BK Precision BK Precision 4085 Manual

INSTRUCTION
MANUAL
Function and Arbitrary/Function Generators
Model 4084AWG, 4086AWG, 4084, 4085, 4086, 4087
99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
d
r
g
C
d
Safety Summary
The following safety precautions apply to both operating and maintenance personnel and must be observed during all phases of operation, service, and repair of this instrument. Before applying power, follow the installation instruc tions and become familiar with the operating instruc tions for this instrument.
Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrum ent. B&K Precision Corpor ation assumes no liability for a customer’s failure to comply with these requirements. This is a Safety Class I instrument.
GROUND THE INSTRUMENT
To minimize shock hazard, the instrument chassis and c abinet must be connecte d to an electrical groun d. This instrument is grounded through the ground conductor of the supplied, three-conduct or ac power cable. The power cable must be plugged into an appr oved three-con ductor elec trical outlet. D o not alter the ground connection. Without the protective ground connect ion, all acc essible con duc tive parts (including control knobs) can render an elect r ic shock. The power jack and mating plug of the power cable meet IEC safety standards.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
KEEP AWAY FROM LIVE CIRCUITS
Instrument covers must not be removed by operating personnel. Component replacement and interna l adjustments must be made by qualified maintenance personnel. Discon ne ct the power cord before removing the instrument covers and replacing components. Under certain conditions, even with the power cable removed, dangerous voltages may exist. To avoid i njuries, always disconnect power and discharge circuits before touching them.
WARNINGS AND CAUT I O N S
WARNING and CAUTION statements denote a hazard. Follow all instructions contained in these
statements. A WARNING statement calls attention to an operating procedure, practice, or conditi on, whic h, if not
followed correctly, could result in injury or death to personnel. A CAUTION statement calls attention to an operating procedure, practice, or condition, which, if not
followed correctly, could result in damage to or destruction of part or all of the product.
WARNING:
WARNING:
CAUTION:
Do not alter the ground connection. Without the protective groun connection, all accessible conductive parts (including control knobs) can render an electric shock. The power jack and mating plug of the powe cable meet IEC safety standards.
To avoid electrical shock hazard, disconnect power cord before removin covers. Refer servicin g to qualified personnel.
Before connecting the line cord to the AC mains, check the rear panel A line voltage indicator. Applying a line voltage other than the indicate voltage can destroy the AC line fuses. For continued fire protection, replace fuses only with those of the specified voltage and current ratings.
2
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Table of Contents
1. PRODUCT INTRODUCTION................................................................................................5
1.1 Description .......................................................... .............................. ...............................5
1.2 Key features..................... ... .. ... .. ............................ ... .. ... .. ... ..............................................5
2. SPECIFICATIONS..................................................................................................................6
2.1 Function Generator... .. ... .. ........................................................ .. ... .. ... ............................. ..6
2.2 Universal Counter .. ...........................................................................................................9
2.3 General ... ............................ .. ... .. ... ....................................................................................10
3. PANEL DESCRIPTION..........................................................................................................11
3.1 Front Panel.................. ... .. ... ..............................................................................................11
3.1.1 Overview of Keys ....................................................................................................11
3.1.2 Display Annunciators..................... ... .. ... .. ............................ .. ... .. ... ..........................12
3.1.3 Description of front panel keys................................................................................13
3.1.3 Description of menu parameters ........................... .. ... .. ... .........................................14
3.2 Rear Panel.............. ............................... .......................................................... ..................17
4. OPERATING INSTRUCTIONS.................. ............................................................................18
4.1 Installation.................................... .............................. ............................... .......................18
4.2 Main operating modes............................ ... .............................. .............................. ...........18
4.3 W aveform Selection..........................................................................................................19
4.4 Data entry .........................................................................................................................20
4.5 Output Configuration..... .. ............................ ... .. ... ........................... ... .. ... ..........................21
4.5.1 Set Frequency and Period.........................................................................................21
4.5.2 Set Amplitude...........................................................................................................21
4.5.3 Set DC Offset Voltage..............................................................................................22
4.5.4 Adjust duty cycle......................................................................................................22
4.5.5 Sync/TTL Signal......................................................................................................23
4.5.6 Signal Store and Recall ............................................................................................23
4.6 Set Modulation and Sweep Parameters ............................................................................24
4.6.1. Sweep mode............................................................................................................24
4.6.2 FM modulation.........................................................................................................26
4.6.3 AM modulation ........................................................................................................27
4.6.4 Burst modulation......................................................................................................28
4.6.5 FSK modulation.......................................................................................................30
4.6.6 PSK modulation.......................................................................................................31
4.7 Set System Parameters............................... .. .............................. ... .. ... .............................. .32
4.8 Universal Counter .. ...........................................................................................................34
5. REMOTE INTERFACE REFERENCE...................................................................................35
5.1 Introduction ........... ...........................................................................................................35
5.2 Overview and syntax of SCPI instructions.......................................................................35
5.3 Detailed description of SCPI Instructions ........................................................................39
6. USER PROGRAMMABLE ARBITRARY WAVEFORM MODULE ............... ... .. ... .. ... ........47
7. APPENDIX..............................................................................................................................57
A) Declaration of conformity .......................... .. .............................. ... .............................. ... .. .58
B) Service and Warranty Information.....................................................................................57
3
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Intentionally left blank
4
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
1. PRODUCT INTRODUCTION
1.1 Description
The B+K Precision 4080 Series are laboratory grade synthesized function generators with a wide frequency range of up to 120 MHz. Direct digital synthesis (DDS) techniques are used to create stable, accurate output signals fo r clean, low distortion sine waves and an extensive selection of built-in standard and arbitrary waveform s. The ins tr ument supports AM, FM, FSK, PSK and pu lse modulation and linear and logarithmic sweep. Modulation parameters can be set precisely and are adjustable over a wide range. Th e 4080 Series supports internal an d externa l modulat ion sources as well as internal, external and gated trigger sources.
All models are capable of generating complex, predefined arbitrary waveforms. Additionally, models 4084AWG and 4086AWG provide the flexibility to create custom waveforms. The AWG module includes an intuitive, graphical Windows based software tool for creating and editing custom arbitrary waveforms and transferring the waveforms to the instrument’s non-volatile memory. The software also provides a direct interface to Tektronix® TDS1000, TDS2000 TPS2000 and TDS3000 series digital storage oscilloscopes offering users a convenient means to recreate waveforms originating from the DSO’s display or internal memory.
The 4080 Series front-panel operation is straightforward. Parameters can be entered using the knob or directly via the numerical keypad and unit keys.
The instruments are fully programmable via the standard RS232 interface. The command set is SCPI (standard Commands for Prog rammable Instruments) compatible.
The combination of classical function and arbitrary waveform generator makes this series a versatile solution for many applications in Electronic Test and Design, Sensor Simulation, Education and Training.
1.2 Features
Direct Digital Synthesis (DDS) architecture
Wide frequency range of 1µHz ~ 120MHz (model 4087, sine wave only)
• Clean and stable output of very small signals down to 1mV (50)
27 build-in standard and complex waveforms.
Eight downloadable 16000 point memories for custom arbitrary waveforms (models
4084AWG and 4086AWG only)
Graphical Arbitrary Waveform Generation Software tool for Microsoft
(models 4084AWG and 4086AWG only)
Convenient data input via knob or numerical keypad.
Bright, easy to read display using VFD (Visible Vacuum Fluorescent) technology
Fully programmable via SCPI compatible command set
100 MHz Universal Counter with frequency measurement and totalize function
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
5
®
WindowsTM
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
2. SPECIFICATIONS
2.1 Function Generator
Waveform Characteristics
Main Waveforms
W aveform Am plitude resolution: 12 bits Sample Rate: 200MSa/s (4084, 4084AWG, 4085, 4086, 4086AWG)
Sine
:
Square
Rise and Fall Time of Square Wave: 15ns
Note: Test conditions for harmonic distortion, sine distortion, rise/fall time: Output Amplitude 2Vp-p, Environmental temperature: 25℃±5℃
Build in standard and complex (arbitrary) waveforms 27 build-in standard and complex waveforms.
Waveform Length: 4096 dots
Amplitude Resolution: 10 bits
Pulse W ave :
DC signal characteristics:
Module for user defined arbitrary waveform generation (models 4084AWG and 4086AWG only)
Number of memory locations for arbitrary waveforms: 8 Length of waveforms8~16 000 points Resolution of wavefo rm a m p litude10 bits Frequency range:100µHz~100kHz Sample rate:200MSa/s
: Sine, square
300MSa/s (4087)
Harmonic Distortion of Sine Wave:
- 50dBc (frequency 5MHz) - 45dBc (frequency 10MHz) - 40dBc (frequency 20MHz) - 35dBc (frequency 40MHz) - 30dBc (frequency > 40MHz)
Total Harmonic distortion: 0.1% (20Hz ~ 100kHz)
:
:
Sine, Square, Triangle, Positive Ramp, Falling Ramp, Noise, Positive Pulse, Negative Pulse, Positive DC, Negative DC, Stair wave, Coded Pulse, Full wave rectified, Half-wave rectified, Sine transverse cut, Sine vertical cut, Sine phase modulation, Logarithmic, Exponent ial, Half-round, SINX/X, Square root, Tangent, Cardiac, Earthquake, Combination
Duty Cycle: 0.1% ~ 99.9% (below 10kHz), 1% ~ 99% (10kHz ~ 100kHz) Rise/Fall Time: 100ns Duty cycle 20%
DC range: 10mV – 10V (high impedance) DC Accuracy: ±5% of setting +10mV (high impedance)
6
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Frequency Characteristics
Frequency Range:
Main waveforms (sine, square)
: Model 4084/4084AWG: 1µHz ~ 20MHz Model 4085: 1µHz ~ 40MHz Model 4086/4086AWG 1µHz ~ 80MHz (sine wave)
1µHz ~ 40MHz (square wave)
Model 4087 1µHz ~ 120MHz (sine wave)
1µHz ~ 40MHz (square wave)
All other waveforms
:
