The following are general safety precautions that are not necessarily related to any specific part or procedure, and
do not necessarily appear elsewhere in this publication. These precautions must be thoroughly understood and
apply to all phases of operation and maintenance.
WARNING
Keep Away From Live Circuits
Operating Personnel must at all times observe general safety precautions. Do not
replace components or make adjustments to the inside of the test equipment with the
high voltage supply turned on. To avoid casualties, always remove power.
WARNING
Shock Hazard
Do not attempt to remove the RF transmission line while RF power is present.
WARNING
Do Not Service Or Adjust Alone
Under no circumstances should any person reach into an enclosure for the purpose of
service or adjustment of equipment except in the presence of someone who is
capable of rendering aid.
WARNING
Safety Earth Ground
An uniterruptible earth safety ground must be supplied from the main power source
to test instruments. Grounding one conductor of a two conductor power cable is not
sufficient protection. Serious injury or death can occur if this grounding is not properly
supplied.
WARNING
Chemical Hazard
Dry cleaning solvents for cleaning parts may be potentially dangerous. Avoid
inhalation of fumes or prolonged contact with skin.
WARNING
Resuscitation
Personnel working with or near high voltages should be familiar with modern
methods of resuscitation.
WARNING
Remove Power
Observe general safety precautions. Do not open the instrument with the power on.
i
Page 3
Safety Symbols
Warning notes call attention to a procedure which, if not correctly performed, could
Caution notes call attention to a procedure which, if not correctly performed, could
The caution symbol appears on the equipment indicating there is important
information in the instruction manual regarding that particular area.
Note: Calls attention to supplemental information.
Warning Statements
Safety Precautions
WARNING
result in personal injury.
CAUTION
result in damage to the instrument.
The following safety warnings appear in the text where there is danger to operating and maintenance personnel
and are repeated here for emphasis.
WARNING
Never attempt to connect or disconnect RF equipment from the transmission line
while RF power is being applied. Leaking RF energy is a potential health hazard.
See page 6, 34, and 43.
WARNING
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement or changing the line voltage
setting.
See page 34, 37, 37, and 43.
WARNING
Use only Nickel Metal Hydride (NiMH) batteries that have a minimum capacity of 4500
milliampere hours (mAh). Do not install batteries that are not Nickel Metal Hydride.
Do not install NiMH batteries that have less than 4500 mAh capacity. Failure to
comply may result in damage to the batteries, damage to the instrument, and injury
to personnel from battery chemicals.
See page 39 and 40.
WARNING
Heavy load. Do not attempt to lift unaided.
See page 43.
ii
Page 4
Thruline RF Power Meter, Model 4421
Caution Statements
The following equipment cautions appear in the text whenever the equipment is in danger of damage and are
repeated here for emphasis.
CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may reduce the
maximum power rating of the unit.
See page 3, 43, 46, 48, and 52.
CAUTION
The Bird 4421 must be powered off when connecting or disconnecting the power
sensor from the power meter.
See page 6 and 43.
CAUTION
Do not use the power sensor with a load VSWR greater then 2:1. Damage to the
power meter, power sensor, or the RF power source could occur.
See page 6.
Long-term storage of this instrument can affect battery performance and reduce
battery life. Do not store the instrument for long periods of time without recharging
See page 8
During remote operation, periodically monitor the bus service request line. Failure to
See page 11 and 16.
maintenance should not be attempted. Removal or disturbance of the power sensor
See page 34 and 43.
Failure to install the properly rated fuse may result in equipment damage or nuisance
CAUTION
the batteries (refer to page 39). Failure to comply may result in reduced battery
charge and shortened battery life.
CAUTION
detect a service request could result in equipment damage.
CAUTION
Due to the complexity of the Bird Power Sensor, field repairs beyond general
cover can result in cancellation of lifetime warranty.
CAUTION
failures.
See page 37.
iii
Page 5
Safety Precautions
Safety Statements
USAGE
ANY USE OF THIS INSTRUMENT IN A MANNER NOT SPECIFIED BY THE MANUFACTURER MAY
IMPAIR THE INSTRUMENT’S SAFETY PROTECTION.
USO
EL USO DE ESTE INSTRUMENTO DE MANERA NO ESPECIFICADA POR EL FABRICANTE, PUEDE
ANULAR LA PROTECCIÓN DE SEGURIDAD DEL INSTRUMENTO.
BENUTZUNG
WIRD DAS GERÄT AUF ANDERE WEISE VERWENDET ALS VOM HERSTELLER BESCHRIEBEN,
KANN DIE GERÄTESICHERHEIT BEEINTRÄCHTIGT WERDEN.
UTILISATION
TOUTE UTILISATION DE CET INSTRUMENT QUI N’EST PAS EXPLICITEMENT PRÉVUE PAR LE
FABRICANT PEUT ENDOMMAGER LE DISPOSITIF DE PROTECTION DE L’INSTRUMENT.
IMPRIEGO
QUALORA QUESTO STRUMENTO VENISSE UTILIZZATO IN MODO DIVERSO DA COME
SPECIFICATO DAL PRODUTTORE LA PROZIONE DI SICUREZZA POTREBBE VENIRNE
COMPROMESSA.
iv
Page 6
Thruline RF Power Meter, Model 4421
SERVICE
SERVICING INSTRUCTIONS ARE FOR USE BY SERVICE-TRAINED PERSONNEL ONLY. TO AVOID
DANGEROUS ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING UNLESS QUALIFIED TO DO
SO.
SERVICIO
LAS INSTRUCCIONES DE SERVICIO SON PARA USO EXCLUSIVO DEL PERSONAL DE SERVICIO
CAPACITADO. PARA EVITAR EL PELIGRO DE DESCARGAS ELÉCTRICAS, NO REALICE NINGÚN
SERVICIO A MENOS QUE ESTÉ CAPACITADO PARA HACERIO.
WARTUNG
ANWEISUNGEN FÜR DIE WARTUNG DES GERÄTES GELTEN NUR FÜR GESCHULTES
FACHPERSONAL.
ZUR VERMEIDUNG GEFÄHRLICHE, ELEKTRISCHE SCHOCKS, SID WARTUNGSARBEITEN
AUSSCHLIEßLICH VON QUALIFIZIERTEM SERVICEPERSONAL DURCHZUFÜHREN.
ENTRENTIEN
L’EMPLOI DES INSTRUCTIONS D’ENTRETIEN DOIT ÊTRE RÉSERVÉ AU PERSONNEL FORMÉ AUX
OPÉRATIONS D’ENTRETIEN. POUR PRÉVENIR UN CHOC ÉLECTRIQUE DANGEREUX, NE PAS
EFFECTUER D’ENTRETIEN SI L’ON N’A PAS ÉTÉ QUALIFIÉ POUR CE FAIRE.
ASSISTENZA TECNICA
LE ISTRUZIONI RELATIVE ALL’ASSISTENZA SONO PREVISTE ESCLUSIVAMENTE PER IL
PERSONALE OPPORTUNAMENTE ADDESTRATO. PER EVITARE PERICOLOSE SCOSSE
ELETTRICHE NON EFFETTUARRE ALCUNA RIPARAZIONE A MENO CHE QUALIFICATI A FARLA.
v
Page 7
Safety Precautions
UNITS ARE EQUIPPED WITH RECHAREABLE BATTERIES. THESE ARE TO BE REPLACED BY
AUTHORIZED SERVICE PERSONNEL ONLY!!!
LAS UNIDADES VIENEN EQUIPADAS CON BATERIAS RECARGABLES. ¡¡¡Y SOLAMENTE EL
PERSONAL DE SERVICIO AUTORIZADO PUEDE REEMPLAZARLAS!!!
GERÄTE SIND MIT WIEDER AUFLADBAREN BATTERIEN BESTÜCKT. BATTERIEN SIND NUR VON
QUALIFIZIERTEM SERICE PERSONAL AUSZUWECHSELN!!!
CES DISPOSITIFS SONT ÉQUIPÉS DE BATTERIES RECHARGEABLES. SEUL LE PERSONNEL
D’ENTRETIEN AUTORISÉ EST HABILITÉ À LES REMPLACER !
LE UNITÀ SONO DOTATE DI BATTERIE RICARICABILI, CHE DEVONO DA COME SPECIFICATO
DAL PRODUTTORE LA PROTEZIONE DI SICUREZZA POTREBBE VENIRNE COMPROMESSA.
vi
Page 8
Thruline RF Power Meter, Model 4421
BE SURE THE 115/230V AC VOLTAGE SELECTOR IS SET TO THE PROPER LINE VOLTAGE, AND
THE CORRECT AC LINE FUSE IS INSTALLED BEFORE AC POWER IS APPLIED.
S’ASSURER QUE LE SÉLECTEUR DE TENSION 115/230V C.A. EST BIEN RÉGLÉ POUR LA
TENSION DU RÉSEAU ET QUE LE FUSIBLE DE LIGNE C.A. CORRECT EST EN PLACE AVANT DE
METTRE SOUS TENSION C.A.
