BIOTRONIK Eluna 8, Eluna 8 SR, Eluna 8 DR-T, Eluna 8 HF-T, Eluna 8 SR-T Technical Manual

...
Page 1
BIOTRONIK SE & Co. KG Woermannkehre 1 12359 Berlin · Germany Tel +49 (0) 30 68905-0 Fax +49 (0) 30 6852804 sales@biotronik.com www.biotronik.com
SR(-T), DR(-T), HF-T (Pr oMRI
®
)
Pacemaker | Bradyarrhythmia Therapy | Cardiac Resynchronization Therapy
Technical Manual
403587
Revision: F (2014-08-04)
© BIOTRONIK SE & Co. KG
All rights reserved. Specification subject to modification, revision and improvement.
® Registered trademarks of BIOTRONIK SE & Co. KG:
BIOTRONIK Home Monitoring, Eluna, ProMRI
Index 403587Technical ManualEluna SR(-T), DR(-T), HF(-T)
0123 0681 2014
Page 2
2 Table of Contents
Table of Contents
Table of Contents
Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Intended Medical Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Indications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Diagnostic and Therapy Functions . . . . . . . . . . . . . . . . . . . . . 8
General Safety Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Operating Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Possible Complications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Possible Risks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Implantation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Implantation Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Precautionary Measures while Programming. . . . . . . . . . . 17
Magnet Response. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Follow-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Patient Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Replacement Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Explantation and Device Replacement . . . . . . . . . . . . . . . . . 22
Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Timing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Pacing and Sensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Rate Adaptation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
MRI Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Preset Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Tolerances of Parameter Values . . . . . . . . . . . . . . . . . . . . . 30
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Mechanical Characteristics. . . . . . . . . . . . . . . . . . . . . . . . . . 31
Electrical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Battery Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Legend for the Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Page 3
3 Product Description
1 Product Description
Product Description1403587Technical ManualEluna SR(-T), DR(-T), HF(-T)
Intended Medical Use
Intended use Eluna is a family of implantable pacemakers that may be implanted for all brady-
cardia arrhythmia indications. The primary objective of the therapy consists of improving patients' symptoms that can be clinically manifested. The implantation of the pacemaker is a symptomatic therapy with the following objective:
• Compensation of bradycardia by atrial, ventricular, or AV sequential pacing
• Additional triple-chamber features: Resynchronization of ventricular chamber contraction via biventricular pacing
Diagnosis and
therapy forms
The cardiac rhythm is automatically monitored and bradycardia arrhythmias are treated. All major therapeutic approaches from the field of cardiology and electro-
physiology are unified in this pacemaker family. BIOTRONIK Home Monitoring
®
enables physicians to perform therapy management at any time.
Required expertise In addition to having basic medical knowledge, the user must be thoroughly familiar
with the operation of a device system.
• Only qualified medical specialists having the special knowledge required for the proper use of implanted devices are permitted to use them.
• If users do not possess this knowledge, they must be trained accordingly.
Indications
Guidelines of
cardiological societies
Generally approved differential diagnostic methods, indications, and recommenda­tions for pacemaker therapy apply to BIOTRONIK devices.
The guidelines provided by cardiology associations offer decisive information:
• We recommend observing the indications published by the German Cardiac Society (Deutsche Gesellschaft für Kardiologie, Herz- und Kreislaufforschung) and the ESC (European Society of Cardiology).
• This also applies to the guidelines published by the Heart Rhythm Society (HRS), the American College of Cardiology (ACC), the American Heart Association (AHA), and other national cardiology associations.
Device types For the following symptoms/expectations, the following device types are indicated:
Symptom/expectation SR DR HF
Disorientation due to bradycardia x x x Presyncope x x x Benefit from resynchronization of the right and left
ventricles
x
Syncope x x x
Page 4
4 Product Description
Pacing modes For the following symptomatic, the following pacing modes are indicated:
MR conditional
ProMRI® labeled MRI conditional pacemakers are safe for use in the MRI environ­ment when used in conjunction with a complete MRI conditional pacing system and according to the instructions given in the ProMRI® manual.
Contraindications
Guidelines No contraindications are known for the implantation of multifunctional single-
chamber, dual-chamber or triple-chamber devices, provided differential diagnos­tics precedes implantation according to the appropriate guidelines and no modes or parameter combinations are configured that pose a risk to the patient.
Pacing modes
and parameters
The compatibility and effectiveness of parameter combinations must be checked and, as the case may be, adapted after programming.
Symptom/expectation Pacing mode
Sick sinus syndrome Dual-chamber pacing Chronic, symptomatic second and third-degree AV block Dual-chamber pacing Adams-Stokes syndrome Dual-chamber pacing Symptomatic bilateral bundle branch block when tach-
yarrhythmia and other causes have been ruled out
Dual-chamber pacing
• Chronotropic incompetence
• Benefit from increased pacing rate with physical activity
R mode or CLS
Sinus node dysfunction in the presence of normal AV and intraventricular conduction
Atrial pacing
Bradycardia in conjunction with the following:
• Normal sinus rhythms with only rare episodes of AV block or sinus arrest
• Chronic atrial fibrillation
• Severe physical disability
Ventricular pacing
Set of facts Contraindicated pacing mode
Additionally implanted ICD Unipolar pacing
Set of facts Inappropriate pacing mode
Chronic atrial tachycardia, chronic atrial fibrillation or flutter
Atrial-controlled modes (DDD, VDD, AAI)
Poor tolerance of pacing rates above the basic rate, e.g., angina pectoris
AV conduction disorder Atrial single-chamber pacing Failing AV conduction
Set of facts Adapt parameters
Slow retrograde conduction after ventricular pacing: Risk of pacemaker­mediated tachycardia
• Extend atrial refractory period (ARP) and/or:
• Shorten AV delay
•Rarely: Program to DDI, DVI or VVI
Poor tolerance of pacing rates above the basic rate, e.g., angina pectoris
• Lower atrial upper rate
• Lower maximum sensor rate
• Deploy atrial overdrive pacing
Page 5
5 Product Description
System Overview
Note: Use only adapters approved by BIOTRONIK for leads with different connec-
tions.
• If you have any questions concerning the compatibility of other manufacturers' leads, please contact BIOTRONIK.
Device family This device family consists of single-chamber, dual-chamber and triple-chamber
devices with or without Home Monitoring. Not all device types are available in every country.
The following device variants are available:
Device The device's housing is made of biocompatible titanium, welded from the outside
and therefore hermetically sealed. The ellipsoid shape facilitates ingrowth into the pectoral muscle area. The housing serves as an antipole in the case of unipolar lead configuration.
IS-1 lead connection The device labeling provides information pertaining to the connection assignment:
Leads BIOTRONIK leads are sheathed in biocompatible silicone. They can be flexibly
maneuvered, are stable long-term, and are equipped for active or passive fixation. They are implanted using a lead introducer set. Some leads are coated with poly­urethane to increase the gliding properties of the lead. Leads with steroids reduce inflammatory processes. The fractal design of the leads allows for low pacing thresholds, high pacing impedance, and a low risk of oversensing.
BIOTRONIK provides adapters to connect already implanted leads to new devices.
Telemetry Telemetric communication between the device and the programmer can be carried
out following initialization either by applying the programming head (PGH) to the device or by using radio frequency (RF) telemetry in the programmer. BIOTRONIK
calls this function SafeSync
®
.
Device type Variant with
Home Monitoring
Variant without Home Monitoring
Single-chamber Eluna 8 SR-T Eluna 8 SR Dual-chamber Eluna 8 DR-T Eluna 8 DR Triple-chamber Eluna 8 HF-T
SR DR HF
Connector port
Lead connector
Configuration Implantation site Device type
RA IS-1 Unipolar, bipolar Atrium DR, HF RV IS-1 Unipolar, bipolar Right ventricle SR, DR, HF LV IS-1 Unipolar, bipolar Left ventricle HF
VVIR /AAIR
IS-1
DDDR
A
IS-1
DDDR
A
IS-1
V
LV
RV
Page 6
Note: Home Monitoring is possible in all modes.
