BK Precision BK Precision 4047B Manual

Model: 4047B
Dual Channel Function/Arbitrary Waveform Generator
USER MANUAL

Safety Summary

The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
Read and understand the safety and operational information in this manual.Apply all the listed safety precautions.Verify that the voltage selector at the line power cord input is set to the correct line voltage.
Operating the instrument at an incorrect line voltage will void the warranty.
Make all connections to the instrument before applying power.Do not operate the instrument in ways not specified by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design, manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply with these requirements.
Category rating
The IEC 61010 standard defines safety category ratings that specify the amount of electrical energy available and the voltage impulses that may occur on electrical conductors associated with these category ratings. The category rating is a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of the circuit to be tested, which defines the voltage impulses expected and required insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected to the mains supply. The voltages in the environment are typically derived from a limited­energy transformer or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the mains supply at a standard wall outlet or similar sources. Example measurement environments are portable tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be connected to the mains installation of a building. Examples are measurements inside a building's circuit breaker panel or the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to the primary power entering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what is
2
specified in this manual for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the instrument's category rating to maintain the instrument's category rating. Failure to do so will lower the category rating of the measuring system.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be 120 V RMS or 240 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This instrument is grounded through the ground conductor of the supplied, three­conductor AC line power cable. The power cable must be plugged into an approved three-conductor electrical outlet. The power jack and mating plug of the power cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive parts (including control knobs) may provide an electric shock. Failure to use a properly­grounded approved outlet and the recommended three-conductor AC line power cable may result in injury or death.
Unless otherwise stated, a ground connection on the instrument's front or rear panel is for a reference of potential only and is not to be used as a safety ground.
Do not operate in an explosive or flammable atmosphere
Do not operate the instrument in the presence of flammable gases or vapors, fumes, or finely-divided particulates.
3
The instrument is designed to be used in office-type indoor environments. Do not operate the instrument
In the presence of noxious, corrosive, or flammable fumes, gases, vapors, chemicals, or finely-
divided particulates.
In relative humidity conditions outside the instrument's specifications.In environments where there is a danger of any liquid being spilled on the instrument or
where any liquid can condense on the instrument.
In air temperatures exceeding the specified operating temperatures.In atmospheric pressures outside the specified altitude limits or where the surrounding gas is
not air.
In environments with restricted cooling air flow, even if the air temperatures are within
specifications.
In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature range is 0 °C to 50 °C and the operating humidity is ≤ 95 % relative humidity at < 30 °C, with no condensation allowed. Measurements made by this instrument may be outside specifications if the instrument is used in non-office-type environments. Such environments may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise, strong electric fields, or strong magnetic fields.
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the instrument, remove the instrument's power cord, remove the instrument from service, label it as not to be operated, and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the instrument.
Clean the instrument only as instructed
4
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual.
Not for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a life-support device or system.
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must be made by qualified service-trained maintenance personnel who are aware of the hazards involved when the instrument's covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may exist when the covers are removed. To avoid injuries, always disconnect the power cord from the instrument, disconnect all other connections (for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there are no hazardous voltages present on any conductors by measurements with a properly­operating voltage-sensing device before touching any internal parts. Verify the voltage-sensing device is working properly before and after making the measurements by testing with known-operating voltage sources and test for both DC and AC voltages. Do not attempt any service or adjustment unless another person capable of rendering first aid and resuscitation is present. Do not insert any object into an instrument's ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit exists.
Fuse replacement
Fuse replacement must be done by qualified service-trained maintenance personnel who are aware of the instrument's fuse requirements and safe replacement procedures. Disconnect the instrument from the power line before replacing fuses. Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings specified in this manual or on the back of the instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a fire. Failure to use the specified fuses will void the warranty.
5
Servicing
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to B&K Precision for service and repair to ensure that safety and performance features are maintained.
Cooling fans
This instrument contains one or more cooling fans. For continued safe operation of the instrument, the air inlet and exhaust openings for these fans must not be blocked nor must accumulated dust or other debris be allowed to reduce air flow. Maintain at least 25 mm clearance around the sides of the instrument that contain air inlet and exhaust ports. If mounted in a rack, position power devices in the rack above the instrument to minimize instrument heating while rack mounted. Do not continue to operate the instrument if you cannot verify the fan is operating (note some fans may have intermittent duty cycles). Do not insert any object into the fan's inlet or outlet.
For continued safe use of the instrument
Do not place heavy objects on the instrument. Do not obstruct cooling air flow to the instrument. Do not place a hot soldering iron on the instrument. Do not pull the instrument with the power cord, connected probe, or connected test lead. Do not move the instrument when a probe is connected to a circuit being tested.
6
This product is subject to Directive 2002/96/EC of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste. Please utilize your local WEEE collection facilities in the disposition of this product and otherwise observe all applicable requirements.