CERCIORESE QUE EL SELECTOR DE VOLTAJE DE 115/230V CA ESTE COLOCADO A LA LINEA
DE VOLTAJE APROPIADA Y QUE EL FUSIBLE ESTE INSTALADO A LA LINEA CA ANTES DE
APLICAR LA CORRIENTE ALTERNA.
VOR EINSCHALTEN DER WECHSELSTROMZUFUHR SICHERSTELLEN, DASS DER 115/230V
WECHSELSPANNUNGS-SELEKTOR AUF DIE VORSCHRIFTSMÄSSIGE LEITUNGSSPANNUNG
EINGESTELLT UND DIE RICHTIGE WECHSELSTROM-HAUPTSICHERUNG EINGESETZT IST.
PRIMA DI EROGARE CORRENTE, ASSICURARSI CHE IL SELETTORE DI VOLTAGGIO 115/230
V.C.A. SIA REGOLATO CORRETTAMENTE E CHE IL FUSIBLE ADATTO ALLA LINEA DI
ALIMENTAZIONE C.A. SIA INSTALLATO.
vii
Page 9
About This Manual
About This Manual
This manual covers the Bird 4421 RF Power Meter and the its sensors. This manual covers the operating and
maintenance instructions for the following models:
We have made every effort to ensure this manual is accurate. If you should discover any errors, or if you have
suggestions for improving this manual, please send your comments to our factory. This manual may be periodically
updated. When inquiring about updates to this manual refer to the part number and revision level on the title page.
Contents
Chapter Layout
Introduction — Describes the purpose and function of the Power Meter as well as a general overview of the
product.
Installation — Describes the installation instructions for the power meter models covered in this manual.
Operating Instructions — Describes the features of the power meters and provides power-up instructions.
IEEE-488 GPIB Interface — Describes the features of the IEEE-488 GPIB Interface.
RS-232 Interface — Describes the features of the RS-232 Interface.
Maintenance — Contains preventative maintenance information, troubleshooting and repair sections, and a list
The Bird 4421 RF Power Meter is one component of a complete RF power measurement system. An RF power
sensor such as a Bird 4021 is also required. The system can be controlled with the front panel buttons, or remotely
through an RS-232 connection or a GPIB-488 connection. See
Items Supplied
Bird 4421 RF Power Meter
AC Power Cord
Sensor Cable
Instruction Manual
Optional Accessories
Panel Mount Kit (P/N 4421-250) — Allows the Bird 4421 to be installed in a standard 19" panel for rack mount
applications.
Null Modem Kit (P/N 4380-250) — Contains the hardware necessary to allow the 4421 to be remotely
controlled by controllers with different wiring arrangements. Requires an RS-232 interface module.
Figure 1 on page 2
.
1
Page 13
Figure 1 Bird 4421 Meter Outline Drawing
Introduction
1. Operating Push Buttons
2. LCD
3. ON/OFF Push Button
4. Handle
5. Central Button
6. Fuse Access Door
7. AC Line Module
8. Master ON/OFF Switch
9. GPIB connector
10. GPIB DIP switches
* A shorter case was introduced in 2016,
4421 length dimension has changed for
models built in 2016 and later.
11. RS-232 connector
12. RS-232 DIP switches
13. Power Sensor Socket
2
Page 14
Thruline RF Power Meter, Model 4421
Power Sensors
Power sensors are available with a variety of connectors; see
complete list. Since the accuracy is critically dependent on the connectors used at calibration, do not remove or
change the connectors.
Changing the sensor’s connectors will invalidate calibration data, and may reduce the
maximum power rating of the unit.
4020 Series
Bird 4020 Series Power Sensors are designed for lab or field use and are accurate to within ± 3%(1) of reading.
4027A Series
Bird 4027A Series Power Sensors are designed for use in semiconductor processing and calibration applications.
Stringent calibration provides long-term unit-to-unit repeatability, allowing consistent amounts of RF energy to be
applied to the etch process over many etch cycles. 4027A Sensors are accurate to ± 1%(1) at specified calibration
frequencies and power levels.
4027F Series
Bird 4027F Series Power Sensors are similar to the 4027A series. However, additional filtering allows the 4027F to
ignore harmonics of the signal being measured. The 4027F is also less sensitive to AM components of the signal.
4027F Sensors are accurate to ± 1%(2) at specified calibration frequencies and power levels.
"Available Connectors" on page 52
CAUTION
for a
Figure 2 Power Sensor Outline Drawing, 4020, 4027A, 4027F, & 4028A Series
4028 Series
Bird 4028 Series Power Sensors are high power sensors otherwise similar to the 4027A series. 4028A sensors are
based on a 7/8” line, 4028B sensors use a 1-5/8” line, and 4028C sensors use a 3-1/8” line. 4028 Sensors are
accurate to ± 2% at specified calibration frequencies and power levels.
3
Page 15
Figure 3 Power Sensor Outline Drawing 4028B Series only
Figure 4 Power Sensor Outline Drawing 4028C Series only
This chapter provides information about preparing the Bird 4421 for use.
Sensor Connection
CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may reduce the
maximum power rating of the unit.
CAUTION
The Bird 4421 must be powered off when connecting or disconnecting the power
sensor from the power meter.
1.Turn OFF the ON/OFF rocker switch on the meter’s rear panel.
2.Align the latch on the cable with the notch of the “Power Sensor” socket on the power meter’s rear panel.
3.Insert the cable until it clicks into place.
4.Connect the other end of the sensor cable to the sensor.
RF Line Connection
WARNING
Never attempt to connect or disconnect RF equipment from the transmission line
while RF power is being applied.
Leaking RF energy is a potential health hazard.
CAUTION
Do not use the power sensor with a load VSWR greater then 2:1. Damage to the power
meter, power sensor, or the RF power source could occur.
Connect the end of the power sensor labeled “SOURCE” to the RF source. Connect the end labeled “LOAD” to the
load or antenna. Reversing these connections will cause measurement errors.
Panel Mounting the 4421 Power Meter
You can install the 4421 Power Meter in an equipment rack if you have the optional panel mount kit (refer to
"Optional Accessories" on page 1
on page 7
kit.
shows the overall dimensions and mounting points for a 4421 Power Meter installed in a panel mount
Note:The power supply interrupt switch for the 4421 Power Meter is located on the rear panel. When
you install a unit in a panel mount kit, you need to provide a means to interrupt the power supply that
is easily accessible to the user (such as a switch mounted in the panel).
). The panel mount kit includes complete installation instructions.
Figure 5
6
Page 18
Thruline RF Power Meter, Model 4421
Figure 5 Panel Mounting Dimensions
Handle Operation
The handle on the Bird 4421 can be set to four different positions (see
center buttons on both sides. Releasing the buttons will lock the handle into position.
Figure 6 Handle Positions
7
Figure 6
). To adjust the handle, press the
Page 19
Installation
115/230V AC Input Power
The internal power supply can operate at 115 VAC or 230 VAC. There is no user setting to select the input voltage
because the power supply automatically senses the input voltage. However, you must make sure that you have the
correct power cord plug and fuses installed for the line voltage you are using (refer to
page 37
AC Line Connectors
To make the AC line cord compatible with non-U.S. voltages, users must install the appropriate connector on the
power cord.
).
"Replacing Fuses" on
Batteries
The Bird 4421 RF Power Meter is completely portable and is powered from internal rechargeable nickel metal
hydride batteries.
The batteries are shipped in a low charge state. It is, therefore, recommended that you charge the instrument for
approximately 28 hours before using it for continuous operation.
Note:Battery charging at temperatures greater than 45° C (113° F) can result in reduced operational
time.
CAUTION
Long-term storage of this instrument can affect battery performance and reduce
battery life. Do not store the instrument for long periods of time without recharging
the batteries (refer to page 39). Failure to comply may result in reduced battery
charge and shortened battery life.
8
Page 20
Chapter 3Operating Instructions
This chapter describes operator controls and indicators on the Bird 4421 RF Power Meter. For remote operation
using a IEEE-488 (GPIB) or RS-232 controller, refer to the instructions in
Push Button Functions
Figure 7 Push Buttons
Push ButtonDescription
FWD, RFL
SWR
MIN, MAX
dBm
LIGHT
AUTO
UP, DOWN
ON/OFF
Press to measure forward or reflected RF power. FWD or RFL
indicator and current unit of measure turn on.
Press to measure standing wave ratio. SWR indicator turns on. Value
displayed will be between 1.0 and 199.9
Used after pressing FWD, RFL, SWR, or dBm. Displays the minimum
(maximum) measured value of the previous function as long as MIN
(MAX) is held down.
Used after pressing FWD or RFL. dBm indicator turns on. Power is
displayed in dBm units.
Note:Used after pressing SWR. Return loss is displayed.
Press to turn on or turn off the display’s backlight. If left on, the light
automatically shuts off after 30 minutes.
Press to automatically set the scale. AUTO turns on.
Press to select the next higher (lower) scale. If the scale is too high
for the power sensor, an error will be displayed.
Note:Used while AUTO indicator is on. Stops automatic scaling. AUTO indicator turns off.
Press to turn the power meter on or off. This switch will not work if
the master power switch on the rear panel is OFF.
Note:The display flashes for about 30 seconds when
the unit is first turned on.
Chapter 4
or
Chapter 5
respectively.
