The Lexos family of Implantable Cardioverter Defibrillators
(ICDs) detect and treat ventricular tachyarrhythmias and provide
rate adaptive bradycardia pacing support. The ICDs are
designed to collect diagnostic data to aid the physician’s
assessment of a patient’s condition and the performance of the
implanted device.
The Lexos ICDs provide therapy for ventricular tachyarrhythmias
with a sophisticated range of programmable anti-tachycardia
pacing (ATP), and/or defibrillation therapy. The shock polarity
and energy may be programmed to tailor the therapy to
appropriately treat each patient's tachyarrhythmias. The ICDs
provide shock therapies with programmable energies from 5 to
30 joules.
2 Lexos Technical Manual
The Lexos family of ICDs includes the following members:
• Lexos DR provides dual chamber rate adaptive
bradycardia pacing support. The ICD uses atrial and
ventricular sensing/pacing leads to provide enhanced
atrial and ventricular tachyarrhythmia discrimination
through BIOTRONIK’s SMART Detection
TM
algorithm.
• Lexos DR-T is identical to the Lexos DR with the added
functionality of BIOTRONIK’s Home Monitoring system.
The Home Monitoring System enables automatic
exchange of information about a patient’s cardiac status
from the implant to the physician remotely.
• Lexos VR-T is identical to the Lexos VR with the added
functionality of BIOTRONIK’s Home Monitoring system.
The Home Monitoring System enables automatic
exchange of information about a patient’s cardiac status
from the implant to the physician remotely.
The Lexos DR and DR-T have two DF-1 defibrillation/
cardioversion and two IS-1 pacing/sensing header ports. The
Lexos VR and VR-T have two DF-1 defibrillation/ cardioversion
and one IS-1 pacing/sensing header ports. IS-1 refers to the
international standard whereby leads and generators from
different manufacturers are assured a basic fit [Reference ISO
5841-3:1992]. DF-1 refers to the international standard for
defibrillation lead connectors [Reference ISO 11318:1993].
External devices that interact with and test the implantable
devices are also part of the ICD System. These external devices
include the TMS 1000
and the EPR 1000
PLUS
PLUS
Tachyarrhythmia Monitoring System
Programming and Monitoring System.
These programmers are used to interrogate and program the
ICD.
1.2 Indications and Usage
The Lexos Implantable Cardioverter Defibrillators (ICDs) are
intended to provide ventricular anti-tachycardia pacing and
ventricular defibrillation, for automated treatment of lifethreatening ventricular arrhythmias.
Lexos Technical Manual 3
1.3 Contraindications
Do not use the Lexos Implantable Cardioverter Defibrillators
(ICDs) in patients:
• Whose ventricular tachyarrhythmias may have transient
or reversible causes including:
− acute myocardial infarction
− digitalis intoxication
− drowning
− electrocution
− electrolyte imbalance
− sepsis
− hypoxia
• Patients with incessant VT of VF
• Patients with unipolar pacemaker
• Patients whose only disorder is bradyarrhythmia or atrial
arrhythmia
1.4 Warnings and Precautions
MRI (Magnetic Resonance Imaging) - Do not expose a patient
to MRI device scanning. Strong magnetic fields may damage
the device and cause injury to the patient.
Electrical Isolation - To prevent inadvertent arrhythmia
induction, electrically isolate the patient during the implant
procedure from potentially hazardous leakage currents.
Lead Systems - The use of another manufacturer’s ICD lead
system may cause potential adverse consequences such as
under sensing of cardiac activity and failure to deliver necessary
therapy.
Resuscitation Availability - Do not perform induction testing
unless an alternate source of patient defibrillation such as an
external defibrillator is readily available. In order to implant the
ICD system, it is necessary to induce and convert the patient’s
ventricular tachyarrhythmias.
4 Lexos Technical Manual
Unwanted Shocks – Always program the VT/VF Detection and
Therapy status to DISABLED prior to handling the device to
prevent the delivery of serious shocks to the patient or the
person handling the device during the implant procedure.
Rate-Adaptive Pacing – Use rate-adaptive pacing with care in
patients unable to tolerate increased pacing rates.
1.4.1 Sterilization, Storage, and Handling
Device Packaging - Do not use the device if the device’s
packaging is wet, punctured, opened or damaged because the
integrity of the sterile packaging may be compromised. Return
the device to BIOTRONIK.
Re-sterilization - Do not re-sterilize and re-implant explanted
devices.
Storage (temperature) - Store the device between 5° to 55°C
(41° - 131° F) because temperatures outside this range could
damage the device.
Storage (magnets) - To avoid damage to the device, store the
device in a clean area, away from magnets, kits containing
magnets, and sources of electromagnetic interference (EMI).
Temperature Stabilization - Allow the device to reach room
temperature before programming or implanting the device
because temperature extremes may affect initial device function.
Use Before Date - Do not implant the device after the USE
BEFORE DATE because the device may have reduced
longevity.
1.4.2 Device Implantation and Programming
Blind Plug - A blind plug must be inserted and firmly connected
into any unused header port to prevent chronic fluid influx and
possible shunting of high energy therapy.
Lexos Technical Manual 5
Capacitor Reformation - Infrequent charging of the high voltage
capacitors may extend the charge times of the ICD. The
capacitors may be reformed manually, or the ICD may be
programmed to reform the capacitors automatically. For further
information, please refer to
Section 2.8.4
, Capacitor
Reformation.
Connector Compatibility - ICD and lead system compatibility
should be confirmed prior to the implant procedure. Consult
your BIOTRONIK representative regarding lead/pulse generator
compatibility prior to the implantation of an ICD system. For
further information, please refer to
Appendix A
.
ERI (Elective Replacement Indicator) - Upon reaching ERI, the
battery has sufficient energy remaining to continue monitoring for
at least three months and to deliver a minimum of six 30 joule
shocks. After this period, all tachyarrhythmia detection and
therapy is disabled. Bradycardia functions are still active at
programmed values until the battery voltage drops below
3.0 volts.
Magnets - Positioning of a magnet or the programming wand
over the ICD will suspend tachycardia detection and treatment.
The minimum magnet strength required to suspend tachycardia
treatment is 1.8 mT. When the magnet strength decreases to
less than 1 mT, the reed contact is reopened.
Pacemaker/ICD Interaction - In situations where an ICD and a
pacemaker are implanted in the same patient, interaction testing
should be completed. If the interaction between the ICD and the
pacemaker cannot be resolved through repositioning of the
leads or reprogramming of either the pacemaker or the ICD, the
pacemaker should not be implanted (or explanted if previously
implanted).
Programmed Parameters – Program the device parameters to
appropriate values based on the patient’s specific arrhythmias
and condition.
Programmers - Use only BIOTRONIK programmers to
communicate with the device (TMS 1000
PLUS
or EPR 1000
PLUS
).
6 Lexos Technical Manual
Sealing System - Failure to properly insert the torque wrench
into the perforation at an angle perpendicular to the connector
receptacle may result in damage to the sealing system and its
self-sealing properties.
Programming Wand Separation Distance – The wand must
not be placed closer than 2 cm to the device (implanted or out of
the box). Programming wand distance closer than 2 cm may
damage the device.
Defibrillation Threshold - Be aware that the changes in the
patient’s condition, drug regimen, and other factors may change
the defibrillation threshold (DFT) which may result in nonconversion of the arrhythmia post-operatively. Successful
conversion of ventricular fibrillation or ventricular tachycardia
during arrhythmia conversion testing is no assurance that
conversion will occur post-operatively.
Manual Shocks – User-commanded shocks may be withheld if
the ICD is already busy processing a manual command or the
Battery Status is low.
Charge Time - When preparing a high energy shock the charge
circuit stops charging the capacitors after 20 seconds, and
delivers the stored energy as shock therapy. After the device
reaches ERI the stored energy may be less than 30 joules per
shock.
Shock Therapy Confirmation – Programming
CONFIRMATION to OFF may increase the incidence of the ICD
delivering inappropriate shocks.
Shock Impedance - If the shock impedance is less than twentyfive ohms, reposition the lead system to allow a greater distance
between the electrodes. Never implant the device with a lead
system that has measured shock impedance as less than
twenty-five ohms. Damage to the device may result.
1.4.3 Lead Evaluation and Connection
Capping Leads - If a lead is abandoned rather than removed, it
must be capped to ensure that it is not a pathway for currents to
or from the heart.
Lexos Technical Manual 7
Gripping Leads - Do not grip the lead with surgical instruments
or use excessive force or surgical instruments to insert a stylet
into a lead.
Kinking Leads - Do not kink leads. This may cause additional
stress on the leads that can result in damage to the lead.
Liquid Immersion - Do not immerse leads in mineral oil, silicone
oil, or any other liquid.
Short Circuit - Ensure that none of the lead electrodes are in
contact (a short circuit) during delivery of shock therapy as this
may cause current to bypass the heart or cause damage to the
ICD system.
Far-field sensing of signals from the atrium in the ventricular
channel or ventricular signals in the atrial channel should be
avoided by appropriate lead placement, programming of
pacing/sensing parameters, and maximum sensitivity settings. If
it is necessary to modify the Far Field Blanking parameter, the
parameter should be lengthened only long enough to eliminate
far-field sensing as evidenced on the IEGMs. Extending the
parameter unnecessarily may cause under sensing of actual
atrial or ventricular events.
Suturing Leads - Do not suture directly over the lead body as
this may cause structural damage. Use the appropriate suture
sleeve to immobilize the lead and protect it against damage from
ligatures.
Tricuspid Valve Bioprosthesis - Use ventricular transvenous
leads with caution in patients with a tricuspid valvular
bioprosthesis.
Setscrew Adjustment – Back-off the setscrew(s) prior to
insertion of lead connector(s) as failure to do so may result in
damage to the lead(s), and/or difficulty connecting lead(s).
Cross Threading Setscrew(s) – To prevent cross threading
the setscrew(s), do not back the setscrew(s) completely out of
the threaded hole. Leave the torque wrench in the slot of the
setscrew(s) while the lead is inserted.
Tightening Setscrew(s) – Do not overtighten the setscrew(s).
Use only the BIOTRONIK supplied torque wrench.
8 Lexos Technical Manual
Sealing System – Be sure to properly insert the torque
wrench into the perforation at an angle perpendicular to the
connector receptacle. Failure to do so may result in damage to
the plug and its self-sealing properties.
1.4.4 Follow-up Testing
Defibrillation Threshold - Be aware that changes in the
patient’s condition, drug regimen, and other factors may change
the defibrillation threshold (DFT), which may result in nonconversion of the arrhythmia post-operatively. Successful
conversion of ventricular fibrillation or ventricular tachycardia
during arrhythmia conversion testing is no assurance that
conversion will occur post-operatively.
Resuscitation Availability - Ensure that an external defibrillator
and medical personnel skilled in cardiopulmonary resuscitation
(CPR) are present during post-implant device testing should the
patient require external rescue.
Safe Program – Within the EP Test screen, pressing the “Safe
Program” key on the programmer head does not immediately
send the safe program to the ICD. Pressing the “Safe Program”
key activates the emergency function screen, but an additional
screen touch is required to send the safe program to the ICD.
1.4.5 Pulse Generator Explant and Disposal
Device Incineration – Never incinerate the ICD due to the
potential for explosion. The ICD must be explanted prior to
cremation.
Explanted Devices – Return all explanted devices to
BIOTRONIK.
Unwanted Shocks – Always program the therapy status to
DISABLED prior to handling the device to prevent the delivery of
serious shocks to the patient or the person handling the device
during the implant procedure.
Lexos Technical Manual 9
1.4.6 Hospital and Medical Hazards
Electromagnetic interference (EMI) signals present in hospital
and medical environments may affect the function of any ICD or
pacemaker. The ICD is designed to selectively filter out EMI
noise. However, due to the variety of EMI signals, absolute
protection from EMI is not possible with this or any other ICD.
The ICD system should have detection and therapy disabled
prior to performing any of the following medical procedures. In
addition, the ICD should be checked after the procedures to
assure proper programming:
Diathermy - Diathermy therapy is not recommended for ICD
patients due to possible heating effects of the pulse generator
and at the implant site. If diathermy therapy must be used, it
should not be applied in the immediate vicinity of the pulse
generator or lead system.
Electrocautery - Electrosurgical cautery could induce ventricular
arrhythmias and/or fibrillation, or may cause device malfunction
or damage. If use of electrocautery is necessary, the current
path and ground plate should be kept as far away from the pulse
generator and leads as possible (at least 6 inches (15 cm)).
External Defibrillation - The device is protected against energy
normally encountered from external defibrillation. However, any
implanted device may be damaged by external defibrillation
procedures. In addition, external defibrillation may also result in
permanent myocardial damage at the electrode-tissue interface as
well as temporary or permanent elevated pacing thresholds. When
possible, observe the following precautions:
10 Lexos Technical Manual
• Position the adhesive electrodes or defibrillation paddles
of the external defibrillator anterior-posterior or along a
line perpendicular to the axis formed by the implanted
device and the heart.
• Set the energy to a level not higher than is required to
achieve defibrillation.
• Place the paddles as far as possible away from the
implanted device and lead system.
• After delivery of an external defibrillation shock,
interrogate the ICD to confirm device status and proper
function.
Lithotripsy - Lithotripsy may damage the ICD. If lithotripsy must
be used, avoid focusing near the ICD implant site.
MRI (Magnetic Resonance Imaging) - Do not expose a patient
to MRI device scanning. Strong magnetic fields may damage
the device and cause injury to the patient.
Radiation - High radiation sources such as cobalt 60 or gamma
radiation should not be directed at the pulse generator. If a
patient requires radiation therapy in the vicinity of the pulse
generator, place lead shielding over the device to prevent
radiation damage and confirm its function after treatment.
Radio Frequency Ablation - Prior to performing an ablation
procedure, deactivate the ICD during the procedure. Avoid
applying ablation energy near the implanted lead system
whenever possible.
1.4.7 Home and Occupational Hazards
Patients should be directed to avoid devices that generate strong
electromagnetic interference (EMI) or magnetic fields. EMI could
cause device malfunction or damage resulting in non-detection
or delivery of unneeded therapy. Moving away from the source
or turning it off will usually allow the ICD to return to its normal
mode of operation.
Lexos Technical Manual 11
The following equipment (and similar devices) may affect normal
ICD operation: electric arc or resistance welders, electric melting
fu r na c es , r ad i o/t e le v isi on and radar t r an s mi t ter s ,
power-generating facilities, high-voltage transmission lines, and
electrical ignition systems (of gasoline-powered devices) if
protective hoods, shrouds, etc., are removed.
