99 Washington Street
Melrose, MA 02176
Fax 781-665-0780
TestEquipmentDepot.com
Safety information
Please review the following safety precautions before operating our equipment.
General information
The following safety precautions should be observed before using this product and any associated
instrumentations.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar
with the safety precautions required to avoid possible injury. Read and follow all installation, operation,
and maintenance information carefully before using the product. Refer to this manual for complete
product specifications.
If the product is used in a manner not specified, the protection provided by the product may be
impaired.
Before performing any maintenance, disconnect the line cord and all test cables.
Protection from electric shock
Operators of this instrument must be protected from electric shock at all times. The responsible body
must ensure that operators are prevented access and/or insulated from every connection point. In
some cases, connections must be exposed to potential human contact.
Under these circumstances personnel must be trained to protect themselves from the risk of electric
shock. If the circuit is capable of operating at or above 1000 volts, no conductive part of the circuit may
be exposed.
Definition of users
Responsible body is the individual or group responsible for the use and maintenance of equipment is
operated within its specifications and operating limits, and for ensuring that operators are adequately
trained.
This product should only be used as intended. Users must be trained in electrical safety procedures
and proper use of the instrument. Users must be protected from electric shock and contact with
hazardous live circuits.
Service is only to be performed by qualified service personnel.
Safety symbols and terms
Connect it to safety earth ground using the wire recommended in the user manual.
The symbol on an instrument indicates that the user should refer to the operating
instructions located in the manual.
Complies with the essential requirements of the European directives.
Certification
We certify that this product met its published specifications at time of shipment from the factory.
2.1 Front & Rear Panel.......................................................................................................................................4
1.Check the list of supplied items..........................................................................................................................5
2.Connect the power cord and turn on the power supply.......................................................................................6
3. System Checkout..................................................................................................................................................6
5.If the Power Supply Does Not Turn On..............................................................................................................8
6.To Adjust the Carrying Handle...........................................................................................................................8
7. To Rack Mount the Instrument.............................................................................................................................9
4.3 VFD Description and Wiring Diagram.......................................................................................................13
4.4 Menu Descriptions......................................................................................................................................13
4.6 Menu Function............................................................................................................................................17
System Set ..........................................................................................................................................................18
5.1 Communication cable.................................................................................................................................22
5.2 Communication between Power Supply and PC ........................................................................................23
6.3.1 Save and Open ...............................................................................................................................................39
6.3.3 Status bar.......................................................................................................................................................40
6.4.1 Configure the system.....................................................................................................................................41
6.4.2 Setting Voltage and Current...........................................................................................................................41
6.5 Test sequence generation............................................................................................................................42
6.5.2 Program .........................................................................................................................................................42
The 9130 is a fully programmable triple Output DC Power Supply delivering 0-30V/0-3A on 2
outputs and 0-5V/0-3A on 1 output. Each output is fully floating and outputs can be adjusted
independently or connected in series or parallel to produce higher voltages or currents. The front
panel keys and the control knob provide a convenient means to adjust the voltage and current of a
selected output; enable or disable parallel or series tracking mode; store and recall operating states
or enable/disable individual outputs. The 9130 is ideally suited for applications in Electronic Test,
Production and Service, where multiple independent DC supplies are required and bench space is
at a premium.
Key Features
z 3 independent, fully programmable and electrically isolated outputs
z Series or parallel operation
z Display Voltage and Current for of all 3 channels simultaneously
z Over Current (OCP) and Over Temperature (OHT) protection
z Very compact foot print ( 2U x 1/2)
z Bright and easy to read display (VFD technology)
z High resolution, accuracy and stability
z Output on/off control
z Low ripple and low noise
z SCPI compatible command set, communicate via USB or RS232 port
z Application Software for front panel emulation and simple test sequence generation
z 50 memory locations for instrument state storage & recall
z Convenient data entry via knob or numerical key pad
z Rack mount kit available
3
2. Quick Start
This chapter will help you getting acquainted with the front panel operation of this power supply.
2.1 Front & Rear Panel
Front panel
1 2
3 4 5 6 7
1
VFD display
2
Rotary knob
3
Main power switch
4
Numerical key pad and “Esc” key
5
Function keys
6
Cursor position & Channel Selection keys, “Enter” key
7
Output terminals
4
Rear Panel
1 2
3 4 5 6
1) Terminal block for GND reference and additional negative output terminal
2) AC receptacle
3) Ventilation holes
4) Interface connector for remote control of power supply
5) 110V/220V AC Power selection switch
6) Fuse compartment
2.2 Preliminary Checkout
The following steps will help you verify that the power supply is ready for use.
1.Check the list of supplied items
Verify that you have received the following items with your power supply. If anything is missing, contact your
authorized B+K Precision distributor.
□ Power cord
□ Instruction manual
□ Calibration Report
□ Communication cable
□ Software installation disk
5
2.Connect the power cord and turn on the power supply
When you turn on the power supply, the front-panel display will light up briefly while the power supply
performs its power-on self-test. All the VFD annunciators will turn on at once. Check for any missing display
segments. In the event that there is no response when you turn on the power supply, refer to section 5 of this
chapter for additional information.
3. System Checkout
At power up, the instrument will automatically perform a self test routine. During this time, the following
should be displayed:
System Test, Please wait!
followed by
0.000V 0.000V 0.000V
<OFF> <OFF> <OFF>
Or:
Note: The first row is the output voltage for channels 1 – 3. The second row will display the on/off state or the
output current of each channel.
In case the self test routine is not successful, you may see one of the following:
If the EEPROM is damaged, the VFD will display the following:
10.000V 11.000V 3.000V
2.000A 3.000A 3.000A
EEPROM Error
If the last operation data which should be stored in the EEPROM was lost you should see
Data Check Error
If the latest data about off-time in EEPROM was lost, the VFD will display
Load OffTime Fail
If the calibration data stored in the EEPROM was lost, the VFD will display
6
CH X Lost Calibration...
