Biotronik BELOS-T User Manual

(Draft)
Cardiac Airbag /
Cardiac Airbag-T
Family of Implantable Cardioverter
Defibrillators and Software Cartridge for
TMS 1000
PLUS
and EPR 1000
PLUS
Technical Manual
X-ray Identification
Cardiac Airbag/Cardiac Airbag-T
Implantable Cardioverter Defibrillator
Inside the housing, top left-hand side:
Year of manufacture
X-Ray identification
EI
CAUTION
Federal (U.S.A.) law restricts this device to sale by, or on the order of, a physician.
2003 BIOTRONIK, Inc., all rights reserved.
Cardiac Airbag Technical Manual i
Contents
1. General...........................................................................1
1.1 System Description....................................................1
1.2 Indications and Usage ...............................................2
1.3 Contraindications .......................................................2
1.4 Warnings and Precautions.........................................3
1.4.1 Sterilization, Storage, and Handling ..................3
1.4.2 Device Implantation and Programming..............4
1.4.3 Lead Evaluation and Connection.......................6
1.4.4 Follow-up Testing...............................................7
1.4.5 Pulse Generator Explant and Disposal..............8
1.4.6 Hospital and Medical Hazards ...........................9
1.4.7 Home and Occupational Hazards......................11
1.4.8 Cellular Phones..................................................11
1.4.9 Electronic Article Surveillance (EAS).................12
1.4.10 Home Appliances ...............................................12
1.5 Adverse Events..........................................................13
1.5.1 Potential Adverse Events...................................13
1.5.2 Observed Adverse Events .................................14
1.6 Clinical Studies ..........................................................17
1.6.1 Patients Studied.................................................17
1.6.2 Methods .............................................................18
1.6.3 Results ...............................................................18
1.7 Patient Selection and Treatment ...............................20
1.7.1 Individualization of Treatment............................20
1.7.2 Specific Patient Populations ..............................20
1.8 Patient Counseling Information .................................20
1.9 Evaluating Prospective ICD Patients.........................21
2. Device Features ............................................................22
2.1 Sensing ......................................................................22
2.1.1 Ventricular Sensitivity Settings ..........................22
2.1.2 Minimum Ventricular Threshold .........................25
2.2 Ventricular Tachyarrhythmia Detection ......................25
2.2.1 VF Classifications ..............................................26
2.2.2 VT Interval Counters ..........................................26
2.2.3 VT Classification ................................................26
2.2.4 Onset and Stability.............................................27
ii Cardiac Airbag Technical Manual
2.3 Tachyarrhythmia Redetection ....................................28
2.3.1 VT Redetection ..................................................28
2.3.2 VF Redetection ..................................................28
2.3.3 Tachyarrhythmia Termination ............................28
2.4 Tachyarrhythmia Therapy ..........................................28
2.4.1 Shock Therapy...................................................29
2.5 Bradycardia Therapy .................................................31
2.5.1 Bradycardia Pacing Modes................................31
2.5.2 Basic Rate..........................................................31
2.5.3 Rate Adaptation .................................................31
2.5.4 Gain and Threshold ...........................................32
2.5.5 Rate Increase / Decrease ..................................32
2.5.6 Maximum Sensor Rate ......................................32
2.5.7 Pulse Amplitude .................................................32
2.5.8 Pulse Width........................................................33
2.5.9 Noise Response.................................................33
2.5.10 Post Shock Pacing.............................................33
2.6 Special Features........................................................33
2.6.1 Home Monitoring (Cardiac Airbag-T Only) ........33
2.6.2 Real-time IEGM Transmission...........................37
2.6.3 Capacitor Reformation .......................................37
2.6.4 Patient and Implant Data ...................................38
2.6.5 System Status....................................................39
2.6.6 Holter Memory ...................................................40
2.6.7 Arrhythmia Induction Features...........................42
2.6.8 Manual Shock ....................................................43
2.6.9 Test Shock .........................................................43
3. Software Features.........................................................44
3.1 Follow-Up Assistant (FAST) Window.........................44
3.1.1 Interrogate ICD without Follow-up .....................45
3.2 Main Function Keys ...................................................45
3.3 Parameter Window ....................................................47
4. Sterilization and Storage..............................................50
