BK Precision BK Precision 878B, BK Precision 879B-220V Manual

Model 878B, 879B
Dual Display LCR
INSTRUCTION MANUAL
METER
Visit us at www.TestEquipmentDepot.com
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431

Safety Summary

The following safety precautions apply to both operating and maintenance personnel and must be observed during all phases of operation, service, and repair of this instrument.
DO NOT OPERATE IN AN EXPLOSIVE
ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
KEEP AWAY FROM LIVE CIRCUITS
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must be made by qualified maintenance personnel.
DO NOT SUBSTITUTE PARTS OR MODIFY
THE INSTRUMENT
Do not install substitute parts or perform any unauthorized modifications to this instrument. Return the instrument to B&K Precisi on for
1
service and repair to ensure that safety features are maintained.
WARNINGS AND CAUTIONS
WARNING and CAUTION statements, such
as the following examples, denote a hazard and appear throughout this manual. Follow all instructions contained in these statements.
A WARNING statement calls attention to an operating procedure, practice, or condition, which, if not followed correctly, could result in injury or death to personnel.
A CAUTION statement calls attention to an operating procedure, practice, or condition, which, if not followed correctly, could result in damage to or destruction of part or all of the product.

Safety Guidelines

To ensure that you use this device safely, follow the safety guidelines listed below:
2
This meter is for indoor use, altitude up to
2,000 m.
The warnings and precautions should be
read and well understood before the instrument is used.
When measuring in-circuit components, first
de-energize the circuits before connecting to the test leads.
Discharge capacitor before testing.
The meter is safety-certified in compliance
with EN61010 (IEC 1010-1) Installation Category II (CAT. II) 50 V, Pollution Degree 2 environment.
Use the meter only as specified in this
manual. Otherwise, the protection provided by the meter may be impaired.
The power for the meter is supplied with a
single standard 9V battery. But also a line operation is possible using a 12V AC to DC adaptor. If a power adaptor is selected, please be sure to use fulfilled the safety requirements of a relevant IEC standard.
3
outside is negative (-)
Safety Symbols
This symbol is a warning and indicates that the user should refer to the operating instructions located in the manual.
DC Current
Indicates inside pin is positive (+),

Compliance Statements

Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with separate collection systems)
4
This product is subject
Council of the European
equipment (WEEE) , and
to Directive 2002/96/EC of the European Parliament and the
Union on waste electrical and electronic
in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste. Please utilize your local WEEE collection facilities in the disposition of this product and otherwise observe all applicable requirements.
5
Storage Humidity
0 – 80% R.H.
Storage Environment
-20 °C to +50 °C
Pollution degree
Pollution degree 2
Environmental Con diti ons
Operating Environment
0 °C to 40 °C
6
TABLE OF CONTENTS
Safety Summar y ........................................................ 1
Safety Guidelines ...................................................... 2
Compliance Statements ........................................... 4
INTRODUCTION ....................................................... 10
PACKAGE CONTENTS ........................................... 11
FRONT PANEL OVERVIEW .................................... 12
Front Panel Display Descriptions .................................. 13
Front Panel Buttons ...................................................... 15
LCD DISPLAY OVERVIEW ...................................... 17
LCD Display Descriptions ............................................. 17
Special Display Indicators ............................................. 19
POWERING INSTRUMENT ..................................... 20
Installing Battery ........................................................... 20
Connecting External Power Source .............................. 22
Low Battery Indication .................................................. 24
Backlit Display (model 879B only) ................................ 25
OPERATION INSTRUCTIONS ................................. 27
Data Hold ...................................................................... 27
Static Recording ........................................................... 27
7
L/C/R/Z Select Mode .................................................... 30
D/Q/θ/ESR Select Mode ............................................... 31
Test Frequency ............................................................. 31
Relative Mode ............................................................... 50
Tolerance ...................................................................... 52
Utility Menu ................................................................... 56
Parallel and Series Measurement Mode ....................... 67
Calibration..................................................................... 85
USB .............................................................................. 90
Automatic Fuse Detection ............................................. 91
QUICK START GUIDE ............................................. 93
CAUTION...................................................................... 93
Inductance Measurement ............................................. 94
Capacitance Measurement ........................................... 96
Resistance Measurement ............................................. 98
Impedance Measurement (Model 879B only) ............... 99
REMOTE COMMUNICATION ................................ 101
Connecting Instrument to PC ...................................... 101
USB (Virtual COM) Configuration ............................... 103
USB Operation ............................................................ 103
Command Protocols ................................................... 105
8
SUPPLEMENTAL INFORMATION ........................ 118
Selecting Test Frequency ........................................... 118
Selecting Series or Parallel Mode ............................... 120
Accuracy Discrepancies.............................................. 121
Guard Terminal ........................................................... 123
SPECIFICATIONS .................................................. 125
General Specifications ................................................ 126
Accuracy Specifications .............................................. 127
MAINTENANCE ...................................................... 133
Service ........................................................................ 133
Cleaning...................................................................... 133
SERVICE INFORMATION ...................................... 135
LIMITED WARRANTY ............................................ 136
9