6 Product Description
Programmer Implantation and follow-up are performed with BIOTRONIK's portable
programmer: There are programmers with integrated or external SafeSync Module for RF telemetry.
The programmer is used during implantation to transfer the current device program to the device. The pacing thresholds can be determined and all tests can be performed during in-office follow-up. In addition to this, the programmer is used to set mode and parameter combinations, as well as for interrogation and saving of data from the device. Leadless ECG, IEGM, markers and functions are displayed simultaneously on the color display.
Modes The mode setting depends on the individual diagnosis:
NBG codes AAIR or VVIR is the NBG code for the antibradycardia mode of the single-chamber
device:
DDDR is the NBG code for the antibradycardia mode of the dual-chamber device:
Device type Modes Standard
SR(-T) • VVI-CLS (8 series only)
• VVIR; V00R; AAIR; A00R
• VVI; VVT; V00; AAI; AAT; A00
•OFF
VVIR
DR(-T) • VVI-CLS; DDD-CLS (8 series only)
• DDDR; DDIR; DVIR; D00R VDDR; VDIR; VVIR; V00R
AAIR; A00R
• DDD; DDT; DDI; DVI; D00 VDD; VDI; VVI; VVT; V00
AAI; AAT; A00
• DDD-ADI; DDDR-ADIR
•OFF
DDDR
HF-T • DDD-CLS; VVI-CLS
• DDDR; DDIR; DVIR; D00R VDDR; VDIR; VVIR; V00R
AAIR; A00R
• DDD; DDT; DDI; DVI; D00 VDD; VDI; VVI; VVT; V00
AAI; AAT; A00
• DDD-ADI; DDDR-ADIR
•OFF
DDDR
A/V Pacing in the atrium or ventricle A/V Sensing in the atrium or ventricle I Pulse inhibition in the atrium and ventricle R Rate adaptation
D Pacing in the atrium and ventricle D Sensing in the atrium and ventricle D Pulse inhibition and pulse triggering R Rate adaptation
Page 7
7 Product Description
DDDRV is the NBG code for the antibradycardia mode of the triple-chamber device:
BIOTRONIK
Home Monitoring
®
In addition to effective pacing therapy, BIOTRONIK provides a complete therapy management system:
• With Home Monitoring, diagnostic and therapeutic information and technical data are automatically sent to a stationary or mobile transmitter via an antenna in the device header. The data are encrypted and sent from the transmitter to the BIOTRONIK Service Center via the cellular phone network.
• The received data are deciphered and evaluated. Each physician can set the criteria for evaluation to be used for each patient and can configure the time of notification via e-mail, SMS or fax.
• A clear overview of the results of this analysis is displayed for the attending physicians on the protected internet pl
atform Home Monitoring Service Center
(HMSC).
• Data transmission from the device is performed with a daily device message.
• Device messages which indicate special events in the patient's heart or in the device are forwarded immediately.
• A test message can be initiated at any time using the programmer to immedi­ately check the Home Monitoring function.
Order numbers Eluna The devices can be obtained as follows:
Package contents The storage package includes the following:
• Sterile packaging with device
• Serial number label
• Patient ID card
• Warranty booklet
Note: The technical manual pertaining to the device is either included in hard copy form in the storage package or in digital form on the internet.
The sterile container includes the following:
•Device
• Screwdriver
D Pacing in the atrium and ventricle D Sensing in the atrium and ventricle D Pulse inhibition and pulse triggering R Rate adaptation V Multisite pacing in both ventricles
Device type Eluna 8 Eluna 8 ProMRI
SR 394939 394972 SR-T 394934 394971 DR 394927 394970 DR-T 394929 394969 HF-T 394917 394968
Page 8
8 Product Description
Diagnostic and Therapy Functions
General overview All the systems have extensive features that allow quick diagnosis and delivery of
safe therapy for bradycardia conditions.
• Automatic functions make it easy and fast to implant, configure, and check the pacemaker.
• Auto-initialization after implantation: The device recognizes the implanted leads autonomously and sets the polarity. The automatic functions of the software are activated after 10
min.
Diagnostic functions • Data from the last 10 interrogations and follow-ups are recorded as well as
arrhythmia episodes; they are stored together with other data to assess patients and the state of the device at any time.
• Continuous automatic below-threshold impedance measurements are performed in the device independent of the pacing pulse in order to check the lead for proper functioning.
• Once a telemetry connection has been established during a test procedure in an in-office follow-up, the IEGM is displayed with markers.
Antibradycardia pacing • Sensing: The amplitudes of the P and R waves are measured in the implanted
device fully automatically to record varying amplitudes. The sensitivity for the atrium and ventricle is adapted automatically on an ongoing basis. The measurement data are averaged and the trend can be displayed.
• Thresholds: atrial as well as ventricular pacing thresholds are automatically determined in the device. Capture control is used to set the pulse amplitudes so that pacing is performed with the optimum atrial and ventricular amplitude for the patients with each change of the pacing threshold.
• Timing: Pacing in the atrium is checked particularly carefully by an automatic adaptation of the atrial refractory period in order to avoid pacemaker-mediated tachycardia.(Auto PVARP function: the postventricular atrial refractory period is adapted automatically).
• Additional, special form of rate adaptation with devices from the 8 series: an increased cardiac output requirement is detected using physiological imped
­ance measurement. The measuring principle is based on contractile changes (ionotropy) of the myocardium (CLS function: Closed Loop Stimulation). Rate adaptation is automatically initialized and optimized in CLS mode.
• Ventricular pacing suppression with devices from the 8 series: unnecessary ventricular pacing is avoided by promoting intrinsic conduction (Vp suppression function). The device can adapt itself to conduction changes. In the case of intrinsic conduction, the device switches to a DDD(R)-ADI(R) mode.
Resynchronisation therapy Triple chamber devices have functions to configure different VV delays in order to
resynchronize the ventricles.
• Capture Control is available for the left ventricle with automated tracking of the pacing threshold or automatic threshold monitoring (ATM) for trend analysis.
• To ensure that no additional surgery is necessary in case of a left-sided increase of pacing threshold or undesired phrenic nerve stimulation, different pacing polarities can be set for the left ventricular lead with a triple-chamber device.
• An additional diagnostic function with biventricular pacing: variability of the heart rate, the patient activity and the thoracic impedance are monitored on a continual basis.
Page 9
9 Product Description
Programs There are two types of therapy programs:
• Default parameters are offered for the most common indications (Program Consult function).
• Individual settings can be saved in 3 individual therapy programs.
Home Monitoring functions The device automatically sends information to the transmitter once a day. Addition-
ally, the test messages can be initiated using the programmer. Important medical information includes, among others, the following:
• Ongoing atrial and ventricular arrhythmia
• Parameters relevant to leads in the atrium and ventricle: thresholds, sensing amplitudes, impedances
• Current statistics on bradycardia therapy
• Individually adjustable timing interval for device messages which provide addi­tional information pertaining to the device messages
• IEGM online HD® with up to 3 high definition channels
• Transmission of these IEGM recordings along with device messages
Page 10
10 General Safety Instructions
2 General Safety Instructions
General Safety Instructions2403587Technical ManualEluna SR(-T), DR(-T), HF(-T)
Operating Conditions
Technical manuals The following technical manuals provide information about usage of the device
systems:
— Technical manual for the device
— Technical manual for the HMSC
— Technical manuals for the programmer and its accessories
— Technical manuals for the user interface
— Technical manuals for cables, adapters and accessories
• Technical manuals are either included in hard copy form in the storage package or in digital form on the internet:
manuals.biotronik.com
• Follow all relevant technical manuals.
• Preserve technical manuals for later use.
Care during shipping and
storage
• Devices are not to be stored close to magnets or sources of electromagnetic interference.
• Note the effects of the storage duration; see Battery Data.
Temperature Extremely low and high temperatures affect the service time of the battery in the
device.
• The following temperatures are permitted for transport, storage, and use: –10°C to 45°C
Sterile delivery The device and the screwdriver have been gas-sterilized. Sterility is guaranteed only
if the blister and quality control seal have not been damaged.
