OPERATION
The Mini PowerZone II unit is supplied with a timer. The timer allows the user to set the treatment time, start the
treatment, and return after the area has been treated.
Setting the Mini PowerZone Timer Plug unit into a 10amp power point
Fig:1
Press the Up Arrow (Fig: 1) once to increase time by 10 minute steps.
Press and hold the Up Arrow to rapidly increase the time. Press and hold the Up
Arrow until the display shows: “C:_:_ _ _” to operate the unit continuously.
Fig:2
Press and hold the Down Arrow (Fig: 2) to rapidly decrease time.
Press and hold the Down Arrow until the display shows: “10.00.0” to
set the minimum time of 10 minutes.
Fig:3
Start the Mini PowerZone timer and Mini PowerZone II
operation by pressing the ON/OFF button (Fig: 3). Pause the
MPZ timer and stop the Mini PowerZone II operation by
pressing the ON/OFF button while the unit is running.
Note: When the Mini PowerZone is plugged in, the timer
remembers the previously used time setting.
MAINTENANCE
The grills/ air vents and fan must be kept free and clear from dust and dirt.
The only serviceable / changeable part is the lamp. If the device is located in a highly contaminated greased,
dusty, humid or smoke filled area, it is recommended that the lamp and plastic end grilles be cleaned as
required. This will significantly lengthen the life of the lamp.
The lamps must be changed maximum 8000 hours, or once a year if continuous operation
Hourly Lamp data available on display
Fig:4
Press the up and down arrows (Fig:4) simultaneously for 5
seconds to show the number of hours of operation (dXXXX)
and the number of on/off cycles (CXXXX).
Eg. (d 1245 , C 896) = 1245 hours of operation, 896
on/off cycles.
Lamp Cleaning and Replacement Procedure – See website www.biozonescientific.com.au for detail
1) Disconnect the power supply from the Mini PowerZone & Allow 10 minutes for the lamp(s) to cool.
2) Unlock and open the cover with the BioZone Key, (separate the cover from the base with the flat section of this key).
3) Unplug the lamp connector(s) from the lamp(s)
4) Remove the lamp(s) from the lamp holders.
5) Wipe and clean the lamp(s) surface with an alcohol wipe.
6) Place the cleaned lamp or new lamp into the lamp holders.
7) Connect the lamp connector observing the correct orientation of the pins
8) Re-install the cover and press firmly on the top near the lock to ensure it
is completely closed. (Ensure that the lamp connector wires do not interfere with the
cover closing)
9) Close and lock the Mini PowerZone with the BioZone Key.
(Make sure the ring around the lock lights up when turned back on after installation)
WARRANTY - 12 Months Parts & Labour (Return to Base)
The provisions of this warranty are in lieu of any other written warranty, whether expressed or implied, written or oral, including any warranty
of merchantability or fitness for a particular purpose. The manufacturer's maximum liability shall not exceed the actual purchase price paid
by you for the product. In no event shall the manufacturer be liable for special, incidental, consequential or indirect damages. This warranty
does not cover a defect that resulted from improper or unreasonable use or maintenance, accident, improper packing, or unauthorised
tampering, alteration or modification as determined solely by us. This warranty is void if label bearing the serial number has been removed
or defaced.
Please contact BioZone Scientific for replacement lamps. To reset these indicators, put the end of a
paperclip into the small hole above the ON/OFF button for 3 seconds.