1.4.8 Cellular Phones
Testing has indicated there may be a potential interaction
between cellular phones and BIOTRONIK ICD systems.
Potential effects may be due to either the cellular phone signal or
the magnet within the telephone and may include inhibition of
therapy when the telephone is within 6 inches (15 cm) of the
ICD, when the ICD is programmed to standard sensitivity.
Patients having an implanted BIOTRONIK ICD who operate a
cellular telephone should:
• Maintain a minimum separation of 6 inches (15 cm)
between a hand-held personal cellular telephone and
the implanted device.
• Set the telephone to the lowest available power setting,
if possible.
• Patients should hold the phone to the ear opposite the
side of the implanted device. Patients should not carry
the telephone in a breast pocket or on a belt over or
within 6 inches (15 cm) of the implanted device as some
telephones emit signals when they are turned ON, but
not in use (i.e., in the listen or stand-by mode). Store
the telephone in a location opposite the side of implant.
Based on results to date, adverse effects resulting from
interactions between cellular telephones and implanted ICDs
have been transitory. The potential adverse effects could
include inhibition or delivery of additional therapies. If
electromagnetic interference (EMI) emitting from a telephone
does adversely affect an implanted ICD, moving the telephone
away from the immediate vicinity of the ICD should restore
normal operation. A recommendation to address every specific
interaction of EMI with implanted ICDs is not possible due to the
disparate nature of EMI.
12 Lexos Technical Manual
1.4.9 Electronic Article Surveillance (EAS)
Equipment such as retail theft prevention systems may interact
with pulse generators. Patients should be advised to walk
directly through and not to remain near an EAS system longer
than necessary.
1.4.10 Home Appliances
Home appliances normally do not affect ICD operation if the
appliances are in proper working condition and correctly
grounded and shielded. There have been reports of the
interaction of electric tools or other external devices (e.g. electric
drills, older models of microwave ovens, electric razors, etc.)
with ICDs when they are placed in close proximity to the device.
Lexos Technical Manual 13
1.5 Adverse Events
1.5.1 Potential Adverse Events
The following is a list of the potential risks that may occur with
this device:
• Acceleration of arrhythmias
• Air embolism
• Bleeding
• Chronic nerve damage
• Erosion
• Excessive fibrotic tissue growth
• Extrusion
• Fluid accumulation
• Formation of hematomas or cysts
• Inappropriate shocks
• Infection
• Keloid formation
• Lead abrasion and discontinuity
• Lead migration / dislodgment
• Myocardial damage
• Pneumothorax
• Shunting current or insulating myocardium during
defibrillation with internal or external paddles
• Potential mortality due to inability to defibrillate or pace
• Thromboemboli
• Venous occlusion
• Venous or cardiac perforation
Patients susceptible to frequent shocks despite antiarrhythmic
medical management may develop psychological intolerance to
an ICD system that may include the following:
• Dependency
• Depression
14 Lexos Technical Manual
• Fear of premature battery depletion
• Fear of shocking while conscious
• Fear that shocking capability may be lost
• Imagined shocking (phantom shock)
There may be other risks associated with this device that are
currently unforeseeable.
1.5.2 Observed Adverse Events
A clinical study of the Phylax AV involved 128 devices implanted
in 126 patients with a cumulative implant duration of 795.5
months (mean implant duration 6.3 months).
N
There were a total of two deaths during the course of the trial;
neither of which was judged by the clinical study investigator to
be device related. The two deaths were related to heart failure
and pneumonia. Both of the deaths occurred more than three
months post implant.
Three devices were explanted during the trial. One device was
explanted secondary to the patient reporting pain at the implant
site; the patient was subsequently implanted with another
device. One device was explanted due to a random component
failure, and the other device was explanted after reaching ERI,
which was anticipated based on the number of shocks delivered.
These two patients were subsequently implanted with other
Phylax AV ICDs.
Table 1
reported during the clinical study regardless of whether or not
the event was related to the ICD system. A complication was
defined as a clinical event that resulted in additional invasive
intervention, injury, or death. An observation was defined as a
clinical event that did not result in additional invasive
intervention, injury, or death.
:
OTE
The Phylax AV ICD is an earlier generation of BIOTRONIK
devices. The Lexos family is based upon the Phylax AV and
other BIOTRONIK ICDs (i.e., Phylax, Tachos, and Belos
families of ICDs).
provides a summary of the adverse events that were
Lexos Technical Manual 15
Table 1: Reported Adverse Events
# of
Complications Total
Patients
with AEs
14 11.1% 18 0.27
% of
Patients
with AEs
# of
AEs
AE /
pt-yrs
Lead Repositioning 10 7.9% 12 0.18
Discomfort at Implant
1 0.8% 1 0.02
Site
Infection 1 0.8% 1 0.02
Thrombus 1 0.8% 1 0.02
Pneumothorax 1 0.8% 1 0.02
ERI 1 0.8% 1 0.02
Random Component
1 0.8% 1 0.02
Failure
Observations Total
47 37.3% 74 1.12
T-wave Oversensing 7 5.6% 7 0.11
Increased Pacing
7 5.6% 7 0.11
Threshold
Required antiarrhythmic
7 5.6% 7 0.11
drug therapy
SVT Therapy-Unrelated
6 4.8% 8 0.12
to SMART
Software version IGAV.1.U
1
6 4.8% 6 0.09
Detection 5 4.0% 5 0.08
Lead revision at implant 5 4.0% 5 0.08
TMS 10002 4 3.2% 4 0.06
Lead difficulties at
Number of Patients = 126, Number of Patient-Years = 66.3,
see next page for notes on table.
Lexos Technical Manual 17
1. This category includes various anomalies that were
related to the programmer software used in the clinical study,
I-GAV.1.U. Each of these events has been resolved
through revisions to the programmer software resulting
in version I-GAV.2.U.
2. This category includes any difficulties encountered while
using the TMS 1000
System, which is a commercially available device that
was used during the clinical investigation.
PLUS
Tachyarrhythmia Monitoring
1.6 Clinical Studies
1.6.1 Tachos DR
The Tachos DR clinical evaluation involved 57 patients
implanted with a Tachos DR outside of the United States.
N
:
OTE
The clinical study information included in this technical
manual was performed with the Phylax AV and Tachos DR
ICDs. The Lexos DR is a downsized version of the
Belos DR, which was also based on and approved with this
data. The clinical study data presented here is applicable
because the Lexos ICDs are downsized versions of the
Belos and Tachos families of ICDs. The Lexos ICDs are
slightly different as compared to the Belos ICDs in the
following areas:
− Reduced size from 39 cc to 32 cc
− Additional shock waveform - Biphasic 2ms (see
Section 2.5.3.3
− Upper Tracking Rate (UTR) programmable in the
VT-1 therapy zone
− Minimum shock energy is 5 Joules
Due to the similarities between the Lexos family and Phylax,
Tachos, and Belos families of ICDs and the limited nature of
these changes, a clinical study of the Lexos DR/DR-T ICD
was determined to be unnecessary.
)
18 Lexos Technical Manual
1.6.1.1 Study Objectives
The objective of the clinical evaluation was to gather basic
information about the function and performance of the
Tachos DR ICD in patients with standard ICD indications.
1.6.1.2 Results
The mean implant duration was 5.7 months with cumulative
implant duration of 323 patient months. No unanticipated events
were reported during the evaluation. A summary of the results
obtained during the evaluation is provided in the following table.
1.6.1.3 Survival
During the initial experience outside of the United States with the
Tachos DR, there have been no sudden cardiac deaths
reported. There was one death reported, which was unrelated to
the implanted device.
Table 2: Tachos DR Study Results
Evaluation Results
Appropriate Atrial Sensing and Pacing 99% (126/127)
Appropriate Ventricular Sensing and
Pacing
99% (122/123)
Appropriate Ventricular
Tachyarrhythmia Detection and
96% (116/121)
Conversion
Complication Rate (per patient) 3.5%
Complication Rate (per patient-year) 0.074
Sudden Cardiac Death Survival Rate 100%
Overall Survival Rate 98%
Lexos Technical Manual 19
1.6.2 Phylax AV
The Phylax AV clinical study involved 126 patients (111 males
(88.1%) and 15 females (11.9%) with a mean age of 66 years
(range: 22-87 years) and a left ventricular ejection fraction of
31% (range: 10-60%). Most patients (80.2%) presented with
coronary artery disease / ischemic cardiomyopathy; 65.1%
presented with monomorphic ventricular tachycardia (MVT) as
their primary tachyarrhythmia.
N
1.6.2.1 Methods
The multi-center, non-randomized clinical investigation was
designed to validate the safety and effectiveness of the
Phylax AV through an analysis of the unanticipated adverse
device effect (UADE) rate. The specific predefined objectives of
the investigation included UADE-free survival rate, morbidity
rate, sudden cardiac death (SCD) survival rate, the appropriate
sensing and pacing rate, detection and conversion of ventricular
tachyarrhythmias, and the appropriate rejection of atrial
tachyarrhythmias.
1.6.2.2 Results
The mean implant duration was 6.3 ± 0.4 months with
cumulative implant duration of 795.5 months. There were 20
patients followed for over twelve months and 62 patients
followed for over six months during the study period from
February 5, 1999 to April 15, 2000. The patient follow-up
compliance rate was 98.4% out of 319 required follow-ups.
Table 3
the predefined endpoints.
:
OTE
The Phylax AV ICD is an earlier generation of BIOTRONIK
devices. The Lexos family is based upon the Phylax AV and
other BIOTRONIK ICDs (i.e., Phylax, Tachos, and Belos
families of ICDs). Therefore, the clinical data from the
Phylax AV was used to support the safety and effectiveness
of the Lexos DR.
provides a summary of the results of the study group for
20 Lexos Technical Manual
Table 3: Clinical Study Results
Description
UADE-free Survival Rate
(patients with at least 3 months
follow-up)
Complication Rate
Sudden Cardiac Death Survival Rate
Appropriate Bradycardia Sensing and
Pacing Rate
Detection and Conversion of
Ventricular Tachyarrhythmias
Appropriate Rejection of
Atrial Tachyarrhythmias
Study Group
[95% CI]
100% (85/85)
[96.5%, 100%]
11.1% (14/126)
[0%, 16.8%]
100% (124/124)
[97.6% 100%]
96.2%(1141/1186
)
[95.2%, 100%]
98.2% (650/662)
[97.1%, 100%]
94% (138/147)
[89.6%, 100%]
1.6.2.3 SMART Detection Algorithm
The SMART Detection algorithm is an integral function of
BIOTRONIK’s dual chamber ICD product line (i.e., Phylax AV,
Tachos DR, and Belos DR) and is designed to discriminate lifethreatening ventricular tachycardias from relatively harmless
atrial tachyarrhythmias. This algorithm uses information about
the signals from the atrial and ventricular lead systems and is
designed to reduce the amount of inappropriate therapy that
might be delivered as a result of a supraventricular tachycardia
(SVT). Neither the SMART Detection algorithm nor the ICDs are
designed to detect or deliver therapy to terminate atrial
arrhythmias, and therefore this is not the purpose of the
algorithm or the device.
Lexos Technical Manual 21
During the Phylax AV clinical study, specific data was collected
to demonstrate the ability of the SMART Detection algorithm to
discriminate between SVT and VT. The Phylax AV
demonstrated the ability to withhold inappropriate therapy in
approximately 94% of the SVT episodes that were reported
during the study. In addition, the SMART Detection algorithm
appropriately delivered therapy in 100% of the ventricular
episodes in which the feature was activated. At routine followups, the algorithm was activated in 80% of patients enrolled into
the study, which further supports the overall ability of the
algorithm to appropriately discriminate between SVT and VT. In
addition, during the clinical study, the investigators indicated that
the primary reason for selecting a dual-chamber ICD was SVT
discrimination for 70% of the patients enrolled.
1.7 Patient Selection and Treatment
1.7.1 Individualization of Treatment
• Determine whether the expected device benefits
outweigh the possibility of early device replacement for
patients whose ventricular tachyarrhythmias require
frequent shocks.
• Determine if the device and programmable options are
appropriate for patients with drug-resistant
supraventricular tachyarrhythmias (SVTs), because
drug-resistant SVTs can initiate unwanted device
therapy.
• Direct any questions regarding individualization of
patient therapy to your BIOTRONIK representative or
BIOTRONIK technical services at 1-800-547-0394.
The prospective patient’s size and activity level should be
evaluated to determine whether a pectoral or abdominal implant
is suitable. It is strongly recommended that candidates for an
ICD have a complete cardiac evaluation including EP testing
prior to device implant to gather electrophysiologic information,
including the rates and classifications of all the patient’s cardiac
rhythms. When gathering this information, delineate all clinically
significant ventricular and atrial arrhythmias, whether they occur
spontaneously or during EP testing.
22 Lexos Technical Manual
If the patient’s condition permits, use exercise stress testing to
do the following:
• Determine the maximum rate of the patient’s normal
rhythm.
• Identify any supraventricular tachyarrhythmias.
• Identify exercise-induced tachyarrhythmias.
The maximum exercise rate or the presence of supraventricular
tachyarrhythmias may influence selection of programmable
parameters. Holter monitoring or other extended ECG
monitoring also may be helpful.
If the patient is being treated with antiarrhythmic or cardiac
drugs, the patient should be on a maintenance drug dose rather
than a loading dose at the time of pulse generator implantation.
If changes to drug therapy are made, repeated arrhythmia
inductions are recommended to verify pulse generator detection
and conversion. The pulse generator also may need to be
reprogrammed.
Changes in a patient’s antiarrhythmic drug or any other
medication that affect the patient’s normal cardiac rate or
conduction can affect the rate of tachyarrhythmias and/or
efficacy of therapy.
If another cardiac surgical procedure is performed prior to
implanting the pulse generator, it may be preferable to implant
the lead system at that time. This may prevent the need for an
additional thoracic operation.
1.7.2 Specific Patient Populations
Pregnancy - If there is a need to image the device, care should
be taken to minimize radiation exposure to the fetus and the
mother.
Nursing Mothers - Although appropriate biocompatibility testing
has been conducted for this implant device, there has been no
quantitative assessment of the presence of leachables in breast
milk.
Geriatric Patients - Most (72%) of the patients receiving an ICD
in the Phylax AV clinical study were over the age of 60 years
(see Clinical Studies).
