The following safety precautions apply to both operating and maintenance personnel and must be followed during all
phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
Read and understand the safety and operational information in this manual.
Apply all the listed safety precautions.
Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating the instrument
at an incorrect line voltage will void the warranty.
Make all connections to the instrument before applying power.
Do not operate the instrument in ways not specied by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design,
manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply
with these requirements.
Category rating
The IEC 61010 standard denes safety category ratings that specify the amount of electrical energy available and the
voltage impulses that may occur on electrical conductors associated with these category ratings. The category rating is
a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of the circuit to be tested,
which denes the voltage impulses expected and required insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected to the
mains supply. The voltages in the environment are typically derived from a limited-energy transformer or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the mains
supply at a standard wall outlet or similar sources. Example measurement environments are portable
tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be connected to the mains
installation of a building. Examples are measurements inside a building’s circuit breaker panel
or the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to the primary
power entering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what is specied in this manual
for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the
instrument’s category rating to maintain the instrument’s category rating. Failure to do so will lower the category rating
of the measuring system.
2
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be
115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your
country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This
instrument is grounded through the ground conductor of the supplied, three-conductor AC line power cable. The power
cable must be plugged into an approved three-conductor electrical outlet. The power jack and mating plug of the power
cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive parts
(including control knobs) may provide an electric shock. Failure to use a properly-grounded approved outlet and the
recommended three-conductor AC line power cable may result in injury or death.
Unless otherwise stated, a ground connection on the instrument’s front or rear panel is for a reference of potential only
and is not to be used as a safety ground. Do not operate in an explosive or ammable atmosphere.
Do not operate the instrument in the presence of ammable gases or vapors, fumes, or nely-divided particulates.
The instrument is designed to be used in oce-type indoor environments. Do not operate the instrument
In the presence of noxious, corrosive, or ammable fumes, gases, vapors, chemicals, or nely-divided particulates.
In relative humidity conditions outside the instrument’s specications.
In environments where there is a danger of any liquid being spilled on the instrument or where any liquid can condense
on the instrument.
In air temperatures exceeding the specied operating temperatures.
In atmospheric pressures outside the specied altitude limits or where the surrounding gas is not air.
In environments with restricted cooling air ow, even if the air temperatures are within specications.
In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature range is
0∘C to 40∘C and 20% to 80% relative humidity, with no condensation allowed. Measurements made by this instrument
may be outside specications if the instrument is used in non-oce-type environments. Such environments may include
rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise,
strong electric elds, or strong magnetic elds.
3
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the
instrument, remove the instrument’s power cord, remove the instrument from service, label it as not to be operated,
and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the
instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases,
or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual. Not
for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a life-support
device or system.
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must
be made by qualied service-trained maintenance personnel who are aware of the hazards involved when the instrument’s
covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may
exist when the covers are removed. To avoid injuries, always disconnect the power cord from the instrument, disconnect
all other connections (for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there
are no hazardous voltages present on any conductors by measurements with a properly-operating voltage-sensing device
before touching any internal parts. Verify the voltage-sensing device is working properly before and after making the
measurements by testing with known-operating voltage sources and test for both DC and AC voltages. Do not attempt
any service or adjustment unless another person capable of rendering rst aid and resuscitation is present.
Do not insert any object into an instrument’s ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit
exists.
Fuse replacement must be done by qualied service-trained maintenance personnel who are aware of the instrument’s fuse
requirements and safe replacement procedures. Disconnect the instrument from the power line before replacing fuses.
Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings specied in this manual or on
the back of the instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a re. Failure
to use the specied fuses will void the warranty.
Servicing
4
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to
B&K Precision for service and repair to ensure that safety and performance features are maintained.
For continued safe use of the instrument
Do not place heavy objects on the instrument.
Do not obstruct cooling air ow to the instrument.
Do not place a hot soldering iron on the instrument.
Do not pull the instrument with the power cord, connected probe, or connected test lead.
Do not move the instrument when a probe is connected to a circuit being tested.
Safety Symbols
SymbolDescription
indicates a hazardous situation which, if not avoided, will result in death or serious injury.
indicates a hazardous situation which, if not avoided, could result in death or serious injury
indicates a hazardous situation which, if not avoided, will result in minor or moderate injury
Refer to the text near the symbol.
