Biosan MMS-3000, MS-3000 User Instructions

2
3
Contents
1. About this edition of user instructions ....................................................................... 3
2. Safety precautions ................................................................................................... 4
3. General information .................................................................................................. 5
4. Getting started .......................................................................................................... 6
5. Operation ................................................................................................................. 7
6. Specifications ........................................................................................................... 8
7. Ordering information ................................................................................................ 9
8. Care and maintenance ............................................................................................. 9
9. Warranty ................................................................................................................. 10
10. EU Declaration of conformity .................................................................................. 11
1. About this edition of user instructions
MS-3000 ............................................................................................. version V.2AW
MMS-3000 .......................................................................................... version V.3AW
4
2. Safety precautions
Caution! Make sure you have fully read and understood the present Manual
before using the equipment. Please pay special attention to sections marked by this symbol.
Attention! Magnetism! Effects of a strong magnetic field on the biological systems
have to be taken in to account. Magnetic fields can affect heart pace­makers, data carriers, etc.
GENERAL SAFETY
The protection provided can be ineffective if the operation of the appliance does not
comply with the manufacturer's requirements.
Save the unit from shocks and falling. Store and transport the unit at ambient temperatures between -20°C and +60°C and
maximum relative humidity of 80%.
After transportation or storage and before connecting it to the electric circuit, keep the
unit under room temperature for 2-3 hrs.
Before using any cleaning or decontamination methods except those recommended
by the manufacturer, check with the manufacturer that the proposed method will not damage the equipment.
Do not make modifications in design of the unit.
ELECTRICAL SAFETY
Connect only to the mains with voltage corresponding to that on the serial number
label.
Use only the external power supply provided with this product. Ensure that the power plug is easily accessible during use. Do not plug the unit into an ungrounded power socket, and do not use an ungrounded
extension lead.
Disconnect the unit from the mains before moving. If liquid penetrates into the unit, disconnect it from the mains and have it checked by
a repair and maintenance technician.
Do not operate the unit in premises where condensation can form. Operating condi-
tions of the unit are defined in the Specifications section.
DURING OPERATION
Do not start operation at maximum speed. Do not operate the unit in environments with aggressive or explosive chemical mix-
tures. Please contact manufacturer for possible operation of the unit in specific at­mospheres.
Do not operate the unit if it is faulty or has been installed incorrectly. Do not use outside laboratory rooms.
BIOLOGICAL SAFETY
The user is responsible to carry out appropriate decontamination if hazardous mate-
rial spills on or penetrates into the equipment.
5
3. General information
MS-3000/MMS-3000 magnetic stirrer is designed for effective stirring of different vis-
cosity liquids.
MS-3000/MMS-3000 is a compact magnetic stirrer with the stainless steel working surface. It provides liquid stirring with the magnetic element rotation speed up to 3000 rpm (max. speed depends on the magnetic element size, stirred volume, viscosity, glassware shape, etc.). The unit is designed for operation with different size magnetic stirring elements (20-50 mm long for MS-3000 and 20-70 mm long for MMS-3000). Other size magnetic ele­ments may not provide appropriate operation.
MMS-3000 is equipped with an attachable stand that allows inserting different sen­sors (temperature, pH etc.) inside the liquid.
Application fields:
Chemistry
Stirring reaction ingredients during the fine organic synthesis, re-
search in the organic catalysis field, different viscosity chemical rea-
gents.
Biochemistry
Solutions preparation, dialyze, salt and alcohol sedimentation of mac-
romolecules, gradient forming in the column chromatography, etc.
Soil science
Biological and chemical substances and samples extraction, research
of the soil and ground chemical and biochemical compounds.
Biotechnology
Using as a minireactor in the micro-organism cells cultivation, culture
medium preparation, titration, etc.
6
4. Getting started
4.1. Unpacking. Remove packing materials carefully and retain them for future shipment
or storage of the unit. Examine the unit carefully for any damage incurred during
transit. The warranty does not cover in-transit damage. Warranty covers only the units
transported in the original package.
4.2. Complete set. Package contents:
4.2.1. MS-3000:
- MS-3000, Magnetic Stirrer .............................................................................. 1 pce.
- Magnetic stirring element
1
................................................................................ 1 pce.
- External power supply ..................................................................................... 1 pce.
- Operating manual, declaration of conformity .................................................. 1 copy
4.2.2. MMS-3000:
- MMS-3000, Magnetic Stirrer ........................................................................... 1 pce.
- SR-1, attachable stand ..................................................................................... 1 pce.
- Magnetic stirring element
1
................................................................................ 1 pce.
- External power supply ..................................................................................... 1 pce.
- Operating manual, declaration of conformity .................................................. 1 copy
4.3. Setup.
- Place the unit on horizontal even working surface.
- Connect the external power supply unit into the socket at the rear side of the unit
and position the unit for an easy access to the external power supply and the power
switch.
4.4. SR-1 stand installation on MMS-3000. Remove the screw on the fixing socket at the
back of the stirrer (fig. 1/1). Screw in the part of the stand with the counter-nut into
the fixing socket and secure with the counter-nut. Screw in the second part of the
stand into the attached first part.
Figure 1. SR-1 stand for MMS-3000
1
Cylindrical (6x25 mm) and encapsulated in PTFE
7
5. Operation
5.1. Connect the external power supply to the mains. Note! Vessel must be flat-bottomed and fit tightly to the working surface of
the magnetic stirrer (figure 2).
Figure 2.
5.2. Position the vessel on the center of the platform. Immerse the magnetic stirring ele-
ment.