INTRODUCTION

B&K Precision’s 878B and 879B 40,000-count L/C/R hand-held meter is designed ideally for measuring inductance, capacitance and resistance components. Simple to operate, the instrument not only takes absolute parallel m ode m easur em ents, but a lso series mode measurement. The meter provides direct and accurate measurements with selectable testing frequencies.
Front panel push buttons maximize the convenience of function and feature selection such as data hold, maximum, minimum and average record mode, relative mode, tolerance s orting mode, fr equency and L/C/R selection.
The test data can be transferred to PC through a Mini USB connection, great for applications that require data logging
A tilt stand provides position flexibility for viewing and operating the meter. The over-molding rubber case protects the meter for better durability. Additionally, top rubber visor molding above the screen is designed to prevent scratches on the display when meter is positioned upside down.
10
A single 9V battery or the included DC 12V power adaptor (model 879B only) can be used to power the meter. This gives user flexibility for portable or bench­top use.

PACKAGE CONTENTS

Each 878B and 879B LCR meters are shipped with the following contents.
878B or 879B LCR meter
Instruction Manual
Mini USB Interface Cable
Red & Black Banana to Alligator Test Leads
9V Batt ery
*AC Ad apter (m odel 879 B only)
*This can be purchased as an optional accessory for model 878B.
Please locate them from the original packaging to ensure nothing is missing. If in the case that an item is missing, please contact B&K Precision immediately.
11
+ -

FRONT PANEL OVERVIEW

LCR Meter
REL
LCR
Z
ESR
MAX
MIN
AVG
¦ÈQ
DH
AUTO
TOL
D
1%5%10%20%
OFF
@
1
2
3
POWER
D/Q/
P
/ESR
FREQ
S
4
5
L/C/R/Z
USB
TOL
879B
deg % kHz
PAL SER
n
¦ÌH
n
p F
¦Ì Mk RMT
R
HOLD
REC
UTIL
REL
CAL
7
8
9
10
6
13
!
11
12
Figure 1 - Front Panel Display (model 879B shown)
12

Front Panel Display Descriptions

1. LCD Display
2. Power ON/OFF Button
3. USB Communication / *Back light button
4. Secondary Display mode (for dissipation
factor(D), quality factor (Q), *phase angle (θ),
*equivalent series resistance (ESR) measurement) selection button
5. Primary Display mode (for inductance, capacitance, resistance, and *impedance measurements) / Parallel or Series measurement method selection button
6. Tolerance mode / Utility down arrow selection button
7. Hold Display mode / Record mode selection button
8. Utility menu button
9. Test Frequency / Utility up arrow select ion button
10. Relative mode / Calibration mode selection button
11. 12V DC adapter input (use with an external power adapter (rated 12VDC, 150mA, 4mm power plug))
Note: Use with included power adapter only. Use with improper power adapters may damage instrument.
13
WARNING: Before connecting an external power adapter, please check the battery compartment in the rear side of the unit. If a battery is installed, be sure that the polarity matches the (+) and (-) labels as indicated inside the battery compartment. If it is not installed correctly , please remove the battery and install it with correct matching polarity as indicated in the compartment. See “Installing Battery section for details. DO NOT, at any time, connect an external power adapter when a battery is installed incorrectly. Doing so will damage the instrument and void its warranty.
Input sockets (banana jack inputs) and terminals for positive, negative, and guard (see “Guard Terminal” in “
14
12. SUPPLEMENTAL INFORMATION” section for details)
13. Standard mini USB port (for remote controllability)
For model 879B only. Not included on model 878B.
*

