BK Precision BK Precision 2194 Manual

5 Commonwealth Ave Woburn, MA 01801 Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Digital Storage Oscilloscope
User Manual
Model 2194







4

I ,.
.








Norm
_
gle
Sin
t
 



al
,




Safety Summary
The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
• Read and understand the safety and operational information in this manual.
• Apply all the listed safety precautions.
• Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating the instrument at an incorrect line voltage will void the warranty.
• Make all connections to the instrument before applying power.
• Do not operate the instrument in ways not specied by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design, manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply with these requirements.
2
Category rating
The IEC 61010 standard denes safety category ratings that specify the amount of electrical energy available and the voltage impulses that may occur on electrical conductors associated with these category ratings. The category rating is a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of the circuit to be tested, which denes the voltage impulses expected and required insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected to the
mains supply. The voltages in the environment are typically derived from a limited-energy trans­former or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the mains
supply at a standard wall outlet or similar sources. Example measurement environments are portable tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be connected to the mains
installation of a building. Examples are measurements inside a building’s circuit breaker panel or the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to the primary
power entering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what is specied in this manual for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the instrument’s category rating to maintain the instrument’s category rating. Failure to do so will lower the category rating of the measuring system.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be 115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This instrument is grounded through the ground conductor of the supplied, three-conductor AC line power cable. The power cable must be plugged into an approved three-conductor electrical outlet. The power jack and mating plug of the power cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive parts (including control knobs) may provide an electric shock. Failure to use a properly-grounded approved outlet and the recommended three-conductor AC line power cable may result in injury or death.
3
Unless otherwise stated, a ground connection on the instrument’s front or rear panel is for a reference of potential only and is not to be used as a safety ground. Do not operate in an explosive or ammable atmosphere.
Do not operate the instrument in the presence of ammable gases or vapors, fumes, or nely-divided particulates.
The instrument is designed to be used in oce-type indoor environments. Do not operate the instrument
• In the presence of noxious, corrosive, or ammable fumes, gases, vapors, chemicals, or nely-divided particulates.
• In relative humidity conditions outside the instrument’s specications.
• In environments where there is a danger of any liquid being spilled on the instrument or where any liquid can condense on the instrument.
• In air temperatures exceeding the specied operating temperatures.
• In atmospheric pressures outside the specied altitude limits or where the surrounding gas is not air.
• In environments with restricted cooling air ow, even if the air temperatures are within specications.
• In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature range is 0∘C to 40∘C and 20% to 80% relative humidity, with no condensation allowed. Measurements made by this instrument may be outside specications if the instrument is used in non-oce-type environments. Such environments may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise, strong electric elds, or strong magnetic elds.
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the instrument, remove the instrument’s power cord, remove the instrument from service, label it as not to be operated, and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual. Not for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a life-support device or system.
4
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must be made by qualied service-trained maintenance personnel who are aware of the hazards involved when the instrument’s covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may exist when the covers are removed. To avoid injuries, always disconnect the power cord from the instrument, disconnect all other connections (for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there are no hazardous voltages present on any conductors by measurements with a properly-operating voltage-sensing device before touching any internal parts. Verify the voltage-sensing device is working properly before and after making the measurements by testing with known-operating voltage sources and test for both DC and AC voltages. Do not attempt any service or adjustment unless another person capable of rendering rst aid and resuscitation is present.
Do not insert any object into an instrument’s ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit exists.
Fuse replacement must be done by qualied service-trained maintenance personnel who are aware of the instrument’s fuse requirements and safe replacement procedures. Disconnect the instrument from the power line before replacing fuses. Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings specied in this manual or on the back of the instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a re. Failure to use the specied fuses will void the warranty.
Servicing
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to B&K Precision for service and repair to ensure that safety and performance features are maintained.
For continued safe use of the instrument
• Do not place heavy objects on the instrument.
• Do not obstruct cooling air ow to the instrument.
• Do not place a hot soldering iron on the instrument.
• Do not pull the instrument with the power cord, connected probe, or connected test lead.
• Do not move the instrument when a probe is connected to a circuit being tested.
Working Environment
5
Environment
This instrument is intended for indoor use and should be operated in a clean, dry environment.
Temperature
Operating: 0℃ to +40℃ Non-operation:-20℃ to +60℃
Note:
Direct sunlight, radiators, and other heat sources should be taken into account when assessing the ambient temperature.
Humidity
Operating: 85% RH, 40 ℃, 24 hours Non-operating: 85% RH, 65 ℃, 24 hours
Altitude
Operating: less than 3 Km Non-operation: less than 15 Km
Installation (overvoltage) Category
This product is powered by mains conforming to installation (overvoltage) category II.
Degree of Pollution
The oscilloscopes may be operated in environments of Pollution Degree II.
Note:
Degree of Pollution II refers to a working environment which is dry and non-conductive pollution occurs. Occasional temporary conductivity caused by condensation is expected.
IP Rating
IP20 (as dened in IEC 60529).
Compliance Statements
Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with separate collection systems)
This product is subject to Directive 2002/96/EC of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste. Please utilize your local WEEE collection facilities in the disposition of this product and otherwise observe all applicable requirements.
Safety Symbols
6
Symbol Description
indicates a hazardous situation which, if not avoided, will result in death or serious injury.
indicates a hazardous situation which, if not avoided, could result in death or serious injury
indicates a hazardous situation which, if not avoided, will result in minor or moderate injury
Refer to the text near the symbol.
Electric Shock hazard
Alternating current (AC)
Chassis ground
Earth ground
This is the In position of the power switch when instrument is ON.
This is the Out position of the power switch when instrument is OFF.
is used to address practices not related to physical injury.
Contents
1 General Information 11
1.1 Product Overview 11
1.2 Features 11
1.3 Contents 11
1.4 Dimensions 12
1.5 Front Panel Overview 13
1.6 Rear Panel Overview 14
1.7 Display Overview 15
2 Getting Started 16
2.1 Input Power Requirements 16
2.2 Fuse Requirements and Replacement 16
2.3 Preliminary Check 17
2.3.1 Verify AC Input Voltage 17
2.3.2 Connect Power 17
2.3.3 Self-Test 18
2.3.4 Self-Cal 18
2.3.5 Check Model and Firmware Version 18
2.3.6 Function Check 19
2.4 Probe Safety 20
3 Vertical Controls 22
3.1 Enable Channel 22
3.2 Channe Menu 22
3.2.1 Channel Coupling 23
3.2.2 Bandwidth Limit 23
3.2.3 Adjust 23
3.2.4 Probe 24
3.2.5 Unit 24
3.2.6 Deskew 25
3.2.7 Invert 25
3.2.8 Oset 25
3.2.9 Trace Visible/Hidden 26
4 Horizontal Control 27
4.1 Horizontal Scale 27
4.2 Zoom 27
4.3 Roll Mode 28
4.4 Trigger Delay 28
5 Sample Control 29
5.1 Run Control 29
5.2 Sampling Theory 29
5.3 Sample Rate 29
5.4 Bandwidth and Sample Rate 30
5.5 Memory Depth 31
5.6 Sampling Mode 32
5.7 Interpolation Method 32
5.8 Acquisition Mode 33
5.9 Average 35
5.10 Eres Acquisition 35
5.11 Horizontal Format 35
5.12 Sequence Mode 37
6 Trigger 38
6.1 Trigger Source 38
6.2 Trigger Mode 39
6.3 Trigger Level 40
6.4 Trigger Coupling 41
6.5 Trigger Holdo 42
6.6 Noise Rejection 43
6.7 Trigger Types 43
6.7.1 Edge Trigger 44
6.7.2 Slope Trigger 45
6.7.3
46
6.7.4 Video Trigger 48
6.7.5 Window Trigger 51
6.7.6 Interval Trigger 54
6.7.7 Dropout Trigger 56
6.7.8 Runt Trigger 58
6.7.9 Pattern Trigger 60
7 Serial Trigger and Decode 62
7.1 I2C Trigger and Serial Decode 62
7.1.1 Setup for I2C Signals 62
7.1.2 I2C Trigger 63
7.1.3 I2C Serial Decode 66
7.2 SPI Trigger and Serial Decode 67
7.2.1 Setup for SPI Signals 67
7.2.2 SPI Trigger 71
7.2.3 SPI Serial Decode 72
7.3 UART Trigger and Serial Decode 73
7.3.1 Setup for UART Signals 73
7.3.2 UART Trigger 74
7.3.3 UART Serial Decode 75
7.4 CAN Trigger and Serial Decode 77
7.4.1 Setup for CAN Signals 77
7.4.2 CAN Trigger 77
7.4.1 CAN Serial Decode 79
7.5 LIN Trigger and Serial Decode 80
7.5.1 Setup for LIN Signals 80
7.5.2 LIN Trigger 81
7.5.1 Interpreting LIN Decode 83
8
8 Reference Waveform 84
8.1 Save REF Waveform to Internal Memory 84
8.2 Display REF Waveform 84
8.3 Adjust REF Waveform 85
8.4 Clear Ref Waveform 85
9 Math 86
9.1 Units for Math Waveforms 86
9.2 Math Operators 87
9.2.1 Addition or Subtraction 87
9.2.2 Multiplication and Division 88
9.2.3 FFT Operation 89
9.3 Math Function Operation 93
9.3.1 Dierentiate 93
9.3.2 Integrate 94
9.3.3 Square Root 94
10 Cursors 96
10.1 X Cursors 96
10.2 Y Cursors 97
10.3 Make Cursor Measurements 98
11 Measure 99
11.1 Type of Measurement 99
11.1.1 Voltage Measurements 99
11.1.2 Time Measurements 101
11.1.3 Delay Measurements 101
11.2 Automatic Measurement 102
11.3 All Measurement 104
11.4 Gate Measurement 105
11.5 Clear Measurement 105
12 Display 106
12.1 Display Type 106
12.2 Color Display 107
12.3 Persistence 108
12.4 Clear Display 109
12.5 Grid Type 109
12.6 Intensity 109
12.7 Grid Brightness 109
12.8 Transparence 110
9
13 Save and Recall 111
13.1 Save Type 111
13.2 Internal Save and Recall 112
13.3 External Save and Recall 113
13.4 Disk Management 115
13.4.1 Create a New File or Folder 115
13.4.2 Delete a File or Folder 116
13.4.3 Rename a File or Folder 116
14 System Settings 117
14.1 View System Status 117
14.2 Self Cal 118
14.3 Quick-Cal 118
14.4 Sound 119
14.5 Language 119
14.6 Pass/Fail Test 119
14.6.1 Set and Perform a Pass/Fail Test 120
14.6.2 Save and Recall Test Mask 121
14.7 IO Set 123
14.7.1 LAN 123
14.7.2 USB Device 124
14.8 Update Firmware and Conguration 124
14.9 Do Self-Test 125
14.9.1 Screen Test 125
14.9.2 Keyboard Test 126
14.9.3 LED Test 127
14.10 Screen Saver 128
14.11 Reference Position 129
14.12 Power On Line 129
15 Search 130
15.1 Setting 130
15.2 Results 131
10
16 Navigate 133
16.1 Time Navigate 133
16.2 History Frame Navigate 133
16.3 Search Event Navigate 133
17 History 134
18 Factory Setup 135
19 Troubleshooting 136
20 Service Information 138
21 LIMITED THREE-YEAR WARRANTY 139

General Information

1.1 Product Overview

Figure 1.1 2194 Front View
The B&K Precision 2194 digital storage oscilloscope (DSO) is a portable benchtop instrument used for making measure­ments of signals and waveforms.
This oscilloscope provides 100 MHz of bandwidth in a 4-channel conguration with a maximum sample rate of 1 GSa/s and best-in class memory depth of 14 Mpts.

1.2 Features

– 4 channels with 100 MHz bandwidth
– Single channel real-time sampling rate of up to 1 GSa/s
– 14 Mpts memory depth
– Standard USB host, USBTMC device, and LAN ports

1.3 Contents

Inspect the instrument mechanically and electrically upon receiving it. Unpack all items from the shipping carton, and check for any obvious signs of physical damage that may have occurred during transportation. Report any damage to the shipping agent immediately. Save the original packing carton for possible future reshipment. Every oscilloscope is shipped with the following contents:
– 1 x 2194 Digital Storage Oscilloscope
– AC Power Cord
– USB type A to type B cable.
– 4 x 1:1/10:1 Passive Oscilloscope Probes
– Certicate of Calibration
– Test Report
General Information 12
Note:
Ensure the presence of all the items above. Contact the distributor if anything is missing.

