BIOTRONIK SE and KG TACHNXT Users Manual

Iforia
( )
ICD Family • Tachyarrhythmia Therapy • Cardiac Resynchronization Therapy
xxx xxx ICD Familie Tachyarrhythmietherapie Kardiale Resynchronisationstherapie xxx xxx xxx xxx xxx xxx xxx xxx
Technical Manual
Technická příručka
Brugermanual
Gebrauchsanweisung
Käyttöohje
Manuel technique
Manuale tecnico di istruzione
Gebruikshandleiding
Instrukcja obsługi
Bruksanvisning
• en
• cs
• da
• de
• es
• 
• fr
• it
• nl
• pl
• pt
• sv
0123
© BIOTRONIK SE & Co. KG All rights reserved. Specications subject to modication, revision and improvement.
® BIOTRONIK Home Monitoring, (ProMRI), IEGM-Online HD and SMART Detecton are registered trademarks of BIOTRONIK SE & Co. KG
0681 2012
12-D-xx Revision: B (2012-xx-xx)
BIOTRONIK SE & Co. KG Woermannkehre 1 12359 Berlin · Germany Tel +49 (0) 30 68905-0 Fax +49 (0) 30 6852804 sales@biotronik.com www.biotronik.com
en • English ................................................................................................................................................................. 2
de • Deutsch ................................................................................................................................................................ 24
393467--B
en • English

Product Description

Intended Medical Use

Intended use

Iforia 3/5/7 is part of a family of implantable cardioverter-defibrillators (ICDs). Primary objective of the therapy is to prevent sudden cardiac death. Furthermore, the device is capable of treating bradycardia arrhythmias and cardiac resynchronization therapy with multisite ventricular pacing.
The implantation of an ICD is a symptomatic therapy with the following objectives:
Termination of spontaneous ventricular fibrillation (VF) through shock delivery
Termination of spontaneous ventricular tachycardia (VT) through antitachycardia pacing (ATP); in case of ineffective ATP or hemodynamically not tolerated VT, with shock delivery
Cardiac resynchronization through multisite ventricular pacing (triple-chamber devices)
Compensation of bradycardia through ventricular (single-chamber devices) or AV sequential pacing (DX, dual- and triple-chamber devices)

Diagnosis and therapy forms

The device monitors the heart rhythm and automatically detects and terminates cardiac arrest resulting from ventricular tachyarrhythmia. All major therapeutic approaches from the field of cardiology and electrophysiology are included. BIOTRONIK Home Monitoring® enables physicians to perform therapy management at any time.

Required expertise

In addition to having basic medical knowledge, the user must be thoroughly familiar with the operation and the operation conditions of a device system.
Only qualified medical specialists having this special knowledge required are permitted to use implantable devices.
If users do not possess this knowledge, they must be trained accordingly.

Indications

Iforia can treat life-threatening ventricular arrhythmias with antitachycardia pacing and defibrillation.
Generally approved differential diagnostics methods, indications, and recommenda­tions for ICD therapy apply to BIOTRONIK devices. See the guidelines of cardiology associations for guidance. We recommend observing the indications published by the German Cardiac Society (Deutsche Gesellschaft für Kardiologie, Herz- und Kreislaufforschung) and the ESC (European Society of Cardiology). This also applies to the guidelines published by the Heart Rhythm Society (HRS), the American College of Cardiology (ACC), the American Heart Association (AHA), and other national cardiology associations.

Single-chamber and dual-chamber

Single-chamber and dual-chamber ICDs are indicated for patients with the following risk:
Sudden cardiac death caused by ventricular arrhythmias

Triple-chamber

Triple-chamber ICDs are indicated for patients with the following risks:
Sudden cardiac death caused by ventricular arrhythmias
Congestive heart failure with ventricular asynchrony
Iforia is also indicated for primary prophylaxis in congestive heart failure patients.

Contraindications

Known contraindications:
Tachyarrhythmia caused by temporary or reversible irritation, e.g. poisoning, electrolyte imbalance, hypoxia, sepsis or acute myocardial infarction
Such frequent VT or VF that the therapies would cause an unacceptably rapid depletion of the device batteries
VT with few or without clinically relevant symptoms
VT or VF treatable by surgery
Concomitant diseases that would substantially limit a positive prognosis
Accelerated idioventricular rhythm

System Overview

Device family

The complete Iforia 3/5/7 device familyconsists of several device types with a DF-1/IS-1connection. Single-chamber: VR-T and VR-T DX (only devices with a DF-1/IS-1 connection); dual-chamber: DR-T; triple-chamber: HF-T. Not all device types are available in every country.
2

