BK Precision 1476 Service manual

@L476
10 MHz,
Dua
l-T
TRIGGERED
race
Osci
SWEEP
@"o#6FmN
WARNING
Normal electrical shock present. stop voltages or
AC RMS current under certain conditions. Your practices away significantly reduce the risk factor
precautions:
1.
Don't
necessary.
2. Use insulated work surface on which to are
3.
Use instrument that could
4.
Always equipment, where one side of the This transformer,
60 Hz in USA), depending upon which way only to test instruments or the equipment under
ground oscilloscopes and most
ground.
use of test equipment exposes
because
electrical shock
An
human heartbeats.
most
more easily
can
should be considered dangerous
that will
from
expose high voltage needlessly. Remove
an insulated floor
not
the time-proven "one hand in the
includes most recent television sets and audio
does
prevent
your
heart in
Turn
or wet. Where insulated floor surface is
damp
probe.
provide a good ground
use an
the
present
this
lead of some
testing
causing l0 milliamps
Higher
produce
contact with
case
equipment
off
material
particularly
Be
isolation transformer to
chassis
of
such equipment may be
a
dangerous
test
instruments
other test instruments with 3-wire
often be
must
vollages
a lethal current. However,
normal work
exposed high voltage, and
of accidental contact
you
if
while making
or a large, insulated
place equipment; and
careful
return
AC
shock
you
to a certain performed
of
current
pose
an even
and hazardous
habits should include all accepted
know and observe the
housings and
test
connections
pocket" technique while handling an
to avoid
path.
power
the
hazard if the
to a "hot"
contacting
power
is
line
connected directly
equipment. Without an isolation
floating
2-wire AC
test
may
chassis. The
amount of where exposed voltage pass
to
through the heart will
greater
voltage
since it
with a
floor
not
available, wear
transformerless
chassis
threat
as
as low
produce
can
that
will
high
voltage. You will
following
covers
in high voltage circuits.
to stand on, and an
mat
make certain
nearby
a
"hot
at line voltage
plug
power
is touched,
result from
power plugs
is inserted.
connecting
ground
danger
because
steer
such surfaces heavy
to the
from
such
volts
35
only when
metal
but damage
lead
DC
a lethal
current
safety
gloves.
object
chassis"
chassis.
(120
VAC,
Not
the
of
most
is an earth
is
test instruments or any equipment with a 3-wire AC
5.
On outlet. This ground.
possible,
If
6. its high voltage unexpected
7.
Also such as on-off switches, fuses, turned off.
8. Never work CPR
is
a safety feature to keep the housing
familiarize
points
remember that
alone.
(cardio-pulmonary
yourself
points.
in defective equipment.
AC
Someone
resuscitation)
with the equipment being
However, remember
present
line voltage
should be
is
power
transformers, etc.,
nearby to render aid
first aid is highly recommended.
power plug,
or other
that high
some
on
even
exposed
tested
and the location
voltage
power
input
when
if
necessary.
use only a 3-wire
elements at
may appear
circuit
the
equipment
earth
of
points
Training
at
is
in
INSTRUCTION
FOR
MANUAL
B
10
MHz,
DUAL."TRACE
K.PRECISION
&
MODEL
L476
TRIGGERED
OSCILLOSGOPE
SWEEP
@BMFffiN
6460 West
Chicago, Illinois
C.ortland
60635
Street
INTRODUCTION
FEATURES SPECIFICATIONS
TABLE
.
OF
CONTENTS
Page
4
OPERATOR'S
OPERATING
Initial Starting Procedure Single-Trace Calibrated Voltage Measurement Calibrated Time Measurement
External Z-Axis
Dual-Trace
DUAL-TRACE
Introduction FrequencyDividerWaveforms Dvide-by-8CircuitWaveforms DigitalCircuitTimeRelationships GatedRingingCircuit
DelaylineTests...'....
StereoAmplifierServicing
VideoEquipmentServicing
SINGLETRACE APPLICATIONS
Introduction
Video Equipment
Sigral Tracing Composite Sync VITS Vectorscope
CONTROIS, INDICATORS
INSTRUCTIONS
Waveform
Horizontal
lnput
Waveform
APPLICATIONS
Video
Pulse
Analysis
(Vertical
Operation
.
and
lnterval Test Sigral)
Observation
Input
Observation
Servicing Peak-to-Peak
Waveform Analysis
(X-Y
AND FACILITIES
Operation)
Voltage Readings
.
9
9 l0 ll t2 t2 t2
.
.. . .. 14
....l4
....14
.......14
.....l4
........16
.......16
.......17
l8 l8 l9 l9 l9
2l 23
TELEVISION ALIGNMENT
Introduction lmportance of Sweep Aligrrment
Sweep Aligrment Methods
Tuner Alignment
IF Alignment
Chroma Alignment
26
26 27 28 29 29
TABLE OF
CONTENTS
page
FMRECEIVERALIGNMENT..
PHASE
FREQUENCYMEASUREMENT
SQUARE
CIRCUIT
MEASUREMENT
WAVE Introduction Testinghocedure
AnalyzingtheWaveforms....
DESCRIPTION General VerticalPreamplifiers Mode
logic VerticalAmplifier TriggerCircuit SyncAmplilierandlnverter AutoSweep HorizontalAmplifier ChoppingOscillator I VoltCalSignal PowerSuppty BlockDagram
TESTING
OF AMPLIFIERS
...
....3I
....