1.4 Dimensions

The 2194 digital storage oscilloscope’s dimensions are approximately: 312.00 mm (12.28 in) x 151.00 mm (5.94 in) x
132.60 mm (5.22 in) (W x H x D).
Figure 1.2 Front View Dimensions
Figure 1.3 Top View Dimensions
General Information 13

1.5 Front Panel Overview

The front panel interface allows for control of the unit.
Figure 1.4 Front Panel
Item Name Description
1 LCD Display Visual presentation of the device function and measurements.
2 Intensity Adjust Universal knob.
3
4 Numeric Keypad Used to enter precise values
5 Rotary Knob Used to navigate menus or congure parameters
6 Navigation Keys
7 CH 2 Terminals Serves as output or input terminals of CH 2 depending on the set functionality
8 Function Keys Frequently used function such as Home, Trig, Menu, ESC, and On/O keys
9 CH 1 Terminals Serves as output or input of CH 1 depending on the set functionality
10 Softkeys Used to invoke any functions displayed above them.
11 Power Switch Power the unit ON or OFF
12 Horizontal Control
13 Auto Set the trigger mode to auto.
14 Menu On/O Enable/disable the menu bar.
15 Softkeys Used to invoke any functions displayed above them.
16 Print Shortcut key for the save function. 17 Input Channels Input channels (1 MΩ BNC)
18 Power Button Power the unit ON or OFF.
19 USB Host Port USB port used to connect ash drives. (Type A)
20
Common
Function Keys
Probe
Compensation
Used to invoke the functions displayed above them.
Used to navigate menus. The enter key can be used to select a menu or enter a para­meter
Probe compensation/ground terminal.
Table 1.1 Front Panel
General Information 14

1.6 Rear Panel Overview

Figure 1.5 Rear Panel Overview
Item Name Description
1 Handle Handle for easy carrying of the instrument.
2 Safety Lock Hole
3 LAN Connect an ethernet cable to remotely control the unit over the network.
4 USB Interface Connect a USB type B to type A to remotely control the unit.
5
6
Pass/Fail or Trigger Out
AC Power Input
& Fuse Box
Locks the instrument to a xed location using the security lock via the lock hole. The lock is not included.
Output a signal that reects the current waveform capture rate of the oscilloscope at each trigger or a pass/fail test pulse.
Houses the fuse as well as the AC input .
Table 1.2 Rear Panel
General Information 15

1.7 Display Overview

Figure 1.6 Display Overview
Item Name Description
1 Trigger Status Displays the trigger status.
2
3
4 Menu Bar Displays the available options in the selected menu.
USB Host
Port Indicator
LAN Port
Indicator
Indicates that a USB is connected to the instrument.
Indicates the status of the LAN connection.
Table 1.3 Display Overview

Getting Started

Before connecting and powering up the instrument, review the instructions in this chapter.

2.1 Input Power Requirements

The oscilloscope has a universal AC input that accepts line voltage and frequency input within:
100 - 240 V (+/- 10%), 50/60 Hz (+/- 5%)
100 - 127 B, 400 Hz
50 W Max
Before connecting to an AC outlet or external power source, be sure that the power switch is in the OFF position and verify that the AC power cord, including the extension line, is compatible with the rated voltage/current and that there is sucient circuit capacity for the power supply. Once veried, connect the cable rmly.
The included AC power cord is safety certied for this instrument operating in rated range. To change a cable or add an extension cable, be sure that it can meet the required power ratings for this instrument. Any misuse with wrong or unsafe cables will void the warranty.
SHOCK HAZARD:
The power cord provides a chassis ground through a third conductor. Verify that your power outlet is of the three conductor type with the correct pin connected to earth ground.

2.2 Fuse Requirements and Replacement

For continued re protection at all line voltages replace only with a 1.25 A / 250 V "F" rated, 5 x 20 mm fuse.
For safety, no power should be applied to the instrument while changing line voltage operation. Disconnect all cables connected to the instrument before proceeding.
Getting Started 17
Check and/or Change Fuse
– Locate the fuse box next to the AC input connector in the rear panel. (See gure 1.5)
– Insert a small athead screwdriver into the fuse box slit to pull and slide out the fuse box as indicated below.
– Check and replace fuse if necessary. (See gure 2.1)
Figure 2.1 Fuse Removal
Any disassembling of the case or changing the fuse not performed by an authorized service technician will void the warranty of the instrument

2.3 Preliminary Check

Complete the following steps to verify that the oscilloscope is ready for use.

2.3.1 Verify AC Input Voltage

Verify proper AC voltages are available to power the instrument.
The AC voltage range must meet the acceptable specication stated in section Input Power Requirements.

2.3.2 Connect Power

Connect the AC power cord to the AC receptacle in the rear panel and press the power switch to turn on the instrument.
The instrument will have a boot up screen while loading, after which the main screen will be displayed.
Getting Started 18

2.3.3 Self-Test

The instrument has 3 self-test option to test the screen ,keyboard, and the LED back light. To perform the self-test, please refer to the Self Test section for further instructions.

2.3.4 Self-Cal

Self option runs an internal self-calibration procedure that will check and adjust the instrument. To perform the self­calibration refer to the Self-Calibration section for further instructions.

2.3.5 Check Model and Firmware Version

The model and rmware version can be veried from within the menu system. To view the model and rmware version: Press the Utility button and use the softkeys to select the System Status option. The following information will be displayed:
Startup Times
Software Version
Uboot-Os Version
FPGA Version
Hardware Version
Product Type
Serial NO
Press the Single key to exit.
Figure 2.2 System Status
Getting Started 19

2.3.6 Function Check

Follow the steps below to do a quick check of the oscilloscope’s functionality.
1. Power on the oscilloscope. Press Default Setup to show the result of the self-check.
– The probe default attenuation is 1X.
2. Set the switch to 1X on the probe and connect the probe to channel 1.
– To do this align the slot in the probe connector with the key on the CH1 BNC, push to connect, and twist to the
right to lock the probe in place.
– Connect the probe tip and reference lead to the Probe Comp connectors.
3. Press the AUTO button to show the square wave with 1 kHz frequency and 3V peak to peak .
Figure 2.3 3 Vpp Square Wave
4. Repeat steps 1 to 3 for the remaining channels.
Getting Started 20

2.4 Probe Safety

A guard around the probe body provides a nger barrier for protection from electric shock.
Figure 2.4 Probe
Connect the probe to the oscilloscope and connect the ground terminal to the ground before you take any measurements.
Shock Hazard:
To avoid electric shock when using the probe, keep ngers behind the guard on the probe body. To avoid electric shock while using the probe, do not touch metallic portions of the probe head while it is connected to a voltage source. Connect the probe to the oscilloscope and connect the ground terminal to ground before you take any measurements.
Getting Started 21
Probe Attenuation
Probes are available with various attenuation factors which aect the vertical scale of the signal. The Probe Check function veries that the probe attenuation option matches the attenuation of the probe.
Press CH 1 once to open the channel menu. Use the softkeys to navigate to page 1/2 and select the Probe option. Select the probe option that matches the attenuation of the probe.
Note:
The default setting for the Probe option is 1 X.
Verify that the attenuation switch on the probe matches the Probe option in the oscilloscope. Switch settings are 1 X and 10 X.
Probe Compensation
Before taking any measurements using a probe, verify the compensation of the probe and adjust it to match the channel inputs. To match your probe to the input channel:
1. Set the channel’s probe attenuation to 10X.
– Press the CH # key corresponding to the channel the probe is connected to.
– Use the softkeys to navigate to page 1.
– Use the softkeys to select Probe.
– Use the Intensity Adjust knob to select 10X.
2. Attach the probe tip to the Compensation Signal Output Terminal 3 V(Cal) connector and the reference lead to the Probe Ground terminal connector.
– Press the Auto Setup key to display the square wave.
3. Check the shape of the displayed waveform.
Undercompensated Correctly Compensated Overcompensated
Figure 2.5 Probe Compensation
4. If necessary, adjust your probe’s compensation trimmer pot.

Vertical Controls

3.1 Enable Channel

The 2194 provides 4 analog input channels. To enable a channel press the corresponding channel button located on the vertical controls.
The enabled channels can be veried on the right side of the display screen.
To disable a channel:
Press the correponding channel key. Once the key has been highlighted by the LED press the channel key again.
– Pressing the channel key of the currently selected channel once will
disable the channel.
and 2 Enabled

3.2 Channe Menu

Figure 3.2 shows the channel 2 menu that is displayed after pressing the CH 2 key.
Channel Menu Page 1/2
Channel Menu Page 2/2
Figure 3.2 CH 2 Menu
Channels 1
Figure 3.1 Enabled Channels
All Channels
Enabled
Vertical Controls 23

3.2.1 Channel Coupling

Coupling mode lters out the undesired signals.
Press the corresponding CH button, then use the softkeys to select Coupling.
Turn the Universal Knob to select the desired coupling method.
Note:
The current coupling method is displayed in the channel label at the right side of the screen. Pressing the Coupling softkey continuously switches between the available coupling method.
DC Coupling: The DC and AC components of the signal under test are both passed.
AC Coupling: The DC components of the signal under test are blocked.
GND Coupling: The DC and AC components of the signal under test are both blocked.

3.2.2 Bandwidth Limit

Sets the bandwidth limit to reduce display noise.
Press the CH button of the channel to be congured.
Use the softkeys to select BW Limit. (The bandwidth limit will alternate between Full and 20 M)
Full: The high frequency components of the signal under test can pass the channel.
20 M: The high frequency components exceeding 20 MHz are attenuated.

3.2.3 Adjust

Adjust the vertical scale sensitivity of the selected channel. The vertical scale is adjusted using the Vertical Variable Knob.
Press the CH button of the channel to be congured.
Use the softkeys to select Adjust. (The scale will alternate between Fine and Coarse)
If the amplitude of the input waveform is a little bit greater than the full scale under the current scale and the amplitude would be a little bit lower if the next scale is used, ne adjustment can be used to improve the amplitude of waveform display to view signal details.
Fine adjustment: Adjust the vertical scale within a relatively smaller range to improve vertical resolution.
– For example: 2 V/div, 1.98V/div, 1.96V/div, 1.94 V/div, ...1 V/div.
Coarse: Adjust the vertical scale in a 1-2-5 step.
– For example: 1 mV/div, 2 mV/div, 5 mV/div, 10 mV/div 200 mV/div, 500 mV/div,... 10 V/div.
The scale information in the channel label at the right side of the screen will change accordingly during the adjustment. The adjustable range of the vertical scale is related to the probe ratio currently set.
Vertical Controls 24
Note:
Push the Vertical Variable Knob to quickly switch between Coarse and Fine adjustment.

3.2.4 Probe

Sets the probe attenuation factor to match the type of probe being used.
1. Press the CH button of the channel to be congured.
2. Use the softkeys to select Probe.
3. Use the softkeys to select Probe once more.
4. Use the Universal Knob to select the probe attenuation.
Table 3.1 shows the probe attenuation factors.
Setting Description
0.1X .01 : 1
0.2X .02 : 1
0.5X .05 : 1 1X 1 : 1 2X 2 : 1 5X 5 : 1
10X 10 : 1
... ...
10000X 10000 : 1
Table 3.1 Attenuation Factor
To customize the probe attenuation factor:
Press the Probe softkey, select Custom, and then press the Custom softkey.
Use the Universal Knob to set the desired probe attenuation ratio.
The range is [1E-6,1E6].

3.2.5 Unit

Selects the amplitude display unit for the selected channel.
The available units are V and A.
1. Press the CH button of the channel to be congured.
2. Use the softkeys to navigate to page 2/2.
3. Use the softkeys to select Unit and alternate between V and A.
The default unit is V.
Vertical Controls 25

3.2.6 Deskew

Adjust the dierence of phase between the channel.
The Valid range of each channel is±100 ns.
1. Press the CH button of the channel to be congured.
2. Use the softkeys to navigate to page 2/2.
3. Use the softkeys to select Deskew.
4. Turn the Universal Knob to change deskew.
– Pushing the Universal Knob open the keypad.
Figure 3.3 Deskew Keypad

3.2.7 Invert

Invert the voltage values of the displayed waveform.
Inverting a channel aects how the channel is displayed, all the results of any math function selected, and measurement functions. To invert the waveform:
– Press the CH button of the channel to be congured.
– Use the softkeys to navigate to page 2/2.
– Use the softkeys to toggle Invert On and O.
3.2.8 Oset
Oset the vertical position of the displayed waveform.
The Valid range of each channel is±100 V.
1. Press the CH button of the channel to be congured.
2. Use the softkeys to navigate to page 2/2.
Vertical Controls 26
3. Use the softkeys to select Oset.
4. Turn the Universal Knob to change deskew.
– Pushing the Universal Knob open the keypad.
Figure 3.4 Oset Keypad
Note:
The Vertical Position Knob can be used to oset the waveform’s vertical position without having to enter the chan­nel’s menu. Pushing the Vertical Position Knob will zero vertical position.

3.2.9 Trace Visible/Hidden

Sets whether waveform of the selected channel is visible or hidden. To toggle between visible and hidden:
1. Press the CH button of the channel to be congured.
2. Use the softkeys to navigate to page 2/2.
3. Use the softkeys to select Trace.

Horizontal Control

4.1 Horizontal Scale

Turn the Horizontal Scale Knob to adjust the horizontal time base. Turning the knob clockwise reduces the horizontal time base. Turning the knob counterclockwise increases the time base.
The time base information at the upper left corner of the screen will change accordingly during the adjustment. The 2194 horizontal scale has a range from 2ns/div to 100s/div.
The Horizontal Scale Knob works (in the Normal time mode) while acquisitions are running or when they are stopped.
When in run mode, adjusting the horizontal scale knob changes the sample rate.
When stopped, adjusting the horizontal scale knob lets you zoom into acquired data.

4.2 Zoom

Zoom is a horizontally expanded version of the normal display. You can use Zoom to locate and horizontally expand part of the normal window for a more detailed (higher- resolution) analysis of signals.
Press the Horizontal Scale Knob to enable the zoom function, and press the button again to turn disable the function.
When Zoom enabled, the display divides in half. The top half of the display shows the normal time base window and the bottom half displays a faster Zoom time base window.
Figure 4.1 Zoom Mode
The area of the normal display that is expanded is outlined with a box and the rest of the normal display is ghosted. The box shows the portion of the normal sweep that is expanded in the lower half.
Horizontal Control 28
To change the time base for the zoom window, turn the Horizontal Scale Knob. The Horizontal Position Knob sets the left- to- right position of the zoom window.
The delay value, which is the time displayed relative to the trigger point is momentarily displayed in the upper right corner of the display when the Horizontal Position Knob is turned. Negative delay values indicate you’re looking at a portion of the waveform before the trigger event, and positive values indicate you’re looking at the waveform after the trigger event.
To change the time base of the normal window, disable Zoom, then turn the Horizontal Scale Knob.

