BIRD 4028A, 4421, 4028B, 4028C, 4021 Operation Manual

...
Page 1
Thruline® RF Power Meter Model 4421
and
®
Thruline
Directional
Rf Power Sensors
4020 Series, 4027A Series,
4027F Series, and 4028 Series
©Copyright 2016 by Bird Electronic Corporation
Instruction Book Part Number 920-4421 Rev W
®
Thruline
is a Registered Trademark
of Bird Electronic Corporation
Page 2
Safety Precautions
The following are general safety precautions that are not necessarily related to any specific part or procedure, and do not necessarily appear elsewhere in this publication. These precautions must be thoroughly understood and apply to all phases of operation and maintenance.
WARNING
Keep Away From Live Circuits
Operating Personnel must at all times observe general safety precautions. Do not
replace components or make adjustments to the inside of the test equipment with the
high voltage supply turned on. To avoid casualties, always remove power.
WARNING
Shock Hazard
Do not attempt to remove the RF transmission line while RF power is present.
WARNING
Do Not Service Or Adjust Alone
Under no circumstances should any person reach into an enclosure for the purpose of
service or adjustment of equipment except in the presence of someone who is
capable of rendering aid.
WARNING
Safety Earth Ground
An uniterruptible earth safety ground must be supplied from the main power source to test instruments. Grounding one conductor of a two conductor power cable is not
sufficient protection. Serious injury or death can occur if this grounding is not properly
supplied.
WARNING
Chemical Hazard
Dry cleaning solvents for cleaning parts may be potentially dangerous. Avoid
inhalation of fumes or prolonged contact with skin.
WARNING
Resuscitation
Personnel working with or near high voltages should be familiar with modern
methods of resuscitation.
WARNING
Remove Power
Observe general safety precautions. Do not open the instrument with the power on.
i
Page 3
Safety Symbols
Warning notes call attention to a procedure which, if not correctly performed, could
Caution notes call attention to a procedure which, if not correctly performed, could
The caution symbol appears on the equipment indicating there is important information in the instruction manual regarding that particular area.
Note: Calls attention to supplemental information.
Warning Statements
Safety Precautions
WARNING
result in personal injury.
CAUTION
result in damage to the instrument.
The following safety warnings appear in the text where there is danger to operating and maintenance personnel and are repeated here for emphasis.
WARNING
Never attempt to connect or disconnect RF equipment from the transmission line
while RF power is being applied. Leaking RF energy is a potential health hazard.
See page 6, 34, and 43.
WARNING
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement or changing the line voltage
setting.
See page 34, 37, 37, and 43.
WARNING
Use only Nickel Metal Hydride (NiMH) batteries that have a minimum capacity of 4500
milliampere hours (mAh). Do not install batteries that are not Nickel Metal Hydride.
Do not install NiMH batteries that have less than 4500 mAh capacity. Failure to
comply may result in damage to the batteries, damage to the instrument, and injury
to personnel from battery chemicals.
See page 39 and 40.
WARNING
Heavy load. Do not attempt to lift unaided.
See page 43.
ii
Page 4
Thruline RF Power Meter, Model 4421
Caution Statements
The following equipment cautions appear in the text whenever the equipment is in danger of damage and are repeated here for emphasis.
CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may reduce the
maximum power rating of the unit.
See page 3, 43, 46, 48, and 52.
CAUTION
The Bird 4421 must be powered off when connecting or disconnecting the power
sensor from the power meter.
See page 6 and 43.
CAUTION
Do not use the power sensor with a load VSWR greater then 2:1. Damage to the
power meter, power sensor, or the RF power source could occur.
See page 6.
Long-term storage of this instrument can affect battery performance and reduce
battery life. Do not store the instrument for long periods of time without recharging
See page 8
During remote operation, periodically monitor the bus service request line. Failure to
See page 11 and 16.
maintenance should not be attempted. Removal or disturbance of the power sensor
See page 34 and 43.
Failure to install the properly rated fuse may result in equipment damage or nuisance
CAUTION
the batteries (refer to page 39). Failure to comply may result in reduced battery
charge and shortened battery life.
CAUTION
detect a service request could result in equipment damage.
CAUTION
Due to the complexity of the Bird Power Sensor, field repairs beyond general
cover can result in cancellation of lifetime warranty.
CAUTION
failures.
See page 37.
iii
Page 5
Safety Precautions
Safety Statements
USAGE
ANY USE OF THIS INSTRUMENT IN A MANNER NOT SPECIFIED BY THE MANUFACTURER MAY IMPAIR THE INSTRUMENT’S SAFETY PROTECTION.
USO
EL USO DE ESTE INSTRUMENTO DE MANERA NO ESPECIFICADA POR EL FABRICANTE, PUEDE ANULAR LA PROTECCIÓN DE SEGURIDAD DEL INSTRUMENTO.
BENUTZUNG
WIRD DAS GERÄT AUF ANDERE WEISE VERWENDET ALS VOM HERSTELLER BESCHRIEBEN, KANN DIE GERÄTESICHERHEIT BEEINTRÄCHTIGT WERDEN.
UTILISATION
TOUTE UTILISATION DE CET INSTRUMENT QUI N’EST PAS EXPLICITEMENT PRÉVUE PAR LE FABRICANT PEUT ENDOMMAGER LE DISPOSITIF DE PROTECTION DE L’INSTRUMENT.
IMPRIEGO
QUALORA QUESTO STRUMENTO VENISSE UTILIZZATO IN MODO DIVERSO DA COME SPECIFICATO DAL PRODUTTORE LA PROZIONE DI SICUREZZA POTREBBE VENIRNE COMPROMESSA.
iv
Page 6
Thruline RF Power Meter, Model 4421
SERVICE
SERVICING INSTRUCTIONS ARE FOR USE BY SERVICE-TRAINED PERSONNEL ONLY. TO AVOID DANGEROUS ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING UNLESS QUALIFIED TO DO SO.
SERVICIO
LAS INSTRUCCIONES DE SERVICIO SON PARA USO EXCLUSIVO DEL PERSONAL DE SERVICIO CAPACITADO. PARA EVITAR EL PELIGRO DE DESCARGAS ELÉCTRICAS, NO REALICE NINGÚN SERVICIO A MENOS QUE ESTÉ CAPACITADO PARA HACERIO.
WARTUNG
ANWEISUNGEN FÜR DIE WARTUNG DES GERÄTES GELTEN NUR FÜR GESCHULTES FACHPERSONAL.
ZUR VERMEIDUNG GEFÄHRLICHE, ELEKTRISCHE SCHOCKS, SID WARTUNGSARBEITEN AUSSCHLIEßLICH VON QUALIFIZIERTEM SERVICEPERSONAL DURCHZUFÜHREN.
ENTRENTIEN
L’EMPLOI DES INSTRUCTIONS D’ENTRETIEN DOIT ÊTRE RÉSERVÉ AU PERSONNEL FORMÉ AUX OPÉRATIONS D’ENTRETIEN. POUR PRÉVENIR UN CHOC ÉLECTRIQUE DANGEREUX, NE PAS EFFECTUER D’ENTRETIEN SI L’ON N’A PAS ÉTÉ QUALIFIÉ POUR CE FAIRE.
ASSISTENZA TECNICA
LE ISTRUZIONI RELATIVE ALL’ASSISTENZA SONO PREVISTE ESCLUSIVAMENTE PER IL PERSONALE OPPORTUNAMENTE ADDESTRATO. PER EVITARE PERICOLOSE SCOSSE ELETTRICHE NON EFFETTUARRE ALCUNA RIPARAZIONE A MENO CHE QUALIFICATI A FARLA.
v
Page 7
Safety Precautions
UNITS ARE EQUIPPED WITH RECHAREABLE BATTERIES. THESE ARE TO BE REPLACED BY AUTHORIZED SERVICE PERSONNEL ONLY!!!
LAS UNIDADES VIENEN EQUIPADAS CON BATERIAS RECARGABLES. ¡¡¡Y SOLAMENTE EL PERSONAL DE SERVICIO AUTORIZADO PUEDE REEMPLAZARLAS!!!
GERÄTE SIND MIT WIEDER AUFLADBAREN BATTERIEN BESTÜCKT. BATTERIEN SIND NUR VON QUALIFIZIERTEM SERICE PERSONAL AUSZUWECHSELN!!!
CES DISPOSITIFS SONT ÉQUIPÉS DE BATTERIES RECHARGEABLES. SEUL LE PERSONNEL D’ENTRETIEN AUTORISÉ EST HABILITÉ À LES REMPLACER !
LE UNITÀ SONO DOTATE DI BATTERIE RICARICABILI, CHE DEVONO DA COME SPECIFICATO DAL PRODUTTORE LA PROTEZIONE DI SICUREZZA POTREBBE VENIRNE COMPROMESSA.
vi
Page 8
Thruline RF Power Meter, Model 4421
BE SURE THE 115/230V AC VOLTAGE SELECTOR IS SET TO THE PROPER LINE VOLTAGE, AND THE CORRECT AC LINE FUSE IS INSTALLED BEFORE AC POWER IS APPLIED.
S’ASSURER QUE LE SÉLECTEUR DE TENSION 115/230V C.A. EST BIEN RÉGLÉ POUR LA TENSION DU RÉSEAU ET QUE LE FUSIBLE DE LIGNE C.A. CORRECT EST EN PLACE AVANT DE METTRE SOUS TENSION C.A.
CERCIORESE QUE EL SELECTOR DE VOLTAJE DE 115/230V CA ESTE COLOCADO A LA LINEA DE VOLTAJE APROPIADA Y QUE EL FUSIBLE ESTE INSTALADO A LA LINEA CA ANTES DE APLICAR LA CORRIENTE ALTERNA.
VOR EINSCHALTEN DER WECHSELSTROMZUFUHR SICHERSTELLEN, DASS DER 115/230V WECHSELSPANNUNGS-SELEKTOR AUF DIE VORSCHRIFTSMÄSSIGE LEITUNGSSPANNUNG EINGESTELLT UND DIE RICHTIGE WECHSELSTROM-HAUPTSICHERUNG EINGESETZT IST.
PRIMA DI EROGARE CORRENTE, ASSICURARSI CHE IL SELETTORE DI VOLTAGGIO 115/230 V.C.A. SIA REGOLATO CORRETTAMENTE E CHE IL FUSIBLE ADATTO ALLA LINEA DI ALIMENTAZIONE C.A. SIA INSTALLATO.
vii
Page 9
About This Manual
About This Manual
This manual covers the Bird 4421 RF Power Meter and the its sensors. This manual covers the operating and maintenance instructions for the following models:
Power Meter 4421
4020 Series
Sensors
4027A Series
Sensors
4027F Series
Sensors
4028A Series
Sensors
4028B Series
Sensors
4028C Series
Sensors
4021 4022 4024 4025
4027A250K 4027A400K 4027A800K 4027A2M 4027A4M 4027A10M 4027A12M 4027A25M 4027A35M 4027A60M 4027A100M 4027A150M 4027F2M 4027F10M 4027F60M
4028A250K 4028A400K 4028A2M 4028A3M 4028A4M 4028A10M 4028A25M
4028B10M 4028B3M
4028C10M
Changes to this Manual
We have made every effort to ensure this manual is accurate. If you should discover any errors, or if you have suggestions for improving this manual, please send your comments to our factory. This manual may be periodically updated. When inquiring about updates to this manual refer to the part number and revision level on the title page.
