BJS Biotechnologies XXPRESS NGx User Manual

The World’s Fastest Real-Time PCR
Thermal Cycler
User Manual
This Manual Copyright © 2018 BJS Biotechnologies Ltd.
Page | 2
XXPRESS® NGX USER MANUAL – XP-5001-01-5
xxpress, xxplates, xxsealer, and their respective logos are registered trademarks of BJS Biotechnologies Ltd, or one of its affiliated companies in the UK and other countries. All rights reserved.
No part of this manual may be reproduced, stored in a retrieval system, or transmitted, by any means or in any form, without the prior permission of BJS Biotechnologies Ltd (BJS Bio).
The information contained in this manual is subject to change without notice. BJS Bio assumes no responsibility for any errors that may appear in this or related documentation.
Document: XP-5001-01-5
xxpress, xxplates, and xxsealer are the brand names for the BJS Bio. range of PCR thermal cycler products and accessories. For further details please visit our web site: www.xxpressPCR.com.
All sales and technical enquiries should be addressed to: BJS Biotechnologies Ltd
65 Bideford Avenue Perivale Greenford Middlesex UB6 7PP United Kingdom
+44 (0)203 021 3750
sales@xxpressPCR.com support@xxpressPCR.com
Page | 3
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Contents
1 Glossary of Terms and Abbreviations ..................................................... 4
2 Symbols Used in this Instruction Manual ................................................ 5
3 Unit Description ....................................................................................... 6
4 Technical Specification ............................................................................ 9
5 Safety Precautions and Limitations of Use ............................................ 11
6 Unit Installation ...................................................................................... 15
7 Unit Operation ....................................................................................... 18
7.1 User Touch Screen Display and Controls ............................................. 18
7.2 Operation of the xxpress®NGx qPCR System ...................................... 18
7.2.1 Turning on xxpress®NGx........................................................... 18
7.2.2 Designing a New PCR Experiment ............................................ 20
7.2.3 Designing a New Melt-Curve Experiment .................................. 37
7.2.4 Running the Experiment ............................................................ 41
7.2.5 Experiment Results and Analysis .............................................. 43
7.2.6 Repeating Experiment and Reviewing Previous Results ........... 47
8 Maintenance and Servicing ................................................................... 50
8.1 Replacing the Unit Fuse ........................................................................ 50
8.2 Routine Cleaning and Inspection .......................................................... 51
8.3 Decontamination Procedure .................................................................. 52
8.4 Transportation and Storage ................................................................... 52
9 Warranty and Returns ........................................................................... 54
10 Product Disposal ................................................................................... 55
Tables
Table 1: Advisory Symbol Meanings ................................................................. 5
Table 2: Unit Features ....................................................................................... 7
Table 3: xxpress® NGxSample Plates and Accessories .................................. 8
Page | 4
XXPRESS® NGX USER MANUAL – XP-5001-01-5
1 Glossary of Terms and Abbreviations
ANSI
American National Standards Institute
BJS Bio
BJS Biotechnologies Limited, UK Company Number 02959160
Deepwell Plate
Plate with an SBS footprint featuring 48, 96 or 384 wells with a larger volume than microplates
DWP
Deepwell plate
EMC
Electro-Magnetic Compatibility
Microtitre Plate
Plate with an SBS footprint featuring 24, 48, 96 or 384 wells
MTP
Microtiter plate
PCR
Polymerase Chain Reaction
PPE
Personal Protective Equipment
Real-Time PCR
Polymerase Chain Reaction where the result is delivered throughout the PCR process by the capturing of fluorescence emitted by the reaction.
SBS
Society for Biomolecular Screening
Thermal Cycler
A machine that moves samples from one precise temperature to another repeatedly
Well
A single sample cavity in a Microtiter plate, PCR plate or Deepwell plate
xxplate™
BJS Bio PCR test plate for the xxpress® NGxreal-time thermal cycler. Available in 24, 54,
and 96 well formats.
xxsealer SA
A semi-automatic heat sealer used to weld a
sealing film onto the top of xxplates™ and
prevent sample evaporation during the PCR.
Page | 5
XXPRESS® NGX USER MANUAL – XP-5001-01-5
2 Symbols Used in this Instruction Manual
The following advisory symbols are used in this manual.
Table 1: Advisory Symbol Meanings
DANGER
Indicates a Risk of Electric Shock which could, if not avoided, result in serve injury or death.
DANGER
Indicates a Burn Hazard which could, if not avoided, result in serve injury or death.
DANGER
Indicates a Risk of Explosion which could, if not avoided, result in serve injury or death.
WARNING
Indicates a hazardous situation which could, if not avoided, result in serve injury or death; or severely damage the unit.
CAUTION
Indicates a hazardous situation which could, if not avoided, result in minor or moderate injury; or degrade or impair the functionality of the unit.
CAUTION
Indicates a Risk of Crush hazard due to moving parts which could, if not avoided, result in minor or moderate injury.
Advisory note or other useful information.
NN
Refer to “Section NN” for more details.
Page | 6
XXPRESS® NGX USER MANUAL – XP-5001-01-5
3 Unit Description
The xxpress®NGx Real-Time Thermal Cycler provides a safe and controlled method of heating and cooling samples placed within xxplates, whilst measuring the level of fluorescence emitted by the samples.
xxpress®NGx has five fluorescence measurement channels that enable fluorophores with a wide range of excitation and collection frequencies to be detected.
