B&K Precision 2682 User's Manual

Safety Summary
The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
Read and understand the safety and operational information in this manual.
Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating the instrument
at an incorrect line voltage will void the warranty.
Make all connections to the instrument before applying power.
Do not operate the instrument in ways not specied by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design, manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply with these requirements.
Category rating
The IEC 61010 standard denes safety category ratings that specify the amount of electrical energy available and the voltage impulses that may occur on electrical conductors associated with these category ratings. The category rating is a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of the circuit to be tested, which denes the voltage impulses expected and required insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected to the mains sup
ply. The voltages in the environment are typically derived from a limited-energy transformer or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the mains supply at
a standard wall outlet or similar sources. Example measurement environments are portable tools and house hold appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be connected to the mains installa
tion of a building. Examples are measurements inside a building’s circuit breaker panel or the wiring of permanently­installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to the primary power en
tering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what is specied in this manual for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the instrument’s category rating to maintain the instrument’s category rating. Failure to do so will lower the category rating of the measuring system.
1
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be 115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This instrument is grounded through the ground conductor of the supplied, three-conductor AC line power cable. The power cable must be plugged into an approved three-conductor electrical outlet. The power jack and mating plug of the power cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive parts (including control knobs) may provide an electric shock. Failure to use a properly-grounded approved outlet and the recommended three-conductor AC line power cable may result in injury or death.
Unless otherwise stated, a ground connection on the instrument’s front or rear panel is for a reference of potential only and is not to be used as a safety ground. Do not operate in an explosive or ammable atmosphere.
Do not operate the instrument in the presence of ammable gases or vapors, fumes, or nely-divided particulates.
The instrument is designed to be used in oce-type indoor environments. Do not operate the instrument
In the presence of noxious, corrosive, or ammable fumes, gases, vapors, chemicals, or nely-divided particulates.
In relative humidity conditions outside the instrument’s specications.
In environments where there is a danger of any liquid being spilled on the instrument or where any liquid can condense
on the instrument.
In air temperatures exceeding the specied operating temperatures.
In atmospheric pressures outside the specied altitude limits or where the surrounding gas is not air.
In environments with restricted cooling air ow, even if the air temperatures are within specications.
In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature range is 0∘C to 40∘C and 20% to 80% relative humidity, with no condensation allowed. Measurements made by this instrument may be outside specications if the instrument is used in non-oce-type environments. Such environments may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise, strong electric elds, or strong magnetic elds.
2
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the instrument, remove the instrument’s power cord, remove the instrument from service, label it as not to be operated, and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual. Not for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a life-support device or system.
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must be made by qualied service-trained maintenance personnel who are aware of the hazards involved when the instrument’s covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may exist when the covers are removed. To avoid injuries, always disconnect the power cord from the instrument, disconnect all other connections (for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there are no hazardous voltages present on any conductors by measurements with a properly-operating voltage-sensing device before touching any internal parts. Verify the voltage-sensing device is working properly before and after making the measurements by testing with known-operating voltage sources and test for both DC and AC voltages. Do not attempt any service or adjustment unless another person capable of rendering rst aid and resuscitation is present. Do not insert any object into an instrument’s ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit exists.
Fuse replacement must be done by qualied service-trained maintenance personnel who are aware of the instrument’s fuse requirements and safe replacement procedures. Disconnect the instrument from the power line before replacing fuses. Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings specied in this manual or on the back of the instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a re. Failure to use the specied fuses will void the warranty.
Servicing
3
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to B&K Precision for service and repair to ensure that safety and performance features are maintained.
For continued safe use of the instrument
Do not place heavy objects on the instrument.
Do not obstruct cooling air ow to the instrument.
Do not place a hot soldering iron on the instrument.
Do not pull the instrument with the power cord, connected probe, or connected test lead.
Do not move the instrument when a probe is connected to a circuit being tested.
Safety Symbols
Symbol Name Description
DANGER indicates a hazardous situation which, if not avoided, will result in death or serious
injury.
WARNING indicates a hazardous situation which, if not avoided, could result in death or serious
injury
CAUTION indicates a hazardous situation which, if not avoided, will result in minor or moderate
injury
Caution Refer to the text near the symbol.
Electric Shock hazard
Alternating current (AC)
Chassis ground
Earth ground
On (Power) This is the In position of the power switch when instrument is ON.
Off (Power) This is the Out position of the power switch when instrument is OFF.
NOTICE is used to address practices not related to physical injury.
4
Contents
1 Product Overview 8
1.1 Package Contents 8
1.2 Dimensions 8
1.3 Input Power Requirements 8
1.4 Fuse 9
1.4.1 Fuse Replacement 9
1.5 Power-up 10
1.6 Front Panel 10
1.7 Rear Panel 11
1.8 Built-in Help 11
2 Front Panel 13
3 Display 15
4 General Spectrum Analysis 16
4.1 Frequency 16
4.1.1 Center Frequency 16
4.1.2 Start/Stop Frequency 17
4.1.3 Freq Step 17
4.1.4 Peak to CF 17
4.1.5 CF to Step 17
4.2 Span 17
4.2.1 Span 17
4.3 Amplitude 18
4.3.1 Ref Level 18
4.3.2 Attenuator 18
4.3.3 RF Preamp 19
4.3.4 Units 19
4.3.5 Scale 19
4.3.6 Scale Type 19
4.3.7 Ref Offset 19
4.3.8 Corrections 19
4.4 Auto Tune 20
5 Settings 22
5.1 Bandwidth Menu 22
5.1.1 Resolution Bandwidth 22
5.1.2 Video Bandwidth 22
5.1.3 Video to Resolution Bandwidth Ratio 22
5.1.4 Average Type 22
5.2 Trace 23
5.2.1 Trace Averaging 24
5.2.2 Math 24
5.3 Detect 25
5.4 Sweep 25
5.4.1 Sweep Time 26
5.4.2 Sweep Rule 26
5.4.3 Single/Continuous sweeping 26
5.4.4 26
5.5 Trigger 26
5
5.5.1 Free Run 26
5.5.2 Video 26
5.5.3 External 27
5.6 Limit 27
5.6.1 Limit Switch 27
5.6.2 Edit 27
5.6.3 Test 27
5.6.4 Setup 27
5.7 TG 28
5.7.1 TG 28
5.7.2 TG Level 28
5.7.3 TG Level Offset 28
5.7.4 Normalize 28
5.8 Demod 29
5.8.1 Demod (AM/FM) 29
5.8.2 Earphone 29
5.8.3 Volume 29
5.8.4 Demod Time 29
6 Measurement Setup 30
6.1 Channel Power 30
6.2 ACPR 30
6.3 Occupied BW 30
6.4 T-Power 30
6.5 Meas Off 30
6.6 ACPR 31
6.6.1 OBW 31
7 Markers 33
7.1 Spec Analyzer Mode 33
7.1.1 Normal Marker 34
7.1.2 Delta Markers 34
7.1.3 Delta Pair Markers 34
7.1.4 Off 34
7.2 Reection Measurement 35
7.3 Marker Table 35
7.4 Marker To 35
7.5 Marker Functions 36
7.5.1 Noise Marker 36
7.5.2 N dB BW 37
7.5.3 Frequency Counter 37
7.5.4 Read Out 37
7.6 peak 37
7.6.1 Search Cong 38
8 System Settings 40
8.1 Language 40
8.2 Power On/Preset 40
8.3 Interface Cong 40
8.4 System Info 40
8.5 Data and Time 41
8.6 Self Test 41
8.1 display 41
8.2 le 41
8.2.1 Browser 41
8.2.2 Open/Load 41
6
8.2.3 Type 42
8.2.4 Save 42
8.2.5 Create Folder 42
8.2.6 Delete 42
8.2.7 Operate 42
8.3 preset 42
8.4 couple 42
8.5 help 43
8.6 save-1 43
9 LIMITED THREE-YEAR WARRANTY 46
10 Service Information 47
7
Product Overview
The series spectrum analyzer has a frequency range from 9 kHz up to 2.1 GHz (Model 2682) /3.2 GHz (Model 2683), numerous measurement functions are included.