Note: “X” denotes the channel for which calibration data was lost.
If the calibration data in the EEPROM is corrupted or the factory calibration values are lost, the VFD will display
the following:
Lost Factory Calibration
Note: You will see a “?” on the VFD, in case of any errors are encountered during the self
test routine.
4.Output Verification
The following procedures verify that the power supply outputs the correct voltage and current levels and properly
responds to entries from the front panel.
Voltage Output Checkout
The following steps verify basic voltage functions without load.
1) Turn on the power supply.
2) Turn on the outputs using the
Note: Flashing voltage values indicates that the power supply is in “Set mode, ‘‘Set mode’’ means that
the VFD display shows the set values for voltage and current. In “meter mode” the display will
indicate the actual output voltage and current.
3) Check that the front-panel voltmeter properly responds to number key entries
Enter a different voltage value and wait a few seconds until the meter mode activates. Verify that the
actual output value voltage is identical to the set value. Also verify that the displayed current value is
close to zero.
4) Ensure that the voltage level can be adjusted from zero to the maximum rated value.
5) Verify channels 2 & 3 according to steps 1) – 4)
Current Output Current
The following steps check basic current functions with a short across the power supply’s output.
1) Turn on the power supply
2) Turn the output off
Press
3) Connect the (+) and (-) output terminals of channel 1 with a short, insulated test lead. Use a wire size
sufficient to handle the maximum current.
4) Set the voltage value with 1V
5) Turn on the output using the
6) Adjust the current value
Enter a different current value, wait until the instrument is in meter mode, then make sure that the
On/Off
key to ensure that the output s are off. (bottom row indicates “OFF”)
On/Off
On/Off
key
key.
7
displayed current value (actual output value) is the same as the set value.
7) Ensure that the current can be adjusted from zero to the maximum rated value.
8) Turn off the power supply and remove the short wire from the output terminals.
9) Verify channels 2 & 3 according to steps 1) – 8)
5.If the Power Supply Does Not Turn On
Use the following steps to help resolve problems you might encounter when turning on the instrument.
1). Verify that there is AC power applied to the power supply.
Verify that the power cord is firmly plugged into the power receptacle on the rear panel of the power supply.
Make sure the power outlet you are using is working properly and verify that the power supply is turned on.
2). Verify the power-line voltage setting.
Make sure the voltage selector switch is set according to the present line voltage (110VAC or 220VAC).
Change the voltage setting if it’s not correct.
3). Verify that the correct power-line fuse is installed.
If the fuse is blown replace it according to the table above.
6.To Adjust the Carrying Handle
To adjust the position, grasp the handle on each side and pull outwards, then rotate the handle to the desired
position.
Bench-top viewing positions Carrying position
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
8
7. To Rack Mount the Instrument
You can mount the power supply in a standard 19-inch rack cabinet using the IT-E151 rack mount kit.
Note: Remove the carrying handle and the two plastic ears before rack-mounting the instrument. To remove
the handle, grasp the handle on the side, pull outwards and rotate it to a special position where the
arrow on the handle and the arrow on the plastic ears are in opposite directions. Now you can pull
the handle outwards. After removing the handle, you can remove the two plastic ears with a screw
driver.
9
3. Specifications
3.1 Specifications
Parameter 9130
Output ratings
±(%of output+offset)
Voltage 0~30V×2,0~5V×1
Current 0~3A×2,0~3A×1
LVP*
Voltage ≤0.01%+3mV Load Regulation
Current ≤0.01%+3mA
0~31V×2,0~6V×1
Line Regulation
±(%of output+offset)
Resolution
Readback
Resolution
Programming accuracy
(12 months)
±(%of output+offset)
Readback accuracy
(25 °C ± 5 °C)
±(%of output+offset)
Ripple & noise Voltage
Temperature coefficient
(0 °C ~ 40 °C)
±(%of output+offset)
Readback Temperature
Coefficient
±(%of output+offset)
Tracking Accuracy
Series Operation
Voltage ≤0.01%+3mV
Current ≤0.1%+3mA
Voltage 1mV Programming
Current 1mA
Voltage 1mV
Current 1mA
Voltage ≤0.03%+10mV
Current ≤0.1%+5mA
Voltage ≤0.03%+10mV
Current ≤0.1%+5mA
Ripple: ≤1mVrms/3mVp-p
Noise: ≤3mVrms
Voltage ≤0.03%+10mV
Current ≤0.1%+5mA
Voltage ≤0.03%+10mV
Current ≤0.1%+5mA
Current ≤0.05%+10mA
Voltage ≤0.02%+5mV Tracking Accuracy
Parallel Operation
Output Timer
Weight 19.8 lbs, (9 kg)
Dimensions
*) LVP: Limit Voltage Protection. Limits the voltage than can be set either via the front panel or remote control command.
NOTE: Specifications and information are subject to change without notice. Please visit www.bkprecision.com for the most
current product information.
Current ≤0.1%+20mA
Time set 1s~999999s
Resolution 1s
214.5mm(W) × 88.2mm (H) × 354.6mm (D)
8.45in(W) x 3.8in(H) x 13.9in(D)
10
3.2 Supplemental Characteristics
State Storage Memory
50 user-configurable stored states
Recommended Calibration Interval
1 year
AC Input Ratings (selectable via switch on the rear panel)
220AV±10%,47~63Hz
110AV±10%,47~63Hz
Maximum power consumption
Model 9130
Power 750VA
Cooling
Fan cooled
Operating Temperature
0 to 40 °C for max rated output
Storage Temperature
-20 to 70 °C for storage environment.