To protect the samples from evaporation and contamination during the heating and cooling of a PCR process, the xxplates™ are sealed using a clear film.
The xxpress®NGx unit has the following external features:
5 3 4 2 1
6
8
7
9
Page | 7
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Table 2: Unit Features
1
xxpress®NGx Power On/Off Button
2 Front Door
3
Front Door Release Button
4 xxplate™ Location
5
xxpress®NGx Display and Touch Screen
6 Front USB Connector
7 Rear Cover
8
Air Intake Vents and Filters (under unit)
9
Power on LED
10
Emergency Release Button (ECR)
11
Battery Compartment for Emergency Release Battery – (PP9)
12
DVI-D Connector for External Monitor
13
RJ45 Connector (Network)
14
Rear USB Connectors
15
Rear Label
16
Unit Air Expelling Fan
17
Mains Input Connector Fuse and Switch
12
14
13
10
11
17
15
16
Page | 8
XXPRESS® NGX USER MANUAL – XP-5001-01-5
The xxpress®NGx is designed to be used with a range of sample plates and accessories – some of which are listed below:
Table 3: xxpress® NGx Sample Plates and Accessories
Picture
Description
24 well xxplate™ SBS
(sample volume 5-40 µL)
54 well xxplate™ SBS
(sample volume 5-15 µL)
96 well xxplate™ SBS
(sample volume 1-5 µL)
xxsealer SA (semi-automatic)
xxcentrifuge
Other specialist accessories may be available at request. Please contact your distributor for details.
Page | 9
XXPRESS® NGX USER MANUAL – XP-5001-01-5
4 Technical Specification
Model Type
Model Name Model Numbers
xxpress®NGx Thermal Cycler XP-1003-CC1 (220V-240V) XP-1004-CC1 (100V-120V)
Physical Unit Properties
Dimensions (W x D x H) Weight
300 mm x 520 mm x 315 mm 35 kg
Mains Supply
Power Cord Rating
Inlet Module Type Supply Voltage
Supply Frequency Range Power Consumption Fuse Ratings and Sizes Mains Supply capability
IEC C13, 3-Core, 10A min Only use an appropriately rated and
approved supply cordset in accordance with the regulations of the country in which the equipment is put into service
IEC C14, DPST, Single Fuse 100V-120V AC (XP-1002-01) 220V-240V AC (XP-1001-01) 50Hz/60Hz 1000W max T8A/250V HRC 20x5mm The power for this unit must come from
a stable mains supply. Do not power from inverters, frequency convertors, or generators. (See Section 5 for further details)
Operating Environment
Temperature Range
Relative Humidity Range Maximum Operating Altitude
+18 to +30 C (Ambient temperatures above 25C
can reduce the rate of cooling) 20% to 80% non-condensing 2000m above sea-level
Page | 10
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Storage and Transportation
Temperature Range Relative Humidity Range
PCR Performance
Thermal Ramp Rate
Thermal Uniformity (Steady state)
Fluorescence Detection
-10 to +50 C 20% to 95% non-condensing
Maximum Ramp Rate 10C per second Adjustable from 0.1C to 10C per
second in 0.1C steps
±0.3C well to well across the test plate Uniformity achieved after 0.5 seconds
5 Channel Fluorescence Detection LED illumination CCD image capture
Detection Channels and Frequencies
Channel
Excitation (nm)
Emission (nm)
Common Dyes
Channel 1 Green
460-488
507-527
FAM, SYBR Green
Channel 2 Yellow
524-546
567-583
JOE, HEX, TAM,
VIC
Channel 3 Orange
567-589
617-633
ROX, Texas Red
Channel 4 Red
626-644
667-679
Cy5
Channel 5 Deep
Red
656-668
698-724
Cy5.5
Page | 11
XXPRESS® NGX USER MANUAL – XP-5001-01-5
5 Safety Precautions and Limitations of Use
It is essential that all users of this equipment have fully read and understood the following safety precautions and limitations of use before installing or operating the xxpress®NGx unit.
IMPORTANT
WARNING
The protection provided by this equipment may be impaired if it is not used in a manner described in this manual.
WARNING
It is essential that the user of this equipment is aware of the potential hazards associated with the unit and its accessories.
All operators should be familiar with the safety precautions and warnings given in these instructions before attempting to operate the unit.
Improper use of this unit or its accessories may impair their functionality and invalidate the manufacturer’s warranty.
Unit Handling Precautions
CAUTION
Care should be taken not to drop the unit or subject it to rough physical handling, both during normal use and during transportation and storage.
WARNING
The unit should be supported using the base of the unit when being lifted or moved. Do not lift the unit by any other part of the casework.
Care should be taken when lifting the unit due to its weight.
CAUTION
Care should be taken not to knock the LCD display. Do not use excessive force when touching the touch screen or
operating the buttons.
Page | 12
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Unit Installation and Operating Environment
DANGER
WARNING
The xxpress®NGx unit is designed for indoor laboratory use only.
The acceptable operating temperature range is 18ºC to 30ºC, with a relative humidity of 20% to 80% non-condensing, at a maximum altitude of 2000m above sea level.
If the unit is stored in conditions outside of these ranges, it must be left to stand unpowered until it has acclimatised to within these environmental limits before being powered.
DANGER
Use an appropriately rated and approved cordset as detailed in Section 4.
The unit must be connected to a suitably earthed mains supply, with appropriate earth-leakage and over-current protection.