All-Digital IF Technology
Frequency Range from 9 kHz up to 3.2 GHz
10 Hz Minimum Resolution Bandwidth(RBW)
Preamplier
Up to 3.2 GHz Tracking Generator
Reection Measurement Kit (Opt.)
Advanced Measurements
10.1 inch display
1.1 Package Contents
1 x Model 2680 Spectrum Analyzer
Power cord
1.2 Dimensions
Figure 1.1 Front View
1.3 Input Power Requirements
The instrument has a selectable AC input that accepts line voltage and frequency input within:
8
Figure 1.2 Top View
Figure 1.3 Side View
1.4 Fuse
An AC input fuse is necessary when powering the instrument. The fuse is located at the back of the instrument. If the fuse needs to be replaced, ensure the AC input power cord is disconnected from the instrument prior to replacement. Refer to Table 1.2 for fuse requirements.
Before replacing fuse, disconnect the AC power cord rst to prevent electric shock. Only use same rating of the fuse. Using a different fuse may damage the instrument.
1.4.1 Fuse Replacement
Check and/or Change Fuse
Locate the fuse box above the AC input in the rear panel.
Remove the small insert. This piece holds the fuse.
Check and replace fuse for the desired line voltage.
9
Frequency Voltage
47 to 60 Hz 100 to 220 V (±10%)
400 Hz 110 V (±10%)
Table 1.1 Input
Power Requirements
Line Voltage Fuse
110 V T 2 AL, 250 V
220 V T 1 AL, 250 V
Table 1.2 Fuses
1.5 Power-up
1. Verify AC Input Voltage Verify and check to make sure proper AC voltages are available to power the instrument. The AC voltage range must meet the acceptable specication as explained in Section 1.3.
2. Connect Power and turn on analyzer Connect AC power cord to the AC receptacle in the rear panel. The power button on the front panel should fade in and out. Press the power button to turn ON the instrument. The button should illuminate continuously and show the startup splash screen.
1.6 Front Panel
Front Panel Items
1 User Graphical
Interface
3 Function Keys 4 Knob
5 Arrow Keys 6 RF Input
7 Numeric Keyboard 8 TG Output
9 Earphone interface 10 USB Host
11 Power Switch
2 Menu Control Keys
Figure 1.4 Front View
10
1.7 Rear Panel
Item Description
1 Handle
2 USB Interface supporting:USB Interface supporting: USBTMC, USBCDC (virtual COM)
3 LAN Interface
4 REF IN 10 MHz
5 REF OUT 10 MHz
6 Trigger in
7 Kensington Lock
8 AC Power Input
Figure 1.5 Rear Panel
1. Handle
2. USB Interface (supports USBTMC, and USBCDC (virtual COM))
3. Ethernet
4. Reference in 10 MHz
5. Reference out 10 MHz
6. Trigger input
7. Security lock
8. AC power connector
1.8 Built-in Help
The built-in help system provides information about every function key at the front panel and every menu soft key. How to acquire built-in help
11
Press the “Help” key and a prompt about how to obtain help information will appear. Then press another key and relevant help information will be shown if available. Close the current help information When the help information show at the center of the screen. Press the the help information again to close it.
12
Front Panel
Figure 2.1 Function Keys
13
Function keys Description
Frequency Frequency menu (start, stop, center)
Span Frequency Span menu
Amplitude Signal amplitude menu (ref level, attenuator, pre-amp, scale, corrections)
Auto Tune Automatically setup the instrument
BW Bandwidth menu (resolution, video BW, averaging, lter)
Trace Trace setup menu and math menu
Sweep Sweep menu (time, number of sweeps, mode)
Detect Detector menu (peak, sample, average)
Trigger Trigger menu (Video Trigger/ External Trigger)
Limit Pass, Fail Limit setup menu
TG Tracking Generator menu
Demod Demodulation menu (AM, FM)
Marker Marker setup menu
Marker-> Apply marker to other analyzer functions (e.g. set center frequency to marker point)
Peak Set the marker to the highest peak
Marker Fn Marker function menu (enable, type, counter)
Meas Measurement menu (power, BW, spectrum monitor, etc...)
Meas Setup Conguration menu for an active measurements
System System setup menu
Mode Mode selection - spectrum or reection
Display Display setup (grid brightness, text, line presence)
File File browser
Preset Return to the default instrument state
Couple Control RBW, VBW, Attenuator, Freq Step, and Sweep time
Help Help menu
Save Screenshot save key
Table 2.1 Function Keys
14
Display
Figure 3.1 LCD Display
15
General Spectrum Analysis
Central to the use of the spetrum analyzer is the ability see a large swath of frequencies and zoom into areas of interest within the range of the analyzer. By using the frequency, span, and amplitude menus provides the fundamental means of moving around the sprectrum, and tuning to desired frequencies. In each menu, consistent with other menus, the values of each setting are adjusted by selecting them via the soft keys, and using the knob and keypad to make changes.
Frequency Menu Change the center, start, and stop frequencies.
Span Dene the frequency range of interest.
Amplitude Adjust the gain/attenuation of the system front end. The Pre-amplier is congured in this menu.