Sterile packaging The device and screwdriver are each packaged in 2 separately sealed blisters. The
inner blister is also sterile on the outside so that it can be transferred in a sterile state during implantation.
Single use only The device and screwdriver are intended for single use only.
• Do not use the device if the package is damaged.
• The device must not be resterilized and reused.
Page 11
11 General Safety Instructions
Possible Complications
General information on
medical complications
Complications for patients and device systems generally recognized among practi­tioners also apply to BIOTRONIK devices.
• Normal complications may include fluid accumulation within the device pocket, infections, or tissue reactions. Primary sources of complication information include current scientific and technological knowledge.
• It is not possible to guarantee the efficacy of antitachycardia therapy, even if the programs have proven successful during tests or subsequent electrophysiolog
­ical examinations. In rare cases the set parameters may become ineffective. In particular it is inevitable that tachyarrhythmias may be induced.
Skeletal myopotentials Bipolar sensing and control of sensitivity are adapted by the device to the rate range
of intrinsic events so that skeletal myopotentials are usually not sensed. Skeletal myopotentials can nonetheless be classified as intrinsic events especially with a unipolar configuration and/or very high sensitivity and, depending on the interfer­ence, may cause inhibition or antiarrhythmia therapy.
Nerve and muscle
stimulation
A device system consisting of a unipolar lead and an uncoated device may result in undesirable pacing of the diaphragm in the case of an initial or permanent high setting of the pulse amplitude.
Possible technical failures Technical failure of a device system cannot be entirely ruled out. Possible causes
may include the following:
• Lead dislodgement
• Lead fracture
• Insulation defects
• Device component failures
• Battery depletion
Electromagnetic
interference (EMI)
Any device can be sensitive to interference, for example, when external signals are sensed as intrinsic rhythm or if measurements prevent rate adaptation.
• BIOTRONIK devices have been designed so that their susceptibility to EMI is minimal.
• Due to the intensity and variety of EMI, there is no guarantee for safety. It is generally assumed that EMI produces only minor symptoms in patients - if any.
• Depending on the pacing mode and the type of interference, sources of interfer­ence may lead to pulse inhibition or triggering, an increase in the sensor-depen­dent pacing rate or asynchronous pacing.
• Under unfavorable conditions, for example during diagnostic or therapeutic procedures, interference sources may induce such a high level of energy into the pacing system that the cardiac tissue surrounding the lead tip is damaged.
Device behavior
in case of EMI
In the case of electromagnetic interference or undesired myopotentials, the device switches to asynchronous pacing for the duration of the time that the interference rate is exceeded.
Static magnetic fields The reed switch in the pacemaker starts to close at a field strength of 1.5 mT.
Page 12
12 General Safety Instructions
Possible Risks
Procedures to avoid The following procedures must be avoided as they may cause harm to the patient or
damage the device and, as a result, put the system functionality at risk:
• Therapeutic ultrasound
• Transcutaneous electrical nerve stimulation
• Hyperbaric oxygen therapy
• Applied pressures higher than normal pressure
Potentially risky therapeutic
and diagnostic procedures
If electrical current from an external source is conducted through the body for diag­nostic or therapeutic purposes, then the device can be subjected to interference and the patient placed at risk.
Arrhythmia or ventricular fibrillation can be induced during diathermic procedures such as electrocautery, HF ablation or HF surgery. For example, damaging pres­sure levels may arise during lithotripsy. Influences on the device are not always immediately clear.
If potentially risky procedures cannot be avoided, the following should be observed at all times:
• Electrically insulate patients.
• Switch the pacemaker function to asynchronous modes if needed.
• Do not introduce energy near the device system.
• Check the peripheral pulse of the patient.
• Monitor the patient during and after every intervention.
External defibrillation The device is protected against the energy that is normally induced by external defi-
brillation. Nevertheless, any implanted device may be damaged by external defibril­lation. Specifically, the current induced in the implanted leads may result in necrotic tissue formation close to the electrode/tissue interface. As a result, sensing prop­erties and pacing thresholds may change.
• Place adhesive electrodes anterior-posterior or perpendicular to the axis formed by the device to the heart at least 10 cm away from the device and from implanted leads.
Radiation therapy The use of radiation therapy must be avoided due to possible damage to the device
and the resulting impaired functional safety. If this type of therapy is to be used anyway, prior risk/benefit analysis is absolutely necessary. The complexity of influ­encing factors such as different sources of radiation, a variety of devices and therapy conditions makes it impossible to issue directives that guarantee radiation therapy without an impact on the device. The EN 45502 standard pertaining to active implantable medical devices requires the following measures during the adminis­tration of therapeutic ionizing radiation:
• Adhere to instructions for potentially risky therapeutic and diagnostic proce­dures.
• Shield device against radiation.
• After applying radiation, double-check the device system to make sure it is func­tioning properly.
Note: Please contact BIOTRONIK with questions on the risk/benefit analysis.
Page 13
13 General Safety Instructions
Magnetic resonance
imaging
Magnetic resonance imaging must be avoided due to the associated high frequency fields and magnetic flux density: Damage or destruction of the device system by strong magnetic interaction and damage to the patient by excessive warming of the body tissue in the area surrounding the device system.
Under certain conditions and when maintaining mandatory measures, magnetic resonance imaging can be performed to protect patient and device system. BIOTRONIK devices with the "MR conditional function bear the identification ProMRI
®
.
•The ProMRI® manual – MR conditional device systems – contains detailed infor­mation on safely conducting an MRI.
— Download the digital manual from the web site:
manuals.biotronik.com
— Order the printed manual at BIOTRONIK.
• Does approval as "MR-Conditional" apply in your country or region? Ask for current information at BIOTRONIK.
Page 14
14 Implantation
3 Implantation
Implantation3403587Technical ManualEluna SR(-T), DR(-T), HF(-T)
Implantation Procedure
Having parts ready The following parts that correspond to the requirements of the EC Directive 90/385/
EEC are required:
• Device with screwdriver from BIOTRONIK
• BIOTRONIK leads and lead introducer set — Single-chamber device: unipolar or bipolar lead for the right ventricle
— Double-chamber device: one unipolar or bipolar lead each for the atrium
and for the right ventricle
— Triple-chamber device: an additional unipolar or bipolar LV lead
• Approved connections are IS-1: Use only adapters approved by BIOTRONIK for leads with different connections or leads from other manufacturers.
• BIOTRONIK programmer (with integrated SafeSync RF telemetry or with sepa­rate SafeSync Module) and approved cables
• External multi-channel ECG device
• Keep spare parts for all sterile components.
Keeping an external
defibrillator ready
In order to be able to respond to unforeseeable emergencies or possible technical failures of the device:
• Keep an external defibrillator and paddles or patch electrodes ready.
!
!
WARNING Inadequate therapy due to defective device
If an unpacked device is dropped on a hard surface during handling, electronic parts could be damaged.
• Use a replacement device.
• Return the damaged device to BIOTRONIK.
Unpacking the device
• Peel the sealing paper off of the outer blister at the marked position in the direc­tion indicated by the arrow. The inner blister must not come into contact with persons who have not sterilized their hands or gloves, nor with non-sterile instruments!
• Use the gripping tab on the inner plastic container to remove it from the outer plastic container.
• Peel the sealing paper off of the sterile inner blister at the marked position in the direction indicated by the arrow.
Note: The device is disabled on delivery and can be implanted immediately after unpack
ing without manual activation.
Checking parts Damage to any of the parts can result in complications or technical failures.
• Check for damage before and after unpacking all parts.
• Replace damaged parts.
Page 15
Note: If necessary, the device can also be programmed before or during auto­initialization.
15 Implantation
Implantation site In general, the pacemaker is implanted on the right side subcutaneously or subpec-
torally, depending on the lead configuration as well as the anatomy of the patient.
Overview: Implanting
Avoid damage to the header Set screws must be tightened or loosened with care.
