 15 Display Stored Recording
6 Startup Screen 16 Numeric Display
7 Setup Screen 17 Logic Channel Status
8 Channel Setup 18 Analog Channel Status
9 Triggers” 19 Screen Capture Button
10 Channel On/O Menu
Figure 3.1 Home Screen
Pressing the buttons on screen activates the dierent modes and menus. Some areas, like the Analog Channel Status on the left side of the screen, also have areas within them to open conguration screens. For example when not in Power Analysis mode, the analog channels are shown. Pressing in one of the channels opens a conguration screen for that channel like Figure 3.2.
17
Figure 3.2 Single Analog Channel Conguration Screen
18
Setup Screen
General conguration of the device, monitoring of the alarm outputs, network address TCP/IP, calibration of the channels, update of the internal software.
Figure 4.1 Setup Screen
1. Language:
2. Screen shuto time setting
3. System time settings
4. Graphic Options: The display orientation of the data in the bargraph (Item 15)
Right Max value to the right
Left Max value to the left
19
5. Additional option: More settings
6. Software update: update of the internal software (see chapter Presentation)
7. Default setup: Set all settings to default values
8. Load on disk: load a stored conguration from memory (internal or USB drive)
9. Save on disk: save the current conguration to memory (internal or USB drive)
10. Export to a text le
11. Network: Setup and status of network communication.
12. Alarm: Setup and status of the alarm outputs A, B, C or D.
13. Alarm setup keys
14. Logic channel status and states
15. Bargraph showing live measurements
4.1 Logic channels
The 16 logic channels can be recorded simultaneously with the analog channels. To set the parameters of the logic channels, you can press the « logic channels » menu on the main window or directly press the logic channels part at the bottom of each conguration window:
You access the conguration page of the logic channels:
Figure 4.2
20
Figure 4.3
1. Selection of the name of each channel on screen
2. Selection of the color of each channel on screen
3. Logic validity: validation of the data acquisition and plotting of the logic channels
4. Validation of the channels
You can validate or not the channels to record or print out in the various conguration menus of the recorder. In order to validate or not the record or the printing of a channel, you must check (to validate) or uncheck (not to validate) the checkbox right to the bargraph of the channel.
4. « Direct display » key
4.2 Real time on-screen display of the measurements
1. Screen conguration: conguration of the display of measurements on screen. Selection of the mode of graphic display:
F(t)XYDigital measures (full screen)Full screen to display the reticule on the whole screen (mode f(t))Colors to customize the display (background, reticule, cursor)Graphic optionCustom reticule
2. Channels validation: Validation of the channels to display
If the logic channels are actives, you shall select whether the diagram of the logic channels should be distinct from the analog channels or not (see chapter 4.3.3). Freeze screen: freeze the measurements on screen to make measurements by cursors, make calculations, save or print the measures on screen (1000 dots) in F(t) mode.
21
You get access to:
Figure 4.4
Figure 4.5
22
Relaunch – relaunches the sweep
Time Cursors displays the vertical cursors (2) to make measurements on the screen; move the cursor by select-
ing it with your ngers or the mouse.
Voltages Cursors displays the horizontal cursors (2) to make amplitude measurements on the screen; pro-
ceed like for the time cursors to move them. You can also change the caliber / zero to expand and move your measurement on screen.
1. Autoset: This function allows you to have automatically the caliber of the active channels and the time base compatible with the frequency of the signal.
2. Time base: Allows you to settle the time base.
3. Math: Makes it possible to display values of mathematical calculations on screen.
4. Access to the channel conguration page:
Figure 4.6
1. Screen setup: conguration of display of measurements on screen Selection of the graphic display mode:
F(t)XYDigital measures (full screen)Full screen to display the reticule on the whole screen (mode f(t)) Colors to customize the display (background,
reticule, cursor).
Graphic optionCustom reticle
2. Channels validation: Allows you to activate or disactivate the channels.
3. Math: Displays values of mathematical calculations on screen.
4. Selection of the channel to modify (with the « + », « - » and central keys).
5. Modication of the range of the selected channel (with the « + », « - » and central keys).
6. Modication of the position of zero (with the « + », « - » and central keys).
23
7. Modication of the position of the selected channel (with the « + », « - » and central keys). 8- Time base: Allows you to adjust the time base.
8. Back to the previous page.
4.3 XY display
The XY display mode makes it possible to display the validated channels in real time one against the other. One of the channels denes the horizontal axis; the other channels provide dots on the vertical axis.
You can access the XY mode via the main menu by pressing
Figure 4.7
1. Screen setup: conguration of the display of measurements on screen. Selection of the mode of graphic display:
F(t)XYDigital measures (full screen)Full screen to display the reticule on the whole screen (mode f(t))Colors to customize the display (background, reticule, cursor)Graphic optionCustom reticule
2. Plot: starts or stops of the plot
3. Delete: erases the screen
4. X channel: selection of the channel on the horizontal axis (sweep) 5- Y channel: the Y channels are the valid channels
4.4 Digital display
The Digital display mode displays the digital values of all channels in real time on screen. When pressing one of the channels, you access directly to the parameter page of the corresponding channel.
24
Figure 4.8
4.5 Trigger menu
Programming the start and stop conditions for the paper print in Direct mode, and the data acquisition conditions for the channels in Memory, File and Template modes. Selection of the actions after data acquisition or plot and validation of the save in real time.
Figure 4.9
The program of triggers is dierent according to the pending mode (Direct, Memory, File or Template). See the chapter of the pending mode for more details.
4.6 Memory output key
Display on screen of the data acquisitions available from the internal memory or from les of the internal memory or an USB stick.
25
Figure 4.10
This function includes the same commands as the « Direct display » function. The « Read block or le » command allows you to select the memory block (zone inside the internal memory) or the le to display. When pressing this key, the following window pops up:
Figure 4.11
Block number: number of the memory block to displayRead le: selection of the le to displayWrite data to le: save the current display into a le
When the size of the data acquisition to display is big, it may take time to recover and display the dots.
The display is made in 2 phases:
A quick phase to display the frame of the data acquisition: some dots may not appearA slower phase where all data acquisition dots are displayed: the advancement percentage is displayed at the bottom
of the screen.
4.7 Record key
This key has various eects according to the current mode of the device.
Direct mode: launches the paper printing if the trigger is in manual Start; else, the device waits for the
start condition.
26
+ 58 hidden pages