Device

DF-1 RV
DF-1
SVC
IS-1
RV
DF-1
RV
DF-1
SVC
IS-1
RA
IS-1
RV
DF-1
RV
DF-1
SVC
IS-1
RA
IS-1
RV
DF-1
RV
DF-1
SVC
IS-1
RA
IS-1
RV
IS-1
LV
The device's housing is made of biocompatible titanium, welded from outside and thus hermetically sealed. The ellipsoid shape facilitates implantation in the pectoral muscle area. The connections for bipolar pacing and sensing (and unipolar connections for the triple-chamber device) as well as for shock delivery are found in the device header. The housing serves as a potential antipole during shock delivery or in the case of unipolar lead configuration.

DF-1/IS-1 lead connection

The device labeling provides information pertaining to possible lead connections depending on the device type and pertaining to connection assignment:
VR DX DR HF
Connector
Lead
port
Configuration Implantation site Device type
connector
RV DF-1 Shock coil Right ventricle VR, DX, DR, HF SVC DF-1 Shock coil Superior vena cava VR, DX, DR, HF RA IS-1 Bipolar Atrium DX, DR, HF (R)V IS-1 Bipolar (Right) ventricle VR, DX, DR, HF LV IS-1 Unipolar, Bipolar Left ventricle HF

Leads

BIOTRONIK leads are sheathed with biocompatible silicone. They can be flexibly maneuvered, are stable long-term, and are equipped for active or passive fixation. They are implanted using a lead introducer set. Some leads are coated with polyurethane which is known to increase the sliding properties for the lead. Leads with steroids reduce inflammatory processes. The fractal design of the electrodes provides for low pacing thresholds. BIOTRONIK provides adapters to connect already implanted leads to new devices.
en • English

Telemetry

Telemetric communication between the device and the programmer can be carried out following initialization either by applying the programming head (PGH) to the device or by using radio frequency (RF) telemetry in the programmer. BIOTRONIK calls this func­tion SafeSync®.

Programmer

Implantation and follow-up are performed with BIOTRONIK's portable programmer. There is one with integrated RF telemetry and one with a separate SafeSync Module. The programmer is used during implantation to transfer the current device program to the device. The pacing thresholds can be determined and all tests can be performed during in-office follow-up. In addition to this, the programmer is used to set mode and parameter combinations, as well as for interrogation and saving of data from the device. Leadless ECG, IEGM, markers and functions are displayed simultaneously on the color display.

Modes

The mode setting depends on the individual diagnosis:
Device type Modes
VR VVI; VVIR; V00; OFF DX VDD; VDDR; VDI; VDIR; VVI; VVIR; V00; OFF DR, HF DDD; DDDR; DDI; DDIR; VDD; VDDR; VDI; VDIR
VVI; VVIR; AAI; AAIR; V00; D00; OFF

NBD and NBG codes

VVE is the NBD code for the antitachycardia mode of the single-chamber, dual­chamber, and triple-chamber devices:
V Shock in the ventricle V Antitachycardia pacing (ATP) in the ventricle E Detection via IEGM analysis
DDDR is the NBG code for the antibradycardia mode of the dual-chamber device:
D Pacing in the atrium and ventricle D Sensing in the atrium and ventricle D Pulse inhibition and pulse triggering R Rate adaptation
3
DDDRV is the NBG code for the antibradycardia mode of the triple-chamber device:
D Pacing in the atrium and ventricle D Sensing in the atrium and ventricle D Pulse inhibition and pulse triggering R Rate adaptation V Multisite pacing in both ventricles
VDDR is the NBG code for the antibradycardia mode of the single-chamber DX device:
VVentricular pacing D Sensing in the atrium and ventricle D Pulse inhibition and pulse triggering R Rate adaptation
VVIR is the NBG code for the antibradycardia pacing modes of the single-chamber device:
VVentricular pacing V Sensing in the ventricle I Pulse inhibition in the ventricle R Rate adaptation
BIOTRONIK Home Monitoring
In addition to effective pacing therapy, BIOTRONIK provides a complete therapy management system:
With Home Monitoring, diagnostic and therapeutic information as well as technical data are automatically sent to a stationary or mobile transmitter via an antenna in the device header. The data are encrypted and sent from the transmitter to the BIOTRONIK Service Center via the cellular phone network.
The received data are deciphered and evaluated. Each physician can set the criteria for evaluation to be used for each patient and can configure the time of notification via E-mail, SMS or fax.
A clear overview of the results of this analysis is displayed for the attending physi­cians on the protected Internet platform Home Monitoring Service Center (HMSC).
®
Data transmission from the device is performed with a daily device message.
Device messages, which indicate special events in the heart or in the device, are forwarded immediately.
A test message can be initiated at any time using the programmer to immediately check the Home Monitoring function.