All models: 1µHz ~ 100kHz
Frequency Stability: ±1×10
-6
(22℃±5℃)
Resolution: 1µHz Frequency Accuracy: ± 5×10
-6
(22℃±5℃)
Data entry Units: s, ms, Hz, kHz, MHz
Amplitude Characteristics
Amplitude Range:
4084, 4084AWG: 2mV ~ 20Vpp (open circuit), 1mV ~ 10Vpp (50) 4085
4086, 4086AWG: for Freq 40MHz: 2mV ~ 20Vpp (open circuit), 1mV ~ 10Vpp (50)
for Freq > 40MHz: 2mV ~ 4Vp-p (open circuit), 1mV ~ 2Vpp (50)
4087: for Freq  40MHz: 2mV ~ 20Vpp (open circuit), 1mV ~ 10Vpp (50)
for Freq > 40MHz: -76dBm ~ +13.5 dBm(50or 0.1mV ~ 3Vpp(50
Max. Resolution: 2µVpp (open circuit), 1µVpp (50)
Amplitude Accuracy: ± 1%+0.2mV (sine wave relative to 1kHz) Amplitude Stability: ±0.5 % /3 hours
Flatness:
For amplitude 2Vpp: ±3% (frequency5MHz), ±10% (5MHz<frequency40MHz) For amplitude >2Vpp: ±5% (frequency5MHz), ±10% (5MHz<frequency20MHz)
±20% (frequency>20MHz)
Models 4086/AWG, 4087 only: ±1dBm (frequency>40MHz)
Output Impedance: 50 Output Units: Vpp, mVpp, Vrms, mVrms, dBm
7
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
DC Offset Characteristics
Offset Range (open circuit)
Freq 40MHz): ±10Vpk ac + dc (Offset 2×peak-to peak amplitude) Freq > 40MHz): ±2Vpk ac + dc (Offset 2×peak-to peak amplitude)
Resolution: 2µV (open circuit), 1µV (50)
Offset Error: ±5% of setting +10mV (Ampl. 2Vpp into open circuit)
±5% of setting +20mV (Ampl. > 2Vpp into open circuit)
AM Characteristics
Carrier Waveforms: sine or square Carrier Frequency Range: same as main waveforms Modulation Source internal or external Modulating Waveform: 5 internal waveforms (sine, square, triangle, rising/falling ramp) Frequency of modulating signal: 100µHz ~ 20kHz Distortion: 2% Modulation Depth: 1% ~ 120%
1% ~ 80% (frequency>40MHz, Ampl > 2Vpp into open circuit)
Modulation Error: ± 5%+0.2% (100µHz < frequency  10kHz)
±10%+2% (10kHz < frequency 20kHz)
Amplitude of ext. input signal: 3Vp-p (-1.5V~ +1.5V)
FM Characteristics
Carrier Waveforms sine or square Carrier Frequency Range: same as main waveforms Modulation Source: internal or external Modulating Waveform: 5 internal waveforms (sine, square, triangle, rising/falling ramp) Frequency of modulating signal: 100µHz ~ 10kHz Peak Frequency Deviation: Max. 50% of carrier frequency for internal FM
Max 100kHz (carrier frequency5MHz) for external FM, with input signal voltage 3Vp-p ( -1.5V~+1.5V)
FSK Characteristics
Carrier Waveform sine or square Carrier Frequency Range: same as main waveforms Control Mode internal or external trigger (external: TTL level, low level F1,
high level F2)
FSK Rate: 0.1ms ~ 800s
PSK Characteristics
Waveform: sine or square Frequency Range: same as main waveforms PSK: Phase 1 (P1) and Phase 2 (P2), range: 0.0 ~ 360.0° Resolution: 0.1° PSK rate: 0.1ms ~ 800s
8
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Control Mode: internal or external trigger (external: TTL level, low level P1,
high level P2)
Burst Characteristics
Waveform: sine or square Frequency Range: same as main waveforms Burst Counts : 1 ~ 10000 cycles Time interval between bursts: 0.1ms ~ 800s Control Mode: internal, single or external gated trigger
Frequency Sweep Characteristics
Waveform: sine or square Start/Stop Freq.: same as main waveforms Sweep Time: 1ms ~ 800s (linear), 100ms ~ 800s (log) Sweep Mode: Linear or Logarithmic External trigger signal frequency: DC ~ 1kHz (linear) DC~10Hz (log) Control Mode: internal or external trigger
Rear Panel Terminals (for modulation and sweep)
Output MOD OUT
Frequency: 100µHz ~ 20kHz Waveform: sine, square, triangle, rising/falling ramp Amplitude: 5Vp-p ± 5% Output Impedance: 600
Modulation IN
3 Vpp = 100% Modulation
External Input Trig/FSK/Burst
Level: TTL
Main OUTput
Impedance: 50 Protection: Short circuit and overload protected
State Storage Characteristics
Storage Parameters: frequency, amplitu de, waveform, DC offset values, modula tion
parameters Storage Capacity: 10 user configurable stored states Storage Time: more than 10 years
2.2 Universal Counter
Frequency Range
Frequency Measurement: 1Hz ~ 100MHz Totalize mode: 50MHz max
9
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Input Characteristics
Sensitivity
Input attenuator disabled: 50mVrms (f: 10Hz ~ 50MHz), 100mVrms (f: 1Hz ~ 100MHz) Input attenuator enabled: 0.5Vrms (f: 10Hz ~ 50MHz), 1Vrms (f: 1Hz ~ 100MHz)
Max. Input Voltage Allowed: 100Vp-p (f100kHz), 20Vp-p (1Hz~100MHz) Input Impedance: R>500k, C<30pF Coupling: AC Waveform: sine or square Low Pass Filter: cut off frequency about 100kHz (with internal attenuation: -3dB) with external attenuation:  -30 dB (f >1MHz) Gate Time Setting: 10ms ~ 10s continuous ly adjustable Display Bits: 8 (for gate Time>5s) To talize Capacity: 4.29×10
9
Control Mode: manual or external gate control Accuracy: time base error ± trigger error (when signal SNR > 40dB,
trigger error 0.3) Time base:
Type: small TCXO Frequency: 10MHz Stability: ±1 × 10
-6
(22°C±5°C)
2.3 General
Power Supply 198~242V or 99~121V, Frequency: 47~ 63Hz Power Consumption: <35VA Operating Te mperature
0° to +40 °C
Operating Humidity 80% R.H
Storage Temperature
-40 to 70
Dimensions (W x H x D): 255 mm x 100 mm x 370 mm (10.0 x 3.93 x 14.56 inch) Weight: 3 kg (6.6 lbs): models 4084, 4084AWG, 4085, 4086, 4086AWG
3.5 kg (7.7 lbs) model 4087 Remote Interface RS232
Accessories included BNC to alligator cable BNC to BNC cable RS232 communication cable Power line cord Test report Spare fuse Installation disk for Arbitrary Waveform Creation Software (models 4084A WG or 4086AWG only)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
10
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
3. PANEL DESCRIPTION
3.1 Front Panel
3.1.1 Overview of Keys:
Summary: Most keys have multiple functions. Primary functions are written on the key. Sim ply press the respective key to enable a primary functions. All Function/Mode keys and some of the numerical entry keys have associated secondary functions indicated above each key in blue. To activate a secondary function, press the 【shift】key followed by the desired key. The bottom row of the Function/Mode keys can be used to enter units directly. The unit is indicated below each key. To enter a unit, enter a numerical value via the numerical keypad then press the corresponding unit key. The unit key also serves as Enter function.
Data entry keys
Key
Name
0 Input Digit 0 Enter SW mode 7 Input Digit 7
1 Input Digit 1
2 Input Digit 2
3 Input Digit 3
4 Input Digit 4
5 Input Digit 5
6 Input Digit 6
*: Direct number entry: Press this key to clear the least significant bit of the displayed number.
Useful for correcting number entry before entry is confirmed with unit key. External totalize mode: Press this key to stop counting and display present counting value. Press again to resume counting.
:
Main
Function
Secondary
Function
Enable ARB1 waveform *** Enable ARB2 waveform *** Enable ARB3 wave *** Enable ARB4 waveform*** Enable ARB5 waveform *** Enable ARB6 waveform***
Key
Name
Main function
Enable ARB7 waveform ***.
8 Input Digit 8
9 Input Digit 9 Not available
Input decimal point Reset Unit
Input negative symbol Move arrow key to left * Move arrow key to right **
Enable ARB8 waveform ***
Enter system menu
Select pulse Select Arblist
waveform
Secondary
Function
11
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
**: External totalize mode: Press this key to reset and resume event counter *** models 4084AWG and 4086AWG only
Function/Mode Keys:
Key name Main Function
Freq./Period
Ampl./Pulse
width
FSK/PSK
Menu Menu Selection
FM Enable FM mode
AM Enable AM mode Enter Recall menu Low Pass Select MHz, Vrms
Sweep
Burst
Other Keys
Key Name Main Function Other Function
Output Main OUTPUT signal On/Off Shift Select secondary function Enter units in s, Vpp, N
:
Toggle between Frequency & Period.
Amplitude Select. FSK/PSK
Function Select
Enable Sweep mode Enable Burst mode
Secondary
Function
Enable Sine Wa ve Not A vailable Not A vailable
Enable Square Wave. Enable Triangle Wave Enable positive ramp Enter Storage menu
Enter Counter Mode
DC Offset Select Gate Select Hz, dBm
Secondary
Function for
Counter Mode
Not A vailable Not A vailable
Not A vailable Not A vailable
Not A vailable Not A vailable Attenuation
Selection
Freq. Meas./ Totalize Enable
Generate single trigger in sweep and burst mode
3.1.2 Display Annunciators:
Unit Entry
ms, mVpp
kHz, mVrms
Display areas:
Waveform Indication Main Alphanumeric Display Frequency Measurement/Totalize Settings Instrument states
12
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Waveform states
Sine waveform is enabled (,main waveform) Square waveform is enabled (main waveform) Triangle waveform is enabled Ramp waveform is enabled
Arb Arbitrary waveform mode is enabled
Frequency Measurement/Totalize states Filter: Low Pass Filter is enabled ATT: Input Attenuator is enabled GATE: Gate received trigger
Function Generator states Adrs: The instrument is in remote state. Trig: Function Generator is waiti ng for a single trigger or external trigger. FM: FM modulation is enabled AM: AM modulation is enabled Sweep: Sweep mode is enabled Ext: Generator is configured for external signal input Freq: Frequency measurement function is enabled Count: Totalize function is enabled FSK: FSK (frequency shift keying) modulation is enabled FSK: PSK modulation is enabled Burst: Burst mode enabled. Offset: DC offset of output signal is not 0. Shift: Shift key has been pressed, Shift mode is active Rmt: The instrument is in remote state.
3.13 Description of front panel keys
shiftkey
Access secondary functions or to enter units “sec onds”, Vpp and “N”
Numerical key pad:
Keys are used for direct entry of numerica l values. Keys【0】【】【-】are assigned secondary functions to enter the systems menu, reset the unit to its default values or revert to Standard Waveform (“SW”) mode.
】【arrow keys
The primary function is to move th e flas hing di git lef t or right or to s elec t the desired ar bitrary waveform from the Arb List (secondary function). When in Counter Mode, these 2 keys are used to start/stop or reset/resume the counter.
Freq/Periodkey:
Toggle between frequency and period display (primary function) or to enable sine waveform.
Ampl/Pulse Widthkey:
Display and adjust the amplitude of waveforms or, when in pulse mode, toggle the display
13
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
between amplitude and pulse width. The secondary function activates the square waveform
FSK/PSKkey:
Toggle between FSK and PSK modulation (primary function). Activate triangle waveform
Menukey:
Enter modulation parameters for FSK, PSK , FM, AM and burst modulation and for sweep mode. When in standard wa veform mode (no modulation, main waveforms enabled), this key can also be used to toggle the units for the amplitude display value bet w een Vpp, Vrms and dBm. (Press the Amplitude key first, then the Menu key to toggle between the units). The secondary function enables the ramp waveform.
FMkey:
Activate FM modulation (primary function). Enter Storage mode (secondary function). Enter units ms or mVpp after entering the desired value by nu merical key pad. In Coun ter Mode, this key turns the input attenuator o n or off.
AMkey:
Activate AM modulation (primary function). The secondary function is used to recall and recreate signals stored in status memory. Enter units “MHz and “Vrms” after entering the desired value by numerical key pad. Enable the low pass filter when in Counter mode.
Sweep key:
Activate sweep mode. Select frequency measuring and totalize mode (secondary function) . Enter units kHz or mVrms after entering the desired numerical value directly via the keypad. Use the Shift key to toggle between frequency measurements and totalize mode.
Burstkey:
Activate burst mode (primary function). Enter DC offset mode (secondary function). Enter units Hz or dBm. When in frequency measurem en t m ode, press t his key to enter the g ate tim e.
Outputkey
Press this key to toggle the main OUTPUT signal between the ON and OFF state. By default the output is turned on, indicated by the green LED and the currently active wave form is available at the OUTPUT terminal. In Burst or Sweep m ode, this key is also used to generate a single trigger.
3.14 Description of menu parameters
Use this key to configure modulation parameters, sweep mode parameters and system parameters. Modulation and Sweep mode: After enabling modulation or sweep mode, press “menu” to configure the related parameters. Each time you press the menu key, the parameter will flash for 1 second, followed by the currently act ive value of that param eter. Use the knob or numerical keys to enter a new value. Once the parameter is set, press 【menu】 to advance to the next parameter. Continue pressing the menu key to cycle throug h all parameters. Press 【Shift】【SW】to return to the main waveform mode and to set the carrier waveform parameters.
Sweep Mode:
MODE > START F > STOP F > TIME >TRIG
14
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
MODE: Select LINEAR or LOGarithmic sweep START F :Sweep start frequency STOP F: Sweep stop frequency TIME: Sweep time TRIG: Select trigger source, INTernal or EXTernal
FM modulation:
FM DEVIA> FM FREQ > FM WAVE > FM SOURCE
FM DEVIA: Peak frequency deviation FM FREQ: Modulating signal frequency FM WAVE: Modulating signal waveform (sine, square, triangle rising or falling ramp) FM SOURCE: Toggle between internal and external modulating signal.