5. Implant Procedure ........................................................52
5.1 Implant Preparation ...................................................52
5.2 Lead System Evaluation ............................................52
5.3 Opening the Sterile Container ...................................53
5.4 Pocket Preparation ....................................................53
5.5 Lead to Device Connection .......................................54
Cardiac Airbag Technical Manual iii
5.6 Blind Plug Connection ...............................................55
5.7 Program the ICD........................................................56
5.8 Implant the ICD..........................................................56
5.9 Suggested Cardiac Airbag Implant Procedure ..........57
6. Follow-up Procedures ..................................................68
6.1 General Considerations.............................................68
6.2 Suggested Cardiac Airbag Follow-Up Procedure......68
6.3 Longevity....................................................................74
6.3.1 Standard ERI Method ........................................74
6.3.2 Treated VF Episode ERI Method.......................76
6.4 Explantation ...............................................................78
7. Technical Specifications ..............................................80
Appendix A...........................................................................85
iv Cardiac Airbag Technical Manual
Cardiac Airbag Specifications
Battery Voltage: 6.3 Volts Maximum Shock Energy: 30 joules Defibrillation Lead Ports Two DF-1 (3.2 mm) Pacing Lead Ports One IS-1 (3.2 mm) Dimension: 55 x 67 x 13 mm Volume: 39 cc Mass: 73 g
Housing Material: Titanium Header Material: Epoxy Resin Sealing Plug Material: Silicone Battery Composition Li / MnO2
Cardiac Airbag Technical Manual 1
1. General
1.1 System Description
The Cardiac Airbag family of Implantable Cardioverter Defibrillators (ICDs) detects and treats ventricular tachyarrhythmias as well as provides rate adaptive bradycardia pacing support. The ICDs are designed to collect diagnostic data to aid the physician’s assessment of a patient’s condition and the performance of the implanted device. The Cardiac Airbag ICDs are specifically designed to have reduced complexity for implant and follow-up, yet provide essential therapies for conversion of life threatening ventricular tachyarrhythmias.
There are 10 programmable parameters to simplify the implant procedure, and detailed diagnostic information is stored for up to 10 ventricular tachycardia (VT) episodes and 3 treated ventricular fibrillation (VF) episodes. There are 30 minutes of single-channel IEGM storage available to record spontaneous and induced ventricular tachyarrhythmias. The Cardiac Airbag is restricted to storage of diagnostic information up to and including 3 treated ventricular fibrillation episodes.
The Cardiac Airbag ICDs provide therapy for ventricular tachyarrhythmias with programmable defibrillation therapy. The ICDs provide high energy biphasic shocks with the first shock having with programmable energies of 20 or 30 joules and up to 8 shocks per VF episode. The remaining 7 shocks in the therapy progression are pre-set at 30 joules.
The Cardiac Airbag family of ICDs includes the following members:
Cardiac Airbag provides therapies for ventricular tachyarrhythmias and single chamber rate adaptive bradycardia pacing support.
Cardiac Airbag-T is identical to the Cardiac Airbag with the added functionality of BIOTRONIK’s Home Monitoring system. The Home Monitoring System enables automatic exchange of information about a patient’s cardiac status from the implant to the physician remotely.
2 Cardiac Airbag Technical Manual
The Cardiac Airbag and Cardiac Airbag-T have two DF-1 defibrillation / cardioversion and one IS-1 pacing/sensing header ports. IS-1 refers to the international standard whereby leads and generators from different manufacturers are assured a basic fit [Reference ISO 5841-3:1992]. DF-1 refers to the international standard for defibrillation lead connectors [Reference ISO 11318:1993].
External devices that interact with and test the implantable devices are also part of the ICD System. These external devices include the TMS 1000 and the EPR 1000 These programmers are used to interrogate and program the ICDs. In addition, the programmer software is used to perform the interrogation and programming of the ICDs during implant and follow-up testing.
PLUS
PLUS
Tachyarrhythmia Monitoring System
Programming and Monitoring System.
1.2 Indications and Usage
The Cardiac Airbag Implantable Cardioverter Defibrillators (ICDs) are intended to provide ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.
1.3 Contraindications
Do not use the Cardiac Airbag Implantable Cardioverter Defibrillators (ICDs) in patients:
Whose ventricular tachyarrhythmias may have transient or reversible causes including:
- acute myocardial infarction
- digitalis intoxication
- drowning
- electrocution
- electrolyte imbalance
- sepsis
- hypoxia
Patients with incessant VT of VF
Patients with unipolar pacemaker
Patients whose only disorder is brady arrhythmia or
atrial arrhythmia
Cardiac Airbag Technical Manual 3
1.4 Warnings and Precautions
ATP (Anti-Tachycardia Pacing) – The Cardiac Airbag ICD
does not provide ATP therapy. Do not implant this ICD in patients with documented ventricular tachycardias unless high energy defibrillation is desired for treatment of the ventricular arrhythmia.
