BIOTRONIK Philos DR-T Technical Manual

Page 1
Philos DR-T
DDDR Dual Chamber Pulse Generator
with Home Monitoring
Technical Manual
Page 2
Philos DR-T
Implantable Pulse Generator
Radiopaque Identification
A radiopaque identification code is visible on standard x-ray, and identifies the pulse generator:
Philos DR-T
VV
CAUTION
Because of the numerous available 3.2-mm configurations (e.g., the IS-1 and VS-1 standards), lead/pulse generator compatibility should be confirmed with the pulse generator and/or lead manufacturer prior to the implantation of a pacing system.
IS-1, wherever stated in this manual, refers to the international standard, whereby leads and generators from different manufacturers are assured a basic fit. [Reference
ISO 5841-3:1992(E)].
CAUTION
Federal (U.S.A.) law restricts this device to sale by or on the order of, a physician (or properly licensed practitioner).
©2004 BIOTRONIK, Inc., all rights reserved.
Page 3
Philos DR-T Technical Manual i
Contents
1. Home Monitoring-Overview ................................................1
1.1 Home Monitoring ..............................................................1
1.2 Transmission of Information .............................................1
1.3 Patient Device with Components......................................2
1.4 Receiving Patient Data .....................................................3
2. Indications and Contraindications.....................................5
3. Warnings and Precautions..................................................7
3.1 Home Monitoring ..............................................................7
4. Types of Messages ............................................................11
4.1 Event Message ............................................................... 11
4.2 Trend Message...............................................................13
4.3 Patient Message .............................................................13
5. Description of Transmitted Data.......................................15
5.1 The Monitoring Interval ...................................................15
5.2 Heart Rate.......................................................................15
5.3 Atrial Rhythm ..................................................................15
5.4 Ventricular Rhythm .........................................................15
5.5 AV Conduction ................................................................16
5.6 System Status.................................................................16
6. Technical Data ....................................................................17
6.1 Modes .............................................................................17
6.2 Home Monitoring Parameters.........................................17
6.3 Pulse and Control Parameters .......................................18
6.3.1 Rate Adaptation ......................................................21
6.3.2 Parameters at Replacement Indication ..................21
6.3.3 Additional Functions................................................23
6.4 Programmers ..................................................................24
6.5 Materials in Contact with Human Tissue ........................24
6.6 Electrical Data/Battery ....................................................24
6.7 Mechanical Data .............................................................25
7. Order Information ..............................................................27
Page 4
ii Philos DR-T Technical Manual
CAUTION
Federal (U.S.A.) law restricts this device to sale by, or on the
order of, a physician (or properly licensed practitioner).
Page 5
Philos DR-T Technical Manual 1
1. Home Monitoring-Overview
Philos DR-T offers the complete functionality of a DDDR pacemaker while being equipped with the additional features associated with Home Monitoring. Consult the Philos technical manual for a description and overview of the standard pacemaker functionality of the Philos DR-T.
1.1 Home Monitoring
Home Monitoring is a novel system, which enables the exchange of information about a patient’s cardiac status between implant, patient, and physician. Home Monitoring can be used to provide the physician with advance reports from the implant and process them into graphical and tabular formats. This information helps the physician optimize the therapy process, as it may result in the patient being scheduled for additional clinical appointments between regular follow-up visits if necessary.
The implant’s Home Monitoring function can be used for the entire operational life of the implant (prior to ERI) or for shorter periods, such as several weeks or months.
1.2 Transmission of Information
The implant transmits information with a small transmitter, which has a range of about 2 meters. The patient’s implant data are sent daily to the corresponding patient device (i.e., CardioMessenger) at a configurable time. The transmissions may also be activated by the patient with the application of a magnet over the implant and by certain cardiac events, as programmed. The types of transmissions are discussed in
Section 4
The minimal distance between the implant and the patient device must be 15 cm.
.