Many settings have a multiplier that needs to be applied to set the correct value. For example, setting the center frequency to 100 MHz can be done with the keypad by selecting “Center Freq” from the “Frequency” menu, entering 100 via the keypad and selecting the MHz soft-key on screen.
4.1 Frequency
Set the frequency parameters and functions of the analyzer. Each setting change restarts the frequency sweep. The displayed frequency range depends on the Start Frequency, Center Frequency and Stop Frequency. Changing any of them causes the others to change accordingly. The value of each setting is modied 3 different ways:
Rotary knob Rotate to change the setting
Numeric Keys Enter the frequency value directly and select the multiplier (Hz, MHz, GHz)
Up/Down Keys Keys below the knob increase or decrease the frequency setting in increments of the ``Freq Step’’ setting.
Figure 4.1 Frequency Menu
4.1.1 Center Frequency
In the Frequency menu, select the button beside the Center Frequency menu item to congure the center frequency. If the span setting does not hit the limits of the analyzer, the start and stop frequencies move along with the center frequency. That is, the span is not changed by changing the center frequency generally.
16
4.1.2 Start/Stop Frequency
Set the start or stop frequency of the current channel. Changing the start or stop frequencies does not change the other, but does change the span and center frequencies.
4.1.3 Freq Step
The frequency step setting denes the frequency jump applied when moving the center, start, and stop frequencies by the Up/Down keys below the knob. Use this setting to quickly move up and down the spectrum or to jump useful amounts.
4.1.4 Peak to CF
Execute a peak search and make the peak’s frequency the center frequency.
4.1.5 CF to Step
Set the current center frequency to be the “Freq Step” value. This may be useful in stepping through the harmonics of a signal.
4.2 Span
Set the span of the analyzer. The change of this parameter will affect the frequency parameters and restart the sweep.
Figure 4.2 Span Menu
4.2.1 Span
The span setting denes the range of frequencies shown. It is another representation of the “Stop Frequency” subtracted from the “Start Frequency”. Changing the span value primarily changes the start and stop frequencies, but can also change the center frequency if required to make the settings consistent. As a reminder, the center frequency and other settings are shown at the bottom of the display. Note the following additional consequences of using the span setting:
17
In manual span mode. The span can be set down to 100Hz and up to the full span
Modifying the span in non-zero span mode may cause an automatic change in both CF step and RBW if they are in
Auto mode. Besides, the change of RBW may inuence VBW (in Auto VBW mode).
Variation in the span, RBW or VBW would cause a change in the sweep time.
The remaining settings:
Full Span Set the span of the analyzer to the maximum.
Zero Span Analyze one frequency. Set the span of the analyzer to 0Hz. In this mode, the horizontal axis represents
time. See Section ?? for more information.
Zoom In Set the span to half of its current value.
Zoom Out Set the span to twice the current value.
Last Span Set the span to the previous span setting.
X-Scale Set the horizontal axis to be linear or logarythmic.
4.3 Amplitude
Set the signal amplitude setting of the analyzer with this menu. See Figure 4.3.
Figure 4.3 Amplitude Setting Menu
4.3.1 Ref Level
This setting is the upper signal limit of the display. The value is displayed at the upper left corner of the screen. The maximum reference level available is affected by the maximum mixing level, input attenuation is adjusted under a constant maximum mixing level in order to fulll the following condition: LRef <= aRF-aPA-30dBm.
4.3.2 Attenuator
This setting denes the input attenuator level. This can be set manually or automatically determined by the analyzer.
18
4.3.3 RF Preamp
For small signals, the analyzer also includes an amplier. For example, to clearly see radio stations far from the trans mitters, it is necessary to enable the pre-amplifer. Used in concert with the reference level and attenuator, clearer signals can be found and measured.
4.3.4 Units
Set the Y-axis scale units to dBm, dBmV, dBuV, dBuA, Volts and Watts. The default setting is dBm.
4.3.5 Scale
Set the logarithmic units per vertical grid division on the display. This function is only available when the scale type is set to ``log’’.
4.3.6 Scale Type
Set the scale type of Y-axis to linear or logarhythmic. The default setting is the logarhythmic scale. In Lin mode, Scale value cannot be changed.
4.3.7 Ref Offset
Assign an offset to the reference level to compensate for gains or losses generated between the device under measurement and the analyzer. Changing this value changes both the reference level readout and the amplitude readout of the marker, but does not impact the position of the curve on the screen.
4.3.8 Corrections
Amplitude corrections are available in this sub-menu in order to compensate for the gain or loss from external devices such as Antennae and Cables. When amplitude correction is enabled, both the trace and related measurement results are corrected. Corrections for RF Input impedance and at specic points are congured here. Correction settings and points are also loaded and saved from this menu by pressing the “Save/Load” menu and working with the le browser. 4 correction tables may be loaded as well as a correction for signal impedance (50/75 Ohms). See Figures 4.4, 4.5, and
4.6 for the menus shown.
If multiple correction factors are enabled, all related data will be used for amplitude correction.
Figure 4.4 Corrections Setting Menu
19
Figure 4.5 Correction Point Menu
Figure 4.6 Correction Add Point Menu
4.4 Auto Tune
Search for signals automatically throughout the full frequency range, and adjust the frequency and amplitude for optimum display of one-key signal found in the search. While actively searching, “Auto Tune” is shown in the status bar on screen until the searching is nished.
20
Figure 4.7 Before Auto Tune
Figure 4.8 After Auto Tune
21
Settings
5.1 Bandwidth Menu
Settings for the “Resolution Bandwidth”, “Video Bandwidth”, and their ratio is set from this menu. Settings for the type of average and lter to use are also contained here. Access this menu by pressing the BW button on the instrument within the Setting area. See Figure 5.1.
Figure 5.1 Bandwidth Menu
5.1.1 Resolution Bandwidth
Set the desired resolution bandwidth in order to distinguish between signals which are close in frequency. Reducing RBW will increase the frequency resolution, but will increase the sweep time.
5.1.2 Video Bandwidth
Set the desired video bandwidth in order to lter out the noise outside the video band. Reducing the VBW will smooth the spectrum line to highlight small signals from noise, but will increase the sweep time.
5.1.3 Video to Resolution Bandwidth Ratio
Set the ratio of Video and resolution bandwidth. Improved measurements can be had depending on the signal being measured:
Sine signal Use 1 to 3 (for faster sweeps)
Pulse signal Use 10 (to reduce the inuence on the amplitude of transient signals)
Noise signal Generally use 0.1 (to obtain the noise average)
5.1.4 Average Type
3 types of averaging are available: log power (video), power (RMS), or voltage. When trace averaging is enabled, the average type is shown on the left side of the display.