MRI (Magne
to MRI device scanning. Strong magnetic fields may damage the device and cause injury to the patient.
tic Resonance Imaging) - Do not expose a patient
Electrical Isolation - To prevent ina
induction, electrically isolate the patient during the implant procedure from potentially hazardous leakage currents.
Lead Systems - The use of another manufacturer’s I system may cause potential adverse consequences such as under sensing of cardiac activity and failure to deliver necessary therapy.
Resuscit
unless an alternate source of patient defibrillation such as an external defibrillator is readily available. In order to implant the ICD system, it is necessary to induce and convert the patient’s ventricular tachyarrhythmias.
Unwanted Shocks – Always Therapy status to DISABLED prior to handling the device to prevent the delivery of serious shocks to the patient or the person handling the device during the implant procedure.
Rate-Adaptive Pacing – Use rate-adaptive pacing with care patients unable to tolerate increased pacing rates.
ation Availability - Do not perform induction testing
program the VT/VF Detection and
dvertent arrhythmia
CD lead
in
1.4.1 Sterilization, Storage, and Handling
Device Packaging - Do not use the device if the
packaging is wet, punctured, opened or damaged because the integrity of the sterile packaging may be compromised. Return the device to BIOTRONIK.
device’s
Re-sterilization - Do not r devices.
e-sterilize and re-implant explanted
4 Cardiac Airbag Technical Manual
Storage (temperature) - Store the device between 5° to 55°C
(41° - 131° F) because temperatures outside this range could damage the device.
Storage (magnets) - To avoid damage to the device, store the device in a clean area, away from magnets, kits containing magnets, and sources of electromagnetic interference (EMI).
Temperature Stabilization - Allow the device to reach room temperature before programming or implanting the device because temperature extremes may affect initial device function.
Use Before Date - Do not implant the device after the USE BEFORE DATE because the device may have reduced longevity.
1.4.2 Device Implantation and Programming
Blind Plug - A blind plug must be inserted and firmly conn
into any unused header port to prevent chronic fluid influx and possible shunting of high energy therapy.
Capacitor Reformation - Infrequent charg capacitors may extend the charge times of the ICD. The capacitors may be reformed manually, or the ICD may be programmed to reform the capacitors automatically. For further information, please refer to Section 2.6.3
Connector Compatibility - ICD and lead system compatibility should be confirmed prior to the implant procedure. Consult your BIOTRONIK representative regarding lead/pulse generator compatibility prior to the implantation of an ICD system. For further information, please refer to Appendix A
ERI (Elective Replacement Indicator) - Upon reaching ERI, the battery has sufficient energy remaining to continue monitoring for at least three months and to deliver a minimum of six 30 joule shocks. After this period, tachyarrhythmia detection and therapy will proceed until EOS is declared. Bradycardia functions are still active at programmed values until the battery voltage drops below 3.0 volts.
ing of the high voltage
, Capacitor Reforming.
.
ected
Cardiac Airbag Technical Manual 5
Magnets - Positioning of a magnet or the programming wand over the ICD will suspend tachycardia detection and treatment. The minimum magnet strength required to suspend tachycardia treatment is 1.8 mT. When the magnet strength decreases to less than 1 mT, the reed contact is reopened.
Pacemaker/ICD Interaction - In situations where an ICD and a pacemaker are implanted in the same patient, interaction testing should be completed. If the interaction between the ICD and the pacemaker cannot be resolved through repositioning of the leads or reprogramming of either the pacemaker or the ICD, the pacemaker should not be implanted (or explanted if previously implanted).
Programmed Parameters – Program the device parameters to appropriate values based on the patient’s specific arrhythmias and condition.
Programmers - Use only BIOTRONIK programmers to communicate with the device (TMS 1000 EPR 1000
Sealing Sy m
PLUS
).
ste - Failure to properly insert the torque wrench
PLUS
, or
into the perforation at an angle perpendicular to the connector receptacle may result in damage to the sealing system and its self-sealing properties.
Defibrillation Thresho
ld - Be aware that the changes in the
patient’s condition, drug regimen, and other factors may change the defibrillation threshold (DFT) which may result in non­conversion of the arrhythmia post-operatively. Successful conversion of ventricular fibrillation or ventricular tachycardia during arrhythmia conversion testing is no assurance that conversion will occur post-operatively.
Manual Shocks – User-commanded s
hocks may be withheld if the ICD is already busy processing a manual command or the Battery Status is low.