Technical manuals

The following technical manuals provide information about usage of the device systems:
Technical manual for the device
Technical manual for the HMSC
Technical manuals for the programmer and the SafeSync Module
Technical manual for device programs as online help on the user interface and as a PDF file in the Manual Library at www.BIOTRONIK.com
Technical manuals for the leads
Technical manuals for cables, adapters and accessories

Order numbers for Iforia with DF-1/IS-1 connection

Not all device types are available in all countries:
VR-T 383586 383583 390081 VR-T DX 383597 390093 DR-T 383570 383567 390067 HF-T 383554 383551 390054

Order numbers for Iforia ProMRI with DF-1/IS-1 connection

Not all device types are available in all countries:
VR-T 390119 390083 VR-T DX 390123 390095 DR-T 390115 390069 HF-T 390111 390056
Iforia 3 Iforia 5 Iforia 7
Iforia 5 ProMRI Iforia 7 ProMRI
4

Scope of delivery

The storage package includes the following:
Sterile container with device
Serial number label
Patient ID card
Warranty booklet
Technical manual for the device
The sterile container includes the following:
Device, blind plug DF-1 (if applicable) and blind plug IS-1 for device type HF
Screwdriver

Therapeutic and Diagnostic Functions

Diagnostic functions

Data from implantation and the most recent interrogations and follow-ups are recorded as well as arrhythmia episodes; they are stored together with other data to assess patients and the state of the device at any time.
To check the lead for proper functioning, an automatic impedance measurement using subthreshold pacing pulses is performed in the device.
Leadless ECG function: For all device types, far-field derivation can be measured without external leads between the right ventricular shock coil and housing, which, depending on the implantation site, corresponds to ECG derivation II or III (Einthoven).
Once a telemetry connection has been established during a test procedure in an in-office follow-up, the leadless ECG and the IEGM are displayed with markers.

Antitachycardia pacing

The ICD can treat ventricular tachycardia with antitachycardia pacing (ATP); ATP can also be delivered in the VF zone (ATP One Shot) when the stability criterion indi­cating that this will be effective before shock delivery (monomorphic rapid VTs) is met.
Depending on the device type, the device program contains not only the ICD func­tions but also all pacemaker functions for 1, 2, or 3 chambers. The heart rhythm is continuously monitored; each arrhythmia is classified according to the heart rate and the adjustable detection criteria. Depending on the preset values, antibrady­cardia as well as antitachycardia therapy is inhibited or delivered.
en • English

Cardioversion, defibrillation

The ICD can treat ventricular tachyarrhythmia with cardioversion and/or defibrilla­tion. Shock polarity and energy can be programmed individually. Shock energies between 2.0 and 40 J are possible. Before delivery of the shock, the ICD can be set to only deliver a shock when ongoing tachyarrhythmia is confirmed; during this time period the device can identify spontaneous conversion of the tachyarrhythmia and cancel the charging process if necessary.
The shock paths can be set between the different shock coils (SVC/RV) and/or the housing.

Antibradycardia pacing and CRT

Innovative rate hystereses, automatic sensor functions, and a night program promote the patient's intrinsic rhythm, avoid overdrive pacing, and facilitate adap­tation of the device to the individual needs of the patient.
Setting an upper tracking rate for the atrium prevents unspecific atrial pacing, thus reducing the risk of pacemaker-mediated tachycardia.
Positive AV hysteresis functions support the intrinsic conduction and thus the natural contraction sequence. Negative AV hysteresis functions support the cardiac resynchronization therapy by maintaining pacing in stressful situations.
For resynchronization of the ventricles, triple-chamber devices have functions for multisite ventricular pacing with possible VV delays in either direction.
To ensure that no additional surgery is necessary in case of a left-sided increase of pacing threshold or undesired phrenic nerve stimulation, different paci ng polarities can be set for the left ventricular lead with a triple-chamber device.
Automatic active capture control is available for the right and left ventricle with automated tracking of the pacing threshold or automatic threshold monitoring (ATM) for trend analysis.
With the Iforia 3, the pacing threshold is only measured.

Storing programs

The parameter settings can be saved in 3 individual therapy programs.

