BIOTRONIK Lexos DR-T, Lexos DR, Lexos VR-T, Lexos VR Technical Manual

Lexos
Family of Implantable Cardioverter
Defibrillators
Technical Manual
X-ray Identification
Lexos DR, DR-T, VR and VR-T
Implantable Cardioverter Defibrillator
Inside the housing, right-hand side:
X-Ray identification
Year of manufacture
C
AUTION
Federal (U.S.A.) law restricts this device to sale by, or on the order of, a physician.
2003 BIOTRONIK, Inc., all rights reserved.
KV
nn
Contents
Lexos Technical Manual i
1. General...........................................................................1
1.1 System Description....................................................1
1.2 Indications and Usage ...............................................2
1.3 Contraindications .......................................................3
1.4 Warnings and Precautions.........................................3
1.4.1 Sterilization, Storage, and Handling ..................4
1.4.2 Device Implantation and Programming..............4
1.4.3 Lead Evaluation and Connection.......................6
1.4.4 Follow-up Testing...............................................8
1.4.5 Pulse Generator Explant and Disposal..............8
1.4.6 Hospital and Medical Hazards ...........................9
1.4.7 Home and Occupational Hazards......................10
1.4.8 Cellular Phones..................................................11
1.4.9 Electronic Article Surveillance (EAS).................12
1.4.10 Home Appliances ...............................................12
1.5 Adverse Events..........................................................13
1.5.1 Potential Adverse Events...................................13
1.5.2 Observed Adverse Events .................................14
1.6 Clinical Studies ..........................................................17
1.6.1 Tachos DR .........................................................17
1.6.2 Phylax AV...........................................................19
1.7 Patient Selection and Treatment ...............................21
1.7.1 Individualization of Treatment............................21
1.7.2 Specific Patient Populations ..............................22
1.8 Patient Counseling Information .................................23
1.9 Evaluating Prospective ICD Patients.........................23
2. Device Features ............................................................25
2.1 Sensing (Automatic Sensitivity Control) ....................25
2.1.1 Ventricular Sensitivity Settings ..........................25
2.1.2 Minimum Ventricular Threshold .........................28
2.1.3 Atrial Sensitivity Settings....................................28
2.1.4 Minimum Atrial Threshold ..................................29
2.1.5 Far Field Blanking ..............................................29
ii Lexos Technical Manual
2.2 Ventricular Tachyarrhythmia Detection ......................30
2.2.1 VF Classifications ..............................................31
2.2.2 VT Interval Counters ..........................................31
2.2.3 VT Classification ................................................31
2.2.4 SMART Detection™...........................................32
2.2.5 Onset..................................................................33
2.2.6 Stability...............................................................33
2.2.7 Sustained VT Timer ...........................................34
2.3 Tachyarrhythmia Redetection ....................................34
2.3.1 VT Redetection ..................................................35
2.3.2 SMART Redetection ..........................................35
2.3.3 VF Redetection ..................................................35
2.4 Tachyarrhythmia Termination.....................................35
2.5 Tachyarrhythmia Therapy ..........................................36
2.5.1 Therapy Options ................................................36
2.5.2 Anti-Tachycardia Pacing (ATP) .........................36
2.5.3 Shock Therapy...................................................39
2.5.4 Progressive Course of Therapy .........................42
2.6 Bradycardia Therapy .................................................43
2.6.1 Bradycardia Pacing Modes................................43
2.6.2 Basic Rate..........................................................44
2.6.3 Night Rate ..........................................................44
2.6.4 Rate Hysteresis..................................................44
2.6.5 Dynamic AV Delay .............................................48
2.6.6 Upper Tracking Rate..........................................49
2.6.7 Mode Switching..................................................50
2.6.8 PMT Management .............................................51
2.6.9 Rate Adaptive Pacing ........................................51
2.6.10 Pulse Amplitude .................................................53
2.6.11 Pulse Width ........................................................53
2.6.12 Post Ventricular Atrial Refractory Period ...........53
2.6.13 PVARP Extension ..............................................53
2.6.14 Noise Response.................................................54
2.6.15 Post Shock Pacing.............................................54
2.7 EP Test Functions ......................................................54
Lexos Technical Manual iii
2.7.1 P and R-wave Amplitude Measurments ............54
2.7.2 Testing for Retrograde Conduction....................55
2.7.3 Pacing Threshold ...............................................56
2.7.4 Arrhythmia Induction Features...........................56
2.7.5 Manual Shock ....................................................57
2.7.6 Manual ATP .......................................................57
2.7.7 Test Shock .........................................................57