AM modulation:
AM LEVEL > AM FREQ > AM W AVE > AM SOURCE
AM LEVEL: Modulation depth AM FREQ: Frequency of modulating signal AM WAVE: Modulating signal waveform (sine, squar e, triangle rising or falling ramp) AM SOURCE Select internal or external modulating signal
Burst modulation:
TRIG > COUNT —> SPACE T > PHASE
TRIG Select trigger source, internal or external COUNT: Number of burst cycles SPACE T:Burst time spacing PHASE: starting phase of the burst
FSK modulation:
ST ART F > STOP F> SPACE T > TRIG
START F Primary frequency (same as carrier wave) STOP F the second frequency (hop frequency) SPACE T FSK rate TRIG Select trigger source, external or internal
PSK modulation
P1 > P2 > SPACE T > TRIG
P1 phase value #1 P2: phase value #2
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
15
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
SPACE T: PSK rate TRIG: trigger mode, internal or external
System Function Mode:
POWER ON > ADDRESS > OUT Z > INTERFACE>…..
…………BAUD > PARITY
POWER ON “Power on” state ADDRESS: set GPIB address (option) OUT Z: Configure amplitude display valu e for 50 or high impedance load termination INTERFACE: Select RS232 or GP-IB (IEEE-488) interface (option) BAUD: Baud rate for the RS232 interface PARITY: Parity and Data Bits configuration for RS232
16
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
3.2 REAR PANEL
Ext. Trig/FSK/Burst :
Input Terminal for external trigger signals for FSK/PSK Burst modulation and sweep mode
MOD In
Apply modulating signal for AM and FM to this Input terminal
Meas Freq/TOT In
Input terminal for Universal Counter which operates in frequency measurement or Totalize mode
MOD OUT
The internally generated modulating signal when in AM mode will be available at this output
RS232C
RS232 interface for remote control of instrument (all models) or for download of custom arbitrary waveforms (models 4084AWG and 4086AWG only)
AC Socket and fuse co mpartment
Connect the supplied power line cord to this receptacle. Make sure to install the appropriate fuse according to the selected AC voltage.
AC Input Selector switch
Set this switch to the corresponding AC voltage in your area
17
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
4. OPERATING INSTRUCTIONS
4.1 Installation
This section contains installation information, power requirements and initial inspection and signal connections for the 4080 series generators.
Mechanical Inspection This instrument was carefully inspected before shipment. Upon receipt, inspect the instrument for damage that might have occurred in transit. If any sign of damage is found, notify your B+K Precision distributor .
Instrument Mounting The 4080 Series is intended for bench use. The instrument includes a handle adjustable for optimum panel viewing angle. The instrument does not require special cooling when operated within conventional temperature limits.
Power Requirements The 4080 Series can be operated from any source of 99V to 242V AC and frequency of 48Hz to 63Hz. The maximum power con sumption is 35 VA. Replace fuses with the same type, according to the rating indicated on the rear panel of the instrument. The instrument power fuse is located in a fuse compartment below the AC input receptacle. To access the fuse, first disconnect the power cord and then remove the fuse cartridge.
Power-on procedure Turn on the instrument by pressing the power key on the front panel of the unit. The display should be flashing “BK PRECISION” for 2 seconds followed by the model number. e.g. .“4084AWG” for 1 second. By default, the instrument will enter the standard waveform (“SW”) mode with the frequency set to 10.00000000 kHz and the waveform annunciator displaying the “~ “ symbol. If the Power ON configuration in the systems menu was modified, the parameters of th e last operat ion before power-down will be displayed.
4.2. Main operating modes
“SW” standard waveform mode In this mode, the generator outputs any of the 27 build-in waveforms, including the 2 “main waveforms” sine and square. This also includes the 8 user programmable arbitrary waveform s (models 4084AWG and 4086AWG only). For most waveforms, you can set frequency, amplitude and DC offset. In this mode, modulation and sweep is not active and all annunciators in status area 4 are turned off . When modulation or sw ee p is en abled, press【shift】【SW】to return to this mode.
Modulation, Pulse and Sweep Mode In this mode, the generator operates like a conventional function generator and the Arb annu nci ator is turned off. The user can selec t one of t he modu lation m odes AM, FM, FSK, PSK, Burst or sweep mode. The 2 main waveforms sine and square can be used as carrier signal.
There is a close, reciprocal relationship between the modulation/sweep mode and the standard
18
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
waveform mode used to set the sine or sq uare wave parameter s. When one of the modulation m odes is active, the parameters of the carrier wave are carried over from the Standard Waveform Settings (for sine or square). Press 【shift】【SW】to toggle between any of the m odulation/sweep m odes (to set the modulation parameters) and the standard waveform mode (to define the carrier settings for frequency , am plitude a nd DC of fset). On the ot her ha nd, the car rier frequ ency c an also be set in th e modulation/sweep menus and if changed, will overwrite the standard waveform frequency setting.
Example: When transitioning f rom standard wavefor m mode to FM modulation mode , (by pressing the FM button) the parameters set previous ly in “SW” mode car ry over to the F M mode and are now defining the FM carrier. If the FM carrier frequency in the FM modulation menu is changed, it will also “overwrite” the frequency setting of the standard waveform” settings.
To adjust the duty cy cle, enable pulse m ode by pressing 【shift】 】. Note that the duty cycle can only be adjusted in pulse mode.
4.3 W aveform Selection:
Enabling common waveforms Press【shift】 followed by the waveform key to select one of the 5 common wavefo rms sine, square,
triangle, ramp and pulse. The corresponding waveform annunciator will be displayed in the waveform display area. Note that the instrument operates in Function Generator mode when selecting sine and square and in ARB mode for the remaining functions. (Arb annunciator is lit). Note: The 5 common waveforms can also be output by selecting the corresponding number from the ArbList in this section. Example: Press 【Shift】【FSK/PSK】to enable the triangle waveform
Enabling stored waveforms from the Arb List Make sure you are in “SW” mode. (All area 4 status annunciators are of f) . If modul ation or sweep is active, pressshift】 【SWto return to the standard waveform mode. Press【shift】 】to enter the ArbList: W aveform,”6: NOISE” is enabled and appears on the display. Use the knob or the numeric keypad to select one of the 27 stored waveforms from the ArbList according to the table in this section. Example: To select negative ramp DOWN_RAMP, do the following: Press 【shift】 【】(enter ArbList)
Press 【5】【N】(enable DOWN_RAMP)
or
select waveform with the knob
Models 4084AWG and 4086AWG only: Locations 28-35 are reserved for the storage of user defined arbitrary waveforms. The display name for these waveforms is ARB1-ARB8. These memory locations can be accessed in one of 2 ways: a) Enter the ArbList, then use the knob or the numeric keypad to enable waveforms ARB1 – ARB8 b) Press 【Shift】followed by any number between 1-8. Example: Press 【Shift】【2】to enable waveform stored in memory location ARB2
19
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Table of stored waveforms (ArbList) and their memory locations
No. Waveform Display Name No. Waveform Display Name
1 Sine wave SINE 15 Half-wave rectified COMMUT_HA 2 Square wave SQUARE 16 Sine transverse cut SINE_TRA 3 Triangle wave TRIANG 17 Sine vertical cut SINE_VER 4 Ramp UP_RAMP 18 Sine phase modulation SINE_PM 5 Falling ramp DOWN_RAMP 19 Logarithmic function LOG 6 Noise NOISE 20 Exponential function EXP 7 Pulse wave PULSE 21 Half-round function ROUND_HAL 8 Positive pulse P_PULSE 22 SINX/X function SINX/X
9 Negative pulse N_PULSE 23 Square root function SQU_ROOT 10 Positive DC P_DC 24 Tangent function TANGENT 11 Negative DC N_DC 25 Cardiac wave CARDIO 12 Stair wave STAIR 26 Earthquake wave QUAKE 13 Coded pulse C_PULSE 27 Combinati on wave COMBIN
14
Full wave rectified
COMMUT_FU 28~35*
User programmable Arbitrary waveforms
ARB1~ARB8
Note:
*No 28~35: memory location for user programmable waveforms, models 4084AW G and 4086AW G only. Refer to chapter 7 for more details
4.4 Data entry
Using the arrow keys and the knob
Use the knob and arrow keys to modify the displayed number.
Use the 】【】 keys to move the flashing digits left or right then adjust the value with the knob. Using this method of entry, the output signal will be updated immediately. Move the arrow keys left for coarse adjustment and right for fine adjustment.
To disable the knob, use the】【】 key to move the cursor all the way to the left or right until the digits stop flashing. Now data entry via the knob is disabled.
Direct entry using the numerical key pad
Use the numerical keypad to enter a number with the appropriate unit.
Enter numbers from left to right. Use the 】key to enter a decimal point. Enter 【-】for negative numbers. Repeatedly pressing this key will toggle between positive (no sign visible) and negative numbers. Numerical entries do no t updat e the o utput si gnal until a unit k ey has bee n pressed. O nce you entered the correct numerical value, press the appropriate unit key to assign a unit and to make the entry effective. The instrument will now output a signal according to the displayed data. For entries not associated with a unit, press the shift key (“N” = “no unit”) to make the entry effective.
Note: Numerical values entered via the keypad ar e not effective has been pressed.
20
until a valid unit key or shift (“N”)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Entry of invalid key presses or invalid values If the entered value exceeds the rated range, a beep sound will be heard. If the entered value is below the lower limit, the instrument will automatically change the entry to the lowest possible value. If the entered value exceeds the upper limit, the instrument will automatically revert to the maximum value. Example: When trying to entering 90MHz in a model 4086 (80 MHz max), a beep will be heard and the value will be forced to 80 MHz.
Invalid keys: A beep sound will inform the user that the key entry is invalid. The instrument will simply ignore the key pressed. Example: Key【-】is pressed when trying to enter a frequency value. The instrument will ignore the entry and respond with a “beep” sound.
4.5 Output Configuration
4.5.1 Set Frequency and Period
The【frequency/period】key is used to toggle between Frequency and Period display of the standard waveform or the carrier wa veform.
Frequency Press the 【frequency】 key to display the current frequency val ue. The value can be m odified using
the numerical keypad or the knob. Example: To set a frequency value of 5.8 kHz, enter the following key sequence:
The display will be 5.80000000 kHz.
Period Setting The signal frequency can also be displayed or entered as a period value. If the current display is frequency, press the【frequency/period】key to display the current period value. Values can be entered with the numerical key pad or using the knob. Example: To set a period value of 10ms, enter the following sequence of keys:
4.5.2 Set Amplitude
Press the 【Amplitude】key to display the current amplitude value. Modify the value using the knob or the numerical keypad. Example: Set amplitude to 4.6V peak-to-peak:
frequency】【5】【】【8】【kHz
or
frequency】【5】【8】【0】【0】【Hz
or
use the knob and 】【】 keys
:
Period】【1】【0】【ms Or
use the knob and】【】 keys
PressAmplitude】【4】【】【6】【Vpp】 or modify values using the knob and arrow keys
21
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
In case of standard waveforms “sine”, “square”, “triangle”, “rising ramp” and “pulse”, num bers can be entered and displayed as Peak-to-peak value (Vpp or m Vpp) , root me an square value (Vrms and mVrms) or dBm value. All other waveforms can only be edited or displayed using Vpp or mVpp units only.
4.5.3 Set DC Offset Voltage
Pressshift】【offset】to display the current DC offset value. If the current DC offset value is not equal to zero, the annunciator “Offset” will turn on. The DC offset value can be entered directly or using the knob. Example: Set an offset value -1.6V peak-to-peak
Pressshift】【offset】【-】【1】【】【6】【Vpp】
or shift】【offset】【1】【】【6】【-】【Vpp or use the knob for value entry
Zero Point Adjustme nt For zero point adjustment of the output signal, using the knob is more convenient than direct entry via the numerical keypad. The transition of the DC offset Voltage from plus to minus sign will be automatic when passing through the zero point. The input range of the signal amplitude and DC offset should satisfy the following equation: |Voffset| + Vpp/2 Vmax, with the parameters defined as followed:
Vpp is the peak-to-peak value of the amplitude
|Voffset| is the absolute value of the DC offset
Vmax is 10V at high impedance and 5V at 50Ω load.