Compliance Statements

Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with separate collection systems)
7
CE Declaration of Conformity
This instrument meets the requirements of 2006/95/EC Low Voltage Directive and 2004/108/EC Electromagnetic Compatibility Directive with the following standards.
Low Voltage Directive
- EN61010-1: 2001
EMC Directive
- EN 61000-3-2: 2006
- EN 61000-3-3: 1995+A1: 2001+A2: 2005
- EN 61000-4-2 / -3 / -4 / -5 / -6 / -11
- EN 61326-1: 2006
8
Refer to the user manual for warning information to avoid hazard or personal injury and prevent damage to instrument.
Electric Shock hazard
Alternating current (AC)
Chassis (earth ground) symbol.
Ground terminal
On (Power). This is the In position of the power switch when instrument is ON.
Off (Power). This is the Out position of the power switch when instrument is OFF.
Power Switch (On/Off). This is the power switch located in front of the instrument.
CAUTION indicates a hazardous situation which, if not avoided, will result in minor or moderate injury
WARNING indicates a hazardous situation which, if not avoided, could result in death or serious injury
DANGER indicates a hazardous situation which, if not avoided, will result in death or serious injury.

Safety Symbols

9
Contents
Safety Summary ............................................................................................. 2
Compliance Statements ......................................................................................................... 7
Safety Symbols ....................................................................................................................... 9
1 General Information ......................................................................... 13
1.1 Product Overview ......................................................................................................... 13
1.2 Package Contents .......................................................................................................... 13
1.3 Front Panel Overview .................................................................................................... 14
1.4 Rear Panel Overview ..................................................................................................... 15
1.5 Display Overview........................................................................................................... 16
2 Getting Started ................................................................................. 17
2.1 Input Power and Fuse Requirements ............................................................................ 17
2.2 Connectors .................................................................................................................... 18
2.3 Output Connections ...................................................................................................... 18
2.4 Preliminary Check ......................................................................................................... 19
3 Operating Instructions ...................................................................... 21
3.1 Menu Keys ..................................................................................................................... 21
3.1.1 WAVEFORM Keys ............................................................................................. 21
3.1.2 MODE Key ........................................................................................................ 22
3.1.3 UTILITY Key ...................................................................................................... 24
3.1.4 SWEEP Key ....................................................................................................... 25
3.1.5 MODULATION Key............................................................................................ 26
3.1.6 ARBITRARY Key ................................................................................................ 31
3.1.7 Arbitrary EDIT Menu ........................................................................................ 33
3.2 ON Key........................................................................................................................... 36
3.3 Cursor Keys .................................................................................................................... 36
3.4 Rotary Input Knob ......................................................................................................... 36
3.5 Power-On Settings ........................................................................................................ 36
3.6 Memory......................................................................................................................... 37
3.7 Displaying Errors ........................................................................................................... 37
3.8 Quick Start ..................................................................................................................... 38
3.8.1 Selecting a Standard Waveform ..................................................................... 38
10
3.8.2 Setting the Output .......................................................................................... 38
3.8.3 Using Voltage Offset ....................................................................................... 38
3.8.4 Storing and Recalling a Waveform Generator Setup ...................................... 39
3.8.5 Creating an Arbitrary Waveform ..................................................................... 40
3.8.6 Entering Individual Data Points ....................................................................... 40
3.8.7 Setting the Arbitrary Frequency ...................................................................... 41
3.8.8 Setting the Amplitude ...................................................................................... 41
3.8.9 Executing an Arbitrary Waveform .................................................................... 42
4 Programming .................................................................................... 43
4.1 Overview ...................................................................................................................... 43
4.1.1 Connecting to USB (Virtual COM) Interface..................................................... 43
4.1.2 USB (Virtual COM) Settings ............................................................................. 46
4.2 Device States ................................................................................................................ 47
4.2.1 Local State (LOCS) ............................................................................................ 47
4.2.2 Remote State (REMS) ....................................................................................... 47
4.3 Message Exchange Protocol ......................................................................................... 47
4.3.1 The Input Buffer ............................................................................................... 47
4.3.2 The Output Queue ........................................................................................... 48
4.3.3 Response Messages ......................................................................................... 48
4.4 Instrument Identification ............................................................................................. 48
4.5 Instrument Reset .......................................................................................................... 48
4.6 Command Syntax ........................................................................................................ 48
4.6.1 General Command Structure ........................................................................... 48
4.7 Status Reporting ........................................................................................................... 51
4.7.1 The Error Queue ............................................................................................... 51
4.7.2 Error Codes ...................................................................................................... 51
4.8 Common Commands ................................................................................................... 54
4.8.1 System Data Commands .................................................................................. 54
4.8.2 Internal Operation Commands ........................................................................ 55
4.8.3 Device Trigger Commands ............................................................................... 55
4.8.4 Stored Settings Commands .............................................................................. 55
4.9 Instrument Control Commands ................................................................................... 56
4.9.1 Default Subsystem ........................................................................................... 57
4.9.2 Arbitrary Subsystem ........................................................................................ 67
11
5 Troubleshooting Guide ...................................................................... 72
6 Specifications .................................................................................... 74
SERVICE INFORMATION ............................................................................... 78
LIMITED THREE-YEAR WARRANTY ................................................................ 79
12