Contents
Chapter Layout
Introduction — Describes the purpose and function of the Power Meter as well as a general overview of the
product.
Installation — Describes the installation instructions for the power meter models covered in this manual. Operating Instructions — Describes the features of the power meters and provides power-up instructions. IEEE-488 GPIB Interface — Describes the features of the IEEE-488 GPIB Interface. RS-232 Interface — Describes the features of the RS-232 Interface. Maintenance — Contains preventative maintenance information, troubleshooting and repair sections, and a list
of replacement parts with part numbers.
viii
Page 10
Table of Contents
Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .i
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Warning Statements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Caution Statements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Safety Statements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .viii
Changes to this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
Chapter Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
Chapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Items Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Power Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
4020 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
4027A Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
4027F Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
4028 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Frequency and Power Ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
4020 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
4027A Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4027F Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
4028 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Chapter 2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Sensor Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
RF Line Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Panel Mounting the 4421 Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Handle Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
115/230V AC Input Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
AC Line Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Chapter 3 Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Push Button Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Audible Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Chapter 4 IEEE-488 GPIB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Cable Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Interface Capabilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Dip Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Talker-Only Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Command Syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
General Bus Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Device Dependent Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
ix
Page 11
Table of Contents
Chapter 5 RS-232 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Cable Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
DIP Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Auto Baud . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Talker-Only Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Command Syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
General Bus Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Device Dependent Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Chapter 6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Functional Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Push Button Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Power Meter Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Replacing Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Replace Batteries (Older models) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Replace Batteries (Newer Models) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Long Term Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Cal Cart Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
Front Panel (Cal Cart only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Cord Reel (Cal Cart only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Power Sensor (Cal Cart only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Casters (Cal Cart only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Handle (Cal Cart only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Storage and Shipment (Cal Cart only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Specifications, Cal Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Max. Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Specifications, 4421 RF Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Specifications, Power Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Specifications, RS-232 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Specifications, IEEE-488 GPIB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Available Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
x
Page 12
Chapter 1 Introduction
Power Meter
The Bird 4421 RF Power Meter is one component of a complete RF power measurement system. An RF power sensor such as a Bird 4021 is also required. The system can be controlled with the front panel buttons, or remotely through an RS-232 connection or a GPIB-488 connection. See
Items Supplied
Bird 4421 RF Power Meter AC Power Cord Sensor Cable Instruction Manual
Optional Accessories
Panel Mount Kit (P/N 4421-250) — Allows the Bird 4421 to be installed in a standard 19" panel for rack mount
applications.
Null Modem Kit (P/N 4380-250) — Contains the hardware necessary to allow the 4421 to be remotely
controlled by controllers with different wiring arrangements. Requires an RS-232 interface module.
Figure 1 on page 2
.
1
Page 13
Figure 1 Bird 4421 Meter Outline Drawing
Introduction
1. Operating Push Buttons
2. LCD
3. ON/OFF Push Button
4. Handle
5. Central Button
6. Fuse Access Door
7. AC Line Module
8. Master ON/OFF Switch
9. GPIB connector
10. GPIB DIP switches
* A shorter case was introduced in 2016, 4421 length dimension has changed for models built in 2016 and later.
11. RS-232 connector
12. RS-232 DIP switches
13. Power Sensor Socket
2
Page 14
Thruline RF Power Meter, Model 4421
Power Sensors
Power sensors are available with a variety of connectors; see complete list. Since the accuracy is critically dependent on the connectors used at calibration, do not remove or change the connectors.
Changing the sensor’s connectors will invalidate calibration data, and may reduce the
maximum power rating of the unit.
4020 Series
Bird 4020 Series Power Sensors are designed for lab or field use and are accurate to within ± 3%(1) of reading.
4027A Series
Bird 4027A Series Power Sensors are designed for use in semiconductor processing and calibration applications. Stringent calibration provides long-term unit-to-unit repeatability, allowing consistent amounts of RF energy to be applied to the etch process over many etch cycles. 4027A Sensors are accurate to ± 1%(1) at specified calibration frequencies and power levels.
4027F Series
Bird 4027F Series Power Sensors are similar to the 4027A series. However, additional filtering allows the 4027F to ignore harmonics of the signal being measured. The 4027F is also less sensitive to AM components of the signal. 4027F Sensors are accurate to ± 1%(2) at specified calibration frequencies and power levels.
"Available Connectors" on page 52
CAUTION
for a
Figure 2 Power Sensor Outline Drawing, 4020, 4027A, 4027F, & 4028A Series
4028 Series
Bird 4028 Series Power Sensors are high power sensors otherwise similar to the 4027A series. 4028A sensors are based on a 7/8” line, 4028B sensors use a 1-5/8” line, and 4028C sensors use a 3-1/8” line. 4028 Sensors are accurate to ± 2% at specified calibration frequencies and power levels.
3
Page 15
Figure 3 Power Sensor Outline Drawing 4028B Series only
Figure 4 Power Sensor Outline Drawing 4028C Series only
Introduction
Frequency and Power Ranges
4020 Series
Model Frequency Range RF Power Range
4021 1.8 – 32 MHz 300 mW – 1 kW 4022 25 – 1000 MHz 300 mW – 1 kW 4024 1.5 – 32 MHz 3 W – 10 kW 4025 100 kHz – 2.5 MHz 3 W – 10 kW
4
Page 16
Thruline RF Power Meter, Model 4421
4027A Series
Model Frequency Range RF Power Range
4027A250K 250 – 400 kHz 3 W – 10 kW 4027A400K 400 – 550 kHz 3 W – 10 kW 4027A800K 800 – 950 kHz 3 W – 10 kW 4027A2M 1.5 – 2.5 MHz 3 W – 10 kW 4027A4M 3 – 5 MHz 3 W – 10 kW 4027A10M 10 – 15 MHz 3 W – 10 kW 4027A12M 10 – 15 MHz 300 mW – 1 kW 4027A25M 25 – 30 MHz 3 W – 9 kW 4027A35M 35 – 45 MHz 3 W – 7.5 kW 4027A60M 45 – 65 MHz 3 W – 6 kW 4027A100M 95 – 105 MHz 3 W – 4 kW 4027A150M 150 – 170 MHz 3.75 W – 3.75kW
4027F Series
4028 Series
Model Frequency Range RF Power Range
4027F2M 1.8 – 2.2 MHz 100 W – 10 kW 4027F10M 12 – 15 MHz 3 W – 4 kW 4027F60M 57 – 63 MHz 3.75 W – 3.75 kW
Model Frequency Range RF Power Range
4028A250K 250 – 400 kHz 1 kW – 20 kW 4028A400K 400 – 550 kHz 1 kW – 20 kW 4028A2M 1.5 – 2.5 MHz 1 kW – 25 kW 4028A3M 2.5 – 3.5 MHz 1 kW – 25 kW 4028A4M 3.5 – 4.5 MHz 1 kW – 25 kW 4028A10M 10 – 15 MHz 1 kW – 25 kW 4028A25M 25 – 30 MHz 1 kW – 25 kW 4028B3M 2.5 – 4 MHz 1 kW – 25 kW 4028B10M 10 – 15 MHz 1 kW – 25 kW 4028C10M 10 – 15 MHz 500W – 50 kW
5
Page 17
Chapter 2 Installation
This chapter provides information about preparing the Bird 4421 for use.
Sensor Connection
CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may reduce the
maximum power rating of the unit.
CAUTION
The Bird 4421 must be powered off when connecting or disconnecting the power
sensor from the power meter.
1. Turn OFF the ON/OFF rocker switch on the meter’s rear panel.
2. Align the latch on the cable with the notch of the “Power Sensor” socket on the power meter’s rear panel.
3. Insert the cable until it clicks into place.
4. Connect the other end of the sensor cable to the sensor.
RF Line Connection
WARNING
Never attempt to connect or disconnect RF equipment from the transmission line
while RF power is being applied.
Leaking RF energy is a potential health hazard.
CAUTION
Do not use the power sensor with a load VSWR greater then 2:1. Damage to the power
meter, power sensor, or the RF power source could occur.
Connect the end of the power sensor labeled “SOURCE” to the RF source. Connect the end labeled “LOAD” to the load or antenna. Reversing these connections will cause measurement errors.
Panel Mounting the 4421 Power Meter
You can install the 4421 Power Meter in an equipment rack if you have the optional panel mount kit (refer to
"Optional Accessories" on page 1 on page 7
kit.
shows the overall dimensions and mounting points for a 4421 Power Meter installed in a panel mount
Note: The power supply interrupt switch for the 4421 Power Meter is located on the rear panel. When you install a unit in a panel mount kit, you need to provide a means to interrupt the power supply that is easily accessible to the user (such as a switch mounted in the panel).
). The panel mount kit includes complete installation instructions.
Figure 5
6
Page 18
Thruline RF Power Meter, Model 4421
Figure 5 Panel Mounting Dimensions
Handle Operation
The handle on the Bird 4421 can be set to four different positions (see center buttons on both sides. Releasing the buttons will lock the handle into position.
Figure 6 Handle Positions
7
Figure 6
). To adjust the handle, press the
Page 19
Installation
115/230V AC Input Power
The internal power supply can operate at 115 VAC or 230 VAC. There is no user setting to select the input voltage because the power supply automatically senses the input voltage. However, you must make sure that you have the correct power cord plug and fuses installed for the line voltage you are using (refer to
page 37
AC Line Connectors
To make the AC line cord compatible with non-U.S. voltages, users must install the appropriate connector on the power cord.
).
"Replacing Fuses" on
Batteries
The Bird 4421 RF Power Meter is completely portable and is powered from internal rechargeable nickel metal hydride batteries.
The batteries are shipped in a low charge state. It is, therefore, recommended that you charge the instrument for approximately 28 hours before using it for continuous operation.
Note: Battery charging at temperatures greater than 45° C (113° F) can result in reduced operational time.
CAUTION
Long-term storage of this instrument can affect battery performance and reduce
battery life. Do not store the instrument for long periods of time without recharging
the batteries (refer to page 39). Failure to comply may result in reduced battery
charge and shortened battery life.
8
Page 20
Chapter 3 Operating Instructions
This chapter describes operator controls and indicators on the Bird 4421 RF Power Meter. For remote operation using a IEEE-488 (GPIB) or RS-232 controller, refer to the instructions in
Push Button Functions
Figure 7 Push Buttons
Push Button Description
FWD, RFL
SWR
MIN, MAX
dBm
LIGHT
AUTO
UP, DOWN
ON/OFF
Press to measure forward or reflected RF power. FWD or RFL indicator and current unit of measure turn on.
