B&K Precision CAL73 User Manual

CONTENTS
Title Page
1. Safety Information………………………….. 1
Environment Conditions…………………………………. 1 Maintenance & Clearing…………………………………. 1 Safety symbols…………………………………………… 1
2. Description ……………….….……………… 2
3. Features……………… ……..…..…………….2
4. Specifications………..……..……………….. 2
Output sound pressure levels…………………………... 2 Output frequency…………………………………………. 2 Reference conditions…………………………………….. 2 Influence of ambient conditions………………………… 2 Total Harmonic Distortion (THD)………………………… 2 Accuracy of sound pressure level……………………… 2 Power……………………………………………………… 2 Battery life………………………………………………….2 Battery test ……………………………………………….. 3 Dimensions……………………………………………….. 3 Weight……………………………………………………... 3 Ambient conditions………………………………………. 3 Storage temperature & humidity………………………... 3 Accessories……………………………………………….. 3
5. Nomenclature And Function.……………... 4
6. Operating Preparation…….….……………. 5
7. Calibration Procedure….…….…………….. 5
1. Safety Information
Read the following safety information carefully before attempting to operate or service the meter. Use the meter only as specified in this manual; otherwise, the protection provided by the meter may be impaired.
Environment Conditions
§ Altitude up to 2000 meters
§ Relative humidity 90% max.
§ Operation Ambient 0° to 40°C
Maintenance & Clearing
Repairs or servicing not covered in this manual should only be performed by qualified personnel. Periodically wipe the case with a dry cloth. Do not use abrasives or solvents on this instrument.
Safety symbols
Comply with EMC When servicing, use only specified replacement parts.
2. Description
Sound level calibrator is used to calibrate sound level meters and other sound measurement equipment. You can calibrate 1 inch diameter microphones directly and 1/2 inch microphones using 1/2 inch adaptor supplied with the calibrator. With available adaptors, you can calibrate other microphones and instruments.
1
3. Features
Conforms to ANSI S1.4-1984 and IEC 60942-2003 Class 2. Calibration levels of 94dB and 114dB. Fits 1 inch and 1/2 inch diameter microphone.
4. Specifications
Output sound pressure levels:
94dB and 114dB re 20 uPa under reference conditions.
Output frequency: 1000Hz +/- 2% Reference conditions:
Temperature :23°C (74°F) Relative humidity: 50% Atmospheric pres sure: 1013hpa
Influence of ambient conditions:
Temperature coefficient: 0.005dB/? Humidity coefficient: 0.005dB/%RH
Total Harmonic Distortion (THD): <3% Accuracy of sound pressure level: +/- 0.5dB Power: one 9V battery 006P or IEC 6F22 or
NEDA 1604.
Battery life: approx.40 hours (Alkaline Battery) Battery test:
Internal circuitry checks condition of battery continuously.
2