Page 6
2 Philos DR-T Technical Manual
1.3 Patient Device with Components
The patient device (Figure 1) is designed for use in the home and is comprised of the mobile device and the associated charging station. The patient can carry the mobile device with them during his or her occupational and leisure activities. The patient device comes with a rechargeable battery that has an approximate operational time of 24 hours after a charge time of 5 hours. It receives information from the implant and forwards it via the mobile network to a BIOTRONIK Service Center.
For additional information about the patient device, please refer to its manual.
Figure 1: Patient Device with Charging Stand (CardioMessenger)
Page 7
Philos DR-T Technical Manual 3
1.4 Receiving Patient Data
The implant’s information is digitally formatted by the BIOTRONIK Service Center and processed into a concise report called a Cardio Report. The Cardio Report, which is adjusted to the individual needs of the patient, contains current and previous implant data. The Cardio Report is sent to the attending physician via fax or is available on the Internet, which is selected during registration of the patient. For more information on registering for Home Monitoring, contact your BIOTRONIK sales representative.
The password protected BIOTRONIK Home Monitoring website can be accessed by registered users at the following URL:
www.biotronik-homemonitoring.com
An online help menu is available in order to assist with the use of the Home Monitoring website.
Use of the Internet for reviewing Home Monitoring data must be in conjunction with the system requirements listed in Table 1 Additionally, Table 1 recommended for optimizing usage of the Internet.
provides system specifications that are
.
Page 8
4 Philos DR-T Technical Manual
Table 1: System Requirements / Recommendations
System
Requirements
System
Recommendations
(for Optimal Usage)
Screen
800 x 600 1024 x 768
Resolution Internet
Bandwidth PC 600 MHz, 128
56 kB/sec 128 kB/sec
(DSL, cable modem) N/A
MB RAM
Internet Browser MS Internet
Explorer 5.0
- or ­Netscape
MS Internet Explorer
5.5
- or ­ Netscape 7/Mozilla
Navigator 4.72 Acrobat Reader Version 4 Version 5 or higher Communication
Channel
Fax (G3) or
e-mail
Fax (G3), e-mail or mobile phone
Additionally, the attending physician may register to be informed of the occurrence of an Event Triggered Message through email or SMS (i.e., mobile phone) with a brief text message. If registered for Internet availability, the patient’s detailed implant data can then be viewed by logging onto the Home Monitoring website.
Page 9
Philos DR-T Technical Manual 5
2. Indications and Contraindications
For the general indications and contraindications, please refer to the Philos Family technical manual. The indications and contraindications of the Philos DR-T are identical to those of the rate adaptive dual chamber Philos DR pulse generator.
Page 10
6 Philos DR-T Technical Manual
Page 11
Philos DR-T Technical Manual 7
3. Warnings and Precautions
Certain therapeutic and diagnostic procedures may cause undetected damage to a pulse generator, resulting in malfunction or failure at a later time. Please note the following warnings and precautions:
Magnetic Resonance Imaging (MRI) – Avoid use of magnetic resonance imaging as it has been shown to cause movement of the pulse generator within the subcutaneous pocket and may cause pain and injury to the patient and damage to the pulse generator. If the procedure must be used, constant monitoring is recommended, including monitoring the peripheral pulse.
Rate-Adaptive Pacing – Use rate-adaptive pacing with care in patients unable to tolerate increased pacing rates.
High Output Settings – High output settings combined with extremely low lead impedance may reduce the life expectancy of the pulse generator to less than 1 year. Programming of pulse amplitudes, higher than 4.8 V, in combination with long pulse widths and/or high pacing rates may lead to premature activation of the replacement indicator.
3.1 Home Monitoring
Patient’s Ability – Use of Home Monitoring requires the patient
and/or caregiver to follow the system instructions and cooperate fully when transmitting data.
If the patient cannot understand or follow the instructions because of physical or mental challenges, another adult who can follow the instructions will be necessary for proper transmission.
Cellular Phone Availability – Home Monitoring is not practical for patients who live in areas where cellular telephone networks, utilizing the GSM standard, are not available or are not likely to become available in the near future.
