BIOTRONIK Cardiac Airbag, Cardiac Airbag-T Technical Manual

Page 1
(Draft)
Cardiac Airbag /
Cardiac Airbag-T
Family of Implantable Cardioverter
Defibrillators and Software Cartridge for
TMS 1000
PLUS
and EPR 1000
PLUS
Technical Manual
Page 2
X-ray Identification
Cardiac Airbag/Cardiac Airbag-T
Implantable Cardioverter Defibrillator
Inside the housing, top left-hand side:
Year of manufacture
X-Ray identification
EI
CAUTION
Federal (U.S.A.) law restricts this device to sale by, or on the order of, a physician.
2003 BIOTRONIK, Inc., all rights reserved.
Page 3
Cardiac Airbag Technical Manual i
Contents
1. General...........................................................................1
1.1 System Description....................................................1
1.2 Indications and Usage ...............................................2
1.3 Contraindications .......................................................2
1.4 Warnings and Precautions.........................................3
1.4.1 Sterilization, Storage, and Handling ..................3
1.4.2 Device Implantation and Programming..............4
1.4.3 Lead Evaluation and Connection.......................6
1.4.4 Follow-up Testing...............................................7
1.4.5 Pulse Generator Explant and Disposal..............8
1.4.6 Hospital and Medical Hazards ...........................9
1.4.7 Home and Occupational Hazards......................11
1.4.8 Cellular Phones..................................................11
1.4.9 Electronic Article Surveillance (EAS).................12
1.4.10 Home Appliances ...............................................12
1.5 Adverse Events..........................................................13
1.5.1 Potential Adverse Events...................................13
1.5.2 Observed Adverse Events .................................14
1.6 Clinical Studies ..........................................................17
1.6.1 Patients Studied.................................................17
1.6.2 Methods .............................................................18
1.6.3 Results ...............................................................18
1.7 Patient Selection and Treatment ...............................20
1.7.1 Individualization of Treatment............................20
1.7.2 Specific Patient Populations ..............................20
1.8 Patient Counseling Information .................................20
1.9 Evaluating Prospective ICD Patients.........................21
2. Device Features ............................................................22
2.1 Sensing ......................................................................22
2.1.1 Ventricular Sensitivity Settings ..........................22
2.1.2 Minimum Ventricular Threshold .........................25
2.2 Ventricular Tachyarrhythmia Detection ......................25
2.2.1 VF Classifications ..............................................26
2.2.2 VT Interval Counters ..........................................26
2.2.3 VT Classification ................................................26
2.2.4 Onset and Stability.............................................27
Page 4
ii Cardiac Airbag Technical Manual
2.3 Tachyarrhythmia Redetection ....................................28
2.3.1 VT Redetection ..................................................28
2.3.2 VF Redetection ..................................................28
2.3.3 Tachyarrhythmia Termination ............................28
2.4 Tachyarrhythmia Therapy ..........................................28
2.4.1 Shock Therapy...................................................29
2.5 Bradycardia Therapy .................................................31
2.5.1 Bradycardia Pacing Modes................................31
2.5.2 Basic Rate..........................................................31
2.5.3 Rate Adaptation .................................................31
2.5.4 Gain and Threshold ...........................................32
2.5.5 Rate Increase / Decrease ..................................32
2.5.6 Maximum Sensor Rate ......................................32
2.5.7 Pulse Amplitude .................................................32
2.5.8 Pulse Width........................................................33
2.5.9 Noise Response.................................................33
2.5.10 Post Shock Pacing.............................................33
2.6 Special Features........................................................33
2.6.1 Home Monitoring (Cardiac Airbag-T Only) ........33
2.6.2 Real-time IEGM Transmission...........................37
2.6.3 Capacitor Reformation .......................................37
2.6.4 Patient and Implant Data ...................................38
2.6.5 System Status....................................................39
2.6.6 Holter Memory ...................................................40
2.6.7 Arrhythmia Induction Features...........................42
2.6.8 Manual Shock ....................................................43
2.6.9 Test Shock .........................................................43
3. Software Features.........................................................44
3.1 Follow-Up Assistant (FAST) Window.........................44
3.1.1 Interrogate ICD without Follow-up .....................45
3.2 Main Function Keys ...................................................45
3.3 Parameter Window ....................................................47
4. Sterilization and Storage..............................................50
5. Implant Procedure ........................................................52
5.1 Implant Preparation ...................................................52
5.2 Lead System Evaluation ............................................52
5.3 Opening the Sterile Container ...................................53
5.4 Pocket Preparation ....................................................53
5.5 Lead to Device Connection .......................................54
Page 5
Cardiac Airbag Technical Manual iii
5.6 Blind Plug Connection ...............................................55
5.7 Program the ICD........................................................56
5.8 Implant the ICD..........................................................56
5.9 Suggested Cardiac Airbag Implant Procedure ..........57
6. Follow-up Procedures ..................................................68
6.1 General Considerations.............................................68
6.2 Suggested Cardiac Airbag Follow-Up Procedure......68
6.3 Longevity....................................................................74
6.3.1 Standard ERI Method ........................................74
6.3.2 Treated VF Episode ERI Method.......................76
6.4 Explantation ...............................................................78
7. Technical Specifications ..............................................80
Appendix A...........................................................................85
Page 6
iv Cardiac Airbag Technical Manual
Cardiac Airbag Specifications
Battery Voltage: 6.3 Volts Maximum Shock Energy: 30 joules Defibrillation Lead Ports Two DF-1 (3.2 mm) Pacing Lead Ports One IS-1 (3.2 mm) Dimension: 55 x 67 x 13 mm Volume: 39 cc Mass: 73 g
Housing Material: Titanium Header Material: Epoxy Resin Sealing Plug Material: Silicone Battery Composition Li / MnO2
Page 7
Cardiac Airbag Technical Manual 1
1. General
1.1 System Description
The Cardiac Airbag family of Implantable Cardioverter Defibrillators (ICDs) detects and treats ventricular tachyarrhythmias as well as provides rate adaptive bradycardia pacing support. The ICDs are designed to collect diagnostic data to aid the physician’s assessment of a patient’s condition and the performance of the implanted device. The Cardiac Airbag ICDs are specifically designed to have reduced complexity for implant and follow-up, yet provide essential therapies for conversion of life threatening ventricular tachyarrhythmias.
There are 10 programmable parameters to simplify the implant procedure, and detailed diagnostic information is stored for up to 10 ventricular tachycardia (VT) episodes and 3 treated ventricular fibrillation (VF) episodes. There are 30 minutes of single-channel IEGM storage available to record spontaneous and induced ventricular tachyarrhythmias. The Cardiac Airbag is restricted to storage of diagnostic information up to and including 3 treated ventricular fibrillation episodes.
The Cardiac Airbag ICDs provide therapy for ventricular tachyarrhythmias with programmable defibrillation therapy. The ICDs provide high energy biphasic shocks with the first shock having with programmable energies of 20 or 30 joules and up to 8 shocks per VF episode. The remaining 7 shocks in the therapy progression are pre-set at 30 joules.
The Cardiac Airbag family of ICDs includes the following members:
Cardiac Airbag provides therapies for ventricular tachyarrhythmias and single chamber rate adaptive bradycardia pacing support.
Cardiac Airbag-T is identical to the Cardiac Airbag with the added functionality of BIOTRONIK’s Home Monitoring system. The Home Monitoring System enables automatic exchange of information about a patient’s cardiac status from the implant to the physician remotely.
Page 8
2 Cardiac Airbag Technical Manual
The Cardiac Airbag and Cardiac Airbag-T have two DF-1 defibrillation / cardioversion and one IS-1 pacing/sensing header ports. IS-1 refers to the international standard whereby leads and generators from different manufacturers are assured a basic fit [Reference ISO 5841-3:1992]. DF-1 refers to the international standard for defibrillation lead connectors [Reference ISO 11318:1993].
External devices that interact with and test the implantable devices are also part of the ICD System. These external devices include the TMS 1000 and the EPR 1000
PLUS
PLUS
Tachyarrhythmia Monitoring System
Programming and Monitoring System. These programmers are used to interrogate and program the ICDs. In addition, the programmer software is used to perform the interrogation and programming of the ICDs during implant and follow-up testing.
1.2 Indications and Usage
The Cardiac Airbag Implantable Cardioverter Defibrillators (ICDs) are intended to provide ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.
1.3 Contraindications
Do not use the Cardiac Airbag Implantable Cardioverter Defibrillators (ICDs) in patients:
Whose ventricular tachyarrhythmias may have transient or reversible causes including:
- acute myocardial infarction
- digitalis intoxication
- drowning
- electrocution
- electrolyte imbalance
- sepsis
- hypoxia
Patients with incessant VT of VF
Patients with unipolar pacemaker
Patients whose only disorder is brady arrhythmia or
atrial arrhythmia
Page 9
Cardiac Airbag Technical Manual 3
1.4 Warnings and Precautions
ATP (Anti-Tachycardia Pacing) – The Cardiac Airbag ICD
does not provide ATP therapy. Do not implant this ICD in patients with documented ventricular tachycardias unless high energy defibrillation is desired for treatment of the ventricular arrhythmia.
MRI (Magne
to MRI device scanning. Strong magnetic fields may damage the device and cause injury to the patient.
tic Resonance Imaging) - Do not expose a patient
Electrical Isolation - To prevent ina
induction, electrically isolate the patient during the implant procedure from potentially hazardous leakage currents.
Lead Systems - The use of another manufacturer’s I system may cause potential adverse consequences such as under sensing of cardiac activity and failure to deliver necessary therapy.
Resuscit
unless an alternate source of patient defibrillation such as an external defibrillator is readily available. In order to implant the ICD system, it is necessary to induce and convert the patient’s ventricular tachyarrhythmias.
Unwanted Shocks – Always Therapy status to DISABLED prior to handling the device to prevent the delivery of serious shocks to the patient or the person handling the device during the implant procedure.
Rate-Adaptive Pacing – Use rate-adaptive pacing with care patients unable to tolerate increased pacing rates.
ation Availability - Do not perform induction testing
program the VT/VF Detection and
dvertent arrhythmia
CD lead
in
1.4.1 Sterilization, Storage, and Handling
Device Packaging - Do not use the device if the
packaging is wet, punctured, opened or damaged because the integrity of the sterile packaging may be compromised. Return the device to BIOTRONIK.
device’s
Re-sterilization - Do not r devices.
e-sterilize and re-implant explanted
Page 10
4 Cardiac Airbag Technical Manual
Storage (temperature) - Store the device between 5° to 55°C
(41° - 131° F) because temperatures outside this range could damage the device.
Storage (magnets) - To avoid damage to the device, store the device in a clean area, away from magnets, kits containing magnets, and sources of electromagnetic interference (EMI).
Temperature Stabilization - Allow the device to reach room temperature before programming or implanting the device because temperature extremes may affect initial device function.
Use Before Date - Do not implant the device after the USE BEFORE DATE because the device may have reduced longevity.
1.4.2 Device Implantation and Programming
Blind Plug - A blind plug must be inserted and firmly conn
ected into any unused header port to prevent chronic fluid influx and possible shunting of high energy therapy.
Capacitor Reformation - Infrequent charg
ing of the high voltage capacitors may extend the charge times of the ICD. The capacitors may be reformed manually, or the ICD may be programmed to reform the capacitors automatically. For further information, please refer to Section 2.6.3
, Capacitor Reforming.