Charge Time - When
preparing a high energy shock the charge circuit stops charging the capacitors after 16 seconds, and delivers the stored energy as shock therapy. After the device reaches ERI the stored energy may be less than 30 joules per shock.
6 Cardiac Airbag Technical Manual
Shock Impedance - If the shock impedance is less than twenty-
five ohms, reposition the lead system to allow a greater distance between the electrodes. Never implant the device with a lead system that has measured shock impedance as less than twenty-five ohms. Damage to the device may result.
Programming Wand - Throughout the EP Test session, the programming wand must be positioned and remain directly over the device. If appropriate arrhythmia detection does not occur shortly after induction, remove the programming wand from the ICD and perform external defibrillation.
Data Transmission - Data collection and transmission may take up to 30 seconds. The ICD cannot be reprogrammed during this time even if the [Emergency] key is pressed. Remove the programming wand immediately to restore the permanent program.
EP Test Functions - Ensure that cardiac resuscitation equipment is available during all EP Test Function operations. Physicians should be trained and experienced in tachyarrhythmia induction, conversion protocols, and have adequate training and experience with this device prior to use.
Potential side effects include:
Non-terminable arrhythmia’s that result in death
Complications from hypoxia due to prolonged
arrhythmia’s
Arrhythmia induction that requires cardioversion or defibrillation
Arrhythmia induction that requires pharmacologic treatment, to which the patient could have an adverse reaction
1.4.3 Lead Evaluation and Connection
Capping Leads - If a lead is abandoned rather than removed, it
must be capped to ensure that it is not a pathway for currents to or from the heart.
Gripping Leads - Do not grip the lead with surgical instruments or use excessive force or surgical instruments to insert a stylet into a lead.
Cardiac Airbag Technical Manual 7
Kinking Leads - Do not kink leads. This may cause additional stress on the leads that can result in damage to the lead.
Liquid Immersion - Do not immerse leads in mineral oil, silicone oil, or any other liquid.
Short Circuit - Ensure that none of the lead electrodes are in contact (a short circuit) during delivery of shock therapy as this may cause current to bypass the heart or cause damage to the ICD system.
Suturing Leads - Do not suture directly over the lead body as this may cause structural damage. Use the appropriate suture sleeve to immobilize the lead and protect it against damage from ligatures.
Tricuspid Valve Bioprosthesis - Use ventricular transvenous leads with caution in patients with a tricuspid valvular bioprosthesis.
Setscrew Adjustment – Back-off the setscrew(s) prior to insertion of lead connector(s) as failure to do so may result in damage to the lead(s), and/or difficulty connecting lead(s).
Cross Threading Setscrew(s) – To prevent cross threading the setscrew(s), do not back the setscrew(s) completely out of the threaded hole. Leave the torque wrench in the slot of the setscrew(s) while the lead is inserted.
Tightening Setscrew(s) – Do not overtighten the setscrew(s). Use only the BIOTRONIK supplied torque wrench.
Sealing System – Be sure to properly insert the torque wrench into the perforation at an angle perpendicular to the connector receptacle. Failure to do so may result in damage to the plug and its self-sealing properties.
1.4.4 Follow-up Testing
Defibrillation Threshold - Be aware that changes in the
patient’s condition, drug regimen, and other factors may change the defibrillation threshold (DFT), which may result in non­conversion of the arrhythmia post-operatively. Successful conversion of ventricular fibrillation or ventricular tachycardia during arrhythmia conversion testing is no assurance that conversion will occur post-operatively.
8 Cardiac Airbag Technical Manual
Resuscitation Availability - Ensure that an external defibrillator
and medical personnel skilled in cardiopulmonary resuscitation (CPR) are present during post-implant device testing should the patient require external rescue.
Safe Program – Within the EP Test screen, pressing the “Safe Program” key on the programmer head does not immediately send the safe program to the ICD. Pressing the “Safe Program” key activates the emergency function screen touch is required to send the safe program to the ICD.
Date and Time Values - If date and time values are incorrect, the system may, as a result, generate false system status information for the implant.
Impedance Measurement - During the impedance measurement with high stimulation amplitudes, nerve or skeletal muscles may be briefly stimulated.
Threshold Test - A minimum 2:1 voltage safety margin should be permanently programmed any time capture thresholds a assessed. Monitor the ECG display closely with pacer­dependent patients. The test should be terminated immediately upon loss of capture.
screen, but an additional
re
Inadvertent Programming - The programmer utilizes a touch sensitive screen for menu selections. Care must be used to avoid inadvertent menu selection screen.
by accidentally touching the
1.4.5 Pulse Generator Explant and Disposal
Device Incineration – Never incinerate the ICD due to the
potential for explosio cremation.