2.8 Special Features........................................................58
2.8.1 Detection and Therapy Status ...........................58
2.8.2 Home Monitoring (Lexos DR-T and VR-T Only) 59
2.8.3 Real-time IEGM Transmission...........................64
2.8.4 Capacitor Reformation .......................................64
2.8.5 Patient and Implant Data ...................................65
2.8.6 System Status....................................................66
2.8.7 Holter Memory ...................................................66
2.8.8 Real-time IEGM .................................................69
2.8.9 Brady Diagnostics ..............................................70
3. Sterilization and Storage..............................................72
4. Implant Procedure ........................................................74
4.1 Implant Preparation ...................................................74
4.2 Lead System Evaluation ............................................76
4.3 Opening the Sterile Container ...................................76
4.4 Pocket Preparation ....................................................77
4.5 Lead to Device Connection .......................................78
4.6 Blind Plug Connection ...............................................81
4.7 Pacemaker Interaction Testing ..................................83
4.8 Program the ICD........................................................86
4.9 Implant the ICD..........................................................87
5. Follow-up Procedures ..................................................91
5.1 General Considerations.............................................91
5.2 Longevity....................................................................92
5.3 Explantation ...............................................................94
6. Technical Specifications ..............................................95
Appendix A...........................................................................103
iv Lexos Technical Manual
*Lexos VR and VR-T ICDs do not have an atrial pace/sense port
Lexos Specifications and Description
Battery Voltage: 6.3 Volts Maximum Shock Energy: 30 joules Defibrillation Lead Ports Two DF-1 (3.2 mm) Pacing Lead Ports Two IS-1 (3.2 mm) (one
for Lexos VR and VR-T) Dimension: 67 x 55 x 12 mm Volume: 32 cc Mass: 80 g
Housing Material: Titanium Header Material: Epoxy Resin Sealing Plug Material: Silicone Battery Composition Li / MnO2
Lexos Technical Manual 1
1. General
1.1 System Description
The Lexos family of Implantable Cardioverter Defibrillators (ICDs) detect and treat ventricular tachyarrhythmias and provide rate adaptive bradycardia pacing support. The ICDs are designed to collect diagnostic data to aid the physician’s assessment of a patient’s condition and the performance of the implanted device.
The Lexos ICDs provide therapy for ventricular tachyarrhythmias with a sophisticated range of programmable anti-tachycardia pacing (ATP), and/or defibrillation therapy. The shock polarity and energy may be programmed to tailor the therapy to appropriately treat each patient's tachyarrhythmias. The ICDs provide shock therapies with programmable energies from 5 to 30 joules.
2 Lexos Technical Manual
The Lexos family of ICDs includes the following members:
Lexos DR provides dual chamber rate adaptive bradycardia pacing support. The ICD uses atrial and ventricular sensing/pacing leads to provide enhanced atrial and ventricular tachyarrhythmia discrimination through BIOTRONIK’s SMART Detection
TM
algorithm.
Lexos DR-T is identical to the Lexos DR with the added functionality of BIOTRONIK’s Home Monitoring system. The Home Monitoring System enables automatic exchange of information about a patient’s cardiac status from the implant to the physician remotely.
Lexos VR provides single chamber rate adaptive bradycardia pacing support.
Lexos VR-T is identical to the Lexos VR with the added functionality of BIOTRONIK’s Home Monitoring system. The Home Monitoring System enables automatic exchange of information about a patient’s cardiac status from the implant to the physician remotely.
The Lexos DR and DR-T have two DF-1 defibrillation/ cardioversion and two IS-1 pacing/sensing header ports. The Lexos VR and VR-T have two DF-1 defibrillation/ cardioversion and one IS-1 pacing/sensing header ports. IS-1 refers to the international standard whereby leads and generators from different manufacturers are assured a basic fit [Reference ISO 5841-3:1992]. DF-1 refers to the international standard for defibrillation lead connectors [Reference ISO 11318:1993].
External devices that interact with and test the implantable devices are also part of the ICD System. These external devices include the TMS 1000 and the EPR 1000
PLUS
PLUS
Tachyarrhythmia Monitoring System
Programming and Monitoring System. These programmers are used to interrogate and program the ICD.
1.2 Indications and Usage
The Lexos Implantable Cardioverter Defibrillators (ICDs) are intended to provide ventricular anti-tachycardia pacing and ventricular defibrillation, for automated treatment of life­threatening ventricular arrhythmias.
Lexos Technical Manual 3
1.3 Contraindications
Do not use the Lexos Implantable Cardioverter Defibrillators (ICDs) in patients:
Whose ventricular tachyarrhythmias may have transient or reversible causes including:
acute myocardial infarction
digitalis intoxication
drowning
electrocution
electrolyte imbalance
sepsis
hypoxia
Patients with incessant VT of VF
Patients with unipolar pacemaker
Patients whose only disorder is bradyarrhythmia or atrial
arrhythmia
1.4 Warnings and Precautions
MRI (Magnetic Resonance Imaging) - Do not expose a patient
to MRI device scanning. Strong magnetic fields may damage the device and cause injury to the patient.
