Biotek MicroFlo select Operator's Manual

Page 1
Operator’s Manual
Dispenser
MicroFlo
Select
Page 2
Page 3
Microplate Dispenser Operator’s Manual
February 2010 © 2010 Part Number 7171000 Revision C BioTek® Instruments, Inc.
Page 4
ii | MicroFlo Select™

Notices

BioTek® Instruments, Inc.
Highland Park, P.O. Box 998 Winooski, Vermont 05404-0998 USA
All Rights Reserved
© 2010, BioTek® Instruments, Inc. All rights are reserved. No part of this manual may be reproduced or transmitted in any form, or by any means electronic or mechanical, including photocopying and recording for any purpose other than the purchaser’s personal use without written permission of BioTek
Trademarks
BioTek is a registered trademark, and MicroFlo™ Select, BioStack™ , and Liquid Handling Control™ Software are trademarks of BioTek Instruments, Inc.
®
Instruments, Inc.
Microsoft
®
, Windows® , Windows XP, Windows 2000, Windows Vista™, and Excel® are either registered trademarks or trademarks of Microsoft Corporation in the United States and/or other countries.
All other trademarks are the property of their respective holders.
Restrictions and Liabilities
Information in this document is subject to change and does not represent a commitment by BioTek. Changes made to the information in this document will be incorporated in new editions of the publication. No responsibility is assumed by BioTek for the use or reliability of software or equipment that is not supplied by BioTek or its affiliated dealers.
BioTek Instruments, Inc.
Page 5