Environmental Conditions
Designed for indoor use in an installation category II, pollution degree 2 environment. Designed to operate at
maximum relative humidity of 95% and at altitudes of up to 2000 meters.
11
4. Front-panel Operation
So far we have covered the quick start chapter which briefly introduced the front panel operation and how to
check basic voltage and current functionality. This chapter describes in detail how to operate the instrument
manually via the front-panel keys.
4.1 Front-panel Operation Overview
z The power supply is shipped from the factory ready for front-panel operation mode. At power-on, the power
supply will automatically enter the front-panel operation mode and the instrument can be controlled via the
font panel keys and knob.
The power supply enters remote mode as soon as a valid remote command is received via the
communication connector in the rear. Switching to remote mode does not impact the supply’s output
parameters. In remote mode, front-panel operation is disabled. If the power supply is in remote mode, and
the [LOCAL] key is not disabled, you can revert to manual mode by pressing the [LOCAL] key .
z The power supply is in Meter mode when it is powered on. In this mode, the VFD will display the actual
voltage and current output value. If the user adjusts the knob while in this mode, the power supply will
automatically enter Set mode then revert back to meter mode and display the modified voltage and current
values.
z The output of the power supply can be enabled/disabled from the front panel by pressing the
z The VFD also displays operation states or error information. “”means the power supply is in remote
mode. If an error occurred, ”?” will be displayed.
z If the power supply is in Set mode, you can modify parameters using the knob. If the power supply is in
menu operation, the knob is used for menu selection. If the power supply is in the regular output mode you
can use the knob to set the voltage value.
z If “?” is displayed on the VFD, please consult the submenu “error information” for more details (see chapter
4.4).
On/Off
●●
key.
12
4.2 Panel Description
12
4
5
7
0~9 Use keys 1~3 to control the output state of the 3 channels。Use keys 4~6 to set the
voltage and keys 7~9 to set the current value for each channel.
V-setV-set
I-setI-set
Save
Recall
Menu
On/Off
Set the voltage value
Set the current value
Save the current setup of the power supply to internal memory
Recall the setup from internal memory
Set general parameters of the power supply
Set the output state of the power supply
3
6
98
Esc
0
●
V-setV-set
Save
Menu
I-setI-set
Recall
On/Off
Enter
Up arrow key, select a menu or channel Down arrow key, select a menu or a channel
4.3 VFD Description and Wiring Diagram
Explanation of annunciators on the display
OFF
The output is off.
The key panel is locked.
The power supply is in remote mode.
There is some error or fault with the power supply.
?
Indicates the channel currently selected
4.4 Menu Descriptions
Press
Menu
to indicate operation mode. View the menu on the VFD, and use the keys or the knob to
13
scroll through the complete menu listed below. Press
Enter
to enter the selected menu function, pre ss
Esc
to return to the previous menu. A ”↑↓” symbol displayed on the left means that you are currently in the middle of
a menu and you can now use the
are at the end of the currently a ctive menu and you can only press the
keys to select other menu parameter. If there is only”↑” displayed, you
key to select a di fferent parameter.
Accordingly, ”↓” indicated that you can use the
enclose the currently selected menu.
Menu
Power Config...
Reset Config System reset
Out State Set “Remember” Output On/Off state before power off
Out Parameter Set “Remember” parameters before power off
Key Sound Set Knob Function Set Enable/disable knob
Screen Brightness Baud Rate Set Set baud rate
Communication Parity System Wait Time determines duration of set mode
Local Address Set communication address
Key Lock Set Set password
Exit
System Set… Out Series Set Configure for series operation
Out Parallel Set Configure for parallel operation
Max Voltage Set… Set the max. voltage for each channel (LVP)
Set First Channel Set Second Channel Set Third Channel Out Time Set… Set output time for each channel
Set First Channel Set Second Channel Set Third Channel Exit
Power Information… Power Model Display the power rating
Power SN display the serial number
Soft Version Display the firmware version
Cal Information Error Information Exit
Exit Menu
key to select another parameter. Symbols …….
14
4.5 Panel Operation
Channel Operation
When the power supply is in “METER” mode, you can select a channel using the arrow keys .
OUT ON/OFF
Pressing
press it, to turn the output state t o OFF. While th e output state is OFF, press
output will turn ON.
To control channels individually, press one of the number keys
1
Key
and key
When the power supply is in remote mode, you can set the output state by sending SCPI command(OUTPut:
ON | OFF)The Output state operation doesn’t affect any other parameter.
controls the output state of the first channel, key 2 controls the output state of t he second cha nnel
3
Note: The
If you want to control the output state of individual channels, use the number keys 1-3.
Timer Operation
On/Off
controls the output state of the third channel.
toggles the output state of all 3 channels of the power supply. If the output state is ON,
On/Off
key controls the output state of all 3 channels simultaneously.
On/Off
1
, 2 3 corresponding to each channel.
and the power supply
If you have set the output timer and the power supply is in “METER” mode, you can check the remaining time
by pressing
0
. When the timer expires, the power supply will turn off the channel automatically.
Set Voltage
Solution 1: Press
Solution 2: Press
Solution 3: Press one of the keys 4-6, corresponding to the channel which you want to adjust (Press 4 to
V-setV-set
V-setV-set
adjust the value with the knob. (Make sure the knob function is enabled in the systems menu,,
otherwise use solution 1). Press
cursor will remain visible indefinitely or until the system Wait time timer expires).
, enter a numeric value (via key pad) then press
, then use the arrow keys to move the cursor to the desired position and
Esc
or
15
Enter
to return to meter mode. (Otherwise the
Enter
to confirm.
adjust the first channel, press 5 to control the second channel and press 6 to control the
third channel).