Electrical Isolation - To prevent inadvertent arrhythmia induction, electrically isolate the patient during the implant procedure from potentially hazardous leakage currents.
Lead Systems - The use of another manufacturer’s ICD lead system may cause potential adverse consequences such as under sensing of cardiac activity and failure to deliver necessary therapy.
Resuscitation Availability - Do not perform induction testing unless an alternate source of patient defibrillation such as an external defibrillator is readily available. In order to implant the ICD system, it is necessary to induce and convert the patient’s ventricular tachyarrhythmias.
4 Lexos Technical Manual
Unwanted Shocks – Always program the VT/VF Detection and
Therapy status to DISABLED prior to handling the device to prevent the delivery of serious shocks to the patient or the person handling the device during the implant procedure.
Rate-Adaptive Pacing – Use rate-adaptive pacing with care in patients unable to tolerate increased pacing rates.
1.4.1 Sterilization, Storage, and Handling
Device Packaging - Do not use the device if the device’s
packaging is wet, punctured, opened or damaged because the integrity of the sterile packaging may be compromised. Return the device to BIOTRONIK.
Re-sterilization - Do not re-sterilize and re-implant explanted devices.
Storage (temperature) - Store the device between 5° to 55°C (41° - 131° F) because temperatures outside this range could damage the device.
Storage (magnets) - To avoid damage to the device, store the device in a clean area, away from magnets, kits containing magnets, and sources of electromagnetic interference (EMI).
Temperature Stabilization - Allow the device to reach room temperature before programming or implanting the device because temperature extremes may affect initial device function.
Use Before Date - Do not implant the device after the USE BEFORE DATE because the device may have reduced longevity.
1.4.2 Device Implantation and Programming
Blind Plug - A blind plug must be inserted and firmly connected
into any unused header port to prevent chronic fluid influx and possible shunting of high energy therapy.
Lexos Technical Manual 5
Capacitor Reformation - Infrequent charging of the high voltage
capacitors may extend the charge times of the ICD. The capacitors may be reformed manually, or the ICD may be programmed to reform the capacitors automatically. For further information, please refer to
Section 2.8.4
, Capacitor
Reformation.
Connector Compatibility - ICD and lead system compatibility should be confirmed prior to the implant procedure. Consult your BIOTRONIK representative regarding lead/pulse generator compatibility prior to the implantation of an ICD system. For further information, please refer to
Appendix A
.
ERI (Elective Replacement Indicator) - Upon reaching ERI, the
battery has sufficient energy remaining to continue monitoring for at least three months and to deliver a minimum of six 30 joule shocks. After this period, all tachyarrhythmia detection and therapy is disabled. Bradycardia functions are still active at programmed values until the battery voltage drops below
3.0 volts.
Magnets - Positioning of a magnet or the programming wand over the ICD will suspend tachycardia detection and treatment. The minimum magnet strength required to suspend tachycardia treatment is 1.8 mT. When the magnet strength decreases to less than 1 mT, the reed contact is reopened.
Pacemaker/ICD Interaction - In situations where an ICD and a pacemaker are implanted in the same patient, interaction testing should be completed. If the interaction between the ICD and the pacemaker cannot be resolved through repositioning of the leads or reprogramming of either the pacemaker or the ICD, the pacemaker should not be implanted (or explanted if previously implanted).
Programmed Parameters – Program the device parameters to appropriate values based on the patient’s specific arrhythmias and condition.
Programmers - Use only BIOTRONIK programmers to communicate with the device (TMS 1000
PLUS
or EPR 1000
PLUS
).
6 Lexos Technical Manual
Sealing System - Failure to properly insert the torque wrench
into the perforation at an angle perpendicular to the connector receptacle may result in damage to the sealing system and its self-sealing properties.
Programming Wand Separation Distance – The wand must not be placed closer than 2 cm to the device (implanted or out of the box). Programming wand distance closer than 2 cm may damage the device.
Defibrillation Threshold - Be aware that the changes in the patient’s condition, drug regimen, and other factors may change the defibrillation threshold (DFT) which may result in non­conversion of the arrhythmia post-operatively. Successful conversion of ventricular fibrillation or ventricular tachycardia during arrhythmia conversion testing is no assurance that conversion will occur post-operatively.
Manual Shocks – User-commanded shocks may be withheld if the ICD is already busy processing a manual command or the Battery Status is low.
Charge Time - When preparing a high energy shock the charge circuit stops charging the capacitors after 20 seconds, and delivers the stored energy as shock therapy. After the device reaches ERI the stored energy may be less than 30 joules per shock.
Shock Therapy Confirmation – Programming CONFIRMATION to OFF may increase the incidence of the ICD delivering inappropriate shocks.