Electric Shock hazard
Alternating current (AC)
Chassis ground
Earth ground
This is the In position of the power switch when instrument is ON.
This is the Out position of the power switch when instrument is OFF.
is used to address practices not related to physical injury.
5
Contents
1Introduction10
1.1Recording Modes10
1.1.1Direct mode10
1.1.2Memory mode10
1.1.3Go No-Go mode10
1.1.4File mode10
1.1.5Power Analysis mode10
2Getting Started11
2.1Measurement Types11
2.2Powering the Recorder11
2.3Connections11
2.3.1Analog Inputs12
2.4Ground the Recorder12
2.5Thermal Printer13
2.5.1Loading Paper13
2.5.2Storing Printed Recordings13
2.6Power-up Conguration13
2.7Routine Maintenance13
2.8Battery14
2.8.1To recharge the battery14
2.8.2Safety precautions14
2.9Oset calibration14
2.10Default calibration15
2.11Screen brightness15
2.11.1Locking of the recorder16
2.12User mode16
2.13Update of the internal software16
3Home Screen17
4Setup Screen19
4.1Logic channels20
4.2Real time on-screen display of the measurements21
4.3XY display24
4.4Digital display24
4.5Trigger menu25
4.6Memory output key25
4.7Record key26
4.8Screen copy27
5FILE MANAGEMENT28
5.1Save setup les29
5.2Load setup les29
5.3Save data acquisitions29
5.4Recovery of the data acquisition les30
5.5Transfer via FTP30
5.6Display under SeframViewer31
5.7Managing with SeframPilot31
6
6Direct Mode33
6.1Launching the printing33
6.2Conguration of print34
6.2.1Relaunching the printing35
6.2.2Writing information35
7Memory Mode36
7.1Setup and start of the data acquisition36
7.2Sampling period37
7.3Internal memory, blocks37
7.4Trigger position38
7.5Double Trigger mode38
7.6Recording38
7.7Memory output40
8File Mode42
8.1Setup and triggering of the data acquisition42
8.1Restrictions43
9Go/No Go Mode44
9.1Setup and launching of the data acquisition44
9.2Blocks: parting of the internal memory in blocks44
9.3Real time save: simultaneous record of the data acquisition into a le44
9.4Start: start condition of the data acquisition45
9.5Automatic: automatic stop when the block is full45
9.5.1Creation of the template45
9.6Use of the template46
10 Power Analysis Mode47
10.1Power Analysis47
10.2Network setup48
10.3Wiring and check49
10.4Fresnel diagram50
10.5Warning messages50
10.6Recording additional channels50
10.7Real time display of the signals50
10.8Power analysis setup52
10.9Parameter acquisition le52
10.10 Record launch52
10.10.1 Manual launch52
10.10.2 Automatic trigger53
10.10.3 Parametric trigger53
10.11 You can select a stop condition (recording time or manual stop)53
10.11.1 Trigger according to an inrush53
10.11.2 In version 1.1 of the rmware or later54
10.11.2.1Trigger according to a duration or a date54
10.12 Recordable parameters54
10.12.1 Voltage and Current54
10.12.2 Power54
10.12.3 Power Factor54
10.13 Record55
10.14 Measurement process56
7
11 Channel Setup57
11.1Analog channels58
11.2Type of measurement V/ t59
11.2.1Example60
11.2.2Type of measurement
60
11.2.3Example61
11.3Additional functions between channels62
12 Triggers64
12.1Analog channel trigger64
12.1.1Single threshold64
12.1.2Multiple thresholds65
12.1.3Example66
12.1.4Parasites66
12.1.5Trigger on Logic Channels66
13 Math Functions68
13.1Denitions68
13.2Types of calculations68
13.2.1Math Functions69
14 Input - Output71
14.1Additional Input/Output Connectors71
14.2Logic inputs71
14.2.1Use72
14.3Alarm outputs72
14.3.1Use72
14.4Power supply output72
14.5External voltage box input/output interface73
15 Ethernet Interface75
15.1General75
15.1.1Crossover cable75
15.2WIFI interface75
15.3FTP transfer76
15.4Managing with VNC77
15.4.1Changing the password77
15.4.2External software VNC Viewer77
16 Programming78
16.1Message Format78
16.2Response Format78
16.3Message Syntax78
16.3.1Header78
16.3.2Header separator78
16.3.3One or several pieces of data78
16.3.4Data separator79
16.3.5Data79
16.3.5.1Alphanumeric data79
16.3.5.2Decimal digital data79
16.3.5.3Text79
16.4Message Response Format79
16.4.1Syntax79
16.4.1.1Single Header79
16.4.2Data80
8
17 Specications81
18 LIMITED TWO-YEAR WARRANTY82
19 Service Information83
9
Introduction
The DAS30, DAS50 and DAS60 programmable recorders measures and records voltage, current, temperature, etc. on up
to 6 channels. The recorders also include digital input capabilities using the 12 logic input channels. An optional feature
of the recorders is a port for measuring temperature with Pt100 or Pt1000 temperature sensors.