4
For example, to set voltage for the first channel: Press
, then enter a numerical value
followed by
using the knob. Press
Enter
. Alternatively, press to move the cursor adjust the voltage value
Esc
o r
Enter
to escape and return to meter mode.
Set Current
Solution 1: Press
Solution 2: Press
I-setI-set
I-setI-set
, enter a numeric value then press
, the press to move the cursor position and adjust the voltage value using the
knob. (Make sure the knob function is enabled in the systems menu,, otherwise use solution 1).
Press
Esc
or
Enter
to exit.
Solution 3: Press one of the number keys 7 – 9 corresponding to the channel for which you want to set the
7
current value. (Press key
9
key
channel, press key
to control the third channel). For example, if you want to set the current for the first
7
to control the first channel, 8 for the second channel or press
, then enter a numeric value followed by the
Enter
to confirm.
Enter
key. Alternatively,
or press
to move the cursor and ad just the value with the knob. Press
Esc
or
Enter
to escape and return to meter mode. (Otherwise the cursor will remain visible indefinitely or
until the system Wait time timer expires).
Save and Recall Operation
You can store up to 50 different operating states in memory locations 1 through 50. Each operating state
includes a constant voltage value, constant current value, maximum output voltage value and voltage step
value. Press
Press
Recall
You can also use the SCPI command:*SAV、*RCL to save and recall.
Over Temperature Protection
If the internal temperature of the power supply exceeds 80 °C, the instrument will protect itself by
automatically turning power OFF. When this happens you will hear a buzzer and the display will indicate the
following:
Save
followed by a number key to save the current operating state to non volatile memory.
followed by a number 1 – 50 to recall operating state assigned to this location.
16
Over Temp
4.6 Menu Function
In menu operation, the keys or the knob can be used to select the menu.
Esc
selection. The
key is used to exit the current menu
Power Config
Reset Config
If you enter this menu and select “YES”, all parameters will be set to their default values.
Out State Set
This parameter sets the Output On/Off state at power up. If you select “Last Set”, the power supply will save
the output state prior to power down and revert to that state at po wer up. If you select “Off”, the output state is
always “OFF” when the power supply is turned on. The recommend setting is “OFF”.
If Out Parameter Set is set to “Default”, th e out pu t stat e o f the po wer sup pl y afte r powe r –up wi ll alwa ys be off,
regardless of the settings of
Out Parameter Set
This parameter determines the state of the power supply after power up. If you select “Last Set”, the
power supply will automatically restore (“remember”) the last output parameters setting prior to power
down. If you select “default”, the default output parameter settings will be active after power up.
Recommend setting is “Last Set”.
Key Sound set
This parameter turns the beep sound for key presses “On” or “Off”.
Knob Function Set
This parameter enables (“On”) or disables (“Off”) the knob.
Baud Rate Set
this parameter
Enter
is used to confirm a
This parameter configures the baud rate for serial communication. Possible values are 4800,9600,19200 or
38400. When operating the power supply in remote mode, make sure that you configure identical baud rate
settings for the power supply and the computer. The default setting is 9600.
Communication Parity
This parameter sets the parity bit for serial communication. Possible values are NONE, ODD and EVEN.
Default setting is NONE.
System W ait T ime
This parameter determines how long the power supply will remain in set mode until it automatically reverts
17
back to meter mode or output mode The minimum is 4 seconds, and the maximum is 9999 seconds. Enter a
numerical value followed by
wait time function, set the wait time to 0 seconds. (In this case, the power supply will remain in set mode until
the ESC or ENTER key is pressed). Set mode is active when a cursor is visible. Default setting is 0 seconds.
Enter
or press +
Enter
to set the wait-time. To disable the system
Note: the wait-time range is 4~9999S, if you set it with 1~3S, the wait-time will be 4S automatically.
Local Address
With this parameter, it is possible to address each instrument. The address range is 0 to 30.
Key Lock Set
A password can be assigned (1 - 4 digits) to lock the function key operation. After setting the password, there
is a sign displayed on the VFD and all the function keys on the front panel will be locked except of the
On/Off
unlock the keys. If you don’t want to lock the function keys, please set the password to 0 when entering the
>SET KEYLOCK function.
When the knob function is enabled:
You can press the keys to move the cursor position, then enter a number and press
key (if the knob is enabled,keys can also can be used). You must enter the correct password to
Enter
to set
the password. Or press
knob. Press +
Esc
or
to select the cursor position and change the value using the password using the
Enter
to exit.
When the knob function is disabled:
Press number key+
Enter
to set the password or use the keys.
System Set
Out Series Set
This function configures the instrument for series operation of 2 channels. “None” means that each channel
operates independently (power supply not in series mode). “1+2” means that channel 1 and channel 2 operate
in series, 1+3 means that channel 1 and channel 3 operate in series.
Note: Channels 2 and 3 can NOT be configured for series operation.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
18
Connection diagram example: Channels 1 and 2 connected in series
+
Load
Out Parallel Set
This function configures the instrument for parallel operation. “None” means that each channel operates
independently. “1+2” means that channel 1 and channel 2 operate in parallel, “1+3” and 2+3 means that
channels 1 and 3 or channels 2 and 3 respectively operate in parallel. ALL indicates that all 3 channels
operate in parallel.
Note: The output state of all channels will automatically be set to OFF and the voltage value
will be 0V after configuring the instrument for series or parallel operation, therefore you have to
set the parameter again after reconfiguration.
19
Connection diagram example: Channels 1 and 2 connected in parallel
+
Load
Note: When configuring for series or parallel operation, the user must wire
the terminals to match the configuration. The power supply does not
automatically connect the appropriate terminals internally based on the
configuration selected.