9
Page 21
Operating Instructions
Error Codes
The Bird 4421 displays error codes when the RF power is either below the selected range (underrange) or above the
selected range (overrange).
Table 1 Error Codes
Table 2 Function Limits
FWD, RFL
FWD dBm, RFL dBm
SWR
Return Loss
Table 1
SymbolExplanation
FunctionLimitError
displays the error codes and
Value greater than overrange limit of function
Value less than underrange limit of function
Power > 199.9 % of full scale or 120 % of
top range
Power > 120 % of full scaleOverrange
Power < 3 % of low rangeUnderrange
FWD < 20 % of low rangeUnderrange
FWD – RFL = 0Overrange
FWD < 20 % of low rangeUnderrange
RFL < 20 % of low rangeUnderrange
Return Loss > 40 dBUnderrange
Table 2
lists the function limits.
Overrange
Audible Warning
If the RF power level exceeds 120 % of the power sensor’s maximum power capability, the power meter will sound a
warning buzzer.
10
Page 22
Chapter 4IEEE-488 GPIB Interface
This chapter discusses setup of the IEEE-488 interface feature and describes the IEEE commands that apply to the
Bird 4421. Operators should understand IEEE standard 488-1978 and have basic computer programming skills
before attempting to write any programs.
CAUTION
During remote operation, periodically monitor the bus
service request line. Failure to detect a service request
could result in equipment damage.
Description
The Bird 4421 IEEE-488 (GPIB) interface has an eight-position DIP switch that sets operational conditions and
interface addresses. The bottom line of the display indicates the current bus status.
Cable Connector
The interface uses a standard IEEE-488 cable connector. Pin assignments are listed in
Can handshake data or command bytes when the unit is
acting as a source.
Can handshake the bus when it is acting as the acceptor of
data or commands.
Can send data over the bus to other devices. This capability
T5
Talker
exists only after the instrument has been addressed to talk,
or after a reading in talk-only mode.
Can receive device-dependent data over the bus. This
L4
Listener
capability exists only after the unit has been addressed to
listen.
SR1
Service RequestCan request service from the controller.
RL1
Remote-LocalCan be placed in remote or local mode.
PP0
Parallel PollDoes not have parallel polling capability.
DC1
Device ClearCan be reset to factory settings.
DT1
Device TriggerCan have its readings triggered.
C0
ControllerDoes not have controller capability.
E1
Bus Driver TypeHas open-collector bus drivers.
TE0
Extended TalkerDoes not have extended talker capability.
LE0
Extended ListenerDoes not have extended listener capability.
IEEE-488 GPIB Interface
Indicators
The bottom line of the power meter’s display shows indicators describing the status of the Bird 4421 when used
with the IEEE interface. These are:
REMOTE — When REMOTE is displayed, the power meter is being controlled through the interface.
Measurements, units of measure, and certain other parameters may be changed from a remote location.
LISTEN — When LISTEN is displayed, the power meter is receiving data.
TALK — When TALK is displayed, the power meter is transmitting data. This is always shown when the unit is in
“talker-only” mode.
LOCAL LOCKOUT — When LOCAL LOCKOUT is displayed, the push buttons are disabled and the power meter’s
functions are being remotely controlled.
Setup
Dip Switch
Set the interface to ADDR (Addressable) by positioning DIP switch 1 to ON. This makes the 4421 respond to
controller commands.
Note:There are several button styles on DIP switches (slide, rocker, lever). Examine the DIP switch to
determine the ON and OFF positions.
Set the primary address using the DIP switches. The primary address is factory set to 6, but can be set to any value
between 1 and 31 (0 is reserved for the controller) To set the primary address, turn OFF switches 4 – 8 so that the
sum of the bits turned off equals the desired primary address. In
off switches have values of 4 and 2).
Note:The primary address is used by the controller to refer to specific devices on the bus. When
programming the controller, the address in the
program must be the same as the address set on
the interface module. Each device on the bus must
have a different primary address.
Talker-Only Mode
The Bird 4421 can be set up for manual operation while automatically sending data to an output device (Talker-Only
Mode). To do so, turn DIP switch 1 OFF and cycle the power. TALK will be displayed.
In Talker-Only mode, pressing a button on the meter triggers a measurement. When the measurement is complete,
the information is sent to the bus and LISTEN turns on momentarily. A listen-only device on the bus, such as a
printer, can read the value. The power meter is then ready to accept another button press.
Command Syntax
The Bird 4421 accepts two types of commands. General bus commands are commands, such as Device Clear (DCL),
that apply to any IEEE interfaced device. Device-dependent commands are specific to the 4421.
If an invalid command is sent to the unit, an error condition is placed in the serial poll byte and the offending
command is not executed.
A group of device-dependent commands can be sent as a single string as long as like command categories are not
repeated, for example: “PNFCFDT3TRG”. This string sets up the 4421 to send no prefixes, read forward dBm, make
one reading on “TRG”, and triggers a measurement.
Note: Commands can be entered in either upper or lower case.
Note: Only the last command entered of each category will be executed. As a command string is pro-
cessed by the 4421, each category of command is stored in a separate location. Two commands of the
same category will be stored in the same location, so that the second will overwrite and erase the first
one.
General Bus Commands
The general bus commands supported by the IEEE-488 interface feature are listed in
executing general commands varies among controllers; check the documentation supplied with your controller for
the proper command structure.
Table 5 IEEE-488 General Bus Commands
CommandEffect on Bird 4421
IDN?Gives product identification
RENGoes into remote mode when next addressed
GTLCancels remote mode, restores local operation
LLOLocks out local operation
IFCGoes into talker and listener-idle status
DCLReturns to default conditions
SDCReturns to default conditions
GETTriggers reading in T2 and T3 modes
SPE, SPDPuts the status byte on the bus
13
Table 5
. The syntax for
Page 25
IDentificatioN(IDN?)
IEEE-488 GPIB Interface
Function
Remarks
Product identifies itself
Replies command same as U2
Remote ENable (REN)
Function
Remarks
Enables remote operation.
The unit must be addressed to listen after setting REN true.
The REMOTE indicator turns on when this command is received.
Go To Local (GTL)
Function
Remarks
Returns device to local operation.
Issuing a GTL command while the device is in Local Lockout mode does not clear the
lockout condition.
The REMOTE indicator turns off.
The LISTEN indicator remains on.
Local LOckout (LLO)
Function
Remarks
Disables local operation of all devices on the bus.
REN must be true to use LLO.
LLO is cleared by setting REN false.
InterFace Clear (IFC)
Function
Remarks
Terminates all bus activity and passes control to the system controller.
All devices are set to talker and listener idle states.
LISTEN mode is canceled, and its indicator is turned off.
14
Page 26
Thruline RF Power Meter, Model 4421
Device CLear (DCL)
Function
Remarks
Table 6 IEEE-488 Default Conditions
F
Auto Range ONRYY
Two Terminators (CR LF)YT
Prefixes YESPY
Trigger One Shot on Talk AddressT1
All SRQ’s OFFM00
Send EOI with last byte of messageKY
Resets the status of all devices to an initialized state.
Does not change the current interface mode.
The 4421 returns to the factory default condition listed in
Default ConditionRelated Command
orward Carrier WaveFC
Selective Device Clear (SDC)
Function
Remarks
Resets the status of a selected device to an initialized state.
Only the device addressed will be cleared.
The 4421 returns to the factory default condition.
Table 6
.
Group Execute Trigger (GET)
Function
Remarks
15
Initiates a measurement for all devices set to trigger on GET.
The 4421 must be already set to trigger on GET.
Used to synchronize measurements of multiple instruments.
Page 27
Serial Polling Enable/Disable (SPE/SPD)
IEEE-488 GPIB Interface
Function
Enables or disables the serial polling sequence.
During remote operation, periodically monitor the bus service request line. Failure to
detect a service request could result in equipment damage.
Remarks
The SPE command puts all devices in serial poll mode waiting to be addressed.
The SPD command clears the SRQ bit (bit 6) and ends the polling sequence.
When addressed a device sends its status byte to the controller. A value of 1 for a bit
means tha
t the device condition that bit refers to is true. A value of 0 means that the
condition is false.
The 4421 does not use all bits of the status byte.
a description and how to reset them.
Table 7 IEEE Status Byte Description
BitNameCondition
Set if a service request is generated by the 4421. If an SRQ
has been received by the controller and this bit is cleared,
6
SRQ
other instruments on the bus should be checked to
determine where the SRQ occurred.
Cleared by a serial poll of the 4421.
Measurement
3
Complete
Set when the power meter has completed a reading.
Cleared by requesting a reading over the bus.
Set when the RF power is underrange and a reading has
2
Reading Underflow
been completed.
Cleared by requesting a reading over the bus.
Set when the RF power is overrange and a reading has been
1
Reading Overflow
completed.
Cleared by requesting a reading over the bus.
Set if an illegal device-dependent command (IDDC) or illegal
device-dependent command option (IDDCO) was received,
0
Error
or if the power meter fails the self test.
Cleared by reading the U1 status word. The U1 word
contains details on the error, see
page 21
CAUTION
.