Press to measure standing wave ratio. SWR indicator turns on. Value displayed will be between 1.0 and 199.9
Used after pressing FWD, RFL, SWR, or dBm. Displays the minimum (maximum) measured value of the previous function as long as MIN (MAX) is held down.
Used after pressing FWD or RFL. dBm indicator turns on. Power is displayed in dBm units.
Note: Used after pressing SWR. Return loss is dis­played.
Press to turn on or turn off the display’s backlight. If left on, the light automatically shuts off after 30 minutes.
Press to automatically set the scale. AUTO turns on. Press to select the next higher (lower) scale. If the scale is too high
for the power sensor, an error will be displayed.
Note: Used while AUTO indicator is on. Stops auto­matic scaling. AUTO indicator turns off.
Press to turn the power meter on or off. This switch will not work if the master power switch on the rear panel is OFF.
Note: The display flashes for about 30 seconds when the unit is first turned on.
Chapter 4
or
Chapter 5
respectively.
9
Page 21
Operating Instructions
Error Codes
The Bird 4421 displays error codes when the RF power is either below the selected range (underrange) or above the selected range (overrange).
Table 1 Error Codes
Table 2 Function Limits
FWD, RFL
FWD dBm, RFL dBm
SWR
Return Loss
Table 1
Symbol Explanation
Function Limit Error
displays the error codes and
Value greater than overrange limit of function
Value less than underrange limit of function
Power > 199.9 % of full scale or 120 % of top range
Power > 120 % of full scale Overrange Power < 3 % of low range Underrange FWD < 20 % of low range Underrange FWD – RFL = 0 Overrange FWD < 20 % of low range Underrange RFL < 20 % of low range Underrange Return Loss > 40 dB Underrange
Table 2
lists the function limits.
Overrange
Audible Warning
If the RF power level exceeds 120 % of the power sensor’s maximum power capability, the power meter will sound a warning buzzer.
10
Page 22
Chapter 4 IEEE-488 GPIB Interface
This chapter discusses setup of the IEEE-488 interface feature and describes the IEEE commands that apply to the Bird 4421. Operators should understand IEEE standard 488-1978 and have basic computer programming skills before attempting to write any programs.
CAUTION
During remote operation, periodically monitor the bus
service request line. Failure to detect a service request
could result in equipment damage.
Description
The Bird 4421 IEEE-488 (GPIB) interface has an eight-position DIP switch that sets operational conditions and interface addresses. The bottom line of the display indicates the current bus status.
Cable Connector
The interface uses a standard IEEE-488 cable connector. Pin assignments are listed in
Table 3 IEEE-488 Pin Assignments
Pin Designation Type
1 D101 Data 2 D102 Data 3 D103 Data 4 D104 Data 5 EOI Management 6 DAV Handshake 7 NRFD Handshake 8 NDAC Handshake
9 IFC Management 10 SRQ Management 11 ATN Management 12 SHIELD Ground 13 D105 Data 14 D106 Data 15 D107 Data 16 D108 Data 17 REN Management 18 GROUND Ground 19 GROUND Ground 20 GROUND Ground 21 GROUND Ground 22 GROUND Ground 23 GROUND Ground 24 GROUND, LOGIC Ground
Table 3
.
11
Page 23
Interface Capabilities
Table 4 IEEE-488 Interface Module Capabilities
Code Name Description
SH1
Source Handshake
AH1
Acceptor
Can handshake data or command bytes when the unit is acting as a source.
Can handshake the bus when it is acting as the acceptor of data or commands.
Can send data over the bus to other devices. This capability
T5
Talker
exists only after the instrument has been addressed to talk, or after a reading in talk-only mode.
Can receive device-dependent data over the bus. This
L4
Listener
capability exists only after the unit has been addressed to listen.
SR1
Service Request Can request service from the controller.
RL1
Remote-Local Can be placed in remote or local mode.
PP0
Parallel Poll Does not have parallel polling capability.
DC1
Device Clear Can be reset to factory settings.
DT1
Device Trigger Can have its readings triggered.
C0
Controller Does not have controller capability.
E1
Bus Driver Type Has open-collector bus drivers.
TE0
Extended Talker Does not have extended talker capability.
LE0
Extended Listener Does not have extended listener capability.
IEEE-488 GPIB Interface
Indicators
The bottom line of the power meter’s display shows indicators describing the status of the Bird 4421 when used with the IEEE interface. These are:
REMOTE — When REMOTE is displayed, the power meter is being controlled through the interface.
Measurements, units of measure, and certain other parameters may be changed from a remote location.
LISTEN — When LISTEN is displayed, the power meter is receiving data. TALK — When TALK is displayed, the power meter is transmitting data. This is always shown when the unit is in
“talker-only” mode.
LOCAL LOCKOUT — When LOCAL LOCKOUT is displayed, the push buttons are disabled and the power meter’s
functions are being remotely controlled.
Setup
Dip Switch
Set the interface to ADDR (Addressable) by positioning DIP switch 1 to ON. This makes the 4421 respond to controller commands.
Note: There are several button styles on DIP switches (slide, rocker, lever). Examine the DIP switch to determine the ON and OFF positions.
Set the primary address using the DIP switches. The primary address is factory set to 6, but can be set to any value between 1 and 31 (0 is reserved for the controller) To set the primary address, turn OFF switches 4 – 8 so that the sum of the bits turned off equals the desired primary address. In off switches have values of 4 and 2).
Figure 8 on page 13
, the address is set to 6 (the
12
Page 24
Thruline RF Power Meter, Model 4421
Figure 8 IEEE Interface Default DIP Switch Settings
Note: The primary address is used by the control­ler to refer to specific devices on the bus. When programming the controller, the address in the program must be the same as the address set on the interface module. Each device on the bus must have a different primary address.
Talker-Only Mode
The Bird 4421 can be set up for manual operation while automatically sending data to an output device (Talker-Only Mode). To do so, turn DIP switch 1 OFF and cycle the power. TALK will be displayed.
In Talker-Only mode, pressing a button on the meter triggers a measurement. When the measurement is complete, the information is sent to the bus and LISTEN turns on momentarily. A listen-only device on the bus, such as a printer, can read the value. The power meter is then ready to accept another button press.
Command Syntax
The Bird 4421 accepts two types of commands. General bus commands are commands, such as Device Clear (DCL), that apply to any IEEE interfaced device. Device-dependent commands are specific to the 4421.
If an invalid command is sent to the unit, an error condition is placed in the serial poll byte and the offending command is not executed.
A group of device-dependent commands can be sent as a single string as long as like command categories are not repeated, for example: “PNFCFDT3TRG”. This string sets up the 4421 to send no prefixes, read forward dBm, make one reading on “TRG”, and triggers a measurement.
Note: Commands can be entered in either upper or lower case. Note: Only the last command entered of each category will be executed. As a command string is pro-
cessed by the 4421, each category of command is stored in a separate location. Two commands of the same category will be stored in the same location, so that the second will overwrite and erase the first one.
General Bus Commands
The general bus commands supported by the IEEE-488 interface feature are listed in executing general commands varies among controllers; check the documentation supplied with your controller for the proper command structure.
Table 5 IEEE-488 General Bus Commands
Command Effect on Bird 4421
IDN? Gives product identification
REN Goes into remote mode when next addressed
GTL Cancels remote mode, restores local operation
LLO Locks out local operation
IFC Goes into talker and listener-idle status
DCL Returns to default conditions
SDC Returns to default conditions
GET Triggers reading in T2 and T3 modes
SPE, SPD Puts the status byte on the bus
13
Table 5
. The syntax for
Page 25
IDentificatioN (IDN?)
IEEE-488 GPIB Interface
Function Remarks
Product identifies itself
Replies command same as U2
Remote ENable (REN)
Function Remarks
Enables remote operation.
The unit must be addressed to listen after setting REN true.
The REMOTE indicator turns on when this command is received.
Go To Local (GTL)
Function Remarks
Returns device to local operation.
Issuing a GTL command while the device is in Local Lockout mode does not clear the lockout condition.
The REMOTE indicator turns off.
The LISTEN indicator remains on.
Local LOckout (LLO)
Function Remarks
Disables local operation of all devices on the bus.
REN must be true to use LLO.
LLO is cleared by setting REN false.
InterFace Clear (IFC)
Function Remarks
Terminates all bus activity and passes control to the system controller.
All devices are set to talker and listener idle states.
LISTEN mode is canceled, and its indicator is turned off.
14
Page 26
Thruline RF Power Meter, Model 4421
Device CLear (DCL)
Function Remarks
Table 6 IEEE-488 Default Conditions
F
Auto Range ON RYY
Two Terminators (CR LF) YT
Prefixes YES PY
Trigger One Shot on Talk Address T1
All SRQ’s OFF M00
Send EOI with last byte of message KY
Resets the status of all devices to an initialized state.
Does not change the current interface mode.
The 4421 returns to the factory default condition listed in
Default Condition Related Command
orward Carrier Wave FC
Selective Device Clear (SDC)
Function Remarks
Resets the status of a selected device to an initialized state.
Only the device addressed will be cleared.
The 4421 returns to the factory default condition.
Table 6
.
Group Execute Trigger (GET)
Function Remarks
15
Initiates a measurement for all devices set to trigger on GET.
The 4421 must be already set to trigger on GET.
Used to synchronize measurements of multiple instruments.
Page 27
Serial Polling Enable/Disable (SPE/SPD)
IEEE-488 GPIB Interface
Function
Enables or disables the serial polling sequence.
During remote operation, periodically monitor the bus service request line. Failure to
detect a service request could result in equipment damage.
Remarks
The SPE command puts all devices in serial poll mode waiting to be addressed. The SPD command clears the SRQ bit (bit 6) and ends the polling sequence.
When addressed a device sends its status byte to the controller. A value of 1 for a bit means tha
t the device condition that bit refers to is true. A value of 0 means that the
condition is false. The 4421 does not use all bits of the status byte.
a description and how to reset them.
Table 7 IEEE Status Byte Description
Bit Name Condition
Set if a service request is generated by the 4421. If an SRQ has been received by the controller and this bit is cleared,
6
SRQ
other instruments on the bus should be checked to determine where the SRQ occurred.
Cleared by a serial poll of the 4421.
Measurement
3
Complete
Set when the power meter has completed a reading. Cleared by requesting a reading over the bus. Set when the RF power is underrange and a reading has
2
Reading Underflow
been completed. Cleared by requesting a reading over the bus. Set when the RF power is overrange and a reading has been
1
Reading Overflow
completed. Cleared by requesting a reading over the bus. Set if an illegal device-dependent command (IDDC) or illegal
device-dependent command option (IDDCO) was received,
0
Error
or if the power meter fails the self test. Cleared by reading the U1 status word. The U1 word
contains details on the error, see
page 21
CAUTION
.