MaxVolt Set (LVP)
The max voltage you set should be within the range of 0V to the maximum rated voltage. You can edit this
value using the
maximum rated voltage for each channel.
keys or via numerical key pad followed by
Out Time Set
This parameter sets the output timer for each channel. The range is 1~999999 seconds. If you enable this
function, and the output state of all channels is on, the timer will start counting down immediately. Once the
timer expires (count down from set value to zero) the output of the assigned channel will turn off. To disable
the timer, set the output time to zero.( 0 seconds). Default setting is 0 Seconds.
Enter
. The default setting is the
Power Information
The following menus provide information about the specific power supply unit.
20
Power Model: the model of power supply
or example: 30V,3A*2CH 5V,3A*1CH
F
Power SN: the serial number
For example: 001001156074001165
Soft Version: the version number of the power supply
For example: Soft Version=1.00
Cal Information: calibration information of the power supply (last calibration date)
For example: 2005-8-26 17:46:13
Error Information: error information of power supply
For example: 0,No Error
Exit Menu
Note: after the error information has been displayed, you can press
Esc
or
Enter
to exit.
Afterwards the error information will be cleared out although the error condition still exists. Consult
the table below for an explanation of error codes.
0 'No Error' There is no error.
1 'Too Many Num Suf' The number in the ROM is too many to deal with.
10 'No Command' The command is invalid.
14 'Num Suf Invalid' The subscript of the number is invalid.
16 'Invalid Number' The number is invalid.
17 'Invalid Dims' The data dimension is invalid.
20 'Param Overflow' Parameter overflow.
30 'Error Para Units' Parameter unit error.
40 'Error Para Type' Parameter type error.
50 'Error Para Count' Parameter count error.
60 'Unmatched Quote' The sign quoted by parameter is unmatched.
65 'Unmatched Bracket' The bracket doesn’t match the parameter.
70 'Invalid Command' The command is invalid.
80 'No Entry' Command entry can not be found.
90 'Too Many Dims' Too many data dimension.
100 'Too Many Command' Too many commands.
101 'Command Exec Err' An error occurred during command execution.
110 'Error Rxd Parity'
120 'Error EEPROM' There is error when checking the EEPROM.
121 'Error Config Data' Configuration data error.
122 'Error Cal. Data' Calibration data error.
123 'Error Factory Data' Calibration data supplied by factory is error.
21
5. Remote Operation
The DB9 interface connector on the rear panel of the power supply can be connected to a RS-232 or USB
interface via the included adapter cable. This chapter describes how to use a computer to control the output of the
power supply.
5.1 Communication cable
RS232 Communication cable IT-E131
The DB9 interface connector on the rear panel of the power supply provides a TTL voltage level interface. Use
the communication cable (IT-E131) to connect the DB9 interface connector of the power supply to the RS-232
interface connector of the computer
IT-E131 communication cable
Power
PC
PCLoad
COMPUTER
COMPUTER
IT-E131 ISOLATED
IT-E131 ISOLATED
IT
IT
COMMUNICATION CABLE
COMMUNICATION CABLE
859666668889942311
859666668889942311
RX
RX
TTL(5V)RS232ISOLATION
TTL(5V)RS232 ISOLATION
TX
TX
INSTRUMENT
INSTRUMENT
supply
Note: It is not possible to connect the DB9 connector of the power supply directly to a PC’s
RS232 or USB port.
USB Communication cable IT-E132
The DB9 interface connector on the rear panel of the power supply provides a TTL voltage level interface. Use
the communication cable (IT-E132) to connect the DB9 interface connector of the power supply to the USB
interface connector of the computer.
IT-E131 communicationcable
IT-E132 communication
Power
PC
PCLoad
COMPUTER
COMPUTER
IT-E131ISOLATED
IT-E131 ISOLATED
IT
IT
COMMUNICATION CABLE
COMMUNICATION CABLE
859666668889942311
859666668889942311
RX
RX
TTL(5V)RS232ISOLATION
TTL(5V)RS232 ISOLATION
TX
TX
INSTRUMENT
INSTRUMENT
supply
Note: Before you can use the USB communication cable, you must install the USB driver which
can be found on the included installation disk.
22
5.2 Communication between Power Supply and PC
Before putting the instrument into remote operation, make sure that the baud rate, parity bit and communication
address settings on the power supply and computer side are identical, otherwise communication will not be
possible.
1. Address Range is 0 to 30. Default setting is 0.
2. Baud rate: 4800,9600,19200 and 38400 are selectable, default setting is 9600
3. Parity and Data bits:None/8 bits (default setting)
Even/8 bits
Odd/8 bits
4. Stop bits:1 (fixed)
5. Start Bits: 1 (fixed)
Data Frame Format
Parity=None Start
Bit
Parity=Odd,
Even
Start
Bit
8 Data Bits
8 Data Bits
Stop
Bit
Parity
Bit
Stop
Bit
End of String character is ’\n’ (0x0a)
DB9 Interface Details
The DB9 connector in the rear panel of the power supply provides a TTL level signal .It can be connected to a
standard PC interface via the IT-E131 or IT-E132 isolated communication cable.
Note: Setting of the address is optional. It is not required to communicate with the instrument. The
address can be set from the front panel and is stored in non volatile memory. This feature is
useful when connecting several instruments, e.g. via USB hub. In this scenario, Windows assigns
a virtual COM port to each device which is unknown prior to establishing communications with
the instrument (could be different each time). In this case, the address will allow the user to
correlate each virtual COM port with the address assigned to each instrument.