4.3 Roll Mode

In Roll mode the waveform moves slowly across the screen from right to left. It operates on time base settings of 50 ms/div and slower. If the current time base setting is faster than the 50 ms/div limit, it will be set to 50 ms/div when Roll mode is entered.
In Roll mode there is no trigger. The xed reference point on the screen is the right edge of the screen and refers to the current moment in time. Events that have occurred are scrolled to the left of the reference point. Since there is no trigger, no pre- trigger information is available.
To enter Roll mode press the Roll button.
To stop the display, press the Run/Stop button.
To clear the display and restart an acquisition in Roll mode, press the Run/Stop button again.
To exit Roll mode press the Roll button.
Note:
Use Roll mode on low- frequency waveforms to yield a display much like a strip chart recorder.

4.4 Trigger Delay

Turn the Horizontal Position Knob on the front panel to adjust the trigger delay of the waveform. During the modication, waveforms of all the channels would move left or right and the trigger delay message at the upper-right corner of the screen would change accordingly. Press down this knob to quickly reset the trigger delay.
Changing the delay time moves the trigger point (solid inverted triangle) horizontally and indicates how far it is from the time reference point. These reference points are indicated along the top of the display grid.
All events displayed left of the trigger point happened before the trigger occurred. These events are called pre- trigger information, and they show events that led up to the trigger point.
Everything to the right of the trigger point is called post- trigger information. The amount of delay range (pre- trigger and post- trigger information) available depends on the time/div selected and memory depth.
The position knob works (in Normal time mode) while acquisitions are running or when they are stopped.

Sample Control

5.1 Run Control

Press the Run/Stop or the Single key to stop the sampling system of the scope.
Running: When the Run/Stop key is green, the oscilloscope is continuously acquiring data.
– To stop acquiring data, press the Run/Stop key.
– When the Run/Stop button is red, data acquisition is stopped.
– Red "Stop" text is displayed next to the trademark logo in the status line at the top of the display.
– To start acquiring data, press Run/Stop.
Single: Clears the display, the trigger mode is temporarily set to Normal (to keep the oscilloscope from auto- triggering immediately), the trigger circuitry is armed, the Single key is illuminated, and the oscilloscope waits until a user dened trigger condition occurs before it displays a waveform.
– When the oscilloscope triggers, the single acquisition is displayed and the oscilloscope is stopped (the Run/Stop
button is illuminated in red).
– Press the Single key again to clear the current waveform and acquire a new one.
Note:
The Single run control lets you view a single shot events without subsequent waveform data overwriting the display. Use Single when you want maximum memory depth for pan and zoom.

5.2 Sampling Theory

The Nyquist sampling theorem states that for a limited bandwidth (band- limited) signal with maximum frequency 𝑓 the equally spaced sampling frequency 𝑓𝑆must be greater than twice the maximum frequency 𝑓 the signal be uniquely reconstructed without aliasing.
𝑓
𝑀𝐴𝑋
= 𝐹
= 𝑁𝑦𝑞𝑢𝑖𝑠𝑡 𝑓𝑟𝑒𝑞𝑢𝑒𝑛𝑐𝑦(𝑓𝑁) = 𝑓𝑜𝑙𝑑𝑖𝑛𝑔 𝑓𝑟𝑒𝑞𝑢𝑒𝑛𝑐𝑦
𝑆/2
, in order to have
𝑀𝐴𝑋
𝑀𝐴𝑋

5.3 Sample Rate

The maximum sample rate of the oscilloscope is 1G Sa/s. The actual sample rate of the oscilloscope is determined by the horizontal scale. See section Horizontal Scale
The actual sample rate is displayed in the information area at the upper- right corner of the screen.
,
Figure 5.1 Actual Sample Rate
Sample Control 30
The sample rate aect the waveform in the following manner :
Waveform Aliasing: Aliasing occurs when the signal is under-sampled. The signal is distorted by low frequencies falsely being reconstructed from an insucient number of sample points.
Figure 5.2 Low Sample Rate

5.4 Bandwidth and Sample Rate

An oscilloscope’s bandwidth is typically described as the lowest frequency at which input signal sine waves are attenuated by 3 dB (-30% amplitude error).
The sampling theory requires the sample rate to be 𝑓𝑆= 2 ∗ 𝑓𝐵𝑊. However, the theory assumes there are no frequency components above 𝑓
𝑀𝐴𝑋(𝑓𝐵𝑊
in this case) and it requires a system with an ideal brick-wall frequency response.
Figure 5.3 Brick-Wall Frequency Response
Digital signals have frequency components about the fundamental frequency (Square waves are made up of sine waves at the fundamental frequency and an innite number of odd harmonics), and typically, for 500 MHz bandwidths and below, oscilloscopes have a Gaussian frequency response.
Sample Control 31
Figure 5.4 Bandwidth Limiting
In practice, an oscilloscope’s sample rate should be four or more times its bandwidth: 𝑓𝑠= 4𝑥𝑓𝐵𝑊. Doing so causes less aliasing, and aliased frequency components have a great amount of attenuation.

5.5 Memory Depth

Memory Depth refers to the number of waveform points that the oscilloscope can store in a single trigger sample. It
reects the storage ability of the sample memory. To set the Memory Depth:
1. Press the Acquire key.
2. Use the softkeys to select Mem Depth (Page 1/2).
3. Turn the Universal Knob to navigate the available option.
4. Push the Universal Knob to set the selected option.
The actual memory depth is displayed in the information area at the upper right corner of the screen.
Single Channel
Mode
14 k 7 k 3.5 k
140 k 70 k 35 k
1.4 M 700 k 350 k 14 M 7 M 3.5 M
Table 5.1 Maximum Storage Depth
Dual Channel
Mode
Three or Four
Channel Mode
Sample Control 32

5.6 Sampling Mode

The oscilloscope only supports real-time sampling. In this mode, the oscilloscope samples and displays waveform within a trigger event. The maximum real-time sample rate is 1GSa/s.
Press the RUN/STOP button to stop the sample, the oscilloscope will hold the last display. When stopped the vertical control and horizontal control are used to pan and zoom the waveform.

5.7 Interpolation Method

Under real-time sampling, the oscilloscope acquires the discrete sample values of the waveform being displayed. In general, a waveform of dots display type is very dicult to observe. In order to increase the visibility of the signal, the digital oscilloscope usually uses the interpolation method to display a waveform.
Interpolation method is a processing method to “connect all the sampling points”, and using some points to calculate the whole appearance of the waveform. For real-time sampling interpolation method is used, even if the oscilloscope in a single captures only a small number of sampling points. The oscilloscope can use interpolation method for lling out the gaps between points, to reconstruct an accurate waveform.
To set the Interpolation Method: Press the Acquire button on the front panel to enter the Acquire Function menu.
Press the Interpolation softkey to toggle between Sinx/x and X.
X: In the adjacent sample points are directly connected on a straight line. This method is only conned to rebuild on the edge of signals, such as square wave.
Figure 5.5 Interpolation Method X
Sample Control 33
Sinx/x Connects the sample points with a curve that has stronger versatility. Sinx interpolation method uses mathe­matical processing to calculation results in the actual sample interval. This method produces a more realistic regular shape than pure square wave and pulse.
When the sampling rate is 3 to 5 times the bandwidth of the system, the Sinx/s interpolation method is recommended.
Figure 5.6 Interpolation Method Sinx/x

5.8 Acquisition Mode

The acquisition mode is used to control how to generate waveform points from sample points. The oscilloscope provides the following acquisition mode: Normal, Peak Detect, Average and Eres.
To set the Acquisition Mode:
– Press the Acquire key to enter the Acquire Function menu.
– Press the Acquisition softkey to view the available options.
– Turn the Universal Knob to navigate the available acquisition mode.
– Push the Universal Knob to set the selected method.
Note:
Default mode is Normal Acquisition
Sample Control 34
Normal
In Normal mode the oscilloscope samples the signal at equal time interval to rebuild the waveform. For most waveforms, the best display eect can be obtained using this mode.
Figure 5.7 Normal Mode
Peak Detect
In Peak Detect mode, the oscilloscope acquires the maximum and minimum values of the signal within the sample interval to get the envelope of the signal or the narrow pulse of the signal that might be lost. Peak Detect can prevent aliasing, but the signal is more susceptible to noise.
In Peak Detect mode the oscilloscope can display all the pulses with pulse width at least as wide as the sample period.
Normal Acquisition Peak Detect Acquisition Pulse Width .1%
Figure 5.8 Peak Detect Acquisition
Sample Control 35

5.9 Average

In Average mode, the oscilloscope averages the waveforms from multiple samples to reduce the random noise of the input signal and improve the vertical resolution. The greater the number of averages is, the lower the noise will be and the higher the vertical resolution will be.
Average mode slows the response of the displayed waveform.
Normal Acquisition Average Acquisition
Figure 5.9 Average Acquisition

5.10 Eres Acquisition

Eres mode uses an ultra-sampling technique to average the neighboring points of the sample waveform to reduce the
random noise on the input signal and generate smoother waveforms.
Eres mode is generally used when the sample rate of the digital converter is higher than the storage rate of the acquisition memory.
Eres mode can be used on both single-shot and repetitive signals. It does not slow down the waveform update, but it does limit the oscilloscope’s real-time bandwidth, because it acts like a low-pass lter.
Note:
Average and Eres mode use dierent averaging methods. The former uses a Waveform Average method, the latter uses a Dot Average method.

5.11 Horizontal Format

The XY mode can be used to compare frequency and phase relationships between two signals.
XY mode can also be used with transducers to display strain versus displacement, ow versus pressure, volts versus current, or voltage versus frequency.
To enable XY mode:
1. Press the Acquire key.
2. Press the XY softkey to toggle XY mode.
Sample Control 36
Disabled: When XY is disabled, YT mode is set. In YT mode the display is set to a volt versus time graph, and signal events occurring before the trigger are plotted to the left of the trigger point and signal events after the trigger plotted to the right of the trigger point.
Enabled: XY mode changes the display to a volt versus volt graph. Channel’s 1 amplitude is plotted on the x-axis and channel’s
2 amplitude is plotted on the Y-axis.
The phase deviation between two signals with the same frequency can be measured via the Lissajous method. The gure below shows the measurement schematic diagram of the phase deviation.
Figure 5.10 Phase Deviation
) 𝑜𝑟 (
𝐶
) (where is the phase deviation angle between the two channels and the denitions of A,
𝐷
𝐴
) 𝑜𝑟±𝑎𝑟𝑐𝑠𝑖𝑛 (
𝐵
𝐶 𝐷
According to 𝑠𝑖𝑛(𝜃) = (
𝐴 𝐵
B, C and D are as shown in gure 5.10), the phase deviation angle is obtained using 𝜃 =±𝑎𝑟𝑐𝑠𝑖𝑛 (
If the principal axis of the ellipse is within quadrant I and III, the phase deviation angle obtained should be within quadrant I and IV, namely within (0 to π/2) or (3π /2 to 2π).
If the principal axis of the ellipse is within quadrant II and IV, the phase deviation angle obtained should be within quadrant II and III, namely within (π /2 to π) or (π to 3π/2).
).
Sample Control 37

5.12 Sequence Mode

Sequence is an acquisition mode, which does not display waveform during sampling process. It improves the waveform
capture rate, the maximal capture rate is 400,000 wfs/s. This allows the 2194 to capture small probability event eectively.
The 2194 runs and lls a memory segment for each trigger event. The oscilloscope continues to trigger until memory is lled, and then display the waveforms on the screen.
Note:
To use the sequence mode, Acquisition Mode must be set to Normal or Peak Detect and Horizontal Format must be set to YT mode.
To enable sequence mode:
1. Press the Acquire.
2. Press the Sequence softkey to enter the Sequence function menu.
3. Press the Max Segments softkey.
– Turn the Universal Knob to select the desired value.
– Pressing the Universal Knob will open to numeric keypad to facilitate the input of the desired value.
To replay the acquired sequence:
1. Press the History key.
2. Press the List softkey to enable the list display.
– The list records the acquisition time and ΔT of every frame.
3. Press the Frame No. softkey.
– Turn the Universal Knob to select the frame to be displayed.
4. Press the
5. Press the softkey to stop the replay.
6. Press the
softkey to replay the waveforms from the current frame to frame 1.
softkey to replay the waveforms from the current frame to the last frame.

Trigger

For triggering, certain condition can be set according to the requirement and when a waveform in the waveform stream meets this condition. Digital oscilloscope, display a waveform continuously regardless of the trigger stability, but only stable trigger can ensures stable display.
The trigger circuit ensures that every time base sweep or acquisition starts from the input signal and the user-dened trigger condition, namely every sweep is synchronous to the acquisition and the waveforms acquired overlap to display stable waveform. Figure 6.1 demonstrates how the position of the trigger event determines the reference time point and the delay setting.
Figure 6.1 Acquisition Memory
A trigger setup tells the oscilloscope when to acquire and display data. Trigger setting are based on the features of the input signal, therefore knowledge of the signal under test is required to quickly capture the desired waveform.

