1. About this edition of user instructions
The current edition of the user instructions applies to the following models and ver-
sions of Compact CO2 Incubator:
S-Bt Smart Biotherm ........ V.1AW, V.1AE
2. Safety precautions
Caution!Make sure you have fully read and understood the present Manual
before using the equipment. Please pay special attention to sections
marked by this symbol.
Caution!Biohazard. Incubator operations should comply criteria of biosafety in
microbiological and biomedical laboratories, according to working
material biohazard level.
Caution!Incubator is equipped with an audible signal to draw attention to sen-
sor data and warnings. Please follow the information on display. Error
and alarm descriptions can be found in the Fault Diagnostics section.
Caution!The incubator contains a source of UV radiation. Do not start the unit
without the cover of the UV recirculator. Exposure to UV light is harmful and can cause damage to unprotected eyes and skin.
GENERAL SAFETY
The protection provided can be ineffective if the operation of the appliance does not
comply with the manufacturer's requirements.
Save the unit from shocks and falling.
Do not use the unit if it has visible mechanical damage.
Store and transport the unit at ambient temperatures between -20°C and +60°C and
maximum relative humidity of 80%.
After transportation or storage and before connecting it to the electric circuit, keep the
unit under room temperature for 2-3 hrs.
Before using any cleaning or decontamination methods except those recommended
by the manufacturer, check with the manufacturer that the proposed method will not
damage the equipment.
Do not make modifications in design of the unit.
When handling hazardous substances, follow the instructions contained in the safety
data sheets for the individual substances that are used and observe the relevant accident prevention regulations.
Follow respective safety datasheet when installing and handling the CO
reservoir.
2
4
ELECTRICAL SAFETY
Connect only to the mains with voltage corresponding to that on the serial number
label.
Do not plug the unit into an ungrounded power socket, and do not use an ungrounded
extension lead.
Ensure that the power plug is easily accessible during use.
Disconnect the unit from the mains before moving.
If liquid penetrates into the unit, disconnect it from the mains and have it checked by
a repair and maintenance technician.
Do not operate the unit in premises where condensation can form. Operating condi-
tions of the unit are defined in the Specifications section.
DURING OPERATION
Do not operate the unit in environments with aggressive or explosive chemical mix-
tures. Please contact manufacturer for possible operation of the unit in specific atmospheres.
Do not operate the unit if it is faulty or has been installed incorrectly.
Do not use outside laboratory rooms.
BIOLOGICAL SAFETY
The user is responsible to carry out appropriate decontamination if hazardous mate-
rial spills on or penetrates into the equipment.
The user is responsible for decontaminate the unit before its decommissioning and
utilization.
5
3. General information
S-Bt Smart BioTherm is designed for work in the areas of cell biology (operations
with animal cell cultures and tissues), molecular biology (DNA/RNA reaction analysis, hybridization reactions), biotechnology (synthesis of target proteins and other molecules) and
immunology (synthesis of antibodies and other proteins of immune system).
Unit provides a six-sided heating: the heating elements are located on the walls and
on the door, thus providing excellent uniform temperature distribution, regardless of external
factors, such as ambient temperature and positioning of the device.
Built-in infrared CO2 sensor allows accurate control of the CO2 level. The sensor
makes measurement non-sensitive to changes in temperature and humidity inside the incubator.
The chamber is made of stainless steel with smoothed seams to minimize contamination and to facilitate cleaning.
S-Bt is equipped with a UV air recirculation system — 1 UV lamp and a fan are
mounted behind the rear wall, providing decontamination of the working volume.
A convenient access port is built in the wall of the incubator for easy output of wire
sensors or devices' installed inside. The access port is heated independently to prevent formation of condensate.
Unit is equipped with error tracing and alarm systems, which significantly lower potential risks during operation.
Unit is equipped with a "black box" system that records temperature, humidity and
CO2 levels as well as statuses for door opening, UV lamp, fan and errors to the inner
memory.
Bluetooth connection to PC is available.
Application areas:
Cell biology: operations with animal cell cultures and tissues
Molecular biology: DNA/RNA reaction analysis, hybridization reactions
Biotechnology: synthesis of target proteins and other molecules
Immunology: synthesis of antibodies and other proteins of immune system
Features:
Six-sided heating provides uniform distribution of the temperature inside the chamber
Infrared CO
UV recirculation system for decontamination cycles
Bluetooth data transfer to PC
"Black box" parameter logging system
Error tracing and alarm system
Separately heated sealable port for chamber access for cables.
sensor, non-sensitive to temperature and humidity changes
2
6
4. Getting started
4.1. Unpacking. Remove packing materials carefully and retain them for future shipment
or storage of the unit. Examine the unit carefully for any damage incurred during
transit. The warranty does not cover in-transit damage. Warranty covers only the units
transported in the original package.