WARNING
Always ensure that the mains power connector is securely inserted into the rear of the unit, and any excess power cord does not pose a potential trip or pull hazard.
Always ensure that the mains plug is easily accessible to allow disconnection from the supply.
WARNING
The power for this unit must come from a stable mains supply. Do not power from inverters, frequency convertors, or generators.
This unit requires a power supply with both a stable voltage and stable frequency. Inverters, frequency convertors, and generators do not provide acceptable stability and the unit can perform outside its specification, or fail to perform, or be damaged by the use of these forms of power supply.
DANGER
Do not operate the unit in any area which is, or has been, or is thought to have been exposed to explosive or flammable gases, vapours or liquids.
WARNING
The unit must be installed and operated on a solid, stable and level working surface; ensuring that the ventilation holes under the unit and at the rear of the unit are not obstructed.
Page | 13
XXPRESS® NGX USER MANUAL – XP-5001-01-5
General Operating Precautions
DANGER
Ensure that the power is switched off at both the AC mains supply outlet and at the back of the unit before inserting or removing the mains power cord.
DANGER
The xxplate™ can reach temperatures of over 100°C and can remain hot for a time after the xxpress® NGx system is turned off.
If conditions are set such that the experiment is finished with the plate at a high temperature it will exit the unit with the potential to burn the operator and care should be taken.
DANGER
The unit is intended for use with plates containing biological samples only.
Never use the unit to process any explosive, volatile or highly reactive substances or chemicals.
CAUTION
There is a possible finger crush hazard due to the moving parts of the door. Care should be taken when closing the door.
Page | 14
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Unit Maintenance and Serviceability
DANGER
WARNING
There are no user or operator serviceable parts inside the unit. Do not remove the unit casework.
Removal of the unit's casework will void the manufacturer’s
warranty and may expose the user to a Risk of Electric Shock resulting in serious injury or death.
DANGER
Once installed, the externally accessible unit fuse will only blow under a fault condition. This fuse should only be changed after the unit has been thoroughly inspected by a qualified engineer. See Section 8.1 for details.
DANGER
Always switch off the unit and disconnect the power cord before performing any cleaning or decontamination procedure.
If liquid is spilt into or over the unit, switch off and disconnect the power from the AC mains outlet before attempting to deal with the spillage.
CAUTION
Ensure that the xxpress®NGxunit has cooled down to room temperature before performing any cleaning operation and before moving or storing the unit.
CAUTION
The use of harsh chemicals and cleaning agents may damage the unit and degrade its performance.
Always follow the cleaning and decontamination procedures specified in Sections 8.2 and 8.3 of this instruction manual.
CAUTION
Do not autoclave any part of the unit or its accessories.
Page | 15
XXPRESS® NGX USER MANUAL – XP-5001-01-5
6 Unit Installation
Before installing the xxpress®NGx unit, please check that the delivery is complete and that the unit and any accessory parts are intact and free from any signs of transportation damage. Also ensure that all external and internal packaging has been removed from the unit before installation.
Please retain all packaging for future transportation and storage of the system.
The xxpress®NGx unit should be installed in a location which meets the following requirements:
Safe and suitable operating environment (see Section 5)
Solid, stable, level working surface
At least 10cm clearance around the unit to adjacent objects and
walls
Earthed AC mains power connection (see Section 4)
WARNING
Please also observe and abide by the Unit Installation and Operating Environment safety precautions and preconditions
listed in Section 5.
WARNING
The power for this unit must come from a stable mains supply that matches the specification on the back panel of the unit. Do not power from inverters, frequency convertors, or generators (see Section 5).
WARNING
Ensure that the correct fuse type has been fitted in the mains inlet fuse holder as detailed on the back of the unit. See Section 4 for details.
WARNING
If the unit has been stored in a cool environment, it must be left to stand unplugged until it has acclimatised to the new room temperature before being powered.
Page | 16
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Install and test the xxpress®NGx unit using the following procedure:
1) Place the unit on the suitably selected working surface (as specified above), ensuring that the ventilation holes underneath and on the rear of the unit are not obstructed.
2) Open the battery compartment and insert the 9 V battery as shown below. Close the battery compartment.
3) Connect the unit to the AC mains power outlet using an appropriately rated and approved cordset as detailed in Section 4.
4) Switch the mains power on at the supply outlet. Then switch the unit on using the power switch located at the rear of the unit.
5) Press the blue xxpress®NGx Power On/Off button, located on the front of the unit and wait whilst it starts up. This can take 30 seconds as the unit initialises. The system will then run a self-test (see Quick Start Sheet for details).
Insert 9Volt Battery
Page | 17
XXPRESS® NGX USER MANUAL – XP-5001-01-5
6) Start-up is complete when the LOGON screen is shown, see below.
Page | 18
XXPRESS® NGX USER MANUAL – XP-5001-01-5
7 Unit Operation
WARNING
Please ensure that you have read and fully understood all the Safety Precautions and Limitations of Use listed in Section 5 before attempting to operate the xxpress®NGx unit.
7.1 User Touch Screen Display and Controls
The xxpress®NGx user interface consists of a large, high definition, colour LCD display. It has an integrated touch screen that supports standard PC multi-finger gestures such as scrolling. The touch screen is designed to operate when wearing standard, clean, solvent-free laboratory gloves.
The unit is controlled through an application that runs on the xxpress®NGx. It has been designed to match experimental procedures used by Biotechnologists, making it intuitive and easy to use.