Contents

Notices ......................................................................................ii
Contents................................................................................... iii
Contacting BioTek Instruments, Inc. ............................................ vii
Revision History....................................................................... viii
Document Conventions ............................................................. viii
Intended Use Statement .............................................................ix
Quality Control...........................................................................ix
Warranty & Product Registration....................................................x
Cassette Warranty.......................................................................x
Warnings ...................................................................................x
Hazards .....................................................................................x
Precautions ...............................................................................xi
CE Mark................................................................................... xii
Electromagnetic Interference and Susceptibility............................ xiv
User Safety ............................................................................. xiv
Safety Symbols......................................................................... xv
Introduction..............................................................................1
The MicroFlo Select Microplate Dispenser....................................... 2
Hardware Features..................................................................... 4
Package Contents....................................................................... 5
Optional Accessories................................................................... 6
Specifications ............................................................................ 7
Contents | iii
All Rights Reserved.................................................................ii
Trademarks...........................................................................ii
Restrictions and Liabilities........................................................ii
Customer Service and Sales................................................... vii
Service/TAC ........................................................................ vii
European Coordination Center/ Authorized European
Representative..................................................................... vii
Regulatory Agency Approvals..................................................ix
Directive 2004/108/EC: Electromagnetic Compatibility.............. xiii
Directive 73/23/EEC: Low Voltage (Safety) ............................. xiii
Directive 2002/96/EC: Waste Electrical and Electronic Equipmentxiii Directive 2002/95/EC: Reduction of Hazardous Substances (RoHS)xiv
Directive 98/79/EC: In Vitro Diagnostics................................. xiv
USA FCC CLASS A ............................................................... xiv
Canadian Department of Communications Class A.................... xiv
Software.............................................................................. 3
Compatibility with the BioStack Microplate Stacker..................... 3
MicroFlo Select Operator's Manual
Page 6
iv | MicroFlo Select™
Installation ...............................................................................9
1: Unpack and Inspect the Instrument .........................................10
Accessories Checklist ............................................................10
2. Remove the shipping hardware................................................12
3: Install the MicroFlo’s Components............................................14
Install the Tubing Cassette ....................................................14
Tubing Cassette Diagram.......................................................15
Insert the Prime Trough Drain Tube ........................................18
Prime Trough Insert..............................................................19
Power Supply and Cords........................................................20
Data Communication ............................................................20
Operating the MicroFlo with the BioStack.................................20
4: Verify Performance................................................................21
Repacking and shipping the MicroFlo............................................22
Before shipping the instrument: .............................................22
Repackaging the instrument and its accessories: ......................22
Repacking Illustration ...........................................................25
Essential Concepts..................................................................27
How it works ............................................................................28
How to optimize performance......................................................29
Quick Dispense or Protocol?........................................................31
Handling Special Plates and Test tube racks..................................33
6-, 12-, 24-, and 48-Well Plates .............................................33
Test Tube Racks...................................................................34
Chemical Compatibility...............................................................35
Table 1. Material/Where Used List...........................................35
Table 2. Chemical Compatibility Ratings...................................36
Operation................................................................................37
Basic Operation.........................................................................38
Turn On/Off the MicroFlo.......................................................38
Using the Keypad.................................................................38
Main Menu ..........................................................................39
Dispenser Run Menu.............................................................40
Dispenser Action Menu..........................................................40
Run a Quick Dispense ................................................................42
About Dispense Protocols ...........................................................43
Run a Dispense Protocol.............................................................47
Prime the MicroFlo.....................................................................48
Purge the MicroFlo.....................................................................49
Define the Cassette ...................................................................50
Removing the Tubing Cassette....................................................50
Define the Plate Type and Plate Map ............................................51
BioTek Instruments, Inc.
Page 7
Contents | v
About Plate Types and Plate Maps...........................................51
Setup Menu Options ..................................................................53
Utility Menu Options ..................................................................54
Using the BioStack ....................................................................55
Preventive Maintenance..........................................................57
Overview .................................................................................58
Recommended Maintenance Schedule..........................................58
Flush the Tubing Cassette...........................................................59
Record the number of plates processed ........................................60
Clean Exterior Surfaces..............................................................61
Clean Priming Trough and Drain Hose ..........................................61
Verify Performance....................................................................62
Unclogging the Dispense Tips......................................................63
Replace the Tubing....................................................................65
Recalibrate the Cassette.............................................................65
Decontamination.......................................................................66
Purpose ..............................................................................66
Tools and Supplies................................................................67
Decontamination Procedure ...................................................67
Instrument Qualification.........................................................69
IQ/OQ/PQ ................................................................................70
Recommended Qualification Schedule ..........................................71
System Self-Check and Checksum Test ........................................72
System Self-Check ...............................................................72
Checksum Test ....................................................................72
Dispense Precision and Accuracy Tests.........................................73
Materials.............................................................................75
Test Solutions......................................................................76
Performing the Precision and Accuracy Tests.................................77
Procedure: ..........................................................................77
Documenting Test Results .....................................................79
MicroFlo Select Dispense Precision & Accuracy Test Worksheet 1 µL
Cassette Test.......................................................................80
MicroFlo Select Dispense Precision & Accuracy Test Worksheet 5 µL
Cassette Test.......................................................................81
MicroFlo Select Dispense Precision & Accuracy Test Worksheet 1 µL
Cassette Test.......................................................................82
MicroFlo Select Dispense Precision & Accuracy Test Worksheet 10
µL Cassette Test ..................................................................83
MicroFlo Select Dispense Precision & Accuracy Test Worksheet 5 µL
Cassette Test.......................................................................84
MicroFlo Select Dispense Precision & Accuracy Test Worksheet 10
µL Cassette Test ..................................................................85
MicroFlo Select Operator's Manual
Page 8
vi | MicroFlo Select™
MicroFlo Select Dispense Precision & Accuracy Test Worksheet
1536 Cassette Test...............................................................86
Troubleshooting......................................................................87
Performance Issues ...................................................................88
Error Codes..............................................................................89
Contact BioTek Technical Assistance Center ..................................90
Service/Technical Assistance Center (TAC)...............................90
European Coordination Center/ Authorized European
Representative.....................................................................90
Error Codes.............................................................................91
Error Codes..............................................................................92
Instrument Error Codes.........................................................92
Software Error Codes............................................................94
Dispensing to Test Tubes ........................................................95
Install the test tube rack ............................................................96
Dispensing to the test tube rack..................................................97
Gravimetric Calibration Instructions.......................................99
Calibrating the Tubing Cassette Gravimetrically........................... 100
Procedure: ........................................................................ 101
Gravimetric Cassette Calibration Worksheet...........................104
About Calibrating Cassettes Gravimetrically ................................105
Index ....................................................................................109
BioTek Instruments, Inc.
Page 9
Contacting BioTek Instruments, Inc. | vii

Contacting BioTek Instruments, Inc.