You can obtain the state of the power supply and read parameters from the operation register. The different states
of the power supply can be read from 7 condition registers. These registers are status byte register, standard event
register, quest condition register and operation status register. The status byte register stores the information of 3
other registers. The following table provides the details on each register’s meaning.
bit code meaning
Operation status
sub register
register
Standard event
register
Status byte
register
operation status
register
0 CAL The power supply is calculating new calibration param eters.
1 UNR The status of the power supply is unregulated.
2 CV The power supply is in constant voltage condition.
3 CC The power supply is in constant current condition.
4 RI Not used
0 OV Not used Quest condition
1 OT Over temperature
0 OPC Operation of power supply is completed.
2 QYE Query error. Data of output array is missing.
3 DDE Device-dependent error. Data stored in register is missing or error occurs
in preliminary checkout.
4 EXE Execution error. Command parameter overflows or the condition is not
right.
5 CME Command error. Syntax or semantic error occurs when receiving
information.
7 PON Power on. It is 1 when power supply is reset.
3 QUES If a quest enable condition changes, QUES is 1.
4
5 ESB If a standard event status enable register changes, ESB is 1.
6 MSS
7 OPER If a operation event enable register changes, OPER is 1.
1 INST1 If the status of one operation status sub register changes, INST is 1
2 INST2 As above
3 INST3 As above
26
The Structure of the condition register is as followed:
Operation status sub register(channel 1) Operation status sub register(channel 2) Operation status sub register(channel 3)
condition event enable
INST1 INST1 INST1
INST2 INST2 INST2
INST3 INST3 INST3
Operation status register
condition event enable event enable
OPC OPC OPC
QYE QYE QYE
DDE DDE DDE QUES QUES
EXE EXE EXE
CME CME CME ESB ESB
RQS RQS
PON
Standard event register Status byte register
condition event enable
OV OV OV
OT OT OT
Second bit: 2
Third bit: 3
Forth bit: 4
Fifth bit: 5
Sixth bit: 6
Quest condition register
PON
or
CC
CV
or
CV CV CV
CC CC CC
or
PON
or
or
or
OPER
OPER
Lowest bit: 0
First bit: 1
Note:
The bit array of each
register is as right
table:
Highest bit: 7
R
or
27
5.4.2 SCPI Command Description
Common IEEE488.2 Commands
*CLS This command clears the following registers:
Standard event status register
Quest condition register
Operation status register
Operation status sub register
Status byte register
Error code
Command syntax:*CLS
Parameter:None
*ESE
This command sets the parameter of the standard event enable register. The value determines which bit value of
the standard event register is set to 1. The ESB bit in the of status byte register will change to 1 to reflect the
changes.
This command queries the standard event status register. After executing this command, the standard event status
register is reset. The bit definition of the standard event status register is identical to the standard event status
enable register
Query syntax:*ESR?
Return parameter:<NR1>
Example: *ESR?
*IDN?
This command reads information about the power supply. The return value contains 4 segments divided by a
comma.
Query syntax:*IDN?
Return parameter:<AARD> segment description
BK Precision manufacturer
XXXX model
XXXXXX product serial number
VX.XX software version number
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
28
For example:BK PRECISION, 9130, 000000000000111101,V1.68
*OPC
When all commands sent to the instrument prior to this command have been executed, OPC of the
This command controls if the power supply sends a query or not when it is reset.
1|ON: When power supply is reset, operation event enable register, quest event enable register and standard
event status register are allreset.
0|OFF:The data of operation event enable register, Quest event enable register and standard event status
enable register is stored in a nonvolatile register, and is recalled when power supply is reset.
Command syntax:*PSC
Parameter:0|1|ON|OFF
Return parameter:data stored
Example:*PSC ON
*RST This command resets the power supply to its default setting.
Command syntax:*RST
Example:*RST
*SRE
This command can set the parameter of the status byte enable register. The value of this parameter determines
which bit value of the status byte register is 1 and the byte will enable RQS of status byte register is 1. The bit
definition of the status byte enable register is the same as the status byte register.
Command syntax:*SRE <parameter>
Parameter:0~255
Reset value:consult command *PSC
Example:*SRE 110
*STB?
This command reads data from the status byte register. After executing this command, the status byte register
is reset.
Command syntax:*STB?
Return parameter:<NR1>
Example:*STB?
*SAV
This command saves the operating parameters of the power supply to non-volatile memory. The
parameters include constant current, constant voltage, maximum voltage value and step voltage values.
29
Command syntax:*SAV
Parameter:0~49
Example:*SAV 10
*RCL This command recalls parameter saved previously (using the *SAV command).
SYSTem:ERRor This command is used to obtain error information from the power supply.
Command syntax:SYSTem:ERRor?
Return parameter:please consult the error information table
Example:SYST:ERR?
SYSTem:VERSion This command queries the software version.
Command syntax:SYSTem:VERSion?
Return parameter:software version
Example:SYST:VERS?
SYSTem:BEEPer This command controls the buzzer
Command syntax:SYSTem:BEEPer[:IMMediate]
Example YST:BEEP
SYSTem:LOCal
This command can set the instrument to local (operate from front panel) mode
Command syntax:SYSTem:LOCal
Example:SYST:LOC
SYSTem:REMote This command configures the power supply for remote control mode.
Command syntax:SYSTem:REMote
Example:SYST:REM
SYST:RWLock
This command also sets the instrument to remote control mode. When using this command, it is not
possible to press LOCAL key on the front panel to revert back to manual mode.
Command syntax:SYSTem:RWLock
Example:SYST:RWL
30
SYSTem:ADDRess This command reads the communication address of the power supply.
Command syntax:SYSTem:ADDRess?
Example:SYST:ADDR?
STATus:QUEStionable:ENABle
This command sets the parameter of the quest event enable register. This parameter determines which bit of
the quest event register is set to 1. If a QUES condition changes, the QUES bit of status byte register will
be set to 1.