6.1 Trigger Source

The 2194 trigger source includes four analog channels and AC line.
To set the trigger source:
1. Press the Setup key to enter the Trigger menu.
2. Press the Source softkey to display the available trigger sources.
– Turn the Universal Knob or continuously press the Source softkey to navigate the available sources.
Available Trigger Sources Selected Trigger Source
Figure 6.2 Trigger Source
The currently selected trigger source is displayed at the upper right corner of the screen.
Analog Channel Input
Signals input from the analog channels can all be used as a trigger source.
Trigger 39
AC Line
The trigger signal is obtained from the AC power input of the oscilloscope. This kind of signals can be used to display the relationship between signals (such as illuminating device) and power (power supply device). For example, it is mainly used in related measurement of the power industry to stably trigger the waveform output from the transformer of a transformer substation.
Note:
To select stable channel waveform as the trigger source to stabilize the display.

6.2 Trigger Mode

The oscilloscope’s trigger mode includes Auto, Normal, and Single. Trigger mode aects the way in which the oscilloscope searches for the trigger.
After the oscilloscope starts running, the oscilloscope operates by rst lling the pre-trigger buer. It starts searching for a trigger after the pre-trigger buer is lled and continues to ow data through this buer while it searches for the trigger. While searching for the trigger, the oscilloscope overows the pre-trigger buer and the rst data put into the buer is rst pushed out (First Input First Out, FIFO).
When a trigger is found, the pre- trigger buer contains the events that occurred just before the trigger. Then, the oscilloscope lls the post- trigger buer and displays the acquisition memory.
To select the trigger mode press the key corresponding to the desired mode.
Auto
If the specied trigger conditions are not found, the triggers are forced and acquisitions are made so that signal activity is displayed on the oscilloscope.
Auto trigger mode is appropriate when:
• Checking DDC signals or signals with unknown levels or activity.
• When trigger conditions occur often enough that forced triggers are unnecessary.
Normal
Triggers and acquisitions only occur when the specied trigger conditions are found. Otherwise, the oscilloscope holds the original waveform and waits for the next trigger.
Normal trigger mode is appropriate when:
• Only specic events specied by the trigger settings are to be acquired.
• Triggering on an infrequent signal from a serial bus such as I2C, SPI, CAN, LIN, etc. or another signal that arrives in bursts.
Note:
Normal trigger mode stabilizes the display by preventing the oscilloscope from auto-triggering.
Trigger 40
Single
The oscilloscope waits for a trigger and displays the waveform when the trigger condition is met, the acquisition is stopped when the trigger conditions are met.
Single trigger mode is appropriate when:
• Capturing a single event or a periodic signal.
• Capturing a burst or other unusual signals.
Note:
The oscilloscope can be forced to trigger by pressing the Single button twice. The trigger status in the upper left corner of the screen will be displayed as "FStop".

6.3 Trigger Level

Trigger level and slope dene the trigger point.
Figure 6.3 Trigger Point
Turn the Trigger Level Knob to adjust the trigger level for the selected analog channel.
Pushing the Trigger Level Knob sets the trigger level to 0 when AC Coupling is selected.
The position of the trigger level for the analog channel is indicated by the trigger level icon (if the analog channel is on). The value of the analog channel trigger level is displayed in the upper- right corner of the display.
The example in gure 6.3 highlights the Positive edge slope at a trigger level of 2.00 mV.
Trigger 41

6.4 Trigger Coupling

To set the trigger coupling:
1. Press the Setup key to enter the Trigger menu.
2. Press the Coupling softkey to display the available options.
3. Turn the Universal Knob or press the Coupling softkey continually to navigate the available modes.
4. Push the Universal Knob to set the selected mode.
Note:
Trigger coupling has nothing to do with the channel coupling.
The oscilloscope provides 4 kinds of trigger coupling modes:
Figure 6.4 Trigger Coupling
DC
Allows both AC and DC components of the waveform into the trigger path.
AC
Blocks all the DC components and attenuate signals lower than 8 Hz. Use AC coupling to get a stable edge trigger when your waveform has a large DC oset.
LF Reject
Block the DC components and reject the low frequency components lower than 2 MHz. Low frequency reject removes any unwanted low frequency components from a trigger waveform, such as power line frequencies, etc. that can interfere with proper triggering. Use LF Reject coupling to get a stable edge trigger when your waveform has low frequency noise.
HF Reject
Reject the high frequency components higher than 1.2 MHz.
Trigger 42
6.5 Trigger Holdo
Trigger holdo can be used to add an additional, user-dened delay to the re-arming of the trigger circuit. This provides control over how rapidly, or how often, the oscilloscope can be triggered. The oscilloscope will not trigger until the holdo time expires.
Use the holdo to trigger on repetitive waveforms that have multiple edges (or other events) between waveform repetitions. You can also use holdo to trigger on the rst edge of a burst when you know the minimum time between bursts.
For example, to get a stable trigger on the repetitive pulse burst shown in gure 6.5 set the holdo time to be >200 ns but <400 ns.
Figure 6.5 Holdo
The correct holdo setting is typically slightly less than one repetition of the waveform. Set the holdo to this time to generate a unique trigger point for a repetitive waveform.
Note:
Only edge trigger and serial trigger have holdo option. The holdo is adjustable from 100 ns to 1.5 s.
To nd the repetition of the waveform:
1. Press the Stop key to stop data acquisition.
2. Turn the Horizontal Position Knob and the Horizontal Scale Knob to nd where the waveform repeats.
3. Measure the time using the cursors function.
4. Set the Holdo time.
To set the Holdo time.
1. Press the Setup key to enter the Trigger menu.
2. Press the Holdo Close softkey to enable Holdo.
3. Turn the Universal Knob to set the desired holdo time.
– Push the Universal Knob to open the numeric keypad.
Note:
Adjusting the time scale and horizontal position will not aect the holdo time.
Trigger 43

6.6 Noise Rejection

Noise Rejection adds additional hysteresis to the trigger circuitry. By increasing the trigger hysteresis band, the possibility
of triggering on noise is decreased, however the trigger sensitivity is also decreased. A larger signal is required to trigger the oscilloscope when Noise Rejection is enabled.
To enable Noise Rejection:
1. Press the Setup key to enter the Trigger menu.
2. Press the Noise Reject softkey to toggle between On and O.
Disabled Enabled
Figure 6.6 Noise Reject
If the signal being probed is noisy, set up the oscilloscope to reduce the noise in the trigger path and on the displayed waveform. First, stabilize the displayed waveform by removing the noise from the trigger path. Second, reduce the noise on the displayed waveform.
• Obtain a stable display.
• Remove the noise from the rtigger path setting Trigger coupling to LF Reject, HF Reject, or enabling Noise Reject.
• Set the Acquisition mode to Average to reduce noise.

6.7 Trigger Types

The 2194 provides the following trigger types :
Edge Trigger • Slope Trigger
Pulse Trigger • Video Trigger
Window Trigger • Interval Trigger
Dropout Trigger Runt Trigger
Pattern Trigger • Serial Trigger
Figure 6.7 Trigger Type
Trigger 44

6.7.1 Edge Trigger

Edge trigger distinguishes the trigger points by seeking the specied edge (rising, falling, alter) and trigger level.
Figure 6.8 Edge Trigger Point
1. Press the Setup key to enter the Trigger menu.
2. Press the Type softkey to display the available trigger types.
3. Turn the Universal Knob or press the Type softkey to navigate the available options.
4. Press the Slope softkey.
– Turn the Universal Knob to set the desired trigger edge (rising, falling or alter)
5. Turn the Trigger Level Knob to adjust the trigger level.
Figure 6.9 Edge Trigger
Holdo, coupling and noise reject can be set in edge trigger.
Trigger 45

6.7.2 Slope Trigger

The slope trigger looks for a rising or falling transition from one level to another level in greater than or less than a certain amount of time.
In the oscilloscope, positive slope time is dened as the time dierence between the two crossing points of trigger level A and B with the positive edge as shown in the gure below.
Figure 6.10 Slope Trigger
1. Press the Setup key to enter the Trigger Menu.
2. Press the Type softkey to view the available trigger types.
3. Turn the Universal Knob
4. Select Slope and push the Universal Knob to select the trigger source.
5. Press the Slope softkey and turn the Universal Knob to set select the desired trigger edge (rising or falling)
– Push down the knob to conrm.
6. Press the Lower Upper softkey to select the Lower or Upper trigger level.
– Turn the Trigger Level Knob to adjust the position.
– The lower trigger level cannot be higher than the upper trigger level.
Note:
In the trigger state message box, L1 means the upper trigger lever while L2 means the lower trigger level
7. Press the Limit Range softkey.
8. Turn the Universal Knob to select the desired slope condition, and push the knob to conrm.
<= (less than a time value): Trigger when the positive or negative slope time of the input signal is lower than
the specied time value.
>= (greater than a time value): Trigger when the positive or negative slope time of the input signal is greater
than the specied time value.
[- - . - -] (within a range of time value):Trigger when the positive or negative slope time of the input signal
is greater than the specied lower limit of time and lower than the specied upper limit of time value.
- -][- - (outside a range of time value): Trigger when the positive or negative slope time of the input signal is
greater than the specied upper limit of time and lower than the specied lower limit of time value.
Trigger 46
6.7.3
The Pulse Trigger type triggers on the positive or negative pulse with a specied width.
Figure 6.11 Pulse Trigger
1. Press the Setup key to enter the Trigger Menu.
2. Press the Type softkey to view the available trigger types.
3. Turn the Universal Knob
4. Select Pulse and push the Universal Knob to select the trigger source.
5. Turn the Trigger Level Knob to adjust the trigger level to the desired place.
6. Press the Polarity softkey to toggle between Positive and Negative pulse.
7. Press the Limit Range softkey and turn the Universal Knob to select the desired condition.
<= (less than a time value): Trigger when the positive or negative pulse time of the input signal is lower than
the specied time value. For example, for a positive pulse, if you set t (pulse real width) 100ns, the waveform will trigger.
Figure 6.12 Less than a Time Value
>= (greater than a time value): Trigger when the positive or negative pulse time of the input signal is greater
than the specied time value. For example, for a positive pulse, if you set t (pulse real width) 100ns, the waveform will trigger.
Figure 6.13 Greater than a Time Value
[- -.- -] (within a range of time value): Trigger when the positive or negative pulse time of the input signal is
greater than the specied lower limit of time and lower than the specied upper limit of time value.
Trigger 47
Figure 6.14 Within a Range of Time Value
- -][- - (outside a range of time value): Trigger when the positive or negative pulse time of the input signal is
greater than the specied upper limit of time and lower than the specied lower limit of time value.
Figure 6.15 Pulse Trigger Example
Coupling and noise reject can be set in pulse trigger, see the sections Trigger Coupling and Noise Rejection for details.
Trigger 48

6.7.4 Video Trigger

Video triggering can be used to capture the complicated waveforms of most standard analog video signals. The trigger circuitry detects the vertical and horizontal interval of the waveform and produces triggers based on the video trigger settings you have selected.
The oscilloscope supports standard video signal eld or line of NTSC (National Television Standards Committee), PAL (Phase Alternating Line) HDTV (High Denition Television) and custom video signal trigger. To setup Video Triggering:
1. Press the Setup key to enter the Trigger Menu.
2. Press the Type softkey to view the available trigger types.
3. Turn the Universal Knob
4. Select Pulse and push the Universal Knob to select the trigger source.
Note:
The trigger level is automatically set to sync pulse. Turning the Trigger Level Knob does not change the trigger level.
5. Press the Standard softkey to select the desired video standard. ( 6.1.)
Standard Type Sync Pulse
NTSC Interlaced BI-level
PAL Interlaced BI-level HDTV 720P/50 Progressive Tri-level HDTV 720P/60 Progressive Tri-level
HDTV 1080P/50 Progressive Tri-level HDTV 1080P/60 Progressive Tri-level
HDTV 1080iP/50 Progressive Tri-level
HDTV 1080i/50 Progressive Tri-level
Custom
Table 6.1 Video Standards
The parameters of Custom Video Trigger are shown in table 6.2.
Frame Rate 25 Hz, 30 Hz, 50 Hz, 60 Hz
Of Lines 300 to 2000
PAL 1, 2, 3, 4
HDTV 720P/50 1:1, 2:1, 4:1, 8:1
Trigger Position Line Field
(line value)/1 1 (line value)/2 2 (line value)/3 3 (line value)/4 4 (line value)/5 5 (line value)/6 6 (line value)/7 7 (line value)/8 8
Table 6.2 Custom Video Trigger Parameters
Trigger 49
Table 6.3 explains the relation between Of Lines, Of Fields, Interlace, Trigger Line and Trigger Field using an Of Lines value of 800.
Of Lines Of Fields Interlace Trigger Line Trigger Field
800 1 1:1 800 1 800 1, 2, 4 or 8 2:1 400 1, 12, 14, 18 800 1, 2, 4 or 8 4:1 200 1, 12, 14, 18 800 1, 2, 4 or 8 8:1 100 1, 12, 14, 18
Table 6.3 Parameters Relations
6. Press the Sync softkey to select Any or Select trigger mode.
Any: Trigger on any of the horizontal sync pulses.
Select: Trigger on the appointed line and eld you have set.
– Press the Line or Field softkey.
– Turn the Universal Knob to set the value.
Table 6.4 list the line numbers per eld for each video standard.
Standard Field 1 Filed 2
NTSC 1 to 262 1 to 263
PAL 1 to 312 1 to 313
HDTV 720P/50, HDTV 720P/60 1 to 750 HDTV 1080P/50, HDTV 1080P/60 1 to 1125 HDTV 1080iP/50, HDTV 1080i/60 1 to 562 1 to 563
Table 6.4 Line Numbers Per Field
Triggering on a Specic Line of Video
Video triggering requires greater than 1/2 division of sync amplitude with any analog channel as the trigger source.
The example below is set to trigger on eld 2, line 124 using the NTSC video standard.
1. Press the Setup key to enter the Trigger menu.
2. Press the Type softkey, then use the Universal Knob to select Video.
3. Press the Source softkey, then use the Universal Knob to select CH 1 as the trigger source.
4. Press the Standard softkey, then use the Universal Knob to select NTSC as the trigger source.
5. Press the Sync softkey and set the option to Select; press the Line softkey and then turn the universal to set 22.
6. Press the Line softkey, then use the Universal Knob to select 022 as the trigger source.
7. Press the Field softkey, then use the Universal Knob to select 1 as the trigger source.
Trigger 50
Use a Custom Video Trigger
Custom video triggering supports frame rate of 25Hz, 30Hz, 50Hz and 60Hz, and the line range is available from 300 to
2000. The steps below show how to set custom trigger.
1. Press the Setup key to enter the Trigger menu.
2. Press the Type softkey, then use the Universal Knob to select Video.
3. Press the Source softkey, then use the Universal Knob to select CH 1 as the trigger source.
4. Press the Standard softkey, then use the Universal Knob to select Custom as the trigger source.
5. Press the Setting softkey to enter the custom setting function menu.
6. Press the Interlace softkey, then use the Universal Knob to set the desired value.
7. Press the Of Field softkey, then use the Universal Knob to set the desired value.
8. Press the Sync softkey to enter the TRIG ON menu.
9. Press the Type softkey to select Any or Select.
• If Select was chosen:
– Press the Line softkey, then use the Universal Knob to set the desired value.
– Press the Field softkey, then use the Universal Knob to set the desired value.
Trigger 51