Front Panel Buttons

All front panel buttons have specific colored labels on them. They are all marked in white, blue, or yellow color. Each color has a specific representation, as described below:
White – With the exception of the
button, all white colored labels represent the primary function of that button; meaning that function will be set or configured upon pressing it.
Blue – Some of the buttons have a blue label
underneath a white label. This means the function indicated by the blue label will be set or configured if that button is pressed and hold down for 2
seconds.
Yellow – There are total of 3 buttons with
yellow labels. They are
15
exclusively for use when entering UTIL menu only. See “Utility Menu” section for details.
. These functions are
16
LCR
REL
ESR
MAX
AVG
MIN
¦ÈQ
AUTO
D
1%5%10%20%
deg kHz
PAL SER
¦ÌH
p F
¦Ì
RMT
Mk
TOL
DH
n
n
@
OFF
Z
1
2 5 9 10
12
14
16
202126
27
28
83
232425
4
15
11
13
6 7
22
17 18
19
%
29

LCD DISPLAY OVERVIEW

Figure 2 - LCD Indicator Display

LCD Display Descriptions

1. LCRZ – Primary display funct ion indic ator (*Z
2. MAX – Maximum reading indicator
3. AVG – Average reading indicator
4. MIN – Minimum reading indicator
5. REL – Relative mode indicator
display)
17
6. Θ – *Phase angle indicator f or sec ondary display
7. Q – Quality factor indicator
8. ESR – *Equivalent series resistance indicator
9. – Secondary display
10. – Beeper tone indicator for tolerance mode
11. deg – *Phase angle degree indicator
12. – *ESR(ohm) units indicator
13. % - Tolerance percentage indicator
14. kHz – Frequency units indicator
15. PAL – Parallel mode indicator
16. SER – Series mode indicator
17. – Inductance units (Henry) indicator
18. – Capacitance units (Farad) indicator
19. MkΩ – Resistance units (Ohm) indicator
20. RMT – Remote mode indicator
21. – Primary display
22. D – Dissipation factor indicator
23. DH – Data hold indicator
24. AUTO – Auto-ranging indicator
25. TOL – Tolerance mode indicator
26.
27. @OFF – Auto power-off indicator
28. 1%5%10%20% - Tolerance sorting
– Low battery indicator
percentage indicator
18
29. MAX AVG MIN – Recording mode indicators
For model 879B only. Not included on model 878B.
*

Special Display Indicators

Indicates short connectors
Indicates open connectors
Error indication
Indicates calibration mode
Indicates damaged or open fuse
AD converter error
AD converter error
19

POWERING INSTRUMENT

Before beginning to operate the instrument, a power source is necessary for it to turn on. There are two methods to power the instrument: Battery and external source.

Installing Battery

The 878B and 879B LCR meters can use a battery to provide power to the instrument so that it can be portable.
The meters use a standard 9V size battery (or NEDA 1604, JIS006P, IEC6F22 carbon-zinc or alkaline battery).
To install the battery:
1. Place the meter upside down. Open up the back-flip stand, and locate the screw that tightens the battery compartment cover as indicated in Figure 3. Use a screwdriver to unscrew and remove the cover.
20
Figure 3 - Back Cover
2. Insert 9V battery into compartment. Note the positive (+) and negative (-) terminals as indicated inside the battery compartment (See Figure 4). Be sure to insert the batter y with matching polarity.
21
Figure 4 - Battery Compartment
3. Place the battery compartment cover piece by sliding it into the top slid first. Place screw at the bottom of the cover piece and tighten down with a screw driver.
4. Push and hold down the button for 2 seconds to turn on the instrument.