The following table shows the corresponding relationship between the Vp-p value of the amplitude and the absolute DC offset value at high impedance:
1.001 V ~ 20.00 V 0 ~ (10.000-Vpp/2) V
316.1 mV ~ 1.000 V 0 ~ 2.000 V
100.1 mV ~ 316.0 mV 0 ~ 632.9 mV
31.01 mV ~ 100.0 mV 0 ~ 200.9 mV
2.000 mV ~ 31.00 mV 0 ~ 62.99 mV
4.5.4 Adjustment of duty cycle
If the current waveform is pulse, and the current display value is amplitude, use the 【Ampl/Duty】 button to toggle between Amplitude and Duty cycle display. When the pulse width is displayed, enter a value via the numeric keypad or the knob. The valid range is 0.1% ~ 99.9% for frequencies below 10kHz with a maximum resolution of 0.1%. For frequencies between 10kHz~100kHz the range is 1% ~ 99% and the maximum resolution is 1%. Example: Enter a duty cycle value of 60.5%
Press Pulse】【6】【0】【】【5】【N】 or use the knob and arrow keys
:
Vp-p value of AC signal Absolute value of DC offset
22
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
4.5.5 Sync/TTL Signal
A TTL signal output is provided on the front-panel SYNC terminal. All signals, including modulated signals and arbitrary waveforms (except DC and noise) have an associated TTL signal.
This signal is a TTL “high” when the waveform’s output (on the main terminal) is positive, relative to zero volts. The signal is a TTL “low” when the output is negative. The TTL signal is generated by passing the main output signal through a comparator configured as a Schmitt Trigger. The TTL signal will transition to high or low once the main output signal exceeds a certain threshol d value, which causes a time delay between the main output signal and the TTL signal.
4.5.6 Signal Store and Recall
Up to 10 different instrument states can be stored in non-volatile memory. T his enables you to rec all the entire instrument configuration with a single command from the remote interface or with just a few key presses from the front panel.
The state storage feature “remembers” the function (including arbitrary waveforms), frequency, amplitude, DC offset, duty cycle, as well as any modulation param eters. To recall a stored state, you must use the same memo r y l ocation used previously to store the state.
The last state of the instrument before power-off is automatically stored in location #0, therefore a total of 11 groups locations ranging from 0 ~ 10 can be recalled.
Example: Store the current output signal in group location #1: Shift】【Store】【1】【N The following prompt will be displayed for a few seconds: “STORE 1” Previously stored signal configurations will be overwritten.
To recall group location #1 and make it the active output signal: PressShift】【recall】【1】【N】
The following prompt will be displayed for a few secon ds: “RECALL: 1”
The stored signals can be continuously recalled and reproduced by scrolling through each location using the knob.
23
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
4.6 Set Modulation and Sweep Parameters
4.6.1. Sweep mode
In the frequency sweep mode, the function generator “steps” from the start frequency to the stop frequency at a sweep rate which you specify. You can sweep up or down in frequency, and with either linear or logarithmic spacing. Y ou can also configure the function generator to output a single sweep (one pass from start frequency to stop frequency) by applying an external trigger. The function generator can produce a frequency sweep for sine or square waveforms.
Press Sweep to enable the frequency sweep mode. The display shows a preset frequency (start frequency) and the “Sweep” symbol appears i n the state display area. The carrier frequency will be displayed and the frequency, amplitude, waveform and DC offset of the carrier signal can be set as described in previous sections of this chapter. The main functions sine or square wave can be selected as a carrier signal.
Press the 【menu】key repeatedly to cycle through the sweep menu parameters listed below.
MODE > START F > STOP F > TIME >TRIG
MODE Select LINEAR or LOGarithmic sweep mode START F sweep start frequency STOP F: sweep stop frequency TIME: sweep time TRIG: Select trigger source INTernal or EXTernal
Each time you press the menu key, the parameter will flash for 1 second, followed b y the value of that parameter. Sweep mode parameters can be entered via numerical keypad or the knob. Once the parameter is set, press 【menu】 to advance to the next parameter.
Sweep mode MODE parameter: Select LINear (No.1) or LOGarthimic (No.2). In linear sweep mode, the output frequency changes in a linear fashion during a sweep, whereas in LOG mode the frequency changes exponentially. The spacing is calculated automatically based on start frequencies, stop frequencies and sweep time.
Start frequency “START F” The frequency where the sweep starts is the start frequency. After displaying “START F” for 1 seco nd, the curr ent start frequency is displayed automatically and can be modified via knob or direct data entry. By default, the start frequency will carry over from the setting for the main wave form
. Stop frequency “STOP F” The sweep stops at the frequency “STOP F” When the start frequency is lower than the stop frequency , the fre que ncy sweep incr eases gradu ally from start frequency (low frequency) to stop frequency (high frequency); When the start frequency is higher than stop frequency, the frequency sweep decreases gradually from the start frequency (high frequency) to stop frequency (low frequency).
:
24
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
The frequency range from start f requency to stop frequency is 1µHz ~ Fmax in linear sweep mode and 1mHz ~ Fmax in log sweep mode. (Fmax see specification sect ion for main waveform)
Sweep time “TIME” The time needed to complete one sweep from start to stop frequency is called sweep time. The sweep time range is 1ms ~ 800s in linear sweep mode or 100ms ~ 800s in LOG sweep mode.
Trigger mode “TRIG”
: A sweep can be triggered internally or externally. The corresponding display values are “1: INT” and “2: EXT”. The default value is INTernal trigger . In internal trigger mode, the instrument will continuously sweeps from the start frequency to the stop frequency according to the Sweep parameters. An external trigger signal can be generated in one of two ways. a) Press the 【output】 key to trigger a single sweep. The signal frequency will in/decrease fr om start frequency to stop frequency, then sweep will stop. b) Apply an input trigger signal to the “EXT Trig” terminal on the rear panel. A rising edge will trigger a single sweep. In external trigger mode, symbols “Trig” and “Ext” are displayed in the status field.
Start and stop of sweep The sweep will start automatically once sweep mode is selected. If you don’t want to output the sweep signal, simply press the 【output】key to disable the signal output (LED above Output button is OFF). Once all the parameters are set, you can make the sweep signal available at the output by pressing the Output button again. In external trigger mode, the 【Output】key functions only as a single pulse trigger key, the Output On/Off functionality is now disabled and the output LED will always be lit.
Sweep configuration example Sweep parameters are as followed: Frequency range 100Hz~200kHz, sweep time 10s, linear sweep, internal trigger mode. Set the carrier frequency parameters
Presssweep Pressmenu Select “MODE”, wait, then enter【1】【N】(linear sweep) Pressmenuto select “START F”, press【1】【0】【0】【Hz】 Pressmenuto select “STOP F”, press【2】【0】【0】【kHz】 Pressmenuto select “TIME, press【1】【0】【s】 Pressmenu to select “TRIG”, press【1】【N】
Hint: The fr equency displayed immediately after pressing [Sweep] is the start freq uency. Any changes to that value will automatically update parameter “START F”
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
25
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
4.6.2 FM modulation
A mod ulated waveform consists of a carrier waveform and a modulating waveform. In FM, the frequency of the carrier is varied by the amplitude of the modulating waveform. The function generator will accept an internal or external FM modulating signal.
Press the 【FM】key to enter into FM mode. The display will indicate the carrier frequency and the “FM” annunciator will be lit. The frequency, amplitude, waveform and DC offset of the carrier signal can be set as described in th e previous section. Only the main functions sine or square wave can be selected as a carrier signal.
Press the 【menu】 key repeatedly to cycle through the FM modulation parameters listed below.
FM DEVIA > FM FREQ > FM WAVE > FM SOURCE
FM DEVIA Peak frequency deviation FM FREQ: Frequency of modulating signal FM WAVE: Waveform of modulating signal FM SOURCE Toggle between internal or external modulating signal
Each time you press the menu key, the parameter will flash for 1 second, followed by the val ue of that parameter. FM modulation parameters can be entered directly or via the knob. Once the parameter is set, press 【menu】 to advance to the next parameter.
Peak frequency deviation ”FM DEVIA” Explanation: The variation in freq uency of the modulating waveform from the carrier freque ncy (center frequency). Range of deviation: For internal FM modulation, th e maximum value s hould not exceed 50 % of the carrier frequency. In external FM mode, the maximum deviation is 10% of the carrier frequency. Additionally, the frequency deviation plus carrier frequency should not excee d the maximum operating frequency of the instrument. (F specifications for main waveforms. This parameter applies only when “FM SOURCE” is set to “INTernal”.
Modulating signal frequency “FM FREQ The frequency range of the modulating signal is 100µHz ~ 10kHz. This parameter applies only when “FM SOURCE” is set to “INTernal”.
Modulating signal waveform “FM WAVE” Waveform of the modulating signal: Waveforms sine, square, triangle, rising and falling ramp can be used as the modulating signal. Waveforms are selected by entering the corresponding number, numbers 1 – 5. This parameter applies only when “FM SOURCE” is set to “INTernal”.
Modulating signal source “FM SOURCE” The modulating signal could be an internal signal and external input signal. The corresponding prompts are “1: INT” and “2: EXT”, the default is INTernal signal. The external modulating signal is applied to terminal “Mod In” on the rear panel (max signal amplitude is 3Vp-p). When
+ Fd <= F
c
”:
:
+ 100 kHz) For Fmax refer to the
max
26
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
modulating signal source “external” is selected, the symbol “Ext” is displaye d and parameters FM DEVIA, FM FREQ and FM WAVE do not apply (disabled)
Start and stop of FM modulation An FM signal is generated once FM function mode is selected. The instrument will automatically output a signal according to the preset parameters. If you do not want to output the FM signal, simply press the 【output】key to disable the signal output (LED above Output button is OFF). Once all the parameters are set, you can turn make the FM signal available at the output by pressing the Output button again.
FM example: Example configuration: Carrier signal is square, frequency is 1MHz, amplitude is 2V, modulating signal is generated internally . Carrier waveform is sine (N o. 1), Frequ ency is 5kH z. Peak fr equen cy deviation is 200kHz.
PressFM Pressfrequencythen【1】【MHz】 (set carrier frequency) Pressamplitude, then【2】【V】 (set carrier amplitude) PressShiftand square (set carrier waveform) Pressmenu, select “FM DEVIA”, enter【2】【0】【0】【kHz】 (set FM deviation) Pressmenu, select “FM FREQ”, press【5】【kHz】 (set FM frequency) Pressmenu, select “FM WAVE, press【1】【N】 (set FM waveform as sine) Pressmenu, select “FM SOURCE”, press【1】【N】 (set FM source as internal)
4.6.3 AM modulation
AM stands for amplitude modulation.
Press【AM】to enable AM modulation. The carrier frequency and the AM annunciator is displayed. Frequency, amplitude, waveform and DC offset of the carrier signal can be set as described in the previous section of this chapter. The parameters carry over from the parameter settings of the main waveforms sine and square. In AM only sine and square waves can be selected for carrier.
Press the 【menu】 key repeatedly to cycle through the AM modulation parameters listed below.
AM LEVEL> AM FREQ > AM W AVE>AM SOURCE
AM LEVEL: Modulation depth AM FREQ: Frequency of modulating signal AM WAVE: Waveform of modulating signal AM SOURCE: internal or external modulating signal
In AM mode, to e nsure n or m al signa l out pu t at 100 % modulation depth, the instrument reduces th e peak-to-peak value of the carrier by 50%. Only sine and square waves can be selected as AM carrier.
27
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Modulation depth “AM LEVEL”: The range of modulation depth is 1% ~ 120%.
Modulating frequency “AM FREQ” Frequency of the modulating signal. Range is 100µHz ~ 20kHz. Parameter does not apply when “AM SOURCE” is set to “E XTernal”
Modulating signal waveform “AM WAVE”: Waveforms sine, square, triangle, rising and falling ramp can be used as the modulating signal . Waveforms are selected by entering the corresponding number, numbers 1 – 5. Parameter does not apply when “AM SOURCE” is set to “EXTernal”
Modulating signal source “AM SOURCE” Select internal signal and external input signal. The number and prompt symbols are 1: INT, 2: EXT . The default of the instrument is internal signal. The external modulating signal is input through the rear panel “Modulation Input” terminal (with a maximum signal amplitude of 3Vp-p).