Connector Compatibility - ICD and lead system compatibility should be confirmed prior to the implant procedure. Consult your BIOTRONIK representative regarding lead/pulse generator compatibility prior to the implantation of an ICD system. For further information, please refer to Appendix A
.
ERI (Elective Replacement Indicator) - Upon reaching ERI, the
battery has sufficient energy remaining to continue monitoring for at least three months and to deliver a minimum of six 30 joule shocks. After this period, tachyarrhythmia detection and therapy will proceed until EOS is declared. Bradycardia functions are still active at programmed values until the battery voltage drops below 3.0 volts.
Page 11
Cardiac Airbag Technical Manual 5
Magnets - Positioning of a magnet or the programming wand
over the ICD will suspend tachycardia detection and treatment. The minimum magnet strength required to suspend tachycardia treatment is 1.8 mT. When the magnet strength decreases to less than 1 mT, the reed contact is reopened.
Pacemaker/ICD Interaction - In situations where an ICD and a pacemaker are implanted in the same patient, interaction testing should be completed. If the interaction between the ICD and the pacemaker cannot be resolved through repositioning of the leads or reprogramming of either the pacemaker or the ICD, the pacemaker should not be implanted (or explanted if previously implanted).
Programmed Parameters – Program the device parameters to appropriate values based on the patient’s specific arrhythmias and condition.
Programmers - Use only BIOTRONIK programmers to communicate with the device (TMS 1000 EPR 1000
Sealing Sy m
PLUS
).
ste - Failure to properly insert the torque wrench
PLUS
, or
into the perforation at an angle perpendicular to the connector receptacle may result in damage to the sealing system and its self-sealing properties.
Defibrillation Thresho
ld - Be aware that the changes in the
patient’s condition, drug regimen, and other factors may change the defibrillation threshold (DFT) which may result in non­conversion of the arrhythmia post-operatively. Successful conversion of ventricular fibrillation or ventricular tachycardia during arrhythmia conversion testing is no assurance that conversion will occur post-operatively.
Manual Shocks – User-commanded s
hocks may be withheld if the ICD is already busy processing a manual command or the Battery Status is low.
Charge Time - When
preparing a high energy shock the charge circuit stops charging the capacitors after 16 seconds, and delivers the stored energy as shock therapy. After the device reaches ERI the stored energy may be less than 30 joules per shock.
Page 12
6 Cardiac Airbag Technical Manual
Shock Impedance - If the shock impedance is less than twenty-
five ohms, reposition the lead system to allow a greater distance between the electrodes. Never implant the device with a lead system that has measured shock impedance as less than twenty-five ohms. Damage to the device may result.
Programming Wand - Throughout the EP Test session, the programming wand must be positioned and remain directly over the device. If appropriate arrhythmia detection does not occur shortly after induction, remove the programming wand from the ICD and perform external defibrillation.
Data Transmission - Data collection and transmission may take up to 30 seconds. The ICD cannot be reprogrammed during this time even if the [Emergency] key is pressed. Remove the programming wand immediately to restore the permanent program.
EP Test Functions - Ensure that cardiac resuscitation equipment is available during all EP Test Function operations. Physicians should be trained and experienced in tachyarrhythmia induction, conversion protocols, and have adequate training and experience with this device prior to use.
Potential side effects include:
Non-terminable arrhythmia’s that result in death
Complications from hypoxia due to prolonged
arrhythmia’s
Arrhythmia induction that requires cardioversion or defibrillation
Arrhythmia induction that requires pharmacologic treatment, to which the patient could have an adverse reaction
1.4.3 Lead Evaluation and Connection
Capping Leads - If a lead is abandoned rather than removed, it
must be capped to ensure that it is not a pathway for currents to or from the heart.
Gripping Leads - Do not grip the lead with surgical instruments or use excessive force or surgical instruments to insert a stylet into a lead.
Page 13
Cardiac Airbag Technical Manual 7
Kinking Leads - Do not kink leads. This may cause additional
stress on the leads that can result in damage to the lead.
Liquid Immersion - Do not immerse leads in mineral oil, silicone oil, or any other liquid.
Short Circuit - Ensure that none of the lead electrodes are in contact (a short circuit) during delivery of shock therapy as this may cause current to bypass the heart or cause damage to the ICD system.
Suturing Leads - Do not suture directly over the lead body as this may cause structural damage. Use the appropriate suture sleeve to immobilize the lead and protect it against damage from ligatures.
Tricuspid Valve Bioprosthesis - Use ventricular transvenous leads with caution in patients with a tricuspid valvular bioprosthesis.
Setscrew Adjustment – Back-off the setscrew(s) prior to insertion of lead connector(s) as failure to do so may result in damage to the lead(s), and/or difficulty connecting lead(s).
Cross Threading Setscrew(s) – To prevent cross threading the setscrew(s), do not back the setscrew(s) completely out of the threaded hole. Leave the torque wrench in the slot of the setscrew(s) while the lead is inserted.
Tightening Setscrew(s) – Do not overtighten the setscrew(s). Use only the BIOTRONIK supplied torque wrench.
Sealing System – Be sure to properly insert the torque wrench into the perforation at an angle perpendicular to the connector receptacle. Failure to do so may result in damage to the plug and its self-sealing properties.
1.4.4 Follow-up Testing
Defibrillation Threshold - Be aware that changes in the
patient’s condition, drug regimen, and other factors may change the defibrillation threshold (DFT), which may result in non­conversion of the arrhythmia post-operatively. Successful conversion of ventricular fibrillation or ventricular tachycardia during arrhythmia conversion testing is no assurance that conversion will occur post-operatively.
Page 14
8 Cardiac Airbag Technical Manual
Resuscitation Availability - Ensure that an external defibrillator
and medical personnel skilled in cardiopulmonary resuscitation (CPR) are present during post-implant device testing should the patient require external rescue.
Safe Program – Within the EP Test screen, pressing the “Safe Program” key on the programmer head does not immediately send the safe program to the ICD. Pressing the “Safe Program” key activates the emergency function
screen, but an additional
screen touch is required to send the safe program to the ICD.
Date and Time Values - If date and time values are incorrect, the system may, as a result, generate false system status information for the implant.
Impedance Measurement - During the impedance measurement with high stimulation amplitudes, nerve or skeletal muscles may be briefly stimulated.
Threshold Test - A minimum 2:1 voltage safety margin should be permanently programmed any time capture thresholds a
re assessed. Monitor the ECG display closely with pacer­dependent patients. The test should be terminated immediately upon loss of capture.
Inadvertent Programming - The programmer utilizes a touch sensitive screen for menu selections. Care must be used to avoid inadvertent menu selection
by accidentally touching the
screen.
1.4.5 Pulse Generator Explant and Disposal
Device Incineration – Never incinerate the ICD due to the
potential for explosio cremation.
Explanted Devices – Return all explanted devices to BIOTRONIK.
Unwante
d Shocks – Always program the therapy status to
DISABLED prior to handling the device to prevent the deliv serious shocks to the patient or the person handling the during the implant procedure.
n. The ICD must be explanted prior to
ery of
device
Page 15
Cardiac Airbag Technical Manual 9
1.4.6 Hospital and Medical Hazards
Electromagnetic interference (EMI) signals present in hospital and medical environments may affect the function of any ICD or pacemaker. The ICD is designed to selectively filter out EMI noise. However, due to the variety of EMI signals, absolute protection from EMI is not poss
The ICD system should have detection and prior to performing any of the following medica addition, the ICD should be checked after the procedures to assure proper programming:
Diathermy - Diathermy therapy is not recommended for ICD patients due to possible heating effects of the pulse generator and at the implant site. If diathermy therapy must be used, should not be applied in the immediate vicinity of the pulse generator or lead system.
Electrocautery - Electrosurgical cautery could induce ventricular arrhythmias and/or fibrillation, or damage. If use of electrocautery is necessary, the current path and ground plate should be kept as far away from the pulse generator and leads as possible (at least 6 inches (15 cm)).
ible with this or any other ICD.
therapy disabled
l procedures. In
it
or may cause device malfunction
Page 16
10 Cardiac Airbag Technical Manual
External Defibrillation - The device is protected against energy
normally encountered from external defibrillation. However, any implanted device may be damaged by external defibrillation procedures. In addition, external defibrillation may also result in permanent myocardial damage at the electrode-tissue interface as well as temporary or permanent elevated pacing thresholds. When possible, observe the following precautions:
Position the adhesive electrodes or defibrillation paddles of the external defibrillator anterior-posterior or along a line perpendicular to the axis formed by the implanted device and the heart.
Set the energy to a level not higher than is required to achieve defibrillation.
Place the paddles as far as possible away from the implanted device and lead system.
After delivery of an external defibrillation shock, interrogate the ICD to confirm device status and proper function.
Lithotripsy - Lithotripsy may damage the ICD. If lithotripsy must be used, avoid focusing near the ICD implant site.
MRI (Magnetic Resonance Imaging) - Do not expose a patient to MRI device scanning. Strong magnetic fields may damage the device and cause injury to the patient.
Radiation - High radiation sources such as cobalt 60 or gamma radiation should not be directed at the pulse generator. If a patient requires radiation therapy in the vicinity of the pulse generator, place lead shielding over the device to prevent radiation damage and confirm its function after treatment.
Radio Frequency Ablation - Prior to performing an ablation procedure, deactivate the ICD during the procedure. Avoid applying ablation energy near the implanted lead system whenever possible.
Page 17
Cardiac Airbag Technical Manual 11
1.4.7 Home and Occupational Hazards
Patients should be directed to avoid devices that generate strong electromagnetic interference (EMI) or magnetic fields. EMI could cause device malfunction or damage resulting in non-detection or delivery of unneeded therapy. Moving away from the source or turning it off will usually allow the ICD to return to its normal mode of operation.
The following equipment (and similar devices) may affect normal ICD operation: electric arc or resistance welders, electric melting furnaces , r a d i o / t e l e v i s i o n and radar t r a n s m i t t e r s , power-generating facilities, high-voltage transmission lines, and electrical ignition systems (of gasoline-powered devices) if protective hoods, shrouds, etc., are removed.
1.4.8 Cellular Phones
Testing has indicated there may be a potential interaction between cellular phones and BIOTRONIK ICD systems. Potential effects may be due to either the cellular phone signal or the magnet within the telephone and may include inhibition of therapy when the telephone is within 6 inches (15 centimeters) of the ICD, when the ICD is programmed to standard sensitivity.
Patients having an implanted BIOTRONIK ICD who operate a cellular telephone should:
Maintain a minimum separation of 6 inches (15 centimeters) between a hand-held personal cellular telephone and the implanted device.
Set the telephone to the lowest available power setting, if possible.
Patients should hold the phone to the ear opposite the side of the implanted device. Patients should not carry the telephone in a breast pocket or on a belt over or within 6 inches (15 centimeters) of the implanted device as some telephones emit signals when they are turned ON, but not in use (i.e., in the listen or stand-by mode). Store the telephone in a location opposite the side of implant.
Page 18
12 Cardiac Airbag Technical Manual
Based on results to date, adverse effects resulting from interactions between cellular telephones and implanted ICDs have been transitory. The potential adverse effects could include inhibition or delivery of additional therapies. If electromagnetic interference (EMI) emitting from a telephone does adversely affect an implanted ICD, moving the telephone away from the immediate vicinity of the ICD should restore normal operation. A recommendation to address every specific interaction of EMI with implanted ICDs is not possible due to the disparate nature of EMI.
1.4.9 Electronic Article Surveillance (EAS)
Equipment such as retail theft prevention systems may interact with pulse generators. Patients should be advised to walk directly through and not to remain near an EAS system longer than necessary.