Lexos Technical Manual 23
Handicapped and Disabled Patients - Special care is needed
in using this device for patients using an electrical wheel chair or
other electrical (external or implanted) devices.
1.8 Patient Counseling Information
The pulse generator is subject to random component failure.
Such failure could cause inappropriate shocks, induction of
arrhythmias or inability to sense arrhythmias, and could lead to
the patient’s death.
Persons administering CPR may experience the presence of
voltage on the patient’s body surface (tingling) when the patient’s
ICD system delivers a shock.
A patient manual is available for the patient, patient’s relatives,
and other interested people. Discuss the information in the
manual with concerned individuals both before and after pulse
generator implantation so they are fully familiar with operation of
the device. (For additional copies of the patient manual, contact
the BIOTRONIK at the address listed in this manual.)
1.9 Evaluating Prospective ICD Patients
The prospective ICD implant candidate should undergo a
cardiac evaluation to classify any and all tachyarrhythmias. In
addition, other patient specific cardiac information will help in
selecting the optimal device settings. This evaluation may
include, but is not limited to:
• an evaluation of the specific tachycardia rate(s)
• the confirmation and/or evaluation of any
supraventricular arrhythmias or bradyarrhythmias
• the evaluation of various ATP and cardioversion
therapies
• the presence of any post-shock arrhythmias, and
• an evaluation of the maximum sinus rate during exercise
If a patient’s drug regimen is changed or adjusted while the ICD
is implanted, additional EP testing may be required to determine
if detection or therapy parameter settings are relevant and
appropriate.
24 Lexos Technical Manual
Empirical changes to the detection or therapy parameters should
be assessed based on patient safety. Some changes may
necessitate a re-assessment of sensing, pacing, or arrhythmia
conversion treatment. Thorough technical knowledge of
BIOTRONIK ICDs, additional ICD experience, and individual
medical judgment will aid in determining the need for additional
testing and follow-up.
Lexos Technical Manual 25
2. Device Features
The Lexos family feature set is presented under the following
sub-headings: Tachyarrhythmia Detection, Tachyarrhythmia
Redetection / Acceleration, Tachyarrhythmia Therapy,
Tachyarrhythmia Termination, Bradycardia Therapy, EP Test
Functions and Special Features. The features apply to all
members of the Lexos family except where specifically
referenced differently.
C
AUTION
Programmed Parameters – Program the device
parameters to appropriate values based on the patient’s
specific arrhythmias and condition.
2.1 Sensing (Automatic Sensitivity
Control)
The Lexos ICDs use Automatic Sensitivity Control (ASC) to
adjust the input stage sensitivity threshold to appropriately detect
the various cardiac signals. The characteristics of the sensing
circuitry have been optimized to ensure appropriate sensing
during all potential cardiac rhythms.
Cardiac signals vary in amplitude; therefore detection thresholds
cannot be static. With the Automatic Sensitivity Control (ASC)
every sensed event is measured, and the upper and lower
thresholds are re-set accordingly (also known as beat-by-beat
adaptation). The ASC begins by tracking the cardiac signals (R
and P-waves) during the sensed refractory periods. The peak
values measured during this time are used to set the sensing
thresholds during the active detection periods.
2.1.1 Ventricular Sensitivity Settings
There are three programmable preset options for setting the
sensitivity of the input stage. The sensitivity selections are
designed to adapt the parameters of the input stage to various
signal conditions. The predefined parameter sets are described
in
Table 4
.
26 Lexos Technical Manual
Table 4: Sensitivity Settings
Setting Definition for Use
Standard This setting is recommended for most
patients, especially for those with
measured R-wave amplitude of ≥3 mV.
Enhanced
T Wave
Suppression
This setting offers suppression of T-wave
oversensing. This mode should not to be
used on patients with the following
conditions:
• Sinus rhythms with small signal
amplitudes, R-waves <4 mV
• VF with highly fluctuating signal
amplitudes.
Enhanced
VF Sensitivity
This setting enhances VF detection, in
cases of highly fluctuating signal
amplitudes. It is not to be used for patients
that have sinus rhythms containing large
amplitude T-waves.
Free This parameter configuration is only
accessed by code and is not available in
the US.
Typically, the upper threshold is reset with each sensed R-wave,
but in order to ensure that pacing does not occur during an
episode of VF, the ASC behaves differently with paced events.
Each paced event is followed by a paced refractory period after
which the ventricular threshold is set to the minimum
programmed value.
The upper threshold is set at 50% of the measured R-wave for
the Standard sensitivity setting following the 100 ms sensed
refractory period. The upper threshold decays 0.125 mV every
250 ms through the T-wave discrimination period (350 ms).
After the T-wave discrimination period, the threshold is
decreased to the lower threshold. The lower threshold is set to
25% of the measured peak R-wave. The lower threshold then
decreases 0.125 mV every 500 ms until the Minimum Threshold
is reach or until the next sensed (or paced) event.
Lexos Technical Manual 27
Figure 1 Automatic Sensitivity Control with Standard Setting
Figure 1 provides an illustration of Automatic Sensitivity Control
with the sensitivity programmed to Standard. The tracked R –
wave is measured to be 6.0 mV, following the sensed refractory
period, the upper threshold is set to 3.0 mV. After the T-wave
discrimination period, the threshold is further reduced to 1.5 mV.
Both the Upper and Lower Thresholds decay over time, but the
Minimum Threshold is never violated. Nominally, the minimum
threshold is set to 0.8 mV, but it can be adjusted by the user.
The Enhanced VF Sensitivity setting is specifically designed to
improve VF detection when the VF signal is very small. Two
adjustments are made to ASC with this setting;
• The T-wave discrimination period is decreased to
100 ms, thus eliminating the Upper Threshold
• The decay rate of the Lower Threshold is increased to
0.125 mV every 250 ms.
These adjustments ensure that the threshold reaches the lower
values more quickly in order to assure that all VF signals are
sensed appropriately.
28 Lexos Technical Manual
The Enhanced T-Wave Suppression setting is specifically
designed to avoid double counting of each QRS-T complexes
during normal sinus rhythms. With sensitivity programmed to
Enhanced T-Wave Suppression;
• High pass filtering is increased to reduce low frequency
signal components such as T-waves and respiratory
artifacts.
• The Upper Threshold is increased to 75% of the
measured R-wave.
• The Upper Threshold may not retrigger with each
sensed event, it is only triggered when the new sensed
R-wave crosses the 50% point of the previous measured
R-wave.
2.1.2 Minimum Ventricular Threshold
This parameter limits the minimum sensitivity of the ICD to a
programmable value. Nominally, the minimum threshold is set to
0.8 mV, but it can be adjusted from 0.5 to 2.5 mV.
2.1.3 Atrial Sensitivity Settings
There is only one option for setting the sensitivity of the atrial
input stage. When atrial sensing is active, the sensitivity is set to
“Standard” which is designed to adapt the parameters of the
input stage to various signal conditions. The available settings
are described in Table 5.
Table 5: Atrial Sensitivity Settings
Setting Definition for Use
Standard This setting is recommended for all
patients with a functioning atrial lead.
Inactive This setting deactivates the atrial channel
for sensing, EGM telemetry and Holter
recording and is typically used when no
atrial lead is implanted.
Free This parameter configuration is only
accessed by code and is not available in
the US.
Lexos Technical Manual 29
Typically, the upper threshold is reset with each sensed P-wave,
but in order to ensure that pacing does not occur during an
episode of AF/VF, the ASC behaves differently with paced
events. Each paced event is followed by a paced refractory
period after which the atrial threshold is set to the minimum
programmed value.
2.1.4 Minimum Atrial Threshold
This parameter limits the minimum sensitivity of the ICD to a
programmable value. Nominally, the minimum threshold is set to
0.4 mV, but it can be adjusted from 0.2 to 2.5 mV.
2.1.5 Far Field Blanking
This parameter blanks the atrial channel of the ICD to the period
before and after each ventricular sensed event. The first number
is the pre-event blanking period and the second number is the
post-event blanking period.
C
AUTION
Far-field sensing of signals from the atrium in the
ventricular channel or ventricular signals in the atrial channel
should be avoided by appropriate lead placement,
programming of pacing/sensing parameters, and maximum
sensitivity settings. If it is necessary to modify the Far Field
Blanking parameter, the parameter should be lengthened
only long enough to eliminate far-field sensing as evidenced
on the IEGMs. Extending the parameter unnecessarily may
cause undersensing of actual atrial or ventricular events.
30 Lexos Technical Manual
2.2 Ventricular Tachyarrhythmia Detection
The Lexos ICDs detect and measure the rate of sensed cardiac
signals to discriminate ventricular tachyarrhythmias from
supraventricular tachycardias, sinus rhythm or sinus
bradycardia. This is accomplished through programmable rate
detection parameters in the device. When a tachyarrhythmia is
present, the ICD classifies the arrhythmia and delivers the
appropriate therapy. If a tachyarrhythmia continues following the
first therapy attempt, then the ICD will redetect the
tachyarrhythmia and deliver subsequent therapies as necessary.
W
ARNING
Unwanted Shocks – Always program the VT/VF Detection
and Therapy status to DISABLED prior to handling the
device to prevent the delivery of serious shocks to the
patient or the person handling the device during the implant
procedure.
Classification of cardiac signals is accomplished primarily by
measuring the cardiac cycle length (R-R, P-R and P-P). In
addition, the ICD can also utilize abrupt changes in rate or
irregularity of the cardiac signal to further differentiate ventricular
tachyarrhythmias. Each detected ventricular tachyarrhythmia is
classified into one of the following zones:
• VT-1 Lower rate ventricular tachycardia
• VT-2 Higher rate ventricular tachycardia
• VF Ventricular fibrillation
Each rhythm class is programmable to a separate rate with the
zone limit defining the lowest rate in each class. The upper rate
limit of each class is equal to the zone limit of the next higher
class, creating a continuous range of rate classes.
Lexos Technical Manual 31
2.2.1 VF Classifications
Detection of ventricular fibrillation (VF) utilizes programmable X
out of Y criterion. Both X and Y are programmable. If X number
of intervals within the sliding window (defined by Y) are shorter
than the programmed VF rate interval (>bpm), VF is detected.
After fibrillation is detected, the programmed therapy sequence
for VF is initiated.
Nominal settings for classification of ventricular fibrillation (VF)
are 8 of 12 intervals; meaning that within a sample window of 12
intervals, 8 intervals must meet or exceed the VF zone rate
criteria.
2.2.2 VT Interval Counters
The VT Interval Counters are separately programmable for VT-1
and VT-2 rate classifications. The Detection Counter is the
number of intervals required to declare a tachyarrhythmia as VT.
A tachyarrhythmia must meet both the rate/interval criteria and
the programmed Detection Counter criteria, in addition to any
other detection enhancements to be declared a tachycardia.
2.2.3 VT Classification
Both VT-1 and VT-2 classification zones utilize separately
programmable detection parameters. Classification of VT-1 or
VT-2 is based on the last interval average preceding declaration
of tachyarrhythmia detection. If this average falls within the
VT-1, the programmed VT-1 therapy is delivered. If the average
falls within the VT-2 limits, the programmed VT-2 therapy is
delivered. If additional detection parameters are activated, each
of these supplemental criteria must also be satisfied before a VT
rhythm can be classified.
32 Lexos Technical Manual
The ICDs may be programmed to use ventricular-only
information, or both atrial and ventricular information for the
discrimination of ventricular tachycardias. With SMART
Detection™ turned ON, the Lexos ICDs use atrial and ventricular
signals for discrimination of fast heart rhythms. With SMART
Detection™ turned OFF, only the ventricular rate is used to
discriminate between ventricular rhythm classes. If SMART
Detection is enabled, this algorithm evaluates all cardiac
signals within the VT range and increments the VT Sample
Count for all intervals that are deemed VT. A full description of
SMART Detection™ is provided in the following text.
In addition, when the termination criteria are met, all
tachyarrhythmia detection criteria, including the VT sample
counters are reset. (See section 2.4).
2.2.4 SMART Detection™
This discrimination algorithm enhances VT-1 and VT-2 detection
by applying a series of tests to the sensed cardiac signal.
SMART Detection™ is intended to discriminate VT from a variety
of supraventricular arrhythmias that are conducted to the
ventricle and that would otherwise satisfy VT-1 or VT-2 rate
detection criteria.
First, the average ventricular rate is compared to the average
atrial rate. In the event that the measured ventricular rate is
faster than the atrial rate, the device immediately declares the
rhythm a VT and delivers programmed ventricular therapy for the
detected VT zone.
In the event that an atrial rate is faster compared to the
ventricular rate one of three tests are performed:
Ventricular rhythm stability, (see Stability on page 33) if the
ventricular signal is unstable, then the rhythm is declared a
supraventricular tachyarrhythmia, (SVT) and ventricular therapy
is typically withheld.
If the ventricular signal is stable, and the atrial rate is a multiple
of the ventricle rate, then the rhythm is declared a
supraventricular tachyarrhythmia (SVT) and ventricular therapy
is typically withheld.
Lexos Technical Manual 33
If the ventricular rhythm is stable and the atrial rate is not a
multiple of the ventricular rate, then the rhythm is declared a VT
and ventricular tachycardia therapy is delivered.
In the event that both the atrial and ventricular signals are
detected at the same rate, a series of additional discrimination
tests are applied.
2.2.5 Onset
Another detection enhancement is ONSET, which may be used
Independently in the VT-1, VT-2, and with SMART Detection™
(deactivated),. In the VT-1 and VT-2 zones, the purpose of
ONSET is to measures abrupt changes in ventricular cycle
length to discriminate between sinus tachycardias and
ventricular and atrial tachyarrhythmias, which characteristically
begin with an abrupt change in cardiac rate.
This feature allows therapy to be appropriately withheld if a sinus
tachycardia rate crosses into one of the VT zones.
The SMART Detection™ algorithm utilizes ONSET as an integral
part of the discrimination algorithm, therefore when
SMART Detection™ is enabled the ONSET parameter must also
remain enabled and set to 20 %.
2.2.6 Stability
In VT-1 and VT-2 zones, the purpose of STABILITY is to assist
in discriminating between stable ventricular tachyarrhythmias
and supraventricular tachyarrhythmias that conduct irregularly to
the ventricles. STABILITY evaluates sudden changes in the
regularity of cardiac events (R-R and P-P intervals) on a beat by
beat basis. The STABILITY criterion compares the current
measured interval with the three preceding cardiac intervals. If a
difference between the current interval and each of the three
preceding intervals is less than the stability range, then the
current intervals are stable.