Shock Impedance - If the shock impedance is less than twenty­five ohms, reposition the lead system to allow a greater distance between the electrodes. Never implant the device with a lead system that has measured shock impedance as less than twenty-five ohms. Damage to the device may result.
1.4.3 Lead Evaluation and Connection
Capping Leads - If a lead is abandoned rather than removed, it
must be capped to ensure that it is not a pathway for currents to or from the heart.
Lexos Technical Manual 7
Gripping Leads - Do not grip the lead with surgical instruments
or use excessive force or surgical instruments to insert a stylet into a lead.
Kinking Leads - Do not kink leads. This may cause additional stress on the leads that can result in damage to the lead.
Liquid Immersion - Do not immerse leads in mineral oil, silicone oil, or any other liquid.
Short Circuit - Ensure that none of the lead electrodes are in contact (a short circuit) during delivery of shock therapy as this may cause current to bypass the heart or cause damage to the ICD system.
Far-field sensing of signals from the atrium in the ventricular channel or ventricular signals in the atrial channel should be avoided by appropriate lead placement, programming of pacing/sensing parameters, and maximum sensitivity settings. If it is necessary to modify the Far Field Blanking parameter, the parameter should be lengthened only long enough to eliminate far-field sensing as evidenced on the IEGMs. Extending the parameter unnecessarily may cause under sensing of actual atrial or ventricular events.
Suturing Leads - Do not suture directly over the lead body as this may cause structural damage. Use the appropriate suture sleeve to immobilize the lead and protect it against damage from ligatures.
Tricuspid Valve Bioprosthesis - Use ventricular transvenous leads with caution in patients with a tricuspid valvular bioprosthesis.
Setscrew Adjustment – Back-off the setscrew(s) prior to insertion of lead connector(s) as failure to do so may result in damage to the lead(s), and/or difficulty connecting lead(s).
Cross Threading Setscrew(s) – To prevent cross threading the setscrew(s), do not back the setscrew(s) completely out of the threaded hole. Leave the torque wrench in the slot of the setscrew(s) while the lead is inserted.
Tightening Setscrew(s) – Do not overtighten the setscrew(s). Use only the BIOTRONIK supplied torque wrench.
8 Lexos Technical Manual
Sealing System – Be sure to properly insert the torque wrench into the perforation at an angle perpendicular to the
connector receptacle. Failure to do so may result in damage to the plug and its self-sealing properties.
1.4.4 Follow-up Testing
Defibrillation Threshold - Be aware that changes in the
patient’s condition, drug regimen, and other factors may change the defibrillation threshold (DFT), which may result in non­conversion of the arrhythmia post-operatively. Successful conversion of ventricular fibrillation or ventricular tachycardia during arrhythmia conversion testing is no assurance that conversion will occur post-operatively.
Resuscitation Availability - Ensure that an external defibrillator and medical personnel skilled in cardiopulmonary resuscitation (CPR) are present during post-implant device testing should the patient require external rescue.
Safe Program – Within the EP Test screen, pressing the “Safe Program” key on the programmer head does not immediately send the safe program to the ICD. Pressing the “Safe Program” key activates the emergency function screen, but an additional screen touch is required to send the safe program to the ICD.
1.4.5 Pulse Generator Explant and Disposal
Device Incineration – Never incinerate the ICD due to the
potential for explosion. The ICD must be explanted prior to cremation.
Explanted Devices – Return all explanted devices to BIOTRONIK.
Unwanted Shocks – Always program the therapy status to DISABLED prior to handling the device to prevent the delivery of serious shocks to the patient or the person handling the device during the implant procedure.
Lexos Technical Manual 9
1.4.6 Hospital and Medical Hazards
Electromagnetic interference (EMI) signals present in hospital and medical environments may affect the function of any ICD or pacemaker. The ICD is designed to selectively filter out EMI noise. However, due to the variety of EMI signals, absolute protection from EMI is not possible with this or any other ICD.
The ICD system should have detection and therapy disabled prior to performing any of the following medical procedures. In addition, the ICD should be checked after the procedures to assure proper programming:
Diathermy - Diathermy therapy is not recommended for ICD patients due to possible heating effects of the pulse generator and at the implant site. If diathermy therapy must be used, it should not be applied in the immediate vicinity of the pulse generator or lead system.
Electrocautery - Electrosurgical cautery could induce ventricular arrhythmias and/or fibrillation, or may cause device malfunction or damage. If use of electrocautery is necessary, the current path and ground plate should be kept as far away from the pulse generator and leads as possible (at least 6 inches (15 cm)).