Table 7
lists the bits used, along with
"Status (Ux)" on
16
Page 28
Thruline RF Power Meter, Model 4421
Device Dependent Commands
The device-dependent commands used by the 4421 Power Meter are listed in
Note: The programming card also has a complete command list.
Table 8 IEEE-488 Device Dependent Command Summary
CategoryCommandDescription
FCForward carrier wave
FDForward dBm
RCReflected carrier wave
Measurement
Range
Terminators
Prefixes
Triggers
Serial Polling
Status
Self-TestJ0Run self-test
EOI Response
Writable StoreWXXXXXXPlace XXXXXX in RAM
RDReflected dBm
SWStanding wave ratio
RLReturn loss
MNMinimum value
MXMaximum value
RYYAuto range on
R00 to R17Manual ranges
RNNAuto range off, stay at present range
YTTwo terminators: CR, LF
YOOne terminator: CR
YNNo terminator
PYPrefix YES
PNPrefix NO
T0Continuous on TALK
T1One shot on TALK
T2Continuous on GET
T3One shot on GET
T4Continuous on measurement command
T5One shot on measurement command
M00Do not generate SRQ
M01Generate SRQ on error
M02Generate SRQ on measurement overrange
M04Generate SRQ on measurement underrange
M08Generate SRQ on operation complete
U0Send back current machine state
U1Send back error conditions
U2Send back revision levels
KYSend EOI on last byte
KNDo not send EOI on last byte
Table 8
, organized by category.
17
Page 29
Forward Carrier Wave (FC)
Forward dBm (FD)
Reflected Carrier W
ave (RC)
Reflected dBm (RD)
IEEE-488 GPIB Interface
Function
Remarks
Selects forward or reflected RF power measurement mode.
Measurement results are returned in Watts or dBm.
Standing Wave Ratio (SW)
Return Loss (RL)
Function
Remarks
Selects SWR or return loss match measurement mode.
Measurement results are returned in VSWR or dB.
MiNimum Value (MN)
MaXimum Value (MX)
Function
Remarks
Selects minimum or maximum measurement mode.
Another measurement must be selected before selecting min or max.
Returns the minimum (or maximum) value of the previous measurement type.
18
Page 30
Thruline RF Power Meter, Model 4421
Range (Rxx)
Function
Remarks
Selects a measurement range listed in
If the selected range is outside the range of the connected power sensor, the command
is ignored.
Turn auto range on
Turn auto range off. Keep present range
18.0 – 199.9 MW
1.80 – 19.99 MW
0.180 – 1.999 MW
18.0 – 199.9 kW
1.80 – 19.99 kW
0.180 – 1.999 kW
18.0 – 199.9 W
1.80 – 19.99 W
0.180 – 1.999 W
18.0 – 199.9 mW
1.80 – 19.99 mW
0.180 – 1.999 mW
18.0 – 199.9 μW
1.80 – 19.99 μW
0.180 – 1.999 μW
18.0 – 199.9 nW
1.80 – 19.99 nW
0.180 – 1.999 nW
Terminators (Yx)
Function
Selects the characters that follow the end of a data string. Set x to:
“T” for two terminators; a carriage return (CR) and a line feed (LF).
“O” for one terminator; a carriage return (CR).
“N” for no terminator; message can be terminated by EOI.
Remarks
19
Many controllers use the terminator sequence to recognize the end of an input string.
Using incorrect terminators can lock the bus.
Page 31
Prefixes (Px)
IEEE-488 GPIB Interface
Function
Turns the prefix mode on or off. Set x to:
“Y” to enable prefixes.
“N” to disable prefixes.
Remarks
Prefixes are sent over the bus with the measurement, and indicate the status of the current
measurement (see
“FC”,”FD”,”RC”,”RD”,”SW”,”RL”,”MN”,”MX” indicates the measurement type.
“U” indicates underflow; the value sent is “.000”.
“O” indicates overflow; the value sent is “199.9”.
“N” indicates normal; the value sent is a normal on-scale reading.
“4421” indicates the Bird model number.
Table 10 Prefix Examples
Data StringDescription
NFC.0.123W(CR)(LF)Normal (N) forward carrier wave (FC), prefixes on
OFC 199.9W(CR)(LF)Overflowed (O) forward carrier wave (FC), prefixes on
199.9W(CR)(LF)Overflowed forward carrier wave, prefixes off
URD .000W(CR)(LF)Underflowed (U) reflected dBm (RD), prefixes on
Table 10
for examples):
Triggers (Tx)
Function
Remarks
Table 11 Trigger Conditions
Selects the condition which will trigger a reading (see
Failure to trigger device before requesting a reading will lock the bus.
T1 halts the bus until a reading is available.
T0 and T1 do not set the measurement complete SRQ.
T2, T3, T4, and T5 set a SRQ when the measurement is complete.
Fastest reading rate is 2.4 readings/second; slowest is 1 reading/sec.
One shot on measurement command (FC, FD,
RC, RD, SW, RL, MN, MX)
).
20
Page 32
Thruline RF Power Meter, Model 4421
SRQ Mask (Mxx)
Function
Remarks
Table 12 SRQ Mask Bits
Selectively masks status bits to prevent unwanted service requests.
Set xx to the sum of the binary values of the desired SRQ trigger bits.
For example, M12 would set the SRQ for both operation complete and underrange
alues 8 and 4). Set xx to “00” to never generate an SRQ.
(v
If a status bit is masked (bit set to 0), SRQs won’t be generated for that condition.
Binary ValueBit NumberMessage
10 (LSB)
21
42
83
Status (Ux)
4
5
6
7 (MSB)
Error (IDDC, IDDCO, self-test fail)
Measurement overrange
Measurement underrange
Operation complete
Not Used
Not Used
Can’t mask
Not Used
Function
Reads a status word and returns the information as a string. Set x to:
“0” for machine status.
“1” for error status.
“2” for revision history.
Remarks
Figure 9 Machine Status Word Format
After sending the status command, a status word is sent the next time the unit is
addressed to talk. To ensure the correct status is transmitted, the status word should be
requested as soon as possible after the command is sent.
Machine Status Word (U0) —
Figure 9
. The default values are also shown.
The format of the machine status word is shown in
21
Page 33
Error Status Word (U1) — The format of the error status word and the possible
error messages are shown in
in the status (serial poll) byte, and a SRQ may be generated (
on page 21
). All flags will revert to their non-error states after the U1 command is
sent.
Figure 10 Error Status Word Format
StatusMeaningDescription
ICMInvalid CommandSet when an illegal device-dependent command
VCMValid CommandSet when no IDDC is received.
ICOInvalid Command OptionSet when an illegal device-dependent command
VCOValid Command OptionSet when no IDDCO is received.
PSSelf-Test PassSet when a self-test has been initiated by the
FLSelf-Test FailSet when the self-test has failed. (This is the
IEEE-488 GPIB Interface
Figure 10
. When an error occurs, an error is also flagged
"SRQ Mask (Mxx)"
(IDDC) such as V2 is received. (V is illegal)
option (IDDCO) such as T6 is received. (6 is illegal)
J0 command and the test result is acceptable.
default condition.)
Revision History Word (U2) — The format of the revision history word is shown in
Figure 11
.
Figure 11 Revision History Word Format
Note: If Writable stores parameter has not been set, this command responds only with -4420-.
22
Page 34
Thruline RF Power Meter, Model 4421
Self Test (J0)
Function
Remarks
Initiates a hardware and software test.
Results are stored in the U1 status word (
“J0” must be sent each time before reading the result.
End Or Identify (Kx)
Function
Enables or disables the End or Identify (EOI) signal. Set x to:
“Y” to enable.
“N” to disable.
Remarks
Writable Store (W x x x x x x)
Function
Remarks
Disabling EOI can cause some controllers to lock unless another terminator is used.
When enabled, EOI is only asserted at the end of a multiple byte string.
Storage for six bytes of ASCII data.
Data stored is lost when the 4421 is turned off.
Data is sent back as part of the U2 status word.
"Status (Ux)" on page 21
).
23
Page 35
Chapter 5RS-232 Interface
This chapter discusses setup of the RS-232 interface feature and describes the RS-232 commands that apply to the
Bird 4421. Operators should understand EIA Standard RS-232-C and have basic computer programming skills before
writing any programs.
Description
The Bird 4421 RS-232 interface feature is an integral part of the I/O hub circuit board inside of the mm21. An eightposition DIP switch is used to set operational conditions such as baud rate, parity, and stop bits. The bottom line of
the display indicates the current bus status.
Cable Connector
The interface uses a standard 25-pin RS-232 connector. Pin assignments are listed in
uses a different wiring arrangement, do not attempt to rewire the interface module’s connector. A null modem kit
should be used for rewiring instead.
Table 13 RS-232 Pin Assignments
PinDesignationNotes
1Protection GroundChassis Ground
2Transmit Data
3Receive Data
4Request to Send(Output) Set true after module power up
5Clear to Send
6Data Set Ready(Input) Set internally true by module
7Signal GroundReturn path for data and control signals
8Receive Signal DET(Input) Set true by module
20Data Terminal Ready(Output) Set true after module power up
Indicators
(Input) Set by input device. When true, it enables the
module to transmit. When false, it disables transmission.