Table 7
lists the bits used, along with
"Status (Ux)" on
16
Page 28
Thruline RF Power Meter, Model 4421
Device Dependent Commands
The device-dependent commands used by the 4421 Power Meter are listed in
Note: The programming card also has a complete command list.
Table 8 IEEE-488 Device Dependent Command Summary
Category Command Description
FC Forward carrier wave
FD Forward dBm
RC Reflected carrier wave
Measurement
Range
Terminators
Prefixes
Triggers
Serial Polling
Status
Self-Test J0 Run self-test
EOI Response
Writable Store WXXXXXX Place XXXXXX in RAM
RD Reflected dBm
SW Standing wave ratio
RL Return loss
MN Minimum value
MX Maximum value
RYY Auto range on
R00 to R17 Manual ranges
RNN Auto range off, stay at present range
YT Two terminators: CR, LF
YO One terminator: CR
YN No terminator
PY Prefix YES
PN Prefix NO
T0 Continuous on TALK
T1 One shot on TALK
T2 Continuous on GET
T3 One shot on GET
T4 Continuous on measurement command
T5 One shot on measurement command
M00 Do not generate SRQ
M01 Generate SRQ on error
M02 Generate SRQ on measurement overrange
M04 Generate SRQ on measurement underrange
M08 Generate SRQ on operation complete
U0 Send back current machine state
U1 Send back error conditions
U2 Send back revision levels
KY Send EOI on last byte
KN Do not send EOI on last byte
Table 8
, organized by category.
17
Page 29
Forward Carrier Wave (FC) Forward dBm (FD) Reflected Carrier W
ave (RC)
Reflected dBm (RD)
IEEE-488 GPIB Interface
Function Remarks
Selects forward or reflected RF power measurement mode.
Measurement results are returned in Watts or dBm.
Standing Wave Ratio (SW) Return Loss (RL)
Function Remarks
Selects SWR or return loss match measurement mode.
Measurement results are returned in VSWR or dB.
MiNimum Value (MN) MaXimum Value (MX)
Function Remarks
Selects minimum or maximum measurement mode.
Another measurement must be selected before selecting min or max.
Returns the minimum (or maximum) value of the previous measurement type.
18
Page 30
Thruline RF Power Meter, Model 4421
Range (Rxx)
Function Remarks
Selects a measurement range listed in
If the selected range is outside the range of the connected power sensor, the command is ignored.
Table 9 Measurement Ranges
Command Power Range
RYY RNN
R17 R16 R15 R14 R13 R12 R11 R10 R09 R08 R07 R06 R05 R04 R03 R02 R01 R00
Table 9
.
Turn auto range on Turn auto range off. Keep present range
18.0 – 199.9 MW
1.80 – 19.99 MW
0.180 – 1.999 MW
18.0 – 199.9 kW
1.80 – 19.99 kW
0.180 – 1.999 kW
18.0 – 199.9 W
1.80 – 19.99 W
0.180 – 1.999 W
18.0 – 199.9 mW
1.80 – 19.99 mW
0.180 – 1.999 mW
18.0 – 199.9 μW
1.80 – 19.99 μW
0.180 – 1.999 μW
18.0 – 199.9 nW
1.80 – 19.99 nW
0.180 – 1.999 nW
Terminators (Yx)
Function
Selects the characters that follow the end of a data string. Set x to:
“T” for two terminators; a carriage return (CR) and a line feed (LF). “O” for one terminator; a carriage return (CR).
“N” for no terminator; message can be terminated by EOI.
Remarks
19
Many controllers use the terminator sequence to recognize the end of an input string. Using incorrect terminators can lock the bus.
Page 31
Prefixes (Px)
IEEE-488 GPIB Interface
Function
Turns the prefix mode on or off. Set x to:
“Y” to enable prefixes. “N” to disable prefixes.
Remarks
Prefixes are sent over the bus with the measurement, and indicate the status of the current measurement (see
“FC”,”FD”,”RC”,”RD”,”SW”,”RL”,”MN”,”MX” indicates the measurement type. “U” indicates underflow; the value sent is “.000”. “O” indicates overflow; the value sent is “199.9”. “N” indicates normal; the value sent is a normal on-scale reading. “4421” indicates the Bird model number.
Table 10 Prefix Examples
Data String Description
NFC.0.123W(CR)(LF) Normal (N) forward carrier wave (FC), prefixes on
OFC 199.9W(CR)(LF) Overflowed (O) forward carrier wave (FC), prefixes on
199.9W(CR)(LF) Overflowed forward carrier wave, prefixes off
URD .000W(CR)(LF) Underflowed (U) reflected dBm (RD), prefixes on
Table 10
for examples):
Triggers (Tx)
Function Remarks
Table 11 Trigger Conditions
Selects the condition which will trigger a reading (see
Failure to trigger device before requesting a reading will lock the bus.
T1 halts the bus until a reading is available.
T0 and T1 do not set the measurement complete SRQ.
T2, T3, T4, and T5 set a SRQ when the measurement is complete.
Fastest reading rate is 2.4 readings/second; slowest is 1 reading/sec.
Command Trigger Condition
T0 T1 T2 T3
T4
T5
Table 11
Continuous on talk
One shot on talk
Continuous on GET
One shot on GET
Continuous on measurement command (FC,FD, RC, RD, SW, RL, MN, MX)
One shot on measurement command (FC, FD, RC, RD, SW, RL, MN, MX)
).
20
Page 32
Thruline RF Power Meter, Model 4421
SRQ Mask (Mxx)
Function
Remarks
Table 12 SRQ Mask Bits
Selectively masks status bits to prevent unwanted service requests.
Set xx to the sum of the binary values of the desired SRQ trigger bits. For example, M12 would set the SRQ for both operation complete and underrange
alues 8 and 4). Set xx to “00” to never generate an SRQ.
(v If a status bit is masked (bit set to 0), SRQs won’t be generated for that condition.
Binary Value Bit Number Message
1 0 (LSB) 21 42 83
Status (Ux)
4 5 6
7 (MSB)
Error (IDDC, IDDCO, self-test fail) Measurement overrange Measurement underrange Operation complete Not Used Not Used Can’t mask Not Used
Function
Reads a status word and returns the information as a string. Set x to:
“0” for machine status. “1” for error status. “2” for revision history.
Remarks
Figure 9 Machine Status Word Format
After sending the status command, a status word is sent the next time the unit is addressed to talk. To ensure the correct status is transmitted, the status word should be requested as soon as possible after the command is sent.
Machine Status Word (U0) —
Figure 9
. The default values are also shown.
The format of the machine status word is shown in
21
Page 33
Error Status Word (U1) — The format of the error status word and the possible
error messages are shown in in the status (serial poll) byte, and a SRQ may be generated (
on page 21
). All flags will revert to their non-error states after the U1 command is
sent.
Figure 10 Error Status Word Format
Status Meaning Description
ICM Invalid Command Set when an illegal device-dependent command
VCM Valid Command Set when no IDDC is received. ICO Invalid Command Option Set when an illegal device-dependent command
VCO Valid Command Option Set when no IDDCO is received. PS Self-Test Pass Set when a self-test has been initiated by the
FL Self-Test Fail Set when the self-test has failed. (This is the
IEEE-488 GPIB Interface
Figure 10
. When an error occurs, an error is also flagged
"SRQ Mask (Mxx)"
(IDDC) such as V2 is received. (V is illegal)
option (IDDCO) such as T6 is received. (6 is illegal)
J0 command and the test result is acceptable.
default condition.)
Revision History Word (U2) — The format of the revision history word is shown in
Figure 11
.
Figure 11 Revision History Word Format
Note: If Writable stores parameter has not been set, this command responds only with -4420-.
22
Page 34
Thruline RF Power Meter, Model 4421
Self Test (J0)
Function Remarks
Initiates a hardware and software test.
Results are stored in the U1 status word (
“J0” must be sent each time before reading the result.
End Or Identify (Kx)
Function
Enables or disables the End or Identify (EOI) signal. Set x to:
“Y” to enable. “N” to disable.
Remarks
Writable Store (W x x x x x x)
Function Remarks
Disabling EOI can cause some controllers to lock unless another terminator is used.
When enabled, EOI is only asserted at the end of a multiple byte string.
Storage for six bytes of ASCII data.
Data stored is lost when the 4421 is turned off.
Data is sent back as part of the U2 status word.
"Status (Ux)" on page 21
).
23
Page 35
Chapter 5 RS-232 Interface
This chapter discusses setup of the RS-232 interface feature and describes the RS-232 commands that apply to the Bird 4421. Operators should understand EIA Standard RS-232-C and have basic computer programming skills before writing any programs.
Description
The Bird 4421 RS-232 interface feature is an integral part of the I/O hub circuit board inside of the mm21. An eight­position DIP switch is used to set operational conditions such as baud rate, parity, and stop bits. The bottom line of the display indicates the current bus status.
Cable Connector
The interface uses a standard 25-pin RS-232 connector. Pin assignments are listed in uses a different wiring arrangement, do not attempt to rewire the interface module’s connector. A null modem kit should be used for rewiring instead.
Table 13 RS-232 Pin Assignments
Pin Designation Notes
1 Protection Ground Chassis Ground
2 Transmit Data
3 Receive Data
4 Request to Send (Output) Set true after module power up
5 Clear to Send
6 Data Set Ready (Input) Set internally true by module
7 Signal Ground Return path for data and control signals
8 Receive Signal DET (Input) Set true by module
20 Data Terminal Ready (Output) Set true after module power up
Indicators
(Input) Set by input device. When true, it enables the module to transmit. When false, it disables transmission.
Table 13
. If the controller
The bottom line of the power meter’s display shows indicators describing the status of the Bird 4421 when used with the RS-232 interface. These are:
TALK — When TALK is displayed, the power meter is transmitting data. This is always shown when the unit is in
“talker-only” mode.
LISTEN — When LISTEN is displayed, the power meter is receiving data.
24
Page 36
Thruline RF Power Meter, Model 4421
Setup
DIP Switch
Set the DIP switches according to application needs and the requirements of the controller. Available settings and factory defaults are listed in
Note: There are several button styles on DIP switches (slide, rocker, lever). Examine the DIP switch to determine the ON and OFF positions.
Figure 12 RS-232 Interface Default DIP Switch Settings
Note: If you change switch settings when the unit is ON, the new settings will not become effective until power is turned OFF and then back ON.
Switch Function Description
1 Stop Bit
2 Command Mode
3 Word Length
4,5 Parity
6,7,8 Baud Rate
Figure 12
.