...... 33
........34
....34
....38
.....38
.. .. . . 38
.......38
....38
...38
......40
.....4O
......4O
........40
.....4O
....50
3l
33
I',IAINTENANCE
WARRANTY
WARRANTY
OPTIONALACCESSORIES...
AND
CALIBRATION....
INSTRUCTIONS
........
........ 43
.....44
.....
40
43
The B & K-Precision loscope observing Dual vertical of waveforms simultaneous permits rate waveforrns.
is a laboratory-quality,
and measuring
inputs
waveforms.
two
are
chopped
viewing.
simultaneous
DUAL TRACE
FULLY
SOLID STATE
TRIGGERED
SWEEP
LARGE
SCREEN
CALIBRATED
VOLTAGE SCALES
CALIBRATED
SWEEP SPEED
WIDE BANDWIDTH
INTRODUCTION
Dual-Trace
Model
waveforms in electronic
provided
are
Low-frequency, low repetition rate
at a 200
Alternate sweep
viewing of high-speed, high repetition-
1476
professional
for
kHz rate to
instrument
simultaneous viewing
of
the two inputs
Oscil-
circuits.
provide
for
for
FEATURES
input waveforms can be
Two viewed
taneously, as
Only
filament. All transistors,
e ffect transistors) and
The 1476's
pres€ntations is beyond son oscilloscopes. at rest being they
Fully allows waveforms ing.
synchronized
trigger.
The 130 mm
diameter cathode easy-to-read
8 x l0 cm area.
Accurate measurement of the stantaneous
ferent attenuator Channel A and Channel B.
Accurate time
l8
DC to
nSEC rise time assure distortion-
free,
at
either
the cathode
with non-triggered
until
observed,
are
adjustable
the desired
Waveforms can
different ranges.
l0
high
high frequencies.
singly
desired.
ray tube uses
other
diodes, FET's
stability
sweeps
The
triggered by
to assure
always synchronized.
trigger
portign
to be used
to an
(approx.
ray tube
presentation
rectangular viewing
voltaees on I I
ranges for
measurements
MHz bandwidth
resolution
or
simul-
stages
of waveform
for trigger-
external sync
5.1 inches)
presentation
use
(field
IC's.
compari-
sweeP
remain
the
signal
that
threshold
of the
be
also
gives
on an
ln-
dif-
both
on
35
and
dual-trace
The width, wide range of sweep speeds, and provided, of electronic equipment, research and laboratory instruction.
make this the ideal oscilloscope
applications,
EXPANDED
SCALE
a
HIGH
SENSITIVITY
VIDEO
SYNC
VECTORSCOPE
CALIBRATION
SOURCE
Z-AXIS
INPUT
feature,
including troubleshooting and
together
A five time rtragrification the horizontal sweep allows close­up the magnification mum
gSEC/cm. Permits
impedance, l0: I
probes measurements, thus assuring less circuit
A
built-in sync separator circuit included television VIDEO
are automatically selected at sweep
times of 0.5 SEC/cm to 0.
mSEC/cm for viewing
frames. Horizontal sync
automatically times of 50 for viewing television lines.
The unit may be used as a vector-
scope to
exactly vision manufacturers.
A
built-in calibrated I volt t o -peak square wave checking and vertical tional
Intensity modulation cluded for time or frequency kers.
brightness increases in state, decreases in logic
with the l0
development, and
examination of waveform. In
provides
sweep speed
the low-capacitance, high-
to be used for
loading.
specifically
signals.
SYNC,
selected at sweep
pSEC/cm
provide
as speciiied by color tele-
recalibration of the
amplifiers
equipment.
Compatible
MHz
high
sensitivrt)
for
a broad
portion
a
addition,
attenuation virtually
for viewing
When
verticai
sync
I
to
a color display
without
capability
with
TTL logic;
logic high
low state.
band­range
repairing
(5X)
of of
the 5X
a maxi-
of
0.2
all
is
using
pulses
television
pulses
are
pSEC/cm
peak-
permits
addi-
in-
mar-
1
WIDE
RANGE OF
SWEEP SPEEDS
Sweep speed to 0.5 SEC/cm speed forms from DC
range of
necessary for
to l0
pSEC/cm
1
provides
viewing
MHz.
every
wave-
MECHANICAL
FEATURES
Carrying molded fingergrip also doubles as tilt
stand. Locking detent each
allows adjustment of Rear feet face-up wrap for bezel over face of dard oscilloscope camera for making
waveform
handle with comfortable
22-l/2"
support
position
and double as
power
storing
photographs.
viewing
oscilloscope in
CRT
angle.
Slotted
cord.
mounts stan-
cord
SPECIFICATIONS
VERTICAL AMPLIFIERS
Deflection
Calibration
Frequency
Risetime Overshoot
Ringing
Input Input Capacity Max.
Operating
Chop Frequency Channel Separation
Factor
Accuracy
Response
Resistance
Voltage
Input
Modes
0.01 calibrated
quence. ! DC: DC to l0 MHz
AC:2Hz to l0 35 nanoseconds. 3%
display. 3% or less
display.
I megohm, t 57o. 22
300 p'p.
Channel Channel B Dual-trace automatically
at slower; alternate trace cally selected times.
2OOkHz(t2Vo).
Better than 60 dB
(CH
A and CH
V/cm to
Variable
5%
on all ranges.
or less at I @ kHz
(t3
pF
(DC +
V
A only.
all sweep times of I mSicm
B)
V/cm in I I
20
ranges
at
pF).
only.
in l-2-5 se-
between
dB)
G3
(-3
MHz
100 kHz squarewave
AC
for
@ |
dB).
squarewave
peak)
or 600
chopped
automati-
all faster sweep
kHz.
ranges.