1.4.10 Home Appliances
Home appliances normally do not affect ICD operation if the appliances are in proper working condition and correctly grounded and shielded. There have been reports of the interaction of electric tools or other external devices (e.g. electric drills, older models of microwave ovens, electric razors, etc.) with ICDs when they are placed in close proximity to the device.
Page 19
Cardiac Airbag Technical Manual 13
1.5 Adverse Events
1.5.1 Potential Adverse Events
The following is a list of the potential risks that may occur with this device:
Acceleration of arrhythmias
Air embolism
Bleeding
Chronic nerve damage
Erosion
Excessive fibrotic tissue growth
Extrusion
Fluid accumulation
Formation of hematomas or cysts
Inappropriate shocks
Infection
Keloid formation
Lead abrasion and discontinuity
Lead migration / dislodgment
Myocardial damage
Pneumothorax
Shunting current or insulating myocardium during
defibrillation with internal or external paddles
Potential mortality due to inability to defibrillate or pace
Thromboemboli
Venous occlusion
Venous or cardiac perforation
Page 20
14 Cardiac Airbag Technical Manual
Patients susceptible to frequent shocks despite antiarrhythmic medical management may develop psychological intolerance to an ICD system that may include the following:
Dependency
Depression
Fear of premature battery depletion
Fear of shocking while conscious
Fear that shocking capability may be lost
Imagined shocking (phantom shock)
There may be other risks associated with this device that are currently unforeseeable.
1.5.2 Observed Adverse Events
A clinical study of the Phylax XM involved 155 devices implanted in 154 patients with cumulative implant duration of 1286 months (mean implant duration 8.3 months). This clinical study was performed with the Phylax XM and Phylax 06 ICDs, which are earlier versions of the Cardiac Airbag ICDs. The observed adverse events are applicable because the Cardiac Airbag ICD is a downsized version of the Phylax XM with rate adaptive pacing capabilities.
NOTE:
The Phylax XM ICD is an earlier generation of BIOTRONIK devices. The Cardiac Airbag family is based upon the Phylax XM and other BIOTRONIK ICDs (i.e., Belos VR and Belos VR-T).
There were a total of five deaths during the course of the trial; none of the deaths were judged by the clinical study investigator to be device related. Heart failure was a major factor in two deaths. The other three deaths were related to renal failure, lung disease, and septic shock secondary to an ischemic bowel, respectively. All five of the deaths occurred more than one month post implant.
Page 21
Cardiac Airbag Technical Manual 15
Two ICDs were explanted during the trial. One was secondary to the patient being unable to tolerate further testing required by the clinical protocol. The other was secondary to a systemic infection; the patient was subsequently implanted with another device.
Table 1 provides a summary of the adverse events that were reported during the clinical study regardless of whether or not the event was related to the ICD system. A complication is defined as a clinical event that results in invasive intervention, injury, or death. An observation is defined as a clinical event that does not result in invasive intervention, injury, or death.
Table 1: Reported Adverse Events (AEs)
Number of Patients = 154, Number of Patient-Years = 107.1
Event # of pts
with AEs
Complications (total) 7 8 0.07 4.5%
% of
pts
with
AEs
# of
AEs
AE/
pt-
yrs
Lead repositioning 2 1.3% 2 0.02 Hematoma 1 0.6% 1 0.01 Systemic infection 1 0.6% 1 0.01 Explant (did not to tolerate
1 0.6% 1 0.01 testing) Insertion of separate sensing 1 0.6% 2 0.02 lead ICD/lead connection 1 0.6% 1 0.01
Observations (total) 79 51.3% 89 0.83
Inappropriate therapy (SVT) 18 11.7% 20 0.19 ICD response to magnet in
1
wand Software messages and 11 7.1% 13 0.12
2
errors
13 8.4% 15 0.14
Increased pacing threshold 7 4.5% 9 0.08 Decreased R-wave amplitude
Frequent VT 5 3.2% 5 0.05
7 4.5% 7 0.07
Oversensing 3 1.9% 3 0.03 TMS 1000 difficulties3 3 1.9% 3 0.03 VT below rate cut-off 2 1.3% 3 0.03
Page 22
16 Cardiac Airbag Technical Manual
Event # of pts
with AEs
High DFT’s 1 0.6% 2 0.02
% of
pts
with
AEs
# of
AEs
AE/
pt-
yrs
Minor stroke 1 0.6% 1 0.01 Renal failure 1 0.6% 1 0.01 Required additional drug
1 0.6% 1 0.01 therapy ICD/lead connection 1 0.6% 1 0.01 ICD therapy during lead
1 0.6% 1 0.01 connection Non-sustained VT 1 0.6% 1 0.01 Non-conversion of arrhythmia
Interpretation of real-time 1 0.6% 1 0.01
1 0.6% 1 0.01
markers Reconfirmation algorithm 1 0.6% 1 0.01
1. This category inclu the
programmer wand that caused the reed switch to
toggle during hig age pacito ha ng or
des is s re o m em f
sue lated t ov ent o
h volt ca r c rgi tachyarrhythmia detection. As a result, appropriate therapy was not delivered in a timely manner. The orientation of the reed switch was optimized and is being monitored as part of the manufacturing process to prevent future occurrences of this type of event.
2.
This category includes various software “anomalies” that were related to error messages or the retrieval of diagnostic information. Each of these events has been resolved through revisions made to the software.
3. This category includes any difficulties encountered while using the TMS 1000 Tachyarrhythmia Monitoring System. Each of these events has been resolved through revisions to the software and hardware
of the
system.
Page 23
Cardiac Airbag Technical Manual 17
1.6 Clinical Studies
N
OTE:
The Phylax XM ICD is an earlier generation of BIOTRONIK devices. The Cardiac Airbag family is based upon the Phylax XM and other BIOTRONIK ICDs (i.e., Belos VR and Belos VR-T).
This clinical study was performed on the Phylax XM and Phylax 06 ICDs, which are earlier versions of the Cardiac Airbag ICD. The clinical study data presented here is applicable because the Cardiac Airbag / Cardiac Airbag-T is a downsized version of the Phylax XM with the addition of rate adaptive pacing capabilities. The Cardiac Airbag / Cardiac Airbag-T ICDs are slightly different as compared to the Phylax XM in the following areas:
Motion based rate adaptive pacing
Reduced programmable feature set
Minor adjustments to therapy delivery options including
no availability of ATP
Reduced size from 69 cc to 39 cc
Addition of Home Monitoring functionality
The rate adaptive pacing circuitry of Cardiac Airbag / Cardiac Airbag-T ICD is based on other US distributed BIOTRONIK products. Due to the similarities between the Cardiac Airbag / Cardiac Airbag-T, Belos VR / VR-T, and Phylax XM and the limited nature of these changes, a clinical study of the Cardiac Airbag / Cardiac Airbag-T ICD was determined to be unnecessary.
1.6.1 Patients Studied
The clinical study involved 154 patients (121 male and 33 female) with a mean age of 64.9 years (range: 26 to 95 years) and a left ventricular ejection fraction of 33% (range: 10% to 80%). Most (72%) presented with coronary artery disease / ischemic cardiomyopathy; 71% presented with monomorphic ventricular tachycardia (MVT) as their primary tachyarrhythmia.
Page 24
18 Cardiac Airbag Technical Manual
1.6.2 Methods
The multicenter clinical investigation was designed to validate the safety and effectiveness of the ICD system to detect and treat monomorphic ventricular tachycardia (MVT), polymorphic ventricular tachycardia (PVT), ventricular fibrillation (VF), and bradycardia. The specific predefined objectives of the investigation included the determination of ventricular tachyarrhythmia conversion rate, sudden cardiac death (SCD) survival rate, morbidity rate, and the appropriate sensing and pacing rate.
The primary endpoint of the study was to evaluate the ventricular tachyarrhythmia conversion rate. Patients underwent standard ICD implantation and then were evaluated at predischarge and regular follow-ups every three months. Induction and conversion of the patient’s tachyarrhythmias was required at the implant procedure and predischarge follow-up.
1.6.3 Results
The mean implant duration was 8.3 ± 0.4 months with cumulative implant duration of 1286 months. There were 39 patients followed for over twelve months and 108 patients followed for over six months. The patient follow-up compliance rate was 99.6% out of 473 follow-up procedures.
2 provides a summary of the results of the study group for
Table
the predefined endpoints.
Page 25
Cardiac Airbag Technical Manual 19
Table 2: Clinical Study Results
Description Study Group
[95% CI]
Tachyarrhythmia Conversion Rate Induced
1
95.8% (496/518) [93.6%, 97.3%]
Spontaneous 99.7% (1540/1544)
[99.3%, 99.9%]
Total 98.7% (2036/2062)
[98.2%, 99.2%]
Sudden Cardiac Death Survival (at one year)
Complication Rate (per total number of patients)
Appropriate Sensing and Pacing Rate
100.0% (39/39)
[91.0%, 100.0%]
5.2% (8/154)
[2.3%, 10.0%]
2
98.0% (703/717) [96.8%, 98.9%]
1. Conversion data were collected in the clinical study for both induced and spontaneous tachyarrhythmia episodes. Therefore, both types of tachyarrhythmia episodes were included in the analysis.
2. The investigator determined the appropriateness of bradycardia sensing and pacing. The rate will be determined by the number of appropriate bradycardia sensing and pacing evaluations divided by the total number of evaluations.
Page 26
20 Cardiac Airbag Technical Manual
1.7 Patient Selection and Treatment
1.7.1 Individualization of Treatment
Determine whether the expected device benefits outweigh the possibility of early device replacement for patients whose ventricular tachyarrhythmias require frequent shocks.
Determine if the device and programmable options are appropriate for patients with drug-resistant supraventricular tachyarrhythmias (SVTs), because drug-resistant SVTs can initiate unwanted device therapy.
Direct any questions regarding individualization of patient therapy to your BIOTRONIK representative or BIOTRONIK technical services at 1-800-547-0394.
1.7.2 Specific Patient Populations
Pregnancy - If there is a need to image the device, care should
be taken to minimize radiation exposure to the fetus and the mother.
Nursing Mothers - Although appropriate biocompatibility testing has been conducted for this implant device, there has been no quantitative assessment of the presence of leachables in breast milk.
Geriatric Patients - Most (72%) of the patients receiving an ICD in the Phylax XM clinical study were over the age of 60 years (see Clinical Studies).
Handicapped and Disabled Patients - Special care is needed in using this device for patients using electrical wheel chair or other electrical (external or implanted devices).
1.8 Patient Counseling Information
The pulse generator is subject to random component failure. Such failure could cause inappropriate shocks, induction of arrhythmias or inability to sense arrhythmias, and could lead to the patient’s death.
Page 27
Cardiac Airbag Technical Manual 21
Persons administering CPR may experience the presence of voltage on the patient’s body surface (tingling) when the patient’s ICD system delivers a shock.
A patient manual is available for the patient, patient’s relatives, and other interested people. Discuss the information in the manual with concerned individuals both before and after pulse generator implantation so they are fully familiar with operation of the device. (For additional copies of the patient manual, contact the BIOTRONIK at the address listed in this manual.)