ON OFF
ON
OFF
ON OFF
(4) ON ON OFF OFF
(6) ON ON ON OFF OFF OFF OFF
1 Stop Bit 2 Stop Bits
ENT command needed before sending reading to controller† Trigger automatically sends reading Reading also sent when front panel push button pressed
8 Data Bits† 7 Data Bits
(5) ON OFF ON OFF
(7) ON
OFF OFF ON ON OFF OFF
No Parity† Odd Parity Even Parity Mark Parity
(8) ON ON OFF ON OFF ON OFF
Auto Baud††
300 600 1200
2400† 4800 9600
† Default factory setting †† For more information see “Auto Baud”
25
Page 37
RS-232 Interface
Auto Baud
Auto Baud is used to automatically determine the correct transmission rate. After setting DIP switches 1 through 5, follow the steps below to use auto baud:
1. Set DIP switches 6, 7, and 8 to ON.
2. Connect a controller to the power meter.
3. Turn the power meter on.
4. Send the character U (hexadecimal 55) from the controller to the power meter.
5. Wait 1 second.
6. If the power meter’s front panel displays LISTEN, the baud rate has been determined. If not, repeat steps 4 and
5.
If, after 30 seconds, the module cannot absolutely determine a baud rate, it will choose a rate based on data acquired during the test.
Note: Auto Baud is the only automatically chosen setting. Other items such as parity and stop bits must be manually selected.
Talker-Only Mode
The Bird 4421 can be set up for manual operation while automatically sending data to an output device (Talker-Only Mode). To do so, turn DIP switch 2 OFF and cycle the power. TALK will be displayed.
In Talker-Only mode, pressing a button on the meter triggers a measurement. When the measurement is complete, the information is sent to the bus.
Command Syntax
The Bird 4421 accepts two types of commands. General bus commands are commands, such as Initialize (INT), that apply to any RS-232 interfaced device. Device-dependent commands are specific to the 4421.
If an invalid command is sent to the unit, an error condition is placed in the serial poll byte and the offending command is not executed.
A group of device-dependent commands can be sent as a single string as long as like command categories are not repeated. For example: “PNFCFDT3TRG”. This string sets up the 4421 to send no prefixes, read forward dBm, make one reading on “TRG”, and triggers a measurement.
Note: Commands can be entered in either upper or lower case. Note: Only the last command entered of each category will be executed. As a command string is pro-
cessed by the 4421, each category of command is stored in a separate location. Two commands of the same category will be stored in the same location, so that the second will overwrite and erase the first one.
General Bus Commands
The general bus commands supported by the RS-232 interface module are listed in
Table 14 RS-232 General Bus Commands
Command Effect on Bird 4421
INT ENT TRG
B1 to B7
XO/XF
Returns to default conditions Sends a reading to the controller Triggers reading in T3 mode Selects a baud rate Enables/disables software handshake
Table 14
.
26
Page 38
Thruline RF Power Meter, Model 4421
INiTialize (INT)
Function Remarks
Resets the Bird 4421 and returns it to the factory defaults.
If INT is linked with any other command within a string, it must be separated from that command by a space.
ENTer (ENT)
Function Remarks
Makes the power meter transmit a reading to the controller.
A measurement must have already been triggered, placing a reading in the output buffer.
To send a reading whenever a measurement is triggered, set DIP switch 2 to OFF. The ENT c
TRiGger (TRG)
Function
Initiates a measurement if the power meter is in trigger mode (T3).
Baud Select (Bx)
Function Remarks
Selects a baud rate listed in
When the meter recognizes a valid Bx command, its baud rate is immediately changed. (The controller is assumed to be transmitting at the new rate; otherwise sending commands would not be possible.)
This command overrides the DIP switch setting.
ommand will not need to be sent.
Table 15
.
Table 15 Baud Rates
27
Command Baud Rate
B2 300 B3 600 B4 1200 B5 2400 B6 4800 B7 9600
Page 39
Xmission Flow Control (XO/XF)
RS-232 Interface
Function
Enables or disables the XON/XOFF flow control.
XO enables flow control. XF disables flow control.
Remarks
Device Dependent Commands
The device-dependent commands used by the 4421 Power Meter are listed in
Table 16 RS-232 Device Dependent Command Summary
Category Command Description
Measurement
Range
Terminators
Prefixes
Triggers
Status
Self-Test Writable Store
When data is being sent from the power meter to the computer and flow control is enabled, data transmission will be suspended when the XOFF character (hexadecimal
13) is sent by the computer. Transmission will resume when XON (hex 11) is sent by the computer.
When data is being sent from the computer to the power meter, XOFF will be sent to
omputer when the input buffer fills up. XON will be sent to the computer when the
the c buffer has emptied.
FC Forward carrier wave FD Forward dBm RC Reflected carrier wave RD Reflected dBm SW Standing wave ratio RL Return loss MN Minimum value MX Maximum value RYY Auto range on R00 to R17 Manual ranges RNN Auto range off, stay at present range YT Two terminators: CR, LF YO One terminator: CR YN No terminator PY Prefix YES PN Prefix NO T0 Continuous on ENT T1 One shot on ENT T3 One shot on TRG T5 One shot on measurement command U0 Send back current machine state U1 Send back error conditions U2 Send back revision levels J0 Run self-test WXXXXXX Place XXXXXX in RAM
Table 16
, organized by category.
28
Page 40
Thruline RF Power Meter, Model 4421
Forward Carrier Wave (FC) Forward dBm (FD) Reflected Carrier W
ave (RC)
Reflected dBm (RD)
Function Remarks
Selects forward or reflected RF power measurement mode.
Measurement results are returned in Watts or dBm.
Standing Wave Ratio (SW) Return Loss (RL)
Function Remarks
Selects SWR or return loss match measurement mode.
Measurement results are returned in VSWR or dB.
MiNimum Value (MN) MaXimum Value (MX)
Function Remarks
Selects minimum or maximum measurement mode.
Another measurement must be selected before selecting min or max.
Returns the minimum (or maximum) value of the previous measurement type.
29
Page 41
Range (Rxx)
RS-232 Interface
Function Remarks
Selects a measurement range listed in
If the selected range is outside the range of the connected power sensor, the command is ignored.
Table 17 Measurement Ranges
Command Power Range
RYY
RNN
R17 R16 R15 R14 R13 R12
R11
R10 R09 R08 R07 R06 R05 R04 R03 R02 R01 R00
Table 17
.
Turn auto range on Turn auto range off. Keep present range
18.0 – 199.9 MW
1.80 – 19.99 MW
0.180 – 1.999 MW
18.0 – 199.9 kW
1.80 – 19.99 kW
0.180 – 1.999 kW
18.0 – 199.9 W
1.80 – 19.99 W
0.180 – 1.999 W
18.0 – 199.9 mW
1.80 – 19.99 mW
0.180 – 1.999 mW
18.0 – 199.9 μW
1.80 – 19.99 μW
0.180 – 1.999 μW
18.0 – 199.9 nW
1.80 – 19.99 nW
0.180 – 1.999 nW
Terminators (Yx)
Function
Selects the characters that follow the end of a data string. Set x to:
“T” for two terminators; a carriage return (CR) and a line feed (LF). “O” for one terminator; a carriage return (CR). “N” for no terminator.
Remarks
Many controllers use the terminator sequence to recognize the end of an input string. Using incorrect terminators can lock the bus.
30
Page 42
Thruline RF Power Meter, Model 4421
Prefixes (Px)
Function
Turns the prefix mode on or off. Set x to:
“Y” to enable prefixes. “N” to disable prefixes.
Remarks
Prefixes are sent over the bus with the measurement, and indicate the status of the current measurement (see
“FC”,”FD”,”RC”,”RD”,”SW”,”RL”,”MN”,”MX” indicates the measurement type. “U” indicates underflow; the value sent is “.000”. “O” indicates overflow; the value sent is “199.9”. “N” indicates normal; the value sent is a normal on-scale reading. “4421” indicates the Bird model number.
Table 18 Prefix Examples
Data String Description
NFC.0.123W(CR)(LF) Normal (N) forward carrier wave (FC), prefixes on
OFC 199.9W(CR)(LF) Overflowed (O) forward carrier wave (FC), prefixes on
199.9W(CR)(LF) Overflowed forward carrier wave, prefixes off
URD .000W(CR)(LF) Underflowed (U) reflected dBm (RD), prefixes on
Table 18
for examples):
Triggers (Tx)
Function Remarks
Table 19 Trigger Conditions
Selects the condition which will trigger a reading (see
Failure to trigger device before requesting a reading will lock the bus.
Fastest reading rate is 2.4 readings/second; slowest is 1 reading/sec.
Command Trigger Condition
T0
T1
T3
T5
Table 19
Continuous on ENT
One shot on ENT
One shot on TRG
One shot on measurement command (FC, FD, RC, RD, SW, RL, MN, MX)
).
31
Page 43
Status (Ux)
RS-232 Interface
Function
Reads a status word and returns the information as a string. Set x to:
“0” for machine status. “1” for error status. “2” for revision history.
Remarks
Figure 13 Machine Status Word Format
Note: This command sets the unit to the settings that were last issued remotely (through RS-232 or GPIB). Therefore, it will over ride any manual changes selected at the front panel of the unit.
After sending the status command, a status word is sent the next time the unit is addressed to talk. To ensure the correct status is transmitted, the status word should be requested as soon as possible after the command is sent.
Machine Status Word (U0) — The format of the machine status word is shown in
Figure 13
. The default values are also shown.
Error Status Word (U1) — The format of the error status word and the possible
error messages are shown in after the U1 command is sent.
Figure 14 Error Status Word Format
Status Meaning Description
ICM Invalid Command Set when an illegal device-dependent command
VCM Valid Command Set when no IDDC is received. ICO Invalid Command
Option
VCO Valid Command Option Set when no IDDCO is received. PS Self-test Pass Set when a self-test has been initiated by the J0
FL Self-test Fail Set when the self-test has failed. (This is the
Figure 14
(IDDC) such as V2 is received. (V is illegal)
Set when an illegal device-dependent command option (IDDCO) such as T6 is received. (6 is illegal)
command and the test result is acceptable.
default condition.)
. All flags will revert to their non-error states
32
Page 44
Thruline RF Power Meter, Model 4421
Revision History Word (U2) — The format of the revision history word is shown in
Figure 15
Figure 15 Revision History Word Format
.
Self Test (J0)
Function Remarks
Initiates a hardware and software test.
Results are stored in the U1 status word (see
“J0” must be sent each time before reading the result.
Writable Store (W x x x x x x)
Function Remarks
Storage for six bytes of ASCII data.
Data stored is lost when the 4421 is turned off.
Data is sent back as part of the U2 status word
"Status (Ux)" on page 32
).
33
Page 45
Chapter 6 Maintenance
This chapter describes routine maintenance, along with troubleshooting instructions for the power meter and power sensor. Disassembly instructions for the Bird Cal Cart are also provided. For service beyond this level, return the unit to a qualified service center.
WARNING
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement or changing the line voltage
setting.
WARNING
Never attempt to connect or disconnect RF equipment from the transmission line
while RF power is being applied.
Leaking RF energy is a potential health hazard.
The Bird 4421 Power Meter requires only simple, routine maintenance.
Wipe off dust and dirt regularly. Check the connectors and cables for damage. Clean the connector contacts with alcohol or dry cleaning solvent.
Cleaning
CAUTION
Do not use harsh or abrasive detergents for cleaning.