Page 12
8 Philos DR-T Technical Manual
Electromagnetic Interference (EMI) – Precautions for EMI
interference with the Philos DR-T pulse generator are provided
in the Philos technical manual in section 4.5. Sources of EMI
including cellular telephones, electronic article surveillance systems, and others are discussed therein.
Use in Cellular Phone Restricted Areas – The mobile patient device (transmitter/receiver) should not be utilized in areas where cellular phones are restricted or prohibited (i.e., commercial aircraft).
Event Triggered Message – A timely receipt of the corresponding event report cannot be guaranteed. The receipt is also dependent on whether the patient was physically situated in the required coverage range of the patient device at the time the event information was sent.
Patient-Activated Message – The magnet effect must be programmed “synchronous” if the [Patient Message] function is activated. Otherwise, this function will not be available.
Not for Diagnosis – The data transmitted by Home Monitoring are not suitable for diagnosis, because not all information available in the implant is being transmitted.
Follow-Ups – When using Home Monitoring, the time period between follow-up visits may not be extended.
The use of Home Monitoring does not replace regular follow-up examinations. The data transmitted via Home Monitoring are not suitable for a conclusive diagnosis.
Magnet Effect – The magnet effect must be programmed “synchronous” if the attending physician enables the patient to transmit messages.
Lead Connection – Because of the numerous available 3.2-mm configurations (e.g., the IS-1 and VS-1 standards), lead/pulse generator compatibility should be confirmed with the pulse generator and/or lead manufacturer prior to the implantation of a pacing system.
IS-1, wherever stated in this manual, refers to the international standard, whereby leads and generators from different manufacturers are assured a basic fit. [Reference ISO 5841-3:1992(E)].
Page 13
Philos DR-T Technical Manual 9
Consult the Philos manual for additional warnings and precautions associated with this device.
Page 14
10 Philos DR-T Technical Manual Philos DR-T Technical Manual 11
Page 15
4. Types of Messages
When the Home Monitoring function is activated, the transmission of a message from the implant can be triggered as follows:
Trend Message – the time period (daily) initiates the report
Event Message – the pacemaker detects certain events, which initiate a message immediately
Patient Message – the patient initiates transmission of a message
The attending physician decides whether or not the patient will be able to initiate the transmission of a message on their own through application of a magnet. Transmissions initiated through magnet application do not affect event or trend message transmissions.
4.1 Event Message
When certain cardiac and technical events are detected by the implant, a message transmission is automatically triggered. This is described as an “event message”.
Page 16
12 Philos DR-T Technical Manual
The following cardiac and technical events initiate a report:
Atrial Lead Check < 300 and > 3000 Ohm
Ventricular Lead Check < 300 and > 3000 Ohm
Ventricular Bursts (Runs)
Ventricular Events (Episodes)
Low P-Wave Amplitude
Low R-Wave Amplitude
1
(< 50% safety margin)
1
(< 50% safety margin)
ERI Activation
st
1
Mode Switch/24 hours
WARNING
A timely receipt of the corresponding event report cannot be guaranteed. The receipt is also dependent on whether the patient was physically situated in the required coverage range of the patient device at the time the event information was sent.
NOTE:
When ERI mode is reached, this status is transmitted. Further measurements and transmissions of Home Monitoring data are no longer possible.
NOTE:
The attending physician must notify the BIOTRONIK Service Center about which of these events he/she wishes to be informed.
1
Examples: The programmed sensitivity is 1.0 mV. A) Average of the measured P/R-Wave amplitudes is 2.6 mV. Therefore, measured value is greater than 100% of the safety margin. Event message is not triggered. B) Average of the measured P/R-Wave amplitudes is 1.9 mV. Therefore, measured value is less than 100%, but greater than 50% of the safety margin. Event message is not triggered. C) Average of the measured P/R-Wave amplitudes is 1.4 mV. Therefore, measured value is smaller than 50% of the safety margin. As a result, an event message is triggered.