BioTek Instruments, Inc.
Highland Park, P.O. Box 998 Winooski, Vermont 05404-0998 USA
Customer Service and Sales
Internet: www.biotek.com
Phone: 888-451-5171 (toll free in the U.S.) 802-655-4740 (outside the U.S.)
Fax: 802-655-7941
E-Mail: customercare@biotek.com
Service/TAC
Phone: 800-242-4685 (toll free in the U.S.) 802-655-4740 (outside the U.S.)
Fax: 802-654-0638
E-Mail: tac@biotek.com
European Coordination Center/ Authorized European Representative
BioTek Instruments GmbH
Kocherwaldstrasse 34 D-74177 Bad Friedrichshall Germany
Internet: www.biotek.de
Phone: +49 (0) 7136 9680
Fax: +49 (0) 7136 968 111
E-Mail: info@biotek.de
MicroFlo Select Operator's Manual
Page 10
viii | MicroFlo Select™

Revision History

Rev Date Changes
A 10/2007 First Issue B 08/2008
C 02/2010
Revised the repackaging artwork and instructions in the Installation chapter to reflect new packaging materials. Removed the hyphen from BioStack. Updated the Trademarks section by adding the BioStack, Liquid Handling Control™ Software, Microsoft, and Excel. Added a Cassette Warranty to the Preface. Updated the Registration section in the Preface by adding an instruction to review the Warranty for the instrument and register online through the Customer Resource Center. Updated the TAC fax number and added “ATTN: RMA # xxxxx” to the BioTek shipping address (for returning the instrument). Added a part number disclaimer in the Package Contents and Optional Accessories sections.
Description and pictures of the Dispense Cassette were revised to match new design. Added descriptions of the Tip Guard and break-away tabs. Added description of the new Plate Clearance Height parameter.
Optional accessories were updated with angled dispensing and tubing extensions products.
Revised the Intended Use statement for IVD.

Document Conventions

The intent of this operator’s manual is to instruct the new user how to set up and operate BioTek’s MicroFlo Select Microplate Dispenser. To help you read and understand this manual, certain document conventions have been used.
This manual uses the following typographic conventions:
Example Description
This icon calls attention to important safety notes.
Warning!
Caution
A Warning indicates the potential for bodily harm and tells you how to avoid the problem.
A Caution indicates potential damage to the instrument and tells you how to avoid the problem.
BioTek Instruments, Inc.
Page 11
Example Description
Intended Use Statement | ix
The menu path to open the feature described is shown in this format.
When applicable, the above format shows you how to navigate to the feature described in that section.
Note: Information of particular importance or usefulness is presented
in this format.
Tips or other helpful information is presented in this format.
This icon calls attention to important informat ion.

Intended Use Statement

BioTek’s MicroFlo Select™ Microplate Dispenser accommodates all standard microplates and deep well blocks for bulk reagent dispensing applications. The dispense tubing cassettes are autoclavable for those applications requiring a sterile fluid path.
The MicroFlo Select™ can operate with standard robotic systems, such as BioTek’s BioStack™ Microplate Stacker.
This product may be used for In Vitro Diagnostic, research and development, or other non-clinical purposes.

Quality Control

It is considered good laboratory practice to run laboratory samples according to instructions and specific recommendations included in the assay package insert or standard laboratory protocol for the test to be conducted. Failure to conduct Quality Control checks could result in erroneous test data.
Regulatory Agency Approvals
The MicroFlo Select Microplate Dispenser has been tested, approved, and labeled to show compliance to the test standards indicated on later pages in this section.
MicroFlo Select Operator's Manual
Page 12
x | MicroFlo Select™

Warranty & Product Registration

Please take a moment to review the Warranty information that shipped with your product. Please also important information and updates about the product(s) you have purchased.
You can register online through BioTek’s Customer Resource Center (CRC) at
www.biotek.com or by calling 888-451-5171 or 802-655-4740.
register your product with BioTek to ensure that you receive

Cassette Warranty

MicroFlo Select dispense cassettes are backed by BioTek's Confidence-Plus lifetime warranty. For the lifetime of the cassette, BioTek guarantees the cassette components (Tip Holder, Center Holder, Tube Tensioner, and Tube Organizer) will withstand steam autoclave conditions of 121°C and 1 bar (750 mmHg) without adversely affecting dispense performance. In the event of a failure of any cassette component previously listed, BioTek will replace the cassette components free of charge.

Warnings

Operate the instrument on a flat surface away from excessive humidity.
When operated in a safe environment according to the instructions in this document, there are no known hazards associated with the instrument. However, the operator should be aware of certain situations that could result in serious
injury; these vary depending on the instrument type. See Hazards and Precautions.

Hazards

The following hazards are provided to help avoid injury:
Warning! Power Rating.
be connected to a power receptacle that provides voltage and current within the specified rating for the system. Use of an incompatible power receptacle may produce electrical shock and fire hazards.
Warning! Electrical Grounding.
connect primary power to the external power supply. Use of a two-prong adapter disconnects the utility ground, creating a severe shock hazard. Always connect the power cord directly to an appropriate receptacle with a functional ground.
The instrument’s power supply or power cord must
Never use a two-prong plug adapter to
BioTek Instruments, Inc.
Loading...