Clean the Bird 4421 Power Meter and its display with a soft cloth dampened with mild detergent and water only. Clean sensors with a dry cleaning solvent that leaves no residue.
Troubleshooting
Since the power meter and power sensor can only work together, the first step is to determine which is malfunctioning. Connect the power sensor to the meter and perform the power meter is malfunctioning, refer to the troubleshooting table below. If the power sensor is malfunctioning, return it for service.
CAUTION
Due to the complexity of the Bird Power Sensor, field repairs beyond general
maintenance should not be attempted.
Removal or disturbance of the power sensor cover can result in cancellation of lifetime
warranty.
This manual cannot list all malfunctions that may occur, or corrective actions. If a malfunction is not listed or is not corrected by the listed corrective actions, contact a qualified service center.
Note: For power sensors in Cal Carts, following the test procedure will check the sensor not covered by the safety cover. To test the other sensor slide the safety cover over, then proceed as normal.
"Functional Test" on page 35
. If the
34
Page 46
Thruline RF Power Meter, Model 4421
PROBLEM POSSIBLE CAUSE CORRECTION
Have the batteries been charged? Recharge the batteries.
Power meter has no power
Is the power meter’s ac power cord connected?
Is the ON/OFF rocker switch on the rear panel set to OFF?
(Cal Cart only)
Is the Cal Cart’s AC power cord
Connect the power cord.
Set the switch to ON.
Connect AC power.
connected to the AC line?
Check fuse rating and replace fuse
Blown fuse?
(See
"Replacing Fuses" on
page 37
).
Is the sensor cable connected to
Dash moves across the display
both the power meter and power sensor?
Connect sensor cable.
Is the sensor cable defective? Replace sensor cable.
Display blank or not updating Have the batteries been charged?
Power meter turns off while on battery power
Push buttons do not respond
Is “LO BAT” displayed?
Test the push buttons (See
Button Test" on page 36
"Push
). Are
NO: Recharge battery YES: Return meter for service.
YES: Recharge battery NO: Return meter for service.
Return meter for service.
they defective?
Every segment on the display is lit Return meter for service.
IEEE-488 Interface does not respond to the interface link
Are the address in the interface program and the address setting on the DIP switches on the interface module the same?
Send the “J0” self test command, then check the status. Does the
Change the program or DIP switch setting so that the addresses are the same (See
page 12
).
Replace I/O hub PCB.
power meter fail the self test? Is the IEEE cable defective? Replace IEEE interface cable.
RS-232 Interface does not respond to the interface link. Fails J0 self­test command.
Are the DIP switches set correctly? Set DIP switches.
Is the RS-232 cable defective? Replace RS-232 interface cable.
For “talk/listen” operation, set
With the RS-232, the power meter displays “TALK” but does not operate as expected.
Is DIP switch 2 set correctly?
Is DIP switch 1 set correctly?
this switch to ON. For “talk always” set this switch to OFF.
Set DIP switch 1 as indicated in
Figure 12 on page 25
"Dip Switch" on
.
Functional Test
1. Turn the power meter off. The ON/OFF switch on the rear panel should be OFF and the ac power cable should be connected.
2. Turn ON the ON/OFF switch on the rear panel of the meter.
3. While holding down the FWD and SWR push buttons, press the ON/OFF button on the front panel of the power meter. Immediately release all three.
35
Page 47
Maintenance
Note: The unit tests the display on power up cycling through and activating each segment two times and then activating all segments at once two times.
4. The power meter’s model number and revision date should scroll across the display. If a dash “–” is displayed instead, then the meter is malfunctioning.
5. The power sensor’s model number and revision date should scroll across the display. If a dash is displayed after the power meter data, then the power sensor is malfunctioning.
Figure 16 Test Display, No Malfunction
Push Button Test
Note: If a push button is malfunctioning, return the power meter.
1. Disconnect the power sensor.
2. Turn the power meter ON.
Note: After the power up display disappears, a dash “–” should scroll across the display.
3. Turn the power meter OFF.
4. Connect the power sensor.
5. Turn the power meter ON.
Note: “AUTO” and “FWD” should be displayed, and a reading of “0.00 W” (or a very low number).
6. Press RFL.
Note: “FWD” should change to “RFL” on the display. The reading should remain the same.
7. Press SWR.
Note: “RFL” should change to “SWR”. “0.00 W” should change to “ ” (underrange error).
8. Hold down MIN.
Note: ” should change to “ ” (overrange error).
9. Release MIN.
Note: ” should change to “ ”.
10. Press dBm.
Note: “SWR” should turn OFF and “dBm” should turn ON.
36
Page 48
Thruline RF Power Meter, Model 4421
11. Press dBm.
Note: “dBm” should change to “SWR”.
12. Press FWD.
Note: “SWR” should change to “FWD” and “ ” to “.0000 W” (or a very low number).
13. Hold down MAX.
Note: “.0000” should change to “1.0000”.
14. Release MAX.
Note: “1.0000” should change to “.0000” (or a very low number).
15. Press LIGHT.
Note: The back-light should turn on.
16. Press LIGHT.
Note: The back-light should turn off.
17. Press s (up).
Note: The power meter should change ranges each time it is pressed until it reaches “0.000 KW”.
18. Press t (down).
Note: The power meter should change ranges each time it is pressed until it reaches “0.00 W”.
19. Turn the power meter OFF.
Power Meter Repair
WARNING
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement or changing the line voltage
setting.
Replacing Fuses
WARNING
To avoid personal injury, disconnect the power cord from the ac line before
performing any maintenance, including fuse replacement.
CAUTION
Failure to install the properly rated fuse may result in equipment damage or nuisance
failures.
1. Gently pry the fuse drawer out of the AC module. The fuse holder does not detach from the AC module. See
Figure 17
.
2. Install the replacement fuses then close and secure the fuse holder. Fuse 1 is a spare fuse and fuse 2 is the active fuse.
37
Page 49
AC Line Voltage Fuse Rating
115 Vac T630 mA, 5x20 mm Time Lag Fuse 230 Vac T315 mA, 5x20 mm Time Lag Fuse
Figure 17 AC Line Fuse
Maintenance
38
Page 50
Thruline RF Power Meter, Model 4421
Replacing Batteries
There are two battery configurations in the 4421 Power Meter. In the larger case models (before 2016), the batteries are horizontally arranged, in units with a smaller case the batteries are vertically arranged in a battery assembly.
Replace Batteries (Older models)
WARNING
Exposed AC line voltage (115 VAC or 230 VAC). Disconnect the power cord from the
AC line before replacing the batteries. Failure to comply may result in severe
electrical shock or death.
WARNING
Use only Nickel Metal Hydride (NiMH) batteries that have a minimum capacity of 4500
milliampere hours (mAh). Do not install batteries that are not Nickel Metal Hydride.
Do not install NiMH batteries that have less than 4500 mAh capacity. Failure to
comply may result in damage to the batteries, damage to the instrument, and injury
to personnel from battery chemicals.
1. Disconnect the AC power cord from the AC mains supply.
2. Remove the four screws that secure the top cover.
3. Lift the top cover to access the attached ground wire.
4. Disconnect the ground wire.
5. Remove the cover.
6. Unfasten the battery retaining belts (
Figure 18
).
7. Remove the battery tubes.
Note: Be sure to note the polarity and orientation of the battery tubes before removing them.
8. Remove the batteries from each tube.
9. Insert replacement batteries.
Note: Be sure to note the polarity (positive and negative) arrangement of the batteries.
10. Install the battery tubes into the unit and secure with the retaining belts.
Note: Be sure to position the retaining belts as they were before removal and tighten them securely.
11. Connect the ground wire to the top cover.
12. Install the top cover.
13. Secure top cover with the four screws removed earlier.
39
Page 51
Figure 18 Batteries
Maintenance
Item Description
1 Battery retaining belt
2
Battery tube (batteries inside)
Replace Batteries (Newer Models)
WARNING
Exposed AC line voltage (115 VAC or 230 VAC). Disconnect the power cord from the
AC line before replacing the batteries. Failure to comply may result in severe
electrical shock or death.
WARNING
Use only Nickel Metal Hydride (NiMH) batteries that have a minimum capacity of 4500
milliampere hours (mAh). Do not install batteries that are not Nickel Metal Hydride.
Do not install NiMH batteries that have less than 4500 mAh capacity. Failure to
comply may result in damage to the batteries, damage to the instrument, and injury
to personnel from battery chemicals.
1. Disconnect the AC power cord from the ac mains supply.
2. Remove the four screws that secure the top cover.
3. Lift the top cover to access the attached ground wire then disconnect the ground wire and remove the cover.
4. Remove two retaining screws securing the battery assembly. See
Figure 19
.
5. Disconnect battery assembly power supply connector.
40
Page 52
Thruline RF Power Meter, Model 4421
Figure 19 Battery Assembly Removal
6. Lift the battery assembly out of the Power Meter.
7. Unfasten the battery retaining belts then remove the battery tubes (
Figure 20
). Be sure to note the polarity
and orientation of the battery tubes before removing them.
Figure 20 Battery Assembly
8. Remove the batteries from each tube and insert replacement batteries. Be sure to note the polarity (positive and negative) arrangement of the batteries.
9. Install the battery tubes into the battery assembly and secure with the retaining belts. Be sure to position the retaining belts as they were before removal and tighten them securely.
10. Install battery assembly in the Power Meter. Align tabs on battery assembly with slots in Power Meter.
11. Install two screws to secure battery assembly to Power Meter.
12. Connect the battery assembly power supply connector to the main board.
13. Connect the ground wire to the top cover.
14. Install the top cover and secure it with the four screws removed earlier.
41
Page 53
Maintenance
Long Term Storage
Do not store the instrument for long periods of time without recharging the batteries. When the instrument is stored for long periods of time without use, the batteries will loose their charge and also loose the ability to reach full charge when put into service. To restore the battery charge, perform a full charge for 28 hours. If the batteries are not fully charged after 28 hours, completely discharge the batteries then perform a full charge again. If necessary, repeat this discharge and charge cycle up to three times. If the batteries do not remain charged after three discharge-charge cycles, replace the batteries.
Cal Cart Repair
It is recommended that you reference the specific cal cart manual for procedures tailored to each Cal Cart type.
Front Panel (Cal Cart only)
Note: If the power meter is mounted in a Cal Cart, to access the power sensor or the ac connections, it
will be necessary to remove the front panel of the Cal Cart.
1. Remove both screws on the front panel, between the RF connector and the handle.
2. Pull on the handle set into the base of the front panel to remove it.
Note: When making AC connections, refer to the following wiring diagram.
Cord Reel (Cal Cart only)
1. Disconnect the cord reel from the terminal strip.
2. Unscrew the AC connector.
3. Unscrew and remove the cord reel.
4. Screw the replacement cord reel into place.
5. Thread the AC connector through the grommet, and connect the wires on the other end to the terminal strip (See
Figure 21 on page 42
Figure 21 Cal Cart Wiring Schematic
).