1.1 Recording Modes
Finally, the optional thermal printer adds the ability to print recorded signals (saved and in real-time).
1.1.1 Direct mode
Print measured signals to the thermal printer
Use Print readings and/or charts. Useful for recording long duration signals.
Enables Complex triggers of the drawing, action after the end of the drawing, simultaneous recording to
a le.
1
1.1.2 Memory mode
Quick acquisition into internal memory of the measured signals
Use quick short data acquisition (transient)
Enables complex triggers of the data acquisition, action after the end of the data acquisition, simulta-
neous record on le
1.1.3 Go No-Go mode
Quick acquisition into internal memory of the measured signals
Use quick short data acquisition (transient) for capture of non repetitive events
Enables complex triggers, particularly when outrunning a pre-recorded template from a channel, action
after the end of the data acquisition, simultaneous record on le
1.1.4 File mode
Quick acquisition into internal memory of the measured signals
Use quick protracted data acquisition (only restricted by the size of the saving means)
Enables complex triggers, action after the end of the data acquisition, acquisition of very big amounts
of data.
1.1.5 Power Analysis mode
single-phase network analysis
three-phase analysis, Aaron mode, Star and Delta. (two-wattmeter method) .
Data acquisition on internal memory.
1
Requires Printer option installed at time of order
10
Getting Started
You turn on the recorder by pressing the button at the top of the device.
When the device is on, the ON/OFF button is lit blue.
After launching the software, the recorder displays a homepage that species the hardware version, then switches automatically to the « Direct display » mode (oscilloscope).
You can display this homepage any time in the main menu by pressing.
To turn o the device, you only have to press the ON/OFF button twice.
If the device does not stop correctly, you must press the power button for 5 seconds and the recorder will shut o; in this
case, the instrument conguration will not be saved.
2.1 Measurement Types
Voltage The voltage is measured between the red and black terminals of the inputs with wires and male
security pin plugs (according to CEI 1010).
Thermocouple Temp The voltage resulting from thermocouple eect must be measured between the red and black ter-
minals of the relevant input. To provide right measure values, connect the two ends of the thermocouple cord directly onto the quick fastening socket. Plug the two pin plugs onto the selected
input with respect of the polarity. (Do not use welding male pin plugs: the welding would alter
the thermocouple eect.)
Platinum RTD The Pt100 probe must be connected on the « I+ » and « I- » terminals.
The voltage from the Pt100 must be measured between the « + » and « -» terminals with one
of the two following mountings: 2, 3 or 4 wires. The 3 or 4-wire mountings make the measurement independent from the resistance of the line.
Current You can make current measurements by shunt between the red and black terminals of the rel-
evant input.
In this case, select the "current" type among the parameters of the relevant channel. Connect
the measurement wires to the terminals of the shunt.
The results are directly displayed in Amperes or milliamperes according to the caliber of the selected channel.
Logic Using the Logic input (25 pin connector) digital signals can included in the recording or used as
triggers. The logic connector also includes pins that can be used as output signals (alarms) for
other equipment. Up to 16 signals are available for recording and triggering and 4 signals for alarm
outputs.
2.2 Powering the Recorder
Each recorder includes an internal rechargeable Lithium-Ion battery. The recorder is powered and charged by a 15 V 4
A charger. The charger included with the recorder is powered by 100-240 VAC at 50-60 Hz. The recorder runs with or
without the power supply connected and provides up to 8 hours of power when running on battery. When the power
supply is connected a green LED lights beside the 25-pin logic input connector.
To turn on the recorder, press the button labeled “ON/OFF” located on the top surface of the recorder. When on, the
ring around the button lights up blue. Battery charge state is shown on the front screen.