The SMART Detection™ algorithm utilizes both atrial and
ventricular STABILITY as integral parts of the discrimination
algorithm. Therefore, when SMART Detection™ is enabled, the
STABILITY parameter must also remain enabled and set to
12%.
34 Lexos Technical Manual
2.2.7 Sustained VT Timer
The Sustained VT Timer can be programmed between 30
seconds and 30 minutes (or to OFF). When the timer expires,
therapy is initiated regardless of the detection enhancements.
The Sustained VT parameter is intended to force tachycardia
therapy in cases where a cardiac rhythm meets the VT rate
criteria but does not satisfy one or more detection enhancement
criterion (Onset, SMART Detection, or Stability) for an extended
duration. A “safety” timer is initiated within one of the VT zones.
If the programmed Sustained VT time period expires without
tachycardia detection, redetection is initiated without utilizing the
detection enhancements.
A simple up/down counter is used to initiate the safety timer.
The counter is incremented by one when an interval falls into the
VT zone, and decrements by one when an interval falls into the
sinus zone. When the counter reaches a number equal to the
programmed VT detection counter, the safety timer starts. The
timer runs until the programmed time expires and therapy is
delivered or until the timer is reset. The timer is reset with initial
detection or VT termination.
The safety timer is not used in redetection. If initial detection
was due to the safety timeout and SMART Redetection is
programmed “ON”, then SMART Detection will not be used for
redetection.
2.3 Tachyarrhythmia Redetection
The Lexos ICDs offer independently programmable settings for
determining if tachyarrhythmias remain after therapy has been
delivered. The redetection routine allows the ICDs to determine
whether further therapy is required when the initial therapy was
unsuccessful in terminating the arrhythmia.
Tachyarrhythmia redetection criteria are based on cardiac cycle
length and number of intervals. The number of intervals is
distinct and independent of the initial detection criteria.
Lexos Technical Manual 35
2.3.1 VT Redetection
The Redetection Counter parameter may be programmed
separately for each arrhythmia class, independent of the initial
detection parameters:
Redetection of an ongoing tachyarrhythmia is declared when the
Redetection Counter is satisfied (based on individual cycles). If
a sensed cardiac signal meets any VT rate criteria, following
therapy, that signal is counted and compared to the programmed
Redetection Counter setting. Tachycardia redetection is
declared when the programmed number of VT samples
(Redetection Counter) is satisfied.
Redetection functions identically to initial VT detection in regards
to the Stability and Onset detection enhancements and it is
based on individual cycle lengths (not averages).
2.3.2 SMART Redetection
With SMART Redetection programmed ON, both atrial and
ventricular signals are used for redetection after initial detection
and therapy for a VT. SMART Detection™ will function
identically as in initial VT detection.
2.3.3 VF Redetection
VF redetection uses the same X out of Y criterion as initial
detection. The X and Y values for initial detection are also used
for redetection to ensure consistent classification of VF.
2.4 Tachyarrhythmia Termination
Termination of a ventricular tachyarrhythmia episode is declared
when 12 out of 16 consecutive sensed intervals are longer than
the VT-interval parameter of the lowest VT class (sinus or
bradycardia rhythm).
36 Lexos Technical Manual
2.5 Tachyarrhythmia Therapy
The Lexos ICDs offer a variety of therapy options that can be
tailored to meet a patient’s specific anti-tachycardia or
defibrillation therapy requirements. Anti-tachycardia pacing
(ATP) therapies can be combined with defibrillation therapies to
provide a broad spectrum of tachyarrhythmia treatment options.
2.5.1 Therapy Options
The Lexos ICDs offer multiple therapy options for each
tachyarrhythmia class (VT1, VT2, VF). Therapy options (up to
20 ATPs and 8 shocks) are available for the VT1 and VT2
zones, whereas up to 8 shock therapies are available for the VF
zone. The specific characteristics of an ATP and shock therapy
are independently programmed for each VT zone.
The ATP and shock therapy options are discussed in detail in
the following sections.
2.5.2 Anti-Tachycardia Pacing (ATP)
Anti-tachycardia pacing (ATP) therapy is available in both VT
detection zones. Available modes of ATP include Burst, Ramp,
and Burst + PES (Programmed Extra Stimuli). In addition, the
Burst and Ramp modes allow interval scanning of the R-S1
interval, the S1-S1 interval, or both. The Attempts parameter
determines the number of burst schemes to be delivered before
the scan parameter is incremented.
Burst – This mode will deliver a series of pacing stimuli with
user defined duration of the burst (number of S1), coupling cycle
length (R-S1) and burst rate (S1-S1). The coupling interval and
the start interval are calculated from the intrinsic R-R average.
Ramp - This mode will deliver a series of pacing stimuli with the
above options including a parameter which decrements each
successive stimuli interval in the burst.
Burst + PES - This mode provides a pulse train followed by one
or more (up to three) additional timed stimuli. The coupling cycle
length of the burst and each extra stimulus is individually
programmed either as an adaptive value (as a percentage) or as
an absolute value (expressed in milliseconds).
Lexos Technical Manual 37
Add S1 - This feature can be programmed with any Burst,
Ramp, or Burst + PES scheme. When “Add S1” is “ON,” the
number of S1 intervals is incremented by one on each
successive ATP therapy. The new S1-S1 interval is dependent
on the initial start interval (S1 decrement) and the programmed
scan decrement (if activated).
S1 Count - The S1 count parameter defines the number of
stimuli of an ATP. For Burst + PES, an extra stimulus with a
separate parameter is coupled.
R-S1 - The R-S1 programmable coupling interval occurs at the
beginning of each ATP. It defines the interval between the last
R-wave signal and the first stimulus (S1). The second stimulus
always follows the first one with the same interval.
S1 Decrement - The S1 decrement continuously reduces the
pulse intervals of the ATP from the second pulse onwards.
Minimum ATP Interval - The programmed minimum interval
prevents ATPs from being given with stimulation values less
than the minimum interval. When the ATP interval reaches the
value of the minimum interval with the S1 decrement or scan
decrement, it then assumes this value and remains constant.
2.5.2.1 ATP Help
ATP help is a useful tool to assist the physician in choosing and
confirming appropriate ATP programming. The “ATP Help”
button is displayed in each ATP therapy option. When the ATP
help button is pressed, a histogram of the chosen therapy
scheme is shown. The histogram displays the intervals for the
programmed ATP scheme. When rate adaptive intervals are
programmed, the displayed intervals are based on the
programmed R-R average.
38 Lexos Technical Manual
2.5.2.2 ATP Therapy Optimization
In order to optimize future therapies, the ICD will store the
parameter configuration of the last successful ATP attempt in
each the VT1 and VT2 classes. The last successful stored ATP
attempt is then used as the starting point for the next detected
episode of the same arrhythmia class. If the stored parameter
configuration is not successful, it is deleted from the ATP
optimization memory of the respective arrhythmia class and
subsequent therapy sequences will begin with ATP1 for the next
detected episode.
ATP Optimization is programmable ON or OFF for all ATP
therapies and VT zones with one parameter.
N
2.5.2.3 ATP Timeout
ATP Timeout is a timer that decrements after the initial
ventricular ATP is delivered (VT-1 zone) and limits the additional
ATP therapies that may be delivered. Once the timer expires, all
further ATP therapies in the sequence are blocked. If further
therapy is required after the timer has expired, the system
advances to the first programmed shock therapy for the
applicable VT zone. Therapy continues until arrhythmia
termination or all programmed therapy (in the applicable zone)
has been delivered. The ATP Timeout is reset each time an
arrhythmia is terminated.
:
OTE
In VT zones, the ICD stores successful ATP therapies only.
The stored information includes not only the number of the
ATP therapy (e.g., ATP 2), but also the successful
configuration in detail (for example: Burst; R-S1 Interval: 320
ms, S1-S1 Interval: 320 ms; etc.).
Lexos Technical Manual 39
2.5.3 Shock Therapy
Shock Therapy can be delivered as defibrillation shocks with or
without confirmation (while the capacitors are being charged).
The first shock energy in each shock module has independently
programmable Shock Energy, and confirmation programmable
ON/OFF. The remaining shock therapies are nonprogrammable and predetermined to deliver 30 joules using
defibrillation without confirmation. All shock therapies may be
programmed to normal or reversed polarity.
2.5.3.1 Number of Shocks
The number of shocks defines the total number of shock
attempts per therapy zone (VT-1, VT-2 or VF). Up to 8 shocks
are available in each therapy zone. The first shock energy and
confirmation parameters are independently programmable, while
the remaining shocks are fixed at 30 joules with Confirmation
turned OFF.
2.5.3.2 Confirmation
The Confirmation parameter is used to verify the presence of a
tachyarrhythmia during the charging of the capacitors. This
function is designed to avoid delivery of inappropriate therapy if
a tachyarrhythmia has spontaneously terminated. The
programmed shock will be delivered unless bradycardia or a
normal sinus rhythm is detected during the Confirmation period.
Confirmation may be programmed ON or OFF for the first shock
therapy and is always OFF for remaining shock therapies.
ONFIRMATION
C
OFF
When Confirmation is programmed OFF, shock therapy will be
delivered to the patient during the synchronization period
regardless of the detected cardiac signal.
C
ONFIRMATION
ON
If the tachyarrhythmia spontaneously converts to bradycardia or
a normal sinus rhythm during the confirmation period, shock
therapy is aborted. However if the device confirms the presence
of the tachyarrhythmia, the device will deliver the programmed
shock therapy.
40 Lexos Technical Manual
C
AUTION
Shock Therapy Confirmation – Programming
CONFIRMATION to OFF may increase the incidence of the
ICD delivering inappropriate shocks.
S
YNCHRONIZATION
A synchronization window is started at the end of the charging
period. During this window, the device will attempt to
synchronize the shock therapy to an R-wave. If no R-wave is
detected, the shock will be delivered asynchronously at the end
of the synchronization period.
2.5.3.3 Shock Waveform
Two waveforms of shock therapy are available with the Lexos
ICDs, Biphasic and Biphasic 2ms. The following diagram
describes each of the shock waveforms.
Figure 2. Biphasic Waveforms
The waveform starts at the calculated voltage, based on the
programmed energy level. After an exponential discharge
through the lead system to 40% of the initial charge voltage, both
shock waveforms switch polarity. At the second phase the:
• Biphasic waveform discharges to 20% of the initial charge
voltage before the waveform is truncated.
• Biphasic 2ms waveform discharges the remaining energy
for two milliseconds.
Figure 2
waveforms.
provides a pictorial representation of both biphasic
Lexos Technical Manual 41
BIOTRONIK recommends use of the standard Biphasic shock
waveform for initial defibrillation threshold testing. If testing
demonstrates high defibrillation thresholds, testing with the
Biphasic 2ms waveform is offered as a therapeutic alternative to
the standard Biphasic shock.
2.5.3.4 Shock Energy
The Lexos ICDs are designed to ensure that the energy
programmed for therapy is the same as what is actually
delivered to the patient regardless of the lead impedance. The
first shock energy in each therapy class is programmable
between 5 and 30 joules. The remaining shock energies will be
delivered at 30 joules.
C
AUTION
Shock Impedance - If the shock impedance is less than
twenty-five ohms, reposition the lead system to allow a
greater distance between the electrodes. Never implant the
device with a lead system that has measured shock
impedance as less than twenty-five ohms. Damage to the
device may result.
Defibrillation Threshold - Be aware that changes in the
patient’s condition, drug regimen, and other factors may
change the defibrillation threshold (DFT), which may result in
non-conversion of the arrhythmia post-operatively.
Successful conversion of ventricular fibrillation or ventricular
tachycardia during arrhythmia conversion testing is no
assurance that conversion will occur post-operatively.
Shock Therapy Confirmation – Programming
CONFIRMATION to OFF may increase the incidence of the
ICD delivering inappropriate shocks.
42 Lexos Technical Manual
2.5.3.5 Shock Polarity
The polarity of the shock therapy may be programmed and
changed non-invasively. The Normal polarity configures the
HV 1 connector port as the negative electrode and the HV 2
connector port and the outer housing of the ICD as the positive
electrode for the first phase of the shock. Reversed polarity will
switch the electrical polarity of the connector ports and housing.
The shock polarity is separately programmable for each
arrhythmia zone.
2.5.4 Progressive Course of Therapy
By design, the Lexos ICDs will deliver more aggressive therapy
for each successive attempt within a single detected episode.
Therefore, the device will not deliver ATP1 therapy following
ATP2 therapy, and will not deliver ATP therapy following a high
voltage defibrillation shock.
When Progressive Course of Therapy is turned ON, the ICD will
always deliver a maximum energy shock upon re-detecting in an
arrhythmia class with a programmed therapy with energy less
than or equal to the previously delivered therapy. In addition, the
ICD blocks all ATP therapy during the current episode if a shock
has already been delivered during the episode.
Furthermore, the ICD prevents therapies of different arrhythmia
classes from permanently retarding or accelerating a VT in such
a way that the cardiac rhythm fluctuates between the different
arrhythmia classes without achieving termination of the
arrhythmia regardless of the Progressive Course of Therapy
setting.
For example, a 10-joule defibrillation shock is delivered for an
arrhythmia detected in the VT-2 zone and results in a
deceleration of the VT so that it is subsequently redetected in the
VT-1 zone. At that point, the Lexos ICDs would continue with
shock therapy, but all shocks programmed at less than 10 joules
would be delivered at an energy of 10 joules.
If a defibrillation shock is delivered but does not terminate the
arrhythmia, the next shock will always have the same or higher
energy than the last delivered shock. Beginning with the third
shock, the shocks are delivered at maximum energy (30 joules).
Lexos Technical Manual 43
2.6 Bradycardia Therapy
The Lexos ICDs have independently programmable dual
chamber bradycardia and post-shock bradycardia pacing
functions. The post-shock bradycardia parameters may be
programmed to higher rates or output values for the period
following a delivered shock, without compromising the longevity
of the ICD for patients who require chronic bradycardia pacing.
The post-shock programmable values are presented in a
separate subsection from the normal bradycardia support
values.