External Defibrillation - The device is protected against energy normally encountered from external defibrillation. However, any implanted device may be damaged by external defibrillation procedures. In addition, external defibrillation may also result in permanent myocardial damage at the electrode-tissue interface as well as temporary or permanent elevated pacing thresholds. When possible, observe the following precautions:
10 Lexos Technical Manual
Position the adhesive electrodes or defibrillation paddles of the external defibrillator anterior-posterior or along a line perpendicular to the axis formed by the implanted device and the heart.
Set the energy to a level not higher than is required to achieve defibrillation.
Place the paddles as far as possible away from the implanted device and lead system.
After delivery of an external defibrillation shock, interrogate the ICD to confirm device status and proper function.
Lithotripsy - Lithotripsy may damage the ICD. If lithotripsy must be used, avoid focusing near the ICD implant site.
MRI (Magnetic Resonance Imaging) - Do not expose a patient to MRI device scanning. Strong magnetic fields may damage the device and cause injury to the patient.
Radiation - High radiation sources such as cobalt 60 or gamma radiation should not be directed at the pulse generator. If a patient requires radiation therapy in the vicinity of the pulse generator, place lead shielding over the device to prevent radiation damage and confirm its function after treatment.
Radio Frequency Ablation - Prior to performing an ablation procedure, deactivate the ICD during the procedure. Avoid applying ablation energy near the implanted lead system whenever possible.
1.4.7 Home and Occupational Hazards
Patients should be directed to avoid devices that generate strong electromagnetic interference (EMI) or magnetic fields. EMI could cause device malfunction or damage resulting in non-detection or delivery of unneeded therapy. Moving away from the source or turning it off will usually allow the ICD to return to its normal mode of operation.
Lexos Technical Manual 11
The following equipment (and similar devices) may affect normal ICD operation: electric arc or resistance welders, electric melting fu r na c es , r ad i o/t e le v isi on and radar t r an s mi t ter s , power-generating facilities, high-voltage transmission lines, and electrical ignition systems (of gasoline-powered devices) if protective hoods, shrouds, etc., are removed.
1.4.8 Cellular Phones
Testing has indicated there may be a potential interaction between cellular phones and BIOTRONIK ICD systems. Potential effects may be due to either the cellular phone signal or the magnet within the telephone and may include inhibition of therapy when the telephone is within 6 inches (15 cm) of the ICD, when the ICD is programmed to standard sensitivity.
Patients having an implanted BIOTRONIK ICD who operate a cellular telephone should:
Maintain a minimum separation of 6 inches (15 cm) between a hand-held personal cellular telephone and the implanted device.
Set the telephone to the lowest available power setting, if possible.
Patients should hold the phone to the ear opposite the side of the implanted device. Patients should not carry the telephone in a breast pocket or on a belt over or within 6 inches (15 cm) of the implanted device as some telephones emit signals when they are turned ON, but not in use (i.e., in the listen or stand-by mode). Store the telephone in a location opposite the side of implant.
Based on results to date, adverse effects resulting from interactions between cellular telephones and implanted ICDs have been transitory. The potential adverse effects could include inhibition or delivery of additional therapies. If electromagnetic interference (EMI) emitting from a telephone does adversely affect an implanted ICD, moving the telephone away from the immediate vicinity of the ICD should restore normal operation. A recommendation to address every specific interaction of EMI with implanted ICDs is not possible due to the disparate nature of EMI.
12 Lexos Technical Manual
1.4.9 Electronic Article Surveillance (EAS)
Equipment such as retail theft prevention systems may interact with pulse generators. Patients should be advised to walk directly through and not to remain near an EAS system longer than necessary.
1.4.10 Home Appliances
Home appliances normally do not affect ICD operation if the appliances are in proper working condition and correctly grounded and shielded. There have been reports of the interaction of electric tools or other external devices (e.g. electric drills, older models of microwave ovens, electric razors, etc.) with ICDs when they are placed in close proximity to the device.
Lexos Technical Manual 13
1.5 Adverse Events
1.5.1 Potential Adverse Events
The following is a list of the potential risks that may occur with this device:
Acceleration of arrhythmias
Air embolism
Bleeding
Chronic nerve damage
Erosion
Excessive fibrotic tissue growth
Extrusion
Fluid accumulation
Formation of hematomas or cysts
Inappropriate shocks
Infection
Keloid formation
Lead abrasion and discontinuity
Lead migration / dislodgment
Myocardial damage
Pneumothorax
Shunting current or insulating myocardium during
defibrillation with internal or external paddles
Potential mortality due to inability to defibrillate or pace
Thromboemboli
Venous occlusion
Venous or cardiac perforation
Patients susceptible to frequent shocks despite antiarrhythmic medical management may develop psychological intolerance to an ICD system that may include the following:
Dependency
Depression
14 Lexos Technical Manual
Fear of premature battery depletion
Fear of shocking while conscious
Fear that shocking capability may be lost
Imagined shocking (phantom shock)
There may be other risks associated with this device that are currently unforeseeable.