The xxpress®NGx software and user instructions will be regularly updated. You can find details of the latest version from your distributor or at www.xxpressPCR.com.
7.2 Operation of the xxpress®NGx qPCR System
7.2.1 Turning on xxpress®NGx
Before connecting xxpress®NGx to the mains electrical connection, make sure that the voltage and frequency detailed on the rear label of the unit (see table 2) matches the voltage supply.
Connect the supply using an appropriately rated and approved cordset as detailed in Section 4. Turn on the power on at the wall, then turn on the power input switch on the mains input connector (see table 2).
The unit is now powered and the rear fan may begin to operate depending on the temperature of the unit and the environment.
Page | 19
XXPRESS® NGX USER MANUAL – XP-5001-01-5
To wake-up the xxpress®NGx, press the blue Power On/Off Button on the front panel once. The xxpress®NGx will wake up, run a self­test and initialise its system. Figure 1 is the HOME screen. Once this screen is visible the xxpress®NGx is ready to use. You can always return to this screen by touching the logo in the bottom left corner of the screen.
From the HOME screen it is possible to create new experiments, run previously designed experiments, and analyse the results of previously run experiments. For each of the operations the
xxpress®NGx will guide the user through the steps required.
Figure 1
To check the software version running on the xxpress®NGx, go to the HOME screen and touch the xxpress®NGx logo in the bottom left corner of the screen. A pop-up window will appear listing the version and the software licence terms. See Figure 2.
Page | 20
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Figure 2
Close the screen by touching the symbol in the top right corner of the pop-up window.
The following sections of this manual will guide the user through the regular operations that can be completed using xxpress®NGx.
7.2.2 Designing a New PCR Experiment
To create a new test method or experiment touch the DESIGN
EXPERIMENT tile on the screen (the xxpress
®
NGx is fitted with a
touch screen which is sensitive and doesn’t need significant force to operate).
The next screen asks for the type of PCR experiment you wish to perform. Touch the desired experiment type, it will become highlighted as shown in Figure 3. The selection can by altered by touching a new PCR type.
Once the desired experiment type has been chosen, touch the arrow symbol at the bottom right of the screen to proceed to the next screen.
Page | 21
XXPRESS® NGX USER MANUAL – XP-5001-01-5
(Choosing the PCR type enables the xxpress®NGx to structure the starting thermal cycling profile shown in later screens to match the experimental need).
Figure 3
In this example, we have selected a 3-Step Real-Time PCR.
You will notice that there is a navigation bar running along the
bottom of the screen. The highlighted cyan text indicates the current stage.
Touching the arrow advances the user to the next stage (Figure 4), where details of the chemistry type used in the experiment are chosen.
Page | 22
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Figure 4
Touch to select whether PROBE or SYBR GREEN based chemistry is being used. This information enables the xxpress®NGx to preselect the camera sensitivity.
Figure 4a
In Figure 4a SYBR GREEN has been selected.
This brings up a list of some kit manufacturers that have been tested
by BJS Bio. Touch to select which kit manufacturer is being used. This brings up a list of their products, touch to select which product is being used. This information is used to programme the thermal profile with a suggested set of parameters.
Page | 23
XXPRESS® NGX USER MANUAL – XP-5001-01-5
As the xxpress®NGx is much faster than traditional thermal cyclers, the BJS Biotechnologies team have developed optimised profiles for running PCR assays at ultra-fast speeds. The figures are only a starting point and all the parameters are adjustable in later screens.
If the user wishes to use a product that isn’t listed, chose a similar product as a starting point and edit the thermal parameters at a later stage. With subsequent software releases, BJS Bio will add new chemistry options. Between releases, application notes for other chemistries can be found on the xxpress®NGx website, www.xxpressPCR.com. If a user wants advice on the thermal profile for a specific chemistry, please contact support@xxpressPCR.com.
Once the supplier and type of chemistry is selected (Figure 4b), the user proceeds to the next screen by touching the arrow at the bottom right of the screen as before.
Figure 4b
Page | 24
XXPRESS® NGX USER MANUAL – XP-5001-01-5
The next screen controls the thermal profile required for the experiment, see Figure 5. From the information already input into xxpress®NGx, the system automatically completes this screen with the correct experiment structure and suggested starting values.
Figure 5
Looking at our example 3-Step PCR, we chose a SYBR Green based test and Kapa SYBR 2G Fast Hot Start chemistry. In Figure 5 the thermal profile reflects this with an initial hot start at 95°C for 20 seconds followed by a 3-Step cycle which is repeated 40 times. Finally, there is a cooling step to bring the plate down to ambient temperature enabling safe removal from the unit.
Touching each part of the profile will produce a pop-up window allowing the parameters to be modified, see Figure 5a. Here we touched the Cycle 1 section.
Page | 25
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Figure 5a
Select the parameter to be changed. In figure 5a we have selected the dwell time on step 1 of the PCR cycle. On selecting, the parameter becomes highlighted. Type in the desired value using the keyboard on the right of the display. Once the user is happy with their changes they can save them by touching the tick symbol. Or if the user wishes to cancel and lose their changes, touch the cross.
In the example in Figure 5a there are 3 steps to the PCR. In step 1 the temperature is raised to 95°C using a ramp rate of 10°C per second. This temperature is then held for 1 second.
In the second step, the temperature is reduced to 60°C, using a ramp rate of 8°C per second. Once the temperature is at 60°C, this is held for 5 seconds.