2.6.1 Bradycardia Pacing Modes
The available bradycardia pacing modes for each member of the
Lexos ICD family are listed in
Table 6
Table 6 Lexos Pacing Modes
Mode Lexos DR and DR-T Lexos VR and VR-T
DDDR X N/A
DDIR X N/A
VDDR X N/A
VDIR X N/A
AAIR X N/A
VVIR X X
DDD X N/A
DDI X N/A
VDD X N/A
VDI X N/A
AAI X N/A
VVI X X
OFF X X
The basic rate timer is started by a sensed or paced event. A
sensed event outside of the refractory period inhibits pacing and
resets the lower rate time; in the absence of a sensed event, a
pacing pulse will be delivered at the end of the lower rate
interval.
.
44 Lexos Technical Manual
The pacing modes with an “R” indicate rate adaptive pacing
controlled by a motion based capacitive sensor. These modes
are functionally the same as the corresponding non-rateadaptive modes, except that the pacing rate is increased based
on physical activity.
2.6.2 Basic Rate
The basic rate is the pacing rate in the absence of a patient’s
intrinsic rhythm. This rate may be independently programmed
for normal and post-shock bradycardia pacing.
2.6.3 Night Rate
The Night Rate is the effective basic rate during the programmed
“sleep” period for the patient. This parameter provides a lower
pacing rate during the patient’s normal sleep time in an attempt
to match the decreased metabolic needs during sleep. When
Night Rate is active, the basic rate automatically decreases to
the programmed pacing Night Rate during the nighttime hours.
At the programmed start time (Begin of Night), the rate gradually
decreases to the night rate. When the internal clock reaches the
programmed end time (End of Night), the pacing rate gradually
changes to the programmed basic rate. The rate changes at the
same rate as the Sensor Gain decrease and increase
parameters.
N
:
OTE
The Night Rate time is based on the programmer clock.
Therefore, the programmer time should be checked prior to
device programming. If a patient travels across different
time zones, the Night Rate time may require adjustment.
2.6.4 Rate Hysteresis
The ability to decrease the effective lower rate through
Hysteresis is intended to preserve a spontaneous rhythm. The
pulse generator operates by waiting for a sensed event throughout
the effective lower rate interval (Hysteresis interval). If no
sensed event occurs, a pacing pulse is emitted following the
Hysteresis interval.
Lexos Technical Manual 45
Hysteresis can be programmed OFF or to values as low as
-65 bpm of the basic rate. Hysteresis is initiated by a sensed
event. The resulting Hysteresis rate is always less than the
lower rate. The Hysteresis rate is limited to only providing a
basic rate that is 30 bpm or greater. Therefore, if the basic rate is
80 bpm and the hysteresis value is programmed to –65 bpm, the
ICD paces at a rate of 30 bpm because this is the minimum at
which it can function.
N
OTES
2.6.4.1 Repetitive Hysteresis
Repetitive hysteresis is expanded programmability of the
Hysteresis feature. Repetitive hysteresis searches for an
underlying intrinsic cardiac rhythm, which may exist slightly
below the programmed lower rate (or sensor-indicated rate).
Following 180 consecutive sensed events, this feature allows the
intrinsic rhythm to drop to or below the hysteresis rate. During
the time when the intrinsic rate is at or below the hysteresis rate,
pacing occurs at the hysteresis rate for the programmed number
of beats (up to 10). Should the number of programmed beats be
exceeded, the stimulation rate returns to the lower rate (or
sensor-indicated rate).
:
If rate adaptation is active, the Hysteresis rate is based on
the current sensor-indicated rate and the value of the
programmable parameter.
If Hysteresis is used in the DDI mode, the AV delay must be
programmed shorter than the spontaneous AV conduction
time. Otherwise, stimulation in the absence of spontaneous
activity occurs at the hysteresis rate instead of the lower
rate.
Hysteresis is suspended during the programmed pacing
Night Rate. Programming conflicts arise when the total
decrease in rate is below 30 ppm. Care should be exercised
to avoid programming a Night Rate and hysteresis that is
below what is appropriate and may be tolerated by the
individual patient.
If an intrinsic cardiac rhythm is detected within the programmed
number of beats between the hysteresis rate and the lower rate,
the intrinsic rhythm is allowed and inhibits the pulse generator.
46 Lexos Technical Manual
Figure 3. Repetitive Hysteresis
Repetitive hysteresis has been incorporated to promote
spontaneous cardiac rhythm and may reduce pulse generator
energy consumption.
N
:
OTE
Repetitive and Scan Hysteresis are not active during the
programmed Night Rate and are only available when
Hysteresis is selected on.
Magnet application (closing of reed switch) suspends 180
consecutive event counter independent of synchronous or
asynchronous magnet effect.
There is one Standard Hysteresis interval which occurs
before the programmable number of Repetitive Hysteresis.
Lexos Technical Manual 47
2.6.4.2 Scan Hysteresis
Scan hysteresis is expanded programmability of the Hysteresis
feature. Scan hysteresis searches for an underlying intrinsic
cardiac rhythm, which may exist slightly below the programmed
lower rate (or sensor-indicated rate) of the pulse generator.
Following 180 consecutive paced events, the stimulation rate is
temporarily decreased to the hysteresis rate for a programmed
number of beats. If a cardiac rhythm is not detected within the
programmed number of beats at the hysteresis rate, the
stimulation rate returns back to the original lower rate (or sensorindicated rate). Several programmable beat intervals are
available to allow a greater probability of detecting a
spontaneous rhythm.
If an intrinsic cardiac rhythm is detected within the programmed
number of beats between the hysteresis rate and the lower rate,
the intrinsic rhythm is allowed and the pulse generator inhibits.
Figure 4. Scan Hysteresis
Scan hysteresis has been incorporated to promote intrinsic
cardiac rhythm and may reduce pulse generator energy
consumption.
48 Lexos Technical Manual
N
:
OTE
Magnet application (closing of reed switch) suspends 180
consecutive event counter independent of the magnet effect.
2.6.5 Dynamic AV Delay
The AV Delay defines the interval between an atrial paced or
sensed event and the ventricular pacing pulse. If the pulse
generator is programmed to a dual chamber sensing mode, an
intrinsic ventricular event falling within the AV Delay will inhibit
the ventricular pacing pulse. If not contraindicated, a longer
AV Delay can be selected to preserve intrinsic AV conduction.
Dynamic AV Delay is where the AV Delay is varied depending
on the spontaneous atrial rate. Dynamic AV Delay provides
independent selection of AV Delays from five rate ranges at
preset AV Delay values. In addition, the AV Delay after atrial
pace events can be differentiated from the atrial sense events for
dual chamber pacing modes.
In addition to selecting the preset values (Low, Medium, and High) with the Dynamic AV Delay window, the Dynamic
AV Delays may be programmed individually (Individual) for
each rate zone or to a fixed AV Delay (Dynamic).
The AV Delay feature includes an AV Delay shortening option
(Sense Compensation) for dual chamber pacing modes. When
enabled, the AV Delay is shortened by the programmed value
(20 to 120 ms) from the programmed AV Delay after an intrinsic
atrial sensed event.
The Dynamic AV Delay is intended to mimic physiologicshortening of the AV Delay with increasing heart rate. It also
serves for automatic prevention and termination of “circus
movement” pacemaker mediated tachycardia and for prevention
of reentrant supraventricular tachycardias. Dynamic AV Delay is
only available with the Lexos DR and DR-T ICDs.
Lexos Technical Manual 49
2.6.6 Upper Tracking Rate
In the atrial tracking modes (DDDR, DDIR, VDDR, VDIR, DDD,
DDI, VDD and VDI), ventricular pacing tracks atrial pace/sense
events. The maximum tracking rate (ventricular pacing rate) is
limited by the Upper Rate parameter.
The UTR response will automatically toggle between 2:1 and
WKB (Wenkebach) depending on the relative programmed
values for upper rate and atrial refractory period.
If the UTR is less than the maximum sensed atrial rate, defined
by the atrial refractory period (60,000/ARP), the WKB response
is utilized. Atrial rates exceeding the selected upper rate will
result in a Wenckebach-type pacing pattern. This is
accomplished by progressively lengthening the AV delay to keep
the ventricular pacing rate at the upper rate. Lengthening of the
AV interval is interrupted as soon as: 1) a P-wave falls within the
atrial blanking period and is not detected; or 2) a succeeding
P-wave is detected before the end of the AV delay previously
started. In the second case, the corresponding ventricular
pacing pulse is suppressed. If the atrial rate is just above the
upper rate, a low degree (i.e. 6:5) block results. Higher atrial
rates result in higher degrees of AV block until the intrinsic atrial
cycle length violates the programmed atrial refractory period
causing a 2:1 or greater block.
The 2:1 response is utilized when the rate defined by the atrial
refractory period is less than the upper rate and Automatic Mode
Conversion is OFF. In such a case, the maximum pacing rate is
regulated by the inability to respond to P-waves falling within the
atrial refractory period.
If the resulting length of the spontaneous atrial cycle is shorter
than the atrial refractory period in a rate-adaptive mode, the
resulting pacing rate will depend on whether the 2:1 rate has
been exceeded. If this is the case, the pulse generator will use
the sensor rate as the pacing rate. If the 2:1 rate is not
exceeded, the pulse generator will use a rate that lies between
the sensor rate and the rate determined by the atrial refractory
period.
50 Lexos Technical Manual
N
:
OTE
The Lexos DR and Lexos DR-T ICDs the UTR is
programmable within the VT-1 zone. This feature is for
patients that are active and have exercise and VT rates that
overlap. This may be desirable in young active patients.
2.6.7 Mode Switching
Mode switching is designed to avoid tracking of atrial
arrhythmias. In the presence of a high atrial rate, the
bradycardia pacing mode is automatically reprogrammed to a
non-atrial tracking mode. The modes available during mode
switching are as shown in Table 7. Mode switching is not
available during the post-shock pacing period. Mode Switching
is only available with the Lexos DR and DR-T ICDs.
Table 7: Mode Switching Modes
Programmed Mode Converted Mode
DDDR
DDD
VDDR
DDIR
DDI
DDIR
DDI
VDIR
VDI
VDD
Mode switching is initiated in atrial tracking modes when the
atrial rate, defined by the programmable mode switch
Intervention Rate is achieved. However, mode switching will
not occur until the Mode Switch Criteria for Activation is also
met. The criteria for activation is a programmable X
After switching to a non-atrial tracking mode, the ICD activates a
Y out of 8 counter that deactivates mode switching when Y
number of cardiac cycles out of the last 8 are below the
Intervention Rate. When this Criteria for Deactivation
parameter is fulfilled, the ICD returns to the normal programmed
pacing mode.
VDIR
VDI
Lexos Technical Manual 51
2.6.8 PMT Management
The following features provide prevention, detection, and
termination of pacemaker-mediated tachycardias (PMT):
• PMT prevention by extending PVARP
• PMT protection (PMT detection and termination)
PMT Prevention - To prevent PMT, the pacemaker restarts the
basic rate and PVARP when there is ventricular sensing but no
atrial event preceding it. If PVARP extension has been
programmed, the PVARP is additionally prolonged after a VES.
A retrograde P wave with a VA conduction time that is shorter
than the PVARP is not able to trigger a ventricular pace, and
thus no PMT.
PMT Termination - Pacemaker-mediated tachycardias can also
be caused by artifacts and atrial extrasystoles. In such cases,
the PMT protection algorithm features provide both detection
and termination of PMTs. In this way, the more hemodynamically
favorable AV synchronization can be quickly re-established.
PMT may be suspected if 16 consecutive cycles of atrial sensing
and ventricular pacing occur and these are within the range of
the PVARP and the PVARP + PVARP extension and are stable.
PMT Protection can be programmed ON or OFF. PMT
Termination features are only available with the Lexos DR and
DR-T ICDs.
2.6.9 Rate Adaptive Pacing
W
ARNING
Rate-Adaptive Pacing – Use rate-adaptive pacing with care
in patients unable to tolerate increased pacing rates.
52 Lexos Technical Manual
Lexos ICDs allow the selection of rate-responsive pacing modes.
These modes allow the ICD’s bradycardia therapy function to
adapt the pacing rate to increasing or decreasing patient
physical activity, based on data collected from a motion based
sensor within the ICD. Separately programmable criteria allow
the clinician to control the rate of increase and decrease of
pacing, as well as the sensitivity of the sensors in response to
motion.
2.6.9.1 Sensor Gain and Threshold
The Sensor Gain defines how much the sensor signal is
amplified before it is transformed to a rate change. When the
Sensor Gain is low (e.g., 2), a great deal of exertion is needed to
cause a significant change in sensor output (and an equal
change in the pacing rate). When the Sensor Gain is high
(e.g., 18), little exertion is needed to increase the sensor output.
Ideally, the gain is programmed so the maximum desired pacing
rate during exercise occurs at a maximum exertion level.
The device ignores all activity that occurs below the Sensor
Threshold because the Sensor Threshold defines the lowest
sensor output that initiates a change in the pacing rate. Five
different threshold settings are available including; VERY LOW,
LOW, MEAN, HIGH, and VERY HIGH. When the threshold is
programmed optimally, the basic rate is the effective rate while
the patient is not moving (at rest).
2.6.9.2 Rate Increase / Decrease
The Rate Increase and Decrease parameters work with the
Sensor Gain to determine how quickly the pacing rate will
increase or decrease during changes in the sensor output.
2.6.9.3 Maximum Sensor Rate
Regardless of the sensor output, the sensor-driven pacing rate
never exceeds the programmable Max. Sensor Rate. The
maximum sensor rate only limits the pacing rate during sensordriven pacing.
2.6.9.4 Auto Sensor Gain
The Lexos ICDs offer Automatic Sensor Gain settings, which
allow the Auto Gain parameter to be adjusted automatically.
Lexos Technical Manual 53
When the Automatic Sensor Gain is activated, the pulse
generator samples the sensor-indicated rate. If, during the
24 hour period beginning at midnight, the total time recorded at
maximum sensor rate exceeds 90 seconds, the sensor gain
setting is reduced by one step. The sensor gain will be
increased by one step after 7 consecutive days during which the
time recorded at maximum sensor rate is less than 90 seconds
each day.
2.6.10 Pulse Amplitude
The Pulse Amplitude parameters, both atrial and ventricular,
define the amplitude in volts of the pacing pulses. The pulse
amplitude is independently programmed for normal and postshock bradycardia pacing.
In addition, if the time received at maximum sensor rate is more
than 90 seconds, the sensor gain is decreased by 1 step.
2.6.11 Pulse Width
The Pulse Width parameters, both atrial and ventricular, define
the duration of the pacing pulses. The pulse width is
independently programmed for normal and post-shock
bradycardia pacing.