• Loosen set screws with the supplied screwdriver. Use only BIOTRONIK screw­drivers with torque control!
• If lead revision is necessary, re-order sterile screwdrivers from BIOTRONIK.
!
!
WARNING Short circuit due to open lead connector ports
Connector ports in the header which are open and thus not electrolyte-proof may cause undesired current flows to the body and penetration of body fluid into the device.
• Close unused connections with IS-1 blind plugs.
!
!
WARNING Inadequate therapy
Insufficient lead spacing or inappropriate lead positioning may lead to far field sensing.
• Tip and ring electrodes must not have contact with each other.
Preventing short circuits
in the header
Keeping distance
between leads
1 Shape the device pocket and prepare the vein. 2 Implant the leads and perform measurements. 3 Connect device and leads.
The device starts auto-initialization on its own. 4 Insert the device. 5 Guide the fixation suture through the opening in the header and fixate the
device in the prepared device pocket. 6 Close the device pocket. 7 Prior to testing and configuration, wait for the successful completion of auto-
matic device initialization.
Page 16
16 Implantation
Connecting the lead
connector to the device
Applying the programming
head
The programming head (PGH) features a diagram of the device. This is used to assist in positioning the head to ensure proper telemetry.
• Make sure the PGH is positioned correctly.
Establishing telemetry
contact
The programmer (or the SafeSync Module) can be no less than 20 cm and no more than 3 m from the device; ideally there should be no hindrances between the patient and the programmer.
• Switch on RF telemetry on the programmer.
• Apply the programming head for about 2 s until successful initialization is displayed on the programmer:
The SafeSync symbol is displayed in the navigator and the signal strength is displayed in the status line.
• Remove the programming head.
Note: After auto-initialization, all parameters are activated as in the factory settings.
Auto-initialization Auto-initialization begins automatically once the first connected lead is sensed.
Auto-initialization is terminated 10 minutes after connection of the first lead. If no other program has been transferred in the meantime, the device subsequently works with active automatic functions in the factory settings or with the preset program of the user.
Manual setting of the lead polarity or measurement of lead impedances is not necessary.
Behavior during
auto-initialization
• During transmission of a permanent program: Auto-initialization is terminated and the transferred program is active.
• During testing: Tests cannot be performed during auto-initialization; stop it beforehand. Auto-
initialization will not be continued upon completion of the test.
1 Disconnect stylets and stylet guides. 2 • Connect the unipolar or bipolar IS-1 lead connector ventricle to RV.
• Connect the unipolar or bipolar IS-1 lead connector atrium to A.
• Connect the unipolar or bipolar IS-1 lead connector ventricle to LV.
3 Push the lead connector into the header without bending the conductor until
the connector tip becomes visible behind the set screw block.
4 If the lead connector cannot be inserted completely, the set screw may be
protruding into the drill hole of the set screw block. Carefully loosen the set screw without completely unscrewing it, so that it does not become tilted upon retightening.
5 Use the screwdriver to perpendicularly pierce through the slitting in the
center of the silicone plug until it reaches the set screw.
6 Turn the set screw clockwise until the torque control starts (you will hear a
clicking sound).
7 Carefully withdraw the screwdriver without retracting the set screw.
• When the screwdriver is withdrawn, the silicone plug automatically seals the lead connection safely.
Page 17
17 Implantation
Precautionary Measures while Programming
Checking the device system • After auto-initialization, perform a follow-up to see if the device system is func-
tioning properly.
• Perform a pacing threshold test to determine the pacing threshold.
Performing standard tests and monitoring the patient
Critical conditions can occur for the patient even during standard tests due to inad­equate parameter settings or interrupted telemetry.
• Ensure sufficient patient care even during tests.
• After the threshold test, check to determine whether the threshold is clinically and technically justifiable.
• Continuously monitor the ECG and the patient's condition.
• Cancel testing if necessary.
Do not interrupt telemetry
during a treatment.
Disconnecting the SafeSync Module from the programmer can result in interfer­ence with or termination of the SafeSync wandless telemetry.
• Do not disconnect the SafeSync Module from the programmer.
• Do not take the Operation Module off the ICS 3000.
Cancelling telemetry Programmer interference or interrupted telemetry during performance of tempo-
rary programs (follow-up tests) can result in inadequate pacing of the patient. This is the case if the programmer can no longer be operated due to a program error or a defective touch screen and therefore the temporary program cannot be termi­nated. Under these circumstances, it is helpful to cancel telemetry, in which case the device automatically switches to the permanent program.
• In the case of telemetry with PGH: lift the programming head by at least 30 cm.
• In the case of RF telemetry: switch off and reposition the programmer.
• Turn off possible sources of interference.
Avoiding critical
parameter settings
No modes and parameter combinations that pose a risk to the patient should be set.
• Prior to setting rate adaptation, determine the patient's capacity for strain.
• Check compatibility and effectiveness of parameter combinations after making settings.
Manually setting
lead polarity
Due to the risk of an entrance/exit block, bipolar lead polarity (sensing/pacing) should only be set if bipolar leads are implanted.
Setting triggered mode Triggered modes perform pacing regardless of intrinsic cardiac events. To prevent
undersensing due to electromagnetic interference in special cases, a triggered mode can be displayed.
Setting sensing Manually set parameters can be unsafe. For example, unsuitable far-field protec-
tion may impede sensing of intrinsic pulses.
• Use automatic sensitivity control.
Setting the sensitivity A value set to < 2.5 mV/unipolar for device sensitivity may result in noise caused by
electromagnetic fields.
• Therefore, it is recommended that a value of ≥ 2.5 mV/unipolar be set according to paragraph 28.22.1 of the EN
45502-2-1 standard. Setting sensitivity values
<
2.5 mV/unipolar requires explicit clinical need. Values like this can only be set
and retained with physician supervision.
Page 18
18 Implantation
Preventing device-induced
complications
BIOTRONIK devices are equipped with several functions to prevent device-induced complications to the greatest extent possible:
• Measure the retrograde conduction time.
• If the function is not yet automatically set: activate PMT protection.
• Set the VA criterion.
Preventing conduction
of atrial tachycardia
BIOTRONIK devices are equipped with several functions to prevent conduction of atrial tachycardia to the ventricle(s):
•Set Mode Switching for indicated patients.
• Set the upper rate and the refractory periods to prevent abrupt ventricular rate switching.
• Prefer Wenckebach response and avoid 2:1 behavior.
• Set all parameters so as to prevent constant changing between atrial and ventricular-controlled modes.
Phrenic nerve stimulation that cannot be terminated
With LV pacing, chronic phrenic nerve stimulation can in rare cases not be termi­nated by reprogramming the available left ventricular pacing configurations or by other measures.
• Possibly set a right ventricular mode both in the permanent program and for Mode
Switching.
Avoiding risks in the case
of exclusive LV pacing
Lead dislodgement in the case of exclusive left ventricular pacing could pose the following risks: loss of ventricular pacing as well as induction of atrial arrhythmia.
• Consider sensing and pacing parameters with reference to loss of therapy.
• Exclusive LV pacing is not recommended for patients who depend on the device.
• Take possible interruption of automatic Active Capture Control into consider­ation.
• In the case of follow-ups and threshold tests, take loss of synchronized ventric­ular pacing into consideration.
• Mode Switching does not allow exclusive LV pacing; consider the consequences when setting Mode Switching parameters.
If an ICD is implanted at
the same time, do not
permit unipolar pacing
If an ICD is implanted in addition to a pacemaker and a lead failure occurs, it is possible to switch to unipolar pacing after resetting the pacemaker or using the automatic lead check. As a result, the ICD could falsely inhibit or trigger tachyar­rhythmia therapy activity.
• Unipolar leads are not permitted in this configuration.
Recognizing lead failure Automatic impedance measurement is always switched on.
• Impedance values that indicate technical failure of a lead are documented in the event list.
Consider power
consumption and service
time
The pacemaker permits programming of high pulse amplitudes with long pulse widths at high rates to be able to adequately treat even rare diagnoses. In combina­tion with low lead impedance, this results in a very high level of power consumption.