V
and
Video Sync
HORIZONTAL
B Input)
CH Deflection Frequency Input Resistance I Input Capacity
Maximum
Voltage
X-Y Operation With
Factor .01
Response
Input 300
Vertical and circuit composite chronized and LINE sync switched automatically SWEEP FRAME 0.5 SEC/cm mSEC/cm LINE (horizontal
AMPUFIER
DC
22
p-p.
CH B becomes the input tion tal
horizontal
provided
(horiz.)
:
(Horizontal
V/cm to 20 V/cm.
to I MHz ( - 3 dB).
megohm
pF
V
SWEEP
CH
(horizontal).
control
position
so that any
video waveform can be
expanded
and
TIME/CM switch.
(vertical
50
sync
(t3
(DC
position,
the B input
sync
pSEC/cm
pulses).
(nominal).
pF).
+
AC
TIMEICM
the CH A input
Y
input
becomes the horizon-
control.
sync s€parator
portion
syn-
viewing.
for
FRAME
to
lnput through
peak)
(vertical)
becomes
The CH B
(vert.)
to 0.1
pulses).
pSEC/cm
I
or 600
switch in
the
posi-
and
of
by
V
X
SWEEP CIRCUITS
Sweep System
Sweep Time
Sweep Time Accuracy
Magnification
Sweep
Horizontal Linearity
TRIGGERING
Source
Slope
Triggering
Range
(Common
to CH A and CH B)
Triggered and automatic. In matic mode, without
I calibrated quence.
+ Obtained by enlarging the
sweep mum
gSEC/cm.
3% or
INT sensitivity). MODE sourceofinternaltrigger; signal is the the the
source in the Positive
ly
AUTO. INT: 20 Hz
EXT:
input
sigral.
pSEC/cm
<q^
CH A and DUAL
CH B signal
variable
to 0.5
ranges,
Variable
5 times from
speed becomes
sweep
less
distortion.
and EXT
triggering
CH B mode.
and negative,
level
to l0 MHz deflection cathode ray DC to l0
is obtained
sweep
SEC/cm in 18
in l-2-5
between ranges.
center.
(l
V
switch selects
the CH A
source in
modes,
is the triggering
continuous-
control;
as
tube).
MHz.
pull
(min.
measured
auto-
se-
above
Maxi-
0.2
p-p
and
for
I cm
on
CALIBRATION
INTENSITY MODULATION
Voltage
Input Resistance
POWER
Input
PROBES
Model Attenuation Input
Connector Tip
REQUIREMENTS
No.
lmpedances
VOLTACE
I
frequency.
TTL logic-compatible, increases decreases
l0 kO
120 20 approved for oscilloscopes.)
Fig.
PR-31,
Combination
l0:l = l0 megohms, Direct
BNC
Spring-loaded,
p-p
V
square
brightness; brightness.
(nominal).
or 240 V AC,
(3-wire
watts.
51.
PR-35, PR-36,
=
I megohm,
(t
wave
high logic
low
tlo7o,
line
cord, CSA-
PR-37.
l0: I
and direct.
18
120
hook-on tip.
5%)
at
logic
5O/6O Hz,
pF.
pF.
line
-
-
See
Fig.
|. Front panel
controls
7 I
9
r0
12
and indicators.
OPERATOR'S
Cathode
l.
the Scale. The
2.
marks rneasurements.
Pilot Lamp.
3.
4.
SWEEP selector. Selects calibrated sweep
(microsecond per
steps. In internal sweep
provide
CAL
5.
peak-to-peak
cy. attenuators.
< >
6. position dual cation when
7
TRIGGERING LEVEL Control. Sync level adjustment determines starts;
(+)
switch selects automatic
(PULL
generated EXT TRIG Jack.
E.
signal.
SYNC
9.
following
SLOPE. waveforms except television composite
(+)
(-)
VIDEO. In television composite
10.
SOURCE sweep.
INT Sweep is triggered by
Ray Tube
waveforms
for voltage
Lights
TIME/CM Switch.
the CH B
horizontal sweep.
|.-P Terminal. Provides
This is
POSITION
equals most
sweep; the vertical sync selected and horizontal sync selected
(+) (-)
EXT Sweep is triggered by an external signal applied
used
of traces
trace mode).
when
pushed
points
(-)
equals
AUTO).
even without an input signal.
Switch. Four-position
positions.
The SLOPE
Sweep is triggered waveform.
Sweep waveform.
lor
for sweep Sweep is triggered on Sweep is triggered
Switch. Selects
switch is in CH A or Sweep is triggered by Channel
MODE
at the
(CRT).
viewed.
are
8 x
l0 cm
when oscilloscope is tumed on.
centimeter)
generator
wave input
square
for
Control. Rotatitin adiusts horizontal
(both
Push-pull
pulled
in.
on
most
positive point
When automatic
Input
is triggered on
VIDEO
the
sweep times
times
switch is in CH B
EXT TRIG
This is
graticule
(vertical)
position,
calibration
out
waveform
negative
positions
video signal are used to trigger the
pulses
and time
Horizontal
this switch disables
permits
and
signal at the
of the vertical amplifier
traces
switch
(PULL
triggering
terminals for external trigger
positive-going
on
negative-going slope of
positions,
(frame)
pulses
of
0.5 SEC/cm (line)
pSEC/cm
of 50
positive-going
on negative-going
triggering source
CH A sigrral
DUAL
jack
8.