Command syntax:STATus:QUEStionable:ENABle <parameter>
Parameter:0~255
Reset value:consult PSC command
Example:STAT:QUES:ENAB 110
STATus:QUEStionable:ENABle?
This query returns a value which corresponds to the binary weighted sum of all bits enabled by
STAT:QUIES:ENAB command. After this command executed, quest event enable register is reset.
Command syntax:STATus:QUEStionable:ENABle?
Return parameter:<NR1>
Example:STAT:QUES:ENAB?
STATus:QUEStionable?
This command reads parameters from the quest event register. After this command executed, quest event
register is reset.
Command syntax:STATus:QUEStionable[:EVENt]?
Return parameter:<NR1>。
Example:STAT:QUES?
STATus:QUEStionable:CONDition?
This command queries the parameter of the quest condition register. When a bit valu e of the quest condit ion
parameter changes, the corresponding bit value in the quest event register will be set to 1.
Command syntax::STATus:QUEStionable:CONDition?
Return parameter:<NR1>
Example:STAT:QUES:COND?
STATus:OPERation ENABle
This command sets the parameter of the operation event enable register. The parameter determines which bit
value of quest event registe r is set to 1. If a OPER condition changes, the OPER bit of the status byte
Return parameter:<parameter of operation event sub register>
Example:STAT:OPER:INST?
STATus:OPERation:INSTrument:ENABle
This command sets the parameter of operation event enable sub register. Setting parameter can determine
which bit value of operation event enable sub register is 1 and the bit will enable OPER of status byte
register is 1.
Note: this command is only valid for current channel.
This command queries the parameter of the operation event enable sub register. After this command
executed, operation event enable sub register is reset.
Note: this command is only valid for the currently active channel.
This command returns the number of the selected channel
Command syntax:INSTrument[:SELect]?
Return parameter:FIRst|SECOnd|THIrd
Example:INST?
INSTrument:NSELect
This command is similar to the command “INSTrument[SELect]”, the only difference is that this command
uses numbers to denote the channel.
Command syntax:INSTrument:NSELect
Parameter:1~3
Reset value:1
Example:INST:NSEL?
OUTPut[:STATe]
This command sets the output state for the current channel to ON or OFF
Command syntax:OUTPut[:STATe] <parameter>
Parameter:1(ON)|0(OFF)
Example:OUTP 1
OUTPut[:STATe]?
This command queries the output state of the current channel is ON or OFF
Command syntax:OUTPut[:STATe]?
Return parameter:0|1
Example:OUTP?
OUTPut:TIMer:DATA
This command sets the time of the output-timer for the current channel.
Command syntax:OUTPut:TIMer:DATA <parameter>
Parameter:0~999999S (if you set 0S,the output-timer will be disabled.)
Reset value:it is concerned about the parameter you set in the menu.
33
Example:OUTPut:TIMer:DATA 100
OUTPut:TIMer:DATA? This command queries the time of the output-timer for the current channel.
This command sets the current value of the power supply.
Command syntax:[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <parameter>
Parameter:MIN|MAX|MIN TO MAX
Unit:A mA uA
Reset value:it is concerned about the parameter you set in the menu.
*RST value:MAX
Example:CURR 2A
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? This command checks the current value of the power supply.
Command syntax:[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? <parameter>
Parameter:MIN | MAX |None
Return parameter:MIN T O MAX
Unit:A
Example:CURR?
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] This command sets the voltage value of the power supply.
Command syntax:[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
Parameter:MIN|MAX|MIN TO MAX
Parameter:V mV uV kV
Reset value:IN TO MAX|MIN|MAX
*RST value:MIN
Example:VOLT 10V
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? This command reads the voltage value of the power supply.
Command syntax:[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? <parameter>
Parameter:MIN|MAX|None
Reset value:MIN TO MAX
Unit:V
Example:VOLT?
34
[SOURce:]VOLTage:PROTection[:LEVel][:IMMediate][:AMPLitude] This command sets the max. voltage for the current channel.
Reset value:it is concerned about the parameter you set in the menu.
*RST value:MAX
Example:VOLT:PROT 20V
MEASure[:SCALer]:CURRent[:DC]?
This command reads the actual output current of the power supply.
Command syntax:MEASure[:SCALer]:CURRent[:DC]?
Return parameter:the actual output current
Unit:A
Example:MEAS:CURR?
MEAS[:SCALer][:VOLTage][:DC]?
This command reads the actual output voltage value.
Command syntax:MEAS[:SCALer][:VOLTage][:DC]?
Return parameter:the actual output voltage
Unit:V
Example:MEAS?
MEASure[:SCALer]:POWer[:DC]?
This command reads the actual output power value.
Command syntax:MEASure[:SCALer]:POWer[:DC]?
Return parameter:the actual output power
Unit:W
Example:MEAS:POW?
DISPlay[:WINDow][:STATe]
This command turns the VFD display on/off.
Command syntax:DISPlay[:WINDow][:STATe] <parameter>
Parameter:0(OFF)|1(ON)
Reset value:1(ON)
*RST value:1(ON)
Example:DISP 1
DISPlay[:WINDow][:STATe]? Thi s command checks the state of the display.
Return parameter:1(ON)|0(OFF)
Example:DISP?
35
CALibration:SECure[:STATe]
This command sets the protection mode to “enable” or “disable”. The power supply will be calibrated only if
the calibration protection is disabled and other commands are disabled, either.
This command sets the new calibration password.
Command syntax:CALibration:SECure:CODE <parameter>
Parameter:the password (1 through 4digits)
Example:CAL:SEC:CODE "1234"
CALibration:STRing
This command sets the calibration information.
Command syntax:CALibration:STRing <parameter>
Parameter:the calibration information about time and so on.