6.7.5 Window Trigger

Windows trigger provides a high trigger level and a low trigger level. The instrument triggers when the input signal passes through the high trigger level or the low trigger level.
There are two kinds of window types: Absolute and Relative. They have dierent trigger level adjustment methods.
Under Absolute window type, the lower and the upper trigger levels can be adjusted respectively via the Level knob.
Under Relative window type the Center value is adjusted to set the window center, and the Delta value is adjusted to set the window range. The lower and the upper trigger levels always move together.
Figure 6.16 Window Trigger
• If the lower and the upper trigger levels are both within the waveform amplitude range, the oscilloscope will trigger on both rising and falling edge.
• If the upper trigger level is within the waveform amplitude range while the lower trigger level is out of the waveform amplitude range, the oscilloscope will trigger on rising edge only.
• If the lower trigger level is within the waveform amplitude range while the upper trigger level is out of the waveform amplitude range, the oscilloscope will trigger on falling edge only.
Trigger 52
Set Window Trigger Via Absolute Window Type
1. Press the Setup key to enter the Trigger menu.
2. Press the Type softkey, then use the Universal Knob to select Window.
3. Press the Source softkey, then use the Universal Knob to select CH1 or CH2 as the trigger source.
4. Press the Window Type softkey to select Absolute.
5. Press the Lower Upper softkey to select Lower or Upper trigger level; then turn the Trigger Level Knob to adjust the position.
Figure 6.17 Absolute Window Trigger
Note:
The Lower trigger level cannot be higher than the upper trigger level. In the trigger state message box, L1 means the upper trigger level while L2 means the lower trigger level.
Trigger 53
Set Window Trigger Via Relative Window Type
1. Press the Setup key to enter the Trigger menu.
2. Press the Type softkey, then use the Universal Knob to select Window.
3. Press the Source softkey, then use the Universal Knob to select CH1 or CH2 as the trigger source.
4. Press the Window Type softkey to select Relative.
5. Press the Center Delta softkey to select Center or Delta trigger level mode, then use the Universal Knob to adjust the position.
Note:
In the trigger state message box, C means Center, the center value of the lower and upper trigger levels; D means Delta, the dierence between the lower (or upper) trigger level and the trigger level center.
Figure 6.18 Relative Window Trigger
Note:
Coupling and noise reject can be set in Window trigger, see the sections Trigger Coupling and Noise Rejection for details.
Trigger 54

6.7.6 Interval Trigger

Trigger when the times dierence between the neighboring rising or falling edges meets the time limit (< =, > =, [ - - . - - ], - - ][ - - ).
Figure 6.19 Interval Trigger
To set an interval trigger:
1. Press the Setup key to enter the Trigger menu.
2. Press the Type softkey, then use the Universal Knob to select Interval.
3. Press the Source softkey, then use the Universal Knob to select CH1 or CH2 as the trigger source
4. Press the Slope softkey to select rising or falling edge.
5. Press the Limit Range softkey, then use the Universal Knob to set the desired condition.
< = (less than a time value): Triggers when the positive or negative pulse time of the input signal is lower
than the specied time value.
> = (greater than a time value): Triggers when the positive or negative pulse time of the input signal is
greater than the specied time value.
[ - - . - - ] (within a range of time value): Triggers when the positive or negative pulse time of the input
signal is greater than the specied lower limit of time and lower than the specied upper limit of time value.
- - ] [ - - (outside a range of time value): Triggers when the positive or negative pulse time of the input signal
is greater than the specied upper limit of time and lower than the specied lower limit of time value.
6. Press the Time Setting softkey (< = , > = , [ - - . - - ] , - - ] [ - -), turn the Universal Knob to select the desired value.
Trigger 55
Figure 6.20 Interval Trigger Example
Note:
Coupling and noise reject can be set in interval trigger, see the sections Trigger Coupling and Noise Rejection for details.
Trigger 56

6.7.7 Dropout Trigger

Dropout trigger includes two types: edge and state.
Edge
Triggers when an edge followed by a specied time with no edges is detected. This is useful for triggering on the end of a pulse train.
Figure 6.21 Dropout Trigger Edge
To set an edge dropout trigger:
1. Press the Setup key to enter the Trigger menu.
2. Press the Type softkey, then use the Universal Knob to select Dropout.
3. Press the Source softkey, then use the Universal Knob to select CH1 or CH2 as the trigger source.
4. Press the Slope softkey to select rising or falling edge.
5. Press the OverTime Type softkey to select Edge.
6. Press the Time softkey, then use the Universal Knob to set the desired value.
Note:
Coupling and noise reject can be set in interval trigger, see the sections Trigger Coupling and Noise Rejection for details.
Trigger 57
State
Triggers when the signal enters or leaves a voltage level and stays there for a specied time. This is useful for detecting when a signal gets stuck at a particular level.
Figure 6.22 Dropout Trigger State
To set a state dropout trigger:
1. Press the Setup key to enter the Trigger menu.
2. Press the Type softkey, then use the Universal Knob to select Dropout.
3. Press the Source softkey, then use the Universal Knob to select CH1 or CH2 as the trigger source
4. Press the Slope softkey to select rising or falling edge.
5. Press the OverTime Type softkey to select State.
6. Press the Time softkey, then use the Universal Knob to set the desired value.
Note:
Coupling and noise reject can be set in interval trigger, see the sections Trigger Coupling and Noise Rejection for details.
Trigger 58

6.7.8 Runt Trigger

The runt trigger detects a pulse that crosses the rst threshold but not the second. It can occur when a logic driver has insucient slew rate to reach a valid logic level in the time available.
Figure 6.23 Runt Trigger
• A positive runt pulse crosses through a lower threshold but not an upper threshold.
• A negative runt pulse crosses through an upper threshold but not a lower threshold.
To setup a runt trigger:
1. Press the Setup key to enter the Trigger menu.
2. Press the Type softkey, then use the Universal Knob to select Runt.
3. Press the Source softkey, then use the Universal Knob to select CH1 or CH2 as the trigger source.
4. Press the Polarity softkey to select Positive or Negative pulse to trigger.
5. Press the Slope softkey to select rising or falling edge.
6. Press the OverTime Type softkey to select State.
7. Press the Limit Range softkey, then use the Universal Knob to select the desired condition (< =, > =, [ - - . - - ] or - - ] [ - - ).
8. Press the Time Setting softkey, then use the Universal Knob to select the desired value.
9. Press the Next Page softkey to enter the second page of the Trigger menu.
10. Press the Lower Upper softkey to select Lower or Upper trigger level.
– Use the Universal Knob to select the desired value.
Trigger 59
Figure 6.24 Runt Trigger Example
Note:
Coupling and noise reject can be set in interval trigger, see the sections Trigger Coupling and Noise Rejection for details.
Trigger 60

6.7.9 Pattern Trigger

The Pattern trigger identies a trigger condition by looking for a specied pattern. The pattern trigger can be expanded to incorporate delays.
Pattern durations are evaluated using a timer. The timer starts on the last edge that makes the pattern “true”. Potential triggers occur on the rst edge that makes the pattern false, provided that the time qualier criterion has been met.
The oscilloscope provides 4 patterns: logical AND, OR, NAND and NOR combination of the channels. Each channel can set to low, high or invalid.
Figure 6.25 Pattern Trigger
To setup a pattern trigger:
1. Press the Setup key to enter the Trigger menu.
2. Press the Type softkey, then use the Universal Knob to select Pattern.
3. Press the Source softkey, then use the Universal Knob to select CH1 or CH2 as the trigger source.
4. Press the Don’t care softkey to assign Don’t care, High or Low to the channel.
Low: Sets the pattern to low on the selected channel. A low is a voltage level that is less than the channel’s
trigger level or threshold level.
• High sets the pattern to high on the selected channel. A high is a voltage level that is greater than the channel’s
trigger level or threshold level.
• Don’t care sets the pattern to don’t care on the selected channel. Any channel set to don’t care is ignored and is
not used as part of the pattern.
– If all channels in the pattern are set to Don’t care, the oscilloscope will not trigger.
Trigger 61
Note:
Adjust the trigger level for the selected analog channel by turning the Trigger Level knob while high or low are selected. Invalid doesn’t need to set trigger level.
5. Press the Next Page softkey to enter the second page of the pattern trigger menu.
6. Press the Logic softkey and then turn the Universal Knob to select the desired logic combination AND, OR, NAND or NOR.
7. Press the Time softkey, then use the Universal Knob to select the desired time value.
8. Press the Holdo Close softkey to enable/disable the Holdo function
– Use the Universal Knob to set the desired value.
Note:
Figure 6.26 Pattern Trigger Example
Holdo can be set in pattern trigger, see the sections Trigger Holdo for details.

Serial Trigger and Decode

The oscilloscope provides I2C, SPI, UART, CAN and LIN serial trigger and decode. This chapter introduces the method of triggering and decoding these serial signals in details.

7.1 I2C Trigger and Serial Decode

7.1.1 Setup for I2C Signals

To set the I2C (Inter-IC bus) signal rst, connect the serial data signal (SDA) and serial clock signal (SCK) to oscilloscope, then specify the threshold voltage of each input signal.
1. Touch Decode key to enter the Decode menu.
2. Touch Decode and select the desired slot (Decode1 or Decode2).
3. Touch Protocol and then select I2C by turning the Universal Knob.
4. Touch Signal to enter the Signal menu.
5. Set the SCL (I2C’s clock signal):
• Touch SCL to select the channel that is connected to the I2C clock signal.
• Touch rst Threshold softkey, then use the Universal Knob to set the I2C clock signal’s threshold voltage level .
– The threshold voltage level is for decoding, and it will be regarded as the trigger voltage level when the trigger
type is set to serial.
6. Set SDA (I2C’s data signal):
• Touch SDA to select the channel that is connected to the I2C data signal.
• Touch second Threshold softkey, then use the Universal Knob to set the I2C clock signal’s threshold voltage level
.
– The threshold voltage level is for decoding, and it will be regarded as the trigger voltage level when the trigger
type is set to serial.
Note:
SDA should keep stable during the whole high clock cycle, otherwise it will be interpreted as a start or stop condition (data transitioning while the clock is high)
7. Touch to return previous menu.
Serial Trigger and Decode 63