Enabling and disabling the AM signal: An AM signal is generated once AM modulation is selected. The instrument will automatically output a signal according to the preset parameters. If you do not want to output the AM signal, simply press the 【output】key to disable the signal output (LED above Output button is OFF). Once all the parameters are set, you can m ake the AM signal a vailable at the outpu t by pressing the Output button again.
AM example
: Configuration: carrier signal is square wave, frequency 1MHz, amplitude 2V, internal signal, modulating waveform sine wave (No.1), modulating signal frequency 5kHz, modulation depth 50%:
PressAM (activate AM modulation) Pressfrequency, then【1】【MHz】 (set carrier frequency) Pressamplitude, then【2】【V】 (set carrier amplitude) Pressshift thensquare (set carrier waveform) Pressmenu, select “AM LEVEL”, press【5】【0】【N】 (set modulation depth) Pressmenu, select “AM FREQ”, press【5】【kHz】 (set modulating signal frequency) Pressmenu, select “AM WAVE”, press【1】【N】 (set AM wave as sine) Pressmenu select “AM SOURCE”, press【1】【N】 (set AM source to internal)
4.6.4 Burst modulation
You can configure the function generator to output a waveform with a specified number of cycles, called a burst. You can output the burst at a rate determined by the internal rate generator or an external signal applied to the rear-panel connector. The function generator can produce a burst using sine or square waveforms.
PressBurstto enter into burst mode. The carrier frequency and the Burst annunciator are displayed. Frequency, amplitude, waveform and DC offset of the carrier signal can be set as
28
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
described in the previous se ction of this chapter. The parameters carry over from the paramete r settings of the main waveforms sine and square. Press the 【menu】 key repeatedly to cy cle through
the burst menu parameters listed below.
TRIG > COUNT > SPACE T > PHASE
TRIG: Trigger source COUNT: Number of cycles SPACE T: Burst time spacing PHASE: The starting phase of the burs t
Select trigger source “TRIG” Three trigger sources are available. Internal trigger and 2 types of external triggers sources, external gated and single. The corresponding menu parameters are 1: INT, 2: EXT and 3: SINGLE. The default is internal triggering.
a) Internal trigger When the internal trigger source is selected, the frequency at which the burst is generated depends on parameter SPACE_T, burst count and the burst carrier frequency.
b) Single trigger mode Generate a single trigger event by pushing the 【output】key once or by applying a single TTL pulse at the “Ext.Trig” input t erm inal in the rear panel. A single burst with predefined parameters will be generated.
c) External gated burst mode The TTL signal applied to “Ext. Trig” input terminal in the rear panel will enable or disable the output of the generator. When the TTL level of the gate signal is high, the generated will generate burst pulses according to the parameters of the set carrier frequency . Burst param eters COUNT and SP ACE_T are ignored in this mode. Anunciators “T rig” and “E xt” t urn on whe n this mo de is active .
Burst count “COUNT”: Definition: The number of cycles to be output per burst. Range: 1 ~10000 cycles in 1 cycle increments. Additionally, the minimum number of cycles must also satisfy F
/50khz +1Parameter does not apply when “TRIG” is set to “EXTernal”.
C
Burst spacing time “SPACE T”: Definition: Tim e interval between the 2 consecutive groups of bursts (time interval during which there is no burst signal present) Parameter does not apply when “TRIG” is set to “EXTernal” or “SINGle”.
Range: 0.1ms ~ 800s.
Starting phase of the burst “PHASE” Burst phase: 0.0º ~ 360.0º in 0.1º increments
29
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Enabling/disabling the burst signal The generator outputs a burst as soon as the burst key is pressed. The instrument will automatically output a signal according to the preset parameters. If you do not want to output the burst signal, simply press the 【output】key to disable the signal output (LED above Output button is OFF). Once all the parameters are set, you can make the burst signal available at the output by pressing the Output button again. Note that the output on/off function does not apply when single trigger mode is selected.
Burst example: Configuration: Burst is sine wave with frequency 20kH z, amplitude 2V, 10 cycles per group, spacing time between each group 10 ms, start phase 90.0º Pressburst (enter into burst mode) Pressfrequency, press【2】【0】【kHz】 (set wave form frequency) Pressamplitude, press【2】【V】 (set waveform amplitude) Pressshiftandsine (set waveform) Pressmenu, select “TRIG”, press【1】【N】 (set trigger mode as internal) Pressmenu, select “COUNT”, press【1】【0】【N】 (set number of bursts/cycles) Pressmenu, select “SP ACE T”, press【1】【0】【ms】 (set space time) Pressmenu, select “PHASE”, press【9】【0】【N】
(set the start phase of burst)
4.6.5 FSK modulation
You can configure the function generator to “shift” its output frequency between two preset values using FSK modulation. The rate at whic h the output shifts between the two frequencies (“carrier frequency” and “hop frequency”) is determined by the internal rate generator or the signal level on the rear-panel ext Trig/FSK/Burst terminal.
PressFSK/PSK to enable FSK modulation. A preset frequency and the FSK annunciator is displayed. The preset frequency is identical to the “START F” parameter. Any changes automatically update “START-F” and vice versa. Frequency , amplitude , waveform and DC of fset of the carrier signal can be set as described in the previous section of this chapter. In FSK/PSK only sine and square waves can be selected as carrier wave.
Press the 【menu】 key repeatedly to cycle through the FSK menu parameters listed below.
ST ART F > STOP F> SPACE T > TRIG
START F: First frequency or carrier frequency STOP F: Second frequency (hop frequency) SP ACE T: FSK rate TRIG: Trigger source
Trigger source “TRIG” Select internal signal and external input signal, 1: INT or 2: EXT. The default of the instrument is
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
30
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
“internal”. In internal mode, the rate at which the signal shifts between the carrier and hop frequency is determined by the FSK rate. In external trigger mode, the trigger signal applied to the trigger input terminal on the rear panel determines the FSK rate. Logical “high” of the trigger signal is associated with frequency 2, while the “Low” level of the trigger signal is associated with frequency 1.
Frequency 1 “START F” The first frequency or carrier frequency.
Frequency 2 “STOP F” Second frequency or hop frequency Frequency input range of frequency 1 and frequency 2 is 1µHz ~ Fmax.
Spacing time “SPACE T”
:
This parameter sets the FSK rate with a range of 1m s ~ 800s.
FSK example
: Configuration: Sine signal of 2V output amplitude, carrier frequency 20kH z and hopping frequency 600 kHz, FSK rate 10ms PressFSK/PSK (enter into FSK function mode) PressAmpl/Duty, press【2】【V】 (set waveform amplitude) Pressshiftandsine (set waveform) Pressmenu, select “TRIG”, enter【1】【N】 (set trigger mode as internal) Pressmenu, select “START F”, enter【2】【0】【kHz】 (set carrier frequency F1) Pressmenu, select “STOP F”, enter【6】【0】【0】【kHz】 (set hop frequency F2) Pressmenu, select spacing time “SPACE T”, press【1】【0】【ms】 (set spacing time)
4.6.6 PSK modulation
You can configure the functi on generator t o “shift” its ou tput phase bet ween two preset values using P
SK modulation. The rate at which the output shifts between the two phases is determined by the
internal rate generator or the signal level applied to the trigger input terminal on the rear-panel.
Press FSK/PSK twice to enter phase shift keying (PSK) mode. The carrier frequency will be displayed along with the symbol “” and “FSK”. Frequency, amplitude, waveform and DC offset of the carrier signal can be set as described in the previous section of this chapter. The parameters carry over from the parameter settings of the ma in waveforms sine and square. If FSK mode was already active, press this key only once, (this key toggles the modulation between FSK and PSK mode). In FSK mode only sine and square waves can be selected as a carrier signal.
Press the 【menu】 key repeatedly to cycle through the burst menu parameters listed below.
P1 > P2 > SPACE T > TRIG
P1 Phase value #1
31
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
P2: phase value #2 SP ACE T PSK rate TRIG Trigger source for PSK
Trigger source “TRIG” Select internal signal and external input signal 1: INT or 2: EXT. The default of the instrument is internal. In internal mode, the phase of t he output si gna l shifts according to the PSK rate parameter. In external trigger mode, the trigger s ignal is applied to the Ext. Trig” Input terminal on the rear panel. In external trigger mode, the trigger signal applied to the trigger i nput term inal on the rear panel determines the PSK rate. Logical “hi gh” of the trigger signal is associated with phase value #2, while the “Low” level of the trigger signal is associated with phase value #1.
Phase 1 “P1” The first start phase value of the PSK output signal. Input range of phase 1 and phase 2 is º0.0º~ 360.0º.
Phase 2 “P2” The second start phase value of the PSK output signal. Input range of phase 1 and phase 2 is 0.0º~ 360.0º.
Spacing time “SPACE T” PSK rate: Range is 0.1ms ~ 800s.
PSK example Configuration: Sine signals with 600 kHz output frequency, 2V amplitude, phase alternating between 90.0º and 180.0º, PSK rate is 10ms PressFSK/PSK (2 times if necessary to enter PSK mode) PressFreq/Period, enter【6】【0】【0】【kHz】 (set waveform frequency) PressAmpl/Duty, enter【2】【V】 (set waveform amplitude) Pressshift】【sine】 (set waveform) Pressmenu, select “TRIG”, enter【1】【N】, (set trigger mode as internal) Pressmenu, select “P1”, enter【9】【0】【N】 (set phase 1) Pressmenukey, select “P2”, enter【1】【8】【0】【N】 (set phase 2) Pressmenu, select “SPA CE T” , e nter【1】【
0】【ms】 (set PSK rate)
4.7 Set System Parameters
PressShiftandsystemto enter the Systems menu. “SYSTEM” will flash on the display. Repeatedly press【menu】to cycle through the system parameters indicated below.
POWER ON > ADDRESS > OUT Z —> INTERFACE>….
…………BAUD > PARITY
POWER ON “Power on” state
32
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
ADDRESS: set GPIB address (option) OUT Z: Configure amplitude display for 50Ω or high impedance load termination INTERFACE: Select RS232 or GP-IB (IEEE-488) interface (option) BAUD: Baud rate for the RS232 interface PARITY: Parity and Data Bits configuration for RS232
After the selected system parameter f lashes f or 1 second, the parameter value will be displayed and the value can be edited using the knob or via direct entry using the numerical keypad.
Power on state “POWER ON: This parameter has 2 possible values, “1: DEFAULT” or “2: LAST STATE”. In determines the initial setting of the instrument after power up. The factory setting is “1: DEFAULT”. The default state is main waveform sine wave, 10kHz, 2Vpp which is identical to the state the instrument is in after pressing “Rese t”. Whe n selecti ng “2: LAST STATE”, the instrument will “remember” the l ast state the unit was in before power was turned off. The power on state is stored in non-volatile memory, location 0 (see Store/Recall section)
GPIB address “ADDRESS” The default of GP-IB interface address is 1. It can be set within 0 ~ 30.
Output impedance “OUT Z” The generator has a fixed output impedance of 50 ohms on the OU TPUT terminal. You can specify whether you are terminating the output into a 50 ohm load or an open circuit. Incorrect impedance matching between the function generator and y our load will res ult in amplitude or offset which does not match the specified signal level.
Select “1: HIGH Z” or “2: 50 OHM” according to your load configuration.
“INTERFACE” selection Select RS232 (“2: RS232”), GPIB (“1: GP-IB”) or USB (“3: USB”). The default is RS232.
“BAUD” rate setting Set the baud rate for the RS232 interface. Possi bl e baud rates are 9600, 4800, 2400, 1200, 60 0 a nd
300. The default rate is 9600 (“1: 9600”)
“PARITY” setting Set parameters parity bit and n umber of data bits. Possible configurat ions are:
None/ 8 data bits (“1: NONE 8 BITS”) Odd/ 7 data bits (“2: ODD 7 BITS” Even/ 7 data bits (“3: EVEN 7 BITS”).
The default setting is None/ 8 data bits.