Home Monitoring functions

The device automatically sends information to the transmitter once a day. It also sends messages related to events, which are immediately forwarded to the Service Center. In addition to this, test messages can be initiated using the programmer.
Appointments for Home Monitoring-supported follow-ups can be scheduled via the HMSC. This applies to Iforia 5/7.
5
Important medical information in the device messages include the following: —
Atrial and ventricular arrhythmias
Parameters relevant to leads in the atrium and ventricle: pacing thresholds, sensing amplitudes, impedances
Current statistics
IEGM online HD® with up to 3 high definition channels

General Safety Instructions

Operating Conditions

Care during shipping and storage

Devices are not to be stored or transported close to magnets or sources of electro­magnetic interference.
Note the effects of the storage duration; see Battery Data.

Delivery in shipment mode

The device is delivered in shipment mode to protect the battery; capacitor reforming required during storage could result in controlled extended charge times of the shock capacitors.
The shipment mode is displayed on the programmer after loading the device program (it is deactivated during implantation on initial measurement of the pacing impedance).

Temperature

Extremely low and high temperatures affect the service time of the battery in the device.
Temperatures of 5°C to 45°C are permitted for transport, storage, and use.

Sterile delivery

The device and the screwdriver have been gas-sterilized. Sterility is guaranteed only if the blister and quality control seal have not been damaged.

Sterile container

The device and screwdriver are packaged in two separately sealed blisters. The inner blister is also sterile on the outside so that it can be transferred in a sterile state during implantation.

Single use only

The device and screwdriver are intended for single use only.
Do not use the device if the package is damaged.
The device must not be resterilized and reused.

Possible Complications

General information on medical complications

Complications for patients and device systems generally recognized among practitio­ners also apply to BIOTRONIK devices.
Normal complications may include fluid accumulation within the device pocket, infections, or tissue reactions. Primary sources of complication information include current scientific and technological knowledge.
It is impossible to guarantee the efficacy of antitachycardia therapy, even if the programs have proven successful during tests or subsequent electrophysiological examinations. In rare cases the set parameters may become ineffective. It is possible for therapies to induce or accelerate tachycardia and cause sustained ventricular flutter or fibrillation.

Skeletal myopotentials

Bipolar sensing and control of sensitivity are adapted by the device to the rate spectrum of intrinsic events so that skeletal myopotentials are usually not recorded. Skeletal myopotentials can nonetheless be classified as intrinsic events especially at very high sensing sensitivity and, depending on the interference, may cause inhibition or anti­arrhythmia therapy. In the case of undesired myopotentials, the device switches to asynchronous pacing if the interference rate is exceeded.

Possible technical failures

Technical failure of a device system cannot be entirely ruled out. Possible causes can include the following:
Lead dislodgement, lead fracture
Insulation defects
Device component failures
Battery depletion
Interrupted telemetry
6

Electromagnetic interference (EMI)

Any device can be sensitive to interference if external signals are sensed as intrinsic rhythm or if measurements prevent rate adaptation.
BIOTRONIK devices have been designed so that their susceptibility to EMI is min­imal.
Due to the intensity and variety of EMI, there is no guarantee for safety. It is gener­ally assumed that EMI produces only minor symptoms, if any, in patients.
Depending on the pacing mode and the type of interference, sources of interference may lead to pulse inhibition or triggering, an increase in the sensor-dependent pacing rate or asynchronous pacing.
Under unfavorable conditions, for example during therapeutic or diagnostic proce­dures, interference sources may induce such a high level of energy into the pacing system that the cardiac tissue surrounding the lead tip is damaged.

Device behavior in case of EMI

In case of electromagnetic interference, the device switches to asynchronous pacing for as long as the interference rate is exceeded.

Static magnetic fields

The reed switch in the device closes starting at a field strength of 1.8 mT. The reed switch opens if the magnetic field falls below 1 mT.

Possible Risks

Contraindicated procedures

The following procedures are contraindicated as they may cause harm to the patient or damage the device and, as a result, put the system functionality at risk:
Therapeutic ultrasound: Harm to the patient via excess warming of body tissue near the device system
Transcutaneous electrical nerve stimulation
Hyperbaric oxygen therapy
Applied pressures higher than normal pressure

Risky therapeutic and diagnostic procedures

If electrical current from an external source is conducted through the body for diag­nostic or therapeutic purposes, then the device can be subjected to interference, which can place the patient at risk.
Arrhythmia or ventricular fibrillation can be induced during diathermic procedures such as electrocautery, HF ablation or HF surgery. For example, damaging heat can result during lithotripsy. Influences on the device are not always immediately clear. If risky procedures cannot be avoided, the following should be observed at all times:
Electrically insulate the patient.
Switch off the ICD's detection function. The pacemaker function can remain active. The device may need to be switched to asynchronous modes for this.
Do not introduce energy near the device system.
Additionally check the peripheral pulse of the patient.
Monitor the patient during and after every intervention.