1.9 Evaluating Prospective ICD Patients
The prospective ICD implant candidate should undergo a cardiac evaluation to classify any and all tachyarrhythmias. In addition, other patient specific cardiac information will help in selecting the optimal device settings. This evaluation may include, but is not limited to:
an evaluation of the specific tachycardia rate(s)
the confirmation and/or evaluation of any
supraventricular arrhythmias or bradyarrhythmias
the evaluation of various ATP and cardioversion therapies
the presence of any post-shock arrhythmias, and
an evaluation of the maximum sinus rate during exercise
If a patient’s drug regimen is changed or adjusted while the ICD is implanted, additional EP testing may be required to determine if detection or therapy parameter settings are relevant and appropriate.
Page 28
22 Cardiac Airbag Technical Manual
2. Device Features
The Cardiac Airbag family feature set is presented under the following sub-headings: Sensing, Tachyarrhythmia Detection, Tachyarrhythmia Redetection, Tachyarrhythmia Therapy, Bradycardia Therapy, and Special Features. The features apply to all members of the Cardiac Airbag family except where specifically referenced differently.
2.1 Sensing
The Cardiac Airbag ICDs use Automatic Sensitivity Control (ASC) to adjust the sensitivity characteristics to appropriately detect the various cardiac signals. The characteristics of the sensing circuitry have been optimized to ensure appropriate sensing during all potential cardiac rhythms.
Cardiac signals vary in amplitude; therefore detection thresholds cannot be static. The Automatic Sensitivity Control (ASC) utilizes an automatic step-down threshold for sensing ventricular signals. The ASC begins by tracking the cardiac signals (R­waves) during the sensed refractory periods. The peak values measured during this time are used to set the sensing thresholds during the active detection periods.
2.1.1 Ventricular Sensitivity Settings
There are three programmable options for setting the sensitivity of the input stage. The sensitivity selections are designed to adapt the parameters of the input stage to various signal conditions. The predefined parameter sets are described in
Table 3
.
Page 29
Cardiac Airbag Technical Manual 23
Table 3: Sensitivity Settings
Setting Definition for Use
Standard This setting is recommended for most
patients, especially for those with
measured R-wave amplitude of 3 mV. Enhanced T Wave Suppression
This setting offers suppression of T-wave
oversensing. This mode should not to be
used on patien
onditions:
c
ts with the following
Sinus rhythms with small signal amplitudes, R-waves <4 mV
VF with highly fluctuating signal amplitudes.
Enhanced This setting enhances VF VF Sensitivity
cases of highly fluctuating signal
detection, in
amplitudes. It is not to be used for patients that have sinus rhythms containing large amplitude T-waves.
Free This parameter configuration is only
accessed by code and is not available in the US.
Typically, the upper threshold (UT) is reset with each sensed R-wave, but in order to ensure that pacing does n
ot occur during an episode of VF, the ASC behaves differently with paced events. Each paced event is followed by a paced refractory period (250 ms) after which the ventricular threshold is set to the minimum programmed value.
TANDARD
S
the Standard sensitivity setting following the 100 ms sensed refractory period. The UT decays 0.125 mV every 250 ms through the T-wave discrimination period (350 ms). After the T­wave discrimination period, the threshold is decreased to the lower threshold (LT). The LT is set to 25% of the measured peak R-wave. The LT then decreases 0.125 mV every 500 ms until the Minimum Threshold is reach or until the next sensed (or paced) event.
- The UT is set at 50% of the measured R-wave for
Page 30
24 Cardiac Airbag Technical Manual
ic Sensitivity Control with Standard Figure 1. Automat
Setting
Figure
1 provides a trol
n illustration of Automatic Sensitivity Con
with the sensitivity programmed to Standard. The tracked R –
ave is measured to be 6.0 mV following the sensed refractory
w period the UT is set to 3.0 mV. After the T-wave discrimination period, the threshold is further reduced to 1.5 mV. Both the Upper and Lower Thresholds decay over time, but the Minimum Threshold is never violated. Nominally, the minimum threshold is set to 0.8 mV, but it can be adjusted by the user.
NHANCED VF SENSITIVIT
E
Y - The Enhanced VF Sensitivity
setting is specifically designed to improve VF detection when the VF signal is very small. Two adjustments are made to ASC with this setting:
The T-wave discrimination period is decreased to 100 ms, thus eliminating the UT
The decay rate of the LT is increased to 0.125 mV every 250 ms.
These adjustme
nts ensure that the threshold reaches the lower values more quickly in order to ensure that all VF signals are sensed appropriately.
Page 31
Cardiac Airbag Technical Manual 25
ENHANCED T-WAVE SUPPRESSION - The Enhanced T-Wave
Suppression setting is specifically designed to avoid double counting of each QRS-T complex during normal sinus rhythms. Two adjustments are made to ASC with this setting:
High pass filtering is increased to reduce low frequency signal components such as T-waves and respiratory artifacts.
The UT is increased to 75% of the measured R-wave.
The UT may not retrigger with each sensed event, it is
only triggered when the new sensed R-wave crosses the 50% point of the previous measured R-wave.
2.1.2 Minimum Ventricular Threshold
This parameter limits the minimum sensitivity of the ICD to programmable value. Nominally, the minimum threshold is set to
0.8 mV, but it can be adjusted from 0.5 to 2.5 mV.
2.2 V
The Cardiac Airbag ICDs detect and measure the rate of sensed cardiac signals to discriminate ventricular tachyarrhythmias from sinus rhythm or sinus bradycardia. This is accomplished through programmable rate detection parameters in the device. When a tachyarrhythmia is present, the ICD classifies the arrhythmia and delivers the appropriate therapy. If a tachyarrhythmia continues following the first therapy attempt, then the ICD will redetect the tachyarrhythmia subsequent therapies as necessary.
Classification of cardiac signals is accomplished primarily by measuring the cardiac cycle length (R-R intervals). In addition, the ICD can also utilize abrupt changes in rate or irregularity of the cardiac tachyarrhythmias. Each detected ventricular tachyarrhythmia is clas
Each rhythm class is set to a separate rate with the zone limit defining the lowest rate in each class. The upper rate limit of the VT zone is equal to the
entricular Tachyarrhythmia Detection
and deliver
signal to further differentiate ventricular
sified into one of the following zones:
VT Ventricular Tachycardia Monitoring Zone
entricular Fibrillation
VF V
VF zone limit.
a
Page 32
26 Cardiac Airbag Technical Manual
2.2.1 VF Classifications
Detection of ventricular fibrillation (VF) utilizes a non­programmable X out of Y criterion. If X number of intervals within the sliding window (defined by Y) are shorter than the prog m
ra med VF rate interval in ms (> in bpm), VF is detected. After fib for VF is initiated.
Pre s re 8 o 2 intervals criterion
rillation is detected, the programmed therapy sequence
set ettings for classification of ventricular fibrillation (VF) a f 1 intervals; meaning that within a sample window of 12
, 8 intervals must meet or exceed the VF zone rate
.
2.2.2 VT Interval Counters
The VT Interval Counters utilize non-programmable VT rate classifications. The Interval Counter is the number of intervals required to declare a tachyarrhythmia as VT. A ta must meet both the rate/interval criteria and the preset Interval Counter, in addition to other detection enhancements (onset and stability) to be declared a tachycardia.
chyarrhythmia
2.2.3 VT Classification
The VT classification zone utilizes a non-programmable detection parameter (VT interval counter) that is different from the VF zone. Classification of VT is based on the last interval average preceding declaration of tachyarrhythmia detection. If this average falls within the VT zone, an IEGM is stored since the VT zone is designed as a “Mon therapies are available.
In addition, when the Cardiac Airbag senses the programmed number of consecutive intervals (termination count) within the sinus rate zone, all tachyarrhythmia detection criteria, including the VT sample counters are reset.
itoring Zone” only and no
Page 33
Cardiac Airbag Technical Manual 27
2.2.4 Onset and Stability
In addition to the standard tachycardia detection parameters previously described, the VT Monitoring Zone incorporates two additional detection enhancements: Onset and Stability. Both Onset and Stability are preset to standard values and are not programmable for the VT Monitoring Zone.
2.2.4.1 Onset
The Onset function provides an additional discrimination test that must be satisfied before a VT tachyarrhythmia can be declared. The purpose of this detection parameter is to discriminate between s increase) and a ventricular tachycardia, which typically begins with an abrupt rate change.
Onset criterion evaluates the most recently sensed cardiac intervals and compares it to the previous four-interval sliding average. Onset will be satisfied if a change in cycle length exceeds the preset Onset value (as compared to the average) and is followed by a cycle that lies within the corresponding VT zone. Onset criterion is defined as (expressed as a percentage of the latest cardiac cycle length). VT is not declared until Onse criteria are satisfied.
inus tachycardia (often characterized by a gradual rate
a 20% adaptive value
t and any additional detection
2.2.4.2 Stability
The purpose of Stability is to assist in the discrimination of stable ventricular tachyarrhythmias from SVTs that conduct irregularly down to the ventricles (i.e., atrial fibrillation). Stability evaluates sudden changes in the cardiac cycle length. The Stability criterion compares each in
terval with the three preceding cardiac cycles to determine if they remain within the Stability range (as defined by the parameter setting of ± 24 ms). A rhythm is declared stable after the number of intervals (equal to the Interval Count) is found to be stable
within the range.
Page 34
28 Cardiac Airbag Technical Manual
2.3 Tachyarrhythmia Redetection
The Cardiac Airbag ICDs incorporate settings for determining if tachyarrhythmias remain after therapy has been delivered. The redetection routine allows the ICDs to determine whether further therapy is required when the initial therapy was unsuccessful in terminating the arrhythmia.
Tachyarrhythmia re length and number of intervals. The number of intervals is distinct and independent of the initial detection criteria.
2.3.1 VT Redetection
The Redetection Count is not programmable and remains independent of the initial dete
Redetection of an ongoing tachyarrhythmia is declared when the Redetection Count is satisfied (based on individual cycles). If a sensed cardiac signal meets the VT rate criterion, following initial detection, that signal is counted and compared to the Redetection Count. Tachycardia redetection is declared when the number of VT samples (Redetection Count) is satisfied.
Redetection functions identically to initial VT detection in regards to the Stability and On on individual cycle lengths (not averages).
detection criteria are based on cardiac cycle
ction parameters:
set detection enhancements and is based
2.3.2 VF Redetection
VF redetection uses the same X out of Y algorithm as initial detection. The 8 out of 12 criterion for initial detection is used for redetection to ensure consistent classification of VF.
2.3.3 Tachyarrhythmia Termination
Termination of a ventricular tachyarrhythmia episode is declared when 12 out of 16 consecutive sensed intervals are l the VT-interval counter.
onger than
2.4 Tachyarrhythmia Therapy
The Cardiac Airbag ICDs offers only defibrillation therapy for the treatment of ventricular tachyarrhythmias classified as VF.
Page 35
Cardiac Airbag Technical Manual 29
2.4.1 Shock Therapy
The Cardiac Airbag ICDs offer shock therapy only for the VF rate classifications. Up to 8 shocks are available for the VF zone for each episode detected.
The first defibrillation shock in the therapy sequence is delivered with confirmation (while the first shock energy is programmable to 20 or 30 joules and is delivered following confirmation of the arrhythmia. The remaining shock energies are non-programma predetermined to deliver 30 joules using defibrillation without confirmation. All shocks utiliz and normal polarity.
2.4.1.1 Number of Shocks
The number of shocks defines the total number of shock attempts for each VF detection. Up to 8 shocks are available in this therapy zone. The first shock energy parameter is programmable to 20 or 30 joules, while the remaining shocks are fixed at 30 joules.