• When programming large parameter values, take into account that the battery depletion indicator ERI will be reached very early because the service time of the battery may be reduced to less than 1
year.
• Home Monitoring ON reduces the service time by 3 months approximately.
RF telemetry requires somewhat more power: More frequent use of RF telemetry than assumed during service time calculation (20 min per year) reduces the service time by about 7 days for the SR(-T), 6 days for the DR(-T), and 5 days for the HF-T device.
• Do not establish unnecessary RF telemetry.
•After 5 minutes without input, SafeSync switches to the economy mode.
• Check the battery capacity of the device at regular intervals.
Page 19
19 Implantation
Magnet Response
Programming head
application
When the programming head is applied, time remains for device interrogation before the device switches back to the previously set permanent therapy mode. The same applies to programming head application to establish RF telemetry contact.
Magnet response in
standard program
Applying a magnet or the programming head can result in an unphysiological rhythm change and asynchronous pacing. The magnet response is set as follows in the standard program of BIOTRONIK pacemakers:
• Asynchronous: For the duration of the magnet application – mode D00 (possibly V00 / A00)
without rate adaptation; Magnet rate: 90 bpm
•Automatic: For 10 cycles – mode D00, subsequently mode DDD without rate adapta
tion;
Magnet rate: 10 cycles with 90 bpm, subsequently set basic rate
• Synchronous:
Note: See also the replacement indication information.
Mode DDD (where applicable: VVI) without rate adaptation; Magnet rate: set basic rate
Magnet response with ERI After reaching ERI, pacing is performed as follows after applying the magnet or
programming head:
Magnet application
by patients
If patients are performing their own magnet application, the synchronous magnet response has to have been programmed. Patients should also know the following:
• When may the magnet be used? In cases of severe dizziness and indisposition.
• How long is the magnet placed on the pacemaker? 1 to 2 s.
• What happens when the magnet is applied? The IEGM of the last 10 seconds is stored.
• What has to happen after magnet application? The patient has to contact the physician for a follow-up.
Magnet response
Cycles 1 to 10 After 10th cyle
Automatic Asynchronous with 80 bpm Synchronous with basic rate
reduced by 4.5 to 11% Asynchronous Asynchronous with 80 bpm Asynchronous with 80 bpm Synchronous Synchronous with basic rate
reduced by 4.5 to 11%
Synchronous with basic rate
reduced by 4.5 to 11%
Page 20
20 Implantation
Follow-up
Follow-up intervals Follow-ups must be performed at regular, agreed intervals.
• Following the lead ingrowth phase, approximately 3 months after implantation, the first follow-up should be carried out by the physician using the programmer (in-office follow-up).
• The next in-office follow-up should be carried out once a year and no later than 12 months after the last in-office follow-up.
Follow-up with BIOTRONIK
Home Monitoring
®
Monitoring using the Home Monitoring function does not serve to replace regular in-office appointments with the physician required for other medical reasons.
Follow-up supported by Home Monitoring can be used to functionally replace in­office follow-up under the following conditions:
• The patient was informed that the physician must be contacted if symptoms worsen or if new symptoms arise despite the use of the Home
Monitoring func-
tion.
• Device messages are transmitted regularly.
• The physician decides whether the data transmitted via Home Monitoring with regard to the patient's clinical condition as well as the technical state of the device system are sufficient. If not, an in-office follow-up has to be carried out.
Possible early detection due to information gained via Home Monitoring may neces­sitate an additional in-office follow-up. For example, the data may indicate at an early stage lead problems or a foreseeable end of service time (ERI). Furthermore, the data could provide indications of previously unrecognized arrhythmias or modi­fication of therapy by reprogramming the device.
Follow-up with the
programmer
Use the following procedure for in-house follow-up:
Patient Information
Patient ID card A patient ID card is included in delivery.
• Provide the patient with the patient ID card.
• Request that patients contact the physician in case of uncertainties.
Prohibitive signs Premises with prohibitive signs must be avoided.
• Draw the patient's attention to prohibitory signs.
Possible sources of
interference
Electromagnetic interference should be avoided in daily activities. Sources of inter­ference should not be brought into close proximity with the device.
• Draw the patient's attention to special household appliances, security check­points, anti-theft alarm systems, strong electromagnetic fields, cell phones, and transmitters among other things.
• Request patients to do the following: — Use cell phones on the opposite side of their body from the device.
— Keep the cell phone at least 15 cm away from the device both during use and
when stowing.
1 Record and evaluate the ECG. 2 Interrogate the device. 3 Evaluate the status and automatically measured follow-up data. 4 Check the sensing and pacing functions. 5 Possibly evaluate statistics and IEGM recordings. 6 Manually perform standard tests if necessary. 7 Possibly customize program functions and parameters. 8 Transmit the program permanently to the device. 9 Print and document follow-up data (print report).
10 Finish the follow-up for this patient.
Page 21
21 Implantation
Replacement Indications
Possible battery levels The time span from the beginning of service (BOS) to elective replacement indica-
tion (ERI) is determined by, among others, the following:
• Battery capacity
• Lead impedance
• Pacing program
• Pacing to inhibition ratio
• Pacemaker circuit properties
The following are the defined pacemaker operational statuses:
ERI activation ERI detection is automatically activated after the following events:
• Successful auto-initialization
• Storage for longer than 24 months
ERI display ERI is displayed as follows:
• On the programmer after interrogation of the pacemaker
• By a defined decrease in the basic rate as well as the magnet rate
Change of the mode
with ERI
This change depends on the mode which is set. It is displayed on the programmer.
• Single-chamber modes: VVI
• Double-chamber modes: VDD
• Triple-chamber modes: Double-chamber pacing, one biventricular setting is kept
Deactivated functions
with ERI
The following functions are deactivated:
• Atrial pacing
• Night program
• Rate adaptation
• Atrial and ventricular capture control
• Rate fading
• Atrial overdrive pacing
• IEGM recordings
• Statistics
• Home Monitoring
• Rate hysteresis
• Ventricular pacing suppression
Rate decrease The decrease of basic rate and magnet rate is defined as follows:
• In the following modes, the pacing rate decreases by 11%: DDD(R); DDT(R); D00(R); VDD(R); VDI(R); VVI(R); VVT(R); AAI(R); AAT(R); A00(R)
• In the modes DDI(R) and DVI(R), only the VA interval is extended by 11%. This reduces the pacing rate by 4.5 to 11%, depending on the configured AV delay.
BOS Beginning of service Battery is in good condition; normal
follow-up.
ERI Elective replacement
indication
The replacement time has been reached. The pacemaker must be replaced.
EOS End of service End of service time with regular pace-
maker activity
Page 22
22 Implantation
Expected service time
after ERI
• The information is based on a lead impedance of 500 Ω at 100% pacing and the data of the battery manufacturer.
• Parameter with high pacing energy: 110 bpm; 4.6 V; 1.5 ms; 500 Ω
• Parameters with low pacing energy: 30 bpm; 0.2 V; 0.1 ms; 500 Ω
• Interval between ERI and EOS for the single-chamber device in AAI(R)/VVI(R) mode, for the double and triple chamber device in DDD(R) mode, in standard program and with both high and low pacing energy:
— Mean value: 8 months — Minimum value: 6 months
Explantation and Device Replacement
Explantation • Disconnect the leads from the header.
• Remove the device and, if necessary, leads using state-of-the-art technology.
• Explants are biologically contaminated and must be disposed of safely due to risk of infection.
Device replacement The following applies to leads from a previous device that are intended for further
use:
• Check the leads prior to connecting to the new device.
If, upon replacing the device, already implanted leads are no longer used but left in the patient, then an additional uncontrolled current path to the heart can result.
• Insulate connections that are not used.
Basic principles:
• The device must not be resterilized and reused.
Cremation Devices should not be cremated.
• Explant the device before the cremation of a deceased patient.
Disposal BIOTRONIK takes back used products for the purpose of environmentally safe
disposal.
• Clean the explant with a solution of at least 1% sodium hypochlorite.