In the third step, the temperature is increased to 72°C, using a ramp rate of 10°C per second. Once the temperature is at 72°C this is held for 5 seconds. At the end of this dwell there is a fluorescence measurement made using the channel 1 filter settings.
The temperatures for each step can be changed to any value between 40°C and 99°C in 0.1°C steps.
The thermal ramp rate can be specified for each step independently or set as a GLOBAL RAMP rate using the tile on the right of the pop-
Page | 26
XXPRESS® NGX USER MANUAL – XP-5001-01-5
up screen. This value can be set from zero to 10°C per second in
0.1°C steps.
The number of times the cycle should be run is set using the REPEAT tile. In Figure 5a the 3-step cycle is repeated forty times.
Touch the tick to accept these conditions or the cross to cancel any changes made.
For experiments where multiple cycles are required these can be added by touching the “plus” symbol and editing them in the same manner. Unwanted steps can be deleted using the “minus” symbol.
The final segment sees the temperature being reduced to 50°C and held for 10 seconds. This is done to cool the plate enough to touch when removed from the xxpress®NGx.
Based on the chemistry selection, xxpress®NGx will select a recommended sensitivity for the camera that measures the fluorescence, this is a combination of gain and exposure time. Touching the blue CAMERA tile will pop-up a window that will allow the sensitivity of the measurements in each of the five fluorescence channels to be individually adjusted, see Figure 5b.
To adjust touch the circle and then drag it to the position along the line that gives the desired value. The value range is from zero to 100 (to drag the circle maintain finger contact with the touch screen).
Page | 27
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Figure 5b
Touch the tick or cross to accept or cancel the changes respectively.
At this stage if a Melt Curve is to be run as part of the experiment it can be added as an additional cycle. It will then be run immediately after the PCR, see Figure 5c. BJS recommend this approach, see section 0 for ‘how to set-up a melt.
Figure 5c
Touch the plus symbol to add the extra cycle, see Figure 5c
Page | 28
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Figure 5d
Touch the MELT tile to add a Melt Curve cycle to the experiment, see Figure 5d.
The Melt cycle is then added to the list of cycles and can be edited by touching. See section 0 for how to set-up a Melt.
Figure 5d
Once all the parameters have been set, touch the forward arrow on the bottom right of the screen to advance to the Tray Selection.
Page | 29
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Figure 6
Chose the desired xxplate SBS size by touching it. The xxpress® NGx has three xxplate™ sizes: 24 well, 54 well, and 96 well. See
Table 3 for the reaction volumes they can accept.
Touch the forward arrow at the bottom right of the once the selection
is correct. The xxpress®NGx will recognise the xxplate™SBS size and check it matches this selection.
Figure 7
Page | 30
XXPRESS® NGX USER MANUAL – XP-5001-01-5
The next set-up screen, seen in figure 7, shows the chosen xxplate™ SBS well layout.
This is an optional page in which the wells can be labelled with information to help with the results analysis. To edit a group of wells, touch to select them, see Figure 7a. Touching them again will de­select them, alternatively you can touch DESELECT ALL to clear the complete selection.
Figure 7a
With the required wells highlighted touch EDIT and a pop-up window will appear, see Figure 7b.
Figure 7b
Page | 31
XXPRESS® NGX USER MANUAL – XP-5001-01-5
In the screen that pops up the type of sample that is to be put in the selected wells can be chosen.
SAMPLE refers to the Sample to be tested. ▪ NTC to a No Template Control sample. ▪ REFERENCE GENE a sample of known response for
comparison.
STANDARD CURVE where that process is being completed
You can also choose the display colours for the amplification and melt graphs. The xxpress®NGx will suggest a colour but this can be changed. Touch the coloured rectangle and a pop-up window will appear as shown in Figure 7c. Now touch the desired colour to select it.
Figure 7c
The samples can also be individually named. Touching the white tile will bring up a touchscreen keyboard so you can type a name for the sample type, see Figure 7d. Close the keyboard by pressing the X in the top right-hand corner, pressing enter, or touching the screen at a site other from the keyboard.
Page | 32
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Figure 7d
Touch the tick will accept the information. The process can be repeated as necessary to label other wells on the plate.
You can clear the well selection or even reset the plate completely using the tiles on the right.
Figure 7e
To make the selection of smaller wells easier, i.e. when using a 54 well xxplate™ or 96 well xxplate™, use two fingers to zoom into the plate area and pan around it. Move fingers in the direction of the blue arrows to zoom in (as shown in Figure 7e), pinch together to zoom out, or drag a finger across the screen to pan. These movements also work on the graph screens.
Page | 33
XXPRESS® NGX USER MANUAL – XP-5001-01-5
The cell references on the edge of the window make it clear which wells are being viewed, see Figure 7f.
Figure 7f
Once all the information is entered as shown in Figure 7f, touch the forward arrow at the bottom right of the screen to move to the review screen.
Figure 8
Figure 8 shows the review screen. This gives the user the opportunity to review the experiment set-up prior to running it. If anything needs editing, touch the blue navigation bar on the bottom on the screen. This will take the user back to the desired page.
Page | 34
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Once the user is happy that the experiment is correctly set-up, place the prepared, sealed, centrifuged xxplate™ SBS (see Section 7.2.4) into the xxpress®NGx. To do this press the Front Door Release Button (See table 2).