1 General Information

1.1 Product Overview

The 4047B is a versatile dual-channel function generator with arbitrary capabilities. Implemented using a DDS (direct digital synthesis) architecture, this instrument generates stable and precise sine, square, triangle, and arbitrary waveforms. The unit also provides linear and logarithmic sweep for users needing sweep capability. An auxiliary TTL output at the generator’s set frequency is available to synchronize external devices. The instrument can also be remotely operated via the USB interface and is SCPI compatible.

1.2 Package Contents

Please inspect the instrument mechanically and electrically upon receiving it. Unpack all items from the shipping carton, and check for any obvious signs of physical damage that may have occurred during transportation. Report any damage to the shipping agent immediately. Save the original packing carton for possible future reshipment. Every generator is shipped with the following contents:
4047B DDS function generator AC Power Cord USB (type A to B) interface cable Certificate of Calibration
Verify that all items above are included in the shipping container. If anything is missing, please contact B&K Precision.
13
Power On/Off Switch
Function Keys (F1-F4)
Mode Key
Sweep Key
Channel Key
Modulation Key
Utility Key
Enter Key
Output BNCs
Output On/Off Key
Rotary Knob and Cursor Keys
Units Keys
Numeric Keypad
Waveform Keys
LCD Display
1 2 3 4 5
6 7 8 9 10
11
12
13
14
15
1
2
3 5 8
11
12
13
15
14 9 10
4 6 7

1.3 Front Panel Overview

Front Panel Description
Figure 1 - Front Panel Overview
14
Modulation Input BNC
External Trigger Input and Counter BNC
CH1 Sync Output BNC
CH2 Sync Output BNC
USB Interface
Earth Ground
AC Power Connector and Fuse Box
Rear Cooling Fan
1 2 3
4
5 6 7
8
8
7
1
2 3 4
5
6
The front-panel controls select, display, and change parameter, function, and mode settings. Use the numerical keypad, rotary input knob and the cursor movement keys to enter data into the waveform generator.
To change a setting:
1. Press the key that leads to the parameter to change.
2. Move cursor using cursor keys to the appropriate position in the numeric field (if applicable).
3. Use the rotary input or the numerical keyboard to change the value of the displayed parameter. Changes take effect immediately.

1.4 Rear Panel Overview

Rear Panel Description
Figure 2 - Rear Panel Overview
15
Wave Type
Trigger Mode
Frequency
Menu Parameters Values
Menu Options
Menu Title
General Waveform Display
Channel Selection Indicator
1 2 3 4 5
6 7 8
5
4 8 2 1 3 7 6

1.5 Display Overview

The function generator has a color LCD display that can display up to 400 x 240 dots. When powering on the unit, sine waveform is selected and current settings will appear in the display. The bottom of the display shows a menu (selectable with function keys) that corresponds to the function, parameter, or mode display selected.
Display Description
Figure 3 - Display Overview
16
The included AC power cord is safety certified for this instrument operating in rated range. To change a cable or add an extension cable, be sure that it can meet the required power ratings for this instrument. Any misuse with wrong or unsafe cables will void the warranty.