Explanted Devices – Return all explanted devices to BIOTRONIK.
Unwante
DISABLED prior to handling the device to prevent the deliv serious shocks to the patient or the person handling the during the implant procedure.
d Shocks – Always program the therapy status to
n. The ICD must be explanted prior to
ery of
device
Cardiac Airbag Technical Manual 9
1.4.6 Hospital and Medical Hazards
Electromagnetic interference (EMI) signals present in hospital and medical environments may affect the function of any ICD or pacemaker. The ICD is designed to selectively filter out EMI noise. However, due to the variety of EMI signals, absolute protection from EMI is not poss
The ICD system should have detection and prior to performing any of the following medica addition, the ICD should be checked after the procedures to assure proper programming:
Diathermy - Diathermy therapy is not recommended for ICD patients due to possible heating effects of the pulse generator and at the implant site. If diathermy therapy must be used, should not be applied in the immediate vicinity of the pulse generator or lead system.
Electrocautery - Electrosurgical cautery could induce ventricular arrhythmias and/or fibrillation, or damage. If use of electrocautery is necessary, the current path and ground plate should be kept as far away from the pulse generator and leads as possible (at least 6 inches (15 cm)).
ible with this or any other ICD.
therapy disabled
l procedures. In
it
or may cause device malfunction
10 Cardiac Airbag Technical Manual
External Defibrillation - The device is protected against energy
normally encountered from external defibrillation. However, any implanted device may be damaged by external defibrillation procedures. In addition, external defibrillation may also result in permanent myocardial damage at the electrode-tissue interface as well as temporary or permanent elevated pacing thresholds. When possible, observe the following precautions:
Position the adhesive electrodes or defibrillation paddles of the external defibrillator anterior-posterior or along a line perpendicular to the axis formed by the implanted device and the heart.
Set the energy to a level not higher than is required to achieve defibrillation.
Place the paddles as far as possible away from the implanted device and lead system.
After delivery of an external defibrillation shock, interrogate the ICD to confirm device status and proper function.
Lithotripsy - Lithotripsy may damage the ICD. If lithotripsy must be used, avoid focusing near the ICD implant site.
MRI (Magnetic Resonance Imaging) - Do not expose a patient to MRI device scanning. Strong magnetic fields may damage the device and cause injury to the patient.
Radiation - High radiation sources such as cobalt 60 or gamma radiation should not be directed at the pulse generator. If a patient requires radiation therapy in the vicinity of the pulse generator, place lead shielding over the device to prevent radiation damage and confirm its function after treatment.
Radio Frequency Ablation - Prior to performing an ablation procedure, deactivate the ICD during the procedure. Avoid applying ablation energy near the implanted lead system whenever possible.
Cardiac Airbag Technical Manual 11
1.4.7 Home and Occupational Hazards
Patients should be directed to avoid devices that generate strong electromagnetic interference (EMI) or magnetic fields. EMI could cause device malfunction or damage resulting in non-detection or delivery of unneeded therapy. Moving away from the source or turning it off will usually allow the ICD to return to its normal mode of operation.
The following equipment (and similar devices) may affect normal ICD operation: electric arc or resistance welders, electric melting furnaces , r a d i o / t e l e v i s i o n and radar t r a n s m i t t e r s , power-generating facilities, high-voltage transmission lines, and electrical ignition systems (of gasoline-powered devices) if protective hoods, shrouds, etc., are removed.
1.4.8 Cellular Phones
Testing has indicated there may be a potential interaction between cellular phones and BIOTRONIK ICD systems. Potential effects may be due to either the cellular phone signal or the magnet within the telephone and may include inhibition of therapy when the telephone is within 6 inches (15 centimeters) of the ICD, when the ICD is programmed to standard sensitivity.
Patients having an implanted BIOTRONIK ICD who operate a cellular telephone should:
Maintain a minimum separation of 6 inches (15 centimeters) between a hand-held personal cellular telephone and the implanted device.
Set the telephone to the lowest available power setting, if possible.
Patients should hold the phone to the ear opposite the side of the implanted device. Patients should not carry the telephone in a breast pocket or on a belt over or within 6 inches (15 centimeters) of the implanted device as some telephones emit signals when they are turned ON, but not in use (i.e., in the listen or stand-by mode). Store the telephone in a location opposite the side of implant.