CONTROLS.
the
screen
provides
times of I
to
0.5 SEC/cm
when
selects 5X magrifi-
5X MAG): normal
slope
point
of triggering.
triggering,
lever switch
used
are
the sync
position.
position.
calibration
(horizontal)
sweep time
pSEC/cm
the CH B input to
calibrated
line frequen-
operated in the
where
of triggering
when pulled
a sweep
for
viewing all
video sigtals.
pulses
are automatically
to 0. I mSEC/cm,
are automatically
prSEC/cm.
to 1
sync
sync
when MODE
B
sigral
which
on
in
l8
the
I volt
sweep
and
Push-pull
out
is
with the
of
slope
of
pulse. pulse.
for the
when
INDICATORS
Channel
ll.
ment adjustment CH
Channel B
12.
adjustment
13.
Channel B. Jack SWEEP
Ctrannel B DC-GNDAC Switch.
14.
DC Direct input of AC and DC
GND Opens sigral
AC
Channel B.
15.
Channel B. from horizontal sensitivity switch
MODE
16.
basic
CH A Only
DUAL Dual-trace operation; both
CH B Only the input signal to Channel B is
17. Channel Channel A. from
18. Channel A DC-GNDAC
DC Direct input of AC and DC component of input
a
GND Opens sigrral
AC Blocks DC
19. Channel A A.
l). Channel
adjustment for
21. Channel A POSITION justment
FOCUS
22.
INTENSITY
23.
brightness
FOWER
24.
FACILITIES
AND
POSITION Control.
B
for Channel position.
B
B INPUT
becomes external
TIME/CM switch 4 is
signal.
amplifier. the when
Blocks
to
.01
4 is in the CH B
Switch. Three-position lever switch;
operating modes of
played
Channel B input signals are separate
played
A VOLTS/CM
to 20 volts
.01
sigral.
amplifier. This
the
when
A DC BAL
for
Control.
Switch. Turns
B trace. Becomes
when SWEEP
DC BAL Adjustment. for
Channel B
Jack. Vertical input
path
provides a zero-signal
This
position
VOLTS/CM Vertical
Vertical
position
INPUT Jack. Vertical inputjack ofChannel
of trace.
ofwhich can
performing
DC component
sensitivity
20 volts
position.
the
input signal
as a single
traces.
as a single trace.
sensitivity per
path
provides
which
of
performing
component of input signal.
Adjustment. Vertical
Channel A trace.
Channel A trace.
Control.
Vertical
TIME/CM switch
Vertical
trace.
horizontal
in the CH B
component
grounds
and
DC measurements.
Switch.
per
when the
the
Switch.
cm. Switch.
and
DC
Control.
Clockwise
unit on and
be used as a
of input
Vertical
is calibrated
This
cm.
oscilloscope.
to Channel A is dis-
trace.
Vertical
is
calibrated
grounds
a zero-sigral base line,
can be
measurements.
Vertical position
rotation increases
position
horizontal
jack
input to vertical
signal.
control
SWEEP
the Channel A and
displayed on two
input to vertical
used
off.
adjust-
position
in the
4 is
Balance
DC
of Channel
input when
position.
of input
line,
base
reference
attenuator
in I I steps
attenuator
in I I steps,
as a reference
DC balance
for
adjusts
TIME/CM
selects the
dis-
for
ad-
the
VARIABLE
25.
the extreme is calibrated.
VARIABLE
CH B
26.
adiustment. In-the extreme attenrator
CH A VARIABLE
n.
adjustment. In-the extreme attenuator
Control.
Fine sweep
clockwise
Control.
Fine control
clockwise
is calibrated.
Control.
control
Fine
clockwise
is calibrated.
of
(CAL),
B
CH
B vertical
CH
position
(CAL) position,
A
CH
A
of CH
(CAL)
time adjustment.
the sweep
vertical
attenuator
In
time
sensitivity'
vertical
the
vertical
attenuator
vertical sensitivity-
position, the vertical
29.
Line Cord
AC loscopes.
(See
Probe
30.
combination for
use designed input impedance capable
Vector Overlay
31. for
of operation
vectorscope
(See
Fig. 3).
l0: l/Direct
with this
for use with
of
(Not
operation.
Fig. 2). CSA-approved
The B & K-Precision
probes
oscilloscope.
an oscilloscope
I megohm
up to 10
Shorvn).
have been
However,
having a nominal
shunted by
MHz, can be
Interchanges with
for oscil-
ModelPR-31
designed
probe
any
pF
35
and
used.
scale
INT MOD
28.
Jack. Intensity
modulation
DC BALANCE CONTROL
(One
on each
1.
side
PULL
)
APART
(Z'axis) input.
Fig.
2. Rear
CRT
ADJUSTMENT
COMBINATION AND
CORD
and side
panel
facilities.
ROTATION
FEET
WRAP
2829
t-r,_tfj1-
--
2. ROTATE 180"
PUSH
3.
1O:1
ATTENUATION
BACK
TOGETHER
Fig.
3.
PROEE
hobe
P ROBE
COMPENSATION
(PR-31
)
probe
Adlust
square wave display while observing
from
signal
COMPENSATION ADJUSTMENT
ADJUSTMENT
compensation
cal J1 1V
details.
p-p
for
ACK.
(PR-35)
best
OPERATING
INSTRUCTIONS
INITIAL STARTING PROCEDURE
i Set PO\l ER switch 24 to OFF
l. Connect
3. Set control their ranges.
PUII
4. position.