To turn o the recorder press the power button. A window will appear on-screen prompting for conrmation; press the
green “Check Mark” to shut down the recorder. If the red “X” is pressed or if the green “check mark” is not pressed
within 10 seconds, the recorder will not turn o.
2.3 Connections
11
DAS60DAS50
DAS30
Figure 2.1 Recorder Connections
2.3.1 Analog Inputs
Isolated inputs have 2 shielded banana jack type terminals for each input:
Red Positive Input
Black Negative Input
Item Description
1DC Power
2PT100-PT1000 (optional)
3Power Switch
4USB Ports
5Ethernet Port
6Logic Signal Connector
7Analog Inputs
8Grounding Terminal
The inputs are capable of measuring500 VDC or 440 VAC.
For best results and to eliminate noise, twisting or shielding the leads is recommended. If the source of the signal you
have to record has low internal impedance, use twisted wires. For high impedance signal sources, using shielded wires
is recommended.
When gathering grounds from the various items on the measurement line, it is good to check that there is no voltage
dierence between them, in order to prevent any shortcut. If any doubt, make a measurement with a voltmeter on a
low resistance (i.e. 1k) between the terminals.
2.4 Ground the Recorder
For safety, the unit should be connected to Earth ground. The recorder includes a terminal for this purpose located on
the top surface of the DAS60 and on the right-hand side of the DAS30 and DAS50.
12
2.5 Thermal Printer
The DAS30, 50 and 60 recorders may be ordered with a thermal printer. The printer prints to 110mm wide thermal
paper. Paper must always be loaded in the printer to avoid printing without paper and risking damage to the thermal
print heads.
2.5.1 Loading Paper
Lay the device on the right side
Press the black button to open the door to the printer
Install the paper into the receptacle ("no de ref" must be on the side of the plugs) and insert the paper through the
slot in the door.
Close the lid securely
Note: If the printer outputs paper and the paper is blank, it is possible that the paper is loaded upside down.
Note: Only one face of the paper is heat-sensitive. Reversing the paper at loading will result in the absence of any
printing out.
2.5.2 Storing Printed Recordings
In order to maintain the quality of the paper recordings, the following is recommended:
Avoid using laminated sleeves
Store away from light, in a dry and cool place.
Store in a carboard sleeve
2.6 Power-up Conguration
At power-up, the recorder will be congured how it was when it was shut down. If it becomes inconsistent, the recorder
may need to be repaired. Contact B&K Precision for service information.
A default setup can be loaded from the “Setup Screen”.
2.7 Routine Maintenance
Maintenance simply consists of cleaning the outside the device. The recorder should be periodically cleaned.
Use water and soap to clean the metal surfaces
never use any product that contains petroleum products, benzine, or alcohol. These may damage screen printed
details.
Wipe all surfaces with a soft lint-free cloth
Use some antistatic screen cleaner for the touchscreen
13
2.8 Battery
The device is equipped with a lithium-ion battery (Li-ion) that should last at least 200 full charge cycles. It is shipped
charged. Upon reciept or if the device has remained unused for more than one month, check its status and recharge if
necessary. When the battery is nearly discharged, the device will close all open les and shut itself o.
To extend the life of the battery:
Avoid complete discharge
Do not store the batteries too long without using them
Store the battery charged to 40%
Do not fully charge or discharge the battery before storage.
Note: Any service on the battery requires dismantling of the device and should be done by a B&K sta.
2.8.1 To recharge the battery
1. Connect the provided external power supply on the jack plug of the device
2. Connect AC power to the charger
When the internal charger starts loading the battery; the green light of battery charge will light. Charging may be done
when it is on or o. When the battery is full, the green light turns o automatically. The recorder can run from AC
power indenitely.
2.8.2 Safety precautions
Never heat up or expose the battery to re
Never shunt parts of the battery: risk of explosion!
Do not puncture the battery
Do not dismantle the battery pack
Do not reverse the polarities of the battery
This battery pack includes a protective item that shall not be damaged or removed
Do not store the pack in a place exposed to excessive heat
Do not store the device in a vehicle in direct sunlight
Warning: When the charger is connected to the device, the metal frame is grounded.
2.9 Oset calibration
Calibration of the recorder inputs for voltage and thermocouple osets can help remove any osets present on each
channel. All enabled analog channels are calibrated when the oset calibration is performed. Disable any channels that
should not be calibrated.