Place the prepared xxplate™ onto the slides ensuring that it correctly located so the slides engage into all three holes on either side of the xxplate™, see Figure 8a.
NEVER PUT A XXPLATE™ IN THE XXPRESS®NGx SYSTEM WITHOUT IT BEING HEAT SEALED. Only heat seal the xxplates SBS with the recommended heat sealing film.
NEVER PUT A LABEL ON THE BOTTOM OF THE XXPLATE™. Keep the bottom “black” surface of the xxplate™ clean and as
the unit measures the temperature of the samples using the radiation emitted from this surface.
NEVER WRITE ON THE PLASTIC WELLS OF THE XXPLATE™. The xxpress®NGx needs to be able to recognise the well boundaries. Labelling can be done using the plate layout section of the user interface or written in marker pen on the non-corrugated aluminium of the xxplate™.
Figure 8a
Page | 35
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Once the xxplate™SBS is correctly inserted, push the door closed.
Touch the forward arrow on the bottom right of the screen to start
the experiment.
Figure 9
Once the experiment is initiated the xxpress®NGx performs several checks on the plate. There will then be a short whirring sound as the clamps lower onto the plate before it starts the heat cycling.
As the cycles complete the fluorescence measurements will appear on the graph. The graph data can be adjusted using the filters on the right. Touch the tiles to toggle the filters on and off.
BASELINE SUBTRACT uses the fluorescence readings of the
first 5 cycles to set the initial trace level to zero. This makes it easier to compare multiple traces.
SMOOTH creates a best fit curve to the data. ▪ COLOUR COMPENSATION is used for multiplex experiments. It
reduces the effects of cross-talk (interference) from adjacent channels. It is not required for a single channel experiment.
Once all the thermal cycles have finished the xxpress®NGx will display a message confirming that the experiment is complete, see Figure 9a. Touch the tick to accept this. The results are saved
Page | 36
XXPRESS® NGX USER MANUAL – XP-5001-01-5
automatically and can now be analysed and/or exported (see Section 7.2.5).
Figure 9a
Page | 37
XXPRESS® NGX USER MANUAL – XP-5001-01-5
7.2.3 Designing a New Melt-Curve Experiment
Creating a Melt-Curve Experiment is very similar to creating the PCR Experiment. Select MELT from the PCR menu, see Figure 10.
Figure 10
Touch the arrow at the bottom right corner to advance to the Chemistry screen.
Figure 11
The Chemistry screen is simpler. Only the signalling type needs to be defined to allow xxpress®NGx to set the sensitivity of the camera, see Figure 11.
Advance to the Thermal Profile screen using the arrow.
Page | 38
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Figure 12
In the Thermal Profile screen, see figure 12, the cycles’ section
shows the temperature range for the melt. The user can adjust this and the number of measurements taken per degree.
Figure 12a
The example in Figure 12a above shows a Melt that will hold at 40°C for 5 seconds before raising the temperature from 40°C to 95°C taking 1 measurement per degree celcius.
The more measurements per degree, the greater the resolution on the melt temperatures and the longer the experiment will take.
Advance to the Tray Selection screen using the arrow.
Page | 39
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Select the type of xxplate™ SBS to be used, see Figure 13.
Figure 13
Advance to the Plate Layout screen using the arrow.
Figure 14
This is an optional page. A plate layout can be created as shown earlier, or alternatively it can be recalled from the memory. Touch the tile LOAD PLATE LAYOUT to select a previously created plate layout. This is shown in Figure 14.
Page | 40
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Touch the desired layout and then accept the selection by touching the tick, see Figure 14a.
Figure 14a
Advance to the Confirm and Run screen using the arrow.
Figure 14b
Figure 14b shows the confirmation screen that gives the user the opportunity to review their experiment prior to running it. If anything needs editing then touching the blue navigation bar on the bottom to return to earlier screens.
Page | 41
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Once the user is happy that the melt is correctly set-up they can insert the prepared xxplateSBS (see Section 7.2.4). To do this touch the Front Door Release Button (See table 2) to open the loading door.
Place the prepared xxplate™ onto the slides ensuring that the slides engage into all three holes on either side of the xxplate™SBS.
Close the door by pushing the front of it until it latches shut.
Touch the forward arrow on the bottom right of the screen to start
the experiment.
There are many other types of experiments that can be created on the xxpress®NGx unit, however they all follow the same format and at each stage all the parameters are available to edit.
7.2.4 Running the Experiment
Before running an experiment, the sample xxplate™ needs to be prepared. All xxplatesare required to have a heat sealed film applied to them. This seals the contents and prevents evaporation during the experiment. Details of how to seal the xxplates can be found in the xxsealer SA manual.
NEVER PUT A XXPLATE™SBS IN THE XXPRESS®NGx SYSTEM WITHOUT IT BEING HEAT SEALED. Only heat seal the xxplatesSBS with the recommended heat sealing film.
NEVER PUT A LABEL ON THE BOTTOM OF THE
XXPLATE™SBS. Keep the bottom “black” surface of the xxplate™ SBS clean and as the unit measures the temperature
of the samples using the radiation emitted from this surface.
NEVER WRITE ON THE PLASTIC WELLS OF THE XXPLATE™SBS . The xxpress®NGx needs to be able to recognise the well boundaries. Labelling can be done using the plate layout section of the user interface or written in marker pen on the non-corrugated aluminium of the xxplate™.