5. Set DC-GND-AC
6. Set MODE single-trace operation trace operation.
7. Tum on oscilloscope ON
8. Wait
warm mode)
9. If INTENSITY observed.
10. Adjust FOCUS for
I 1. Readjust
center the
12. Check for
and 20 CALIBRATION ments
The oscilloscope
measurements.
power
cord 29 to a I
CH A
position.
no trace appears, increase
the
POSITION
I I
and<>POSITION
TRIGGERING
CH A DC-GND-AC
switch switch 16
Pilot lamp 3 will light.
few
a
seconds for
up. A trace
should
thinnest,
position
traces.
as described in the MAINTENANCE
require
(two
appear on the face of
control 23
control 22and INTENSITY
sharpest trace.
controls 6,21 andl I if necessary,
proper
portion
checking
is now ready
LEVEL
14
or the DUAL
by turning
adjustment of
l7-volt,
control
control 6
control 7 to
switch 18
to the
GND
to the CH A position
the cathode ray
traces
if operating
(clockwise)
until the trace
of this
periodically.
only
position.
50160 Hz outlet.
21,
CH B
to the centers of
and CH B
positions.
position
POWER
DC
manual. These
for making
switch
tube
in the DUAL
the CRT.
the
BAL controls 12
POSITION
the AUTO
for
for dual-
to
24
(CRT)
to
setting of
is easilv
control23
to
AND
adjust-
waveform
the l0:l capacitance impedance when
J.
SetCH only the AC most measurements being measured the and the DC reference, frequency The GND ground readings.
4.
Connect equipment the measured.
a. If
b. The
5.
Set CH A VOLTS/CM
gives vertical deflection. probably concemed speed,
desired duces
jicient
position
in the DIRect (low-capacity
possible,
A
DC
point
the less AC former to prevent
measurement when
2 to
which presents
at least I
tigger signal
to decrease circuit loading.
DC-GND-AC
component
position
reference
ground
number
includes a large DC
for measuring
waveform
position
clip
under
in the
peak-to-peak
using the DIRect
be
test. Connect
circuit
equipment
powered
(two
6
cm
unsynchronized.
with
adjusting synchronization of waveforms.
cm vertical
and I megohm with
is required only when
is required, such
WARNING
should not exceed
to synchronize
position.
position)
switch
and
(below
of
under test is a transformer-
item, use
dangerous electrical
to six large
The
l8 to
(this
the
is
must
be used if the
both the
any
and
Hz)
5
probe to
where the waveform
voltage at the
switch 17
display on the
a
chassis
the
an
position
The remaining
stable display
Any signal
deflection
l20pF shunt
The higher input
should
AC for measuring
normal
component). Use
AC component
time a very low
is
to be observed.
as for DC voltage
ground
tip of the
isolation
point
600 volts
of the
position
to
a
squares on the scale)
develops suf-
the
sweep.
be used
position
a zero-signal
shock.
probe.
screen will
and
showing
point
of the
probe
is
to be
trans-
of
that
steps are
sweep
pro-
that
for
to
the
Never
allow remain few
seconds. burned. Reduce motion
SINGI,E-TRACE
Either
single-trace
following
the
Perform
l.
with
the connect the The following
B & K-Precision Model PR-3 For
2.
all except low-amplitude
are set for
(below
forms DIRect.
DIRect, or
to input
impedance
a small spot of
stationary on the
The screen mav
by
Channel A or Channel
operation. For simplicity,
instructions.
the
MODE
See Fig. 3
intensity
causing it to
WAYEFORM
steps
switch
probe
instructions assume
l0: I attenuation. For low-amplitude
volt
0.5 vice
with
CAUTION
high brilliance
screen for
become
-or
sweep.
OBSERVATION
of
the "Initial Starting Procedure"
16 in
the
to the CH A
cable
i combination
peak-to-peak),
for
changing
versa.
only l8
The
probe
more permanentlv
keep
ihe
spot
B
can be used for
Channel A
CH A position.
INPUT
the
waveforms,
set the
the probes
has
shunt
capacitance
a l0 megohm
pF
than a
is used in
jack
use of the
probes.
the
probe
from
to
iir
Then
19.
probes
wave-
for
l0:l
in
6.
Set SOURCE provides observed
waveforms
external
an
switch
cable should to the external
Set SYNC switch 9 to the
for observing to all
the
the negative-going
unknown, the
8.
Readjust TRIGGERING
synchronized
the
that will in the
disappear sweep,
amplitude waveforms.
should
the SLOPE
other
sweep is to
(-)
control may
such
switch l0 to
internal
is
should
types
position
produce
center
if there
sync
also used to
be viewed
sync
placed
be
be connected from
sync source.
television
(+)
or SLOPE
of waveforms. Use
be triggered
if the
wave.
SLOPE
display
be
a sweep,
portion
when
as
INT
the
that the waveform
so
trigger the
source is required,
without
pushed
is inadequate
using internal
in the EXT position
the EXT
VIDEO(+)
composite
by a
sweep is to
If the type
(+)
position
LEVEL control
which
of
measuring
If no sweep
video
(-) positions
the
positive-going
be triggered
may be used.
jitter.
in and rotated
its
As a
is usually
range.
signal
DC or
or
of waveform
can be obtained,
position.
sweep. Most
sync. When
the
SOURCE
and
jack
TRIG
(-)
positions
waveforms
for
observing
(+)
position_if
wave,
7 to obtain
starting
The trace
to trigger
extremely
point,
to
any
somewhere
This
being
or
or
by
is
point
will
the
low
a
8
a
a
pull
the
control
out
(PULL
AUTO) for
triggering. Set SWEEP
9. of waveforms. This
portion
a progressively
is
This
because the
repetition rate
TIME/CM switch 4
control may be set for viewing
of a waveform,
dimmer as a smaller
sweep speed
does not change.
for
the desired
the
but
portion
increases
trace becomes
but the sweep
NOTE
When
using very
rates, the operator
fast
sweep speed
may
low repetition
at
wish to operate
intensity control toward maximum. Under
conditions, a
extreme left of the
retrace "pip"
This does
trace.
may appear at the
not in any way
affect the oscilloscope operation and may be
disregarded.