Page 17
Philos DR-T Technical Manual 13
4.2 Trend Message
An additional type of message is the programmable “Trend Message”. Trend messages occur at a programmable time of transmission (i.e., at the end of the Monitoring Interval). The time can be set anywhere between 0:00 and 23:50 hours. It is recommended that you select a time during the late night or early morning hours (between 0:00 and 4:00), or other time when the patient is usually in his or her home.
The length of the time interval (monitoring interval) is not programmable – it is preset to “daily”.
4.3 Patient Message
It is possible to trigger a transmission through magnet application over the pacemaker. The attending physician must inform the patient in detail about operating the device and about the physical symptoms which would warrant a magnet application by the patient.
WARNING
The magnet effect must be programmed “synchronous” if the [Patient Message] function is activated. Otherwise, this function will not be available.
Page 18
14 Philos DR-T Technical Manual Philos DR-T Technical Manual 15
Page 19
5. Description of Transmitted
Data
The following data are transmitted by Home Monitoring, when activated. In addition to the medical data, the serial number of the implant is also transmitted.
5.1 The Monitoring Interval
The monitoring interval is considered the time period since the last trend message was transmitted. For a trend message, the monitoring interval since the previous trend message is set at 24 hours. For an event or patient message, the monitoring interval would typically be less than 24 hours. This occurs when these messages are sent after the programmed transmission time of the trend message.
5.2 Heart Rate
Average (mean) ventricular heart rate (bpm)
5.3 Atrial Rhythm
Intrinsic rhythm (As / Ax) (%)
Number of Mode Switching
Duration of Mode Switching (%)
5.4 Ventricular Rhythm
Intrinsic rhythm (Vs/ Vx) (%)
Ventricular rate at Mode Switching (bpm)
Number of ventricular runs (4...8 sequential VES)
Number of ventricular events (more than 8 sequential
VES), defined as a ventricular episode
Duration of the longest ventricular event (sec)
Page 20
16 Philos DR-T Technical Manual
5.5 AV Conduction
AV Synchronicity (Ax Vx/Vx) (%)
with intrinsic rhythm (AsVs) (%)
with atrial stimulation (ApVs) (%)
with ventricular stimulation (AsVp) (%)
with dual-chamber stimulation (ApVp) (%)
5.6 System Status
Atrial lead check
Ventricular lead check
Mean P-Wave amplitude / programmed sensitivity (%)
Mean R-Wave amplitude / programmed sensitivity (%)
Battery status
NOTE:
Atrial and Ventricular Lead Check must be programmed ON to enable Home Monitoring functionality of this data.
Page 21
Philos DR-T Technical Manual 17
6. Technical Data
6.1 Modes
The following modes are available in the Philos DR-T when Home Monitoring is deactivated:
DDDR, DDTR/A, DDTR/V, DDTR, DDIR, DDIR/T, DVIR, DVTR, DOOR, VDDR, VDTR, VDIR, VVIR, VVTR, VOOR, AAIR, AATR, AOOR, DDD, DDT/A, DDT/V, DDT, DDI, DDI/T, DVI, DVT, DOO, VDD, VDT, VDI, VVI, VVT, VOO, AAI, AAT, AOO, OFF
The Home Monitoring function is available for the following pacing modes:
DDDR, DDTR/A, DDTR/V, DDTR, DDIR, DDIR/T, VDDR, VDTR, VDIR, DDD, DDT/A, DDT/V, DDT, DDI, DDI/T, VDD, VDT, VDI
NOTE:
Bold parameters indicate factory settings.
6.2 Home Monitoring Parameters
Home Monitoring
Off, On
Monitoring Interval
1 day
Time of Transmission
0:00...(10)...23:50 hours
Patient Message
Off, On
Event Message
Off, On
Page 22
18 Philos DR-T Technical Manual
6.3 Pulse and Control Parameters
Basic rate
30...(1)...60...(1)...88...(2)...122...(3)...140...(5)...180 ppm
Night rate
Off, 30...(1)...60...(1)...88...(2)...122...(3)...140...(5)...180 ppm
Rate Hysteresis
Off; -5...(5)...-50 bpm
Repetitive Hysteresis
Off; 1…(1)…10
Scan Hysteresis
Off: 1…(1)…10
Upper Tracking Rate (UTR)
100; 110; 120; 130; 140; 160; 185 ppm
UTR Response
2:1; WRL (automatic selection)
2,3,4
Rate Limitation
190…220 ppm
Dynamic AV Delay (Dual chamber only)
low; medium; high; individual; fixed
AV Delay Values (Dual chamber only)
15; 50; 75; 100; 120...(10)...200; 225; 250; 300 ms
(Programmable in 5 ranges)
2
The corresponding intervals t correlate with the rates f by the formula t = 60.000 / f (t in ms, f in ppm).