Table 13
. If the controller
The bottom line of the power meter’s display shows indicators describing the status of the Bird 4421 when used
with the RS-232 interface. These are:
TALK — When TALK is displayed, the power meter is transmitting data. This is always shown when the unit is in
“talker-only” mode.
LISTEN — When LISTEN is displayed, the power meter is receiving data.
24
Page 36
Thruline RF Power Meter, Model 4421
Setup
DIP Switch
Set the DIP switches according to application needs and the requirements of the controller. Available settings and
factory defaults are listed in
Note:There are several button styles on DIP switches (slide, rocker, lever). Examine the DIP switch to
determine the ON and OFF positions.
Note:If you change switch settings when the unit is ON, the new settings will not become effective
until power is turned OFF and then back ON.
SwitchFunctionDescription
1Stop Bit
2Command Mode
3Word Length
4,5Parity
6,7,8Baud Rate
Figure 12
.
ON
OFF
ON
OFF
ON
OFF
(4)
ON
ON
OFF
OFF
(6)
ON
ON
ON
OFF
OFF
OFF
OFF
1 Stop Bit
2 Stop Bits
†
ENT command needed before sending reading to controller†
Trigger automatically sends reading
Reading also sent when front panel push button pressed
8 Data Bits†
7 Data Bits
(5)
ON
OFF
ON
OFF
(7)
ON
OFF
OFF
ON
ON
OFF
OFF
No Parity†
Odd Parity
Even Parity
Mark Parity
(8)
ON
ON
OFF
ON
OFF
ON
OFF
Auto Baud††
300
600
1200
2400†
4800
9600
† Default factory setting
†† For more information see “Auto Baud”
25
Page 37
RS-232 Interface
Auto Baud
Auto Baud is used to automatically determine the correct transmission rate. After setting DIP switches 1 through 5,
follow the steps below to use auto baud:
1.Set DIP switches 6, 7, and 8 to ON.
2.Connect a controller to the power meter.
3.Turn the power meter on.
4.Send the character U (hexadecimal 55) from the controller to the power meter.
5.Wait 1 second.
6.If the power meter’s front panel displays LISTEN, the baud rate has been determined. If not, repeat steps 4 and
5.
If, after 30 seconds, the module cannot absolutely determine a baud rate, it will choose a rate based on data
acquired during the test.
Note:Auto Baud is the only automatically chosen setting. Other items such as parity and stop bits
must be manually selected.
Talker-Only Mode
The Bird 4421 can be set up for manual operation while automatically sending data to an output device (Talker-Only
Mode). To do so, turn DIP switch 2 OFF and cycle the power. TALK will be displayed.
In Talker-Only mode, pressing a button on the meter triggers a measurement. When the measurement is complete,
the information is sent to the bus.
Command Syntax
The Bird 4421 accepts two types of commands. General bus commands are commands, such as Initialize (INT), that
apply to any RS-232 interfaced device. Device-dependent commands are specific to the 4421.
If an invalid command is sent to the unit, an error condition is placed in the serial poll byte and the offending
command is not executed.
A group of device-dependent commands can be sent as a single string as long as like command categories are not
repeated. For example: “PNFCFDT3TRG”. This string sets up the 4421 to send no prefixes, read forward dBm, make
one reading on “TRG”, and triggers a measurement.
Note: Commands can be entered in either upper or lower case.
Note: Only the last command entered of each category will be executed. As a command string is pro-
cessed by the 4421, each category of command is stored in a separate location. Two commands of the
same category will be stored in the same location, so that the second will overwrite and erase the first
one.
General Bus Commands
The general bus commands supported by the RS-232 interface module are listed in
Table 14 RS-232 General Bus Commands
CommandEffect on Bird 4421
INT
ENT
TRG
B1 to B7
XO/XF
Returns to default conditions
Sends a reading to the controller
Triggers reading in T3 mode
Selects a baud rate
Enables/disables software handshake
Table 14
.
26
Page 38
Thruline RF Power Meter, Model 4421
INiTialize (INT)
Function
Remarks
Resets the Bird 4421 and returns it to the factory defaults.
If INT is linked with any other command within a string, it must be separated from that
command by a space.
ENTer (ENT)
Function
Remarks
Makes the power meter transmit a reading to the controller.
A measurement must have already been triggered, placing a reading in the output
buffer.
To send a reading whenever a measurement is triggered, set DIP switch 2 to OFF. The
ENT c
TRiGger (TRG)
Function
Initiates a measurement if the power meter is in trigger mode (T3).
Baud Select (Bx)
Function
Remarks
Selects a baud rate listed in
When the meter recognizes a valid Bx command, its baud rate is immediately changed.
(The controller is assumed to be transmitting at the new rate; otherwise sending
commands would not be possible.)
The device-dependent commands used by the 4421 Power Meter are listed in
Table 16 RS-232 Device Dependent Command Summary
CategoryCommandDescription
Measurement
Range
Terminators
Prefixes
Triggers
Status
Self-Test
Writable Store
When data is being sent from the power meter to the computer and flow control is
enabled, data transmission will be suspended when the XOFF character (hexadecimal
13) is sent by the computer. Transmission will resume when XON (hex 11) is sent by the
computer.
When data is being sent from the computer to the power meter, XOFF will be sent to
omputer when the input buffer fills up. XON will be sent to the computer when the
the c
buffer has emptied.
FCForward carrier wave
FDForward dBm
RCReflected carrier wave
RDReflected dBm
SWStanding wave ratio
RLReturn loss
MNMinimum value
MXMaximum value
RYYAuto range on
R00 to R17Manual ranges
RNNAuto range off, stay at present range
YTTwo terminators: CR, LF
YOOne terminator: CR
YNNo terminator
PYPrefix YES
PNPrefix NO
T0Continuous on ENT
T1One shot on ENT
T3One shot on TRG
T5One shot on measurement command
U0Send back current machine state
U1Send back error conditions
U2Send back revision levels
J0Run self-test
WXXXXXXPlace XXXXXX in RAM
Table 16
, organized by category.
28
Page 40
Thruline RF Power Meter, Model 4421
Forward Carrier Wave (FC)
Forward dBm (FD)
Reflected Carrier W
ave (RC)
Reflected dBm (RD)
Function
Remarks
Selects forward or reflected RF power measurement mode.
Measurement results are returned in Watts or dBm.
Standing Wave Ratio (SW)
Return Loss (RL)
Function
Remarks
Selects SWR or return loss match measurement mode.
Measurement results are returned in VSWR or dB.
MiNimum Value (MN)
MaXimum Value (MX)
Function
Remarks
Selects minimum or maximum measurement mode.
Another measurement must be selected before selecting min or max.
Returns the minimum (or maximum) value of the previous measurement type.
29
Page 41
Range (Rxx)
RS-232 Interface
Function
Remarks
Selects a measurement range listed in
If the selected range is outside the range of the connected power sensor, the command
is ignored.
Table 17 Measurement Ranges
CommandPower Range
RYY
RNN
R17
R16
R15
R14
R13
R12
R11
R10
R09
R08
R07
R06
R05
R04
R03
R02
R01
R00
Table 17
.
Turn auto range on
Turn auto range off. Keep present range
18.0 – 199.9 MW
1.80 – 19.99 MW
0.180 – 1.999 MW
18.0 – 199.9 kW
1.80 – 19.99 kW
0.180 – 1.999 kW
18.0 – 199.9 W
1.80 – 19.99 W
0.180 – 1.999 W
18.0 – 199.9 mW
1.80 – 19.99 mW
0.180 – 1.999 mW
18.0 – 199.9 μW
1.80 – 19.99 μW
0.180 – 1.999 μW
18.0 – 199.9 nW
1.80 – 19.99 nW
0.180 – 1.999 nW
Terminators (Yx)
Function
Selects the characters that follow the end of a data string. Set x to:
“T” for two terminators; a carriage return (CR) and a line feed (LF).
“O” for one terminator; a carriage return (CR).
“N” for no terminator.
Remarks
Many controllers use the terminator sequence to recognize the end of an input string.
Using incorrect terminators can lock the bus.
30
Page 42
Thruline RF Power Meter, Model 4421
Prefixes (Px)
Function
Turns the prefix mode on or off. Set x to:
“Y” to enable prefixes.
“N” to disable prefixes.
Remarks
Prefixes are sent over the bus with the measurement, and indicate the status of the
current measurement (see
“FC”,”FD”,”RC”,”RD”,”SW”,”RL”,”MN”,”MX” indicates the measurement type.
“U” indicates underflow; the value sent is “.000”.
“O” indicates overflow; the value sent is “199.9”.
“N” indicates normal; the value sent is a normal on-scale reading.
“4421” indicates the Bird model number.
Table 18 Prefix Examples
Data StringDescription
NFC.0.123W(CR)(LF)Normal (N) forward carrier wave (FC), prefixes on
OFC 199.9W(CR)(LF)Overflowed (O) forward carrier wave (FC), prefixes on
199.9W(CR)(LF)Overflowed forward carrier wave, prefixes off
URD .000W(CR)(LF)Underflowed (U) reflected dBm (RD), prefixes on
Table 18
for examples):
Triggers (Tx)
Function
Remarks
Table 19 Trigger Conditions
Selects the condition which will trigger a reading (see
Failure to trigger device before requesting a reading will lock the bus.