10. After
obtaining
9,
step
it is sometimes desirable to make a final
adjustment
(-)
The waveform (+)
direction
at
direction selects
the desired number of waveforms,
of
the TRIGGERING LEVEL
which
the most negative
selects
sweep triggering will occur
positive point
most
the
automatic
number
is
displayed.
with the
these
control 7.
point
and the
only
as in
on the
on the
waveform control desired
I l. For
outward
a close-up view of a portion
which
at
may
be adjusted to start the
portion
of the waveform.
on the<)POSITION the sweep by a factor displays only
portion
a control clockwise, center, turn the
on the
control to return the
non-magnified
CALIBRATED
Peak voltages, peak-to-peak
-
voltages easily Dual-Trace
of a specific portion
and
accurately
Triggered
Adjust
l.
controls
waveform to
2. Set
CH A
vertical
deflection
center
the
the
to
left
and to
control
condition.
VOLTAGE
Sweep
previously
as
be measured.
VOLTS/CM
limits of the vertical
sweep triggering
of the
will occur.
sweep on
waveform,
The
any
pull
control 6. This expands
(5X
of five
partion
magrification)
of
the
sweep. To view
and
of center, turn the<>POSITION
portion
view
counterclockwise.
sweep to
MEASLJR-EMENT (See
voltages,
of the
measured
complex
on the
to
the right of
Push
inward
the normal,
Fig. a)
DC voltages
waveform Model
1476
and
Oscilloscope.
instructed
switch 17 for
possible
without exceeding
to
display the
the maximum
scale.
are
the
POSITI
ON CONTROL ADJUSTE SO THAT FORM A HORIZONTAL
BOTTOM OF
ALIGNS
EXACTLY
REFERENCE
EXAMPLE:
VERTICAL
PROBE
PEAK-TO-PEAK
.} POSITION
TOP OF WAVEFORM
VERTICALSCALE MARKER AND
WAVE-
DEF
VOLTAGE/CM
ATTENUATION
WAVEFORM
CONTROL ADJUSTED
EASE
WITH
LECTI ON
.O2Y x
D
LINE
4.2
READING
OF
4.2cm .O2
.084V
0.84V
CROSSES
v
10
SO THAT
CENTER OF
FOR ACCURACY
qTo1"t
t
VOLTS/cm
SET
.o2v
PROBE
TO
10:1
ATTENUATION
l0
Fig.
4.
Typical
voltage measurement.
vertical
Read the amount
3. the s;ale. The rea.ltusted :erJrng ::.r' reierence G\D deflected The and
4. Calculate
ri desired.
CH A
with the
position,
when the switch
trace deflects upward
downward
the voltage
vertical deflection
setting
(see
l7
ro
of
CH A
the reference
shift
When measuring a DC voltage, adjust
POSITION
CH A DC-GND-AC
then note
for a negative voltage input.
(in
example
voltage reading displayed
l/l0th the actual
probe
is
set
displayed
when
voltage being measured
for l0: I attenuation.
probe
the
ment,
5. Calibration
accuracy occasionally checked by peak
square
jack
5. This calibrated peak-to-peak.
volt dicated, see the
wave
signal
If a
"MAINTENANCE AND
TION" section of the manual
deflection
POSITION control
point
for easier scale
control 2l
to
switch
the amount the
placed
is
for a
in the
positive
reading as follows:
VOLTS/CM control
by the
cm)
in Fig.4) Don't
on the
forget that the
oscilloscope is only
(in
a convenient
DC
voltage
Multiply the
from
cm)
2l may be
l8
in the
trace is
position.
input
when the
voltage is
measure-
is set
The actual
for
DIRect
of this oscilloscope may be
observing
available at
the I volt
the CAL lV
peak-to-
P-P
source should read exactly I
need for recalibration
is in-
CALIBRA-
for
complete
procedures.
MEASUREMENT
CALIBRATED
Pulse width,
waveform time
other measured on ments
from possible. visible at one time. moving the beginning and
TIME
waveform
this oscilloscope.
second
.5
At low sweep speeds,
However,
from left to right
ending
to spot.
Adjust
l.
controls
display of
2. Set the SWEEP possible
display
as
the desired
TIME/CM control
of the
measured, usually one
3. If necessary, readjust trol 7 for the most
Read the
4. between the control 6 measurement
amount of
points
may
be readjusted
points
easier reading.
5. Calculate the
time duration
(See
Fig. 5)
periods,
durations
down
across
points
previously described
circuit delays easily and accurately
are Calibrated to 0.1
entire
the
the bright
tinp rneasure'
microsecond are
waveform is not
spot
the screen, which
the measurement easy
of
can
for
waveform.
4 for
the
waveform
segment
cycle. the TRIGGERING
display.
stable
horizontal
of measurement.
to aligrr
LEVEL con-
deflection
ThecPOSITION
one of the
with a vertical scale marker
follows: Multiply
as
and all
seen
be
makes
stable
a
largest
to be
(in
cm)
for
the
<)
POSITION THAT
FORM REFERENCE LINE. EDGE MAY NOT BE VISIBLE ON FAST PULSES; IN ALIGN BEGINS.
