Bischoff & Bischoff RMS - 20, WMS-5 Instruction Manual

Page 1
Enjoy mobility.
Bischoff & Bischoff
Instruction manual
RMS - 20 WMS - 5
Page 2
2
(Mattress pad with tube system)
WMS-5
(Mattress pad with honeycomb
structure)
Contents:
1| Foreword 3
1.1 Signs and symbols
2| Important safety instructions 4
2.1 General safety instructions
2.2 Purpose
2.3 Indications
2.4 Contraindications
2.5 Responsibility
2.6 Declaration of Conformity
3| Overview of product and accessories 8
3.1 Checking the delivery
3.2 Package
4| Installation and use 9 5| Technical data 11 6| User care instructions 13
6.1 Cleaning and disinfection
7| Repair and maintenance instructions 13 8| Disposal 14 9| Maintenance 14
10| Warranty 14
Page 3
RMS-20
WMS-5
Bi s c h o ff & B i s choff GmbH • 04 . 0 5 . 2011 3
English
1. Foreword
Dear user,
You have selected a high-quality product from Bischoff & Bischoff. Thank you for your trust in our products.
The mattress pad with micropores is used on the existing hospital mattress and does not replace it.
The continuous controlled exchange of air provided by the micropores creates a favourable microclimate in the area between the pad and the patient. Both mattress pads are designed for use with one patient.
The products are not suitable for reuse. Bischoff & Bischoff GmbH cannot accept liability for use other than the intended use.
Please read and follow the instruction manual carefully before using your new alternating pressure system for the first time. The manual is an integral and important com­ponent of the system. Keep it where you can refer to it.
Repairs and adjustments require specialised technical training and must be carried out by Bischoff & Bischoff authorised dealers only.
Page 4
1.1 Signs and symbols
Caution! Indicates specially safety-relevant instructions. Follow instruction manual.
Important! Indicates particularly useful information in the specific context.
Do not dispose of product with domestic waste. The product will be accepted by a recycling centre free of charge.
Suitable for indoor use only
Safety Class II (double insulation)
Type BF Applied Part
4
Page 5
English
Bi s c h o ff & B i s choff GmbH • 04 . 0 5 . 2011 5
2. Important safety instructions
2.1 General safety instructions
Before using the alternating pressure system check that the pad is correctly fastened in position and all parts are correctly connected. Note that the position of the patient must still be changed when using alternating pressure systems.
When using the alternating pressure system in adjustable hospital beds make sure that the air hoses are not jammed. Do not use open flames and heat (cigarettes) near the system.
Do not place flammable materials near the system. Keep sharp and pointed objects away from the mattress. The alternating pressure system is designed for persons with a body weight of 40 to max. 100 kg. For safety reasons the distance between the top edge of the closed bed rail and the top edge of the pad must be at least 22 cm when the alternating pressure system is in use. See the specifi­cations for the height of the mattress pad when inflated in Technical Data and the maximum permitted thickness of the mattress pad specified by the bed manufacturer.
When operating several electrical devices there is a possi­bility of unwanted interference (EMC radiation etc.).
RMS-20
WMS-5
Page 6
6
This may occur even with full compliance with current laws and standards. Devices such as mobile telephones may interfere with the system. If problems with interference are encountered disconnected the system from the mains and contact the technician.
If resuscitation is required, disconnect the hoses from the pump immediately. The RMS-20 also has a CPR valve for fast air discharge.
2.2 Purpose
Your new alternating pressure system from Bischoff & Bischoff is intended for professional use as equipment fo prevention and treatment of decubitus.
2.3 Indications
RMS-20: Decubitus prophylaxis and therapy up to and including stage 2 (as per EPUAP)
WMS-5: Decubitus prophylaxis and therapy up to and including stage 1 (as per EPUAP)
2.4 Contraindications
Use of the alternating pressure system is contraindicated with:
- unstable spinal and bone fractures
- patients sensitive to pain
- patients after myocardial infarction
Page 7
English
Bi s c h o ff & B i s choff GmbH • 04 . 0 5 . 2011 7
RMS-20
WMS-5
2.5 Responsibility
A warranty claim can only be accepted if the product is used under the specified conditions and for the specified purposes.
- changes, extensions, repairs and maintenance work must be carried out only by technicians authorised by Bischoff & Bischoff
- the alternating pressure system must be used as directed by all manuals and instructions for use
2.6 Declaration of Conformity
Bischoff & Bischoff as the manufacturer of the product declares in sole responsibility that the RMS-20 and WMS-5 alternating pressure systems fully comply with the require­ments of Directive 93/42/EEC.
Page 8
8
3. Overview of product and accessories
3.1 Checking the delivery
The alternating pressure system is delivered fully assembled in a carton. The authorised dealer provides instruction in its use.
If the goods are delivered by rail or road transport they must be inspected for shipping damage in the presence of the delivery person. If any damage is detected, contact our customer service department immediately (Tel.: +49 (0)700/6000 7070). Check the contents of the package for completeness and damage. If any discrepancies or damage are detected, contact our customer service department immediately (Tel.: +49 (0)700/6000 7070).
3.2 Package
The package contains the following components:
RMS-20
alternating pressure system
- RM-20 CPR pad (incl. hose system)
- RMC-20 cover
- WMK-10 control unit
- instructions for use
WMS-5
alternating pressure system
- WM-5 pad (incl. hose system)
- WMC-5 cover (optional)
- WMK-10 control unit
- instructions for use
Page 9
English
Bi s c h o ff & B i s choff GmbH • 04 . 0 5 . 2011 9
RMS-20
WMS-5
4. Installation and use
4.1. Place the control unit on a stable and horizontal sur­face or hang the pump on the hook provided at the end of the bed.