• Rinse with water.
• Fill out explantation form and send to BIOTRONIK together with the cleaned device.
Page 23
23 Parameters
4 Parameters
Parameters4403587Technical ManualEluna SR(-T), DR(-T), HF(-T)
Timing
Basic rate day/night
Rate hysteresises
AV delay
AV hystereses
Parameter Range of values Standard SR DR HF
Basic rate 30 ... (5) ... 100 ... (10) ... 200 bpm 60 bpm x x
50 bpm x
Night rate OFF; 30 ... (5) ... 100 ... (10)
... 190 bpm
OFF x x x
Night begins
00:00 ... (10 min) ... 11:50 PM hh:mm 22:00 hh:mm x x x
Night ends 00:00 ... (10 min) ... 11:50 PM hh:mm 06:00 hh:mm x x x
Parameter Range of values Standard SR DR HF
Hysteresis OFF;
-5 ... (-5) ... -25 ... (-20) ... -65 bpm
OFF x x x
Repetitive / search cycles
OFF; ON OFF x x x
Parameter Range of values Standard SR DR HF
AV delay Low; Medium; High; Fixed;
Individual
Low x x
20 ... (5) ... 350 ms (in 6 rate ranges)
180-170­160-150­140 ms
x
150-140­130-120 ms
x
Sense compensation OFF; -10 ... (-5) ... -120 ms -45 ms x x AV safety interval 100 ms 100 ms x x
Parameter Range of values Standard SR DR HF
AV hysteresis mode OFF; Positive; Negative
HF in RV modes: IRSplus
OFF x x
Positive modes: AV hysteresis
70; 110; 150; 200 ms 70 ms
CLS modes: 110 ms
xx
Negative modes: AV hysteresis
10 ... (10) ... 150 ms 50 ms x x
AV repetetive / scan cyles
ON; OFF ON x x
Page 24
24 Parameters
Ventricular pacing
Upper rate
Mode switching
Ventricular pacing
suppression
Parameters valid for devices in DDD-ADI or DDDR-ADIR modes:
Parameter Range of values Standard SR DR HF
Ventricular pacing BiV BiV x Triggering OFF; RVs; RVs + PVC RVs x LV T-wave protection ON; OFF ON x Maximum
trigger rate
AUTO; 90 ... (10) ... 160 bpm
AUTO x
Initially paced chamber
RV; LV LV x
VV delay after Vp 0 ... (5) ... 80 ... (10) ... 100 ms 0 ms x VV delay after sense 0 ms 0 ms x
Parameter Range of values Standard SR DR HF
Upper rate SR: in VVT mode
90 ... (10) ... 200 bpm 130 bpm x x x
Wenckebach response Automatically set x x Atrial upper rate OFF;
175; 200; 240 bpm
240 bpm x x
Parameter Range of values Standard SR DR HF
Mode switching OFF; ON ON x x Intervention rate 100 ... (10) ... 250 bpm 160 bpm x x Switch to (mode) DDI; DDI(R) when
permanent DDD(R) VDI; VDI(R) when permanent VDD(R)
DDI(R) VDI(R)
xx
Ventricular pacing RV; BiV BiV x Onset criterion 3 ... (1) ... 8 5 x x Resolution criterion 3 ... (1) ... 8 5 x x Change of the basic rate
with mode switching
OFF; +5 ... (5) ... +30 bpm
+10 bpm x x
Rate stabilization with mode switching
OFF; ON OFF x x
2:1 lock-in protection OFF; ON DR: ON
HF: only when RV modes ON
xx
Parameter Range of values Standard SR DR HF
Vp suppression ON; OFF OFF x x Pacing suppression
after consecutive Vs
1 ... (1) ... 8 6 x x
Pacing supported after X-out-of-8 cycles
1; 2; 3; 4 3 x x
Page 25
25 Parameters
Refractory periods
Blanking periods
PMT protection
Pacing and Sensing
Pulse amplitude
and pulse width
Sensitivity
Parameter Range of values Standard SR DR HF
Refractory period 200 ... (25) ... 500 ms 250 ms x Atrial refractory
period
AUTO AUTO x x
Atrial refractory period in the modes AAI(R); AAT(R); DDT
300 ... (25) ... 775 ms 350 ms x x
AUTO PVARP ON; OFF ON x x PVARP Auto PVARP OFF:
175 ... (25) ... 600 ms
Auto PVARP ON: Automatically set
xx
PVARP after PVC PVARP + 150 ms
(max: 600 ms) is auto­matically programmed
400 ms x x
Right ventricular refractory period
200 ... (25) ... 500 ms 250 ms x x
Left ventricular refractory period
200 ms 200 ms x
Parameter Range of values Standard SR DR HF
Far-field protection after Vs
100 ... (10) ... 220 ms 100 ms x x
Far-field protection after Vp
100 ... (10) ... 220 ms 150 ms x x
Ventricular blanking period after Ap
30 ... (5) ... 70 ms 30 ms x x
Parameter Range of values Standard SR DR HF
PMT protection OFF; ON ON x x VA criterion 250 ... (25) ... 500 ms 350 ms x x
Parameter Range of values Standard SR DR HF
Pulse amplitude A/ RV/LV
0.2 ... (0.2) ... 6.0 ... (0.5) ... 7.5 V
3.0 V x x x
Pulse width A/RV/LV 0,1 ...(0,1) ... 0.5 ... (0.25)
... 1.5 ms
0.4 ms x x x
Parameter Range of values Standard SR DR HF
Sensitivity AUTO; 1.5 ... (0.5) ... 7.5 mV AUTO x Sensitivity A AUTO; 0.1 ... (0.1) ... 1.5
... (0.5) ... 7.5 mV
AUTO x x
RV sensitivity AUTO; 0.5 ... (0.5) ... 7.5 mV AUTO x x LV sensitivity OFF; AUTO; 0.5 ... (0.5)
... 7.5 mV
AUTO x
Page 26
26 Parameters
Atrial capture control
Ventricular capture control
Atrial overdrive pacing
Lead configuration
Parameter Range of values Standard SR DR HF
Atrial capture control ATM (monitoring only);
ON; OFF
ON x x
Minimum amplitude 0.5 ... (0.1) ... 4.8 V 1.0 V x x Threshold test start 2.4 ... (0.6) ... 4.8 V 3.0 V x x Safety margin 0.5 ... (0.1) ... 1.2 V 1.0 V x x Search type Interval; time of day Time of day x x Interval 0.1; 0.3; 1; 3; 6; 12; 24 h 24 h x x Time of day 00:00 ... (10 min) ... 23:50 2:00 AM hh:mm x x
Parameter Range of values Standard SR DR HF
Ventricular capture control RV, LV
ATM (monitoring only); ON; OFF
ON x x
Minimum amplitude 0.7 V 0.7 V x x Threshold test start 2.4 ... (0.6) ... 4.8 V 3.0 V x x RV safety margin 0.3 ... (0.1) ... 1.2 V 0.5 V x x LV safety margin 1.0; 1.2 V 1.0 V x Search type Interval; time of day Time of day x x Interval 0.1; 0.3; 1; 3; 6; 12; 24 h 24 h x x Time of day 00:00 ... (00:10)
... 23:50 hh:mm
00:30 hh:mm x x
Parameter Range of values Standard SR DR HF
Atrial overdrive pacing
ON; OFF With ON: maximum overpacing rate 120 bpm, mean rate increase approximately 8 bpm, rate decrease after 20 cycles
OFF x x
Parameter Range of values Standard SR DR HF
Sensing polarity A Unipolar, Bipolar Unipolar x x Sensing polarity RV Unipolar, Bipolar Unipolar x x x Sensing polarity LV Unipolar, Bipolar Unipolar x Pacing polarity A Unipolar, Bipolar Unipolar x x Pacing polarity RV Unipolar, Bipolar Unipolar x x x Pacing polarity LV LV tip -> LV ring
LV tip -> RV ring LV ring -> LV tip LV ring -> RV ring LV tip -> housing LV ring -> housing
LV tip –> housing
x
Page 27
27 Parameters
IEGM recordings
Rates for statistics
Rate Adaptation
CLS modes: closed loop
stimulation
Parameters valid for 8 series devices:
R modes: Accelerometer
Parameter Range of values Standard SR DR HF
IEGM recordings 8 series:
20 (quantity); each max. 10 s 6 series: 12 (quantity); each max. 10 s
—xxx
Types of IEGM recordings
High atrial rate (HAR) HAR x x x Mode switching x x x High ventricular rate (HVR) ON x x x 8 series:
Patient triggering (triggered by patient)
OFF x x x
Pre-trigger recording
0; 25; 50; 75; 100% 75 % x x x
IEGM signal Filtered; Unfiltered Filtered x x x
Parameter Range of values Standard SR DR HF
High atrial rate: HAR limit
100 ... (5) ... 250 bpm 200 bpm x x
High ventricular rate: HVR limit
150 ... (5) ... 200 bpm 180 bpm x x x
HVR counter 4; 8; 12; 16 cycles 8 cycles x x x Start resting period 00:00 ... (00:10) ...