External defibrillation

The device is protected against the energy that is normally induced by external defibril­lation. Nevertheless, any implanted device may be damaged by external defibrillation. Specifically, the current induced in the implanted leads may result in necrotic tissue formation close to the electrode/tissue interface. As a result, sensing properties and pacing thresholds may change.
Place adhesive electrodes anterior-posterior or perpendicular to the axis formed by the device to the heart at least 10 cm away from the device and from implanted leads.

Radiation therapy

The use of radiation therapy is contraindicated due to possible damage to the device and the resulting impaired functional safety. If this type of therapy is to be used anyway, prior risk/benefit analysis is absolutely necessary. The complexity of influencing factors such as different sources of radiation, a variety of devices and therapy condi­tions makes it impossible to issue directives that guarantee radiation therapy without an impact on the device. The EN 45502 standard pertaining to active implantable medical devices requires the following measures during the administration of thera­peutic ionizing radiation:
Adhere to instructions for risky therapy and diagnosis procedures.
Shield device against radiation.
After applying radiation, double-check the device system to make sure it is func­tioning properly.
Note:
Please contact BIOTRONIK with questions during the risk/benefit analysis.
en • English
7

Magnetic resonance imaging

Magnetic resonance imaging (MRI) is contraindicated due to the high frequency fields and the associated magnetic flux density: damage or destruction of the device system by strong magnetic interaction and damage to the patient by excessive warming of the body tissue in the area surrounding the device system.
Under certain conditions and when maintaining mandatory measures, magnetic reso­nance imaging can be performed to protect the patient and device system. BIOTRONIK devices with the "MR conditional" function bear the identification ProMRI®.
The ProMRI® manual – MR conditional device systems – contains detailed informa­tion on safely conducting an MRI. —
Download the digital manual from the web site: www.biotronik.com/manuals/manualselection
Order the printed manual from BIOTRONIK.
Does approval as "MR-Conditional" apply in your country or region? Request current information from BIOTRONIK.

Implantation

Implantation Procedure

Having parts ready

The following parts that correspond to the requirements of the EC Directive 90/385/EEC are required:
BIOTRONIK device with blind plug and screwdriver
BIOTRONIK leads and lead introducer set —
Single-chamber device: One bipolar ICD lead with 1 or 2 shock coils for the ventricle
Dual-chamber device: One bipolar lead for the atrium and one bipolar ICD lead for the ventricle with 1 or 2 shock coils
Triple-chamber device: an additional unipolar or bipolar LV lead
DF-1 and IS-1 connections are approved. For leads with a different connection or leads from other manufacturers, use adapters approved by BIOTRONIK only.
BIOTRONIK programmer (with integrated SafeSync RF telemetry or with separate SafeSync Module) and approved cable
External multi-channel ECG device
Keep spare parts for all sterile components.

Keeping an external defibrillator ready

In order to be able to respond to unforeseeable emergencies or possible technical failures of the device:
Keep an external defibrillator and paddles or patch electrodes ready.

Unpacking the device

WARNING
Inadequate therapy due to defective device
If an unpacked device is dropped on a hard surface during handling, electronic parts could be damaged.
Use a replacement device.
Return the damaged device to BIOTRONIK.
Peel the sealing paper off of the outer blister at the marked position in the direction indicated by the arrow. The inner blister may not come into contact with persons who have not sterilized their hands or gloves, nor with non-sterile instruments.
Take hold of the inner blister by the gripping tab and take it out of the outer blister.
Peel the sealing paper off of the sterile inner blister at the marked position in the direction indicated by the arrow.

Checking parts

Damage to any of the parts can result in complications or technical failures.
Check for damage before and after unpacking all parts.
Replace damaged parts.
The ICD is shipped with tachyarrhythmia therapy deactivated and is only to be connected and implanted in this state.
Leads may not be shortened.

Implantation site

Depending on lead configuration and the patient's anatomy, the ICD is generally implanted subpectorally on the left side.

Preventing leakage currents

Leakage currents between the tools and the device must be prevented during implanta­tion.
Electrically insulate the patient.
8

Preventing unintentional shock delivery

WARNING
Shock delivery with activated ICD
There is a risk of unintended shock delivery when handling an activated ICD.
Deactivate ICD therapy before touching the device during implantation, device replacement and explantation.