Fastest reading rate is 2.4 readings/second; slowest is 1 reading/sec.
CommandTrigger Condition
T0
T1
T3
T5
Table 19
Continuous on ENT
One shot on ENT
One shot on TRG
One shot on measurement command (FC, FD,
RC, RD, SW, RL, MN, MX)
).
31
Page 43
Status (Ux)
RS-232 Interface
Function
Reads a status word and returns the information as a string. Set x to:
“0” for machine status.
“1” for error status.
“2” for revision history.
Remarks
Figure 13 Machine Status Word Format
Note:This command sets the unit to the settings that were last issued remotely (through RS-232 or
GPIB). Therefore, it will over ride any manual changes selected at the front panel of the unit.
After sending the status command, a status word is sent the next time the unit is
addressed to talk. To ensure the correct status is transmitted, the status word should be
requested as soon as possible after the command is sent.
Machine Status Word (U0) — The format of the machine status word is shown in
Figure 13
. The default values are also shown.
Error Status Word (U1) — The format of the error status word and the possible
error messages are shown in
after the U1 command is sent.
Figure 14 Error Status Word Format
StatusMeaningDescription
ICMInvalid CommandSet when an illegal device-dependent command
VCMValid CommandSet when no IDDC is received.
ICOInvalid Command
Option
VCOValid Command OptionSet when no IDDCO is received.
PSSelf-test PassSet when a self-test has been initiated by the J0
FLSelf-test FailSet when the self-test has failed. (This is the
Figure 14
(IDDC) such as V2 is received. (V is illegal)
Set when an illegal device-dependent command
option (IDDCO) such as T6 is received. (6 is
illegal)
command and the test result is acceptable.
default condition.)
. All flags will revert to their non-error states
32
Page 44
Thruline RF Power Meter, Model 4421
Revision History Word (U2) — The format of the revision history word is shown in
Figure 15
Figure 15 Revision History Word Format
.
Self Test (J0)
Function
Remarks
Initiates a hardware and software test.
Results are stored in the U1 status word (see
“J0” must be sent each time before reading the result.
Writable Store (W x x x x x x)
Function
Remarks
Storage for six bytes of ASCII data.
Data stored is lost when the 4421 is turned off.
Data is sent back as part of the U2 status word
"Status (Ux)" on page 32
).
33
Page 45
Chapter 6Maintenance
This chapter describes routine maintenance, along with troubleshooting instructions for the power meter and
power sensor. Disassembly instructions for the Bird Cal Cart are also provided. For service beyond this level, return
the unit to a qualified service center.
WARNING
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement or changing the line voltage
setting.
WARNING
Never attempt to connect or disconnect RF equipment from the transmission line
while RF power is being applied.
Leaking RF energy is a potential health hazard.
The Bird 4421 Power Meter requires only simple, routine maintenance.
Wipe off dust and dirt regularly.
Check the connectors and cables for damage.
Clean the connector contacts with alcohol or dry cleaning solvent.
Cleaning
CAUTION
Do not use harsh or abrasive detergents for cleaning.
Clean the Bird 4421 Power Meter and its display with a soft cloth dampened with mild detergent and water only.
Clean sensors with a dry cleaning solvent that leaves no residue.
Troubleshooting
Since the power meter and power sensor can only work together, the first step is to determine which is
malfunctioning. Connect the power sensor to the meter and perform the
power meter is malfunctioning, refer to the troubleshooting table below. If the power sensor is malfunctioning,
return it for service.
CAUTION
Due to the complexity of the Bird Power Sensor, field repairs beyond general
maintenance should not be attempted.
Removal or disturbance of the power sensor cover can result in cancellation of lifetime
warranty.
This manual cannot list all malfunctions that may occur, or corrective actions. If a malfunction is not listed or is not
corrected by the listed corrective actions, contact a qualified service center.
Note:For power sensors in Cal Carts, following the test procedure will check the sensor not covered
by the safety cover. To test the other sensor slide the safety cover over, then proceed as normal.
"Functional Test" on page 35
. If the
34
Page 46
Thruline RF Power Meter, Model 4421
PROBLEMPOSSIBLE CAUSECORRECTION
Have the batteries been charged? Recharge the batteries.
Power meter has no power
Is the power meter’s ac power
cord connected?
Is the ON/OFF rocker switch on
the rear panel set to OFF?
(Cal Cart only)
Is the Cal Cart’s AC power cord
Connect the power cord.
Set the switch to ON.
Connect AC power.
connected to the AC line?
Check fuse rating and replace fuse
Blown fuse?
(See
"Replacing Fuses" on
page 37
).
Is the sensor cable connected to
Dash moves across the display
both the power meter and power
sensor?
Connect sensor cable.
Is the sensor cable defective?Replace sensor cable.
Display blank or not updatingHave the batteries been charged?
Power meter turns off while on
battery power
Push buttons do not respond
Is “LO BAT” displayed?
Test the push buttons (See
Button Test" on page 36
"Push
). Are
NO: Recharge battery
YES: Return meter for service.
YES: Recharge battery
NO: Return meter for service.
Return meter for service.
they defective?
Every segment on the display is litReturn meter for service.
IEEE-488 Interface does not
respond to the interface link
Are the address in the interface
program and the address setting
on the DIP switches on the
interface module the same?
Send the “J0” self test command,
then check the status. Does the
Change the program or DIP switch
setting so that the addresses are
the same (See
page 12
).
Replace I/O hub PCB.
power meter fail the self test?
Is the IEEE cable defective?Replace IEEE interface cable.
RS-232 Interface does not respond
to the interface link. Fails J0 selftest command.
Are the DIP switches set correctly? Set DIP switches.
Is the RS-232 cable defective?Replace RS-232 interface cable.
For “talk/listen” operation, set
With the RS-232, the power meter
displays “TALK” but does not
operate as expected.
Is DIP switch 2 set correctly?
Is DIP switch 1 set correctly?
this switch to ON. For “talk
always” set this switch to OFF.
Set DIP switch 1 as indicated in
Figure 12 on page 25
"Dip Switch" on
.
Functional Test
1.Turn the power meter off. The ON/OFF switch on the rear panel should be OFF and the ac power cable should
be connected.
2.Turn ON the ON/OFF switch on the rear panel of the meter.
3.While holding down the FWD and SWR push buttons, press the ON/OFF button on the front panel of the power
meter. Immediately release all three.
35
Page 47
Maintenance
Note:The unit tests the display on power up cycling through and activating each segment two times
and then activating all segments at once two times.
4.The power meter’s model number and revision date should scroll across the display. If a dash “–” is displayed
instead, then the meter is malfunctioning.
5.The power sensor’s model number and revision date should scroll across the display. If a dash is displayed after
the power meter data, then the power sensor is malfunctioning.
Figure 16 Test Display, No Malfunction
Push Button Test
Note: If a push button is malfunctioning, return the power meter.
1. Disconnect the power sensor.
2. Turn the power meter ON.
Note: After the power up display disappears, a dash “–” should scroll across the display.
3. Turn the power meter OFF.
4. Connect the power sensor.
5. Turn the power meter ON.
Note: “AUTO” and “FWD” should be displayed, and a reading of “0.00 W” (or a very low number).
6. Press RFL.
Note: “FWD” should change to “RFL” on the display. The reading should remain the same.
7. Press SWR.
Note: “RFL” should change to “SWR”. “0.00 W” should change to “” (underrange error).
8. Hold down MIN.
Note: “” should change to “” (overrange error).
9. Release MIN.
Note: “” should change to “”.
10. Press dBm.
Note: “SWR” should turn OFF and “dBm” should turn ON.
36
Page 48
Thruline RF Power Meter, Model 4421
11. Press dBm.
Note: “dBm” should change to “SWR”.
12. Press FWD.
Note: “SWR” should change to “FWD” and “ ” to “.0000 W” (or a very low number).
13. Hold down MAX.
Note: “.0000” should change to “1.0000”.
14. Release MAX.
Note: “1.0000” should change to “.0000” (or a very low number).
15. Press LIGHT.
Note: The back-light should turn on.
16. Press LIGHT.
Note: The back-light should turn off.
17. Press s (up).
Note: The power meter should change ranges each time it is pressed until it reaches “0.000 KW”.
18. Press t (down).
Note: The power meter should change ranges each time it is pressed until it reaches “0.00 W”.
19. Turn the power meter OFF.
Power Meter Repair
WARNING
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement or changing the line voltage
setting.
Replacing Fuses
WARNING
To avoid personal injury, disconnect the power cord from the ac line before
performing any maintenance, including fuse replacement.
CAUTION
Failure to install the properly rated fuse may result in equipment damage or nuisance
failures.
1. Gently pry the fuse drawer out of the AC module. The fuse holder does not detach from
the AC module. See
Figure 17
.
2. Install the replacement fuses then close and secure the fuse holder. Fuse 1 is a spare
fuse and fuse 2 is the active fuse.
