BIOTRONIK SE and KG TACH70 Users Manual

Inventra 7
Iperia 5/7
Itrevia 5/7
VR-T, VR-T DX
DR-T
HF-T, HF-T QP
ICD-Families Tachyarrhythmia Therapy Cardiac Resynchronization Therapy
Technical Manual Draft
Subject to completion and modification

1 Product Description

Intended Medical Use

Intended use

Inventra/Iperia/Itrevia are parts of a familiy of implantable cardioverter-defibrillators (ICDs). Primary objective of the therapy is to prevent sudden cardiac death. Further­more, the device is capable of treating bradycardia arrhythmias and cardiac resynchro­nization therapy with multisite ventricular pacing.
The implantation of an ICD is a symptomatic therapy with the following objectives:
Termination of spontaneous ventricular fibrillation (VF) through shock delivery
Termination of spontaneous ventricular tachycardia (VT) through antitachycardia pacing (ATP); in case of ineffective ATP or hemodynamically not tolerated VT, with shock delivery
Cardiac resynchronization through multisite ventricular pacing (triple-chamber devices)
Compensation of bradycardia through ventricular (single-chamber devices) or AV sequential pacing (DX, dual and triple-chamber devices)

Diagnosis and therapy forms

The device monitors the heart rhythm and automatically detects and terminates cardiac arrest resulting from ventricular tachyarrhythmia. All major therapeutic approaches from the field of cardiology and electrophysiology are included. BIOTRONIK Home Monitoring® enables physicians to perform therapy management at any time.

Required expertise

In addition to having basic medical knowledge, the user must be thoroughly familiar with the operation and the operation conditions of a device system.
Only qualified medical specialists having this special knowledge required are permitted to use implantable devices.
If users do not possess this knowledge, they must be trained accordingly.
1

Indications

DF-1
RV
DF-1
SVC
IS-1
RV
DF-1
RV
DF-1
SVC
IS-1
RA
IS-1
RV
DF-1
RV
DF-1
SVC
IS-1
RA
IS-1
RV
DF-1
RV
DF-1
SVC
IS-1
RA
IS-1
RV
IS-1
LV
Inventra/Iperia/Itrevia can treat life-threatening ventricular arrhythmias with anti­tachycardia pacing and defibrillation. Generally approved differential diagnostics methods, indications, and recommenda­tions for ICD therapy apply to BIOTRONIK devices. See the guidelines of cardiology associations for guidance. We recommend observing the indications published by the German Cardiac Society (Deutsche Gesellschaft für Kardiologie, Herz- und Kreislaufforschung) and the ESC (European Society of Cardiology). This also applies to the guidelines published by the Heart Rhythm Society (HRS), the American College of Cardiology (ACC), the American Heart Association (AHA), and other national cardiology associations.

Single-chamber and dual-chamber

Single-chamber and dual-chamber ICDs are indicated for patients with the following risk:
Sudden cardiac death caused by ventricular arrhythmias

Triple-chamber

Triple-chamber ICDs are indicated for patients with the following risks:
Sudden cardiac death caused by ventricular arrhythmias
Congestive heart failure with ventricular asynchrony

Contraindications

Known contraindications:
Tachyarrhythmia caused by temporary or reversible irritation, e.g. poisoning, elec­trolyte imbalance, hypoxia, sepsis or acute myocardial infarction
Such frequent VT or VF that the therapies would cause an unacceptably rapid depletion of the device batteries
VT with few or without clinically relevant symptoms
VT or VF treatable by surgery
Concomitant diseases that would substantially limit a positive prognosis
Accelerated idioventricular rhythm

System Overview

Device family

This device families consist of several device types with different lead connections: DF-1/IS-1, DF4/IS-1 or DF4/IS4/IS-1: single-chamber: VR-T and VR-T DX (only devices with a DF-1/IS-1 connection) dual-chamber: DR-T triple-chamber: HF-T and HF-T QP (only devices with a DF4 connection) All devices include BIOTRONIK Home Monitoring. Not all device types are available in every country.

Device

The device's housing is made of biocompatible titanium, welded from outside and thus hermetically sealed. The ellipsoid shape facilitates implantation in the pectoral muscle area. The connections for bipolar pacing and sensing (and unipolar connections for the triple-chamber device) as well as for shock delivery ar e found in the device header. The housing serves as a potential antipole during shock delivery or in the case of unipolar lead configuration.