2.6.12 Post Ventricular Atrial Refractory Period
Immediately following a each sensed or paced ventricular event,
an atrial refractory period is started, this period is called Post
Ventricular Atrial Refractory Period or PVARP. Atrial signals are
ignored during this time for bradycardia timing purposes to
prevent the ICD from sensing inappropriate signals. PVARP is
only available with the Lexos DR and DR-T ICDs.
2.6.13 PVARP Extension
Extends the Post Ventricular Atrial Refractory Period by the
programmed interval. PVARP Extension is only available with
the Lexos DR and DR-T ICDs.
54 Lexos Technical Manual
2.6.14 Noise Response
The Lexos ICD’s response to detected noise is to deliver
asynchronous pacing in the affected channel.
2.6.15 Post Shock Pacing
Separately programmable bradycardia pacing support is
available with the ICD following shock therapy delivery.
Because a delay in bradycardia pacing may avoid re-initiation of
a tachyarrhythmia after a short blanking period (1 second) the
ICD will begin bradycardia therapy at the post shock pacing rate,
amplitude, and pulse width for the programmed Post Shock Duration.
Separate post shock programming of the following parameters is
available:
• Basic Rate
• Hysteresis
• AV Delay
If bradycardia pacing is still required after the post shock
duration expires, standard bradycardia pacing parameters will be
active.
2.7 EP Test Functions
Several EP test functions are available with the Lexos family of
ICDs including; P and R-wave amplitude, pacing and shock
impedance, retrograde conduction and pacing threshold
measurements. Extensive testing of defibrillation thresholds as
well as the ability to verify the effectiveness of anti-tachycardia
pacing and defibrillation shocks are also available.
2.7.1 P and R-wave Amplitude Measurments
The Lexos ICDs provide a P-/R-wave test for measuring the
amplitude of intrinsic events during follow-up examination. The
test determines the amplitudes with a predetermined temporary
pacing mode.
Lexos Technical Manual 55
To permit evaluation of the sensing function, the pacing rate
must be lower than the patient's intrinsic rate. In demand
pacing, the proper sensing function can be recognized if the
interval between intrinsic events and the following pacing pulse
equals the basic interval (if no Hysteresis is programmed). The
following parameters are programmable when performing the
measurements:
• Pacing Mode
• Pacing Rate
• AV Delay
• Automatic Printing Function
For evaluation of the sensing function, the pulse generator
features an intracardiac electrogram (IEGM) with marker signals
to indicate sensed and paced events.
2.7.2 Testing for Retrograde Conduction
Retrograde conduction from the ventricle to the atrium can be
assumed when a 1:1 relationship between the ventricular
stimulation and atrial depolarization has been obtained with a
constant coupling interval during ventricular stimulation. The
ICD features a test for measuring retrograde conduction time.
During operation of this test, the patient is paced (in VDI mode)
at an increased ventricular rate over several cycles while the
retrograde conduction time is measured. Therefore, the pacing
Rate must be programmed at a rate higher than the patient’s
intrinsic rhythm.
Both the programmer display and printout provide measured
retrograde conduction times. The duration of time that the test is
conducted is based on how long the Measure button is
depressed. The paper speed for the test printout is also
programmable for this test.
To prevent retrograde P-waves from triggering ventricular
pulses, thereby mediating a “re-entry” tachycardia (pacemaker
mediated tachycardia, PMT), it is recommended that the
programmed post-ventricular atrial refractory period be
programmed longer than the retrograde conduction time.
56 Lexos Technical Manual
2.7.3 Pacing Threshold
The test is activated as a temporary program, and removal of the
programmer head immediately stops the test and reactivates the
permanent program.
The following parameters are programmable during the pacing
threshold test: Appropriate chamber and pacing mode, pacing
rate, AV Delay (if appropriate), pulse amplitude and pulse width,
number of pulses for each test voltage and automatic printing
capabilities. The pacing modes available for the threshold test
are AAI, VVI, DDI, and DDD. The pulse amplitude is easily
adjustable during the threshold testing by selecting the desired
value from the table.
W
ARNING
Resuscitation Availability - Do not perform induction testing
unless an alternate source of patient defibrillation such as an
external defibrillator is readily available. In order to implant
the ICD system, it is necessary to induce and convert the
patient’s ventricular tachyarrhythmias.
C
AUTION
Manual Shocks – User-commanded shocks may be withheld
if the ICD is already busy processing a manual command or
the Battery Status is low.
2.7.4 Arrhythmia Induction Features
The ICD offers three arrhythmia induction methods for noninvasive EP testing. These include the following:
HF Burst Induction This feature consists of a large number of
pulses delivered in rapid succession over a period of several
seconds. The frequency of the pulses and the duration of the
burst are defined by the user.
Lexos Technical Manual 57
Burst + PES Induction delivers a programmed number of burst
of pacing stimuli followed by a programmable number of timed
extra stimuli. The burst rate is independently programmable, as
is the Number S1. The interval between S1s and the remaining
programmed extra stimuli (PES: S1 through S4 possible) is also
programmable.
Shock on T induction mode allows tachyarrhythmia induction by
means of a timed T wave shock delivered after a series of paced
stimuli. Energy of the T wave shock, number of pulses
(Number S1) in the pulse train, synchronization interval (R-S1)
and the shock Coupling interval are all user programmable.
2.7.5 Manual Shock
The ICD can deliver a manual shock on demand through a
programmer command in the EP test menu. To deliver a shock,
place the wand over the device and select the Start Shock
button. A confirmation menu will appear and the shock
command will be delivered upon selecting the OK button in this
screen. After each manual shock, the EP test screen will display
the shock energy, lead impedance and charge time.
2.7.6 Manual ATP
The ICD can deliver a manual ATP on demand through a
programmer command in the EP test menu. To deliver an ATP
sequence, place the wand over the device and select the Start ATP button. A confirmation menu will appear and the
programmed pacing sequence command will be delivered upon
selecting the OK button in this screen. Programming of the
manual ATP is similar to the programming available for
automatic ATP therapy as described in
Section 2.5.2
.
2.7.7 Test Shock
The ICD can deliver a 1 joule (R-wave synchronous) test shock
on demand through a programmer command in the EP test
menu. This shock is designed to measure the shock impedance
and test the integrity of the shock electrodes of an implanted ICD
lead.
58 Lexos Technical Manual
W
ARNING
Resuscitation Availability - Do not perform induction testing
unless an alternate source of patient defibrillation such as an
external defibrillator is readily available. In order to implant
the ICD system, it is necessary to induce and convert the
patient’s ventricular tachyarrhythmias.
C
AUTION
Defibrillation Threshold - Be aware that the changes in the
patient’s condition, drug regimen, and other factors may
change the defibrillation threshold (DFT) which may result in
non-conversion of the arrhythmia post-operatively.
Successful conversion of ventricular fibrillation or ventricular
tachycardia during arrhythmia conversion testing is no
assurance that conversion will occur post-operatively.
2.8 Special Features
W
ARNING
Unwanted Shocks – Always program the VT/VF Detection
and Therapy status to DISABLED prior to handling the device
to prevent the delivery of serious shocks to the patient or the
person handling the device during the implant procedure.
2.8.1 Detection and Therapy Status
Interrogating the device and observing the detection and therapy
section (upper right hand corner) of the main programming
screen indicates the detection and therapy status (either
ENABLED or DISABLED). The status can be changed by
selecting the Detection section of the main programming screen
and selecting the desired setting from the resulting pop-up
screen.
Lexos Technical Manual 59
2.8.2 Home Monitoring (Lexos DR-T and VR-T
Only)
Home Monitoring enables the exchange of information about a
patient’s cardiac status from the implant to the physician. Home
Monitoring can be used to provide the physician with advance
reports from the implant and process them into graphical and
tabular format called a Cardio Report. This information helps the
physician optimize the therapy process, as it allows the patient to
be scheduled for additional clinical appointments between
regular follow-up visits if necessary.
W
ARNING
The use of Home Monitoring does not replace regular followup examinations. Therefore, when using Home Monitoring,
the time period between follow-up visits may not be extended.
The implant’s Home Monitoring function can be used for the
entire operational life of the implant (prior to ERI) or for shorter
periods, such as several weeks or months.
N
2.8.2.1 Transmission of Information
The implant transmits information with a small transmitter, which
has a range of about 6 feet (2 meters). The patient’s implant
data are sent to the corresponding patient device upon the
detection of an arrhythmia episode, as programmed. The types
of transmissions are discussed in Section 4.
The minimal distance between the implant and the patient device
must be 6 inches (15 cm).
:
OTE
When ERI mode is reached, this status is transmitted.
Further measurements and transmissions of Home
Monitoring data are no longer possible.
60 Lexos Technical Manual
2.8.2.2 Patient Device
The patient device (Figure 5
) is designed for use in or away
from the home and is comprised of the mobile unit and the
associated charging station. The patient can carry the mobile
unit with them during his or her occupational and leisure
activities. The patient device is rechargeable, allowing for an
approximate operational time of 24 hours. It receives
information from the implant and forwards it via the mobile
network to a BIOTRONIK Service Center.
For additional information about the patient device, please refer
to its manual.
Figure 5: Example of Patient Device with Charging Stand
Lexos Technical Manual 61
2.8.2.3 Cardio Report
The implant’s information is digitally formatted by the
BIOTRONIK Service Center and processed into a concise report
called a Cardio Report. The Cardio Report is available in two
formats, which are titled depending on the type of report
transmission – trend and event. This Cardio Report, which is
adjusted to the individual needs of the patient, contains current
and previous implant data. The Cardio Report is sent to the
attending physician via fax. All reports use the same report
format.
2.8.2.4 Types of Report Transmissions
When the Home Monitoring function is activated, the
transmission of a report (Cardio Report) from the implant can be
triggered as follows:
• Event report – the ICD detects certain events, which initiate a
report immediately
• Trend Event report – the ICD detects certain events, which
initiate a report at the programmed time of trend transmission.
To assure successful transmission of the patient data, the
Lexos VR-T and Lexos DR-T send 4 repetitive transmissions of
identical data at a 60 minute time interval.
N
Event Report
When certain cardiac and technical events are detected by the
implant, a report transmission is automatically triggered. This is
described as an “event message”.
:
OTE
Battery voltage and pace/sense lead impedance are
measured before the first transmission of the day.
Therefore, the first transmission may occur 2 minutes after
the programmed transmission time.
62 Lexos Technical Manual
The following cardiac and technical events initiate an immediate
message transmission:
• Special Device Status
• Detected and terminated VT-1
• Detected and terminated VT-2
• Detected and terminated VF
• First ineffective 30 J shock detected
• Pace impedance < 300 Ohm or > 1 K Ohm
• Shock impedance < 30 Ohm or > 110 Ohm
• Device status – ERI
Trend Event Report
When certain cardiac and technical events are detected by the
implant, a report transmission is triggered at the time of trend
transmission. This is described as a “trend event message”.
The following cardiac and technical events initiate an immediate
message transmission:
• Change of lead impedance (delta limits are
programmable by the physician)
W
ARNING
A timely receipt of the event report cannot be guaranteed.
The receipt is also dependent on whether the patient was
physically situated in the required coverage range of the
patient device at the time the event information was sent.
2.8.2.5 Description of Transmitted Data
The following data are transmitted by the Home Monitoring
system, when activated. In addition to the medical data, the
serial number of the implant is also transmitted.
The Monitoring Interval
Lexos Technical Manual 63
The monitoring interval is considered the time period since the
last trend information was sampled. In the absence of an event
report, the monitoring interval would be 24 hours. For an event
report, the monitoring interval would be less than 24 hours, since
these reports are sent after the programmed time of trend
transmission.
Detection
• # of Episodes in VT1 Zone
• # of Episodes in VT2 Zone
• # of Episodes in VF Zone
• # of SVT events*
Therapy
• ATPs delivered
• ATPs successful
• Shocks delivered
• Shocks successful
• Shocks aborted
• 30J Shock without Success
Intrinsic Rhythm*
• Percentage of Atrial Senses (A
s/Ax
• Percentage of Ventricular Senses (V
Battery
• Status (i.e., BOL, MOL1, MOL2, ERI, EOS)
• Battery Voltage
• Date of voltage measurement
Leads
• Pace Impedance (atrial*, ventricular)
• Shock Impedance
• Date of impedance measurements
Device Status Summary
)
)
s/Vx
64 Lexos Technical Manual
• Status
• Remarks
*Available only with dual chamber ICDs.
2.8.3 Real-time IEGM Transmission
The pulse generators provide real time transmission of the
unfiltered intracardiac electrogram (IEGM) to the programmer.
During dual chamber operation, IEGMs from the atrium and
ventricle can be simultaneously recorded with a bandwidth of 0.5
to 200 Hz. During single chamber operation, a far field
ventricular electrogram can be simultaneously recorded. The
IEGMs may be transmitted to the programmer via the
programming head positioned over the implanted pulse
generator. They are then displayed together with surface ECG
and markers on the programmer screen and printed on the ECG
recorder. Likewise, intracardiac signals and markers identifying
atrial/ventricular paced and sensed events are received via the
programming head, and may be displayed on the programmer
screen and printed on the ECG recorder.
To determine the amplitudes of intracardiac signals (P-/Rwaves) the automatic P/R-wave measurement function may be
used.
Please refer to the appropriate software technical manual for a
description of marker signal operation.
2.8.4 Capacitor Reformation
Shock charge times may be prolonged if the high voltage
capacitors remain uncharged for an extended period of time.
Conditioning (or re-forming) the capacitors by periodically
charging them will help assure shorter charge times in those
patients that do not regularly receive shock therapy. The ICD
may be programmed to automatically re-form the capacitors after
every 3, or 6 months or not at all (OFF). The capacitor
reformation clock is reset following an automatic or manual
capacitor re-form, or any device initiated maximum charging of
the high voltage capacitors.
Lexos Technical Manual 65
An automatic or manually initiated capacitor reform fully charges
the capacitors with a specific sequence and then allows the
capacitors to discharge into an internal resistor. No shock will
be delivered to the patient. Throughout the re-formation process
the ICD will provide bradycardia pacing support and
tachyarrhythmia sensing and detection as programmed. If a
tachyarrhythmia is detected during capacitor re-formation, the
process is aborted and therapy is available if required.
C
AUTION
Capacitor Reformation - Infrequent charging of the high
voltage capacitors may extend the charge times of the ICD.
