VIP System
PARAMETER SPECIFICATIONS
Frequency Range 307 kHz - 150 MHz (Sensor Dependent)
Frequency Resolution 100 Hz
Frequency Accuracy ± 1 kHz
Harmonics 10 maximum, up to 150 MHz (Sensor Dependent)
Number of fundamentals (F0) Maximum of 3 simultaneously
Digital > Voltage, current, phase, frequency, impedance, power at frequencies selected by user
Analog 5 Outputs, 0-10Vdc, 1000Ω-source
Update Rates 60 Hz typical for 1 fundamental (Note 1)
Network Protocol DeviceNet, Ethernet
RF Power, Max 10 kW or maximum power limit of RF connector
RF Connector Custom or QC
Receiver Operating +20 to +40 °C (68 to 104 °F)
Receiver Storage -20 to +80 °C (-4 to +176 °F)
Cable Operating 0 to +100 °C (32 to 212 °F)
Cable Storage -20 to +100 °C (-4 to 212 °F)
Sensor Operating/Storage Refer to Sensor Specication
Humidity, Max 85% Non-condensing
Air Pressure, min 745 mbar (equivalent to 2,500 m / 8,200 ft. max altitude)
Operating Power 11-24 Vdc, 1.4-3A input to receiver
PARAMETER VOLTAGE CURRENT PHASE ANGLE
Measurement RF: 1 to 3000V
rms
(Note 2) 0.1 to 100 A
rms
(Note 2) -180° to + 180°
Resolution IEEE 754 Single Precision Floating Point
Uncertainty 307 kHz-1 MHz
(Note 3)
for F°, ± 0.5 V or 1% of reading
whichever is greater for F
n
, ± 0.1 V or
2% of reading, whichever is greater
(95% condence intreval)
for F°, ± 0.05 A or 1% of reading
whichever is greater for F
n
, ± 0.10 A
or 2% of reading, whichever is greater
(95% condence intreval)
Absolute Angle:
for F°, ≥ 10 V, 1A:, ±1°
for F°, < 10 V, 1A:, ±4°
for F
n
, ≥ 10 V, 1A:, ±2°
for F
n
, < 10 V, 1A:, ±6°
(95% condence intreval)
Uncertainty 1-100 MHz (Note 3) for F°, ± 0.1 V or 1% of reading
whichever is greater for Fn, ± 0.2 V or
2% of reading, whichever is greater
(95% condence intreval)
for F°, ± 0.01A or 1% of reading
whichever is greater for Fn, ± 0.02 A
or 2% of reading, whichever is greater
(95% condence intreval)
Absolute Angle:
for F°, ≥ 10 V, 1A:, ±1°
for F°, < 10 V, 1A:, ±4°
for F
n
, ≥ 10 V, 1A:, ±2°
for F
n
, < 10 V, 1A:, ±6°
(95% condence intreval)
Uncertainty 100-150 MHz
(Note 3)
for F°, ± 0.2 V or 2% of reading
whichever is greater for Fn, ± 0.4 V or
4% of reading, whichever is greater
(95% condence intreval)
for F°, ± 0.02 A or 2% of reading
whichever is greater for Fn, ± 0.04 A
or 4% of reading, whichever is greater
(95% condence intreval)
Absolute Angle:
for F°, ≥ 10 V, 1A:, ±2°
for F°, < 10 V, 1A:, ±8°
for F
n
, ≥ 10 V, 1A:, ±4°
for F
n
, < 10 V, 1A:, ±12°
(95% condence intreval)
Receiver Temperature
Derating - From 25 °C
± 0.05% / °C ± 0.05% / °C ± 0.1% / °C
30303 Aurora Rd. Solon, OH 44139 866.695.4569 www.bird-technologies.com
Note 1: Typical data rate for 1 fundamental, 10 harmonics, auto-ADC mode, and no averaging is 60 Hz. Data rate can vary significantly depending
on conguration, network trac, and host performance.
Note 2: Maximum power is limited by the size of the sensor line section and connectors. See sensor specication document.
Note 3: At customer specied frequencies.