Example for system function setting Configuration: Set power on state [POWER ON] to default and output impedance to 50
PressShift】【system】 (enter systems menu) Pressmenu, “POWER ON”, press【1】【N】 (set power on state to default)
33
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Pressmenu, select “OUT Z”, press【2】【N】 (configure for load impedance of 50)
4.8 Universal Counter
Overview of Counter functions This instrument contains a counter with frequency measurement and totalize functionality which is completely independent from the Arbitrary/function generator section. The range of the frequency measurement function is 1Hz ~ 100MHz. Repeatedly pressing 【Shift】【Count】will toggle between Frequency measurement and Totalize mode.
PressShiftandF.C./TOTonce to activate the frequency measurem ent mode . Symbols “E xt” and “Freq” will be displayed in the state field and “WAITING” will flash on the main display. The unit is now ready to measure and display t he frequency of a signal applied to the input term inal “Meas Freq/TOT IN” in the rear panel. When totalize mode is active, sym bols “Ext” and “Count” will be displayed on the status field along with the result for the count on the main display (“0” is displayed until a valid signal is detected). The unit is now ready to totalize events represented by the signal applied to input terminal “Meas Freq/TOT IN” in the rear panel
In event counting mode, press the key to start and stop the event counter . When stopped, the current count will be displayed. Press】key to reset the event counter to zero and resume counting.
Gate time Only applies to frequency measurement mode. Press【Shift】【gate】to enter a gate time setting.
Either the numerical keypad or the knob can be used to enter a gate time value. If the gate is open, the symbol “GATE” is displayed in the counter state field on the right. The range of gate times is 10ms ~ 10s
Low pass filte
r
By pressing 【Shift】【LPF】, the input signal will be passed through a low pass filter before being passed onto the counter circuit. T he word “Filter ” ap pears on t he righ t dis play fiel d to ind ica te that the Low Pass Filter is active.
Attenuation PressShift】【ATT】 to activate the attenuator which attenuates the input signal by a factor 10:1.
“ATT” is displayed on the right to indicate that the input attenuator is active.
34
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
5. REMOTE INTERFACE REFERENCE
5.1 Introduction
An Introduction to the SCPI Language SCPI (Standard Commands for Programmable Instruments) is an ASCII-based instrument command language designed for test and m easurement instr uments . SCPI comm ands are based on a hierarchical structure, also known as a tree system. In this system, associated commands are grouped together under a common node or root, thus forming subsystems. A portion of the SOURCE subsystem is shown below to illustrate the tree system.
SOURce:
FREQuency:
STARt {<frequency>|MINimum|MAXimum}
FREQuency:
STARt? [MINimum|MAXimum]
FREQuency:
STOP {<frequency>|MINimum|MAXimum}
FREQuency:
STOP? [MINimum|MAXimum]
SOURce is the root keyword of the command, FREQuency is the second-level keywords, and STARt and STOP are third-level keywords. A colon (: ) separates a command keyword from a lower-level keyword.
Preparations for Remote Operation
The Arbitrary/Function Generator series 4080 can be remotely controlled from a PC via the standard RS232 interface. Make s ure to configure t he RS232 interfac e to match the se ttings on your PC by setting the parameters in the Systems menu accordingly. The instrument enters the remote state after any rem ote command is re ceived at which time all keys except of the 【Local】 key (Shift) are locked When in Remote mode, the instrument can be returned to Local mode by pressing the Local key or by sending a “LOCAL instruction via the RS232 interface.
5.2 Overview and syntax of SCPI instructions
This section summarizes the SCPI commands available to program the function generator over the remote interface. Refer to the later sections in this chapter for more details on each command.
Throughout this manual, the following conventions are used for SCPI command syntax.
Square brackets ([ ]) indicate optional keywords or parameters.
Braces ({ }) enclose parameters within a command string.
Triangle brackets (< >) indicate that you must substitute a value for the enclosed parameter.
A vertical bar (|) separates multiple parameter choices.
Command Summary
APPLY instruction
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
35
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
APPLy:SINusoid [<frequency>, <amplitude>, <offset>] APPLy:SQUare [<frequency>, <amplitude>, <offset>] APPLy:TRIangle [<frequency>, <amplitude>, <offset>] APPLy:UP_RAMP [<frequency>, <amplitude>, <offset>] APPLy:DOWN_RAMP [<frequency>, <amplitude>, <offset>] APPLy:NOISe [<frequency>*, <amplitude>, <offset>] APPLy:P_DC [<frequency>,* <amplitude>, <offset>] APPLy?
*: if waveform is Noise or DC, the frequency data contained in the command is not invalid.
OUTPUT instructions
[SOURce:] FUNCtion SHAPe {SINusoid|SQUare|TRIangle|UP_RAMP|DOWN_RAMP|
NOISe|PULSe|P_PULSE|N_PULSE|P_DC|N_DC|STAIR|
C_PULSE|COMMUT_FU|COMMUT_HA|SINE_TRA|
SINE_VER|SINE_PM|LOG|EXP|ROUND_HAL|
SINX/X|SQU_ROOT|TANGent|CARDIO|QUAKE|COMBIN} FUNCtion:SHAPe?
[SOURce:] FREQuency <frequency> FREQuency?
[SOURce:] PULSe:DCYCle <percent> PULSe:DCYCle?
[SOURce:] VOLTage <amplitude> VOLTage? VOLTage: OFFSet <offset> VOLTage: OFFSet?
[SOURce:] OUTPut:LOAD {50|INFinity} OUTPut:LOAD? INPUT:FILTER <on|off> INPUT:FILTER? INPUT:ATTenuator <on|off> INPUT:ATTenuator?
*SAV {0|1|2|3|4|5|6|7|8|9|10}. State 0 is the working state of the instrument before power off *RCL {0|1|2|3|4|5|6|7|8|9|10}. States 1—10 are user defined working states
MODULATION instructions
[SOURce:] AM:DEPTh <depth in percent> AM:DEPTh? AM:INTernal:FUNCtion {SINusoid|SQUare|TRIangle|UP_RAMP|DOWN_RAMP|} AM:INTernal:FUNCtion?
36
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
AM:INTernal:FREQuency <frequency> AM:INTernal:FREQuency? AM:SOURce {INTernal|EX Ternal} AM:SOURce? AM:STATe {ON|OFF} AM:STATe?
[SOURce:] FM:DEViation <peak deviation in Hz> FM:DEViation? FM:INTernal: FUNCtion {SINusoid|SQUare|TRIangle|U P_RAMP|DOWN_RAMP|……} FM:INTernal: FUNCtion? FM:INTernal: FREQuency <frequency> FM:INTernal: FREQuency? FM:SOURce {INTernal|EXTernal} FM:SOURce? FM:STATe {ON|OFF} FM:STATe?
[SOURce:] BM:NCYCles <#cycles> BM:NCYCles? BM:PHASe <degrees> BM:PHASe? BM:INTernal: Space <time in second> BM:INTernal: Space? BM:SOURce {INTernal|EXTernal|SINGle} BM:SOURce? BM:STATe {ON|OFF} BM:STATe?
FSK instructions
[SOURce:] FSKey:FREQuency <frequency> FSKey:FREQuency? FSKey:INTernal: Space <time in second> FSKey:INTernal: Space? FSKey:SOURce {INTernal|EXTernal} FSKey:SOURce?
FSKey:STATe {ON|OFF} FSKey:STATe?
PSK instructions
[SOURce:] PSKey:PHASe1 <degrees> PSKey:PHASe1? PSKey:PHASe2 <degrees> PSKey:PHASe2? PSKey:INTernal: Space <time in second> PSKey:INTernal: Space?
37
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
PSKey:SOURce {INTernal|EXTernal} PSKey:SOURce? PSKey:STATe {ON|OFF} PSKey:STATe?
SWEEP instructions
[SOURce:] FREQuency:STARt <frequency> FREQuency:STARt? FREQuency:STOP <frequency> FREQuency:STOP?
[SOURce:] SWEep:SPACing {LINear|LOGarithmic} SWEep:SPACing? SWEep:TIME <time in second> SWEep:TIME? SWEep:SOURce {INTernal|EXTernal} SWEep:SOURce? SWEep:STATe {ON|OFF} SWEep:STATe?
COUNT instructions
[SOURce:] FUNCtion:TOTal INITial FUNCtion:TOTal STARt FUNCtion:TOTal STOP FUNCtion:TOTal CLEAr FUNCtion:TOTal?
FREQUENCY MEASUREMENT instructions
[SOURce:] FUNCtion:FREQuency MEASure FUNCtion:FREQuency? FUNCtion:FREQuency GATE <time> FUNCtion:FREQuency GATE?
TRIGGER instructions
TRIGger:SOURce {IMMediate|EXTernal|BUS} TRIGger:SOURce?
SYSTEM instructions
38
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
*IDN *RST *SAV {0|1|2|3|4|5|6|7|8|9|10} *RCL {0|1|2|3|4|5|6|7|8|9|10}
RS232 specific instructions
SYSTem:LOCAL SYSTem:REMOTE
5.3 Detailed description of SCPI Instructions
APPLy instructions APPLy instructions are used to directly set the output of the function generato r throu gh the remote interface including waveform, frequ ency and offset. For example, to output a sine wave of 5kHz, 3Vp-p with –2.5V DC off set, the following commands need to be sent: “APPL:SIN 5 KHZ, 3.0 VPP, -2.5 V” “APPL:SIN 5.0E+3,3.0, -2.5”
Instructions of lower level can also be used: “FUNC:SHAPE SIN”, to output a sine wave “FREQ 5.0 KHZ”, to set frequency as 5kHz “VOLT 3.0”, to set voltage as 3 VPP “VOLT:OFFSET -2.5”, to set output offset as -2.5 V
APPLy? Queries the waveform, frequency, amplitude and offset of current funct ion output. The r eturned data format is: “SIN 5.000000000000E+03, 3.000000E+00, -2.500000E+00”
OUTPUT instructions FUNCtion:SHAPe {SINusoid|SQUare|TRIangle|UP_RAMP|DOWN_RAMP|NOISe……} Select output wavefor m. 27 wave forms are availa ble. In fixed freq uency mode, each of th e 27 kinds can be selected. In other modes, only sine and square waves can be selected.
FUNCtion:SHAPe? Check the waveform of current function output. The returned data are SIN, SQU, TRI, UP_RAMP, etc.
FREQuency <frequency> Set the frequency of function output
FREQuency Check the frequency of current ma in function output. The returned data unit is Hz.
PULSe:DCYCle <percent> Set the duty cycle of pulse output. Duty cycle: 0.1%99.9%, in step of 0.1% (f 10 KHz)
1%99% in step of 1% (10 kHz < f 100 kHz), default value 20%.
PULSe:DCYCle? Check the duty cycle of current pulse output. The returned value is percentage.
VOLTage <amplitude>
39
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Set the amplitude of the currently active waveform. The amplitude setting should be consistent with the technical specifications. The output amplitude ranges of P_PULSE, N_PULSE, P_DC and N_DC should be 2mVpp ampl 10Vpp in high impedance output setting and 1mVpp ampl 5Vpp in 50 setting. The above value is given based on a DC offset of 0V. If a DC offset is used, refer to the setting of DC offset described in Chapter 4 in this manual
VOLTage Check the output amplitude of current main function out put.
VOLTage OFFSet <offset> Set the DC offset. For the relationship between the DC of fset and output am plitude, refer to section
4.5.3.
VOLTage OFFSet Check the DC offset of main functi on outp ut.
OUTPut:LOAD {50|INFinity} Configure the amplitude display value for 50 or high impedance (I NFinity) load termination. The output amplitude should be consistent with the technical specifications.
OUTPut:LOAD Read the current load impedance setting of the function generator.
INPUT: FILTER {ON|OFF} Enable/disable the low pass filter for the counter
INPUT:FILTER? Read the current “on/off” state of the low pass filter.
INPUT:ATTenuator {ON|OFF} Enable/disable Set the input attenuator of the counter
INPUT:ATTenuator? Read the state of the input attenuator (on/off)
*SAV {0|1|2|3|4|5|6|7|8|9|10} Save as many as 11 input states of the instrument. Among the 11 states, state 0 is used to automatically store the power off setting of the instrument. States 1~10 are user defined.
*RCL {0|1|2|3|4|5|6|7|8|9|10} Recall one of the states 0 -10.