22
Log Power Logarithmic (decibel) scale for all ltering and averaging processes. This scale is helps nd CW signals near
noise.
Power Average In this average type, all ltering and averaging processes work on the power (the square of the magnitude)
of the signal, instead of its log or envelope voltage. This scale is best for measuring the true time power of complex signals.
Voltage Average In this Average type, all ltering and averaging processes apply to the voltage of the envelope of the sig
nal. This scale is good for observing rise and fall behavior of AM or pulse-modulated signals such as radar and cellular transmitters.
5.2 Trace
Up to 4 different traces can be shown on screen at a time. Markers may be congured to make measurements relative to the traces as well. Figure 5.2 the “Trace” menu is shown as well as 3 on screen traces. The trace in lavender is an averaged max-hold, yellow is the average signal, and blue is the averaged minimum-hold trace. Labels for each trace are shown in the trace’s color on the lower left side of the display.
Figure 5.2 Trace Menu
Spectrum Analyzer allows for up to four traces to be displayed at the same time. Each trace has its own color. All traces can be independently congured. By default, spectrum analyzer will choose Trace A and set the type to Clear Write.
Clear Write Erases any data previously stored in the selected trace, and display the data sampled in real-time of each
point on the trace.
Max Hold Retains the maximum level for each trace point of the selected trace. Updates the data if a new maximum
level is detected in successive sweeps.
Min Hold Display the minimum from multiple sweeps for each point of the trace and update the data if a new min
imum is generated in successive sweeps.
View Holds and displays the amplitude data of the selected trace. The trace data is not updated as the ana
lyzer sweeps.
Blank Disable the trace display and all related measurements.
23
Figure 5.3 Trace Type
5.2.1 Trace Averaging
Trace signal averaging is set by the Avg Times setting in the Trace menu. Set the number of averages of the selected trace. A larger number of averages can reduce the noise and the inuence of other random signals; thus highlighting the stable signal characteristics.
5.2.2 Math
Additionally, a set of math functions can be applied to the measured signals. This can be useful for nding small signals near the noise oor. For example, using the difference between Max and Min hold traces and adding an offset to move the trace into the screen. To do so, open the Trace menu and select the Math function. The math conguration menu will be shown as in Figure 5.4.
Figure 5.4 Trace Math Menu
Variable X, Y Select A, B, or C as the math source data
24
Output Z The trace to use as the math output
Power: X-Y+Offset The difference in between the 2 selected signals. This compares the absolute power difference.
For example, 2 signals of -20 and -26 dBm (A and B respectively) yield approximately -21.2 dBm.
Power: X+Y+Offset The sum of 2 signals. This combines the absolute power levels.
Log: X+Offset Take a signal and offset it a value in dB. The “ref” is the offset of the signal from 0 dB.
Log: X-Y+Ref Take the relative difference between 2 signals.
Ex. A is -20 dB and B is -26 dB, the value of this will be 6dB. Use the “ref” value to move the signal.
5.3 Detect
The analyzer displays the sweep signal on the screen in the form of trace. For each trace point, the analyzer always captures all the data within a specic time interval and processes (Peak, Average, etc.). The captured data uses the detector currently selected and then displays the processed data on the screen. Selecting an appropriate detector type according to the actual application ensures the best measurement accuracy. The available detector types are:
Positive Peak Displays the maximum value of data sampled within the corresponding time interval.
Negative Peak Displays the minimum value of data sampled within the corresponding time interval.
Sample Displays the transient level corresponding to the central time point of the corresponding time interval. This
detector type is applicable to noise or noise-like signals.
Normal (also called a Rosenfell or Min/Max detector) Displays the maximum value and the minimum value of the
sample data segment in turn. For an odd-numbered data point, the maximum value is displayed, and for an even-numbered point, the minimum value. This helps to show the amplitude variation range of the sig nal.
Average Displays the average value of data sampled within the corresponding time interval.
5.4 Sweep
As signals are discriminated across the spectrum, the analyzers performs a frequency sweep. As signals are measured, they are displayed on screen. For a fast sweep time, this happens quickly. For a long sweep, a progress indicator is shown at the bottom of the screen. The corresponding waveform is updated accordingly. See Figure 5.5.
Figure 5.5 Sweep Menu
25
5.4.1 Sweep Time
Set the time needed for the spectrum analyzer to nish a sweep. The sweep time can be set automatically or manually. In zero span mode, the sweep time corresponds to the time shown on screen. Also note that while decreasing the sweep time speeds up measurement, it also can cause error if the specied sweep time is less than the minimum sweep time in Auto coupling. Aat this point, “UNCAL” is shown in the status bar of the screen.
5.4.2 Sweep Rule
The analyzer provides two sweep goals that help determine automatic sweep time:
Speed Activates the default fast sweep time rule.
Accuracy Activates the normal sweep time rule to ensure the measurement accuracy. Speed sweep time rule pro
vides a fast measurement function that decreases the sweep time. Using Fast Sweep will decrease the mea surement accuracy.
5.4.3 Single/Continuous sweeping
By default, the analyzer is set to continuously start a new sweep after the current one is complete. For some cases it may be helpful to stop the instrument after a single or set of sweeps. For continuous sweeping, set the Sweep setting to “Continuous”. To run a prescribed number of sweeps, set the Sweep setting to single and set the number of sweeps to run in the Numbers eld. In “Single” mode, press the Single soft-key to start sweeping. All sweeping is started when enabled and the conditions dened by the “Trigger” settings are met.
5.4.4
Sweep mode includes auto, sweep and FFT. Auto Mode - When the sweep mode is auto-coupled, the analyzer selects the sweep mode automatically between Sweep and FFT Mode. Sweep Mode - Work in point-by-point scanning, slow and time-consuming, ts when RBW is more than 10k Hz. FFT Mode - Work in Parallel scans, fast and short time. This FFT mode is only available when RBW is set to 1 kHz, 3 kHz, or 10 kHz.
5.5 Trigger
The trigger type can be Free Run, Video or External.
5.5.1 Free Run
The trigger conditions are satised at any time and the analyzer generates trigger signals continuously.
5.5.2 Video
A trigger signal will be generated when the system detects a video signal of which the voltage exceeds the specied video trigger level. Note: This function is invalid in Avg detection in zero span mode. Trigger level - Set the trigger level in video trigger. At this point, the trigger level line (TL) and value are displayed on the screen. You can use the numeric keys, knob or direction keys to modify this parameter. Please refer to “Parameter Setting” for more details.