POSITION
THATTRAILING EDGE
CROSSES HORIZONTAL FOR ACCURACY
EXAMPLE: HORIZONTAL
c.
CONTROL
LEADING EDGE OF WAVE-
ALIGNS
WHEREVER WAVEFORM
CONTROL
SWEEP
TIME
PEBIOD)
OF WAVEFORM
ADJUSTED
WITH l. r ERTICAL
THIS CASE
ADJUSTED
AND EASE
DEFLECTION
TIME/CM
DURATION
SO
VERY
OF WAVEFORM
SCALE
SO
MARKER
READING
OF
=
6.35cm
1 OpSEC
­=
63.SpSEC
HORIZONTAL
DEF LECTI ON
O.JJ
cm
f
ol-ol"t
SWEEP TIME/CM
'l0l,l
set to
SLOPE
tto start syveep
negativejoing
INT
sec
set to
edgel
Q on
FREouENCY -
=:-
IIME
=
.0000635 sEc
15,750 Hz
Fig.
5.
Typical time
DISPLAYS RECEIVER HORIZONTAL
measurement.
SHOW TYPICAL TELEVISION
WAVEFORM AT GRID
OUTPUT TUBE
OF
ll
horizontal
deflection switch 4 setting when the divided
5X magnification is used, the result must be
by 5 to obtain the actual time duration.
6. Time measurements especially started waveform
true when measuring delays. The
by a
measured in a subsequent measurement and the
subsequent
(in
(see
example in Fig. 5).
often
sync sigral
of the
display
waveform.
cm) by the SWEEP TIME/CM
require external
from
one
circuit. This allows
between the
perform
To
urements using external sync, use the following a. Set the b. Connect a
source
c. Set the SYNC
position
d. Readjust
necessary for
e.
If measuring start of the
7. Another excellent method for with
dual-trace operation. The
SOURCE switch l0 to the EXT
from the EXT TRIG
cable
of sync sigral. Use a
switch
for
proper polarity
the
short shielded cable.
9 to the
SLOPE
for the
the TRIGGERING LEVEL
a stable
a delay,
sweep
waveform.
measure
to the
the
start of the waveform.
measuring
procedures the "DUAL-TRACE APPLICATIONS" manual,
EXTERNAL HORTZONTAL
For some measurements,
tion sigral is required. measurement, tion
and the X input horizontal phase
measurement,
where the Y
input may be a sinusoidal wave,
input must be 10 mV
voltage of
any satisfactory operation. To
the following
use
This is also referred to
provides
or an external sweep
per
cm
mV or
100
use an external horizontal
procedure:
1. Set the SWEEP TIME/CM the CH B
2. Use Channel B
3. Adjust the CH B VOLTS/CM
position.
the Channel A
probe
for
amount of horizontal
probe
the horizontal input.
control 15.
4. The CH B POSITION
horizontal
control
position
is disabled.
control, and the
input
control I I
INPUT
extemal
an
provides
horizontal
of
deflection or greater
switch
for
the vertical
(X-y
OPERATTON)
horizontal
is sufficient for
4 fully
deflection with the
now serves
NOTE
Do NOT
X-Y operation.
to adjust
5. All
Z.NilS
.The
modulated rygks 99
INT MOD oscilloscope will govide of the trace.
horizontal
sync
controls
INPUT
trace
(Z-aas
required.
(intensity
See Fig.
use
the
Use
displayed
input) where A TTL
modulation)
6.
PULL
5X MAG
the
CH B VOLTS/CM
gain.
are disconnected
on the
screen
frequency-
compatible
allernate
sigrral
j
ack 24 on
brightness
control durine
and have
may
t2
Remember,
sync. This
sweep is
circuit and the
pulse
sync such meas-
steps:
position.
jack
8 to the
(+)
(-)
or
sync signal.
control 7 if
from
time
the
time delays is
given
are
in
section of the
deflec-
as an X-Y
vertical
deflection.
such
deflec-
The
for
as
voltage. This
greater;
thus
input,
clockwise
to
input and the
the
as
POSITION
<>
controT
no effect.
be intensity
or
time-scall
applied
th-e-rear
and
at the
of
the
blanking
is
f
\
l
t
Fig.
6.
Oscilloscope
DUALTRACE (Refer
to Fig.
In observing and B, frequency the
other although An example mulliplier. on Channel of
this reference
In
this way, synchronized, with
the Channel
phase
or frequency simultaneously, both
waveforms
To display
tion,
use the following procedure:
Perform
l.
2.
Connect and CH B IMUT
If
the recommended
3.
WAVEFORM
7) simultaneous
it is necessary
or
that one
the basic frequencies
of this
The
A, for
is in
reference,
example,
frequency
when the
thq
display
A
display.
relationship
it will
in
sync for
two waveforms
the
steps of
oscilloscope probe
jacks
oscilloscope probes
should peak-to-peak forms for
be used
except for
or
less. For
the DIRect position
changing the
probe Whenever possible, capacity l0: I
4. Set MODE traces should
5.
Adjust
place
to trace,
and
mark
on the
6.
Set both and 14 to most
measurements
being measured
position
switch 16
appear on
CH
A
and CH B POSITION
the
Channel
adjust
both
scale.
the
CH A
the
position.