AM instructions 1To set the carrier of output using APPLy instructions or FUNCtion, FREQuency, VOLTage,
VOLTAGE: OFFSet instructions.
2T o set modulating waveform of AM using AM:INTernal: FUNCtion
{SINusoid|SQUare|TRIangle……} instructions 3.To set modulatin g frequency of AM usi ng AM: INT e rnal: FREQuency <frequency> instruction s. 4To set the m odu lat io n dept h of AM using AM: DEPTh {depth in percent} instructions 5To set modulating source of AM using AM: SOURce {INTernal|EXTernal} instructions 6To start AM using AM: STATe ON inst ructions.
AM: DEPTh {depth in percent} Set modulation depth of AM. MIN=1%, MAX=120%.
AM: DEPTh
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
40
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Read the current modulation depth of AM.
AM:INTernal:FUNCtion {SINusoid|SQUare|TRIangle……} Set the modulating waveform for AM
DOWN_RAMP.
,using modulating waveforms S IN, SQU, TRI UP, RAMP and
AM:INTernal:FUNCtion Check current modulating waveform of AM. Re turn parameters are “SIN”, “SQU”, TRI”, “UP_RAMP”, “DOWN_RAMP”
AM:INTernal: FREQuency <frequency> Set frequency of modulating signal of AM. When internal modulating source is selected, the modulating frequency should be≤10kHz.
AM:INTernal:FREQuency? Read the frequency of the current modulating signal for AM.
AM:SOURce {INTernal|EX Ternal} Set modulating source of modulating signal of AM as internal (INTernal) or external (EXTernal).
AM:SOURce Read the current modulating source of AM.
AM:STATe {ON|OFF} Turn AM mode on/off
AM:STATe Check the “on/off” state of AM mode. The returned data are “0” (OFF) or “1” (ON).In “of f” state , the instrument reverts back to fixed frequency mode.
FM instructions 1To set the car rier of output usin g APPLy or FUNCtion, FREQuency, VOLTage, VOLTAGE:
OFFSet instructions. 2To set modulating waveform of FM us ing FM: INTernal: FUNCtion
{SINusoid|SQUare|TRIangle……} instructions 3To set modulating frequency of FM using instruction FM:INTernal: FREQuency <frequency> 4To set the modulatin g deviation of FM using instructi on FM: DEViation <peak deviation in Hz> 5To set modulating source of FM using the FM: SOURce {INTernal|EXTernal} instruction 6To start FM using FM: STATe ON instruction.
FM:DEViation <peak deviation in Hz> Set the modulating deviation of FM. The maximum value should be half of carrier frequency in internal modulating sour ce and 10% of carrier freque ncy in exter nal modulating fr equency. (Should also satisfy the following equation: F
+ Fd <= F
c
+100khz
max
FM:DEViation Read the current modulating deviation of FM.
FM:INTernal:FUNCtion {SINusoid|SQUare|TRIangle……} Set the modulating wavef orm of FM u sin g
SIN, SQU, TRI UP, RAMP or DOWN_RAMP waveforms.
FM:INTernal:FUNCtion Read the current modulating waveform of FM. The returned data are “SIN”, “SQU”, “TRI”, “UP_RAMP”, “DOWN_RAMP”.
41
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
FM:INTernal:FREQuency <frequency> Set the frequency of modulating signal of FM. The modula ting frequency should be 10KHz when internal modulating source is selected.
FM:INTernal:FREQuency? Read the frequency of current modulating signal of FM.
FM:SOURce {INTernal|EXTernal} Set modulating source of FM signal as internal (INTernal) or external (EXTernal).
FM:SOURce Read the current modulating source of FM.
FM:STATe {ON|OFF} Turn on FM mode (O N), turn off FM mode (OFF).
FM:STATe Check the “on/off” state of FM mode. The returned data are “0” (OFF) or “1” (ON). It returns to fixed frequency mode in “off” state.
Burst Mode instructions
1. Set the carrier using APPLy instructions or FUNCtion, FREQuency, VOLTage, VOLTAGE:
OFFSet instruction.
2. Set number of burst using BM: NCYCles <#cycles> instruction.
3. Set start phase of burst using BM: PHASe <degrees> instruction.
4. Set time interval between two bursts using BM:INTernal: SPACe <time> instruction.
5. Set trigger source of burst using BM: SOURce {INTernal|EXTernal|SINGle} instruction.
6. To turn on burst use the BM: STATe ON instruction.
BM:NCYCles <#cycles> Set waveform numbers of burst. MAX10000. BM:NCYCles Check number of bursts.
BM:PHASe <degrees> Set start phase of burst in 0360º in step of 0.1º.
BM:PHASe Check the start phase of burst.
BM:INTernal: SPACe <time> Set time interval between two bursts.
BM:INTernal: SPACe? Check time interval between two bursts.
BM:SOURce {INTernal|EXTernal|SINGle} Set trigger source of burst as internal (INTernal), external (EXTernal) or single (SINGle).
BM:SOURce Check current trigger source of b urst.
BM:STATe {ON|OFF} Turn on burst m ode (ON) or turn off burst mode (OFF).
BM:STATe
42
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Check “on/off” state of burst mode. The returned data are “0” (OFF) or “1” (ON). The instrument reverts back to fixed frequency mode in “off” state.
FSK Mode instructions 1To set the output waveform, amplitude, DC offset and freque ncy 1 using APPLy instructions or
FUNCtion, FREQuency, VOLTage, VOLTAGE:OFFSet instructions. 2set frequency 2 of FSK using FSKey: FREQuency <frequ ency> instruction. 3set time interval of frequency switching using FSKey:INTernal: SPACe <time> instruction. 4set trigger source of FSK using FSKey: SOURce {INTernal|EXTernal} instructions. 5turn on FSK using FSKey: STATe ON instruction.
FSKey:FREQuency <frequency> Set frequency 2 of FSK.
FSKey:FREQuency Check frequency 2 of FSK.
FSKey:INTernal:SPACe <time> Set time interval of frequency switching
FSKey:INTernal:SPACe? Check time interval of frequency switchi ng
FSKey:SOURce {INTernal|EXTernal} Set trigger source of FSK as internal (INTernal) or external (EXTernal).
FSKey:SOURce Check current trigger source of FSK.
FSKey:STATe {ON|OFF} Turn on FSK mode ( ON) or turn off FSK mode (OFF).
FSKey:STATe Check the “on/off” state of FSK mode. The returned data are “0” (OFF) or “1” (ON). It returns to fixed frequency mode in “off” state.
PSK Mode instructions
1. To set the output waveform, amplitude, DC offset and frequency using APPLy instructions or
FUNCtion, FREQuency, VOLTage, VOLTAGE:OFFSet instructions.
2. To set phase 1of PSK using PSKey: PHASe1 <degrees> instruction.
3. To set phase 2 of PSK using PSKey: PHASe2 <degrees> instruction.
4. To set time interval of phase switching using PSKey:INTernal: SPACe <time> instruction.
5. To set trigger source of PSK using PSKey: SOURce {INTernal|EXTernal} instructions.
6. To turn on PSK using PSKey: STATe ON instruction.
FSKey:PHASe1 <degrees> Set phase 1 of PSK in 0--360º.
PSKey:PHASe1 Check phase 1of PSK
PSKey:PHASe2 <degrees> Set phase 2 of PSK in 0--360º.
PSKey:PHASe2 Check phase 2 of PSK.
43
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
PSKey:INTernal:SPACe <time> Set time inter val of phase switching
PSKey:INTernal:SPACe? Check time interval of phase switchi ng
PSKey: SOURce {INT erna l|EXTernal} Set trigger source of PSK as internal (INTernal) or external (EXTernal).
PSKey:SOURce Check current trigger source of PSK.
PSKey:STATe {ON|OFF} Turn on PSK mode (ON) or turn of f PSK mode (OFF).
PSKey:STATe Check the “on/off” state of PSK mode. The returned data are “0” (OFF) or “1” (ON). It returns to fixed frequency mode in “off” state.
Sweep Mode instructions
1. To set the output waveform, amplitude, DC offset and frequency using APPLy instructions or
FUNCtion, FREQuency, VOLTage, VOLTAGE:OFFSet instructions.
2. To set start frequency of sweep using FREQuency: STARt <frequency> instruction.
3. To set stop frequency of sweep us ing P FREQuency: STOP<frequency> instruction.
4. To set sweep mode using SWEep: SPACing {LINear|LOGarithmic} instruction.
5. To set sweep time using SWEep: TIME <time> instructions.
6. To set trigger source of sweep using SWEep: SOURce {INTernal|EXT ernal}instructions.
7. To turn on sweep using SWEep: STATe ON instruction.
FREQuency:STARt <frequency> Set start frequency of sweep.
FREQuency:STARt Check startting frequency of sweep.
FREQuency:STOP <frequency> Set stop frequency of sweep.
FREQuency:STOP Check stop frequency of sweep.
SWEep:SPACing {LINear|LOGarithmic} Set sweep mode as linear (LINear) or logarithmic (LOGarithmic)
SWEep:SPACing Check sweep mode.
SWEep:TIME <time> Set sweep time.
SWEep:TIME? Check the sweep time.
SWEep:SOURce {INTernal|EXTernal} Set trigger source of sweep as internal (INTernal) or external (EXTernal).
SWEep:SOURce
44
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Check current trigger source.
SWEep:STATe {ON|OFF} Turn on sweep mode (ON) or turn off sweep mode (OFF).
SWEep:STATe Check the “on/off” state of sweep mode. The returned data are “0” (OFF) or “1” (ON). It returns to fixed frequency mode in “off” state.
Counting instructions FUNCtion:TOTal INITial Initialize the counter.
FUNCtion:TOTal ST ARt Set the counter to start counting.
FUNCtion:TOTal STOP Set the counter to stop counting.
FUNCtion:TOTal CLEAr Clear the counted value of the counter.
FUNCtion:TOTal? Record the current value of the counter.
Frequency Measurement instructions FUNCtion:FREQuency MEASure Enable frequency measurement function for the instrument to start a new frequency measurement.
FUNCtion:FREQuency? Record last frequency value after frequency measurement is finished.
FUNCtion:FREQuency GATE <time> Set gate time in the range of 10ms to 10s for frequency measurement.
FUNCtion:FREQuency GATE? Check current gate time for frequency meadurement.
Trigger instructions TRIGger:SOURce {IMMediate|EXTernal|BUS} Set trigger source or modulating source for available functions as internal (IMMediate), external (EXTernal) or single (BUS). TRIGger:SOURce? Check trigger source or modulating source of current function as internal (IMM), external, (EXT) or single (BUS).
System related instructions:
*IDN This query will return the instrument’s manufacturer, model number and firmware version. Example: B+K Precision 4084AWG V1.00
*RST Reset instrument to the default state
*SAV {0|1|2|3|4|5|6|7|8|9|10} Save as many as 11 input states of the instrument. Among the 11 states, state 0 is used to automatically store the power off setting of the instrument. States 1~10 are user defined.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
45
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
*RCL {0|1|2|3|4|5|6|7|8|9|10} Recall on of the states 0 -10.
Special instructions for RS232
SYSTem:LOCAL This instruction cau ses the instrument to enter into local state. After receipt of this command, all front panel keys are enabled.
SYSTem:REMOTE This instruction causes the instrument enter remote state from local state. After receipt of this command, all front panel keys are disabled except of the local key (shift key) which can be us ed to manually revert to local state.
NOTE: All command strings send to the instrument must terminate with a <new line> character. (0x0A or “\n”).
Wiring diagram of RS232 cable required for remote control (null modem configuration)
DB9 connector
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
DB9 connector
46
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
6. USER PROGRAMMABLE ARBITRARY W AVEFORM MODULE
Product Overview
In addition to the 27 build-in waveforms, models 4084AW G and 4086AWG also provide the capability to generate and store up to 8 user-defined waveforms in non-volatile memory.
This Win dows-based software tool offers a graphical user interface that you can use to create, edit, and transfer custom waveforms to the internal memory of the 4084AWG or 4086AWG Function/Arbitrary Generator.
Waveforms can be created in several ways: Users can draw new waveforms with a mouse, edit one of the predefined waveforms, rea d in waveform param eters from a text f ile or downloa d waveforms from a TEKTRONIX transmitted to the generator and stored in non- volat ile memory.