DF-1/IS-1 or DF4/IS-1 or DF4/IS4

BIOTRONIK provides ICDs with headers for different standardized lead connections: DF-1/IS-1, DF4/IS-1 and DF4/IS4.
Note:
The device type DX can only be connected using a DF-1/IS-1 connector.
The device type HF QP can only be connected using a DF4/IS-1 or DF4/IS4 connector.

DF-1/IS-1 lead connection

The device labeling provides information pertaining to possible lead connections depending on the device type and pertaining to connection assignment:
VR DX DR HF
2
Connector
DF4-LLHH
RV
DF4-LLHH
RV
IS-1
RA
DF4-LLHH
RV
IS-1
RA
IS-1
LV
DF4-LLHH
RV
IS4-LLLL
RA
IS-1
LV
Lead
port
Configuration Implantation site Device type
connector
RV DF-1 Shock coil Right ventricle VR, DX, DR, HF SVC DF-1 Shock coil Superior vena
cava
VR, DX, DR, HF
RA IS-1 Bipolar Atrium DX, DR, HF (R)V IS-1 Bipolar (Right) ventricle VR, DX, DR, HF LV IS-1 Unipolar, bipolar Left ventricle HF

DF4/IS-1 lead connection

The device labeling provides information pertaining to possible lead connections depending on the device type and pertaining to connection assignment:
VR DR HF
Connector
Lead
port
Configuration Implantation site Device type
connector
RA IS-1 Bipolar Atrium DR, HF LV IS-1 Unipolar, bipolar Left ventricle HF RV, SVC DF4 Bipolar and shock Right ventricle VR, DR, HF
Note:
The device's DF4 connector port may only be used for connecting leads with a
DF4 connector that conform to ISO 27186.

DF4/IS4/IS-1 lead connection

The device labeling provides information pertaining to possible lead connections depending on the device type and pertaining to connection assignment:
HF QP
Connector
Lead
port
Configuration Implantation site Device type
connector
RA IS-1 Bipolar Atrium HF QP LV IS4 Unipolar, bipolar Left ventricle HF QP RV, SVC DF4 Bipolar and shock Right ventricle HF QP
Note:
The device's DF4/IS4 connector port may only be used for connecting leads w ith
a DF4/IS4 connector that conform to ISO 27186.

Leads

BIOTRONIK leads are sheathed with biocompatible silicone. They can be flexibly maneuvered, are stable long-term, and are equipped for active or passive fixation. They are implanted using a lead introducer set. Some leads are coated with polyurethane which is known to increase the sliding properties for the lead. Leads with steroids reduce inflammatory processes. The fractal design of the electrodes provides for low pacing thresholds. BIOTRONIK provides adapters to connect already implanted leads to new devices.

Telemetry

Telemetric communication between the device and the programmer can be carried out following initialization either by applying the programming head (PGH) to the device or by using radio frequency (RF) telemetry in the programmer. BIOTRONIK calls this func­tion SafeSync®.
3

Programmer

Implantation and follow-up are performed with BIOTRONIK's portable programmer: There is one with integrated RF telemetry and one with a separate SafeSync Module. The programmer is used during implantation to transfer the current device program to the device. The pacing thresholds can be determined and all tests can be performed during in-office follow-up. In addition to this, the programmer is used to set mode and parameter combinations, as well as for interrogation and saving of data from the device. Leadless ECG, IEGM, markers and functions are displayed simultaneously on the color display.