Page | 42
XXPRESS® NGX USER MANUAL – XP-5001-01-5
To place the prepared and sealed xxplate™SBS into the xxpress®NGx, touch the Front Door Release Button (See table 2) to open the loading door.
Figure 15
Place the xxplate™SBS onto the slides ensuring that the slides engage into all three holes on either side of the xxplate™SBS, see Figure 15.
When correctly located close the door by pushing on the front of it until it latches shut.
Once the experiment has been started by touching the forward arrow on the RUN screen, the xxpress®NGx will run a number of checks on the plate. These are listed in an information pop-up window on the display. These include the recognition of the plate to make sure it is the right size and in the correct position. The recognition of the plate also enables accurate fluorescence measurement.
The unit also clamps the plate in-situ so that it can control the heating and cooling precisely. The clamping of the plate can be recognised by the whirring of the motors that drive the clamps. Once all the checks are done the experiment will start and a results screen will be presented.
Page | 43
XXPRESS® NGX USER MANUAL – XP-5001-01-5
During operation, it is normal to hear a short hum during each cycle. This is caused by the pressurised air pumps used to cool the xxplate™.
Once the experiment is complete the clamps will release and an information window will pop up to confirm this.
Figure 16a
7.2.5 Experiment Results and Analysis
At the end of an experiment the results are displayed as shown in Figure 17. Depending on the type of experiment different parameters are shown. For a PCR run the Cq value is calculated whereas for a MELT the Tm value is calculated.
The results graph has several options for displaying the results. The curves can be smoothed or just show the raw data. Similarly, the data can be shown with the background fluorescence removed (BASELINE SUBTRACT) or not. Touch the tiles on the right to toggle these filters on and off.
The PCR or MELT and ANALYSIS tab tiles on the top left of the screen change the view from the graph to a results template.
Page | 44
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Figure 17
The Cq is automatically calculated and shown in for each well in the
ANALYSIS tab, see Figure 17a. The threshold for the Cq
measurements is also calculated and is shown in a blue box on the
PCR tab. This can be manually adjusted by sliding the red circle up
or down. This will result in the Cq values in the ANALYSIS tab being updated.
Page | 45
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Figure 17a
In the melt section, you have the option to differentiate the graph. This helps identify the Tm more easily. Figure 17b shows an example of a Melt in it standard form.
Figure 17b
Touching the DIFFERENTIATE tile on the right will give a display like the one shown in Figure 17c.
Page | 46
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Figure 17c
Results can be exported in an RDML or .csv format ready for further analysis. Select the blue EXPORT tile to do this. Note, if the smoothing filter or background filter is active when the data is exported the values will include this filtering.
Select the data to be exported and the export format using the screen shown in Figure 17d.
Figure 17d
Page | 47
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Note:
The RDML export format can be used to directly transfer information from xxpress®NGx to the BioGazelle PCR analysis program, qBase+. Contact your distributor for more information on qBase+ or go to the BioGazelle web-site, www.biogazelle.com
Figure 17e
For Multiplex experiments, where data is gathered on more than one fluorescence channel, the data can be viewed using the coloured channel tiles at the bottom of the screen, see Figure 17e.
Where there may be cross talk between adjacent fluorescence channels touching the COLOUR COMPENSATE tile activates the filter and will enhance the results.
7.2.6 Repeating Experiment and Reviewing Previous Results
Repeating a previously run experiment is straightforward on the
xxpress®NGx . Once an experiment has been run it appears in the
History tab and can be selected to re-run at any time. From the Home screen touch the HISTORY tile, see Figure 18.
Page | 48
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Figure 18
Then chose the previous experiment that is to be repeated from the pop-up selection box. Touch the to confirm your selection and open the file. The experiment will open at the Confirm and Run screen, see Figure 18a.
Figure 18a
Touching the menu along the bottom of the screen enables the selection of any of the experiment set up screens. These can be edited as required for the new experiment; any changes made will not affect the original experiment. Once all the parameters are as required and the xxplate™ has been put into xxpress®NGx , start
Page | 49
XXPRESS® NGX USER MANUAL – XP-5001-01-5
the experiment by touching the arrow on the bottom right of the screen.
This method of editing and re-running a previous experiment is a quick way of creating new experiments which are similar to ones previously run.
Figure 18b
All the results from the experiments that have been completed are stored on the xxpress®NGx. These can be reviewed at any time. To review the results of a completed experiment, touch the
ANALYSIS tile on the Home screen, see Figure 18.
Then chose the experiment of interest from the pop-up window, see Figure 18b. Touch the to open the results file.
It is also possible to import previously export data using this menu using the LOAD FROM CSV tile.
Page | 50
XXPRESS® NGX USER MANUAL – XP-5001-01-5
8 Maintenance and Servicing
Although the xxpress®NGx unit does not require any scheduled maintenance or servicing, the operator should regularly clean and inspect the unit for any defects, as described in Section 8.2 below.
DANGER
WARNING
Please observe and comply with all of the Unit Maintenance and Serviceability precautions listed in
Section 8. Never removed the unit casework. There are no user or
operator serviceable parts inside the unit. Always switch off and unplug the unit before performing
any cleaning or disinfecting tasks.
For technical and service related enquiries, please contact your distributor or BJS Bio at the address given on Page 2 of this manual.
8.1 Replacing the Unit Fuse
The unit fuse should only be replaced by a suitably qualified technician.
DANGER
The unit fuse will only blow as a result of an internal unit fault or if the voltage selector switch has been incorrectly set (see Section 5). This fuse should only be changed after the unit has been thoroughly inspected, and must be replaced with the exact type specified in Section 4.