2.4.1.2 Confirmation
Confirmation is used to verify the presence of a tachyarrhythmia during the charging of the capacitors. This function is designed to avoid delivery of inappropriate therapy has spontaneously terminated. The programmed shock will be delivered unless bradycardia detected during the Confirmation period. Confirmation is always ON for the first shock therapy and is always OFF for remaining shock therapies.
Confirmation OFF - When Confirmation is OFF, shock therapy will be delivered to the patient during the synchr regardless of the detected cardiac signal.
capacitors are being charged). The
ble and
e a standard biphasic waveform
if a tachyarrhythmia
or a normal sinus rhythm is
onization period
Confirmation ON - If the tachyarrhythmia spontaneously converts to bradycardia or a normal sinus rhythm during the confirmation period, sho device confirms the presence of the tachyarrhy will deliver the programmed shock therapy.
ck therapy is aborted. However if the
thmia, the device
Page 36
30 Cardiac Airbag Technical Manual
Synchronization - A synchronization window is started at the
end of the charging period. During this window, the device will attempt to synchronize the shock therapy to an R-wave. If no R­wave is detected, the shock will be delivered asynchronously at the end of the synchroniza
tion period.
2.4.1.3 Shock Waveform
All shocks utilize a standard biphasic waveform. The waveform starts at the calculated voltage, based on the programmed energy level. After an exponential discharge through the lead system to 40% of the initial charge voltage, the shock switches polarity. At that point, it discharges to 20% of the initial charge voltage before the w
aveform is truncated. Figure 2
provides a
pictorial representation of the biphasic waveform.
Phase 1 Phase 2
Begin 100% 40%
End 40% 20%
Figure 2. Biphasic Waveform
2.4.1.4 Shock Energy
The Cardiac Airbag ICDs are designed to ensure that the energy programmed for therapy is the same as what is actually delivered to the patient regardless of the lead impedance.
2.4.1.5 Shock Polarity
The polarity of the shock therapy is non-programmable and preset to Norma
l. This polarity configures the HV 1 connector port as the negative electrode and the HV 2 connector port and the outer housing of the ICD as the positive electrode for the first phase of the shock.
Page 37
Cardiac Airbag Technical Manual 31
2.5 Bradycardia Therapy
The Cardiac Airbag ICDs have programmable bradycardia and post-shock bradycardia pacing functions. The post-shock bradycardia parameters are preset to a higher rate and output values following a delivered shock, without compromising the longevity of the ICD for patients pacing. The post-shock values are presented in the following subsections after the chronic bradycardia support values.
2.5.1 Bradycardia Pacing Modes
The bradycardia pacing mode may be programmed to VVI or VVIR for bradycardia pacing support or to OFF (OVO). The basic rate timer is initiated by a sensed or paced event. A
ensed event outside of the refrac
s tory period inhibits pacing and resets the lower rate timer. In the absence a pacin will be delivered at the end of the lower rate
g pulse
interva e OFF disabl dyc
l. Th mode es bra
tachycardia sensing and therapy may remain active.
The mode that contains an “R adaptive mode. This mode is corresponding non-rate ada rate will be automatically adjusted to take into account the current load on the patient’s heart in response to increased physical activity.
who require chronic bradycardia
of a sensed event,
ardia pacing; however
” in its designation is a rate
functionally the same as the
ptive mode; except that the pacing
2.5.2 Basic Rate
The basic rate is the rate at which bradycardia pacing will occur in the absence of a patient’s intrinsic rhythm. This rate may be individually programmed for normal bradycardia pacing.
2.5.3 Rate Adapt
Cardiac Airbag / Cardiac Airbag-T ICDs allow the selection of a rate responsive pacing mode (VVIR). This mode allows the ICDs bradycardia therapy function to adapt the pacing rate to increasing or decreasing patient activity, based on data collected from a motion sensor within the ICD. Separate criteria controls the rate of increase and decrease of pacing, as well as the sensitivity of the sensors.
ation
Page 38
32 Cardiac Airbag Technical Manual
2.5.4 Gain and Threshold
The Gain defines how much the sensor signal is amplified before it is used by the rate adaptive algorithm. The Gain is programmed so the maximum desired pacing rate during exercise occurs at a maximum exertion level. The Gain is preset to 4 when programmed to the VVIR mode.
The Sensor Threshold defines the lowest sensor output that initiates a change in the pacing rate and all motion belo threshold is ignored by the algorithm. The Sensor Threshold is preset to Mean when programmed to the VV
IR mode.
w this
2.5.5 Rate Increase / Decrease
The Rate Increase and Rate Decrease parameters work together with the Gain to determine how quickly pacing rate increases or decreases to occur with changes in the sensor output. A rate increase of 2 ppm per second would take 45 seconds to change from a pacing rate of 60 ppm to 150 ppm. The Rate Increase is preset to 2 ppm/sec when prog the VVIR mode. A rate decrease setting of 0.4 ppm per second will take 225 seconds to decrease a pacing rate of 150 ppm to 60 ppm. The Rate Decrease is preset to 0.4 ppm/sec when programmed to the VVIR mode.
rammed to
2.5.6 Maximu
Regardless of the sensor output, the sensor-driven pacing rate never exceeds the progr maximum sensor rate only limits the pacing rate during sensor­driven pacing. The Maximum Sensor Rate is programmable to either 100 or 125 ppm when programmed to the VVIR mode.
m Sensor Rate
ammable Maximum Sensor Rate. The
2.5.7 Pulse Amplitude
The Pulse Amplitude parameter defines the amplitude in volts of the pacing pulses. The pulse amplitude is independently set for normal and post-shock bradycardia pacing.
Page 39
Cardiac Airbag Technical Manual 33
2.5.8 Pulse Width
The Pulse Width parameter defines the duration of the pacing pulses. The pulse width is independently set for normal and post-shock bradycardia pacing.
2.5.9 Noise Response
The Cardiac Airbag ICD’s response to detected noise is to deliver asynchronous pacing in ventricular channel.
2.5.10 Post Shock Pacing
Separately, bradycardia pacing support is available with the ICD following shock therapy delivery. After a short blanking period (1 second), the ICD will begin bradycardia therapy at the post shock pacing rate, amplitude, and pulse width for the post shock duration.
If bradycardia pacing is still required after the post shock duration expires, standard bradycardia pacing parameters will be utilized.
2.6 Special Features
2.6.1 Home Monitoring (C
Home Monitoring enables the exchan patient’s cardiac status from the impla Monitoring can be used to provide the physician with advance reports from the implant and process them into graphs and tables. This information helps the physician optimize the therapy process, as it allows the patient to be scheduled for additional clinical appointments between regular follow-up visits necessary.
The implant’s Home Monitoring entire operational life of the implant (prior to ERI) or for shorter periods, such as several weeks or months.
OTE
N :
When ERI mode is reached, this status is transmitted. Further measurements and transmissions of Home Monitoring data are no longer possible.
ardiac Airbag-T Only)
ge of information about a
nt to the physician. Home
if
function can be used for the
Page 40
34 Cardiac Airbag Technical Manual
2.6.1.1 Transmission of Information
The implant transmits information with a small transmitter, which has a range of about 2 meters. The transmissions are activated by the detection of an arrhythmia episode, as programmed. The types of transmissions are discus
sed in Section 2.6.1.4
.
The minimal distance between the
implant and the patient device
must be 15 cm.
2.6.1.2 Patient Device
The patient device (Figure and is comprised of the mobile
3) is designed for use in the home device and the associated
charging station. The patient can carry the mobile device with them during his or her occupational and leisure activities. The patient device is rechargeable, allowing for an approximate operational time of 24 hours. It receives information from the implant and forwards it via a GSM mobile cell phone network to a BIOTRONIK Service Center.
For additional information about the patient device, please refer to its man
ual.
Page 41
Cardiac Airbag Technical Manual 35
Figure 3: Example of Patient D (CardioMessenger)
evice with Charging Stand
2.6.1.3 Card
io Report
The implant’s information is digitally formatted by the BIOTRONIK Service Center and processed into a concise report called a Cardio Report. The Cardio Report is titled depending on the type of event transmission. This Cardio Report contains current and previous implant data. The Cardio Report is sent to the attending physician via fax. All reports use the same report format.
2.6.1.4 Types of Report Transmissions
When the Home Monitoring function is activated, the transmission of a report (Cardio Report) from the implant can be triggered as follows:
Event report – the ICD detects certain events, which initiate a report
Page 42
36 Cardiac Airbag Technical Manual
of the patient data,
To ensure successful transmission the
Cardiac Airbag-T is programmed to send up to 10 repetitive transmissions of identical data at an time interval.
hourly
Event Report events are detected by the implant, t transmission is automatically triggered. This is descr
The following cardiac and technical events initiate
- When certain cardiac and technical a repor
ibed as an “event message”.
a message
transmission:
Special device status (errors)
Detected
Detected and terminated VF
First ineffective shock detected
Pace
Shock
Device st
2.6.1.5
Description of Transmitted Data
The following data are transmitted by the Home Monitor system, when activated. In addition to the medical data, serial number of the im
VT
impedance < 200 Ohm or > 3 K Ohm
impedance < 25 Ohm or > 150 Ohm
atus - ERI
ing the
plant is also transmitted.
Detection
# of episodes in VT Monitoring Zone
# of episodes in VF Zone
Therapy
Shocks delivered
Shocks successful
Shocks aborted
1st Shock without success
Battery
Status (i.e., OK, ERI, EOS)
measurement
Date of voltage
Page 43
Cardiac Airbag Technical Manual 37
Leads
Pace impedance (ventricular)
Shock impeda
nce
Date of impedance measurements
Device Status Summary
Status
Remarks
2.6.2 ission
The pulse generators provide real time transmission of the unfi e EGM) to the programmer. IEGMs SVC) and ventricle can be sim n th of 0.5 to 200 Hz. The IEG er via the prog ey are then displ kers on the programmer screen and printed on the ECG recorder. Likewise, intracardiac signals and markers identifying and sensed events are received via the programming head, and may be displayed on the programmer screen and printed on the ECG recorder. IEGM markers are available for all sensed and paced events.
To determin the automatic R-wave measurement function may be used.
Please refer to the appropriate technical manual for a description of marker signal operation.
2.6 ation
Sho prolonged if the high voltage capacitors remain uncharged for an extended period of time. Conditioning (or reforming) the capacitors by periodically charging them will help to ensure shorter charge times in those patients t preset to The p eset following an automatic or manual capacitor reform, or any device initiated maximum charging of the high voltage capacitors.
Real-time IEGM Transm
lter d intracardiac electrogram (I
from the proximal shock coil (
ulta eously recorded with a bandwid
Ms may be transmitted to the programm
ramming head positioned over the ICD. Th
ayed together with the surface ECG and mar
ventricular paced
e the amplitudes of intracardiac signals (R-waves)
.3 Capacitor Reform
ck charge times may be
hat do not regularly receive shock therapy. The ICD is
automatically re-form the capacitors every 3 months.
ca acitor reformation clock is r
Page 44
38 Cardiac Airbag Technical Manual
An automatic or manually initiated capacitor reform fully charges the capacitors and then allows the capacitors to drain off through the internal circuitry of the ICD. No shock will be delivered to the pati ation process the ICD will provide
ent. Throughout the reform bradycardia pacing support and tachyarrhythmia sensing and detection as programmed. If a tachyarrhythmia is detected during capacitor reformatio
n, the process is aborted and therapy
is available if required.