4.2. Position the deflated pad on the mattress with the hose connections towards the end of the bed. Fasten the:
- RMS-20 with the rubber straps at all four corners
- WMS-5 by folding the flaps under the mattress (see photo at sec. 1).
4.3. Now connect the mattress cover hoses to the con­trol unit. Then place the cover over the mattress. Make sure that the hoses are not kinked or jammed.
Fig. 1
Fig. 2
Fig. 3
Page 10
10
Fig. 4
Fig. 5
Fig. 6
4.4. Turn the rotary switch to 5 MAX and switch on the pump. The red indicator light comes on and the alarm sound is activated. For more information see Section 5.
4.5. If the alarm sound is disturbing, press the button on the case to the right until the pad is fully inflated.
4.6. When the indicator light switches off, set the air pressure for the weight of the patient with the rotary switch.
Switch position Patient weight 1 40 to 55 kg 2 55 to 65 kg 3 65 to 75 kg 4 75 to 85 kg 5 85 to 100 kg
The control unit has an optical and an acoustic alarm system. The alarm sound is activated as soon as the mattress pad loses pressure and the patient is in danger of lying on the mattress. Check the hose lines and/or the mattress pad for damage.
Page 11
English
Bi s c h o ff & B i s choff GmbH • 04 . 0 5 . 2011 11
RMS-20
WMS-5
5. Technical data
RM-20 mattress pad
Model: tubes
Material: mattress pad / Nylon cover / Nylon
Number of cells: 20
Micropores: 4 x 3 laser-cut
Dimensions: (LxW) 200 x 90 cm
Height, inflated: 10 cm
Weight: 3.8 kg
WM-5 mattress pad
Model: honeycomb
Material: mattress pad / Nylon cover / Nylon
Number of cells: 130
Micropores: 4 x 2 laser-cut
Dimensions: (LxW) 200 x 90 cm
Height, inflated: 7 cm
Weight: 2.1 kg
Page 12
12
WMK - 10 control unit
Voltage: 230 V AC
Frequency / power: 50-60 Hz / 8W
Max. pressure: 120 mmHg
Pressure adjustment: automatic depending on
specified patient weight
Alternating cycle: two-chamber system
Total / relaxation time: 10 / 5 minutes
Weight: 1.8 kg
Dimensions: (LxWxH) 13.5x 27x 11.5 cm
Alarms: optical and acoustic Warning
on pressure loss in the mattress pad RMS-20
Pressure progress
Cycle time in minutes
Cycle chart RMS-20 and WMS-5
Chamber 1
Chamber 2 Chamber 1 Chamber 2
Page 13
English
Bi s c h o ff & B i s choff GmbH • 04 . 0 5 . 2011 13
RMS-20
WMS-5
6. Care instructions for the user
6.1 Cleaning and disinfection
Cover: the cover can be machine-washed at 65°C. Use a mild detergent and rinse with clean water after. For disin­fection add a standard disinfectant detergent to the rinse cycle that will eliminate bacteria and germs reliably at a wash temperature of 65°C.Leave the cover to air-dry.
Mattress pad: clean the mattress pad with a mild soap solution and a soft brush. Rinse with clean water after. For disinfection wipe the surface with a wipe disinfectant. Follow the manufacturer's directions.
See the B+B instructions for care, hygiene and maintenan­ce at www.bischoff-bischoff.com
Do not use high pressure cleaners or steam cleaners.
7. Repair and maintenance instructions
Warranty claims can be made for all defects of the product that can be confirmed as due to material or manufacturing faults. If you make a claim under the warranty, please submit the completed warranty certificate. Our alternating pressure systems are guaranteed for two years. If the instruction manual is not followed or repairs and maintenance work are done incorrectly any claims under the warranty will not be accepted and no liability will be accepted.
Our customer service is available to answer questions at +49 (0)700/6000 7070.
Page 14
14
The dealer can receive the required information, documentation and circuit diagrams for repair and service on request.
8. Disposal
If your alternating pressure system is no longer used and you wish to dispose of it, please con­tact your dealer. If you wish to dispose of it your­self, please contact a local recycling company for the disposal regulations of your jurisdiction.
9. Maintenance
Our customer service can provide a maintenance schedule for the product.
10. Warranty
The alternating pressure systems are generally maintenan­ce-free. They should be checked regularly by technicians. See our maintenance instructions for information. We recommend having the alternating pressure systems inspected by the dealer at least once a year.
If faults or defects in the alternating pressure system are found, it must be returned to the dealer for repair imme­diately.
Our customer service can provide a mainten­ance schedule for the product.
To find a dealer in your area please contact our customer service department at +49 (0)700/6000 7070.
Page 15
English
Bi s c h o ff & B i s choff GmbH • 04 . 0 5 . 2011 15
RMS-20
WMS-5
Warranty certificate
(Please note our general business terms and conditions)
Model: RMS-20 / WMS-5
Serial number:*
(to be entered by dealer)
Dealer:
Date and stamp
* See the side frame for information stickers
Page 16
Bischoff & Bischoff GmbH
Becker-Goering-Strasse 13 • D-76307 Karlsbad
Tel.: +49 700/6000 7070 • Fax: +49 700/6000 8080
www.bischoff-bischoff.com
info@bischoff-bischoff.de
Loading...