Thoroughly inspect the unit for any signs of damage, loose components or liquid spillage or ingress. If in doubt, please contact BJS Bio on the number given on Page 2 of this manual.
DANGER
The fuse holder is removed by disconnecting the mains cord and then using a small flat bladed screwdriver to carefully pull open the fuse holder (see Figure 19) and remove the old fuse.
After replacing the fuse with the correct rated one for the operating voltage being used (Section 4), push the fuse holder firmly back into the inlet module.
The unit must be electrically safety tested for excess leakage current before being repowered from the mains supply.
Page | 51
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Figure 19
1. Remove the AC plug from the power inlet module
2. Use a small flat bladed screwdriver to carefully pull out the fuse holder
3. Fit the correct fuse in the far position only
4. Push the holder back into the inlet mode
8.2 Routine Cleaning and Inspection
The unit casework should be cleaned and inspected at regulator internals, and whenever contamination or spillage occurs, as follows:
1. Switch off the unit and disconnect the power before performing any inspection checks or cleaning.
2. Before cleaning, always inspect the unit casework, heater plate and moving parts for any signs of wear, damage, cracks or other defects.
3. Clean the casework using a damp cloth soaked with a disinfectant solution (such as Virkon), whilst wearing suitable PPE.
4. Remove any debris or fluff from around or between the moving parts of the door mechanism and sliding plate carrier.
5. Clean the display and buttons, taking care to avoid over wetting.
6. Check that the ventilation holes underneath and on the rear of the unit are clear of dust and fluff build-up. Replace the air filter as necessary.
DANGER
After cleaning, ensure that the unit is thoroughly dry, especially around the mains power inlet, before reconnecting the power cord and switching the unit on.
Page | 52
XXPRESS® NGX USER MANUAL – XP-5001-01-5
8.3 Decontamination Procedure
The unit and accessories should be decontaminated using the following procedure before being stored or transported.
Decontamination Procedure
Thoroughly clean all outside surfaces of the product (including any accessories, power cords, manuals, packaging, etc) with a damp cloth soaked with suitable disinfectant solution (such as Virkon).
8.4 Transportation and Storage
The xxpress®NGx unit and its accessories should be thoroughly decontaminated using the procedure detailed in Section 8.3 before being placed in its original packaging for transportation or storage.
WARNING
Refer to Section 4 for the acceptable range of Storage and Transportation environmental conditions.
Always ensure that the unit and accessories are completely dry and free of any condensation before being packed.
Page | 53
XXPRESS® NGX USER MANUAL – XP-5001-01-5
Certificate of Decontamination
We respect the health and safety of our clients and employees, and request that any products or accessories being returned are decontaminated in accordance with the procedure below.
1. Decontamination Procedure
Thoroughly clean all outside surfaces of the product (including any accessories, power cords, manuals, packaging, etc) with a damp cloth soaked with suitable disinfectant solution (such as Virkon).
Allow to dry fully before packing.
2. Decontamination Declaration
Company Name:
Address:
Product Code:
Serial Number:
Reason For Return:
Where Product Used:
Please tick the appropriate option(s) below: I certify that I have decontaminated the product as per the above procedure.
Decontaminant Used:
I certify that the product has not been exposed to any chemical or biological
materials.
Title:
Name:
Signature:
Date:
Telephone:
Email:
Page | 54
XXPRESS® NGX USER MANUAL – XP-5001-01-5
9 Warranty and Returns
BJS Biotechnologies Ltd. warrants the xxpress®NGx product, when purchased new and installed and operated in accordance with the instructions of this manual, to be free from defects in materials and workmanship, and will repair or replace, at their discretion, any unit or accessory which exhibits such defects.
In no event will BJS Biotechnologies be liable for any indirect, incidental or consequential damages resulting from any defect or warranty claim.
NOTE
Unspecified use or unauthorised modification of any part of the xxpress®NGx unit or its accessories or the use or attachment of any adaptor or peripheral not supplied, specified or sanctioned by BJS Biotechnologies will invalidate this warranty.
This warranty is provided to the original purchaser of the product for one year from the date of purchase.
Under the terms of this warranty, the product must be returned in its original packaging, transportation prepaid by the sender, with a copy of the Proof of Purchase and a detailed description of the problem.
WARNING
The product must be decontaminated using the procedure detailed in Section 8.3 and a Certificate of Decontamination supplied with any return. If the product is considered too hazardous to be shipped, please contact BJS Biotechnologies on the number given on Page 2 of this manual for further instructions.
Please contact your distributor (or BJS Bio on the number given on Page 2 of this manual) to assess the defect and arrange the return the product if necessary.
Page | 55
XXPRESS® NGX USER MANUAL – XP-5001-01-5
10 Product Disposal
At end-of-life, this product must be disposed of in accordance with your local authority regulations for the disposal of potentially hazardous waste and electronic equipment.
The unit and its accessories should be decontaminated using the procedure detailed in Section 8.3 before disposal or shipping.
Do not dispose of this product into unsorted municipal waste or public landfill.
Please contact your distributor (or BJS Bio at the address on Page 2 of this manual) for details of how to correctly dispose of this product.
BJS Biotechnologies Ltd
65 Bideford Avenue
Perivale Greenford Middlesex
UB6 7PP
United Kingdom
www.xxpressPCR.com
Loading...