42
Page 54
Thruline RF Power Meter, Model 4421
Power Sensor (Cal Cart only)
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement or changing the line voltage
Never attempt to connect or disconnect RF equipment from the transmission line
while RF power is being applied. Leaking RF energy is a potential health hazard.
The Bird 4421 must be powered off when connecting or disconnecting the power
Changing the sensor’s connectors will invalidate calibration data, and may reduce the
WARNING
setting.
WARNING
CAUTION
sensor from the power meter.
CAUTION
maximum power rating of the unit.
CAUTION
Due to the complexity of the Bird Power Sensor, field repairs beyond general
maintenance should not be attempted. Removal or disturbance of the power sensor
cover can result in cancellation of lifetime warranty.
1. Disconnect the RF line from the Cal Cart.
2. Remove the Cal Cart front panel (see
3. Disconnect the sensor cable from the power sensor.
4. Disconnect the RF cable from the sensor output port.
5. Remove the screws on the sensor mounting bracket.
6. Remove the sensor from the mounting bracket.
7. Put the new sensor in the bracket and screw it into place.
"Front Panel (Cal Cart only)" on page 42
).
Note: Make sure the arrow on the side of the sensor points towards the load, and that the end labeled “SOURCE”
points towards the front of the Cal Cart.
8. Connect the RF cable to the sensor end labeled “LOAD”. Connect the sensor cable.
9. Replace the Cal Cart front panel.
Load
WARNING
Heavy load. Do not attempt to lift unaided.
1. Disconnect all RF cables from the RF connectors on the load.
2. Unplug the blower assembly on blower-equipped loads.
3. Replace the vent plugs with the shipping plugs.
4. Remove the mounting bolts connecting the load to the Cal Cart.
5. Remove the load.
43
Page 55
Casters (Cal Cart only)
Maintenance
1. Remove the load (See
2. Flip the cart over.
3. Unscrew the defective caster.
4. Screw the new caster into place and replace the load.
Handle (Cal Cart only)
1. Unscrew the handle
2. Remove the old handle.
3. Screw the new handle into place.
"Load" on page 43
).
Storage and Shipment (Cal Cart only)
Cover the unit before storing to keep out dust and dirt. It is not necessary to install the shipping plug. Store in a dry, dust-free environment where the ambient temperature will remain between –20 and +50 °C (–4 to +122 °F).
Before shipping the Bird Cal Cart, take the following precautions:
Remove the vent plugs and replace them with the shipping plugs. Wrap the vent plugs with padding and
tape them to the side of the load for protection.
Note: With the shipping plugs installed, it is not necessary to empty out the coolant.
Repack in the original carton or contact Bird for a transit case.
44
Page 56
Thruline RF Power Meter, Model 4421
Specifications, Cal Cart
To determine the model numbers of the Cal Cart components, refer to specific cal cart manual. For load specifications, refer to the load manual included.
Note: Certain sensors or connectors may be incompatible with some models. Refer to specific cal cart manual for a list of available components, or contact Customer Service.
Max. Power
Under normal operating conditions, the Cal Cart’s maximum power is the MINIMUM of the max. power of the sensor and the max. power of the load. The following modifiers apply:
Above 65 MHz, derate the load’s max. power. See
Figure 22 on page 51
.
Due to the wide variety of available connectors, frequency range and maximum power may be reduced.
Insertion loss is specified with female N connectors. Choose connectors appropriate for the frequency and power of operation.
Derate RF power by 2.5% for every 305m (1,000 ft.) above 1,520m (5,000 ft.).
Table 20 Bird Cal Cart
Frequency Range Sensor dependent, 30 MHz max. Power Range Dependent on sensor, sensor connectors, and load
Connectors
Dimensions
Short Cart Long Cart
Weight, Nominal
Short Cart Long Cart
Temperature Range
Operating
Storage Altitude 1520 m (5000 ft.) Humidity 85% noncondensing max
Calibration Cycle, Nominal
CE
Customer specified (See
page 52
42”L x 20”W x 42”H (1067 x 508 x 1067 mm) 52”L x 20”W x 42”H (1321 x 508 x 1067 mm)
175 lbs. (80 kg) 290 lbs. (114 kg)
0 to +35 °C (32 to +95 °F) – 20 to +70 °C (– 4 °F to +158 °F)
1 year CE Compliant. Refer to Declaration of Conformity for
specific standards.
)
"Available Connectors" on
45
Page 57
Specifications, 4421 RF Power Meter
Table 21 Bird 4421 RF Power Meter
Frequency Range Sensor dependent Power Range Sensor dependent VSWR Display 1.0 – 199.9 max Return Loss Display 0 to 40 dB max Display Accuracy ± 1 on least significant digit AC Power 115 / 230 Vac @ 50 / 60 Hz; 60 W Batteries 8 Nickel Metal Hydride 1.2 v cells, rechargeable
Battery Life
Battery Charger
Display
Interfaces
Fuse Rating IEC (5 x 20 mm) Time Lag Type T 115 VAC
230 VAC
CE
Humidity 95% max. (noncondensing) Altitude Up to 10,000 feet (3,000 m) Temperature Range Operating
Storage Dimensions
2016 and prior
2016 and newer Weight, Nominal 9.5 lbs. (4.3 kg)
Approximately 8 hours continuous usage (reduced after high temperature charging)
Built-in battery charger. Drained batteries require approximately 28 hours to recharge.
LCD, 4½ digit display. Indicates mode, measurement units, battery condition, remote status, and signal increase / decrease. Self contained backlight.
IEEE-488 GPIB RS-232
T630 mA T315 mA
CE Compliant. Refer to Declaration of Conformity for specific standards.
0 to 50 °C (32 to 122 °F) –20 to +50 °C (–4 to +122 °F)
12.3”L x 12.2”W x 4.2”H (312 x 309 x 103 mm)
10.7”L x 12.2”W x 4.2”H (271 x 309 x 103 mm)
Maintenance
Changing the sensor’s connectors will invalidate calibration data, and may reduce the
CAUTION
maximum power rating of the unit.
46
Page 58
Thruline RF Power Meter, Model 4421
Specifications, Power Sensors
Table 22 Specifications Common to all Sensors
Impedance, Nominal 50 ohms Max. Allowable Terminating VSWR 2.00:1
Frequency-specific calibration factors stored in
Calibration Technique
nonvolatile memory in each sensor. Sensor output corrected for frequency and temperature within specified ranges.
Calibration Cycle, Nominal
4027A and 4027F, 6 months All other models, 1 year
Calculated from FWD accuracy and FWD power
Accuracy, Reflected
RFL Accuracy FWD Accuracy
Calculated from FWD and RFL power
Accuracy, VSWR
Sampling Rate, Nominal 2 readings / second Operating Power Supplied by power meter via sensor cable
Connectors
4028B10M 4028C10M 4028A Series All other models
CE
1-5/8” EIA Flanged 3-1/8” EIA Flanged 7-16 DIN, LC, HN, or 7/8” Customer specified from QC list, appropriate for frequency and power.
CE Compliant. Refer to Declaration of Conformity
for specific standards. Humidity, Max. 95% (noncondensing) Altitude, Max. 10,000 feet (3,000 m) Temperature Range
Operating Storage
Dimensions
4028B3M, 4028B10M 4028A Series All other models
Weight, Nominal
4028B3M, 4028B10M, 4028C10M
4028A2M, 3M, 4M, 10M, and 25M
All other models
0 to 50 °C (32 to 122 °F)
–20 to +70 °C (–4 to +158 °F)
6.75”L x 3.5”W x 4.75”H (175 x 89 x 121 mm)
4.7”L x 3.2”W x 3.8”H (120 x 82 x 97mm)
5.2”L x 2.5”W x 3.25”H (137 x 64 x 83 mm)
5 lb. 2 oz. (2.33 kg)
3 lb. 5 oz. (1.5 kg)
1 lb. 13 oz. (0.8 kg)
VSWR 1
  
FWD Power
---------------------------------+= 10

P
R
1
=
-------+

P

F
Directivity 10/
P
R
-------
P
F
47
Page 59
Table 23 Bird 4020 Series RF Power Sensors
Maintenance
RF Power Range
4021, 4022 4024, 4025
Frequency Range
4021 4022 4024 4025
Accuracy, Fwd, Best Case
VSWR, Max.
4021, 4024, 4025 4022
Insertion Loss, Max.
4021, 4024, 4025 4022
Directivity, Min.
4021 4022
4024 4025
† For rated accuracy, no more than 1% AM; Harmonics –50 dBc or less
Derate accuracy by 3.0% (1
300 mW – 1 kW 3 W – 10 kW
1.8 – 32 MHz 25 MHz – 1 GHz
1.5 – 32 MHz 100 kHz – 2.5 MHz
± 3% (1)
1.05:1
1.05:1, 25 – 512 MHz and 1.10:1, 512 MHz – 1 GHz
0.05 dB
0.05 dB, 25 – 512 MHz and 0.13 dB, 512 MHz – 1 GHz
30 dB 26 dB (28 dB typ), 400 – 1000 MHz 30 dB, 25 – 400 MHz 28 dB, 1.5 – 2.5 and 25 – 32 MHz and 30 dB, 2.5 – 25 MHz 28 dB, 100 – 125 kHz and 30 dB, 125 – 2500 kHz
) below 15 °C and above 35 °C
CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may reduce the
maximum power rating of the unit.
Table 24 Bird 4027A Series RF Power Sensors
Frequency Range 4027A250K 250 – 400 kHz
4027A400K 400 – 550 kHz 4027A800K 800 – 950 kHz 4027A2M 1.5 – 2.5 MHz 4027A4M 3 – 5 MHz 4027A10M 10 – 15 MHz
RF Power Range 4027A12M 300 mW – 1 kW 4027A25M 3 W – 9 kW 4027A35M 3 W – 7.5 kW 4027A60M 3 W – 6 kW
Accuracy, Fwd, Best Case
Calibration Frequencies, Typical (MHz)
4027A250K 0.25, 0.40 4027A400K 0.40 4027A800K 0.90 4027A2M 1.8, 2.0, 2.17 4027A4M 4.0, 5.0 4027A10M 10.0, 13.56, 15.0
Calibration Power, Typical
4027A12M All other models
VSWR, Max. Insertion Loss, Max. 0.05 dB (with female “N” connectors)
††
4027A12M 10 – 15 MHz 4027A25M 25 – 30 MHz 4027A35M 35 – 45 MHz 4027A60M 45 – 65 MHz 4027A100M 95 – 105 MHz 4027A150M 150 – 170 MHz
4027A100M 3 W – 4 kW 4027A150M 3.75 W – 3.75 kW All other models 3 W – 10 kW
± 1.0% (1)
4027A12M 10.0, 13.56, 15.0 4027A25M 25.76, 27.12, 28.48 4027A35M 40.68 4027A60M 55.0, 60.0 4027A100M 95.0, 100.0 4027A150M 162.0
700 W
1.7 kW
1.05:1
48
Page 60
Thruline RF Power Meter, Model 4421
Directivity, Min.