23:50 hh:mm
02:00 hh:mmxxx
Duration of resting period
00:30 ... (00:30) ... 12:00 hh:mm
04:00 hh:mm
xxx
Enable lead check ON; OFF ON x x x
Parameter Range of values Standard SR DR HF
Maximum CLS rate 80 ... (5) ... 160 bpm 120 bpm x x x CLS response Very low; Low; Medium;
High; Very high
Medium x x x
CLS resting rate control
OFF; +10 ... (10) ... +50 bpm +20 bpm x x x
Vp required Yes; No No
BiV modes: Yes
xxx
Parameter Range of values Standard SR DR HF
Sensor gain AUTO; Very low; Low;
Medium; High; Very high
AUTO x x x
Max. sensor rate 80 ... (5) ... 180 bpm 120 bpm x x x Sensor threshold Very low; Low; Medium;
High; Very high
Medium x x x
Rate fading OFF; ON OFF x x x Rate increase 1; 2; 4; 8 bpm/cycle 2 bpm/cycle x x x Rate decrease 0.1; 0.2; 0.5; 1.0 bpm/cycle 0.5 bpm/
cycle
xxx
Page 28
28 Parameters
MRI Program
MRI modes Modes valid for devices marked ProMRI
MRI parameters Preset parameters in the MRI program:
Preset Programs
Standard and safe program Mode after auto-initialization:
Lead configuration, determined and set immediately after connection (auto lead check)
Parameters after auto-initialization:
Mode Range of values Standard SR DR HF
MRI Program ON; OFF OFF x x x MRI mode OFF; A00; V00 x
OFF; D00; A00; V00 x OFF; D00; A00; V00;
D00-BiV; V00-BiV
—x
Parameter Range of values Standard SR DR HF
Basic rate 70 ... (10) ... 160 bpm x x x AV delay 110 ms x x Pulse amplitude A/RV4.8 V x x x
Pulse width A/RV 1.0 ms Pulse amplitude LV As in permanent program x Pulse width LV VV delay 0 ms x
Parameter Factory
settings
Standard program
Safe program SR DR HF
Mode VVI VVIR VVI
In the AAI mode, the safe program is also AAI.
x
Mode DDD DDDR VVI x x
Parameter Factory
settings
Standard program
Safe program
SR DR HF
Pacing polarity A/RV Unipolar Unipolar Unipolar x x x Pacing polarity LV TCUP TCUP TCUP x Sensing polarity A/RV Unipolar Unipolar Unipolar x x x Sensing polarity LV Unipolar Unipolar Unipolar x Automatic lead check ON ON x x x
Parameter Factory
settings
Standard program
Safe program
SR DR HF
Basic rate 60 bpm 60 bpm 70 bpm x x
50 bpm x Night program OFF OFF OFF x x x Rate hysteresis OFF OFF OFF x x x Upper rate 130 bpm 130 bpm x x Dynamic AV delay Low Low x x
Page 29
29 Parameters
AV hysteresis OFF OFF x x Sense compensation -45 ms -45 ms x x AV safety delay 100 ms 100 ms x x VV delay 0 0 0 x LV T-wave protection ON ON ON x Far-field protection
after Vs
100 ms 100 ms x x
Far-field protection after Vp
150 ms 150 ms x x
Ventricular blanking period after Ap
30 ms 30 ms x x
PMT protection ON ON x x VA criterion 350 ms 350 ms x x Magnet response AUTO AUTO AUTO x x x Pulse amplitude A 3.0 V 3.0 V x x Pulse amplitude RV 3.0 V 3.0 V 4.8 V x x x Pulse amplitude LV 3.0 V 3.0 V 4.8 V x Pulse width A 0.4 ms 0.4 ms x x Pulse width RV 0.4 ms 0.4 ms 1.0 ms x x x Pulse width LV 0.4 ms 0.4 ms 1.0 ms x Sensitivity A AUTOAUTO— xx Sensitivity RV AUTO AUTO 2.5 mV x x x Sensitivity LV AUTO AUTO 2.5 mV x Refractory period A AUTO AUTO x x Refractory period RV 250 ms 250 ms 300 ms x x x Refractory period LV 200 ms 200 ms 200 ms x Mode switching ON ON x x Onset criterion 5-out-of 8 5-out-of 8 x x Resolution criterion 5-out-of 8 5-out-of 8 x x Intervention rate 160 bpm 160 bpm x x Switches to DDIR DDIR x x The basic rate with mode
switching
+10 bpm +10 bpm x x
Rate stabilization with mode switching
OFF OFF x x
PVARP AUTO
(Start
250 ms)
AUTO (Start 250 ms)
—xx
PVARP after PVC 400 ms 400 ms x x Capture control A ON ON OFF x x x Capture control RV ON ON OFF x x Capture control LV ON ON OFF x Atrial overdrive pacing OFF OFF x x Vp suppression OFF OFF x IEGM recording (HAR, HVR) ON ON OFF x x x Home Monitoring OFF OFF OFF x x x
Parameter Factory
settings
Standard program
Safe program
SR DR HF
Page 30
30 Parameters
Tolerances of Parameter Values
Parameter Range of values Tolerance
Basic rate 30 ... (5) ... 100 ...
(10) ... 200 bpm
± 20 ms
Basic interval 1000 ms ± 20 ms Magnet rate (magnet interval) 90 bpm (664 ms) ± 20 ms Pulse amplitude 0.3 ... 7.5 V The greater value of
±50 mV or +20/-25%
Pulse width 0.1 ... 0.4 ms ± 0.04 ms
0.5 ... 1.0 ms ± 0.10 ms
1.25 ... 1.5 ms ± 0.15 ms
Sensitivity A EN 45502-2-1 triangle pulse
0.1 ... 0.2 mV ±0,05 mV
0.3 ... 7.5 mV ±20%
RV/LV sensitivity EN 45502-2-1 triangle pulse
0.5 ... 7.5 mV ±50%
Refractory period 200 ... 500 ms + 10/-30 ms Maximum activity rate 80 ... 180 bpm ± 20 ms Lead impedance 100 ... 200 Ω ±50 Ω
201 ... 2500 Ω ±25%
Page 31
31 Technical Data
5 Technical Data
Technical Data5403587Technical ManualEluna SR(-T), DR(-T), HF(-T)
Mechanical Characteristics
Note: D = housing without header
Measurements
for the housing
X-ray identification BIO SF
Materials in contact
with body tissue
• Housing: titanium
• Header: epoxy resin
• Plugs in the header: silicone
Electrical Characteristics
Electrically conductive
surface
The device housing has the form of a flattened ellipsoid. The electrically conductive area is 33 cm
2
.
Components and
input values
Electrical characteristics determined at 37°C, 500 Ω:
Telemetry data • Nominal carrier frequency: 403.62 MHz
• Maximum power of transmission: < 25 µW (–16 dBm)
Pulse form The pacing pulse has the following form:
The pulse amplitude reaches its maximum value at the beginning of the pulse (Ua). With increasing pacing duration (tb), the pulse amplitude is reduced dependent on the pacing impedance.