2 Getting Started

Before connecting and powering up the instrument, please review and go through the instructions in this chapter.

2.1 Input Power and Fuse Requirements

Input Power
The instrument has a universal AC input that accepts line voltage and frequency input within:
90 V to 264 V, 48 Hz to 66 Hz
The maximum power consumption is 30 VA. Use a slow blow fuse of 1A, UL/CSA approved as indicated on the rear panel of the instrument.
Before connecting to an AC outlet or external power source, be sure that the power switch is in the OFF position and verify that the AC power cord, including the extension line, is compatible with the rated voltage/current and that there is sufficient circuit capacity for the power supply. Once verified, connect the cable firmly.
Fuse Requirements
The instrument power fuse is located in the AC input plug. To access the fuse, first disconnect the power cord and then remove the fuse box.
There is a 1A, 250V rated slow blow fuse at the AC input. Should the fuse ever get blown, follow the steps below to replace:
1. Locate the fuse box next to the AC input connector in the rear panel.
2. With a small flat blade screwdriver, insert into the fuse box slit to pull and slide out the fuse
box as indicated below.
3. Check and replace fuse if necessary.
17
Fuse box slit
Fuse box
Check/Remove Fuse
Figure 4 - Fuse Replacement

2.2 Connectors

The function generator has two BNC connectors on the front panel where you can connect coaxial cables. These coaxial cables serve as carrier lines for output signals delivered from the function generator.
Output Connector
Use this connector to transfer the main output signal from the function generator. One connector is available for each channel.
Trig In Connector
Use this connector to apply an external trigger or gate signal, depending on the waveform generator setting, to the generator. When the built-in frequency counter is enabled, this connector becomes an input for the counter. The connector is common to both channels acting as the MODE selected for each channel.
Sync Out Connector Use this connector to output a positive TTL sync pulse generated at each waveform cycle. One connector is available for each channel.
Modulation In Connector
5V p-p signal for 100% modulation, 10Kohms input impedance with DC - >20 KHz bandwidth. The connector is common to both channels acting as the MODULATION function selected for each channel.

2.3 Output Connections

The waveform generator output circuits operate as a 50 Ω voltage source working into a 50 Ω load. At higher frequencies, non-terminated or improperly terminated output causes aberrations on the output waveform. In addition, loads less than 50 Ω reduce the waveform amplitude, while loads more than 50 Ω increase waveform amplitude.
18
Excessive distortion or aberrations caused by improper termination are less noticeable at lower frequencies, especially with sine and triangle waveforms. To ensure waveform integrity, follow these precautions:
1. Use good quality 50 Ω coaxial cable and connectors.
2. Make all connections tight and as short as possible.
3. Use good quality attenuators if it is necessary to reduce waveform amplitudes applied to sensitive circuits.
4. Use termination or impedance-matching devices to avoid reflections.
5. Ensure that attenuators and terminations have adequate power handling capabilities.
If there is a DC voltage across the output load, use a coupling capacitor in series with the load. The time constant of the coupling capacitor and load must be long enough to maintain pulse flatness.
Impedance Matching
If the waveform generator is driving a high impedance, such as a 1 MΩ input impedance (paralleled by a stated capacitance) of an oscilloscope vertical input, connect the transmission line to a 50 Ω attenuator, a 50 Ω termination and to the oscilloscope input. The attenuator isolates the input capacitance of the device and terminates the waveform generator properly.