Caution!Due to the high weight of the unit, its unpacking and installing must
be carried out by two persons.
4.2. Complete set. Package contents:
4.2.1. Standard set:
- S-Bt Smart Biotherm, Compact CO
- Power cable .................................................................................................... 1 pce.
5.9.5.1 All View Plot options use following interface. Figure 13 shows the temperature plot
in relation to time.
Use Left and Right buttons to move the plot along the x axis
Use Up and Down buttons to move the plot along the y axis.
Use <-xx-> button to choose the x axis navigational step. Available scales are 15
minutes, 1 hour, 1 day, 1 week and 1 month.
5.9.5.2 Clear Log button clears the logger’s data.
5.9.5.3 Logger On/Off button triggers the automatic logging of the data.
5.9.5.4 Memory write location indicator. When the data logger uses up all allocated
memory, it begins overwriting data from beginning. 100% of memory is used up in
approximately 1 month.
5.10. After stabilization of temperature and CO2 level, open the doors and load the materi-
als in the incubator according to methodological recommendations.
Note.If the door is left open for more than a minute, a short signal is
sounded every 20 seconds until the door is closed.
9
5.11. Incubator air sampling. Sampling can be performed through the capped opening
marked as “air sample”, on the rear access panel.
5.12. To turn off the incubator, set the power switch into position О (off) and detach the
mains plug from the mains outlet.
5.13. Operation recommendations.
The incubator saves settings even when switched off.
During decontamination treat every visible surface and door. When working with po-
tentially pathogenic material, use more potent decontamination method. See 9.3.
Figure 1. Working screen Figure 2. Menu
Figure 3. Timer menu Figure 4. UV recirculator is on
Figure 5. Date & Time menu Figure 6. Service menu
10
Figure 7. Current Errors menu Figure 8. Settings menu
Figure 9. Alarm Settings menu Figure 10. Auto Lock Delay menu
Figure 11. Active Alarms menu Figure 12. Data Logger
Figure 13. Temperature plot
11
6. Fault diagnostics
CO2 Sens. Err.
CO2 sensor error
Back
Back temperature sensor error
Bott.
Bottom temperature sensor error
Left
Left wall temperature sensor error
Right
Right wall temperature sensor error
Up
Ceiling temperature sensor error
Door
Door temperature sensor error
Chmb.
Temperature sensor in the incubator error
Perimeter T.Sens. Err.
Perimeter temperature sensor error
Recirc. Lamp Err.
UV lamp error
Communication Err.
Error connecting to power board
CO2.Sens.Comm.Err.
Error connecting to CO2 sensor board
Temp. High Alarm
Adjustable limit is exceeded.
Temp. Low Alarm
CO2 High Alarm
CO2 Low Alarm
Low Humidity
UV Lamp Hours > 9000h
Lamp reaches end of lifetime. Consider a replacement.
6.1. The incubator is equipped with an alarm and error system. Active alarms and errors
are appropriately indicated on the display. Alarms and errors are followed by sound
signal that can be turned off by pressing any button.
6.2. Error text can be found in Menu > Service > Settings > Current Errors, see 5.9.4.
Choose Errors Reset and confirm to reset the error list.
6.3. Error list can be found in the table below:
6.4. Alarm text can be found in Menu > Active Alarms. Choose Alarms Reset and con-
firm to reset the alarm list.
6.5. Alarms list can be found in the table below. Alarms are adjustable, see 5.9.3.2.
12
7. Specifications
1
2
The unit is designed for operation in cold rooms, incubators (except CO2 incubators)
and closed laboratory rooms at ambient temperature from +4°C to +40°C in a non-condens-
ing atmosphere and maximum relative humidity 80% for temperatures up to 31°C decreasing
linearly to 50% relative humidity at 40°C.
Biosan is committed to a continuous programme of improvement and reserves the
right to alter design and specifications of the equipment without additional notice.
7.1. Chamber Material ............................................................................... Stainless Steel
7.2. Temperature setting range ................................................................+25°C ... +60°C
7.3. Temperature resolution ..................................................................................... 0.1°C
7.4. Temperature stability ....................................................................................... ±0.1°C
7.5. Temperature uniformity @37°C....................................................................... ±0.3°C
7.6. Working volume ............................................................................................ 46 litres
7.7. Number of shelves ...................................................................... 6 max. (3 provided)
7.8. Outer door ............................................................................... Heated, made of steel
7.9. Inner door .................................... Lockable glass with UV protection film, 4 mil clear
7.24. Working voltage ........................................................................ 230/115 V, 50/60 Hz
7.25. Weight2 ........................................................................................................... 37.7 kg
At temperatures below 50°CAccurate within ±10%
13
8. Ordering information
Model
Version
Description
Catalogue number
S-Bt Smart Biotherm,
Compact CO2 Incubator
V.1AW
230 V, 50/60 Hz
BS-010425-A01
V.1AE
120 V, 50/60 Hz
Optional accessories
Catalogue number
Shelf
BS-010425-AK
Shelf rack
on request
Bluetooth dongle and software for PC
BS-010425-BK
Incubator stacking device
BS-010425-CK
Replacement parts
Catalogue number
Hydrophobic filter with tubing
BS-010425-S01
Water pan and lid
BS-010425-S02
Access port seal, foamed PE
BS-010425-S03
8.1. Models and versions available:
8.2. To inquire about or order the optional accessories or the replacement parts, contact
Biosan or your local Biosan representative.