12 Cardiac Airbag Technical Manual
Based on results to date, adverse effects resulting from interactions between cellular telephones and implanted ICDs have been transitory. The potential adverse effects could include inhibition or delivery of additional therapies. If electromagnetic interference (EMI) emitting from a telephone does adversely affect an implanted ICD, moving the telephone away from the immediate vicinity of the ICD should restore normal operation. A recommendation to address every specific interaction of EMI with implanted ICDs is not possible due to the disparate nature of EMI.
1.4.9 Electronic Article Surveillance (EAS)
Equipment such as retail theft prevention systems may interact with pulse generators. Patients should be advised to walk directly through and not to remain near an EAS system longer than necessary.
1.4.10 Home Appliances
Home appliances normally do not affect ICD operation if the appliances are in proper working condition and correctly grounded and shielded. There have been reports of the interaction of electric tools or other external devices (e.g. electric drills, older models of microwave ovens, electric razors, etc.) with ICDs when they are placed in close proximity to the device.
Cardiac Airbag Technical Manual 13
1.5 Adverse Events
1.5.1 Potential Adverse Events
The following is a list of the potential risks that may occur with this device:
Acceleration of arrhythmias
Air embolism
Bleeding
Chronic nerve damage
Erosion
Excessive fibrotic tissue growth
Extrusion
Fluid accumulation
Formation of hematomas or cysts
Inappropriate shocks
Infection
Keloid formation
Lead abrasion and discontinuity
Lead migration / dislodgment
Myocardial damage
Pneumothorax
Shunting current or insulating myocardium during
defibrillation with internal or external paddles
Potential mortality due to inability to defibrillate or pace
Thromboemboli
Venous occlusion
Venous or cardiac perforation
14 Cardiac Airbag Technical Manual
Patients susceptible to frequent shocks despite antiarrhythmic medical management may develop psychological intolerance to an ICD system that may include the following:
Dependency
Depression
Fear of premature battery depletion
Fear of shocking while conscious
Fear that shocking capability may be lost
Imagined shocking (phantom shock)
There may be other risks associated with this device that are currently unforeseeable.
1.5.2 Observed Adverse Events
A clinical study of the Phylax XM involved 155 devices implanted in 154 patients with cumulative implant duration of 1286 months (mean implant duration 8.3 months). This clinical study was performed with the Phylax XM and Phylax 06 ICDs, which are earlier versions of the Cardiac Airbag ICDs. The observed adverse events are applicable because the Cardiac Airbag ICD is a downsized version of the Phylax XM with rate adaptive pacing capabilities.
NOTE:
The Phylax XM ICD is an earlier generation of BIOTRONIK devices. The Cardiac Airbag family is based upon the Phylax XM and other BIOTRONIK ICDs (i.e., Belos VR and Belos VR-T).
There were a total of five deaths during the course of the trial; none of the deaths were judged by the clinical study investigator to be device related. Heart failure was a major factor in two deaths. The other three deaths were related to renal failure, lung disease, and septic shock secondary to an ischemic bowel, respectively. All five of the deaths occurred more than one month post implant.
Cardiac Airbag Technical Manual 15
Two ICDs were explanted during the trial. One was secondary to the patient being unable to tolerate further testing required by the clinical protocol. The other was secondary to a systemic infection; the patient was subsequently implanted with another device.
Table 1 provides a summary of the adverse events that were reported during the clinical study regardless of whether or not the event was related to the ICD system. A complication is defined as a clinical event that results in invasive intervention, injury, or death. An observation is defined as a clinical event that does not result in invasive intervention, injury, or death.
Table 1: Reported Adverse Events (AEs)
Number of Patients = 154, Number of Patient-Years = 107.1
Event # of pts
with AEs
Complications (total) 7 8 0.07 4.5%
% of
pts
with
AEs
# of
AEs
AE/
pt-
yrs
Lead repositioning 2 1.3% 2 0.02 Hematoma 1 0.6% 1 0.01 Systemic infection 1 0.6% 1 0.01 Explant (did not to tolerate
1 0.6% 1 0.01 testing) Insertion of separate sensing 1 0.6% 2 0.02 lead ICD/lead connection 1 0.6% 1 0.01
Observations (total) 79 51.3% 89 0.83
Inappropriate therapy (SVT) 18 11.7% 20 0.19 ICD response to magnet in
1
wand Software messages and 11 7.1% 13 0.12
2
errors
13 8.4% 15 0.14
Increased pacing threshold 7 4.5% 9 0.08 Decreased R-wave amplitude
Frequent VT 5 3.2% 5 0.05
7 4.5% 7 0.07
Oversensing 3 1.9% 3 0.03 TMS 1000 difficulties3 3 1.9% 3 0.03 VT below rate cut-off 2 1.3% 3 0.03
16 Cardiac Airbag Technical Manual
Event # of pts
with AEs
High DFT’s 1 0.6% 2 0.02
% of
pts
with
AEs
# of
AEs
AE/
pt-
yrs
Minor stroke 1 0.6% 1 0.01 Renal failure 1 0.6% 1 0.01 Required additional drug
1 0.6% 1 0.01 therapy ICD/lead connection 1 0.6% 1 0.01 ICD therapy during lead
1 0.6% 1 0.01 connection Non-sustained VT 1 0.6% 1 0.01 Non-conversion of arrhythmia
Interpretation of real-time 1 0.6% 1 0.01
1 0.6% 1 0.01
markers Reconfirmation algorithm 1 0.6% 1 0.01
1. This category inclu the
programmer wand that caused the reed switch to
toggle during hig age pacito ha ng or
des is s re o m em f
sue lated t ov ent o
h volt ca r c rgi tachyarrhythmia detection. As a result, appropriate therapy was not delivered in a timely manner. The orientation of the reed switch was optimized and is being monitored as part of the manufacturing process to prevent future occurrences of this type of event.