2.4 Preliminary Check

The 4047B is intended for bench use. The instrument includes a front feet tilt mechanism for optimum panel viewing angle. The instrument does not require special cooling when operated within conventional temperature limits. It may be installed in a closed rack or test station if proper air flow can assure removing about 15 W of power dissipation.
Complete the following steps to verify that the generator is ready for use.
1. Verify AC Input Voltage
Verify and check to make sure proper AC voltages are available to power the instrument. The AC voltage range must meet the acceptable specification as explained in 2.1 Input Power and Fuse Requirements.
2. Connect Power
Connect AC power cord to the AC receptacle in the rear panel and press the power switch to the ON position to turn ON the instrument. The instrument will have a boot screen while loading, after which the main screen will be displayed.
3. Self Adjust
At power-on, the waveform generator performs a diagnostic self-test procedure to check
itself for errors. If it finds an error, an error code and text will appear in the display window. Other error codes appear when you enter an invalid front-panel setting. For more information on error codes, see the 3.7 Displaying Errors. When the waveform generator
19
finishes the diagnostic self-test routine, it enters the local state (LOCS) and assumes power-on default settings.
Output Check
Follow the steps below to do a quick check of the settings and waveform output.
1. Turn on the instrument and set the instrument to default settings. To set to default, press
Setups, press Recall, and select 0 Default Setup. The instrument will set both channels with the following parameters: Waveform Shape: Sine Frequency: 1.000000000 kHz Amplitude: 5 .000 Vpp Offset: 0.000 Vdc Phase: 0.0 ° Output Impedance: 50 OHM
2. Connect the BNC output of CH1 into an oscilloscope.
3. Press the On button on top of CH1 output BNC to turn on the output and observe a sine
wave with the parameters above.
4. Press the Freq option in the menu and use the rotary knob or the numeric keypad to
change frequency. Observe the changes on the oscilloscope display.
5. Press the Ampl option in the menu and use the rotary knob or the numeric keypad to
change the amplitude. Observe the changes on the oscilloscope display.
6. Press the Offset option in the menu and use the rotary knob or the numeric keypad to
change the DC offset. With the oscilloscope set for DC coupling, observe the changes on the display.
7. Now, connect the BNC output of CH2 into an oscilloscope and follow steps 3 to 6 to check
its output.
20

3 Operating Instructions

3.1 Menu Keys

These keys select the main menus for displaying or changing a parameter, function, or mode.

3.1.1 WAVEFORM Keys

These keys select the waveform output and display the waveform parameter menu (frequency, amplitude and offset).
F1: Frequency – Selects and displays the frequency. Change the frequency setting using the
cursor keys, rotary knob, or numerical keys.
F2: Amplitude – Selects and displays the amplitude. Change the amplitude setting using the
cursor keys, rotary knob, or numerical keys.
F3: Offset – Selects and displays the offset parameter. Change the offset by using the
cursor keys, rotary knob, or numerical keys. If a certain setting cannot be produced, the waveform generator will display a “Setting Conflict” message.
Amplitude and offset settings interact and are bound by hardware restrictions. To obtain the desired waveform, the amplitude and offset must satisfy the following formula:
Figure 5 - Sine Menu
21
(Vp-p)/2 + |offset| <= 5 volts
F4: Symmetry – When the Square or Triangle waveforms are selected, the SYMMETRY (duty
cycle) parameter is available. Change the symmetry (Triangle) or duty cycle (Square) by using the cursor keys, rotary knob, or numerical keys. If a certain setting cannot be produced, the waveform generator will display a warning message.
Figure 6 - Square Menu

3.1.2 MODE Key

Selects the output mode: CONT (Continuous), TRIG (Triggered), GATE (Gated), and BRST (Burst). To select the output mode, press MODE, then press the function key that corresponds to the desired Mode menu option, as shown:
Figure 7 - Mode Menu
22
F1: Continuous – Selects continuous output.
F2: Triggered – Triggers one output cycle of the selected waveform for each trigger event.
F3: Gated – Triggers output cycles as long as the trigger source asserts the gate signal.
F4: Burst – Triggers ‘N’ number of output cycles for each trigger event, where N ranges
from 2 to 65,535.
After selecting the TRIGGERED, GATED, or BURST menu, the trigger source menu is available:
Figure 8 - Trigger Menu
F1: Manual – Selects manual as the trigger source. To trigger the waveform generator, press
this MANUAL trigger button again.
F2: Internal – Selects the internal trigger generator as the trigger source. Change the
internal trigger rate displayed with the rotary input knob.
F3: External – Selects the external trigger signal as the trigger source. The trigger source is
supplied through the TRIG IN connector.
In BURST MODE, the F4 key displays N-Burst, representing the number of burst cycles to output with each trigger. The N value can be changed from 2 to 65,535.
23
Loading...
+ 53 hidden pages