AC
include
't
t
I
t
\
trace with Z-axis
OBSERVATION
waveforms
that
the waveforms
of the
waveforms
checking a frequency
or
"clock" frequency
and the
will
be displayed
waveform
on Channel
If two waveforms
to
be difficult if
any useful
simultaneously
the "Initial
19
and 13.
B
& K-Precision
are used,
the lower
should be
from
use the
to minimize
to the DUAL
the
screen
A
trace
traces to
and CH B
This is the
and must
a large
t
\
I
t
t
input.
on
be
be
synchronized
may
be different.
can be used
multiple
or
submultiple
on
display
each
of Channel
B will
other
also be in
are displayed
not impossible
observation.
for
Starting Procedure."
cables
to both
Model PR-31
10 :
waveforms
I attenuation
of
amplitude
used.
l0:l
to DIR
high
or vice
impedance,
circuit
loading.
position.
controls above the a convenient
DC-GND-AC
switches
position
be used
DC
if
component.
channels
related
divider
ChanneiB.
A
sync
having
no
tb lirck
observa-
the
CH A
0.5 volt wave-
See Fig.
veisa.
low
Two
2l and
ll
Channel
reference
l8
for
used
points
the
A
in
to or
is
3
B
gound
Connect g:ound
-.i *aveiorms
*aveform will
the
the equipment
of
probes
the
are to be measured. The sipal to which the
be synchronized
Channel A input
to
for
of the
clips
under
points
in the
intemal sync operation.
WARNING
the equipment under test is a transformer-
a. If
less AC unit. use an isolation transformer to
prevent
b. The
dangerous
peak-to-peak
measurement should not
probe
the
R
Set the VOLTS/CM controls
B to
and
deflection.
is
position
a
The displays on the screen will
unsynchronized. The
those outlined for single-trace operation, describe
to
procedure
the
electrical shock.
voltage at the
in the DIR
used
that
remaining steps,
for
obtaining
exceed
l5
gives
displays.
Set the SOURCE switch
9. provides
internal sync so that
l0 to
the Channel A waveform being observed is also used to in dual-trace operation, a sync source
point
measurement
set the SOURCE switch to the EXT
case position
to
8
the sync source.
the SYNC switch
Set
10.
positions waveforms, or positions
and connect a cable
for observing
for observing
for Channel A is required. In this
from the EXT TRIG
9 to the VIDEO(+) or
te.-vrsion composite video
to the SLOPE
other types
all
probes
to the
test.
Connect
circuit
must
be applied to
600
the tips
where the
point
volts.
chassis
of
position.
17 for
and
Channels A
2 to 3 cm vertical
probably
although similar
stable,
the
synchronized
position.
INT
trigger the sweep.
other
than the
(external)
(+)
or SLOPE
waveforms.
of
the
if
be
This
Often
jack
(-)
(-)
(+)
Use
the
positive-going
a sweep
I l.
Adjust stable, synchronized sweep. control
produce
will the
center
if there is inadequate signal
when measuring
as
no sweep can be obtained,
positions
if the sweep is
wave, or to the
is
triggered by a negative-going
to be
TRIGGERING LEVEL may be
pushed
a sweep,
portion
in and
which is usually somewhere
of its range. The
extremely low amplitude signals. If
(-)
control 7
As a
rotated to
trace will disappear
to
trigger
pull
out the control
AUTO) for automatic triggering.
t2.
Set SWEEP of waveforms. This
portion
a progessively
13.
After obtaining the desired number of
TIME/CM
switch
control
of a waveform, but the trace becomes
dimnrer
a smaller
as
step 12, it is sometimes
for the
4
may be set
portion
desirable to make a final
adjustment of the TRIGGERING
(-)
The
point will occur
positive point triggering will occur. The start the sweep
direction of rotation selects
on the sync waveform at
and
on
the
the sync
direction selects the most
waveform
(+)
control
on
any desired
which
may be
portion
waveform.
The obsewed waveforms of
14.
expanded
be
on the <>POSITION
rotated
be left
desired.
right extremes of
and
Push
by a factor of 5 by
control
clockwise or counterclockwise
inward on the control to retum the
Channels A and B can
This control can then
6. waveform
the
sweep to the normal, non-magrrified condition. Calibrated voltage measurements,
15.
calibrated
urements and operation with Z-axis
to be triggered
position
if the
wave.
to obtain a
any
the
point,
point
sweep,
that
such
starting
(PULL
desired number
for
viewing
only
is displayed.
waveforms
LEVEL
the
most
sweep
which
at
as
control 7.
negative
triggering
sweep
adjusted
of the sync
pulling
outward
to
view the
displays as
time meas-
input are identical
by
the
in
in
to
""o
A. REFERENCE FREOUENCY
a",
B.
PULSES PER
.(1OOO
LEADING VISIBLE
DIVIDE LEADING
.BY .TWO
SECOND)
EDGES MAY NOT BE
AT FAST SWEEP
EDGE
OUTPUT
REFERENCE PULSE
OF
I l*ra
."t
l
DIVIDE
C.
TRAILING
.BY .TWO
EDGE
OF
OUTPUT SYNCHRONIZED
REFERENCE
CHANNEL A WAVEFORM
filii,t?3Hfi
PULSE
TRAIN
RATES
SYNCHRONIZED
PULSE
Fig. 7. Waveforms
TO
TO
WAVEFORM
A
HEIGHT
in
divide-by-two
@
@
o o
WAVEFORM
B
HEIGHT
circuit.
l mS/cm
SLOPE+
INT
WAVEFORM
B
WAVEFORM
A
l3
Loading...
+ 32 hidden pages