1.5.2 Observed Adverse Events
A clinical study of the Phylax AV involved 128 devices implanted in 126 patients with a cumulative implant duration of 795.5 months (mean implant duration 6.3 months).
N
There were a total of two deaths during the course of the trial; neither of which was judged by the clinical study investigator to be device related. The two deaths were related to heart failure and pneumonia. Both of the deaths occurred more than three months post implant.
Three devices were explanted during the trial. One device was explanted secondary to the patient reporting pain at the implant site; the patient was subsequently implanted with another device. One device was explanted due to a random component failure, and the other device was explanted after reaching ERI, which was anticipated based on the number of shocks delivered. These two patients were subsequently implanted with other Phylax AV ICDs.
Table 1
reported during the clinical study regardless of whether or not the event was related to the ICD system. A complication was defined as a clinical event that resulted in additional invasive intervention, injury, or death. An observation was defined as a clinical event that did not result in additional invasive intervention, injury, or death.
:
OTE
The Phylax AV ICD is an earlier generation of BIOTRONIK devices. The Lexos family is based upon the Phylax AV and other BIOTRONIK ICDs (i.e., Phylax, Tachos, and Belos families of ICDs).
provides a summary of the adverse events that were
Lexos Technical Manual 15
Table 1: Reported Adverse Events
# of
Complications Total
Patients
with AEs
14 11.1% 18 0.27
% of
Patients
with AEs
# of
AEs
AE /
pt-yrs
Lead Repositioning 10 7.9% 12 0.18 Discomfort at Implant
1 0.8% 1 0.02 Site Infection 1 0.8% 1 0.02 Thrombus 1 0.8% 1 0.02 Pneumothorax 1 0.8% 1 0.02 ERI 1 0.8% 1 0.02 Random Component
1 0.8% 1 0.02 Failure
Observations Total
47 37.3% 74 1.12
T-wave Oversensing 7 5.6% 7 0.11 Increased Pacing
7 5.6% 7 0.11 Threshold Required antiarrhythmic
7 5.6% 7 0.11 drug therapy SVT Therapy-Unrelated
6 4.8% 8 0.12 to SMART Software version I­GAV.1.U
1
6 4.8% 6 0.09
Detection 5 4.0% 5 0.08 Lead revision at implant 5 4.0% 5 0.08 TMS 10002 4 3.2% 4 0.06 Lead difficulties at
3 2.4% 3 0.05 Implant Difficulties with
3 2.4% 3 0.05 Telemetry Atrial Lead Dislodgment 2 1.6% 2 0.03
16 Lexos Technical Manual
# of
SVT Therapy-Related to
Patients
with AEs
2 1.6% 4 0.06
% of
Patients
with AEs
# of
AEs
AE /
pt-yrs
SMART Initial therapy did not
2 1.6% 2 0.03 convert VT/VF Low P/R-Wave
2 1.6% 2 0.03 Amplitude Intermittent Under /
2 1.6% 2 0.03 Oversensing Lead Repositioning at
2 1.6% 2 0.03 implant Asynchronous Pacing 2 1.6% 2 0.03 Atrial Arrhythmias 2 1.6% 2 0.03 Atrial arrhythmia with
1 0.8% 1 0.02 ventrical tracking External cardioversion
1 0.8% 1 0.02 due to AT P-wave changes with
1 0.8% 1 0.02 position Patient Symptomatic at
1 0.8% 1 0.02 Upper Tracking Rate Diaphragmatic Pacing 1 0.8% 1 0.02 Myocardial Infarction 1 0.8% 1 0.02 Cautery caused Shock
1 0.8% 1 0.02 Delivery Phantom programming 1 0.8% 1 0.02
Number of Patients = 126, Number of Patient-Years = 66.3,
see next page for notes on table.
Lexos Technical Manual 17
1. This category includes various anomalies that were related to the programmer software used in the clinical study, I-GAV.1.U. Each of these events has been resolved through revisions to the programmer software resulting in version I-GAV.2.U.
2. This category includes any difficulties encountered while using the TMS 1000 System, which is a commercially available device that was used during the clinical investigation.
PLUS
Tachyarrhythmia Monitoring
1.6 Clinical Studies
1.6.1 Tachos DR
The Tachos DR clinical evaluation involved 57 patients implanted with a Tachos DR outside of the United States.
N
:
OTE
The clinical study information included in this technical manual was performed with the Phylax AV and Tachos DR ICDs. The Lexos DR is a downsized version of the Belos DR, which was also based on and approved with this data. The clinical study data presented here is applicable because the Lexos ICDs are downsized versions of the Belos and Tachos families of ICDs. The Lexos ICDs are slightly different as compared to the Belos ICDs in the following areas:
Reduced size from 39 cc to 32 cc
Additional shock waveform - Biphasic 2ms (see
Section 2.5.3.3
Upper Tracking Rate (UTR) programmable in the
VT-1 therapy zone
Minimum shock energy is 5 Joules
Due to the similarities between the Lexos family and Phylax, Tachos, and Belos families of ICDs and the limited nature of these changes, a clinical study of the Lexos DR/DR-T ICD was determined to be unnecessary.