26
5.5.3 External
In this mode, an external signal (TTL signal) is input from the TRIGGER IN connector at the rear panel and trigger signals are generated when this signal fullls the specied trigger edge condition. Set the trigger edge in external trigger to the rising (Pos) or falling (Neg) edge of the pulse. Note: When “Free Run” is selected, Trig Setup is grayed out and disabled.
5.6 Limit
Spectrum Analyzer supports the Pass/Fail test function. In this function, the measured curve is compared with the pre-edited curve. If the related rules are met, the result is “Pass”; or else is “Fail”.
5.6.1 Limit Switch
Select the desired limit line (upper or lower) for editing. Each limit line has its own switch. You can test according to both together, either one or neither.
5.6.2 Edit
Edit the properties of the limit lines. Table 2- 18 Pass/Fail Menu
. Explanation
Upper/Lower Select the desired limit line (upper or lower) for editing
Mode Select the line or point for editing. Set the number of the point to be edited if you selected the point
type. The range is from 1 to 100
Add point Add a new point for editing.
X-axis Edit the X-axis value (frequency or time) of the current point. If the X-axis unit is frequency and the
Ref Freq is enabled, edit the frequency difference between the frequency of the current point and the center frequency.
Amplitude Edit the amplitude of the current point or line. If the Ref AMPT is enabled, edit the amplitude
difference between the amplitude of the current point and the reference level.
Del Point Delete the point you are editing.
Del All Delete all point.
Save/Recall Save or load the limit le.
Table 5.1 Pass/Fail Menu
5.6.3 Test
Enable or disable the limit test function.
5.6.4 Setup
Fail to stop Select whether the instrument will continue or stop operation when a failure occurs. Buzzer Turn on or off the buzzer. When the buzzer is on, it beeps when a failure occurs. X Axis Set the X-axis unit to frequency or time unit. Note that all the points of the current limit line will be deleted when the X-axis unit changes.
27
5.7 TG
Set the parameter related to the tracking generator (TG).
5.7.1 TG
TG is used to enable or disable the TG. When the TG is enabled, a signal with the same frequency of the current sweep signal will be output from the [GEN OUTPUT 50 Ω] connector at the front panel. The power of the signal could be set through the menu.
5.7.2 TG Level
Set the output power of the signal of the tracking generator. You can use the numeric keys, knob or direction keys to modify this parameter. For more details, please refer to “Parameter Setting”.
5.7.3 TG Level Offset
Assign a certain offset to the output power of the TG when gains or losses occur between the TG output and external device in order to display the actual power value. Key Points:
This parameter only changes the readout of the TG output power, rather than the actual value of it.
The offset could be either a positive (gain in the external output) or a negative (loss in the external output).
You can use the numeric keys, knob or direction keys to modify this parameter. For more details, please refer to
“Parameter Setting”.
5.7.4 Normalize
Normalization can eliminate the error of TG Level. Before using this function, connect the [GEN OUTPUT 50 Ω] output terminal of the TG with the [RF INPUT 50 Ω] input terminal of the analyzer.
Normalize - Enable or disable the normalization. When enabled, the reference trace will be stored automatically after the current sweep nishes if no reference trace is stored before. During the reference trace storage, the corresponding prompt message is displayed. When normalization is enabled, the corresponding value of the reference trace will be subtracted from the trace data after every sweep.
Norm Ref Lvl - Adjust the vertical position of the trace on the screen by adjusting the reference level when normalization is enabled. Being different from the Ref Level function in the AMPT menu, this parameter has no inuence on the reference level of the analyzer. You can use the numeric keys, knob or direction keys to modify this parameter. For more details, please refer to “Parameter Setting”.
Norm Ref Pos - Adjust the vertical position of the normalization reference level on the screen by adjusting the reference position when normalization is enabled.
The function of this menu is similar to that of Norm Ref Lvl. When it is set to 0%, the normalization reference level is displayed at the bottom of the screen grid and at the top when it is set to 100%.
You can use the numeric keys, knob or direction keys to modify this parameter. For more details, please refer to “Parameter Setting”.
Ref Trace - Set whether to display the reference trace or not. If “View” is selected, the reference trace saved (Trace D) will be shown in “FreezeView” type.
28
Note: When normalization is enabled, the unit of Y-axis is “dB” and will not be inuenced by the denition in AMPT->Units. At this point, “(dB)” is displayed under the Y-axis scale in the user interface.
5.8 Demod
Press Demod at the front panel to enter the demodulation setting menu. Both AM and FM demodulations are available in this device.
5.8.1 Demod (AM/FM)
Set the demodulation type to AM or FM; or disable the demodulation function. The default is off. Key Points:
The system will enable a marker automatically, place it at the center frequency and perform AM (or FM) demodulation on this frequency point after you enable AM (or FM) demodulation.
Analyzer provides earphone jack and the demodulated signal can be output in audio frequency (AF) mode through the earphone. The frequency and intensity of AF denotes the frequency and amplitude of the signal respectively.
5.8.2 Earphone
Set the status of the earphone. When it is on, the demodulated signal can be heard through the earphone during the demodulation. By default, it is off.
5.8.3 Volume
Set the volume of the earphones.
5.8.4 Demod Time
Set the time for the analyzer to complete a signal demodulation after each sweep. If Earphone is set to “On”, you will hear the demodulated signal through the earphone during the demodulation. You can use the numeric keys, knob or direction keys to modify this parameter. For more details, please refer to “Parameter Setting”.
29
Measurement Setup
Provide measurement function, the screen will be divided into two parts, the above part is measure screen, displaying trace, the other part is used to display result of a measurement.
6.1 Channel Power
Measure the power and power density within the specied channel bandwidth. When this function is enabled, the span and resolution bandwidth are automatically adjusted to smaller values. Select Channel Power and press Meas Setup to set the corresponding parameters.
6.2 ACPR
Measure the powers of the main channel and adjacent channels as well as the power difference between the main channel and each of the adjacent channels. When this function is enabled, the span and resolution bandwidth of the analyzer are adjusted to smaller values automatically. Select ACPR and press Meas Setup to set the corresponding parameters.
6.3 Occupied BW
Integrate the power within the whole span and calculate the bandwidth occupied by this power according to the specied power ratio. The OBW function also indicates the difference (namely “Transmit Freq Error”) between the center frequency of the channel under measurement and the center frequency of the analyzer. Select Occupied BW and press Meas Setup to set the corresponding parameters.
6.4 T-Power
The system enters zero span mode and calculates the power within the time domain. The types of powers available include Peak, Average and RMS. Select T-Power and press Meas Setup to set the corresponding parameters.
6.5 Meas Off
Turn off all the Meas function.