2.
This category includes various software “anomalies” that were related to error messages or the retrieval of diagnostic information. Each of these events has been resolved through revisions made to the software.
3. This category includes any difficulties encountered while using the TMS 1000 Tachyarrhythmia Monitoring System. Each of these events has been resolved through revisions to the software and hardware
of the
system.
Cardiac Airbag Technical Manual 17
1.6 Clinical Studies
N
OTE:
The Phylax XM ICD is an earlier generation of BIOTRONIK devices. The Cardiac Airbag family is based upon the Phylax XM and other BIOTRONIK ICDs (i.e., Belos VR and Belos VR-T).
This clinical study was performed on the Phylax XM and Phylax 06 ICDs, which are earlier versions of the Cardiac Airbag ICD. The clinical study data presented here is applicable because the Cardiac Airbag / Cardiac Airbag-T is a downsized version of the Phylax XM with the addition of rate adaptive pacing capabilities. The Cardiac Airbag / Cardiac Airbag-T ICDs are slightly different as compared to the Phylax XM in the following areas:
Motion based rate adaptive pacing
Reduced programmable feature set
Minor adjustments to therapy delivery options including
no availability of ATP
Reduced size from 69 cc to 39 cc
Addition of Home Monitoring functionality
The rate adaptive pacing circuitry of Cardiac Airbag / Cardiac Airbag-T ICD is based on other US distributed BIOTRONIK products. Due to the similarities between the Cardiac Airbag / Cardiac Airbag-T, Belos VR / VR-T, and Phylax XM and the limited nature of these changes, a clinical study of the Cardiac Airbag / Cardiac Airbag-T ICD was determined to be unnecessary.
1.6.1 Patients Studied
The clinical study involved 154 patients (121 male and 33 female) with a mean age of 64.9 years (range: 26 to 95 years) and a left ventricular ejection fraction of 33% (range: 10% to 80%). Most (72%) presented with coronary artery disease / ischemic cardiomyopathy; 71% presented with monomorphic ventricular tachycardia (MVT) as their primary tachyarrhythmia.
18 Cardiac Airbag Technical Manual
1.6.2 Methods
The multicenter clinical investigation was designed to validate the safety and effectiveness of the ICD system to detect and treat monomorphic ventricular tachycardia (MVT), polymorphic ventricular tachycardia (PVT), ventricular fibrillation (VF), and bradycardia. The specific predefined objectives of the investigation included the determination of ventricular tachyarrhythmia conversion rate, sudden cardiac death (SCD) survival rate, morbidity rate, and the appropriate sensing and pacing rate.
The primary endpoint of the study was to evaluate the ventricular tachyarrhythmia conversion rate. Patients underwent standard ICD implantation and then were evaluated at predischarge and regular follow-ups every three months. Induction and conversion of the patient’s tachyarrhythmias was required at the implant procedure and predischarge follow-up.
1.6.3 Results
The mean implant duration was 8.3 ± 0.4 months with cumulative implant duration of 1286 months. There were 39 patients followed for over twelve months and 108 patients followed for over six months. The patient follow-up compliance rate was 99.6% out of 473 follow-up procedures.
2 provides a summary of the results of the study group for
Table
the predefined endpoints.