7.1.2 I2C Trigger

The I2C trigger has nine trigger conditions: (Start, Stop, Restart, No Ack, EEPROM, 7 Addr&Data, 10 Addr&Data and Data Length)
Start Condition: The oscilloscope will be triggered when the SDA signal transitions from high to low while the SCL clock is high. If it is chosen as the condition of trigger (including frame triggers), a restart will be treated as a “Start condition”.
Stop Condition: The oscilloscope will be triggered when SDA transitions from low to high while the SCL is high.
Figure 7.1 Start and Stop Conditions
Restart: The oscilloscope will be triggered when another “Start condition” occurs before a “Stop condition”.
No Ack: The oscilloscope will be triggered when SDA data is high during any SCL’s ACK bit.
EEPROM: The trigger searches for EEPROM control byte (the value is 1010xxx) on the SDA bus where there is a Read bit and an ACK bit behind EEPROM.
– Use the Limit Range to set the qualier.
– Use the Data1 to set the data’s value.
Note:
If EEPROM’s data is greater than Data1, the oscilloscope will be triggered at the edge of ACK bit behind Data byte. It’s unnecessary that the Data byte musts follow the EEPROM.
Figure 7.2 Trigger Point
7 address & Data: The oscilloscope will be triggered when the following conditions are satised:
– The address’s length must be 7 bits and the address’s value must be the same as the set value.
– Data1’s or Data2’s value is set to the same value of the signal’s data.
– Both Data1’s and Data2’s value are set and the signal has two consecutive data values, the rst is value is Data1
and the second data value is Data2.
Serial Trigger and Decode 64
Note:
If the data’s value is 0xXX, any data value will be matched.
Figure 7.3 7 Address & Data
10 address & Data: The oscilloscope will be triggered when the following conditions are satised:
– The address’s length must be 10 bits and the address’s value must be the same as the set value.
– Data1’s or Data2’s value is set to the same value of the signal’s data.
– Both Data1’s and Data2’s value are set and the signal has two consecutive data values, the rst is value is Data1
and the second data value is Data2.
Note:
If the data’s value is 0xXX, any data value will be matched.
Figure 7.4 10 Address & Data
Data Length: When SDA data’s length is equal to the value of Byte Length and address’s length is the same as set value, the oscilloscope will be triggered. Byte length is in the range of 1 to 12 bits.
Serial Trigger and Decode 65
Operation Step
1. Touch Setup key to enter the Trigger menu.
2. Touch Type and select Serial.
3. Touch Protocol and select I2C.
4. Touch Trigger Setting softkey.
5. Touch Condition and turn the Universal Knob to select the trigger condition:
Figure 7.5 Condition Options
• If EEPROM is selected :
– Touch Limit Range to set the qualier (=, < or >).
– Touch Data1 and set its value by turning the Universal Knob/
• If 7 Addr & Data or 10 Addr & Data is selected:
– Touch Addr and turn the Universal Knob to select the 7- bit or 10- bit device address.
– Touch Data1 or Data2 and set the value.
– Touch R/W bit and select write-frame or read-frame to trigger the oscilloscope.
Note:
If the device’s address is 7-bit, the value of address is in range of 0x00 to 0x7F. If device’s address is 10-bit, the value of address is in range of 0x00 to 0x3FF.
• If Data Length is selected:
– Touch Address to set the SDA address length 7bit or 10 bit.
– Touch Byte Length and set the byte length by Universal Knob. The range of the Byte Length is 1 to 12.
Serial Trigger and Decode 66

7.1.3 I2C Serial Decode

Once the setup for I2C signal and trigger has been completed, the decoding operation must be setup.
1. Touch Decode followed by the Decode softkey.
– Select one of the options from the Decode1 and Decode2.
2. Touch Display and select On to display the result of decoding.
3. Touch Next Page to view page 2/2 of the Decode menu.
4. Touch Format to change the character encoding format of the decoding’s result.
5. Touch List to enter the List menu.
6. Touch Display and choose the Decode option selected in step 2.
7. Touch Lines and set the number of lines by Universal Knob.
– The range of the lines is 1 to 7.
8. Press the
9. Touch Scroll and turn the Universal Knob to view all frames.
to return to the previous menu.
Interpreting I2C Decode
The frames of decoding result:
• The address value is displayed at the beginning of a frame. The write address is displayed in green, and the read address is displayed in yellow.
• W/R bit is represented by (W) and (R), following the address value.
• The data value is displayed in white.
• "˜A" after a data or address bits indicates no acknowledgement. For example, DB˜A.
• A red dot indicates there is not enough space on the display to show the complete content of a frame.
Figure 7.6 I2C Decode Bus Display
The lists of decoding result:
Time: The horizontal displacement between current frame and trigger position.
Address: The address of a frame.
R/W: The type of a frame (write or read).
Data: The value of data.
Figure 7.7 Decode List Display
Serial Trigger and Decode 67

7.2 SPI Trigger and Serial Decode

7.2.1 Setup for SPI Signals

Setting the SPI (Serial Peripheral Interface) signal includes two steps: Connecting the CLK, MISO, MISO and CS signal to oscilloscope, specifying the parameters of each input signal.
1. Touch Decode key to enter the Decode menu.
2. Touch Decode and select the desired slot (Decode1 or Decode2).
3. Touch Protocol softkey, then use the Universal Knob to select SPI.
4. Touch Signal to enter the Signal menu.
Figure 7.8 Signal Menu
5. Set CLK (clock signal):
• Touch CLK to enter CLK menu.
• Touch CLK to select the channel that is connected to the SPI clock signal.
• Touch Threshold to set the SPI clock signal’s threshold voltage level by Universal Knob.
– The threshold voltage level is for decoding, and it will be regard as the trigger voltage level when set the
trigger type to serial.
• Touch Edge Select to set the oscilloscope will samples at clock signal’s rising edge or falling edge.
• Touch
6. Set MISO:
• Touch MISO to enter the MISO menu.
• Touch MISO to select the channel that is connected to the SPI MISO signal.
• Touch Threshold softkey, then use the Universal Knob to set the SPI MISO signal’s threshold voltage level.
to return previous menu.
Figure 7.9 CLK Menu
– The threshold voltage level is for decoding, and it will be regard as the trigger voltage level when set the
trigger type to serial.
• Touch
to return to the previous menu.
Figure 7.10 MISO Menu
Serial Trigger and Decode 68
7. Set MOSI:
• Press the MOSI to enter the MOSI menu.
• Touch MOSI to select the channel that is connected to the SPI MOSI signal.
• Press the Threshold softkey, then use the Universal Knob to set the SPI MOSI signal’s threshold voltage level.
– The threshold voltage level is for decoding, and it will be regarded as the trigger voltage level when the trigger
type is set serial.
• Touch
8. Set CS:
• Touch CS to enter the MOSI menu.
• Touch CS Type to select the chip select type.
• Modify the CD value.
• Touch
Function Menu Settings Description
CS Type
to return to the previous menu.
Figure 7.11 MOSI Menu
to return to the previous menu.
˜CS Low voltage level of CS signal is available
CS High voltage level of CS signal is available
If the time between two edges of the clock signal is less than (or equal to) the value of timeout, the signal between the two edges is treated as a frame. The range of clock timeout is 100 ns - 5 ms.
CLK Timeout
This setting is suitable for case where the CS signal is not connected, or the number of oscilloscope channels is insucient (such as 2-channel oscilloscopes).
Table 7.1 CS Type Paramters
Example
Connect the data, CLK and ˜CS signals of a SPI bus respectively to C1, C2, and C3. Data width = 8-bit, Bit order = MSB, CS polarity = CS, and 12 data bytes are transmitted in one frame.
In the SPI trigger signal menu, set the source and threshold of CLK, MISO and CS signals, then copy the trigger settings to decoding. Adjust the time base, so that the falling edge of the CS signal is shown in the display:
Serial Trigger and Decode 69
Figure 7.12 Example
When the CS type is set to Clock Timeout, the clock idle time between frames is T3, the clock period is T1, then set the timeout to a value between T1 and T3
Figure 7.13 Example 2
Serial Trigger and Decode 70
If the data width is set to be greater than 8 bits (such as 16 bits), the clock idle time between 8-bit data packets T2, and then set the timeout time to a value between T1/2+T2 and T3.
Figure 7.14 Example 3
Serial Trigger and Decode 71

7.2.2 SPI Trigger

This section provides a brief introduction for the operation of the SPI trigger.
1. Touch Setup button to enter the TRIGGER function menu.
2. Touch Type and select Serial.
3. Touch Protocol and select SPI.
4. Touch Trigger Setting softkey.
Figure 7.15 SPI Trigger Setup Menu
5. Touch Trigger Type to select the trigger condition.
Function Menu Settings Explanation
Trigger Type MISO Master-In, Slave-Out
MOSI Master-Out, Slave-In
Table 7.2 SPI Trigger Type
6. Touch Data Length softkey, and turn the Universal Knob to set the length of a data. The range of data length is 4 to 96 bits.
7. Set the value of the trigger data.
• Set the value of a bit:
a. Touch Bit Roll to select a bit in data.
b. Touch Bit Value to set the value of the selected bit.
• Set the value of all bits:
a. Touch All Same to set the value of all bits.
Function Menu Settings Explanation
Bit Value 0 High voltage level
1 Low voltage level
X Don’t care the voltage level
Table 7.3 SPI Bit Value
8. Touch Next Page softkey.
9. Touch Bit Order to set the bit order (MSB or LSB).
Serial Trigger and Decode 72

7.2.3 SPI Serial Decode

Once the setup of SPI signal and trigger is complete, the SPI signals can be decoded. Operation steps as follows:
1. Touch Decode –>Decode. Select one of the options from the Decode1 and Decode2.
2. Touch Display and select On to display the result of decoding.
3. Touch List to enter the LIST function menu.
4. Touch Display and choose the same options as the rst step.
5. Touch Lines and set the number of lines by Universal Knob. The range of the lines is 1 to 7.
6. Touch Format to change the character encoding format of the decoder’s result.
7. Touch Scroll and turn the Universal Knob to view all frames.
Interpreting SPI Decode
The frames of decoding result:
• The data values are displayed in frames and are shown in white. Support 4 96 bit data display.
MISO
MOSI — the decoding result of “Master-Out, Slave-In” line.
Indicates there is not enough space on the display to show the complete content of a frame, and some content is
The lists of decoding result:
Time
MISO — the decoding result of “Master-In, Slave-Out” line.
MOSI — the decoding result of “Master-Out, Slave-In” line.
— the decoding result of “Master-In, Slave-Out” line.
hidden.
Figure 7.16 SPI Decode Bus Display
— the horizontal displacement between current frame and trigger position.
Figure 7.17 SPI Decode List Display
Serial Trigger and Decode 73

7.3 UART Trigger and Serial Decode

7.3.1 Setup for UART Signals

1. Touch Decode key to enter the DECODE function menu.
2. Touch Decode and select the desired slot (Decode1 or Decode2).
3. Touch Protocol and then select UART by turning Universal Knob.
4. Touch Signal to enter the SIGNAL menu as below shows.
5. Set RX:
a. Touch RX to select the channel that is connected to the RX signal.
b. Touch rst Threshold key to set the RX signal’s threshold voltage level by Universal Knob. The threshold
voltage level is for decoding, and it will be regard as the trigger voltage level when set the trigger type to serial.
6. Set TX:
a. Touch TX to select the channel that is connected to the TX signal.
b. Touch second Threshold key to set the TX signal’s threshold voltage level by Universal Knob. The threshold
voltage level is for decoding, and it will be regard as the trigger voltage level when set the trigger type to serial.
7. Touch
8. Touch Congure to enter BUS CONFIG menu.
9. Touch Baud to set baud rate.
• The baud rate can be set as predened value.
• If the desired baud rate is not listed, press the Baud and select custom option, press the Custom and turn the
Universal Knob to set the desired baud rate.
10. Touch Data Length and set byte bits (5-8) by Universal Knob.
11. Touch Parity Check to set the type of parity check (Even, Odd, Mark, Space or None).
to return previous menu.
Figure 7.18 Bus Cong Menu
12. Touch Stop Bit to set the length of stop bit (1, 1.5 or 2 bits).
13. Touch Next Page softkey.
14. Touch Bit Order to select the bit order (LSB or MSB).
15. Touch Idle Level to set the idle level (LOW or HIGH).
Serial Trigger and Decode 74

7.3.2 UART Trigger

This section includes an introduction and description for the operation of the UART trigger.
1. Touch Setup key to enter the TRIGGER function menu.
2. Touch Type and select Serial.
3. Touch Protocol and select UART.
4. Touch Trigger Setting to enter UART TRIG SET menu.
Figure 7.19 UART Trig Set Menu
5. Touch Source Type to select the source of trigger (RX or TX).
6. Touch Condition and set up the desired trigger condition:
Start
Stop —the oscilloscope will be triggered at the position of stop bits.
Data — the oscilloscope will be triggered when found a byte which is equal to (greater or less than) the specied
ERROR
— the oscilloscope will be triggered at the position of start bit.
data.
a. Touch Compare Type and choose an equality qualier (>, < or =).
b. Touch Value to set data’s value. Data’s value is in range of 0x00 to 0x.
— if the parity check has been set, and the bit of parity check is error, the oscilloscope will be triggered.
Figure 7.20 UART Trigger
Serial Trigger and Decode 75

7.3.3 UART Serial Decode

Upon completing the setup of UART signal and trigger, the UART signals can be decoded. Operation steps as follows:
1. Press Decode –> Decode. Select one of the options from the Decode1 and Decode2.
2. Touch Display and select On to display the result of decoding.
3. Touch List to enter the LIST function menu.
4. Touch Display key and choose the same options as the rst step.
5. Touch Lines and set the number of lines by Universal Knob. The range of the lines is 1 to 7.
6. Touch Format softkey to change the character encoding format of the decoding’s result.
7. Turn the Universal Knob to view all frames.
Figure 7.21 UART Decode
Serial Trigger and Decode 76
Interpreting UART Decode
The frames of decoding result:
RX — the decoding result of the data received.
TX — the decoding result of the data transmitted.
Indicates there is not enough space on the display to show the complete content of a frame, and some content is hidden.
Figure 7.22 UART Decode Bus Display
The lists of decoding result:
Time — the horizontal displacement between current frame and trigger position.
RX — the receiving channel.
TX — the transmitting channel.
RX Err — Parity error or unknown error in the data received.
TX Err
— Parity error or unknown error in the data transmitted.
Figure 7.23 UART Decode List Display
Serial Trigger and Decode 77

7.4 CAN Trigger and Serial Decode

Placed in order of Setup for CAN Signals, CAN Trigger, CAN Serial Decode to trigger and decode the signals.

7.4.1 Setup for CAN Signals

1. Touch Decode key to enter the DECODE function menu.
2. Touch Decode and select the desired slot (Decode1 or Decode2).
3. Touch Protocol and then select CAN by turning Universal Knob.
4. Touch Signal to enter the SIGNAL menu as shown in gure ??.
Figure 7.24 CAN Signal Menu
a. Touch Source to select the channel that is connected to the CAN signal.
b. Touch Threshold key to set the CAN signal’s threshold voltage level by Universal Knob. The threshold voltage
level is for decoding, and it will be regard as the trigger voltage level when set the trigger type to serial.
5. Touch Congure to enter the BUS CONFIG menu.
6. Touch Baud to set baud rate by rotating the Universal Knob.
• The baud rate can be set as predened value (from 5kb/s to 1Mb/s) or custom value (from 5kb/s to 1Mb/s).
• If the desired baud rate is not listed, press Baud and select the custom option, press the Custom and turn the
Universal Knob to set the desired baud rate.