37
Page 49
AC Line VoltageFuse Rating
115 VacT630 mA, 5x20 mm Time Lag Fuse
230 VacT315 mA, 5x20 mm Time Lag Fuse
Figure 17 AC Line Fuse
Maintenance
38
Page 50
Thruline RF Power Meter, Model 4421
Replacing Batteries
There are two battery configurations in the 4421 Power Meter. In the larger case models (before 2016), the
batteries are horizontally arranged, in units with a smaller case the batteries are vertically arranged in a battery
assembly.
Replace Batteries (Older models)
WARNING
Exposed AC line voltage (115 VAC or 230 VAC). Disconnect the power cord from the
AC line before replacing the batteries. Failure to comply may result in severe
electrical shock or death.
WARNING
Use only Nickel Metal Hydride (NiMH) batteries that have a minimum capacity of 4500
milliampere hours (mAh). Do not install batteries that are not Nickel Metal Hydride.
Do not install NiMH batteries that have less than 4500 mAh capacity. Failure to
comply may result in damage to the batteries, damage to the instrument, and injury
to personnel from battery chemicals.
1.Disconnect the AC power cord from the AC mains supply.
2.Remove the four screws that secure the top cover.
3.Lift the top cover to access the attached ground wire.
4.Disconnect the ground wire.
5.Remove the cover.
6.Unfasten the battery retaining belts (
Figure 18
).
7.Remove the battery tubes.
Note: Be sure to note the polarity and orientation of the battery tubes before removing them.
8.Remove the batteries from each tube.
9.Insert replacement batteries.
Note: Be sure to note the polarity (positive and negative) arrangement of the batteries.
10. Install the battery tubes into the unit and secure with the retaining belts.
Note: Be sure to position the retaining belts as they were before removal and tighten them securely.
11. Connect the ground wire to the top cover.
12. Install the top cover.
13. Secure top cover with the four screws removed earlier.
39
Page 51
Figure 18 Batteries
Maintenance
ItemDescription
1Battery retaining belt
2
Battery tube (batteries
inside)
Replace Batteries (Newer Models)
WARNING
Exposed AC line voltage (115 VAC or 230 VAC). Disconnect the power cord from the
AC line before replacing the batteries. Failure to comply may result in severe
electrical shock or death.
WARNING
Use only Nickel Metal Hydride (NiMH) batteries that have a minimum capacity of 4500
milliampere hours (mAh). Do not install batteries that are not Nickel Metal Hydride.
Do not install NiMH batteries that have less than 4500 mAh capacity. Failure to
comply may result in damage to the batteries, damage to the instrument, and injury
to personnel from battery chemicals.
1.Disconnect the AC power cord from the ac mains supply.
2.Remove the four screws that secure the top cover.
3.Lift the top cover to access the attached ground wire then disconnect the ground wire and remove the cover.
4.Remove two retaining screws securing the battery assembly. See
Figure 19
.
5.Disconnect battery assembly power supply connector.
40
Page 52
Thruline RF Power Meter, Model 4421
Figure 19 Battery Assembly Removal
6.Lift the battery assembly out of the Power Meter.
7.Unfasten the battery retaining belts then remove the battery tubes (
Figure 20
). Be sure to note the polarity
and orientation of the battery tubes before removing them.
Figure 20 Battery Assembly
8.Remove the batteries from each tube and insert replacement batteries. Be sure to note the polarity (positive
and negative) arrangement of the batteries.
9.Install the battery tubes into the battery assembly and secure with the retaining belts. Be sure to position the
retaining belts as they were before removal and tighten them securely.
10. Install battery assembly in the Power Meter. Align tabs on battery assembly with slots in Power Meter.
11. Install two screws to secure battery assembly to Power Meter.
12. Connect the battery assembly power supply connector to the main board.
13. Connect the ground wire to the top cover.
14. Install the top cover and secure it with the four screws removed earlier.
41
Page 53
Maintenance
Long Term Storage
Do not store the instrument for long periods of time without recharging the batteries. When the instrument is
stored for long periods of time without use, the batteries will loose their charge and also loose the ability to reach
full charge when put into service. To restore the battery charge, perform a full charge for 28 hours. If the batteries
are not fully charged after 28 hours, completely discharge the batteries then perform a full charge again. If
necessary, repeat this discharge and charge cycle up to three times. If the batteries do not remain charged after
three discharge-charge cycles, replace the batteries.
Cal Cart Repair
It is recommended that you reference the specific cal cart manual for procedures tailored to each Cal Cart type.
Front Panel (Cal Cart only)
Note: If the power meter is mounted in a Cal Cart, to access the power sensor or the ac connections, it
will be necessary to remove the front panel of the Cal Cart.
1.Remove both screws on the front panel, between the RF connector and the handle.
2.Pull on the handle set into the base of the front panel to remove it.
Note: When making AC connections, refer to the following wiring diagram.
Cord Reel (Cal Cart only)
1.Disconnect the cord reel from the terminal strip.
2.Unscrew the AC connector.
3.Unscrew and remove the cord reel.
4.Screw the replacement cord reel into place.
5.Thread the AC connector through the grommet, and connect the wires on the other end to the terminal strip
(See
Figure 21 on page 42
Figure 21 Cal Cart Wiring Schematic
).
42
Page 54
Thruline RF Power Meter, Model 4421
Power Sensor (Cal Cart only)
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement or changing the line voltage
Never attempt to connect or disconnect RF equipment from the transmission line
while RF power is being applied. Leaking RF energy is a potential health hazard.
The Bird 4421 must be powered off when connecting or disconnecting the power
Changing the sensor’s connectors will invalidate calibration data, and may reduce the
WARNING
setting.
WARNING
CAUTION
sensor from the power meter.
CAUTION
maximum power rating of the unit.
CAUTION
Due to the complexity of the Bird Power Sensor, field repairs beyond general
maintenance should not be attempted. Removal or disturbance of the power sensor
cover can result in cancellation of lifetime warranty.
1.Disconnect the RF line from the Cal Cart.
2.Remove the Cal Cart front panel (see
3.Disconnect the sensor cable from the power sensor.
4.Disconnect the RF cable from the sensor output port.
5.Remove the screws on the sensor mounting bracket.
6.Remove the sensor from the mounting bracket.
7.Put the new sensor in the bracket and screw it into place.
"Front Panel (Cal Cart only)" on page 42
).
Note: Make sure the arrow on the side of the sensor points towards the load, and that the end labeled “SOURCE”
points towards the front of the Cal Cart.
8.Connect the RF cable to the sensor end labeled “LOAD”. Connect the sensor cable.
9.Replace the Cal Cart front panel.
Load
WARNING
Heavy load. Do not attempt to lift unaided.
1.Disconnect all RF cables from the RF connectors on the load.
2.Unplug the blower assembly on blower-equipped loads.
3.Replace the vent plugs with the shipping plugs.
4.Remove the mounting bolts connecting the load to the Cal Cart.
5.Remove the load.
43
Page 55
Casters (Cal Cart only)
Maintenance
1.Remove the load (See
2.Flip the cart over.
3.Unscrew the defective caster.
4.Screw the new caster into place and replace the load.
Handle (Cal Cart only)
1.Unscrew the handle
2.Remove the old handle.
3.Screw the new handle into place.
"Load" on page 43
).
Storage and Shipment (Cal Cart only)
Cover the unit before storing to keep out dust and dirt. It is not necessary to install the shipping plug. Store in a dry,
dust-free environment where the ambient temperature will remain between –20 and +50 °C (–4 to +122 °F).
Before shipping the Bird Cal Cart, take the following precautions:
Remove the vent plugs and replace them with the shipping plugs. Wrap the vent plugs with padding and
tape them to the side of the load for protection.
Note: With the shipping plugs installed, it is not necessary to empty out the coolant.
Repack in the original carton or contact Bird for a transit case.
44
Page 56
Thruline RF Power Meter, Model 4421
Specifications, Cal Cart
To determine the model numbers of the Cal Cart components, refer to specific cal cart manual. For load
specifications, refer to the load manual included.
Note:Certain sensors or connectors may be incompatible with some models. Refer to specific cal cart
manual for a list of available components, or contact Customer Service.
Max. Power
Under normal operating conditions, the Cal Cart’s maximum power is the MINIMUM of the max. power of the
sensor and the max. power of the load. The following modifiers apply:
Above 65 MHz, derate the load’s max. power. See
Figure 22 on page 51
.
Due to the wide variety of available connectors, frequency range and maximum power may be reduced.
Insertion loss is specified with female N connectors. Choose connectors appropriate for the frequency and
power of operation.
Derate RF power by 2.5% for every 305m (1,000 ft.) above 1,520m (5,000 ft.).
Table 20 Bird Cal Cart
Frequency RangeSensor dependent, 30 MHz max.
Power RangeDependent on sensor, sensor connectors, and load
Connectors
Dimensions
Short Cart
Long Cart
Weight, Nominal
Short Cart
Long Cart
Temperature Range
Operating
Storage
Altitude1520 m (5000 ft.)