The capacitors may be reformed manually, or the ICD may
be programmed to reform the capacitors automatically.
N
Biotronik recommends automatic reforming capacitors, because
disabling this function may reduce the benefits associated with
frequent capacitor charging (e.g., short charge times in patients
who have infrequent shock therapy).
OTE
:
2.8.5 Patient and Implant Data
The Patient and Implant data screens allow input of data
regarding the patient name, demographics, implanting physician,
date, devices implanted, location of the implant, and various
conditions related to the patient. This information is transmitted
to the ICD and resides in the device memory for later recall if
needed.
66 Lexos Technical Manual
2.8.6 System Status
Various device parameters can be monitored through the Status
section of the programmer screen. Displayed data includes ICD
information, charge circuit parameters, capacitor reform
information, battery status and voltage, and lead information.
The system status screen presents a large variety of information
about the Lexos ICDs including:
• Serial number (always displayed after interrogation)
• Software Release
• Device status
• Battery voltage
• Battery status
• Last charge event
− Date
− Energy
− Charge time
• Total number of charges
• Last P and R-wave measurements
• Last pacing lead impedance (atrial and ventricle)
• Last pacing threshold measurement with pulse width
(atrial and ventricle)
• Last shock impedance measurement and date
2.8.7 Holter Memory
Various device information is available within the Holter memory.
The Holter memory can be configured a number of different
ways depending on the physician’s preference.
2.8.7.1 Episode List
The ICD stores a variety of useful diagnostic data about
tachyarrhythmia episodes, which may be used to optimize
tachyarrhythmia detection and therapy parameters. This
diagnostic data includes detection counters, therapy counters,
last delivered ATP and shock therapy, shock data memory,
therapy history, and stored intracardiac electrograms.
Lexos Technical Manual 67
E
PISODE DETAILS
Detailed information about each individual episode presented as
a table of events ordered from most recently delivered to the first
delivered. Each IEGM segment can be viewed from the episode
detail sub-menu by selecting the EGM button. From this screen,
an IEGM can be expanded and scrolled to assist in a more
accurate IEGM interpretation by enabling a closer examination of
specific segments.
68 Lexos Technical Manual
S
TORED
IEGM
The ICD can store up to 30 minutes of dual chamber intracardiac
electrograms (IEGMs) including the history and prehistory of the
following events:
• Detection
• Redetection
• Terminations
• Manual Shocks
• SVT Success and/or Termination
TORED INTERVALS
S
The ICD can store a large number of cardiac intervals including
the history and prehistory of events. The graphic display of
these intervals includes significant events to provide a base
timeline (e.g., shocks, detections, terminations). The following
intervals are available:
• R-R
• P-P
• P-R
• R-P
2.8.7.2 Shocks
The device history regarding high energy shocks is presented in
a table format with the following information:
• Shock Number
• Date
• Time
• Energy
• Charge time
• Impedance
• Type of shock
2.8.7.3 Counters
Lexos Technical Manual 69
The device history regarding several therapy and detection
parameters is presented in the “Counters” screen. This screen
contains both the number of events since the last ICD follow-up
and totals since the device was implanted. The available
parameters include:
D
ETECTION EPISODES
• SVT
• VT-1
• VT-2
• VF
HERAPIES
T
• Successful ATP Therapies
• Unsuccessful ATP Therapies
• Successful ATP Therapies
• Unsuccessful ATP Therapies
SVT D
ETAILS
• AFlut
• AFib
• Sinus T
• 1:1
2.8.8 Real-time IEGM
The surface ECG is continuously displayed in the Overview
screen, the Sensing screen and the EP test functions module.
Real-time IEGMs are available in the EP tests and sensing /
impedance screens.
The sensing / impedance screen allows automatic measurement
of P-waves and R-waves. The sensing / impedance screen also
allows a temporary bradycardia program to be sent to the ICD
for evaluation of pacing parameters. IEGM markers are
available for all sensed and paced events.
70 Lexos Technical Manual
2.8.9 Brady Diagnostics
The ICD stores a variety of useful diagnostic data of the
bradycardia history as described in the following sections.
2.8.9.1 Event Counters
The total number of atrial sensed, atrial paced, ventricular sense
and ventricular paced events since the statistics package was
initiated are available. The total percentage of time for each of
the above listed events is also available.
2.8.9.2 Activity Report
The activity report provides information that can assist the
physician in optimizing pacing and/or sensor parameters. This
report contains the maximum sensor rate attained and the mean
sensor rate.
2.8.9.3 Ventricular Rate Histogram
The ventricular rate histogram shows the percentage of time the
ventricular rate lies within given heart rate bins regardless if the
sensor is used or not. The heart rate range is divided into
sixteen segments ranging from less than 40 to greater than
179 bpm.
2.8.9.4 Sensor Rate Histogram
The sensor rate histogram shows the percentage of time the
sensor rate lies within given heart rate bins regardless if the
sensor is used or not. The heart rate range is divided into
sixteen segments ranging from less than 40 to greater than
179 bpm.
Lexos Technical Manual 71
72 Lexos Technical Manual
3. Sterilization and Storage
The ICD is shipped in a storage box, equipped with a quality
control seal and product information label. The label contains
the model specifications, technical data, serial number, use
before date, and information on sterilization and storage.
The ICD and its accessories have been sealed in a container
and gas sterilized with ethylene oxide. To assure sterility, the
container should be checked for integrity prior to opening.
C
AUTION
Device Packaging - Do not use the device if the device’s
packaging is wet, punctured, opened or damaged because
the integrity of the sterile packaging may be compromised.
Return the device to BIOTRONIK.
Re-sterilization - Do not re-sterilize and re-implant
explanted devices.
Storage (temperature) - Store the device between 5° to
55°C (41° - 131° F) because temperatures outside this range
could damage the device.
Storage (magnets) - To avoid damage to the device, store
the device in a clean area, away from magnets, kits
containing magnets, and sources of electromagnetic
interference (EMI).
Temperature Stabilization - Allow the device to reach room
temperature before programming or implanting the device
because temperature extremes may affect initial device
function.
Use Before Date - Do not implant the device after the USE
BEFORE DATE because the device may have reduced
longevity.
Lexos Technical Manual 73
74 Lexos Technical Manual
4. Implant Procedure
4.1 Implant Preparation
Prior to beginning the ICD implant procedure; ensure that all
necessary equipment is available. The implant procedure
requires the selected lead system (including sterile back-ups),
the programmer with appropriate software, and the necessary
cabling and accessories.
PLUS
For TMS 1000
accessories are available:
PK44 - used to connect the TMS 1000
systems for complete testing of the lead systems during the
implant procedure. The following adapters may be necessary:
• Adapters PA-2/PA-3 - The PA-2 adapter is used to
connect IS-1 compatible leads to the PK-44 cable. The
PA-3 adapter is used to connect DF-1 compatible leads
to the PK-44 cable.
• Adapter PA-4 - used to connect the PK-44 cable to
sensing and pacing leads while the stylet is still inserted.
The ICD System also has the following accessory available (at
the discretion of the physician) for the implant procedure:
based testing, the following cabling and
PLUS
to implanted lead
Test housing that allows acute testing of the lead system prior to
opening the sterile package.
Perform an interrogation of the ICD. Ensure programmer
operation, nominal device parameters and battery status is
appropriate for a new Lexos ICD. Note that the battery status
may appear lower than its true value when the ICD is not at body
temperature. Program detection and therapy to “Disabled” prior
to handling the Lexos ICD.
Sufficient training on the device and its associated components
is required prior to implanting the ICD. For additional
information, training and training materials contact your
BIOTRONIK representative.
Lexos Technical Manual 75
W
ARNING
Lead Systems - The use of another manufacturer’s ICD lead
system may cause potential adverse consequences such as
undersensing of cardiac activity and failure to deliver
necessary therapy.
C
AUTION
Blind Plug - A blind plug must be inserted and firmly
connected into any unused header port to prevent chronic
fluid influx and possible shunting of high energy therapy.
Connector Compatibility - ICD and lead system
compatibility should be confirmed prior to the implant
procedure. Consult your BIOTRONIK representative
regarding lead/pulse generator compatibility prior to the
implantation of an ICD system. For further information,
please refer to
Pacemaker/ICD Interaction - In situations where an ICD
and a pacemaker are implanted in the same patient,
interaction testing should be completed. If the interaction
between the ICD and the pacemaker cannot be resolved
through repositioning of the leads or reprogramming of either
the pacemaker or the ICD, the pacemaker should not be
implanted (or explanted if previously implanted).
Programmed Parameters – Program the device parameters
to appropriate values based on the patient’s specific
arrhythmias and condition.
Appendix A
.
76 Lexos Technical Manual
C
AUTION
Shock Impedance - If the shock impedance is less than
twenty-five ohms, reposition the lead system to allow a
greater distance between the electrodes. Never implant the
device with a lead system that has a measured shock
impedance of less than twenty-five ohms. Damage to the
device may result.
Far-field sensing of signals from the atrium in the ventricular
channel or ventricular signals in the atrial channel should be
avoided by appropriate lead placement, programming of
pacing/sensing parameters, and maximum sensitivity
settings. If it is necessary to modify the Far Field Blanking
parameter, the parameter should be lengthened only long
enough to eliminate far-field sensing as evidenced on the
IEGMs. Extending the parameter unnecessarily may cause
undersensing of actual atrial or ventricular events.
4.2 Lead System Evaluation
The ICD is mechanically compatible with DF-1 defibrillation lead
connectors and IS-1 sensing and pacing lead connectors. IS-1,
wherever stated in this manual, refers to the international
standard, whereby leads and pulse generators from different
manufacturers are assured a basic fit [Reference ISO
5841-3:1992]. DF-1, wherever stated in this manual, refers to
the international standard [Reference ISO 11318:1993].
Refer to the appropriate lead system technical manual.
4.3 Opening the Sterile Container
The Lexos ICDs are packaged in two plastic containers, one
within the other. Each is individually sealed and then sterilized
with ethylene oxide.
Due to the double packing, the outside of the inner container is
sterile and can be removed using standard aseptic technique
and placed on the sterile field.
Lexos Technical Manual 77
Peel off the sealing paper of the outer
container as indicated by the arrow.
Do not contaminate the inner tray.
ST_01
Take out the inner sterile tray by
gripping the tab. Open the inner tray
by peeling the sealing paper as
indicated by the arrow.
C
AUTION
Device Packaging - Do not use the device if the device’s
packaging is wet, punctured, opened or damaged because
the integrity of the sterile packaging may be compromised.
Return the device to BIOTRONIK.
4.4 Pocket Preparation
Using standard surgical technique, create a pocket for the
device either in the patient’s pectoral or abdominal region
dependent on patient anatomy. The device may be implanted
either below the subcutaneous tissue or in the muscle tissue.
The ICD should be implanted with the etched side facing up.
The leads should be tunneled or surgically brought into the
device pocket. If lead tunneling is performed, re-evaluation of
the baseline lead signals, after tunneling is recommended.
78 Lexos Technical Manual
C
AUTION
The ICD system should have detection and therapy disabled
prior to performing medical procedures. In addition, the ICD
should be checked after the procedures to assure proper
programming:
Electrocautery - Electrosurgical cautery could induce
ventricular arrhythmias and/or fibrillation, or may cause
device malfunction or damage. If use of electrocautery is
necessary, the current path and ground plate should be kept
as far away from the pulse generator and leads as possible
(at least 6 inches (15 cm)).
4.5 Lead to Device Connection
The Lexos ICDs have been designed and are recommended for
use with a defibrillation lead systems having one IS-1 connector
for ventricular sensing and pacing and up to two DF-1
connectors for delivery of shock therapy. A separate bipolar
atrial lead with IS-1 connector is required for atrial sensing and
pacing functions (Lexos DR and DR-T only).
the configuration of the header ports on the Lexos DR , where
HV1 and HV2 are for DF-1 connectors, and A P/S and V P/S are
for IS-1 connectors. Lexos VR and VR-T ICDs are identical
except for the absence of the Atrial pace/sense port.
Figure 6
depicts
Figure 6. Header Ports
Lexos Technical Manual 79
80 Lexos Technical Manual
C
AUTION
Connector Compatibility - ICD and lead system
compatibility should be confirmed prior to the implant
procedure. Consult your BIOTRONIK representative
regarding lead/pulse generator compatibility prior to the
implantation of an ICD system. For further information,
please refer to
Setscrew Adjustment – Back-off the setscrew(s) prior to
insertion of lead connector(s) as failure to do so may result in
damage to the lead(s), and/or difficulty connecting lead(s).
Cross Threading Setscrew(s) – To prevent cross threading
the setscrew(s), do not back the setscrew(s) completely out
of the threaded hole. Leave the torque wrench in the slot of
the setscrew(s) while the lead is inserted.
Tightening Setscrew(s) – Do not overtighten the
setscrew(s). Use only the BIOTRONIK supplied torque
wrench.
Sealing System – Be sure to properly insert the torque
wrench into the perforation at an angle perpendicular to the
connector receptacle. Failure to do so may result in damage
to the plug and its self-sealing properties.
Far-field sensing of signals from the atrium in the ventricular
channel or ventricular signals in the atrial channel should be
avoided by appropriate lead placement, programming of
pacing/sensing parameters, and maximum sensitivity
settings. If it is necessary to modify the Far Field Blanking
parameter, the parameter should be lengthened only long
enough to eliminate far-field sensing as evidenced on the
IEGMs. Extending the parameter unnecessarily may cause
undersensing of actual atrial or ventricular events.
Appendix A
.
Lexos Technical Manual 81
Refer to the following steps when connecting the leads to the
device.
1. Confirm that the setscrews are not protruding into the
connector receptacles. To retract a setscrew, insert the
enclosed torque wrench through the perforation in the
self-sealing plug at an angle perpendicular to the lead
connector until it is firmly placed in the setscrew. Rotate
the wrench counterclockwise until the receptacle is clear
of obstruction.
2. Insert the lead connector into the connector port of the
ICD without bending the lead until the connector pin
becomes visible behind the setscrew. Hold the
connector in this position. If necessary, apply silicone oil
only to the o-rings on the connector (not the connector
pin).
3. Insert the enclosed torque wrench through the
perforation in the self-sealing plug at an angle
perpendicular to the lead connector until it is firmly
placed in the setscrew.
4. Securely tighten the setscrew of the connector clockwise
with the torque wrench until torque transmission is
limited by the wrench.