1. meas-setup
2. Channel Power
Measurement Results: channel power and power spectral density.
Channel Power: power within the integration bandwidth.
Power Spectral Density: power (in dBm/Hz) normalized to 1 Hz within the integration bandwidth.
Measurement Parameters: Center Freq, integration bandwidth , Span, Span power 1 Center Freq Sets the center frequency of display this CF which is the same with the CF of the analyzer. Modifying this parameter will change the CF of the analyzer. Integration bandwidth
30
Set the frequency width of the channel to be tested and the power of the channel is the power integral within this bandwidth. You can use the numeric keys, knob or direction keys to modify this parameter. Span Set the frequency range of the channel. This span which is the same with the span of the analyzer is the frequency range of the sweep. Modifying this parameter will change the span of the analyzer.
The channel power span is related to the integration bandwidth and the range available is from integration bandwidth to integration bandwidth×20.
You can use the numeric keys, knob or direction keys to modify this parameter.
4Span Power
Set the integrated bandwidth to the sweep span of display. The channel power and power spectral density display on the screen simultaneously.
6.6 ACPR
Adjacent Channel Power Measurement: Main CH Power, Left channel power and Right channel power.
Main CH Power: display the power within the bandwidth of the main power
Left channel power : display the power of left channel and the power difference between the left channel and the
main channel (in dBc)
Right channel power: display the power of the right channel and the power difference between the right channel and the main channelin dBc
Measurement parameter: center frequency, main channel bandwidth, adjacent channel bandwidth and channel spacing 1 Center Frequency Sets the center frequency of display this CF which is the same with the CF of the analyzer. Modifying this parameter will change the CF of the analyzer. Main channel bandwidth Set the bandwidth of the main channel and the power of the main channel is the power integral within this bandwidth. You can use the numeric keys, knob and direction keys to modify this parameter. Adjacent channel bandwidth Set the frequency width of the adjacent channels.
The adjacent channel bandwidth is related to the main channel bandwidth and the range available is from main channel bandwidth/20 to main channel bandwidth×20.
You can use the numeric keys, knob or direction keys to modify this parameter.
Adjacent Channel space Set the difference between the center frequency of the main channel and the center frequency of the adjacent channels.
Adjusting this parameter will also adjust the distance between the upper/lower channel and the main channel.
You can use the numeric keys, knob or direction keys to modify this parameter.
6.6.1 OBW
OBW measurement: occupied bandwidth and transmit frequency error.
Occupied Bandwidth: integrate the power within the whole span and then calculate the bandwidth occupied by the power according to the specied power ratio.
31
Transmit Frequency Error: difference between the center frequency of the channel and the center frequency of the analyzer.
1. T-Power
T-Power: namely the power of the signal from the start line to the stop line. Measurement Parameter: center frequency, start line, stop line Center Frequency Sets the center frequency of display, this CF which is the same with the CF of the analyzer. Modifying this parameter will change the CF of the analyzer. Start line Set the left margin (in time unit) of T-Power measurement. The data calculated under this measurement is between the start line and stop line. You can use the numeric keys, knob or direction keys to modify this parameter. Stop line Set the right margin (in time unit) of T-Power measurement. The data calculated under this measurement is between the start line and stop line. You can use the numeric keys, knob or direction keys to modify this parameter.
32
Markers
Four main markers are available for making measurements of signal level, signal period, bandwidth, peaks, relative values, noise level, and others. Marker measurements apply to all traces captured on screen. For example, set Trace B to Max- Hold and make relative measurements to a trace set to Clear Write (live data). To congure these setting use the 4 buttons within the blue square labeled “Marker”. Each marker is shown in orange, often with the marker number above it. See Figure 7.1. The type and availabilty of markers depends on what spectrum analyzer mode is used, and the menus change accordingly.
Figure 7.1 Markers
7.1 Spec Analyzer Mode
Figure 7.2 Marker Menu
When markers are active, the data is displayed at the top of the screen. Also, by activating the “Marker Table”, data for each marker is shown on screen. See Figure 7.2.
33
Figure 7.3 Marker Table
7.1.1 Normal Marker
Mark an X position representing frequency or time, and the corresponding signal amplitude at that point In the Marker Table, the value shows the marker values.
7.1.2 Delta Markers
Mark and measure the difference between 2 points In the Marker Table, 2 points are shown.
The [1-4]r entry is the reference point and the data shown is the measurement value at that point.
• ΔMarker[1-4] is the offset point and the values are the change from [1-4]r.
The reference marker is placed at the center frequency on screen at the time of acivation. If another location is
desired, activate a normal marker, set the desired point, then change the marker type to delta.
When the Noise Marker function under the Marker Fn menu is activated, the result of the noise measurement is automatically corrected for and normalized to 1 Hz.
7.1.3 Delta Pair Markers
This setting creates a pair of markers and measures relative values between them, one is denoted the reference [1-4]r and the other is realtive to the reference labeled [1-4]. Marker movement is dened in 4 different ways:
Delta The relative marker [1-4] moves
Ref The reference marker [1-4]r moves
Span Both markers move toward and away from eachother equally
Center Both markers move together and maintain their relative spacing horizontally
7.1.4 Off
This marker mode disables the selected marker.
34
7.2 Reection Measurement
Markers available for reection measurements are much like Normal markers in spectrum analyzer mode. Marker measurements are shown in a table like Table ??. The return loss, reection coecient, and Voltage Standing Wave Ratio (VSWR) is shown for each marker. The frequency (shown in Hz or period (1/f)) of the marker or the time from the sweep start is also shown. This is the X-Axis value shown in the table.
Figure 7.4 Reection Measurement Marker Table
7.3 Marker Table
Display all the markers enabled on the lower portion of the screen, including marker number, trace number, marker readout type, X-axis readout and amplitude. Through this table you can view the measurement values of multiple points. The table allows for up to eight markers to be displayed at one time.
Figure 7.5 Marker Table
7.4 Marker To
Markers may be used to center, and change the sweep frequencies. Effective use of this function can help to quickly focus in on desired signals. Access this menu by pressing the “Marker→”.
35
If a Normal marker is selected, the center frequency will be set to the frequency of the current marker.
If a Delta, or Delta Pair marker is selected, the setting is set to the frequency of the Delta Marker.
Most of the functions do not apply to “Zero Span” mode.
The functions available here are shown in Figure ?? and described below:
MCF Set the center frequency of the analyzer to the frequency of the current marker
MCF Step Set the center frequency step of the analyzer to the frequency of the current marker.
MStart Freq Set the start frequency of the analyzer to the frequency of the current marker.
MStop Freq Set the stop frequency of the analyzer to the frequency of the current marker.