Humidity85% noncondensing max
Calibration Cycle, Nominal
CE
Customer specified (See
page 52
42”L x 20”W x 42”H (1067 x 508 x 1067 mm)
52”L x 20”W x 42”H (1321 x 508 x 1067 mm)
175 lbs. (80 kg)
290 lbs. (114 kg)
0 to +35 °C (32 to +95 °F)
– 20 to +70 °C (– 4 °F to +158 °F)
1 year
CE Compliant. Refer to Declaration of Conformity for
specific standards.
)
"Available Connectors" on
45
Page 57
Specifications, 4421 RF Power Meter
Table 21 Bird 4421 RF Power Meter
Frequency RangeSensor dependent
Power RangeSensor dependent
VSWR Display1.0 – 199.9 max
Return Loss Display0 to 40 dB max
Display Accuracy± 1 on least significant digit
AC Power115 / 230 Vac @ 50 / 60 Hz; 60 W
Batteries8 Nickel Metal Hydride 1.2 v cells, rechargeable
Battery Life
Battery Charger
Display
Interfaces
Fuse RatingIEC (5 x 20 mm) Time Lag Type T
115 VAC
230 VAC
CE
Humidity95% max. (noncondensing)
AltitudeUp to 10,000 feet (3,000 m)
Temperature Range
Operating
Storage
Dimensions
2016 and prior
2016 and newer
Weight, Nominal9.5 lbs. (4.3 kg)
Approximately 8 hours continuous usage
(reduced after high temperature charging)
Built-in battery charger. Drained batteries require
approximately 28 hours to recharge.
LCD, 4½ digit display. Indicates mode, measurement units,
battery condition, remote status, and signal increase /
decrease. Self contained backlight.
IEEE-488 GPIB
RS-232
T630 mA
T315 mA
CE Compliant. Refer to Declaration of Conformity for
specific standards.
0 to 50 °C (32 to 122 °F)
–20 to +50 °C (–4 to +122 °F)
12.3”L x 12.2”W x 4.2”H (312 x 309 x 103 mm)
10.7”L x 12.2”W x 4.2”H (271 x 309 x 103 mm)
Maintenance
Changing the sensor’s connectors will invalidate calibration data, and may reduce the
Calibration Power, Typical3.5 kW
VSWR, Max.1.05:1
Insertion Loss, Max.0.05 dB (with female 7-16 DIN connectors)
Directivity, Min.28 dB
† For rated accuracy, no more than 1% AM; Harmonics –50 dBc or less
Derate accuracy by 2% (2
Derate accuracy by 2% (2
†† Other calibration frequencies available upon request
Specifications, RS-232 Interface
Table 27 RS - 232 Interface
Logic LevelsMeets all EIA Standard RS-232-C specifications
Modes of OperationSwitch and bus selectable
Talk AlwaysAllows the 4421 to send to the bus keyboard-initiated
AddressableAllows the 4421 to be commanded by an RS-232 interface controller
ConnectorRS-232 Interface Type
Humidity, Max95% (non-condensing)
Altitude, Max10,000 feet (3,000 m)
Temperature Range
Operating
Storage
Dimensions6.5”L x 4.5” W (165 x 115 mm)
Weight0.5 lbs (0.23 kg) nominal
measurements only
0 to 50º C (32 to 122º F)
–20 to +50º C (–4 to +122º F)
0.25, 0.40
0.40
1.8, 2.0, 2.17
2.5, 3.2, 3.5
3.5, 4.0
10.0, 13.56, 15.0
25.76, 27.12, 28.48
2.5, 3.2, 3.5, 4.0
) outside cal. power or cal. frequency
) below 15 °C and above 35 °C
Specifications, IEEE-488 GPIB Interface
Table 28 IEEE - 488 Interface
Logic LevelsMeets all IEEE Standard 488-1978 specifications
Modes of OperationSwitch and bus selectable
Talk OnlyAllows the 4421 to send to the bus keyboard-initiated measurements
AddressableAllows the 4422 to be addressed as talker or listener under the
ConnectorStandard IEEE-488 bus type
Humidity, Max95% (non-condensing)
Altitude, Max10,000 feet (3,000 m)
Temperature Range
Operating
Storage
Dimensions6.5”L x 4.5” W (165 x 115 mm)
Weight, Nominal0.5 lbs (0.23 kg)
only
command of an IEEE-488 bus controller
0 to 50º C (32 to 122º F)
–20 to +50º C (–4 to +122º F)
5 ft.
Null Modem Kit (RS-232 only)14380-250
Panel Mount Kit14421-250
Battery, C size, NiMH85A1230
Maintenance
1
1
1
1
5A2257-14
5A2257-11
5-1286
5A2416
5-1317-2
5-1317-1
5-1662-2
5-1662-1
Available Connectors
Changing the sensor’s connectors will invalidate calibration data, and may reduce the
CAUTION
maximum power rating of the unit.
ConnectorP/N
HN (F)4240-268
LC (F)4240-031
BNC-Female4240-125
BNC-Male4240-132
C-Female4240-100
C-Male4240-110
HN-Female4240-268
HN-Male4240-278
LC-Female4240-031
LC-Male4240-025
Open Term. # 10-32 Nut4240-080
N (F)4240-062
52
Page 64
Thruline RF Power Meter, Model 4421
7/16 Jack, IEC Type 169-44240-344
LT-Female4240-018
LT-Male4240-012
N-Female4240-062
N-Male4240-063
SC-Female4240-090
SMA-Female4240-336
SMA-Male4240-334
7/16 Jack, IEC Type 169-44240-344
7/16 Plug, IEC Type 169-44240-363
TRU 6934 (F)4240-371
TRU 7958 (F)4240-372
Mini UHF-Female4240-346
UHF-Female4240-050
UHF-Male4240-179
1-5/8” EIA Fixed4240-096
1-5/8” EIA Swivel4240-208
7/8” EIA4240-002
TNC-Female4240-156
TNC-Male4240-160
Customer Service
Any maintenance or service procedure beyond the scope of those in this chapter should be referred to a qualified
service center.
If the unit needs to be returned for any reason, request an Return Material Authorization (RMA) through the Bird
Technologies website. All instruments returned must be shipped prepaid and to the attention of the RMA number.
For the location of the Sales Office nearest you, visit our Web site at:
http://www.birdrf.com
53
Page 65
Limited Warranty
All products manufactured by Seller are warranted to be free from defects in material and workmanship for a
period of one (1) year, unless otherwise specified, from date of shipment and to conform to applicable specifications, drawings, blueprints and/or samples. Seller’s sole obligation under these warranties shall be to issue credit,
repair or replace any item or part thereof which is proved to be other than as warranted; no allowance shall be
made for any labor charges of Buyer for replacement of parts, adjustment or repairs, or any other work, unless such
charges are authorized in advance by Seller.
If Seller’s products are claimed to be defective in material or workmanship or not to conform to specifications,
drawings, blueprints and/or samples, Seller shall, upon prompt notice thereof, either examine the products where
they are located or issue shipping instructions for return to Seller (transportation charges prepaid by Buyer). In the
event any of our products are proved to be other than as warranted, transportation costs (cheapest way) to and
from Seller’s plant, will be borne by Seller and reimbursement or credit will be made for amounts so expended by
Buyer. Every such claim for breach of these warranties shall be deemed to be waived by Buyer unless made in writing within ten days from the date of discovery of the defect.
The above warranties shall not extend to any products or parts thereof which have been subjected to any misuse or
neglect, damaged by accident, rendered defective by reason of improper installation or by the performance of
repairs or alterations outside of our plant, and shall not apply to any goods or parts thereof furnished by Buyer or
acquired from others at Buyer’s request and/or to Buyer’s specifications. Routine (regularly required) calibration is
not covered under this limited warranty. In addition, Seller’s warranties do not extend to the failure of tubes, transistors, fuses and batteries, or to other equipment and parts manufactured by others except to the extent of the
original manufacturer’s warranty to Seller.
The obligations under the foregoing warranties are limited to the precise terms thereof. These warranties provide
exclusive remedies, expressly in lieu of all other remedies including claims for special or consequential damages.
SELLER NEITHER MAKES NOR ASSUMES ANY OTHER WARRANTY WHATSOEVER, WHETHER EXPRESS, STATUTORY,
OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY AND FITNESS, AND NO PERSON IS AUTHORIZED TO
ASSUME FOR SELLER ANY OBLIGATION OR LIABILITY NOT STRICTLY IN ACCORDANCE WITH THE FOREGOING.
Special Lifetime Warranty - Series 4020, Series 4027A,
Series 4027F, and Series 4028 Power Sensor Head
In addition to its standard warranty, the Bird Electronic Corporation warrants its Series 4020, Series 4027A, Series
4027F, and Series 4028 Thruline Power Sensor Heads for lifetime to original purchaser. This extended warranty is
against burnout. For the warranty to apply, the Sensor Head must be used with the correct Bird Electronic Corporation Display Unit, the maximum power rating of the Sensor must not be exceeded, the Sensor RF circuit must be
properly terminated and the Sensor not subjected to physical abuse.
Bird Electronic Corporation, at its option, will repair or replace the defective Sensor at its world Headquarters at
30303 Aurora Road, Solon, Ohio 44139.
The customer is responsible to pay transportation charges to return the defective sensor to Bird.
54
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.