4027A12M All other models
† For rated accuracy, no more than 1% AM; Harmonics –50 dBc or less
Derate accuracy by 1% (1 Derate accuracy by 1% (1
†† Other calibration frequencies available upon request
) outside cal. power or cal. frequency ) below 15 °C and above 35 °C
Table 25 Bird 4027F Series RF Power Sensors
Frequency Range
4027F2M 4027F10M 4027F60M
RF Power Range
4027F2M, 4027F10M 4027F60M
Accuracy, Fwd, Best Case ± 1.0% (2 Calibration Frequencies, Typical
4027F2M 4027F10M 4027F60M
Calibration Power, Typical 1.7 kW Harmonic Rejection, Min.
4027F2M 4027F10M 4027F60M
Low Frequency Rejection, Min.
4027F10M 4027F60M
Max Error Induced by 10% AM
4027F2M, 4027F10M 4027F60M
VSWR, Max. 1.05:1 Insertion Loss, Max. 0.05 dB (with female “N” connectors) Directivity, Min. 28 dB
1.8 – 2.2 MHz 12 – 15 MHz 57 – 63 MHz
0.1 – 10 kW
0.1 – 3 kW
1.8, 2.0, 2.17 MHz
12.0, 12.5, 13.56, 14.0, 15.0 MHz
57.0, 58.5, 60.0, 61.5, 63.0 MHz
26 dB @ 3.6 – 3.8 MHz, 30 dB @ > 3.8 MHz 30 dB @ > 25 MHz 30 dB @ > 114 MHz
30 dB @ < 1 MHz 30 dB @ < 15 MHz
0.2% @ < 5 kW, 1.0% @ 5 – 10 kW
0.2% @ < 1.5 kW, 1.0% @ 1.5 – 3 kW
30 dB 28 dB
)
† Other calibration frequencies available upon request
Table 26 Bird 4028A and 4028B Series RF Power Sensors
Frequency Range
4028A250K 4028A400K 4028A2M 4028A3M 4028A4M 4028A10M, 4028B10M, 4028C10M 4028A25M 4028B3M
RF Power Range
4028A250K, 4028A400K 4028C10M
All other models
Accuracy, Fwd, Best Case
49
250 – 400 kHz 400 – 550 kHz
1.5 – 2.5 MHz
2.5 – 3.5 MHz
3.5 – 4.5 MHz
10 – 15 MHz 25 – 30 MHz
2.5 – 4 MHz
1 kW – 20 kW 500W – 50 kW
1 kW – 25 kW ± 2.0% (2)
Page 61
Maintenance
Calibration Frequencies, Typical (MHz)††
4028A250K 4028A400K 4028A2M 4028A3M 4028A4M 4028A10M, 4028B10M, 4028C10M 4028A25M 4028B3M
Calibration Power, Typical 3.5 kW VSWR, Max. 1.05:1 Insertion Loss, Max. 0.05 dB (with female 7-16 DIN connectors) Directivity, Min. 28 dB
† For rated accuracy, no more than 1% AM; Harmonics –50 dBc or less
Derate accuracy by 2% (2 Derate accuracy by 2% (2
†† Other calibration frequencies available upon request
Specifications, RS-232 Interface
Table 27 RS - 232 Interface
Logic Levels Meets all EIA Standard RS-232-C specifications Modes of Operation Switch and bus selectable Talk Always Allows the 4421 to send to the bus keyboard-initiated
Addressable Allows the 4421 to be commanded by an RS-232 interface controller Connector RS-232 Interface Type Humidity, Max 95% (non-condensing) Altitude, Max 10,000 feet (3,000 m) Temperature Range
Operating Storage
Dimensions 6.5”L x 4.5” W (165 x 115 mm) Weight 0.5 lbs (0.23 kg) nominal
measurements only
0 to 50º C (32 to 122º F) –20 to +50º C (–4 to +122º F)
0.25, 0.40
0.40
1.8, 2.0, 2.17
2.5, 3.2, 3.5
3.5, 4.0
10.0, 13.56, 15.0
25.76, 27.12, 28.48
2.5, 3.2, 3.5, 4.0
) outside cal. power or cal. frequency ) below 15 °C and above 35 °C
Specifications, IEEE-488 GPIB Interface
Table 28 IEEE - 488 Interface
Logic Levels Meets all IEEE Standard 488-1978 specifications Modes of Operation Switch and bus selectable Talk Only Allows the 4421 to send to the bus keyboard-initiated measurements
Addressable Allows the 4422 to be addressed as talker or listener under the
Connector Standard IEEE-488 bus type Humidity, Max 95% (non-condensing) Altitude, Max 10,000 feet (3,000 m) Temperature Range
Operating Storage
Dimensions 6.5”L x 4.5” W (165 x 115 mm) Weight, Nominal 0.5 lbs (0.23 kg)
only
command of an IEEE-488 bus controller
0 to 50º C (32 to 122º F) –20 to +50º C (–4 to +122º F)
50
Page 62
Thruline RF Power Meter, Model 4421
Figure 22 High Frequency Derating 1
Figure 23 High Frequency Derating 2
51
Page 63
Replacement Parts
Table 29 Replacement Parts
Description Qty Part Number
Fuse, IEC (5 x 20 mm) Time Lag Type T
115 VAC, T630 mA 230 VAC, T315 mA
Cord, AC Power
115 VAC
230 VAC Harmonized Plug, 115 VAC 1 5A2626 Cable, Sensor 1 4421-038 Grommet 1 4421A372-1 Cordreel (Cal Cart) 1 4421A383 Casters (Cal Cart) 4 4421A384 Handle (Cal Cart) 1 4421A385 Cable, IEEE-488
2 m
1 m Cable, RS-232
10 ft.
5 ft. Null Modem Kit (RS-232 only) 1 4380-250 Panel Mount Kit 1 4421-250 Battery, C size, NiMH 8 5A1230
Maintenance
1
1
1
1
5A2257-14 5A2257-11
5-1286 5A2416
5-1317-2 5-1317-1
5-1662-2 5-1662-1
Available Connectors
Changing the sensor’s connectors will invalidate calibration data, and may reduce the
CAUTION
maximum power rating of the unit.
Connector P/N
HN (F) 4240-268 LC (F) 4240-031 BNC-Female 4240-125 BNC-Male 4240-132 C-Female 4240-100 C-Male 4240-110 HN-Female 4240-268 HN-Male 4240-278 LC-Female 4240-031 LC-Male 4240-025 Open Term. # 10-32 Nut 4240-080 N (F) 4240-062
52
Page 64
Thruline RF Power Meter, Model 4421
7/16 Jack, IEC Type 169-4 4240-344 LT-Female 4240-018 LT-Male 4240-012 N-Female 4240-062 N-Male 4240-063 SC-Female 4240-090 SMA-Female 4240-336 SMA-Male 4240-334 7/16 Jack, IEC Type 169-4 4240-344 7/16 Plug, IEC Type 169-4 4240-363 TRU 6934 (F) 4240-371 TRU 7958 (F) 4240-372 Mini UHF-Female 4240-346 UHF-Female 4240-050 UHF-Male 4240-179 1-5/8” EIA Fixed 4240-096 1-5/8” EIA Swivel 4240-208 7/8” EIA 4240-002 TNC-Female 4240-156 TNC-Male 4240-160
Customer Service
Any maintenance or service procedure beyond the scope of those in this chapter should be referred to a qualified service center.
If the unit needs to be returned for any reason, request an Return Material Authorization (RMA) through the Bird Technologies website. All instruments returned must be shipped prepaid and to the attention of the RMA number.
Bird Service Center
30303 Aurora Road Cleveland (Solon), Ohio 44139-2794 Fax: (440) 248-5426 E-mail: bsc@birdrf.com
For the location of the Sales Office nearest you, visit our Web site at:
http://www.birdrf.com
53
Page 65
Limited Warranty
All products manufactured by Seller are warranted to be free from defects in material and workmanship for a period of one (1) year, unless otherwise specified, from date of shipment and to conform to applicable specifica­tions, drawings, blueprints and/or samples. Seller’s sole obligation under these warranties shall be to issue credit, repair or replace any item or part thereof which is proved to be other than as warranted; no allowance shall be made for any labor charges of Buyer for replacement of parts, adjustment or repairs, or any other work, unless such charges are authorized in advance by Seller.
If Seller’s products are claimed to be defective in material or workmanship or not to conform to specifications, drawings, blueprints and/or samples, Seller shall, upon prompt notice thereof, either examine the products where they are located or issue shipping instructions for return to Seller (transportation charges prepaid by Buyer). In the event any of our products are proved to be other than as warranted, transportation costs (cheapest way) to and from Seller’s plant, will be borne by Seller and reimbursement or credit will be made for amounts so expended by Buyer. Every such claim for breach of these warranties shall be deemed to be waived by Buyer unless made in writ­ing within ten days from the date of discovery of the defect.
The above warranties shall not extend to any products or parts thereof which have been subjected to any misuse or neglect, damaged by accident, rendered defective by reason of improper installation or by the performance of repairs or alterations outside of our plant, and shall not apply to any goods or parts thereof furnished by Buyer or acquired from others at Buyer’s request and/or to Buyer’s specifications. Routine (regularly required) calibration is not covered under this limited warranty. In addition, Seller’s warranties do not extend to the failure of tubes, tran­sistors, fuses and batteries, or to other equipment and parts manufactured by others except to the extent of the original manufacturer’s warranty to Seller.
The obligations under the foregoing warranties are limited to the precise terms thereof. These warranties provide exclusive remedies, expressly in lieu of all other remedies including claims for special or consequential damages. SELLER NEITHER MAKES NOR ASSUMES ANY OTHER WARRANTY WHATSOEVER, WHETHER EXPRESS, STATUTORY, OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY AND FITNESS, AND NO PERSON IS AUTHORIZED TO ASSUME FOR SELLER ANY OBLIGATION OR LIABILITY NOT STRICTLY IN ACCORDANCE WITH THE FOREGOING.
Special Lifetime Warranty - Series 4020, Series 4027A,
Series 4027F, and Series 4028 Power Sensor Head
In addition to its standard warranty, the Bird Electronic Corporation warrants its Series 4020, Series 4027A, Series 4027F, and Series 4028 Thruline Power Sensor Heads for lifetime to original purchaser. This extended warranty is against burnout. For the warranty to apply, the Sensor Head must be used with the correct Bird Electronic Corpora­tion Display Unit, the maximum power rating of the Sensor must not be exceeded, the Sensor RF circuit must be properly terminated and the Sensor not subjected to physical abuse.
Bird Electronic Corporation, at its option, will repair or replace the defective Sensor at its world Headquarters at 30303 Aurora Road, Solon, Ohio 44139.
The customer is responsible to pay transportation charges to return the defective sensor to Bird.
54
Loading...