Modes

The mode setting depends on the individual diagnosis:
Device type Modes
VR VVI; VVIR; V00; OFF
DX VDD; VDDR; VDI; VDIR; VVI; VVIR; V00; OFF
DR, HF, QP DDD; DDDR; DDD-ADI; DDDR-ADIR; DDI; DDIR;

NBD and NBG codes

VVE is the NBD code for the antitachycardia mode of the single-chamber, dual­chamber, and triple-chamber devices:
V Shock in the ventricle V Antitachycardia pacing (ATP) in the ventricle E Detection via IEGM analysis
VDE is the NBD code for the antitachycardia mode of the dual-chamber, and triple­chamber devices:
V Shock in the ventricle D Antitachycardia pacing (ATP) in the atrium and ventricle E Detection via IEGM analysis
Series 7: VVI-CLS
Series 7: VVI-CLS
VDD; VDDR; VDI; VDIR VVI; VVIR; AAI; AAIR; V00; D00; OFF Series 7: VVI-CLS; DDD-CLS
DDDR is the NBG code for the antibradycardia mode of the dual-chamber device:
D Pacing in the atrium and ventricle D Sensing in the atrium and ventricle D Pulse inhibition and pulse triggering R Rate adaptation
DDDRV is the NBG code for the antibradycardia mode of the triple-chamber device:
D Pacing in the atrium and ventricle D Sensing in the atrium and ventricle D Pulse inhibition and pulse triggering R Rate adaptation V Multisite pacing in both ventricles
VDDR is the NBG code for the antibradycardia mode of the single-chamber DX device:
VVentricular pacing D Sensing in the atrium and ventricle D Pulse inhibition and pulse triggering R Rate adaptation
VVIR is the NBG code for the antibradycardia pacing modes of the single-chamber device:
VVentricular pacing V Sensing in the ventricle I Pulse inhibition in the ventricle R Rate adaptation
4
BIOTRONIK Home Monitoring
In addition to effective pacing therapy, BIOTRONIK provides a complete therapy management system:
With Home Monitoring, diagnostic and therapeutic information as well as technical data are automatically sent to a stationary or mobile transmitter via an antenna in the device header. The data are encrypted and sent from the transmitter to the BIOTRONIK Service Center via the cellular phone network.
The received data are deciphered and evaluated. Each physician can set the criteria for evaluation to be used for each patient and can configure the time of notification via E-mail, SMS or fax.
A clear overview of the results of this analysis is displayed for the attending physi­cians on the protected Internet platform Home Monitoring Service Center (HMSC).
Data transmission from the device is performed with a daily device message.
Device messages which indicate special events in the heart or in the device are forwarded immediately.
A test message can be initiated at any time using the programmer to immediately check the Home Monitoring function.
®
Order numbers for Iperia with DF-1/IS-1, DF4/IS-1 or DF4/IS4/IS-1 connection
Order numbers for Inventra with DF-1/IS-1, DF4/IS-1 or DF4/IS4/IS-1 connection
Not all device types are available in every country:
Iperia 5 Iperia 7
DF-1/IS-1 DF4/IS-1 DF4/IS4 DF-1/IS-1 DF4/IS-1 DF4/IS4
VR-T 393052 393053 393035 3393031
VR-T DX 393049 393033
DR-T 392415 392420 392410 392424
HF-T 393028 393026 393008 393010
HF-T QP — 402658 401658
Order numbers for Iperia ProMRI with DF-1/IS-1, DF4/IS-1 or DF4/IS4/IS-1 connection
Not all device types are available in every country:
Iperia 5 ProMRI Iperia 7 ProMRI
DF-1/IS-1 DF4/IS-1 DF4/IS4 DF-1/IS-1 DF4/IS-1 DF4/IS4
VR-T 393034 3393051 — 393050 3393030
VR-T DX 393048 393032
DR-T 392418 392419 392409 392423
HF-T 393027 393025 393007 393009
HF-T QP — 402656 401657
5
Order numbers for Itrevia with DF-1/IS-1, DF4/IS-1 or DF4/IS4/IS-1 connection Not all device types are available in every country:
Itrevia 5 Itrevia 7
DF-1/IS-1 DF4/IS-1 DF4/IS4 DF-1/IS-1 DF4/IS-1 DF4/IS4
VR-T 393058 393059 393040 393041
VR-T DX 393055 393037
DR-T 392417 392422 392412 392426
HF-T 393066 393064 393014 393016
HF-T QP — 402659 401662
Order numbers for Inventra with DF-1/IS-1, DF4/IS-1 or DF4/IS4/IS-1 connection Not all device types are available in every country:
Inventra 7
DF-1/IS-1 DF4/IS-1 DF4/IS4
VR-T 399443 399441
VR-T DX 399437
DR-T 399431 399429
HF-T 399423 399422
HF-T QP 393012
Order numbers for Itrevia ProMRI with DF-1/IS-1, DF4/IS-1 or DF4/IS4/IS-1 connection
Not all device types are available in every country:
Itrevia 5 ProMRI Itrevia 7 ProMRI
DF-1/IS-1 DF4/IS-1 DF4/IS4 DF-1/IS-1 DF4/IS-1 DF4/IS4
VR-T 393056 393057 393038 393039
VR-T DX 393054 393036
DR-T 392416 392421 392411 392425
HF-T 393065 393063 393013 393015
HF-T QP — 402657 401661
Order numbers for Inventra ProMRI with DF-1/IS-1, DF4/IS-1 or DF4/IS4/IS-1 connection
Not all device types are available in every country:
Inventra 7 ProMRI
DF-1/IS-1 DF4/IS-1 DF4/IS4
VR-T 399442 399440
VR-T DX 399436
DR-T 399430 399428
HF-T 399419 393020
HF-T QP 393011
6