2.6.4 Patient and Implant Data
The Patient and Implant data screens allow
input of data regarding the patient name, demographics, implanting physician, date, devices implanted, location of the implant, and various conditions related to the patient. This information is transmitted to the ICD and resides in the device memory for later retrieval if needed.
Page 45
Cardiac Airbag Technical Manual 39
2.6.5 System Status
Various device parameters can be monitored through the Status section of the programmer screen. (See Figure data includes ICD information, charge circuit parameters, capacitor reformation data, battery status, and lead information. The system status screen presents a large variety of information about the Cardiac Airba
g ICDs including:
Serial number (always displayed after interrogation)
Software release
Device status
Battery status
- BOL (Begin of Life)
- ERI (Elective Replacement Indication)
- EOS
(End of Service)
Last charge event
- Date
- Energy
- Charge time
Total number of charges
Last R-wave measurements
Last pacing lead impedance (ventricle)
Last pacing threshold measurement with pulse width
(ventricle)
Last shock impedance measurement and date
4) Displayed
Page 46
40 Cardiac Airbag Technical Manual
Figure 4. Sys
tem Status
2.6.6 Holter Me
Imp a in the Holter memory. The
ort nt information is available with Holter m ration to provide the most criti i
2.6.6
The D rrhythmia epis e rrhythmia
emory has a preset configu
cal nformation to the physician.
.1 Episode List
IC stores essential diagnostic data about tachya
od s that may be used to optimize tachya
mory
detection and therapy parameters. This diagnostic data includes a therapy history and stored intracardiac electrograms.
Episode Details - Detailed information about each individual episode is presented as a table of events with the most recent episode listed first. Each IEGM segment can be viewed from the episode detail sub-menu by selecting the EGM button. From this screen, an IEGM can be expanded and scrolled to assist in a more accurate IEGM interpretation and a closer examination of specific segments. (See Figure
5)
Page 47
Cardiac Airbag Technical Manual 41
Figure 5. Episode List
Stored IEGM - The ICD can store up to 30 minutes of two
channel intracardiac electrogr and prehistory of the following
ams (IEGMs) including the history
events:
Detection
Redetection
Terminations
Delivered Shocks
The ICD can store IEGMs for the following events prior to ERI:
3 spontaneous VF episodes treated with shock therapy
Non-sustained VF episodes without shock thera
py
10 VT monitoring zone episodes
Induced episodes while the programmer wand is over
the implanted ICD
Following ERI declaration, no further EGMs are stored in the ICD. However, the episode counters continue to update until EOS is declared. (See Figure
6)
Page 48
42 Cardiac Airbag Technical Manual
Figure 6. Stored IEGM
2.6.7 Arrhythmia Induction Features
The D offers two arrhythmia induction methods
Cardiac Airbag IC for non-invasive EP testing. These include the following:
HF Burst Induction This feature consists of a large number of pulses delivered in rapid succession over a period of several seconds. The frequency of the pulses and the duration of the burst are defined by the user.
Sho
ck on T induction mode allows tachyarrhythmia induction by
means aced stim . lses (Nu e ization interval (R-S1) and
of a timed T wave shock delivered after a series of p
uli Energy of the T wave shock, number of pu
mb r S1) in the pulse train, synchron
the shock Coupling interval are all user programmable.
Page 49
Cardiac Airbag Technical Manual 43
2.6.8 Manual Shock
The Cardiac Airbag ICD can deliver a manual shock on demand through a programmer command in the EP test menu. To deliver a shock, place the wand over the device and select the Start Shock button. A confirmation menu will appear and the shock command will be delivered upon selecting the OK button in this screen. After each manual shock, the EP test screen will display the shock energy, lead impedance and charge time.
2.6.9 Test Shock
The Cardiac Airbag ICD can deliver a low-energy 1 joule (R­wave synchronous) test shock on demand through a programmer command in the EP test menu. This shock is designed to measure the shock impedance and test the integrity of the shock electrodes of an implanted ICD lead.
Page 50
44 Cardiac Airbag Technical Manual
3. Software Features
This section describes the features available with A-K00.0.U programmer software and the procedures necessary for interrogating and programming Cardiac Airbag ICDs. All references to Cardiac Airbag are also applicable to the Cardiac Airbag-T.
ws.
PLUS
or EPR 1000
Please refer to the TMS 1000 manuals for detailed descriptions of how to operate within the specific menus and windo
3.1 Follow-Up Assistant (FAST) Window
After interrogating the ICD, the programmer initially displays the Follow-Up Assistant window. (See Figure 7
PLUS
techn
)
ical
The Follow-up Assistant (FAST) is a program inco
e Cardiac Airbag applications for a user-defined guided
th
llow-up. The “guided” follow-up function was developed to allow
fo the user to be directed through the required program modules with the press of a single button.
Interrogation of the device
Measurement of the battery status
Interrogation of the episode counters
Measurement of the R-wave amplitude
Measurement of the ventricular pacing impedance
Measurement of pacing threshold - Threshold test
screen appears and waits for the user to perform the ventricular threshold test.
Printing of the follow-up data, which contains the results of the performed interrogations.
The routines are activated in the order displayed on the programmer screen. Functions belonging to a common topic are grouped together. Use the check boxes to control whether a specific procedure is performed or not.
When all of the functions have been completed the system opens the Follow-Up Assistant window displaying the measured values (See Figure 7 the measured values listed in the completed window.
). The Printed Follow-Up report contains
rporated into
Page 51
Cardiac Airbag Technical Manual 45
Figure 7. Follow-Up Assistant w
ith measured values
3.1.1 t Follow-up
To perform an interrogation without using the Follow-Up Assis n ate ICD. Once interrogation is completed all current programmed parameters are e
3.2
A general description the main function keys is presented below:
[VT
or deac y in the ICD. The user may select either Enable Detection or Disable Detection from the pop-up menu. Detection is automatically enabled following selection of the Enable Detection command. The user is required to confirm before the Disable Detection is tr
Interrogate ICD withou
ta t, select Skip Follow-Up Interrog
list d within the Parameter screen.
Main Function Keys
/VF Detection] - The VT/VF Detection key is used to activate
tivate the programmed VF therap
ansmitted.
Page 52
46 Cardiac Airbag Technical Manual
[Follow-Up] - The Follow-up Assistant (FAST) is incorporated
into the programmer applications to provide a user-defined system guided follow-up. The “guided” follow-up function was developed to allow consistent and quick actions by the user through a single command.
[Parameter] - The [Parameter] window serves for programming all parameters for sensing, detection and therapies.
[Status] - The [Status] window displays information on the interrogated ICD and the connected lead system. The window
is
grouped under two tabbed sections: Status and Measurement
Trend.
[Holter] - The [Holter] functions are grouped within the following
tabs: Episode List and Holter Configuration. Episode List displays a list of all stored episodes. Holter Configuration contains Holter configuration and a function for modifying the date and time setting of the programmer. Holter Time Setting clears the date and time from the implant memory and sets the internal clock of the implant to the time of the programming device.
[Tests] - The [Tests] are grouped within the follo Amplitude / Impedance, Threshold, and DFT Tes
wing tabs:
t. Each of
these tabs activates windows for various tests used at implant and follow-up procedures.
[Export] - The [Export] selection is used to copy the complete database of the implant to a separate interface and BIOTRONIK’s CDM 300
computer via a serial
0 software which is not be
available in the US.
[Options] - The [Options] window contains the following three tabs:
Options: for manually reformation of the high voltage capacitors contained within the ICD, to activate Home Monitoring, and for programmer control functions.
Patient Data: For storing information regarding the p
atient and
their physician.
Reset: To re-initialize the ICD. This feature is locked out by code, if it is necessary to reset the ICD, please contact your BIOTRONIK representative.
Page 53
Cardiac Airbag Technical Manual 47
[Emergency] - The [Emergency] function key is always present.
Activating this key produces a screen display that turns ventricular tachyarrhythmia Detection ON and OFF, activates emergency bradycardia pacing parameters, or activates a programmer commanded s
[Print] - This function key appears in the lower right side of the screen in windows with printing options. Activation immediately prints the relevant data from the open window.
hock.
of this key
3.3 Parameter Window
The [Pa sensing, detection and therapies. All programming of detection, therapies, pacing are completed from this Parameter screen (see Figure displayed only if VT/VF Detection is enabled. In order to provide easily understandable parameter displays, the numeric values of the detection parameters can be shown as an Interval (ms) or as a Rate (bpm).
Two ma parameter window regardless of the tab selected.
[Interrogate ICD] - This function key interrogates the ICD and loads the current programmed parameters into the programmer memory.
[Transmit] - Activation of this key transmits the displayed parameters as a permanent program to the ICD. This key also appears in the [Options] window as [Transmit Settings].
rameter] window serves for programming parameters for
8). Detection and Therapy parameters are
in function keys are located at the bottom of the
Page 54
48 Cardiac Airbag Technical Manual
Figure 8. [Parameter] Window
Page 55
Cardiac Airbag Technical Manual 49
Page 56
50 Cardiac Airbag Technical Manual
4. Sterilization and Storage
The ICD is shipped in a storage box, equipped with a quality control seal and product information label. The label contains the model specifications, technical data, serial number, use before date, and sterilization and storage information.
The ICD and its accessories have been sealed in a container and gas sterilized with ethylene oxide. To ensure sterility, the container should be checked for integrity prior to opening.
Page 57
Cardiac Airbag Technical Manual 51
Page 58
52 Cardiac Airbag Technical Manual
5. Implant Procedure
5.1 Implant Preparation
Prior to beginning the ICD implant procedure, ensure that all necessary equipment is available. The implant procedure requires the selected lead system (including sterile back-ups), the programmer with appropriate software, and the necessary cabling and accessories.
For TMS 1000 accessories are available:
PK44 - used to connect the TMS 1000 systems for complete testing of the lead systems during the implant procedure. The following adapters may be necessary:
Adapters PA-2/PA-3 - The PA-2 adapter is used to connect IS-1 compatible leads to the PK-44 cable. The PA-3 adapter is used to connect DF-1 compatible leads to the PK-44 cable.
Adapter PA-4 - used to connect the PK-44 cable to sensing and pacing leads while the stylet is still inserted.
Perform an interrogation of the ICD. Ensure programmer operation, nominal device parameters and battery status is appropriate for a new Cardiac Airbag ICD. Program Detection and Therapy to “Disabled” prior to handling the Cardiac Airbag ICD.
Sufficient training on the device and its associated components is required prior to implanting the ICD. For additional information, training and training materials contact your BIOTRONIK representative.
PLUS
based testing, the following cabling and
PLUS
to implanted lead
5.2 Lead System Evaluation
The ICD is mechanically compatible with DF-1 defibrillation lead connectors and IS-1 sensing and pacing lead connectors. IS-1, wherever stated in this manual, refers to the international standard, whereby leads and pulse generators from different manufacturers are assured a basic fit [Reference ISO 5841-3:1992]. DF-1, wherever stated in this manual, refers to the international standard [Reference ISO 11318:1993].
Page 59
Cardiac Airbag Technical Manual 53
Refer to the appropriate lead system technical
5.3 Opening the Sterile C
The Cardiac Airbag ICDs are packaged in two plastic containers, one within the other. Each is individually sealed and then sterilized with ethylene oxide.
Due to the double packing, the outside of the inner container is sterile and can be remove and placed on the sterile field.
d using standard aseptic technique
Peel off the sealing paper of the outer container as indicated by the arrow. Do not contaminate the inner tray.