5. After carefully retracting the torque wrench, the
perforation will self-seal.
4.6 Blind Plug Connection
The Lexos DR and DR-T ICDs come with a blind plug (pre
inserted) in an unused header port. Refer to the following steps
when connecting blind plugs to the device.
82 Lexos Technical Manual
1. Confirm that the setscrews are not protruding into the
connector receptacles. To retract a setscrew, insert the
enclosed torque wrench through the perforation in the
self-sealing plug at an angle perpendicular to the lead
connector until it is firmly placed in the setscrew. Rotate
the wrench counterclockwise until the receptacle is clear
of obstruction.
2. Insert the blind plug into the connector port of the ICD
until the connector pin becomes visible behind the
setscrew.
3. Insert the enclosed torque wrench through the
perforation in the self-sealing plug at an angle
perpendicular to the connector until it is firmly placed in
the setscrew.
4. Securely tighten the setscrew of the connector clockwise
with the torque wrench until torque transmission is
limited by the wrench.
5. After carefully retracting the torque wrench, the
perforation will self-seal.
C
AUTION
Blind Plug - A blind plug must be inserted and firmly
connected into any unused header port to prevent chronic
fluid influx and possible shunting of high energy therapy.
Lexos Technical Manual 83
4.7 Pacemaker Interaction Testing
There are three situations in which pacemaker/ICD interaction
testing is appropriate when:
• a pacemaker and an ICD are implanted at the same
procedure
• an ICD is implanted in a patient with a chronic
pacemaker
• a pacemaker is implanted in a patient with a chronic ICD
In each of these cases, the pacemaker and ICD may interact in
such a way that the pacemaker could interfere with the
classification of tachyarrhythmias by the ICD. The three possible
mechanisms of interaction are listed below:
• During a tachyarrhythmia episode, the pacemaker may
not detect the patient’s tachyarrhythmia. In addition, the
amplitude of the pacemaker pacing pulses may be large
enough to cause the ICD to detect only the pacing
pulses and not sense the underlying tachyarrhythmia.
Therefore, the ICD would not provide appropriate antitachyarrhythmia therapy.
• The ICD may detect both the pacing pulses and the
resulting ventricular response as separate signals
(doubled count). The ICD might then classify the normal
paced rhythm as a tachyarrhythmia and subsequently
deliver therapy inappropriately.
• If the pacemaker experiences a sensing failure, a lead
dislodgment, or lack of capture the ICD could sense the
asynchronous pacing pulses along with the patient’s
normal sinus rhythm. The ICD may then classify the
rhythm as a tachyarrhythmia and deliver inappropriate
therapy.
The following test procedures should be performed during
implantation of the ICD with a concomitant pacemaker. There
are two separate procedures that must be completed.
84 Lexos Technical Manual
Part 1
Verify that inappropriate therapy will not be initiated by
oversensing of pacemaker pulses.
1. Program the detection status and magnet mode of the
ICD to “DISABLED”.
2. Keep the programming wand in place over the ICD to
observe the intracardiac electrograms and markers
when the pacemaker is inhibited.
3. Program the pacemaker’s lower rate and AV Delay, if
applicable, to values that ensure consistent pacing.
Program the pacemaker to unipolar (or bipolar) pacing
with the pacing amplitude and pulse width parameters at
maximum values.
4. Observe the intracardiac electrograms and markers
again. If either signal shows events that are
oversensed, the ICD or pacemaker leads should be
repositioned in order to minimize the amplitude of the
pacing artifacts.
5. It may be necessary to reduce the pacing amplitude and
pulse width settings of the pacemaker during testing to
eliminate interaction with the ICD. If testing indicates a
set of maximum allowable programmable parameters, it
should be recorded in the patient’s record for future
reference, in the event that reprogramming is required.
Part 2
Verify that oversensing of pacemaker pulses during a
tachyarrhythmia episode will not inhibit tachyarrhythmia therapy.
1. Program the pacemaker to a unipolar asynchronous
pacing mode (V00 or D00) at maximum pacing
amplitude and pulse width settings.
2. Program the detection status of the ICD to “ENABLED”.
3. Induce ventricular fibrillation from the EP Test screen
4. Observe the intracardiac electrograms and the markers.
BE PREPARED TO DELIVER AN EMERGENCY
SHOCK IF THE TACHYARRHYTHMIA IS NOT
DETECTED AND TERMINATED BY THE ICD.
Lexos Technical Manual 85
5. If the ICD did not detect the tachyarrhythmia, reduce the
pacemaker’s output settings and repeat step 4 until
maximum allowable pacemaker output settings are
defined. The maximum allowable programming set
should be recorded in the patient’s records for future
reference, should reprogramming be required.
6. After conversion testing is complete, interrogate the
pacemaker to ensure that its programmed parameters
have not been changed and that no damage was
caused by delivery of therapy by the ICD.
7. Program the pacemaker to the appropriate pacing
parameters based on the completed testing.
To reduce the possibilities of pacemaker/ICD interaction, it is
recommended that:
• the ICD and pacemaker leads be placed as far away as
possible from one another
• the pacemaker leads with a short inter-electrode spacing
be used
• the pacemaker be programmed to the lowest allowable
amplitude and pulse width to ensure consistent, chronic
capture
• the pacemaker must be programmed to the maximum
sensitivity (without oversensing during a normal rhythm)
to ensure pacing is inhibited during tachyarrhythmia
episodes.
• the pacemaker be programmed to the minimum lower
rate sufficient for the patient.
Pacemaker/ICD Interaction - In situations where an ICD
and a pacemaker are implanted in the same patient,
interaction testing should be completed. If the interaction
between the ICD and the pacemaker cannot be resolved
through repositioning of the leads or reprogramming of either
the pacemaker or the ICD, the pacemaker should not be
implanted (or explanted if previously implanted).
C
AUTION
86 Lexos Technical Manual
4.8 Program the ICD
Program the ICD to appropriately treat the patient’s arrhythmias
and other therapy needs. The information obtained during the
lead system evaluation should be helpful in tailoring the various
parameters of the ICD to treat each individual patient. The
detection and therapy status of the ICD may be activated for
testing purposes once all of the lead connectors have been
securely fastened in the device header ports. The physician
shall be made aware of the program that is in effect after the
patient leaves the office, by viewing the parameters displayed on
the programmer screen after the device has been programmed
and interrogated.
C
AUTION
Programmed Parameters – Program the device parameters
to appropriate values based on the patient’s specific
arrhythmias and condition.
Programmers - Use only BIOTRONIK programmers to
communicate with the device (TMS 1000
EPR 1000
Defibrillation Threshold - Be aware that the changes in the
patient’s condition, drug regimen, and other factors may
change the defibrillation threshold (DFT) which may result in
non-conversion of the arrhythmia post-operatively.
Successful conversion of ventricular fibrillation or ventricular
tachycardia during arrhythmia conversion testing is no
assurance that conversion will occur post-operatively.
Unwanted Shocks – Always program the VT/VF Detection
and Therapy status to DISABLED prior to handling the device
to prevent the delivery of serious shocks to the patient or the
person handling the device during the implant procedure.
PLUS
).
W
ARNING
PLUS
, or
Lexos Technical Manual 87
4.9 Implant the ICD
The ICD may be placed in the pocket at this time. Place the
device into the pocket with the etched side facing up. Carefully
coil any excess lead length beside or above the ICD.
The pacing and sensing functions of the device should be
evaluated. It is also recommended that at least one induction
and device conversion be done prior to closing the pocket. This
will ensure that the lead system has been securely connected to
the device and has not changed position.
C
AUTION
Connector Compatibility - ICD and lead system
compatibility should be confirmed prior to the implant
procedure. Consult your BIOTRONIK representative
regarding lead/pulse generator compatibility prior to the
implantation of an ICD system. For further information,
please refer to
Pacemaker/ICD Interaction - In situations where an ICD
and a pacemaker are implanted in the same patient,
interaction testing should be completed. If the interaction
between the ICD and the pacemaker cannot be resolved
through repositioning of the leads or reprogramming of either
the pacemaker or the ICD, the pacemaker should not be
implanted (or explanted if previously implanted).
Shock Impedance - If the shock impedance is less than
twenty-five ohms, reposition the lead system to allow a
greater distance between the electrodes. Never implant the
device with a lead system that has a measured shock
impedance of less than twenty-five ohms. Damage to the
device may result.
Appendix A
.
88 Lexos Technical Manual
W
ARNING
Resuscitation Availability - Do not perform induction testing
unless an alternate source of patient defibrillation such as an
external defibrillator is readily available. In order to implant
the ICD system, it is necessary to induce and convert the
patient’s ventricular tachyarrhythmias.
C
AUTION
Pacing Threshold - Testing of the pacing threshold by the
ICD system should be performed with the pacing rate
programmed to a value at least 20 ppm higher than the
patient's intrinsic rate.
Defibrillation Threshold - Be aware that the changes in the
patient’s condition, drug regimen, and other factors may
change the defibrillation threshold (DFT) which may result in
non-conversion of the arrhythmia post-operatively.
Successful conversion of ventricular fibrillation or ventricular
tachycardia during arrhythmia conversion testing is no
assurance that conversion will occur post-operatively.
Electromagnetic interference (EMI) signals present in
hospital and medical environments may affect the function of
any ICD or pacemaker. The ICD is designed to selectively
filter out EMI noise. However, due to the variety of EMI
signals, absolute protection from EMI is not possible with this
or any other ICD.
The ICD system should have detection and therapy disabled
prior to performing any of the following medical procedures.
In addition, the ICD should be checked after the procedures
to assure proper programming:
Electrocautery - Electrosurgical cautery could induce
ventricular arrhythmias and/or fibrillation, or may cause
device malfunction or damage. If use of electrocautery is
necessary, the current path and ground plate should be kept
as far away from the pulse generator and leads as possible
(at least 6 inches (15 cm)).
Lexos Technical Manual 89
Prior to surgically closing the pocket, the telemetry contact
should be evaluated to help ensure chronic programmer
communication. Close the device pocket using standard
surgical technique. As the final step at device implant and each
patient follow-up, the permanent program should be
retransmitted to the ICD.
Complete the Medical Device Registration Form provided with
the ICD and return it to BIOTRONIK.
90 Lexos Technical Manual
Lexos Technical Manual 91
5. Follow-up Procedures
5.1 General Considerations
An ICD follow-up serves to verify appropriate function of the ICD
system, and to optimize the programmable parameter settings.
In addition to evaluating the patient’s stored therapy history and
electrograms, acute testing of sensing and pacing is
recommended. The physician shall be made aware of the
program that is in effect after the patient leaves the office after
each follow-up, by viewing the parameters displayed on the
programmer screen after the device has been programmed and
interrogated. As the final step at device implant and each patient
follow-up, the permanent program should be retransmitted to the
ICD. Due to longevity concerns, it is recommended the
physician schedule a patient follow-up visit every 3 months.
W
ARNING
Resuscitation Availability - Do not perform induction
testing unless an alternate source of patient defibrillation
such as an external defibrillator is readily available. In order
to implant the ICD system, it is necessary to induce and
convert the patient’s ventricular tachyarrhythmias.
92 Lexos Technical Manual
5.2 Longevity
The service time of an ICD can vary based on several factors,
including the number of charge sequences, programmed
parameters, number of tachyarrhythmias detected, relative
amount of bradycardia pacing required, pacing lead impedance,
storage time, battery properties, and circuit operating
characteristics. Service time is the time from beginning of
service (BOS) to the elective replacement indication (ERI). To
assist the physician in determining the optimum time for ICD
replacement, a replacement indicator is provided that notifies the
user that replacement within a certain period of time is required.
Upon reaching ERI, the battery has at least enough energy left
to continue monitoring for three months along with the ability to
deliver six high-energy shocks. After this period, all
tachyarrhythmia detection and tachyarrhythmia therapy is
disabled.
C
AUTION
Charge Time - When preparing a high energy shock the
charge circuit stops charging the capacitors after 20 seconds,
and delivers the stored energy as shock therapy. After the
device reaches ERI the stored energy may be less than
30 joules per shock.
The service times from beginning of service (BOS) to elective
replacement indication (ERI) are listed below in
estimates assume pacing rate of 50 ppm with a pulse width of
0.5 ms and pulse amplitude of 2.4 volts and 500 ohm pacing
impedance with all shocks at maximum energy (30 joules) at
37C. It is assumed that the shocks are equally spaced
throughout the life of the ICD. The estimates associated with
0% pacing support assume the ICD is sensing an intrinsic sinus
rhythm at a rate of 70 bpm.
Table 8
. All
Lexos Technical Manual 93
Table 8: Longevity Estimates
DDD Pacing
Support
Shocks
Per Year
Years
12 3.5
100 %
4 4.3
1 4.6
0 4.8
12 4.0
50 %
4 5.0
1 5.5
0 5.7
12 4.4
15 %
4 5.6
1 6.3
0 6.6
12 4.6
0 %
4 6.0
1 6.8
0 7.1
Each maximum energy (30 joule), high voltage charging
sequence reduces the longevity of the device by approximately
21 days.
Upon reaching ERI, the battery has enough energy left to
continue monitoring for three months and to deliver six high
energy shocks. The estimates associated with duration of ERI
assume the ICD is sensing an intrinsic sinus rhythm at a rate of
70 bpm. After this period the device is at EOS (End of Service)
and requires explantation. Once at EOS, all tachyarrhythmia
detection and therapy is disabled.
94 Lexos Technical Manual
5.3 Explantation
Explanted ICDs, lead systems, and accessories may not be
reused. Please complete the appropriate out of service (OOS)
form and return it to BIOTRONIK with the explanted devices. All
explanted devices should be sent either to the local BIOTRONIK
representative or the BIOTRONIK home office for expert
disposal. Contact BIOTRONIK if you need assistance with
returning explanted devices. If possible, the explanted devices
should be cleaned with a sodium-hyperchlorine solution of at
least 1% chlorine and then washed with water prior to shipping.
The pulse generator should be explanted before the cremation of
a deceased patient.
W
ARNING
Unwanted Shocks – Always program the VT/VF Detection
and Therapy status to DISABLED prior to handling the
device to prevent the delivery of serious shocks to the
patient or the person handling the device during the implant
procedure.
C
AUTION
Device Incineration – Never incinerate the ICD due to the
potential for explosion. The ICD must be explanted prior to
cremation.
Explanted Devices – Return all explanted devices to
BIOTRONIK.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.