Avoiding damage to the header

There is a blind plug for DF-1 and IS-1 connections in the header. The provided set screws must be carefully loosened or tightened.
Loosen set screws with the supplied screwdriver. Use only BIOTRONIK screw­drivers with torque control!
Do not forcibly pull out the blind plug!
If lead repositioning is necessary, re-order sterile screwdrivers from BIOTRONIK.

Preventing short circuits in the header

WARNING
Short circuit due to open lead connector ports
Connector ports in the header which are open and thus not electrolyte-proof may cause undesired current flows to the body and penetration of body fluid into the device.
Either leave unused ports closed with the premounted blind plugs, or close them using the supplied blind plugs.

Connecting the lead connector to the device

1 Disconnect stylets and stylet guides. 2 DF-1/IS-1 connection:
Connect the DF-1 connector for the right-ventricular shock coil to RV.
Connect the DF-1 connector for the supraventricular shock coil to SVC. Or connect a subcutaneous array to SVC.
3 DF-1/IS-1 connection:
Connect the bipolar IS-1 lead connector for the atrium to RA.
Connect the IS-1 lead connector for the right ventricle to RV.
Connect the unipolar or the bipolar IS-1 lead connector for the left ventricle to LV.
4 Push the lead connector into the header without twisting or bending the con-
nector or conductor until the connector tip (on the DF-1 connector) (on the DF4 connector) becomes visible behind the set screw block.
5 If you cannot easily plug the lead connector into the connection:
Use only sterile water as lubricant.
6 If the lead connector cannot be inserted completely, the set screw may be pro-
truding into the drill hole of the set screw block.
Use the screwdriver to perpendicularly pierce through the slitted point in the center of the silicone plug until it reaches the set screw.
Carefully loosen the set screw without completely unscrewing it, so that it does not become tilted upon retightening.
7 Turn the set screw clockwise u ntil torque control starts (you will he ar a clicking
sound).
8 Carefully withdraw the screwdriver without retracting the set screw.
In case of IS-1 connections with two set screws, tighten both screws!
When the screwdriver is withdrawn, the silicone plug automatically safely seals the lead connector port.

Keeping distance between leads

WARNING
Inadequate therapy
When leads are not spaced sufficiently apart or are positioned inappropriately, this can lead to far-field sensing or insufficient defibrillation.
The distance between 2 shock coils must be greater than 6 cm.
Tip and ring electrodes must not have contact with each other.
en • English
9

Implanting

1Prepare the vein. 2 Implant the leads, perform the measurements, and fixate the leads. 3 Form the device pocket. 4 Connect the lead connector to the device. 5 Insert the device. 6 Guide the fixation suture through the opening in the header and fixate the
device in the prepared device pocket. 7 Close the device pocket. 8 Check the device with standard tests.

Applying the programming head

The programming head (PGH) features a diagram of the device. This is used to assist in positioning the head to ensure proper telemetry.
Make sure the PGH is positioned correctly.

Establishing telemetry contact

The programmer (or the SafeSync Module) can be no more than 3 m from the device; ideally there should be no hindrances between the patient and the programmer.
Switch on RF telemetry on the programmer.
Apply the programming head for about 2 s until successful initialization is displayed on the programmer:
The SafeSync symbol is displayed in the navigator and the signal strength is displayed in the status line.
Remove the programming head.

Activating ICD therapy

Load the device program that is suitable for the device type in the programmer.
Activate ICD therapy.
Shipment mode is permanently deactivated once the leads have been connected and initial measurement of the pacing impedance has been performed. The device data are saved.
Take precautionary measures while programming.
If the device induces tachycardia while programming ATPs or does not deliver ade­quate therapy in the DFT test: use emergency shock or an external defibrillator.

Precautionary Measures while Programming

Performing standard tests and monitoring the patient

Critical conditions can occur for the patient even during standard tests due to inade­quate parameter settings or interrupted telemetry.
Ensure sufficient patient care even during tests.
After the threshold test, check to determine whether the threshold is clinically and technically justifiable.
Continuously monitor the ECG and the patient's condition.
Cancel testing if necessary.

Cancelling telemetry

Programmer interference or interrupted telemetry during performance of temporary programs (follow-up tests) can result in inadequate pacing of the patient. This is the case if the programmer can no longer be operated due to a program error or a defective touch screen and therefore the temporary program cannot be terminated. Under these circumstances, it is helpful to cancel telemetry, in which case the device automatically switches to the permanent program.
In the case of telemetry with programming head: lift the PGH by at least 30 cm.
In the case of RF telemetry: switch off and reposition the programmer.
Turn off possible sources of interference.