Connecting External Power Source

The 878B and 879B can also be powered using an external AC adapter. The model 879B comes with this adapter included in the pac kage, while it is optional
for model 878B.
22
For external power, use AC adapter rate d for output 12VDC, 150mA, with a standard 4mm connector only.
WARNING: Use of incorrect adapters may damage
the instrument. Please use B&K Precision’s adapter only.
To connect the adapter, do the following:
1. If a battery is installed, please check the battery compartment again that the polarity of the battery matches the polarity as indicated by the labels inside the compartment. If it is not, please remove and insert the battery with matching polarity. If a battery is not installed, continue to the next step.
WARNING: DO NOT, at any time, connect an external power adapter when a battery is installed incorrectly (reverse polarity or non-matching polarity to indicator of battery compartment). Doing so will damage the instrument and void its warranty.
2. Connect the AC adapter connector into the right side panel of the instrument. See Figure 5 below.
23
AC Adapter
Figure 5 - Connecting AC Adapter to Meter
3. Now, connect the AC Adapter socket into an electrical outlet.
4. Push and hold down the button for 2 seconds to turn on the instrument.
12VDC Input
Note: The meter can be operated with a battery installed while an AC adapter is plugged in at the same time (As long as the battery is inserted properly with correct polarity). In this event, the meter will automatically switch to consume power from the AC adapter instead of the battery to preserve battery life.

Low Battery Indication

The LCR meter has a low battery indicator to notify the user when to replace battery. When the display
24
starts flashing the indicator, the battery voltage is below normal working voltage. In this case, accuracy of the meter will also decrease. It is recommended that the battery be replaced as soon as possible before continuing operation. See “Installing Battery” for instructions.

Backlit Display (model 879B only)

Model 879B LCR meter has a backlit display that allows you to see the LCD display in dark conditions.
To turn on the back light, press and hold down button for 2 seconds. Back light will turn on and brighten the LCD display.
To turn off the back light at any time, press and hold down button for 2 seconds again. Back light
will turn off and return to normal display.

When Using Battery Power

When the meter is powered using 9V battery, the back light display will turn on upon holding down the
button for 2 seconds. It will stay at maximum
25
brightness for 15 seconds. After another 15 seconds (30 seconds total from the time of turning on), the back light will automatically turn off to conserve battery power.

When Using External Power

When the meter is powered using an external AC adapter, the back light display will turn on upon
holding down the button for 2 seconds. It will stay at maximum brightness continuously until the
user presses and holds down the button for 2 seconds again.
Note: If a battery is installed while using an AC adapter simultaneously, un plu ggi ng the AC ada pter will automatically turn off the back light after it has been lit for 30 seconds.
26

OPERATION INSTRUCTIONS

Data Hold

The data hold function allows the user to freeze the display when pressed, holding the measured value until data hold is turned off.

Turn On Data Hold

To use data hold, press the button once. The “DH” indicator will display on the screen when data hold is active.

Turn Off Data Hold

To disable the data hold, press again. The “DH” indicator will disappear on the screen, and meter will remain in normal operation mode.
Note: Changing the primary function, secondary function, or test frequency will automatically turn off the data hold.

Static Recording

This mode is used for recording maximum, minim um, and average values. It is often useful for testing the range of values in which a component is expected to fall within upon measurement.
27

Enable Static Recording

Press and hold down the button for two seconds to enter the static recording mode. The display should indicate “MAX AVG MIN” simultaneously. This indicates the meter is in static recording mode and recording is performed immediately.

Using Static Recording

There are four different modes that can be selected in static recording. They are indicated by the descriptions below. These modes can be changed
with each button press. Per each press of the
button, the modes will change and repeat in
the following order:
Recording Mode Maximum Mode Minimum Mode Average Mode
Recording Mode
This is the default mode when first enabling static recording. In this mode, the screen will display “MAX AVG MIN” indicator. At this point, the meter will start making recordings based on measured values from
28
Loading...
+ 108 hidden pages