3
In the event of electronic defect.
4
Rate Limitation changes as the Pacemaker approaches End of Service. The Rate Limitation is nominally 190 ppm at Beginning of Service (BOS) and can reach 220 ppm at End of Service (EOS) due to battery depletion.
Page 23
Philos DR-T Technical Manual 19
AV Hysteresis
Off; low; medium; high
AV Repetitive Hysteresis
Off; 1...(1)...6
AV Scan Hysteresis
Off; 1...(1)...6
AV safety delay (Dual chamber only)
100 ms
Sense Compensation
Off; -15...(15)...-120 ms
Ventricular Blanking Time
16; 24; 32; 40; 48; 56; 72 ms
Magnet effect
Automatic; asynchronous; synchronous
Asynchronous Magnet Effect: paces at 90 ppm.
Automatic Magnet Effect; 10 cycles at 90 ppm asynchronous; thereafter synchronous with the programmed basic rate
Synchronous Magnet Effect; synchronous with programmed basic rate
Pulse amplitude
A 0.1...(0.1)...3.6...(0.1)...4.8...(1.2)...8.4 V
V 0.1...(0.1)...3.6...(0.1)...4.8...(1.2)...8.4 V
Ventricular Pulse Amplitude for Safe Program
4.8V
Pulse width
A 0.1; 0.2; 0.3; 0.4; 0.5; 0.75; 1.0; 1.5 ms
V 0.1; 0.2; 0.3; 0.4; 0.5; 0.75; 1.0; 1.5 ms
Page 24
20 Philos DR-T Technical Manual
Sensitivity
A 0.1...(0.1)...1.0...(0.1)…1.5…(0.5)...7.5 mV
V 0.5...(0.5)...2.5...(0.5)...7.5 mV
5
Refractory period
A 200...(25)…425...(25)...775 ms
V 170, 195, 220, 250; 300; 350; 400 ms
PVARP (minimum)
Off, dependent on TARP and AV Delay settings
ARP Extension
0...(50)...350 ms
Lead Check
Off; On
Automatic Mode Conversion
Off; On (in modes DDD(R), DDT(R)/A, DDT(R)/V, and VDD(R))
Mode Switch (X out of Y)
Off; On (in modes DDD(R), DDT(R)/A, DDT(R)/V and VDD(R))
X = 3...(1)...8
Z = 3...(1)...8
Lead Polarity
Pace: A unipolar; bipolar
V unipolar; bipolar
Sense: A unipolar; bipolar
V unipolar; bipolar
Far-Field Blanking
50...(25)...200 ms
5
In the DDIR, VVIR and VOOR modes, lower maximum sensor rates result than indicated
here (partly depending on the selected AV interval). The correct values are indicated by the programmer.
Page 25
Philos DR-T Technical Manual 21
PMT Management
Off; On
6.3.1 Rate Adaptation
Sensor gain
1.0, 1.1, 1.3, 1.4, 1.6, 1.8, 2.0, 2.2, 2.6, 3.0, 3.3, 3.7, 4.0, 4.5,
5.0, 6.0, 7.0, 8.0, 8.5, 10, 11, 12, 14, 16, 18, 20, 22, 24, 28, 32, 35, 40
Sensor threshold
very low; low; mean; high; very high
Rate increase
1, 2, 4, 8 ppm/s
Maximum sensor rate
80...(5)...180 ppm
Rate decrease
0.1, 0.2, 0.4, 0.8 ppm/s
Automatic Sensor Gain
Off; On
6.3.2 Parameters at Replacement Indication
Basic Rate
Programmed value minus 11%
(in modes DVI(R), DDI(R), DVT(R), DDI/T(R) minus 4.5–11%, depending on programmed AV Delay)
Magnet Rate
The magnet rate in all modes decreases as shown in the following table.