ST_01
Take out the inner sterile tray by gripping the tab. Open the inner tray by peeling the sealing paper as indicated by the arrow.
ontai
manual.
ner
5.4 Pocket Preparation
Using standard surgical technique, create a pocket for the device in the patient’s pect implanted either below the subcutaneous tis tissue. The ICD should be implanted with th up. The leads should be tunneled or surgically brought into the device pocket. If lead tunneling is performed, re-evaluation of the baseline lead signals, after tunneling is recommended.
oral region. The device may be
sue or in the muscle
e etched side facing
Page 60
54 Cardiac Airbag Technical Manual
5.5 Lead to Device Connection
The Cardiac Airbag ICDs have been designed and are recommended for use with a defibrillation lead systems having one IS-1 connector for ventricular sensing and pacing and up to two DF-1 connectors for delivery of shock therapy. Figure depicts the configuration of the header ports on the Cardiac Airbag, where HV1 and HV2 are for DF-1 connectors, and V P/S is for IS-1 connectors.
9
Figure 9. Header Ports
Page 61
Cardiac Airbag Technical Manual 55
Refer to the following steps when connecting device.
1. Confirm that the setscrews are not protruding into the connector receptacles. To retract a setscrew, insert the enclosed torque wrench through the perforation in the self-sealing plug at an angle perpendicular to the lead connector until it is firmly placed in the setscrew. Rotate the wrench counterclockwise until the receptacle is clear of obstruction.
2.
Insert the lead connector into the connector port of the ICD without bending the lead until the connector pin becomes visible behind the setscrew. connector in this position. If necessary, apply silicone oil only to the o-rings on the connector (not the connector pin).
3. the
Insert the enclosed torque wrench through perforation in the self-sealing plug at an angle perpendicular to the lead connector until it is firmly placed in the setscrew.
4.
Securely tighten the setscrew of the connector clockwise with the torque wrench until torque limited by the wrench.
5. wrench, the
After carefully retracting the torque perforation will self-seal.
5.6 Blind Plug Co
nnection
the leads to the
Hold the
transmission is
he Cardiac Airbag ICDs come with a blind plug (pre inserted) in
T an unused header port. Refer to the following steps when connecting blind plugs to the device.
Page 62
56 Cardiac Airbag Technical Manual
1. Confirm that the setscrews are not protruding into the connector receptacles. To retract a setscrew, insert the enclosed torque wrench through the perforation in the self-sealing plug at an angle perpendicular to the lead connector until it is firmly placed in the setscrew. Rotate the wrench counterclockwise until the receptacle is clear of obstruction.
2.
Insert the blind plug into the connector port of the ICD until the conne setscrew.
3.
Insert the enclosed torque wrench through the perforation in the self-sealing plug at an angle perpendicular to the connector until it is firmly placed in the setscrew.
4. ely tighten the setscrew of the connector clockwise
Secur with the torque wrench until torque transmission is limited by the wrench.
5.
After carefully retracting the torque wrench, the perforation will self-seal.
5.7 P
rogram the ICD
ctor pin becomes visible behind the
Program ely treat the patient’s arrhythmias and lead system evaluation should be helpful in tailoring the various
arameters of the ICD to treat each individual patient. The
p detection and therapy status of the ICD testing purposes once all of the lead connectors have been securely fastened in the device header ports.
the ICD to appropriat
other therapy needs. The information obtained during the
may be activated for
5.8 Implant the ICD
The ICD may be placed in the pocket at this time. Place the device into the pocket with the etched side facing up. Carefully coil any excess lead length beside or above the ICD.
The pacing and sensing functions of the device should be evaluated. It is also recommended that at least one induction and device conversion be done prior to closing the pocket. This will ensure that the lead system has been securely connected to the device and has not changed position.
Page 63
Cardiac Airbag Technical Manual 57
Prio
r to surgically closing the pocket, the telemetry contact should commu surgica patient retransm
be evaluated to help ensure chronic programmer nication. Close the device pocket using standard l technique. As the final step at device implant and each
follow-up, the permanent program should be
itted to the ICD.
Complete the Medical Device Registration Form provided with the
ICD and return it to BIOTRONIK.
5.9 Suggested Cardiac Airbag Implant Procedure
Pre-O
perative steps
1. Check paper supply TMS 1000 . Ensure both the programmer date and time are
2
correct. (To change
the date and time, access
More Preferences PC)
3 G cable and verify the signal
. Connect the PK-44 EC
before the patient is prepped and draped.
4. Run a baseline ECG
5. Ensure that an external defibrillator is connected to
the patient.
PLUS
or EPR 1000
PLUS
Preparing the ICD
1. Select a Cardiac Airbag ICD package.
2. Place the programming wand over the ICD. A green
flashing light indicates a g
ood telemetry
communication.
3. Cardiac Airbag is
automatically interrogated with
the wand over the ICD.
4. Press Status, (See Figure
10) located on the right side of the screen. The button performs a complete interrogation of the ICD.
Verify Implant Status OK. Verify Battery Status OK. Verify Episode Gauge is full.
Page 64
58 Cardiac Airbag Technical Manual
Figure 10
. System Status
Performing a manual capacitor reformation
1. Press Options button.
2. Formation (See Figure
Select Start 11.) and then OK.
3. During charging, an ICD Charging displayed in the ICD Status box.
4. Return to Status screen.
5. arge Time.
Verify the ICD Ch
meter is
Page 65
Cardiac Airbag Technical Manual 59
Figure 11. Manual Cap Reform
Acc for lead and device-basedessing the TMS 1000
ing
test
PLUS
1. Press Options button.
2. In the Programmer area, press Implant List button to return to the main programmer screen.
3. Select TMS 1000 from the
Implant List.
Connecting the ventricular lead for testing
1. Connect the terminal pins of the ventricular lead to
the appropriate port of the PK-44 Cable.
Pace/Sense pin = PA-2 block
Shock pins = PA-3 block
Page 66
60 Cardiac Airbag Technical Manual
Measuring R-wave Amplitude
1. Select Intracardiac Measurements
2. Press and hold the Record Data (See Figure 12) button to measure the R-wave amplitude. There is an audible tone when the ICD has detected a sensed R-wave.
3. In general, R-wave amplitudes should be greater than 5 mV.
4. Press Print to obtain a printout of the measurement.
Figure 12. R-Wave Measurement
Page 67
Cardiac Airbag Technical Manual 61
Determining ventricular capture
thresholds
1. Verify VVI mode is highlighted.
2. Set the Lower Rate at 5 to 10 ppm above the patient’s intrinsic rate.
Press Start Pacing to begin threshold testing at the
3. displayed parame
ters.
4. Decrease Pulse Amplitudes (See Figure loss-of-capture occurs.
5. Press Stop Pacing to return to the patient’s intrinsic rhythm.
13) until
Figure 13. Pacing Threshold Test
6. In general, ventricular pacing thresholds should be
7. Press Print to obtain a printout of the threshold.
less than 1.0 V.
Page 68
62 Cardiac Airbag Technical Manual
Testing for diaphragmatic stimulation
1. Select 10.0 V in the Pulse Amplitude pop-up menu to test for diaphragmatic stimulation.
2.
Press Start Pacing.
3. Check for diaphragmatic stimulation.
4. eturn to the patient’s intrinsic
Press Stop Pacing to r rhythm.
Connecting the ICD to the leads
5. Verify VT/VF Detection is disabled.
6. Hand off the pre-programmed ICD to the implanting physician.
7. Disconnect the ventricular lead pins from the PK-44 cable and insert each lead pin into the appropriate ICD header port.
Pace/Sense pin = P/S V {IS-1} HV 1 Shock pin = HV 1 {distal coil} HV 2 Shock pin = HV 2 {proximal coil}
8. Tighten each set-screw using a BIOTRONIK torque wrench until a clicking sound is heard.
9. Insert the ICD into the pocket with etching facing up.
Page 69
Cardiac Airbag Technical Manual 63
Performing device-based testing measurements
1. Press Implant List button.
2. Place a sterile cover over the programming wand and instruct the implanting physician wand over the device
.
3. A green flashing light on the wand
to position the
indicates good
telemetry communication
4. Airbag is automatically interrogated.
Cardiac
5. Press Start Follow-Up to automatically interrogate the device. Battery status, sensing, lead
impedance, and e
pisode counters are
automatically verified.
6. Check R-wave Amplitudes (greater than 5 mV).
7. ing Impedance values are within an
Verify pac acceptable range (300-1000Ω).
8.
Verify Battery Status is OK.
9. test.
Perform a
10. a summary of
Press Resume Follow-up to print out
Pacing Threshold
these baseline measurements.
11. Enable VT/VF Detection.
12. ician.
Verify ICD settings with implanting phys
Performing 1 J test shock
1. Verify external defibrillation is available.
2. Verify testing sequence with implanting physician.
3. Press Tests button.
4. Select DFT Test tab.
5. Ensure wand is over the device and good telemetry communication is established.
6. Press 1 J Test Shock and press OK to deliver a synchronized low-energy shock (1 Joule test shock) to confirm adequate shock coil impedance.
7. Confirm Shock Impedance values are within an acceptable range (30-120Ω).
Page 70
64 Cardiac Airbag Technical Manual
DFT testing
1. Select method to induce ventricular fibrillation with the ICD: Shock on T wave or HF burst.
Press Print button to be
2. gin printing induction.
3. Press Start VT/VF Induction (See Figure OK to induce VF.
4. Observe the IEGM, markers, and ICD charging indicator on the programmer screen.
14) and
Figure 14. VF Induction Test
5. After successful delivery of therapy, press Print to
stop printing. (See Figure
15)
Page 71
Cardiac Airbag Technical Manual 65
Figure 15. VF Induction Test
6. Read Shock Data is automatically updated with shock values.
7. Press Print to print out the DFT Test Screen.
8. Select Holter and review EGM to ensure proper sensing during VF.
9. Repeat induction process for a second therapy success or until the DFT is found.
10. After testing is complete, select VT/VF Detection to disable detection an
d prevent any inappropriate shocks during the remainder of the implantation procedure.
Page 72
66 Cardiac Airbag Technical Manual
Final Programming
1. Verify final programming of the following parameters: detection intervals, pacing, sensing, and therapy.
2. Select Transmit to permanently program final parameters.
3. Verify VT/VF Detection is Enabled.
4. Select Parameter button and press Interrogate then Print to print out final programming and system status.
Page 73
Cardiac Airbag Technical Manual 67
Page 74
68 Cardiac Airbag Technical Manual
6. Follow-up Procedures
6.1 General Considerations
An ICD follow-up serves to verify appropriate function of the ICD system, and to optimize the programmable parameter settings.
In addition to evaluating the patient’s stored therapy history and electrograms, acute testing of sensing and pacing is recommended. As the final step during the patient follow-up, the programmed parameters should be verified with the physicians and the permanent program should be retransmitted to the ICD. ACC/AHA/NASPE Guidelines recommend the physician perform a patient follow-up visit every 3 months.
6.2 Suggested Cardiac Airbag Follow-Up Procedure
Setting up the programmer
1. Turn power ON.
2. Ensure the programmer date and time are correct.
3. To change the date and time, access More Preferences PC from the implant screen
4. Connect the patient to the surface ECG cable.
5. Place the programming wand over the ICD. A green flashing light indicates a good telemetry communication.
6. The Cardiac Airbag is automatically interrogated with the wand over the ICD.
Page 75
Cardiac Airbag Technical Manual 69
Starting Follow-Up Assistant
1. The software begins by displaying the Follow-Up Assistant screen. (See Figure
16) (To fully activate the Follow-Up Assistant program, ensure that all boxes are checked.)