Avoiding critical parameter settings

No modes and parameter combinations that pose a risk to the patient should be set.
Prior to setting rate adaptation, determine the patient's capacity for strain.
Check compatibility and effectiveness of parameter combinations after making settings.

Check for leads suitable for shock path

Three shock paths can be set, two of which form an electrical path to the device housing.
A second shock coil (dual shock coil) must be available for the shock path RV -> SVC.
10

Monitoring the patient when setting asynchronous modes

The asynchronous modes V00 and D00 can only be set if tachyarrhythmia sensing is deactivated. This would leave the patient without sensing and therefore without ICD therapy.
Continually monitor the patient.
Keep an external defibrillator ready.

Setting sensing

Manually set parameters can be unsafe. For example, unsuitable far-field protection may impede sensing of intrinsic pulses.
Note automatic sensitivity control.

Preventing device-induced complications

BIOTRONIK devices feature several functions to prevent device-induced complications to the greatest extent possible:
Measure the retrograde conduction time.
Set PMT protection.
Set the VA criterion.

Preventing conduction of atrial tachycardia

BIOTRONIK devices feature several functions to prevent conduction of atrial tachy­cardia to the ventricle(s):
Set mode switching for indicated patients.
Set the upper rate and the refractory periods to prevent abrupt ventricular rate switching.
Give preference to Wenckebach response and avoid 2:1 behavior.
Set all parameters so as to prevent constant changing between atrial and ventricular-controlled modes.

Observing the shock impedance limit

The implanted device could be damaged if the shock impedance is too low.
The shock impedance must be > 25 Ω.

Preventing recurrence after therapy shock

After a therapy shock, pacing can be performed with a post-shock program if there is no intrinsic rhythm.
Permanent program Post-shock program
DDD, DDI, AAI DDI VDD, VDI VDI VVI and OFF VVI
The following post-shock program parameters can be adjusted: post-shock duration, basic rate, rate hysteresis, ventricular pacing, LV-T-wave protection, triggering, AV delay (fixed, not dynamic).
The default settings for the post-shock program are as follows: A and RV: 7.5 V, 1.5 ms LV: settings from the permanent program

Phrenic nerve stimulation that cannot be terminated

In rare cases, chronic phrenic nerve stimulation cannot be terminated by reprogram­ming of the available left ventricular pacing configurations or by other measures.
As the case may be, set a right ventricular mode both in the permanent program as well as the ATP, in the post-shock program and for mode switching.

Avoiding risks in the case of exclusive LV pacing

Lead dislodgement in the case of exclusive left ventricular pacing could pose the following risks: loss of ventricular pacing and ATP therapy, induction of atrial arrhyth­mias.
Consider sensing and pacing parameters with reference to loss of therapy.
Exclusive LV pacing is not recommended for patients who depend on the device.
Take non-availability of automatic active capture control into consideration.
In the case of follow-ups and threshold tests, take loss of synchronized ventricular pacing into consideration.
Mode switching and post-shock do not allow for exclusive LV pacing. Also take the effects into account when setting the mode switching and post-shock parameters.

Recognizing lead failure

Automatic impedance measurement is always switched on.
Impedance values that indicate technical failure of a lead are documented in the event list.
en • English
11

Considering power consumption and service time

RF telemetry requires somewhat more power: Consumption during implantation corre­sponds to approximately 10 days of service time and consumption during a 20-minute follow-up corresponds to approximately 3 days.
Do not establish unnecessary RF telemetry.
After 5 minutes without input, SafeSync switches to the economy mode.
Check the battery capacity of the device at regular intervals.

Magnet Response

Application of the programming head when ICD therapy is set

If a connected programming head is applied and is communicating with the pro­grammer and ICD therapy is permanently set, detection and therapy remain intact except during the diagnostic tests. If ICD therapy is not set as permanent, no therapy is delivered when the programming head is applied.

Programming head application

When the programming head is applied, time remains for device interrogation and for manual activation or deactivation of the therapy before the device switches back to the previously set permanent therapy mode. The same applies to programming head appli­cation to establish RF telemetry contact.

Application of a permanent magnet

Applying a permanent magnet interrupts detection and therapy of tachycardia events. After 8 hours of this type of deactivation, the device automatically reactivates the therapy functions to prevent accidental permanent deactivation.
If detection interruptions of longer than 8 hours are required, the magnet has to be briefly removed from the device. The 8 hour countdown restarts when the magnet is applied again.
Use BIOTRONIK magnets: type M-50 permanent magnets.