MRef Level Set the reference level of the analyzer to the amplitude of the current marker. This action also works in
Zero Span” mode.
ΔMSpan Set the span of the analyzer to the frequency difference between the two markers in Delta, or Delta Pair.
This function is not available for “Normal” markers.
ΔMCF Set the center frequency of the analyzer to the frequency difference between two Delta, or Delta Pair mark
ers
7.5 Marker Functions
In addition to showing frequency and level, markers may be congured to display more information. This information can be the noise at a point, noise bandwidth, and a frequency counter representaton of the point. To access the Marker
Function menu, press the “Marker Fn” button. Settings for each of the 4 available markers is opened on screen. See
Figure 7.6. Special marker functions include Noise Marker, N dB BW and Freq Counter.
Setting the marker function to off disables the functions listed in this section and does not disable the marker itself.
Figure 7.6 Marker Function Menu
7.5.1 Noise Marker
The Noise marker function displays the noise power spectral density measured in dBm/Hz. Activating the Noise Marker will rst set its to Normal type automatically. The noise measurement is the average noise level at the marked point
36
normalized to 1 Hz bandwidth. The measured value is effected by the detection and trace types used. More precise measurement is made if RMS Avg or Sample detection is used. This function can be used for measuring the C/N ratio.
7.5.2 N dB BW
N dB BW represents the frequency difference between two points that are located on either side of the current marker that fall within the dB fall (N<0) or rise (N>0) in amplitude as shown in Figure 7.7. When the measurement starts, the analyzer will search for the two points which are located at both sides of the current point with N dB fall or rise in amplitude and display the frequency difference between the two points in the active function area. "—-" would be displayed if the search fails.
Figure 7.7 N dB BW
7.5.3 Frequency Counter
The marker function can lastly be set to read the frequency at the marker’s location. Use this when the marker is set to jump to a peak and read the peak frequency. The frequency counter function is only applicable to marker 1.
7.5.4 Read Out
The marker x-axis position is represented in 3 ways: frequency, period, and ΔT. This setting changes the readout but does not change the actual measured value. The frequency and period represent the conventional idea, but ΔT shows the time difference between the marker and the start of the sweep. The Delta marker and Delta Pair marker show the sweep time difference between the delta marker and reference marker.
7.6 peak
The remaining button in the Marker section is the peak detector. Peak detection points may be used to set the center frequency, and easily list the 16 highest peaks shown on screen. By setting the “Search” settings, positive or negative peaks with a prescribed level may be found automatically. Negative peaks may be especially useful when using the tracking generator. Figure 7.8 shows the “Peak” menu screen and a scan of radio stations in Los Angeles.
PeakCF Set the rst
Next Peak Search for and mark the peak whose amplitude is closest to that of the current peak that meets the peak
search conditions.
37
Figure 7.8 Peak detector menu
Left Peak Move to the next peak to the left of the current peak.
Right Peak Move to the next peak to the left of the current peak.
Peak Peak Search for the higest/lowest peak and the minimum measured trace value and mark them with relative value
markers. The signal level difference is shown in the upper right corner of the screen.
Count Peak Enable continuous peak searching. When enabled, the system will always execute a peak search automat
ically after each sweep.
Peak Table Show the table of up to 16 peaks that meet the search condition settings.
7.6.1 Search Cong
Dene the conditions of peak search for various peak searches. A real peak should meet the requirements of both the “Peak Excursion” and “Peak Threshold” settings. See Figure 7.9, only the 4 peaks that meet the search conditions are marked and listed in the “Peak Table” below the spectrum plot.
Figure 7.9 Peak search menu
38
Peak Threshold Assign a minimum for the peak amplitude. Peaks whose amplitudes are greater than the specied peak
threshold are treated as real peaks.
Peak Excursion Set the excursion between the peak and the minimum amplitude on both sides of it. Peaks whose excur
sions are beyond the specied excursion are treated as real peaks.
Peak Type Set the peak search condition. The available options are Maximum and Minimum
39
System Settings
Set the system parameters.
1 Language
The spectrum Analyzer supports multi-language menu, Chinese and English build-in help and popup messages. Press this key to select the desired display language.
2 Power On/Preset
1Power On/Preset Set the power on setting to” factory (default)”, “Last” or one of “user”. Set the preset type to factory (default), “Last” or one of “User”. Last When “Last” is selected, settings before the last power-off would be recalled automatically at power on. User Select user-dened conguration from “user1” to “user6” If power on set to user, the device will recall the specied conguration after power on and press Preset at the front panel under any operation interface will recall the specied preset type.
2. USER-DEFINED
Save the current instrument setting as user-dened setting into the internal non-volatile memory. You can store and name up to 6 system states, and name every STA le. 3Factory When “factory” is selected, the device will recall the initial cong.
3 Interface Cong
The spectrum Analyzer support communications through LAN, USB as standard interface.
1. LAN
Cong or reset corresponding parameters of LAN.
1. Calibration
Auto Cal When “Auto Cal” open, spectrum Analyzer will process self-calibration regularly. Within half an hour after power-on, the device executes a self-calibration every 10 minutes.
4 System Info
1. System Info
Product Model , Serial and HOST ID
Software Version and Hardware Version
Calibration Date
Option Information
2. Firmware Update
Update rmware
40
5 Data and Time
The system time is displayed in “ymd”, “mdy”, “dmy” format in user interface.
6 Self Test
Screen Test Test whether the screen has a dot defect using ve colors: White, Red, Green, Blue and Black. Press any key to switch the screen color and exit the test. Keyboard Test Enter the keyboard test interface. Press the function keys at the front panel one by one and observe whether the corresponding key be checked. If not, an error may have occurred in that key. To exit the test, press “<-“ four times. LCD Test If the keys at the front panel are transparent, when the key is pressed , the corresponding backlight will turn on when testing it.
8.1 display
Control the screen display of the analyzer, such as setting the display grid brightness. 1Grid brightness Control the display grid brightness. Table 2- 36 Grid brightness
Parameter Explanation
Default 30%
Range 0 ~ 100%
Unit None
Knob Step 1%
Direction Key Step 1%
2Display Line Open or Close Display Line or move the location.
1.
8.2 le
1.
8.2.1 Browser
Browser type including “Dir” and “File”, Dir: when selected, use the knob or direction keys to browser directory in system File: when selected, use the knob or direction keys to switch among les or folders under the current directory.
8.2.2 Open/Load
Open the current folder or load current le
41
8.2.3 Type
user can dene the le type be browsed and saved, included “all type”, “STA”, “SET”, “TRC”, “COR”, “LIM”, “BMP”, CSV”.