Page 26
22 Philos DR-T Technical Manual
Magnet Mode Cycles 1-10 after
magnet
application
Automatic Asynchronous,
basic rate at 80 ppm
Asynchronous Asynchronous,
basic rate at 80 ppm
Synchronous Synchronized with
basic rate reduced
by 4.5 - 11%
Pulse Widths
Programmed values
Sensitivities
Programmed values
After Cycle 10
Synchronized with basic rate reduced
by 4.5 - 11%
Asynchronous with
basic rate at 80
Synchronized with basic rate reduced
by 4.5 - 11%
Page 27
Philos DR-T Technical Manual 23
6.3.3 Additional Functions
Home Monitoring
Additional functions conform with Philos DR:
Temporary Program Activation
High Precision Threshold test in the range of 0.1 up to
4.8 V with 0.1 V resolution
PAC (pulse amplitude control) system produces consistent pulses
Analog Telemetry with measuring of battery, pulse and lead data
Two channel Real Time IEGM Transmission with markers
Patient Data Memory
Sensor Simulation
Position Indicator for the programmer head
24 hour Trend
Heart Rate Histogram
Sensor Rate Histogram
Sensor Test Trend with complete Rate Forecast
Automatic Sensor Gain with Trend Monitor
VES Analysis
AES Analysis
Retrograde Conduction Test
Automatic Mode Conversion
Mode Switching
PMT Management
Activity Report
Event counter
P-/R-wave Tests with Trend Data
External Pulse Control up to 800 ppm
Night Program
Page 28
24 Philos DR-T Technical Manual
Arrhythmia Detection Recordings (note that only 3 ADRs can be stored by the Philos DR-T; however, 10 can be stored by the Philos DR)
Lead Impedance Trends
Lead Check
6.4 Programmers
Home Monitoring functions and their parameters can be configured with the following programming devices:
EPR 1000
PLUS
, TMS 1000
PLUS
6.5 Materials in Contact with Human Tissue
Housing: Titanium
Connector receptacle: Epoxy resin
Sealing Plugs: Silicone Rubber
Coating for unipolar devices: Silicone Rubber
6.6 Electrical Data/Battery
N
OTE:
At 37ºC, with pacing impedance of 500 Ohms.
LECTRICAL DATA
E
Circuit
Hybrid electronics with VLSI CMOS Chip
Input Impedance
A 240 kOhm
V 240 kOhm
Pulse Form
biphasic, asymmetric
Page 29
Philos DR-T Technical Manual 25
Polarity
cathodic
Power Consumption
BOS, inhibited: 14 µA
BOS, 100 % pacing: 22 µA
Conducting Surface
2
uncoated: 32.8 cm
coated: 7.23 cm
2
Conducting Shape
uncoated: flattened ellipsoidal
coated: ellipsoidal
BATTERY
Power Source
Li/I
Open-Circuit Voltage
2.8 V
Voltage at ERI
2.5 V
6
Nominal Capacity
1.3 Ah
6.7 Mechanical Data
Lead Connector
IS-1 (accepts unipolar and bipolar)
Weight
27 g
6
Battery manufacturer’s specification.
Page 30
26 Philos DR-T Technical Manual
Volume
3
12 cm
Dimensions
6 x 44 x 51 mm
X-Ray Identification
VV
Page 31
Philos DR-T Technical Manual 27
7. Order Information
Product Name Order Number
Philos DR-T 331 440
Page 32
Distributed by:
BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035-5369 (800) 547-0394 (24-hour) (503) 635-9936 (FAX)
M4064-C 7/04
Manufactured by:
BIOTRONIK GmbH & Co. KG
Woermannkehre 1
D-12359 Berlin Germany
Loading...