Follow-up

Follow-up intervals

Follow-ups must be performed at regular, agreed intervals.
The first follow-up should be carried out by the physician using the programmer (in-office follow-up) approximately 3 months after implantation following the lead ingrowth phase.
The next in-office follow-up should be carried out once a year and no later than 12 months after the last in-office follow-up.

Follow-up with BIOTRONIK Home Monitoring®

Monitoring using the Home Monitoring function does not serve to replace regular in­office appointments with the physician required for other medical reasons. Follow-up supported by Home Monitoring can be used to functionally replace in-office follow-up under the following conditions:
The patient was informed that the physician must be contacted despite use of the Home Monitoring function if symptoms worsen or if new symptoms arise.
Device messages are transmitted regularly.
The physician decides whether the data transmitted via Home Monitoring with regard to the patient's clinical condition as well as the technical state of the device system are sufficient. If not, an in-office follow-up has to be carried out.
Possible early detection due to information gained via Home Monitoring may necessi­tate an additional in-office follow-up. For example, the data may indicate at an early stage lead problems or a foreseeable end of service time (ERI). Furthermore, the data could provide indications of previously unrecognized arrhythmias or modification of the therapy by reprogramming the device.

Follow-up with the programmer

Use the following procedure for in-office follow-up:
1 Record and evaluate the ECG. 2 Interrogate the device. 3 Evaluate the status and automatically measured follow-up data. 4 Check the sensing and pacing functions. 5 Possibly evaluate statistics and IEGM recording. 6 Manually perform standard tests if necessary. 7 Possibly customize program functions and parameters. 8 Transmit the program permanently to the device. 9 Print and document follow-up data (print report).
10 Finish the follow-up for this patient.

Patient Information

Patient ID card

A patient ID card is included in delivery.
Provide the patient with the patient ID.
Request that patients contact the physician in case of uncertainties.
12

Prohibitory signs

Places with prohibitory signs must be avoided.
Draw the patient's attention to prohibitory signs.

Possible sources of interference

Electromagnetic interference should be avoided in daily activities. Sources of inter­ference should not be brought into close proximity with the device.
Draw the patient's attention to special household appliances, security checkpoints, anti-theft alarm systems, strong electromagnetic fields, cell phones, and transmit­ters among other things.
Request patients to do the following: —
Use cell phones on the side of their body that is opposite of the device.
Keep the cell phone at least 15 cm away from the device both during use and when stowing.

Replacement Indications

Possible battery levels

BOS: Beginning of Service: > 70% charge
MOS 1: Middle of Service: 70% to 40% residual charge
MOS 2: Middle of Service: < 40% residual charge
ERI: Elective Replacement Indication, (i.e. RRT: Recommended Replacement Time)
EOS: End of Service

Elective Replacement Indication (ERI)

Elective Replacement Indication can be detected by Home Monitoring.
CAUTION
Temporally limited therapy
If ERI occurs shortly after follow-up and is only detected during the subsequent follow-up, then the remaining service time can be much less than 3 months.
Replace device soon.
The device can monitor the heart rhythm for at least 3 more months.
At least 6 maximum energy shocks can be delivered until EOS occurs.
The selected parameters in the device program do not change.
en • English

EOS replacement indication

End of Service can be detected by Home Monitoring.
WARNING
Patient at risk of death
If EOS replacement indication occurs before replacement of the device, then the patient is without therapy.
Replace device immediately.
Monitor patient constantly until immediate replacement of the device!
VT and VF detection and all therapies are deactivated!
The antibradycardia function remains active in the VVI mode: —
Ventricular pacing: RV; basic rate 50 bpm; without special pacemaker functions such as hysteresis, etc.
Pulse amplitude of 6 V; pulse width of 1.5 ms
Time of transmission for Home Monitoring: 90 days

Explantation and Device Replacement

Explantation

Interrogate the device status.
Deactivate VT and VF therapies prior to explantation.
Remove the leads from the header. Do not simply cut them loose.
Use state-of-the-art techniques to remove the device and, if necessary, the leads.
Note:
Normal oxidation processes may cause ICD housing discolorations. This is
neither a device defect nor does it influence device functionality.
Explants are biologically contaminated and must be disposed of safely due to risk of infection.

Device replacement

If, upon replacing the device, already implanted leads are no longer used but left in the patient, then an additional uncontrolled current path to the heart can result.
Deactivate VT and VF therapies prior to device replacement.
Insulate connections that are not used.
13
Loading...
+ 33 hidden pages