)
18 Lexos Technical Manual
1.6.1.1 Study Objectives
The objective of the clinical evaluation was to gather basic information about the function and performance of the Tachos DR ICD in patients with standard ICD indications.
1.6.1.2 Results
The mean implant duration was 5.7 months with cumulative implant duration of 323 patient months. No unanticipated events were reported during the evaluation. A summary of the results obtained during the evaluation is provided in the following table.
1.6.1.3 Survival
During the initial experience outside of the United States with the Tachos DR, there have been no sudden cardiac deaths reported. There was one death reported, which was unrelated to the implanted device.
Table 2: Tachos DR Study Results
Evaluation Results
Appropriate Atrial Sensing and Pacing 99% (126/127)
Appropriate Ventricular Sensing and Pacing
99% (122/123)
Appropriate Ventricular Tachyarrhythmia Detection and
96% (116/121)
Conversion
Complication Rate (per patient) 3.5%
Complication Rate (per patient-year) 0.074
Sudden Cardiac Death Survival Rate 100%
Overall Survival Rate 98%
Lexos Technical Manual 19
1.6.2 Phylax AV
The Phylax AV clinical study involved 126 patients (111 males (88.1%) and 15 females (11.9%) with a mean age of 66 years (range: 22-87 years) and a left ventricular ejection fraction of 31% (range: 10-60%). Most patients (80.2%) presented with coronary artery disease / ischemic cardiomyopathy; 65.1% presented with monomorphic ventricular tachycardia (MVT) as their primary tachyarrhythmia.
N
1.6.2.1 Methods
The multi-center, non-randomized clinical investigation was designed to validate the safety and effectiveness of the Phylax AV through an analysis of the unanticipated adverse device effect (UADE) rate. The specific predefined objectives of the investigation included UADE-free survival rate, morbidity rate, sudden cardiac death (SCD) survival rate, the appropriate sensing and pacing rate, detection and conversion of ventricular tachyarrhythmias, and the appropriate rejection of atrial tachyarrhythmias.
1.6.2.2 Results
The mean implant duration was 6.3 ± 0.4 months with cumulative implant duration of 795.5 months. There were 20 patients followed for over twelve months and 62 patients followed for over six months during the study period from February 5, 1999 to April 15, 2000. The patient follow-up compliance rate was 98.4% out of 319 required follow-ups.
Table 3
the predefined endpoints.
:
OTE
The Phylax AV ICD is an earlier generation of BIOTRONIK devices. The Lexos family is based upon the Phylax AV and other BIOTRONIK ICDs (i.e., Phylax, Tachos, and Belos families of ICDs). Therefore, the clinical data from the Phylax AV was used to support the safety and effectiveness of the Lexos DR.
provides a summary of the results of the study group for
20 Lexos Technical Manual
Table 3: Clinical Study Results
Description
UADE-free Survival Rate (patients with at least 3 months follow-up)
Complication Rate
Sudden Cardiac Death Survival Rate
Appropriate Bradycardia Sensing and Pacing Rate
Detection and Conversion of Ventricular Tachyarrhythmias
Appropriate Rejection of Atrial Tachyarrhythmias
Study Group
[95% CI]
100% (85/85)
[96.5%, 100%]
11.1% (14/126) [0%, 16.8%]
100% (124/124)
[97.6% 100%]
96.2%(1141/1186 )
[95.2%, 100%]
98.2% (650/662) [97.1%, 100%]
94% (138/147) [89.6%, 100%]
1.6.2.3 SMART Detection Algorithm
The SMART Detection algorithm is an integral function of BIOTRONIK’s dual chamber ICD product line (i.e., Phylax AV, Tachos DR, and Belos DR) and is designed to discriminate life­threatening ventricular tachycardias from relatively harmless atrial tachyarrhythmias. This algorithm uses information about the signals from the atrial and ventricular lead systems and is designed to reduce the amount of inappropriate therapy that might be delivered as a result of a supraventricular tachycardia (SVT). Neither the SMART Detection algorithm nor the ICDs are designed to detect or deliver therapy to terminate atrial arrhythmias, and therefore this is not the purpose of the algorithm or the device.