Package contents

The storage package includes the following:
Sterile packaging with device
Serial number label
Patient ID card
Warranty booklet
Note:
Technical manuals are available either printed in the storage package or digitally
in the internet. The sterile container includes the following:
Device, blind plug DF-1 (if applicable) and blind plug IS-1 for device type HF
Screwdriver

Therapeutic and Diagnostic Functions

Diagnostic functions

Data from implantation and the most recent interrogations and follow-ups are recorded as well as arrhythmia episodes; they are stored together with other data to assess patients and the state of the device at any time.
To check the lead for proper functioning, an automatic impedance measurement using subthreshold pacing pulses is performed in the device.
Leadless ECG function: For all device types, far-field derivation can be measured without external leads between the right ventricular shock coil and housing, which, depending on the implantation site, corresponds to ECG derivation II or III (Einthoven).
Once a telemetry connection has been established during a test procedure in an in­office follow-up, the leadless ECG and the IEGM are displayed with markers.

Antitachycardia pacing

The ICD can treat ventricular tachycardia with antitachycardia pacing (ATP); ATP can also be delivered in the VF zone (ATP One Shot) when the stability criterion indicating that this will be effective before shock delivery (monomorphic rapid VTs) is met.
Arial tachycardia can be treated with antitachycardia pacing (atrial ATP) at stable heart rhythms and with high frequency bursts (HF burst) at instabil heart rhythms.
Depending on the device type, the device program contains not only the ICD func­tions but also all pacemaker functions for 1, 2, or 3 chambers. The heart rhythm is continuously monitored; each arrhythmia is classified according to the heart rate and the adjustable detection criteria. Depending on the preset values, antibrady­cardia as well as antitachycardia therapy is inhibited or delivered.

Cardioversion, defibrillation

The ICD can treat ventricular tachyarrhythmia with cardioversion and/or defibrilla­tion. Shock polarity and energy can be programmed individually. Shock energies between 2.0 and 45 J are possible. Before delivery of the shock, the ICD can be set to only deliver a shock when ongoing tachyarrhythmia is confirmed; during this time period the device can identify spontaneous conversion of the tachyarrhythmia and cancel the charging process if necessary.
The shock paths can be set between the different shock coils (SVC/RV) and/or the housing.
7