14
1. Let the device work for 20 minutes (ambient temperature 68 - 77 F)
2. Enable all desired channels in the channel conguration menu
3. Short the terminals of the active channels
4. Go to the setup menu
1. Press
2. Press
5. From setup, go to calibration
1. Press
2. Press
6. Press. The recorder will perform the oset adjustment and save the values when complete.
2.10 Default calibration
To return the instrument to the factory calibration values, use the “Recover factory calibration” function.
1. Go to the setup menu
1. Press
2. Press
2. From setup, go to calibration
1. Press
2. Press
3. Press “Recover factory calibration” and conrm the operation by pressing the green check mark.
2.11 Screen brightness
Screen brightness is set in the setup screen under “Additional Options”.
1. Go to the setup menu
1. Press
2. Press
2. From setup, press
3. Press
15
4. Use the slider to adjust screen brightness.
5. Press the “Return” button to accept the change.
2.11.1 Locking of the recorder
The touchscreen of the recorder can be disabled by locking the recorder. Pressand then the green
check mark to lock the recorder. The recorder will enter lock mode in 10 seconds.
To unlock the recorder, press5 times. This button is located at the top right of the screen.
2.12 User mode
User mode disables most conguration settings leaving other menus available.
The password is: Sefram.
2.13 Update of the internal software
Software updates to the recorders may be available. These updates are available on the Sefram website in the “Software
Updates” section. Only update the recorder when connected to power and/or with a charged battery.
http://www.sefram.com/en/software-updates.html
1. Copy the le to USB stick
2. Plug the USB stick into one of the USB connectors
3. Press
4. Press
5. Press
6. Select the update le
7. Initiate the update and wait till it is complete. The internal software automatically copies the les required for the
new version.
8. Restart the recorder
16
Home Screen
The home screen provides access to all of the settings and functions of the recorder. It is accessed by pressing.
Item DescriptionItem Description
1“Direct Mode”11Logic Setup
2“Memory Mode”12X-Y Plotting
3“File Mode”13Start Recording
4“Go/No Go Mode”14Chart Mode
5“Power Analysis Mode” 15Display Stored Recording
6Startup Screen16Numeric Display
7“Setup Screen”17Logic Channel Status
8“Channel Setup”18Analog Channel Status
9“Triggers”19Screen Capture Button
10Channel On/O Menu
Figure 3.1 Home Screen
Pressing the buttons on screen activates the dierent modes and menus. Some areas, like the Analog Channel Status on
the left side of the screen, also have areas within them to open conguration screens. For example when not in Power
Analysis mode, the analog channels are shown. Pressing in one of the channels opens a conguration screen for that
channel like Figure 3.2.
17
Figure 3.2 Single Analog Channel Conguration Screen
18
Setup Screen
General conguration of the device, monitoring of the alarm outputs, network address TCP/IP, calibration of the channels,
update of the internal software.
Figure 4.1 Setup Screen
1. Language:
2. Screen shuto time setting
3. System time settings
4. Graphic Options: The display orientation of the data in the bargraph (Item 15)
Right Max value to the right
Left Max value to the left
19
5. Additional option: More settings
6. Software update: update of the internal software (see chapter Presentation)
7. Default setup: Set all settings to default values
8. Load on disk: load a stored conguration from memory (internal or USB drive)
9. Save on disk: save the current conguration to memory (internal or USB drive)
10. Export to a text le
11. Network: Setup and status of network communication.
12. Alarm: Setup and status of the alarm outputs A, B, C or D.
13. Alarm setup keys
14. Logic channel status and states
15. Bargraph showing live measurements
4.1 Logic channels
The 16 logic channels can be recorded simultaneously with the analog channels.
To set the parameters of the logic channels, you can press the « logic channels » menu on the main window or directly
press the logic channels part at the bottom of each conguration window:
You access the conguration page of the logic channels:
Figure 4.2
20
Figure 4.3
1. Selection of the name of each channel on screen
2. Selection of the color of each channel on screen
3. Logic validity: validation of the data acquisition and plotting of the logic channels
4. Validation of the channels
You can validate or not the channels to record or print out in the various conguration menus of the recorder.
In order to validate or not the record or the printing of a channel, you must check (to validate) or uncheck (not to
validate) the checkbox right to the bargraph of the channel.