Lexos Technical Manual 21
During the Phylax AV clinical study, specific data was collected to demonstrate the ability of the SMART Detection algorithm to discriminate between SVT and VT. The Phylax AV demonstrated the ability to withhold inappropriate therapy in approximately 94% of the SVT episodes that were reported during the study. In addition, the SMART Detection algorithm appropriately delivered therapy in 100% of the ventricular episodes in which the feature was activated. At routine follow­ups, the algorithm was activated in 80% of patients enrolled into the study, which further supports the overall ability of the algorithm to appropriately discriminate between SVT and VT. In addition, during the clinical study, the investigators indicated that the primary reason for selecting a dual-chamber ICD was SVT discrimination for 70% of the patients enrolled.
1.7 Patient Selection and Treatment
1.7.1 Individualization of Treatment
Determine whether the expected device benefits outweigh the possibility of early device replacement for patients whose ventricular tachyarrhythmias require frequent shocks.
Determine if the device and programmable options are appropriate for patients with drug-resistant supraventricular tachyarrhythmias (SVTs), because drug-resistant SVTs can initiate unwanted device therapy.
Direct any questions regarding individualization of patient therapy to your BIOTRONIK representative or BIOTRONIK technical services at 1-800-547-0394.
The prospective patient’s size and activity level should be evaluated to determine whether a pectoral or abdominal implant is suitable. It is strongly recommended that candidates for an ICD have a complete cardiac evaluation including EP testing prior to device implant to gather electrophysiologic information, including the rates and classifications of all the patient’s cardiac rhythms. When gathering this information, delineate all clinically significant ventricular and atrial arrhythmias, whether they occur spontaneously or during EP testing.
22 Lexos Technical Manual
If the patient’s condition permits, use exercise stress testing to do the following:
Determine the maximum rate of the patient’s normal rhythm.
Identify any supraventricular tachyarrhythmias.
Identify exercise-induced tachyarrhythmias.
The maximum exercise rate or the presence of supraventricular tachyarrhythmias may influence selection of programmable parameters. Holter monitoring or other extended ECG monitoring also may be helpful.
If the patient is being treated with antiarrhythmic or cardiac drugs, the patient should be on a maintenance drug dose rather than a loading dose at the time of pulse generator implantation. If changes to drug therapy are made, repeated arrhythmia inductions are recommended to verify pulse generator detection and conversion. The pulse generator also may need to be reprogrammed.
Changes in a patient’s antiarrhythmic drug or any other medication that affect the patient’s normal cardiac rate or conduction can affect the rate of tachyarrhythmias and/or efficacy of therapy.
If another cardiac surgical procedure is performed prior to implanting the pulse generator, it may be preferable to implant the lead system at that time. This may prevent the need for an additional thoracic operation.
1.7.2 Specific Patient Populations
Pregnancy - If there is a need to image the device, care should
be taken to minimize radiation exposure to the fetus and the mother.
Nursing Mothers - Although appropriate biocompatibility testing has been conducted for this implant device, there has been no quantitative assessment of the presence of leachables in breast milk.
Geriatric Patients - Most (72%) of the patients receiving an ICD in the Phylax AV clinical study were over the age of 60 years (see Clinical Studies).
Lexos Technical Manual 23
Handicapped and Disabled Patients - Special care is needed
in using this device for patients using an electrical wheel chair or other electrical (external or implanted) devices.
1.8 Patient Counseling Information
The pulse generator is subject to random component failure. Such failure could cause inappropriate shocks, induction of arrhythmias or inability to sense arrhythmias, and could lead to the patient’s death.
Persons administering CPR may experience the presence of voltage on the patient’s body surface (tingling) when the patient’s ICD system delivers a shock.
A patient manual is available for the patient, patient’s relatives, and other interested people. Discuss the information in the manual with concerned individuals both before and after pulse generator implantation so they are fully familiar with operation of the device. (For additional copies of the patient manual, contact the BIOTRONIK at the address listed in this manual.)
1.9 Evaluating Prospective ICD Patients
The prospective ICD implant candidate should undergo a cardiac evaluation to classify any and all tachyarrhythmias. In addition, other patient specific cardiac information will help in selecting the optimal device settings. This evaluation may include, but is not limited to:
an evaluation of the specific tachycardia rate(s)
the confirmation and/or evaluation of any
supraventricular arrhythmias or bradyarrhythmias
the evaluation of various ATP and cardioversion therapies
the presence of any post-shock arrhythmias, and
an evaluation of the maximum sinus rate during exercise
If a patient’s drug regimen is changed or adjusted while the ICD is implanted, additional EP testing may be required to determine if detection or therapy parameter settings are relevant and appropriate.
24 Lexos Technical Manual
Empirical changes to the detection or therapy parameters should be assessed based on patient safety. Some changes may necessitate a re-assessment of sensing, pacing, or arrhythmia conversion treatment. Thorough technical knowledge of BIOTRONIK ICDs, additional ICD experience, and individual medical judgment will aid in determining the need for additional testing and follow-up.
Loading...
+ 80 hidden pages