Antibradycardia pacing and CRT

Innovative rate hystereses, automatic sensor functions, and a night program promote the patient's intrinsic rhythm, avoid overdrive pacing, and facilitate adap­tation of the device to the individual needs of the patient.
Thresholds: atrial as well as ventricular pacing thresholds are automatically deter­mined in the device. Capture control is used to set the pulse amplitudes so that pacing is performed with the optimum atrial and ventricular amplitude for the patients with each change of the pacing threshold.
Setting an upper tracking rate for the atrium prevents unspecific atrial pacing, thus reducing the risk of pacemaker-mediated tachycardia.
Positive AV hysteresis functions support the intrinsic conduction and thus the natural contraction sequence. Negative AV hysteresis functions support the cardiac resynchronization therapy by maintaining pacing in stressful situations.
For resynchronization of the ventricles, triple-chamber devices have functions for multisite ventricular pacing with possible VV delays in either direction.
To ensure that no additional surgery is necessary in case of a left-sided increase of the pacing threshold or undesired phrenic nerve stimulation, different pacing polarities can be set for the left ventricular lead with a triple-chamber device. With the HF-T QP device up to 12 vectors are possible.
Automatic active capture control is available for the right and left ventricle with automated tracking of the pacing threshold or automatic threshold monitoring (ATM) for trend analysis.
Series 7: additional, special form of rate adaptation: an increased cardiac output requirement is detected using physiological impedance measurement. The measuring principle is based on contractile changes (inotropy) of the myocardium (CLS function: Closed Loop Stimulation). The suitable rate adaptation is automati­cally initialized and optimized in CLS mode.
Ventricular pacing suppression: unnecessary ventricular pacing is avoided by promoting intrinsic conduction (Vp suppression function). The device can adapt itself to conduction changes. In the case of intrinsic conduction, the device switches to a DDD(R)-ADI(R) mode.

Storing programs

There are two types of therapy programs: — Default parameters are offered for the most common indications (BradyProgram­Consult function). — Individual settings can be saved in 3 individual therapy programs

Home Monitoring functions

The device automatically sends information to the transmitter once a day. It also sends messages related to events, which are immediately forwarded to the Service Center. In addition to this, test messages can be initiated using the programmer.
Appointments for Home Monitoring-supported follow-ups can be scheduled via the HMSC.
Important medical information in the device messages include the following: —
Atrial and ventricular arrhythmias
Parameters relevant to leads in the atrium and ventricle: pacing thresholds, sensing amplitudes, impedances
Current statistics
IEGM online HD® with up to 3 high definition channels
8

2 General Safety Instructions

Operating Conditions

Technical manuals

Folgende Gebrauchsanweisungen informieren über die Anwendung von Implantatsys­temen: — Technical manual for the device — Technical manual for the HMSC — Technical manuals for the programmer and the SafeSync Module — Technical manuals for the user interface — Technical manuals for cables, adapters and accessories
Technical manual s are available either prin ted in the storage package or dig itally in the internet: https://manuals.biotronik.com/manuals/home
Consider all relevant technical manuals.
Keep technical manuals for further use.

Care during shipping and storage

Devices must not be stored or transported close to magnets or sources of electro­magnetic interference.
Note the effects of the storage duration; see Battery Data.

Delivery in shipment mode

The device is delivered in shipment mode to protect the battery; capacitor reforming required during storage could result in controlled extended charge times of the shock capacitors.
The shipment mode is displayed on the programmer after loading the device program (it is deactivated during implantation on initial measurement of the pacing impedance).

Temperature

Extremely low and high temperatures affect the service time of the battery in the device.
Temperatures of 5°C to 45°C are permitted for transport, storage, and use.

Sterile delivery

The device and the screwdriver have been gas-sterilized. Sterility is guaranteed only if the blister and quality control seal have not been damaged.

Sterile packaging

The device and screwdriver are packaged in two separately sealed blisters. The inner blister is also sterile on the outside so that it can be transferred in a sterile state during implantation.

Single use only

The device and screwdriver are intended for single use only.
Do not use the device if the package is damaged.
The device must not be resterilized and reused.

Possible Complications

General information on medical complications

Complications for patients and device systems generally recognized among practitio­ners also apply to BIOTRONIK devices.
Normal complications may include fluid accumulation within the device pocket, infections, or tissue reactions. Primary sources of complication information include current scientific and technological knowledge.
It is impossible to guarantee the efficacy of antitachycardia therapy, even if the programs have proven successful during tests or subsequent electrophysiological examinations. In rare cases the set parameters may become ineffective. It is possible for therapies to induce or accelerate tachycardia and cause sustained ventricular flutter or fibrillation.

Skeletal myopotentials

Bipolar sensing and control of sensitivity are adapted by the device to the rate spectrum of intrinsic events so that skeletal myopotentials are usually not recorded. Skeletal myopotentials can nonetheless be classified as intrinsic events especially at very high sensing sensitivity and, depending on the interference, may cause inhibition or antiar­rhythmia therapy. In the case of undesired myopotentials, the device switches to asynchronous pacing if the interference rate is exceeded.
9
Loading...
+ 19 hidden pages