Resistance to interference All variants of BIOTRONIK devices comply with the requirements of
prEN 45502-2-1: 2006, § 27.5.1 at the highest sensitivity.
Device W x H x D [mm] Volume [cm3] Mass [g]
Single-chamber SR(-T) 53 x 39 x 6.5 11 24 Dual-chamber DR(-T) 53 x 44.5 x 6.5 12 25 Triple-chamber HF-T 53 x 49 x 6.5 14 27
Circuit Hybrid electronics with VLSI-CMOS chip Input impedance > 10 kΩ Pulse form Biphasic, asymmetric Polarity Cathodic
Page 32
32 Technical Data
Battery Data
Battery characteristics The following data is provided by the manufacturers:
Shortening of the service
time after long storage
period
Depending on the storage period, the service time from the beginning of service BOS to the replacement time ERI decreases as follows:
•After 1 year by three months
• After 1.5 years by four months
Power consumption • BOS, inhibited: SR(-T), DR(-T) 6 µA; HF-T: 7 µA
• BOS, 100% pacing: SR(-T) 9 µA; DR(-T) 14 µA; HF-T: 18 µA
Calculation of service times Mean service times are valid for devices with and without Home Monitoring; they
are pre-estimated from the following and other data:
• Technical data of the battery manufacturer:
• Basic rate of 60 bpm in AAIR/VVIR modes (single-chamber devices) or DDDR modes (dual-chamber and triple-chamber devices)
•Home Monitoring configuration: OFF
• With 10-minute RF telemetry twice a year as well as without RF telemetry
• Configuration of different pulse amplitudes and lead impedances
Mean service times SR(-T) For single-chamber devices, the following times (in years) result:
Using RF telemetry:
Manufacturer LITRONIK GmbH, 01796 Pirna,
Germany
Battery type LiS 3150M System LiMn0
2
Device type SR(-T); DR(-T); HF-T Battery voltage at BOS 3.1 V Open-circuit voltage 3.1 V Nominal capacity 1.2 Ah Remaining capacity at ERI 0.2 Ah Usable capacity until EOS 1.0 Ah
Amplitude Impedance [Ω] Pacing
10% 50% 100%
1.5 V 500 14.7 14.4 13.8 1000 14.9 14.8 14.4
2.5 V 500 14.1 13.5 11.8 1000 14.5 14.2 13.1
3.0 V 500 13.5 13.1 10.8 1000 14.2 13.6 12.4
3.5 V 500 13.1 12.8 10.1 1000 13.9 13.3 11.8
5.0 V 500 10.4 8.7 6.2 1000 12.1 10.8 8.3
Page 33
33 Technical Data
Without using RF telemetry:
Mean service times DR(-T) For dual-chamber devices, the following times (in years) result:
Using RF telemetry:
Without using RF telemetry:
Amplitude Impedance [Ω] Pacing
10% 50% 100%
1.5 V 500 >15 >15 >15 1000 >15 >15 >15
2.5 V 500 >15 >15 13.3 1000 >15 >15 >15
3.0 V 500 >15 14.2 11.2 1000 >15 >15 14.1
3.5 V 500 14.9 12.9 10.1 1000 >15 >15 12.9
5.0 V 500 11.3 9.1 6.2 1000 14.5 12.5 9.4
Amplitude Impedance [Ω] Pacing
10% 50% 100%
1.5 V 500 12.7 12.3 11.4 1000 12.9 12.8 12.2
2.5 V 500 11.8 11.0 8.9 1000 12.4 11.9 10.4
3.0 V 500 11.0 9.9 7.9 1000 11.9 11.2 9.7
3.5 V 500 10.5 9.3 7.1 1000 11.6 10.7 8.9
5.0 V 500 7.5 5.8 3.8 1000 9.3 7.8 5.5
Amplitude Impedance [Ω] Pacing
10% 50% 100%
1.5 V 500 >15 14.8 13.0 1000 >15 >15 14.7
2.5 V 500 13.6 12.1 9.4 1000 >15 14.1 12.0
3.0 V 500 12.1 10.2 7.9 1000 14.2 12.7 10.1
3.5 V 500 11.1 9.3 7.1 1000 13.4 11.7 8.9
5.0 V 500 7.8 5.8 3.8 1000 10.7 8.7 5.9
Page 34
34 Technical Data
Mean service times HF-T For triple-chamber devices, the following times (in years) result:
Using RF telemetry:
Without using RF telemetry:
Pulse width:
0.4 ms
Amplitude:
Pacing Impedance: atrial + biven-
tricular 100%
A+RV+LV: 500 Ω
A+RV: 500 Ω LV: 800 Ω
A+RV+LV: 1000 Ω
A: 2.5 V RV: 2.5 V LV: 3.5 V
0% 7.8 8.4 9.3 30% 7.5 8.1 9.1 50% 7.3 7.9 8.9 100% 6.9 7.5 8.6
A: 2.5 V RV: 2.5 V LV: 2.5 V
0% 8.9 9.4 10.3 30% 8.6 9.0 9.9 50% 8.3 8.8 9.8 100% 7.8 8.2 9.3
A: 3.5 V RV: 3.5 V LV: 5 V
0% 4.8 5.7 6.6 30% 4.6 5.3 6.3 50% 4.5 5.2 6.2 100% 4.2 4.8 5.8
A: 2.0 V RV: 2.2 V LV: 2.4 V
0% 9.1 9.5 10.3 30% 8.8 9.2 10.0 50% 8.6 8.9 9.9 100% 8.1 8.4 9.6
A: 3.0 V RV: 3.0 V LV: 3.0 V
0% 7.6 8.1 9.2 30% 7.2 7.6 8.8 50% 6.9 7.3 8.6 100% 6.3 6.7 8.1
Pulse width:
0.4 ms
Amplitude:
Pacing Impedance: atrial + biven-
tricular 100%
A+RV+LV: 500 Ω
A+RV: 500 Ω LV: 800 Ω
A+RV+LV: 1000 Ω
A: 2.0 V RV: 2.2 V LV: 2.4 V
0% 10.6 11.1 12.2 30% 10.2 10.8 11.9 50% 9.9 10.4 11.8 100% 9.4 9.8 11.3
A: 2.5 V RV: 2.5 V LV: 2.5 V
0% 10.3 10.8 12.0 30% 9.8 10.4 11.7 50% 9.6 10.1 11.4 100% 9.0 9.4 10.9
A: 2.5 V RV: 2.5 V LV: 3.5 V
0% 8.8 9.8 10.8 30% 8.5 9.3 10.6 50% 8.3 9.1 10.4 100% 7.8 8.6 10.0
A: 3.0 V RV: 3.0 V LV: 3.0 V
0% 8.8 9.4 10.8 30% 8.3 8.8 10.3 50% 7.9 8.5 10.1 100% 7.3 7.8 9.5
A: 3.5 V RV: 3.5 V LV: 5 V
0% 5.3 6.4 7.7 30% 5.1 6.1 7.3 50% 4.9 5.8 7.2 100% 4.6 5.3 6.8
Page 35
35 Technical Data
Legend for the Label
The label icons symbolize the following:
Manufacturing date Use by
Storage temperature Order number
Serial number Product identification
number
CE mark
Contents Follow the instructions for
use
Sterilized with ethylene oxide
Do not resterilize Do not reuse
Do not use if packaging is damaged
Non-sterile
Transmitter with non-ionizing radiation at designated frequency
Label icon on devices with ProMRI
®
:
MR conditional: Patients having a device system implanted whose components are labeled with this symbol on the packaging can be examined using an MR scan under precisely defined conditions.
TP2 Compatibility with telemetry protocol version 2
of BIOTRONIK Home Monitoring
Example
Uncoated device: NBG code and compatible leads
Example
Coated device: NBG code and compat­ible leads
Screwdriver
Example
Header
Bipolar IS-1 connector
Unipolar IS-1 connector
STERILIZE
2
NON
STERILE
Loading...