5.3. Turn the Power switch (fig. 3/1 or 3/3) into the position On.
5.4. Set the required speed using the Speed knob (fig. 3/2 or 3/4). Increase the speed. Note! A maximum efficient speed is shown in Table 1 for different stirring
element lengths and mixing volumes. Beyond these numbers, mixing might be unstable.
Note! In the model MMS-3000, if the stirring element loses positioning, the
motor drops the speed, positions the element and slowly picks up the speed. If mixing is still unstable due to high viscosity or large mixing volume, consider lowering mixing speed.
5.5. After finishing the operation, decrease the rotation speed to the minimum and turn
the Power switch into position Off.
5.6. Disconnect the external power supply from the mains outlet.
MS-3000 MMS-3000
Figure 3. Control panel
8
6. Specifications
The unit is designed for operation in cold rooms, incubators (except CO2 incubators)
and closed laboratory rooms at ambient temperature from +4°C to +40°C in a non-condens-
ing atmosphere and maximum relative humidity 80% for temperatures up to 31°C decreasing
linearly to 50% relative humidity at 40°C.
Biosan is committed to a continuous programme of improvement and reserves the right to alter design and specifications of the equipment without additional notice.
MS-3000
MMS-3000
Speed control range
0–3000 rpm
Maximum stirring volume (H2O)
5 L
20 L
Working surface dimensions
110x110 mm
Ø160 mm
Working surface material
Stainless steel
SR-1 detachable stand size
Ø8x320 mm
Maximum length of magnetic
stirring element
50 mm
70 mm
Maximum stirred liquid viscosity
Up to 1170 mPa*s
Maximum continuous operation time1
24 h
Working current / Power consumption
220 mA / 2.6 W
250 mA / 3 W
External power supply
in AC 100-240 V, 50/60 Hz, out DC 12 V
Weight2
0.8 kg
1.5 kg
Dimensions
120x150x65 mm
185x230x75 mm
Table 1. MS-3000 and MMS-3000 maximal speed in RPM, depending on H2O volume and mag-
netic element length
Magnetic element length, mm
MS-3000
MMS-3000
2 L
5 L
2 L
5 L
15 L
20 L
25
3000
2400
3000
2800
2650
2600
50
700
700
1700
1500
1300
1250
70
– – 620
530
440
360
1
Maintain at least 1 hour long pause between prolonged continuous operations
2
Accurate within ± 10%
9
7. Ordering information
7.1. Models and versions available:
Model
Version
Catalogue number
MS-3000
V.2AW
BS-010301-AAF
MMS-3000
V.3AW
BS-010305-AAF
7.2. To inquire about or order the replacement parts, contact Biosan or your local Biosan
representative.
7.3. Replacement parts:
Replacement parts
Description
Catalogue number
SR-1
Detachable stand for MMS-3000
BS-010302-AK
Magnetic stirring
element
Cylindrical shape (6x25 mm)
and PTFE encapsulated
BS-010302-S12
8. Care and maintenance
8.1. If the unit requires maintenance, disconnect the unit from the mains and contact Bio-
san or your local Biosan representative.
8.2. All maintenance and repair operations must be performed only by qualified and spe-
cially trained personnel.
8.3. Standard ethanol (75%) or other cleaning agents recommended for cleaning of labor-
atory equipment can be used for cleaning and decontamination of the unit.
10
9. Warranty
9.1. The Manufacturer guarantees the compliance of the unit with the requirements of
Specifications, provided the Customer follows the operation, storage and transporta-
tion instructions.
9.2. The warranted service life of the unit from the date of its delivery to the Customer is
24 months. For extended warranty, see 9.5.
9.3. Warranty covers only the units transported in the original package.
9.4. If any manufacturing defects are discovered by the Customer, an unsatisfactory
equipment report shall be compiled, certified and sent to the local distributor address.
To obtain the claim form, visit Technical support page on our website at link below.
9.5. Extended warranty. For MS-3000 MMS-3000, the Basic Plus class models, extended
warranty is a paid service. Contact your local Biosan representative or our service
department through the Technical support section on our website at the link below.
9.6. Description of the classes of our products is available in the Product class descrip-
tion section on our website at the link below.
Technical support
Product class description
biosan.lv/en/support
biosan.lv/classes-en
9.7. The following information will be required in the event that warranty or post-warranty
service comes necessary. Complete the table below and retain for your records.
Model
MS-3000 MMS-3000, Magnetic stirrers
Serial number
Date of sale
11
10. EU Declaration of conformity
Unit type Magnetic stirrers Models MS-3000, MMS-3000 Serial number 14 digits styled XXXXXXYYMMZZZZ, where XXXXXX is
model code, YY and MM – year and month of production, ZZZZ – unit number.
Manufacturer SIA BIOSAN
Latvia, LV-1067, Riga, Ratsupites str. 7/2
Applicable Directives EMC Directive 2014/30/EU
LVD Directive 2014/35/EU RoHS2 2011/65/EU WEEE 2012/19/EU
Applicable Standards LVS EN 61326-1: 2013
Electrical equipment for measurement, control and laboratory use. EMC requirements. General requirements.
LVS EN 61010-1: 2011
Safety requirements for electrical equipment for measurement, control, and laboratory use. General requirements.
LVS EN 61010-2-051: 2015
Particular requirements for laboratory equipment for mixing and stirring.
We declare that this product conforms to the requirements of the above Directives
____________________ ____________________
Signature Signature
Svetlana Bankovska Aleksandr Shevchik Managing director Engineer of R&D
____________________ ____________________
Date Date
Biosan SIA Ratsupites 7, build.2, Riga, LV-1067, Latvia Phone: +371 67426137 Fax: +371 67428101
http://www.biosan.lv
Edition 2.-3.01 – October 2018