4. « Direct display » key
4.2 Real time on-screen display of the measurements
1. Screen conguration: conguration of the display of measurements on screen. Selection of the mode of graphic
display:
F(t)
XY
Digital measures (full screen)
Full screen to display the reticule on the whole screen (mode f(t))
Colors to customize the display (background, reticule, cursor)
Graphic option
Custom reticule
2. Channels validation: Validation of the channels to display
If the logic channels are actives, you shall select whether the diagram of the logic channels should be distinct from the
analog channels or not (see chapter 4.3.3).
Freeze screen: freeze the measurements on screen to make measurements by cursors, make calculations, save or print
the measures on screen (1000 dots) in F(t) mode.
21
You get access to:
Figure 4.4
Figure 4.5
22
Relaunch – relaunches the sweep
Time Cursors displays the vertical cursors (2) to make measurements on the screen; move the cursor by select-
ing it with your ngers or the mouse.
Voltages Cursors displays the horizontal cursors (2) to make amplitude measurements on the screen; pro-
ceed like for the time cursors to move them. You can also change the caliber / zero to expand
and move your measurement on screen.
1. Autoset: This function allows you to have automatically the caliber of the active channels and the time base compatible
with the frequency of the signal.
2. Time base: Allows you to settle the time base.
3. Math: Makes it possible to display values of mathematical calculations on screen.
4. Access to the channel conguration page:
Figure 4.6
1. Screen setup: conguration of display of measurements on screen Selection of the graphic display mode:
F(t)
XY
Digital measures (full screen)
Full screen to display the reticule on the whole screen (mode f(t)) Colors to customize the display (background,
reticule, cursor).
Graphic option
Custom reticle
2. Channels validation: Allows you to activate or disactivate the channels.
3. Math: Displays values of mathematical calculations on screen.
4. Selection of the channel to modify (with the « + », « - » and central keys).
5. Modication of the range of the selected channel (with the « + », « - » and central keys).
6. Modication of the position of zero (with the « + », « - » and central keys).
23
7. Modication of the position of the selected channel (with the « + », « - » and central keys). 8- Time base: Allows
you to adjust the time base.
8. Back to the previous page.
4.3 XY display
The XY display mode makes it possible to display the validated channels in real time one against the other.
One of the channels denes the horizontal axis; the other channels provide dots on the vertical axis.
You can access the XY mode via the main menu by pressing
Figure 4.7
1. Screen setup: conguration of the display of measurements on screen. Selection of the mode of graphic display:
F(t)
XY
Digital measures (full screen)
Full screen to display the reticule on the whole screen (mode f(t))
Colors to customize the display (background, reticule, cursor)
Graphic option
Custom reticule
2. Plot: starts or stops of the plot
3. Delete: erases the screen
4. X channel: selection of the channel on the horizontal axis (sweep) 5- Y channel: the Y channels are the valid channels
4.4 Digital display
The Digital display mode displays the digital values of all channels in real time on screen. When pressing one of the
channels, you access directly to the parameter page of the corresponding channel.
24
Figure 4.8
4.5 Trigger menu
Programming the start and stop conditions for the paper print in Direct mode, and the data acquisition conditions for
the channels in Memory, File and Template modes.
Selection of the actions after data acquisition or plot and validation of the save in real time.
Figure 4.9
The program of triggers is dierent according to the pending mode (Direct, Memory, File or Template).
See the chapter of the pending mode for more details.
4.6 Memory output key
Display on screen of the data acquisitions available from the internal memory or from les of the internal memory or an
USB stick.
25
Figure 4.10
This function includes the same commands as the « Direct display » function.
The « Read block or le » command allows you to select the memory block (zone inside the internal memory) or the
le to display.
When pressing this key, the following window pops up:
Figure 4.11
Block number: number of the memory block to display
Read le: selection of the le to display
Write data to le: save the current display into a le
When the size of the data acquisition to display is big, it may take time to recover and display the dots.
The display is made in 2 phases:
A quick phase to display the frame of the data acquisition: some dots may not appear
A slower phase where all data acquisition dots are displayed: the advancement percentage is displayed at the bottom
of the screen.
4.7 Record key
This key has various eects according to the current mode of the device.
Direct mode: launches the paper printing if the trigger is in manual Start; else, the device waits for the
start condition.
26
Loading...
+ 58 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.