Cardiac Airbag Technical Manual 19
Table 2: Clinical Study Results
Description Study Group
[95% CI]
Tachyarrhythmia Conversion Rate Induced
1
95.8% (496/518) [93.6%, 97.3%]
Spontaneous 99.7% (1540/1544)
[99.3%, 99.9%]
Total 98.7% (2036/2062)
[98.2%, 99.2%]
Sudden Cardiac Death Survival (at one year)
Complication Rate (per total number of patients)
Appropriate Sensing and Pacing Rate
100.0% (39/39)
[91.0%, 100.0%]
5.2% (8/154)
[2.3%, 10.0%]
2
98.0% (703/717) [96.8%, 98.9%]
1. Conversion data were collected in the clinical study for both induced and spontaneous tachyarrhythmia episodes. Therefore, both types of tachyarrhythmia episodes were included in the analysis.
2. The investigator determined the appropriateness of bradycardia sensing and pacing. The rate will be determined by the number of appropriate bradycardia sensing and pacing evaluations divided by the total number of evaluations.
20 Cardiac Airbag Technical Manual
1.7 Patient Selection and Treatment
1.7.1 Individualization of Treatment
Determine whether the expected device benefits outweigh the possibility of early device replacement for patients whose ventricular tachyarrhythmias require frequent shocks.
Determine if the device and programmable options are appropriate for patients with drug-resistant supraventricular tachyarrhythmias (SVTs), because drug-resistant SVTs can initiate unwanted device therapy.
Direct any questions regarding individualization of patient therapy to your BIOTRONIK representative or BIOTRONIK technical services at 1-800-547-0394.
1.7.2 Specific Patient Populations
Pregnancy - If there is a need to image the device, care should
be taken to minimize radiation exposure to the fetus and the mother.
Nursing Mothers - Although appropriate biocompatibility testing has been conducted for this implant device, there has been no quantitative assessment of the presence of leachables in breast milk.
Geriatric Patients - Most (72%) of the patients receiving an ICD in the Phylax XM clinical study were over the age of 60 years (see Clinical Studies).
Handicapped and Disabled Patients - Special care is needed in using this device for patients using electrical wheel chair or other electrical (external or implanted devices).
1.8 Patient Counseling Information
The pulse generator is subject to random component failure. Such failure could cause inappropriate shocks, induction of arrhythmias or inability to sense arrhythmias, and could lead to the patient’s death.
Cardiac Airbag Technical Manual 21
Persons administering CPR may experience the presence of voltage on the patient’s body surface (tingling) when the patient’s ICD system delivers a shock.
A patient manual is available for the patient, patient’s relatives, and other interested people. Discuss the information in the manual with concerned individuals both before and after pulse generator implantation so they are fully familiar with operation of the device. (For additional copies of the patient manual, contact the BIOTRONIK at the address listed in this manual.)
1.9 Evaluating Prospective ICD Patients
The prospective ICD implant candidate should undergo a cardiac evaluation to classify any and all tachyarrhythmias. In addition, other patient specific cardiac information will help in selecting the optimal device settings. This evaluation may include, but is not limited to:
an evaluation of the specific tachycardia rate(s)
the confirmation and/or evaluation of any
supraventricular arrhythmias or bradyarrhythmias
the evaluation of various ATP and cardioversion therapies
the presence of any post-shock arrhythmias, and
an evaluation of the maximum sinus rate during exercise
If a patient’s drug regimen is changed or adjusted while the ICD is implanted, additional EP testing may be required to determine if detection or therapy parameter settings are relevant and appropriate.
22 Cardiac Airbag Technical Manual
2. Device Features
The Cardiac Airbag family feature set is presented under the following sub-headings: Sensing, Tachyarrhythmia Detection, Tachyarrhythmia Redetection, Tachyarrhythmia Therapy, Bradycardia Therapy, and Special Features. The features apply to all members of the Cardiac Airbag family except where specifically referenced differently.
2.1 Sensing
The Cardiac Airbag ICDs use Automatic Sensitivity Control (ASC) to adjust the sensitivity characteristics to appropriately detect the various cardiac signals. The characteristics of the sensing circuitry have been optimized to ensure appropriate sensing during all potential cardiac rhythms.
Cardiac signals vary in amplitude; therefore detection thresholds cannot be static. The Automatic Sensitivity Control (ASC) utilizes an automatic step-down threshold for sensing ventricular signals. The ASC begins by tracking the cardiac signals (R­waves) during the sensed refractory periods. The peak values measured during this time are used to set the sensing thresholds during the active detection periods.
2.1.1 Ventricular Sensitivity Settings
There are three programmable options for setting the sensitivity of the input stage. The sensitivity selections are designed to adapt the parameters of the input stage to various signal conditions. The predefined parameter sets are described in
Table 3
.
Loading...
+ 64 hidden pages