7.4.2 CAN Trigger

This section provides a brief introduction and description for the operation of the CAN trigger.
Trigger Conditions
Start — the oscilloscope will be triggered at the start bit of a frame.
Remote — the oscilloscope will be triggered by a remote frame with specied ID.
— the oscilloscope will be triggered by a remote or data frame that have specied ID.
ID
ID+DATA — the oscilloscope will be triggered by data frame that have specied ID and data.
Error — the oscilloscope will be triggered by an error frame.
Serial Trigger and Decode 78
Operation Steps
1. Touch Setup to enter the TRIGGER function menu.
2. Touch Type and select Serial.
3. Touch Protocol and select CAN.
4. Touch Trigger Setting to enter the CAN TRIG SET menu.
5. Touch Condition and select the trigger condition by turning the Universal Knob:
• If you select the REMOTE and ID condition:
a. Touch ID Bits to set the length of ID (11bits or 29 bits).
b. Touch Curr ID Byte and turn the Universal Knob to select the byte that you want to set.
c. Touch ID and set the ID’s value by turning the Universal Knob. (Tips: In order to make is convenient for
the operator to set the parameters, ID b is split into several bytes. For example, if the ID’s length is 11 bits, it will be split into two bytes, a byte includes 8 bits. If “1st byte” is selected, only the 8 least signicant bits can be changed.)
• If you select the ID+DATA condition:
a. Touch ID bits to select the ID’s length (11 or 29 bits).
b. Touch Curr ID Byte and turn the Universal Knob to select the byte that you want to modify.
c. Touch ID and set the ID’s value by turning the Universal Knob.
d. Touch Data and set the value of the rst byte by turning the Universal Knob.
Figure 7.25 CAN Trigger
Serial Trigger and Decode 79

7.4.1 CAN Serial Decode

Upon completing the setup of can signal and trigger, the CAN signals can be decoded.
1. Press Decode –> Decode. Select one of the options from the Decode1 and Decode2.
2. Touch Display and select On to display the result of decoding.
3. Touch List to enter the LIST function menu.
4. Touch Display and choose the same options as the rst step.
5. Touch Lines and set the number of lines by Universal Knob. The range of the lines is 1 to 7.
6. Touch Format softkey to change the character encoding format of the decoder’s result.
7. Touch Scroll and turn the Universal Knob to view all frames.
Interpreting CAN Decode
The frame of decoding result:
• Arbitration eld is displayed in frame
• Control eld is displayed in frame
• Data eld is displayed in frame
• CRC eld is displayed in frame
Indicates there is not enough space on the display to show complete content of a frame and some content is hidden.
Figure 7.26 CAN Decode Bus Display
The list of decoding result:
Time — the horizontal displacement between current frame and trigger position.
Type
ID — the id of frames, the oscilloscope can automatically detect the length of frame’s id (11 bits or 27 bits).
Length — the length of data eld.
Data — the value of data eld.
— the type of frames, “D” represents data frame, “R” represents remote frame.
CRC
ACK — Acknowledgment bit.
— the value of CRC (Cyclic Redundancy Check) eld.
Figure 7.27 CAN Decode List Display
Serial Trigger and Decode 80

7.5 LIN Trigger and Serial Decode

7.5.1 Setup for LIN Signals

There are two steps of setting the LIN signal, connecting the signal to oscilloscope, specifying the parameters of each input signal.
1. Touch Decode button to enter the DECODE function menu.
2. Touch Decode and select the desired slot (Decode1 or Decode2).
3. Touch Protocol and then select LIN by turning Universal Knob.
4. Touch Signal to enter the SIGNAL menu as shown in gure 7.22.
Figure 7.28 CAN Signal Menu
5. Touch Source to select the channel that is connected to the LIN signal.
6. Touch Threshold and set the LIN signal’s threshold voltage level by Universal Knob. The threshold voltage level is for decoding, and it will be regard as the trigger voltage level when set the trigger type to serial.
7. Touch
8. Touch Congure to enter the BUS CONFIG menu.
9. Touch Baud to set baud rate.
• The baud rate can be set as predened value.
• If the desired baud rate is not listed, select custom option, press the Custom and turn the Universal Knob to
set the desired baud rate.
to return previous menu.
Serial Trigger and Decode 81

7.5.2 LIN Trigger

This section provides a brief introduction and description for the operation of the LIN trigger.
Trigger Condition
Break — the oscilloscope will be triggered at the position of break eld’s break delimiter.
ID (Frame ID) — the oscilloscope will be triggered at the position of identier eld’s stop bit, if the value of a frame’s ID is equal to specied value. (Note: If the data’s value is 0xXX, any data value will be matched)
ID + Data (Frame ID and Data) — the oscilloscope triggers when a frame with an ID and data equal to the selected values is detected. Use the Universal Knob to select the value for the ID, Data1 and Data2.
a. The ID’s value is the same as set value.
b. If you have set either Data1’s or Data2’s value, and the signal has a data is the same as that value. If you have
set both Data1’s and Data2’s value, the signal should has two consecutive data, the rst data’s value is Data1, second data value is Data2.
Note:
If the data’s value is 0xXX, any data value will be matched
Data Error —the oscilloscope will be triggered when errors (such as ID check error, checksum error, sync byte eld error) are detected.
Operation Steps
• Touch Setup key to enter the TRIGGER function menu.
• Touch Type and select Serial.
• Touch Protocol and select I2C.
• Touch Trigger Setting to enter LIN TRIG SET menu.
• Touch Condition and select the trigger condition by Universal Knob:
• If you select ID condition:
– Touch ID and set its value by turning the Universal Knob.
• If you select ID+DATA condition:
– Touch ID and set its value by turning the Universal Knob.
– Touch DATA1 and set its value by turning the Universal Knob.
– Touch DATA2 and set its value by turning the Universal Knob.
Serial Trigger and Decode 82
Figure 7.29 LIN Trigger
Serial Trigger and Decode 83

7.5.1 Interpreting LIN Decode

The frame of decoding result:
• Protected Identier Field is displayed in frame
• Data Length is displayed in frame
• Data Field is displayed in frame.
• Checksum Field is displayed in frame.
Indicates there is not enough space on the display to show complete content of a frame and some content is hidden.
Figure 7.30 LIN Decode Bus Display
The list of decoding result:
Time — the horizontal displacement between current frame and trigger position.
— the value of frame’s Protected Identier Field.
ID
Data Length — the length of Data Field.
ID Parity — the two check bits of Protected Identier Field.
Data — the value of Data Field.
Checksum — the value of Checksum Field.
Figure 7.31 LIN Decode List Display

Reference Waveform

The oscilloscope can save analog channel or math waveforms to the reference waveform locations in the oscilloscope. Then, a reference waveform can be displayed and compared against other waveforms. All reference waveforms can be displayed at a time.

8.1 Save REF Waveform to Internal Memory

Do the following steps to save the REF waveform to internal memory:
1. Press the REF button on the front to enter the REF WAVE function menu. Note that when the time horizontal format is in X-Y mode, REF function cannot be enabled.
2. Press the Source softkey; then, turn the Universal Knob to select the source of reference channel. The source includes analog channel and math waveforms.
3. Press the Location softkey; then, turn the Universal Knob to select the position to save the REF waveform. The source includes analog channel and math waveforms.
4. Press the Save softkey to save the channel or math waveform to the appointed location. The vertical scale information and the vertical oset of the waveform will be saved at the same time. It will pop out the message “Store Data
Success” when the waveform has been saved successfully.
Note:
The REF waveforms are non-volatile. The REF waveform can still be saved after restarts or default operation.

8.2 Display REF Waveform

To the following steps to display REF waveform:
1. Press the REF button on the front to enter the REF WAVE function menu.
2. Press the Location softkey; then, turn the Universal Knob to select the REF waveform that you want to display.
3. Press the Display softkey to select On to display the REF waveform on the screen. Only saved location can be displayed. The oscilloscope can display all four reference waveforms at a time.
Reference Waveform 85

8.3 Adjust REF Waveform

1. Refer to the Display REF Waveform section above to display the desired reference waveform.
2. Press the Scale and Position softkey and turn the Universal Knob to adjust the vertical scale and position of the reference waveform. The vertical scale and position information display at the middle of the screen. The initial values display at the middle of the screen is the setup that when the reference waveform been saved.
The initial values displayed in the middle of the screen is the setup of the reference saved waveform.
Figure 8.1 Reference Waveform

8.4 Clear Ref Waveform

The oscilloscope does not have the “Clear” option under the REF WAVE function menu.
To clear the appointed reference waveform, you can save a new reference waveform to the same location to cover it.
Or follow the Save/Recall –> Recall –> type and select Security Erase to clear the stored waveform.

Math

The oscilloscope supports many math operations between analog channels including addition (+), subtraction (-), multiplication (*), division (/), FFT, dierential (d/dt), integral (∫𝑑𝑡), square root (√). The resulting math waveform is displayed in white and labeled with “M”. You can use cursors to measure it.
Note:
If the analog channel or the math function is cut o (waveforms do not display on the screen completely), the resulting math will also be cut o.

9.1 Units for Math Waveforms

Use the channel function menu to set the unit of each channel to “V” or “A”.
The oscilloscope math operation includes units as below:
Math Operation Unit
Addition (+)or subtraction (-) V, A
multiplication (*) 𝑉2, 𝐴2𝑜𝑟𝑊
division (/) None, 𝑉 ∗𝐴−1𝑜𝑟 𝑉−1∗𝐴
FFT dBVrms, Vrms, dBm, dBArms, Arms
dierential (d/dt) 𝑉 ∗ 𝑆−1𝑜𝑟 𝐴 ∗ 𝑆
integral (dt) VS or AS (V * second or A * second)
square root ( ) 𝑉
1/2
𝑜𝑟 𝐴
1/2
−1
Table 9.1 Math Operations Units
Math 87

9.2 Math Operators

The oscilloscope supports math count operation (Addition, subtraction, multiplication, division), FFT (Fourier transform) operation, and math function operation (dierential, integral, square root).

9.2.1 Addition or Subtraction

Math operators perform arithmetic operations add or subtract operation on any two analog input channels. When you select addition or subtraction, the Source A and Source B values are added or subtracted point by point, and the result is displayed.
1. Press the Math key on the front panel to enter the MATH function menu.
2. Press the Source A and Source B softkey respectively, and then turn the Universal Knob to select the source to do math operation. Addition or Subtraction can be applied between analog channels.
3. Press the Operation softkey and then turn the universal to select + or - to make addition or subtraction operation. The resulting math waveform is displayed in white and labeled with “M”.
Figure 9.1 CH 1 + CH 2
4. To invert the math waveform, press the Invert button and set the option to On to invert the display of the math waveform.
Math 88

9.2.2 Multiplication and Division

Math operators perform arithmetic operations add or subtract operation on any two analog input channels. When you select multiplication or division, the Source A and Source B values are added or subtracted point by point, and the result is displayed.
1. Press the Math key on the front panel to enter the MATH function menu.
2. Press the Source A and Source B softkey respectively, and then turn the Universal Knob to select the source to do math operation. Addition or Subtraction can be applied between analog channels.
3. Press the Operation softkey and then turn the universal to select * or / to make addition or subtraction operation. The resulting math waveform is displayed in white and labeled with “M”.
Figure 9.2 CH 1 * CH 2
4. To invert the math waveform, press the Invert button and set the option to On to invert the display of the math waveform.
Math 89