2. Press Start Follow-Up, located in the lower left corner of the screen. This button begins the follow­up procedure by performing a complete interrogation of the ICD. Battery status, sensing, lead impedance, and episode counters are automatically verified.
Figure 16. Follow-up Assistant
Page 76
70 Cardiac Airbag Technical Manual
Determining the ventricular pa
cing threshold
1. Enter TestsThreshold (See Figure (automatic through Follow-up Assistant.)
Set Rate at 5 to 10 ppm above the patient’s intrinsic
2. rate.
3. If desired, turn Print Test ON to run paper during testing.
4.
Press Start Test to begin threshold testing at displayed parameters.
5. litudes until loss-of-capture
Decrease Test Amp occurs.
6. Press Stop Test. (Permanently programmed pacing values are immediately restored.)
7. Select the threshold value from the pop-up menu and press OK.
8. Select Resume Follow-up to return to the Follow- up Assistant screen.
9. A Follow-up Report will be printed automatically.
17)
Page 77
Cardiac Airbag Technical Manual 71
Figure 17. Pacing Threshold Test
Printing and analyzing detailed Holter data information
1. Press the Holter (See Figure
18) button to view
episodes.
2. From the Episode List, select the episode EGM you wish to view and/or print.
3. Press Print to obtain a printed record of all events since implantation.
Page 78
72 Cardiac Airbag Technical Manual
Figure 18. Holter EGM Episode
Verify system status
1. Press Follow-up button.
2. mplitudes (greater than 5 mV)
Check R A
3. Verify Pacing Impedance values are within an acceptable range (300-10
00Ω)
4. Verify Battery Status is OK.
Verifying remaining shocks to ERI
1. Press Status button. (See Figure
19)
2. Review the date of last charge event, charge time, delivered energy, and shock impedance.
Page 79
Cardiac Airbag Technical Manual 73
Figure 19. System Status
3. Verify remaining shocks to ERI by checking the level of the episode gauge.
4. ERI is reached when 3 treated VF episo occurred.
5. At ERI, contact your local BIOTRONIK representative.
des have
Making parameter changes and obtaining final printouts
1. Touch the parameter to be
2. Permanently program any changes by pressing
3. Interrogate the device and print final summary by
changed and select the
new value from the pop-up menu.
Transmit.
selecting the Print button from the Follow-Up screen.
Page 80
74 Cardiac Airbag Technical Manual
6.3 Longevity
The service time of an ICD can vary based on several factors, including the number of charge sequences, programmed parameters, number of tachyarrhythmias detected, relative amount of bradycardia pacing required, pacing lead impedance, storage time, battery properties, and circuit operating characteristics. For the Cardiac Airbag ICD, there are two methods for reaching ERI. One method is standard battery depletion over the life of the time with no delivered therapies and the other method is based on the number of treated VF episodes. Both methods are described in detail below.
Service time is the time from beginning of service (BOS) to the elective replacement indication (ERI). To assist the physician in determining the optimum time for ICD replacement, a replacement indicator is provided that notifies the user that replacement within a certain period of time is required. Upon reaching ERI, the battery has enough energy left to continue monitoring for at least three months along with the ability to deliver a minimum of six high-energy shocks. Upon reaching end of service (EOS) all tachyarrhythmia detection and therapy
disa d
is ble .
6.3.1 Standard ERI Method
The servi m beginning of service (BOS) to elective replac estimates rate of 50 ppm with a pulse width of
.5 ms an de of 2.4 volts and 500 ohm pacing
0
pedance and all shocks at maximum energy (30 joules) at
im 37C. The estimates for service time are based on a 24 month sh prior
elf life, therefore, if the Cardiac Airbag ICD is implanted to the what is de
ce times fro
ement indication (ERI) are listed below in Table 4
assume pacing
d pulse amplitu
24 month period, the service time may be longer than
picted in Table 4
.
. All
Page 81
Cardiac Airbag Technical Manual 75
In this table, it is assum treat ventricular tachyar
ed that the device delivers no shocks to
rhythmias, however, automatic capacitor reformations are equally spaced on a quarterly (every 3 months) basis throughout the life of the ICD. Therefore, the table starts at a minimum of 4 shocks per year, because capacitor reformations are equivalent to shocks. The estimates associated with 0% pacing support assume the ICD is sensing an intrinsic sinus rhythm at a rate of 70 bpm.
Table 4: Longevity Estimates
VVI Pacing
Support
Shocks Per Year
(includes cap reforms)
Years
12 3.5 10 3.7
100 %
8 3.9 6 4.1
4 4.4 12 3.7 10 3.9
50 %
8 4.1
6 4.4
4 4.7 12 3.8 10 4.0
15 %
8 4.3
6 4.6
4 4.9 12 3.9 10 4.0
0 %
8 4.4
6 4.7
4 5.0
After the ERI period the device is at EOS (End of Service) and requires explantation. Upon reaching EOS all tachyarrhythmia detection and therapy is disabled.
Page 82
76 Cardiac Airbag Technical Manual
6.3.2 Treated VF Episode ERI Method
The Cardiac Airbag ICD is designed to treat a limited number (3) of spontaneous VF episodes prior to reaching ERI. Induced VF episodes via the BIOTRONIK programmer that receive shock therapy with the programmer wand in place do not increment against the 3 treated VF episode limit. Only spontaneous VF episodes that receive shock ther episode count. Any VF episode that is detected by the ICD and the resulting charge is aborted prior to receiving a shock does not count against the 3 treated VF episode limit.
The er screen and printouts have been designed to
programm display the current status of the Cardiac Airbag ICD upon interrogation. Figure after ontaneous ther was delivered. ure
one sp apy Fig
20 displays the VF treated episode Gauge
displays the ERI status after the Cardiac Airbag has treated 3 VF episodes.
apy increment the treated VF
21
Figure 20. Status Following 1 VF Episode
Page 83
Cardiac Airbag Technical Manual 77
Figure 21. Status at ERI
Upon reaching ERI, the battery has enough energy left to continue monitoring for at least three months and to deliver
a minimum of six high energy shocks. The estimates associated with duration of ERI assume the ICD is sensing an intrinsic sinus rhythm at a rate of 70 bpm. After this period the device is at EOS and should be explanted. Upon reaching EOS
all
tachyarrhythmia detection and therapy is disabled.
Page 84
78 Cardiac Airbag Technical Manual
6.4 Explantation
Explanted ICDs, lead systems, and accessories may not be reused. Please complete the appropriate out of service (OOS) form and return it to BIOTRONIK with the explanted devices. All explanted devices should be sent packaged in a bio-hazard container. These may be delivered to either the local BIOTRONIK representative or the BIOTRONIK home office for expert disposal. Contact BIOTRONIK if you need assistance with returning explanted devices. If possible, the explanted devices should be cleaned with a sodium-hyperchlorine solution of at least 1% chlorine and then washed with water prior to shipping.
The pulse generator should be explanted before the cremation of a deceased patient.
Page 85
Cardiac Airbag Technical Manual 79
Page 86
80 Cardiac Airbag Technical Manual
7. Technical Specifications
The following are the technical specifications for the Cardiac Airbag ICDs. The ranges are presented in the format:
x…(y)…z
where x = the lowest value, y = the increment, and z = the largest value.
NOTE:
Values depicted in gray are preset in the ICD and are not programmable values.
Mechanical Properties
Parameter Value Range
Dimensions 67 x 55 x 13 mm Conducting Surface Area 67 cm2 Volume 39 cc Mass 73 g Housing Material Titanium
Header Material Epoxy resin Seal Plug Material Silicone Cardiac Airbag and Cardiac Airbag-T Lead Ports
1 x 3.2 mm IS-1 Bipolar 2 x 3.2 mm DF-1
Page 87
Cardiac Airbag Technical Manual 81
Parameters - Tachyarrhythmias
Parameter Value Range
Detection Parameters for VT Monitoring Zone
Rate (VT Monitoring) OFF; 270…(10)…600 ms
100…
222 bpm
Interval Counter: Initial
16 Detection Interval Counter: Redetection 12 Onset 20 % (adaptive) Stability ±24 ms (absolute)
Detection and Redetection Parameters for VF Zone
Inte OFF; 200 …(10)…400 ms
rval / Rate
150…300 bp
m Number of X 8 Number of Y 12 Termination Detection 12 in 16
Shock Therapy
Parameter Value Range
Number of Shocks 6...(1)...8 (VF) 1st Shock Energy ules 20, 30 Jo Further Shocks 30 Joules Shock Waveform Biphasic Confirmation ON Polarity Normal
Sensing
Parameter Value Range
Sensitivity Standard
Enhanced T wave suppression Enhanced VF sensitivity Free (This feature is locked-out in the US)
Minimum Threshold 0.5…(0.1)…2.5 mV
Page 88
82 Cardiac Airbag Technical Manual
Bradycardia Therapy
Parameter Value Range
Mode VVI, VVIR, OVO (OFF) Rate 30…(5)…120 ppm Amplitude 7.5 V 0.2…(0.1)…6.2, Pulse Width 0.5, 1.0, 1.5 ms Maximum Sensor Rate 100, 125 ppm Sensor Gain 4 Sensor Threshold mean Rate Increase 2 ppm/s Rate Decrease 0.4 ppm/s
Post-Shock Brad ycardia Therapy
Parameter Value Range
Mode VVI Rate 70 ppm Amplitude 7.5 V Pulse Width 1.5 ms Duration 30 seconds
Home Monitoring (Ca Only) rdiac Airbag-T
Parameter Value Range
Home Monitoring Event Report On Repeated Interval 60 minutes
OFF, ON
Page 89
Cardiac Airbag Technical Manual 83
Federal Communications Commission Disclosure
The Belos uipped with an RF wireless communications. T
not cause harmful interference to 4 band in the Meteoro tters and
06.000 MHz
r to communicate w Meteorological
eceivers used eather data), the
S oration S and must accept
atellite, or the Earth Expl atellite Services interference that may be caused b aids, including interference t ired operation nsmitter shall be used only
hat may cause undes
i he FCC Rule g the Medical Implant
n accordance w
C Service. Analog and digital voice communications are
ommunications
prohibited. Although this transmitter has been approved by the Federal
ommunications Commission, there is no guarantee that it will not
C receive interfer ion from this transmitter interference.
The FCC : PG6
-T ICD is eq transmitter for his transmitter der the
Implant Communications S nd must
l
ith t
ence or that any particular transmiss
will be free from
ID number for this device is BELOS-T.
is authorized by rule un
95) aMedica ervice (47 CFR Part
s 00.150 -
tations operating in the 4
logical Aids (i.e., transmi
y such
. This tra
s governin
Page 90
84 Cardiac Airbag Technical Manual
Page 91
Cardiac Airbag Technical Manual 85
Appendix A
Connector Compatibility
Cardiac Airbag ICDs are indicated for use only with commercially available BIOTRONIK bipolar ICD lead systems or other lead systems with which it has been tested. The Cardiac Airbag family of ICDs is mechanically compatible with:
IS-1 sensing/pacing lead connectors
DF-1 defibrillation lead connectors.
The Cardiac Airbag and Cardiac Airbag-T ICDs have a single IS-1 header port and two DF-1 header ports.
Page 92
y
Distributed by:
BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035-5369 (800) 547-0394 (24-hour) (503) 635-9936 (FAX)
M4088-A 3/03 German
Manufactured by:
BIOTRONIK GmbH & Co.
Woermannkehre 1
D-12359 Berlin
Loading...