Software Installation
This section describes how to install the Arbitrar y Wa veform Creation Software using the Install Shield Wizard.
1) Insert the Installation CD into the CD-ROM drive.
2) Locate setup.exe on the CD-ROM and double-click it.
3) Follow the instructions given by the Install Shield wizard.
Launching the Software
To start the Software you can do one of the following: a) Run the application from the Windows Start menu. Select Start> Programs> BK_Precision>4080 Series Arbitrary Waveform Software\Arb_Process.exe b) Go to the folder where you have installed the Arbitrary Waveform Creation Software and double-click on Arb_process.exe.
To exit the application, select File
®
digital storage oscilloscope. Once completed, the waveform can be
> Exit from the pull down menu
47
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
CREA TING AND EDITING W AVEFORMS
Create waveforms by loading waveform data from a text file or by using the graphical tools of this Software.
Working with text files
Waveforms can be imported to or exported from a text file with the following format: The first line contains the number of data points, the consecutive lines the Y axis value of the sampling points.
Example wave1.txt:
8 (Line 1length of the waveform)
-0.666 (Line 2waveform value of po in t 0)
-0.555 (Line 3waveform value of po in t 1)
-0.111 (Line 4waveform value of point 2)
0.001 (Line 5waveform value of po int 3)
0.123 (Line 6waveform value of po int 4)
0.235 (Line 7waveform value of po int 5)
0.378 (Line 8waveform value of po int 6)
0.654 (Line 9waveform value of po int 7)
Only use text editors such as NOTEPAD to create or edit this file.
Saving and opening text files To save the current waveform as a text file, Select File>Save to and enter a filename. To import waveform data, select “Open” from the File menu. The waveform data will be automatically displayed in the editing window.
48
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Working with graphical tools
Waveform Parameters The waveform parameters are represented in the following format: Values of the vertical axis (Y) are normalized to -1.000 ~ 1.000. Values of the horizontal axis (X axis) are 0 ~ length of waveform minus 1.
Setting the number of points The waveform length can be set as required. Input the required number of points in the waveform setup area then press ENTER. The X-axis will be updated accordingly. The minimum number of points is 8, the maximum number is 16000.
Drawing waveforms
Click on the left mouse key then grab and move the default center line to the desired location. The coordinates of the cursor position will be displayed while “drawing”. Release the left mouse key to stop drawing.
49
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Clearing the waveform To clear all waveforms in the editing area, click the “Clear” waveform button.
Zoom function To en large an area of the waveform, click on the “zoom in” sy mbol then us e the scroll b ar below the editing window to view the desired area. You can also use V iew>Zoom in. To zoom out, click on the “zoom out” symbol on the tool bar or select use View>Zoom.
Click the “11”symbol to restore the default view.
Direct waveform entry
Accurate input of point values Define a point by entering the (X) and (Y) coor dinates in the start point field then click “Set point” to complete the entry. The value of this point will be recorded and displayed in the drawing area.
Accurate input of line segments To define a straight line segment, enter the corresponding (X) and (Y) coordinates of the starting point of this line segment in the “start point” field. Enter the (X) and (Y) coordinates for the stop point of the line in the”stop point” field. Click “set line”. The value of this line segm ent will be automatically calculated and displayed in the drawing area.
Generating standard waveform 6 standard waveforms can be generated:squaresinecosinetrianglerisi ng and falling ram p.
Select a waveform from the standard waveforms menu. The waveform will be displayed immediately in the editing window.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
50
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Modifying existing waveforms with mathematical Functions
Waveforms can be modified by applying one of the following mathematical functions: Addition, subtraction of waveforms, symmetry of Y-axis and X-axis, division and multiplication by a constant.
Add Add waveform data stored in a text file to the current waveform (displayed on the screen). The
51
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
resulting new waveform is automatically normalized by dividing the new waveform by 2. To look at it differently, each amplitude value is weighted by a factor 0.5 before the add operation is performed. If the lengths of the two waveforms differ, the longer length is taken as the length of the new waveform.
Subtract Subtract waveform data stored in a text file from the current waveform (displayed on the screen). The resulting new waveform is automatically normalized by dividing the new waveform by 2. To look at it differently, each amplitude value is weighted by a factor 0.5 before the subtract operation is performed. If the lengths of the two waveforms differ, the longer length is taken as the length of the new waveform.
Symmetry of waveform X-axis Flip the current image around the X-axis to create a mirror image of the previous waveform.
Symmetry of waveform Y-axis Flip the current image around the vertical center line, which divides the editing screen in half.
Amplitude Divide Divide the amplitude of the current waveform by an integer value. The value can be positive or negative (but not zero)
Amplitude multiply Multiply the amplitude of the current waveform by an integer value. The value can be positive or negative (but not zero)
Frequency multiply Multiply the timebase value of the current waveform by an integer value. The value must be positive (but not zero). An integer value of 2 will double the frequency
Formula of sine add Add 2 sine waves according to the predef ined form ula 1*sin(x) + a*sin (b*x). Choose parameters a and b. The first sine wave in the equation cannot be edited.
52
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
TRANSFERING WAVEFORM DATA T O THE 408XAWG
RS232 communication setup
1) Connect the computer to the 4084AWG or 4086AWG using the RS232 cable included.
2) Set the RS232 parameters of the Arbitrary Waveform Creation Tool
3) Set the RS232 parameters in the Systems menu of the 4080 series generator. Make sure the “COM port”, “Baud rate” and “parity” and Flow control parameters match that of the Software settings.
Select a memory location Models 4084AWG and 4086AWG offer 8 memory locations for storing user defined arbitrary waveforms. Numbers 1 – 8 in the waveform number field of the software correspond to “ARB1” – “ARB8” on the instrument side which ar e located at positions 28 to 35 on the Arblis t (see chapter 4, section 3).
Arblist location Press <Shift> + Display Name
28 1 ARB1 29 2 ARB2 30 3 ARB3 31 4 ARB4 32 5 ARB5 33 6 ARB6 34 7 ARB7 35 8 ARB8
Transfer of waveform in current editing area Initiate a waveform transfer to the 4080 series generator by clicking the “Send Data” button or Selecting Other>Send Data from the menu. A pop-up window will indicate the progress of the data transfer. The window will automatically disappear once the data was downloaded successfully. If
53
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
the download was unsuccessful due to communication problems, a pop message will appear prompting the user to verify the setup and settings of the communication link.
Example
Draw a waveform with a 1024 point length and transfer the data to memory location “ARB1”.
1) Connect the computer and the instrument us ing the supplie d RS232 ca ble and config ure the
RS232 parameters on the instrument and software side. Make sure the parameters match
2) Set waveform length as 1024 points.
3) Create a waveform
4) Set “Waveform Number” to 1
5) Click on the “Send Data” button
6) After the transfer completed successfully, press [Shift][Shift] [1] on the instrument to select
and enable waveform “ARB1” (the first Shift key press puts the instrument into Local
mode) Note: If ARB1 was already enabled, the output of the generator will be updated immediately with the new waveform data as soon as the download is complete
READING DATA FROM A TEKTRONIX® DIGITAL STORAGE OSCILLOSCOPE
This tool allows the direct transfer of waveform data from a Tektronix digital storage oscilloscope to this Software. The data can originate from the oscilloscope’s screen display (ch annel 1 or channel 2) or internal memory where the reference waveform from channel 1 or 2 is stored. The Software is compatible with TDS1000, TDS2000, T PS2000 a nd TDS300 0 series oscilloscopes equipped wi th a RS232 interface. Download instructions:
1) Connect the computer and digital storage oscilloscope (DSO) using an RS232 cable.
2) Set the communications parameters for the DSO. Refer to the oscilloscope’s manual for
setup instructions. Make sure that the RS232 parameters under “menu RS232 set” match
that of the Arbitrary waveform Creation Software. Additionally set the “EOL” or “End of
Line” string in the oscilloscopes menu. Supported termination strings are <CR>, LF or a
combination of the two, <CR><LF> or <LF><CR>
3) Set the RS232 parameters and EOL string on the Software side. Make sure parameters
“communications address”, “Baud rate”, “parity”, “flow control” and“EOF character
string” matches that of the oscilloscope settings.
54
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
4) Click “Set Link” in the “oscilloscope” menu of the software to check the communications
link between software and oscilloscope. A pop-up window will indicate successful
communication; otherwise, the operation will result in an error message. This operation
will also automatically populate the waveform length field based on the oscilloscope’s
internal storage memory.
5) Select the Source of the data to be read from the oscilloscope
6) Select the desired data source from oscilloscope menu>Source. Refer to your
oscilloscope’s user manual for data sources available.
7) Read the oscilloscope data from the source selected: Click on “read data” in sub-menu
oscilloscope”. A pop-up window will appear to display the current status of the transfer.
The pop up window will automatically disappear once transmission was completed
successfully and the downloaded data will automatically be displayed in the Software’s
editing window.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
55
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Requirement for RS232 cable connecting the TEKTRONIX oscilloscope and 408xAWG Refer to the oscilloscope’s instruction manual. For the TEKTRO NIX series digital storage
oscilloscopes, a null modem (cross over) using 2 handshake lines is recommended.
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
56
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
7. APPENDIX
Responsible Party Alternate Manufacturi ng Sit e
Manufacturer’s Name: Manufacture’s Address:
B&K Precision Corporation B&K China 1364
22820 Savi Ranch Pkwy. Yorba Linda, CA 92887-4610 USA
Declares that the below mentio ned product Product Name: Function/ Arbitrary Generator with Counter Part Numbers: 4084, 4084AWG, 4085, 4086, 4086AWG, 4087
complies with the essential requirements of the following applicable European Directives:
Low Voltage Directive 73/23/EEC (19.02.73) amended by 93/68/EEC (22.07.93) Electromagnetic Compatibility (EMC) 89/336/EEC (03.05.88) amended by 92/68/EEC (22.07.93)
and conforms with the following product standards:
LVD EN61010-1:2001, EN61010-031:2002
EMC
EN61326: 1997+A1: 1998+A2: 2001+A3: 2003 EN55022: 1998+A1:2000+A2 EN61000-3-2: 2000 EN61000-3-3: 1995+A1:2001 EN55024: 1998+A1: 2001+A2:2003
This Declaration of Conformity applies to above listed products place on the EU market after:
President
January 31 2007
Date Victor Tolan
57
4080 series Arbitrary/Function Generator with Counter –Instruction Manual
Service Information
Warranty Service: Pl ease return the produ ct in the original packagi ng with proof of purch ase to the address below.
Clearly state in writing the performance problem and return any leads, probes, connectors and accessories that you are using with the device.
Non-Warranty Service: Return the product in the original packaging to the address below. Clearly state in writing the performance problem and return any leads, probes, con nectors and accessories that you are using with the devi ce. Customers not on open account must include payment in the form of a money order or credit card. For the most current repair charges please visit www.bkprecision.com
Return all merchandise to B&K Precision Corp. with pre-paid shipping. The flat-rate repair charge for Non-W arrant y Service does not include return shipping. Return shipping to locations in North American is included for Warranty Service. For overnight shipments and non-North American shipping fees please contact B&K Precision Corp.
B&K Precision Corp.
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
www.bkprecision.com
Include with the returned instrument your complete return shipping address, contact name, phone number and description of problem.
and click on “service/repair”.
714-921-9095
Limited One-Year Warranty
B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of one year from d ate of purchase.
B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt.
To o btain warranty coverage in the U.S.A., this product must be registered by completing a warranty registration form on www.bkprecision.com
Exclusions: This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized alterations or repairs. The warranty is void if the serial number is alter ed, defaced or r emoved.
B&K Precision Corp. shall not be liable for any consequential damages, including without limitation damages resulting from loss of use. Some states do not allow limitations of incidental or consequential damages. So the above limitation or exclusion may not apply to you.
This warranty gives you specific rights and you may have other rights, which vary from state-to-state.
B&K Precision Corp. 22820 Savi Ranch Parkway Yorba Linda, CA 92887
www.bkprecision.com 714-921-9095
within fifteen (15) days of purchase.
58
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
Loading...