9.2.3 FFT Operation

FFT is used to compute the fast Fourier transform using analog input channels. FFT takes the digitized time record of the specied source and transforms it to the frequency domain. When the FFT function is selected, the FFT spectrum is plotted on the oscilloscope display as magnitude in dBV versus frequency. The readout for the horizontal axis changes from time to frequency (Hertz) and the vertical readout changes from V to dB.
FFT operation can facilitate the following works:
• Measure harmonic components and distortion in the device under test
• Measure the characteristics of the noise in DC power
• Analyze vibration
To display a FFT waveform
1. Press the Math key on the front panel to open the MATH function menu.
2. Press the Operator softkey and then turn the Universal Knob to select FFT. The resulting math waveform is displayed in white and labeled with “M”.
Figure 9.3 FFT Menu
3. Press the Source softkey, and then turn the Universal Knob to select the source to do FFT operation. Analog channels can be used as the source.
4. Press the Cong softkey to enter CONFIG menu.
Figure 9.4 FFT Cong Menu
• Press the Maximum points softkey, and then turn the Universal Knob to select the Maximum points.
• Press the Window softkey, and then turn the Universal Knob to select an appropriate window.
Spectral leakage can be considerably decreased when a window function is used. The oscilloscope provides ve windows (Rectangle, Blackman, Hanning, Hamming and Flattop) which have dierent characteristics and are applicable to measure dierent waveforms. Select the window function according to dierent waveforms and their characteristics. Refer to table 9.2 to make an appropriate option according to the input signal.
Math 90
Window Applications and Characteristics
These are normally used when the signal is transient (completely contained in the time­domain window) or known to have a fundamental frequency component that is an integer
Rectangle
multiple of the fundamental frequency of the window. Signals other than these types will show varying amounts of spectral leakage and scallop loss, which can be corrected by selecting another type of window.
Hanning
Hamming
Flat Top
These reduce leakage and improve amplitude accuracy. However, frequency resolution is also reduced.
These reduce leakage and improve amplitude accuracy. However, frequency resolution is also reduced.
This window provides excellent amplitude accuracy with moderate reduction of leakage, but with reduced frequency resolution.
Blackman It reduces the leakage to a minimum, but with reduced frequency resolution.
Table 9.2 FFT Window
• Press the Auto set softkey to automatically set the appropriate parameters for the FFT measurement.
• Press the Display softkey to select Split, Full Screen or Exclusive display mode.
Split: Time domain waveform and frequency domain waveform are displayed separately. The time domain wave­form is on the upper half screen, while the frequency domain waveform is located within the lower half of the display. In Split mode, if Zoom is enabled, the zoom waveform and the frequency domain waveform are displayed on the lower half screen together.
Full Screen Exclusive
: Time-domain waveform and frequency-domain waveform are displayed together.
: Only the frequency-domain waveform is displayed.
• Press the Mode softkey to select Normal, Max-Hold or Average. When you select Average, it is necessary to
set the average times.
5. Press the Vertical softkey to enter VERTICAL menu.
Figure 9.5 Vertical Menu
• Press the Scale softkey, and then turn the Universal Knob to select the desired vertical FFT scale
• Press the Ref Level softkey, and then turn the Universal Knob to select the desired vertical FFT oset.
• Press the Unit softkey to select the unit of vertical axis. The units of the vertical axis can be dBVrms, dBm,
Vrms or dBArms, Arms, which use a logarithmic scale or a linear scale to display vertical amplitude respectively.
• Press the Ext Lode softkey and then turn the Universal Knob to select the external load value.
5. Press the Horizontal softkey to enter HORIZONTAL menu.
Figure 9.6 Horizontal Menu
• Press the Center softkey, and then turn the Universal Knob to select the desired center frequency.
• Press the Hz/div softkey, and then turn the Universal Knob to select the desired resolution frequency.
Math 91
6. Press the Tools softkey to enter TOOLS menu. Press the Type softkey to select the type of tools. The type of the tools can be Peaks, Markers, or O. Peaks: Automatically mark the peak of the current FFT waveform according to the search conguration.
Figure 9.7 FTT Peaks Menu
• Press the Show Table softkey to display a table of peak, and press Show Frequency softkey to display the
frequency value of the peak. Press the Sort By softkey continuously to set the table sorting by Amplitude or Frequency.
• Press the Search Cong softkey to set search conguration.
Figure 9.8 Search Cong Menu
a. Press the Threshold softkey and turn the Universal Knob to set the minimum peak amplitude. Only peaks
larger than the peak limit can be judged as peaks.
b. Press the Excursion softkey and turn the Universal Knob to set the dierence between the peak value and
the minimum Amplitude on both sides. Only when the dierence is greater than the peak value of peak oset can the peak value be determined.
Markers
• Press Markers Control softkey to enter MARKER Menu.
a. Press the Marker softkey continuously to select the marker number from NO.1 NO.8.
b. Press the Show Marker softkey continuously to turn on or o the selected mark.
c. Press the Frequency softkey to set the frequency value of the selected marker.
: Customize the marker locations on the FFT waveform based on search conguration.
Figure 9.9 FFT Markers Menu
Figure 9.10 Marker Control Menu
d. Press the Next Peak softkey to move the selected mark to the next peak. And press the Next Amplitude
softkey to move the selected marker to the next peak with lower amplitude.(Up to 10 peaks are supported)
• Press the Search Cong softkey to enter the SEARCH menu. Similar to the search conguration for peak
markers.
Math 92
• Press the Markers on Peaks softkey to set the markers on peaks, and press the Markers on harmonics softkey
to set the markers on harmonics.
• Press the Show Table softkey to show the amplitude of the markers; press the Show Frequency softkey to
show the frequency value of the markers, and press the Show Delta softkey to show the delta amplitude between markers.
Figure 9.11 shows the FFT waveform displayed on the split screen and the peak tool is turned on:
Figure 9.11 FFT Waveform
Note:
• Signals with DC components or deviation would cause error or deviation of the FFT waveform components. To re­duce the DC components, set the Channel Coupling to AC.
• To reduce the random noise and aliasing frequency components of repetitive or single pulse, set the Acquisition of the oscilloscope to Average.
Measure FFT Waveform
To make cursor measurements, press the Cursors key, and then press the Mode softkey to select On to turn the cursors, Use the X1 and X2 cursors to measure frequency values and the dierence between two frequency values (ΔX). Use the Y1 and Y2 cursors to measure amplitude in dB and dierence in amplitude (ΔY). You can nd the frequency value at the rst occurrence of the waveform maximum by using the X at Max Y measurement.
Note:
Refer to the Cursors section for a description on using cursors.
Math 93

9.3 Math Function Operation

The oscilloscope supports math function operation including dierential (d/dt), integral (∫𝑑𝑡) and square root (√).
9.3.1 Dierentiate
𝑦(𝑖 +Δ𝑡) − 𝑦(𝑖 − Δ𝑡)
𝑑𝑖 =
d/dt (dierentiate) calculates the discrete time derivative of the selected source. Where:
• d = dierential waveform
• y = channel 1, 2, 3, or 4 data points
• i = data point index
• Δ t = point- to- point time dierence
The dx option under d/dt math function operation menu shows the point- to- point time dierence, and it ranges from
0.02div to 0.40div. “div” indicates the number of the pixel points that each division has. The oscilloscope has 50 pixel points per division. Take 0.2div as an example: 0.2*50 = 10. It means to calculate the ten point’s discrete time derivative of the selected source, and the Δ t is the ten point’s point-to-point time dierence.
2Δ𝑡
Figure 9.12 Dierence Function Operation
You can use dierentiate to measure the instantaneous slope of a waveform. For example, the slew rate of an operational amplier may be measured using the dierentiate function
Note:
Because dierentiation is very sensitive to noise, it is helpful to set acquisition mode to Average.
Math 94

9.3.2 Integrate

dt (integrate) calculates the integral of the selected source. You can use integrate to calculate the energy of a pulse in volt-seconds or measure the area under a waveform.
dt plots the integral of the source using the "Trapezoidal Rule". The equation is:
𝐼𝑛= 𝑐0+Δ𝑡 +𝑛∑
𝑖=0
𝑦
𝑖
Where:
• I = integrated waveform
• Δt = point- to- point time dierence
• y = channel 1, 2, 3, or 4 data points
• co = arbitrary constant
• i = data point index
The integrate operator provides an Oset softkey that lets you enter a DC oset correction factor for the input signal. Small DC oset in the integrate function input (or even small oscilloscope calibration errors) can cause the integrate function output to "ramp" up or down. This DC oset correction lets you level the integrate waveform.
Figure 9.13 Integral with Oset

9.3.3 Square Root

Square root (√) calculates the square root of the selected source. Where the transform is undened for a particular input, holes (zero values) appear in the function output.
Math 95
Figure 9.14 Square Root

Cursors

Cursors are horizontal and vertical markers that indicate X-axis values and Y-axis values on a selected waveform source. Cursors can be used to make custom voltage and time measurements on the oscilloscope signals.

10.1 X Cursors

X cursors are vertical dashed lines that adjust horizontally and can be used to measure time (when the source is FFT waveform, X cursors measure frequency).
X1 cursor is the left (default position) vertical dotted line; it can be moved to any place of the screen.
X2 cursor is the right (default position) vertical dotted line; it can be moved to any place of the screen.
Use the Universal Knob to set the X1 and X2 cursor values. The values are displayed in the cursors box along with the dierence between X1 and X2 (T) and 1/T. (See gure 10.1)
When cursor type is set to X2-X1, use Universal Knob to move the X1 and X2 cursors together. The value under the menu option is the dierence between the X1 and X2 cursors.
Figure 10.1 X Cursors Box
Cursors 97

10.2 Y Cursors

Y cursors are horizontal dotted lines that adjust vertically and can be used to measure voltage (V) or current (A). When the cursors source is the math function, the unit will match the math function.
Y1 cursor is the top (default position) horizontal dotted line; it can be moved to any vertical place of the screen.
Y2 cursor is the down (default position) horizontal dotted line; it can be moved to any vertical place of the screen.
Use the Universal Knob to set the Y1 and Y2 cursor values and the values are displayed in the cursors box along with the dierence between Y1 and Y2 (Y). (See gure 10.2)
When set cursor type to Y2-Y1, use Universal Knob will move the Y1 and Y2 cursors together. The value under the menu option is the dierence between the Y1 and Y2 cursors.
Figure 10.2 Y Cursors Box
Cursors 98

10.3 Make Cursor Measurements

1. Press the Cursors key on the front panel to enter the CURSOR function menu.
2. Press the Mode softkey and set the cursors mode to Manual or Track.
3. Press the Source softkey, and then use the Universal Knob to select the desired source. Only analog channels, math waveforms and reference waveforms that are displayed are available for cursors.
4. Press the X Ref and Y Ref softkey to set the reference of X cursors and Y cursors.
Position
display.
Delay/Oset: when the horizontal/vertical is changed, the value of X/Y cursors remain xed.
5. To make cursor measurements:
• To measure the horizontal time, use the Universal Knob to move the X1 and X2 cursors to desired place. If
necessary, set the cursor type to X2-X1, move X1 and X2 cursors together.
• To measure vertical voltage or current, use the Universal Knob to move the Y1 and Y2 cursors to desired place.
If necessary, set the cursor type to “Y2-Y1”, move Y1 and Y2 cursors together.
• To adjust the transparence of the cursors message box, press the Display/Persist softkey and go to the second
page, press the Transparence (20% to 80%) softkey and then turn the Universal Knob to adjust the transparence to the desired value.
Cursor examples:
: when the horizontal/vertical scale is changed, the X/Y cursors remain xed to the grid position on the
Figure 10.3 Cursor Track Mode

Measure

The oscilloscope provides measurements of 36 waveform parameters and the statistics. It contains voltage, time, and delay parameters.
Voltage and time parameters are under Type option. The results of the last four selected measurements are displayed at the bottom of the screen and above the menu. Delay parameters are under All Measure submenu. Set the Delay option to On to display all the delay parameters.

11.1 Type of Measurement

11.1.1 Voltage Measurements

Voltage measurements include 15 kinds of voltage parameters measurements.
Figure 11.1 Voltage Measurements
1. Peak-Peak: Dierence between maximum and minimum data values.
2. Maximum
3. Minimum: Lowest value in input waveform.
4. Amplitude: Dierence between top and base in a bimodal signal, or between max and min in a unimodal signal.
5. Top: Value of most probable higher state in a bimodal waveform.
6. Base: Value of most probable lower state in a bimodal waveform.
7. Mean
8. Cycle mean: Average of data values in the rst cycle.
9. Stdev: Standard deviation of all data values
10. Cycle Stdev: Standard deviation of all data values in the rst cycle
11. Rms: Root mean square of all data values.
12. Cycle RMS
: Highest value in input waveform.
: Average of all data values
: Root mean square of all data values in the rst cycle.
Measure 100
13. Overshoot means rising edge overshoot and FOV means falling edge overshoot.
14. Preshoot X cursors show which edge is being measured (edge closest to the trigger reference point).
: Overshoot is distortion that follows a major edge transition expressed as a percentage of Amplitude. ROV
Figure 11.2 Overshoot
𝑅𝑖𝑠𝑖𝑛𝑔 𝑒𝑑𝑔𝑒 𝑜𝑣𝑒𝑟𝑠ℎ𝑜𝑜𝑡 =
𝐹𝑎𝑙𝑙𝑖𝑛𝑔 𝑒𝑑𝑔𝑒 𝑜𝑣𝑒𝑟𝑠ℎ𝑜𝑜𝑡 =
: Preshoot is distortion that precedes a major edge transition expressed as a percentage of Amplitude. The
𝑙𝑜𝑐𝑎𝑙 𝑀𝑎𝑥𝑖𝑚𝑢𝑚− 𝐷 𝑇𝑜𝑝
𝐴𝑚𝑝𝑙𝑖𝑡𝑢𝑑𝑒
𝐵𝑎𝑠𝑒 −𝐷 𝑙𝑜𝑐𝑎𝑙 𝑀𝑖𝑛𝑖𝑚𝑢𝑚
𝐴𝑚𝑝𝑙𝑖𝑡𝑢𝑑𝑒
𝑋 100
𝑋 100
Figure 11.3 Preshoot
𝑅𝑖𝑠𝑖𝑛𝑔 𝑒𝑑𝑔𝑒 𝑝𝑟𝑒𝑠ℎ𝑜𝑜𝑡 =
𝐹𝑎𝑙𝑙𝑖𝑛𝑔 𝑒𝑑𝑔𝑒 𝑝𝑟𝑒𝑠ℎ𝑜𝑜𝑡 =
15. Level@X: The voltage value between the trigger point and the vertical position of the channel
𝑙𝑜𝑐𝑎𝑙 𝑀𝑖𝑛𝑖𝑚𝑢𝑚− 𝐷 𝑇𝑜𝑝
𝐴𝑚𝑝𝑙𝑖𝑡𝑢𝑑𝑒
𝐵𝑎𝑠𝑒 −𝐷 𝑙𝑜𝑐𝑎𝑙 𝑀𝑎𝑥𝑖𝑚𝑢𝑚
𝐴𝑚𝑝𝑙𝑖𝑡𝑢𝑑𝑒
𝑋 100
𝑋 100
Loading...