8.3. Optional accessories:
8.4. Replacement parts:
14
9. Care and maintenance
Name
Description
UV lamp
TUV 6W, T5, G6
Fuse
M 3.15 A
Hydrophobic filter
CO2 filter
Hydrophobic filter set
Filter and tubing
Access port seal
Made from foamed PE
9.1. If the unit requires maintenance, disconnect the unit from the mains and contact Bio-
san or your local Biosan representative.
9.2. All maintenance and repair operations must be performed only by qualified and spe-
cially trained personnel.
9.3. Regular service schedule
9.3.1. After operation. Switch on the recirculator running for 1 to 3 hours in an empty incu-
bator to decontaminate air in the incubator. To switch on the UV recirculator, see
5.9.1.1. After decontamination, slightly open the incubator doors – UV lamp has been
known to produce reactive oxygen species in minor concentrations.
9.3.2. Monthly (more frequently, if necessary):
- Rub the inner surfaces, water pan with lid and shelves of the incubator with 75%
ethanol solution or other acceptably disinfectant such as sodium hypochlorite solu-
tion, and allow them to dry.
- After cleaning the glass door, rinse with water and dry the excess to avoid streaks
and possible corrosion.
- Remove dust with moist cloth piece from outer incubator surface, especially front
screen and upper part of the incubator.
- Visually check the hydrophobic filter (see 9.7) and access port seal (see 9.8) and
replace if necessary.
- Check the accuracy of CO
tion with a reference CO2 analyser.
9.4. Expendable materials:
sensor system by measuring the actual CO2 concentra-
2
9.5. Fuse replacement. Disconnect incubator from the mains. Disconnect power plug
from power socket. Open fuse holder. Check and replace if necessary, 230 V – М
3.15 А (type М - time lag: Medium).
15
9.6. UV lamp replacement. Average lifetime of the supplied UV lamps is 9000 h. Re-
placement is necessary after lamp stops functioning or at the end of manufacturer
specified lifetime. Only the persons who have completed special training are allowed
to perform lamp replacement. UV lamp operation lifetime is checked and tracked by
the incubator.
9.7. Hydrophobic filter replacement. The filter should be replaced when it becomes
glazed with dirt or with sucked liquid. Maximum time of use is 30 days. We recom-
mend replacing the filter immediately after full cycle of maintenance.
Close the CO2 valve on the gas reservoir and turn off the unit. Remove the filter from
the tubing. Insert a new filter into the socket and tighten it. Pay attention that “IN”
marking must be facing right side when viewing the incubator from the front.
9.8. Access port seal replacement. To prevent possible contamination of access port
seal (made from foamed polyethylene) and to ensure complete sealing of the port,
the seal should be replaced after 30 days from the first operation.
9.9. Calibration. Recalibrate the unit after the date on calibration certificate. To complete
the process, contact Biosan or your local Biosan representative.
16
10. Warranty
Technical support
Product class description
biosan.lv/en/support
biosan.lv/classes-en
Model
S-Bt Smart Biotherm, Compact CO2 Incubator
Serial number
Date of sale
10.1. The Manufacturer guarantees the compliance of the unit with the requirements of
Specifications, provided the Customer follows the operation, storage and transporta-
tion instructions.
10.2. The warranted service life of the unit from the date of its delivery to the Customer is
24 months. For extended warranty, see 10.5.
10.3. Warranty covers only the units transported in the original package.
10.4. If any manufacturing defects are discovered by the Customer, an unsatisfactory
equipment report shall be compiled, certified and sent to the local distributor address.
To obtain the claim form, visit Technical support page on our website at link below.
10.5. Extended warranty. For S-Bt Smart Biotherm, the Smart class model, extended war-
ranty is a paid service. Contact your local Biosan representative or our service de-
partment through the Technical support section on our website at the link below.
10.6. Description of the classes of our products is available in the Product class descrip-
tion section on our website at the link below.
10.7. The following information will be required in the event that warranty or post-warranty
service comes necessary. Complete the table below and retain for your records.
17
11. Notes
18
12. EU Declaration of conformity
Unit type Compact CO2 incubator
Models S-Bt Smart BioTherm
Serial number 14 digits styled XXXXXXYYMMZZZZ, where XXXXXX is
model code, YY and MM – year and month of production,
ZZZZ – unit number.