Example:CAL:STR "2005-1-9 20:12"
CALibration:STRing?
This command reads the cali bration information.
Command syntax:CALibration:STRing?
Return parameter:the calibration informati on s ave d in the EEPRO M.
Example:CAL:STR?
CALibration:SAVe
This command saves calibration coefficients into EEPROM
Command syntax:CALibration:SAVe
Example:CAL:SAV
Note:the calibration parameter will take effect only after it is saved.
CALibration:INITital
This command renews the calibration coefficients and set the default values.
Command syntax:CALibration:INITital
Example:CAL:INIT
Note:if the calibration fails, you can use this command to renew.
37
6. PV9130 Software Guide
6.1 Introduction
PV9130 is a software application for the 9130 Triple Output Programmable DC Power Supply. To use this
software communication you need either communication adapter cable IT-E131 or IT-E132. This application
software supports all the functionality available through the front panel. Additionally you can generate and
execute simple test sequences.
6.2 Installation
1.Requirements for computer
Pentium or Pentium processor-based personal computer
Windows 98/2000/XP or Windows NT4.0
2. Installation
Insert the included CD into the CD drive, locate PV9130 and install the software according to the onscreen
instructions.
6.3 PV9130 basics
After launching software PV9130, you should see following screen:
1
2
3
4
5
6
7
89
11
10
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
38
1
Configure the software operating parameters and serial communication parameters.
2
Voltage chart
3
Voltage and current displayed for all 3 channels
4
Set the voltage value
5
Set operation mode: PC Control or Panel Control
6
Set the output state: Output On/Output Off
7
“Local Enable” allows the user to use the front panel keys while the instrument is in remote control mode.
When “Local Disable” is active, the front panel keys of the instrument are disabled.
8
Set the current value
9
Status bar: Displays the power supply model, communication status and operation status.
10
Set voltage and current value quickly and conveniently via predefined hotkeys
11
Current chart
6.3.1 Save and Open
Save the program, Quick Set, voltage sweep and GO/NG settings to a PAR file.
Open the program, Quick Set, voltage sweep and GO/NG settings stored in a PAR file.
39
6.3.2 Chart Description
The voltage and current chart can help you to analyze voltage and current changes more easily.
If you want to observe the voltage and current change curve after you have set the voltage and current value for
one channel, please make sure that the channel’s output state is on . If you want to observe the
voltage and current curve at the same time, you need to set voltage and current value for each channel one by one,
and turn the channel’s output state on.
Note: After you have set voltage/current values or the output state for one channel and you want to set
parameters for another channel, you can select the channel you want to configure directly via the
tabs (blue means the channel is selected). The other channel’s state isn’t
affected.
6.3.3 Status bar
The status bar will give you the communication information. When communication is successful, the status bar
will display the following:
Power mode Communication status Time
1) Power mode: it will display the real part number of the field power supply which detected by the computer.
(9130)
2) Communication status: If Communication waiting appears in the status bar for more than 10 seconds, you
need to check your configuration and connection between computer and power supply. If
necessary, reconfigure the baud rate, address, COM port etc to make sure the parameters on
both sides match.
40
6.4 Operation
6.4.1 Configure the system
The first step is to configure the system and set up communication. Click the button and the
following windows will be displa yed:
1)Comm:Set the communication port and baud rate
2)Voltage/Current Step:Set the step size of the
arrow key, page up/down key, and mouse-wheel.
When you set the value via knob (coarse
adjustment), you can use the arrow keys (↑,↓, ←, →),
page up/down keys or mouse wheel for fine
adjustment.
3)Reload last parameter: When you check this box,
the last parameters from a previous session (before
closing this application) will be loaded.
Note:When you set the communication port, please make sure that the baud rate of the computer
matches that of the power supply. Also make sure that “Parity” in the power supply menu is set to
“None” (default value).
6.4.2 Setting Voltage and Current
Before setting voltage/current values, you need to select a channel . If you want to set voltage and current for
several channels, you can do so by setting one channel at a time
Use the mouse to click on the rotary knob and move the mouse to change the value. You also can use the mouse
wheel, or Page Up/Down keys and arrow keys (↑, ↓,←, →) from the keyboard for fine adjustment of voltage or
current values.. The setup value will be displayed on the second line indicator.
41
6.5 Test sequence generation
6.5.1 Quickset
Hot Keys
Hotkey:Right-click each Hotkey Button, it will display
, Click on it, it will display as follows, change
the voltage/current value as you desired, then click “OK” button to
confirm.
Note: The voltage/current value you set should not exceed the max voltage /current value.
Voltage Sweep
Before this operation, please select the channel you need:.
Enter parameters StartValue, StopValue, voltageStep and Delay. Press “View Scan Panel”, then “Run” to sweep
from the Start to the Stop value once or “Step” to manually execute each test step.
Refer to the on line help for more details
6.5.2 Program
Write a simple sequence of test steps using this tool.
Before programming, please select the channel:.
Right-click in the Program area. The program tools (AppendLine, Insert Line, DeleteLine) will appear. Use
these tools to define each test step and add new test steps.
Click on “View Test Control Panel” to execute your test sequence. Select “Run” to execute the sequence in a
loop, “Once” or Custom (define the number of loops)
For more details, refer to the on-line help.
42
6.5.2 Go/NG
This tool allows you to write a simple test sequence including min/max values. The test program generates a
pass fail verdict and an optional test report.
Before test, please select the channel:.
Right-click in the Program area. The program tools (AppendLine, Insert Line, DeleteLine) will appear. Use
these tools to define each test step and add new test steps.
Click on “View Test Control Panel” to execute your test sequence. Select “Run” to execute the sequence in a
loop, “Once” or Custom (define the number of loops)
For more details, refer to the on-line help.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
43
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.