8.2.4 Save
Save le in current directory, the le type is set in "Type"
8.2.5 Create Folder
Create a new folder in current directory.
8.2.6 Delete
Delete selected le or directory.
1. Rename
Rename the selected le or folder.
8.2.7 Operate
Open/Load: Open the selected folder or directory, Load the selected le Cut : Cut the Selected le or folder, and delete the primary one after paste Copy : Copy the Selected le or folder for paste. Paste : Paste the le cut or copied before into the current. Delete : Delete selected le or directory.
8.3 preset
Recall the preset setting and restore the analyzer to a specied status.
Press System ->Pwr On/Preset .Preset to select “Def”, “Last” or “User”.
Press Preset to load the factory settings listed in the following table (except items marked with “**”) or User-dened
settings.
Table 2-60 Factory Settings
8.4 couple
Set related parameters according to the coupling relationship. Auto all : Set Related parameters automatically according to the coupling relationship. RBW RBW have couple relationship with span. Please refer to the introduction of the "Resolution Bandwidth" VBW VBW have coupling relationship with VBW. Please refer to the introduction of "VBW". 3Attenuation Input attenuation have coupling relation ship with Ref Level, preamp. Please refer to introduction of the "amplitude". CF step CF step have coupling relationship with RBW at the zero span, When in none-zero mode, CF step have coupling relationship with Span. please refer to the introduction of the "Freq Step". Sweep time
42
Parameter Default
Frequency
Center Freq 1.6 GHz (or 1.5 GHz, 1.05 GHz, 900 MHz, 750 MHz, 500 MHz)
Start Freq 0 Hz
Stop Freq 3.2GHz (or 3.0 GHz, 2.1 GHz, 1.8 GHz, 1.5 GHz, 1.0 GHz)
Freq Step Auto, 320MHz (or 300MHz, 210MHz, 180MHz, 150MHz, 100MHz
Span
Span 3.2GHz (or 3.0 GHz, 2.1 GHz, 1.8 GHz, 1.5 GHz ,1.0 GHz)
Amplitude
Ref Level 0 dBm
Attenuator Auto, 20 dB
Preamp Off
Units dBm
Scale/Div 10 dB
Scale Type Log
Ref Offset 0 dBm
Corrections Off
BW
RBW Auto, 1MHz
VBW Auto, 1MHz
VBW/RBW 1
Avg Type Log Pwr
Sweep
Sweep Time Auto, 312.416 ms (or 312.416 ms, 216.288 ms, 192.256 ms, 168.224 ms, 120.160
ms)
Sweep Rule Speed
Sweep Continue
Sweep Mode Auto
Numbers 1
Trig
Trigger Type Free Run
Sweep time have coupling relationship with RBW, VBW and span. Please refer to the introduction of "Sweep Time".
8.5 help
After press “Help”, press any key to help information.
8.6 save-1
Quick saving a le.
43
Video Trigger 0 dBm
External Trigger Rasing
TG
TG Off
TG Level 0 dBm
TG Lvl Offset 0 dB
Normalize Off
Norm Ref Lvl 0 dB
Norm Ref Pos 100%
Ref Trace Blank
Trace
Select Trace A
Trace Type of
Clear Write
Trace A
Avg Times 100
Variable X A
Variable Y B
Constant 0 dB
Output Z C
Math Type Off
Detect
Select Trace A
Detect Type of
Pos Peak
Trace A
Limit
Limit1 Off, Limit
Upper, 0 dBm
Limit2 Off, Limit
Lower, -100 dBm
Test Stop
Fail to stop Off
Buzzer Off
X Axis Freq
Demod
Demod Mode Close
Earphone Off
Volume 6
Demod Time 5 ms
Marker
Select Marker 1
Select Trace A
Marker Type Normal
Delta Pair Delta
Relative To Off
Marker Table Off
Marker Fctn
44
Select Marker 1
Marker Fn Off
N dB BW -3 dB
Read Out Frequency
Peak
Cont Peak Off
Peak Table Off
Peak Threshold -140 dBm
Peak Excursion 10 dB
Peak Type Max
Mode
Mode Spec Analyzer
Measure
Meas Type Off
Measure Setup
Channel Power
Center Freq 1.6 GHz (or 1.5 GHz, 1.05 GHz, 900 MHz, 750 MHz, 500
MHz)
Integration BW 2 MHz
Span 3.2GHz (or 3.0 GHz 2.1 GHz 1.8 GHz 1.5 GHz 1.0 GHz
ACPR
Center Freq 1.6 GHz (or 1.5 GHz, 1.05 GHz, 900 MHz, 750 MHz, 500
MHz)
Main Channel 1 MHz
Adjacent Chn 1 MHz
Adj Chn Space 3 MHz
Occupied BW
Method dBc 26% 99%
T-Power
Center Freq 1.6 GHz (or 1.5 GHz, 1.05 GHz, 900 MHz, 750 MHz, 500
MHz)
Start Line 0 s
Stop Line 20 ms
System**
Language English
Power On Def
Preset Def
IP Cong Static
Auto Cal Close
Tine Date On
Set Format ymd
45
LIMITED THREE-YEAR WARRANTY
B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of three years from date of purchase. B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt. To help us better serve you, please complete the warranty registration for your new instrument via our website www.bk precision.com
Exclusions: This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized alterations or repairs. The warranty is void if the serial number is altered, defaced or removed.
B&K Precision Corp. shall not be liable for any consequential damages, including without limitation damages resulting from loss of use. Some states do not allow limitations of incidental or consequential damages. So the above limitation or exclusion may not apply to you. This warranty gives you specic rights and you may have other rights, which vary from state-to-state.
B&K Precision Corp.
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
www.bkprecision.com
714-921-9095
46
Service Information
Warranty Service: Please go to the support and service section on our website at bkprecision.com to obtain an RMA #. Return the product in the original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Non-Warranty Service: Please go to the support and service section on our website at bkprecision.com to obtain an RMA #. Return the product in the original packaging to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Customers not on an open account must include payment in the form of a money order or credit card. For the most current repair charges please refer to the service and support section on our website. Return all merchandise to B&K Precision Corp. with prepaid shipping. The at-rate repair charge for Non-Warranty Service does not include return shipping. Return shipping to locations in North America is included for Warranty Service. For overnight shipments and non-North American shipping fees please contact B&K Precision Corp.
B&K Precision Corp.
22820 Savi Ranch Parkway
Yorba Linda, CA 92887 http://bkprecision.com
714-921-9095 Include with the returned instrument your complete return shipping address, contact name, phone number and description of problem. Version – March 4, 2019
47
Loading...